Abstract
Objectives
To determine if a ventilator available in an emergency department could quickly be modified to provide ventilation for four adults simultaneously.
Methods
Using lung simulators, readily available plastic tubing, and ventilators (840 Series Ventilator; Puritan‐Bennett), human lung simulators were added in parallel until the ventilator was ventilating the equivalent of four adults. Data collected included peak pressure, positive end‐expiratory pressure, total tidal volume, and total minute ventilation. Any obvious asymmetry in the delivery of gas to the lung simulators was also documented. The ventilator was run for almost 12 consecutive hours (5.5 hours of pressure control and more than six hours of volume control).
Results
Using readily available plastic tubing set up to minimize dead space volume, the four lung simulators were easily ventilated for 12 hours using one ventilator. In pressure control (set at 25 mm H2O), the mean tidal volume was 1,884 mL (approximately 471 mL/lung simulator) with an average minute ventilation of 30.2 L/min (or 7.5 L/min/lung simulator). In volume control (set at 2 L), the mean peak pressure was 28 cm H2O and the minute ventilation was 32.5 L/min total (8.1 L/min/lung simulator).
Conclusions
A single ventilator may be quickly modified to ventilate four simulated adults for a limited time. The volumes delivered in this simulation should be able to sustain four 70‐kg individuals. While further study is necessary, this pilot study suggests significant potential for the expanded use of a single ventilator during cases of disaster surge involving multiple casualties with respiratory failure.
Keywords: disaster, ventilator, respiratory failure, surge capacity
References
- American College of Emergency Physicians. Health care system surge capacity recognition, preparedness, and response Ann Emerg Med 2005;45:239. [DOI] [PubMed] [Google Scholar]
- Hick JL, Hanfling D, Burstein JL, et al. Health care facility and community strategies for patient care surge capacity Ann Emerg Med 2004;44:253–261. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hick JL, O'Laughlin DT. Concept of operations for triage of mechanical ventilation in an epidemic Acad Emerg Med 2006;13:223–229. [DOI] [PubMed] [Google Scholar]
- Yale New Haven Health. Emergency Preparedness Briefing. Available at:http://yalenewhavenhealth.org/emergency/commu/briefings/EPBriefingV3I49.pdf. Accessed Apr 25, 2006.
- VORTRAN Medical Technology 1, Inc. VORTRAN Automatic Resuscitator (VAR) User's Guide. Available at:http://www.vortran.com/VAR_User_Guide_2005A.pdf. p 3. Accessed May 8, 2006.
- Darowski M, Englisz M. Artificial ventilation of the lungs for emergencies Front Med Biol Eng 2000;10:177–183. [DOI] [PubMed] [Google Scholar]
- MacIntyre NR. Principles of mechanical ventilation In: Mason RJ, Murray JF, Broaddus VC, Nadel JA. editors. Murray & Nadel's Textbook of Respiratory Medicine. 4th ed.. Philadelphia : PA, 2005. pp. 2342. [Google Scholar]
- Michaels AJ. Management of post traumatic respiratory failure Crit Care Clin 2004;20:83–99. [DOI] [PubMed] [Google Scholar]