Table 2.

 Oxygen consumption and endurance of original pneumatic ventilator for adverse environments (design i) and a conventional commercial pneumatic ventilator under different conditions.

Ventilator Tidal volume; ml Compliance; ml.cmH2 o −1 Rate; breaths.min−1 Endurance of standard 340‐l D cylinder Oxygen consumption; l.min−1
Test ventilator 800 60 10 5 h 0 min 1.13
Oxylog 800 60 10 1 h 34 min 3.62
Test ventilator 800 40 10 4 h 27 min 1.27
Oxylog 800 40 10 1 h 32 min 3.70
Test ventilator 650 60 10 5 h 47 min 0.98
Oxylog 650 60 10 1 h 50 min 3.09
Test ventilator 650 40 10 5 h 50 min 0.97
Oxylog 650 40 10 1 h 55 min 2.96