Abstract
Low intensity transcranial focused ultrasound (LIFU) is a promising method of non-invasive neuromodulation that uses mechanical energy to affect neuronal excitability. LIFU confers high spatial resolution and adjustable focal lengths for precise neuromodulation of discrete regions in the human brain. Before the full potential of low intensity ultrasound for research and clinical application can be investigated, data on the safety of this technique is indicated. Here, we provide an evaluation of the safety of LIFU for human neuromodulation through participant report and neurological assessment with a comparison of symptomology to other forms of non-invasive brain stimulation. Participants (N = 120) that were enrolled in one of seven human ultrasound neuromodulation studies in one laboratory at the University of Minnesota (2015–2017) were queried to complete a follow-up Participant Report of Symptoms questionnaire assessing their self-reported experience and tolerance to participation in LIFU research (Isppa 11.56–17.12 W/cm2) and the perceived relation of symptoms to LIFU. A total of 64/120 participant (53%) responded to follow-up requests to complete the Participant Report of Symptoms questionnaire. None of the participants experienced serious adverse effects. From the post-hoc assessment of safety using the questionnaire, 7/64 reported mild to moderate symptoms, that were perceived as ‘possibly’ or ‘probably’ related to participation in LIFU experiments. These reports included neck pain, problems with attention, muscle twitches and anxiety. The most common unrelated symptoms included sleepiness and neck pain. There were initial transient reports of mild neck pain, scalp tingling and headache that were extinguished upon follow-up. No new symptoms were reported upon follow up out to 1 month. The profile and incidence of symptoms looks to be similar to other forms of non-invasive brain stimulation.
Subject terms: Neuroscience, Neurology, Risk factors
Introduction
Transcranial focused ultrasound (LIFU) is a new and promising non-surgical low-energy technique that uses mechanical energy to modulate neuronal activity with high spatial resolution and adjustable depth of focus. LIFU has been used safely and effectively for cortical and sub-cortical neuromodulation in mouse1–4, rat5–7, rabbit8, sheep9,10, pig11 and primate12–15, and has also been demonstrated to be an effective method of transient cortical and sub-cortical neuromodulation in humans16,17. In humans, LIFU has been applied to the temporal cortex18, primary somatosensory cortex (S1)16,19, secondary somatosensory cortex (S2)20, primary motor cortex21, primary visual cortex22 and thalamus17,23. LIFU has been shown to affect the amplitude of evoked potentials8,16,19, the power, phase and frequency of the electroencephalogram (EEG)16,24; the blood oxygen level dependent (BOLD) magnetic resonance imaging signal8,25,26, as well as tactile16,19 and reaction time21 behavior.
As currently employed, human neuromodulation with LIFU typically involves coupling one (or more than one20) focused single-elements usually in the ~250 to 600 kHz range (for efficient energy transfer through skull27) to the scalp to target a desired brain region. Transducers for cortical targeting are generally small (~30 mm diameter with a ~30 mm focal length); produce a ~3–4 millimeter lateral and ~1–2 centimeter axial resolution, and can be placed anywhere on the head similar to other current methods of non-invasive brain stimulation. Ultrasound is also capable of reaching brain targets deep to the cortical surface as the acoustic waves can be focused to any desired depth within certain physical limits of the transducer employed. Transducers for deep brain modulation are typically larger (~70 mm diameter)17,23 to achieve this deeper focal length at reasonable axial resolutions17. In addition to adjusting focal lengths, there are a number of parameters that can be manipulated when using ultrasound including the acoustic frequency, amplitude, duration, duty cycle, pulse repetition frequency etc. and the efficacy of some of these for successful neuromodulation has been addressed in small3,4 and large animal studies10 though the mechanism of acoustic energy for neuronal modulation is largely theoretical28–31 and the impact of parameter space in humans is not yet well-described. The bioeffects of ultrasound for neuromodulation in humans as described here are likely largely mechanical as opposed to thermal or cavitational as the parameters used are of low intensity and short duration and generate temperatures insufficient for thermal modulation32. Ultrasound for transient neuromodulation is different from the use of ultrasound for surgery where high intensities are used to thermally ablate tissues33,34 or for transient blood-brain barrier (BBB) opening where high intensities are also used in combination with contrast agents (microbubbles) to intentionally produce cavitation as a means of opening the BBB35,36.
In its current state, ultrasound for neuromodulation generally follows the safety guidelines of the Food and Drug Administration (FDA) for obstetric diagnostic ultrasound and adult cephalic applications37. These include derated limits of spatial peak pulse average (Isppa) of 190 W/cm2, a spatial peak temporal average of 720 mW/cm2 (94 mW/cm2 for adult cephalic) and a mechanical index (MI = peak negative pressure/√fcenter frequency) of 1.9. MI is an indication of the ability to produce cavitation related bio-effects and can be used as an indication for potential micromechanical damage. In addition to FDA guidelines, the IEC 60601 part 2 standard for therapeutic equipment sets 3 W/cm2 as the limit on acoustic intensity38. These levels of energy are generally respected in human ultrasonic neuromodulation studies16,17,19–22,24–26 even though there are no definitive guidelines for energy deposition into the human brain. There is a long history of ultrasound for diagnostic and therapeutic applications, but explicit expository and dosimetry are still largely lacking39,40. There are several thorough reports examining the effect different intensities of ultrasound to affect tissue41,42 and efforts made to develop thresholds for potential hazards43,44 though these studies typically use continuous wave schemes and high intensities well beyond the levels used for transient neuromodulation and therefore are not wholly informative for low intensity applications. As such, it is important to assess the safety of ultrasound for human neuromodulation. It is the purpose of this paper to provide an initial assessment of the symptoms and minor adverse events and safety of single element focused ultrasound for human neuromodulation as there is little research on participant perceived tolerance and report of symptoms. Here, we report on the findings of a variant of the Participant Report of Symptoms questionnaire45,46 assessing participants’ perceived tolerance to participation in LIFU and their perceived relation of any symptoms to the ultrasound intervention. Of a group of 120 queried, a total of (N = 64) consented to completing the questionnaire at various time points from immediately post-experiment out to 22 months.
Results
Follow-up response rate
A total of 64/120 (53.3%) participants responded to the email regarding follow-up questionnaire. The mean age of the participants was 22.96 ± 2.14 years (29 Male, 35 Female). See Table 1 for individual experiment demographics and follow-up response rates for each study. The time of response post experimental participation ranged from 1 month to 22 months after participation for experiments 1–6 (see Fig. 1a). For experiment 7, 17 participants responded to the questionnaire immediately post experiment and then were contacted at a random time post experiment out to one month (Fig. 1b).
Table 1.
Study | Target | Results | Methods | N | Response | Age (Mean ± SD) | M/F | Response Rate |
---|---|---|---|---|---|---|---|---|
117 | Thalamus | P14 SEP ↓ α, β, γ power ↓ | EEG, SEP | 20 | 12 | 23.4 ± 1.8 | 5/7 | 60% |
217 | Thalamus | Tactile discrimination ↓ | Behavior Assessment | 20 | 11 | 22.7 ± 2.5 | 5/6 | 55% |
321 | M1 | Reaction Time ↓ | Behavior Assessment | 25 | 11 | 23.4 ± 1.9 | 6/5 | 44% |
426 | M1 | BOLD volume ↓ | 7 T MRI | 6 | 4 | 22.5 ± 2.4 | 3/1 | 67% |
521 | M1 | MEP amplitude ↓ | Single pulse TMS | 12 | 5 | 24.5 ± 1.0 | 2/3 | 42% |
621 | M1 | SICI–, SICF ↓ | Paired Pulse TMS | 10 | 4 | 23.8 ± 1.5 | 2/2 | 40% |
7** | M1 | Motor Learning– | EEG, EMG, Behavior | 27 | 17 | 21.2 ± 2.2 | 6/11 | 63% |
Total | 120 | 64 | 22.96 ± 2.14 | 29/35 | 53.3% |
Symptoms reported
Data from all seven experiments revealed 7/64 reported mild or moderate symptoms that they felt were ‘possible’ or ‘probably’ related to the ultrasound intervention. These included neck pain, difficulty paying attention, muscles twitches and anxiety. There were no reports of any severe and related symptoms (Fig. 2a). There was however, one report of severe unusual feelings, attitudes or emotions though was reported as being unrelated to LIFU in experiment 3. Of the other reported conditions, participants rated these as unrelated or unlikely related to the ultrasound intervention (see Fig. 2b). The most common reported symptom was sleepiness though this was rated as unrelated or unlikely for all instances. Other responses included headache (n = 4), itchiness (n = 5), tooth pain (n = 1) and forgetfulness (n = 4). No participant rated any reported symptom as definitely related to the ultrasound intervention (see Fig. 2b). See Fig. 3 for a breakdown of symptoms by experiment for experiments 1–6.
Duration of symptoms
In a subset of participants (n = 17) for experiment 7 we collected response to the questionnaire at two time points: immediately following experimentation (~20 minutes) and then at a randomly assigned follow-up at 1 week, 2 weeks, 3 weeks or 1 month (see Fig. 1b). On day zero, there were three reports of neck pain, three reports of sleepiness, one report of scalp tingling, one report of tooth pain, one report of difficulty paying attention and one report of feeling anxious, worried or nervous and one ‘other’ report of mild back pain (see Fig. 4a,b). No participant reported more than one symptom at initial inquiry. Of these reports, neck pain was perceived as unrelated in two instances and possible in one. Sleepiness was perceived as unrelated in two instances and unlikely in the other. Tingling of the scalp was perceived as possibly related to the intervention, difficulty paying attention was unlikely, tooth pain was unlikely and anxiousness was perceived as possibly related (Fig. 4). At follow-up (1 week to 1 month) these participants did not report any persisting or new effects (Fig. 4c). Of the 7 participants who did not report any initial symptoms, none reported additional symptoms at follow-up (Fig. 4c).
Correlation of symptom response to LIFU parameters
To gauge the overall positive symptom rate, we tabulated all positive responses regardless of subjective report on the relation to the experimental intervention. The overall positive report of symptoms for all experiments (1–7) included in our neurological questionnaire was 55/1280 total possible positives for an overall positive response rate of 4.3%. Of the 55 total positive responses 38/55 (69%) were judged by the participants to be unrelated to the LIFU interventions, 10/55 (18%) unlikely, 4/55 (7%) possible, 3/55 (5%) probable and 0/55 definitely related. The positive response rates for experiments 1–7 were: 5.4%, 3.2%, 4.5%, 2.5%, 8%, 6.3% and 2.9% respectively (see Fig. 5a). As a means to begin to relate ultrasound to potential symptomology we performed correlations between ultrasound parameters and the overall response rate. The linear correlation of the response rate percentage and mechanical index (MI) was not significant (r = 0.633, p = 0.13), there was no linear correlation between number of stimulations and response rate; r = 0.285, p = 0.53 however, (Isppa) was found to have a significant positive correlation with response rate; r = 0.797, p = 0.0319 (Fig. 5b).
Comparison to other non-invasive neuromodulation modalities
A total of 948 subjects from repetitive transcranial magnetic stimulation (rTMS) studies, 83 subjects from deep TMS (dTMS), and 554 subjects from transcranial direct current stimulation (tDCS) studies were included in the overall symptom incidence calculations. References for the included papers in this analysis are provided in Supplementary Materials. Only those incidences that could be directly compared (had a similar label or categorization to those used in our questionnaire) to the LIFU symptom incidences from this study are presented below except for the dTMS. The most common symptoms from rTMS included headache (14.6%) and neck pain (3.8%); tDCS reported itchiness (16.6%), tingling (13.2%), and sleepiness (10.8%) most frequently. Ultrasound had no incidences of seizure and lower incidence of headache than both forms of TMS, but higher incidences of neck pain, attention issues, twitching, tingling, itchiness, and sleepiness. Compared to tDCS, ultrasound had similar incidences of headache, attention issues and feelings of nausea and both have no report of seizure. A comparison of symptom profiles is summarized in Table 2. Because ultrasound is capable for sub-cortical neuromodulation, comparison of reported symptoms to dTMS is highly relevant if one is to consider either LIFU or dTMS for sub-cortical neuromodulation. Deep TMS is a method of TMS that uses specific coil geometries to help penetrate and focus the magnetic field deeper into the brain for stimulation of sub-cortical targets47. From our literature review, we found a similar total number of subjects (N = 83) for dTMS as reported here. The most common reported symptom(s) from dTMS were general discomfort and/or unpleasantness (21.7%) and headache (10.8%). A comparison of symptom profiles is shown in summary Table 2. One of the feasibility and safety dTMS studies published was not included in the summary above since the number of symptom occurrences were not reported48. It was reported that some dTMS (H1-coil, H2-coil) subjects experienced dizziness but rTMS and sham TMS subjects did not. Perhaps more strikingly, some H1-coil dTMS subjects experienced feelings of detachment and a transient impairment in spatial recognition memory and Lenoir et al. (2018) recently reported two cases of seizure using dTMS for insular stimulation49.
Table 2.
LIFU | rTMS | dTMS | tDCS | |
---|---|---|---|---|
Number of Subjects | 64 | 948 | 83 | 554 |
Age Range | 18–38 | 18–65 | 20–54 | 18–73 |
% Female | 55.0% | 45.4% | 48.8% | 45.5% |
% Reported during session | 21.0% | 97.2% | 100.0% | 100.0% |
% Reported at a follow up | 79.0% | 6.8% | 27.5% | 6.8% |
Headache | 6.3% | 14.6% | 10.8% | 6.9% |
Neck Paina | 12.5% | 3.8% | 3.6% | NR |
Nausea | 1.6% | 1.4% | NR | 0.9% |
Attention Issuesb | 6.3% | NR | NR | 8.1% |
Seizure | 0.0% | 0.3% | 2.4% | NR |
Twitchingc | 4.7% | 1.7% | NR | NR |
Tingling | 1.6% | NR | NR | 13.2% |
Itchinessd | 7.8% | NR | NR | 16.6% |
Sleepinesse | 23.4% | 0.1% | 5.2% | 10.8% |
Discomfort | NR | 6.98% | 21.7% | 3.61% |
aNeck Pain: includes reported neck pain, neck ache, pain in trigeminal nerve region, or jaw ache.
bAttention Issues: includes reported attention issues or difficulty concentrating.
cTwitching: includes twitching, muscle twitching, or muscular problems.
dItchiness: includes reported itchiness, itching, itching sensation, or contact dermatitis.
eSleepiness: includes reported sleepiness, fatigue, or tiredness.
NR = not reported.
Discussion
In this report, we provide a qualitative assessment of minor adverse events reported from seven experiments conducted in our lab using single element LIFU for human neuromodulation. We collected retrospective data via a participant report of symptoms questionnaire administered over the telephone at varying time points post experiment that ranged from 0 months (day of experiment collected in person) to 22 months post experiment (telephone interview). 64/120 total participants responded to the questionnaire. Symptoms included headache, neck pain, itchiness, sleepiness, problems with attention, tooth pain, muscle twitches, and anxiety. Only one of these reports was rated as severe and none were reported as definitely related to the LIFU intervention. A subset of participants took the questionnaire immediately after experimentation. Immediate symptoms included mild headache, mild neck pain, and tingling in the scalp. None of these symptoms persisted upon follow-up out to one month and no new symptoms were reported. The intensity (Isppa) of ultrasound ranged from 11.56 W/cm2 to 17.12 W/cm2 (in free water) for the experiments included in this study and we found a significant positive linear correlation of the symptom response rate and the ultrasound intensity (Isppa). The intensity (Isppa) used in these studies is considerably lower than FDA thresholds for ultrasound diagnostics. Despite a lack of definitive causation, and the finding that most of the reported symptoms were believed by the participants to be unrelated to the LIFU intervention, this finding nevertheless speaks to limiting the intensity used in future ultrasound experiments and determining as low as reasonably achievable levels for neuromodulation. Despite the Isppa level being below FDA thresholds, the Ispta (spatial peak temporal average) in these studies was above FDA thresholds for diagnostics. Ispta is defined by the Isppa multiplied by the duty factor providing a metric of the average intensity over the duration of the pulse. One of the main concerns of determining safe intensity levels is estimating the intracranial pressure. The intensities presented here were taken from empirical recordings in free water and hence the derated intensities will be considerably lower. The skull is highly attenuative to ultrasound and the in situ derated pressures are not exactly known but can be estimated using either empirical pressure measurements using a hydrophone through skull fragments or through computer modelling that takes into consideration the acoustic properties of bone and tissue16,22,50. For low intensity neuromodulation, in general, ultrasound intensity intracranially is estimated to be attenuated ~3–4 fold from values measured in free water16 and would thus produce Ispta values under the FDA recommended limits for obstetric diagnostics (720 mW/cm2)37.
The experiments documented here had a rather small range of Isppa though used considerably different number of stimulations and different inter-stimulus intervals that would contribute to overall exposure and may influence potential hazard. Indeed, Lee et al. (2016) found that a high number of total stimulations (600) with a low ISI (1 s) resulted in evidence of microhemorrhage in sheep despite the intensity being rather low at 6.6 W/cm2,9. We found no relation of number of stimulations with symptom prevalence though there were other variables that were also manipulated such as the inter-stimulus interval. It is currently unclear what constitutes a high number of total stimulations or a low ISI though taking these experimental parameters into consideration in the planning of experimental design is prudent. Additional metrics like energy density (J/cm2) as has been used in ultrasound sonoporation51 and neuromodulation parameter4 studies, as well as the total experimental energy density (J/cm2) that takes into account intensity, duty cycle as well as total number of stimulations and the ISI would prove an additional valuable safety metric given the results of Lee et al.9 though the relation of total experimental energy density to hazard is not well understood.
In addition to our group, Seung-Schik Yoo’s lab has performed multiple human ultrasound neuromodulation studies and completed thorough safety analysis including similar telephone follow-up as well as neurological assessment pre and post experiment including anatomical MRI and reported zero events in their three studies19,20,22. In the published literature and including the experiments in this study, a total of 260 individuals have participated in human ultrasound neuromodulation experiments to date with no reported serious adverse events and the data from this report is the first to report on minor transient events associated with the LIFU intervention. Caution is always advised when imparting energy into the brain and further research is recommended examining the effect of total number of stimulations and the inter-stimulus interval and the potential interaction of these parameters with intensity and duty cycle. Consideration of these parameters should be undertaken in experimental design to keep total experimental energy levels as low as reasonably achievable. Despite the differences in total number of stimulations and ISI in the experiments reported here, there were no large differences in response rate or in the type of report of symptoms between the seven experiments. It should be noted here too, that the experiments included in this report used additional technologies such as EEG, MRI and TMS in addition to LIFU that could potentially account for some of the positive findings. To this point, in experiments that employed TMS in addition to LIFU, we cannot tease out if symptoms were related to TMS or LIFU interventions. However, the fact that there isn’t a large difference in total symptoms for these studies as compared to the studies that just used LIFU suggests that there are not compounding or additive symptomology with concurrent forms of neuromodulation. To provide additional context to the safety assessments of ultrasound neuromodulation, we have compared our experiments’ symptom response rate and symptom profile to other forms of non-invasive human neuromodulation, including repetitive transcranial magnetic stimulation (rTMS), deep transcranial magnetic stimulation (dTMS), and transcranial direct current stimulation (tDCS).
The overall symptom response rate and type of symptoms is similar but without follow-up with all participants, it is possible that some additional symptoms might have occurred and were not included in the analysis. It should be noted here however, that our comparative analysis has some significant limitations. Firstly, we utilized highly restrictive exclusion criteria in our literature review. By excluding protocols that failed to report at least one event following neuromodulation, the calculated incidences presented may be greater than expected. Secondly, because we aimed to calculate overall incidences for symptoms across subjects, we excluded several studies that provided incidences across treatment sessions, further limiting the data. Thirdly, this study was one of the few that described the extent to which subjects attributed symptoms to the treatment itself though participant subjective assessment of the relation of symptoms to the intervention is speculative. It would be prudent for future studies to include physical examinations by a physician at multiple regular intervals to clarify patterns of symptoms for safety guidelines.
Conclusions
We provide qualitative data of reported symptoms from seven separate experiments from a single research team using different experimental protocols and energy levels of ultrasound for human neuromodulation as assessed by participant report of symptom questionnaire. Symptom rate and type were similar across studies and are similar to other forms of human non-invasive neuromodulation like TMS and tDCS that have a long-standing history of being safe forms of human neuromodulation.
Material and methods
Participants
All experiments were conducted with the approval of the Institutional Review Board at the University of Minnesota. All research was performed in accordance with relevant guidelines/regulations and all participants provided written informed consent to participate. The data presented here is from a one laboratory. A total of 120 volunteer study participants (48 male, 72 female aged 18–38 with a mean age of 22.96 ± 2.14 years) provided written informed consent to participate in one or more of the seven experiments from which the data for this study is taken between 2015 and 2017 at the University of Minnesota. Prior to formal experimental procedures, participants were screened via questionnaire for contraindications to non-invasive neuromodulation and none of the participants reported any neurological impairment or identified any contraindications to non-invasive neuromodulation as outlined by Rossi et al. (2009) as identified for transcranial magnetic stimulation52.
Experiments
The data for this study is a summary of 64 individual participants that participated in one of seven completed experiments conducted in our lab in the Department of Rehabilitation Medicine at the University of Minnesota. Details of the specific objectives or hypotheses of each study are not elaborated upon though it was the purpose of all studies to assess how LIFU targeting either the primary motor cortex or the thalamus affected physiological markers of neuronal activity and behaviour. Specifics of the methods can be found in Legon et al. (2018a) and Legon et al. (2018b). We have included details on the LIFU application including transducer specifics, target (cortical, sub-cortical) and parameters (amplitude, duration, etc.) are enumerated. For the purposes of this report, experiments will be referred to by number (1–7) based upon chronological date of commencement. All experiments were conducted in neurologically healthy volunteer participants to test the effect of LIFU on either cortical or sub-cortical neuronal excitability and/or effect on specific behaviours. The environment of the experiments differed as one experiment (Experiment 4)26 was conducted in a 7 T MRI scanner at the Center for Magnetic Resonance Research at the University of Minnesota (https://www.cmrr.umn.edu) and experiment 6 and 7 also involved transcranial magnetic stimulation (TMS) either concurrent with LIFU or as a pre/post measure of motor cortical excitability21. For all experiments (except fMRI experiment), participants were seated in a dentist-type chair and asked to either perform a simple task or sit passively for the duration of the experimental protocol. Tasks included a sensory discrimination task16 and simple stimulus response tasks on a computer. All experiments were repeated measures design with either one of or both an active or passive sham along with the LIFU condition. See Table 1 for experiment recruitment totals, participant demographics, and cited studies for individual experimental design and targeting details.
Questionnaire and follow-up
For all experiments, participants were retrospectively contacted via email at random time intervals (1 week–22 months) post experiment for their willingness to participate in a follow-up questionnaire on their experience of undergoing LIFU neuromodulation and perceived relation of any reported symptoms to LIFU. For all experiments, participants were contacted via email only once. For experiment 7, participants filled out the questionnaire immediately (~20 minutes after LIFU): (This is designated as 0 months in Figures and following text) after experimentation with follow-up at one of four time points post-experiments (1 week, 2 weeks, 3 weeks and 1 month). Those participants who responded affirmatively via email were subsequently contacted via telephone and asked to respond to 20 questions regarding their subjective assessment of their current neurological health (see Supplementary Material for questionnaire). This questionnaire is a variant of the Participant Report of Symptoms questionnaire that has previously been used in other non-invasive neuromodulation studies45,46. If there was a positive response to a question indicating perceived experience of the symptom, participants were then asked to rank the symptom severity from 2–4 (1 = absent) where 2 = mild, 3 = moderate and 4 = severe. In addition, participants were asked for their subjective assessment of the relation of the symptom to their involvement in the ultrasound experiments. Potential responses were: 1 = unrelated, 2 = unlikely, 3 = possible, 4 = probable and 5 = definite. In instances of positive subjective report – each case was referred to a neurologist (G.M) for medical record review and assessment of reported symptoms. Participants were remunerated for their participation in this telephone interview session. Total phone call discussion time ranged from 5–10 minutes. All phone calls were conducted by one of two lab investigators.
Transcranial focused ultrasound
For all experiments, the LIFU condition involved acoustically coupling the active face of the ultrasound transducer to the scalp at the pre-determined site depending upon the target of interest. Targeting was done using individual participant structural MRI scans paired with a neuronavigation system (Brainsight, Rogue Research). neuronavigation was used for localization for LIFU targets. The safety and comfort of participants was monitored through self-report during LIFU. No cavitation monitoring or temperature measurements were completed for these experiments The passive sham condition involved either placing a high acoustic impedance disk on the face of the transducer (Experiments 1 and 2)17, flipping the transducer over (while on) or simply turning it off during collection (experiment 4 MRI)26. For active sham conditions, ultrasound was delivered to another brain region (Experiment 3)21. Shamming maintained contact of the transducer to the head to mimic the audible sensation of a slight buzzing but attenuate any energy into the head. The audible sound was identical for sham and LIFU conditions and no subjects reported any sensory or perceptual differences between sham and LIFU conditions as previously reported16. This identical audible sound for sham and LIFU was designed to mitigate potential effects of indirect activation of auditory pathways53,54. The active sham condition, when employed, involved delivering LIFU to another scalp site (vertex) with the same parameters as the experimental site21. All experiments employed a repeated measures design where the participant received LIFU and sham stimulation in the same session regrettably making a disassociation of symptoms between real and sham stimulation not possible.
LIFU waveforms
All experiments used a single element 0.5 MHz transducer. Transcranial ultrasonic neuromodulation waveforms were generated using a two-channel, 2-MHz function generator (BK 4078B Precision Instruments). Channel 1 was used to gate channel 2 that was a 500 kHz sine wave. Channel 1 is the pulse repetition frequency (PRF) and was a 5Vp-p square wave burst of 1 kHz (N = 500) resulting in a 0.5 second total duration. The duty cycle was 36%; the period is 1/PRF resulting in a single pulse duration of 360 µs. The output of channel 2 was sent through a 100-W linear RF amplifier (E&I 2100 L; Electronics & Innovation) before being sent to the custom-designed focused ultrasound transducer. A total of 3 different transducers were used across the 7 experiments. Experiments 1 and 2 used the same transducer; experiments 3, 4 and 7 used the same transducer and experiments 5 and 6 used the same transducer. See Table 3 for ultrasound beam parameters and focality, and Table 4 for transducer specifications, including intensity and peak negative pressure for each study.
Table 3.
Experiment | Af (kHz) | Aperture (mm) | Focal length (mm) | F# | Beam Geometry (−3dB) | ||
---|---|---|---|---|---|---|---|
X (mm) | Y (mm) | Z (mm) | |||||
1 | 500 | 63 | 71 | 1.13 | 3.6 | 3.6 | 19.6 |
2 | 500 | 63 | 71 | 1.13 | 3.6 | 3.6 | 19.6 |
3 | 500 | 30 | 24 | 0.8 | 1.8 | 1.7 | 25.2 |
4 | 500 | 30 | 24 | 0.8 | 1.8 | 1.7 | 25.2 |
5 | 500 | 30 | 25 | 0.83 | 2.1 | 2.2 | 25.4 |
6 | 500 | 30 | 25 | 0.83 | 2.1 | 2.2 | 25.4 |
7 | 500 | 30 | 24 | 0.8 | 1.8 | 1.7 | 25.2 |
Table 4.
Experiment | # of stimulations | ISI (sec) | Isppa (W/cm2) | Ispta (W/cm2) | MI |
---|---|---|---|---|---|
1 | 300 | 4 | 14.56 | 5.24 | 0.89 |
2 | 90 | 8 | 14.56 | 5.24 | 0.89 |
3 | 200 | 4 | 11.56 | 4.16 | 0.78 |
4 | 54 | 5.5 | 11.56 | 4.16 | 0.78 |
5 | 80 | 10 | 17.12 | 6.16 | 0.9 |
6 | 150 | 10 | 17.12 | 6.16 | 0.9 |
7 | 44 | 4.75 | 11.56 | 4.16 | 0.78 |
Comparison of LIFU symptoms to other non-invasive brain stimulation modalities
To better assess and relate the reported symptoms from LIFU we compared our results to reported symptoms from other non-invasive brain stimulation modalities including repetitive transcranial magnetic stimulation (rTMS), deep TMS (dTMS) and transcranial electric stimulation (TES). To do so, we performed a literature search on the PubMED database, identifying 22 publications pertaining to rTMS safety in healthy subjects (from 1992 to 2013), 7 publications pertaining to dTMS safety in healthy subject (from 2004 to 2018), and 7 publications pertaining to tDCS safety in healthy subjects (from 2004 to 2015). We utilized the following search criteria: (transcranial magnetic stimulation) AND safety AND Humans; (transcranial magnetic stimulation) AND deep OR H-coil AND Humans; (transcranial direct current simulation) AND safety AND Humans. Articles were excluded if they did not report at least one minor adverse event following neuromodulation; did not include healthy subjects; did not report the total number of healthy subjects or did not report total of incidences of symptoms. Studies focusing on pediatric or geriatric populations were also excluded, as were single patient case studies. Associated article references were also reviewed and included in the analysis if they met the preceding criteria. The number of healthy subjects, incidence of specific symptoms, and pertinent stimulation parameters were obtained from each study. Overall symptom incidences were calculated by summating incidences across all studies for each rTMS, dTMS, and tDCS.
Supplementary information
Acknowledgements
The authors would like to thank Jeff Elias for insightful comments on the manuscript.
Author contributions
W.L., B.T.G. conceived of the study, wrote the manuscript, oversaw data collection and analyzed data. S.A., P.B., P.D.P., L.H., G.M. helped with manuscript preparation and data collection. P.B., L.A., J.K.M. collected and analyzed data
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
is available for this paper at 10.1038/s41598-020-62265-8.
References
- 1.Tufail Y, et al. Transcranial pulsed ultrasound stimulates intact brain circuits. Neuron. 2010;66:681–694. doi: 10.1016/j.neuron.2010.05.008. [DOI] [PubMed] [Google Scholar]
- 2.Mehic E, et al. Increased anatomical specificity of neuromodulation via modulated focused ultrasound. PLoS One. 2014;9:e86939. doi: 10.1371/journal.pone.0086939. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.King RL, Brown JR, Newsome WT, Pauly KB. Effective parameters for ultrasound-induced in vivo neurostimulation. Ultrasound Med. Biol. 2013;39:312–331. doi: 10.1016/j.ultrasmedbio.2012.09.009. [DOI] [PubMed] [Google Scholar]
- 4.Kim H, Chiu A, Lee SD, Fischer K, Yoo SS. Focused ultrasound-mediated non-invasive brain stimulation: examination of sonication parameters. Brain Stimul. 2014;7:748–756. doi: 10.1016/j.brs.2014.06.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Min BK, et al. Focused ultrasound-mediated suppression of chemically-induced acute epileptic EEG activity. BMC Neurosci. 2011;12:23-2202–12-23. doi: 10.1186/1471-2202-12-23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Younan Y, et al. Influence of the pressure field distribution in transcranial ultrasonic neurostimulation. Med. Phys. 2013;40:082902. doi: 10.1118/1.4812423. [DOI] [PubMed] [Google Scholar]
- 7.Yang PS, et al. Transcranial focused ultrasound to the thalamus is associated with reduced extracellular GABA levels in rats. Neuropsychobiology. 2012;65:153–160. doi: 10.1159/000336001. [DOI] [PubMed] [Google Scholar]
- 8.Yoo SS, et al. Focused ultrasound modulates region-specific brain activity. Neuroimage. 2011;56:1267–1275. doi: 10.1016/j.neuroimage.2011.02.058. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Lee W, et al. Image-Guided Focused Ultrasound-Mediated Regional Brain Stimulation in Sheep. Ultrasound Med. Biol. 2016;42:459–470. doi: 10.1016/j.ultrasmedbio.2015.10.001. [DOI] [PubMed] [Google Scholar]
- 10.Yoon K, et al. Effects of sonication parameters on transcranial focused ultrasound brain stimulation in an ovine model. PLoS One. 2019;14:e0224311. doi: 10.1371/journal.pone.0224311. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Dallapiazza, R. F. et al. Noninvasive neuromodulation and thalamic mapping with low-intensity focused ultrasound. J. Neurosurg., 1–10 (2017). [DOI] [PMC free article] [PubMed]
- 12.Folloni D, et al. Manipulation of Subcortical and Deep Cortical Activity in the Primate Brain Using Transcranial Focused Ultrasound Stimulation. Neuron. 2019;101:1109–1116.e5. doi: 10.1016/j.neuron.2019.01.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Verhagen, L. et al. Offline impact of transcranial focused ultrasound on cortical activation in primates. Elife8, 10.7554/eLife.40541 (2019). [DOI] [PMC free article] [PubMed]
- 14.Deffieux T, et al. Low-intensity focused ultrasound modulates monkey visuomotor behavior. Curr. Biol. 2013;23:2430–2433. doi: 10.1016/j.cub.2013.10.029. [DOI] [PubMed] [Google Scholar]
- 15.Wattiez N, et al. Transcranial ultrasonic stimulation modulates single-neuron discharge in macaques performing an antisaccade task. Brain Stimul. 2017;10:1024–1031. doi: 10.1016/j.brs.2017.07.007. [DOI] [PubMed] [Google Scholar]
- 16.Legon W, et al. Transcranial focused ultrasound modulates the activity of primary somatosensory cortex in humans. Nat. Neurosci. 2014;17:322–329. doi: 10.1038/nn.3620. [DOI] [PubMed] [Google Scholar]
- 17.Legon W, Ai L, Bansal P, Mueller JK. Neuromodulation with single-element transcranial focused ultrasound in human thalamus. Hum. Brain Mapp. 2018;39:1995–2006. doi: 10.1002/hbm.23981. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hameroff S, et al. Transcranial ultrasound (TUS) effects on mental states: a pilot study. Brain Stimul. 2013;6:409–415. doi: 10.1016/j.brs.2012.05.002. [DOI] [PubMed] [Google Scholar]
- 19.Lee W, et al. Image-guided transcranial focused ultrasound stimulates human primary somatosensory cortex. Sci. Rep. 2015;5:8743. doi: 10.1038/srep08743. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Lee W, Chung YA, Jung Y, Song IU, Yoo SS. Simultaneous acoustic stimulation of human primary and secondary somatosensory cortices using transcranial focused ultrasound. BMC Neurosci. 2016;17:68. doi: 10.1186/s12868-016-0303-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Legon W, Bansal P, Tyshynsky R, Ai L, Mueller JK. Transcranial focused ultrasound neuromodulation of the human primary motor cortex. Sci. Rep. 2018;8:10007-018–28320-1. doi: 10.1038/s41598-018-28320-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Lee W, et al. Transcranial focused ultrasound stimulation of human primary visual cortex. Sci. Rep. 2016;6:34026. doi: 10.1038/srep34026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Monti MM, Schnakers C, Korb AS, Bystritsky A, Vespa PM. Non-Invasive Ultrasonic Thalamic Stimulation in Disorders of Consciousness after Severe Brain Injury: A First-in-Man Report. Brain Stimul. 2016;9:940–941. doi: 10.1016/j.brs.2016.07.008. [DOI] [PubMed] [Google Scholar]
- 24.Mueller J, Legon W, Opitz A, Sato TF, Tyler WJ. Transcranial focused ultrasound modulates intrinsic and evoked EEG dynamics. Brain Stimul. 2014;7:900–908. doi: 10.1016/j.brs.2014.08.008. [DOI] [PubMed] [Google Scholar]
- 25.Ai, L., Mueller, J. K., Bansal, P. & Legon, W. Transcranial focused ultrasound for BOLD fMRI signal modulation in humans. EMBC, 1758–1761 (2016). [DOI] [PubMed]
- 26.Ai L, Bansal P, Mueller JK, Legon W. Effects of transcranial focused ultrasound on human primary motor cortex using 7T fMRI: a pilot study. BMC Neurosci. 2018;19:56-018–0456-6. doi: 10.1186/s12868-018-0456-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.White PJ, Clement GT, Hynynen K. Longitudinal and shear mode ultrasound propagation in human skull bone. Ultrasound Med. Biol. 2006;32:1085–1096. doi: 10.1016/j.ultrasmedbio.2006.03.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Plaksin, M., Kimmel, E. & Shoham, S. Cell-Type-Selective Effects of Intramembrane Cavitation as a Unifying Theoretical Framework for Ultrasonic Neuromodulation. eNeuro3, 10.1523/ENEURO.0136-15.2016. eCollection 2016 May-Jun (2016). [DOI] [PMC free article] [PubMed]
- 29.Krasovitski B, Frenkel V, Shoham S, Kimmel E. Intramembrane cavitation as a unifying mechanism for ultrasound-induced bioeffects. Proc. Natl. Acad. Sci. USA. 2011;108:3258–3263. doi: 10.1073/pnas.1015771108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Tyler WJ. Noninvasive neuromodulation with ultrasound? A continuum mechanics hypothesis. Neuroscientist. 2011;17:25–36. doi: 10.1177/1073858409348066. [DOI] [PubMed] [Google Scholar]
- 31.Tyler WJ. The mechanobiology of brain function. Nat. Rev. Neurosci. 2012;13:867–878. doi: 10.1038/nrn3383. [DOI] [PubMed] [Google Scholar]
- 32.Mueller, J. K., Ai, L., Bansal, P. & Legon, W. Computational exploration of wave propagation and heating from transcranial focused ultrasound for neuromodulation. J. Neural Eng. 13, 056002–2560/13/5/056002. Epub 2016 Jul 28 (2016). [DOI] [PubMed]
- 33.Elias WJ, et al. A pilot study of focused ultrasound thalamotomy for essential tremor. N. Engl. J. Med. 2013;369:640–648. doi: 10.1056/NEJMoa1300962. [DOI] [PubMed] [Google Scholar]
- 34.Lipsman N, et al. MR-guided focused ultrasound thalamotomy for essential tremor: a proof-of-concept study. Lancet Neurol. 2013;12:462–468. doi: 10.1016/S1474-4422(13)70048-6. [DOI] [PubMed] [Google Scholar]
- 35.Downs ME, et al. Long-Term Safety of Repeated Blood-Brain Barrier Opening via Focused Ultrasound with Microbubbles in Non-Human Primates Performing a Cognitive Task. PLoS One. 2015;10:e0125911. doi: 10.1371/journal.pone.0125911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.McDannold N, Vykhodtseva N, Hynynen K. Targeted disruption of the blood-brain barrier with focused ultrasound: association with cavitation activity. Phys. Med. Biol. 2006;51:793–807. doi: 10.1088/0031-9155/51/4/003. [DOI] [PubMed] [Google Scholar]
- 37.http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM070911.pdf.
- 38.Duck FA. Medical and non-medical protection standards for ultrasound and infrasound. Prog. Biophys. Mol. Biol. 2007;93:176–191. doi: 10.1016/j.pbiomolbio.2006.07.008. [DOI] [PubMed] [Google Scholar]
- 39.Shaw A, ter Haar G, Haller J, Wilkens V. Towards a dosimetric framework for therapeutic ultrasound. Int. J. Hyperthermia. 2015;31:182–192. doi: 10.3109/02656736.2014.997311. [DOI] [PubMed] [Google Scholar]
- 40.O’Brien WD., Jr. Ultrasound-biophysics mechanisms. Prog. Biophys. Mol. Biol. 2007;93:212–255. doi: 10.1016/j.pbiomolbio.2006.07.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Mesiwala AH, et al. High-intensity focused ultrasound selectively disrupts the blood-brain barrier in vivo. Ultrasound Med. Biol. 2002;28:389–400. doi: 10.1016/S0301-5629(01)00521-X. [DOI] [PubMed] [Google Scholar]
- 42.Vykhodtseva NI, Hynynen K, Damianou C. Histologic effects of high intensity pulsed ultrasound exposure with subharmonic emission in rabbit brain in vivo. Ultrasound Med. Biol. 1995;21:969–979. doi: 10.1016/0301-5629(95)00038-S. [DOI] [PubMed] [Google Scholar]
- 43.Dunn F, Fry FJ. Ultrasonic threshold dosages for the mammalian central nervous system. IEEE Trans. Biomed. Eng. 1971;18:253–256. doi: 10.1109/TBME.1971.4502847. [DOI] [PubMed] [Google Scholar]
- 44.Ulrich WD. Ultrasound dosage for nontherapeutic use on human beings–extrapolations from a literature survey. IEEE Trans. Biomed. Eng. 1974;21:48–51. doi: 10.1109/TBME.1974.324362. [DOI] [PubMed] [Google Scholar]
- 45.Gillick, B. et al. Transcranial direct current stimulation and constraint-induced therapy in cerebral palsy: A randomized, blinded, sham-controlled clinical trial. Eur. J. Paediatr. Neurol. (2018). [DOI] [PMC free article] [PubMed]
- 46.Gillick, B. T. et al. Non-Invasive Brain Stimulation in Children With Unilateral Cerebral Palsy: A Protocol and Risk Mitigation Guide. Frontiers in pediatrics March (2018). [DOI] [PMC free article] [PubMed]
- 47.Deng ZD, Lisanby SH, Peterchev AV. Coil design considerations for deep transcranial magnetic stimulation. Clin. Neurophysiol. 2014;125:1202–1212. doi: 10.1016/j.clinph.2013.11.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Levkovitz Y, et al. A randomized controlled feasibility and safety study of deep transcranial magnetic stimulation. Clinical Neurophysiology. 2007;118:2730–2744. doi: 10.1016/j.clinph.2007.09.061. [DOI] [PubMed] [Google Scholar]
- 49.Lenoir, C. et al. Report of one confirmed generalized seizure and one suspected partial seizure induced by deep continuous theta burst stimulation of the right operculo-insular cortex. 11, 1187 (2018). [DOI] [PMC free article] [PubMed]
- 50.Mueller, J. K., Ai, L., Bansal, P. & Legon, W. Numerical evaluation of the skull for human neuromodulation with transcranial focused ultrasound. J. Neural Eng., 10.1088/1741-2552/aa843e (2017). [DOI] [PubMed]
- 51.Karshafian R, Bevan PD, Williams R, Samac S, Burns PN. Sonoporation by ultrasound-activated microbubble contrast agents: effect of acoustic exposure parameters on cell membrane permeability and cell viability. Ultrasound Med. Biol. 2009;35:847–860. doi: 10.1016/j.ultrasmedbio.2008.10.013. [DOI] [PubMed] [Google Scholar]
- 52.Rossi S, Hallett M, Rossini PM, Pascual-Leone A, Safety of TMS Consensus Group. Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research. Clin. Neurophysiol. 2009;120:2008–2039. doi: 10.1016/j.clinph.2009.08.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Sato, T., Shapiro, M. G. & Tsao, D. Y. Ultrasonic neuromodulation in vivo: discovery of a somato-auditory artifact and its implications. 808.1 (2017).
- 54.Guo H, et al. Ultrasound produces extensive brain activation via a cochlear pathway. Neuron. 2018;98:1020–1030. e4. doi: 10.1016/j.neuron.2018.04.036. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.