Is High Definition MLC Dosimetrically Superior to Standard Definition MLC for SIB-SBRT for Carcinoma Prostate

Bijina T K 1,2,*, K M Ganesh 3, Pichandi A 2
PMCID: PMC7173375  PMID: 31870127

Abstract

Objective:

The study was conducted to quantitatively evaluate the dosimetric effects of high definition (2.5 mm) and standard definition (5.0 mm) MLC on the quality of SBRT plans using SIB-IMRT and SIB-VMAT technique for carcinoma prostate and also to evaluate the dosimetric advantage of one technique over the other.

Materials and Methods:

Seventeen annonymized planning CT data sets were used to generate plans for both VMAT and IMRT techniques using 2.5 mm and 5.0 mm MLC.The prescription to the nodule was 45Gy in 5 fractions and to the prostate was 35Gy in 5 fractions.CI, GI, D2%, D98%, D50% and V95% for target; D2%, Dmean, V80%, V20% for OAR’s; V5% of the irradiated volume, and delivered MU’s were analyzed.An independent t-test was used to compare the plans. Patient specific QA for all plans were also performed and analyzed.

Results:

Minor difference in dosimetric indices was observed between 2.5mm and 5mm MLC VMAT plans, except D2% (PTV35) and D98% (GTV45) were better in 2.5mm MLC plans (p<0.05).D2% and D50% to GTV45 with p<0.009 and <0.03 respectively, and D2% of bladder (p<0.03) were significantly lesser for VMAT plans compared to IMRT plans. GI was significantly better for 2.5mm MLC plans in both techniques (p<0.05). 2.5mm MLC plans on an average had 9.9% and 7% more MU for VMAT and IMRT plans respectively.

Conclusion:

All the plans provided adequate dose coverage to target. Highly gradient plans with better conformity were achieved with 2.5mm MLC and VMAT combination with an increase in MU delivered, compared to 5mm MLC plans.

Key Words: IMRT, VMAT, MLC, Prostate, SIB, SBRT

Introduction

Radiation therapy (RT) is one of the important modalities in treatment of cancers, which accounts for approximately 50% of all cancer patients. With the advancements in RT techniques when compared to the older conventional RT treatment techniques, intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT), has resulted in an advantage of higher dose delivery to the cancer cells and minimized dose to the organs at risk (OAR) and healthy tissue. The higher dose delivery is made possible by the use of multileaf collimator (MLC), which generates a very steep dose gradient and non-uniform dose during treatment delivery with a constantly changing field shape in IMRT and with the further addition of a constantly moving gantry in VMAT (Woon et al., 2018).

The prostate is one of the most common tumor sites where dosimetric comparison between VMAT and IMRT is performed. Though with VMAT, shortened treatment time is a common finding, there are inconsistencies noted in the dosimetric outcome. Studies on dosimetric comparison in prostate-only or prostate with seminal vesicles found that one-arc VMAT provided equivalent or superior target coverage and OAR sparing compared with 5-f IMRT or 7-f IMRT (Ishii et al., 2013).

The resolution of the modulated photon beam intensities or fluence, in VMAT is determined by the MLC leaf width, just as in IMRT. Quite a few studies have investigated the relationship between MLC leaf width and the quality of IMRT plans and have shown that finer leaf widths can result in more conformal dose distributions for target volumes and less dose delivered to normal tissue when the target volumes were small (Rodal et al., 2010; Shang et al., 2013; Hong et al., 2014). Similarly, few studies have shown the dosimetric advantage of using MLCs with narrow leaf width (Lafond et al., 2013; Park et al., 2016). However, there are very little data on the comparison of the dosimetric effect of narrow width MLC on the quality of IMRT and VMAT plans with simultaneous integrated boost techniques in our study setting.

Consequently, this study was conducted to quantitatively evaluate the dosimetric effect of 2.5 mm and 5.0 mm MLC on the quality of simultaneous integrated boost (SIB) IMRT and VMAT stereotactic body radiotherapy (SBRT) plans for carcinoma prostate and also to evaluate the dosimetric advantage of one technique over the other.

Materials and Methods

This is a retrospective study using 17 annonymized planning CT data sets previously treated with SBRT at our center. Plans were made for both VMAT and IMRT techniques using 2.5 mm and 5.0 mm MLC.

The entire prostate gland was contoured as CTV with 0.3cm margin given around CTV to create PTV35. The dominant nodule within the prostate gland was contoured as GTV45 with no margins. Organs at risks (OARs) such as the urinary bladder, rectum, penile bulb urethra, small bowel and bilateral femoral heads were delineated.

Treatment planning

The CT data and RT structures were exported to the planning system and the treatment plans were generated with Eclipse version 13.6 (Varian Medical Systems, Palo Alto, CA) TPS for which QA is done periodically. For IMRT plan optimization, Dose Volume Optimizer and for VMAT, Progressive Resolution Optimizer were used. The dose calculation was performed using Anisotropic Analytical Algorithm. The prescribed dose to PTV35 was 35 Gy in 5 fractions with simultaneous integrated boost to GTV45 with 45Gy in 5 fractions. The parameters such as the isocenter location, the number of fields, the MLC margin, the gantry, collimator, and couch angles for each beam and the dose constraints were constant for both MLC plans. The set treatment planning dosimetry goals were enumerated in Table 1. No limit on target dose heterogeneity was specified, but effort was made to limit the maximum dose (Dmax) to GTV <120%. The IMRT plans consisted of single-isocenter, coplanar, and 9 equidistant fields delivered in dynamic MLC mode. The VMAT plans were implemented with a single-isocenter 2-full arc, and without couch rotation. The collimator angles were set to 30° and 330° for each arc. The gantry angle range of each arc was 179.9°-181.1°. Thus, 68 treatment plans were generated for two MLC leaf widths (2.5 mm and 5.0 mm), and two types of planning techniques (VMAT and IMRT). All the plans were normalized to cover 95% of GTV45 and PTV35 to their respective prescription dose. The dosimetric outcomes were compared for both MLC’s and between the treatment techniques.

Table 1.

Dose Constraints of Target and OAR’s

Structure Planning goals
Targets
GTV45 V45Gy ≥95%
PTV35 V35Gy ≥95%
OAR’s
Rectum V50%(22.5Gy)< 50%
V80%(36Gy)< 20%
V100%(45Gy) < 5%
Bladder V50% (22.5Gy)< 40%
V100% (45Gy)<5%
Urethra Max point dose should not exceed 110%
Femoral heads V40%(18Gy) < 5%

*OAR, Organ At Risk; V45Gy, Volume receiving 45 Gy; V35Gy, Volume receiving 35 Gy; V50%, Volume receiving 50% of prescription dose; V80%, Volume receiving 80% of prescription dose; V100%, Volume receiving 100% of prescription dose; V40%, Volume receiving 40% of the prescription dose

Evaluation parameters

The evaluation parameters included dose received by 98% of both GTV45 and PTV35 (D98%), D2%, D50%, the percentage volume of target receiving at least 95% of the prescription dose (V95%), Conformity Index (CI), and Gradient Index (GI). The definition of each index is summarized below.

CI: The ratio used to evaluate the quality of fit of the target volume to the prescription isodose volume. A score of 1 represents a perfectly conformal plan.

CI = (TVPIV/PIV) x (TVPIV/TV)

Wherein, VPIV represents volume of PTV receiving the prescription dose; PIVrepresents prescription isodose volume and TV represents target volume (Van’t Riet et al., 1997). CI is calculated for both PTV35 and GTV45.

GI: The index that represents the degree of dose drop-off outside the target volume. A smaller value of GI indicates a better degree of dose drop-off outside the target volume.

GI = PIV50/PIV

Where PIV50 is the volume receiving at least 50% dose of the prescription dose; PIV is the prescription isodose volume (Paddick, Lippitz 2006).

The irradiated volume receiving more than 5% of the prescription dose (V5%) is recorded for all the plans. The mean dose, D2%and the dose volumes V80% and V20% were analyzed for bladder and rectum. D2% and mean dose were compared for bilateral femoral heads. Plan efficiency was determined using the treatment delivery parameters such as MU and estimated treatment time per fraction.

Statistical analysis

All the data were entered into MS Excel sheet. An independent t-test was used to analyze and compare the dosimetric indices between IMRT and VMAT plans according to the MLC leaf width. The statistical analysis was conducted using SPSS version 16.0 and a p value of <0.05 was considered statistically significant.

Results

Seventeen patients were included in this study. The median volume of GTV45 and PTV35 was 5.2cm3 and 70.8 cm3 respectively. The DVH evaluation parameters averaged over all patients were shown in Table 2, 3 and 4. A comparison of dose distributions for both MLC and treatment technique combination of one representative patient was shown in Figure 1.

Table 2.

DVH Parameters among 5.0mm vs2.5mm MLC and VMAT vs IMRT

Variables Treatment Techniques MLCs leaf width
p-value between MLCs
5.0mm 2.5mm
PTV35
D2% (Gy) VMAT 43.22±0.58 42.68±0.49 0.02*
IMRT 42.97±1.09 42.64±2.64 0.68
P-value between techniques 0.48 0.95
VMAT 34.07±0.35 34.29±0.24 0.07*
D98% (Gy) IMRT 34.02±0.36 33.93±0.47 0.57
P-value between techniques 0.73 0.02*
VMAT 38.51±0.75 38.15±0.56 0.18
D50% (Gy) IMRT 37.47±0.59 37.79±0.56 0.16
P-value between techniques 0.001* 0.12
VMAT 99.46±0.29 99.62±0.15 0.09
V95% (%) IMRT 99.38±0.36 99.41±0.43 0.88
P-value between techniques 0.58 0.11
GTV45
VMAT 47.88±0.57 47.67±0.45 0.31
D2% (Gy) IMRT 48.14±0.52 48.31±0.67 0.48
P-value between techniques 0.25 0.009*
VMAT 44.44±0.14 44.60±0.16 0.01*
D98% (Gy) IMRT 44.39±0.21 44.65±0.79 0.29
P-value between techniques 0.52 0.86
VMAT 46.55±0.24 46.44±0.25 0.26
D50% (Gy) IMRT 46.86±0.44 46.80±0.49 0.75
P-value between techniques 0.03* 0.028*
VMAT 99.89±0.10 99.93±0.09 0.36
V95% (%) IMRT 99.83±0.21 99.87±0.21 0.67
P-value between techniques 0.31 0.32

* indicates differences to be statistically significant (p<0.05); Values represent Mean±SD; D2% , - Dose to 2% volume, D98%- Dose to 98% volume, D50%- Dose to 50% volume,V95% - Volume receiving 95% of prescription dose

Table 3.

Dosimetric Indices among 5.0mm vs 2.5mm MLC and VMAT vs IMRT

Variables Treatment Techniques MLCs leaf width
p-value between MLCs
5.0mm 2.5mm
CIGTV45 VMAT 0.70±0.09 0.75±0.08 0.13
IMRT 0.69±0.09 0.67±0.09 0.45
P-value between techniques 0.88 0.02*
CIPTV35 VMAT 0.85±0.07 0.89±0.03 0.29
IMRT 0.82±0.03 0.82±0.03 0.85
P-value between techniques 0.45 0.009*
GI VMAT 3.81±0.30 3.81±0.24 0.98
IMRT 4.38±0.34 4.04±0.38 0.02*
P-value between techniques 0.09 <0.0001*
MUs VMAT 2806.23±429.50 3085.32±650.25 0.21
IMRT 2739.31±318.09 2930.03±350.49 0.16
P-value between techniques 0.66 0.46
V5Gy (cc) VMAT 2760.68±736.76 3056.49±835.80 0.35
IMRT 2458.29±602.43 2743.91±698.32 0.28
P-value between techniques 0.26 0.31

* indicates differences to be statistically significant (p<0.05); Values represent Mean±SD; CIGTV45- Conformity Index (GTV45); CIPTV35- Conformity Index (PTV35); GI- Gradient Index; MU- Monitor units; V5Gy-Volume receiving 5Gy

Table 4.

Comparison of Dose Received by OAR’s

Variables Treatment Techniques MLC leaf width
p-value between MLCs
5.0mm 2.5mm
Bladder
V80%§ VMAT 0.07 (0-0.67) 0.02 (0-0.59) 0.54
[Median (Range)] IMRT 0.01 (0-0.92) 0.01 (0-0.7) 0.58
P-value between techniques 0.96 0.72
V20% VMAT 35.69±12.22 36.68±12.42 0.84
IMRT 38.42±11.03 44.22±11.64 0.21
P-value between techniques 0.56 0.12
Dmean (Gy) VMAT 8.17±2.26 8.27±2.21 0.91
IMRT 8.69±1.98 9.88±2.40 0.18
P-value between techniques 0.53 0.09
D2% (Gy) VMAT 25.86±5.67 25.62±5.47 0.91
IMRT 28.94±3.72 29.66±3.67 0.63
P-value between techniques 0.11 0.03*
Rectum
V80%§ VMAT 0.15 (0-0.97) 0.15 (0-0.99) 0.72
[Median (Range)] IMRT 0.14 (0-1.02) 0.13 (0-1.34) 0.8
P-value between techniques 0.58 0.42
VMAT 48.64±13.28 50.72±14.10 0.7
V20% IMRT 54.35±15.62 58.72±16.34 0.49
P-value between techniques 0.33 0.19
Dmean (Gy) VMAT 9.85±2.28 9.77±2.22 0.93
IMRT 10.38±2.26 11.43±2.29 0.25
P-value between techniques 0.55 0.07
D2% (Gy) VMAT 28.19±3.27 27.97±3.29 0.87
IMRT 29.56±3.46 30.81±4.09 0.41
P-value between techniques 0.31 0.06
Lt.Femur
Dmean (Gy) VMAT 7.88 ±1.98 8.24±2.22 0.67
IMRT 7.77±2.03 7.46±2.15 0.71
P-value between techniques 0.89 0.37
D2% (Gy) VMAT 12.46±1.85 13.18±5.04 0.63
IMRT 13.54±1.49 13.45±2.04 0.89
P-value between techniques 0.12 0.86
Rt. Femur
Dmean (Gy) VMAT 7.89±1.92 7.98±1.76 0.89
IMRT 8.03±2.09 7.75±2.05 0.73
P-value between techniques 0.86 0.75
D2% (Gy) VMAT 12.06±2.11 13.41±4.85 0.37
IMRT 13.37±1.27 13.28±1.51 0.86
P-value between techniques 0.07 0.93

* indicates differences to be statistically significant (P<0.05); §-Mann Whitney U test applied; V80%- Volume receiving 80% of prescription dose, V20%- Volume receiving 20% of prescription dose; V20%- Volume receiving 20% of prescription dose, D2%- Dose to 2% volume

Figure 1.

Figure 1

Isodose Comparison of One Representative Patient

Dosimetric analysis of 5.0mm v/s2.5mm MLC in VMAT

The minimum dose received by 2% of the PTV35was noted to be significantly higher in 5.0mm MLC (5.0mm v/s 2.5mm: 43.22±0.58 Gy v/s 42.68±0.49 Gy) (p<0.05). The minimum dose received by 98% of the PTV35 (5.0mm v/s 2.5mm: 34.07±0.35 Gy v/s 34.29±0.24 Gy) and GTV45was significantly higher in 2.5mm MLC (5.0mm v/s 2.5mm: 44.44±0.14 Gy v/s 44.60±0.16 Gy) (p<0.05). The CIGTV45 and CIPTV35 did not vary significantly with the width of MLC (p>0.05) as shown in Table 3. Dmean, D2%, V80% and V20% of bladder, rectum, left and right femur were not significant in either of the MLC widths (p>0.05) as shown in Table 4. No statistically significant difference was observed with respect to MU between 5mm and 2.5mm MLC plans.

Dosimetric analysis of 5.0mm v/s 2.5mm MLC in IMRT

Among the dosimetric indices and DVH parameters viz., minimum dose received by 2%, 50% and 98% of the PTV35and GTV35, CI, GI and V5%to irradiated volume, none of the indices varied significantly with the width of MLC (p>0.05) but for the GI which was better in 2.5mm MLC for IMRT wherein it was significantly lower (5.0mm v/s 2.5mm: 4.38±0.34 v/s 4.04±0.38) (p<0.05). Damon and D2% and V80% and V20% of bladder, rectum, left and right femur were not significant in either of the MLC widths (p>0.05) as shown in Table 4. In IMRT technique, 2.5mm MLC plans produced more MU compared to 5mm MLC plans which is not statistically significant (p>0.05).

Dosimetric analysis of VMAT v/s IMRT

The minimum dose received by 50% (VMAT v/s IMRT: 34.29±0.24 Gy v/s 33.93±0.47 Gy) and 98% (VMAT v/s IMRT: 38.51±0.75 Gy v/s 37.47±0.59 Gy) of the PTV35 was significantly higher in VMAT for 2.5mm and 5.0mm MLC respectively. D2% of GTV45 was significantly lesser in VMAT for 2.5mm MLC (VMAT v/s IMRT: 47.67±0.45 Gy v/s 48.31±0.67 Gy) and for D50% both 2.5mm (VMAT v/s IMRT: 46.44±0.25 Gy v/s 46.80±0.49 Gy) and 5.0mm (VMAT v/s IMRT: 46.55±0.24 Gy v/s 46.86±0.44 Gy) MLC respectively (p<0.05). The CIPTV35 and CIGTV45 was no different either between the techniques except in 2.5mm MLC which was significantly better in VMAT compared to IMRT (VMAT v/s IMRT: 0.75±0.08 v/s 0.67±0.09, 0.89±0.03 v/s 0.83±0.03) respectively (p<0.05). The GI was significantly better in VMAT when compared to the IMRT technique for 2.5mm MLC (VMAT v/s IMRT: 3.81±0.24 v/s 4.04±0.38). Damon and D2%and V80% and V20% of bladder, rectum, left and right femur were not significant in either of the techniques except for D2% for bladder in IMRT technique in 2.5mm MLC (VMAT v/s IMRT: 25.62±5.47 Gy v/s 29.66±3.67 Gy) which was significantly lesser in VMAT technique (p<0.05) as shown in Table 4.

For each patient, four QA plans were delivered in PTW Octavius 729 array detector. All plans yielded equivalent results and passed our criteria of gamma index with ≥ 95% with 3%, 3mm distance to agreement as shown in Table 5.

Table 5.

Gamma Analysis Percentage Passing Rate (DD 3%, DTA 3 mm)

Patient No. IMRT
VMAT
2.5 mm MLC 5 mm MLC 2.5 mm MLC 5 mm MLC
1 99.5 99.7 99.2 99.3
2 98.7 99.0 98.2 98.9
3 99.1 99.1 99.3 99.2
4 97.5 98.4 98.6 99.0
5 99.5 100 99.4 99.5
6 96.4 97.1 98.5 98.8
7 97.4 97.8 97.8 98.3
8 97.9 97.8 97.3 97.8
9 98.2 98.1 99.4 99.5
10 99.2 99.8 99.5 99.0
11 100.0 100.0 100.0 100.0
12 95.8 96.5 96.8 98.3
13 98.3 99.0 99.5 99.4
14 95.7 96.3 96.9 97.2
15 99.9 100.0 100.0 100.0
16 96.3 97.5 97.8 97.5
17 98.9 98.5 99.5 100.0

DD, Dose Difference; DTA, Distance to Agreement; IMRT, Intensity Modulated Radiotherapy; VMAT, Volumetric Modulated Arc Therapy; MLC, Multileaf collimator

Discussion

In view of developments in MLCs along with the improvements in radiotherapy planning techniques and delivery methods, radiosurgery has been extensively used and dosimetric superiority and clinical effectiveness have been elicited in the literatures (Hong et al., 2014). However, there has paucity of research on the effects of MLC leaf width on volumetric-modulated arc therapy (VMAT) and also in IMRT in the current study setting. Hence we have compared the dosimetric indices in VMAT and IMRT and also the superiority of one technique over the other. The treatment plans were successful in satisfying the protocol-defined goals in all generated plans. The PTV and GTV coverage was excellent, with 95% of PTV35 and GTV45received the prescription dose in all the plans. All plans used in this study also met the OAR constraints.

The minimum doses received by 98% of the PTV and GTV were significantly higher in 2.5mm MLC VMAT and similar results with respect to dosimetric indices except for the significance has been observed in few studies (Chaeet al., 2014; Serna et al., 2015; Zhang et al., 2018). The minimum doses received by 2% of the PTV were noted to be significantly higher in 5.0mm MLC. The impact of MLC of 5-mm and 10-mm leaf width on VMAT in the case of prostate was found with increased OARs sparing for thinner MLCs (Kesterenet al., 2012).

Jacob et al., (2010) demonstrated only small dosimetric effects of leaf widths between 2.5 mm and 5 mm in regard to target coverage. There was no significant difference in sparing of OAR’s between the IMRT plans with different leaf widths. The mean GI was significantly better in 2.5mm MLC in a study by Chae et al., (2014), and it was correlating to the current study finding where in it was significantly better in 2.5mm.

A study by Quan et al., (2012) revealed that IMRT plan quality was similar or superior to that of VMAT when the number of beams in IMRT was increased to a certain number between 12 to 24.Many studies mention VMAT to be better in target volume coverage (Palma et al., 2008; Kristoffersen et al., 2009; Hardcastle et al., 2011). When compared VMAT with IMRT plans for ten prostate patients, better rectal, bladder and femoral head sparing or normal tissue sparing was observed with the VMAT plans over the IMRT and 3DCRT plans (Palma et al., 2008; Kristoffersenet al., 2009). Similarly, with better gradient index in VMAT, the D2 dose delivery for bladder in 2.5mm MLC was significantly less compared to IMRT and it was achieved in spite of better conformity in IMRT (Palma et al., 2008; Kristoffersenet al., 2009; Hardcastle et al., 2011).

Though statistically insignificant difference in MU delivered was observed between 5mm and 2.5mm leaf width for both techniques in our study, leakage radiation is one of the major components that would directly correlate with increase in MU delivered. Many literatures have reported that leakage radiation would contribute to the integral dose to peripheral normal tissue. This may lead to increased risk of secondary malignancy (Sakthivel et al., 2017).

Highly conformal dose distribution with different dose levels within the primary tumor can be planned and delivered with the help of modern radiotherapy techniques. The combination of two MLC leaf widths with two different techniques was explored in this study. All the plans provided adequate dose coverage to target. Highly gradient plans with better conformity were achieved with 2.5mm MLC and VMAT combination with an increase in MU delivered, compared to 5mm MLC plans. The 2.5mm MLC in combination with VMAT technique resulted in dosimetric benefits in the treatment of prostate SIB SBRT.

Conflict of interest

Authors have declared no conflict of interest.

References

  1. Chae SM, Lee GW, Son SH. The effect of multileaf collimator leaf width on the radiosurgery planning for spine lesion treatment in terms of the modulated techniques and target complexity. Radiat Oncol. 2014;9:72. doi: 10.1186/1748-717X-9-72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Hardcastle N, Tome WA, Foo K, et al. Comparison of prostate IMRT and VMAT biologically optimised treatment plans. Med Dosi. 2011;36:292–8. doi: 10.1016/j.meddos.2010.06.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Hong CS, Ju SG, Kim M, et al. Dosimetric effects of multileaf collimator leaf width on intensity-modulated radiotherapy for head and neck cancer. Med Phys. 2014;41:021712. doi: 10.1118/1.4860155. [DOI] [PubMed] [Google Scholar]
  4. Ishii K, Ogino R, Okada W, et al. A dosimetric comparison of RapidArc and IMRT with hypofractionated simultaneous integrated boost to the prostate for treatment of prostate cancer. Br J Radiol. 2013;86:1030. doi: 10.1259/bjr.20130199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Jacob V, Bayer W, Astner ST, Busch R, Kneschaurek P. A planning comparison of dynamic IMRT for different collimator leaf thicknesses with helical tomotherapy and RapidArc for prostate and head and neck tumors. Strahlenther Onkol. 2010;186:502–10. doi: 10.1007/s00066-010-2124-3. [DOI] [PubMed] [Google Scholar]
  6. Kesteren Z, Janssen TM, Damen E, van Vliet-Vroegindeweij C. The dosimetric impact of leaf interdigitation and leaf width on VMAT treatment planning in Pinnacle: Comparing Pareto fronts. Phys Med Biol. 2012;57:2943–2. doi: 10.1088/0031-9155/57/10/2943. [DOI] [PubMed] [Google Scholar]
  7. Kristoffersen F, Ohlhues L, Medin J, Korreman S. 2009), RapidArc volumetric modulated therapy planning for prostate cancer patients. Acta Oncol. 48:227–32. doi: 10.1080/02841860802266748. [DOI] [PubMed] [Google Scholar]
  8. Lafond C, Chajon E, Devillers A, et al. Impact of MLC leaf width on volumetric-modulated arc therapy planning for head and neck cancers. J Appl Clin Med Phys. 2013;14:4074. doi: 10.1120/jacmp.v14i6.4074. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg. 2006;105:194–201. doi: 10.3171/sup.2006.105.7.194. [DOI] [PubMed] [Google Scholar]
  10. Palma D, Vollans E, James K, et al. Volumetric modulated arc therapy (VMAT) for delivery of prostate radiotherapy: Reduction in treatment time and monitor unit requirements compared to intensity modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2008;72:312. doi: 10.1016/j.ijrobp.2008.02.047. [DOI] [PubMed] [Google Scholar]
  11. Park JM, Park SY, Kim JH, Carlson J, Kim JI. The effect of extremely narrow MLC leaf width on the plan quality of VMAT for prostate cancer. Radiat Oncol. 2016;11:85. doi: 10.1186/s13014-016-0664-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Quan EM, Li X, Li Y, et al. A comprehensive comparison of IMRT and VMAT plan quality for prostate cancer treatment. Int J Radiat Oncol Biol Phys. 2012;83:1169–78. doi: 10.1016/j.ijrobp.2011.09.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Rodal J, Sovik A, Malinen E. Influence of MLC leaf width on biologically adapted IMRT plans. Acta Oncol. 2010;49:1116–23. doi: 10.3109/0284186X.2010.498832. [DOI] [PubMed] [Google Scholar]
  14. Sakthivel V, Kadirampatti Mani G, Mani S, Boopathy R, Selvaraj J. Estimating second malignancy risk in intensity-modulated radiotherapy and volumetric-modulated arc therapy using a mechanistic radiobiological model in radiotherapy for carcinoma of left breast. J Med Phys. 2017;42:234–40. doi: 10.4103/jmp.JMP_89_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Serna A, Puchades V, Mata F, Ramos D, Alcaraz M. Influence of multi-leaf collimator leaf width in radiosurgery via volumetric modulated arc therapy and 3D dynamic conformal arc therapy. Phys Med. 2015;31:293–306. doi: 10.1016/j.ejmp.2015.01.011. [DOI] [PubMed] [Google Scholar]
  16. Shang Q, Qi P, Ferjani S, Xia P. Effect of MLC leaf width on treatment adaptation and accuracy for concurrent irradiation of prostate and pelvic lymph nodes. Med Phys. 2013;40:061701. doi: 10.1118/1.4803499. [DOI] [PubMed] [Google Scholar]
  17. Van’tRiet A, Mak AC, Moerland MA, Elders LH, van der Zee W. A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate. Int J Radiat Oncol Biol Phys. 1997;37:731–6. doi: 10.1016/s0360-3016(96)00601-3. [DOI] [PubMed] [Google Scholar]
  18. Woon WA, Ravindran PB, Ekayanake P, Lim YY. Validation of delivery consistency for intensity-modulated radiation therapy and volumetric-modulated arc therapy plans. J Med Phys. 2018;43:119–28. doi: 10.4103/jmp.JMP_109_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Zhang F, Xu W, Jiang H, et al. Dosimetric evaluation of VMAT radiation therapy technique for breast cancer after conservative surgery based on three different types of multileaf collimators. Oncol Transl Med. 2018;4:208–14. [Google Scholar]