Parallel array with axially coded light-sheet microscope

Pablo Loza-Alvarez 1,
PMCID: PMC7170929  PMID: 32351689

Abstract

A parallel array of frequency modulated light sheets results in a scanning-less light sheet microscope capable of fast volumetric imaging.

Subject terms: Light-sheet microscopy, Imaging and sensing


Light-sheet fluorescence microscopy (LSFM) is a powerful technique to visualize, with high spatial resolution, the dynamics occurring within a volume of a living sample. Many implementations have been presented and optimized to capture faster and faster events occurring in the full 3D volume of the sample. Most of them are normally based on the use of complex setups and synchronization schemes. These unavoidably rely on fast scanning of the light sheet using galvanometric scanning mirrors1,2, acousto-optic deflectors3, or polygonal rotating mirrors4. Then, to properly image the signal generated by the moving light sheet, advanced instrumentation is normally required. These include the use of electrically tunable5 or tunable acoustic gradient3 lenses, the use of deformable mirrors for wavefront-coding strategies6, or even the use of photomultiplier tube arrays7 to keep up with the fast image acquisition rates.

On the other hand, there already exist strategies for imaging full volumes of a sample that need no scanning. These methods rely on decoding the information generated from the sample using Fourier domain approaches8, compressive sensing9, or deconvolution algorithms10 based on the frequency or spatial modulation of the excitation beam. These volume illumination approaches can be later demodulated for 3D image reconstruction of the sample. Those schemes do not rely on any light scanning, as the full volume to be imaged is illuminated at once. Such approaches open the possibility to reduce the number of acquisitions for 3D imaging and consequently the acquisition time and the phototoxic effects induced in the sample.

More recently, Ren et al.11 introduced a hybrid yet simple setup that takes advantage of illuminating the full sample and, at the same time, uses the potential of LSFM without the need to perform any scanning. This concept was named coded light-sheet array microscopy (CLAM). The key element in CLAM is the use of two static and nearly parallel highly reflective mirrors12 that, after a number of reflections, reflect back the incoming beam. This element converts the incident beam into a number of virtual sources that are used to generate a parallel array of light sheets, which are sent into the sample. Then, by using a fast rotating amplitude encoder, each of these light sheets is modulated with a different frequency. The Fourier transform of the signal, captured by each pixel of the camera (in the detection arm of the light-sheet microscope), will then contain frequency components that correspond to a different light sheet and therefore to a different depth in the sample.

In CLAM, by controlling the angle between the two tilted mirrors, the light-sheet array can be reconfigurable in both the number of planes or the array density and coherency. Furthermore, a 100% spatial duty cycle in detection can be achieved, which is translated into a longer voxel dwell time and therefore offers the possibility to increase the signal-to-noise ratio and, therefore, enables a reduced light dose. Finally, this approach can be adapted to any existing LSFM system with minimal hardware or software modification.

The future challenge for CLAM relies on its own nature, as it is based on its capacity to image a number of consecutive parallel light sheets. These require the need for strategies to extend the depth of field, which intrinsically trades-off resolution. Nevertheless, CLAM is a general and versatile technique that can be combined with the multiple strategies normally implemented in LSFM. For example, it can be used in a simultaneous multiview configuration, or can be combined with the use of adaptive optics for aberration correction or, more generally, with wavefront coding/shaping techniques. The additional degree of freedom that it provides allows for the creation of sophisticated patterns, for example, for structured illumination, opening up new possibilities, and applications.

Acknowledgements

PL-A acknowledges support form the spanish Ministry of Economy and Competitiveness through the “Severo Ochoa” program for Centres of Excellence in R&D (SEV-2015-0522), Fundació Privada Cellex, Fundación Mig-Puig and the Generalitat de Catalunya through the CERCA program.

Conflict of interest

The author declares that he has no conflict of interest.

References

  • 1.Kumar S, et al. High-speed 2D and 3D fluorescence microscopy of cardiac myocytes. Opt. Express. 2011;19:13839–13847. doi: 10.1364/OE.19.013839. [DOI] [PubMed] [Google Scholar]
  • 2.Voleti V, et al. Real-time volumetric microscopy of in vivo dynamics and large-scale samples with SCAPE 2.0. Nat. Methods. 2019;16:1054–1062. doi: 10.1038/s41592-019-0579-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Duocastella M, et al. Fast inertia-free volumetric light-sheet microscope. ACS Photonics. 2017;4:1797–1804. doi: 10.1021/acsphotonics.7b00382. [DOI] [Google Scholar]
  • 4.Bouchard MB, et al. Swept confocally-aligned planar excitation (SCAPE) microscopy for high-speed volumetric imaging of behaving organisms. Nat. Photonics. 2015;9:113–119. doi: 10.1038/nphoton.2014.323. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Fahrbach FO, et al. Rapid 3D light-sheet microscopy with a tunable lens. Opt. Express. 2013;21:21010–21026. doi: 10.1364/OE.21.021010. [DOI] [PubMed] [Google Scholar]
  • 6.Olarte OE, et al. Decoupled illumination detection in light sheet microscopy for fast volumetric imaging. Optica. 2015;2:702–705. doi: 10.1364/OPTICA.2.000702. [DOI] [Google Scholar]
  • 7.Martin C, et al. Line excitation array detection fluorescence microscopy at 0.8 million frames per second. Nat. Commun. 2018;9:4499. doi: 10.1038/s41467-018-06775-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Judkewitz B, Yang C. Axial standing-wave illumination frequency-domain imaging (SWIF) Opt. Express. 2014;22:11001–11010. doi: 10.1364/OE.22.011001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Calisesi G, et al. Spatially modulated illumination allows for light sheet fluorescence microscopy with an incoherent source and compressive sensing. Biomed. Opt. Express. 2019;10:5776–5788. doi: 10.1364/BOE.10.005776. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Prevedel R, et al. Simultaneous whole-animal 3D imaging of neuronal activity using light-field microscopy. Nat. Methods. 2014;11:727–730. doi: 10.1038/nmeth.2964. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Ren YX, et al. Parallelized volumetric fluorescence microscopy with a reconfigurable coded incoherent light-sheet array. Light.: Sci. Appl. 2020;9:8. doi: 10.1038/s41377-020-0245-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Wu JL, et al. Ultrafast laser-scanning time-stretch imaging at visible wavelengths. Light.: Sci. Appl. 2017;6:e16196. doi: 10.1038/lsa.2016.196. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Light, Science & Applications are provided here courtesy of Nature Publishing Group

RESOURCES