. 2020 May 19;129:104439. doi: 10.1016/j.jcv.2020.104439

SARS-CoV-2 viral load assessment in respiratory samples

Steven Kleiboeker 1,*, Scott Cowden 1, James Grantham 1, Jamie Nutt 1, Aaron Tyler 1, Amy Berg 1, Michelle Altrich 1
PMCID: PMC7235577  PMID: 32674034

Abstract

Real-time reverse transcriptase polymerase chain reaction (rRTPCR) has been the main method for diagnosis of SARS-CoV-2 infection in the early stages of the COVID-19 pandemic. De-identified results from upper and lower respiratory samples submitted to a reference laboratory demonstrated a positivity rate of 14.9 % (4428 of 29,713 samples tested). Distribution of results by birth year cohort and specimen type suggested general consistency in mean, median and peak values but higher positivity rates in individuals born from 1964 to 1974. Female patients had a significantly lower positivity rate (P < 0.0001), although similar load mean and median values, compared to males. Overall, 15.3 % (676 of 4428 positive results) of positive results had viral loads greater than 8 log10 copies/mL, with occasional samples exceeding 10 log10 copies/mL. These results support quantitative assessment of SARS-CoV-2 viral load in patient testing and efforts to control viral transmission

Keywords: SARS-CoV-2, Viral load diagnostic, RT-PCR


In the early stages of the COVID-19 pandemic, molecular testing by real-time reverse transcriptase polymerase chain reaction (rRT-PCR) has been the main method for diagnosis of SARS-CoV-2 infection in symptomatic patients [1]. Qualitative assays have primarily been used for this purpose. Our reference laboratory developed and validated an assay that targets two regions of the N protein gene using standard rRT-PCR methods. Validation of upper and lower respiratory samples was focused on providing qualitative results. Positivity was defined as a cycle threshold (CT) value < 38.0. This cutoff was derived from the mean CT value at the limit of detection plus approx. two standard deviations. Sensitivity of the assay, expressed as limit of detection, is 73 copies/mL. Although the intended use of the assay is to report qualitative results, linearity and dynamic range were assessed and demonstrated from 2.77 to 8.77 log10 copies/mL (slope = -3.32, y-intercept = 41.2, R2 = 0.9998), with CT values from 31.56 to 12.05 in the linear range of the assay. During review of laboratory results, we occasionally observed very low CT values (i.e. < 10) from patient samples, suggesting extremely high concentrations of SARS-CoV-2 in respiratory specimens. This report characterizes the viral load of all positive samples in an unbiased reference laboratory caseload from the United States.

Using 29,713 de-identified rRT-PCR results, we described the viral load of respiratory samples tested during the first weeks of the COVID-19 pandemic in the United States. Overall 14.9 % (4428) of samples tested were positive for SARS-CoV-2. The distribution of positive samples by log10 copies/mL and CT values is shown in Fig. 1 . The overall viral load range of positive samples was 0.91–10.42 log10 copies/mL (CT 6.16–37.92) with a mean of 5.85 log10 copies/mL and median of 6.05 log10 copies/mL. Samples with greater than 8 log10 copies/mL represented 15.3 % (676 of 4428) of results reported as positive. Samples with less than 3 log10 copies/mL represented 9.89 % (438 of 4428) of results reported as positive.

Fig. 1.

Fig. 1

Frequency distribution histograms of CT (A) and Log10 copies/mL values (B) for SARS-CoV-2 viral loads in respiratory samples tested by a reference laboratory.

Assessment of viral load results by birth year of patients is shown in Table 1 . Patients born from 1955 to 1994 had both the highest number and percentage of positive results, with peak values in patients born from 1960 to 1974. Mean and median viral load values did not vary markedly based on birth cohort, although it was noted that the highest maximum values (>10 log10 copies/mL) were in patients born from 1995 to 2009. In Table 2 , viral load results are shown by specimen type and patient gender. A majority of samples received (69.3 % or 20,603 of 29,713) did not have the specific specimen type identified. When specimen type was identified, NP swabs were the majority of samples tested (84.3 % or 7682 of 9110). Although more females were tested, the percentage of samples tested which were positive was significantly lower (Chi square test, P <  0.0001) compared to males while no significant differences were observed in viral load mean or median values.

Table 1.

SARS-CoV-2 viral load characteristics in respiratory samples by birth cohort.

SARS-CoV-2 Log10 copies/mL
Birth cohort No. tested No. Pos. % of Pos. results Average (SD) Median Minimum Maximum
2010 - 2020 1197 15 0.34% 6.13 (2.07) 6.27 2.04 8.64
2000 - 2009 1157 81 1.83% 5.89 (2.29) 6.38 1.08 10.19
1995 - 1999 1935 248 5.60% 6.03 (2.03) 6.44 0.96 10.42
1990 - 1994 2647 388 8.76% 6.09 (1.97) 6.27 0.91 9.84
1985 - 1989 2994 405 9.15% 5.90 (2.06) 6.15 1.09 9.18
1980 - 1984 2862 380 8.58% 5.77 (2.01) 5.97 1.00 9.62
1975 - 1979 2601 416 9.39% 5.88 (1.94) 6.12 1.05 9.78
1970 - 1974 2548 450 10.16% 5.74 (1.95) 5.76 0.92 9.27
1965 - 1969 2497 463 10.46% 5.82 (1.88) 5.98 0.93 9.44
1960 - 1964 2447 453 10.23% 5.77 (2.00) 6.05 0.94 9.64
1955 - 1959 2180 359 8.11% 5.75 (1.94) 5.91 1.12 9.17
1950 - 1954 1625 264 5.96% 5.81 (1.96) 6.03 1.08 9.43
1945 - 1949 1224 201 4.54% 5.89 (1.94) 6.00 1.13 9.01
1940 - 1944 818 115 2.60% 5.77 (1.87) 5.75 1.44 9.20
1935 - 1939 446 83 1.87% 5.83 (2.06) 6.15 0.94 9.49
1930 - 1934 348 79 1.78% 5.68 (2.00) 6.04 1.20 8.64
1925 - 1929 134 25 0.56% 6.41 (1.77) 6.66 2.48 8.51
1920 - 1924 49 3 0.07% 6.67 (2.74) 7.82 3.54 8.64
1915 - 1919 4 0 0.00% N.A.a N.A. N.A. N.A.
a

N.A., Not applicable.

Table 2.

SARS-CoV-2 viral load characteristics in respiratory samples by specimen type and gender.

SARS-CoV-2 Log10 copies/mL
No. tested No. pos. (% pos) Average (SD) Median Minimum Maximum
Specimen type
BAL 130 8 (6.15 %) 5.37 (2.13) 5.06 2.65 8.90
Nasal swab 677 61 (9.01 %) 5.52 (2.35) 5.99 1.38 9.18
Nasal wash 42 5 (11.90 %) 5.23 (2.64) 5.74 2.36 8.52
NP Swab 7682 711 (9.26 %) 5.66 (2.03) 5.78 0.93 9.49
NP wash 63 5 (7.94 %) 6.09 (2.25) 6.90 2.97 8.59
OP swab 191 8 (4.19 %) 4.52 (1.30) 4.47 2.69 6.53
NP/OP swab 324 19 (5.86 %) 5.66 (1.65) 5.40 2.55 8.36
Not specified 20,603 3611 (17.53 %) 5.90 (1.96) 6.13 0.91 10.42
Gender
Female 17,587 2153 (12.24 %) 5.88 (1.98) 6.09 0.91 10.42
Male 11,766 2261 (19.22 %) 5.82 (1.97) 6.02 0.92 9.84
Not specified 360 14 (3.89 %) 6.13 (2.20) 6.53 1.66 8.63

Recent reports of CT values and viral loads in SARS-CoV-2 infected patients have demonstrated shedding at high viral load levels in individual patients or small groups of patients. While some studies only report CT values, which may vary between methods, these results provide a reasonable approximation of viral load. Lack of a common standard makes comparison of calculated viral load values between studies an approximation as well. A study of 82 SARS-CoV-2 infected patients showed viral loads of 4–8 log10 copies/mL in throat swabs and sputum from -1 to 15 days post-onset of clinical symptoms with occasional sputum samples having 10–11 log10 copies/mL [2]. A second study of 17 symptomatic patients reported CT values as low as 13 in both throat and nasal swabs of one patient, with a majority samples having CT values >20 in the early clinical period [3]. Wang et al. [4] reported CT values > 17 in similarly collected specimens. Viral load kinetics have also been studied with peak reported values of approx. 8 log10 copies/mL [5] and 9 log10 copies/mL [6], both in upper respiratory specimens. Initial and peak mean viral loads of 6.17 and 6.91 log10 copies/mL, respectively, have also been reported for 10 SARS-CoV-2 infected patients severe disease; values were 10-fold lower (initial viral load) and 30-fold lower (peak viral load) in 13 patients with mild disease by [7].

Viral loads in respiratory samples from patients infected with other respiratory coronaviruses have also been reported. In 14 patients infected with SARS-CoV-1, peak viral loads of approx. 5.5–8.5 log10/mL were documented in nasopharyngeal aspirates [8]. For patients with severe MERS, peak viral loads in sputum or tracheal aspirates ranged from 6 to 9 log10 copies/mL while mild cases were at least 100-fold lower [9]. Viral load results for common respiratory coronaviruses such as strains NL63 and OC43 are generally 1000- to 10,000-fold lower compared to peak values for SARS-CoV-1, SARS-CoV-2 and MERS [10].

The limitations of this study are that clinical information was not available from the reference laboratory caseload and full de-identification of samples was required for research use. Most importantly, the proportion of samples tested from symptomatic patients versus asymptomatic individuals was not known, nor is it known when samples were collected relative to the time of symptom onset for those patients suffering from disease. Therefore the overall prevalence of positivity is minimally significant and the prevalence by birth year may well be skewed based on likelihood of sampling due to age or occupation. However, additional study of asymptomatic or symptomatic patient samples with defined clinical characteristics is warranted based on our data.

In summary, our results demonstrate that a significant number of patients have extremely high levels of virus (> 8 log10 copies/mL) in respiratory samples and thus may be responsible for a disproportionate level of transmission to contacts and contamination of environmental surfaces. Additionally, results demonstrated that samples with low viral loads (<1000 copies/mL) are common, and individuals from which these samples were collected may represent a low risk for transmission. This study supports the concept that quantitative viral detection methods for SARS-CoV-2 diagnosis may provide significant benefits in efforts to control viral transmission, and perhaps even in the medical care of these patients.

Declaration of Competing Interest

The authors have no conflicts of interest to declare.

References

  • 1.Corman V.M., Landt O., Kaiser M., Molenkamp R. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020;25(3) doi: 10.2807/1560-7917.ES.2020.25.3.2000045. pii=2000045. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Pan Y., Zhang D., Yang P., Poon L.L.M., Wang Q. Viral load of SARS CoV-2 in clinical samples. Lancet Infect. Dis. 2020;20(4):411–412. doi: 10.1016/S1473-3099(20)30113-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Zou L., Ruan F., Huang M., Liang L. SARS-CoV-2 viral load in upper respiratory specimens of infected patients. N. Engl. J. Med. 2020;382(12):1177–1179. doi: 10.1056/NEJMc2001737. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Wang W., Xu Y., Gao R., Lu R. Detection of SARS-CoV-2 in different types of clinical specimens. JAMA. 2020;(March (11)) doi: 10.1001/jama.2020.3786. [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Kim J.Y., Ko J.H., Kim J.Y., Kim J.Y. Viral load kinetics of SARS-CoV-2 infection in first two patients in Korea. J. Korean Med. Sci. 2020;35(7):e86. doi: 10.3346/jkms.2020.35.e86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Wölfel R., Corman V.M., Guggemos W., Seilmaier M. Virological assessment of hospitalized patients with COVID-2019. Nature. 2020;(April (1)) doi: 10.1038/s41586-020-2196-x. [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
  • 7.To K.K., Tsang O.T., Leung W.S., Tam A.R. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study. Lancet Infect. Dis. 2020 doi: 10.1016/S1473-3099(20)30196-1. pii: S1473-3099(20)30196-1. [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Peiris J.S., Chu C.M., Cheng V.C., Chan K.S. Clinical progression and viral load in a community outbreak of coronavirus-associated SARS pneumonia: a prospective study. Lancet. 2003;361(9371):1767–1772. doi: 10.1016/S0140-6736(03)13412-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Oh M.D., Park W.B., Choe P.G., Choi S.J. Viral load kinetics of MERS coronavirus infection. N. Engl. J. Med. 2016;375(13):1303–1305. doi: 10.1056/NEJMc1511695. [DOI] [PubMed] [Google Scholar]
  • 10.Lassaunière R., Kresfelder T., Venter M. A novel multiplex real-time RT-PCR assay with FRET hybridization probes for the detection and quantitation of 13 respiratory viruses. J. Virol. Methods. 2010;165(2):254–260. doi: 10.1016/j.jviromet.2010.02.005. [DOI] [PMC free article] [PubMed] [Google Scholar]