Skip to main content
Nagoya Journal of Medical Science logoLink to Nagoya Journal of Medical Science
. 2020 Feb;82(1):79–84. doi: 10.18999/nagjms.82.1.79

Compatibility of magnetic resonance imaging in patients with orthopedic implants: manufacturer questionnaires

Katsuyuki Iwatsuki 1, Hidemasa Yoneda 1, Tetsuro Onishi 1, Hisao Ishii 1, Shigeru Kurimoto 1, Michiro Yamamoto 1, Masahiro Tatebe 1, Hitoshi Hirata 1
PMCID: PMC7103874  PMID: 32273635

ABSTRACT

In clinical practice, surgeons have stated that magnetic resonance imaging (MRI) can be performed in patients with titanium alloy implants. However, manufacturers and distributors of many implants may not comply with this common practice. As such, this study aimed to investigate manufacturers’ views on MRI use in patients fitted with their implants.

The questionnaire survey was conducted between May and August 2018.

Is your product compatible with MRI? ( ) Select from (1) to (3).

In case of (1) or (2), up to ( ) Tesla.

(1) MRI can be performed even at the sites of implanted fixators.

(2) MRI can be performed at sites without implanted fixators.

(3) MRI cannot be performed, or the manufacturer does not approve MRI use (cannot issue a certificate).

The questionnaire forms were sent to 12 manufacturers, and the response rate was 100%. Manufacturers responded that they could not publicly allow MRI use in patients with their products.

These findings do not conclude that MRI cannot be performed in such patients. This survey revealed that currently decisions regarding MRI use is left to the treating physicians. This situation poses a great problem for medical safety and imposes a substantial burden on physicians. As many problems remain in the field of orthopedic surgery, manufacturers of implants should proactively manage issues surrounding the usage of MRI.

Key Words: magnetic resonance imaging, implant, total arthroplasty, fracture, patient safety

INTRODUCTION

With advances in medical technology, magnetic resonance imaging (MRI) has recently become essential in clinical settings. In the field of orthopedic surgery, MRI allows assessment of bones, soft tissues, and nerves, and it is a diagnostic imaging alternative to arthrography and myelography. Although MRI is a useful modality that does not result in exposure to radiation, there are problems with cost and safety.1-3 MRI has been known to cause problems such as burn injuries, pain, and tinnitus. Instances of burn injuries have been reported in patients with implants, including pulse oximetry and Swan-Ganz thermodilution catheter devices.4,5 In addition, artifacts can also be a problem.1,2 However, as indicated by the advent of new cardiac pacemaker models, compatible with MRI examinations, technological innovation has driven the progress of implants.6-8 In the field of orthopedic surgery, many implants are now made of titanium alloy and considered to be compatible with MRI. The use of implants by orthopedic surgeons has been studied; however, many studies use MRI at field strengths up to 1.5 T only.3,9-11

In actual clinical practice, surgeons have stated that MRI can be performed in patients with titanium alloy Implants. However, only a few models, such as the Hoffmann III external fixation system (Stryker, Kalamazoo, MI), are sold specifically as MRI-compatible models, whereas the intentions of the majority of manufacturers and distributors of many implants remain unknown. Thus, this study aimed to investigate manufacturers’ views on MRI use in patients with their implants.

MATERIALS AND METHODS

A questionnaire survey was conducted for manufacturers distributing locking plates for distal radius fractures, nails for trochanteric fractures, and joint prostheses for the knees and hips between May and August 2018. As for products manufactured overseas and imported to Japan, the survey was conducted on foreign manufacturers through their Japanese branches or importers of their products. We chose major implant companies who conducted more than 75% of the research share on each implant in our country. We have no conflict of interest to declare. The contents of the questionnaire survey were as follows:

Views of manufacturers on MRI compatibility in patients with their implants

Is your product compatible with MRI? ( ) Select from (1) to (3).

In case of (1) or (2), up to ( ) Tesla.

(1) MRI can be performed even at the sites of implanted fixators.

(2) MRI can be performed at sites without implanted fixators.

(3) MRI cannot be performed, or the manufacturer does not approve MRI use (cannot issue a certificate).

RESULTS

The questionnaire survey was conducted from May to August 2018. The questionnaire forms were sent to 12 manufacturers, and the response rate was 100%. The following manufacturers were surveyed (manufacturer/Japanese distributers):

Acumed (U.S.A.) / Japan Medicalnext

DePuy Synthes, Johnson & Johnson

Kyocera

Medartis (Switzerland) / ME system

Meira

Mizuho

Next OrthoSurgical (U.S.A.) / NextMed International

Ortho Development Corporation (U.S.A.) / Japan Medical Dynamic Marketing

Smith & Nephew

Stryker

Teijin Nakashima Medical

Zimmer Biomet

Zimmer Biomet provided a list of its MRI-compatible products and the following comment instead of responding to the question on the survey:

“We have not evaluated our products that are not included in the table on the previous page (including plates for distal radius fractures, fixators for femoral fractures, and joint prostheses, which are targeted in this survey). Thus, we cannot vouch for the safety and compatibility of them.”

All other responding manufacturers selected option (3). In other words, they not state that their devices are compatible with MRI. The tables show responses according to types of implants (Table 1–4).

Table 1.

Implants for distal radius fractures

COMPANY IMPLANT ANSWER
Acumed (U.S.A.) /Japan Medicalnext Acu-Loc 2 Wrist Plating System 3
DePuy Synthes , Johnson & Johnson 2.4mm LCP Distal Radius Plate 3
Medartis(Switzerland) / ME system APTUS2.5 3
Meira DualLoc Radii system /
Distal Radius Plate System
3 Japan only
Mizuho Hearty Plate 3 Japan only
Next OrthoSurgical (U.S.A.) /
NextMed International
GlobalForm VDR Fixation System 3 Japan only
Ortho Development Corporation (U.S.A.) /
Japan Medical Dynamic Marketing
MODE Distal Radius Plate System 3 Japan only
Stryker VariAx 2 Distal Radius 3
Teijin Nakashima Medical Locking Volar Plate System 3 Japan only

Table 2.

Implants for trochanteric fractures

COMPANY IMPLANT ANSWER
DePuy Synthes , Johnson & Johnson TFN-ADVANCED / PFNA 3
ME system TURIUS Femoral Nail System 3 Japan only
Meira Hook Pin Nail 3 Japan only
Mizuho CHY II Nail 3 Japan only
Next OrthoSurgical (U.S.A.) /
NextMed International
Multi-Fix Femoral Nail 3 Japan only
Smith & Nephew Trigen intertan 3
Stryker Gamma3 3
Teijin Nakashima Medical Inter Blade Nail 3 Japan only

Table 3.

Implants for total hip arthroplasty

COMPANY IMPLANT ANSWER
DePuy Synthes , Johnson & Johnson CORAIL Hip System /
ACTIS Total Hip System etc.
3
Kyocera Initia Total Hip System 3 Japan only
Mizuho MX HIP SYSTEM 3 Japan only
Next OrthoSurgical (U.S.A.) /
NextMed International
NexFlex Bipolar System 3 Japan only
Ortho Development Corporation (U.S.A.) /
Japan Medical Dynamic Marketing
Ovation Hip System/
Escalede Hip System
3 Japan only
Smith & Nephew POLARSTEM / SL-PLUS MIA 3
Stryker Accolade II / Exeter 3
Teijin Nakashima Medical Preserve stem 3 Japan only

Table 4.

Implants for total knee arthroplasty

COMPANY IMPLANT ANSWER
DePuy Synthes , Johnson & Johnson ATTUNE Knee System 3
Kyocera Kyocera Modular Limb Salvage System 3 Japan only
Ortho Development Corporation (U.S.A.) /
Japan Medical Dynamic Marketing
Balanced Knee System 3 Japan only
Smith & Nephew JOURNEY II 3
Stryker Triathlon Total Knee System 3
Teijin Nakashima Medical FINE Total Knee System 3 Japan only

DISCUSSION

Due to a rapid increase in the mean lifespan of the Japanese population, recent medical advances, and a very low birthrate, the ratio of older adults to the total population is rising. Under these circumstances, MRI is used for diagnosing a wide range of diseases and disorders. As cases requiring emergency MRI due to cerebrovascular accidents and other disorders increase,12,13 physicians on the scene are placed in a situation in which they must immediately decide whether MRI can be performed, even though there are many other technologies available to assess brain and other areas.14 Patients with cardiac pacemakers use a pacemaker diary, which indicates whether the device is compatible with MRI or not, thus facilitating these decisions. In contrast, orthopedic implants do not have the same associated information, and physicians do not know which device has been used without contacting the hospital where the implantation was performed.6-8 However, many orthopedic implants are now made of titanium, therefore, MRI presumably causes no major problems. Thus, MRI is currently performed at the discretion of physicians in clinical practice.

Plates used for the co-aptation of bone fragments are typically removed in some countries, including Japan; however, other countries, such as the United States, do not remove them. Differences depending on healthcare systems and cultures have also been reported.15 Even among orthopedic co-aptation devices used in Japan, plates are often removed for those implanted in patients with distal radius fractures, but femoral trochanter nails are rarely removed. Furthermore, once joint prostheses are implanted, they are left in the body permanently. These devices are only removed if problems such as infection develop. Given these circumstances, we believe that guidelines regarding MRI usage in patients with orthopedic implants left in the body for long periods of time are required.

This study has several limitations. First, it is based on a questionnaire survey that did not include all implant manufacturers. Second, this study did not include manufacturers that are currently dealing with this issue or have no plan to do so. Third, the targeted implant types were limited.

CONCLUSION

In this questionnaire survey, manufacturers responded that they could not publicly endorse MRI use in patients with their products. These findings however do not conclude that MRI cannot be performed in such patients. In fact, MRI is frequently performed in orthopedic patients in clinical practice.16 In other words, this survey revealed the current situation in which decisions regarding MRI use is left to physicians on the scene. This situation poses problem with regards to medical safety and imposes substantial burden on physicians. As there are still many problems regarding the field of orthopedic surgery, it is hoped that manufacturers of implants will proactively deal with this MRI issue.

CONFLICT OF INTEREST

We have no conflict of interest to declare.

ACKNOWLEDGEMENTS

We appreciate all manufacturers cooperation.

Abbreviations

MRI

magnetic resonance imaging

REFERENCES

  • 1).Carlson ML, Neff BA, Link MJ, et al. Magnetic resonance imaging with cochlear implant magnet in place: safety and imaging quality. Otol Neurotol. 2015;36:965–971. doi: 10.1097/MAO.0000000000000666. [DOI] [PubMed]
  • 2).Yang CW, Liu L, Wang J, et al. Magnetic resonance imaging of artificial lumbar disks: safety and metal artifacts. Chin Med J (Engl). 2009;122:911–916. [PubMed]
  • 3).Kumar R, Lerski RA, Gandy S, Clift BA, Abboud RJ. Safety of orthopedic implants in magnetic resonance imaging: an experimental verification. J Orthop Res. 2006;24:1799–802. [DOI] [PubMed]
  • 4).Dempsey MF, Condon B. Thermal injuries associated with MRI. Clin Radiol. 2001;56:457–65. [DOI] [PubMed]
  • 5).Friedstat JS, Moore ME, Goverman J, Fagan SP. An unusual burn during routine magnetic resonance imaging. J Burn Care Res. 2013;34:e110–111. doi: 10.1097/BCR.0b013e3182642a40. [DOI] [PubMed]
  • 6).Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376:7557–64. doi: 10.1056/NEJMoa1603265. [DOI] [PubMed]
  • 7).Shinbane JS, Colletti PM, Shellock FG. Magnetic resonance imaging in patients with cardiac pacemakers: era of “MR Conditional” designs. J Cardiovasc Magn Reson. 2011;13:63. doi: 10.1186/1532-429X-13-63. [DOI] [PMC free article] [PubMed]
  • 8).Shurrab M, Kaoutskaia A, Baranchuk A, et al. Are there increased periprocedural complications with the MRI-conditional Medtronic Revo SureScan Pacing System? A meta-analysis. Neth Heart J. 2018;26:233–239. doi: 10.1007/s12471-018-1086-4. [DOI] [PMC free article] [PubMed]
  • 9).Zou YF, Chu B, Wang CB, Hu ZY. Evaluation of MR issues for the latest standard brands of orthopedic metal implants: plates and screws. Eur J Radiol. 2015;84:450–457. doi: 10.1016/j.ejrad.2014.12.001. [DOI] [PubMed]
  • 10).Shellock FG. Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system. J Magn Reson Imaging. 2002;16:721–732. [DOI] [PubMed]
  • 11).Gomez C, Nelson S, Speirs J, Barnes S. Magnetic intramedullary lengthening nails and MRI compatibility. Pediatr Orthop. 2018;38:e584–587. doi: 10.1097/BPO.0000000000001236. [DOI] [PubMed]
  • 12).Iwatsuki K, Deguchi M, Hirata H, Kanamono T. Spontaneously resolved recurrent cervical epidural hematoma in a 37-week primigravida. Global Spine J. 2015;5:e44–47. doi: 10.1055/s-0034-1398489. [DOI] [PMC free article] [PubMed]
  • 13).Takeda, S, Iwatsuki, K, Takeichi, Y, Kano, T, Tabuchi, A, Hirata, H. A man with severe right knee pain. Clin Pract Cases Emerg Med. 2019;3:168–169. doi: 10.5811/cpcem.2019.1.41417. [DOI] [PMC free article] [PubMed]
  • 14).Iwatsuki K, Hoshiyama M, Oyama S, Shimoda S, Hirata H. Magnetoencephalographic evaluation for the myoelectric hand prosthesis with tacit learning system. NeuroRehabilitation. 2019;44:19–23. doi: 10.3233/NRE-182514. [DOI] [PubMed]
  • 15).Iwatsuki K, Tatebe M, Yamamoto M, Shinohara T, Nakamura R, Hirata H. Ulnar impaction syndrome: incidence of lunotriquetral ligament degeneration and outcome of ulnar-shortening osteotomy. J Hand Surg Am. 2014;39:1108–1113. doi: 10.1016/j.jhsa.2014.03.006. [DOI] [PubMed]
  • 16).Mosher ZA, Sawyer JR, Kelly DM. MRI Safety with Orthopedic Implants. Orthop Clin North Am. 2018;49:455–463. doi: 10.1016/j.ocl.2018.05.010. [DOI] [PubMed]

Articles from Nagoya Journal of Medical Science are provided here courtesy of Nagoya University School of Medicine/Graduate School of Medicine

RESOURCES