Skip to main content
Journal of Medical Imaging logoLink to Journal of Medical Imaging
. 2020 May 12;7(3):033501. doi: 10.1117/1.JMI.7.3.033501

Evaluation of a hybrid direct–indirect active matrix flat-panel imager using Monte Carlo simulation

Scott Dow 1,*, Adrian Howansky 1, Anthony R Lubinsky 1, Wei Zhao 1
PMCID: PMC7215239  PMID: 32411813

Abstract.

Purpose: Monte Carlo simulations were used to evaluate the imaging properties of a composite direct–indirect active matrix flat-panel imager (AMFPI) with potentially more favorable tradeoffs between x-ray quantum efficiency and spatial resolution than direct or indirect AMFPIs alone. This configuration, referred to as a hybrid AMFPI, comprises a scintillator that is optically coupled to an a-Se direct AMFPI through a transparent electrode and hole blocking layer, such that a-Se acts as both a direct x-ray converter and an optical sensor.

Approach: GEANT4 was used to simulate x-ray energy deposition, optical transport, and charge signal generation processes in various hybrid AMPFI configurations under RQA5 and RQA9 x-ray beam conditions. The Fujita–Lubberts–Swank method was used to quantify the impact of irradiation geometry, x-ray converter thicknesses, conversion gain of each layer, and x-ray cross talk between layers on detective quantum efficiency (DQE).

Results: Each hybrid configuration had a greater DQE than its direct AMFPI layer alone. The DQE improvement was largest at low spatial frequencies in both front- and back-irradiation (BI) geometries due to increased x-ray quantum efficiency provided by the scintillator. DQE improvements persisted at higher frequencies in BI geometry due to preferential x-ray absorption in a-Se. Matching the x-ray-to-charge conversion gains of a hybrid AMFPI’s direct and indirect detection layers affects its Swank factor and, thus, DQE(0). X-ray cross talk has a negligible impact on the DQE(f) of hybrid AMFPIs with sufficiently high optical quantum efficiency.

Conclusion: An optimized hybrid AMFPI can achieve greater DQE performance than current direct or indirect AMFPIs.

Keywords: digital radiography, flat-panel detector, active matrix flat-panel imager, detective quantum efficiency, amorphous selenium, scintillator

1. Introduction

Active matrix flat-panel imagers (AMFPIs) are currently the dominant technology in digital radiography and fluoroscopy and may be broadly classified as either direct or indirect AMFPIs.1 Direct AMFPIs [Fig. 1(a)] use a photoconductor to convert absorbed x-ray energy into charge, which may be collected by a 2D array of pixel electrodes and read out using an amorphous silicon (a-Si:H) thin-film transistor (TFT) array. Direct AMFPIs possess a high spatial resolution due to the minimal lateral spread of charge within the photoconductor. However, they suffer from a low x-ray quantum efficiency (η) due to the low atomic number (Z=34) of amorphous selenium (a-Se), which is currently the only photoconductor that is commercially feasible for use in AMFPIs. This fundamentally limits detective quantum efficiency (DQE), a detector performance metric that quantifies the spatial-frequency (f) dependent signal-to-noise properties of an x-ray imager. Higher Z alternatives to a-Se, such as PbI2, HgI2, and various perovskites, are currently being researched, but they have not yet been sufficiently developed for practical use.25 Direct AMFPIs using thick a-Se layers (up to 1000  μm)69 have been developed to improve η; however, further increases in thickness are not possible due to the limited mobility–lifetime product of electrons in a-Se.7 As a result, direct AMFPIs are best suited for lower energy (20 to 50 keV) applications, such as mammography and digital breast tomosynthesis.10,11

Fig. 1.

Fig. 1

(a) Schematic of a direct AMFPI. X-rays are converted directly into charge via the photoconductor, which is read out by the TFT array. (b) Schematic of an indirect AMFPI. X-rays are converted into optical photons, which are then converted into charge and read out by a photodiode-TFT array. BI of indirect AMFPIs results in greater light escape efficiency and less blur and image noise than FI.

Indirect AMFPIs [Fig. 1(b)] consist of a scintillator coupled to a pixelated photodiode–TFT array. The scintillator converts the energy of absorbed x-rays into optical photons, which are then converted to charge by the photodiodes. Indirect AMFPIs possess a higher η than direct AMFPIs under most imaging conditions due to the high Z of the inorganic scintillators they use, which are typically powder Gd2O2S:Tb (GOS) or columnar CsI:Tl (herein, referred to as CsI). This advantage is partially offset by the fact that indirect AMFPIs have a lower spatial resolution and more image noise (i.e., Swank noise and the Lubberts effect)12,13 than direct AMFPIs due to light scattering and absorption within the scintillator. Consequently, indirect AMFPIs are best suited for dose-sensitive applications that require higher x-ray energies (e.g., >100  kVp) and heavily filtered beams, such as interventional fluoroscopy and cone-beam CT.

Alternative AMFPI configurations are now being proposed to improve indirect AMFPI performance. For example, Sato et al. proposed applying a back-irradiated (BI) geometry commonly used in screen-film radiography to AMFPIs.14 In a BI geometry [see Fig. 1(b)], x-rays are first transmitted through the photodiode–TFT array before interacting with the scintillator. This results in preferential x-ray absorption near the optical sensor, which increases light escape efficiency and reduces blur and image noise. Lubinsky et al.15 and Howansky et al.16 showed that these advantages increase the DQE of indirect AMFPIs relative to their performance in the usual front-irradiated (FI) geometry, particularly at high f. Further improvements in DQE are expected using a properly optimized “sandwich” AMFPI configuration, comprising an FI and BI scintillator coupled to a bidirectional optical sensor.15 Other investigators have reported the performance of AMFPIs comprising multiple stacks of scintillators and optical sensors. Rottmann et al. investigated an electronic portal imaging device (EPID) comprising four layers, each consisting of a copper build-up layer, a scintillator, and a photodiode–TFT array.17 A substantially higher DQE was reported compared with an EPID using only one layer. Maurino proposed a similar detector for single-shot dual-energy diagnostic imaging, comprising three layers of stacked indirect AMFPIs using CsI.18

In this work, we propose a novel indirect–direct detector configuration referred to as a hybrid AMFPI (Fig. 2), which consists of an x-ray scintillator coupled to an a-Se direct detector through thin, optically transparent intervening layers (i.e., bias electrode and blocking layer).19 In this configuration, both the scintillator and a-Se function as x-ray converters; a-Se generates signal charge via direct interactions, while the scintillator produces optical photons, which are subsequently detected and converted into charge by a-Se. Practical limits on a-Se thickness and, thus, the direct detector’s η, are overcome using the scintillator to absorb x-rays that would otherwise be undetected to achieve both a high η and high spatial resolution. BI geometry is proposed for the hybrid AMFPI both to maximize x-ray attenuation in a-Se, which has more desirable imaging properties than the scintillator, and to maximize the inherent imaging performance of the scintillator (Fig. 2). The purpose of this work is to use Monte Carlo (MC) simulation to evaluate the potential imaging performance of hybrid AMFPIs and the impact of using different design parameters.

Fig. 2.

Fig. 2

Schematic of a hybrid AMFPI. Both a-Se and the scintillator function as x-ray converters. X-rays absorbed in a-Se are converted directly into charge. X-rays absorbed in the scintillator are converted into optical photons, which are detected at the scintillator/a-Se interface and converted into charge. Charge is then read out by the TFT array.

2. Methods

MC simulations were used to model the x-ray imaging performance of different hybrid AMFPI configurations (Sec. 2.1) by generating ensembles of the detector’s 2D response to individual normally incident x-rays, i.e., single x-ray responses (Sec. 2.2). Zero-frequency metrics, such as the pulse height spectrum (PHS), Swank factor, and DQE(0), were found by integrating each single x-ray response in 2D to yield the total signal induced by each x-ray (Sec. 2.3). To derive spatial-frequency dependent metrics, such as MTF(f), NPS(f), and DQE(f), the Fujita–Lubberts–Swank (FLS) method was used (Sec. 2.4). The impact of irradiation geometry, x-ray converter thickness, x-ray conversion gain of each layer (Sec. 2.5), and x-ray cross talk between each layer (Sec. 2.6) on these performance metrics was then investigated.

2.1. Overview of the Hybrid AMFPI Simulation

MC simulations were performed using GEANT4, a C++ based MC package designed to simulate particle transport through matter.20 X-ray and electron transport were modeled using the “G4EmPenelopePhysics” GEANT4 physics list. Optical transport was modeled using the “G4OpticalPhysics” list. As shown in Fig. 3(a), the hybrid AMFPI was modeled as a three-layer structure comprising (1) an x-ray scintillator, (2) a photoconductive layer of a-Se, and (3) TFT substrate glass. An overview of simulation parameters is shown in Table 1.

Fig. 3.

Fig. 3

(a) Schematic of our hybrid AMFPI model, which consists of GOS, a-Se, and TFT substrate glass. Four types of x-ray interactions are possible. Type 1 interactions only deposit energy in a-Se and only result in a direct signal component. Type 2 interactions only deposit energy in GOS and only result in an indirect signal component. Type 3 and 4 interactions result in x-ray cross talk from a-Se into GOS and GOS into a-Se, respectively, which occur due to either x-ray scatter or fluorescence. (b) Relative intensity plots of RQA5 and RQA9 x-ray spectra, which have average energies of 52 and 76 keV, respectively.

Table 1.

Parameters used to simulate the hybrid AMFPI.

Parameter Value
GOS thickness (dGOS) 0 to 1000  μm
a-Se thickness (da-Se) 0 to 1000  μm
TFT substrate thickness 700  μm
GOS density 4.59  gcm3
a-Se density 4.29  gcm3
GOS absorption length 40 mm
GOS scattering length 3.67  μm
Reflectivity of GOS backing 0.88
Optical quantum efficiency (β) 0 to 1
GOS scintillation yield (YGOS) 60  keV1
X-ray energy per detected ehp in a-Se (WSe) 7 eV to WSe
Electric field of a-Se (ESe) 0 to 35  Vμm1
Exponent in WSe versus ESe power law (γ) 0.7
Pixel size 75  μm, 139  μm

The x-ray scintillator material considered in this work was GOS due to the fact that its x-ray imaging properties have been extensively characterized in the literature both experimentally and theoretically. Optical transport in GOS was modeled using the Mie scattering model developed by Liaparinos et al., with a Mie scattering length of 3.67  μm and an anisotropy factor of 0.82.2123 An absorption length of 40 mm was used.24 The density (4.59  gcm3)22 and backing reflectivity (0.88)25 were taken to match those of the Lanex Fast Back GOS screen (Eastman Kodak Company), which has a thickness of 290  μm. However, the thickness of GOS may easily be increased up to 1000  μm.26 Accordingly, we simulated a range of GOS thicknesses between 0 and 1000  μm. a-Se was modeled as a uniform slab of selenium with a density of 4.29  gcm3.27 The thickness of a-Se used in direct AMFPIs is usually 200  μm28 for mammography and up to 1000  μm for general radiography and fluoroscopy. Therefore, we also simulated a-Se using a range of thicknesses from 0 to 1000  μm. The TFT substrate was modeled as borosilicate glass29 with a density of 2.23  gcm3 and a thickness of 700  μm. Each hybrid AMFPI configuration was simulated using both an FI and a BI geometry.

Each simulation comprised a pencil beam of 20,000 input x-rays using either RQA5 or RQA9 x-ray beam qualities [Fig. 3(b)]. These x-ray energy spectra were computed using the TASMIP spectral model30; the energy of each input x-ray in the simulations was determined by randomly drawing from these distributions. X-rays were sent with normal incidence along the z axis toward the detector at the xy plane’s origin, and the 2D spatial distribution of signal induced by each interacting x-ray was recorded. The results were then spatially binned in 2D to represent signal integration on a pixel matrix. A default pixel size of 75  μm (100% fill factor) was considered in this work, although this parameter is readily adjusted during postprocessing.

2.2. Simulation of the Hybrid AMFPI’s Single X-Ray Response

Signal charge is induced directly in a hybrid AMFPI by x-ray energy that is directly converted to charge in the a-Se photoconductor and indirectly by optical photons produced by x-ray interactions in the scintillator that are converted to charge upon reaching the scintillator–selenium interface. The former signal contribution is hereafter referred to as a “direct signal component” and the latter as an “indirect component.” In a hybrid structure, each interacting x-ray may either deposit its entire energy in the a-Se or scintillator layer or deposit a fraction of its energy in both layers through processes such as remote K-fluorescence reabsorption or x-ray scattering [Fig. 3(a)]. Consequently, the 2D spatial distribution of the signal induced by a single interacting x-ray may also be described in terms of its direct and indirect signal components, denoted as Sdirect(r) and Sindirect(r), respectively. The total spatial distribution of signal induced by each detected x-ray is

S(r)=Sdirect(r)+Sindirect(r), (1)

where r=(x,y).

For a given x-ray, Sdirect(r) was derived by collecting raw data sets consisting of the magnitude and location of the energy deposited by primary electrons produced by photoelectric or Compton interactions in a-Se. Charge generation was assumed to be local to each site of energy deposition, and lateral spread of charge during transit to the pixel electrode was assumed to be negligible.31 Due to x-ray and electron scattering, each incoming x-ray results in multiple sites of energy deposition. The number of electron–hole pairs (ehps) generated at the i’th site of energy deposition in a-Se (gSe,i) was modeled as a Poisson-distributed random variable with an average of

gSe,i=1000·EiWSe, (2)

where Ei denotes the amount of energy in keV deposited at i’th deposition site and WSe denotes the average amount of energy in eV required to generate one detected ehp. A default value of WSe=50  eV was assumed, corresponding to an electric field of ESe=10  Vμm1 within a-Se.32 For an x-ray resulting in N energy depositions in a-Se, Sdirect(r) is given by

Sdirect(r)=iNgSe,iδ(rri), (3)

where ri denotes the location of the i’th energy deposition. For each energy deposition, gSe,i was computed during postsimulation processing. This approach was used to evaluate the impact of different values of WSe (e.g., due to differences in applied bias voltage) using otherwise identical data sets.

Sindirect(r) was derived by collecting raw data sets consisting of the magnitude and location of energy deposited by primary electrons in the scintillator. The number of optical photons generated at the i’th site of energy deposition in GOS was approximated as a Poisson-distributed random variable with an average of

gGOS,i=YGOS·Ei, (4)

where YGOS is the scintillation yield of GOS, which was assumed to be 60  keV1 based on reports from the literature.3336 Optical photons created at the i’th energy deposition site in GOS reach the GOS–selenium interface with a depth-dependent probability ξ(zi), which is determined by the extent of light scattering and absorption within the scintillator. This process may be represented by a Bernoulli distributed random variable ξij, which has a value of 1 if the j’th optical photon due to the i’th energy deposition reaches a-Se and 0 otherwise. Each optical photon that reaches a-Se will produce an ehp with a probability β denoting the optical quantum efficiency of a-Se to GOS luminescence. This process is also represented by a Bernoulli distributed random variable βij. For a single x-ray interaction resulting in M energy depositions in GOS, Sindirect(r) is given by

Sindirect(r)=iMjgGOS,iβijξijδ(rrij),. (5)

where rij denotes the location of each optical photon when it reaches the GOS–selenium interface. For each optical photon, βij was computed during postsimulation processing. This approach was used to evaluate the impact of the optical quantum efficiency between the scintillator and a-Se using otherwise identical data sets.

2.3. Zero-Frequency Properties of the Hybrid AMFPI: PHS, AS, and DQE(0)

Zero-frequency detector performance metrics were determined by first integrating S(r) in 2D to find the total signal S induced by each single x-ray. The probability distribution of S is referred to as the PHS and was calculated by creating histograms of all S from a given simulation. The Swank factor, AS, is defined as follows:

AS=S2S2, (6)

and was calculated directly from this equation.12 AS characterizes the relative spread of the PHS; AS=1 corresponds to an infinitely narrow PHS [i.e., δ(S)], while AS=0 corresponds to an infinitely broad PHS. DQE(0) was found according to the following equation:12

DQE(0)=ηAS. (7)

2.4. Spatial-Frequency Dependent Properties of the Hybrid AMFPI: MTF(f), NPS(f), and DQE(f)

DQE(f) of each hybrid configuration was calculated using the FLS simulation approach described in detail by Star-Lack et al.22 Briefly, this approach uses the following equation:

DQE(f)=MTF2(f)qNNPS(f), (8)

where MTF(f) is the detector’s presampling MTF, q is the x-ray input fluence, and NNPS(f) is the noise power spectrum divided by the mean squared signal. MTF(f) is determined using the slanted slit method of Fujita et al.37 We used an infinitely thin slit created during postprocessing by uniformly spacing each S(r) along a line angled at 3.57 deg with respect to the x axis, corresponding to an oversampling rate of 16.37 Next, the product qNNPS(f) is determined by first calculating its frequency dependence. This is done by integrating each S(r) along the y axis to yield the 1D single x-ray imaging response S(x). The frequency dependence is then given by

qNNPS(f)|S(f)|2, (9)

where S(f) denotes the discrete Fourier transform of S(x).13 This expression is then normalized to the correct zero-frequency value, which is found by combining Eqs. (7) and (8) to yield

qNNPS(0)=1ηAS. (10)

The qNNPS(f) may then be calculated as follows (Fig. 4):

qNNPS(f)=qNNPS(0)|S(f)|2|S(0)|2=1ηAS|S(f)|2|S(0)|2. (11)

Fig. 4.

Fig. 4

A schematic depicting the methods used to calculate various detector performance metrics. Row 1: By integrating each single x-ray response, S(r), in 2D, the total signal magnitude due to each x-ray, S, is derived. Each measurement of S is then binned in a histogram to determine the PHS, which is used to calculate the Swank factor AS and DQE(0). Row 2: The S(r) ensemble is spaced uniformly along an infinitely thin slanted slit, from which MTF(f) is computed via Fujita’s method. Row 3: Each S(r) is integrated along the y-axis to derive S(x). Fourier analysis is performed to compute qNNPS(f). DQE(f) is then found from MTF(f) and qNNPS(f).

2.5. X-Ray Cross Talk

Due to the multilayer structure of the hybrid AMFPI, processes such as remote K-fluorescence reabsorption and x-ray scatter can result in x-ray “cross talk,” where an x-ray that initially interacts in one x-ray converter also deposits energy in the other converter. As seen in Fig. 3(a), an x-ray’s direct signal component may be generated in two ways through direct interaction with a-Se (types 1 and 3) or through x-ray cross talk from GOS into a-Se (type 4). Similarly, an x-ray’s indirect signal component may be generated in two ways: through x-ray interactions in GOS that result in optical photons that reach the a-Se/GOS interface (types 2 and 4) or through x-ray cross talk from a-Se into GOS (type 3).

The following approach was used to separately evaluate the impact of cross talk (i.e., interactions of type 3 and type 4, respectively) on the performance of a BI hybrid AMFPI comprising 300  μm of both GOS and a-Se under RQA5 beam conditions. PHS and DQE(f) were calculated using only the indirect component of each x-ray’s signal and were compared for the cases of either including or excluding all signals due to cross talk. Similarly, PHS and DQE(f) were calculated using only the direct component of each x-ray’s signal and were compared for the cases of either including or excluding all signals due to cross talk.

2.6. Gain Matching

Each detected x-ray in a hybrid AMFPI will experience a different x-ray-to-charge conversion gain depending on whether it initially interacts with a-Se or GOS. These random gain fluctuations degrade the detector’s AS, thereby reducing its DQE(0) according to Eq. (7). For a given amount of absorbed x-ray energy, the conversion gain of GOS is determined by YGOS, ξ(z), and β [see Eqs. (4) and (5)], while that of a-Se is determined by WSe [see Eqs. (2) and (3)]. This results in a design consideration referred to as “gain matching” that is unique to multilayer detectors, wherein As may be optimized by adjusting YGOS, ξ(z), β, and WSe to minimize variation in conversion gain between layers. We confined our analysis to β and WSe since both can be easily altered by adjusting the internal electric field of a-Se, ESe, while β can also be altered by doping a-Se. The following approach was used to evaluate the impact of gain matching on the performance of a BI hybrid AMFPI comprising 300  μm of both GOS and a-Se under RQA5 beam conditions.

DQE(0) was calculated as a function of both β and WSe using

DQE(0)=ηS(β,WSe)2S(β,WSe)2, (12)

where S(β,WSe) was computed for each x-ray via postprocessing of simulation data. Due to the excessive computation involved in processing the same raw dataset for each pair (β,WSe), we first calculated Sindirect and Sdirect for β=1 and WSe=50  eV. We then estimated S(β,WSe) as follows:

S(β,WSe)=β·Sindirect+(50/WSe)·Sdirect, (13)

which is valid for large values of Sindirect and Sdirect.

DQE(0) was also calculated as a function of ESe. Values from ESe=0  Vμm1 to ESe=35  Vμm1 were considered. WSe was assumed to follow the empirically derived proportionality:

WSeESeγ, (14)

where γ is a fitting parameter with a value of 0.7.38,39 WSe was normalized to a value of 50 eV at ESe=10  Vμm1. β was assumed to follow the Onsager mechanism, where β(r0,ESe) is a function of both ESe and an initial separation r0 of ehps, which is both wavelength (λ) and temperature (T) dependent, given by

β(r0,E)=eAeBBm=0n=0Amm!Bm+n(m+n)!, (15)

where A=e2/4πκε0kTr0 and B=er0ESe/kT.40 The values of β, WSe, and ESe that resulted in optimal DQE(0) were determined.

Additionally, two PHS were calculated: PHSa-Se which only included x-rays that initially interacted with a-Se (interactions of types 1 and 3), and PHSGOS, which only included those that initially interacted in GOS (interactions of types 2 and 4). The total PHS of the hybrid AMPFI is given by the sum:

PHS=PHSGOS+PHSa-Se. (16)

From Eq. (7), we expected DQE(0) to be highest when values of β and WSe were chosen that resulted in the narrowest combined PHS (i.e., the highest AS).

3. Results

3.1. Effect of Irradiation Geometry on DQE(f)

Figure 5 compares DQE(f) of a hybrid AMFPI using 300  μm of both GOS and a-Se in FI versus BI geometry at both (a) RQA5 and (b) RQA9. Optimal gain matching was assumed, using β=0.44 and WSe=50  eV. For each beam quality and combination of layer thicknesses, BI hybrid AMFPIs had higher DQE(f) than if they were FI, particularly at higher f. We shall, therefore, confine subsequent results to BI hybrid AMFPI configurations.

Fig. 5.

Fig. 5

DQE(f) of a hybrid AMFPI in FI and BI geometries for both (a) RQA5 and (b) RQA9 beam qualities. Both a-Se and GOS thicknesses were 300  μm, and optimal gain matching was assumed using β=0.44 and WSe=50  eV. BI results in a significant increase in DQE(f) at higher frequencies compared to FI.

3.2. Effect of Thickness of Both a-Se and the Scintillator on DQE(f)

Figure 6 shows DQE(f) of hybrid AMFPIs using various a-Se thicknesses and a fixed GOS thickness (300  μm). Optimal gain matching was assumed, using β=0.44 and WSe=50  eV. Increasing a-Se thickness increases DQE(f) at all f, with diminishing returns after 700  μm. Improvement in DQE(f) is greater when using an RQA5 spectrum compared with RQA9.

Fig. 6.

Fig. 6

DQE(f) of hybrid AMFPIs with a GOS thickness of 300  μm and various a-Se thicknesses at (a) RQA5 and (b) RQA9. Optimal gain matching was assumed using β=0.44 and WSe=50  eV. Greater a-Se thickness increases DQE(f) at all frequencies, particularly when using lower energy x-ray spectra such as RQA5.

Figure 7 shows DQE(f) of hybrid AMFPIs using various GOS thicknesses and a fixed a-Se thickness (300  μm). Optimal gain matching was assumed, using β=0.44 and WSe=50  eV. The addition of GOS to a-Se (solid curves) greatly increases DQE(f) at lower f with smaller improvements at higher f. As f increases, the hybrid AMFPI’s DQE(f) approaches that of a-Se alone. As GOS thickness increases, DQE(f) continues to improve, but with diminishing returns; this improvement becomes restricted to increasingly lower f. Improvement of DQE(f) with GOS thickness is more significant when using the RQA9 spectrum as opposed to RQA5. When using an a-Se thickness of 700  μm and a GOS thickness of 1000  μm (dotted curves), hybrid AMFPI DQE(f) outperforms or is comparable to DQE(f) of a typical CsI indirect AMPFI (dashed curve41) at all f; this is particularly the case for low f when using an RQA9 spectrum, for which an improvement from 47% to 76% in DQE(0) is observed.

Fig. 7.

Fig. 7

DQE(f) of hybrid AMFPIs with an a-Se thickness of 300  μm and various GOS thicknesses at (a) RQA5 and (b) RQA9. Optimal gain matching was assumed using β=0.44 and WSe=50  eV. Greater GOS thickness increases DQE(f) at lower frequencies, particularly when using higher energy x-ray spectra such as RQA9.

3.3. Effect of Gain Matching on DQE(0)

Figure 8 shows PHSGOS and PHSa-Se of a hybrid AMFPI (dSe=300  μm, dGOS=300  μm) using different values of β and WSe under RQA5 beam conditions. Average gains are indicated by dashed lines. Reducing β from 1 to 0.44 [Figs. 8(a) and 8(b)] brings both averages closer together and improves DQE(0) from 0.589 to 0.678, a 9% increase on the absolute scale. An identical improvement in DQE(0) is obtained by reducing WSe by the same factor of 0.44 [Fig. 8(c)].

Fig. 8.

Fig. 8

PHSGOS (blue) and PHSa-Se (magenta) using different values of β and WSe. Greater overlap increases DQE(0). Dashed lines indicate averages. (a) β=1, WSe=50  eV; (b) β=0.44, WSe=50  eV; and (c) β=1, WSe=22  eV.

Figure 9(a) shows DQE(0) as a function of β, with WSe=50  eV, compared with DQE(0) of a-Se alone with WSe=50  eV. When β=0, gain matching is poorest, resulting in a DQE(0) of 0.410. DQE(0) then increases with β until it reaches a maximum at β=0.44, after which DQE(0) decreases. Note that, regardless of the value of β, DQE(0) of the hybrid AMFPI is still greater than that of a-Se alone.

Fig. 9.

Fig. 9

(a) Hybrid AMFPI DQE(0) (dark blue) as a function of β, with WSe=50  eV. (b) Hybrid AMFPI DQE(0) (magenta) as a function of WSe with β=1. DQE(0) of a-Se alone with WSe=50  eV is shown in teal in both plots.

Figure 9(b) shows DQE(0) as a function of WSe, with β=1, compared with DQE(0) of a-Se alone with WSe=50  eV. At 1/WSe=0 (i.e., WSe), DQE(0) of the hybrid AMFPI approaches that of GOS alone (with a reduction due to x-ray attenuation by a-Se), which is slightly less than that of a-Se alone. As 1/WSe increases, DQE(0) increases until a maximum is attained when WSe=22  eV, after which DQE(0) decreases.

Figure 10(a) shows a contour plot of DQE(0) as a function of both β and WSe. DQE(0) of a-Se alone (0.410) is attained at β=0, whereas a DQE(0) of GOS alone (0.301) is attained at 1/WSe=0 (with a reduction due to x-ray attenuation by a-Se). A maximum DQE(0) of 0.678 is achieved for all β and WSe combinations lying on the dark red band, corresponding to optimal gain matching. The curves shown in black show β and WSe as functions of ESe at different wavelengths of optical light, λ. ESe intervals of 5  Vμm1 are marked with asterisks on each curve from 0 to 35  Vμm1 (left to right).

Fig. 10.

Fig. 10

(a) A contour plot of DQE(0) as a function of β and WSe. Warmer colors indicate higher values. DQE(0) reaches a maximum of 0.678 (dark red) whenever optimal gain matching is achieved. Black curves trace paths through β and WSe as a function of ESe for various values of λ. 5  Vμm1 intervals are marked by asterisks. (b) DQE(0) versus ESe for various values of λ. When ESe is larger than 5  Vμm1, DQE(0) is primarily a function of λ.

Figure 10(b) shows curves of DQE(0) versus ESe, where the optical emissions of GOS have been artificially set to various values of λ. This is equivalent to taking the value of DQE(0) at each point along the black curve corresponding to λ in Fig. 10(a). When ESe is greater than about 5  Vμm1, DQE(0) primarily depends on λ and remains relatively constant with ESe. At a given ESe, DQE(0) is highest between 400 and 450 nm and decreases with λ.

Figure 11(a) shows the x-ray induced luminescence spectrum of GOS plotted against the spectral sensitivity curves of a-Se both with42,43 and without40 tellurium (Te) doping. While the main peak of GOS is at 550 nm, 38.6% of its emissions are below 500 nm. This results in an effective optical quantum efficiency of β=46.7% and β=13.5% with and without Te doping, respectively. Figure 11(b) shows DQE(f) of 300  μm of a-Se alone (dark blue) and a hybrid AMFPI (teal), where dSe=300  μm, dGOS=300  μm, WSe=50  eV, and β=13.5%. DQE(f) of the same hybrid AMFPI, but with β=46.7%, is shown in red. The addition of GOS results in a large increase in DQE(f) at lower frequencies, particularly for Te-doped a-Se.

Fig. 11.

Fig. 11

(a) The x-ray induced luminescence spectrum of GOS and the spectral sensitivity of a-Se both with and without Te doping. (b) DQE(f) of a-Se alone (blue) and the hybrid AMFPI using GOS both with (red) and without (teal) Te doped a-Se.

3.4. Effect of Cross Talk between a-Se and the Scintillator on DQE(f)

Figure 12(a) shows the PHS of the indirect signal component of a hybrid AMFPI (dSe=300  μm, dGOS=300  μm) using optimal gain matching (β=0.44 and WSe=50  eV) under RQA5 beam conditions. Counts due to x-rays initially interacting in GOS are shown in blue, while the additional counts due to cross talk from a-Se are shown in green. The main peak of GOS is centered at 900  ehp, while the K-fluorescence escape peak is centered at 200  ehp. Cross talk results in an additional peak (shown in green) at 280  ehp, which corresponds to the absorption of Se K-fluorescence in GOS. The impact of x-rays scattering from a-Se into GOS is negligible. When β=1 and WSe=50  eV, cross talk from a-Se into GOS constitutes 0.88% of the total hybrid AMFPI signal (see Table 2). For an optimally gain matched detector, this decreases to 0.59%.

Fig. 12.

Fig. 12

(a) PHS consisting of only the indirect component of signal. Counts due to GOS x-ray interactions are shown in blue, while those due to a-Se cross talk are shown in green. Cross talk results in an additional peak at 280  ehp due to a-Se K-fluorescence, which is negligible in magnitude. (b) PHS consisting of only the direct component of signal. Cross talk results in two additional peaks contained within that of a-Se due to gadolinium K-fluorescence. Counts due to a-Se x-ray interactions are shown in blue, while those due to GOS cross talk are shown in green. (c) Hybrid AMFPI DQE(f) of both direct and indirect signal components is shown with and without cross talk. Only cross talk from GOS into Se was significant, resulting in an increase at low f with a degradation at all other f.

Table 2.

A summary of the percent contribution of various types of energy deposition to total signal magnitude using different gain matching parameters. Energy deposited in layer B due to an x-ray initially interacting with layer A is labeled AB. The contribution of a-SeGOS was negligible. The contribution of GOSa-Se was inversely related to β and reaches a maximum of 11.10% at β=0. The values β=0.44 and WSe=50  eV are an example of parameters that lead to optimal gain matching.

  β=1, WSe=50  eV β=0.44, WSe=50  eV β=0
% GOSGOS 61.19 41.29 0
% a-SeGOS 0.88 0.59 0
% a-Sea-Se 33.72 51.67 88.90
% GOSa-Se 4.22 6.45 11.10

Figure 12(b) shows the PHS of the direct signal component of the same hybrid AMFPI under the same beam conditions. Counts due to x-rays initially interacting in a-Se are shown in blue, while the additional counts due to cross talk from GOS are shown in green. a-Se interactions result in a single peak centered at 925 ehp. K-fluorescence from gadolinium (Kα43  keV and Kβ49  keV) results in two additional peaks, which are contained within the peak of a-Se. Scatter from GOS into a-Se is negligible. When β=1 and WSe=50  eV, cross talk from GOS into a-Se constitutes 4.22% of the total hybrid AMFPI signal (see Table 2). This grows to 6.45% in an optimally gain matched detector and 11.10% when β=0.

Figure 12(c) shows the DQE(f) of direct and indirect signal components both with (solid lines) and without cross talk (dashed lines). The addition of cross talk increases DQEdirect(f) at very low frequencies but slightly degrades DQEdirect(f) for all other frequencies. The impact of cross talk on DQEindirect(f) is negligible at all frequencies.

4. Discussion

4.1. Effect of Irradiation Geometry on DQE(f)

Superior DQE(f) was observed in a BI hybrid AMFPI geometry for two reasons: First, BI increases the fraction of x-rays that interact with a-Se, which has a superior spatial resolution compared with GOS. Second, BI increases the imaging performance of GOS by ensuring that more x-rays are absorbed closer to the optical sensor, i.e., a-Se. This results in a greater light-escape efficiency, less optical blur, and a reduction in Lubberts effect.16

4.2. Effect of Thickness of Both a-Se and Scintillator on DQE(f)

Our results demonstrate that DQE(f) of a-Se alone provides a lower bound for DQE(f) of the hybrid AMFPI, while GOS provides additional low frequency DQE(f). The addition of indirect signal increases the width of a direct AMFPI’s PHS and, thus, decreases its AS (see Fig. 8). However, the improvement in η made through increasing dGOS apparently outweighs this penalty to improve DQE(f) overall. Improvement in DQE(f) with dGOS is greater for RQA9 due to the fact that GOS is more efficient at absorbing the higher energy x-rays. Conversely, improvement in DQE(f) with da-Se is greater for RQA5 due to the fact that x-rays are first incident on a-Se, which is able to absorb the lower energy x-rays in RQA5. Figure 6 suggests that there is little advantage to fabricating hybrid AMFPIs with da-Se>700  μm as DQE(f) does not significantly increase with da-Se beyond this point for either RQA5 or RQA9. Figure 7 suggests that a large GOS thickness of 1000  μm would be optimal for a hybrid AMFPI utilizing GOS.

4.3. Effect of Gain Matching on DQE(0)

Our results showed that the shapes of PHSGOS and PHSa-Se may be stretched along the x axis by β and 1/WSe, respectively. Values of β and WSe that produce the narrowest combined PHS (relative to its average) maximize AS and, thus, DQE(0). While many pairs of β and WSe produce the same DQE(0), the optimal choice for a practical detector at low dose is that which provides the highest average gain to overcome electronic noise. Though this work used an RQA5 spectrum to demonstrate gain matching, optimal values of β and WSe did not significantly vary between RQA5 and RQA9.

The strong λ dependence of β [and thus DQE(0)] highlights the importance of scintillator choice in optimizing hybrid AMFPI performance. As seen in Fig. 10(a), a-Se is predominantly sensitive to blue light (below 500 nm) and has limited sensitivity to green light (500 to 565 nm). An ideal scintillator for the hybrid AMFPI would therefore be a blue-emitting scintillator with a high light yield. Because 38.6 % of GOS’s optical emissions lie below 500 nm, β=13.5% and a large increase in DQE(f) is realized at low f. However, if a scintillator’s emissions have a sufficiently large λ that β approaches 0, merely altering ESe may not yield optimal gain matching [e.g., λ=550  nm, Fig. 10(b)]. In such cases, it is possible to augment β by doping a-Se. For example, the SATICON camera uses Te doping of a-Se to increase β for green and red light, resulting in β=46.7%.42,43 Our model suggests that a hybrid AMFPI using GOS paired with Te doped a-Se achieves near-ideal gain matching.

4.4. Effect of Cross Talk between a-Se and the Scintillator on DQE(f)

Cross talk from a-Se into GOS was not found to be a significant effect for hybrid AMFPI performance. On the other hand, cross talk from GOS into a-Se could potentially impact performance, particularly for low values of β. This is due to the high energy of GOS K-fluorescence (Kα43  keV and Kβ49  keV), resulting in a higher signal magnitude for cross talk. The slight improvement of DQEdirect(0) with cross talk in Fig. 12(c) is due to the additional η provided by the detection of GOS K-fluorescence and scatter. This also explains the fact that DQE(0) improved even when β=0 in Fig. 9(a). While the cross talk analysis presented in this work considered RQA5 beam conditions, the impact of cross talk was found to be similar under RQA9 conditions as well.

4.5. Limitations and Future Work

Our model used GOS as an example of a scintillator that may be used in the hybrid AMFPI. In the future, we will develop and implement MC simulations of other types of scintillators used in AMFPI, e.g., CsI. Our model did not include electronic noise; thus, our results represent a high dose limit of hybrid AMFPI performance. Since the hybrid AMFPI benefits from signal from both direct and indirect detection layers, we expect the degradation effect of electronic noise to be less than that in either the direct or indirect AMFPI alone. Compared to indirect AMFPI, the effect of electronic noise at high spatial frequencies is mitigated by the direct detection events.44 Compared to direct AMFPI, the low dose performance benefits from the boost of x-ray quantum noise at low spatial frequencies, which is afforded by the x-ray absorption in the scintillator.44 Effects such as improved low-frequency DQE(f) via coupling of scintillators to a-Se have been validated in preliminary experimental work using the first ever prototype hybrid AMFPI.45 The impact of factors, such as scintillator choice, x-ray converter thickness, gain matching, x-ray cross talk, and electronic noise, on hybrid AMFPI performance will be the subject of future experimental investigations with prototype detectors.

5. Conclusions

In this work, we used MC simulation to model and evaluate the potential imaging performance of the hybrid AMFPI, a novel hybrid direct–indirect AMFPI comprising a scintillator coupled to an a-Se direct detector. In a BI geometry, the addition of the GOS to a-Se substantially improved DQE(f) at lower f by increasing x-ray quantum efficiency. Larger a-Se thicknesses resulted in better performance, with diminishing returns after 700  μm. Further improvement in DQE(f) was achieved by increasing GOS from 300 to 1000  μm, which yielded superior or comparable performance to that of a typical CsI indirect AMFPI at all f. Gain matching of direct/indirect signal components was shown to play a crucial role in hybrid AMFPI optimization and is affected by choice of scintillator, ESe, and doping of a-Se. When paired with Te doping, GOS achieved near-ideal gain matching with a-Se.

Acknowledgments

The authors gratefully acknowledge financial support from the U.S. National Institutes of Health (Grant Nos. R01 EB002655 and R01 EB026267). We also thank Dr. Kenkichi Tanioka and members of the Digital Radiological Imaging Laboratory (DRIL) at Stony Brook University for useful discussions.

Biographies

Scott Dow has been a graduate student in DRIL Laboratory at Stony Brook University since 2016. His research interests are in medical imaging physics.

Adrian Howansky is a medical physics resident at Stony Brook University Hospital, Department of Radiology. He received his PhD in biomedical engineering from Stony Brook. His research interests are in medical imaging physics.

Anthony R. Lubinsky is a principal investigator at the State University of New York at Stony Brook. He received his PhD in physics from Northwestern University. His research interests are in physics, imaging technology, and materials science for medical imaging systems.

Wei Zhao is a professor of radiology at the State University of New York at Stony Brook. She received her PhD in medical biophysics from the University of Toronto. Her research interests are in medical image sensors and digital breast tomosynthesis.

Disclosures

The authors have no financial interests to disclose.

Contributor Information

Scott Dow, Email: scott.dow.1@stonybrook.edu.

Adrian Howansky, Email: Adrian. Howansky@stonybrookmedicine.edu.

Anthony R. Lubinsky, Email: anthony.lubinsky@stonybrookmedicine.edu.

Wei Zhao, Email: wei.zhao@stonybrook.edu.

References

  • 1.Rowlands J. A., Yorkston J., “Flat panel detectors for digital radiography,” in Handbook of Medical Imaging, Van Metter R. L., Beutel J., Kundel H. L., Eds., Vol. 1, pp. 223–328, SPIE Press, Bellingham, Washington: (2000). [Google Scholar]
  • 2.Kabir M. Z., Kasap S., “Photoconductors for x-ray image detectors,” in Springer Handbook of Electronic and Photonic Materials, Kasap S., Capper P., Eds., pp. 1–1, Springer, Cham: (2017). [Google Scholar]
  • 3.Jiang H., et al. , “Development of active matrix flat panel imagers incorporating thin layers of polycrystalline HgI2 for mammographic x-ray imaging,” Phys. Med. Biol. 58(3), 703 (2013). 10.1088/0031-9155/58/3/703 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Semeniuk O., et al. , “Characterization of polycrystalline lead oxide for application in direct conversion x-ray detectors,” Sci. Rep. 7(1), 8659 (2017). 10.1038/s41598-017-09168-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Yakunin S., et al. , “Detection of x-ray photons by solution-processed lead halide perovskites,” Nat. Photonics 9(7), 444 (2015). 10.1038/nphoton.2015.82 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hunt D. C., Tousignant O., Rowlands J. A., “Evaluation of the imaging properties of an amorphous selenium-based flat panel detector for digital fluoroscopy,” Med. Phys. 31(5), 1166–1175 (2004). 10.1118/1.1707755 [DOI] [PubMed] [Google Scholar]
  • 7.Hunt D. C., Zhao W., Rowlands J. A., “Detective quantum efficiency of direct flat-panel x-ray imaging detectors for fluoroscopy,” Proc. SPIE 3336, 195–201 (1998). 10.1117/12.317041 [DOI] [Google Scholar]
  • 8.Tousignant O., et al. , “Spatial and temporal image characteristics of a real-time large area a-Se x-ray detector,” Proc. SPIE 5745, 207–215 (2005). 10.1117/12.602183 [DOI] [Google Scholar]
  • 9.Choquette M., et al. , “Performance of a real-time selenium-based x-ray detector for fluoroscopy,” Proc. SPIE 4320, 501–508 (2001). 10.1117/12.430873 [DOI] [Google Scholar]
  • 10.Yaffe M. J., Mainprize J. G., “Detectors for digital mammography,” Technol. Cancer Res. Treatment 3(4), 309–324 (2004). 10.1177/153303460400300401 [DOI] [PubMed] [Google Scholar]
  • 11.Kiarashi N., Samei E., “Digital breast tomosynthesis: a concise overview,” Imaging Med. 5(5), 467–476 (2013). 10.2217/iim.13.52 [DOI] [Google Scholar]
  • 12.Swank R. K., “Absorption and noise in x-ray phosphors,” J. Appl. Phys. 44(9), 4199–4203 (1973). 10.1063/1.1662918 [DOI] [Google Scholar]
  • 13.Lubberts G., “Random noise produced by x-ray fluorescent screens,” J. Opt. Soc. Am. 58(11), 1475–1483 (1968). 10.1364/JOSA.58.001475 [DOI] [Google Scholar]
  • 14.Sato K., et al. , “Effect of x-ray incident direction and scintillator layer design on image quality of indirect-conversion flat-panel detector with GOS phosphor,” Proc. SPIE 7961, 79614I (2011). 10.1117/12.877752 [DOI] [Google Scholar]
  • 15.Lubinsky A. R., et al. , “Back-irradiated and dual-screen sandwich detector configurations for radiography,” J. Med. Imaging 6(3), 033501 (2019). 10.1117/1.JMI.6.3.033501 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Howansky A., et al. , “Comparison of CsI: Tl and Gd2O2S: Tb indirect flat panel detector x-ray imaging performance in front-and back-irradiation geometries,” Med. Phys. 46(11), 4857–4868 (2019). 10.1002/mp.13791 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Rottmann J., et al. , “A novel EPID design for enhanced contrast and detective quantum efficiency,” Phys. Med. Biol. 61(17), 6297–6306 (2016). 10.1088/0031-9155/61/17/6297 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Maurino S. L., “Design and optimization of a stacked three-layer x-ray detector for multi-spectral medical imaging applications,” Thesis, University of Waterloo (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Dow S., Howansky A., Zhao W., “Evaluation of novel multilayer x-ray detector designs using rapid Monte Carlo simulation,” Proc. SPIE 10573, 105730K (2018). 10.1117/12.2293944 [DOI] [Google Scholar]
  • 20.Agostinelli S., et al. , “GEANT4—a simulation toolkit,” Nucl. Instrum. Methods Phys. Res. Sect. A 506(3), 250–303 (2003). 10.1016/S0168-9002(03)01368-8 [DOI] [Google Scholar]
  • 21.Liaparinos P. F., et al. , “Modeling granular phosphor screens by Monte Carlo methods,” Med. Phys. 33(12), 4502–4514 (2006). 10.1118/1.2372217 [DOI] [PubMed] [Google Scholar]
  • 22.Star-Lack J., et al. , “Rapid Monte Carlo simulation of detector DQE(f),” Med. Phys. 41(3), 031916 (2014). 10.1118/1.4865761 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Zyazin A. S., Peters I. M., “Complete optical stack modeling for CMOS-based medical x-ray detectors,” Proc. SPIE 9412, 94122H (2015). 10.1117/12.2081836 [DOI] [Google Scholar]
  • 24.Kausch C., et al. , “Monte Carlo simulations of the imaging performance of metal plate/phosphor screens used in radiotherapy,” Med. Phys. 26(10), 2113–2124 (1999). 10.1118/1.598727 [DOI] [PubMed] [Google Scholar]
  • 25.Poludniowski G. G., Evans P. M., “Optical photon transport in powdered-phosphor scintillators. Part II. Calculation of single-scattering transport parameters,” Med. Phys. 40(4), 041905 (2013). 10.1118/1.4794485 [DOI] [PubMed] [Google Scholar]
  • 26.Kingsley J. W., et al. , “Detecting 6 MV X-rays using an organic photovoltaic device,” Org. Electron. 10(6), 1170–1173 (2009). 10.1016/j.orgel.2009.06.006 [DOI] [Google Scholar]
  • 27.Saunders A., “The allotropic forms of selenium,” J. Phys. Chem. 4, 424–513 (1900). 10.1021/j150024a001 [DOI] [Google Scholar]
  • 28.Hu Y. H., Zhao W., “The effect of amorphous selenium detector thickness on dual-energy digital breast imaging,” Med. Phys. 41(11), 111904 (2014). 10.1118/1.4897244 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Kloss T., Lautenschläger G., Schneider K., “Advances in the process of floating borosilicate glasses and some recent applications for specialty borosilicate float glasses,” Glass Technol. 41(6), 177–181 (2000). [Google Scholar]
  • 30.Boone J. M., Seibert J.A., “An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV,” Med. Phys. 24(11), 1661–1670 (1997). 10.1118/1.597953 [DOI] [PubMed] [Google Scholar]
  • 31.Zhao W., et al. , “Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: characterization of a small area prototype detector,” Med. Phys. 30(2), 254–263 (2003). 10.1118/1.1538233 [DOI] [PubMed] [Google Scholar]
  • 32.Blevis I., Hunt D., Rowlands J., “Measurement of x-ray photogeneration in amorphous selenium,” J. Appl. Phys. 85(11), 7958–7963 (1999). 10.1063/1.370615 [DOI] [Google Scholar]
  • 33.Van Eijk C. W., “Inorganic scintillators in medical imaging,” Phys. Med. Biol. 47(8), R85 (2002). 10.1088/0031-9155/47/8/201 [DOI] [PubMed] [Google Scholar]
  • 34.Ludwig G., “X-ray efficiency of powder phosphors,” J. Electrochem. Soc. 118(7), 1152–1159 (1971). 10.1149/1.2408271 [DOI] [Google Scholar]
  • 35.Wickersheim K., Alves R., Buchanan R., “Rare earth oxysulfide x-ray phosphors,” IEEE Trans. Nucl. Sci. 17(1), 57–60 (1970). 10.1109/TNS.1970.4325559 [DOI] [Google Scholar]
  • 36.Howansky A., et al. , “An apparatus and method for directly measuring the depth-dependent gain and spatial resolution of turbid scintillators,” Med. Phys. 45(11), 4927–4941 (2018). 10.1002/mp.13177 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Fujita H., et al. , “A simple method for determining the modulation transfer function in digital radiography,” IEEE Trans. Med. Imaging 11(1), 34–39 (1992). 10.1109/42.126908 [DOI] [PubMed] [Google Scholar]
  • 38.Stone M. F., et al. , “The x-ray sensitivity of amorphous selenium for mammography,” Med. Phys. 29(3), 319–324 (2002). 10.1118/1.1449874 [DOI] [PubMed] [Google Scholar]
  • 39.Que W., Rowlands J. A., “X-ray photogeneration in amorphous selenium: geminate versus columnar recombination,” Phys. Rev. B 51(16), 10500–10507 (1995). 10.1103/PhysRevB.51.10500 [DOI] [PubMed] [Google Scholar]
  • 40.Pai D. M., Enck R. C., “Onsager mechanism of photogeneration in amorphous selenium,” Phys. Rev. B 11(12), 5163–5174 (1975). 10.1103/PhysRevB.11.5163 [DOI] [Google Scholar]
  • 41.Samei E., Murphy S., Christianson O., “DQE of wireless digital detectors: comparative performance with differing filtration schemes,” Med. Phys. 40(8), 081910 (2013). 10.1118/1.4813298 [DOI] [PubMed] [Google Scholar]
  • 42.Jackson K. G., Townsend G. B., TV & Video Engineer’s Reference Book, Elsevier, Amsterdam: (2014). [Google Scholar]
  • 43.Goto N., et al. , “Saticon: a new photoconductive camera tube with Se-As-Te target,” IEEE Trans. Electron Devices 21(11), 662–666 (1974). 10.1109/T-ED.1974.17991 [DOI] [Google Scholar]
  • 44.Reiser I., Glick S., Tomosynthesis Imaging, Taylor & Francis Group, Boca Raton, Florida: (2014). [Google Scholar]
  • 45.Howansky A., et al. , “Initial characterization of a hybrid direct-indirect active matrix flat panel imager for digital radiography,” Proc. SPIE 11312, 113120X (2020). 10.1117/12.2549893 [DOI] [Google Scholar]

Articles from Journal of Medical Imaging are provided here courtesy of Society of Photo-Optical Instrumentation Engineers

RESOURCES