Highlights
-
•
The paper “Intra-Aortic Balloon Counterpulsation Timing: A New Numerical Model for Programming and Training in the Clinical Environment.”, is original research from our ongoing program on numerical simulation applied to the clinical environment.
-
•
In this work we present a new numerical model of the intraaortic balloon pump (IABP). The new model also allows us to simulate patient weaning from IABP assistance.
-
•
Furthermore, the cardiovascular network model in which the IABP is inserted, allows us to simulate also a mechanical ventilatory assistance that is very important in patients affected by COVID-19.
Keywords: Heart failure, Pressure volume loop, Software simulation, IABP, Training, Clinical environment
Abstract
Background and Objective
The intra-aortic balloon pump (IABP) is the most widely available device for short-term mechanical circulatory support, often used to wean off cardiopulmonary bypass or combined with extra-corporeal membrane oxygenation support or as a bridge to a left ventricular assist device. Although based on a relatively simple principle, its complex interaction with the cardiovascular system remains challenging and open to debate. The aim of this work was focused on the development of a new numerical model of IABP.
Methods
The new model was implemented in CARDIOSIM©, which is a modular software simulator of the cardiovascular system used in research and e-learning environment. The IABP is inserted into the systemic bed divided in aortic, thoracic and two abdominal tracts modelled with resistances, inertances and compliances. The effect induced by the balloon is reproduced in each tract of the aorta by the presence of compliances connected to PIABP generator and resistances. PIABP generator reproduces the balloon pressure with the option to change IABP timing. We have used literature data to validate the potential of this new numerical model.
Results
The results have shown that our simulations reproduced the typical effects induced during IABP assistance. We have also simulated the effects induced by the device on the hemodynamic variables when the IABP ratio was set to 1:1, 1:2, 1:4 and 1:8. The outcome of these simulations is in accordance with literature data measured in the clinical environment.
Conclusions
The new IABP module is easy to manage and can be used as a training tool in a clinical setting. Although based on literature data, the outcome of the simulations is encouraging. Additional work is ongoing with a view to further validate its features. The configuration of CARDIOSIM© presented in this work allows the simulation of the effects induced by mechanical ventilatory assistance. This facility may have significant importance in the management of patients affected by COVID-19 when they require mechanical circulatory support devices.
NOMENCLATURE
- ABP (Pas)
Aortic blood pressure [mmHg]
- EDP
End diastolic blood pressure [mmHg]
- PDP
Peak diastolic blood pressure [mmHg]
- LVP (Plv)
Left ventricular pressure [mmHg]
- BSA
Body surface area [m2]
- CI
Cardiac index [L/min/m2]
- CO
Cardiac output [L/min]
- EDV (Ved)
End-diastolic volume [ml]
- ESV (Ves)
End-systolic volume [ml]
- SV
Stroke volume [ml]
- Ea
Arterial elastance [mmHg/ml]
- Ees
Slope of ESPVR [mmHg/ml]
- EDPVR (ESPVR)
End-diastolic (end-systolic) pressure volume relationship
- EF%
Ejection fraction
- MCBF
Mean coronary blood flow [ml/min]
- HR
Heart rate [bpm]
- RAP
Right atrial pressure [mmHg]
- PAP
Pulmonary arterial pressure [mmHg]
- PCWP
Pulmonary capillary wedge pressure [mmHg]
- Ped (Pes)
End-diastolic (end-systolic) ventricular pressure [mmHg]
- PVR
Pulmonary vascular resistance [mmHg · s/ml]
- LVSW
Left ventricular stroke work [g · m−1]
- RVSW
Right ventricular stroke work [g · m−1]
- LVSWI
Left ventricular stroke work index [g · m · m−2]
- RVSWI
Right ventricular stroke work index [g · m · m−2]
- IABP
Intra-aortic balloon pump
- MVA
Mechanical ventilator assistance
1. Introduction
The intra-aortic balloon pump (IABP) is a widely available in-series cardiac assist device. It consists of a double-lumen catheter with a polyurethane balloon attached at its distal end and a mobile pump console, which shuttles helium through the main lumen of the catheter. The tip of the catheter has a pressure sensor to monitor aortic blood pressure. The IABP is now inserted percutaneously through the femoral artery and positioned just below the origin of the left subclavian artery either under fluoroscopic guidance in the cath lab or under trans-oesophageal guidance in theatre.
The IABP is based on the principle of counterpulsation, which aims to optimize the balance between myocardial oxygen supply and demand in terms of endocardial viability ratio (EVR) [1],[2].
The functional relationship between stroke volume (SV), aortic mean diastolic pressure (MDP), tension time index (TTI), aortic end diastolic pressure (EDP), balloon inflation/deflation timing and heart rate (HR) is a key element for optimal pump control [1],[2].
The physiological advantage of intra-aortic balloon counterpulsation is increased aortic diastolic blood pressure. Rapid inflation of the balloon at the beginning of diastole generates proximal and distal blood displacement, which is proportional to the volume of the balloon. Diastolic blood pressure augmentation increases the intrinsic windkessel effect leading to storage of extra potential energy in the aorta and conversion to kinetic energy following the elastic recoil of the vessel [3]. This event has the potential to increase coronary blood flow. Rapid deflation of the balloon in early systole leads to afterload reduction or, to be more precise, to reduction of impedance to ventricular ejection and cardiac work [4],[5]. This is in accordance with a previous analytical model [5].
The ability to model the interactions between IABP and the cardiovascular system and how alterations of specific parameters such as timing can affect their coupling remains a key element for clinical application [6],[7].
Simulations of combined VA-ECMO and IABP support show an increase in pulsatility and LV stroke volume between 5% and 10% due to afterload reduction although PCWP and left ventricular EDV are only marginally affected. Significant LV unloading is achieved during combined VA-ECMO and Impella support although aortic valve opening and improved diastolic coronary perfusion pressure are not observed in comparison with IABP [8]. Nevertheless, the pulse contour is higher and more similar to the physiological pattern during partial ECMO support where some degree of LV ejection is allowed [9].
The concomitant use of IABP and VA-ECMO shows reduced in-hospital mortality in patients with cardiogenic shock secondary to post-cardiotomy failure, ischaemic heart disease and myocarditis [10], [11], which is in contrast with the questionable outcome of the SHOCK II trial [12], [13], [14]. The study was designed as a multicentre, randomised, open-label trial. Between 2009 and 2012, 600 patients with cardiogenic shock following acute myocardial infarction and requiring early revascularisation were randomised to IABP versus control. Long-term follow-up (6.2 years) showed no difference in mortality, recurrent myocardial infarction, stroke, repeat revascularisation or hospital readmission for cardiac reasons between the two groups. Nevertheless, the use of IABP in cardiogenic shock remains the subject of significant debate and controversy.
2. Material And Method
2.1. The heart and circulatory numerical network
The electrical analogue of the cardiovascular system assembled inside the software simulator platform CARDIOSIM© is reported in Fig. 1 . The circuit consists of systemic and venous sections, coronary section and pulmonary arterial and venous sections. Each section is modelled by RLC electrical circuit based on 0-D numerical representation. The systemic venous section consists of a compliance (Cvs) and two variable resistances (Rvs1 and Rvs2); the pulmonary arterial section is modelled with a characteristic resistance (Rcp), a variable pulmonary arterial resistance (Rap), a compliance (Cap) and an inertance (Lap). Finally, the behavior of the pulmonary venous section is reproduced with a resistance (Rvp) and a compliance (Cvp). Pt is the mean intrathoracic pressure.
The following CARDIOSIM© module was selected to simulate the heart activity. The behavior of the left and right native ventricles is reproduced by the time-varying elastance model. The same theory is used to model both left and right atria and the septum [15], [16], [17], [18], [19]. Ventricles, atria and septum activities are synchronized with the electrocardiographic (ECG) signal [19]. The described model allows inter-ventricular and intra-ventricular dyssynchrony to be simulated [19], [20], [21].
2.2. IABP and systemic arterial numerical models
Figure 2 shows the electrical analogue of the systemic circulation and the IABP. The systemic bed is divided in aortic, thoracic and two abdominal tracts. When the IABP is “OFF” (SW1=ON and SW2=OFF), the aortic (thoracic) tract is modelled using resistance RAT (RTT), compliance CAT (CTT) and inertance LAT (LTT). The behavior of the first abdominal tract is reproduced by RLC elements (RABT1, LABT1 and CABT1) when the IABP is disabled. The second abdominal tract is modelled by RABT2, LABT2, CABT2 and by variable systemic arterial resistance (Ras). The switches SW1 and SW2 are set to OFF (ON) and ON (OFF) respectively, when the device is (not) working. The effect induced by the balloon is reproduced in each tract of the aorta by the presence of compliances (CIABP1, CIABP2 andCIABP3) connected to PIABP generator and resistances (RIABP1, RIABP2 and RIABP3). PIABP generator reproduces the balloon pressure with the option to change the IABP timing. The resistances (compliances) RIABP1 (CIABP1), RIABP2 (CIABP2) and RIABP3 (CIABP3) are connected in series (parallel) to RAT (CAT), RTT (CTT) and RABT1 (CABT1) in the aortic, thoracic and first abdominal tract. Figure 3 shows the waveform of the pressure produced by the generator that is defined as follows:
When the IABP is OFF, the network showed in Fig. 2 is solved by the equations:
If the IABP is ON, the equations are:
where Plv is the left ventricular pressure. The resistance Rlo and the diode D1 model the mitral valve (Fig. 2). The software simulator allows the IABP to be synchronized with either the ECG or the aortic pressure waveform. The frequency of balloon-assisted beats can be set from the maintenance 1:1 ratio to a weaning 1:2 ratio (every other systole is assisted). Depending on the clinician's judgment, weaning modes of 1:4 or even 1:8 may be initiated if a more gradual approach is needed. In addition, IABP driving and vacuum pressures can be changed.
The hemodynamic effects induced by the IABP may vary with assisting frequency and depend on balloon inflation/deflation timing. A range of settings T1-T7 is available in the software simulator. The IABP can be triggered to deflate during systole once the peak of the R wave is sensed. IABP inflation may be triggered to occur in the middle of the T wave, which corresponds to diastole. The simulator allows the setting of different delays. Changing IABP compliance and resistance allows the balloon volume to be modified.
2.3. Cardiogenic Shock Patients
Patients with acute myocardial infarction (AMI) and cardiogenic shock (CS) may require intra-aortic balloon counterpulsation as an adjunct to medical treatment [27]. For the purposes of our study, literature data [28], [29], [30] were used to reproduce the baseline conditions of CS patients and those following IABP assistance. The hemodynamic data used in this study have been listed in Table 1 .
Table 1.
Baseline conditions | IABP on [1:1] | |
---|---|---|
PDP [mmHg] | 67±7 | 75±7 |
EDP [mmHg] | 55±6 | 42±9 |
HR [beat/min] | 99±27 | 100±22 |
LV EDV [cc] | 293±35 | 285±36 |
LV ESV [cc] | 263±32 | 259±36 |
SV [cc] | 29±4 | 26±4 |
PAP [mmHg] | 42±11 | 32±6 |
PCWP [mmHg] | 27±8 | 23±6 |
RA [mmHg] | 10±5 | 10±5 |
CI [l·min−1/m2] | 1.56±0.29 | 1.57±0.3 |
SVR [Wood] | 21.67±5.2 | 16.42±11.8 |
PVR [Wood] | 4.95±2.7 | 1.46±1.1 |
BSA [m2] | 1.8±0.13 | 1.8±0.13 |
CO [l · min−1] | 2.8±1.25 | 2.91±1.25 |
2.4. Simulation Protocol
The duration of the whole cardiac cycle was set at 1000 ms for all the simulations. Starting from the reproduced baseline conditions, the IABP was activated with a driving (vacuum) pressure of PDRIVE = 240 mmHg (PVACUUM= -10 mmHg). The plateau IABP pressure was set to PPLATEAU = 150 mmHg. During the simulations, the IABP was synchronized with the ECG and its ratio was set to 1:1, 1:2, 1:4 and 1:8. During baseline conditions and when the IABP ratio was set to 1:1, the mean values for pressure, flow, EDV and ESV (for both ventricles) were calculated for one cardiac cycle. When the IABP ratio was set to 1:2 (1:4 or 1:8), the mean values for pressure, flow, EDV and ESV (for both ventricles) were calculated for two (four or eight) cardiac cycles.
3. Results
Figure 4a shows a screen output produced by CARDIOSIM© when patient's baseline conditions were reproduced using the data reported in Table 1. The left (right) ventricular loop has been plotted in the upper (lower) window. The first column on the left hand side is the command of the software. The middle column shows the mean value of pressures and flows. End systolic volume (ESV Ves), end diastolic volume (EDV Ved), stroke volume and ejection fraction are listed below the middle column. Finally, the last windows on the right hand side show the cardiac and ventricular work index with the left and right ventricular energetic variables.
Figure 4b shows the data obtained during a simulation, which are stored in excel files and subsequently processed.
Table 2 shows the simulation results obtained when the IABP ratio was set to 1:1; 1:2; 1:4 and 1:8. When the IABP ratio was set to 1:2, the mean values of each parameter were calculated during two cardiac cycles (one with IABP set to “ON” and the other with IABP set to “OFF”). When the IABP ratio was set to 1:4 (1:8), the mean values were calculated during four (eight) cardiac cycles i.e. one cardiac cycle with IABP set to “ON” and three (seven) cardiac cycles with IABP set to “OFF”. The results obtained when the IABP ratio was set to 1:1 are comparable with the literature data reported in Table 1. The simulation results presented when the IABP ratio was set to 1:2, 1:4 and 1:8 were obtained after changing the IABP ratio from baseline conditions.
Table 2.
Baseline conditions | IABP on [1:1] | IABP on [1:2] | IABP on [1:4] | IABP on [1:8] | |
---|---|---|---|---|---|
PDP [mmHg] | 70.8 | 91 | 83.7±4.3 | 74.2±12.4 | 70.44±12.25 |
EDP [mmHg] | 57.2 | 41.5 | 40.5±8.5 | 46.18±9.15 | 49.84±9.6 |
HR [beat/min] | 99 | 99 | 99 | 99 | 99 |
LV EDV [cc] | 293.35 | 267.92 | 262.3±0.3 | 270±2.4 | 276.06±4.05 |
LV ESV [cc] | 264.9 | 231.92 | 224.8±8.4 | 236±9.25 | 244.77±10.43 |
RV EDV [cc] | 128.37 | 139.98 | 138±0.6 | 133.46±1.4 | 129.44±2.11 |
RV ESV [cc] | 99.92 | 103.98 | 102±0.3 | 100.0±0.32 | 98.48±0.49 |
SV [cc] | 28.4 | 35.9 | 36.9±3.9 | 33.7±5.35 | 31.13±6.15 |
PAP [mmHg] | 40.3 | 41 | 40.3±0.05 | 39.8±0.1 | 39.45±0.15 |
PCWP [mmHg] | 27.2 | 24.8 | 23.95±0.2 | 24.75±0.3 | 25.3±0.35 |
RA [mmHg] | 8.6 | 9.8 | 9.5±0.1 | 9.0±0.2 | 8.69±0.2 |
CI [l · min−1/m2] | 1.56 | 1.98 | 1.84±0.2 | 1.68±0.3 | 1.56±0.31 |
SVR [Wood] | 18.87 | 15.33 | 14.66±4.7 | 15.48±3.93 | 16.47±3.96 |
PVR [Wood] | 4.64 | 4.64 | 5.11±1.58 | 4.93±1.26 | 4.8±1.2 |
BSA [m2] | 1.8 | 1.8 | 1.8 | 1.8 | 1.8 |
CO [l · min−1] | 2.82 | 3.56 | 3.65±0.4 | 3.34±0.53 | 3.08±0.61 |
Left Ea | 2.49 | 1.71 | 1.66±0.32 | 1.93±0.43 | 2.14±0.51 |
Left Ea/Ees | 8.31 | 5.68 | 5.52±1.05 | 6.42±1.43 | 7.13±1.67 |
Right Ea | 1.51 | 1.22 | 1.18±0.03 | 1.26±0.05 | 1.34±0.07 |
Right Ees/Ea | 0.32 | 0.39 | 0.41±0.01 | 0.38±0.02 | 0.36±0.01 |
LVSW [g · m−1] | 13.4 | 19.3 | 16.25±3.15 | 14.63±4.2 | 13.64±4.6 |
RVSW [g · m−1] | 12.2 | 15.4 | 15.2±0.4 | 14.0±0.45 | 13.06±0.6 |
LVSWI [g · m · m−2] | 7.43 | 10.7 | 9.05±1.75 | 8.13±2.36 | 7.58±2.58 |
RVSWI [g · m · m−2] | 6.8 | 8.55 | 8.72±1.97 | 7.9±2.15 | 7.28±2.37 |
PVC [ml · mmHg−1] | 7.6 | 9.94 | 9.97±0.04 | 9.84±0.04 | 9.74±0.04 |
MCBF [ml/min] | 58.3 | 70.5 | 55.3±4.8 | 53.4±4.4 | 53.6±3.8 |
Left EF% | 9.7 | 13.4 | 14.3±3.3 | 12.58±3.5 | 11.38±3.8 |
Right EF% | 22.2 | 25.7 | 26.25±0.45 | 25.05±0.6 | 24.05±0.85 |
Figure 5a shows a screen output produced by CARDIOSIM© during IABP assistance.
The middle window [A] shows the left ventricular loops following different assisted conditions. The two red loops are obtained when the IABP ratio is set to 1:2; in this case, two different values are obtained for the systemic arterial elastance Ea1 (green lines): the first when the IABP is ON and the second when the IABP is OFF. The elastances have the same slope. When the IABP is active, the simulator reproduces the largest left ventricular loop. When the IABP ratio is set to 1:4, the software simulator plots the blue left ventricular loops. The wider loop corresponds to the cardiac cycle in which IABP is ON. In this case, the systemic arterial elastance Ea2 (lilac lines) assumes four different values. The left ventricular elastance (Ees) does not change during circulatory assistance, so the variation in coupling (LEFT Ea/Ees) depends only on Ea.
Window [B] ([C]) shows the instantaneous left ventricular pressure (LVP Plv, green line) and aortic blood pressure (ABP Pas, red line) when the IABP ratio is set to 1:4 (1:8).
Fig. 5b shows the right ventricular loops during IABP assistance with ratio 1:2 (red loops) and 1:4 (blue loops). The pulmonary arterial elastance (Ea1) assumed two values (green lines) when the ratio of the assistance was set to 1:2. The arterial elastance Ea2 assumed four values (lilac lines) when IABP ratio was set to 1:4. The right ventricular end systolic volume (ESV) presents minimal changes on 1:2 and 1:4 assistance. The right ventricular elastance (Ees) does not change during IABP assistance.
4. Discussion
The numerical model for IABP assistance implemented in CARDIOSIM© produces an increase between 10% (1:8 ratio) and 25% (1:1 ratio) in CO following activation. It is estimated that IABP assistance with 1:1 ratio in the presence of sinus rhythm can increase cardiac output up to 20-25% of its initial value according to the available literature data. We point out that in the transition from IABP setting with 1:1 ratio to that with 1:8 ratio, the parameters of the simulator were not changed. Systemic vascular resistance (SVR) decreases from 23% to 18% when IABP is activated. The assistance induces about 10% reduction in EDV (when the ratio is 1:1) and about 5% reduction in the mean value of EDV (calculated on eight cardiac cycles) when the IABP ratio is set to 1:8. There is about 13% (7%) decrease in ESV when IABP is set to 1:1 (1:8). A 27% (13%) reduction in end diastolic blood pressure (EDP) is observed during IABP activation with 1:1 (1:8) ratio. The simulation results show a reduction in the left ventricular end-diastolic and end-systolic volumes with concomitant decrease in the left atrial filling and in the pulmonary capillary wedge pressures (PCWP) leading to reduced right ventricular afterload. These findings are completely in accordance with literature data. Besides, the PCWP reduction during the simulation can be up to 10%.
The peak diastolic blood pressure (PDP) increases by 30% when the IABP ratio is set to 1:1. The left ventricular arterial coupling (Ea/Eas) decreases by 32% (14%) when the assistance ratio is set to 1:1 (1:8). The mean coronary blood flow (MCBF) increases by 21% when the IABP ratio is set to 1:1 but it decreases by 8% when the ratio is set to 1:8. In this case, the MCBF is calculated over eight cardiac cycles where only one is assisted by IABP activation.
The limitations of this work are related to the use of incomplete data from the available literature, which affects the accurate reproduction of a patient's cardiovascular conditions. Furthermore, we were unable to obtain data related to the trend of the hemodynamic variables during different IABP operating modes. We believe these data would be useful to further validate our model. Nevertheless, the cardiovascular network used in this work has been previously validated [15],[16] and has led to the successful evaluation of the trend of the hemodynamic variables during patient weaning off the device.
We have also observed that the cardiovascular network outlined in Fig. 1 allows changes in the mean value of intrathoracic pressure to be made. This option may be used to evaluate the effects induced by mechanical ventilatory assistance (MVA) during IABP support. Simulating the effects induced by MVA may have significant importance in the management of patients affected by COVID-19 when they require mechanical circulatory support devices.
The recent events leading to the COVID-19 pandemic have confirmed how unprepared we are despite our level of knowledge and technology. Although a certain degree of uncertainty remains, the ability to model and simulate critical clinical conditions in advance with a view to treatment optimization and outcome prediction may be an option to consider on a more routine basis. The severe respiratory impairment caused by COVID-19 requiring mechanical ventilation and ECMO support has generated severe strain on intensive care units capacity and hospital resources. Redeployment of medical and nursing staff in different roles has required crash training for those not accustomed to a certain type of work. The ability to study and train through modelling and simulation may help with organization and rearrangement of working patterns. This may become part of the training of healthcare professionals with a view to focus on specific subjects with an open mind where the clinician remains the ultimate decision-maker.
5. Conclusion
We have developed an innovative IABP numerical model with potential for clinical application. The focus has been on IABP timing and the evaluation of the most appropriate weaning strategy in terms of device assist ratio. Its value as a training tool in a clinical setting has been proposed. Although based on literature data, the outcome of the simulations is encouraging. Additional work is ongoing with a view to further validate its features.
6. Funding And Conflict Of Interest
None
Declaration of Competing Interest
All authors have no conflicts of interest
Acknowledgements
None
Contributor Information
Claudio De Lazzari, Email: claudio.delazzari@ifc.cnr.it.
Beatrice De Lazzari, Email: beatrice.delazzari@gmail.com.
Attilio Iacovoni, Email: aiacovoni@asst-pg23.it.
Silvia Marconi, Email: silvia.marconi@ifc.cnr.it.
Silvia Papa, Email: silvia.papa@uniroma1.it.
Massimo Capoccia, Email: capoccia@doctors.org.uk.
Roberto Badagliacca, Email: roberto.badagliacca@uniroma1.it.
Carmine Dario Vizza, Email: dario.vizza@uniroma1.it.
References
- 1.Jaron D., Moore T., He P., S. Mathematical Modelling of Optimal Intraaortic Balloon Pumping. Progr. Artif. Org. 1983:923–929. [Google Scholar]
- 2.Jaron D., Moore T.W., He P. Control of Intraaortic Balloon Pumping: Theory and Guidelines for Clinical Applications. Ann. Biomed. Eng. 1985;13:155–175. doi: 10.1007/BF02584236. [DOI] [PubMed] [Google Scholar]
- 3.Krishna M., Zacharowski K. Principles of intra-aortic balloon pump counterpulsation, Continuing Education in Anaesthesia. Critical Care & Pain. 2009;9(1):24–28. [Google Scholar]
- 4.Weber K.T., Janicki J.S. Intraaortic Balloon Counterpulsation: A Review of Physiological Principles, Clinical Results and Device Safety”. Ann. Thorac. Surg. 1974;17:602–636. doi: 10.1016/s0003-4975(10)65706-2. [DOI] [PubMed] [Google Scholar]
- 5.Murthy V.S., McMahon T.A., Jaffrin YM.Y., Shapiro A.H. The Intra-Aortic Balloon for Left Heart Assistance: An Analytic Model. J. Biomech. 1971;4:351–367. doi: 10.1016/0021-9290(71)90056-x. [DOI] [PubMed] [Google Scholar]
- 6.Schampaert S., Rutten M.C.M., van't Veer M., van Nunen L.X., Tonino P.A.L., Pijls N.H.J., van de Vosse F.N. Modelling the Interaction Between the Intra-Aortic Balloon Pump and the Cardiovascular System: The Effect of Timing. ASAIO J. 2013;59:30–36. doi: 10.1097/MAT.0b013e3182768ba9. [DOI] [PubMed] [Google Scholar]
- 7.Kolyva C., Pantalos G.M., Pepper J.R., Khir A.W. Does conventional intra-aortic balloon pump trigger timing produce optimal haemodynamic effects in vivo? Int. J. Artif. Organs. 2015;38(3):146–153. doi: 10.5301/ijao.5000385. [DOI] [PubMed] [Google Scholar]
- 8.Donker D.W., Brodie D., Henriques J.P.S., Broomé M. Left Ventricular Unloading During Veno-Arterial ECMO: A Simulation Study. ASAIO J. 2019;65:11–20. doi: 10.1097/MAT.0000000000000755. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Caruso M.V., Gramigna V., Renzulli A., Fragomeni G. Computational analysis of aortic hemodynamics during total and partial extracorporeal membrane oxygenation and intra-aortic balloon pump support”. Acta Bioeng. Biomech. 2016;18(3):3–9. [PubMed] [Google Scholar]
- 10.Li Y., Yan S., Gao S., Liu M., Lou S., Liu G., Ji B., Gao B. Effect of an intra-aortic balloon pump with venoarterial extracorporeal membrane oxygenation on mortality of patients with cardiogenic shock: a systematic review and meta-analysis. Eur. J. Cardio-Thorac. Surg. 2019;55:395–404. doi: 10.1093/ejcts/ezy304. [DOI] [PubMed] [Google Scholar]
- 11.Chen K., Hou J., Tang H., Hu S. Concurrent initiation of intra-aortic balloon pumping with extracorporeal membrane oxygenation reduced in-hospital mortality in postcardiotomy cardiogenic shock. Ann. Intensive Care. 2019;9:16. doi: 10.1186/s13613-019-0496-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Thiele H., Schuler G., Neumann F.J., Hausleiter J., Olbrich H.G., Schwarz B., Hennersdorf M., Empen K., Fuernau G., et al. Intraaortic balloon counterpulsation in acute myocardial infarction complicated by cardiogenic shock: design and rationale of the Intraaortic Balloon Pump in Cardiogenic Shock II (IABP-SHOCK II) trial. Am. Heart J. 2012;163:938–945. doi: 10.1016/j.ahj.2012.03.012. [DOI] [PubMed] [Google Scholar]
- 13.Thiele H., Zeymer U., Neumann F.J., Ferenc M., Olbrich H.G., Hausleiter J., de Waha A., et al. Intraaortic Balloon Pump in cardiogenic shock II (IABP-SHOCK II) trial investigators. Intra-aortic balloon counterpulsation in acute myocardial infarction complicated by cardiogenic shock (IABP-SHOCK II): final 12 month results of a randomised, open-label trial. Lancet. 2013;382:1638–1645. doi: 10.1016/S0140-6736(13)61783-3. [DOI] [PubMed] [Google Scholar]
- 14.Thiele H., Zeymer U., Thelemann N., Neumann F.J., Hausleiter J., Abdel-Wahab M., Meyer-Saraei R., et al. IABPSHOCK II Trial (Intraaortic Balloon Pump in Cardiogenic Shock II) Investigators. Intraaortic Balloon Pump in Cardiogenic Shock Complicating Acute Myocardial Infarction: Long-Term 6-Year Outcome of the Randomized IABP-SHOCK II Trial. Circulation. 2019;139:395–403. doi: 10.1161/CIRCULATIONAHA.118.038201. [DOI] [PubMed] [Google Scholar]
- 15.Capoccia M., Marconi S., Singh S.A., et al. Simulation as a preoperative planning approach in advanced heart failure patients. A retrospective clinical analysis. BioMedical Engineering OnLine. 2018;17(1):52. doi: 10.1186/s12938-018-0491-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Capoccia M., Marconi S., De Lazzari C. Decision-making in advanced heart failure patients requiring LVAD insertion: Can preoperative simulation become the way forward? A case study. Journal of Biomedical Engineering and Informatics. 2018;4(2):8–20. [Google Scholar]
- 17.Suga H., Sagawa K., Shoukas A.A. Load Independence of the Instantaneous Pressure-Volume Ratio of the Canine Left Ventricle and Effects of Epinephrine and Heart Rate on the Ratio. Circ. Res. 1973;32:314–322. doi: 10.1161/01.res.32.3.314. [DOI] [PubMed] [Google Scholar]
- 18.De Lazzari C., Genuini I., et al. Interactive simulator for e-Learning environments: a teaching software for health care professionals. BioMedical Engineering OnLine. 2014;13:172. doi: 10.1186/1475-925X-13-172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.De Lazzari C. Interaction between the septum and the left (right) ventricular free wall in order to evaluate the effects on coronary blood flow: numerical simulation. Comput. Methods. Biomech. Biomed. Eng. 2012;15(12):1359–1368. doi: 10.1080/10255842.2011.597354. [DOI] [PubMed] [Google Scholar]
- 20.Sagawa K., Maughan L., Suga H., Sunagawa K. Oxford University Press; New York: 1988. Cardiac contraction and the Pressure-Volume relationships. [Google Scholar]
- 21.Maughan W.L., Sunagawa K., Sagawa K. Ventricular systolic interdependence: volume elastance model in isolated canine hearts. Am. J. Physiol. Heart Circ. Physiol. 1987;253:H1381–H1390. doi: 10.1152/ajpheart.1987.253.6.H1381. [DOI] [PubMed] [Google Scholar]
- 22.C. De Lazzari, D. Stalteri. 2011-2019, CARDIOSIM© Website. Original website platform regarding the implementation of the cardiovascular software simulator CARDIOSIM© (https://cardiosim.dsb.cnr.it/).
- 23.De Lazzari C., L'Abbate A., Micalizzi M., Trivella M.G., Neglia D. Effects of amlodipine and adenosine on coronary haemodynamics: in vivo study and numerical simulation. Computer Methods in Biomechanics and Biomedical Engineering. 2014;17(15):1642–1652. doi: 10.1080/10255842.2012.761691. [DOI] [PubMed] [Google Scholar]
- 24.De Lazzari C., D'Ambrosi A., Tufano F., Fresiello L., Garante M., Sergiacomi R., F., et al. Cardiac resynchronization therapy: could a numerical simulator be a useful tool in order to predict the response of the biventricular pacemaker synchronization. Eur. Rev. Med. Pharmacol. Sci. 2010;14(11):969–978. [PubMed] [Google Scholar]
- 25.De Lazzari C., Neglia D., Ferrari G., Bernini F., Micalizzi M., L'Abbate A., Trivella M.G. Computer simulation of coronary flow waveforms during caval occlusion. Meth.of Inf. in Med. 2009;48(2):113–122. doi: 10.3414/ME0539. [DOI] [PubMed] [Google Scholar]
- 26.De Lazzari C., Darowski M., Ferrari G., Clemente F. The influence of left ventricle assist device and ventilatory support on energy-related cardiovascular variables. Medical Eng. & Phy. 1998;20(2):83–91. doi: 10.1016/s1350-4533(98)00008-3. [DOI] [PubMed] [Google Scholar]
- 27.Le Bras A. No long-term benefit of IABP in cardiogenic shock. Nature Reviews Cardiology. 2019;16:3. doi: 10.1038/s41569-018-0136-3. [DOI] [PubMed] [Google Scholar]
- 28.den Uil C.A., Galli G., Jewbali S.L., Caliskan K., Manintveld O.C., et al. First-Line Support by Intra-Aortic Balloon Pump in Non-Ischaemic Cardiogenic Shock in the Era of Modern Ventricular Assist Devices. Cardiology. 2017;138:1–8. doi: 10.1159/000471846. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Imamura T., Kinugawa K., Nitta D., et al. Prophylactic Intra-Aortic Balloon Pump Before Ventricular Assist Device Implantation Reduces Perioperative Medical Expenses and Improves Postoperative Clinical Course in INTERMACS Profile 2 Patients. Circulation J. 2015;79(9):1963–1969. doi: 10.1253/circj.CJ-15-0122. [DOI] [PubMed] [Google Scholar]
- 30.Annamalai S.K., Buiten L., Esposito M.L., Paruchuri V., Mullin A., Breton C., Pedicini R., et al. Acute Hemodynamic Effects of Intra-aortic Balloon Counterpulsation Pumps in Advanced Heart Failure. Journal of Cardiac Failure. 2017;23(8):606–614. doi: 10.1016/j.cardfail.2017.05.015. [DOI] [PubMed] [Google Scholar]