Abstract
Background
Postbariatric hypoglycemia (PBH) can threaten safety and reduce quality of life. Current therapies are incompletely effective.
Methods
Patients with PBH were enrolled in a double-blind, placebo-controlled, crossover trial to evaluate a closed-loop glucose-responsive automated glucagon delivery system designed to reduce severe hypoglycemia. A hypoglycemia detection and mitigation algorithm was embedded in the artificial pancreas system connected to a continuous glucose monitor (CGM, Dexcom) driving a patch infusion pump (Insulet) filled with liquid investigational glucagon (Xeris) or placebo (vehicle). Sensor/plasma glucose responses to mixed meal were assessed during 2 study visits. The system delivered up to 2 doses of study drug (300/150 μg glucagon or equal-volume vehicle) if triggered by the algorithm. Rescue dextrose was given for plasma glucose <55 mg/dL or neuroglycopenia.
Results
Twelve participants (11 females/1 male, age 52 ± 2, 8 ± 1 years postsurgery, mean ± SEM) completed all visits. Predictive hypoglycemia alerts prompted automated drug delivery postmeal, when sensor glucose was 114 ± 7 vs 121 ± 5 mg/dL (P = .39). Seven participants required rescue glucose after vehicle but not glucagon (P = .008). Five participants had severe hypoglycemia (<55 mg/dL) after vehicle but not glucagon (P = .03). Nadir plasma glucose was higher with glucagon vs vehicle (67 ± 3 vs 59 ± 2 mg/dL, P = .004). Plasma glucagon rose after glucagon delivery (1231 ± 187 vs 16 ± 1 pg/mL at 30 minutes, P = .001). No rebound hyperglycemia occurred. Transient infusion site discomfort was reported with both glucagon (n = 11/12) and vehicle (n = 10/12). No other adverse events were observed.
Conclusion
A CGM-guided closed-loop rescue system can detect imminent hypoglycemia and deliver glucagon, reducing severe hypoglycemia in PBH.
Clinical Trials Registration
Keywords: hypoglycemia, gastric bypass, glucagon, closed-loop
With the rising use of bariatric surgery, one increasingly recognized complication of Roux-en-Y gastric bypass (RYGB) is hypoglycemia (1). Symptomatic hypoglycemia can also occur after sleeve gastrectomy, but prevalence remains uncertain because the limited number of clinical studies in the postsleeve population have analyzed glucose response after liquid meals or glucose loads (2,3). This condition, termed postbariatric hypoglycemia (PBH), is typically recognized between 1 and 3 years postoperatively, and causes not only distressing adrenergic and cholinergic symptoms but also neuroglycopenia and hypoglycemia unawareness, threatening safety. Episodes of hypoglycemia impair cognition and increase the risk for syncope, arrhythmias, seizures, coma, and even death. Moreover, many patients are disabled by their hypoglycemia because of their inability to perform job-related tasks and to safely operate a motor vehicle.
Patients affected by PBH demonstrate distinct postprandial glucose patterns, with a very rapid rise in glucose shortly after food intake (mean rise rate 9 ± 1 mg/dL/min in a previous study (4)), linked to rapid delivery and intestinal absorption. These high peak glucose levels, together with increased postprandial levels of GLP1 and other intestinally derived hormones (5–7), stimulate excessive insulin secretion and contribute to the very rapid lowering of glucose to hypoglycemic levels 1 to 3 hours after eating. Insulin-independent glucose disposal is also increased in PBH. Moreover, counterregulatory hormones, including glucagon (5), catecholamines, and cortisol, are reduced during experimentally induced hypoglycemia in post-RYGB patients (8).
Given the multiple factors promoting hypoglycemia and reduced counterregulatory responses to hypoglycemia, it is not surprising that management of PBH poses distinct challenges for patients and physicians alike. Initial treatment focuses on reducing intake of simple carbohydrates and use of acarbose to minimize the postmeal glucose “spikes.” Strategies aimed at reducing incretin and insulin responses to meals include somatostatin receptor analogues (eg, octreotide), diazoxide, and even reversal of the surgical procedure for refractory PBH (9,10). Unfortunately, these therapies are limited by side effects or incomplete efficacy, even in combination. Although partial pancreatectomy was used initially to reduce islet mass (11–13), recurrence of hypoglycemia is observed (14), underscoring the complex metabolic interactions beyond insulin secretion that contribute to hypoglycemia. A major challenge for patients affected by PBH is that glucose levels drop very quickly in the postprandial state (mean fall rate 7 ± 1 mg/dL/min in a previous study (4)), making it very difficult to respond to and treat impending hypoglycemia before neuroglycopenia develops; this problem is of even greater magnitude in patients with hypoglycemia unawareness.
With the incomplete efficacy of currently available therapies, there is an urgent need for novel approaches for treatment of severe hypoglycemia to maintain health, improve quality of life, and improve safety. Glucagon can be used successfully to treat acute hypoglycemia in PBH. Currently available glucagon preparations limit utilization because of 1) need to reconstitute powder immediately before use, 2) expense of single-use emergency kits, and 3) side effects such as nausea and rebound hyperglycemia with traditional rescue doses (0.5–1.0 mg) (15). A newly developed stable liquid formulation of native glucagon (16) can be delivered via infusion pump, allowing lower doses of glucagon to be delivered only when hypoglycemia is imminent.
We thus hypothesized that a novel glucose-responsive glucagon delivery system designed to address the unique pathophysiology of severe PBH, including large glycemic excursions, impaired counterregulation, and hypoglycemia unawareness, would be effective to reduce the occurrence and severity of hypoglycemia. In a pilot study, we demonstrated the feasibility of an event-based hypoglycemia prediction algorithm guided by CGM data to direct manual delivery of small glucagon boluses via a patch infusion pump (4). We now report results of a double-blind, placebo-controlled, crossover trial to determine the efficacy of a closed-loop glucagon delivery system to reduce severe hypoglycemia after a mixed meal.
Methods
Materials and methods—clinical
Study design
This double-blind, placebo-controlled crossover trial was designed to evaluate the ability of a novel integrated closed-loop glucagon system to reduce meal-provoked hypoglycemia in PBH (17).
Participants
Participants with a history of RYGB and PBH with neuroglycopenia, uncontrolled on medical nutrition therapy and medications, were recruited from the Joslin Diabetes Center hypoglycemia clinic and other endocrine clinics in the region. Exclusion criteria included major systemic illness, cardiac arrhythmia, hypertension, active coronary artery disease, fasting hypoglycemia, known insulinoma, pregnancy, substance or alcohol abuse, recent steroid or investigational drug exposure, and use of medications (beyond hypoglycemia treatment) known to affect insulin secretion or action. The Food and Drug Administration approved investigational device exemption (IDE G170159) for the closed-loop glucagon studies. The Joslin Diabetes Center Committee on Human Studies approved the study. Written informed consent was obtained from all participants.
Investigational glucagon and vehicle formulation, randomization, and masking
Investigational ready-to-use liquid glucagon and matching vehicle (Xeris Pharmaceuticals, Chicago, IL) were provided in numbered vials and stored at room temperature. Vehicle was visually indistinguishable from active product. Sets of 4 vials, 1 primary and backup vial for each study visit, were grouped in a box for each study participant and labeled with participant number and study visit number. A list of ascending randomization numbers starting with 001 was generated by a Xeris statistician using SAS version 9.4, Proc Procedure set for a block size of 8. The randomization list was provided to the packager so that boxes would be packaged with a 1:1 distribution of active and placebo vehicle vials at each visit. After eligibility was confirmed, qualifying subjects were assigned the lowest number kit remaining in inventory at the time of their initial meal challenge. Participants, study personnel, and Xeris study staff were blinded to study drug identity, and identity of vials was unblinded only after data analysis.
At release for use in the study, samples from glucagon vials were analyzed to determine glucagon concentration using ultra-high-performance liquid chromatography (Pyramid Labs, Costa Mesa, CA). Additional samples were placed in an International Council for Harmonization-compliant stability program for concentration analysis during and following completion of the clinical study (ultra-high-performance liquid chromatography, MRI Global, Kansas City, MO). The stability results at the timepoint following completion of the clinical trial showed that glucagon concentrations remained well within acceptance criteria.
Initiation of glucagon delivery system and mixed-meal tolerance testing
Following screening history, physical examination and laboratory testing, and confirmation of participant eligibility, 2 blinded continuous glucose monitors (CGMs, Dexcom G4 Professional, San Diego, CA) were inserted on the anterior abdominal wall. Participants were instructed to perform calibrations when prompted and to return 48 to 72 hours later for the first of 2 study visits. Medications, including acarbose, short-acting octreotide, and diazoxide, were held for at least 24 hours before study visits; long-acting octreotide was withdrawn at least 1 month before study entry.
Participants arrived after an overnight fast. After IV catheter placement for blood sampling, an Omnipod pump (Insulet Corporation, Acton, MA) filled with either investigational glucagon or vehicle, was placed on the anterior abdominal wall, and the pump catheter was inserted into the subcutaneous space. After calibration, the sensor with glucose values most closely matching the serum glucose was connected to the Windows tablet running the portable Artificial Pancreas System (pAPS) and the PBH detection algorithm (4,18). After baseline blood sampling, a liquid mixed meal (Ensure Compact, containing 64 g carbohydrate, 18 g protein, 12 g fat, 440 kcal, 236 mL volume) was consumed over 3 to 5 minutes. As previously described (4), this high-carbohydrate meal was chosen as a standardized experimental intervention to increase the likelihood of induction of the postprandial glucose and insulin surge and subsequent hypoglycemia typical of PBH. Use of this protocol allowed testing of the efficacy of glucagon vs vehicle to mitigate hypoglycemia.
Sensor glucose and plasma glucose, insulin, C-peptide, and glucagon concentrations were measured at baseline and at predetermined intervals following the mixed meal, at time of hypoglycemia alert, and for at least 2 hours following study drug delivery.
The pAPS system
The pAPS is a drug delivery system including a CGM, an Omnipod pump, and a tablet computer (Dell Venue 10 Pro, Dell Technologies, Round Rock, TX) that centrally manages all the devices (Fig. 1). The user (study physician or engineer) can interact with the system (e.g., set subject parameters, initiate or terminate the session) through a touch graphical user interface implemented on the tablet. The pAPS is responsible for automated delivery of study drug via the pump; this is accomplished by the hypoglycemia detection and mitigation algorithm that resides on the tablet. More specifically, the CGM sensor transmits a glucose value every 5 minutes to the CGM receiver, which is serially connected to the tablet. The transmitted glucose value is used by the algorithm to calculate the dose of study drug, which is then signaled from the tablet to the pump relay via Bluetooth Low Energy communication for pump delivery. The pAPS also provides visual and audible alerts in the tablet and sends wireless alerts to the clinical team in the event of impending hypoglycemia.
Figure 1.
Components of the closed-loop glucagon system. (A) A continuous glucose sensor detects interstitial glucose values, wirelessly transmits data to the continuous glucose monitor receiver (B), which transmits to the pAPS tablet hosting (C) the postbariatric hypoglycemia algorithm. When hypoglycemia is predicted, a command for study drug delivery is sent to an infusion pump containing either (D) study glucagon or vehicle. pAPS, pAPS, portable Artificial Pancreas System.
Hypoglycemia detection and mitigation algorithm
The pAPS system has reliably supported many inpatient and outpatient artificial pancreas clinical studies (19–23) using distinct internal algorithms designed for diabetes management. Given the unique postmeal glycemic patterns in PBH, a novel hypoglycemia detection and mitigation algorithm was developed and tested in our previous open-loop pilot study, described in Laguna Sanz, et al. (4). The algorithm has been further optimized to handle missing CGM values in case of lost connectivity or CGM malfunction. The key components of the algorithm, now utilized in a closed-loop mode (requiring no intervention from the clinical team), are detailed in Supplemental Methods (17).
In brief, the algorithm includes 2 distinct components, designed to permit recognition and response to the unique postprandial glucose metabolism patterns in PBH. For both components, a threshold for drug delivery was chosen that would account for the delay of CGM-derived interstitial glucose values relative to plasma glucose (24) and allow delivery of glucagon before the onset of hypoglycemia. The PBH alert, designed to predict postprandial hypoglycemia, is activated if glucose is falling at a rapid rate after a meal has been detected and hypoglycemia is predicted to occur in <30 minutes; this alert can be activated even if glucose levels are within the normoglycemic range (<150 mg/dL) to allow early detection and response. By contrast, the low glucose prediction component can only be activated at a lower glucose and does not require detection of a meal. If the prespecified conditions were reached, the pAPS system directed the pump to deliver 300 μg of the blinded study drug (glucagon or equivalent volume of vehicle) within 5 minutes. If the system predicted or detected a second hypoglycemic event, 300 μg or 150 μg of blinded study drug were delivered based on filtered glucose levels being below or above 75 mg/dL, respectively. The study team received alerts when a hypoglycemia event was detected or predicted. Though the participants were not directly informed, they could usually feel a sensation of study drug delivery and were aware when the team drew blood samples and began asking about symptoms using the Edinburgh Hypoglycemia Scale.
For all participants, the pump was removed and a standard low-carbohydrate lunch was provided either at 120 minutes following the first hypoglycemia mitigation alert or 30 minutes after a second alert, whichever was later. If at any time during observation, the participant developed severe hypoglycemia (serum glucose ≤55 mg/dL or neuroglycopenia), rescue therapy was provided (IV dextrose bolus, subcutaneous glucagon, or oral glucose) to achieve resolution of neuroglycopenia. The pump was removed and lunch was provided once glucose was >100 mg/dL. All participants were observed for 2 additional hours before discharge.
Outcomes and sample size
The primary endpoint for this study is prevention of meal-provoked hypoglycemia, defined as glucose <65 mg/dL, with the glucagon-containing system but not the vehicle-containing system. Secondary outcomes included prevention of severe hypoglycemia (<60 mg/dL), prevention of rebound hyperglycemia (glucose >180 mg/dL), avoidance of clinically significant hypoglycemia requiring rescue therapy (glucose ≤55 mg/dL and/or severe neuroglycopenia), and improvements in glucose time in goal (65 to 180 mg/dL, reported in minutes). Sample size of 12 subjects was chosen before study initiation to provide power >0.9 to detect significant differences between conditions in the primary outcome at α = 0.05. When 10 participants had completed the trial, a noncomparative (masked) interim analysis was undertaken to determine whether the total number of outcomes thus far provided adequate power to justify stopping enrollment. Discordant responses to treatment between conditions in at least 8 subjects were required to stop enrollment. Because 6 subjects had discordant responses, the trial was continued until the planned enrollment target of 12. Enrollment was halted when the prespecified target sample size was reached.
Additional clinical assessments
The Edinburgh Hypoglycemia Scale was used to assess hypoglycemia symptoms at baseline, at time of hypoglycemia prediction alert, and 15, 30, and 60 minutes after hypoglycemia alert and blinded study drug bolus (25). Scores for the 5 autonomic, 8 neuroglycopenic, and 5 nonspecific symptoms were summed for each time point.
Pain or other symptoms at the infusion site, using a score from 1 (none) to 10 (greatest) were recorded, and the infusion site was examined at 30 and 120 minutes after pump removal using the Draize scale (26).
Hormonal analyses
Plasma glucose was measured by glucose oxidation (YSI 2300 STAT, Yellow Springs, OH). Using solid-phase extraction, a validated HPLC method with tandem mass spectometry detection was used to measure plasma glucagon (Algorithme Pharma, Laval, Quebec). C-peptide was determined using a sandwich electrochemiluminescence immunoassay (Roche Diagnostics) using a biotinylated monoclonal C-peptide–specific antibody and a monoclonal C-peptide–specific labeled antibody (Medpace, Cincinnati, OH). Insulin was assayed using electrochemiluminescence immunoassay insulin reagent kits and an immuno-analyzer instrument (Roche Diagnostics), which uses a competitive test principle using an anti-insulin monoclonal antibody (Medpace).
Statistical analysis
Descriptive statistics and figures were produced using GraphPad Prism (GraphPad Software, La Jolla, CA). All other analyses were completed using SAS, version 9.4, for Windows (SAS Institute, Cary, NC). All data were tested to confirm that they met the assumptions of the planned analyses. Continuous variables are expressed as mean ± SD or as median with interquartile range. Dichotomous and multinomial outcomes between conditions were compared using McNemar’s test, and continuous outcomes using linear mixed effects models to account for correlation within subjects over time.
Results
Between October 2017 and August 2018, 23 individuals were screened and 18 enrolled (consort diagram, Fig. 2). Two participants withdrew before the first mixed-meal study visit because of inability to obtain adequate intravenous access. Sixteen participants completed the first mixed-meal study visit; 12 participants completed both mixed meal study visits and were included in analysis (Table 1). Of the 4 participants who completed only 1 mixed-meal visit, 1 developed intractable nausea and emesis after consuming the mixed meal, and a hypoglycemia alert was not triggered in 3 (and thus no study drug delivery).
Figure 2.
Consort diagram. *Exclusion criteria identified during screening included no IV access (n = 2), screening laboratory abnormalities (hypokalemia, n = 1), no prior trial of medications for treatment of postbariatric hypoglycemia (n = 1), and uncontrolled hypertension (n = 1). **Participants were withdrawn from study because of intractable nausea and emesis after MMTT (n = 1), no hypoglycemia alert or study drug delivery during observation period (n = 3). MMTT, mixed-meal tolerance test.
Table 1.
Participant Characteristics: Demographic Information for 12 Enrolled Participants Who Completed 2 Mixed-Meal Study Visits and Were Included in the Analysis
| Mean | SEM | |||
|---|---|---|---|---|
| Summary | (Median) | (25th-75th percentile) | ||
| Age (y) | 52 | ± | 2.4 | |
| Gender (male:female) | 1:11 | |||
| Body mass index (kg/m2) | 27.6 | ± | 1.5 | |
| Hemoglobin A1c (%) | 5.3 | ± | 0.1 | |
| Years since surgery | 8.4 | ± | 1.5 | |
| Years from surgery to neuroglycopenia | (2.2) | (1.4–6.1) | ||
| On any antihypoglycemic medications | 10 of 12 | |||
| Acarbose | 5 of 12 | |||
| Diazoxide | 4 of 12 | |||
| Octreotide | 3 of 12 | |||
| Pramlintide | 2 of 12 |
The prespecified primary outcome was to test whether the system containing glucagon would be superior to vehicle in preventing hypoglycemia after the mixed meal (moderate hypoglycemia, <65 mg/dL). Avoidance of plasma glucose <65 mg/dL occurred in 4 of 12 participants in response to glucagon but not vehicle study days (P = .18). Similarly, avoidance of sensor glucose <65 mg/dL was successful in response to glucagon but not vehicle study day in 5 of 12 participants (P = .10) (Table 2).
Table 2.
Dichotomous Outcomes: Glucose Values Measured During the Time Defined From the Initial Hypoglycemia Mitigation Alert Until Lunch or Delivery of Rescue Glucose
| Outcome | Source | Met Criterion on Vehicle Only | Met Criterion on Glucagon Only | P Value (G vs V) |
|---|---|---|---|---|
| Glucose <65 mg/dL | Plasma | 4 | 1 | .180 |
| Sensor | 5 | 1 | .103 | |
| Glucose <60 mg/dL | Plasma | 3 | 0 | .083 |
| Sensor | 3 | 1 | .317 | |
| Glucose <55 mg/dL | Plasma | 5 | 0 | .025 |
| Sensor | 0 | 0 | N/A | |
| Rebound hyperglycemia after study drug | Plasma or sensor | 0 | 0 | N/A |
| Rescue needed | 7 | 0 | .008 |
Rebound hyperglycemia defined as plasma glucose >180 mg/dL after study drug delivery. Protocol-specified rescue delivery of IV glucose was performed if plasma glucose <55 mg/dL and/or significant neuroglycopenia developed.
Abbreviations: G, glucagon; N/A, not applicable; V, vehicle.
Although statistical significance was not achieved for the primary endpoint, several clinically meaningful secondary endpoints demonstrating reduction in severe hypoglycemia were achieved. There was a significant difference in avoidance of severe hypoglycemia or neuroglycopenia; 5 participants had plasma glucose <55 mg/dL after vehicle, but plasma glucose levels remained >55 mg/dL after glucagon delivery (P = .03). No participants reached sensor glucose <55 mg/dL after either treatment. Clinically significant hypoglycemia, defined as plasma glucose <55 mg/dL (n = 5) or neuroglycopenia (n = 2) requiring rescue, occurred after vehicle but not glucagon delivery in 7 of 12 participants (P = .008). One participant required rescue after both glucagon and vehicle; importantly, no participants required rescue after only glucagon treatment. Rebound hyperglycemia (glucose >180 mg/dL) did not occur in any study visits.
Nadir plasma glucose was significantly lower after vehicle versus glucagon (58 vs 67 mg/dL, P = .049) (Table 3). Plasma glucose remained in target range (70–180 mg/dL) for a greater percentage of time after glucagon versus vehicle (85 vs 64%, P < .05). A second hypoglycemia mitigation alert was triggered more often during vehicle than glucagon visits (11 of 12 vs 8 of 12 visits), reflecting higher glucose levels after the first glucagon delivery vs vehicle (86 vs 70 mg/dL, P < .01) (Fig. 3). The mean glucose values for participants not requiring rescue glucose was significantly higher at 60 minutes after glucagon compared with vehicle (99 ± 3.5, n = 12 vs 72 ± 5.0, n = 8, P < .001).
Table 3.
Continuous Outcomes: Outcomes Are Based on Glucose Values (Sensor or Capillary, as Indicated) and Time of Key Events During Study Visits
| Outcome | Treatment Condition | Mean | SEM | P Value (G vs. V) |
|---|---|---|---|---|
| % Time plasma glucose in rangea after drug delivery | G | 0.852 | 0.082 | .049 |
| V | 0.645 | 0.074 | ||
| % Time sensor glucose in rangea after drug delivery | G | 0.987 | 0.013 | .056 |
| V | 0.815 | 0.077 | ||
| Nadir plasma glucose (mg/dL) | G | 67.4 | 2.70 | .004 |
| V | 58.5 | 1.87 | ||
| Nadir sensor glucose (mg/dL) | G | 72.7 | 2.21 | .059 |
| V | 65.3 | 1.85 | ||
| Time to nadir plasma glucose after mixed-meal (min) | G | 138 | 11.9 | .195 |
| V | 125 | 7.43 | ||
| Time to nadir sensor glucose after mixed-meal (min) | G | 156 | 11.6 | .043 |
| V | 134 | 6.04 | ||
| Time to alarm (min) | G | 89.9 | 3.90 | .659 |
| P | 87.7 | 4.9975 | ||
| Time to delivery (min) | G | 94 | 5.52 | .406 |
| V | 89.3 | 5.33 | ||
| Alarm 1 sensor glucose (mg/dL) | G | 134 | 5.93 | .611 |
| V | 139 | 4.92 | ||
| Alarm 1 capillary glucose (mg/dL) | G | 98.1 | 7.30 | .183 |
| V | 109 | 5.75 | ||
| Alarm 2 sensor glucose (mg/dL) | G | 85.7 | 4.96 | .009 |
| V | 70.1 | 2.26 | ||
| Alarm 2 capillary glucose (mg/dL) | G | 94.2 | 1.70 | .789 |
| V | 91.7 | 8.62 | ||
| Study drug pain scoreb first dose | G | 4.00 | 0.685 | .865 |
| V | 3.83 | 0.815 | ||
| Study drug pain scoreb second dose | G | 1.14 | 0.615 | .875 |
| V | 1.31 | 0.671 |
Abbreviations: G, glucagon; V, vehicle.
aRange is 65–180 mg/dL.
bPain score: 1 (least severe pain) to 10 (most severe).
Figure 3.
(A) Representative plot of sensor glucose levels for 1 participant during mixed meal tolerance test study visits during which either vehicle (blue) or glucagon (red) was delivered. During the vehicle delivery visit, the participant received 2 doses of vehicle before reaching clinically significant hypoglycemia, at which time IV dextrose was administered and the closed-loop system was discontinued. Sensor glucose values were not plotted after IV dextrose administration because of rapid change in glucose leading to signal loss. During the glucagon treatment visit, the participant received 2 doses of glucagon, and clinically significant hypoglycemia was prevented. (B) Glucose response in the first 60 minutes after the initial vehicle (blue) or glucagon (red) infusions for all study participants. Black triangles denote administration of rescue glucose and premature discontinuation of the system after clinically significant hypoglycemia (defined by plasma glucose) in 7 of 24 visits; 1 participant required rescue glucose after glucagon (red line ending in black triangle).
During both glucagon and vehicle treatment visits, all participants received either 300 μg of glucagon or the equivalent volume of vehicle with the first alarm. During vehicle visits, 9 of 12 participants had second alarms because of persistent glucose values below threshold; 5 received a volume of vehicle equivalent to 300 μg and 4 received a volume of vehicle equivalent to 150 μg. For the 3 remaining participants, second alarms were generated for 2, but study drug was not delivered as the pump had been removed upon rescue glucose treatment for clinically significant hypoglycemia. By contrast, during glucagon treatment visits, 6 of 12 participants did not have a second alarm because higher glucose levels; of the 6 with a second alarm, 5 received 150 μg of glucagon, and 1 received 300 μg of glucagon.
Plasma glucagon levels were similar at baseline, and increased moderately at early time points after the mixed meal, as previously reported (27); postmeal glucagon levels did not differ significantly between study visit days (Fig. 4A) (17). At the time of the first hypoglycemia mitigation alert, glucagon levels were not increased from baseline or different between treatment groups. However, 10 minutes after study drug infusion, significant differences emerged, with serum glucagon levels 417 ± 91 vs. 14 ± 1 pg/mL (glucagon vs vehicle, P < .001). Differences between glucagon and vehicle study days were maximal at 30 minutes after study drug infusion (1231 ± 187 vs 19 ± 3 pg/mL, P < 0.001), representing a 35-fold increase from baseline in the glucagon treatment group (P < .0001).
Figure 4.
(A) Glucagon, (B) insulin, and (C) C-peptide levels. Hormone levels measured at baseline, and at specified times after mixed meal ingestion, at the time of first and second hypoglycemia mitigation alerts and at specified intervals after alert and study drug infusions. Numbers in gray denote number of participants that had samples drawn at each timepoint. These numbers varied, depending on the timing of alerts; for example, if an alert and study drug delivery occurred before 120 minutes after the meal, additional blood sampling was performed as a posthypoglycemia alert sample. *P < .05, **P < .001. MMTT, mixed-meal tolerance test.
Insulin and C-peptide levels were also assessed at baseline and at several intervals following the mixed meal and study drug infusion (Fig. 4B-C) (17). Consistent with prior studies, plasma insulin levels rose dramatically after meal ingestion on both glucagon and vehicle study days, peaking at 30 minutes (327 vs 243 μU/mL, P < .0001 for both). Insulin levels remained elevated (21- and 36-fold above baseline) at the time of hypoglycemia alert, but did not differ between glucagon and vehicle study days. C-peptide levels demonstrated a similar trend for glucagon vs vehicle study days.
After study drug delivery, plasma insulin levels continued to fall, with no significant difference between glucagon and vehicle delivery days at any time point. C-peptide levels were significantly higher at 30 minutes after glucagon delivery (12.2 ± 3.1 vs 6.1 ± 0.9 ng/mL, P = .03), but did not differ at any other time points.
Hypoglycemia symptoms, assessed with the Edinburgh Hypoglycemia Scale, were recorded and analyzed both individually and by category (autonomic, neuroglycopenic, or nonspecific symptoms); these did not differ between study days at baseline, at the time of predictive hypoglycemia alert, or after drug delivery even in participants who had hypoglycemia.
No serious adverse events were observed during this trial. Nausea was experienced after the mixed meal but before study drug delivery in 12 of 24 study visits in the participants who completed 2 study visits. Nausea occurred only after glucagon delivery in 2 of 12 glucagon study visits and none of the vehicle study visits. Discomfort at the infusion pump site did not differ between vehicle and glucagon infusions (17). Mild erythema and edema were self-limited and did not differ between glucagon and vehicle infusions (17).
Discussion
This study tested a new closed-loop glucose-responsive automated glucagon delivery system, which demonstrated superior performance as compared with placebo (vehicle) delivery in prevention of significant postprandial hypoglycemia in patients with PBH, without any serious adverse events. Our CGM-informed hypoglycemia mitigation algorithm, designed to detect and respond to the unique rapid and high-magnitude swings in postprandial glucose in PBH, successfully activated the pump to deliver study drug in all participants who developed clinically significant hypoglycemia. Glucagon levels rose in response to pump activation, indicating successful delivery, and posttreatment glucose levels were significantly higher in response to glucagon. Finally, clinically significant hypoglycemia requiring IV glucose rescue (neuroglycopenia or plasma glucose <55 mg/dL) occurred only after vehicle infusion, except in the 1 participant who did not respond to either vehicle or glucagon infusions. Thus, glucagon delivery via the glucose-responsive automated delivery system provided clinical efficacy in mitigating severe postprandial hypoglycemia in PBH.
Analysis of glucagon levels during the study revealed several key points. First, glucagon levels did not increase above baseline even as glucose levels approached 55 mg/dL. This indicates that the glucagon counterregulatory response to hypoglycemia is impaired in PBH and parallels findings of Abrahamsson et al., who demonstrated blunted counterregulatory response to hypoglycemia at 6 months post-RYGB, even in individuals without known hypoglycemia (8). Thus, glucagon therapy targets a key element of the dysfunctional hormonal response to hypoglycemia in individuals with PBH.
There was substantial interindividual variation in glycemic responses to mixed meal. Indeed, several participants did not develop adequate hypoglycemic risk to trigger an alert or study drug dosing, despite a history of these events in the outpatient setting. Potential contributors to this unexpected finding include the study environment, which differs substantially from the ambulatory environment in several respects. Participants consumed a liquid mixed meal high in simple carbohydrates, a meal typically avoided because of propensity to induce dumping syndrome and hypoglycemia (28). Moreover, inactivity (remaining in bed during the study) and the stress of participation may have increased stress-related hormonal responses (e.g., cortisol) and induced relative insulin resistance, reducing the likelihood of hypoglycemia.
Although the standard emergency rescue dose of glucagon is 1 mg, we used lower doses in this system (300 μg for the first dose, 300 or 150 μg for the second dose, if needed). Similar submaximal doses have been found to be effective to treat insulin- and exercise-induced hypoglycemia in patients with type 1 diabetes, while minimizing adverse effects such as nausea and rebound hyperglycemia (29–31). For example, in studies by Haymond et al. in a pediatric population, age-based microdosing of glucagon (20–150 μg) successfully increased glucose by 3.3 to 5 mmol/L (60–90 mg/dL)(30). Higher doses of glucagon were required following insulin administration, and 52% of participants required 2 or 3 doses to maintain euglycemia. In this study, we used a 300-μg dose for the first delivery, based on a more robust rise in glucose with this dose (as compared with a 150-μg dose) during the pilot open-loop study (4). We observed significantly higher glucose values at 60 minutes after glucagon dosing as compared with vehicle, with minimal side effects. However, there was substantial interindividual variability in glycemic response, and the magnitude of response to glucagon was generally lower than in studies of T1D. Potential factors contributing to these patterns include the very high insulin levels in the postprandial state in PBH; high-magnitude and prolonged insulin action likely contributed to both ongoing tissue uptake of glucose for several hours after meal ingestion and reduced glucagon responsiveness. Moreover, suboptimal glycogen stores could also contribute to reduced glucagon responsiveness, as observed in patients with type 1 diabetes consuming low carbohydrate diets who have inadequate response to glucagon from insulin-induced hypoglycemia (32). Although we recommend a diet that includes complex carbohydrates in controlled portions, some patients restrict total carbohydrates to reduce hypoglycemia frequency (28), potentially contributing to inadequacy of glycogen stores and reduced glucagon responsiveness. Indeed, 1 participant did not avoid clinically significant hypoglycemia during either vehicle or glucagon delivery, despite achieving supraphysiologic glucagon levels after glucagon delivery, and similar insulin and C-peptide levels to other participants who did respond to glucagon. We do not understand the precise nature of this variability in response at present. Future studies could include patient-specific dosing optimization.
Given the challenging nature of PBH, additional pharmacologic treatment strategies are being developed. Given the importance of hyperinsulinemia as a contributor to postprandial hypoglycemia in PBH, blocking insulin secretion and/or action may be effective. For example, blockade of the GLP-1 receptor by exendin 9-39 can reduce insulin secretion and hypoglycemia after a glucose tolerance test in short-term studies (33, 34). Inhibition of insulin action using a monoclonal anti-insulin receptor antibody has been shown to reduce insulin action in healthy individuals (35), but results of a clinical trial in individuals with PBH (NCT02772718) have not yet been published. A study testing efficacy of pump delivery of microdoses of glucagon throughout the day to maintain glycemia also has not yet been published (NCT 02966275). Additional studies will be needed to assess the relative efficacy of such approaches, either in isolation or in combination with use of our integrated delivery system to detect and prevent severe hypoglycemia.
We acknowledge limitations of our study. Our study was performed in a single center. Only 1 male completed both study visits (despite enrolling additional men). This may reflect, in part, the higher number of women undergoing bariatric surgery (36) and the increased risk for PBH in females (37). Future multicenter studies could aim for improved gender parity. We used liquid mixed meal testing to reproducibly provoke hypoglycemia to test our system, but acknowledge that a liquid meal is not physiologic and may provoke some features of the dumping syndrome. Frequency of provoked hypoglycemia would likely be lessened with use of a solid meal, as we recommend to our patients in clinical practice. We used glycemic patterns detected by CGM to predict hypoglycemia, as a closed-loop glucagon rescue system would be most valuable for patients with hypoglycemia unawareness. Although use of CGM in postbariatric surgery patients is a relatively new strategy, comparison of CGM with plasma glucose-based detection of hypoglycemia has been reported in 2 studies (38, 39). Kefurt et al. reported that 75% of post-RYGB patients had glucose <55 mg/dL during blinded CGM vs only 29% with plasma glucose <55 mg/dL during meal testing. These differences could reflect high rates of asymptomatic reductions in glucose in this population, erroneous detection of low glucose by CGM, or higher rates of low glucose levels during daily life compared with postmeal testing (38). Interestingly, Nielsen et al. found that minimum interstitial glucose values were an average of 20 mg/dL higher than plasma values measured concurrently during mixed meal, potentially leading to overestimation of glucose values and underestimation of hypoglycemia by CGM (39). Future studies will be required to fully assess the role of CGM and other methods to detect and define hypoglycemia in this population (38, 39).
Although the study was conducted in a clinical research center to ensure participant safety, outpatient “real-world” experience indicates that activity and ambulation may also increase risk of hypoglycemia. Additional studies will be required to test the efficacy of the closed-loop rescue system to prevent hypoglycemia in response to solid meals, activity, and other aspects of outpatient daily life. Our system delivered a maximum of 2 doses of study drug; future studies will be required to test dose dependency of the glucagon response and whether additional doses could promote improved efficacy. Future studies could investigate a more personalized dose based on weight or glucagon responsiveness. In summary, we report development and implementation of a novel technology-driven approach to address the unique pathophysiology of PBH, and demonstrate that our closed-loop automated glucagon system was effective in prevention of severe postprandial hypoglycemia in participants with PBH.
Acknowledgments
We gratefully acknowledge study volunteers and staff of the Joslin Clinical Research Center for their diligent support, including Jan Willem Middlebeek, MD; Charlene Coneys, RN; Casey Holden, RN; and Jesse Walker. We thank Sunil Deshpande, PhD, for support during study visits and Howard Zisser, MD, for helpful discussion.
Financial Support: We gratefully acknowledge study funding from NIDDK R44DK10711 (SBIR) and supplemental financial study support from Xeris Pharmaceuticals. C.M.M. was supported by the Hearst Foundation.
Glossary
Abbreviations
- CGM
continuous glucose monitor
- pAPS
portable Artificial Pancreas System
- PBH
postbariatric hypoglycemia
- RYGB
Roux-en-Y gastric bypass
Additional Information
Disclosure Summary: M.J.C., B.N., and S.J.P. are employees of Xeris Pharmaceuticals. M.E.P. reports investigator-initiated grants from Janssen Pharmaceuticals, Medimmune, and Dexcom, and personal fees from Eiger Pharmaceuticals, Avolynt, and Fractyl. E.D. and F.J.D. report an issued patent “Health Monitoring System.” E.D., A.J.L.S., and M.E.P. report a patent application “Prevention of Post-Bariatric Hypoglycemia Using a Novel Glucose Prediction Algorithm and Mini-Dose Stable Glucagon.” M.E.P. reports an issued patent PCT/US2017/045061 “Methods and Compositions for Treating Hypoglycemia.” M.J.C., B.N., and S.J.P. acknowledge pending patent PCT/US2019/014815. F.J.D. reports equity, licensed intellectual property, and is a member of the Scientific Advisory Board of Mode AGC. I.A., C.M.M., D.P., S.Z., L.R., and P.W. have nothing to disclose.
Data Availability: The datasets generated during and/or analyzed during the current study are not publicly available but are available from the corresponding author on reasonable request.
References
- 1. Salehi M, Vella A, McLaughlin T, Patti ME. Hypoglycemia after gastric bypass surgery: current concepts and controversies. J Clin Endocrinol Metab. 2018;103(8):2815–2826 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Lee C, Brown T, Schweitzer M, Magnuson T, Clark J. 389-P: comparison of hormonal response to a mixed-meal challenge in individuals with hypoglycemia after sleeve gastrectomy vs. gastric bypass. Diabetes. 2019;68(Supplement 1):389-P. [Google Scholar]
- 3. Capristo E, Panunzi S, De Gaetano A, et al. Incidence of hypoglycemia after gastric bypass vs sleeve gastrectomy: a randomized trial. J Clin Endocrinol Metab. 2018;103(6):2136–2146. [DOI] [PubMed] [Google Scholar]
- 4. Laguna Sanz AJ, Mulla CM, Fowler. et al. Design and clinical evaluation of a novel low-glucose prediction algorithm with mini-dose stable glucagon delivery in post-bariatric hypoglycemia. Diabetes Technol Ther. 2018;20(2):127–139. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Salehi M, Woods SC, D’Alessio DA. Gastric bypass alters both glucose-dependent and glucose-independent regulation of islet hormone secretion. Obesity (Silver Spring). 2015;23(10):2046–2052. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. le Roux CW, Welbourn R, Werling M, et al. Gut hormones as mediators of appetite and weight loss after Roux-en-Y gastric bypass. Ann Surg. 2007;246(5):780–785. [DOI] [PubMed] [Google Scholar]
- 7. Mulla CM, Goldfine AB, Dreyfuss JM, et al. Plasma FGF-19 levels are increased in patients with post-bariatric hypoglycemia. Obesity Surgery 2019;29( 7):2092–2099. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Abrahamsson N, Börjesson JL, Sundbom M, Wiklund U, Karlsson FA, Eriksson JW. Gastric bypass reduces symptoms and hormonal responses in hypoglycemia. Diabetes. 2016;65(9):2667–2675. [DOI] [PubMed] [Google Scholar]
- 9. Arora I, Patti ME. Can reversal of RYGB also reverse hypoglycemia? Mol Metab. 2018;9:1–3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Svane MS, Toft-Nielsen MB, Kristiansen VB, et al. Nutrient re-routing and altered gut-islet cell crosstalk may explain early relief of severe postprandial hypoglycaemia after reversal of Roux-en-Y gastric bypass. Diabet Med. 2017;34(12):1783–1787. [DOI] [PubMed] [Google Scholar]
- 11. Patti ME, McMahon G, Mun EC, et al. Severe hypoglycaemia post-gastric bypass requiring partial pancreatectomy: evidence for inappropriate insulin secretion and pancreatic islet hyperplasia. Diabetologia. 2005;48(11):2236–2240. [DOI] [PubMed] [Google Scholar]
- 12. Service GJ, Thompson GB, Service FJ, et al. Hyperinsulinemic hypoglycemia with nesidioblastosis after gastric-bypass surgery. N Engl J Med. 2005;353(3):249–254. [DOI] [PubMed] [Google Scholar]
- 13. Patti ME, Goldfine AB, Hu J, et al. Heterogeneity of proliferative markers in pancreatic β-cells of patients with severe hypoglycemia following Roux-en-Y gastric bypass. Acta Diabetol. 2017;54(8):737–747. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Vanderveen KA, Grant CS, Thompson GB, et al. Outcomes and quality of life after partial pancreatectomy for noninsulinoma pancreatogenous hypoglycemia from diffuse islet cell disease. Surgery. 2010;148:1237–1245; discussion 1245–1236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Vukmir RB, Paris PM, Yealy DM. Glucagon: prehospital therapy for hypoglycemia. Ann Emerg Med. 1991;20(4):375–379. [DOI] [PubMed] [Google Scholar]
- 16. Castle JR, Youssef JE, Branigan D, et al. Comparative pharmacokinetic/pharmacodynamic study of liquid stable glucagon versus lyophilized glucagon in type 1 diabetes subjects. J Diabetes Sci Technol. 2016;10(5):1101–1107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Patti ME. A Randomized, Placebo-Controlled Double-Blind Trial of a Closed-Loop Glucagon System for Post-Bariatric Hypoglycemia. Vol 2.Cambridge, MA: Harvard Dataverse; 2019. [Google Scholar]
- 18. Dassau E, Zisser H, C Palerm C, A Buckingham B, Jovanovic L, J Doyle F 3rd. Modular artificial beta-cell system: a prototype for clinical research. J Diabetes Sci Technol. 2008;2(5):863–872. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Dassau E, Brown SA, Basu A, et al. Adjustment of open-loop settings to improve closed-loop results in type 1 diabetes: a multicenter randomized trial. J Clin Endocrinol Metab. 2015;100(10):3878–3886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Gondhalekar R, Dassau E, Doyle FJ 3rd. Periodic zone-MPC with asymmetric costs for outpatient-ready safety of an artificial pancreas to treat type 1 diabetes. Automatica (Oxf). 2016;71:237–246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Pinsker JE, Lee JB, Dassau E, et al. Randomized crossover comparison of personalized MPC and PID control algorithms for the artificial pancreas. Diabetes Care. 2016;39(7):1135–1142. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Buckingham BA, Christiansen MP, Forlenza GP, et al. Performance of the omnipod personalized model predictive control algorithm with meal bolus challenges in adults with type 1 diabetes. Diabetes Technol Ther. 2018;20(9):585–595. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Pinsker JE, Laguna Sanz AJ, Lee JB, et al. Evaluation of an artificial pancreas with enhanced model predictive control and a glucose prediction trust index with unannounced exercise. Diabetes Technol Ther. 2018;20(7):455–464. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Schmelzeisen-Redeker G, Schoemaker M, Kirchsteiger H, Freckmann G, Heinemann L, Del Re L. Time delay of CGM sensors: relevance, causes, and countermeasures. J Diabetes Sci Technol. 2015;9(5):1006–1015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Hepburn DA, Deary IJ, Frier BM, Patrick AW, Quinn JD, Fisher BM. Symptoms of acute insulin-induced hypoglycemia in humans with and without IDDM. Factor-analysis approach. Diabetes Care. 1991;14(11):949–957. [DOI] [PubMed] [Google Scholar]
- 26. Draize JH, Woodard G, Calvery HO. Methods for the study of irritation and toxicity of substances applied topically to the skin and mucous membranes. 1944;82:377–390 [Google Scholar]
- 27. Goldfine AB, Mun EC, Devine E, et al. Patients with neuroglycopenia after gastric bypass surgery have exaggerated incretin and insulin secretory responses to a mixed meal. J Clin Endocrinol Metab. 2007;92(12):4678–4685. [DOI] [PubMed] [Google Scholar]
- 28. Suhl E, Anderson-Haynes SE, Mulla C, Patti ME. Medical nutrition therapy for post-bariatric hypoglycemia: practical insights. Surg Obes Relat Dis. 2017;13(5):888–896. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Rickels MR, DuBose SN, Toschi E, et al. T1D Exchange Mini-Dose Glucagon Exercise Study Group Mini-dose glucagon as a novel approach to prevent exercise-induced hypoglycemia in type 1 diabetes. Diabetes Care. 2018;41(9):1909–1916. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Haymond MW, Schreiner B. Mini-dose glucagon rescue for hypoglycemia in children with type 1 diabetes. Diabetes Care. 2001;24(4):643–645. [DOI] [PubMed] [Google Scholar]
- 31. Haymond MW, Redondo MJ, McKay S, et al. Nonaqueous, mini-dose glucagon for treatment of mild hypoglycemia in adults with type 1 diabetes: a dose-seeking study. Diabetes Care. 2016;39(3):465–468. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Ranjan A, Schmidt S, Damm-Frydenberg C, et al. Low-carbohydrate diet impairs the effect of glucagon in the treatment of insulin-induced mild hypoglycemia: a randomized crossover study. Diabetes Care. 2017;40(1):132–135. [DOI] [PubMed] [Google Scholar]
- 33. Craig CM, Liu LF, Nguyen T, Price C, Bingham J, McLaughlin TL. Efficacy and pharmacokinetics of subcutaneous exendin (9-39) in patients with post-bariatric hypoglycaemia. Diabetes Obes Metab. 2018;20(2):352–361. [DOI] [PubMed] [Google Scholar]
- 34. Craig CM, Liu LF, Deacon CF, Holst JJ, McLaughlin TL. Critical role for GLP-1 in symptomatic post-bariatric hypoglycaemia. Diabetologia. 2017;60(3):531–540. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Johnson KW, Neale A, Gordon A, et al. Attenuation of insulin action by an allosteric insulin receptor antibody in healthy volunteers. J Clin Endocrinol Metab. 2017;102(8):3021–3028. [DOI] [PubMed] [Google Scholar]
- 36. Khorgami Z, Shoar S, Andalib A, Aminian A, Brethauer SA, Schauer PR. Trends in utilization of bariatric surgery, 2010–2014: sleeve gastrectomy dominates. Surg Obes Relat Dis 2017;13:774–778 [DOI] [PubMed] [Google Scholar]
- 37. Lee CJ, Clark JM, Schweitzer M, et al. Prevalence of and risk factors for hypoglycemic symptoms after gastric bypass and sleeve gastrectomy. Obesity (Silver Spring). 2015;23(5):1079–1084. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Kefurt R, Langer FB, Schindler K, Shakeri-Leidenmühler S, Ludvik B, Prager G. Hypoglycemia after Roux-En-Y gastric bypass: detection rates of continuous glucose monitoring (CGM) versus mixed meal test. Surg Obes Relat Dis. 2015;11(3):564–569. [DOI] [PubMed] [Google Scholar]
- 39. Nielsen MS, Christensen BJ, Ritz C, et al. Roux-En-Y gastric bypass and sleeve gastrectomy does not affect food preferences when assessed by an ad libitum buffet meal. Obes Surg. 2017;27(10):2599–2605. [DOI] [PubMed] [Google Scholar]