Abstract
A rapid analysis of pesticides using on-line Solid phase extraction LC MS/MS (Agilent Technology) was performed using only 2-mL water samples. SPE cartridge PLRP-s was used for the pre-concentration sample with methanol elution in back flush. Sensitive transitions and mass spectrometry conditions were optimized by direct infusion of individual standard solutions in a positive electrospray mode. Water samples were spiked with internal standards to compensate the matrix effect. The limit of quantification was calculated to be 20 ng L−1 using the standard deviation of blank analysis injected ten times and uncertainties were estimated at less than 20% on concentrations. This method was validated to study leaching water samples for which only small quantities of water were available.
-
•
Only 2 mL water sample was used.
-
•
Samples were filtered at 0.2 µm and spiked with individual standard.
-
•
Compounds were separated in an 18.5-min elution time using the dynamic MRM program.
Keywords: Mass spectrometry, Triple quadrupole, Pesticide, Groundwater
Graphical abstract
Specification Table
| Subject Area | Chemistry |
| More specific subject area: | Analytical chemistry |
| Method name: | A rapid analytical method to analyse pesticides using only 2-mL water samples |
| Name and reference of original method | |
| Resource availability |
Method details
The method developed was applied to investigate atrazine and its main by-products Desethylatrazine (DEA), Deisopropylatrazine (DIA), Hydroxyatrazine (HA), simazine and S-metolachlor in 2-mL water samples. These samples were collected in a quarry located in Saint-Martin-le-Noeud, 70 km north of Paris. This site is an exceptional natural lysimeter, 1.2 km long, with no disturbed soils and an unsaturated layer above the quarry [1,2]. The study sites correspond to 15 different lakes isolated from each other. Both percolation water and groundwater were sampled. The percolation water was sampled in beakers attached at the roof of each site and groundwater directly into the lake. Depending on percolate flow, it is possible to collect from 5 mL to more than 200 mL in the beaker for major element analysis, as well as tritium and other trace elements. A limited water volume was available for pesticide analysis. Pesticide uses have been investigated directly near farms for the last 40 years and we focused on pesticides that were the most widely used: atrazine, simazine, S-metolachlor.
The on-line method was developed to analyze these pesticides using LC-MS/MS (LC-1200, QQQ-G6410B, Agilent Technologies). LC-MS/MS is commonly used to analyze these pesticides in different matrices [3], [4], [5], [6]. On-line solid-phase extraction methods were also developed, using different SPE cartridges such as OASIS HLB, Waters [7], Strata-X extraction cartridge, Phenomenex [8] or PLRPs, Agilent [9]. In these three studies, water sample volumes were 5, 20 and 5 mL and limits of quantification (LOQ) were <1, 6–15, and 57–508 ng L−1 respectively. Highest LOQ [9] were obtain with the residual standard deviation method. An on-line SPE method had already been developed for antibiotic analysis with only 2-mL samples [10]. Such a method is essential because it was not possible to devote a larger volume to the analysis of pesticides in our study, especially for percolate water. Here we adapted this method for pesticides, using only 2 mL water sample filtered and directly analyzed by SPE – LC MS/MS.
Reagents and standards
LC/MS-grade methanol (MeOH) and LC/MS-grade acetonitrile (MeCN) were purchased from Merck (Guyancourt, France) and analytical grade formic acid (99%) from Carlo Erba (Val de Reuil, France). Ultrapure water 18 Mohm (UP water) was dispensed from a Pure Lab Chorus water purification system (Veolia Water STI, France).
Unlabeled standards at 100 µg mL−1 (purity between 96.8 and 99%) were purchased from LGC (Molsheim, France): atrazine, DEA, DIA in MeCN and simazine and HA in MeOH, and from CIL Cluzeau (Saint Foy la Grande, France): S-Metolachlor in acetonitrile. Ring-labeled internal standards at 100 µg mL−1 (purity between 96.8 and 98% and isotopic value between 99 and 99.9%) were purchased from LGC (Molsheim, France): Atrazine Ring-13C3 in acetone, DEA Ring-13C3 in MeCN and HA ring-13C3 in 80% water/20% diethylamine. All individual stock standard solutions were stored at −18 °C.
Mixed unlabeled solution was prepared in UP water containing 1 mg L−1 of each compound. The mixed labeled solution was also prepared at 1 mg L−1 in UP water. Both solutions were stored at 4 °C for 1 week maximum. All working solutions were prepared immediately before each analytical series by appropriate dilution of the mixed solution with UP water.
Sample preparation
Two milliliters of water samples were filtered at 0.2 µm using polytetrafluoroethylene (PTFE) filters (4 mm, 0.2 µm for the syringe, Interchim (Montluçon, France) using a Luer syringe and spiked with 0.1 mL of the internal standard (ISs) mixture at 10 µg L−1. Since water had no suspended matter, there was no need to previously filter it at 0.45 µm.
On-line enrichment
The on-line SPE system consisted of an automated liquid sampler (ALS) fitted with a 900-µL injection loop with multidraw capability. A programable six-port/two-position valve was used to switch between the load or elution modes (Fig. 1). PLRP-s cartridges (15–25 µm, 2 × 10 mm, Serlabo, Entraigues, France) were first conditioned for 10 min of methanol (A) and 10 min of UP water (B) at a flow rate of 0.3 mL min−1 for a series of three blanks, seven standards at different concentration levels and 16 samples.
Fig. 1.
Scheme of sample enrichment method [11].
A quaternary pump delivered the loading buffer at a flow rate of 1 mL min–1 (100% UP-water) and 1800-µL (i.e., twice 900 µL) samples were loaded onto the cartridge (Table 1). At 0.1 min after injection, the valve switch position and the analytes retained on the PLRPs cartridge were progressively transferred to the LC analytical column in back-flush mode, using the LC solvent gradient described below (Table 2).
Table 1.
Quaternary pump time table.
| Time | Flow | Pressure | Solvant ratio B (UP-water)% |
|---|---|---|---|
| 0 | 1 | 400 | 100 |
| 0.1 | 0.5 | 400 | 100 |
| 2 | 0.5 | 400 | 100 |
| 2.1 | 1 | 400 | 0 |
| 8 | 1 | 400 | 0 |
| 8.1 | 0.5 | 400 | 100 |
| 18.5 | 0.5 | 400 | 100 |
Table 2.
LC Gradient to separate analytes.
| Step | Time | A (UP water+ 0.1% formic acid) % | B (Acetonitrile+ 0.1% formic acid) % | Flow mL min−1 |
|---|---|---|---|---|
| 1 | 0.00 | 5% | 95% | 0.300 |
| 2 | 0.50 | 5% | 95% | 0.300 |
| 3 | 10.00 | 100% | 0% | 0.300 |
| 4 | 14.00 | 100% | 0% | 0.300 |
| 5 | 15.00 | 5% | 95% | 0.300 |
| 6 | 18.50 | 5% | 95% | 0.300 |
Chromatographic separation was achieved using an Agilent Zorbax Eclipse XDB-C 18 column (4.6 mm I.D. × 50 mm, 1.8-m particle size; Agilent, Les Ulis, France) with a 0.2-µm prefilter upstream to protect the analytical column. Mobile-phase solvents were UP water + 0.1% formic acid (B) and acetonitrile + 0.1% formic acid (A) in an initial ratio (B:A) of 95/5. Separation was achieved at 35 °C using a flow rate of 0.3 mL min–1 with the gradient described in Table 2. Then the system was equilibrated for 1 min prior to the next injection (total run time: 18.5 min).
During MS/MS analysis, the sample injection loop was flushed to prevent cross-contamination: 100% B for 2 min, 100% A for 6 min (Table 1).
Procedural blanks consisting of UP water spiked with ISs were analyzed as a control of contamination during sample handling in the laboratory and to assess a potential memory effect in the on-line SPE apparatus.
Mass spectrometry
The Agilent 6410B triple quadrupole mass spectrometer was equipped with an electrospray ionization (ESI) source and was operated in positive mode. Nitrogen (99.9%, Air Liquide, Paris, France) was used as collision gas while nitrogen used as the nebulizing gas (11.0 L h−1, nebulizer pressure 35 psi) was produced via a nitrogen generator (Claind, France). The source temperature was set at 350 °C and MS/MS signal acquisition was performed in dynamic multiple reaction monitoring (dMRM) mode. For MS/MS optimization, individual standard solutions were directly infused in the triple quadrupole. Method selectivity was provided by the monitoring of two transitions per compound at the retention time of the analyte, corresponding to the transition between the precursor ion and the two most abundant product ions (Table 3). The most abundant one was used for quantification while the other one was used for confirmation. The ratio of quantification transition area to the confirmation transition area was compared to that obtained with an authentic standard (margin applied: 20%). Data were collected using Mass Hunter software from Agilent Technologies, which was also used for analyte quantification.
Table 3.
MS/MS optimized conditions for selected pesticides and internal standards.
| Compound Name | Prec Ion | Prod Ion | Frag (V) | CE (V) | Ret time |
|---|---|---|---|---|---|
| HydroxyAtrazine 13C | 201.2 | 159.1 (Q) | 110 | 14 | 5.2 |
| HydroxyAtrazine 13C | 201.2 | 116.2 | 110 | 18 | 5.2 |
| Hydroxyatrazine | 198.1 | 142.1 | 115 | 12 | 5.2 |
| Hydroxyatrazine | 198.1 | 86.1 (Q) | 115 | 24 | 5.2 |
| DIA | 174.1 | 96 | 110 | 16 | 5.9 |
| DIA | 174.1 | 68.2 (Q) | 110 | 28 | 5.9 |
| DEA 13C | 191.6 | 149.1 (Q) | 110 | 14 | 6.8 |
| DEA 13C | 191.6 | 106 | 110 | 26 | 6.8 |
| DEA | 188 | 146 (Q) | 105 | 16 | 6.8 |
| DEA | 188 | 104 | 105 | 24 | 6.8 |
| Simazine | 202 | 132.1 | 100 | 20 | 8.05 |
| Simazine | 202 | 104.1 (Q) | 100 | 20 | 8.05 |
| Atrazine 13C | 219.7 | 106 | 110 | 30 | 8.9 |
| Atrazine 13C | 219.7 | 70 (Q) | 110 | 42 | 8.9 |
| Atrazine | 216 | 174.1 (Q) | 130 | 16 | 8.9 |
| Atrazine | 216 | 104.1 | 130 | 30 | 8.9 |
| S-metolachlor | 284 | 252 (Q) | 95 | 8 | 10.65 |
| S-metolachlor | 284 | 176.2 | 95 | 24 | 10.65 |
To correct potential analyte losses, matrix effect or volume variation, ISs were used to quantify one or several analytes: Atrazine ring13C3 was used for atrazine, simazine and S-Metolachlor, HA13C3 was used for HA, and DEA 13C3 was used for DEA and DIA. Quantification was carried out by calculating the response factor of each analyte relative to its corresponding IS and the concentrations were determined using a least-square linear regression analysis of the peak area ratio versus the concentration ratio.
Method validation
As high temporal and spatial variation of concentrations were measured in samples, the linearity was studied by injecting seven different concentration levels in the range 0–0.01–0.05–0.1–0.5–1–5 µg L−1. In all cases, the calibration curves showed correlation coefficients (r) greater than 0.999 with a linear regression, based on relative responses between the analyte peak area and its selected IS peak area.
Limits of detection and quantification were calculated using the blank method. It consists in injecting blank (here UP water spiked with IS) ten times and integrating a peak at the retention time of each analyte [12]. When no signal was detected, the noise was integrated and limits of detection and quantification were calculated according to Eqs. (1) and (2), respectively.
| (1) |
| (2) |
with LOD = limit of detection,
LOQ = limit of quantification,
XBl = the mean concentration calculated for the ten blanks
SD0 = standard deviation from the average concentration
However, if mean values were greater than 10 ng L−1, we considered that the sample was not a blank value and the limits of detection and quantification were determined by Eqs. (3) and (4), respectively:
| (3) |
| (4) |
The mean values and standard deviations are reported in Table 4. For S-metolachlor and HA, mean blanks were 15 ng L−1 and 11 ng L−1, respectively, and the LOD and LOQ were calculated with Eqs. (3) and (4). For S-metolachlor, the concentrations were confirmed by the qualifier transition and a signal-to-noise ratio (SNR) above 10. Since S-metolachlor was also detected in drinking water, we suggest that the UP-water system was not sufficiently efficient for this herbicide and linear regression was forced to zero for sample quantification. For HA, SNR was below 3 and the qualifier was too weak to be detected. Moreover, it was not detected in UP water without ISs, suggesting that ISs were the source of this detection. Isotope Ring-labeled HA13C3 was 99% pure and the concentration of ISs in vials was 500 ng L−1. This detection at 10 ng L−1 in all blanks is certainly due to labeled HA13C3 used as EI, and linear regression was then calculated with the blank offset method.
Table 4.
Mean, standard deviation (SD) and calculated LOD and LOQ.
| Compound | Atrazine | DEA | DIA | HA | Simazine | S-metolachlor |
|---|---|---|---|---|---|---|
| Mean (ng L−1) | 4.52 | 3.56 | 1.73 | 10.86 | 0.99 | 14.75 |
| SD (ng L−1) | 0.59 | 2.75 | 1.01 | 5.26 | 0.58 | 1.01 |
| LOD (ng L−1) | 3 | 10 | 4 | 21 | 2 | 4 |
| LOQ (ng L−1) | 10 | 31 | 12 | 63 | 7 | 12 |
For all the other analytes, blank quantification corresponded to noise integration, and Eqs. (1) and (2) were used to calculate LOD and LOQ. Limits of quantification were higher than using the signal-to-noise ratio.
To achieve accurate quantification of the analytes, triplicates of both natural and spiked samples were injected to determine accuracy and repeatability. To avoid volume variation of the spiked sample, 10 µL of a mix of standard solution (500 µg L−1) was added to the 2-mL sample prior to analysis, which corresponds to a final concentration of 2.5 µg L−1 of each compound (Table 5). This protocol was repeated to obtain a final concentration of 0.25 µg L−1 with another sample (Table 6). A sampling site that was not too contaminated was spiked to avoid concentrations higher than 5 µg L−1.
Table 5.
Average, standard deviation of natural and spiked samples and accuracy at 2.5 µg L−1.
| Compound | Atrazine | DEA | DIA | HA | Simazine | S-metolachlor |
|---|---|---|---|---|---|---|
| Average natural sample (ng L−1) | ND | ND | ND | 1344 | ND | ND |
| SD (%) | 6 | |||||
| Average of spiked samples (ng L−1) | 2121 | 2160 | 2202 | 4038 | 2097 | 2249 |
| Standard deviation | 13 | 13 | 14 | 7 | 13 | 11 |
| Recovery (%) | 85 | 86 | 88 | 92 | 84 | 90 |
Table 6.
Average, standard deviation of natural and spiked samples and accuracy at 0.250 µg L−1.
| Compound | Atrazine | DEA | DIA | HA | Simazine | S-metolachlor |
|---|---|---|---|---|---|---|
| Average natural sample (ng L−1) | 76 | 294 | ND | ND | ND | ND |
| Average of spiked samples (ng L−1) | 257 | 256 | 265 | 271 | 253 | 265 |
| Standard deviation | 1 | 2 | 2 | 2 | 5 | 2 |
| Recovery (%) | 103 | 103 | 106 | 108 | 101 | 106 |
Repeatability was estimated using the standard deviation of triplicates of natural (when available) and spiked samples. Accuracy was the difference between the concentration in spiked samples and the theoretical concentration of 2500 ng L−1. For HA, the average in the natural sample concentration was removed to calculate accuracy.
For all analytes, this method was validated with a mean LOQ of 20 ng L−1. This value is high considering the analytical performance of the LC-MS/MS, due to the calculation method. Lower LOQ can be calculated using SNR, especially in pure or tap water but it does not reflect the accuracy requirements for analytical methods. Concentrations lower than 20 ng L−1 were then considered to be not quantified and the first level of analytical curve was kept at 10 ng L−1. Even if only 2 mL of water is sampled, this method show that we still have a good recovery and accuracy compared to previous studies. Uncertainties are estimated to be less than 20% in both low and high spiked concentrations, including uncertainties on the spiked volume.
Declaration of Competing Interest
Authors declare no conflict of interest.
Acknowledgments
The authors particularly thank the Région Ile-de-France and the CNRS EC2Co program for funding.
References
- 1.Barhoum S., Valdes D., Guerin R., Marlin C., Vitale Q., Benmamar J., Gombert P. Spatial heterogeneity of high-resolution Chalk groundwater geochemistry underground quarry at Saint Martin-le-Noeud. France J. Hydrol. 2014;519(Part A):756–768. [Google Scholar]
- 2.Chen N., Valdes D., Marlin C., Blanchoud H., Guerin R., Rouelle M., Ribstein P. Water, nitrate and atrazine transfer through the unsaturated zone of the Chalk aquifer in northern. France Sci. Total Environ. 2019;652:927–938. doi: 10.1016/j.scitotenv.2018.10.286. [DOI] [PubMed] [Google Scholar]
- 3.Koivunen M.E., Dettmer K., Vermeulen R., Bakke B., Geea S.J., Hammock B.D. Improved methods for urinary atrazine mercapturate analysis—assessment of an enzyme-linked immunosorbent assay (ELISA) and a novel liquid chromatography–mass spectrometry (LC–MS) method utilizing online solid phase extraction (SPE) Anal. Chim. Acta. 2006;572:180–189. doi: 10.1016/j.aca.2006.05.037. [DOI] [PubMed] [Google Scholar]
- 4.Masiá A., Ibánez M., Blasco C., Sancho J.V., Picó Y., Hernández F. Combined use of liquid chromatography triple quadrupole mass spectrometry and liquid chromatography quadrupole time-of-flight mass spectrometry in systematic screening of pesticides and other contaminants in water samples. Anal. Chim. Acta. 2013;761:117–127. doi: 10.1016/j.aca.2012.11.032. [DOI] [PubMed] [Google Scholar]
- 5.Jadhav M.R., Pudale A., Raut P., Utture S., Shabeer T.P.A., Banerjee K. A unified approach for high-throughput quantitative analysis of the residues of multi-class veterinary drugs and pesticides in bovine milk using LC-MS/MS and GC–MS/MS. Food Chem. 2019;272:292–305. doi: 10.1016/j.foodchem.2018.08.033. [DOI] [PubMed] [Google Scholar]
- 6.Camilleri J., Baudot R., Wiest L., Vulliet E., Cren-Olivé C., Daniele G. Multiresidue fully automated online SPE-HPLC-MS/MS method for the quantification of endocrine-disrupting and pharmaceutical compounds at trace level in surface water. Int. J. Environ. Anal. Chem. 2015;95(1):67–81. [Google Scholar]
- 7.Rodríguez-González N., Beceiro-González E., González-Castroa M.J., Alpendurada M.F. On-line solid-phase extraction method for determination of triazine herbicides and degradation products in seawater by ultra-pressure liquid chromatography–tandem mass spectrometry. J. Chromatogr. A. 2016;1470:33–41. doi: 10.1016/j.chroma.2016.10.007. [DOI] [PubMed] [Google Scholar]
- 8.Singer H., Jaus S., Hanke I., Lück A., Hollender J., Alder A.C. Determination of biocides and pesticides by on-line solid phase extraction coupled with mass spectrometry and their behavior in wastewater and surface water. Environ. Poll. 2010;158:3054–3064. doi: 10.1016/j.envpol.2010.06.013. [DOI] [PubMed] [Google Scholar]
- 9.Postigo C., López de Alda M.J., Barceló D., Ginebreda A., Garrido T., Fraile J. Analysis and occurrence of selected medium to highly polar pesticides in groundwater of catalonia (NE Spain): an approach based on on-line solid phase extraction–liquid chromatography–electrospray-tandem mass spectrometry detection. J. Hydrol. 2010;383:83–92. [Google Scholar]
- 10.Dinh Q.T., Alliot F., Moreau-Guigon E., Eurin J., Chevreuil M., Labadie P. Measurement of trace levels of antibiotics in river water using on-line enrichment and triple-quadrupole. LC-MS/MS Talanta. 2011;85(3):1238–1245. doi: 10.1016/j.talanta.2011.05.013. [DOI] [PubMed] [Google Scholar]
- 11.Huber U. Automated sample preparation – sample enrichment and clean-up with the Agilent 1100 series valve solutions application note. Agilent. 2010 5989-0510EN. [Google Scholar]
- 12.J.P. Ghestem, M.P. Strub, P. Bados, J. Cabilic, G. Labarraque, C. BrachPapa Conséquences de la directive européenne 2009/90/CE sur l'estimation des limites de quantification des méthodes d'analyse chimique Rapport final BRGM/RP 58049-FR (2009), 96p. infoterre.brgm.fr/rapports/RP-61807-FR.pdf.