Abstract
Background
Data on the usefulness of cardiovascular magnetic resonance (CMR) imaging for clinical decision making in patients with an implanted cardioverter defibrillator (ICD) are scarce. The present study determined the impact of CMR imaging on diagnostic stratification and treatment decisions in ICD patients presenting with electrical instability or progressive heart failure symptoms.
Methods
212 consecutive ICD patients underwent 1.5 T CMR combining diagnostic imaging modules tailored to the individual clinical indication (ventricular function assessment, myocardial tissue characterization, adenosine stress-perfusion, 3D-contrast-enhanced angiography); four CMR examinations (4/212, 2%) were excluded due to non-diagnostic CMR image quality. The resultant change in diagnosis or clinical management was determined in the overall population and compared between ICD patients for primary (115/208, 55%) or secondary prevention (93/208, 45%). Referral indication consisted of documented ventricular tachycardia, inadequate device therapy or progressive heart failure symptoms.
Results
Overall, CMR imaging data changed diagnosis in 40% (83/208) with a significant difference between primary versus secondary prevention ICD patients (37/115, 32% versus 46/93, 49%, respectively; p = 0.01). The information gain from CMR led to an overall change in treatment in 21% (43/208) with a similar distribution in primary versus secondary prevention ICD patients (25/115,22% versus 18/93,19%, p = 0.67). The effect on treatment change was highest in patients initially scheduled for ventricular tachycardia ablation procedure (18/141, 13%) with revision of the treatment plan to medical therapy or coronary revascularization.
Conclusions
CMR imaging in ICD patients presenting with electrical instability or worsening heart failure symptoms provided diagnostic or management-changing information in a considerable proportion (40% and 21%, respectively).
Keywords: Cardiovascular magnetic resonance imaging, Implanted cardioverter defibrillator, Diagnostic reclassification, Ventricular tachycardia ablation, Treatment decision
Background
Over the past decade, increasing evidence has been accumulated regarding the safety of magnetic resonance (MR) imaging in patients with cardiovascular implantable electronic devices [1–3]. In general, MR scanning has been found to be safe when conducted under careful supervision in both MR conditional and conventional active rhythm devices [4]. However, more recent research addressed the pivotal issue of achieving adequately high diagnostic image quality of cardiovascular magnetic resonance (CMR) imaging in patients with implanted cardioverter defibrillator (ICD) and utilized well-established spoiled-gradient echo cine sequences following gadolinium-containing contrast agent application in combination with wideband late-gadolinium enhancement (LGE) imaging. Thus, the fundamental diagnostic CMR modules of basic ventricular functional assessment and myocardial tissue characterization have eventually become available for routine clinical usage in ICD patients either [5–8]. While previous investigations have largely focused on safety and image quality, the clinical impact and usefulness of CMR imaging in ICD patients have not been evaluated yet. Hence, the current study investigated the impact of CMR imaging on diagnostic stratification and treatment decisions in ICD patients presenting with electrical instability or progressive heart failure symptoms.
Methods
Patient population
The study was conducted in accordance with the local institutional review board and the standards of the University of Leipzig ethics committee. Written informed consent was obtained from all patients prior to CMR examination. All patients with an ICD referred for clinical management of electrical instability and/or worsening heart failure were seen by the cardiologist/electrophysiologist in charge and the decision to undergo CMR imaging was at the discretion of the clinician. Consecutive patients referred for CMR imaging with an ICD between June 2015 and December 2018 formed the study population. The CMR examination rigorously followed a previously published, standardized protocol [5]. All patients underwent chest x–ray prior to the CMR examination and in the presence of abandoned, epicardial or fractured leads, CMR imaging was not performed (i.e. patients were not considered for study inclusion). In addition, the time period between device implantation and CMR examination was > 6 weeks.
Device interrogation and programming
Directly prior to the CMR examination, an experienced electrophysiologist carried out a thorough ICD device interrogation in the scanner console room : battery status and sensing/pacing thresholds of all leads were documented and the device memory was evaluated for events (e.g. appropriately or inappropriately classified arrhythmias; date, number and kind of delivered antitachycardia pacing therapies etc.). Subsequently, devices were programmed for the CMR scanner environment: all tachyarrhythmia functions were switched off; MR–conditional labeled devices were programmed into the manufacturer–provided MR–safe mode settings; non–MR–conditional devices were programmed to pacing off, sensing–only (ODO or OVO) or to asynchronous pacing (VOO) depending on the intrinsic rhythm of the patient. The electrophysiologist and the programming device were present in the scanner console room throughout the entire CMR examination. Directly after completion of the CMR examination, patients underwent a repeat ICD interrogation in the console room and all devices were restored to their original settings.
CMR imaging protocol
A cardiologist with a high level of expertise in CMR imaging (I.P. or C.J., both > 20 years of experience and Society for Cardiovascular Magnetic Resonance and European Society of Cardiology level 3 trained) was present throughout the entire procedure for online evaluation of all CMR images and to guide or adapt a previously established CMR workflow/imaging strategy for device patients if needed to fully address the clinical indication [5]. Briefly, CMR imaging protocols were tailored to the clinical indication and diagnostic imaging modules were combined accordingly (i.e. ventricular function assessment, myocardial tissue characterization using wideband LGE imaging plus/minus T1- or T2-weighted blackblood turbo spin echo sequences, myocardial perfusion assessment using dynamic first-pass perfusion spoiled-gradient echo imaging during adenosine stress and conventional contrast-enhanced three-dimensional angiography scans). Cine imaging was performed using spoiled gradient echo (SGE) sequences with short–axis geometries being acquired before and long–axis geometries after contrast agent application. For LGE imaging, the wideband technique was employed (bandwidth of the inversion prepulse, 3.000 Hz; frequency offset, + 1.000 Hz) [6]. Both, cine and LGE imaging sequences were acquired during end-expiratory breath-holding. All CMR examinations were carried out on a 1.5 T CMR scanner (Ingenia, Philips Healthcare, Best, The Netherlands) using a 28–element array coil with full in–coil signal digitalization and optical transmission. Following current recommendations, whole–body specific absorption rate (SAR) was restricted to 2 watts per kilogram bodyweight. Patients were continuously monitored throughout the procedure based on vector–surface electrocardiogram (ECG), peripheral pulse oximetry, respiratory motion pattern, non–invasive blood pressure measurements as well as continuous visual and voice contact. If online evaluation of initial short–axis cine images demonstrated ≥12 non–evaluable left–ventricular (LV) segments due to severe generator–related artifacts, the CMR examination was considered non–diagnostic and terminated.
CMR image quality
Quality of CMR images (i.e. cine, LGE, T1- and T2-weighted blackblood turbo-spin echo, dynamic perfusion and contrast-enhanced three-dimensional angiographic image sequences) was evaluated offline > 6 weeks later by two experienced CMR imaging experts (I.P., C.J.) in a consensus read; device–related image artifacts were assessed per myocardial segment of the LV and right ventricle (RV) following previously defined classification systems [9, 10]. Myocardial segments were considered evaluable if artifact–free in at least one standard cardiac geometry.
CMR–based change in diagnosis or treatment
In all patients, clinical data were recorded at the time of the CMR examination and the indication for CMR referral was documented: electrical instability was defined as electrical storm (i.e. ≥ 3 separate arrhythmia episodes leading to adequate ICD therapies consisting of anti-tachycardia pacing (ATP) or shocks occurring within a 24-h period), ≤ 2 adequate ATP/shock(s), or sustained ventricular tachycardia (VT) below detection; progressive heart failure was defined as any increase in New York Heart Association stage of at least one functional class during the last three months. The CMR imaging results were determined by the consensus of two experienced CMR imaging experts (I.P., C.J.) within < 1 h following the CMR examination and the CMR–based diagnosis was documented. CMR–based change in diagnosis was defined as a new, previously not considered diagnosis (i.e. by clinical history, other cardiac imaging modalities etc.). CMR–based change in treatment decision constituted the sole responsibility of the electrophysiologist/cardiologist in charge and was defined as change in management consisting of any deviation from the referral to invasive procedures (e.g. electrophysiology study/ablation procedure, cardiac surgery, coronary revascularization etc.) or a change in medical therapy.
Statistical analysis
All analyses were done using SPSS (version 21, International Business Machines, Inc., Armonk, New York, USA). Continuous variables were given as mean ± standard deviation for normally distributed data; frequencies and percentages were used to describe categorical data. Differences between continuous and categorical variables were assessed using Student’s t–test and Chi–square test as appropriate. All tests were two–tailed and a p–value of < 0.05 was considered significant.
Results
Patient characteristics
CMR examinations were carried out in 212 consecutive ICD patients; for patients who underwent repeat/follow-up examinations, only the first CMR dataset was considered for analysis. Four CMR examinations (4/212, 2%) were excluded due to non-diagnostic CMR image quality. Hence, the final dataset for analysis comprised of 208 CMR examinations of ICD patients. Data analysis was conducted for all patients and for the subgroups of primary versus secondary prevention ICD patients (n = 115 versus n = 93, respectively). Overall, 94 patients were equipped with an MR-conditional device (94/208, 45%; see Table 1); patients with MR-conditional devices were younger (58 ± 12 versus 61 ± 12 years; p = 0.023) and had a shorter mean time since ICD implantation (48 ± 43 versus 68 ± 54 months; p = 0.003). Referral indication consisted of documented VT (141/208, 68%), inadequate device therapy (15/208, 7%) or progressive heart failure symptoms; detailed patient characteristics are given in Table 1. In all ICD patients, device interrogation before and after the CMR examination yielded no significant changes (mean change of battery status 0.0 ± 0.1 V; lead threshold, 0.0 ± 0.3 V; lead impedance, − 7.5 ± 52.7 Ω; for all p = ns).
Table 1.
All patients (n = 208) | Primary prevention (n = 115) | Secondary prevention (n = 93) | p–value | |
---|---|---|---|---|
Age, years | 60 ± 12 | 60 ± 11 | 59 ± 13 | 0.532 |
Gender, male, n(%) | 174 (84) | 98 (85) | 76 (82) | 0.498 |
BMI, kg/m2 | 28.6 ± 5.1 | 29.2 ± 5.3 | 27.9 ± 4.8 | 0.062 |
Hypertension, n(%) | 144 (69) | 80 (70) | 64 (69) | 0.907 |
Diabetes, n(%) | 46 (22) | 32 (28) | 14 (15) | 0.027 |
CHA2DS2–VASc Score | 2.7 ± 1.4 | 2.8 ± 1.2 | 2.4 ± 1.5 | 0.006 |
NYHA class | 1.9 ± 0.9 | 2.1 ± 0.9 | 1.6 ± 1.0 | 0.001 |
Known CAD, n(%) | 83 (40) | 52 (45) | 31 (33) | 0.082 |
Severity of CAD | 0.178 | |||
Single–vessel, n(%) | 31 (37) | 18 (35) | 13 (42) | |
Multi–vessel, n(%) | 52 (63) | 34 (65) | 18 (58) | |
Electrical instability | 0.009 | |||
≤ 2 adequate ATP/shock(s), n(%) | 23 (11) | 16 (14) | 7 (8) | |
Electrical storm, n(%) | 107 (51) | 53 (46) | 54 (58) | |
Below detection, n(%) | 11 (5) | 3 (3) | 8 (9) | |
Inadequate ATP/shock(s), n(%) | 15 (7) | 6 (5) | 9 (10) | |
Prior VT ablation, n(%) | 50 (24) | 16 (14) | 34 (37) | < 0.001 |
Antiarrhythmic drugs | ||||
Betablockers, n(%) | 189 (91) | 104 (90) | 85 (91) | 0.811 |
Class I, n(%) | 14 (7) | 5 (4) | 9 (10) | 0.127 |
Class III, n(%) | 60 (29) | 31 (27) | 29 (31) | 0.504 |
Device characteristics | 0.070 | |||
S–ICD, n(%) | 6 (3) | 5 (4) | 1 (1) | |
Single chamber ICD, n(%) | 101 (49) | 55 (48) | 46 (49) | |
Dual chamber ICD, n(%) | 65 (31) | 30 (26) | 35 (38) | |
CRT–D, n(%) | 36 (17) | 25 (22) | 11 (12) | |
MR-conditional devices, n(%) | 94 (45) | 48 (42) | 46 (49) | 0.266 |
Time since ICD implantation, months | 58.9 ± 50.0 | 56.4 ± 42.4 | 61.9 ± 58.2 | 0.436 |
Range, months | 1.4–240.4 | 1.6–225.0 | 1.4–240.4 |
Values are mean ± SD or n (%); p-values are given for the comparison of primary versus secondary prevention. BMI body mass index; CAD coronary artery disease; ATP antitachycardia pacing; VT ventricular tachycardia; ICD implanted cardioverter defibrillator; S–ICD subcutaneous implantable cardioverter defibrillator; CRT–D cardiac resynchronization therapy defibrillator, MR magnetic resonance
CMR imaging protocol
The following CMR diagnostic imaging modules were employed: cine imaging for ventricular function assessment (n = 208), myocardial tissue characterization (LGE, n = 206 + / - T1– or T2–weighted blackblood spin echo imaging, n = 49 or n = 55, respectively), resting or adenosine stress perfusion (n = 27) and contrast–enhanced 3D–angiography (pulmonary veins, n = 25 or thoracic aorta, n = 3). Results of CMR data are given in Table 2.
Table 2.
All patients (n = 208) | Primary prevention (n = 115) | Secondary prevention (n = 93) | p–value | |
---|---|---|---|---|
Heart rate, bpm | 68 ± 14 | 70 ± 14 | 64 ± 13 | 0.002 |
Systolic BP, mmHg | 127 ± 20 | 124 ± 20 | 130 ± 21 | 0.010 |
Diastolic BP, mmHg | 73 ± 12 | 73 ± 12 | 74 ± 12 | 0.888 |
LVEDV, ml | 249 ± 106 | 283 ± 111 | 207 ± 83 | < 0.001 |
LVESV, ml | 170 ± 102 | 207 ± 104 | 125 ± 78 | < 0.001 |
LVEF, % | 36 ± 15 | 30 ± 13 | 44 ± 13 | < 0.001 |
RVEDV, ml | 128 ± 65 | 133 ± 71 | 121 ± 55 | 0.163 |
RVESV, ml | 62 ± 48 | 66 ± 53 | 57 ± 40 | 0.139 |
RVEF, % | 54 ± 11 | 54 ± 12 | 56 ± 10 | 0.186 |
LGE positive (n = 206), n(%) | 135 (66) | 73 (65) | 62 (67) | 0.756 |
LV, LGE positive, n(%) | 132 (64) | 73 (65) | 59 (63) | 0.863 |
LV-LGE pattern (n = 132) | 0.862 | |||
subendocardial, n(%) | 82 (62) | 47 (64) | 35 (59) | |
subepicardial, n(%) | 14 (11) | 7 (10) | 7 (12) | |
midwall striae, n(%) | 23 (17) | 13 (18) | 10 (17) | |
midwall patchy, n(%) | 13 (10) | 6 (8) | 7 (12) | |
LV-LGE, segments | 4.1 ± 2.7 | 4.2 ± 2.8 | 4.0 ± 2.5 | 0.647 |
LV-LGE, % transmurality | 73 ± 25 | 75 ± 24 | 70 ± 26 | 0.243 |
RV, LGE positive, n(%) | 9 (4) | 2 (2) | 7 (8) | 0.044 |
LV thrombus, n(%) | 9 (4) | 5 (4) | 4 (4) | 0.987 |
Values are mean ± SD or n (%); p-values are given for the comparison of primary versus secondary prevention. BP blood pressure; LVEDV left ventricular end–diastolic volume; LVESV left ventricular end–systolic volume; LVEF left ventricular ejection fraction; LGE late gadolinium enhancement; RV right ventricle; LV LGE pattern was categorized according to reference [11]
CMR image quality
Overall, post–contrast cine imaging allowed for evaluation of 96.1 ± 6.1% of LV and 95.6 ± 6.6% of RV segments. For LGE imaging, LV and RV segmental evaluability amounted to 94.2 ± 7.8% and 95.8 ± 6.9%, respectively; T1- and T2 imaging resulted in 94.9 ± 7.2% and 91.3 ± 8.5% evaluable LV segments, respectively; perfusion imaging allowed for evaluation of 98.8 ± 2.2% of LV myocardial segments. In general, artifact occurrence was most common in the anterior LV segments; representative CMR imaging examples are provided in Figs. 1, 2, 3 and 4 and additional files 1–3 (video format: .mp4), respectively. Image quality of contrast-enhanced three-dimensional angiography of the pulmonary veins or thoracic aorta was unimpaired in all cases (100% evaluability).
CMR–based change in diagnosis
Overall, CMR imaging data resulted in a change in diagnosis in 40% (83/208) of ICD patients with a significant difference between primary versus secondary prevention ICD patients (37/115, 32% versus 46/93, 49%; p = 0.01; see Table 3 and Fig. 5); based on CMR imaging data, the percentages reclassified mainly affected the following diagnostic categories: hypertensive heart disease (22/83, 27%), post-acute myocarditis (13/83, 16%), dilated cardiomyopathy (13/83, 16%) and normal (10/83, 12%). In addition to mere reclassification of underlying cardiac disease, newly discovered and important diagnostic findings consisted of detection of LV thrombus (9/208, 4%) and identification of LGE-defined fibrosis in non-ischemic cardiomyopathy (75/208, 36%) representing ventricular substrate as potential target for ablation therapy. Finally, for the group comparison of “electrical instability” versus “worsening heart failure” no significant difference for “change in diagnosis” was found (59/156, 38% versus 24/52, 46%, respectively; p = 0.288).
Table 3.
Post-CMR diagnosis | All patients (n = 208) | All patients Change in diagnosis | Primary prevention (n = 115) | Primary prevention Change in diagnosis | Secondary prevention (n = 93) | Secondary prevention Change in diagnosis |
---|---|---|---|---|---|---|
Acute MI | 2 | 2 (1) | 1 | 1 (1) | 1 | 1 (1) |
Cardiac storage disease | 1 | 0 (0) | 1 | 0 (0) | 0 | 0 (0) |
ARVC | 5 | 3 (1) | 1 | 0 (0) | 4 | 3 (3) |
Idiopathic DCM | 43 | 13 (6) | 33 | 9 (8) | 10 | 4 (4) |
HCM | 7 | 2 (1) | 3 | 1 (1) | 4 | 1 (1) |
HHD | 25 | 22 (11) | 11 | 11 (10) | 14 | 11 (12) |
ICM | 63 | 8 (4) | 40 | 2 (2) | 23 | 6 (6) |
NCCM | 1 | 1 (0.5) | 0 | 0 (0) | 1 | 1 (1) |
Normal (prior LGE (+) VT ablation lesions) | 4 | 4 (2) | 1 | 1 (1) | 3 | 3 (3) |
Normal | 21 | 10 (5) | 5 | 4 (3) | 16 | 6 (6) |
Post–acute myocarditis | 20 | 13 (6) | 9 | 5 (4) | 11 | 8 (9) |
PPCM | 2 | 0 (0) | 2 | 0 (0) | 0 | 0 (0) |
Cardiac sarcoidosis | 7 | 4 (2) | 4 | 2 (2) | 3 | 2 (2) |
Systemic sclerosis | 1 | 0 (0) | 0 | 0 (0) | 1 | 0 (0) |
TOF | 1 | 0 (0) | 0 | 0 (0) | 1 | 0 (0) |
VHD | 5 | 1 (0.5) | 4 | 1 (1) | 1 | 0 (0) |
Total | 208 | 83 (40) | 115 | 37 (32) | 93 | 46 (49)* |
Values are n (%) for all, primary and secondary prevention groups. MI myocardial infarction; ARVC arrhythmogenic right-ventricular cardiomyopathy; DCM dilative cardiomyopathy; HCM hypertrophic cardiomyopathy; HHD hypertensive heart disease; ICM ischemic cardiomyopathy; NCCM non-compaction cardiomyopathy; LGE late gadolinium enhancement; VT ventricular tachycardia; PPCM peripartum cardiomyopathy; TOF tetralogy of Fallot; VHD valvular heart disease. * p = 0.011 for the comparison of primary versus secondary prevention
CMR–based change in treatment
The information gained from CMR imaging resulted in an overall change in treatment in 21% (43/208) with a similar distribution in primary versus secondary prevention ICD patients (25/115, 22% versus 18/93, 19%, p = 0.67; see Table 4 and Fig. 5). The main effect of the CMR results on treatment change can be gauged when focusing on patients referred for an invasive treatment: after CMR, a total of 123 of 141 patients initially scheduled for a VT ablation procedure subsequently underwent VT ablation while in the remaining 18 patients (13%) the treatment categories changed to medical therapy (n = 14) or coronary revascularization (n = 4). In more detail, the CMR findings giving rise to deferral of VT ablation were evidence of myocardial ischemia (n = 4), acute myocardial inflammation/cardiac sarcoidosis (n = 4), newly detected LV thrombus formation (n = 5), absence of CMR-identifiable structural heart disease and, thus, subsequent opportunity to initiate class IC antiarrhythmic drug therapy (n = 2), extreme LV wall thinning after transmural ischemic myocardial infarction (< 2 mm, n = 1), newly diagnosed arrhythmogenic RV dysplasia (end-stage disease with extensive RV fibrosis formation, n = 1) and in one patient the presence of CMR evidence of substantial pericardial fibrotic tissue formation/epicardial substrate after coronary artery bypass grafting who was deemed not suitable for an epicardial approach. Likewise, in patients initially considered for coronary revascularization (n = 14), the results of CMR imaging led to the following treatment strategies: coronary revascularization (n = 5), medical therapy (n = 8) and pulmonary venous isolation procedure (n = 1, negative adenosine stress CMR ruled out progression of known CAD and atrial fibrillation was deemed to represent the cause of progressive dyspnea).
Table 4.
Post-CMR Treatment | All patients (n = 208) | All patients Change in treatment | Primary prevention (n = 115) | Primary prevention Change in treatment | Secondary prevention (n = 93) | Secondary prevention Change in treatment |
---|---|---|---|---|---|---|
Coronary Revasc | 10 | 10 (5) | 5 | 5 (4) | 5 | 5 (5) |
CRT upgrade | 3 | 3 (1) | 2 | 2 (2) | 1 | 1 (1) |
LVAD | 2 | 1 (0.5) | 2 | 1 (1) | 0 | 0 (0) |
Medical therapy | 56 | 27 (13) | 32 | 15 (13) | 24 | 12 (13) |
PVI | 13 | 1 (0.5) | 8 | 1 (1) | 5 | 0 (0) |
PV angioplasty | 1 | 1 (0.5) | 1 | 1 (1) | 0 | 0 (0) |
VT ablation | 123 | 0 (0) | 65 | 0 (0) | 58 | 0 (0) |
Total | 208 | 43 (21) | 115 | 25 (22) | 93 | 18 (19)* |
Values are n (%) for all, primary and secondary prevention groups. Coronary Revasc, coronary revascularization procedure; CRT cardiac resynchronization therapy LVAD left ventricular assist device; PVI pulmonary venous isolation procedure; PV pulmonary vein; VT ventricular tachycardia. * p = 0.673 for the comparison of primary versus secondary prevention
In addition, while n = 2 patients initially scheduled for device upgrade to CRT were not considered for this treatment after CMR detected extensive inferolateral infarction/scar tissue (thereby rendering CRT ineffective), another n = 3 patients were newly stratified for device upgrade based on accurate, CMR-derived volumetric determination of low systolic LV ejection fraction which had been rather overestimated by prior transthoracic echocardiography [12]. For the group comparison of “electrical instability” versus “worsening heart failure”, a statistically significant difference of “change in treatment” was noted (25/156, 16% versus 18/52, 35%, respectively; p = 0.004).
Discussion
The main findings of the current CMR study in ICD patients presenting with electrical instability or worsening heart failure symptoms were as follows: (1) taking advantage of a standardized CMR protocol, a high proportion of diagnostic examinations (98%) with a consistently high ventricular segmental evaluability of cine and LGE imaging modules (94 to 96%) was achieved, (2) CMR testing resulted in a change in diagnosis in 40% of the overall population with a significant difference between primary and secondary prevention ICD patients (32% versus 49%, respectively), (3) CMR testing caused a change in treatment decisions in 21% of patients with no significant difference between primary and secondary prevention ICD patients (22 versus 19%, respectively).
There may be several reasons for the relatively high proportion of change in diagnosis particularly found in secondary prevention ICD patients: while in accordance with current recommendations, the common practice to perform echocardiography and invasive coronary angiography (i.e. assessment of ventricular dimensions/function and exclusion of significant CAD) as the sole diagnostic work-up prior to ICD implantation may not be sufficient to cover the whole range of myocardial tissue alterations possibly being detectable at the time of implantation. In particular, cardiac diseases mimicking LV hypertrophy (sarcoidosis, storage diseases etc.) may be misclassified as hypertensive heart disease in the presence of concomitant arterial hypertension and, thus, may go undetected (see Fig. 2, additional file 2). Notwithstanding, current guidelines regarding non-invasive diagnostic testing in survivors of sudden cardiac death defined the recommendation level for CMR imaging as “moderate” only (IIa C-EO, i.e. the diagnostic test may be useful/beneficial mainly based on clinical experience and/or expert opinion only) [13].
Study data in smaller populations suggested that the addition of CMR testing to the diagnostic work-up (i.e. “baseline” CMR imaging prior to ICD implantation) in patients with documented VT/survivors of sudden cardiac death yielded a high incremental diagnostic value with a resultant reassignment of about 50% of patients to a new or alternate diagnosis [14]. It must be noted, however, that baseline CMR imaging alone may not suffice to fully address the diagnostic dilemma: from a pathophysiological point of view, the cells of the intrinsic electrical conductance system are generally more susceptible to inflammation than myocardial cells [15, 16]. Hence, “indicator” arrhythmias (e.g. higher-grade AV-block, ventricular arrhythmias) may occur in earlier inflammatory disease stages even before any of the structural myocardial cell alterations which are detectable by CMR imaging (= inherent limit of detection of CMR imaging). However, for the time being ICD implantation constitutes the central component of treatment for patients with ischemic or non-ischemic cardiomyopathy, who are at risk for sudden cardiac death due to VT. Consequently, in consideration of the natural progression of myocardial diseases and/or fluctuating disease stages, diagnostic re-evaluation of ICD patients should be strongly considered, especially if life threatening arrhythmias or worsening heart failure symptoms reoccur over time. Previous research reported on the added value of CMR imaging in patients with MR–conditional pacemakers only which resulted in diagnostic or management-changing information in 63% of the patient cohort [17], while the present study data further corroborated the high impact of CMR-based information on clinical decision making in ICD patients.
The high diagnostic performance of CMR-based myocardial tissue characterization offers the unique ability to adequately monitor cardiac disease progression in the individual patient and in the current study, the information gain from CMR testing proved useful to re-route clinical pathways: clinical management/treatment decisions were changed in a high proportion of patients (21%). Most importantly, in the non-ischemic cardiomyopathy group LGE-defined ventricular substrate as a potential target for subsequent VT ablation was newly detected in 36% of ICD patients; since substrate-guided (“fibrosis-guided”) ablation has become a cornerstone for catheter treatment of complex ventricular arrhythmias such CMR-based information is of great clinical usefulness to determine overall suitability of the patient for ventricular ablation (i.e. expected complexity of the ablation procedure), the preferred route of access (i.e. endocardial or epicardial approach only or combined endo−/epicardial approach) and to facilitate a more targeted ablation with knowledge about the segmental location of potential substrate beforehand [18–20]. However, in the current study changes in access route/ablation strategy were not considered a CMR-based change in treatment but rather a readily available and complimentary information gain for the interventional electrophysiologist.
Study limitations
While the study cohort closely reflected the clinical spectrum typically encountered at a highly specialized, tertiary care center, the data as such may only be applicable to a similar clinical scenario. Some referral bias may be present in the study population, since the decision to request CMR imaging has entirely been left to the referring electrophysiologist/cardiologist and critically-ill patients were generally not considered for a CMR study.
Importantly, while at our institution CMR imaging in device carriers is routinely carried out with a high degree of CMR imaging expertise (> 20 years of experience) following a previously established workflow taking advantage of more recent CMR sequence developments and with adherence to current guideline recommendations, local expertise and/or scanner availability may vary considerably and must be taken into account when implementing CMR imaging of ICD patients as an integral component of the diagnostic workup [5, 6, 21].
On a per patient level, a 2% rate of “non-diagnostic” studies was determined (i.e. studies for which there was no interpretable CMR imaging data) and this may be considered the total rate of non-usable study information based on the proposed CMR imaging protocol in ICD patients; in case CMR imaging was carried out successfully, the rates of non-diagnostic segmental evaluability of regional wall motion and wideband LGE were 4 and 6%, respectively.
Conclusions
The present study data demonstrated the high impact of CMR imaging regarding diagnosis and clinical management in ICD patients presenting with electrical instability or worsening heart failure symptoms: CMR-based findings provided diagnostic and treatment changing information in a considerable proportion of ICD patients (40 and 21%, respectively). Hence, this work further contributed to the growing information base for CMR testing in its evolving role as an arbiter for profound clinical decision making and subsequent rerouting of clinical pathways in ICD patients either.
Supplementary information
Acknowledgements
Not applicable.
Authors’ information (optional)
Not applicable.
Abbreviations
- ATP
Antitachycardia pacing
- CAD
Coronary artery disease
- CMR
Cardiovascular magnetic resonance
- CRT
Cardiac resynchronization therapy
- ECG
Electrocardiogram
- ICD
Implanted cardioverter defibrillator
- LGE
Late-gadolinium enhancement
- LV
Left ventricle/left ventricular
- MR
Magnetic resonance
- RV
Right ventricle/right ventricular
- SAR
Specific absorption rate
- SGE
Spoiled gradient echo
- VT
Ventricular tachycardia
Authors’ contributions
FL, SO, IP, AA, ND, SR and BD assisted with the design and the set-up of the study, collected and analysed data and drafted the manuscript. AB and GH contributed to the design of the study and the supervision of the work and assisted with critical revision of the manuscript. IP, CJ, SH, SL, MD and SR performed the cardiovascular magnetic resonance examinations and critically revised the manuscript. CS assisted in the collection of patient data and critically revised the manuscript. CJ designed the study, supervised the work, assisted with drafting and critical revision of the manuscript. All authors read and approved the final manuscript.
Funding
Not applicable.
Availability of data and materials
Not applicable.
Ethics approval and consent to participate
Ethics approval was obtained by the local ethics committee of the faculty of medicine of the University of Leipzig. For each subject written informed consent was obtained.
Consent for publication
All subjects gave written consent to participate and to publish.
Competing interests
Not applicable.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Frank Lindemann and Sabrina Oebel contributed equally to this work and are joint first authors.
Supplementary information
Supplementary information accompanies this paper at 10.1186/s12968-020-00609-z.
References
- 1.Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376(8):755–764. doi: 10.1056/NEJMoa1603265. [DOI] [PubMed] [Google Scholar]
- 2.Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, et al. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011;155(7):415–424. doi: 10.7326/0003-4819-155-7-201110040-00004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Nazarian S, Hansford R, Rahsepar AA, Weltin V, McVeigh D, Gucuk Ipek E, et al. Safety of magnetic resonance imaging in patients with cardiac devices. N Engl J Med. 2017;377(26):2555–2564. doi: 10.1056/NEJMoa1604267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H, et al. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla. Circulation. 2006;114(12):1277–1284. doi: 10.1161/CIRCULATIONAHA.105.607655. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Hilbert S, Jahnke C, Loebe S, Oebel S, Weber A, Spampinato R, et al. Cardiovascular magnetic resonance imaging in patients with cardiac implantable electronic devices: a device-dependent imaging strategy for improved image quality. Eur Heart J Cardiovasc Imaging. 2018;19(9):1051–1061. doi: 10.1093/ehjci/jex243. [DOI] [PubMed] [Google Scholar]
- 6.Hilbert S, Weber A, Nehrke K, Bornert P, Schnackenburg B, Oebel S, et al. Artefact-free late gadolinium enhancement imaging in patients with implanted cardiac devices using a modified broadband sequence: current strategies and results from a real-world patient cohort. Europace. 2018;20(5):801–807. doi: 10.1093/europace/eux016. [DOI] [PubMed] [Google Scholar]
- 7.Schwitter J, Gold MR, Al Fagih A, Lee S, Peterson M, Ciuffo A, et al. Image quality of cardiac magnetic resonance imaging in patients with an implantable cardioverter defibrillator system designed for the magnetic resonance imaging environment. Circ Cardiovasc Imaging. 2016;9(5). [DOI] [PubMed]
- 8.Stevens SM, Tung R, Rashid S, Gima J, Cote S, Pavez G, et al. Device artifact reduction for magnetic resonance imaging of patients with implantable cardioverter-defibrillators and ventricular tachycardia: late gadolinium enhancement correlation with electroanatomic mapping. Heart Rhythm. 2014;11(2):289–298. doi: 10.1016/j.hrthm.2013.10.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the cardiac imaging Committee of the Council on clinical cardiology of the American Heart Association. Circulation. 2002;105(4):539–542. doi: 10.1161/hc0402.102975. [DOI] [PubMed] [Google Scholar]
- 10.Rudski LG, Lai WW, Afilalo J, Hua L, Handschumacher MD, Chandrasekaran K, et al. Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2010;23(7):685–713. doi: 10.1016/j.echo.2010.05.010. [DOI] [PubMed] [Google Scholar]
- 11.Almehmadi F, Joncas SX, Nevis I, Zahrani M, Bokhari M, Stirrat J, et al. Prevalence of myocardial fibrosis patterns in patients with systolic dysfunction: prognostic significance for the prediction of sudden cardiac arrest or appropriate implantable cardiac defibrillator therapy. Circ Cardiovasc Imaging. 2014;7(4):593–600. doi: 10.1161/CIRCIMAGING.113.001768. [DOI] [PubMed] [Google Scholar]
- 12.Bleeker GB, Kaandorp TA, Lamb HJ, Boersma E, Steendijk P, de Roos A, et al. Effect of posterolateral scar tissue on clinical and echocardiographic improvement after cardiac resynchronization therapy. Circulation. 2006;113(7):969–976. doi: 10.1161/CIRCULATIONAHA.105.543678. [DOI] [PubMed] [Google Scholar]
- 13.Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, et al. 2017 AHA/ACC/HRS guideline for Management of Patients with Ventricular Arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association task force on clinical practice guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2018;72(14):e91–e220. doi: 10.1016/j.jacc.2017.10.054. [DOI] [PubMed] [Google Scholar]
- 14.White JA, Fine NM, Gula L, Yee R, Skanes A, Klein G, et al. Utility of cardiovascular magnetic resonance in identifying substrate for malignant ventricular arrhythmias. Circ Cardiovasc Imaging. 2012;5(1):12–20. doi: 10.1161/CIRCIMAGING.111.966085. [DOI] [PubMed] [Google Scholar]
- 15.Lazzerini PE, Capecchi PL, El-Sherif N, Laghi-Pasini F, Boutjdir M. Emerging arrhythmic risk of autoimmune and inflammatory cardiac channelopathies. J Am Heart Assoc. 2018;7(22):e010595. doi: 10.1161/JAHA.118.010595. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Rosenthal N. A Guardian of the heartbeat. N Engl J Med. 2017;377(1):84–86. doi: 10.1056/NEJMcibr1705327. [DOI] [PubMed] [Google Scholar]
- 17.Raphael CE, Vassiliou V, Alpendurada F, Prasad SK, Pennell DJ, Mohiaddin RH. Clinical value of cardiovascular magnetic resonance in patients with MR-conditional pacemakers. Eur Heart J Cardiovasc Imaging. 2016;17(10):1178–1185. doi: 10.1093/ehjci/jev305. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Acosta J, Fernandez-Armenta J, Penela D, Andreu D, Borras R, Vassanelli F, et al. Infarct transmurality as a criterion for first-line endo-epicardial substrate-guided ventricular tachycardia ablation in ischemic cardiomyopathy. Heart Rhythm. 2016;13(1):85–95. doi: 10.1016/j.hrthm.2015.07.010. [DOI] [PubMed] [Google Scholar]
- 19.Andreu D, Ortiz-Perez JT, Boussy T, Fernandez-Armenta J, de Caralt TM, Perea RJ, et al. Usefulness of contrast-enhanced cardiac magnetic resonance in identifying the ventricular arrhythmia substrate and the approach needed for ablation. Eur Heart J. 2014;35(20):1316–1326. doi: 10.1093/eurheartj/eht510. [DOI] [PubMed] [Google Scholar]
- 20.Oebel S, Dinov B, Arya A, Hilbert S, Sommer P, Bollmann A, et al. ECG morphology of premature ventricular contractions predicts the presence of myocardial fibrotic substrate on cardiac magnetic resonance imaging in patients undergoing ablation. J Cardiovasc Electrophysiol. 2017;28(11):1316–1323. doi: 10.1111/jce.13309. [DOI] [PubMed] [Google Scholar]
- 21.Indik JH, Gimbel JR, Abe H, Alkmim-Teixeira R, Birgersdotter-Green U, Clarke GD, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017;14(7):e97–e153. doi: 10.1016/j.hrthm.2017.04.025. [DOI] [PubMed] [Google Scholar]