Abstract
Introduction: Polyethylene wear and subsequent osteolysis remain obstacles to the long-term survivorship of total hip arthroplasty (THA). Highly cross-linked polyethylene (XLPE) with radical quenching represents a massive leap forward with dramatically improved wear rates compared to ultra-high molecular weight polyethylene (UHMWPE). In this study we evaluate the wear of UHMWPE and XLPE coupled with oxidized zirconium (OxZr) femoral heads. Methods: A longitudinal, retrospective analysis was performed identifying consecutive patients who received a 28-mm OxZr-on-polyethylene primary THA from 2003 to 2004 by a single, high-volume arthroplasty surgeon. Patients were divided into two groups: those that received (1) UHMWPE liner and (2) a highly XLPE liner. Patients were included if clinical follow-up was complete to 2014 or later. Radiographic analysis was performed by two blinded observers. Measures included cup position, annual linear wear rate, and presence of osteolysis. Pairwise comparisons, correlations, and inter-rater reliability were calculated. Results: Eighty patients were in the UHMWPE group with an average follow-up of 10 ± 1.23 years and 88 patients in the XLPE group with an average of 10 ± 1.03-year follow-up. Average age (68) was similar between groups (p = 0.288). Observer reliability was excellent for cup abduction (ICC = 0.940), anteversion (ICC = 0.942), and detection of osteolysis (ICC = 0.811). Annual linear wear rates were significantly higher (p = 1 × 10−19) with UHMWPE (0.21 ± 0.12 mm/year) compared to XLPE (0.05 ± 0.03 mm/year). Linear wear rate was significantly correlated to decreasing acetabular abduction (p = 0.035). Osteolysis was noted only in the UHMWPE group, with 17 patients (21.2%) exhibiting acetabular osteolysis and 37 (46.3%) patients exhibiting femoral osteolysis. Conclusions: OxZr coupled with XLPE showed minimal wear and no osteolysis at 10-year follow up. The yearly linear penetration rate is similar to that seen in other studies of XLPE THA. A careful longitudinal follow-up will be required to determine if advanced bearings such as OxZr or ceramic can show improved performance in the second decade of implantation.
Keywords: Oxidized zirconium, Highly cross-linked polyethylene, Total hip arthroplasty, Long-term outcomes, Osteolysis
Introduction
Polyethylene wear and subsequent osteolysis remain the most common reason for revision in the medium to long term following total hip arthroplasty (THA) [1, 2]. Particulate debris generated by polyethylene produces an inflammatory tissue response that leads to retroacetabular and femoral osteolysis, with subsequent loosening of these components [3]. Previous investigations have associated a causal link between wear debris and osteolysis [4] and a polyethylene linear wear rate of 0.1 mm/year is widely quoted to be the threshold for which osteolysis occurs [5]. Efforts to improve polyethylene wear resistance resulted first in the production of ultra-high molecular weight polyethylene (UHMWPE) in 1971 [6], with subsequent refinements occurring throughout the 1990s [7, 8]. The second and most recent major effort to improve polyethylene wear resistance was the production of highly cross-linked polyethylene (XLPE), which was first used clinically in 1998 [9] and was subsequently utilized in over half of all THAs within the United States by 2003 [10]. Several studies have demonstrated that first-generation XLPE is associated with a significantly lower annual wear rate compared to UHMWPE as well as a significantly lower rate of radiographic signs of osteolysis [9, 11–13].
Yet despite promising results with XLPE, concerns have been raised regarding how the material will perform in the long-term, specifically with regard to continued oxidation and delamination [14–16]. Furthermore, cobalt chrome femoral heads in metal-on-polyethylene articulations have been implicated in cases of trunnionosis, an increasingly recognized etiology of hip pain and subsequent revision surgery [17, 18]. Therefore, alternative femoral head bearings have been developed to improve wear characteristics. Oxidized zirconium (OxZr, Oxinium, Smith & Nephew, Memphis, TN) is one such example, combining the surface smoothness of a ceramic component with the internal toughness of cobalt chrome. OxZr has been shown to have better wear properties than cobalt chrome and has not been associated with any reported case of trunnionosis [19, 20]. Furthermore, OxZr has been shown to be equivalent to cobalt chrome in the short- to mid-term [21, 22] with regard to linear wear and peri-implant osteolysis. However, no long-term investigations of OxZR with regard to in-vivo polyethylene wear and incidence of osteolysis exist.
The main objective of the current study is to compare the radiographic wear and incidence of osteolysis in the long-term of a cohort of consecutive THA patients that received OxZr femoral heads articulating with either UHMWPE or HCLPE liners. We hypothesized that OxZr on HCLPE would (1) produce a significantly lower annual wear rate and (2) produce a significantly lower incidence of osteolysis compared to UHMWPE. We also hypothesized that OxZr on UHMWPE would produce a long-term annual wear rate below 0.2 mm/yr, which has been previously reported [23, 24] with cobalt chrome on UHMWPE with the same acetabular component utilized in this study.
Methods
This study received approval from our Institutional Review Board. A retrospective analysis was performed to identify consecutive patients in our institution with minimum 8-year follow-up who received an OxZr-on-polythylene primary THA performed by the senior author (PP) from January 1st, 2003 to December 1st, 2004. The senior author has been utilizing OxZr femoral heads routinely in primary THA since 2003 and the specified follow-up was selected to coincide with the transition from UHMWPE liner to XLPE liner use. Apart from the change in liner composition, all THAs were performed using the posterolateral approach with the same cementless acetabular (Reflection, Smith, & Nephew) and femoral (Synergy, Smith, & Nephew) implants. The most common (>95%) head size utilized at the time was 28 mm and to increase the study’s internal validity, we decided prior to data collection that only THAs performed using a 28-mm head would be included for analysis. Preoperative variables retained for analysis included patient age, gender, and diagnosis. Cup size and head size were recorded from the operative notes. Outpatient notes were reviewed to determine date of most recent follow-up, if revision surgery had been performed surgery and what the indication for revision was. Failure of the bearing surface was defined as the occurrence of revision surgery due to osteolysis or aseptic loosening of the femoral and/or acetabular components.
Radiographic analysis was performed on anteroposterior (AP) pelvis digital X-rays that were taken at the six-week postoperative visit and at the most recent follow-up. X-ray analyses were using the validated Hip Analysis Suite Software (version 8.0.4.5., Martell Hip Analysis Suite™, Chicago, IL) [25–27] and all measures were performed by two experienced observers who were blinded to PE liner type and clinical outcome. The advanced manual mode of the Martell Hip Analysis Software was used to decrease the risk of automated error [28]. Pelvic landmarks, the outline of the femoral head, the center of rotation, and the outline of the acetabular component were manually identified on each AP film. Acetabular component version and inclination were calculated for both initial and final follow-up X-rays, and each AP film’s accompanying shoot-through lateral view was evaluated to assign a positive (anteversion) and negative (retroversion) for version. Annual liner wear rates were calculated through automated comparisons between initial and final follow-up X-rays. Final follow-up X-rays were assessed for evidence of acetabular osteolysis using the DeLee and Charnley method [29].
Statistical analysis
Data were analyzed using SPSS statistical software (SPSS 23, IBM, NY). For all study variables, a Shapiro–Wilk test was performed to determine the presence or absence of a Gaussian distribution. Patient demographics (age, gender, diagnosis, length of follow-up), implant position (version, inclination), annual liner wear rates, and incidence of osteolysis were compared using Student’s t-test, Mann–Whitney U-test, or Fisher’s exact test when indicated. Pearson’s correlational coefficient was calculated to determine if any relationship existed between PE wear rates and acetabular position. Inter-rater reliability for radiographic measurements was determined through calculating the intraclass correlation coefficient using a two-way random-effects model assuming a single measurement and absolute agreement. An intra-class correlational coefficient (ICC) of 1 represents perfect reliability, and any value greater than 0.8 indicates excellent reliability [30]. Kaplan–Meier survival curves were plotted according to revision required due to osteolysis or aseptic loosening. For all statistical comparisons, a p-value < 0.05 was determined as significant.
A post-hoc power analyses was performed to assess the power of the annual wear measurements between the groups. The post hoc power analysis yielded 100% power to detect a clinically meaningful difference in the annual wear of 0.037 mm with a probability of 95%. This clinically meaningful difference was calculated through averaging the mean annual wear measurements of eight recent long-term HCLPE studies [13, 31–37].
Results
The retrospective analysis identified 80 patients in the UHMWPE group with an average follow-up of 10.5 ± 1.23 years and 88 patients in the XLPE group with an average follow-up of 10.3 ± 1.03 years (Table 1). Patients in the XLPE group were significantly younger (58.0 ± 10.4 years; p < 0.01) compared to the UHMWPE group (64.5 ± 8.67 years). No significant differences were noted in gender (p = 0.06) or duration of follow-up (p = 0.373). Five patients (6.25%) with UHMWPE liners required revision specifically for aseptic loosening secondary to osteolysis. A sixth UHMWPE patient was revised for a periprosthetic fracture, with acetabular component revision also being performed for an incidental intraoperative finding of severe osteolysis. Conversely, no patients in the XLPE group underwent revision THA for aseptic loosening or osteolysis.
Table 1.
UHMWPE group (mean + SD) | HCLPE group (mean ± SD) | p-value | |
---|---|---|---|
N | 80 | 88 | – |
Age | 64.5 + 8.67 | 58.0 + 10.40 | 3 × 10−6a |
Gender | 47:34 | 37:49 | 0.06 |
Follow-up (years) | 10.0 + 1.23 | 9.9 + 1.10 | 0.373 |
Inclination | 41.9 + 6.22 | 42.56 + 6.72 | 0.322 |
Ante version | 20.5 + 5.57 | 21.9 + 7.11 | 0.267 |
Annual wear rate (mm/year) | 0.21 ± 0.12 | 0.05 + 0.03 | 1 × 10−19a |
Acetabular osteoly | 17 of 80 | 0 of 88 | 7.4 × 10−7a |
Zone 1 | 2 | – | – |
Zone 2 | 6 | – | – |
Zone 3 | 17 | – | – |
Femoral osteolysis | 37 of 80 | 0 of 88 | 9.4 × 10−15a |
Zone 1 | 21 | – | – |
Zone 6 | 2 | – | – |
Zone7 | 34 | – | – |
Other | 5 | – | – |
UHMWPE: ultra-high molecular weight polyethylene, XLPE: highly cross-linked polyethylene, SD: standard deviation.
Denotes significance of p < 0.01.
With regard to radiographic parameters, UHMWPE liners exhibited a significantly higher annual wear rate (0.21 ± 0.12 mm/year) compared to XLPE liners (0.05 ± 0.03 mm/year, p = 1 × 10−19). These findings were present despite no significant difference existing in implant inclination (p = 0.322) or version (p = 0.267) between groups. A significant, yet weakly positive correlation was found with increasing annual wear and decreasing acetabular inclination (Pearson’s correlational coefficient: 0.165; p = 0.035). Radiographic evidence of osteolysis was noted only in the UHMWPE group (Figure 1). Seventeen (21.2%) UHMWPE patients exhibited acetabular osteolysis at final follow-up, with zone 3 being the most common location (17 of 17) and seven UHMWPE patients demonstrating osteolysis in at least two zones. Thirty-seven (46.3%) UHMWPE patients exhibited femoral osteolysis, with zone 7 being the most common location (24 of 37) and 15 patients demonstrating osteolysis in more than one zone. Inter-observer reliability was excellent for cup abduction (ICC = 0.940), anteversion (ICC = 0.942) and detection of osteolysis (ICC = 0.811).
Discussion
The long-term success of THA is largely dependent on avoiding periprosthetic osteolysis by minimizing polyethylene wear. In this study, the longest follow-up of oxidized zirconium THA to date, we demonstrated that OxZr coupled with XLPE showed minimal wear and no osteolysis at an average of 10 years following implantation. Conversely, use of UHMWPE was associated with significantly higher rates of wear, osteolysis, and revision for aseptic loosening. Our reported linear wear rate for OxZr on XLPE is similar to a smaller long-term study (0.05 mm/year) [22] and mid-term study (0.061 mm/year) [38], and slightly higher than a mid-term randomized control trial (0.01 mm/year) [21] as well as a mid-term study in younger patients (0.022 mm/year) [39]. As in our own cohort, none of these studies reported any cases of revision THA for osteolysis or aseptic loosening in OxZr-on-XLPE patients.
When compared to other bearing surfaces articulating on XLPE, OxZr remains an appealing choice for long-term THA survival. When compared to ceramic heads, OxZr has demonstrated in clinical follow-up to produce either equivalent [38] or two- to four-fold [21, 22] reductions in annual linear wear rates. Furthermore, OxZr has not been associated with higher rates of tribocorrosion compared to ceramic heads, and in one retrieval study no difference was found in macroscopic fretting and corrosion compared to cobalt chrome [40]. Although a recent hip simulator study [41] has raised concerns for increased wear from OxZR heads that can become scratched following recurrent dislocations, this issue was not observed in the current study due to no episodes of instability.
The strengths of our study include long-term follow-up, multiple reliable assessors for radiographic measurements, and consistent internal validity for surgical technique, head size, and cementless component use. We do acknowledge several limitations in the study. Firstly, the annual wear rate comparison of an immediate postoperative film to most recent follow-up imaging does not uniquely represent true wear (removal of polyethylene) and instead if a combination of true wear with the creep or bedding-in that occurs in the first few months due to non-elastic deformation of the polyethyelene [42]. Secondly, we acknowledge that the computer software utilized for wear analysis is not as sensitive as RSA, widely considered to be the gold standard [43]. Nevertheless, the Martell Hip Analysis software has been widely utilized and in a recent comparative study, was found to be less expensive, less cumbersome, and comparably accurate to both RSA and CT-based wear measurement techniques [44]. Thirdly, although the selected time period for case inclusion was decided upon to maximize follow-up time for the XLPE group, we acknowledge that this may introduce selection bias. Finally, the decision to include only 28-mm OxZr heads, while improving internal validity, also introduces selection bias that could mask possible higher rates of osteolysis that could occur with larger head sizes. Although we acknowledge this limitation for conventional UHMWPE, the effect of head size does not appear to be a significant concern with XLPE [45], with Lachiewicz identifying no relationship between head size and the occurrence of osteolytic lesions in long-term follow-up [46].
The present study reaffirms low wear rates when using XLPE liners in primary THA and provides the longest follow-up of OxZr heads to date. Significantly higher rates of linear wear, radiographic osteolysis, and revision for aseptic loosening were noted with the use of non-crosslinked UHMWPE and consequently, we advise that surgeons consider implementing routine surveillance of patients with these liners once they pass ten years postoperatively. The excellent wear performance of OxZr on the first generation of XLPE liners has positive implications for subsequent XLPE generations, yet similar long-term studies will be needed to verify their equivalence or superiority.
Cite this article as: Carli AV, Patel AR, Cross MB, Mayman DJ, Carroll KM, Pellicci PM & Jerabek SA (2020) Long-term performance of oxidized zirconium on conventional and highly cross-linked polyethylene in total hip arthroplasty. SICOT-J 6, 10
Conflicts of interest
MBC has received consultancies from Smith and Nephew outside of this work. PMP has received IP royalties from Smith and Nephew outside of this work. DJM has received consultancies, IP royalties, and research support from Smith and Nephew outside of this work. AVC, ARP, KMC, and SAJ do not have any relevant disclosures to report.
References
- 1.AOANJRR (2013) Australian Orthopaedic Association National Joint Replacement Registry. In: Annual Report. Adelaide, AOA. [Google Scholar]
- 2.Ulrich SD, Seyler TM, Bennett D, Delanois RE, Saleh KJ, Thongtrangan I, Kuskowski M, Cheng EY, Sharkey PF, Parvizi J, Stiehl JB, Mont MA (2008) Total hip arthroplasties: what are the reasons for revision? Int Orthop 32, 597–604. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 17(5), 649–661. [DOI] [PubMed] [Google Scholar]
- 4.Oparaugo PC, Clarke IC, Malchau H, Herberts P (2001) Correlation of wear debris induced osteolysis and revision with volumetric wear-rates of polyethylene: A survey of 8 reports in the literature. Acta Orthop Scand 72(1), 22–28. [DOI] [PubMed] [Google Scholar]
- 5.Sochart DH (1999) Relationship of acetabular wear to osteolysis and loosening in total hip arthroplasty. Clin Orthop Relat Res 363, 135–150. [PubMed] [Google Scholar]
- 6.Oonishi H, Clarke IC, Masuda S, Amino H (2001) Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene. J Arthroplasty 16(8 Suppl 1), 129–133. [DOI] [PubMed] [Google Scholar]
- 7.Oonishi H, Kuno M, Tsuji E, Fujsiawa A (1997) The optimum dose of gamma radiation-heavy doses to low wear polyethylene in total hip prostheses. Mater Sci Mater Med 8, 11–18. [DOI] [PubMed] [Google Scholar]
- 8.Sun DC, Schmidig G, Yau SS, et al. (1997) Correlation between oxidation, cross linking, and wear performance of UHMWPE. 43rd Annual Meeting, Orthopaedic Research Society. San Francisco, CA, p. 783. [Google Scholar]
- 9.Kurtz SM, Gawel HA, Patel JD (2011) History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res 469, 2262–2277. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Kurtz SM (2004) Compendium of highly crosslinked and thermal treated UHMWPEs, in The UHMWPE Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint, 2nd ed. Kurtz SM, Editor. Elsevier Academic Press: San Diego, CA: p. 354. [Google Scholar]
- 11.Callary SA, Solomon LB, Holubowycz OT, Campbell DG, Munn Z, Howie DW (2015) Wear of highly crosslinked polyethylene acetabular components. Acta Orthop 86(2), 159–168. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Takada R, Jinno T, Koga D, Hirao M, Muneta T, Okawa A (2016) Comparison of wear rate and osteolysis between annealed and remelted highly cross-linked polyethylene in total hip arthroplasty. A case control study at 7 to 10 years follow-up. Orthop Traumatol Surg Res 102(6), 717–721. [DOI] [PubMed] [Google Scholar]
- 13.Glyn-Jones S, Thomas GE, Garfjeld-Roberts P, Gundle R, Taylor A, McLardy-Smith P, Murray DW (2015) The John Charnley Award: Highly crosslinked polyethylene in total hip arthroplasty decreases long-term wear: A double-blind randomized trial. Clin Orthop Relat Res 473(2), 432–438. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Citters DW (2007) Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am 89, 2023–2029. [DOI] [PubMed] [Google Scholar]
- 15.Halley D, Glassman A, Crowninshield RD (2004) Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: A case report. J Bone Joint Surg Am 86, 827–830. [DOI] [PubMed] [Google Scholar]
- 16.Hammerberg EM, Zhinian W, Manish D, Lawrence DD (2010) Wear and range of motion of different femoral head sizes. J Arthroplasty 25, 839–843. [DOI] [PubMed] [Google Scholar]
- 17.Mistry JB, Chughtai M, Elmallah RK, Diedrich A, Le S, Thomas M, Mont MA (2016) Trunnionosis in total hip arthroplasty: A review. J Orthop Traumatol 17(1), 1–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Lavernia CJ, Iacobelli DA, Villa JM, Jones K, Gonzalez JL, Jones WK (2015) Trunnion-head stresses in THA: Are big heads trouble? J Arthroplasty 30(6), 1085. [DOI] [PubMed] [Google Scholar]
- 19.Good V, Ries M, Barrack RL, et al. (2003) Reduced wear with oxidized zirconium femoral heads. J Bone Joint Surg [Am] 85-A(Suppl 4), 105–110. [DOI] [PubMed] [Google Scholar]
- 20.Bourne RB, Barrack R, Rorabeck CH, Salehi A, Good V (2005) Arthroplasty options for the young patient: Oxinium on cross-linked polyethylene. Clin Orthop Relat Res 441, 159–167. [DOI] [PubMed] [Google Scholar]
- 21.Jonsson BA, Kadar T, Havelin LI, Haugan K, Espehaug B, Indrekvam K, Furnes O, Hallan G (2015) Oxinium modular femoral heads do not reduce polyethylene wear in cemented total hip arthroplasty at five years: A randomised trial of 120 hips using radiostereometric analysis. Bone Joint J 97-B(11), 1463–1469. [DOI] [PubMed] [Google Scholar]
- 22.Karidakis GK, Karachalios T (2015) Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: A 7- to 12-year in vivo comparative wear study. Clin Orthop Relat Res 473(12), 3836–3845. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Clement ND, Mitchell M, Trayner ME, Porter DE, Lawson GM, Burnett R (2013) Accelerated polyethylene wear and early failure of the uncemented Reflection acetabular component: A ten to fifteen-year follow-up study. Int Orthop 37(12), 2337–2343. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Hengst DM, Thomsen PB, Homilius M, Hansen TB, Stilling M (2014) Higher wear-rate of third-generation metal-backed reflection cups with eto-sterilised UHMWPE at a mean 13 years follow-up. Hip Int 24(6), 587–595. [DOI] [PubMed] [Google Scholar]
- 25.Martell JM, Berdia S (1997) Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am 79, 1635–1641. [DOI] [PubMed] [Google Scholar]
- 26.Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H (2011) The John Charnley Award: risk factors for cup malpositioning: Quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res 469(2), 319–329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Dora C, Burckhardt K, Székely G, Rousson V, Hodler J, Gerber C (2008) A computerized analysis-by-synthesis algorithm improves precision of linear wear measurements in total hip replacements. J Orthop Res 26(8), 1121–1126. [DOI] [PubMed] [Google Scholar]
- 28.Kraay MJ, Moore RD, Martell JM, Rimnac CM (2010) Reassessment of computerized wear measurement for total hip arthroplasty with correction for projectional image distortion: A brief follow-up report. J Bone Joint Surg Am 92(9), 1858–1867. [DOI] [PubMed] [Google Scholar]
- 29.DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res 121, 20–32. [PubMed] [Google Scholar]
- 30.Fleiss J (1981) Statistical methods for rates and proportions, 2nd edn New York, NY, John Wiley and Sons. [Google Scholar]
- 31.Johanson PE, Digas G, Herberts P, Thanner J, Kärrholm J (2012) Highly crosslinked polyethylene does not reduce aseptic loosening in cemented THA 10-year findings of a randomized study. Clin Orthop Relat Res 470(11), 3083–3093. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Geerdink CH, Grimm B, Vencken W, Heyligers IC, Tonino AJ (2009) Cross-linked compared with historical polyethylene in THA: An 8-year clinical study. Clin Orthop Relat Res 467(4), 979–984. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Röhrl SM, Nivbrant B, Nilsson KG (2012) No adverse effects of submelt-annealed highly crosslinked polyethylene in cemented cups: An RSA study of 8 patients 10 yaers after surgery. Acta Orthop 83(2), 148–152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Engh CA Jr, Hopper RH Jr, Huynh C, Ho H, Sritulanondha S, Engh CA Sr (2012) A prospective, randomized study of cross-linked and non-cross-linked polyethylene for total hip arthroplasty at 10-year follow-up. J Arthroplasty 27(8 Suppl), 2–7.e1. [DOI] [PubMed] [Google Scholar]
- 35.García-Rey E, García-Cimbrelo E, Cruz-Pardos A (2013) New polyethylenes in total hip replacement: A ten- to 12-year follow-up study. Bone Joint J 95-B(3), 326–332. [DOI] [PubMed] [Google Scholar]
- 36.Snir N, Kaye ID, Klifto CS, Hamula MJ, Wolfson TS, Schwarzkopf R, Jaffe FF (2014) 10-year follow-up wear analysis of first-generation highly crosslinked polyethylene in primary total hip arthroplasty. J Arthroplasty 29(3), 630–633. [DOI] [PubMed] [Google Scholar]
- 37.Babovic N, Trousdale RT (2013) Total hip arthroplasty using highly cross-linked polyethylene in patients younger than 50 years with minimum 10-year follow-up. J Arthroplasty 28(5), 815–817. [DOI] [PubMed] [Google Scholar]
- 38.Morison ZA, Patil S, Khan HA, Bogoch ER, Schemitsch EH, Waddell JP (2014) A randomized controlled trial comparing Oxinium and cobalt-chrome on standard and cross-linked polyethylene. J Arthroplasty 29(9 Suppl), 164–168. [DOI] [PubMed] [Google Scholar]
- 39.Garvin KL, White TC, Dusad A, Hartman CW, Martell J (2015) Low wear rates seen in THAs with highly crosslinked polyethylene at 9 to 14 years in patients younger than age 50 years. Clin Orthop Relat Res 473(12), 3829–3835. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Tan SC, Lau AC, Del Balso C, Howard JL, Lanting BA, Teeter MG (2016) Tribocorrosion: Ceramic and oxidized zirconium vs cobalt-chromium heads in total hip arthroplasty. J Arthroplasty 31(9), 2064–2071. [DOI] [PubMed] [Google Scholar]
- 41.Carli A, Koch CN, Esposito CI, Wright TM, Padgett DE (2018) Polyethylene wear increases in liners articulating with scratched oxidized zirconium femoral heads. Clin Orthop Relat Res 476(2), 182–192. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Sychterz CJ, Engh CA Jr, Yang A, Engh CA (1999) Analysis of temporal wear patterns of porous-coated acetabular components: Distinguishing between true wear and so-called bedding in. J Bone Joint Surg 81, 821. [DOI] [PubMed] [Google Scholar]
- 43.Callary SA, Solomon LB, Holubowycz OT, Campbell DG, Howie DW (2017) Accuracy of methods to measure femoral head penetration within metal-backed acetabular components. J Orthop Res 35(5), 988–996. [DOI] [PubMed] [Google Scholar]
- 44.Langlois J, Zaoui A, Scemama C, Martell J, Bragdon C, Hamadouche M (2015) Validation of a computer-assisted method for measurement of radiographic wear in total hip arthroplasty using all polyethylene cemented acetabular components. J Orthop Res 33(3), 417–420. [DOI] [PubMed] [Google Scholar]
- 45.Gaudiani MA, White PB, Ghazi N, Ranawat AS, Ranawat CS (2018) Wear rates with large metal and ceramic heads on a second generation highly cross-linked polyethylene at mean 6-year follow-up. J Arthroplasty 33(2), 590–594. [DOI] [PubMed] [Google Scholar]
- 46.Lachiewicz PF, Soileau ES, Martell JM (2016) Wear and osteolysis of highly crosslinked polyethylene at 10 to 14 years: The effect of femoral head size. Clin Orthop Relat Res 474(2), 365–371. [DOI] [PMC free article] [PubMed] [Google Scholar]