Abstract
Ligament tension is an important factor that can affect the success of total knee arthroplasty (TKA) procedures. However, surgeons currently lack objective approaches for assessing tension in a particular ligament intraoperatively. The purpose of this study was to investigate the use of noninvasive shear wave tensiometry to characterize stress in medial and lateral collateral ligaments (MCLs and LCLs) ex vivo and evaluate the capacity of shear wave speed to predict axial load. Nine porcine MCL and LCL specimens were subjected to cyclic axial loading while shear wave speeds were measured using laser vibrometry. We found that squared shear wave speed increased linearly with stress in both the MCL (r2avg = 0.94) and LCL (r2avg = 0.98). Shear wave speeds were slightly lower in the MCL than the LCL when subjected to comparable axial stress (p < 0.001). Specimen-specific calibrations predicted tension within 13.0 N, or 5.2% of the maximum load. A leave-one-out analysis was also performed and showed that calibrations based on ligament type could predict axial tension within 15% of the maximum load. These observations suggest it may be feasible to use noninvasive shear wave speed measures as a proxy of ligament loading, which in the future might enhance decision making during orthopedic procedures such as TKA.
Keywords: Total knee arthroplasty, Shear wave tensiometry, Non-invasive intraoperative sensor development, Orthopedic surgery, Soft tissue balancing
Graphical Abstract
1. Introduction
Ligament balancing is critical to outcomes of total knee arthroplasty (TKA) (Blankevoort et al., 1988; Wilson et al., 1998). To achieve balance, surgeons adjust TKA component alignment, select TKA component size, and/or perform ligament releases to adequately tensions the ligaments without limiting motion. Tension in the superficial medial and lateral collateral ligaments (MCL and LCL, respectively) are of particular interest. The MCL and LCL are primary restraints to rotational motion in the frontal plane (i.e., varus-valgus rotation) and secondary restraints to rotation in the transverse plane (i.e., internal-external rotation) and translation in the sagittal plane (i.e., anterior-posterior translation) (Blankevoort et al., 1991; Grood et al., 1981; Haimes et al., 2007; Lim et al., 2012; Seering et al., 1980). Overly-loose (Abdel and Haas, 2014; Cottino et al., 2016; Vince, 2016), overly-tight (Babazadeh et al., 2009; Laskin and Beksac, 2004), or asymmetrically tensioned collateral ligaments (Vince et al., 2006) are associated with complications following TKA, including joint instability, stiffness, and pain. Hence, ligament tension is an important factor that can affect the success of TKA (Anderson et al., 1996; Harris and Sledge, 1994; Hawker et al., 1998). However, surgeons currently lack objective approaches for assessing tension in a particular ligament intraoperatively, and instead are reliant on qualitative and indirect approaches (Camarata, 2014; Daines and Dennis, 2014; Edwards et al., 1988; Fujimoto et al., 2015; Gustke, 2012; Gustke et al., 2014; Matsumoto et al., 2011; Nowakowski et al., 2012).
Shear wave tensiometry is an emerging, non-invasive approach for gauging soft tissue tension based on the propagation of induced shear waves. A tensioned beam model and experiments have demonstrated that squared shear wave speed increases linearly with axial stress in a tendon under physiological loads (Martin et al., 2019, 2018). This observation results from the anisotropic nature of tendons in which the aligned collagen fibers lead to a high axial stiffness, but a relatively low shear modulus because the fibers are able to slide relative to one another. The shear wave speed-stress relationship may extend to ligaments, which also consist of bundles of collagen fibers that bear physiological loads (Quapp and Weiss, 1998). However, the fibrous structure, boundary conditions, and geometry of ligaments differ from those of tendons in ways that might alter the shear wave speed-stress relationship. At the microstructural level, ligament fibers are generally less well aligned than those in tendons (Amis, 1998). Ligaments are also anchored to the skeleton on both the proximal and distal ends, resulting in boundary conditions which could reflect shear waves. Finally, ligaments like the MCL are sheet-like with large aspect ratios that differ considerably from the LCL as well as from the tendons (Achilles, patellar) that have been tested previously with shear wave tensiometry (Martin et al., 2018).
A fundamental pre-requisite step to using shear wave tensiometry intraoperatively is to determine whether the shear wave speed-stress relationship observed in tendons extends to collateral ligaments. Accordingly, the objectives of this study were (1) to characterize the squared shear wave speed-stress relationship in isolated collateral ligaments, (2) to determine whether the squared shear wave speed-stress relationship differed between the MCL and LCL, and (3) to determine the errors in using shear wave speed to predict axial stress in these specimens.
2. Materials and Methods
2.1. Specimen Preparation
Nine MCLs and nine LCLs were procured from a crossbreed of large white, landrace, and red duroc pigs (weight = 132.5 ± 18.3 kg, age = 6 months). While not skeletally mature, porcine collateral ligaments were used for this study because MCLs from pigs of this size have been previously shown to have an axial loading response similar to that of human MCLs (Germscheid et al., 2011). Each ligament was isolated by first removing superficial and surrounding tissue (Fig. 1a). Knees (i.e., stifle joints) were then flexed to the angle where the fibers across the ligament appeared to be most uniformly taught. This was typically around 80° of flexion (Germscheid et al., 2011). Next, bone blocks containing the bony ligament attachments were cut from the distal femoral condyles and tibia/fibula using an oscillating saw. Each bone block was secured in an aluminum cup using a potting material (Bondo Fiberglass Resin 3M, St. Paul MN). Excess tissue superficial and deep to the outer surface of the ligament was removed prior to experimentation.
Figure 1:
Representative images of (a) dissected porcine knee (i.e., stifle joint) and (b) isolated MCL. (a) The image shows the medial aspect of the knee with the MCL identified by the black dashed rectangle. Excess soft tissue was removed from the intact porcine knee to expose the MCL and LCL. The MCL and LCL were isolated with the knee at the flexion angle at which the fibers across the width of the ligament were most uniformly taught; this was typically at 80° flexion (Germscheid et al., 2011). (b) The image shows the deep surface of a representative MCL mounted in the MTS machine. The bone blocks were potted in aluminum fixtures to enable a rigid fixation within the MTS grips. Two laser vibrometers were used for tracking shear wave propagation.
2.2. Experimental Procedure
Specimens were mounted in an electrodynamic testing system (Acumen 3, MTS, Eden Prairie, MN) and pre-loaded to 10 N of axial force. Specimens were first immersed in an acoustic tile-lined water bath (Aptflex F28, Precision Acoustics, Dorchester, UK) to facilitate ultrasound imaging (SonixTOUCH Research, BK Medical, Peabody, MA). A linear array ultrasound transducer (L14–5W/38, BK Medical, Peabody, MA) was affixed to a motorized linear stage (Newmark Systems Inc, Ranch Santa Margarita, CA) to facilitate collection of cross-sectional images at 0.1 mm increments along the length of the ligament. The B-Mode imaging was performed through an acoustic window (0.01 inch thick polycarbonate). Cross-sectional areas were segmented from the ultrasound images using an open source 3D visualization software (3D Slicer, Cambridge, MA). The aspect ratio was computed as the ratio between the major and minor axes of the ligament cross section (Fig. 2a). The major axis was the greatest edge-to-edge distance of the cross-section that intersected the centroid. The minor axis was the edge-to-edge distance perpendicular to the major axis and coincident with the centroid.
Figure 2:
Diagrams show the testing setups used to determine the cross-sectional geometries of the ligaments and the shear wave speed stress relationships. (a) The cross-sectional area and aspect ratio were measured by segmenting B-mode ultrasound images collected along the length of the ligament. (b) Ligaments underwent cyclic axial loading at 1 Hz between 10 and 250 N. Simultaneously, shear waves were excited using a piezoelectric tapper and transverse ligament motion was measured using two laser Doppler vibrometers. Shear wave propagation speed, c, was computed based on laser spacing, d, and the time delay between the shear wave arrival, t2 - t1, at the two measurement locations.
Cyclic pre-conditioning (100 cycles from 10 to 250 N at 1 Hz) was performed prior to experimentation to mitigate viscoelastic effects (Schatzmann et al., 1998), and thereby result in more consistent specimen excursions during the cyclic loading cycles. Ligaments were then subjected to cyclic axial loading to determine the shear wave speed-stress relationship. Ligament specimens underwent ten loading cycles from 10 to 250 N at 1 Hz in air. The maximum tension was limited to 250 N based on pilot studies that showed failures at the bony attachments of several porcine ligaments when applied loads exceeded 250 N. During cyclic loading, shear waves were excited using a custom piezoelectric (PK4JQP1, Thorlabs Inc., Newton, NJ) device to deliver 20 micron taps across the width of the ligament at 25 Hz. Transverse velocity of the ligament was measured using laser Doppler vibrometers (Polytec Inc., Irvine, CA) at two points on the ligament surface, opposite to that of the tapper. Laser points were spaced 5 mm apart with the first point 5 mm above the tap location (Figs. 1b and 2b) This method of excitation and laser vibrometer measurement was previously shown to be effective for tracking shear wave propagation in tendons (Martin et al., 2019). Three cyclic loading trials for each specimen were recorded, which summed to thirty total loading cycles for each specimen. For each tap, there were 2,000 load measurements, which were averaged to obtain the stress for the corresponding shear wave speed measurement. Ligaments were kept moist with 0.9% saline for the duration of the testing procedure.
2.3. Post-Processing
The induced shear wave speeds for each tap were computed by measuring the time delay between the arrivals of the shear wave at the two successive laser points. Waveforms were filtered using a 2nd order Butterworth bandpass filter (MATLAB R2018) with lower and upper cutoff frequencies of 100 and 5000 Hz, respectively. A time-dependent cross-correlation between the transient response of the two waveforms (i.e., the first 0.6 ms of each signal) was performed to determine the time delay in shear wave arrival between laser vibrometer points (Martin et al., 2018). Sub-sample interpolation was performed using a local 3-point cosine fit of the normalized cross-correlation values (Cespedes et al., 1995). Average axial stresses in ligaments during loading were computed using the applied force measured by the load cell (661.18E-02, MTS, reported hysteresis/nonlinearity = 0.08% full scale) and average cross-sectional area between the measurement locations determined from the segmented ultrasonic scan of each ligament.
2.4. Statistical Analysis
Specimen-specific simple linear regressions were performed to ascertain the linearity of the squared shear wave speed-stress relationship with stress as the independent variable and squared shear wave speed the dependent variable. All statistics were performed on the raw dataset (i.e., not sorted into bins). The level of significance, α, was set to 0.05. With nine ligaments of each type, strong relationships between squared shear wave speed and stress (i.e., r2 ≥ 0.55) could be detected with α = 0.05 and (1 - β) = 0.8 (Faul et al., 2009).
Stress and shear wave speed data were then grouped according to ligament type (i.e., MCL or LCL) to determine whether the squared shear wave speed-stress relationship differed between the MCL and LCL. For each ligament type, a simple linear regression was conducted with the stress as the independent variable and the squared shear wave speed as the dependent variable. A follow-up linear regression (Gujarati, 1970) was performed to determine whether the intercepts and slopes of the squared shear wave speed-stress relationships were different between ligament types (Eq. 1). Again, α was set at 0.05. The dependent variable in this follow up linear regression was the stress, σ. The independent variables were the squared shear wave speed, , a binary indicator variable for ligament type, Di (D = 0 for MCL, D = 1 for LCL), and the interaction term between squared shear wave speed and the indicator variable, . The index, i, indicates the data point.
(1) |
By including the indicator variable in the regression, the slope of the squared shear wave speed-stress relationship is β1 for the MCL and β1 + β3 for the LCL, and the intercept of the squared shear wave speed-stress relationship is β0 for the MCL and β0 + β2 for the LCL. Differences in the slope and intercept of the shear wave speed-stress relationship in the MCL and LCL will be assessed with α = 0.05.
Finally, we performed a leave-one-out regression to evaluate the capacity of using shear wave speed to predict axial ligament stress using a subset of specimens. In this analysis, a linear regression between squared shear wave speed and stress was performed on data from 8 ligaments and then the stress in the 9th ligament was predicted using the regression equation. We also predicted the stress using a linear regression of the entire dataset excluding the predicted ligament, and also using a linear regression of only the predicted ligament. This procedure was repeated for all combinations of 8 ligaments, and the resulting errors in stress were characterized by the mean (bias), standard deviation (precision), and root-mean-square error (RMSE) (ISO 5725–1, ASTM E177–13).
The average lengths and cross-sectional areas of the porcine MCLs and LCLs were compared using two-sample t-tests with α = 0.05. Average ligament aspect ratios were compared using one-sided t-test with α = 0.05 and the hypothesis that the MCL has a larger aspect ratio than the LCL.
3. Results
The shear wave speed of both the MCL and the LCL increased monotonically with axial stress (Fig. 3). There was a strong linear relationship from the specimen-specific simple linear regressions between the squared shear wave speed and axial stress in each MCL and LCL (r2 = 0.94 ± 0.06 (mean ± standard deviation across the 9 ligaments of each type) and 0.98 ± 0.02, respectively). When all of the data for each ligament type was grouped, there were also strong linear relationships (Table 1), but there was specimen-to-specimen variability that decreased the coefficients of determination of each ligament.
Figure 3:
Scatter plots show the monotonically increasing shear wave speed-stress relationship in all nine MCLs (a) and all nine LCLs (b). Data from each MCL and LCL is plotted with a different shade of red or blue, respectively. Data points indicate the mean and error bars indicate the standard deviation of shear wave speeds within 0.5 MPa bins for each specimen.
Table 1:
Comparison of linear regression coefficients (Eq 1.) from the grouped data of each ligament type. The LCL exhibited slightly higher shear wave speeds than the MCL at a comparable stress, resulting in a smaller slope of the squared shear wave speed-stress relationship (Eq. 1). We did not detect a significant difference between intercept terms of each ligament type. The coefficients of determination, r2, for the simple linear regressions of grouped data were less than those for an individual specimen due to specimen-to-specimen differences.
slope (kPa/m2/s2) | Intercept (kPa) | r2 | |
---|---|---|---|
MCL | 0.620 | −645 | 0.81 |
LCL | 0.507 | −616 | 0.88 |
p < 0.001 | p = 0.689 |
From the grouped data collected from each ligament type (Fig. 4, Table 1), the slope of the linear fit was significantly greater in the MCL than that of the LCL (p < 0.001). However, we did not detect a significant difference between the vertical axis intercepts of the squared shear wave speed-stress relationships between the MCL and LCL (p = 0.689) (Table 1).
Figure 4:
Scatter plot represents the mean (squares) and standard deviation (error bars) of the data from every specimen grouped into bins. Data was binned into 5 m2/s2 bins. The linear prediction model characterized over 80% of the variance in axial stress in the MCL and LCL (r2 = 0.81 and 0.88 from grouped simple linear regressions, respectively).
From the leave-one-out regression analysis, we determined that, in general, the errors in predicting stress using measured shear wave speed were reduced using ligament-specific data rather than data from other ligaments (Tables 2 and 3). For the MCL, the RMSE of predicted stress was 1.90 MPa (22% of max load) when the calibration regression was performed on data from all other MCLs and LCLs (Table 2). The RMSE was 1.95 MPa (23% of max load) and 0.68 (7.9% of max) when the calibrated regression was performed on data from just the other MCLs or just the one specific MCL, respectively (Table 2). For the LCL, the RMSE of predicted stress was 1.53 MPa (20% of max load) when the calibration regression was performed on data from all other MCLs and LCLs (Table 3). The RMSE was 1.22 MPa (16% of max) and 0.47 (6.2% of max) when the calibrated regression was performed on data from just the other LCLs or just the one specific LCL, respectively (Table 3).
Table 2:
Leave-one-out analysis of nine subsets of MCL specimens using regression fits from all MCL and LCL specimens (total dataset), fits specific to ligament type, and specimen-specific regression fits. Shown are the mean (b, bias), standard deviation (p, precision), and root-mean-square error (RMSE).
Predicted Specimen | Stress Error Metric [MPa] | ||||||||
---|---|---|---|---|---|---|---|---|---|
Total Dataset | Ligament Type | Specimen Specific | |||||||
b | p | RMSE | b | p | RMSE | b | p | RMSE | |
MCL 1 | 0.55 | 0.47 | 0.72 | 0.46 | 0.37 | 0.60 | 0.44 | 0.35 | 0.56 |
MCL 2 | 0.42 | 0.43 | 0.60 | 0.85 | 0.85 | 1.20 | 0.23 | 0.14 | 0.27 |
MCL 3 | 0.30 | 0.21 | 0.36 | 0.26 | 0.20 | 0.33 | 0.25 | 0.16 | 0.29 |
MCL 4 | 1.11 | 1.02 | 1.51 | 1.56 | 1.49 | 2.16 | 0.45 | 0.23 | 0.51 |
MCL 5 | 0.94 | 0.92 | 1.32 | 1.18 | 0.97 | 1.53 | 1.01 | 0.96 | 1.40 |
MCL 6 | 1.03 | 0.64 | 1.21 | 0.55 | 0.41 | 0.68 | 0.33 | 0.24 | 0.41 |
MCL 7 | 1.11 | 0.67 | 1.30 | 1.81 | 1.04 | 2.09 | 0.27 | 0.23 | 0.25 |
MCL 8 | 3.94 | 2.77 | 4.81 | 3.62 | 2.57 | 4.44 | 0.49 | 0.52 | 0.71 |
MCL 9 | 0.87 | 0.72 | 1.13 | 0.65 | 0.58 | 0.87 | 0.62 | 0.50 | 0.79 |
Average | 1.14 | 1.53 | 1.90 | 1.22 | 1.52 | 1.95 | 0.45 | 0.50 | 0.68 |
Table 3:
Leave-one-out analysis of nine subsets of LCL specimens using regression fits from all MCL and LCL specimens (total dataset), fits specific to ligament type, and specimen-specific regression fits. Shown are the mean (b, bias), standard deviation (p, precision), and root-mean-square error (RMSE).
Predicted Specimen | Stress Error Metric [MPa] | ||||||||
---|---|---|---|---|---|---|---|---|---|
Total Dataset | Ligament Type | Specimen Specific | |||||||
b | p | RMSE | b | p | RMSE | b | p | RMSE | |
LCL 1 | 0.36 | 0.26 | 0.45 | 0.51 | 0.40 | 0.65 | 0.32 | 0.22 | 0.39 |
LCL 2 | 0.69 | 0.46 | 0.83 | 0.20 | 0.13 | 0.24 | 0.13 | 0.10 | 0.16 |
LCL 3 | 0.92 | 0.47 | 1.03 | 0.54 | 0.27 | 0.60 | 0.13 | 0.09 | 0.16 |
LCL 4 | 0.40 | 0.33 | 0.52 | 0.59 | 0.40 | 0.71 | 0.34 | 0.28 | 0.44 |
LCL 5 | 0.42 | 0.34 | 0.54 | 0.33 | 0.30 | 0.44 | 0.23 | 0.27 | 0.36 |
LCL 6 | 0.86 | 0.60 | 1.04 | 0.40 | 0.25 | 0.47 | 0.32 | 0.18 | 0.36 |
LCL 7 | 0.91 | 0.70 | 1.15 | 1.32 | 1.01 | 1.66 | 0.38 | 0.34 | 0.51 |
LCL 8 | 2.73 | 1.98 | 3.37 | 2.08 | 1.57 | 2.61 | 0.35 | 0.34 | 0.49 |
LCL 9 | 1.86 | 1.16 | 2.19 | 1.24 | 0.80 | 1.47 | 0.63 | 0.63 | 0.89 |
Average | 1.02 | 1.14 | 1.53 | 0.80 | 0.92 | 1.22 | 0.32 | 0.35 | 0.47 |
The lengths and cross-sectional areas did not differ between the MCL and LCL (p = 0.698 and p = 0.325, respectively). However, the aspect ratios did differ between the MCL and LCL (p = 0.025), with the MCL (3.0 ± 0.9) being wider and flatter than the LCL (2.4 ± 0.5) (Table 2).
4. Discussion
The objectives of the present study were (1) to characterize the squared shear wave speed-stress relationship in isolated collateral ligaments, (2) to determine whether these relationships differed between the MCL and LCL, and (3) to determine the errors in using shear wave speed to predict axial stress in these specimens. The first key finding is that strong linear relationships exist in both the MCL and LCL. This justifies the future use of shear wave speed measurements, which are known to quantify axial stress in tendons, as a proxy measure for ligament loading. The second key finding is that there were subtle differences in the shear wave speed-stress relationship between the MCL and LCL, which may need to be considered when using shear wave speed to infer loading of individual structures. The third key finding was that, when using specimen-specific calibrations, ligament loading could be predicted from the shear wave speed within 7.9% of the maximum load level.
There was a strong linear relationship between squared shear wave speed and axial stress in both the MCL and LCL (r2 = 0.94 and 0.98, respectively, Fig. 3), which is comparable to that observed previously in porcine digital flexor tendons (r2 = 0.96–0.98) (Martin et al., 2019, 2018). The linear relationship is consistent with a tensioned beam model (Martin et al., 2018), which predicts that shear wave speed depends both on the shear modulus of the material and the applied axial stress. The shear modulus is relatively low in ligaments (Tanaka et al., 2007; Weiss et al., 2002) due to low resistance to interfibrillar sliding. Hence, the dependence of wave speed on axial stress is believed to become dominant as loading of the collagen fibers is increased, giving rise to the linear squared shear wave speed-stress relationship that is observed. While the fundamental shear wave speed-stress relationship is similar between ligaments and tendons, the ligament shear wave speeds were generally higher than those observed in porcine digital flexor tendons (Martin et al., 2019). For example, we observed average ligament shear wave speeds of 128 m/s for an 8 MPa load, which is 45% higher than that observed in tendon under similar loads ex vivo.
Structural differences in the geometry, boundary conditions, and fibrous organization attributed to ligaments are potentially contributing factors to the heightened shear wave speeds observed in this study. Ligaments like the MCL are sheet-like with large aspect ratios that differ from both the LCL (Table 2) and the digital flexor tendons that have been tested previously (Martin et al., 2019, 2018). Moreover, wave-guided behavior in tendon and ligament gives rise to shear wave dispersion, which is known to depend on the thickness of the tissue relative to the shear wave length (Brum et al., 2014). Hence, the MCL and LCL, which both have high aspect ratios, may have a fundamentally different shear wave dispersion profile than that of tendon. A second factor is boundary conditions. The ligaments were tested with intact bony attachments in contrast with previous studies that studied wave speeds in tendons secured in rigid metal grips (Martin et al., 2019, 2018). The ligament bony attachments appear to provide a soft enough interface to limit the reflections of shear waves from the ends of the ligament, and therefore, we did not observe standing waves in this study which were observed in clamped porcine flexor tendons (Martin et al., 2019, 2018). Finally, at the microstructural level, ligament fibers are generally less well aligned than those in tendons (Amis, 1998). This may result in greater resistance to shear along the long axis of the ligament and hence a larger apparent shear modulus. Specimen-to-specimen differences in fiber architecture could also contribute to variability seen in the overall regression (Figure 4).
Although there were strong linear relationships between squared shear wave speed and axial stress in both the MCL and LCL, our results demonstrate that the slopes, and thus the linear relationships, do differ between ligament types (Fig. 4, Table 1). Unlike slope, we did not detect a significant difference between the intercepts of the linear regressions in the MCL and LCL, which indicates that the two ligament types have a similar shear wave speed in an unloaded state, and thus have a similar unloaded shear modulus (Martin et al., 2018). This also indicates that the effective density, which also exists in the analytical relationship governing shear wave speed, was similar between specimens. Moreover, this seems likely because all surrounding tissue was removed from each specimen, which would negate any increase in the effective tissue density.
The difference in slope indicates that ligament-specific calibrations might be necessary to use shear wave speed as a proxy measure of axial stress in ligaments moving forward. The greater slope in the shear wave speed-stress relationship in the MCL indicates that the shear wave speeds are lower in the MCL than those in the LCL for a comparable axial stress. One factor that might bring about this difference is non-uniformity in the axial stress across the width of the MCL. This non-uniformity is more likely to occur in the MCL due to its larger aspect ratio. In a non-uniformly loaded ligament, the local stress might be less than the average stress used to determine the squared shear wave speed-stress relationship due to high stresses near one of the edges of the ligament. It is feasible that these high edge stresses would increase the average stress leading to an over-estimation of the local stress under the laser points. Thus, the discrepancy between average and local stress could give rise to localized shear wave speeds in the MCL that differ from those measured in the LCL at a comparable applied average stress. This effect is likely more prominent in ligaments in vivo, where subtle changes in knee flexion angle are known to produce different stress or strain gradients across the ligament width (Gardiner and Weiss, 2003; Robinson et al., 2004). If shear wave speeds are indeed sensitive to local stress, then ligament shear wave tensiometry might enable clinicians to determine the stress distribution within the ligament rather than just the average axial stress.
The final objective of this study was to evaluate the capacity of using shear wave speed to predict ligament stress. A leave-one-out analysis using the specimens in this experiment showed that, on average, we can detect the axial stress within 1.22 and 0.80 MPa in the MCL and LCL, respectively, using a “group” calibration. Given the average MCL and LCL cross-sectional areas, these errors are relatively large, corresponding to 35.4 and 26.3 N, respectively, and would correspond to changes in shear wave speed of 44.4 and 39.7 m/s, respectively. This shows that there is still potential for error in balancing collateral ligaments during surgery if this technique were to be implemented using a group calibration. However, one large source of error in the leave-one-out analysis is the ligament-to-ligament difference in the shear wave speed-stress relationship. We found that specimen-specific calibrations predicted axial stresses in the MCL and LCL within 0.68 MPa (19.6 N) and 0.47 MPa (15.4 N), respectively, which correspond to within 7.9% of the maximum load. Future studies will focus on developing this specimen-specific calibration procedure.
Although a promising shear wave speed-stress relationship was determined in the porcine MCLs and LCLs, three important challenges in the clinical translation of shear wave tensiometry still remain. One challenge is that the shear wave speed-stress relationship in human ligaments may be unique from the relationship exhibited in porcine ligaments. Porcine collateral ligaments do provide an appropriate animal model for preliminary ex vivo testing (Germscheid et al., 2011; Xerogeanes et al., 1998) because of the similarities in the mechanical response of the human and porcine MCL to loading. While porcine collateral ligaments have aspect ratios that are different between the MCL and LCL (3.0 and 2.4, respectively), there is a greater distinction in aspect ratio between human MCLs and LCLs (16 and 2, respectively) (Wilson et al., 2012). Additional work is needed to ascertain whether these geometric differences lead to more distinct shear wave speed-stress relationships in human MCLs and LCLs. A second challenge is that shear wave speed measurements were taken on isolated collateral ligaments. Interactions with surrounding tissues/implants, including but not limited to the joint capsule, menisci, underlying femoral and tibial bone before TKA, and the underlying femoral and tibial implants after TKA would likely decrease the shear wave speed at a given stress by increasing the effective mass (Martin et al., 2018). Bone insertions of the ligaments were maintained in this study, suggesting that these transient shear wave speed measurements are viable on intact collateral ligaments. A third challenge is creating a handheld shear wave tensiometer suitable for intraoperative use. Here, we showed that transient shear wave tensiometry via laser Doppler vibrometry is a viable approach to gauge isolated ligament stress ex vivo. However, the experimental setup we used is not practical for intraoperative checks made by surgeons. A miniaturized laser Doppler vibrometry approach exists to monitor pulse wave propagations using a handheld scanner (Mancini et al., 2019), which might allow for a more viable measurement within the operating room. Also, using miniature accelerometers to monitor transverse tissue motion might make intraoperative ligament tensiometry possible as this approach is used in vivo for tendons (Martin et al., 2018). Such a handheld device is a primary focus of ongoing research because it would enable the translation to TKA and other intraoperative applications including tendon repair (Beskin et al., 1987; Savin et al., 2017) and muscle lengthening (Novacheck and Gage, 2007; Schwartz et al., 2004).
In conclusion, shear wave tensiometry provides a useful and objective quantification of the axial stress in ligaments as it does in tendons. Both the porcine MCLs and LCLs used in this study have strongly linear but distinct squared shear wave speed-stress relationships. These findings indicate that shear wave speed measurements can be used to predict tension in collateral ligaments.
Table 4:
Comparison of ligament geometry parameters of the MCL and LCL. The MCL had a significantly higher mean aspect ratio than the LCL (*p = 0.025). We did not detect a significant difference between the mean MCL and LCL length or cross-sectional area.
Ligament Geometry Measures (mean ± standard deviation) | ||
---|---|---|
Ligament Geometry Parameter | MCL | LCL |
Length (mm) | 23.3 ± 8.3 | 24.5 ± 4.5 |
Area (mm2) | 29.0 ± 8.9 | 32.9 ± 7.3 |
Aspect Ratio* | 3.0 ± 0.9 | 2.4 ± 0.5 |
Highlights.
Shear wave tensiometry is a promising method for gauging load in fibrous connective tissues.
Porcine medial and lateral collateral ligaments exhibited highly linear (r2 > 0.9) relationships between squared shear wave speed and axial stress.
Subtle differences in the slope of the squared shear wave speed-stress relationship between ligament types may need to be considered when using shear wave tensiometry to infer loading of individual structures.
5. Acknowledgments
This work was supported by the National Institute of Biomedical Imaging and Bioengineering (NIBIB) and the National Institute of Aging (NIA) of the National Institutes of Health under award numbers [NIBIB: R21EB024957; NIA: T32AG000213].
All specimens were procured from the University of Wisconsin-Madison Department of Animal Sciences.
6. Appendices
6.1. Appendix A
The data used to evaluate the capacity of shear wave speed in predicting stress can be observed below (Fig. A.1, A.2) for each specimen of each ligament type.
Figure A.1:
Scatter plots show all shear wave speeds collected for each MCL specimen.
Figure A.2:
Scatter plots show all shear wave speeds collected for each LCL specimen.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Declaration of interests
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
References
- Abdel MP, Haas SB, 2014. The unstable knee: Wobble and Buckle. Bone Jt. J 96B, 112–114. 10.1302/0301-620X.96B11.34325 [DOI] [PubMed] [Google Scholar]
- Amis AA, 1998. Biomechanics of Bone, Tendon, and Ligament, in: Sciences Basic to Orthopaedics. pp. 222–239. [Google Scholar]
- Anderson JG, Wixson RL, Tsai D, Stulberg SD, Chang RW, 1996. Functional outcome and patient satisfaction in total knee patients over the age of 75. J. Arthroplasty 11, 831–840. 10.1016/S0883-5403(96)80183-5 [DOI] [PubMed] [Google Scholar]
- Babazadeh S, Stoney JD, Lim K, Choong PFM, 2009. The relevance of ligament balancing in total knee arthroplasty: how important is it? A systematic review of the literature. Orthop. Rev. (Pavia) 1, 70–78. 10.4081/or.2009.e26 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Beskin JL, Sanders RA, Hunter SC, Hughston JC, 1987. Surgical repair of Achilles tendon ruptures. Am. J. Sports Med 15, 1–8. 10.1177/036354658701500101 [DOI] [PubMed] [Google Scholar]
- Blankevoort L, Huiskes R, de Lange A, 1991. Recruitment of Knee Joint Ligaments. J. Biomech. Eng 113, 94 10.1115/1.2894090 [DOI] [PubMed] [Google Scholar]
- Blankevoort L, Huiskes R, de Lange A, 1988. The envelope of passive knee joint motion. J. Biomech 21, 705–720. 10.1016/0021-9290(88)90280-1 [DOI] [PubMed] [Google Scholar]
- Brum J, Bernal M, Gennisson J, Tanter M, 2014. In vivo evaluation of the elastic anisotropy of the human Achilles tendon using shear wave dispersion analysis. Phys. Med. Biol 59, 505–523. 10.1088/0031-9155/59/3/505 [DOI] [PubMed] [Google Scholar]
- Camarata D, 2014. Soft tissue balance in total knee arthroplasty with a force sensor. Orthop. Clin. North Am 45, 175–184. 10.1016/j.ocl.2013.12.001 [DOI] [PubMed] [Google Scholar]
- Cespedes I, Huang Y, Ophir J, Spratt S, 1995. Methods for estimation of subsample time delays of digitized echo signals. Ultrason. Imaging 17, 142–171. 10.1006/uimg.1995.1007 [DOI] [PubMed] [Google Scholar]
- Cottino U, Sculco P, Sierra R, Abdel M, 2016. Instability After Total Knee Arthroplasty. Orthop. Clin. North Am 47, 311–316. 10.1016/j.ocl.2015.09.007 [DOI] [PubMed] [Google Scholar]
- Daines BK, Dennis DA, 2014. Gap balancing vs. measured resection technique in total knee arthroplasty. Clin. Orthop. Surg 6, 1–8. 10.4055/cios.2014.6.1.1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Edwards E, Miller J, Chan K, 1988. The effect of postoperative collateral ligament laxity in total knee arthroplasty. Clin. Orthop. Relat. Res 236, 44–51. [PubMed] [Google Scholar]
- Faul F, Erdfelder E, Buchner A, Lang AG, 2009. Statistical power analyses using G*Power 3.1: Tests for correlation and regression analyses. Behav. Res. Methods 41, 1149–1160. 10.3758/BRM.41.4.1149 [DOI] [PubMed] [Google Scholar]
- Fujimoto E, Sasashige Y, Tomita T, Sasaki H, Touten Y, Fujiwara Y, Ochi M, 2015. Intra-operative gaps affect outcome and postoperative kinematics in vivo following cruciate-retaining total knee arthroplasty. Int. Orthop 40, 41–49. 10.1007/s00264-015-2847-y [DOI] [PubMed] [Google Scholar]
- Gardiner JC, Weiss JA, 2003. Subject-specific finite element models can predict strain in the human medial collateral ligament. J. Orthop. Res 21, 1098–1106. 10.1016/S0736-0266(03)00113-X [DOI] [PubMed] [Google Scholar]
- Germscheid NM, Thornton GM, Hart DA, Hildebrand KA, 2011. A biomechanical assessment to evaluate breed differences in normal porcine medial collateral ligaments. J. Biomech 44, 725–731. 10.1016/j.jbiomech.2010.10.036 [DOI] [PubMed] [Google Scholar]
- Grood ES, Noyes FR, Bulter DL, Suntay WJ, 1981. Ligamentous and capsular restraints preventing straight medial and lateral laxity in intact human cadaver knees. J. Bone Jt. Surg. - Ser. A 63, 1257–1269. 10.2106/00004623-198163080-00007 [DOI] [PubMed] [Google Scholar]
- Gujarati D, 1970. Use of Dummy Variables in Testing for Equality between Sets of Coefficients in Two Linear Regressions: A Note. Am. Stat 24, 50–52. 10.1080/00031305.1970.10477181 [DOI] [Google Scholar]
- Gustke K, 2012. Use of smart trials for soft-tissue balancing in total knee replacement surgery. J. Bone Jt. Surg 94, 147–150. 10.1302/0301-620X.94B11.30621 [DOI] [PubMed] [Google Scholar]
- Gustke KA, Golladay GJ, Roche MW, Elson LC, Anderson CR, 2014. Primary TKA patients with quantifiably balanced soft-tissue achieve significant clinical gains sooner than unbalanced patients. Adv. Orthop. 2014 10.1155/2014/628695 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Haimes JL, Wroble RR, Grood ES, Noyes FR, 2007. Role of the Medial Structures in the intact and Anterior Cruciate Ligament-Deficient Knee. Am. J. Sports Med 22, 402–409. 10.1177/036354659402200317 [DOI] [PubMed] [Google Scholar]
- Harris WH, Sledge CB, 1994. Total Hip and Total Knee Replacement. N. Engl. J. Med 330, 1776–1781. 10.1056/NEJM199009133231106 [DOI] [PubMed] [Google Scholar]
- Hawker G, Wright J, Coyte P, Paul J, Dittus R, Croxford R, Katz B, Bombardier C, Heck D, Freund D, 1998. Health-related quality of life after knee replacement. J. Bone Joint Surg. Am 80, 163–73. 10.2106/00004623-199802000-00003 [DOI] [PubMed] [Google Scholar]
- Laskin R, Beksac B, 2004. Stiffness after total knee arthroplasty. J. Arthroplasty 19, 41–46. 10.1016/j.arth.2004.02.008 [DOI] [PubMed] [Google Scholar]
- Lim HC, Bae JH, Bae TS, Moon BC, Shyam AK, Wang JH, 2012. Relative role changing of lateral collateral ligament on the posterolateral rotatory instability according to the knee flexion angles: A biomechanical comparative study of role of lateral collateral ligament and popliteofibular ligament. Arch. Orthop. Trauma Surg 132, 1631–1636. 10.1007/s00402-012-1591-7 [DOI] [PubMed] [Google Scholar]
- Mancini V, Tommasin D, Li Y, Reeves J, Baets R, Greenwald S, Segers P, 2019. Detecting carotid stenosis from skin vibrations using Laser Doppler Vibrometry – An in vitro proof-of-concept. PLoS One 14, 1–17. 10.1371/journal.pone.0218317 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martin JA, Brandon SCE, Keuler EM, Hermus JR, Ehlers AC, Segalman DJ, Allen MS, Thelen DG, 2018. Gauging force by tapping tendons. Nat. Commun 9, 2–10. 10.1038/s41467-018-03797-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martin JA, Schmitz DG, Ehlers AC, Allen MS, Thelen DG, 2019. Calibration of the shear wave speed-stress relationship in ex vivo tendons. J. Biomech 90, 9–15. 10.1016/j.jbiomech.2019.04.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Matsumoto T, Muratsu H, Kubo S, Matsushita T, Kurosaka M, Kuroda R, 2011. Soft Tissue Tension in Cruciate-Retaining and Posterior-Stabilized Total Knee Arthroplasty. J. Arthroplasty 26, 788–795. 10.1016/j.arth.2010.06.006 [DOI] [PubMed] [Google Scholar]
- Novacheck TF, Gage JR, 2007. Orthopedic management of spasticity in cerebral palsy. Childs Nerv. Syst 23, 1015–1031. 10.1007/s00381-007-0378-6 [DOI] [PubMed] [Google Scholar]
- Nowakowski A, Majewski M, Muller-Gerbl M, Valderrabano V, 2012. Measurement of knee joint gaps without bone resection: “physiologic” extension and flexion gaps in total knee arthroplasty are asymmetric and unequal and anterior and posterior cruciate ligament resections produce different gap changes. J. Orthop. Res 30, 522–527. 10.1002/jor.21564 [DOI] [PubMed] [Google Scholar]
- Quapp KM, Weiss JA, 1998. Material Characterization of Human Medial Collateral Ligament. J. Biomech. Eng 120, 757 10.1115/1.2834890 [DOI] [PubMed] [Google Scholar]
- Robinson JR, Sanchez-Ballester J, Bull AMJ, Thomas, R. de WM, Amis AA, 2004. The posteromedial corner revisited. J. Bone Joint Surg. Br 86-B, 674–681. 10.1302/0301-620x.86b5.14853 [DOI] [PubMed] [Google Scholar]
- Savin DD, Watson J, Youderian AR, Lee S, Hammarstedt JE, Hutchinson MR, Goldberg BA, 2017. Surgical management of acute distal biceps tendon ruptures. J. Bone Jt. Surg. - Am Vol. 99, 785–796. 10.2106/JBJS.17.00080 [DOI] [PubMed] [Google Scholar]
- Schatzmann L, Brunner P, Stäubli HU, 1998. Effect of cyclic preconditioning on the tensile properties of human quadriceps tendons and patellar ligaments. Knee Surgery, Sport. Traumatol. Arthrosc 6, 56–61. 10.1007/s001670050224 [DOI] [PubMed] [Google Scholar]
- Schwartz MH, Viehweger E, Stout J, Novacheck TF, Gage JR, 2004. Comprehensive Treatment of Ambulatory Children with Cerebral Palsy: An Outcome Assessment. J. Pediatr. Orthop 24, 45–53. 10.1097/01241398-200401000-00009 [DOI] [PubMed] [Google Scholar]
- Seering WP, Piziali RL, Nagel DA, Schurman DJ, 1980. The function of the primary ligaments of the knee in varus-valgus and axial rotation. J. Biomech 13, 785–794. 10.1016/0021-9290(80)90240-7 [DOI] [PubMed] [Google Scholar]
- Tanaka E, Inubushi T, Takahashi K, Shirakura M, Sano R, Dalla-Bona DA, Nakajima A, van Eijden TMGJ, Tanne K, 2007. Dynamic shear properties of the porcine molar periodontal ligament. J. Biomech 40, 1477–1483. 10.1016/j.jbiomech.2006.06.022 [DOI] [PubMed] [Google Scholar]
- Vince K, 2016. Mid-flexion instabiity after total knee arthroplasty: woolly thinking or a real concern? Bone Jt. J 98-B, 84–88. 10.1302/0301-620X.98B1.36445 [DOI] [PubMed] [Google Scholar]
- Vince K, Abdeen A, Sugimori T, 2006. The unstable total knee arthroplasty: causes and cures. J. Arthroplasty 21, 44–49. 10.1016/j.arth.2006.02.101 [DOI] [PubMed] [Google Scholar]
- Weiss JA, Gardiner JC, Bonifasi-Lista C, 2002. Ligament material behavior is nonlinear, viscoelastic and rate-independent under shear loading. J. Biomech 35, 943–950. 10.1016/S0021-9290(02)00041-6 [DOI] [PubMed] [Google Scholar]
- Wilson DR, Feikes JD, O’Connor JJ, 1998. Ligaments and articular contact guide passive knee flexion. J. Biomech 31, 1127–1136. 10.1016/S0021-9290(98)00119-5 [DOI] [PubMed] [Google Scholar]
- Wilson WT, Deakin MAH, Payne AP, Picard F, Wearing SC, 2012. Comparative Analysis of the Structural Properties of the Collateral Ligaments of the Human Knee. J. Orthop. Sports Phys. Ther 42, 345–351. 10.2519/jospt.2012.3919 [DOI] [PubMed] [Google Scholar]
- Xerogeanes JW, Fox RJ, Takeda Y, Kim HS, Ishebashi Y, Carlin GJ, Woo SLY, 1998. A Functional comparison of animal anterior cruciate ligament models to the human anterior cruciate ligament. Ann. Biomed. Eng 26, 345–352. 10.1114/1.91 [DOI] [PubMed] [Google Scholar]