Event-Driven ECG Sensor in Healthcare Devices for Data Transfer Optimization

Manel Ben-Romdhane 1,, Asma Maalej 1, Mariam Tlili 2, Chiheb Rebai 1, François Rivet 3, Dominique Dallet 3
PMCID: PMC7223297  PMID: 32421087

Abstract

The long-term monitoring of cardiovascular signs requires a wearable and connected electrocardiogram (ECG) healthcare device. It increases user’s comfort and diagnosis quality of chronic cardiac and/or high-risk patients. This paper covers the enormous data to be transmitted from the ECG device to the physician’s, namely the cardiologist’s, control unit. Existent ECG devices uniformly sample analog signals and convert them to digital samples which are compressed before data transmission. However, event-driven sampling simultaneously compresses and samples. Therefore, this paper quantitatively compares successive approximation register analog-to-digital converter (SAR ADC) with discrete wavelet transform (DWT) compression and level-crossing analog-to-digital converter (LC-ADC). Evaluation metrics are the percent root-mean-square difference (PRD), bit compression ratio (BCR) and data length in bits. When a 12-bit reconstruction is operated on the outputs of an 8-bit LC-ADC with 12-bit and 10-kHz reference counter, the BCR is equal to 80% for 75% of test ECG signals. That is better than the 71.87% BCR of the 12-bit 1-kHz SAR ADC with DWT compression. The modeled LC-ADC guarantees a signal quality in terms of PRD comparable to the PRD of the SAR ADC with DWT compression. The data length in bits of the LC-ADC is lower than the data length in bits of the SAR ADC with more than 14-bit resolution with DWT compression for 82% of the test ECG signals. However, for lower resolutions, to obtain lower power consumption for radiofrequency transmission, a better alternative remains the SAR ADC with DWT compression.

Keywords: Analog-to-digital converter, Event-driven sampling, Successive approximation of registers, ECG sensor, Discrete wavelet transform, Data compression

Introduction

In health care, there are considerable developments in medical devices in order to provide continuously or on-demand monitoring about the well-being of the chronic cardiac and/or high-risk patients [1]. There are also reported works on electrocardiogram (ECG) biometric mode for the identification and authentication of individuals but also aiming at securing data and accessing services, as medical ones [2, 3]. Additionally, American Heart Association (AHA) statistics show that the coronary heart disease is the leading cause of death in the USA. In fact, 43.8% of deaths are caused by cardiovascular diseases (CVDs) [4]. Therefore, Internet of things (IoT) healthcare devices, namely electrocardiogram (ECG) sensors, become increasingly vital to the twenty-first century patients. Increased need for ECG devices is driven due to rising healthcare costs and increasing rates of CVD worldwide. These factors drive the demand for low-cost, reliable and wearable ECG devices. These devices allow ECG signal acquisition [5], digital signal processing [6] and data transmission to the remote physician, namely the cardiologist [1], for abnormal ECG signal analysis [1]. This leads the authors to focus on ECG signal acquisition but not only ECG features extraction. Commercialized wearable ECG devices as QARDIOCORE by Qardio [7] and wired ECG devices derive from 1 to 12 signals, known as leads in medical context. Each is composed of the P wave, the QRS complex and then the T wave.

Currently, the most recent wearable ECG devices are KardiaMobile by AliveCor [8], QARDIOCORE by Qardio [9] and Apple Watch Series4 by Apple [8]. KardiaMobile records a single lead from 30 s to 5 min to detect atrial fibrillation, bradycardia, tachycardia as well as normal heart rhythm. Apple Watch Series4 also records a single-lead ECG for 30 s and detects only atrial fibrillation or normal heart rhythm. QARDIOCORE, meanwhile, detects heart rate variability without diagnosis prediction. It records 3 ECG signals or leads with a 16-bit analog-to-digital converter (ADC) at 600-Hz sampling frequency. It sends a message via Bluetooth wireless communication to the connected smartphone. Then, the smartphone transfers the message to the physician within few seconds. KardiaMobile’s ADC samples at 300 Hz, then quantizes and codes each sample on 16 bits. No additional information is available regarding Apple Watch Series4’s sampling frequency and ADC resolution.

As an example, the KardiaMobile’s analog-to-digital converter (ADC) delivers a 4.8-kbits message during one second of conversion for a single signal. In the scientific literature, to optimize or reduce data transfer, decrease sensors’ power consumption and increase wireless sensor network lifetime, three research axes have been developed. First, communication protocols consider the optimization of data communication between the healthcare sensors or devices, the patient’s smartphone and the hospital server [10]. Second, radiofrequency (RF) transceiver is miniaturized to realize a comfortable wearable smart sensor as in [11]. Third, message data length of the ECG is reduced in order to optimize the data transfer via a given wireless communication transceiver [12]. In this context, this paper argues that the event-driven ECG sensor should be used to reduce message data length and to optimize data transfer.

Thus, it is interesting to reduce the message data length in bits, which is obtained at the ADC output, to optimize data transfer power consumption [13]. An Lb-bit message is transmitted during a given time duration from the ADC output to the receiver. The transmitter’s data transfer power consumption is consequently Lb multiplied by the wireless transmission power consumption per bit [10, 14]. Besides, the message data length in bits during one second is the message to be transferred during one second and not the wireless communication standard’s throughput.

As far as ECG signal acquisition is concerned, the wearable ECG devices and many wired or wireless ECG acquisition systems use generally Nyquist ADCs and rarely oversampling ADCs. Data are uniformly extracted at every sampling period, considering at least Nyquist frequency according to Shannon theorem [5]. In fact, sigma-delta ADCs offer higher resolutions than Nyquist ADCs, but generally require higher power consumption due to their oversampling technique. Besides, the maximum frequency of ECG signal is 250 Hz [15, 16]. Consequently, the sampling frequency must be higher than 500 Hz according to Shannon theorem. ADC survey shows the successive approximation register ADCs (SAR ADCs) are more suitable for ECG signal acquisition than pipeline, continuous-time sigma-delta (CTSD), switched-capacitor sigma-delta (SCSD) and flash ADCs according to the signal-to-noise and distortion ratio (SNDR) and for a given technology as the 180 nm complementary metal oxide semiconductor (CMOS) [17]. Moreover, SAR ADC architectures use few building blocks and thus consume less power than other ADC architectures [18]. Their propagation delays are lower than those of other architectures and are less sensitive to technology accuracy limits of passive components and jitter.

However, many ADCs need data compression to send or store lower amount of data that represent original samples. The aim is to restore the ECG signal after decompression at the distant ECG control unit [5]. The obtained ECG signal is quasi-identical to the original ECG signal. It is then analyzed by the physician who sends the diagnosis to the patient. Knowing that, off-line QRS complex detection [19], real-time QRS complex detection or even ECG features extraction [1] methods are irreversible compression methods. Indeed, they do not allow to recover the ECG signal. Conversely, in the literature, some proposals are made with event-driven ECG sensors that sample signals regarding levels as in [20]. These sensors require level-crossing analog-to-digital converter (LC-ADC) that combines event-driven sampling and compression simultaneously.

This paper aims at selecting the best-fit discrete wavelet transform (DWT) to compress a successive approximation register analog-to-digital converter (SAR ADC) output in an ECG sensor and to fulfill a quantitative comparative study with the level-crossing ADC (LC-ADC). To the best of the authors’ knowledge, no quantitative comparative study between (A1) the SAR ADC architecture which is followed by a DWT compression stage and (A2) the LC-ADC architecture has been conducted. Thus far, this paper’s main contribution is to achieve a quantitative comparison by:

  • Reviewing on the one hand, the literature of the ADCs that relies on the SAR ADC as the best candidate among Nyquist besides sigma-delta ADCs, and the LC-ADC as the event-driven sampling-based ADC, and on the other, the literature of the data compression methods which decides on DWT compression method,

  • Simulating examples of the architectures (A1) and (A2) where test ECG signals from databases [21] are applied at their inputs,

  • Comparing the percent root-mean-square differences (PRDs), the bit compression ratios (BCRs) and the message data lengths in bits of (A1) and (A2),

  • Selecting the best architecture to reduce the message data length in bits and save power in wireless sensor networks.

Consequently, the remaining of this paper is organized as follows:

  • Section 2 gives a state of the art of ECG signal acquisition and data compression.

  • Section 3 details the performance evaluation test-bench steps of (A1) the SAR ADC with discrete wavelet transform (DWT) compression and states how to select the most suitable DWT to compress the SAR ADC output. It also briefs (A2) the LC-ADC, its performance evaluation test bench, the data transfer optimization and evaluation metrics that are required to compare (A1) and (A2).

  • Section 4 presents and discusses the quantitative comparison of the evaluation metrics as the percent root-mean-square differences (PRD), the bit compression ratios (BCR) and the data length in bits. The objective is to provide a defendable selection of (A1) the SAR ADC with DWT compression or (A2) the LC-ADC.

  • Section 5 summarizes the whole paper and draws future works.

ECG Signal Acquisition and Data Compression

In all manufactured electrocardiogram (ECG) acquisition devices and most ECG acquisition devices in the literature, the analog-to-digital converter (ADC) uniformly samples and quantizes ECG analog signals to obtain ECG data, ECGout. In an ECG signal, the waveform from the T wave to the P wave has usually a lower amplitude and a larger duration than those of the P wave, the QRS complex and the T wave [22]. Thus, data compression of the uniform ECG data is required to reduce the message data length to be transferred from the ECG device to a distant control unit for an ECG diagnosis. However, this paper also deals with the sampling regarding analog signal amplitude events. This signal acquisition is indeed more suitable for the ECG signal. In this section, the ECG monitoring system architecture is introduced to show its key blocks. Then, the authors present a review of ECG signal acquisition, namely the successive approximation register ADCs (SAR ADCs) and level-crossing ADCs (LC-ADCs). Finally, existent data compression methods that follow a SAR ADC are provided. They are compared in terms of bit compression ratio (BCR) in order to evaluate ECG data reduction and percent root-mean-square difference of data compression (PRDcomp) that evaluates the recovered-signal quality.

ECG Monitoring System Architecture

The configuration of an ECG monitoring system, as presented in Fig. 1, is composed of the wearable ECG sensor, the signal acquisition, the data compression and the distant ECG control unit. The obtained ECG signal from two electrodes is filtered, amplified then sampled and quantized in the analog front end to deliver the digitized signal. In commercialized systems [1], ECG signal acquisition is only operated using Nyquist or sigma-delta ADCs. However, in the ECG devices that have been suggested in the literature [20], level-crossing ADCs have been also employed. The compression is required for Nyquist or sigma-delta ADCs but never quantitatively compared to the inherent compression of level-crossing ADCs. In fact, the LC-ADCs decrease the amount of data compared to Nyquist ADCs and considerably decrease the amount of data compared to sigma-delta ADCs. After compression or source encoding, followed by radiofrequency (RF) transmission, the digital reconstruction block performs signal decompression or source decoding. In case of event-driven devices, interpolation is also required to recover uniform samples. The control unit manages the signal decompression, the digital signal processing and the interpolation. The physician also analyzes the ECG signals using one of his communication devices and saves only received data if required.

Fig. 1.

Fig. 1

Configuration of a wearable ECG monitoring system

ECG Analog-to-Digital Converters’ State of the Art

Most of ECG acquisition systems generally digitize analog signals with successive approximation register ADC (SAR ADC) as stated in the commercialized systems or in the scientific literature proposals [1]. Indeed, the SAR ADC structure is simply composed of a sample-and-hold (S/H) circuit, a comparator to compare the sample to the previous estimation, successive approximation of registers (SARs) and a digital-to-analog converter (DAC) [18]. The DAC delivers the analog value of the previous estimation. The SAR ADC converts an analog ECG signal into digital samples. Thus, the T-to-P ECG waveform, which amplitude variations are lower than 0.1 mV during at least 0.3 s, and the QRS complex, which amplitude is around 1 mV during at most 0.1 s [22], are sampled at the same frequency [23, 24]. That delivers an important number of redundant samples in the T-to-P ECG waveform. However, sampling only when events occur makes it possible to reduce the T-to-P ECG data. For instance, in the pacemaker proposal in [25], flexible resolution SAR ADC with asynchronous dynamic logic samples the ECG leads and their features after the QRS complex extraction in the analog domain.

Moreover, in [14], to reduce the data length when sampling an ECG signal with a SAR ADC, the QRS complex is sampled at a frequency higher than the sampling frequency of the remaining parts of the ECG signal. Further compression reduces the number of bits per sample. In fact, after the SAR ADC, a digital adaptive controller is required to compare the difference between successive samples and the predefined threshold to select adequate sampling frequency. The SAR ADC parameters given in [14] are measured only for a test ECG signal. Another SAR ADC-based proposal is to sample the ECG signal at two different sampling frequencies which are selected according to its analog slope extraction [26]. Additionally, authors in [27] propose a high-precision ECG sensor system due to an ADC that integrates a stochastic flash ADC with a digitally controlled variable threshold comparator into a SAR ADC. In addition, a dynamic tracking SAR ADC and a hybrid SAR-sigma-delta ADC are proposed in [28, 29], respectively. Otherwise, sigma-delta converters are rarely used as in [29, 30]. Indeed, they offer higher resolutions, but require important sampling frequencies due to their oversampling technique [17].

Therefore, some of the most recent SAR ADCs are summarized in Table 1 in reference to different resolutions, m, that are sorted from low to high values. In fact, there are proposals to reduce the data length of the SAR ADC output [14, 2426, 28] and others to increase SAR ADC resolutions [27, 29]. For the data length reduction, the SAR ADC in [28] uniformly samples the ECG signals, and then, a digital signal processing is operated. The SAR ADC in [14] samples the signals after a frequency digital selection, whereas the SAR ADC in [26] samples the signals after a frequency analog selection. Other parameters are used to compare SAR ADCs. They are the frequency of the successive approximation of registers (SARs), FSAR, the sample-and-hold frequency or the reference frequency, Fref, the dynamic range, the complementary metal oxide semiconductor (CMOS) technology, the power consumption and the core area, and the effective number of bits (ENOB) is presented in Table 1. The ENOB is given in the reference or is extracted from the signal-to-noise and distortion ratio (SNDR) in decibels (dB) as in (1) [27].

ENOB=SNDR-1.766.02 1

Table 1.

SAR ADC state of the art

References [25] [24] [14] [28] [26] [29] [27]
Resolution, m, from low to high (bits) 7–10 8 10 12 14 16 24
Proposal to reduce SAR ADC output data length Asynchronous dynamic logic and flexible resolution SAR ADC SAR ADC with two sampling frequencies and data compression Dynamic tracking SAR ADC SAR ADC with two sampling frequencies
Proposal to increase SAR ADC resolution Hybrid SAR-sigma delta ADC Successive stochastic approximation ADC
Reference frequency Fref (kHz) 1.024 1–20 0.125 and 1 10 1 20–1600 62.5, 125, 250
SARs’ frequency FSAR (kHz) 32 50,000 200 1040–83,200
Dynamic range (V) 0.11 1 1.8 0.6 1 1.2
CMOS technology

180 nm

1.3–1.8 V

180 nm

1 V

180 nm

1.8 V

130 nm

0.6 V

130 nm

1.5 V

180 nm

1 V

130 nm

1.2 V

Power consumption (µW) 0.884–1.22

1.12 for

1-kHz Fref

78.8 for test ECG signala 0.96 1.3 4.73–482 3.91
Area (mm2) 8.6 0.0325 0.7225 0.126 4.488 0.15 0.342
ENOB (bits) 9 6.93 6.63 for test ECG signala 11.77 12.16 15.15 14.4

aNot tested for sinewave signal

From the SAR ADC state of the art, the outputs require data length reduction by either modifying the SAR ADC architecture or applying a compression technique at the SAR ADC output.

Therefore, the event-driven signal acquisition is a more appealing architecture than the SAR ADC, also known as level-crossing sampling. It has been introduced since the 1980s as a combined compression digital encoding technique of analog signals [31], but its circuit implementation or modeling is recent. In fact, level-crossing analog-to-digital converter (LC-ADC) architectures are proposed to sample signals regarding amplitude levels instead of time instants. That detects the exact amplitude values instead of the quantized data values. The LC-ADC thus guarantees the conversion of the samples at selected levels. However, it requires a quantization of each interval between two successive level-crossing instants. In the last decade, for instance, authors in [32] and [33] studied event-driven speech and biomedical signals, respectively. Different circuit realizations were particularly proposed for bio-signal acquisitions, ranging from LC-ADCs in application-specific integrated circuits (ASICs) [34, 35] to fully integrated system-on-chip (SoC) solutions for wearable sensors [20].

As far as LC-ADC architecture types are concerned, some LC-ADC architectures use the fixed-window level-crossing type [3537], whereas others use the floating-window type [20, 38, 39]. In fact, the floating-window type is the conventional and the widely adopted type in LC-ADC architectures. It consists of two comparators, two digital-to-analog converters (DACs), a logic control and a reference counter clock. First, the comparators detect the input crosses of the analog levels. Second, the DACs create analog levels bounding the input. Third, the logic control obtains up/down slope and evaluates the samples. Fourth, the reference counter clock quantizes the interval between two successive level-crossing instants. Alternatively, the fixed-window type does not require the updating of the window between two successive analog levels, but it requires the subtraction of the DAC output from the analog input. The obtained difference is compared to the analog fixed levels. Instead of DACs in the LC-ADC, some proposals are based on a scaler to select analog levels for the comparators [40]. Furthermore, authors in [35, 38] suggest other solutions with variable resolutions regarding slope threshold which is the window out of the quantized time between two successive samples.

The LC-ADC state of the art is presented in Table 2. The LC-ADC references are organized according to the LC-ADC resolutions, M, that are sorted from 5 to 8 bits. These values are adequate to digitize the test ECG signals according to the literature. In addition, the reference counter frequency, FC, quantizes the time between two successive level-crossing instants. In Table 2, the power consumption and core area values are sometimes proportionally computed from other values in the given reference or in a previous reference of the same authors. Table 2 shows that LC-ADC power consumptions vary from 0.0024 to 106 µW for test sinewave input signals at different frequencies from 1 Hz to 10 kHz. These values are close to SAR ADCs’ values from 0.884 to 78.8 µW when 1 kHz reference frequency is adopted. However, it is important to mention that for low-frequency and low-amplitude inputs, as in T-to-P ECG waveform, the LC-ADC samples less densely in time than for high-frequency and high-amplitude inputs, as in QRS complex. As far as the areas are concerned, they are generally similar because most of SAR ADC and LC-ADC architectures are based on comparators, DACs and digital processing.

Table 2.

LC-ADC state of the art

References [39] [37] [36] [33] [20] [35] [40]
Resolution, M, from low to high (bits) 5 5 5 6 8 8
Type Floating window Fixed window Fixed window Fixed window Floating window Fixed window and variable resolution Fixed window
Reference counter frequency FC (MHz) 0.001 50 0.01–5 10
CMOS technology 130 nm, 0.3 V 130 nm, 1 V 500 nm, 3.3 V 180 nm, 0.8 V 180 nm, 1.25 V 130 nm, 0.8 V 180 nm, 0.55–1 V
Power consumption (µW) @ sinewave input frequency 0.22 @ 50 Hz 0.31 @ 2.6 kHz 106 @ 1 kHz 0.313–0.582@ 110 Hz-1.65 kHz 2.42 @ 1 kHz 7.4 @ 10 kHz

0.0024–0.541

@1 Hz-1 kHz

Area (mm2) 0.36 0.02 0.06 0.045 0.6 1.69 0.0144
ENOB (bits) 4.4 5.39 5.9 6.35–7.84 7.43 7.51–8.67 6.45–7.96

From the current review of the literature, tested circuits and measurement results with a test ECG signal are rarely published. The authors give in SAR ADC state of the art in Table 1, the measured results of [14] for test ECG signal, and in LC-ADC state of the art in Table 2, measured results of [20] are given for 1 kHz sinewave signal. ECG sensor in [20] is tested with an ECG signal, but performances are not measured. In fact, the ENOB of [14] is 6.63 bits for a 10-bit SAR ADC, 1 kHz reference frequency and test ECG signal, whereas it is 7.43 bits [20] for an 8-bit LC-ADC with a sinewave test signal at the input. That ENOB is estimated from SNDR which is computed according to the ECG wireless sensor’s input referred noise and input voltage. Therefore, LC-ADC and SAR ADC have almost the same performances leading to similar sensitivity levels [14, 20, 35]. Thus, aside from the combined signal acquisition and inherent compression of analog signals with LC-ADC, many compression techniques have been proposed in the literature. These techniques were implemented and tested on digital test ECG data at the SAR ADC outputs. In the next subsection, the existent compression methods are introduced and compared.

ECG Data Compression Methods’ State of the Art

Long-term monitoring using ECG electrodes connected to smartphones and/or point-of-care devices has to save and/or send selected ECG data for an early detection of cardiovascular diseases (CVDs). Since 1968, many works have been published covering ECG data compression, redundancy reduction and parameters’ extraction. For example, in ECG acquisition systems, data reduction for the sake of storage is introduced in [41] with amplitude zone time epoch coding (AZTEC). In fact, large amount of ECG data compression is applied either to store or to transmit less data in order to reduce bit rate and time processing while decreasing power consumption [42]. In the present paper, the uniform ECG data, ECGout, are compressed to obtain the data compression output, ECGcomp. To evaluate data compression methods, it is required to measure the data lengths in bits during the same time duration, DT, before and after the data compression. Thus, the bit compression ratio, BCR, is computed as given in (2),

BCR%=100×mL-LbcompmL 2

where m is the SAR ADC resolution, L is the number of samples, ECGout, and Lbcomp is the data length in bits of ECGcomp. The lower the data length in bits after compression is, the higher the BCR of the data compression method is.

Decompression quality can be clinically acceptable although many techniques highly compress ECG data. In this paper, the presented methods are applied after the SAR ADC and require ECG data decompression to recover estimated ECG data. From the recovered ECG signal, the percent root-mean-square difference of data compression, PRDcomp, is assessed, and the physician accomplishes the diagnosis as explained in Fig. 1. Indeed, to evaluate ECG data compression quality, the PRDcomp is computed as in (3) [43],

PRDcomp%=100×i=1LECGouti-ECG^outi2i=1LECGouti-ECG¯out2 3

where ECGout is the original uniform ECG data, ECG^out is the estimated data after decompression from Lbcomp bits, ECG¯out is the mean of the original uniform ECG data and L is the original uniform ECG data length or the number of samples. In order to achieve a correct clinical diagnosis only from the Lbcomp bits [44], at most a 9% PRDcomp is allowed. A 0% PRDcomp reflects a lossless compression and a reversible decompression which means that ECG^out is exactly equal to the original uniform ECG data, ECGout.

Thus, the commonly used compression algorithms are compared in order to select the adequate compression method regarding BCR and PRDcomp aiming at the evaluation of ECG data compression quality. Once the PRDcomp is lower than 9%, the lower the ECG data length in bits after compression, Lbcomp, is, the higher the BCR is and the better the data compression method is. The data compression methods are commonly distinguished in lossless and lossy methods. In fact, with lossless compression methods, reversible decompression algorithms recover the exact original uniform ECG data with a 0% PRDcomp. Opposingly with lossy compression methods, irreversible decompression algorithms remove some samples or bits from original uniform ECG data, and others project them given a transform function and remove some of the obtain coefficients. Consequently, ECG data recovery leads up to data distortion which generates a nonzero PRDcomp.

As far as the lossless data compression methods are concerned, entropy coding is widely used with differential pulse-code modulation (DPCM) regarding predefined error threshold with prediction or interpolation [45]. In [46], the prediction with DPCM is the least complex, and the prediction with more complex algorithms, such as neural network, auto-regression filters or Lempel–Ziv, is unsuitable for real-time transmission. Besides, entropy coding algorithms, such as Huffman, arithmetic, Golomb, Golomb–Rice (GR) and fixed-length packaging, are the commonly used ones. Adopting the lossless compression methods recovers the exact original ECG data, but low BCRs are computed. Combining predictor with entropy coding algorithms makes the bit compression ratios vary between almost 48% and 58% [4649] as shown in Table 3. Combining fuzzy optimization and Huffman region coding offers a BCR of 64.7% [50]. Nonetheless, adaptive trending predictor with modified Huffman and GR coding offers a BCR of 62.5% [51]. In these references, the obtained results are measured for a small number of test ECG signals or for test ECG signals from the same database, generally the MIT-BIH arrhythmia database [21].

Table 3.

Lossless data compression methods’ state of the art

Lossless data compression method BCR%
Linear predictor/Huffman coding [47] 47.9
Delta predictor/Golomb–Rice (GR) coding [46] 57.9
Adaptive predictor/Huffman coding [48] 58.8
Adaptive linear predictor/fixed-length packaging [49] 56.1
Fuzzy optimization/Huffman region coding [50] 64.7
Adaptive trending predictor/modified Huffman and GR coding [51] 62.5

However, nowadays, researchers in the field of smart health sensors adopt higher bit compression ratios with lossy data compression methods that offer a small ECG data distortion. In fact, works in [44, 5254] are based on thresholding-based wavelets or discrete wavelet transform (DWT) as lossy compression methods. The best trade-off combines DWT with shrinkage, Huffman and run-length encoding (RLE) algorithms to offer a 96% BCR and 3% PRDcomp [53]. Other lossy compression methods are proposed in [42, 55]. They offer around 77% and 87% of BCR with an excellent PRDcomp, that is, lower than 2%. In [55], the method consists in ECG data decimation by a fractional factor. In [42], the method uses Fun algorithm that is combined with an on-demand residual Huffman coding. Thus, Huffman compression is only applied on the residual data after online decompression and residual data request.

As far as [56] is concerned, the proposed compression method detects turning angles larger than a threshold. In fact, the useful information for the diagnosis is in the rapid variations of the P wave, the QRS complex and the T wave. Another proposal comes in [12] to reach a 99% BCR, but only an 8% mean PRDcomp is obtained for all test ECG signals of MIT-BIH arrhythmia database. In fact, convolutional auto-encoder is applied to ECG data compression by adopting a spindle structure to extend and then reduce the data length. A more recent work shows a double exponential wavelet design that is combined to compressive sensing [57]. This compression method tries to reach a 90% BCR and a 6.5% PRDcomp, but these are mean values obtained from 20 test ECG signals. Additionally, in [58], the compression algorithm begins by segmentation of the test ECG signal into clinically relevant parts (the QRS complexes and possibly P and T waves) and nonclinically relevant parts (the remaining parts of the ECG signal). The first group is compressed with differential pulse-code modulation (DPCM) and run-length encoding (RLE) that are lossless methods, whereas others are compressed due to DWT with threshold which is a lossy method.

Table 4 summarizes the lossy data compression state of the art. The obtained results are measured for one test ECG signal [56], two test ECG signals [58] or all test ECG signals [12] from MIT-BIH arrhythmia database. Thus, after the SAR ADC, the DWT-based compression methods provide the best trade-off between BCR and PRDcomp [53]. In addition, the DWT compression as implemented in [53] does not require test ECG signal segmentation as in [58], and decompression is always irreversible. However, the LC-ADC seems to be appealing as it simultaneously operates ECG signal acquisition and data compression. Besides, the reconstruction is always irreversible. In fact, although the LC-ADC allows lower amount of data regarding the SAR ADC with DWT compression, quantitative comparative study has not been conducted yet in terms of bit compression ratio, message data length in bits versus signal quality for both architectures. Thus far, the main contribution of this paper is a quantitative comparison between (A1) the SAR ADC architecture with DWT compression and (A2) the LC-ADC architecture. This comparison is important as the lower the message data length in bits is, the lower the ECG sensor power consumption of RF transmission is. In fact, the data length in bits is multiplied by the RF transmission power consumption per bit [14] to obtain the ECG sensor power consumption, knowing that RF transmission results in 63% of the power consumption of a given ECG sensor [13]. Thereafter, the next section describes the evaluation methods for both architectures and defines the evaluation metrics.

Table 4.

Lossy data compression methods’ state of the art

Lossy data compression method BCR% PRDcomp%
Simultaneous orthogonal matching pursuit [52] 86.1 2.57
Compressive sensing [44] 71 9
DWT with shrinkage [53] 68 0.12
DWT with shrinkage/Huffman/RLE [53] 65–96 0–3.11
Two sampling frequencies and Huffman encoding [14] 64 0.82
DWT/RLE with modified threshold [54] 81 2.7
Decimating by a fractional factor [55] 77.7 0.53
Fan algorithm/on-demand residual Huffman coding [42] 87.1 0.51
Turning-angle-based compression [56] 71.08a 5.87a
Spindle convolutional auto-encoder [12] 99.06 2.453–18.2
Compressive sensing and double exponential wavelet transform [57] 10–90b 0.43–6.5b
DPCM/RLE or DWT with threshold according to the part of ECG waves [58]

93a

93.19c

2.81a

2.78c

aRecord 117 from MIT-BIH arrhythmia database

bMean values for 20 ECG signals from MIT-BIH arrhythmia database

cRecord 110 from MIT-BIH arrhythmia database

Evaluation Methods

As provided previously in Table 1, recent SAR ADC architectures add specific compression as in [27, 29] or are combined with a further data compression block. Both must reduce the large amount of data delivered by the SAR ADC as in [25] to reduce the data amount to be sent to the control unit. This section deals with evaluation methods that are required for the quantitative comparison between (A1) the SAR ADC with discrete wavelet transform (DWT) compression and (A2) the LC-ADC. First, the authors acquaint the SAR ADC architecture. The performance evaluation test bench of SAR ADC that is followed by DWT compression is also proposed to test (A1) with many test ECG signals at its input. The DWT is inherently reversible transform. Opposingly, DWT compression is irreversible, and it is a lossy compression. Some of the obtained DWT output coefficients are removed before data transmission to the distant ECG control unit. Second, DWT-based ECG data compression is studied, and the most adequate DWT compression is proposed. Third, (A2) the designed LC-ADC architecture alongside its performance evaluation test bench is introduced. Moreover, optimization of the LC-ADC outputs is stated to reduce the data transfer in healthcare devices. Forth, test ECG signals from Physionet are presented together with the required evaluation metrics to analyze the quantitative comparison of (A1) the SAR ADC with DWT compression and (A2) the LC-ADC.

SAR ADC with DWT Compression Evaluation

The SAR ADC first samples and holds (S/H) the analog input signal, ECGin, at the S/H frequency or the reference frequency, Fref. Second, the obtained sample is compared to the first analog approximated value. This value is delivered after the successive approximation registers (SARs) and the digital-to-analog converter (DAC) [14]. Each cycle takes 1/FSAR where the SARs’ frequency, FSAR, is at least m times the reference frequency, Fref. The SARs’ initial value is set with the most significant bit (MSB) at “1” and other m-1 bits at “0” where m is the SAR ADC resolution. The DAC output is then equal to Vref/2 where Vref is the power supply voltage of the SAR ADC. This first analog approximated value is obtained in the first cycle. Third, if the first sample is higher than Vref/2, the MSB value is maintained at “1,” unless it becomes “0” in the second cycle. In this cycle, still for the same sample, the SARs move to the next bit and initialize it at “1.” The SARs operate similarly as for the MSB. Then, they move to the next bit until they reach the least significant bit (LSB).

After 1/Fref period, the following sample is compared to the first analog approximated value, Vref/2, approximated and converted as it was explained in the previous paragraph for the first sample. The SAR ADC architecture is presented in Fig. 2.

Fig. 2.

Fig. 2

SAR ADC architecture

The SAR ADC is simplified in two basic blocks of sampling and quantization. The S/H or reference frequency, Fref, is 1 kHz as it is the commonly used value to sample ECG signals, whereas the resolution, m, varies from 8 to 16 as presented in Table 1. To obtain test ECG signals that are representative of analog signals, each ECG data vector, ECGdata, from a given database, is up-sampled at 1 MHz to obtain a signal with a fine time axis. The obtained “analog” signal is sampled at Fref. The ith analog sample is quantized to the digital sample, ECGouti, using the data length, m, that is, the SAR ADC resolution. DWT compression is operated on L ECG digital samples and delivers L compressed data, ECGcomp, which length in bits is Lbcomp. In this paper, the performance evaluation test bench of (A1) the SAR ADC with DWT compression is illustrated in Fig. 3 in order to test SAR ADC with DWT compression.

Fig. 3.

Fig. 3

Performance evaluation test bench of (A1) the SAR ADC with DWT compression

To evaluate the quality of the DWT compression regarding the original uniform ECG data, ECGout, DWT irreversible decompression is applied to ECGcomp to estimate ECG^out that is compared to ECGout. However, before dealing with evaluation metrics, the authors describe the DWT compression methodology in the next subsection.

Discrete Wavelet Transform for Compression

In this subsection, first, a brief mathematical definition of the DWT and its relationship with compression is given as reported in the literature. Second, some wavelet matrices are provided as examples for a better understanding of this paper. Third, the methodology of the compression study is introduced.

The signal decomposition is generally performed for lossless transforms using orthogonal wavelets or biorthogonal wavelets that are invertible [59]. The transformed signal, s, has the same length as the original one and is a representation of the original signal in an orthogonal or invertible matrix. The original signal is perfectly recovered as it requires the transposition of an orthogonal decomposition matrix, or the inversion of an invertible decomposition matrix. These transformations are lossless, whereas they separate the signal from nonsignificant data.

Indeed, the DWT is a set of linear operations performed on an L-length vector of sampled signal, ECGout. It generates a vector of discrete data, s, that will be exploited instead of ECGout to perform many tasks in signal processing [60]. The matrix form of these linear projections is given in (4) where Φ is an L-by-L decomposition or analysis matrix with ϕi its ith row and αi the ith obtained coefficient. The coefficient, αi, results from the scalar product of ECGout and ϕi. In fact, each coefficient, αi, compares the signal ECGout to the row ϕi.

Φ·ECGout=s
withϕi,ECGout=αiands=αi,1iL 4

Thus, s is a different representation of ECGout in the analysis matrix described by the rows in Φ. The exact recovery of ECGout from these coefficients is done by multiplying both left sides of (4) with the inverse matrix of Φ that is noted Ψ. This gives (5) where Ψ is the wavelet or synthesis matrix. For orthogonal matrix, Ψ, the columns of Ψ, ψi, are the rows of Φ.

ECGout=Ψs 5

Noting that the DWT compression rounds the small coefficients to zeros, thus invertibility or reversibility is lost and the transformation becomes lossy. Therefore, the number, K, is the number of the most significant coefficients which are higher than a given threshold, Th, and is small compared to the length of s. The set of the coefficients αi>Th corresponds to the DWT compression output, ECGcomp, as presented in Fig. 3. Its data length is K. If αi is higher than Th, the wavelet ψi, which is the ith column in Ψ, is active in ECGout. Furthermore, if ECGout is compressible in Φ, the nonsignificant coefficients in s are rounded to zeros. From (5), the recovery of ECGout is approximated from its projections as in (6),

ECG^out=uαuψu
withui|αi>Th,1iLandK=cardi 6

where ECG^out is the decompressed data after the DWT compression of ECGout and K is the number of the active columns in ECGout regarding αi>Th.

For an orthogonal matrix, ϕi,1iL, the irreversible decompression is done with the K active columns of Ψ that are the rows of Φ. For an orthogonal matrix with no normalized rows to unit, the irreversible decompression is done with the K active columns of Ψ and the matrix ΦΨ is diagonal. For a biorthogonal matrix, the rows of Φ are biorthogonal to the columns of Ψ and the decompression is also irreversible. Thus, each scalar product ϕiψj is equal to 1 only if i is equal to j for 1iL and 1jL. Besides, both orthogonal and biorthogonal wavelet matrices can be implemented due to tree-structured filter banks. The filter coefficients of the analysis tree are deduced from the decomposition matrix factorization. Meanwhile, the filter coefficients of the synthesis tree are deduced from the wavelet matrix factorization [59]. The number of matrices in a factorization corresponds to J which is the wavelet decomposition level. The filter bank is tree-structured with J low-pass (LP) filters and J high-pass (HP) filters in both analysis and synthesis banks. LP filters in the analysis bank select signal information. These LP filters outputs are known as approximations. However, HP filters in the analysis bank select noise part of the signal. These HP filters outputs are known as details. The approximations and details are truncated to m bits that are the SAR ADC resolution. The selected values that are higher than the threshold, Th, constitutes the DWT compression output, ECGcomp. Its length in bits, Lbcomp, is equal to m times K that is the number of the selected coefficients, ECGcomp. Therefore, in this paper, the DWT compression as a lossy compression is selected for implementation with tree-structured filter banks since the authors study an embedded solution for IoT healthcare devices. The online DWT compression of clinically and nonclinically relevant ECG parts is preferred to off-line solution with lossless compression of clinically relevant ECG parts and lossy compression of nonclinically relevant ECG parts after ECG data segmentation.

The DWT decomposition is generally performed due to the orthogonal wavelets such as Coiflet (coif), Daubechies (db), Symlet (sym), Dmeyer (dmey) and Haar (haar) families, or the biorthogonal wavelets such as biorthogonal (bior) and reverse biorthogonal (rbio) families [61]. For each family, there are one or more subfamilies according to wavelet family parameters that impact numbers and values of the filter coefficients. After the SAR ADC, DWT-based ECG data compression often uses Daubechies or biorthogonal wavelets. These two wavelet examples of orthogonal and biorthogonal wavelet families are widely used and yield good compression ratios [53] even though, to quantitatively compare (A1) the SAR ADC with DWT compression to (A2) the LC-ADC, it is necessary to select the best-fit wavelet subfamily. The objective is to select the best DWT compression to be implemented after the SAR ADC considering the compression ratio, the ECG signal distortion and the implementation complexity. Thus, to select the best-fit wavelet subfamily regarding only the compression ratio, BCR, and the ECG signal distortion, PRDcomp, the orthogonal matching pursuit (OMP) algorithm is used to obtain an approximation of the signal in the L-by-L dictionary consisting of wavelet subfamily where L corresponds to the number of the SAR ADC samples. In fact, the OMP algorithm aims at computing all coefficients, αi,1iL, and selecting the K most significant ones, based on an iterative application of the DWT [62]. Thus, K is provided as the convergence criterion and is also defined as the minimum number of iterations to guarantee a given PRDcomp since the OMP selects the most active row of Φ at each iteration. The methodology of the DWT compression study is defined in order to evaluate DWT compression on a uniform ECG data, ECGout, and for a given wavelet subfamily, as shown in Fig. 4. The DWT compression study, organized in 5 steps, is proposed to select the best-fit wavelet subfamily according to the compression ratio and the ECG signal distortion. The proposed study is achieved through MATLAB simulations including the functions of wavelet toolbox, statistics and machine learning toolbox as well as the signal processing toolbox. In fact, the simulations are based on the function of the denoising and compression wavelet family to perform the OMP algorithm. This algorithm is developed for the sake of rapid MATLAB programming and simulation results.

Fig. 4.

Fig. 4

Flowchart of DWT compression study

The orthogonal matching pursuit (OMP) algorithm’s running time is dominated by the comparison of the data ECGout or the residual to the rows ϕi which complexity is OKL2 [63]. First, to reduce this complexity and for given decomposition matrix dimensions, D×D, ECGout is split into n successive segments ECGtoutl,1ln. Each segment, ECGtoutl, is composed of D successive samples, and ECGout is truncated to ECGtout. The ECGtout length, L^, is given in (7), where L is the ECGout data length and . is the floor function.

L^=nDwithn=LD 7

Second, the OMP is applied separately to each segment, ECGtoutl, for a given value of the wavelet decomposition level, J, and for the maximum number of iterations, K. Thus, each segment is compressed to ECGtcompl which length is equal to K. Third, the original segment is recovered to obtain ECGt^outl. In fact, the decomposition matrix is also applied separately to each ECGtcompl using (4) and (6) where ECGout is equal to ECGtoutl, K is equal to K and L is equal to D.

Fourth, from the second segment, also equal to the second iteration, l=2, of the DWT compression algorithm study given in Fig. 4, the recovered segment, ECGt^outl, is concatenated to the previous recovered segment, ECGt^outl-1. At the nth iteration, all the recovered segments are finally concatenated to obtain ECGt^out that represents the recovered data of ECGtout. Consequently, the OMP complexity becomes OnKD2 where K is the number of the selected rows from the decomposition matrix to represent each D-length segment. Fifth, the PRDcomp is computed to evaluate the DWT compression in terms of recovered signal quality for a given signal and a given subfamily.

Thus far, the five steps extract PRDcomp variations versus K for given wavelet subfamily, decomposition matrix dimensions, D×D, and wavelet decomposition level, J. It is important to keep the PRDcomp lower than 9% after sampling, quantizing and compression as it is the criterion for a low distortion and thus for a good diagnosis quality [43]. Due to ECG temporal and amplitude range variations [64], the number of coefficients, K, for a 9% PRDcomp varies from one signal to another. Therefore, the data compression efficiency evaluation is applied to all wavelet subfamilies for given dimension, D, and wavelet decomposition level, J, as far as many test ECG signals are concerned. A statistical comparison is processed, and the number of the optimal coefficients, Kopt, is obtained for the selected best-fit DWT. Aside from compression, transmission then diagnosis purposes as it is in this paper, DWT can also be used in the training process in order to extract significant ECG parameters. These parameters are keys to identify patients [2].

Finally, performances are evaluated after the best-fit wavelet subfamily selection for compression of uniform ECG data. The LC-ADC model is presented, and the performance evaluation test bench is proposed in the next subsection before evaluation metrics definitions to compare them to (A2) the LC-ADC performances.

LC-ADC Model and Evaluation

In this paper, the best candidate of (A2) the LC-ADC to tackle the comparative study to (A1) the SAR ADC with DWT compression is the floating-window type. Indeed, it has been a widely used and tested type during the two last decades [34, 38, 65]. The ECG analog signal, ECGin, is continuously compared to two reference levels. The in-between difference is the LC-ADC quantization step, q, that is given in (8) where FS is the LC-ADC full scale and M is the LC-ADC resolution.

q=FS2M 8

The two comparator outputs are INC and DEC regarding the up- or down-crossing direction, respectively. When an event is detected by the comparators, digital values of the reference levels are updated. Then, the digital-to-analog converter (DAC) delivers the updated analog reference levels. These steps result in increment or decrement of both crossed levels by q. The floating-window type inherits its name from the simultaneous update of both analog levels. Moreover, a logic control sets the digital value of the crossed level, ECGoutLC, as the output sample [66]. In fact, the LC-ADC signal dependency is due to this behavior. Thus, when the signal amplitude is not varying, the LC-ADC is not activated. That makes it more suitable for ECG signal acquisition than a Nyquist or sigma-delta ADC that samples ECG signal at each time period. In order to recover the signal from the LC-ADC samples, ECGoutLC, the time information is needed to time-localize the level crossings (LCs). Therefore, the time-tracking stage includes a synchronized counter at a reference counter frequency, FC. This counter enumerates the number of periods, 1/FC, to obtain the time interval, TI, between two successive LCs. The reference counter resolution is equal to N bits.

The block diagram of the described LC-ADC architecture is shown in Fig. 5. A detailed LC-ADC model is already implemented in MATLAB/SIMULINK and designed according to ECG signal specifications in terms of amplitude and frequency [66]. In order to recover the ECG signal, the samples at the LC-ADC output, ECGoutLC, besides the time intervals at the LC-ADC output, TI, are required. Alternatively, only DEC and INC outputs of both comparators instead of all samples, ECGoutLC, are enough to be transmitted to a distant control unit. In fact, from the first sample, ECGoutLC1, DEC values, INC values together with the LC-ADC quantum, the samples, ECGoutLC, can be computed with digital level updates and the logic control blocks.

Fig. 5.

Fig. 5

LC-ADC block diagram

Behavioral simulations are done with modeled ECG signals for normal, bradycardia and tachycardia [66] and then tested with real ECG records for normal, emotional or physical duress and various pathologies from Physionet database [21]. In fact, the LC-ADC is designed and modeled to cope with all ECG signals that give heart rates from 30 beats per minute (bpm) to 240 bpm [66]. Thus, the LC-ADC full scale, FS, is set to 10 mV, the reference counter frequency, FC, is equal to 10 kHz, and the reference counter resolution, N, is 12 bits for an M-bit LC-ADC where M is equal to 8. In the present paper, the LC-ADC performance evaluation test bench is illustrated in Fig. 6. As in Fig. 3, to obtain test ECG signals that represent analog signals, the input data, ECGdata, are up-sampled at 1 MHz to obtain the “analog” signal, ECGin, with a fine time axis. This “analog” signal is the LC-ADC model input. Besides, the signal reconstruction to obtain uniformly sampled signal at Fref, ECG^outLC, is operated due to linear interpolation with a p-bit resolution. In fact, the 1 kHz reconstruction frequency and the p-bit resolution are the same as the S/H frequency, Fref, and the resolution, m, respectively, in the SAR ADC to perform the quantitative comparative study. Afterward, the LC-ADC evaluation metrics are computed from three vectors of data obtained during the same time duration. They are the samples at the LC-ADC output, ECGoutLC, and the linear interpolation output, ECG^outLC, but also the SAR ADC output after truncation, ECGtout, as described in Fig. 3.

Fig. 6.

Fig. 6

Performance evaluation test bench of (A2) the LC-ADC

To show comparative curves of LC-ADC output and SAR ADC output, SAR ADC is chosen with 16-bit resolution and 1 kHz reference frequency. The test ECG data are the V1 lead of s0044lre from PTB diagnostic ECG (ptbdb) [21]. An example of samples at the LC-ADC output versus the samples at the SAR ADC output is given in Fig. 7. The patient of s0044lre has a mean cardiac rhythm of 67 beats per minute (bpm) and suffers from myocardial infarction which is a challenging medical emergency. In this example, the Q wave, that is, generally before the R wave, is not observed. Between the pointed data cursors, the LC-ADC delivers S samples, equal to 184, whereas the SAR ADC delivers L samples, equal to 782 samples.

Fig. 7.

Fig. 7

LC-ADC output versus ECG SAR ADC output

Between the pointed data cursors in Fig. 7, the SAR ADC delivers L samples, equal to 782 samples of 16 bits each. Alternatively, the LC-ADC delivers S samples. This number S is equal to 184 which needs 3680 bits if 8 bits per ECGoutLC and 12 bits per TI are sent. However, the LC-ADC can deliver only INC and DEC that are the required bits aside with the first crossed level, ECGoutLC1, to update LC-ADC levels and to provide the exact samples at the LC-ADC output before linear interpolation. This optimization makes the amplitude output requires only 376 bits. As far as the time intervals, TI, are concerned, they require 2208 bits for S samples. Thus, instead of 12512 bits at the SAR ADC output, the LC-ADC can send only 2584 bits. Thus, this LC-ADC with data optimization represents the architecture (A2). In case of (A1) the SAR ADC with DWT compression at BCR equal to 70%, the data length of 12,512 bits becomes equal to 3754 bits. That confirms the SAR ADC output compression’s importance of test ECG signals or the use of LC-ADC amplitude output optimization.

For the V1 lead of s0044lre, Fig. 8 shows different values of time, amplitude and total data lengths in order to compare the data lengths at the outputs of (A1) the SAR ADC with DWT compression, (A2) the LC-ADC and other architectures. In fact, six architectures are compared. They are the standalone SAR ADC, (A1) the SAR ADC with DWT compression, the SAR ADC with DWT compression which is followed by Huffman compression, the LC-ADC with complete amplitude output, (A2) the LC-ADC with optimization of amplitude output, the LC-ADC with optimization of amplitude output and Huffman compression which is applied to time intervals. In this paper, (A2) the LC-ADC means that the LC-ADC is tested with optimization of amplitude output. For the LC-ADC with optimization of amplitude output and Huffman compression, the time intervals, TI, could be compressed by a lossless compression method. However, this is not the aim of this paper as Huffman will be a further compression after the DWT compression or of the LC-ADC time interval output. Thus, the authors carry out their comparative study between (A1) and (A2).

Fig. 8.

Fig. 8

ECG sensor with different signal acquisition and data compression architectures, namely (A1) the SAR ADC with DWT compression and (A2) the LC-ADC with optimization of data transmission

This paper focuses on the event-driven ECG sensor to reduce the message data length. The aim is to optimize data transmission from a healthcare device to a distant control unit in order to analyze patients’ ECG signals after ECG signal reconstruction, features’ extraction can be added, and then, patients’ identification and authentication can be operated. The authors detail performance evaluation metrics in the next subsection in order to highlight the advantage of the event-driven ECG sensor regarding SAR ADC-based ECG sensor. Selected ECG signals, ECGdata, for the inputs of performance evaluation test benches are also included.

ECG Signals and Evaluation Metrics

Performance evaluation test benches of (A1) the SAR ADC with DWT compression and (A2) the LC-ADC are illustrated in Figs. 3 and 6, respectively. The test ECG data at their input, ECGdata, are downloaded from Physionet [21]. In this paper, (A1) and (A2) are tested with 50 different normal or pathological ECG signals. These signals are introduced in Annex A. The selected signals cover normal and apnea, ECG signals in addition to many pathological ECG signals but also ECG signals during emotional or physical duress which is described by ST segment variation. Pathologies are from bradycardia, arrhythmia, atrial fibrillation and tachycardia to the most critical one which is myocardial infarction. The selected signals are from many ECG databases as Apnea ECG, MIT-BIH arrhythmia (mitdb), PTB diagnostic ECG (ptbdb), long-term ST (lstdb), MIT-BIH atrial fibrillation (afdb), MGH-MF waveform (mghdb) which have different resolutions and sampling frequencies.

Evaluation metrics of (A1) the SAR ADC with DWT compression and (A2) the LC-ADC performance evaluation test benches are computed during the same given time duration, DT. Evaluation metrics of (A1) the SAR ADC with DWT compression performance evaluation test bench are cited below.

  • The DWT compression percent root-mean-square difference, PRDSAR&DWT, is computed to evaluate the recovered signal quality after DWT decompression.

  • The data length of the SAR ADC with DWT compression, data__lengthSAR&DWT, is the data length in bits of the selected coefficients after DWT compression during DT.

  • The DWT bit compression ratio, BCRSAR&DWT, is the bit compression ratio of DWT compression which is applied on the SAR ADC output.

First, the PRDSAR&DWT is computed as defined in (9) from (3) where L is replaced by nD, ECGout by ECGtout, ECG^out by ECGt^out and ECG¯out by ECGt¯out.

PRDSAR&DWT%=100×i=1nDECGtouti-ECGt^outi2i=1nDECGtouti-ECGt¯out2 9

In fact, the first step of the DWT compression study in Fig. 4 is the segmentation of L-length ECGout into n segments of length equal to D and their concatenation into L^-length ECGtout, as given in (7). When extracting PRDSAR&DWT, results must be discussed according to SAR ADC effective number of bits (ENOB). Indeed, ENOB represents the real resolution of the ADC. Thus, signal quality of the overall SAR ADC with DWT compression is PRDSAR&DWT versus ENOB and not versus the SAR ADC resolution, m.

Second, the DWT compression is applied n times on n segments, ECGtoutl, which length is equal to D. The data length, nK, is thus obtained from the coefficients’ data length of the DWT compression, ECGtcomp. Moreover, as each coefficient is quantized to m bits, the data length in bits of (A1) is given in (10).

data__lengthSAR&DWTbits=mnK 10

Conversely, the bit compression ratio, BCRSAR&DWT, of SAR ADC with DWT compression depends only on the BCR of DWT compression as defined in (2). Indeed, third, BCRSAR&DWT is computed from (2) by replacing L by nD and Lbcomp by data_lengthSAR&DWT which is defined in (10). The third evaluation metric of (A1) is thus given in (11).

BCRSAR&DWT%=mnD-mnKmnD=D-KD 11

If the operated DWT compression corresponds to the selected best-fit DWT that returns Kopt optimal coefficients, data__lengthSAR&DWT in (10) and BCRSAR&DWT in (11) have to be computed for K equal to Kopt.

In this paper, to compare performances of (A1) the SAR ADC with DWT compression and (A2) the LC-ADC, the SAR ADC number of samples, nD, and the LC-ADC number of samples, S, must be taken in a given duration, DT, respectively.

Evaluation metrics of (A2) the LC-ADC performance evaluation test bench are four:

  • The LC-ADC percent root-mean-square difference, PRDLC - ADC, is computed to evaluate the recovered signal quality after linear interpolation.

  • The LC-ADC mean sampling frequency, Fmean, is computed to approximate the data output frequency.

  • The LC-ADC data length, data__lengthLC - ADC, is the data length in bits of the LC-ADC outputs during DT.

  • The LC-ADC bit compression ratio, BCRLC - ADC, is computed to evaluate the data reduction at the LC-ADC outputs.

As far as the PRDLC - ADC is concerned, on the one hand, the ECG signal quality is evaluated after linear interpolation at Fref to recover p-bit uniformly sampled ECG signal, ECG^outLC. On the other hand, during DT, ECGoutLC has S samples. Consequently, the recovered digital signal, ECG^outLC, has exactly nD samples during the same duration, DT. The PRDLC - ADC is hence defined in (12) where ECGtout is the SAR ADC equivalent output after truncation to nD samples as given in Fig. 6.

PRDLC - ADC%=100×i=1nDECGtouti-ECG^outLCi2i=1nDECGtouti-ECGt¯out2 12

However, from the time intervals at the LC-ADC output, the mean sampling frequency, Fmean, is computed as in (13) where TIi is the ith time interval between two successive LCs, FC is the reference counter frequency and S is the number of time intervals.

Fmean=1S.FCi=1STIi-1 13

As far as the LC-ADC data length in bits, data__lengthLC - ADC, is concerned, it is equal to the data length in bits of the samples at the LC-ADC output added to the data length in bits of the time intervals at the LC-ADC output. On the one part, during DT, the number of samples, S, is reduced as in Fig. 8. In fact, only the first M-bit amplitude sample is required besides all the INC and DEC values that describe the other samples, ECGoutLC. Thus, the data length in bits of the samples at the LC-ADC output is equal to M+2S-1. On the other part, the data length in bits of the time intervals is NS with N the reference counter resolution. Finally, the data__lengthLC - ADC is given in (14).

data__lengthLC - ADCbits=N+2S+M-2 14

By the end, the LC-ADC bit compression ratio, BCRLC - ADC, derives from the definition in (2). Indeed, L is replaced by nD because of truncation of the samples at the SAR ADC output, m by p which is the linear interpolation resolution, and Lbcomp by data_lengthLC - ADC. Thus, the last evaluation metric of (A2) is given in (15).

BCRLC - ADC%=pnD-N+2S+M-2pnD 15

The next section introduces MATLAB simulation results regarding the performance evaluation test benches. Then, it deals with quantitative comparison between (A1) the SAR ADC with DWT compression and (A2) the LC-ADC in terms of the defined evaluation metrics.

Evaluation Results and Discussions

To conduct the compression study, 50 ECG signals were applied as test input. They are normal and pathologic ECG signals selected with different shapes, amplitude ranges and pathology types from Physionet database [21]. There are also Apnea-ECG signals and ECG signals during emotional or physical duress which is described by ST segment variation. Each ECG input signal, ECGdata, is applied at the input of the evaluation performance test bench of the SAR ADC with DWT compression. The SAR ADC samples at 1 kHz reference frequency. To compress SAR ADC output, all the 54 available MATLAB wavelet subfamilies of the Coiflet (coif), Daubechies (db), Symlet (sym), DMeyer (dmey), biorthogonal (bior) and reverse biorthogonal (rbio) families are generated using MATLAB functions. Simulations are operated considering three parameters: the decomposition matrix dimensions, D×D, the wavelet decomposition level, J, and the number of selected coefficients to compress the signal, K. The PRDSAR&DWT upper limit is equal to 9%.

This section starts with some preliminary results and best-fit wavelet subfamily with K, D and J selection for the SAR ADC with DWT compression. Afterward, the performances are discussed according to the selected DWT, K, D and J. In addition, each test ECG signal, ECGdata, is also applied at the input of the evaluation performance test bench of the LC-ADC. Then, the results for the LC-ADC output are presented and discussed. Finally, the authors compare the compression and signal quality of (A1) the SAR ADC with DWT compression versus (A2) the LC-ADC in terms of percent root-mean-square difference, bit compression ratio and data length in bits.

Preliminary Results

Preliminary simulations were done for a single test ECG signal which is the I lead of s0044lre from PTB diagnostic ECG (ptbdb) with all the wavelet subfamilies of MATLAB. Each subfamily is set for wavelet decomposition levels, J, from 2 to 4, and for decomposition matrix dimensions, D×D, equal to 64 × 64. Besides, as the decomposition matrix is invertible and matrix dimensions are required to be a power of 2 to design analysis and synthesis filters, the authors observe preliminary results also for D equal to 128, 256 and 512. On the one hand, lower values of D may not show enough decomposition matrix rows with many zeros to obtain data decompositions in small and high coefficients. On the other hand, increasing dimensions from 64 to 512 has a slight effect on compression ratio and higher values of D show precisely the same results. For the selected ECG signal and for a PRDSAR&DWT upper limit that is equal to 9%, Fig. 9 shows that the BCRSAR&DWT varies between 54.7% and 93.5% for many configurations of D and J. For the test ECG signal, the best wavelet subfamilies are the bior3.3, bior3.5, bior3.7 and bior3.9 with D equal to 512 and J equal to 4. Increasing the wavelet decomposition level, J, leads to a better compression performance, namely from 2 to 3. For example, Table 5 gives the BCRSAR&DWT values for bior3.3 and Haar for J from 2 to 5. These help to observe that J is fixed till third wavelet decomposition level because for higher values of J, performances do not vary significantly. In fact, for bior3.3 wavelet subfamily, BCRSAR&DWT has increased from 75 to 86.72% when J varies from 2 to 3,whereas increasing J from 3 to 4 adds almost 5% to the BCRSAR&DWT, and from 4 to 5 adds only 1.56% to the previous BCRSAR&DWT.

Fig. 9.

Fig. 9

Bit compression ratio of DWT compression on the SAR ADC output for the I lead of s0044lre ECG signal from PTB diagnostic ECG (ptbdb) for D=64, D=128, D=256 and D=512 (from top to down) for different values of wavelet decomposition level

Table 5.

Bit compressio ratio,BCRSAR&DWT(%), of SAR ADC with DWT compression for bior3.3 and Haar wavelet subfamilies for D=128

DWT Level
2 3 4 5
bior3.3 75% 86.72% 91.41% 92.97%
Haar 73.44% 82.03% 85.16% 85.94%

The performances in terms of PRDSAR&DWT concerning the number of decomposition matrix rows, K, retained by the OMP are also analyzed for (J, D) equal to (2, 64), (2, 128), (3, 64) or (3, 128) for all test ECG signals. For the sake of clarity, these results are not presented in this paper since that requires four figures for each subfamily. As expected, the PRDSAR&DWT increases when the number of selected rows, K, decreases. In fact, the more decomposition matrix rows are discarded from data representation, the more information is lost. In the case of bior3.3 wavelet subfamily, 18.75–28.12% of the decomposition matrix rows are needed for a 9% PRDSAR&DWT for all the test ECG signals depending on the decomposition matrix dimensions and the wavelet decomposition level. Thus, the correspondent BCRSAR&DWT varies from 71.87 to 81.25%. The maximum value of K is equal to 17 and 36 for J equal to 2 in the case of 64 and 128 matrix dimensions, respectively. The minimum value of BCRSAR&DWT is equal to 73.43% and 71.87%, respectively. For J equal to 2 and D equal to 64, at least 26.56% of bior3.3 wavelet coefficients must not be zeroed to recover ECG signals with 9% PRDSAR&DWT or less. This corresponds to a minimum value of BCRSAR&DWT equal to 73.43%. Furthermore, for J equal to 3, the minimum value of K is equal to 8 and 11 for D equal to 64 and 128, respectively. These are promising values for BCRSAR&DWT equal to 87.5% and 91.4% but guarantee a 9% PRDSAR&DWT only for one test ECG signal. To sum up, preliminary simulation results show that bior3.3 wavelet subfamily is not always the best choice to compress the uniformly sampled ECG signals if the 9% PRDSAR&DWT criterion is respected. The best-fit DWT selection results are presented in the next subsection.

Best-Fit DWT Selection

To achieve a complete study of the DWT compression after SAR ADC, all available MATLAB orthogonal and biorthogonal subfamilies have been applied to the 50 real ECG signals as in the performance evaluation test bench of Fig. 3. Compression results are obtained for (L, D) equal to (2, 64), (2, 128), (3, 64) or (3, 128). Using 54 wavelet subfamilies, compression results are studied with MATLAB boxplot function from statistics and machine learning toolbox. In fact, the boxplot function is a statistical representation that describes the data distribution by delimiting with a box its 25th percentile (first quartile) and 75th percentile (third quartile), and by two horizontal lines its minimum and maximum values. It also indicates with a line inside the box the median value of the distributed data.

In Fig. 10, a statistical variation range of BCRSAR&DWT is presented for the wavelet subfamilies. Using the flowchart in Fig. 4, the number of the most significant coefficients, K, satisfying the 9% PRDSAR&DWT varies according to the wavelet subfamily, J, D and the test ECG signal. Many wavelet subfamilies are not appropriate to the compression of the uniformly sampled ECG signals regarding others, namely the rbio3.1 wavelet. In fact, it presents the highest variation range of K for all decomposition matrix dimensions and wavelet decomposition levels. This choice delivers the lowest variation range of BCRSAR&DWT that was already detected in the preliminary results presented in Fig. 9. In Fig. 10 (a), there are some exceptions, known as outliners, that are the values outside the boxplot range between the minimum and maximum values. The outliners are present for subfamilies which median, first quartile and third quartile are identical. Moreover, the range between the maximum and the minimum is proportional to the range between the first and third quartiles. That hence explains the presence of outliners suggesting that some test ECG signals use more or less coefficients than almost all the test ECG signals. For example, db1 selects 17 coefficients and BCRSAR&DWT is equal to 73.43% for almost all the test ECG signals. However, it rarely selects 13, 15 or 16 coefficients to obtain higher bit compression ratios.

Fig. 10.

Fig. 10

Selected coefficient number after DWT compression of SAR ADC outputs regarding a PRD equal to 9% versus wavelet subfamilies for (2, 64), (2, 128), (3, 64) and (3 ,128) configurations (from top to down)

Still, from Fig. 10, regarding all the subfamilies, the authors focus on the highest value of the first quartile among the 54 subfamilies rather than the highest value of the third quartile, the maximum, the minimum or the median. This ensures that at least 25% of the test ECG signals satisfy a PRDSAR&DWT of at most 9% for a given wavelet subfamily with the lowest value of BCRSAR&DWT as defined in (11).

Thus, according to Fig. 10, for J equal to 2 and D equal to 64, bior3.3, bior3.5, bior3.7 and bior3.9 wavelet subfamilies that ensure Kopt equal to 17 and BCRSAR&DWT equal to 73.43% are the selected ones. These subfamilies can reach a 76.56% BCRSAR&DWT for 75% of the test ECG signals. These results and those for other configurations of J and D are summarized in Table 6.

Table 6.

Wavelet decomposition results for a 9% PRDSAR&DWT

J,D Kopt BCRSAR&DWT (%) Wavelet subfamilies
(2, 64) 17 73.43 bior3.3, bior3.5, bior3.7, bior3.9
(2, 128) 32 75.00 bior3.1, bior3.3, bior3.5, bior3.7, bior3.9
(3, 64) 10 84.37 coif2, rbio1.3, db4, bior2.2, sym5
(3, 128) 20 84.37 bior3.3, bior3.5, bior3.7, bior3.9

In Table 6, it is shown that selected subfamilies are similar except for (J, D) equal to (3, 64). Besides, dimension increase does not importantly affect the compression ratio but requires two times larger number of ECG samples to be compressed. For D equal to 64 and wavelet decomposition level, J, going from 2 to 3 increases 1.5 times the complexity of the analysis tree. In fact, three high-pass (HP) filters and three low-pass (LP) filters are required instead of two HP filters and two LP filters. For example, for bior3.3 subfamily which is the least complex among selected subfamilies for (J, D) equal to (2, 64), each LP analysis filter has 8 coefficients and each HP analysis filter has 4 coefficients. However, in the biorthogonal family, bior2.2 and bior3.1 subfamilies have the least numbers of filter coefficients. The factorized values of filter coefficients are given in Table 7. The number of operations to evaluate filter coefficients is the same, but the bior2.2 filter processing requires more processing cycles as it needs 5 cycles instead of 4 for bior3.1. In addition, from an implementation point of view, using those biorthogonal wavelet subfamilies leads to implement multiplier-free filters in the tree-structured filter banks. As the coefficients of such filters are the addition of powers of 2, there is no added quantization error while performing DWT decomposition. The quantization is only used to code 2 on m bits which is equal to the SAR ADC resolution.

Table 7.

Bior2.2 and bior3.1 analysis filter coefficients

Filter coefficients

bior2.2

LP analysis filter

2×0,-2-3,2-2,2-2+2-1,2-2,-2-3

bior2.2

HP analysis filter

2×0,2-2,-2-1,2-2,0,0

bior3.1

LP analysis filter

2×-2-2,2-2+2-1,2-2+2-1,-2-2

bior3.1

HP analysis filter

2×-2-3,2-3+2-2,-2-3-2-2,2-3

Thereby, following the SAR ADC and for all test ECG signals, the best-fit configuration of DWT compression is obtained for bior3.1 subfamily with J equal to 2, D equal to 64, K equal to 17 and thus BCRSAR&DWT equal to 73.43%. To add some margin to PRDSAR&DWT, the optimal value of K, Kopt, is fixed at 18. Thus, BCRSAR&DWT is always equal to 71.875% as the chosen value of 18 coefficients out of 64 coefficients is obtained from 64 samples at the SAR ADC output regardless SAR ADC resolution. The SAR ADC with DWT compression performance evaluation test bench of Fig. 3 is then processed for different ECGdata and configurations. In fact, the test ECG signal quality in terms of PRDSAR&DWT depends on the effective number of bits, ENOB, equal to 8, 10, 12, 14 and 16. The results are summarized in Fig. 11 and show the best values of DWT compression only for ENOB higher than 8. In fact, for ENOB equal to 8 bits, the PRDSAR&DWT values are roughly between 3.75% and 5.5% for 75% of the test ECG signals. For ENOB equal to 10, 12, 14 and 16, PRDSAR&DWT is generally between 2.5% and 4.5% for 75% of the test ECG signals. In the next subsection, the authors compare LC-ADC evaluation metrics, PRDLC - ADC, Fmean and BCRLC - ADC, for the 50 test ECG signals.

Fig. 11.

Fig. 11

PRDSAR&DWT versus SAR ADC effective number of bits where SAR ADC is followed by bior3.1 compression with BCRSAR&DWT equal to 71.87%

LC-ADC Performances

LC-ADC behavioral simulations are done with the same 50 real ECG records that are used for the SAR ADC with DWT compression performance evaluation test-bench. As explained in Fig. 6, the signals are up-sampled at 1 MHz frequency. This step is done to obtain signals that are representative of analog signals with a fine time axis. Then, since the LC-ADC resolution, M, has already been chosen in [66], the converter model is only simulated with a resolution equal to 8. This value is typical for ECG amplitude quantization and also ensures good quality signal reconstruction which requires at least 8-bit resolution [67, 68]. The LC-ADC full scale is set to 10 mV. As far as the counter is concerned, the resolution, N, is 12 bits and the reference counter frequency, FC, is 10 kHz for the 8-bit LC-ADC. For the LC-ADC performance evaluation test bench, simulations are carried out over the same time duration DT of the SAR ADC with DWT compression. The simulations are done during DT equal to one second for every test ECG signal which contains at least one P wave, one QRS complex and one T wave.

The BCRLC - ADC and the Fmean performances are given in Fig. 12, for all the 50 test ECG signals. Generally, the more the samples at the LC-ADC output are, the more the mean sampling frequency is, the less the bit compression ratio is. The BCRLC - ADC varies from 69.9 to 92.35%. Certainly, and at the opposite of the samples at the SAR ADC output, the results are different from a test ECG signal to another since the number of samples depends on the test ECG signal amplitude and frequency. The 8-bit LC-ADC with 12-bit reference counter samples at a mean sampling frequency between 109 Hz and 348 Hz, whereas generally used SAR ADC sampling frequency is always 1 kHz as provided in the present paper and many works in the literature.

Fig. 12.

Fig. 12

Mean sampling frequency and BCRLC - ADC versus ECG signal number with BCRLC - ADC computed for a 16-bit reconstruction resolution

As far as the PRDLC - ADC values are concerned, they are also evaluated for all test ECG signals and presented with BCRLC - ADC values which are computed for p equal to 16 bits in linear interpolation algorithm. The results in Fig. 13 show no correlation between these two performances. The PRDLC - ADC values vary from 1.25 to 7.25%, whereas BCRLC - ADC values move from 69.9 to 92.35%. To sum up, LC-ADC sampling efficiency depends on the test ECG signal. In fact, for high-frequency and high-amplitude waves, the LC-ADC activity increases.

Fig. 13.

Fig. 13

PRDLC - ADC and BCRLC - ADC versus ECG signal number for LC-ADC with BCRLC - ADC computed for a 16-bit reconstruction resolution

The next subsection compares the percent root-mean-square difference, the bit compression ratio and the data length in bits of (A1) the SAR ADC with DWT compression and (A2) the LC-ADC in ECG sensors.

LC-ADC Versus SAR ADC with bior3.1 Compression

The bit compression ratio of the LC-ADC, BCRLC - ADC, as given in (15), is computed for a p-bit reconstruction resolution and 1 kHz reconstruction frequency. Indeed, the LC-ADC input is an analog signal, the LC-ADC output is a digital compressed data, whereas compression ratio is a digital performance. For a given signal, the BCRLC - ADC values increase when reconstruction resolution, p, varies from 8, 10, 12, 14 to 16. For instance, if p is equal to 16, the reconstruction recovers a 16-bit code every 1 ms independently from their positions in the test ECG signal. Thus, for a given signal, it is obvious that the 16-bit reconstructed LC-ADC output is more performant in terms of bit compression ratio than lower reconstruction resolutions. As far as (A1) is concerned, DWT compresses all the SAR ADC samples including clinically and nonclinically relevant parts in the test ECG signal. Therefore, as shown in Fig. 14, the BCRSAR&DWT is equal to 71.87% for all the test ECG signals. Adversely, LC-ADC compression is signal dependent. More than 50% of the test ECG signals are compressed with BCRLC - ADC higher than 80% if computed regarding a 16-bit reconstruction resolution. For p equal to 8, 75% of test ECG signals offer a BCRLC - ADC equal to 70.17% which is almost the same value as BCRSAR&DWT for the best-fit DWT compression with J equal to 2 and D equal to 64 after 1-kHz SAR ADC.

Fig. 14.

Fig. 14

BCRSAR&DWT for SAR ADC with best-fit DWT compression versus BCRLC - ADC where the BCRLC - ADC is computed regarding p-bit reconstruction resolution

As mentioned in the SAR ADC state of the art, a 10-bit SAR ADC provides an ENOB of 6.63 and a 12-bit SAR ADC provides an ENOB of 11.77. The authors in this subsection are interested only in ENOBs that are equal to 12, 14 and 16. Thus, PRD results are summarized in Fig. 15 where the PRDSAR&DWT results are computed for SAR ADC with the best-fit DWT compression. That is compared to the PRDLC - ADC results according to the reconstruction resolutions, p, which are equal to the selected values of ENOB. In fact, the LC-ADC delivers non-uniform samples and requires linear interpolation to recover p-bit 1 kHz uniformly sampled signal. The recovered signal is used to evaluate the PRDLC - ADC regarding the SAR ADC output for a given value of ENOB equal to p. Figure 15 shows that the LC-ADC almost performs the same as the SAR ADC with DWT compression in terms of PRD. Thus, as this paper aims at optimizing data transmission in order to save data transmission power in RF sensor networks, the authors quantitatively compare data lengths in bits of (A1) the SAR ADC with DWT compression to the ones of (A2) the LC-ADC.

Fig. 15.

Fig. 15

PRDSAR&DWT for SAR ADC of ENOB equal to 12, 14 or 16 with best-fit DWT compression versus PRDLC - ADC which is computed for p=ENOB

Since LC-ADC has non-uniform outputs, the authors need to adopt a same time duration, DT, to compare data length of (A1) to (A2) for all the test ECG signals. If this DT is equal to one second, an m-bit SAR ADC at 1 kHz S/H frequency delivers m kbits during one second of conversion. This amount is reduced to m1-BCRSAR&DWT kbits which is due to the reached compression ratio, BCRSAR&DWT, equal to 71.87%. For a 12-bit SAR ADC with DWT compression, this amount is 3.375 kbits. As far as the proposed 8-bit LC-ADC with 12-bit reference counter is concerned, the mean sampling frequency is between 109 Hz and 348 Hz which depends on the test ECG signal.

Besides, it delivers 3.786 kbits for 75% of test ECG signals during one second of continuous-time sampling as presented in Fig. 16. This data length in bits is located between data lengths at the outputs of 12-bit and 14-bit SAR ADCs with DWT compression. Furthermore, the proposed LC-ADC is more advantageous than a 16-bit SAR ADC with DWT compression for 94% of the test ECG signals. Thus, it will be more beneficial to use the LC-ADC instead of the SAR ADC with DWT compression especially for recent ECG acquisition systems that require 16-bit resolution [7, 9].

Fig. 16.

Fig. 16

Data length for m-bit SAR ADC with best-fit DWT compression versus LC-ADC data length during one second

From this quantitative comparison, the performances of (A1) the SAR ADC with DWT compression show similarities, advantages and disadvantages when compared to (A2) the LC-ADC. In fact, they recover approximately the same ranges of percent root-mean-square differences, PRDSAR&DWT, as the ones of PRDLC - ADC, for different values of ENOB and of linear interpolation resolutions. As far as compression ratio is concerned, (A1) the SAR ADC with DWT compression has a constant value of bit compression ratio,and thus, data length is expected to be reduced before data transmission. However, the LC-ADC and 8-bit linear interpolation deliver a BCRLC - ADC equal to 70.17%, for 75% of the test ECG signals, which is slightly lower than the constant value of compression in (A1). Finally, the drawback of (A1) is that the data length is always constant as the bit compression ratio is constant too. That is considered as a drawback because the data length in bits of (A2) is lower than the data length in bits of:

  • the 16-bit SAR ADC with DWT compression for 94% of the test ECG signals,

  • the 14-bit SAR ADC with DWT compression for 82% of the test ECG signals,

  • the 12-bit SAR ADC with DWT compression for 68% of the test ECG signals.

Consequently, if existent SAR ADC-based ECG sensor requires more than 14-bit resolution, the LC-ADC can be perceived as a better architecture to obtain the same signal quality but with lower message data length to transmit to the physician distant control unit. Nonetheless, if the existent SAR ADC-based ECG sensor requires a resolution which is lower than or equal to 12 bits, the ECG sensor’s production should be maintained if it meets the patient and the physician requirements.

Finally, the authors conducted this quantitative comparative study while considering the ECG signal quality in the presence of noise and the ECG signal acquisition circuit complexity. On the one hand, the authors have already conducted another work in order to assess the proposed design according to systematic time-quantization error and analog components’ non-idealities [66]. The findings of this work can be summarized as follows: First, with offset voltage errors, the PRDLC - ADC is kept lower than 9%, thus reflecting good quality signals. Second, when referring to simulation results with DACs nonlinearities and commercialized DACs specifications, good signal quality was reached.

On the other hand, architecture complexity can be identified according to three performances: circuit area, circuit propagation delay and circuit power consumption. For instance, in 180 nm CMOS technology, the circuit areas of (A1) and (A2) are approximately equal to 0.45 mm2 and 0.2 mm2, respectively. In fact, SAR ADC area is computed from the mixed-signal chip of [14], DWT compression area is computed from the signal processing chip of [53], and estimated LC-ADC area from the layout is already computed in one of the authors’ previous work [66]. Thus, in terms of area the LC-ADC is also a promising architecture. As far as the propagation delay is concerned, the authors evaluate the propagation delays of the comparator which is equal to 5 ns and the digital levels’ update circuit which is equal to 0.8 ns. Added to that, the DAC’s settling time in the 180 nm UMC CMOS technology is estimated to be at most 62 ns. The total delay is at most 68 ns for the LC-ADC. That is in fact lower than the reference counter period equal to 0.1 ms to prevent from the increase in the time offset and hence from harmonic distortion. The LC-ADC requires in its architecture an 8-bit DAC that provides the output 62 ns after the input is injected. Nonetheless, the SAR ADC requires in its architecture a 12-bit DAC that provides the output more than 62 ns after the input is injected. Moreover, the DWT compression will also require more propagation delay to deliver the DWT compression output as it is composed of filters, decimators, detection, estimation and level thresholding [53]. Noting that power consumption is equal to 78.8 µW in case of SAR ADC with two different sampling frequencies and Huffman compression for test ECG signal [14], whereas it is only equal to 2.42 µW in case of LC-ADC for test sinewave signal [20] which is not really concluding. In fact, to the best of the authors’ knowledge, no other works have been conducted to measure power consumption with test ECG signals.

To conclude, PRD of (A1) the SAR ADC with DWT compression and (A2) the LC-ADC are similar despite the presence of noise as in [66]. Distinctively, data length to be transmitted after (A2) is lower than the one to be transmitted after (A1) with 14-bit resolution as shown in Fig. 16 for 82% of the test ECG signals. In addition, (A2) is also preferable than (A1) covering area, propagation delay and power consumption. Thus, the leaders in design and manufacturing of mixed-signal processing integrated circuits should more investigate on the event-driven ECG sensors. For example, an autonomous wireless sensor node with asynchronous ECG monitoring has been proposed with a 3.8-mm2 area and 9.7-µW power consumption in 180 nm CMOS technology [20] but not completely tested with ECG signals yet.

Conclusion

This paper quantitatively compares the SAR ADC with DWT compression to the LC-ADC starting with ECG signal acquisition and data compression state of the art. Then, it reviews the required tools to deal with the quantitative comparative study, namely the SAR ADC architecture, the DWT definition, its parameters and its related compression method. It also explains the performance evaluation test bench of (A1) the SAR ADC with DWT compression architecture. Moreover, (A2) the LC-ADC architecture, the LC-ADC model and performance evaluation test bench are described as well as the evaluation metrics of both compared architectures. Finally, performance evaluation results are discussed to show similarities between the two architectures in terms of signal quality for different configurations. However, the aim is to optimize data transmission and therefore reduce the ECG sensor power consumption of the RF transmission. Accordingly, the proposed 8-bit LC-ADC with 12-bit and 10-kHz reference counter delivers less data than the 14-bit SAR ADC with DWT compression for 82% of the test ECG signals. Consequently, the 16-bit or 14-bit SAR ADC would be replaced by the LC-ADC in the commercialized ECG sensors.

This quantitative comparative study and its uses are limited to models, algorithms and simulations with test ECG signals from Physionet databases. Meanwhile, SAR ADC results should be improved by the acquisition of ECG signals from existent ECG test devices instead of test ECG signals from Physionet databases. Then, DWT compression can be implemented on field programmable gate array (FPGA) and tested on the acquisition of ECG signals. Besides, more digital signal processing and lossless compression will be also implemented and tested. In fact, a lossless compression, such as the Huffman coding, can be adopted for the DWT selected coefficients. As far as the LC-ADC is concerned, the authors have to send to the foundry the CMOS circuit design files. An LC-ADC test platform will be achieved in order to test ECG signal acquisition. Afterward, the comparison of (A1) the SAR ADC with DWT compression to (A2) the LC-ADC will be more effective. Following that, a lossless compression, such as the Huffman coding, can be adopted after the LC-ADC to the values that are higher than a given amplitude level and to all time intervals. Future works will also deal with an event-driven ECG device with online ECG features’ extraction from the LC-ADC optimized outputs, identification and authentication of patients. Additionally to the LC-ADC, the bradycardia requires the highest value of the reference counter resolution that is equal to 12 bits. Meanwhile, other pathologies require lower values of the reference counter resolution. Consequently, pathology-centric event-driven ECG sensors which are different in terms of LC-ADC’s reference counter resolution would be proposed.

Acknowledgements

The authors would like to thank the CMCU (Comité Mixte de Coopération Universitaire) as it financially supported the Wiobio’ACT project. This project involves cooperation between University of Bordeaux and the Higher School of Communications of Tunis (SUP’COM) from 2015 to 2017. The funding program is Partenariat Hubert Curien-Utique (PHC-Utique) 2015 N°32755YC.

Annex A

Test ECG data, ECGdata, are applied at the input of evaluation performances test benches of Figs. 3 and 6. That data are downloaded from Physionet [21], and 50 different normal, apnea or pathological ECG signals are used as well as ECG signals in the presence of emotional and physical duress. The selected signals, as presented in Table 8, cover normal and Apnea-ECG signals in addition to many pathological ECG signals. Pathologies are from bradycardia, arrhythmia, atrial fibrillation and tachycardia to the most critical one which is myocardial infarction.

Table 8.

Test ECG signals from Physionet database

ECG signal number Record/lead Pathology or description or normal Segment duration (s) Resolution/sampling frequency
Physionet database: Apnea-ECG (apnea-ecg)
 1 a03/ECGa Apnea 2.57 12 bits/100 Hz
 2 a05/ECGa
 3 a07/ECGa
 4 a10/ECGa
 5 a11/ECGa
 6 a15/ECGa
 7 a16/ECGa
 8 a17/ECGa
Physionet database: MIT-BIH arrhythmia (mitdb)
 9 202/IIb Bradycardia (49–59 bpm) 2.85 11 bits/360 Hz
 10 217/IIb PVCa
 11 217/V1b
 12 219/IIb Bradycardia (38–59 bpm)
 13 221/V1b Atrial fibrillation (47–110 bpm)
 14
 15 223/IIb PVCa
 16 230/IIb Wolff–Parkinson–White (WPW) syndrome
 17 231/IIb Second-degree heart block
 18 233/IIb PVCa
Physionet database: PTB diagnostic ECG (ptbdb)
 19 s0044lre/Ib Myocardial infarction 2.05 16 bits/1000 Hz
 20 s0044lre/IIb
 21 s0044lre/aVRb
 22 s0044lre/V1b
 23 s0044lre/aVLb
 24 s0101lre/V2b
 25 s0101lre/V3b
 26 s0101lre/V5b
 27 s0101lre/V6b
 28 s0101lre/Vxc
Physionet database: long-term ST (lstdb)
 29 s20011/IIe ST segment variation 4.1 12 bits/250 Hz
 30 s20011/V2e 2.05
 31 s20041/ECGf 1.02
 32 s20111/V2e 4.1
 33 s20141/IIe 2.05
Physionet database: MIT-BIH atrial fibrillation (afdb)
 34 04048/ECG1g Atrial fibrillation 2.05 12 bits/250 Hz
 35 04746/ECG2g
 36 04936/ECG2g
 37 05121/ECG1g
Physionet database: PTB diagnostic ECG (ptbdb)
 38 s0476_re/V2b Normal 2.05 16bits/1000 Hz
 39 s0476_re/V3b
 40 s0476_re/V4b
 41 s0476_re/Vxc
 42 s0476_re/V6b
 43 s0476_re/V5b
Physionet database: MGH-MF waveform (mghdb)
 44 mgh010/Ib Ventricular hypertrophy 1.42 12 bits/360 Hz
 45 mgh017/IIb
 46 mgh027/IIb
 47
Physionet database: MIT-BIH arrhythmia (mitdb)
 48 102/V5b Paced rhythm 2.85 11 bits/360 Hz
 49 103/IIb APC
 50 103/V2b APC

The selected signals are from many ECG databases as Apnea-ECG (apnea-ecg), MIT-BIH arrhythmia (mitdb), PTB diagnostic ECG (ptbdb), long-term ST (lstdb), MIT-BIH atrial fibrillation (afdb), MGH-MF waveform (mghdb) which have different resolutions and sampling frequencies. A record can contain 1 lead as record “a03” from “apnea-ecg,” 2 leads as for record “04048” from “afdb,” 12 leads as for “202” from “mitdb” and 15 leads as for “s0101lre” from “ptbdb.” Thus, Table 8 contains information covering record reference and its lead reference, pathology or record description or normal ECG, time duration of the tested segment that contains at least one P wave, one QRS complex and one T wave, and sampling information according to the database acquisition. These information are provided to all the test ECG signals that are numbered from 1 to 50. Notations are explained as follows:

  1. Only one available ECG lead

  2. Standard 12-lead ECG

  3. Frank vectorcardiographic system (Vx, Vy, Vz)

  4. Premature ventricular contractions (PVC)

  5. Two ECG leads that are obtained by ambulatory records

  6. 24-hour ECG holter monitoring with rhythm changes and ECG signal changes

  7. Only two available ECG leads

  8. Atrial premature complexes (APC)

Contributor Information

Manel Ben-Romdhane, Email: manel.benromdhane@supcom.tn.

Asma Maalej, Email: asma.maalej@supcom.tn.

Mariam Tlili, Email: maryem.tlili@supcom.tn.

Chiheb Rebai, Email: chiheb.rebai@supcom.tn.

François Rivet, Email: francois.rivet@ims-bordeaux.fr.

Dominique Dallet, Email: dominique.dallet@ims-bordeaux.fr.

References

  • 1.Majumder S, Chen L, Marinov O, Chen C-H, Mondal T, Deen MJ. Non-contact wearable wireless ECG systems for long term monitoring. IEEE Rev. Biomed. Eng. 2018;11(5):306–321. doi: 10.1109/RBME.2018.2840336. [DOI] [PubMed] [Google Scholar]
  • 2.Lim CLP, Woo WL, Dlay SS, Gao B. Heartrate-dependent heartwave biometric identification with thresholding-based GMM-HMM methodology. IEEE Trans. Industr. Inf. 2019;15(1):45–53. [Google Scholar]
  • 3.Lim CLP, Woo WL, Dlay SS, Wu D, Gao B. Deep multi-view heartwave authentication. IEEE Trans. Industr. Inf. 2019;15(2):777–786. [Google Scholar]
  • 4.Benjamin EJ, Muntner P, Alonso A, Bittencourt MS, Callaway CW, Carson AP, Chamberlain AM, Chang AR, Cheng S, Das SR, Delling FN, Djousse L, Elkind JFF, Fornage M, Jordan LC, Khan SS, Kissela BM, Knutson KL, Kwan TW, Lackland DT, Lewis TT, Lichtman JH, Longenecker CT, Loop MS, Lutsey PL, Martin SS, Matsushita K, Moran AE, Mussolino ME, O’Flaherty M, Pandey A, Perak AM, Rosamond WD, Roth GA, Sampson UKA, Satou GM, Schroeder EB, Shah SH, Spartano NL, Stokes S, Tirschwell DL, Tsao CW, Turakhia MP, VanWagner LB, Wilkins JT, Wong SS, Virani SS. AHA statistical update, Heart disease and stroke statistics—2019 update: a report from the American Heart Association. Circulation. 2019;139(10):56–528. doi: 10.1161/CIR.0000000000000659. [DOI] [PubMed] [Google Scholar]
  • 5.Habte TT, Saleh H, Mohammad B, Ismail M. Ultra Low Power ECG Processing System for IoT Devices. Basel: Springer; 2019. [Google Scholar]
  • 6.Rana KPS, Kumar V, Singhal A, Chandel A, Pahuja D, Vashisht A. Time-varying pole-radius IIR multi-notch filters with improved performance. Arab. J. Sci. Eng. 2019;44(8):7101–7120. [Google Scholar]
  • 7.QARDIOCORE the world’s first wearable ECG free from patches and wires (CARDIO). http://www.getqardio.com/qardiocore-wearable-ecg-ekg-monitor-iphone. Accessed 30 April 2019 (2019)
  • 8.Apple Watch Series 4 proactive health monitor (Apple). http://www.apple.com/uk/apple-watch-series-4/health. Accessed 30 April 2019 (2019)
  • 9.KardiaMobile personal EKG (AlivCor), http://store.alivecor.com/products/kardiamobile. Accessed 30 April 2019 (2019)
  • 10.Mittal N, Singh U, Singh Sohi B. An energy-aware cluster-based stable protocol for wireless sensor networks. Neural Comput. Appl. 2019;31(11):7269–7286. [Google Scholar]
  • 11.Dong J, Jiang H, Yang K, Weng Z, Li F, Wei J, Ning Y, Chen X, Wang Z. A wireless body sound sensor with a dedicated compact chipset. Circuits Syst. Signal Proc. 2017;36(6):2341–2359. [Google Scholar]
  • 12.Wang F, Ma Q, Liu W, Chang S, Wang H, He J, Huang Q. A novel ECG signal compression method using spindle convolutional auto-encoder. Comput. Methods Programs Biomed. 2019;175:139–150. doi: 10.1016/j.cmpb.2019.03.019. [DOI] [PubMed] [Google Scholar]
  • 13.Yazicioglu RF, Kim S, Torfs T, Kim H, Hoof CV. A 30 µW analog signal processor ASIC for portable biopotential signal monitoring. IEEE J. Solid-State Circuits. 2011;460(1):209–223. [Google Scholar]
  • 14.Chen S-L, Villaverde JF, Lee H-Y, Chung DW-Y, Lin T-L, Tseng C-H, Lo K-A. A power-efficient mixed-signal smart ADC design with adaptive resolution and variable sampling rate for low-power applications. IEEE Sens. J. 2017;17(11):3461–3469. [Google Scholar]
  • 15.Gonzalez R, Perez R, Lopez M, Fernandez I, Espinosa J, Badias L, Fernandez A, Pena Y, Rodriguez G. A new tool for heart disease prognosis in the community. Comput. Cardiol. 2011;38:773–776. [Google Scholar]
  • 16.Rijnbeek PR, Kors JA, Witsenburg M. Minimum bandwidth requirements for recording of pediatric electrocardiograms. Circulation. 2001;104(25):3087–3090. doi: 10.1161/hc5001.101063. [DOI] [PubMed] [Google Scholar]
  • 17.Murmann, B.: ADC Performance Survey 1997–2019. http://web.stanford.edu/~murmann/adcsurvey.html. Accessed 11 Feb 2020
  • 18.Rodriguez-Vazquez A, Medeiro F, Janssens E. CMOS Telecom Data Converters. New York: Springer; 2003. [Google Scholar]
  • 19.Kaur A, Agarwal A, Agarwal R, Kumar S. A novel approach to ECG R-peak detection. Arab. J. Sci. Eng. 2018;44(8):6679–6691. [Google Scholar]
  • 20.Mansano AL, Li Y, Bagga S, Serdijn WA. An autonomous wireless sensor node with asynchronous ECG monitoring in 0.18 μm CMOS. IEEE Trans. Biomed. Circuits Syst. 2016;10(3):602–611. doi: 10.1109/TBCAS.2015.2495272. [DOI] [PubMed] [Google Scholar]
  • 21.Moody GB, Mark RG, Goldberger AL. PhysioNet: a web-based resource for the study of physiologic signals. IEEE Eng. Med. Biol. Mag. 2001;20(3):70–75. doi: 10.1109/51.932728. [DOI] [PubMed] [Google Scholar]
  • 22.Khan MG. Rapid ECG Interpretation. New York: Springer; 2008. [Google Scholar]
  • 23.Wang T-Y, Li H-Y, Ma Z-Y, Huang Y-J, Peng S-Y. A bypass-switching SAR ADC with a dynamic proximity comparator for biomedical applications. IEEE J. Solid State Circuits. 2018;53(6):1743–1754. [Google Scholar]
  • 24.Rahiminejad E, Saberi M, Lotfi R. A power-efficient signal-specific ADC for sensor-interface applications. IEEE Trans. Circuits Syst. II Express Briefs. 2017;64(9):1032–1036. [Google Scholar]
  • 25.Yan L, Harpe P, Pamula VR, Osawa M, Harada Y, Tamiya K, Van Hoof C, Yazicioglu RF. A 680 nA ECG acquisition IC for leadless pacemaker applications. IEEE Trans. Biomed. Circuits Syst. 2014;8(6):779–786. doi: 10.1109/TBCAS.2014.2377073. [DOI] [PubMed] [Google Scholar]
  • 26.Sharma A, Polley A, Lee SB, Narayanan S, Li W, Sculley T, Ramaswamy S. A Sub-60-μA multimodal smart biosensing SoC with > 80-dB SNR, 35μA photoplethysmography signal chain. IEEE J. Solid State Circuits. 2017;52(4):1021–1033. [Google Scholar]
  • 27.Hirai Y, Matsuoka T, Tani S, Isami S, Tatsumi K, Ueda M, Kamata T. A biomedical sensor system with stochastic A/D conversion and error correction by machine learning. IEEE Access. 2019;7:21990–22001. [Google Scholar]
  • 28.Zhang Z, Li J, Zhang Q, Wu K, Ning N, Yu Q. A dynamic tracking algorithm based SAR ADC in bio-related applications. IEEE Access. 2018;6:62166–62173. [Google Scholar]
  • 29.Adimulam MK, Srinivas MB. A 1.0 V, 9.84 fJ/c-s FOM reconfigurable hybrid SAR-sigma delta ADC for signal processing applications. Analog Integr. Circ. Sig. Process. 2019;99(2):261–276. [Google Scholar]
  • 30.Chen C, Chen L, Wang X, Zhang F. A 0.6 V, 8.4 μW AFE circuit for biomedical signal recording. Microelectron. J. 2018;75:105–112. [Google Scholar]
  • 31.Mark JW, Todd TD. A nonuniform sampling approach to data compression. IEEE Trans. Commun. 1981;29(1):24–32. [Google Scholar]
  • 32.Kozmin K, Johansson J, Delsing J. Level-crossing ADC performance evaluation toward ultrasound application. IEEE Trans. Circuits Syst. I Regul. Pap. 2009;56(8):1708–1719. [Google Scholar]
  • 33.Li Y, Zhao D, Serdijn W. A sub-microwatt asynchronous level-crossing ADC for biomedical applications. IEEE Trans. Biomed. Circuits Syst. 2013;7(2):149–157. doi: 10.1109/TBCAS.2013.2254484. [DOI] [PubMed] [Google Scholar]
  • 34.Allier E, Sicard G, Fesquet L, Renaudin M. Asynchronous level crossing analog to digital converters. Measurement. 2005;37(4):296–309. [Google Scholar]
  • 35.Weltin-Wu C, Tsividis Y. An event-driven clockless level-crossing ADC with signal-dependent adaptive resolution. IEEE J. Solid State Circuits. 2013;48(9):2180–2190. [Google Scholar]
  • 36.Tang W, Osman A, Kim D, Goldstein B, Huang C, Martini B, Pieribone VA, Culurciello E. Continuous time level crossing sampling ADC for bio-potential recording systems. IEEE Trans. Circuits Syst. I Regul. Pap. 2013;60(6):1407–1418. doi: 10.1109/TCSI.2012.2220464. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Rovere G, Fateh S, Benini L. A 2.2 μW cognitive always-on wake-up circuit for event-driven duty-cycling of IoT sensor nodes. IEEE J. Emerg. Sel. Top. Circuits Syst. 2018;8(3):543–554. [Google Scholar]
  • 38.Trakimas M, Sonkusale SR. An adaptive resolution asynchronous ADC architecture for data compression in energy constrained sensing applications. IEEE Trans. Circuits Syst. I Regul. Pap. 2011;58(5):921–934. [Google Scholar]
  • 39.Zhang X, Lian Y. A 300-mV 220-nW Event-driven ADC with real-time QRS detection for wearable ECG sensors. IEEE Trans. Biomed. Circuits Syst. 2014;8(6):834–843. doi: 10.1109/TBCAS.2013.2296942. [DOI] [PubMed] [Google Scholar]
  • 40.Hou Y, Qu J, Tian Z, Atef M, Yousef K, Lian Y, Wang G. A 61-nW level-crossing ADC with adaptive sampling for biomedical applications. IEEE Trans. Circuits Syst. II Express Briefs. 2019;66(1):56–60. [Google Scholar]
  • 41.Cox JR, Nolle FM, Fozzard HA, Oliver GC. AZTEC, a preprocessing program for real-time ECG rhythm analysis. IEEE Trans. Biomed. Eng. 1968;15(2):128–129. doi: 10.1109/tbme.1968.4502549. [DOI] [PubMed] [Google Scholar]
  • 42.Deepu CJ, Heng CH, Lian Y. A hybrid data compression scheme for power reduction in wireless sensors for IoT. IEEE Trans. Biomed. Circuits Syst. 2017;11(2):245–254. doi: 10.1109/TBCAS.2016.2591923. [DOI] [PubMed] [Google Scholar]
  • 43.Zigel Y, Cohen A, Katz A. The weighted diagnostic distortion (WDD) measure for ECG signal compression. IEEE Trans. Biomed. Eng. 2000;47(11):1422–1430. doi: 10.1109/TBME.2000.880093. [DOI] [PubMed] [Google Scholar]
  • 44.Mamaghanian H, Khaled N, Atienza D, Vandergheynst P. Compressed sensing for real-time energy-efficient ECG compression on wireless body sensor nodes. IEEE Trans. Biomed. Eng. 2011;58(9):2456–2466. doi: 10.1109/TBME.2011.2156795. [DOI] [PubMed] [Google Scholar]
  • 45.Jalaleddine S, Hutchens CG, Strattan RD, Coberly W. ECG data compression techniques-a unified approach. IEEE Trans. Biomed. Eng. 1990;37(4):329–343. doi: 10.1109/10.52340. [DOI] [PubMed] [Google Scholar]
  • 46.Chua E, Fang WC. Mixed bio-signal lossless data compressor for portable brain-heart monitoring systems. IEEE Trans. Consum. Electron. 2011;57(1):267–273. [Google Scholar]
  • 47.Chen S-L, Lee H-Y, Chen C-A, Huang H-Y, Luo C-H. Wireless body sensor network with adaptive low-power design for biometrics and healthcare applications. IEEE Syst. J. 2009;3(4):398–409. [Google Scholar]
  • 48.Chen S-L, Wang J-G. VLSI implementation of low-power cost-efficient lossless ECG encoder design for wireless healthcare monitoring application. Electron. Lett. 2013;49(2):91–93. [Google Scholar]
  • 49.Deepu CJ, Lian Y. A joint QRS detection and data compression scheme for wearable sensors. IEEE Trans. Biomed. Eng. 2015;62(1):165–175. doi: 10.1109/TBME.2014.2342879. [DOI] [PubMed] [Google Scholar]
  • 50.Chen S-L, Tuan M-C, Chi T-K, Lin T-L. VLSI architecture of lossless ECG compression design based on fuzzy decision and optimization method for wearable devices. Electron. Lett. 2015;51(18):1409–1411. [Google Scholar]
  • 51.Chen S-L, Tuan M-C, Lee H-Y, Lin T-L. VLSI implementation of a cost-efficient micro control unit with an asymmetric encryption for wireless body sensor networks. IEEE Access. 2017;5:4077–4086. [Google Scholar]
  • 52.Polania, L.F.; Carrillo, R.E.; Blanco-Velasco, M.; Barner, K. E.: Compressed sensing based method for ECG compression. In: Proceedings of IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP), pp. 761–764 (2011)
  • 53.Ieong C-I, Li M, Law M-K, Mak P-I, Vai MI, Martins RP. A 0.45 V 147–375 nW ECG compression processor with wavelet shrinkage and adaptive temporal decimation architectures. IEEE Trans. VLSI Syst. 2017;25(4):1307–1319. [Google Scholar]
  • 54.Kumar R, Kumar A, Pandey RK. Beta wavelet based ECG signal compression using lossless encoding with modified thresholding. Comput. Electr. Eng. 2013;39(1):130–140. [Google Scholar]
  • 55.Elgendi M, Mohamed A, Ward R. Efficient ECG compression and QRS detection for e-Health applications. Sci. Rep. 2017;7(1):1–16. doi: 10.1038/s41598-017-00540-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Zhou J, Wang C. An ultra-low power turning angle based biomedical signal compression engine with adaptive threshold tuning. Sensors. 2017;17(8):1–12. doi: 10.3390/s17081809. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Abhishek S, Veni S, Narayanankutty KA. Biorthogonal wavelet filters for compressed sensing ECG reconstruction. Biomed. Signal Process. Control. 2019;47:183–195. [Google Scholar]
  • 58.Abo-Zahhad MM, Abdel-Hamid TK, Mohamed AM. Compression of ECG signals based on DWT and exploiting the correlation between ECG signal samples. Int. J. Commun. Netw. Syst. Sci. 2014;7:53–70. [Google Scholar]
  • 59.Strang G, Nguyen T. Wavelets and Filter Banks. Wellesley: Wellesley-Cambridge Press; 1996. [Google Scholar]
  • 60.Daubechies I. Ten Lectures on Wavelets. Pennsylvania: SIAM; 1992. [Google Scholar]
  • 61.Misiti M, Misiti Y, Oppenheim G, Poggi JM. Wavelet Toolbox User’s Guide. Natick: The MathWorks Inc.; 2009. [Google Scholar]
  • 62.Padhy S, Sharma L, Dandapat S. Multilead ECG data compression using SVD in multiresolution domain. Biomed. Signal Process. Control. 2016;23:10–18. [Google Scholar]
  • 63.Tropp JA, Gilbert AC. Signal recovery from random measurements via orthogonal matching pursuit. IEEE Trans. Inf. Theory. 2007;53(12):4655–4666. [Google Scholar]
  • 64.Lu L, Yan J, de Silva CW. Feature selection for ECG signal processing using improved genetic algorithm and empirical mode decomposition. Measurement. 2016;94:372–381. [Google Scholar]
  • 65.Schell B, Tsividis Y. A continuous-time ADC/DSP/DAC system with no clock and with activity-dependent power dissipation. IEEE J. Solid State Circuits. 2008;43(11):2472–2481. [Google Scholar]
  • 66.Tlili M, Ben-Romdhane M, Maalej A, Rivet F, Dallet D, Rebai C. Level-crossing ADC design and evaluation methodology for normal and pathological electrocardiogram signals measurement. Measurement. 2018;124:413–425. [Google Scholar]
  • 67.Jaw F-S, Tseng Y-L, Jang J-K. Modular design of a long-term portable recorder for physiological signals. Measurement. 2010;43(10):1363–1368. [Google Scholar]
  • 68.Bera P, Gupta R. Hybrid encoding algorithm for real time compressed electrocardiogram acquisition. Measurement. 2016;91:651–660. [Google Scholar]

Articles from Arabian Journal for Science and Engineering are provided here courtesy of Nature Publishing Group