Abstract
Exosomes are nanometer-sized lipid vesicles present in liquid biopsies and used as biomarkers for several diseases including cancer, Alzheimer’s, and central nervous system diseases. Purification and subsequent size and surface characterization are essential to exosome-based diagnostics. Sample purification is, however, time consuming and potentially damaging, and no current method gives the size and zeta potential from a single measurement. Here, we concentrate exosomes from a dilute solution and measure their size and zeta potential in a one-step measurement with a salt gradient in a capillary channel. The salt gradient causes oppositely directed particle and fluid transport that trap particles. Within minutes, the particle concentration increases more than two orders of magnitude. A fit to the spatial distribution of a single or an ensemble of exosomes returns both their size and surface charge. Our method is applicable for other types of nanoparticles. The capillary is fabricated in a low-cost polymer device.
Subject terms: Nanoparticles, Nanofluidics
Exosomes are used as disease biomarkers, but their characterization in biological samples is challenging. Here the authors achieve simultaneous characterization of size and zeta potential of individual nanoparticles and particle mixtures at physiological salinity conditions, exploiting a salt gradient in a capillary channel.
Introduction
Purification and analysis of nanoparticles, as routinely performed when isolating exosomes from liquid biopsies, are often based on size and surface properties, for example, their biochemical components or zeta potential1. The same parameters are also important for the synthesis and functionality of both solid state and soft matter synthetic colloids. These can, for example, be designed to promote interactions with living cells, have active reaction sites on their surfaces, or show specific optical properties.
The zeta potential, which depends on the surface charge, is important for the stability of nanoparticles in suspension2 and is also the major factor in the initial adsorption of nanoparticles onto the cell membrane3. After adsorption, the endocytotic uptake rate depends on the particle size4. The zeta potential and size thus affect nanoparticle toxicity5.
Exosomes, present in most body fluids, have received great interest due to their potential as biomarkers and use in precision medicine6–8. The zeta potentials and sizes of exosomes from various body fluids are significantly different as they interact with different cellular targets. So these parameters reflect both the origin and the endocytotic pathway that the exosomes may use3, and they are integral properties of the cell-to-cell signaling system. Control of the size and zeta potential are thus important factors for the effectiveness of nanoparticles for drug delivery1,9–11, and allows for specifying the cellular targets for, for example, liposomes4,12,13, gold nanoparticles14, and copolymer micelles15.
For soft matter nanoparticles, for example, exosomes or liposomes, traditional purification methods pose a number of challenges. Ultracentrifugation, one of the most commonly used concentration methods, is time consuming and suffers from contamination1,16. Size-based filtration through membranes can result in deformations or even break the particles1,17. Current characterization techniques for nanoparticles also all have challenges, for example, cryo-electron microscopy that is a destructive method18. Measurements of size and zeta potential are routinely performed on the same instrument by dynamic light scattering (DLS) and laser Doppler electrophoresis (LDE), but require multiple experiments, and are challenged by polydisperse samples, for example, exosomes isolated from body fluids.
The fast development within micro- and nanofluidics allows for reduced reagent consumption in lab-on-a-chip devices. Microfluidics-based purification and characterization methods have become increasingly important for nanoparticle research in general, and in particular for soft matter nanoparticles1,19–22. In micro- and nanofluidic devices, nanoparticles can be exposed to gradients in, for example, electric potential23, temperature24, pH25, and solute concentration26, and such gradients induce phoretic transport processes that depend on particle properties27. In an electric field, particles experience electrophoretic transport, which is, for example, utilized in size-dependent separation of DNA molecules in microlithographic arrays28, and for focusing of proteins in nanochannels23. In the presence of a solute gradient, particles will spontaneously migrate along the concentration gradient due to diffusiophoresis26,29. Diffusiophoresis offers much higher transport rates than regular diffusion and has been used for efficient filling of dead-end channels30–32, water filtration33, colloidal particle accumulation34, and active collective colloidal behavior30,35,36. For a Debye length comparable to the particle size, the diffusiophoretic velocity shows much higher particle size dependence than its electrophoretic counterpart, and it can thus be utilized for size-dependent separation of nanoparticles30.
The physical mechanism behind diffusiophoresis also gives rise to a diffusioosmotic fluid flow in the presence of a solute gradient near a charged plane26, for example, in a nanochannel37. This opens up for the possibility to manipulate particles by combining the diffusiophoretic migration with the diffusioosmotic flow of the surrounding fluid. The fluid velocity can be controlled efficiently by the nanochannel geometry.
In this article, we show how to combine diffusiophoretic particle transport with the oppositely directed diffusioosmotic fluid transport in order to trap, concentrate, and characterize nanoparticles in a nanofluidic device. The advantage of our method is that particles are trapped at an equilibrium position, not by a transient fluid flow30. This allows us to track individual particles in dilute suspensions for several minutes and to infer their properties from the trajectories. For ensembles, particle concentrations are much higher in the trap than in the initial sample, and particles are characterized based on the concentration profile. Data analysis is based on a closed set of equations without the need for calibration to simulations19. Our nanofluidic device does not require any electrodes to produce an applied electrical field in combination with the salt gradient23, and the trapping depends solely on the salt gradient and the geometry of the nanochannel. Hence, the device can be fabricated by scalable production processes, for example, polymer injection molding.
We first demonstrate the method on an ensemble of exosomes (Fig. 1). We then determine the size and surface charge of individual liposomes (Fig. 2), establishing the method as a single-particle tool. Next, we characterize ensembles of liposomes with nine different combinations of diameters and lipid compositions (Fig. 3). Finally, we perform on-chip separation and characterization of liposome populations based on size and lipid composition (Fig. 4).
Results
Principle of trapping by a salt gradient in a nanochannel
Our nanofluidic device traps nanoparticles in 16 parallel funnel-shaped nanochannels bridging two microchannels, see Fig. 1a–c. Different salt concentrations in the two microchannels maintain a position-dependent salt concentration C(x) across each nanochannel. The salt concentrations are identical in all nanochannels, that is, there are 16 parallel experiments on the device. The salt gradients in the nanochannels cause a diffusiophoretic particle migration26 and an oppositely directed diffusioosmotic fluid flow37, and they both depend on the relative, local salt gradient . Particles are trapped where the diffusiophoretic transport and the diffusioosmotic fluid flow balance each other, but they diffuse around the trapping position due to Brownian motion. The concentration profile of the trapped nanoparticles Cp(x) depends on their size and zeta potential.
Diffusion determines the salt gradient in each funnel-shaped nanochannel. The boundary conditions are the fixed salt concentrations in the microchannels, that is, CN and CW at the narrow and wide ends of the nanochannel, respectively. So inside the nanochannel, the salt concentration is (see “Methods”)
1 |
where w(x) is the local width of the nanochannel. We parametrize it as w(x) = wN + Δwx/L, with Δw = wW − wN, where wW = 20 μm and wN = 5 μm are the widths of the nanochannel at the wide and narrow ends of the nanochannel, respectively. The length of the nanochannel is L = 440 μm.
The salt gradient results in a diffusiophoretic particle velocity
2 |
Here Γph is the diffusiophoretic mobility, which depends on the particle diameter d and zeta potential ζ (for details, see “Methods”).
The diffusioosmotic fluid flow in the nanochannel is due to the local diffusioosmotic slip velocity νslip(x). For a fixed solute gradient in bulk near a charged wall, the diffusioosmotic slip velocity is26
3 |
where Γos(ζch) is the diffusioosmotic mobility and ζch is the zeta potential of the channel wall (details in “Methods”). The diffusioosmotic slip velocity causes a constant fluid flow rate Q due to conservation of mass, but the fluid velocity varies because of the changing channel width. We introduce a position-dependent diffusioosmotic flow velocity
4 |
where h = 240 nm is the height of the nanochannel.
We then relate the diffusioosmotic flow velocity in the nanochannel νos(x) to the diffusioosmotic slip velocity νslip(x) by assuming that37
5 |
holds locally, as the width is slowly varying compared to the length of the nanochannel (Δw/L ≪ 1). Here ∂xP(x) is the internal pressure gradient in the fluid along the axis of the channel. It is assumed that the pressures are identical at the two ends of the nanochannel, P(0) = P(L). Dividing both sides in Eq. (5) with w(x) and integrating from 0 to L using the boundary conditions on the pressure gives Q = Γos(ζch)ln(CN/CW)hΔw/[L ln(1 + Δw/wN)], and, consequently, vos(x), see Eq. (4). Note that the flow rate only depends on the salt concentrations in the microchannels, not the specific form of the salt concentration in the nanochannel.
Particles are trapped at the position x0, where νos(x0) + νph(x0) = 0. In a straight channel, the diffusioosmotic flow velocity νos(x) is constant along the channel37, but our funnel-shaped nanochannel gives a varying fluid velocity. This results in a tighter trap.
Finally, we calculate the particle concentration in the nanochannel Cp(x). It is related to the particle current density along the x-axis Jx(x) as (Fick’s first law of diffusion with drift)
6 |
Here Dp = Dp(d) is the size-dependent diffusion coefficient of particles in the nanochannel, which differs from its bulk value due to the interactions with the walls of the nanochannel (see “Methods”). The particle current Ip(x) in the nanochannel equals the particle current density times the cross-sectional area of the nanochannel, that is, Ip(x) = hw(x)Jx(x). In steady state, Ip(x) is constant along the nanochannel, and if particles are trapped, it vanishes. So Ip(x) = 0, and, consequently, Jx(x) = 0. For a trapping position x0 far away from both ends of the nanochannel, a solution to the particle concentration is (see Eq. (6))
7 |
Here Cp(x0) = N/[hw(x0)Ltrap], where N is the number of trapped particles and is the effective length of the trap. That is, the particle distribution is normalized such that an integration over the volume of the nanochannel gives the number of trapped particles, . The expression for Cp(x) does not have a simple analytic form, but it is straightforward to solve numerically.
In Eq. (7), the diffusioosmotic flow velocity νos(x) depends solely on the properties of the nanochannel, not the properties of the nanoparticle. So νos(x) can be determined from the known zeta potential of the channel walls or a prior measurement (Supplementary Note 1). Both the diffusion coefficient Dp and the diffusiophoretic velocities νph(x) depend on the diameter of the nanoparticle, and νph(x) also depends on its zeta potential ζ. So a fit of Eq. (7) to an experimentally measured concentration profile has the particle diameter d and zeta potential ζ as the only free parameters, except for an arbitrary scale factor that converts between the particle concentration and measured intensity. In all fits of concentration profiles presented below, the particle diameters, zeta potentials, and scale factors are fitted with no constraints between data sets.
Trapping of exosomes
We first demonstrate trapping, concentration and characterization of exosomes isolated from human blood serum of healthy donors. The stationary salinity gradients across the nanochannels are kept by a continuous flow of phosphate-buffered saline (PBS) through the microchannels. The concentrations are CN and CW at the narrow and wide ends of the nanochannels, respectively. Exosomes stained with the fluorescent dye DiO are introduced in the microchannel connected to the narrow ends of the nanochannels. At these ends of the nanochannels, diffusiophoresis dominates diffusioosmosis and particles migrate into the nanochannels and get trapped (Fig. 1e and Supplementary Movie 1).
After 90 s, 16 exosomes are trapped in the first nanochannel (see Fig. 1a). So, the particle concentration in the trap is ~400 times higher than the initial concentration in the microchannel (from 6 pM to 2.4 nM). Here the trap volume is defined as Vtrap = Ltraphw(x0) = 10.6 fL. From the known flow rate in the microchannel, its cross-sectional area of 150 (μm)2, and a particle concentration equal to 6 pM, we calculate that ~290 exosomes pass the entry of the nanochannel in 90 s. So the trapping efficiency of exosomes for a single nanochannel is 16/290 = 5.5%.
Figure 1e shows trapped exosomes at three different salinity gradients achieved by changing CN while maintaining CW at 10× PBS. Exosomes are trapped at physiologically relevant salinities (0.3–0.8× PBS), where the Debye length is ~1 nm (see “Methods”). The trapping positions and widths depend significantly on the gradient. For each measurement, a fit of the intensity profile gives the particle diameter and zeta potential, see Fig. 1f. Weighted averages over the three measurements at different salinity gradients (Fig. 1g) yields d = (78 ± 7) nm for the particle diameter and ζ = (−18 ± 1) mV for the zeta potential. The latter is confirmed by LDE as ζLDE = (−20 ± 5) mV. Size measurements with DLS were inconclusive due to the polydispersity of the sample, but the measured size agrees with the peak in the size distribution provided by the vendor (Supplementary Note 3).
Characterizing individual liposomes
Liposomes can be made with desired sizes and zeta potentials, so they are well suited for establishing trapping in a salinity gradient and to perform characterization. We use liposomes with three different lipid compositions, that is, different POPC (1-palmitoyl-2-oleoyl-glycero-3-phosphocholine):POPG (1-palmitoyl-2-oleoyl-sn-glycero-3-phosphoglycerol) ratios that are either 3:1, 1:1, or 1:3, and, consequently, different zeta potentials (~−30, ~−36, and ~−48 mV). Liposomes were extruded through membranes with different pore sizes to produce subpopulations with different diameters (~70, ~110, and ~150 nm). Before loading the liposomes in the nanofluidic device, they were analyzed with DLS and LDE to benchmark our results.
We first measure the sizes and zeta potentials of individual liposomes. For a low concentration of liposomes in the microchannel, the filling rate of the trap is so slow that we can capture a single liposome in a nanochannel and track its stochastic motion for up to several minutes. So we load a nanochannel with a single liposome from an ensemble with a mean diameter dDLS = (76 ± 3) nm and zeta potential ζLDE = (−28 ± 1) mV (Supplementary Movie 2). From microscopy images (Fig. 2a), we extract the time-dependent position of the particle in the nanochannel (Fig. 2b) and plot the x-coordinates (Fig. 2c).
We assume that in the long-time limit, the distribution of a single particle’s positions is identical to the spatial distribution for an ensemble of similar particles. So for each particle, we fit Eq. (7) to the histogram of its measured x-coordinates (black, dashed curves in Fig. 2d). For the histogram in the upper panel in Fig. 2d, the fit yields the particle diameter d = (72 ± 9) nm and the zeta potential ζ = (−30 ± 2) mV. This is consistent with the DLS and LDE measurements. Uncertainties are only ~15% and ~6%, respectively, of the measured values for a 40 s measurement. Other panels in Fig. 2d show histograms of x-coordinates from trajectories for two liposomes with different sizes (middle panel) and lipid compositions (lower panel). Notice how the changes in size and zeta potential alter the trapping positions x0 and the widths of the distributions. Figure 2e shows fitted values for the diameters and zeta potentials for particles from the POPC:POPG 3:1 population with ζLDE = (−28 ± 1) mV and diameters dDLS = (76 ± 3) nm (red, dashed lines). The measurements on individual liposomes indicate that particle-to-particle variation can be resolved by tracking individual particles in the nanofluidic trap.
Characterizing liposome ensembles
The trapping of exosomes demonstrates characterization of ensembles (Fig. 1). Here, we further validate the method with nine different combinations of liposome sizes and zeta potentials. We verify that results do not depend on the specific salinity gradient in the experiment, that all results are consistent with DLS and LDE measurement, and show how the trapping position x0 depends on the liposome parameters and the salinity gradient.
Figure 3a shows trapping of POPC:POPG 3:1 liposomes for seven different salinity gradients with ln(CN/CW) from −9.2 to −2.3. For ln(CN/CW) = −9.2, the first nanochannel captures 1000 liposomes in 10 s. That is ~6% of the liposomes that passed its entry, so the trapping efficiency is similar to that measured for exosomes. The 16 nanochannels (Fig. 1a) trap in total ~37% of the liposomes introduced in the microchannel (see Supplementary Note 4).
The widths of the distributions of trapped particles increase as the salinity gradient decreases because the diffusiophoretic and diffusioosmotic velocities decrease. Adjustment of the salt concentrations in the microchannels allows for moving the trapping position to physiological salinity (1× PBS, marked by the horizontal green lines in Fig. 3a), here for ln(CN/CW) between −4.6 and −3.5. Equation (7) is fitted to the intensity profile for all salinity gradients (Fig. 3b), and Fig. 3c displays the fitted values for the diameters and zeta potentials for all seven salinity gradients. No systematic dependence on the salinity gradient is observed for the fitted values and they are consistent with common weighted average values d = (75 ± 3) nm and ζ = (−30 ± 1) mV (dotted lines in Fig. 3c). Measurements identical to those in Fig. 3a–c were performed for the other eight types of liposomes. The results are summarized in Fig. 3d and are consistent with DLS and LDE measurements. For the data in Fig. 3d, the estimated net charges on the liposomes vary between ~1000e and ~7600e (ref. 2, Eq. 2.5.5).
The ability to discriminate particle populations based on size and zeta potential depends on the channel geometry as it modulates the velocity profile along the nanochannel, and hence the trapping position. For the present nanochannel design and experimental conditions, the trapping position is more sensitive to the zeta potential than the size. This is demonstrated by a comparison of the trapping positions for liposome samples with identical zeta potentials (Fig. 3e) or identical sizes (Fig. 3f) at fixed salinity gradients. Solid lines are the theoretically predicted trapping positions versus salt gradient, where the input parameters are the measured sizes and zeta potentials shown in Fig. 3d (trapping method). Notice in both cases the nonmonotonic dependence of the trapping position on the salt gradient. This occurs because the diffusioosmotic flow velocity is proportional to ln(CN/CW), while the diffusiophoretic velocity is proportional to ∂x ln C(x). So, the latter depends nonlinearly on ln(CN/CW) (see Supplementary Note 5).
Separating mixed liposome populations
For a fixed salt gradient, the distance between the trapping positions is only a few microns for the two largest particle sizes (d = 106 and 155 nm), much smaller than the widths of the spatial distributions of particles. The separation of trapping positions for different zeta potentials is significantly larger (see Fig. 3e, f). We therefore demonstrate separation of a liposome mixture based on surface characteristics and introduce an equal mixture of liposomes with identical sizes, but different lipid compositions, in the nanofluidic device. The liposomes are POPC:POPG 3:1 with diameters dDLS = (73 ± 3) nm and zeta potentials ζLDE = (−28 ± 1) mV, and POPC:POPG 1:3 with dDLS = (73 ± 3) nm and ζLDE = (−50 ± 3) mV. Figure 4a shows the separation of two liposome populations marked with different fluorophores (Supplementary Movie 3), where the separation is due to the different zeta potentials. Importantly, the separation is also clearly seen when both populations are marked with the same fluorophore (Fig. 4b). A fit to the particle distribution gives d = (75 ± 3) nm and ζ = (−30 ± 1) mV for POPC:POPG 3:1, and d = (69 ± 3) nm and ζ = (−48 ± 2) mV for POPC:POPG 1:3 (Fig. 4c). Results are consistent with DLS and LDE measurements, but the trapping method requires only a single measurement, and much smaller sample volumes and concentrations.
Discussion
From a single measurement lasting less than minutes, we can accurately determine both the particle size and zeta potential for individual particles, ensembles of particles, or even particle mixtures. Data analysis is based on a closed set of equations and does not require any calibration to simulations. Device fabrication by injection molding, a scalable industrial process, allows for low-cost mass production of the device and integration with microfluidics for sample and cell handling.
With its combination of rapid measurements and a single-use device, the technology can be applied to on-chip concentration of dilute samples in applications where sample amounts are scarce, for example, in single-cell analysis. By tuning the trapping position to physiological salinity, biochemical reactions, for example, immunoreactions, can occur and be monitored in real time. A novel device design may also allow for surface charge characterization of smaller particles, for example, individual proteins20, at physiological salinity. The method is equally viable for other types of nanoparticles, and the integration of this nanofluidic method with microfluidics would, for example, allow for in-line characterization of nanoparticle liquid phases synthesis. Finally, liquid biopsy-based diagnostics can be envisaged. For example, by taking advantage of the high concentrations of particles in the trap for label-tree detection of cancerous extracellular vesicles with Raman spectroscopy38.
Methods
Sample preparation
Purified exosomes derived from human blood serum in PBS (NaCl 137 mM, KCl 2.7 mM, Na2PHO4 10 mM, and KH2PO4 1.8 mM) were purchased from BioCat. The size distribution has its maximum at a diameter of 70 nm (full size distribution is shown in Supplementary Fig. 4). Exosomes were incubated at 37 °C with the green lipofilic fluorophore DiO. Excess fluorophores were removed by spin column purification. Liposomes were prepared by mixing the lipids POPC, POPG, and fluorophore (Texas red and DiO were used) at appropriate ratios and were dissolved in a 9:1 tertiary butanol to water solution. Samples were freeze dried overnight and the lipid film was rehydrated in PBS while being vortexed at 50 °C. All POPC:POPG mixtures were divided into three subpopulations, which were extruded through 30, 50, and 100 nm filters, respectively.
Nanochannel device fabrication
The nanochannel device was made by replicating a nickel master using injection molding of the cyclic olefin copolymer (COC) TOPAS 501339. The nickel master was produced in a two-step ultraviolet (UV) lithography and reactive ion etching process making nanochanels and microchannels with heights of 240 nm and 5 μm, respectively. The injection molded part was sealed with a 150-μm-thick COC foil by UV-assisted thermal bonding.
Particle trapping experiments
The device is coated for 30 min with a phospholipid POPC:POPG mixture dissolved in 70% ethanol to passivate the surface. The ratio of the uncharged POPC to the charged POPG also determines the zeta potential of the channel walls ζch, and, consequently, the diffusioosmotic flow26,37. We used lipid coatings with a surface potential of ζch = (−24 ± 1) mV and ζch = (−30 ± 1) mV for the exosome and liposome experiments, respectively. Characterization of the coatings is shown in Supplementary Note 1. To establish and maintain the salinity gradient in the nanochannels, PBS buffer solutions with two different concentrations were continuously flown through the microchannels by a 5 mbar pressure difference between the in- and outlets (Fig. 1a). Pressures were controlled by a Fluigent MFCS-EX pump with a stated instrumental error of ±0.3 mbar. The resulting flow rates in the microchannels are 450 pL min−1, and as the cross-sections of the microchannels are 150 (μm)2, the flow velocity is 50 μm s−1. Constant and reproducible residual flows of (36±15) fL min−1 were observed in the nanochannels with these pump settings (Supplementary Note 1). They correspond to a residual pressure across the nanochannel of 0.21 mbar, consistent with the instrumental error of the pump. The residual flows in the nanochannels were included in the data analysis by using νos(x) = Q/[hw(x)] with the fitted value of Q, including the residual flow, shown in Supplementary Fig. 1b, rather than the theoretical expression for Q stated in and below Eq. (5). After the nanoparticles are introduced, it takes ~2 min before the salt gradient is established and a sufficient amount of nanoparticles are trapped. The solution is then changed to one with identical salinity, but without nanoparticles. The data acquisition is then performed on the trapped particles. The fluorescence image from the trapped particles was recorded with a Nikon eclipse Ti2 microscope with a Photometrics Evolve 512 electron-multiplying charge-coupled device camera. A CoolLED pE-300 Ultra LED was used as light source.
Salt gradient in a funnel-shaped nanochannel
The salt concentration C(x) in the nanochannel depends on the salt concentrations in the two microchannels, that is, CN at x = 0 and CW at x = L (see Fig. 1). The salt current density along the x-axis is J(x) = −D∂xC(x), where D is the diffusion coefficient of the salt ions. The salt current I(x) in the nanochannel equals the salt current density times the cross-sectional area of the nanochannel, that is, I(x) = hw(x)J(x). In steady state, I(x) is constant along the nanochannel, that is, ∂xI(x) = 0, so ∂x[w(x)∂xC(x)] = 0. The solution for C(x) that fulfills the boundary conditions is shown in Eq. (1).
Diffusiophoretic and diffusioosmotic mobilities
An expansion of the diffusiophoretic mobility Γph to first order in λ = λDB/(d/2) is , where is the Debye length, d is the diameter of the particle, is the permittivity of the medium, η is the dynamic viscosity, Z is the valence of the solute, e is the elementary charge, , and u1 = F0 + βF1 + Pe[F2 + β(F3 + F5) + β2F4]26. Here NA is Avogadro’s number, I the ionic strength, and is the Peclet number. Furthermore, , , β = (D+ − D−)/(D+ + D−), and D = 2D+D−/(D+ + D−), where D+ and D− are the diffusion coefficient of the cations and anions of a monovalent salt, respectively. The dominant ions in PBS are sodium ions, Na+, and chloride ions, Cl−, with the diffusion constants and , respectively30. Thus, β ≃ −0.20 for NaCl. The F-functions depend on and are tabulated in Table 2 in ref. 26.
The diffusioosmotic mobility is 26.
Diffusion between walls
A particle diffusing between walls has a different diffusion coefficient Dp than in bulk, where the diffusion coefficient is D0 = kBT/(6πηd/2). So an expression for the diffusion coefficient of a particle in a nanochannel is required to determine the particle size and zeta potential from a concentration profile, see Eq. (7). We here use Dp = [1 − 1.004(d/h) + 0.418(d/h)3 + 0.21(d/h)4 − 0.169(d/h)5]D040, an expansion in the ratio between the particle diameter d and the channel height h that is only strictly valid for a particle midway between two infinite walls. In Supplementary Note 2, we test experimentally the validity of the expression for our device and nanoparticles.
DLS and LDE
DLS and LDE measurements were performed with a Zetasizer Nano Z instrument to determine the nanoparticles’ diameters and zeta potentials, respectively.
Supplementary information
Acknowledgements
We thank J.B. Larsen for assisting with liposome synthesis.
Author contributions
R.M. and J.N.P. developed the project and led the experimental research. M.K.R and R.M. performed the experiments. M.K.R. and J.N.P. developed the theory to predict the trapping and device behavior. M.K.R., R.M. and J.N.P. analyzed and interpreted the data. All authors co-wrote the paper.
Data availability
The data that support the findings of this study are available from the corresponding authors upon reasonable request.
Code availability
The code used in this study is available from the corresponding authors upon reasonable request.
Competing interests
The authors declare no competing interests.
Footnotes
Peer review information Nature Communications thanks the anonymous reviewers for their contribution to the peer review of this work. Peer reviewer reports are available.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Jonas N. Pedersen, Email: jnpe@dtu.dk
Rodolphe Marie, Email: rcwm@dtu.dk.
Supplementary information
Supplementary information is available for this paper at 10.1038/s41467-020-15889-3.
References
- 1.Li P, Kaslan M, Lee SH, Yao J, Gao Z. Progress in exosome isolation techniques. Theranostics. 2017;7:789. doi: 10.7150/thno.18133. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Hunter, R. Zeta Potential in Colloid Science. Principles and Applications (Academic Press, 1981).
- 3.Honary S, Zahir F. Effect of zeta potential on the properties of nano-drug delivery systems—a review (part 1) Trop. J. Pharm. Res. 2013;12:255–264. [Google Scholar]
- 4.Sharma A, Sharma US. Liposomes in drug delivery: progress and limitations. Int. J. Pharm. 1997;154:123–140. doi: 10.1016/S0378-5173(97)00135-X. [DOI] [Google Scholar]
- 5.Schwegmann H, Feitz AJ, Frimmel FH. Influence of the zeta potential on the sorption and toxicity of iron oxide nanoparticles on s. cerevisiae and e. coli. J. Colloid Interface Sci. 2010;347:43–48. doi: 10.1016/j.jcis.2010.02.028. [DOI] [PubMed] [Google Scholar]
- 6.Raposo G, Stoorvogel W. Extracellular vesicles: exosomes, microvesicles, and friends. J. Cell Biol. 2013;200:373–383. doi: 10.1083/jcb.201211138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Valadi H, et al. Exosome-mediated transfer of mrnas and micrornas is a novel mechanism of genetic exchange between cells. Nat. Cell Biol. 2007;9:654. doi: 10.1038/ncb1596. [DOI] [PubMed] [Google Scholar]
- 8.Théry C, Zitvogel L, Amigorena S. Exosomes: composition, biogenesis and function. Nat. Rev. Immunol. 2002;2:569. doi: 10.1038/nri855. [DOI] [PubMed] [Google Scholar]
- 9.Blanco E, Shen H, Ferrari M. Principles of nanoparticle design for overcoming biological barriers to drug delivery. Nat. Biotechnol. 2015;33:941. doi: 10.1038/nbt.3330. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sercombe L, et al. Advances and challenges of liposome assisted drug delivery. Front. Pharmacol. 2015;6:286. doi: 10.3389/fphar.2015.00286. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Im H, et al. Label-free detection and molecular profiling of exosomes with a nano-plasmonic sensor. Nat. Biotechnol. 2014;32:490. doi: 10.1038/nbt.2886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Farokhzad OC, Langer R. Impact of nanotechnology on drug delivery. ACS Nano. 2009;3:16–20. doi: 10.1021/nn900002m. [DOI] [PubMed] [Google Scholar]
- 13.Al-Jamal W, Kostarelos K. Liposomes: from a clinically established drug delivery system to a nanoparticle platform for theranostic nanomedicine. Acc. Chem. Res. 2011;44:1094–1104. doi: 10.1021/ar200105p. [DOI] [PubMed] [Google Scholar]
- 14.Cho EC, Zhang Q, Xia Y. The effect of sedimentation and diffusion on cellular uptake of gold nanoparticles. Nat. Nanotechnol. 2011;6:385. doi: 10.1038/nnano.2011.58. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Hao W, et al. ph-triggered copolymer micelles as drug nanocarriers for intracellular delivery. RSC Adv. 2016;6:29149–29158. doi: 10.1039/C6RA00673F. [DOI] [Google Scholar]
- 16.Lane RE, Korbie D, Anderson W, Vaidyanathan R, Trau M. Analysis of exosome purification methods using a model liposome system and tunable-resistive pulse sensing. Sci. Rep. 2015;5:7639. doi: 10.1038/srep07639. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Batrakova EV, Kim MS. Using exosomes, naturally-equipped nanocarriers, for drug delivery. J. Control. Rel. 2015;219:396–405. doi: 10.1016/j.jconrel.2015.07.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Sokolova V, et al. Characterisation of exosomes derived from human cells by nanoparticle tracking analysis and scanning electron microscopy. Colloids Surf. B. 2011;87:146–150. doi: 10.1016/j.colsurfb.2011.05.013. [DOI] [PubMed] [Google Scholar]
- 19.Mojarad N, Krishnan M. Measuring the size and charge of single nanoscale objects in solution using an electrostatic fluidic trap. Nat. Nanotechnol. 2012;7:448. doi: 10.1038/nnano.2012.99. [DOI] [PubMed] [Google Scholar]
- 20.Ruggeri F, et al. Single-molecule electrometry. Nat. Nanotechnol. 2017;12:488. doi: 10.1038/nnano.2017.26. [DOI] [PubMed] [Google Scholar]
- 21.Wunsch BH, et al. Nanoscale lateral displacement arrays for the separation of exosomes and colloids down to 20 nm. Nat. Nanotechnol. 2016;11:936. doi: 10.1038/nnano.2016.134. [DOI] [PubMed] [Google Scholar]
- 22.Skaug MJ, Schwemmer C, Fringes S, Rawlings CD, Knoll AW. Nanofluidic rocking brownian motors. Science. 2018;359:1505–1508. doi: 10.1126/science.aal3271. [DOI] [PubMed] [Google Scholar]
- 23.Hsu W-L, et al. Concentration gradient focusing and separation in a silica nanofluidic channel with a non-uniform electroosmotic flow. Lab Chip. 2014;14:3539–3549. doi: 10.1039/C4LC00504J. [DOI] [PubMed] [Google Scholar]
- 24.Ross D, Locascio LE. Microfluidic temperature gradient focusing. Anal. Chem. 2002;74:2556–2564. doi: 10.1021/ac025528w. [DOI] [PubMed] [Google Scholar]
- 25.Startsev MA, Inglis DW, Baker MS, Goldys EM. Nanochannel ph gradient electrofocusing of proteins. Anal. Chem. 2013;85:7133–7138. doi: 10.1021/ac4014447. [DOI] [PubMed] [Google Scholar]
- 26.Prieve DC, Anderson JL, Ebel JP, Lowell ME. Motion of a particle generated by chemical gradients. Part 2. Electrolytes. J. Fluid Mech. 1984;148:247–269. doi: 10.1017/S0022112084002330. [DOI] [Google Scholar]
- 27.Anderson JL. Colloid transport by interfacial forces. Annu. Rev. Fluid Mech. 1989;21:61–99. doi: 10.1146/annurev.fl.21.010189.000425. [DOI] [Google Scholar]
- 28.Volkmuth WD, Austin RH. Dna electrophoresis in microlithographic arrays. Nature. 1992;358:600–602. doi: 10.1038/358600a0. [DOI] [PubMed] [Google Scholar]
- 29.Staffeld PO, Quinn JA. Diffusion-induced banding of colloid particles via diffusiophoresis: 1. Electrolytes. J. Colloid Interface Sci. 1989;130:69–87. doi: 10.1016/0021-9797(89)90079-9. [DOI] [Google Scholar]
- 30.Shin S, et al. Size-dependent control of colloid transport via solute gradients in dead-end channels. Proc. Natl Acad. Sci. USA. 2016;113:257–261. doi: 10.1073/pnas.1511484112. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Kar A, Chiang T-Y, Ortiz Rivera I, Sen A, Velegol D. Enhanced transport into and out of dead-end pores. ACS Nano. 2015;9:746–753. doi: 10.1021/nn506216b. [DOI] [PubMed] [Google Scholar]
- 32.Battat S, Ault JT, Shin S, Khodaparast S, Stone HA. Particle entrainment in dead-end pores by diffusiophoresis. Soft Matter. 2019;15:3879–3885. doi: 10.1039/C9SM00427K. [DOI] [PubMed] [Google Scholar]
- 33.Shin S, Shardt O, Warren PB, Stone HA. Membraneless water filtration using co2. Nat. Commun. 2017;8:15181. doi: 10.1038/ncomms15181. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Shin S, Ault JT, Warren PB, Stone HA. Accumulation of colloidal particles in flow junctions induced by fluid flow and diffusiophoresis. Phys. Rev. X. 2017;7:041038. [Google Scholar]
- 35.Palacci J, Sacanna S, Steinberg AP, Pine DJ, Chaikin PM. Living crystals of light-activated colloidal surfers. Science. 2013;339:936–940. doi: 10.1126/science.1230020. [DOI] [PubMed] [Google Scholar]
- 36.Palacci J, Abécassis B, Cottin-Bizonne C, Ybert C, Bocquet L. Colloidal motility and pattern formation under rectified diffusiophoresis. Phys. Rev. Lett. 2010;104:138302. doi: 10.1103/PhysRevLett.104.138302. [DOI] [PubMed] [Google Scholar]
- 37.Lee C, et al. Osmotic flow through fully permeable nanochannels. Phys. Rev. Lett. 2014;112:244501. doi: 10.1103/PhysRevLett.112.244501. [DOI] [PubMed] [Google Scholar]
- 38.Lee W, et al. Label-free prostate cancer detection by characterization of extracellular vesicles using raman spectroscopy. Anal. Chem. 2018;90:11290–11296. doi: 10.1021/acs.analchem.8b01831. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Utko P, Persson F, Kristensen A, Larsen NB. Injection molded nanofluidic chips: fabrication method and functional tests using single-molecule DNA experiments. Lab Chip. 2011;11:303–308. doi: 10.1039/C0LC00260G. [DOI] [PubMed] [Google Scholar]
- 40.Happel, J. & Brenner, H. Low Reynolds Number Hydrodynamics (Prentice-Hall, 1965).
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study are available from the corresponding authors upon reasonable request.
The code used in this study is available from the corresponding authors upon reasonable request.