Skip to main content
Elsevier - PMC COVID-19 Collection logoLink to Elsevier - PMC COVID-19 Collection
. 2009 Feb 19;46(2):71–75. doi: 10.1016/S1875-4597(08)60029-6

Sunscope: A Video-guided Intubation System Through a Detachable Imaging Probe

Jia-Rong Yeh a, Jiann-Shing Shieh a, Chih-Peng Lin b, Wei-Zen Sun b,*
PMCID: PMC7128121  PMID: 18593652

Abstract

We have designed a novel apparatus, the Sunscope, which integrates a semiconductor image sensor into a compact video-guided intubation system. This device consists of three separate modules: viewer, console and visual tube. The 4-inch LCD viewer panel displays the real-time video image with optimal view angle. The console is designed with respect to ergonomics allowing comfortable manipulation and internally accommodating the power supply, image processing components and connector platform for both viewer and probe. The distal end of the detachable probe is packaged with a high resolution lens, CMOS sensor, and four LEDs. The proximal end is a 6-pin connector which can be readily removed and attached on demand. The probe is detachable and disposable with length and diameter adaptable to the size of the endotracheal tube. In our preliminary test, the video-guided apparatus helped inexperienced performers to identify the vocal cords correctly and improve the success rate of intubation on the simulation model. With further improvements on the miniature design, all captured images could be transmitted to remote devices through standard wireless transmission and could thus be stored in a specific database station. The wireless technique enables image sharing on multiple devices while a powerful database can provide valuable resources for training, data mining and serial case studies. We demonstrate that the CMOS image sensor combined with advanced reduced instruction set computer machine can serve as a visual aid for tracheal intubation. The disposable station will become a revolutionary technology both in clinical practice and medical education.

KEY WORDS: disposable equipment; equipment design: CMOS image sensor (CIS); equipment design: reduced instruction set computer (RISC); intubation, intratracheal; Sunscope

References

  • 1.Donnelly WH. Histopathology of endotracheal intubation: an autopsy study of 99 cases. Arch Pathol. 1969;88:511–520. [PubMed] [Google Scholar]
  • 2.Samsoon GL, Young JRB. Difficult tracheal intubation. Anaesthesia. 1987;42:487–490. doi: 10.1111/j.1365-2044.1987.tb04039.x. [DOI] [PubMed] [Google Scholar]
  • 3.Lasinska-Kowara M, Sulkowski B, Wujtewicz M. Thyromental distance as a predictor of difficult intubation. Anaesthesiol Intensive Ther. 2007;1:8–12. [Google Scholar]
  • 4.Hung OR, Pytka S, Morris I, Murphy M, Launcelott G, Stevens S, MacKay W. Clinical trial of a new lightwand device (Trachlight) to intubate the trachea. Anesthesiology. 1995;83:509–514. doi: 10.1097/00000542-199509000-00009. [DOI] [PubMed] [Google Scholar]
  • 5.Hung OR, Pytka S, Morris I, Murphy M, Stewart RD. Lightwand intubation: II—clinical trial of a new lightwand for tracheal intubation in patients with difficult airways. Can J Anaesth. 1995;42:826–830. doi: 10.1007/BF03011187. [DOI] [PubMed] [Google Scholar]
  • 6.Weiss M. The video-intubating laryngoscope. The Internet Journal of Anesthesiology. 1999;3:N1. [Google Scholar]
  • 7.Bigas M, Cabruja E, Forest J, Salvi J. Review of CMOS image sensors. Microelectronics J. 2006;37:433–452. [Google Scholar]
  • 8.Kempainen S. CMOS image sensors: eclipsing CCDs in visual information? EDN. 1997;42:21. [Google Scholar]
  • 9.Chen CH, Tsai HJ, Huang KS, Liu HT. Study for cross contamination between cmos image sensor and ic product. IEEE/SEMI Advanced Semiconductor Manufacturing Conference. 2001:121–123. [Google Scholar]
  • 10.Wong HSP. CMOS image sensors—recent advances and device scaling considerations. International Electron Devices Meeting. 1997:201–204. [Google Scholar]
  • 11.Hillebrand M, Stevanovic N, Hosticka B, Conde J, Teuner A, Schwarz M. High speed camera system using a cmos image sensor. Proceedings of the IEEE Intelligent Vehicles Symposium. 2000:656–661. [Google Scholar]
  • 12.Chen TH, Tsai SK, Lin CJ, Lu CW, Tsai TP, Sun WZ. Does the suggested lightwand bent length fit every patient? Anesthesiology. 2003;98:1070–1076. doi: 10.1097/00000542-200305000-00008. [DOI] [PubMed] [Google Scholar]

Articles from Acta Anaesthesiologica Taiwanica are provided here courtesy of Elsevier