Abstract
Exosomes are endocytic lipid-membrane bound bodies with the potential to be used as biomarkers in cancer and neurodegenerative disease. The limitations and scarcity of current exosome characterization approaches have led to a growing demand for translational techniques, capable of determining their molecular composition and physical properties in physiological fluids. Here, we investigate label-free immunosensing, using a quartz crystal microbalance with dissipation monitoring (QCM-D), to detect exosomes by exploiting their surface protein profile. Exosomes expressing the transmembrane protein CD63 were isolated by size-exclusion chromatography from cell culture media. QCM-D sensors functionalized with anti-CD63 antibodies formed a direct immunoassay toward CD63-positive exosomes in 75% v/v serum, exhibiting a limit-of-detection of 2.9 × 108 and 1.4 × 108 exosome sized particles (ESPs)/mL for frequency and dissipation response, respectively, i.e., clinically relevant concentrations. Our proof-of-concept findings support the adoption of dual-mode acoustic analysis of exosomes, leveraging both frequency and dissipation monitoring for use in bioanalytical characterization.
Introduction
Extracellular vesicles (EVs) are heterogeneous, biomolecular structures enclosed by a lipid bilayer. They are secreted by nearly all eurkaryotic cells into the extracellular space and most bodily fluids.1 Of particular interest are exosomes, a subset of EVs with a nanoscale size range (30–150 nm) originating from invaginations of early endosomes and released upon the fusion of multivesicular bodies with the cell membrane.2 They are enriched in nucleic acids, surface proteins such as tetraspannins (CD63, CD81, and CD9), and cytosolic proteins including heat shock proteins (HSP90 and HSP70) and TSG101.3,4
Traditionally thought to function as cellular waste bins, the roles of exosomes in intercellular communication,5 disease propagation, and regenerative processes are now well established.6,7 Crucially, exosome concentrations and phenotype have been shown to vary between healthy and diseased states, reflecting their parental cell of origin.8,9 Thus, exosomes have attracted widespread interest as a concentrated source of biomarkers for minimally invasive, point-of-care liquid biopsies.10,11
Typically, exosomes are characterized via nanoparticle tracking analysis (NTA). Here, the imaging of light scattered from particles moving under Brownian diffusion is used to determine the hydrodynamic size and concentration.12 Alternatively, tunable elastomeric pore sensing analyzes individual particles via the electrical impedance they impart at an aperture.13 These methods are often coupled with total protein quantification via colorimetric assays such as microBCA and Bradford.14 One limitation of the above techniques is that they do not selectively distinguish between exosomes and other EVs, protein aggregates, and lipoproteins. This lack of discrimination is compounded by the choice of exosome isolation technique, where commonly adopted centrifugation and polymer precipitation methods coisolate nonexosomal artifacts from complex media.15 Thus, there is a difficulty in defining subsets within a heterogeneous exosome population, which hinders these techniques in sensing specific markers in complex biological matrices.16
By contrast, flow cytometry17,18 and fluorescence-based NTA have been successfully employed to quantify exosomes and determine their phenotypes via selective tagging of their surface epitopes.19 Nonetheless, labeling approaches are restricted by the strength of interaction between the label and exosome. Furthermore, these techniques are largely destructive, limiting downstream application of the analyte. Enzyme-linked immunosorbent assay (ELISA) is the current gold standard for exosomal protein quantification, with sensitivity in the picomolar range.20 However, traditional ELISAs can suffer from a lack of multiplexing, cross-contamination, and limited potential for point-of-care application. Recently, Ren et al. introduced an enzyme-free colorimetric immunoassay toward alpha-fetoprotein (AFP), using an antibody-labeled metal-polydopamine framework that displayed sensitivity down to 2.3 pg mL–1.21 An alternative approach with similar sensitivity (5.3 pg mL–1) was devised by the same group via near-infrared excitation of nanospheres as part of a photoelectrochemical enzyme immunoassay for AFP detection.22
There is increasing interest in automation and miniaturization of exosome screening through microfluidics and lab-on-a-chip approaches to match the clinical demand of minimally invasive patient stratification.23,24 Examples of advanced exosomal analytical approaches include interferometry,25 electrochemistry,26,27 and optical sensors utilizing nanoplasmonics.28,29 Recently, Rupert et al. successfully demonstrated surface plasmon resonance (SPR) based sensing of CD63-positive exosomes through surface based immunocapture.30 Collectively, the above-mentioned techniques provide a sensitive, label-free, and real-time assessment of exosomes. A potential drawback of these methods is the difficulty in distinguishing between exosome and artifactual binding phenomena.31,32 Qiu et al. was able to overcome background fluctuations and interference in a photoelectrochemical biosensor by using a ratiometric aptasensor, which spatially resolved dual signal readouts from two working electrodes.33 Recently, Yu et al. successfully employed a carbon-nanotube modified pressure electrode to discern between human serum biomarkers and the analyte of interest, carcinoembryonic antigen.34 This is an essential consideration, as not all circulating particles may be exosomal in composition, potentially leading to a false positive result if not appropriately distinguished from other colloidal contaminants.
To overcome the issue of specificity, this study employs quartz crystal microbalance with dissipation (QCM-D) monitoring, to leverage differences in mechanical properties between exosomes and associated contaminants in colloidal suspension. QCM-D is capable of characterizing interfacial structure, binding kinetics, molecular affinity, and mechanical properties of the adsorbent.35 Advantages of the acoustic technique include label-free, real-time measurements36 and ease of miniaturization.37 For a 5-MHz resonator in water, the acoustic wave possesses a penetration depth of around 250 nm from the oscillator surface, which matches the size of the exosomes.38 Nonetheless, QCM-D has yet to be fully exploited for exosome characterization. QCM-D transduced immunosensing has been explored to build assays where an antibody serves as a bioreceptor toward a target analyte. For instance, Uludag and Tothill demonstrated its applicability for the detection of a prostate specific antigen.39 Pirincci et al. evaluated a competitive immunoassay for the detection of mycotoxin Ochratoxin A.40 Tang et al. devised an immunoassay, which exploits the frequency change upon glucose displacement of bound concanavalin-A complex to detect brevetoxin down to 0.6 pg mL–1.41
In this work, we establish a direct immunoassay of CD63-positive exosomes using QCM-D. After successfully isolating and characterizing exosomes, we identify an optimal antibody immobilization approach. The sensor performance is then validated by assessing its sensitivity toward spiked samples of CD63 protein. We first evaluate the platform performance toward purified CD63 exosomes in buffer, as isolated by size-exclusion chromatography. We then determine the specificity, limit-of-detection (LOD), and limit-of-quantification (LOQ) of the approach, toward CD63 exosomes spiked in human serum, by comparing the target platform performance to a control surface. Our findings are validated by complementary in-liquid atomic force microscopy (AFM).
Experimental Section
Exosome Isolation and Characterization
Size-Exclusion Chromatography
Size-exclusion chromatography (SEC) was chosen as the isolation technique for exosomes from human umbilical cord mesenchymal stem cell culture media (HUMSCCM), based on previous work in the field.42 HUMSSCM was first filtered with a 0.45 μm filter (Merck Millipore, USA). Thirty milliliters of clarified media was subsequently concentrated using Amicon Ultra-15 centrifugal filters with a 10 kDa pore size cutoff (Merck Millipore, USA). The filters were spun at 4000g for 30 min at 4 °C. Postspin, 0.5 mL of concentrated filtrate was loaded onto a qEV SEC column (Izon Science, UK). Next, 0.2 μM filtered HEPES buffered saline (HBS, 0.01 M HEPES, pH 7.4, 0.15 M NaCl; GE Healthcare Life Sciences, Sweden) was used as the eluting buffer at a flow rate of 1 mL/min. Twenty 1 mL fractions were collected and stored at −80 ◦C.
NTA Analysis of SEC Fractions
The concentration and hydrodynamic size of particulates for each fraction were assessed by nanoparticle tracking analysis (NTA) with the Nanosight LM10 instrument (Malvern Instruments, UK). The machine was calibrated with 100 nm polystyrene beads (Thermofisher Scientific, UK) prior to fraction assessment. Measurement specifications were as follows: 532 nm green laser, five videos per fraction, 60 s video length, shutter speed of 25–32 ms, camera gain of 400, camera level 15, lower threshold of 910, and higher threshold of 11 180. Captured videos were processed with the NTA software (version 3.2), a detection threshold of 5, auto settings for blur, minimum track length, and minimum particle size. Measurements were carried out in static mode at room temperature.
Protein Content Analysis
Total protein concentrations of SEC fractions were determined using the Pierce Micro Bicinchoninic Acid (Micro BCA) protein assay kit (ThermoFisher, UK) per the manufacturer’s instructions (see Supporting Information for details).
Western Blot Analysis of Final Isolate
To validate the SEC isolation, exosome presence was verified through Western blot analysis via capillary gel electrophoresis format, using a WES instrument from Protein Simple (Biotechne Ltd., USA). SEC fraction 4 was selected for analysis as it possessed the highest ESP purity of 1.6 × 1010 ESPs/μg of protein (Figure S1). Exosomal proteins Alix (97 kDa) and tetraspannin CD63 (57 kDa) were probed by chemiluminescent immunoassay, using mouse monoclonal anti-Alix (634502, Biolegend UK) and mouse monoclonal anti-CD63 (353013, Biolegend UK) as primary antibodies. The WES run was conducted per the manufacturer’s instruction (see Supporting Information for details).
Gold Immuno-Electron Microscopy
Microscopy images were captured on a Jeol 2100 TEM instrument (Japan). SEC fraction 4 was diluted 100-fold, and 20 μL was spotted onto parafilm. A Formvar/carbon coated copper grid (200 mesh) was incubated on top of the sample for 7 min. The grid was washed three times with filtered phosphate buffered saline (PBS) buffer (pH 7.4, 137 mM NaCl, 2.7 mM KCl, 10 mM Na2HPO4, 1.8 mM KH2PO4; Sigma-Aldrich, USA). The grid was then incubated with 20 μL of normal serum block (927501, Biolegend UK), matching the species in which the secondary antibody was generated, for 30 min. Grids were subsequently washed three times with PBS. Twenty microliters of a 1:20 mouse monoclonal anti-CD63 solution (353013, Biolegend UK) in PBS/5% normal serum block was incubated with the grids for 1 h. The grid was washed six times with PBS. Thereafter, the grids were incubated with goat antimouse IgG conjugated gold nanoparticles (Insight Biotechnology, UK), at a 1:20 dilution in PBS/5% normal serum block, for 1 h in the dark. The grids were washed six times with PBS and incubated with 20 μL of 1% uranyl acetate negative stain, mixed with 2% phosphotungstic acid in a 1:10 ratio in DI water, for 5 min. Excess dye was removed with filter paper held at 45°, and grids were then left to dry before analysis.
QCM-D Measurements
All QCM-D measurements were carried out using a Q-Sense E4 instrument (Biolin Scientific, Sweden). Analysis of frequency and dissipation response was conducted with the QTools software, version 3.0.17.560 (Biolin Scientific, Sweden). Changes in resonance frequency (Δf) were recorded from the third, fifth, seventh, ninth, and 11th overtones. The presented data relates to the fifth overtone, with variation of (Δf) between overtones being 10% or less. In all instances, samples were degassed prior to exchange in the QCM flow module, and AT-cut, 5-MHz gold coated quartz crystal sensors with a 0.79 cm2 active area (Biolin, Sweden) were used.
An optimal antibody immobilization procedure was first investigated, comparing a covalent- and affinity-based approach. The covalent approach involved the formation of a self-assembled monolayer (SAM). A 1 mM ethanolic solution of SH-PEG(2 kDa)-COOH and SH-PEG(800 Da)–CH3 (Laysan Bio, USA) in a 1:3 mixture was flowed across the sensor surface at 7.5 μL/min overnight. The molarity was used as per previous SAM optimization studies.43 Carboxylic acid end groups were activated with an equal parts solution of 1-ethyl-3-(3-(dimethylamino)propyl)carbodiimide (EDC) cross-linker (0.4 M) and N-hydroxysuccinimide (NHS; 0.1 M) in pH 5.5 2-(N-morpholino)ethanaesulfonic acid (MES) coupling buffer (50 mM) at 7.5 μL per minute. Twenty micrograms per milliliter of mouse monoclonal anti-CD63 (353013, Biolegend UK) was immobilized on the surface by binding the amine group of the antibody to the activated carboxylic acid group on the functionalized sensor. This was conducted at 10 μL/min, in pH 7.4 HBS buffer. Unbound active groups were deactivated with 1 M ethanolamine for 20 min at 40 μL/min. This preceded a rinse step and response stabilization for 30 min prior to sample addition.
The affinity-based approach utilized an alternative SAM. Here, a 1 mM ethanolic solution of SH-PEG(2 kDa)-Biotin (Laysan Bio, USA) and spacer molecule SH-OEG(800 Da)-COOH (Polypure, Norway) at a 1:9 ratio was flowed across the sensor surface at 7.5 μL/min overnight. The molar ratio was based on optimized protocols in previous sensing studies.44,45 A 100 μg/mL solution of streptavidin (SAv; Sigma-Aldrich, USA) was flowed across the sensor surface at 10 μL/min, followed by a rinse step of HBS at 80 μL/min. Twenty micrograms per milliliter of mouse monoclonal biotinylated-anti-CD63 (353017, Biolegend UK) was immobilized on the surface at 10 μL/min, followed by another rinse step and response stabilization for 30 min prior to sample addition.
The affinity-based approach was chosen as the method of choice for the following investigations. Immuno-sensor functionality toward spiked CD63 and exosomal CD63 was assessed. Spiked samples of CD63 (Sinobiological, China) of concentrations 1.25, 2.5, 5, 10, 20, 40, 80, 160, and 320 nM determined the platform’s sensitivity toward the target antigen in HBS buffer. Performance toward exosomal CD63 was initially assessed by measuring responses to SEC fractions 3, 4, 5, 6, and 7, following a 100-factor dilution in HBS buffer. Sensitivity toward CD63-positive exosomes in HBS buffer was tested using dilutions of fraction 4 from SEC isolations with concentrations being verified by NTA. The following concentrations were assessed: 1 × 108, 2.5 × 108, 5 × 108, 7.5 × 108, 1 × 109, 5 × 109, 1 × 1010, and 5 × 1010 ESPs/mL. The same concentrations of ESPs were spiked into 75% v/v serum (Sigma-Aldrich, USA) to determine platform sensitivity toward CD63-positive exosomes in more complex media. Determination of specificity and signal-to-noise ratio (SNR) required response comparison with a control sensor surface. Here, biotin-IgG isotype control antibody (400103, Biolegend UK) was used instead of biotin-anti-CD63. Binding response was compared to a target sensor against a 5 × 109 ESPs/mL sample in HBS buffer and 1 × 109 ESPs/mL, spiked in 0%, 25%, 50%, and 75% v/v serum. For the purpose of this work, LOD and LOQ were defined as the concentration eliciting an SNR of 3 and 10, respectively, as governed by best practice.46 SNR was calculated by a ratio of the response seen on the target sensor and that on a control sensor surface.
To ensure reproducibility of the fabrication process, all analytes were prepared using the same degassed HBS stock solution or serum stock to minimize impact of buffer properties during sample exchange in observed responses. These were prepared to identical volumes (0.25 mL per sensor). All reagents were sourced from the same suppliers throughout the study to avoid influences of differing characteristics or quality. In all cases, the analyte was flowed at 10 μL/min, and a sensor was reserved for baseline measurement, to account for drift and background changes induced by buffer exchange. Frequency and dissipation responses are reported net or post-HBS rinse, to account for the removal of weakly bound analytes.
AFM Measurements
AFM was carried out on ultraflat gold substrates (Platypus Tech, USA) with a Multimode 8 instrument (Bruker, Santa Barbara, USA), using imaging based on fast force spectroscopy (PeakForce Nanomechanical mapping), with 1 kHz oscillation frequency and 5 nm amplitude. Imaging was conducted in solution with a PF-HR-B probe (Bruker, spring constant = 0.12 N/m, resonance frequency = 100 kHz in air). Filtered HBS served as imaging buffer. Substrates were incubated overnight in the SAM solution at room temperature. Surfaces were functionalized with identical streptavidin and antibody solutions as used in the affinity approach, by introducing and washing directly within the liquid cell. For assessment of ESPs in buffer and serum, 100 μL of 5 × 109 ESPs/mL in HBS buffer and 1 × 109 ESPs/mL in 75% v/v serum were introduced into the chamber, respectively, followed by incubation for 30 min, prior to washing with 1 mL of HBS buffer and image capture.
Images were processed with the Gwyddion software (version 2.5) by first aligning rows using a median of differences correction. Images were then flattened using mean plane subtraction. A three-pixel (∼6 nm) Gaussian filter was applied to reduce image noise. Height and diameter profiles of particles were taken as an average from 15 detected particles across three micrographs.
Results and Discussion
SEC Isolation of CD63-Positive Exosomes
The need for advanced analytical techniques is predicated on reproducible and efficient isolation from complex biological matrices. Moreover, the resulting exosomal samples must possess a high degree of structural and biological integrity. Thus, SEC was chosen as the isolation technique for its low impact on the nanosized vesicles.47
NTA analysis identified the fourth out of 20 collected fractions to have the highest yield of ESPs, with 4.4 × 1011 ESPs/mL. The ESP number reduced sequentially in the following fractions, which coincided with an increase in total protein (Figure 1A). The elevated protein concentrations post fraction 6 are attributed to the increasing elution of nonexosomal, colloidal protein present in the starting cell culture media concentrate. The concentration of nonexosomal protein in latter fractions is significantly superior to the total protein quantified from high ESP yield fractions (4 and 5), supporting previous reports on SEC-based exosome isolation.42,48 Overall, these results suggest that SEC successfully isolated the large majority of ESPs from contaminating protein in the HUMSCCM source material, based on molecule size. Size distribution analysis of fraction 4 confirmed that over 90% of the particles were within the ESP size range (Figure 1B). The isolation protocol was validated by Western blot, which identified exosome-enriched proteins, Alix and CD63, consistent with the presence of exosomes in the final sample. As evidenced by immuno-electron microscopy (immuno-EM) analysis (Figure 1C,D), the CD63 protein was shown to be accessible and present at the membrane surface. These results underpin the validity of surface-based immuno-capture of CD63-positive exosomes, as exploited herein.
Figure 1.
Exosome characterization obtained by SEC. (A) ESP and protein concentration profiles across SEC fractions using NTA and microBCA analysis. (B) Particle size distribution of SEC fraction 4. (C) Western blot identification of exosome enriched proteins, Alix and CD63, from fraction 4 and (D) Gold immuno-EM analysis confirming presence of CD63 protein at the exosome membrane.
Immuno-Sensor Performance toward Spiked-CD63
Prior to sensing exosomes, an appropriate antibody immobilization procedure was investigated. A covalent approach utilizing amine-coupling to directly bind anti-CD63 antibodies was compared with an affinity driven approach. Amine-coupling relies on the activation of carboxyl end groups on a SH-PEG-COOH:SH-PEG-CH3 mixed monolayer (formed by chemisorption) using carbodiimide chemistry to cross-link the amine residues found in lysine groups present in the antibody structure.49 The latter approach exploits the binding between SAv and biotin; the strongest naturally occurring noncovalent association (Kd = 10–14 M–1).50 SAv’s four binding sites are used to bind to biotin end groups on an SH-PEG-Biotin:SH-OEG-COOH mixed monolayer and biotin-functionalized anti-CD63 antibodies.51
The frequency response during the layer formation of the respective approaches is shown in Figure 2A,B. Net decreases in frequency are seen upon the addition of anti-CD63, SAv, and biotin-anti-CD63, with no indication of desorption (frequency increase) during the rinse step. Corresponding dissipation response to exosome-free CD63 proteins is shown in Table 1. Collectively, this suggests that the detection system has been successfully fabricated to the sensor surface.
Figure 2.
Example frequency profiles (post-SAM formation) for (A) covalent- and (B) affinity-based methods of antibody immobilization with diagrammatic illustration of surface modification. (C) Comparative frequency responses to 2 μg/mL of CD63 between antibody immobilization approaches. (D) Response curve for spiked-CD63 in buffer using optimized affinity functionalization method.
Table 1. QCM-D Frequency and Dissipation Responses in Figure 2C.
| sensor | analyte | conc | Δf (Hz) | ΔD (1 × 10–6) |
|---|---|---|---|---|
| baseline | running buffer | 0.61 ± 0.13 | 0.16 ± 0.05 | |
| control | CD63 | 2 μg/mL | –0.52 ± 0.11 | 0.15 ± 0.05 |
| covalent | CD63 | 2 μg/mL | –0.82 ± 0.17 | 0.17 ± 0.04 |
| affinity | CD63 | 2 μg/mL | –11.04 ± 1.21 | 0.73 ± 0.12 |
A significant net reduction in frequency of the sensor oscillation was seen for the affinity approach in response to the CD63 sample (Figure 2C). This frequency response is a result of CD63 capture by surface bound anitbodies, effectively increasing the crystal thickness and modifying the crystal resonance during thickness-shear oscillation. With the ratio of dissipation and frequency change being <4 × 10–7 Hz–1, we can confirm the CD63 adsorbate layer is rigid in nature.52 This was contrasted by a negligible response for the covalently functionalized sensor, which represented only a marginally larger net frequency shift than found for the control surface (functionalized with nonspecific antibodies). Both of these responses are only marginally different from frequency drift witnessed with the running buffer alone, suggesting almost no CD63 capture to the sensor surface (Table 1).
Despite exhibiting clear antibody fabrication, the poor performance of the covalent approach may be caused by suboptimal antibody orientation. As previously reported, the amine coupling approach has the risk of random antibody orientation toward the sensor surface, as opposed to oriented covalent and noncovalent approaches. Steric hindrance can prevent optimal binding site exposure for antigen access.53−55 Conversely, the binding observed with the affinity approach was likely due to superior orientation of the antibody via the Fc-bound biotin. This is in line with previous studies, which reported benefits of improved antibody orientation, higher binding site availability, and improved antigen sensitivity by utilizing the SAv–biotin interaction for antibody immobilization as compared to randomly orientated covalent alternatives.56,57
Having identified the affinity immobilization approach as offering optimal performance, this method was employed for the rest of the study. Figure S2 demonstrates the in situ overnight formation of a SAM at the sensor surface. The performance of the immuno-sensor was subsequently assessed toward exosome-free, spiked CD63 samples in HBS buffer (Figure 2D). Our platform displayed a LOD of 3.0 nM and LOQ of 7.9 nM, with a dynamic range of 7.9–160 nM. This demonstrated the high sensitivity of the immuno-sensor toward the target antigen, aligning with sensitivities reported by other practitioners of acoustic biosensing.58,59
Sensitivity and Specificity Assessment toward CD63-Positive Exosomes
For initial screening of CD63-positive exosomes, the various SEC fractions were compared for their acoustic response (Figure 3). QCM-D responses align well with the relative ESP/mL concentrations as identified by NTA (Figure 1A), with higher frequency reductions for those fractions possessing larger ESP concentrations. ESP-rich fractions 4 and 5 exhibited significant responses (∼−100 ± 8 and ∼91 ± 7 Hz, respectively) compared to fractions 3, 6, and 7. These responses are larger than found for fractions 6 and 7, in spite of higher total protein content in these fractions (Figure 1A). The greater responses may be due to (1) the larger size of the adsorbed material causing more liquid at the surface to be moved, (2) a high degree of hydrodynamic cosolvation, and/or (3) the greater mass of the fluid-filled vesicles.
Figure 3.
Example QCM-D profile responses to eluted SEC fractions taken from three independent runs. (A) Frequency response to SEC fractions 3–7 and (B) corresponding dissipation profiles.
An important component of acoustic exosome analysis is dissipation. As exosomes are not inherently rigid and prone to deformation,30 one expects significant friction in the newly formed adlayer during oscillation of the sensor, leading to regions of energy loss. Fractions 4 and 5 exhibited the most pronounced dissipation of 14.7 ± 1.9 and 13.5 ± 1.6, on average, respectively (Figure 3B). This phenomenon has been previously reported by groups sensing synthetic vesicles,60,61 where the viscoelastic structures resulted in energy storage (elastic) and loss (viscous) during oscillation. This is of interest, as it provides another discriminating factor to determine whether bound adsorbates are exosomal (vesicular and dissipative) or artifacts (nonvesicular and rigid). Hence, minimal responses seen for fractions 6 and 7 are likely a combination of smaller-sized vesicles expressing CD63, lower ESP concentrations overall, and the relative rigidity of the abundant nonexosomal protein. Notably, despite the ESP concentration in fraction 4 being 6.3-fold higher than fraction 5 (Figure 1A), the differences in frequency and dissipation response are in relative terms smaller and not proportional. This supports the response curve in Figure S4, as the concentrations of fractions 4 and 5 (4.4 × 1011 and 6.7 × 1010 per ml, respectively) sit outside the dynamic range of the QCM platform, with the immunosensor nearing performance saturation at these concentrations.
Having demonstrated successful immuno-sensing of spiked-CD63 and CD63-positive exosomes, we established the specificity of our sensing platform toward SEC purified ESPs in HBS buffer (see Figure S4). Critically, specificity in more complex biological media was also observed. Figure 4A–D compares the QCM-D response of a target and control surface, toward 1 × 109 ESPs/mL, spiked in increasing concentrations of human serum. Note that the lags in response from ESP addition and rinse procedures are attributed to varying tube lengths to the respective sample chambers.
Figure 4.
Selectivity of the QCM-D immuno-sensing approach toward CD63-postive exosomes in human serum. (A) Example frequency and (B) dissipation profiles from target surfaces (anti-CD63). (C) Bar chart comparing net frequency change and (D) net dissipation change between target and control surfaces (control-isotype) against 1 × 109 ESPs/mL spiked in increasing concentrations of human serum. Average QCM-D data taken from three independent experiments. Student t-test (two-tailed) compared responses from the control and target sensor (***p < 0.001). (E) AFM micrographs comparing control and target surfaces post-ESP incubation in 75% v/v human serum. Scale bar: 200 nm. Color scale (height): 100 nm).
The maximum shift in frequency and dissipation increases along with serum content, likely due to a greater quantity of serum particles available to adsorb at the sensor surface. However, the net changes (postrinse) on both measurements for serum containing samples are similar to ESPs in 0% v/v serum, suggesting a large proportion of the prerinse shift was caused by loosely bound, nonspecific adsorbates, of which the majority is removed upon buffer addition. No significant differences were found in net frequency and dissipation changes between ESPs in buffer alone versus ESPs in serum samples for the 95% confidence interval, using a Student t-test (two-tailed). These results indicate that the presence of complex colloidal contaminants does not hinder the target surface’s ability to sense CD63-positive exosomes. The control surfaces did reveal some signatures of binding, but this response was nearly overcome upon rinsing with buffer. Therefore, it is likely this binding was weak in nature. In 75% v/v serum, an average SNR of 5.8 and 11.4 was demonstrated by the frequency and dissipation modes of measurement, respectively (with exosomes in the picomolar concentration range). Interestingly, our findings highlight dissipation as a more sensitive and selective measurement over frequency. In our view, sensing CD63 protein as part of a vesicular structure provides an amplified and sensitive dissipation response, without the need for a secondary probe.
These QCM-D responses were supported by in-liquid AFM measurements, as particles within the exosome size range were detected (Figure 4E) on the targeted surface, while almost no vesicular material was seen to be bound on the substrate functionalized with control-IgG. Detected particles on the target surface had an average height of 53 ± 13 nm (mean ± standard deviation, n = 15 particles). To this end, surface-induced deformation of exosomes upon adsorption needs to be considered, which likely reduced the observed height from the nominal value.62 The average particle diameter was determined to be 64 ± 9 nm (n = 15 particles, thus suggesting that the majority of exosomes expressing CD63 were smaller than the mean (93.7 nm) and modal (86.5 nm) particle size of the ESP sample (Figure 1B), which supports previous findings by Rupert et al.62 We note that the observed QCM-D responses were generally below the expected values for a densely packed adlayer of particles ranging from 50 to 100 nm.63 On the basis of the evidence obtained by AFM, the lower readings are likely attributed to limited sensor coverage by bound exosomes.
The frequency measurement, in particular, highlights a small degree of residual, nonspecific (nonexosomal) binding that occurred on the control surfaces (Figure 4C,D). This indicates that some rigid adsorbates remained, while dissipative, loosely bound particles were removed from the surface with ease. The source of this phenomenon is likely attributed to the complexity of human serum, which is contaminated with its own proteins, lipids, and electrolytes. Fortunately, this phenomenon does not hinder ESP specificity, with high SNRs and differences between target and control responses remaining statistically significant. This is of analytical importance as it ensures the platform sufficiently discriminates between CD63-positive vesicles and other colloidal contaminants. We note the role of the SAM composition for nonspecific protein binding. As shown in Figure S5, we found a greatly decreased propensity for fouling in samples with a mixed PEG-OEG SAM, which is in line with previous studies.64
In order to elucidate changes in conformation during the ESP binding process, plots that display the change in dissipation as a function of the change in frequency are shown in Figure 5A–D. Moreover, the reaction fingerprint may be ascertained, i.e., how the viscoelastic properties of the layer changed (approximated by dissipation) with adsorbed mass (approximated by frequency). In all instances, viscoelasticity was found to increase with adsorbed mass, suggesting that both the specific binding of ESPs and nonspecific adsorption of serum contaminants induced significant friction in the sensor adlayer. A rinse of the sensor surface elicited significant concomitant decreases in frequency and dissipation for serum containing samples, consistent with dissipative material being removed from the bound layer (Figure 5C,D). Conversely, no change in frequency was observed alongside the small decrease in dissipation for the ESP in HBS buffer sample, suggesting that any removed material may have not been bound to the sensor surface (thus not affecting mass or layer thickness) while weakly contributing to energy dissipation during oscillatory decay (Figure 5A). Tymchenko et al. previously reported on conformational changes (spreading) of cellular layers, which imparted negative dissipation effects but negligible mass changes, suggesting a similar rearrangement of ESPs during this phase.65 Another possibility may lie in the difficulty to interpret the binding of certain viscoelastic microparticles, which elicit both positive and negative effects on oscillatory frequency, a behavior explained by the coupled-resonator model.66
Figure 5.
Dissipation change as a function of frequency change for 1 × 109 ESPs/mL binding to the target immuno-sensing surface in (A) 0% v/v serum, (B) 25% v/v serum, (C) 50% v/v serum, and (D) 75% v/v serum.
Additionally, to verify that the responses were not limited by the binding reaction, the introductory flow rate of a sample was increased from 10 μL/min to 80 μL/min and 120 μL/min (Figure S6A,B). The increase in adsorption rate was proportional to the cube root of the flow rate, indicating mass-transport limited binding (Figure S6C).67 These findings are consistent with diffusion of exosomes to the detection antibody being the rate limiting step during the adsorption not the binding kinetics.
Having established the platform’s specificity, the sensitivity was initially determined using varying concentrations of SEC-purified ESPs in HBS buffer (Figure S3). The same concentrations of EPSs were spiked into 75% v/v serum, revealing marginally elevated LOD and LOQ values as compared to sensing in HBS buffer (Figure 6A,B). Given our findings of a small degree of nonspecific binding from serum matrices (Figure 4), these results are expected and encouraging considering the relatively small increment in LOD and LOQ (Table 2). For both matrices, the dissipation mode of measurement showed a higher sensitivity relative to frequency, particularly in 75% v/v serum, outlining its utility in distinguishing CD63-positive exosomes among complex media. These values may be further improved by increasing the saturation capacity of the sensor, allowing for increased capture and coverage of exosomes.
Figure 6.
QCM-D immuno-sensor performance against varying concentrations of ESPs spiked in 75% v/v human serum. (A) Frequency and (B) corresponding dissipation response curves. Standard deviation determined from three independent experiments.
Table 2. QCM-D Immunosensor LOD and LOQ Values.
| LOD
(ESPs/mL) |
LOQ
(ESPs/mL) |
|||
|---|---|---|---|---|
| sample media | frequency | dissipation | frequency | dissipation |
| HBS buffer | 1.7 × 108 | 1.1 × 108 | 8.2 × 108 | 3.3 × 108 |
| 75% v/v serum | 2.9 × 108 | 1.4 × 108 | 9.4 × 108 | 4.5 × 108 |
Notably, the sensitivity by QCM-D reported herein is superior to some commercially available immunoassays toward exosomal CD63.68 Other recently reported approaches, e.g., interferometric plasmonic imaging of exosomes,69 demonstrated similar limits of detection. Some detection platforms have shown superior sensitivity in terms of particle concentration. Ko et al. devised an optofluidic device with sensitivity of up to 1.1 × 107 total particles/mL.70 Zhou et al. combined electrochemistry with an aptamer probe to sense exosomes down to 1 × 106 total particles/mL.26 Recently, Lim et al. showcased a LOD of just 200 exosomes per milliliter using an amplified plasmonic exosome (APEX) platform. This sensitivity was coupled with the plasmonic shift upon target prefibrillar Aβ localization to surface bound CD63-positive exosomes, enabling measurement comparison and biomarker subtyping between healthy and Alzheimer’s Disease (AD) blood samples.71 Our presented QCM-D platform could be used to further investigate this binding phenomenon of Aβ to surface-captured CD63-positive exosomes, as a complementary sensing technique in the clinical characterization of AD.
The limited sensitivity is a potential disadvantage of the adopted QCM-D immunoassay approach and can be attributed to the entire piezoelectric substrate being used for wave propagation. In contrast, surface acoustic wave (SAW) sensors confine the acoustic wave to a thin surface layer of the substrate and are capable of operating at higher frequencies (50 MHz-GHz), thus offering, in principle, pathways to higher sensitivity.72,73 Nevertheless, with native exosome counts in serum reported in the range of 0.8 × 108 to 1 × 109 particles per milliliter,74,75 the acoustic immunosensor developed in this work can detect exosomes from serum at clinically relevant concentrations. In some biological matrices, exosome concentrations may only reach values as low as 1 × 107 per milliliter.76,77 Further modifications are required to reliably improve the platform sensitivity toward this concentration domain. The process currently consumes 300 μL for analysis, slightly more than commercially available ELISAs. Improvements such as reducing the volume of the sample chamber, tubing dead volume, and blocking nonsensing surfaces by preincubation with bovine serum albumin (BSA) may reduce the sample requirement down to 20 μL, yet above minimal requirements for some SPR and electrochemical setups of less than 5 μL.27
Importantly, the platform offers a dual mode assessment of exosomes, with dissipation describing the stiffness of the adsorbate relative to the frequency change, lending a superior degree of selectivity to the procedure. This is an advantage over conventional immunoassays, which often deliver a single method of discrimination (e.g., colorimetric or electrochemical immunoassays). Another advantage over a traditional, single end point ELISA sandwich is an ability to make measurements in real time, avoiding lags associated with sample manipulation for signal amplification and indirect assessment.
Beyond serving as a useful quality control tool, the clinical utility of sensing exosomal CD63 has been highlighted by earlier work from Logozzi et al., which demonstrated how CD63-positive exosomes may serve as biomarkers due to their increased abundance in melanoma patients.68 More recently, Miki et al. identified exosomal CD63 as a potential prognostic marker in gastric cancer.78 QCM-D offers a robust platform to analytically supplement these research areas, while being applicable to other exosomal biomarkers.
For QCM-D to be adopted in a clinical diagnostic setting, additional efforts should focus on two areas: (i) advancing the current sensitivity to match plasmonic approaches and (ii) converting QCM-D responses into quantitative information about exosome concentrations. The former is desirable but not a necessity for the adsorption of particles in an exosomal size range, providing that robustness of measurement is displayed. Nonetheless, this can be achieved through devising novel surface functionalization with improved binding capacity and exosome entrapment. The latter is met with challenges of current models such as the Sauerbrey or Voigt methods, which both assume homogeneous layer formation.52 The introduction of a formalism to account for the discrete and dissipative nature of exosome binding, along with the coupled solvated mass, would further enhance the analytical insights offered by QCM-D. However, this is a nontrivial task due to multiple measurement parameters, the complexity of sample structure, and the difficulty in distinguishing between coupled water and the bulk solvent. A possible approach is to replace water with D2O. This would increase the shear viscosity of the bulk liquid while not affecting the kinetic or equilibrium state. Hence, discerned differences in frequency between water and D2O would allow one to determine mass contribution by the particle and coupled water fractions, respectively.79 Alternatively, combining the QCM-D measurement with optical techniques, such as SPR, has shown advantages to decoupling the contribution from layer hydration, thus offering pathways to identify the proportion of response attributed to the species of interest.80
Conclusion
This study establishes a QCM-D transduced immuno-sensing approach as a complementary technique for exosome characterization. Unlike other methods, we exploit a combination of mass, viscoelasticity, and surface antigens of exosomes. This allows sensitive and selective detection of CD63-positive exosomes in 75% v/v human serum without the need of a secondary label. At concentrations of 1 × 109 ESPs/mL, we achieved signal-to-noise ratios of 5.8 and 11.4 compared to a control surface for frequency and dissipation measurements, respectively. With a detection limit as low as 1.4 × 108 ESPs/mL, acoustic biosensing presents a direct route to phenotypically subtype exosomes at native concentrations, thus offering advantages in fundamental bioanalytical characterization with a potential to integrate with real-time liquid biopsies for clinical diagnostics in the future.
Acknowledgments
The authors thank the Engineering and Physical Sciences Research Council (EPSRC; EP/L01646X), the Medical Research Council (MRC, MR/R000328/1), University College London, and University of Nottingham CDT in Advanced Therapeutics and Nanomedicines for financial support. The authors also thank Dr. Benjamin Weil and Prof. Mark Lowdell from the Centre for Cell, Gene, and Tissue Therapeutics at the Royal Free London NHS Foundation Trust, for donating the HUMSCCM used in this research.
Supporting Information Available
The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acs.analchem.9b05736.
Cell-culture; microBCA protein assay; Western blot; SEC fraction particle purity; overnight in situ formation of SAM; QCM-D immunosensor performance toward CD63-positive ESPs in HBS buffer; QCM-D immunosensor specificity toward CD63-positive ESPs in HBS buffer; antifouling properties of sensor monolayers; demonstration of diffusion limited conditions for exosome sensing (PDF)
The authors declare no competing financial interest.
Supplementary Material
References
- Théry C.; Boussac M.; Véron P.; Ricciardi-Castagnoli P.; Raposo G.; Garin J.; Amigorena S. Proteomic Analysis of Dendritic Cell-Derived Exosomes: A Secreted Subcellular Compartment Distinct from Apoptotic Vesicles. J. Immunol. 2001, 166, 7309–7318. 10.4049/jimmunol.166.12.7309. [DOI] [PubMed] [Google Scholar]
- Aguilera-Rojas M.; Badewien-Rentzsch B.; Plendl J.; Kohn B.; Einspanier R. Exploration of serum- and cell culture-derived exosomes from dogs. BMC Vet. Res. 2018, 14, 179. 10.1186/s12917-018-1509-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Van Niel G.; Mallegol J.; Bevilacqua C.; Candalh C.; Brugìere S.; Tomaskovic-Crook E.; Heath J. K.; Cerf-Bensussan N.; Heyman M. Intestinal epithelial exosomes carry MHC class II/peptides able to inform the immune system in mice. Gut 2003, 52, 1690–1697. 10.1136/gut.52.12.1690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kowal J.; Arras G.; Colombo M.; Jouve M.; Morath J. P.; Primdal-Bengtson B.; Dingli F.; Loew D.; Tkach M.; Théry C. Proteomic comparison defines novel markers to characterize heterogeneous populations of extracellular vesicle subtypes. Proc. Natl. Acad. Sci. U. S. A. 2016, 113, E968–E977. 10.1073/pnas.1521230113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Colombo M.; Raposo G.; Théry C. Biogenesis, Secretion, and Intercellular Interactions of Exosomes and Other Extracellular Vesicles. Annu. Rev. Cell Dev. Biol. 2014, 30, 255–289. 10.1146/annurev-cellbio-101512-122326. [DOI] [PubMed] [Google Scholar]
- Coleman B. M.; Hill A. F. Extracellular vesicles - Their role in the packaging and spread of misfolded proteins associated with neurodegenerative diseases. Semin. Cell Dev. Biol. 2015, 40, 89–96. 10.1016/j.semcdb.2015.02.007. [DOI] [PubMed] [Google Scholar]
- Ibrahim A. G. E.; Cheng K.; Marbàn E. Exosomes as critical agents of cardiac regeneration triggered by cell therapy. Stem Cell Rep. 2014, 2, 606–619. 10.1016/j.stemcr.2014.04.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Saman S.; Kim W. H.; Raya M.; Visnick Y.; Miro S.; Saman S.; Jackson B.; McKee A. C.; Alvarez V. E.; Lee N. C.; Hall G. F. Exosome-associated tau is secreted in tauopathy models and is selectively phosphorylated in cerebrospinal fluid in early Alzheimer disease. J. Biol. Chem. 2012, 287, 3842–3849. 10.1074/jbc.M111.277061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gauthier S. A.; Perez-Gonzalez R.; Sharma A.; Huang F.-K.; Alldred M. J.; Pawlik M.; Kaur G.; Ginsberg S. D.; Neubert T. A.; Levy E. Enhanced exosome secretion in Down syndrome brain - a protective mechanism to alleviate neuronal endosomal ab- normalities. Acta Neuropathol. Commun. 2017, 5, 65. 10.1186/s40478-017-0466-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Taylor D. D.; Gercel-Taylor C. Exosome platform for diagnosis and monitoring of traumatic brain injury. Philos. Trans. R. Soc., B 2014, 369, 165–171. 10.1098/rstb.2013.0503. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lin S. Y.; Chang C. H.; Wu H. C.; Lin C. C.; Chang K. P.; Yang C. R.; Huang C. P.; Hsu W. H.; Chang C. T.; Chen C. J. Proteome profiling of urinary exosomes identifies alpha 1-antitrypsin and H2B1K as diagnostic and prognostic biomarkers for urothelial carcinoma. Sci. Rep. 2016, 6, 34446. 10.1038/srep34446. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gardiner C.; Ferreira Y. J.; Dragovic R. A.; Redman C. W.; Sargent I. L. Extracellular vesicle sizing and enumeration by nanoparticle tracking analysis. J. Extracell. Vesicles 2013, 2, 19671. 10.3402/jev.v2i0.19671. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vogel R.; et al. A standardized method to determine the concentration of extracellular vesicles using tunable resistive pulse sensing. J. Extracell. Vesicles 2016, 5, 31242. 10.3402/jev.v5.31242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bradford M. M. A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Anal. Biochem. 1976, 72, 248–254. 10.1016/0003-2697(76)90527-3. [DOI] [PubMed] [Google Scholar]
- Lobb R. J.; Becker M.; Wen Wen S.; Wong C. S. F.; Wiegmans A. P.; Leimgruber A.; Moller A. M̈oller, A. Optimized exosome isolation protocol for cell culture supernatant and human plasma. J. Extracell. Vesicles 2015, 4, 27031. 10.3402/jev.v4.27031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Abramowicz A.; Widlak P.; Pietrowska M. Proteomic analysis of exosomal cargo: The challenge of high purity vesicle isolation. Mol. BioSyst. 2016, 12, 1407–1419. 10.1039/C6MB00082G. [DOI] [PubMed] [Google Scholar]
- Erdbrügger U.; Lannigan J. Analytical challenges of extracellular vesicle detection: A comparison of different techniques. Cytometry, Part A 2016, 89, 123–134. 10.1002/cyto.a.22795. [DOI] [PubMed] [Google Scholar]
- Lannigan J.; Erdbruegger U. Imaging flow cytometry for the characterization of extracellular vesicles. Methods 2017, 112, 55–67. 10.1016/j.ymeth.2016.09.018. [DOI] [PubMed] [Google Scholar]
- Lee S.; Zhu X.; Sha Q. Recent studies of Kell and XK: Expression profiles of mouse Kell and XK mRNA. Neuroacanthocytosis Syndr. II 2008, 7, 107–114. 10.1007/978-3-540-71693-8_8. [DOI] [Google Scholar]
- Ferguson S. W.; Nguyen J. Exosomes as therapeutics: The implications of molecular composition and exosomal heterogeneity. J. Controlled Release 2016, 228, 179–190. 10.1016/j.jconrel.2016.02.037. [DOI] [PubMed] [Google Scholar]
- Ren R.; Cai G.; Yu Z.; Zeng Y.; Tang D. Metal-Polydopamine Framework: An Innovative Signal-Generation Tag for Colorimetric Immunoassay. Anal. Chem. 2018, 90, 11099–11105. 10.1021/acs.analchem.8b03538. [DOI] [PubMed] [Google Scholar]
- Luo Z.; Zhang L.; Zeng R.; Su L.; Tang D. Near-Infrared Light-Excited Core- Core-Shell UCNP@Au@CdS Upconversion Nanospheres for Ultrasensitive Photoelec- trochemical Enzyme Immunoassay. Anal. Chem. 2018, 90, 9568–9575. 10.1021/acs.analchem.8b02421. [DOI] [PubMed] [Google Scholar]
- He M.; Crow J.; Roth M.; Zeng Y.; Godwin A. K. Integrated immunoisolation and protein analysis of circulating exosomes using microfluidic technology. Lab Chip 2014, 14, 3773–3780. 10.1039/C4LC00662C. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang Y.; Yuan W.; Kimber M.; Lu M.; Dong L. Rapid Differentiation of Host and Parasitic Exosome Vesicles Using Microfluidic Photonic Crystal Biosensor. ACS Sensors 2018, 3, 1616–1621. 10.1021/acssensors.8b00360. [DOI] [PubMed] [Google Scholar]
- Daaboul G. G.; Gagni P.; Benussi L.; Bettotti P.; Ciani M.; Cretich M.; Freedman D. S.; Ghidoni R.; Ozkumur A. Y.; Piotto C.; Prosperi D.; Santini B.; Ünlü M. S.; Chiari M. Digital Detection of Exosomes by Interferometric Imaging. Sci. Rep. 2016, 6, 37246. 10.1038/srep37246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhou Q.; Rahimian A.; Son K.; Shin D. S.; Patel T.; Revzin A. Development of an aptasensor for electrochemical detection of exosomes. Methods 2016, 97, 88–93. 10.1016/j.ymeth.2015.10.012. [DOI] [PubMed] [Google Scholar]
- Doldán X.; Fagúndez P.; Cayota A.; Láız J.; Tosar J. P. Electrochemical Sandwich Immunosensor for Determination of Exosomes Based on Surface Marker-Mediated Sig- nal Amplification. Anal. Chem. 2016, 88, 10466–10473. 10.1021/acs.analchem.6b02421. [DOI] [PubMed] [Google Scholar]
- Im H.; Shao H.; Park Y. I.; Peterson V. M.; Castro C. M.; Weissleder R.; Lee H. Label-free detection and molecular profiling of exosomes with a nano-plasmonic sensor. Nat. Biotechnol. 2014, 32, 490–495. 10.1038/nbt.2886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Raghu D.; Christodoulides J. A.; Christophersen M.; Liu J. L.; Anderson G. P.; Robitaille M.; Byers J. M.; Raphael M. P. Nanoplasmonic pillars engineered for single exosome detection. PLoS One 2018, 13, e0202773. 10.1371/journal.pone.0202773. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rupert D. L.; Lässer C.; Eldh M.; Block S.; Zhdanov V. P.; Lotvall J. O.; Bally M.; Höök F. Determination of exosome concentration in solution using surface plasmon resonance spectroscopy. Anal. Chem. 2014, 86, 5929–5936. 10.1021/ac500931f. [DOI] [PubMed] [Google Scholar]
- Helmerhorst E.; Chandler D. J.; Nussio M.; Mamotte C. D. Real-time and label- ree bio-sensing of molecular interactions by surface plasmon resonance: A laboratory medicine perspective. Clin. Biochem. 2012, 33, 161–173. [PMC free article] [PubMed] [Google Scholar]
- Nguyen H. H.; Park J.; Kang S.; Kim M. Surface plasmon resonance: A versatile technique for biosensor applications. Sensors 2015, 15, 10481–10510. 10.3390/s150510481. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Qiu Z.; Shu J.; Liu J.; Tang D. Dual-Channel Photoelectrochemical Ratiometric Aptasensor with up-Converting Nanocrystals Using Spatial-Resolved Technique on Homemade 3D Printed Device. Anal. Chem. 2019, 91, 1260–1268. 10.1021/acs.analchem.8b05455. [DOI] [PubMed] [Google Scholar]
- Yu Z.; Tang Y.; Cai G.; Ren R.; Tang D. Paper Electrode-Based Flexible Pressure Sensor for Point-of-Care Immunoassay with Digital Multimeter. Anal. Chem. 2019, 91, 1222–1226. 10.1021/acs.analchem.8b04635. [DOI] [PubMed] [Google Scholar]
- Hook F.; Rodahl M.; Keller C.; Glasmastar K.; Fredriksson C.; Dahiqvist P.; Kasemo B. Dissipative QCM-D technique: Interfacial phenomena and sensor applications for proteins, biomembranes, living cells and polymers. Proc. Annu. IEEE Int. Freq. Control Symp. 1999, 2, 966–972. 10.1109/FREQ.1999.841467. [DOI] [Google Scholar]
- Li X.; Song S.; Shuai Q.; Pei Y.; Aastrup T.; Pei Y.; Pei Z. Real-time and label- free analysis of binding thermodynamics of carbohydrate-protein interactions on unfixed cancer cell surfaces using a QCM biosensor. Sci. Rep. 2015, 5, 14066. 10.1038/srep14066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tao W.; Xie Q.; Wang H.; Ke S.; Lin P.; Zeng X. Integration of a miniature quartz crystal microbalance with a microfluidic chip for amyloid beta-Aβ42 quantitation. Sensors 2015, 15, 25746–25760. 10.3390/s151025746. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Johannsmann D.; Reviakine I.; Richter R. P. Dissipation in films of adsorbed nanospheres studied by quartz crystal microbalance (QCM). Anal. Chem. 2009, 81, 8167–8176. 10.1021/ac901381z. [DOI] [PubMed] [Google Scholar]
- Uludag Y.; Tothill I. E. Development of a sensitive detection method of cancer biomark- ers in human serum (75%) using a quartz crystal microbalance sensor and nanoparticles amplification system. Talanta 2010, 82, 277–282. 10.1016/j.talanta.2010.04.034. [DOI] [PubMed] [Google Scholar]
- Pirincci S. S.; Ertekin O.; Laguna D. E.; Ozen F. S.; Ozturk Z. Z.; Ozturk S. Label-Free QCM Immunosensor for the Detection of Ochratoxin A. Sensors 2018, 18, 1161. 10.3390/s18041161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tang D.; Zhang B.; Tang J.; Hou L.; Chen G. Displacement-type quartz crystal microbalance immunosensing platform for ultrasensitive monitoring of small molecular toxins. Anal. Chem. 2013, 85, 6958–6966. 10.1021/ac401599t. [DOI] [PubMed] [Google Scholar]
- Böing A. N.; van der Pol E.; Grootemaat A. E.; Coumans F. A.; Sturk A.; Nieuwland R. Single-step isolation of extracellular vesicles by size-exclusion chromatography. J. Extracell. Vesicles 2014, 3, 23430. 10.3402/jev.v3.23430. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ayela C.; Roquet F.; Valera L.; Granier C.; Nicu L.; Pugnìere M. Antibody- antigenic peptide interactions monitored by SPR and QCM-D. A model for SPR detec- tion of IA-2 autoantibodies in human serum. Biosens. Bioelectron. 2007, 22, 3113–3119. 10.1016/j.bios.2007.01.020. [DOI] [PubMed] [Google Scholar]
- Kim Y. H.; Kim J. P.; Han S. J.; Sim S. J. Aptamer biosensor for lable-free detection of human immunoglobulin E based on surface plasmon resonance. Sens. Actuators, B 2009, 139, 471–475. 10.1016/j.snb.2009.03.013. [DOI] [Google Scholar]
- Knoll W.; Liley M.; Piscevic D.; Spinke J.; Tarlov M. J. Supramolecular architectures for the functionalization of solid surfaces. Adv. Biophys. 1997, 34, 231–251. 10.1016/S0065-227X(97)89642-6. [DOI] [PubMed] [Google Scholar]
- Shrivastava A.; Gupta V. Methods for the determination of limit of detection and limit of quantitation of the analytical methods. Chronicles Young Sci. 2011, 2, 21. 10.4103/2229-5186.79345. [DOI] [Google Scholar]
- Gámez-Valero A.; Monguió-Tortajada M.; Carreras-Planella L.; Franquesa M.; Beyer K.; Borràs F. E. Size-Exclusion Chromatography-based isolation minimally alters Extracellular Vesicles’ characteristics compared to precipitating agents. Sci. Rep. 2016, 6, 33641. 10.1038/srep33641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Welton J. L.; Webber J. P.; Botos L. A.; Jones M.; Clayton A. Ready-made chromatography columns for extracellular vesicle isolation from plasma. J. Extracell. Vesicles 2015, 4, 27269. 10.3402/jev.v4.27269. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vashist S. K. Comparison of 1-Ethyl-3-(3-Dimethylaminopropyl) Carbodiimide Based Strategies to Crosslink Antibodies on Amine-Functionalized Platforms for Immunodi- agnostic Applications. Diagnostics 2012, 2, 23–33. 10.3390/diagnostics2030023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wilchek M.; Bayer E. A. Foreword and introduction to the book (strept)avidin-biotin system. Biomol. Eng. 1999, 16, 1–4. 10.1016/S1050-3862(99)00032-7. [DOI] [PubMed] [Google Scholar]
- Dubacheva G. V.; Araya-Callis C.; Geert Volbeda A.; Fairhead M.; Codee J.; Howarth M.; Richter R. P. Controlling Multivalent Binding through Surface Chemistry: Model Study on Streptavidin. J. Am. Chem. Soc. 2017, 139, 4157–4167. 10.1021/jacs.7b00540. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Reviakine I.; Johannsmann D.; Richter R. P. Hearing what you cannot see and vi- sualizing what you hear: Interpreting quartz crystal microbalance data from solvated interfaces. Anal. Chem. 2011, 83, 8838–8848. 10.1021/ac201778h. [DOI] [PubMed] [Google Scholar]
- Tan Y. H.; Liu M.; Nolting B.; Go J. G.; Gervay-Hague J.; Liu G. Y. A nanoengi- neering approach for investigation and regulation of protein immobilization. ACS Nano 2008, 2, 2374–2384. 10.1021/nn800508f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Baniukevic J.; Kirlyte J.; Ramanavicius A.; Ramanaviciene A. Application of ori- ented and random antibody immobilization methods in immunosensor design. Sens. Actuators, B 2013, 189, 217–223. 10.1016/j.snb.2013.03.126. [DOI] [Google Scholar]
- Kausaite-Minkstimiene A.; Ramanaviciene A.; Kirlyte J.; Ramanavicius A. Comparative study of random and oriented antibody immobilization techniques on the binding capacity of immunosensor. Anal. Chem. 2010, 82, 6401–6408. 10.1021/ac100468k. [DOI] [PubMed] [Google Scholar]
- Trilling A. K.; Harmsen M. M.; Ruigrok V. J.; Zuilhof H.; Beekwilder J. The effect of uniform capture molecule orientation on biosensor sensitivity: Dependence on analyte properties. Biosens. Bioelectron. 2013, 40, 219–226. 10.1016/j.bios.2012.07.027. [DOI] [PubMed] [Google Scholar]
- Balevicius Z.; Ramanaviciene A.; Baleviciute I.; Makaraviciute A.; Mikoliunaite L.; Ramanavicius A. Evaluation of intact- and fragmented-antibody based immunosensors by total internal reflection ellipsometry. Sens. Actuators, B 2011, 160, 555–562. 10.1016/j.snb.2011.08.029. [DOI] [Google Scholar]
- Hwang S. S.; Chan H.; Sorci M.; Van Deventer J.; Wittrup D.; Belfort G.; Walt D. Detection of amyloid β oligomers toward early diagnosis of Alzheimer’s disease. Anal. Biochem. 2019, 566, 40–45. 10.1016/j.ab.2018.09.011. [DOI] [PubMed] [Google Scholar]
- Chen Q.; Tang W.; Wang D.; Wu X.; Li N.; Liu F. Amplified QCM-D biosensor for protein based on aptamer-functionalized gold nanoparticles. Biosens. Bioelectron. 2010, 26, 575–579. 10.1016/j.bios.2010.07.034. [DOI] [PubMed] [Google Scholar]
- Grieshaber D.; De Lange V.; Hirt T.; Lu Z.; Vörös J. Vesicles for signal amplification in a biosensor for the detection of low antigen concentrations. Sensors 2008, 8, 7894–7903. 10.3390/s8127894. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jackman J. A.; Yorulmaz Avsar S.; Ferhan A. R.; Li D.; Park J. H.; Zhdanov V. P.; Cho N.-J. Quantitative profiling of nanoscale liposome deformation by a localized surface plasmon resonance sensor. Anal. Chem. 2017, 89, 1102–1109. 10.1021/acs.analchem.6b02532. [DOI] [PubMed] [Google Scholar]
- Rupert D. L. M.; Shelke G. V.; Emilsson G.; Claudio V.; Block S.; Lasser C.; Dahlin A.; Lotvall J. O.; Bally M.; Zhdanov V. P.; Hook F. Dual-Wavelength Surface Plasmon Resonance for Determining the Size and Concentration of Sub-Populations of Extracellular Vesicles. Anal. Chem. 2016, 88, 9980–9988. 10.1021/acs.analchem.6b01860. [DOI] [PubMed] [Google Scholar]
- Morita S.; Nukui M.; Kuboi R. Immobilization of liposomes onto quartz crystal microbalance to detect interaction between liposomes and proteins. J. Colloid Interface Sci. 2006, 298, 672–678. 10.1016/j.jcis.2005.12.043. [DOI] [PubMed] [Google Scholar]
- Lokanathan A. R.; Zhang S.; Regina V. R.; Cole M. A.; Ogaki R.; Dong M.; Besen- bacher F.; Meyer R. L.; Kingshott P. Mixed poly (ethylene glycol) and oligo (ethylene glycol) layers on gold as nonfouling surfaces created by backfilling. Biointerphases 2011, 6, 180–188. 10.1116/1.3647506. [DOI] [PubMed] [Google Scholar]
- Tymchenko N.; Nileback E.; Voinova M. V.; Gold J.; Kasemo B.; Svedhem S. Reversible changes in cell morphology due to cytoskeletal rearrangements measured in real-time by QCM-D. Biointerphases 2012, 7, 43. 10.1007/s13758-012-0043-9. [DOI] [PubMed] [Google Scholar]
- Tarnapolsky A.; Freger V. Modeling QCM-D Response to Deposition and Attachment of Microparticles and Living Cells. Anal. Chem. 2018, 90, 13960–13968. 10.1021/acs.analchem.8b03411. [DOI] [PubMed] [Google Scholar]
- Van Der Meulen S. A.; Dubacheva G. V.; Dogterom M.; Richter R. P.; Leunissen M. E. Quartz crystal microbalance with dissipation monitoring and spectroscopic ellipsometry measurements of the phospholipid bilayer anchoring stability and kinetics of hydrophobically modified DNA oligonucleotides. Langmuir 2014, 30, 6525–6533. 10.1021/la500940a. [DOI] [PubMed] [Google Scholar]
- Logozzi M.; De Milito A.; Lugini L.; Borghi M.; Calabro L.; Spada M.; Perdicchio M.; Marino M. L.; Federici C.; Iessi E.; Brambilla D.; Venturi G.; Lozupone F.; Santinami M.; Huber V.; Maio M.; Rivoltini L.; Fais S. High levels of exosomes expressing CD63 and caveolin-1 in plasma of melanoma patients. PLoS One 2009, 4, e5219. 10.1371/journal.pone.0005219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zeng X.; Yang Y.; Zhang N.; Ji D.; Gu X.; Jornet J. M.; Wu Y.; Gan Q. Plasmonic interferometer array biochip as a new mobile medical device for cancer detection. IEEE J. Sel. Top. Quantum Electron. 2019, 25, 25. 10.1109/JSTQE.2018.2865418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ko J.; Hemphill M. A.; Gabrieli D.; Wu L.; Yelleswarapu V.; Lawrence G.; Pennycooke W.; Singh A.; Meaney D. F.; Issadore D. Smartphone-enabled optofluidic exosome diagnostic for concussion recovery. Sci. Rep. 2016, 6, 31215. 10.1038/srep31215. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lim J.; Choi M.; Lee H.; Kim Y. H.; Han J. Y.; Lee E. S.; Cho Y. Direct isolation and characterization of circulating exosomes from biological samples using magnetic nanowires. J. Nanobiotechnol. 2019, 17, 1. 10.1186/s12951-018-0433-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kankare J. Sauerbrey equation of quartz crystal microbalance in liquid medium. Langmuir 2002, 18, 7092–7094. 10.1021/la025911w. [DOI] [Google Scholar]
- Xu Z.; Yuan Y. J. Quantification of: Staphylococcus aureus using surface acoustic wave sensors. RSC Adv. 2019, 9, 8411–8414. 10.1039/C8RA09790A. [DOI] [PMC free article] [PubMed] [Google Scholar]
- De Vrij J.; Maas S. L.; Van Nispen M.; Sena-Esteves M.; Limpens R. W.; Koster A. J.; Leenstra S.; Lamfers M. L.; Broekman M. L. Quantification of nanosized extracellular membrane vesicles with scanning ion occlusion sensing. Nanomedicine 2013, 8, 1443–1458. 10.2217/nnm.12.173. [DOI] [PubMed] [Google Scholar]
- Huang X.; Yuan T.; Tschannen M.; Sun Z.; Jacob H.; Du M.; Liang M.; Dittmar R. L.; Liu Y.; Liang M.; Kohli M.; Thibodeau S. N.; Boardman L.; Wang L. Characterization of human plasma-derived exosomal RNAs by deep sequencing. BMC Genomics 2013, 14, 319. 10.1186/1471-2164-14-319. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wongkaew N. Nanofiber-integrated miniaturized systems: an intelligent platform for cancer diagnosis. Anal. Bioanal. Chem. 2019, 411, 4251–4264. 10.1007/s00216-019-01589-5. [DOI] [PubMed] [Google Scholar]
- Soares Martins T.; Catita J.; Martins Rosa I.; A. B. da Cruz e Silva O.; Henriques A. G. Exosome isolation from distinct biofluids using precipitation and column-based approaches. PLoS One 2018, 13, e0198820. 10.1371/journal.pone.0198820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miki Y.; Yashiro M.; Okuno T.; Kuroda K.; Togano S.; Hirakawa K.; Ohira M. Clinico-pathological significance of exosome marker CD63 expression on cancer cells and stromal cells in gastric cancer. PLoS One 2018, 13, e0202956. 10.1371/journal.pone.0202956. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Craig V. S.; Plunkett M. Determination of coupled solvent mass in quartz crystal microbalance measurements using deuterated solvents. J. Colloid Interface Sci. 2003, 262, 126–129. 10.1016/S0021-9797(03)00210-8. [DOI] [PubMed] [Google Scholar]
- Höök F.; Kasemo B.; Nylander T.; Fant C.; Sott K.; Elwing H. Variations in coupled water, viscoelastic properties, and film thickness of a Mefp-1 protein film during adsorption and cross-linking: A quartz crystal microbalance with dissipation monitoring, ellipsometry, and surface plasmon resonance study. Anal. Chem. 2001, 73, 5796–5804. 10.1021/ac0106501. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.