Ventilatory Ratio in Hypercapnic Mechanically Ventilated Patients with COVID-19–associated Acute Respiratory Distress Syndrome

Xiaoqing Liu 1,2,3,*, Xuesong Liu 1,2,3,*, Yonghao Xu 1,2,3, Zhiheng Xu 1,2,3, Yongbo Huang 1,2,3, Sibei Chen 1,2,3, Shiyue Li 1,2,3, Dongdong Liu 1,2,3, Zhimin Lin 1,2,3, Yimin Li 1,2,3,
PMCID: PMC7233337  PMID: 32203672

To the Editor:

Lung-protective ventilation with low Vt has become a cornerstone of management in patients with acute respiratory distress syndrome (ARDS) (1, 2). However, a consequence of low-Vt ventilation is hypercapnia, which has significant physiological effects and may be associated with higher hospital mortality (2, 3).

Ventilatory ratio (VR), defined as [minute ventilation (ml/min) × PaCO2 (mm Hg)]/[predicted body weight × 100 (ml/min) × 37.5 (mm Hg)] (4), is a simple bedside index of impaired efficiency of ventilation and correlates well with physiological Vd fraction (Vd-to-Vt ratio, Vd/Vt) in patients with ARDS (46). However, the VR and appropriate lung ventilation strategy for coronavirus disease (COVID-19)-associated ARDS remain largely unknown.

Here, we report a case series highlighting ventilatory ratio in hypercapnic mechanically ventilated patients with COVID-19–associated ARDS in our ICU and their individualized ventilation strategies.

Case Series

The study was approved by the ethics committee of the First Affiliated Hospital of Guangzhou Medical University. The requirement for informed consent was waived because the study was observational and the family members were in quarantine.

The First Affiliated Hospital of Guangzhou Medical University is the designated center for patients with COVID-19 in Guangdong, China. We included eight consecutive patients (seven male; mean age, 63.2 ± 11.0 yr) who were intubated in another hospital before being transferred to our ICU. All patients had a history of exposure in Wuhan City or direct contact with patients with confirmed COVID-19. All patients reported fever, cough, shortness of breath, and generalized weakness before hospitalization and tested positive for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) on the basis of real-time PCR of throat swab specimens. All patients were diagnosed with ARDS according to the Berlin definition (7): PaO2/FiO2 ratio, 102.0 ± 29.7 mm Hg (mean ± SD), with Acute Physiology and Chronic Health Evaluation II score 21.6 ± 5.3 and Sequential Organ Failure Assessment score 9.1 ± 2.7 (Table 1).

Table 1.

Baseline Characteristics of Eight Patients with Acute Respiratory Distress Syndrome Infected with SARS-CoV-2

Characteristic Patients (N = 8)
Exposure history 8/8
Age, yr 63.2 ± 11.0
Sex, M 7/8
Body mass index, kg/m2 22.7 ± 2.3
Predicted body weight, kg 64.7 ± 6.0
Chronic medical illness  
 Hypertension 4/8
 Diabetes 3/8
 Coronary heart disease 1/8
 Chronic obstructive pulmonary disease 1/8
 Obstructive sleep apnea syndrome 1/8
 Hepatitis B 1/8
 Smoker 3/8
Presenting symptoms onset  
 Fever 8/8
 Cough 7/8
 Generalized weakness 4/8
 Shortness of breath 3/8
Real-time RT-PCR of throat swab 8/8
Radiologic characteristics  
 Bilateral pneumonia 8/8
 Multiple mottling and ground-glass opacity 8/8
Noninvasive ventilation before intubation 1/8
 Duration of noninvasive ventilation, d 1
HFNC before intubation 7/8
 Duration of HFNC, d 2.6 ± 2.2
PaO2/FiO2 ratio, mm Hg 102.0 ± 29.7
APACHE II score 21.6 ± 5.3
SOFA score 9.1 ± 2.7
Weaning before day 28 at ICU 5/8
Discharge before day 28 at ICU 5/8
28-d mortality at ICU 0/8

Definition of abbreviations: APACHE = Acute Physiology and Chronic Health Evaluation; HFNC = high-flow nasal cannula oxygen therapy; SARS-CoV-2 = severe acute respiratory syndrome coronavirus 2; SOFA = Sequential Organ Failure Assessment.

Data are presented as mean ± SD or n/N unless otherwise noted.

A ventilation strategy using a low Vt of 6.0 ml/kg predicted body weight (PBW) was used in the first four consecutive patients. However, they had respiratory distress with low oxygen saturation as measured by pulse oximetry, so we immediately increased Vt to 7.0 ± 0.6 ml/kg PBW (Table 2). This resulted in an acceptable plateau pressure (23.3 ± 2.2 cm H2O) and driving pressure (12.3 ± 1.7 cm H2O). However, all four patients developed hypercapnia (PaCO2, 57.7 ± 5.2 mm Hg). Respiratory system compliance was only moderately reduced (static respiratory system compliance, 35.7 ± 5.8 ml/cm H2O). To examine this issue, we measured VR; the mean value was 2.1 ± 0.3 in the initial four patients, suggesting high Vd/Vt (46).

Table 2.

Ventilator Settings

Variables Low Vt (Initial 4 Patients) Intermediate Vt (Initial 4 Patients) P Value Intermediate Vt (8 Patients)
Vt, ml/kg PBW 7.0 ± 0.6 7.7 ± 0.8 0.022 7.5 ± 0.6
PaCO2, mm Hg 57.7 ± 5.2 44.1 ± 3.6 0.003 41.8 ± 3.7
PaO2/FiO2 ratio 207 ± 61 241 ± 38 0.402 230 ± 49
RR, beats/min 21.5 ± 2.0 21.0 ± 1.4 0.182 20.1 ± 1.5
Ve, L/min 9.1 ± 1.0 9.8 ± 1.0 0.020 9.3 ± 1.0
Ventilation ratio 2.1 ± 0.3 1.7 ± 0.2 0.018 1.6 ± 0.2
Pplat, cm H2O 23.3 ± 2.2 23.3 ± 3.1 >0.999 23.6 ± 2.7
PEEP, cm H2O 11.0 ± 1.2 10.0 ± 1.4 0.250 9.6 ± 1.2
ΔP, cm H2O 12.3 ± 1.7 13.5 ± 2.7 0.080 14.1 ± 2.5
Cst, ml/cm H2O 35.7 ± 5.8 36.1 ± 7.9 0.595 33.9 ± 7.6
EELV, ml 2,559 ± 61 2,285 ± 355

Definition of abbreviations: Cst = static respiratory system compliance; ΔP = driving pressure; EELV = end-expiratory lung volume; PBW = predicted body weight; PEEP = positive end-expiratory pressure; Pplat = plateau pressure; RR = respiratory rate.

Data are presented as mean ± SD. P value indicates difference between low Vt and intermediate Vt of the initial four patients using a paired t test.

We then performed titration of Vt. An increased Vt (7.7 ± 0.8 ml/kg PBW) was applied to the initial four patients (Table 2). PaCO2 decreased significantly compared with Vt 7.0 ml/kg PBW (57.7 ± 5.2 vs. 44.1 ± 3.6 mm Hg; P = 0.003) with permitted plateau pressure (23.3 ± 3.1 cm H2O) and driving pressure (13.5 ± 2.7 cm H2O). Importantly, VR in the four patients was significantly decreased (1.7 ± 0.2 vs. 2.1 ± 0.3; P = 0.018) and PaO2/FiO2 was slightly improved (241 ± 38 mm Hg vs. 207 ± 61; P = 0.402) compared with Vt 7.0 ml/kg PBW. Therefore, an intermediate Vt of 7.5 ± 0.6 ml/kg PBW was applied to the subsequent four patients with COVID-19 ARDS. The PaCO2 was 41.8 ± 3.7 mm Hg, and VR was 1.6 ± 0.2.

Discussion

We found that hypercapnia was common in patients with COVID-19–related ARDS with low Vt ventilation. High VR was found in these patients, indicating inadequacy of ventilation in patients with ARDS with COVID-19. An intermediate Vt (7–8 ml/kg PBW) ventilation strategy was applied to the first four patients to increase pulmonary efficiency to eliminate CO2, and this was used in the next four patients.

Gas exchange consists of oxygenation and ventilation. Oxygenation is quantified by the PaO2/FiO2 ratio, and this method has gained wide acceptance, particularly since publication of the Berlin definition of ARDS (7). However, the Berlin definition does not include additional pathophysiological information about ARDS, such as alveolar ventilation, as measured by pulmonary dead space, which is an important predictor of outcome (8). Increased pulmonary dead space reflects the inefficiency of the lungs to eliminate CO2, which may lead to hypercapnia.

In our patients with ARDS with COVID-19, hypercapnia was common at ICU admission with low Vt ventilation. Assuming the anatomic portion of dead space is constant, increasing Vt with constant respiratory rate would effectively increase alveolar ventilation. Any such increase in Vt would decrease PaCO2, which would be captured by VR (6). VR, a novel method to monitor ventilatory adequacy at the bedside (46), was very high in our patients, reflecting increased pulmonary dead space and inadequacy of ventilation.

With an acceptable plateau pressure and driving pressure, titration of Vt was performed. PaCO2 and VR were significantly decreased when an intermediate Vt (7–8 ml/kg PBW) was applied. We suggest that intermediate Vt (7–8 ml/kg PBW) is recommended for such patients. Therefore, low Vt may not be the best approach for all patients with ARDS, particularly those with a less severe decrease in respiratory system compliance and inadequacy of ventilation.

In summary, we found that hypercapnia was common in patients with COVID-19–associated ARDS while using low Vt ventilation. VR was increased in these patients, which reflected increased pulmonary dead space and inadequacy of ventilation. An intermediate Vt was used to correct hypercapnia efficiently, while not excessively increasing driving pressure. Clinicians must have a high index of suspicion for increased pulmonary dead space when patients with COVID-19–related ARDS present with hypercapnia.

Supplementary Material

Supplements
Author disclosures

Acknowledgments

Acknowledgment

The authors thank Dr. Arthur S. Slutsky for the invaluable assistance with the manuscript.

Footnotes

Supported by National Science and Technology Major Project (No. 2017ZX10204401), National Natural Science Foundation of China (81970071), the Special Project for Emergency of the Ministry of Science and Technology (2020YFC0841300), and the Special Project of Guangdong Science and Technology Department (2020B111105001).

Author Contributions: Xiaoqing Liu, Xuesong Liu, Y.X., and Y.L. conceived and designed the study; Xiaoqing Liu, Xuesong Liu, Y.X., Z.X., Y.H., and Y.L. analyzed the data and wrote the manuscript; Xiaoqing Liu, Xuesong Liu, Y.X., Z.X., Y.H., S.C., S.L., D.L., Z.L., and Y.L. reviewed and revised the manuscript.

Originally Published in Press as DOI: 10.1164/rccm.202002-0373LE on March 23, 2020

Author disclosures are available with the text of this letter at www.atsjournals.org.

References

  • 1.Brower RG, Matthay MA, Morris A, Schoenfeld D, Thompson BT, Wheeler AAcute Respiratory Distress Syndrome Network Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome N Engl J Med20003421301–1308. [DOI] [PubMed] [Google Scholar]
  • 2.Barnes T, Zochios V, Parhar K.Re-examining permissive hypercapnia in ARDS: a narrative review Chest2018154185–195. [DOI] [PubMed] [Google Scholar]
  • 3.Tiruvoipati R, Pilcher D, Buscher H, Botha J, Bailey M.Effects of hypercapnia and hypercapnic acidosis on hospital mortality in mechanically ventilated patients Crit Care Med201745e649–e656. [DOI] [PubMed] [Google Scholar]
  • 4.Sinha P, Fauvel NJ, Singh S, Soni N.Ventilatory ratio: a simple bedside measure of ventilation Br J Anaesth2009102692–697. [DOI] [PubMed] [Google Scholar]
  • 5.Sinha P, Fauvel NJ, Singh P, Soni N.Analysis of ventilatory ratio as a novel method to monitor ventilatory adequacy at the bedside Crit Care201317R34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Sinha P, Calfee CS, Beitler JR, Soni N, Ho K, Matthay MA, et al. Physiologic analysis and clinical performance of the ventilatory ratio in acute respiratory distress syndrome Am J Respir Crit Care Med2019199333–341. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ranieri VM, Rubenfeld GD, Thompson BT, Ferguson ND, Caldwell E, Fan E, et al. ARDS Definition Task Force Acute respiratory distress syndrome: the Berlin definition JAMA20123072526–2533. [DOI] [PubMed] [Google Scholar]
  • 8.Nuckton TJ, Alonso JA, Kallet RH, Daniel BM, Pittet JF, Eisner MD, et al. Pulmonary dead-space fraction as a risk factor for death in the acute respiratory distress syndrome N Engl J Med20023461281–1286. [DOI] [PubMed] [Google Scholar]