Abstract
Background: The aim of the present study was to compare, in low-density polyurethane blocks, the primary implant stability values (micromobility) and removal torque values of three different implant geometries in two different bone densities representing the structure of the human posterior jaws. Methods: A total of 60 implants were used in the present investigation: twenty implants for each of three groups (group A, group B, and group C), in both polyurethane 10 pcf and 20 pcf densities. The insertion torque, pull-out torque, and implant stability quotient (ISQ) values were obtained. Results: No differences were found in the values of Group A and Group B implants. In both these groups, the insertion torques were quite low in the 10 pcf blocks. Better results were found in the 20 pcf blocks, which showed very good stability of the implants. The pull-out values were slightly lower than the insertion torque values. High ISQ values were found in Group A and B implants. Lower values were present in Group C implants. Conclusions: The present investigation evaluated implants with different geometries that are available on the market, and not experimental implants specifically created for the study. The authors aimed to simulate real clinical conditions (poor-density bone or immediate post-extraction implants) in which knowledge of dental implant features, which may be useful in increasing the primary stability, may help the oral surgeon during the surgery planning.
Keywords: bone density, cylindrical implants, conical implants, implant stability quotient, insertion and pull-out torque, polyurethane foam blocks, primary stability
1. Introduction
Oral rehabilitation with dental implants represents a highly predictable procedure for partial and full edentulism, characterized by a success rate of over 90% [1]. The osseointegration of dental implants is determined by two different processes: the primary stability obtained with the mechanical engagement of the screw with the bone wall of the preparation site, and the secondary stability, due to new bone formation during the healing period [2,3].
The relationship between these two conditions is influenced by the absence of micromovements of the implant after its placement into the preparation site [3].
Implant stability is influenced by several factors: bone density [4], implant macro- and micro-geometry [5,6], and surgical technique [7,8,9]. In the case of poor bone quality, or immediate post-extraction implants, the primary stability could be improved by the selection of a specific implant geometry and thread design.
For this reason, it is important to compare the primary stability of different implant geometries in order to establish which implant shape or geometry could be useful for obtaining high primary fixation in each clinical condition.
Several factors could condition local bone density: sex, age, medical treatments, systemic diseases, but, generally, the posterior regions of the maxilla are clinically characterized by a lower density of bone tissue [10,11,12].
Many different non-invasive tests have been proposed to evaluate, clinically and in laboratory conditions, the primary stability of dental implants.
The resonance frequency analysis (RFA) is a repeatable technique for implant stability evaluation that provides a measurement of the micromovements of the implant positioned in the bone site [13,14,15,16].
On the contrary, the insertion and removal torque evaluation is a non-repeatable measurement of the mechanical friction between an implant and bone walls during implant insertion and unscrewing [16,17].
With the use of polyurethane block sheets, it is possible to standardize the mechanical response to the forces generated during implant positioning [18].
In fact, the in vitro simulation on polyurethane is able to overcome the anatomical limits of implant osteotomies performed on ex-vivo samples, such as human cadaveric bone [19], bovine and pig ribs [19,20,21], and rabbit tibiae [22].
The human cadaver bone or fresh animal bone shows quality modifications over time, and selecting various bone blocks of the same quality represents an operation with high variability. To overcome the possible biases due to differences in bone quality, it is possible to use synthetic bony blocks made of rigid polyurethane [23,24,25]. The uniformity and consistent properties of rigid polyurethane foam make it an ideal material for comparative mechanical tests of different implant designs.
Moreover, as reported in the literature, polyurethane blocks are commercially available in different densities and thicknesses, and as such, are able to simulate the different consistencies of bone tissues present in the different regions of the upper and lower jaws [13,14,26].
The aim of the present study was to compare, in low-density polyurethane blocks, the primary implant stability values (micromobility) and removal torque values of three different implant geometries in two different bone densities representing the structure of the human posterior jaws.
2. Materials and Methods
2.1. Implants
In total, 60 implants were used, divided into three groups: Group A had 20 implants (4 mm diameter and 10 mm length) with a Cone Morse connection and a conical shape (UN II, Implacil De Bortoli, Sao Paulo, Brasil); Group B also had 20 implants (4 mm diameter and 10 mm length) with a Cone Morse connection and a conical shape (UN III, Implacil De Bortoli, Sao Paulo, Brasil) (Figure 1 and Figure 2A,B); and Group C had 20 (4.1 mm × 10 mm length) cylindrical screw-shaped implants (RBM Restore, Keystone Dental, Burlington, MA, USA) (Figure 2C). The UN III macro design differed from the UN II implants regarding its larger thread, the lack of double thread pitch, a round, not self-tapping, apex, and the presence of healing chambers between the cutting surface of the threads. The UN II presented a narrow-thread shape, double thread pitch, and self-tapping apex. The surface treatment used was performed by blasting with microparticles (~100 μm) of titanium dioxide and followed by double etching with maleic acid. The surfaces of the Group C implants were sandblasted with 150 microns hydroxyapatite particles (Figure 1).
2.2. Polyurethane Blocks
The American Society for Testing and Materials (ASTM F-1839-08) has approved the use of polyurethane and has recognized it as a standard for testing instruments and oral implants for comparative testing of bone screws (“Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Test Orthopedic Devices for Instruments”). Different types of solid rigid polyurethane blocks, which were 120 mm × 170 mm × 31 mm foam blocks (SawBones H, Pacific Research Laboratories Inc., Vashon, WA, USA) with homogeneous densities, were selected for the present investigation. The densities of polyurethane foam were 10 pounds per cubic foot (pcf) and 20 pcf. A 1-mm thin layer of 30 pcf was added to all of these blocks in order to simulate real clinical conditions.
2.3. Implant Insertion
Group A and B implants were inserted using a lance drill, then a 2-mm bur at 1200 Rpm, and subsequently a conical 3.5-mm bur at 800 rpm with the implant insertion at 20 rpm. Group C implants were inserted following the protocol of the manufacturer by using an implant lance drill, a 2-mm drill (1600 rpm), and a 3-mm final drill (800 rpm). The handpiece was calibrated at a speed of 70 rpm and a torque of 30 Ncm. (Figure 3A,B). Torque values were taken with software (ImpDat Plus, East Lansing, MI, USA) installed on a digital card.
2.4. Insertion Torque and Removal Torque
The insertion torque (IT, Ncm) values indicated the force of the maximum clockwise movement that stripped bone. The investigation was conducted by a single operator (LC) who compared the torque insertion and the removal strength values of the Group A, B, and C implants. The study was conducted comparing the insertion torque and the removal strength values with a calibrated torque meter with a torque range of 5–80 N/cm. The final 1 mm insertion torque of the implants into the bone sheets was recorded. In the present study, mechanical torque gauges (Implacil De Bortoli, Sao Paulo, Brasil) were used to assess the insertion torque and the removal strength values.
2.5. Resonance Frequency Analysis
After implant insertion, primary stability was measured using resonance frequency analysis (RFA) values expressed in ISQ with hand-screwed Smart-Pegs (number 7 for group A and B implants and number 1 for the group C implants) (Osstell Mentor Device, Integration Diagnostic AB, Savadelen, Swden) (Figure 3B). The implant stability quotient (ISQ) ranged from 0 to 100 (measured between 3500 and 85,000 Hz) and was divided into low (<60 ISQ), medium (60–70 ISQ), and high stability (>70 ISQ). For each specimen, the RFA measurement was repeated two times. Measurements were performed in two orientations separated by a 90-degree angle, and the average ISQ values were calculated.
2.6. Statistical Evaluation
The normal distribution of the data was evaluated by the Shapiro–Wilks test and the differences between the study groups regarding insertion torque, removal, and RFA groups were analyzed by one-way analysis of variance (ANOVA) followed by Tukey post-hoc test. A p-value < 0.05 was considered statistically significant. Data treatment and statistical analysis were performed by Excel origin (Microsoft Company, Redmond, WA, USA) and StatPlus 6 software (AnalystSoft, Walnut, CA, USA). The accuracy of RFA assessment was evaluated by Bland–Altman and linear regression model to determine the agreement between the ISQ measurements.
3. Results
No differences were found in the values of the Group A and Group B implants. Regarding the implants for both of these groups, the insertion torque values were quite low in the 10 pcf blocks (16–28 Ncm) (Figure 4, Table 1).
Table 1.
Insertion Torque | 10 PCF | 20 PCF | ||||
---|---|---|---|---|---|---|
UN II (A) | UN III (B) | RBM (C) | UN II (D) | UN III (E) | RBM (F) | |
Mean | 17.00 | 16.80 | 7.00 | 29.10 | 31.10 | 13.30 |
Std. Deviation | 0.94 | 1.14 | 1.05 | 0.99 | 0.99 | 1.16 |
Better results were found in the 20 pcf blocks, which showed very good stability of the implants. The removal values for both the Group A and B implants were slightly lower than the insertion torque values. High ISQ values were found in both the Group A and B implants (57–80). The Group C implants, on the other hand, presented very low insertion torque values (6–12 Ncm) in both polyurethane densities (10 and 20 pcf) (Figure 4; Table 1 and Table 2). Also, the removal values were very low (5–10 Ncm) (Figure 5; Table 3 and Table 4). Furthermore, the ISQ values were also in the very low range (10–37) (Figure 6; Table 5 and Table 6).
Table 2.
Multiple Comparison Insertion Torque | Mean Diff | 95.00% CI of Diff | Adjusted p Value |
---|---|---|---|
A-B | 0.20 | −1.156 to 1.556 | >0.9999 |
B-C | 9.80 | 8.444 to 11.16 | <0.0001 |
A-C | 10.00 | 8.644 to 11.36 | <0.0001 |
D-E | −2.00 | −3.356 to −0.6439 | 0.0007 |
E-F | 17.80 | 16.44 to 19.16 | <0.0001 |
D-F | 15.80 | 14.44 to 17.16 | <0.0001 |
A-D | −12.10 | −13.46 to −10.74 | <0.0001 |
B-E | −14.30 | −15.66 to −12.94 | <0.0001 |
C-F | −6.30 | −7.656 to −4.944 | <0.0001 |
Table 3.
Removal Torque | 10 PCF | 20 PCF | ||||
---|---|---|---|---|---|---|
UN II (A) | UN III (B) | RBM (C) | UN II (D) | UN III (E) | RBM (F) | |
Mean | 11.10 | 12.00 | 6.20 | 27.10 | 27.10 | 9.40 |
Std. Deviation | 0.99 | 1.25 | 0.79 | 1.10 | 1.29 | 1.35 |
Table 4.
Multiple Comparison Removal | Mean Diff | 95.00% CI of Diff | Adjusted p Value |
---|---|---|---|
A-B | −0.90 | −2.378 to 0.5782 | 0.7585 |
B-C | 5.80 | 4.322 to 7.278 | <0.0001 |
A-C | 4.90 | 3.422 to 6.378 | <0.0001 |
D-E | 0.00 | −1.478 to 1.478 | >0.9999 |
E-F | 17.70 | 16.22 to 19.18 | <0.0001 |
D-F | 17.70 | 16.22 to 19.18 | <0.0001 |
A-D | −16.00 | −17.48 to −14.52 | <0.0001 |
B-E | −15.10 | −16.58 to −13.62 | <0.0001 |
C-F | −3.20 | −4.678 to −1.722 | <0.0001 |
Table 5.
RFA | 10 PCF | 20 PCF | ||||
---|---|---|---|---|---|---|
UN II (A) | UN III (B) | RBM (C) | UN II (D) | UN III (E) | RBM (F) | |
Mean | 58.05 | 57.90 | 12.85 | 76.65 | 76.60 | 36.40 |
Std. Deviation | 0.28 | 0.81 | 0.78 | 1.00 | 0.77 | 0.77 |
Table 6.
Multiple Comparison RFA | Mean Diff | 95.00% CI of Diff | Adjusted p Value |
---|---|---|---|
A-B | −2.25 | 0.15 | >0.9999 |
B-C | 20.80 | 45.05 | <0.0001 |
A-C | 0.15 | 45.20 | <0.0001 |
D-E | 18.70 | 0.05 | >0.9999 |
E-F | 20.95 | 40.20 | <0.0001 |
D-F | 18.55 | 40.25 | <0.0001 |
A-D | −1.55 | −18.60 | <0.0001 |
B-E | 3.35 | −18.70 | <0.0001 |
C-F | 0.05 | −23.55 | <0.0001 |
4. Discussion
Primary stability is defined as the mechanical engagement of the implant during positioning and represents an essential requirement for osseointegration [27].
The present investigation evaluated implants with different geometries that are present on the market and not experimental implants specifically created for the study. The authors aimed to simulate real clinical conditions (poor-density bone or immediate post-extraction implants) in which the knowledge of dental implant features, which may be useful in increasing the primary stability, could help the oral surgeon to plan the surgery.
Regarding the macro-design, a cylindrical shape could lead to a higher bone-to-implant contact percentage, if compared with a conical geometry, providing that local factors such as depth of positioning, bone density, and surgical technique are equal [16,18,28]. Micro-geometry, threads pitch, and surface roughness could also significantly influence the primary stability [5,29,30].
The use of a polyurethane model represents a cost-effective technique for evaluating the mechanical properties of dental implant protocols. This material could be tested in several different forms, such as solid foams [13,31], cellular rigid blocks [32], and blocks with composite densities [31,33,34], thus allowing researchers to analyze experimental measurements recorded on a structurally homogeneous substrate while overcoming the anatomical and ethical limits of ex vivo investigations.
Moreover, the present investigation tested a polyurethane density of 10 pcf and 20 pcf, equivalent, respectively, to 0.16 g/cm3 and 0.32 g/cm3, covered by a layer of cortical thickness to simulate the situation in the posterior region of the maxilla.
In the literature, only a few studies have evaluated dental implant primary stability on low-density double-layered polyurethane, which demonstrates the novelty of the present investigation; the authors have reason to believe that low-density double-layered polyurethane provides a simulation closer to clinical conditions encountered for surgeries on humans [35,36].
In fact, Devlin et al. reported in humans a mean bone mineral density of the posterior maxilla of 0.31 g/cm3 compared to the anterior maxilla, which showed a mean density of 0.55 g/cm3.
In the present study, the Group A and B implants showed high stability in low-density polyurethane blocks, with no statistical differences between the two groups, notwithstanding the diversities of the apex and thread profiles. Group C implants showed lower values.
Within the limitations of the present in vitro model, the investigation outcomes could be interpreted as a dominant effect of the implant macro-design (such as tapered and straight, conical and cylindrical geometry) rather than the effects that cutting chamber and apex profile played on IT and stability as a result of the screw positioning.
Markovic et al. reported in vivo an increased implant stability and RFA values in favor of self-tapping implants than non-self-tapping and round apex implants [37].
The treatment of low-density bone represents a procedure that requires high clinical sensibility and, possibly, as reported in the literature, also implant site under-preparation and condensation of the peripheral bone walls [7,38,39].
In the present study, the Group A and B implants showed a high level of IT and RFA on polyurethane blocks, including for 10 pcf blocks [40,41].
The use of conical shaped implants could represent an advantage in cases of reduced diameter alveolar ridges [42,43,44,45].
5. Conclusions
The present investigation evaluated implants with different geometries that are present on the market, and not experimental implants specifically created for the study. The authors aimed to simulate real clinical conditions (poor-density bone or immediate post-extraction implants) in which the knowledge of dental implant features, which may be useful for increasing the primary stability, may help the oral surgeon during the surgery planning.
Acknowledgments
The authors declare no acknowledgments for this study.
Author Contributions
Conceptualization: A.P. and G.I.; methodology: A.P., L.C., G.I, A.E.P.; software: M.T.; validation: A.P., G.I.; formal analysis: M.T., A.P.; investigation: A.P.; resources: A.P., G.I.; data curation: M.T., L.C., A.E.P.; writing—original draft preparation: A.P., M.T.; writing—review and editing: A.P., M.T.; funding acquisition: A.P. All authors have read and agreed to the published version of the manuscript.
Funding
This work has been supported by the Italian Ministry of Education, University and Research (M.I.U.R.), Rome, Italy, and by a grant from Implacil De Bortoli, Sao Paulo, Brasil.
Conflicts of Interest
The authors declare no conflict of interest.
References
- 1.Tey V.H.S., Phillips R., Tan K. Five-year retrospective study on success, survival and incidence of complications of single crowns supported by dental implants. Clin. Oral Implant. Res. 2017;28:620–625. doi: 10.1111/clr.12843. [DOI] [PubMed] [Google Scholar]
- 2.Degidi M., Daprile G., Piattelli A. Influence of Stepped Osteotomy on Primary Stability of Implants Inserted in Low-Density Bone Sites: An In Vitro Study. Int. J. Oral Maxillofac. Implant. 2017;32:37–41. doi: 10.11607/jomi.4852. [DOI] [PubMed] [Google Scholar]
- 3.Javed F., Ahmed H.B., Crespi R., Romanos G.E. Role of primary stability for successful osseointegration of dental implants: Factors of influence and evaluation. Interv. Med. Appl. Sci. 2013;5:162–167. doi: 10.1556/IMAS.5.2013.4.3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Misch C.E. Bone density: A key determinant for clinical success. Contemp. Implant Dent. 1999;8:109–118. [Google Scholar]
- 5.Javed F., Almas K., Crespi R., Romanos G.E. Implant surface morphology and primary stability: Is there a connection? Implant Dent. 2011;20:40–46. doi: 10.1097/ID.0b013e31820867da. [DOI] [PubMed] [Google Scholar]
- 6.Oue H., Doi K., Oki Y., Makihara Y., Kubo T., Perrotti V., Piattelli A., Akagawa Y., Tsuga K. Influence of implant surface topography on primary stability in a standardized osteoporosis rabbit model study. J. Funct. Biomater. 2015;6:143–152. doi: 10.3390/jfb6010143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Stacchi C., Vercellotti T., Torelli L., Furlan F., Di Lenarda R. Changes in implant stability using different site preparation techniques: Twist drills versus piezosurgery. A single-blinded, randomized, controlled clinical trial. Clin. Implant Dent. Relat. Res. 2013;15:188–197. doi: 10.1111/j.1708-8208.2011.00341.x. [DOI] [PubMed] [Google Scholar]
- 8.Zizzari V.L., Berardi D., Congedi F., Tumedei M., Cataldi A., Perfetti G. Morphological Aspect and iNOS and Bax Expression Modification in Bone Tissue Around Dental Implants Positioned Using Piezoelectric Bone Surgery Versus Conventional Drill Technique. J. Craniofac. Surg. 2015;26:741–744. doi: 10.1097/SCS.0000000000001540. [DOI] [PubMed] [Google Scholar]
- 9.González-García A., Diniz-Freitas M., Somoza-Martín M., García-García A. Piezoelectric and conventional osteotomy in alveolar distraction osteogenesis in a series of 17 patients. Int. J. Oral Maxillofac. Implant. 2008;23:891–896. [PubMed] [Google Scholar]
- 10.Dahiya K., Kumar N., Bajaj P., Sharma A., Sikka R., Dahiya S. Qualitative Assessment of Reliability of Cone-beam Computed Tomography in evaluating Bone Density at Posterior Mandibular Implant Site. J. Contemp. Dent. Pract. 2018;19:426–430. doi: 10.5005/jp-journals-10024-2278. [DOI] [PubMed] [Google Scholar]
- 11.Ko Y.-C., Huang H.-L., Shen Y.-W., Cai J.-Y., Fuh L.-J., Hsu J.-T. Variations in crestal cortical bone thickness at dental implant sites in different regions of the jawbone. Clin. Implant Dent. Relat. Res. 2017;19:440–446. doi: 10.1111/cid.12468. [DOI] [PubMed] [Google Scholar]
- 12.Misch C.E., Judy K.W. Classification of partially edentulous arches for implant dentistry. Int. J. Oral Implantol. 1987;4:7–13. [PubMed] [Google Scholar]
- 13.Comuzzi L., Iezzi G., Piattelli A., Tumedei M. An In Vitro Evaluation, on Polyurethane Foam Sheets, of the Insertion Torque (IT) Values, Pull-Out Torque Values, and Resonance Frequency Analysis (RFA) of NanoShort Dental Implants. Polymers. 2019;11:1020. doi: 10.3390/polym11061020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Comuzzi L., Tumedei M., Piattelli A., Iezzi G. Short vs. Standard Length Cone Morse Connection Implants: An In Vitro Pilot Study in Low Density Polyurethane Foam. Symmetry. 2019;11:1349. doi: 10.3390/sym11111349. [DOI] [Google Scholar]
- 15.Sennerby L., Meredith N. Implant stability measurements using resonance frequency analysis: Biological and biomechanical aspects and clinical implications. Periodontology 2000. 2008;47:51–66. doi: 10.1111/j.1600-0757.2008.00267.x. [DOI] [PubMed] [Google Scholar]
- 16.Degidi M., Daprile G., Piattelli A., Carinci F. Evaluation of factors influencing resonance frequency analysis values, at insertion surgery, of implants placed in sinus-augmented and nongrafted sites. Clin. Implant Dent. Relat. Res. 2007;9:144–149. doi: 10.1111/j.1708-8208.2007.00042.x. [DOI] [PubMed] [Google Scholar]
- 17.Lages F.S., Douglas-de Oliveira D.W., Costa F.O. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: A systematic review. Clin. Implant Dent. Relat. Res. 2018;20:26–33. doi: 10.1111/cid.12565. [DOI] [PubMed] [Google Scholar]
- 18.Möhlhenrich S.C., Kniha K., Heussen N., Hölzle F., Modabber A. Effects on primary stability of three different techniques for implant site preparation in synthetic bone models of different densities. Br. J. Oral Maxillofac. Surg. 2016;54:980–986. doi: 10.1016/j.bjoms.2016.07.004. [DOI] [PubMed] [Google Scholar]
- 19.Szalma J., Lovász B.V., Vajta L., Soós B., Lempel E., Möhlhenrich S.C. The influence of the chosen in vitro bone simulation model on intraosseous temperatures and drilling times. Sci. Rep. 2019;9:11817. doi: 10.1038/s41598-019-48416-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Iezzi G., Scarano A., Di Stefano D.A., Arosio P., Doi K., Ricci L., Piattelli A., Perrotti V. Correlation between the bone density recorded by a computerized implant motor and by a histomorphometric analysis: A preliminary in vitro study on bovine ribs. Clin. Implant Dent. Relat. Res. 2015;17(Suppl. 1):e35–e44. doi: 10.1111/cid.12121. [DOI] [PubMed] [Google Scholar]
- 21.Di Stefano D.A., Arosio P., Perrotti V., Iezzi G., Scarano A., Piattelli A. Correlation between Implant Geometry, Bone Density, and the Insertion Torque/Depth Integral: A Study on Bovine Ribs. Dent. J. 2019;7:25. doi: 10.3390/dj7010025. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Duyck J., Roesems R., Cardoso M.V., Ogawa T., De Villa Camargos G., Vandamme K. Effect of insertion torque on titanium implant osseointegration: An animal experimental study. Clin. Oral Implant. Res. 2015;26:191–196. doi: 10.1111/clr.12316. [DOI] [PubMed] [Google Scholar]
- 23.Petrey J.S., Saunders M.M., Kluemper G.T., Cunningham L.L., Beeman C.S. Temporary anchorage device insertion variables: Effects on retention. Angle Orthod. 2010;80:446–453. doi: 10.2319/070309-376.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Gracco A., Giagnorio C., Incerti Parenti S., Alessandri Bonetti G., Siciliani G. Effects of thread shape on the pullout strength of miniscrews. Am. J. Orthod. Dentofac. Orthop. 2012;142:186–190. doi: 10.1016/j.ajodo.2012.03.023. [DOI] [PubMed] [Google Scholar]
- 25.Shah A.H., Behrents R.G., Kim K.B., Kyung H.-M., Buschang P.H. Effects of screw and host factors on insertion torque and pullout strength. Angle Orthod. 2012;82:603–610. doi: 10.2319/070111-427.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Allen R.F., Baldini N.C., Donofrio P.E., Gutman E.L., Keefe E., Kramer J.G. Annual Book of ASTM Standards, Medical Devices and Services. ASTM International; West Conshohocken, PA, USA: 1998. Standard specification for rigid polyurethane foam for use as a standard material for testing orthopedic devices and instruments (F1839-97) [Google Scholar]
- 27.Scarano A., Degidi M., Iezzi G., Petrone G., Piattelli A. Correlation between implant stability quotient and bone-implant contact: A retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans. Clin. Implant Dent. Relat. Res. 2006;8:218–222. doi: 10.1111/j.1708-8208.2006.00022.x. [DOI] [PubMed] [Google Scholar]
- 28.Turkyilmaz I., Sennerby L., Yilmaz B., Bilecenoglu B., Ozbek E.N. Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: A human cadaver study. Clin. Implant Dent. Relat. Res. 2009;11:52–58. doi: 10.1111/j.1708-8208.2008.00095.x. [DOI] [PubMed] [Google Scholar]
- 29.Gehrke S.A., Eliers Treichel T.L., Pérez-Díaz L., Calvo-Guirado J.L., Aramburú Júnior J., Mazón P., de Aza P.N. Impact of Different Titanium Implant Thread Designs on Bone Healing: A Biomechanical and Histometric Study with an Animal Model. J. Clin. Med. 2019;8:777. doi: 10.3390/jcm8060777. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Romanos G.E., Delgado-Ruiz R.A., Sacks D., Calvo-Guirado J.L. Influence of the implant diameter and bone quality on the primary stability of porous tantalum trabecular metal dental implants: An In Vitro biomechanical study. Clin. Oral Implant. Res. 2018;29:649–655. doi: 10.1111/clr.12792. [DOI] [PubMed] [Google Scholar]
- 31.Di Stefano D.A., Arosio P., Gastaldi G., Gherlone E. The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks. J. Prosthet. Dent. 2018;120:706–714. doi: 10.1016/j.prosdent.2017.04.012. [DOI] [PubMed] [Google Scholar]
- 32.Yamaguchi Y., Shiota M., FuJii M., Sekiya M., Ozeki M. Development and application of a direct method to observe the implant/bone interface using simulated bone. Springerplus. 2016;5:494. doi: 10.1186/s40064-016-2116-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Aleo E., Varvara G., Scarano A., Sinjari B., Murmura G. Comparison of the primary stabilities of conical and cylindrical endosseous dental implants: An in-vitro study. J. Biol. Regul. Homeost. Agents. 2012;26:89–96. [PubMed] [Google Scholar]
- 34.Gehrke S.A., Pérez-Díaz L., Mazón P., De Aza P.N. Biomechanical Effects of a New Macrogeometry Design of Dental Implants: An In Vitro Experimental Analysis. J. Funct. Biomater. 2019;10:47. doi: 10.3390/jfb10040047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Wang T.-M., Lee M.-S., Wang J.-S., Lin L.-D. The effect of implant design and bone quality on insertion torque, resonance frequency analysis, and insertion energy during implant placement in low or low- to medium-density bone. Int. J. Prosthodont. 2015;28:40–47. doi: 10.11607/ijp.4063. [DOI] [PubMed] [Google Scholar]
- 36.Nokar S., Rasouli-Ghahroudi A.R., Shidvash E., Atri F. Comparative investigation of primary stability of four different dental implants in low-density bone model. Dent. Res. J. (Isfahan) 2019;16:18–23. [PMC free article] [PubMed] [Google Scholar]
- 37.Marković A., Calvo-Guirado J.L., Lazić Z., Gómez-Moreno G., Ćalasan D., Guardia J., Čolic S., Aguilar-Salvatierra A., Gačić B., Delgado-Ruiz R., et al. Evaluation of primary stability of self-tapping and non-self-tapping dental implants. A 12-week clinical study. Clin. Implant Dent. Relat. Res. 2013;15:341–349. doi: 10.1111/j.1708-8208.2011.00415.x. [DOI] [PubMed] [Google Scholar]
- 38.Trisi P., Berardini M., Falco A., Podaliri Vulpiani M. New Osseodensification Implant Site Preparation Method to Increase Bone Density in Low-Density Bone: In Vivo Evaluation in Sheep. Implant Dent. 2016;25:24–31. doi: 10.1097/ID.0000000000000358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Falisi G., Severino M., Rastelli C., Bernardi S., Caruso S., Galli M., Lamazza L., Di Paolo C. The effects of surgical preparation techniques and implant macro-geometry on primary stability: An in vitro study. Med. Oral Patol. Oral Cir. Bucal. 2017;22:e201. doi: 10.4317/medoral.21286. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Araújo M.G., Silva C.O., Misawa M., Sukekava F. Alveolar socket healing: What can we learn? Periodontolog 2000. 2015;68:122–134. doi: 10.1111/prd.12082. [DOI] [PubMed] [Google Scholar]
- 41.Van der Weijden F., Dell’Acqua F., Slot D.E. Alveolar bone dimensional changes of post-extraction sockets in humans: A systematic review. J. Clin. Periodontol. 2009;36:1048–1058. doi: 10.1111/j.1600-051X.2009.01482.x. [DOI] [PubMed] [Google Scholar]
- 42.Tumedei M., Savadori P., Del Fabbro M. Synthetic Blocks for Bone Regeneration: A Systematic Review and Meta-Analysis. Int. J. Mol. Sci. 2019;20:4221. doi: 10.3390/ijms20174221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Gehrke A., Mazon P., Del Fabbro M., Tumedei M., Aramburù J., Perez-Diaz L., De Aza P. Histological and Histomorphometric Analyses of Two Bovine Bone Blocks Implanted in Rabbit Calvaria. Symmetry. 2019;11:641–654. doi: 10.3390/sym11050641. [DOI] [Google Scholar]
- 44.Scarano A., Piattelli A., Murmura G., Iezzi G., Assenza B., Mancino C. Delayed expansion of the atrophic mandible by ultrasonic surgery: A clinical and histologic case series. Int. J. Oral Maxillofac. Implant. 2015;30:144–149. doi: 10.11607/jomi.2753. [DOI] [PubMed] [Google Scholar]
- 45.Malchiodi L., Quaranta A., D’Addona A., Scarano A., Quaranta M. Jaw reconstruction with grafted autologous bone: Early insertion of osseointegrated implants and early prosthetic loading. J. Oral Maxillofac. Surg. 2006;64:1190–1198. doi: 10.1016/j.joms.2006.04.015. [DOI] [PubMed] [Google Scholar]