Abstract
Lipidomics is a rapidly growing field that enables the characterization of the entire lipidome either in cells, tissues, or an organism. Changes in lipid metabolism and homeostasis caused by different disease states or drug treatments can be probed by lipidomics experiments, which can aid our understanding of normal physiology and disease pathology at the molecular level. While current technologies using liquid chromatography coupled with high-resolution mass spectrometry has greatly increased the coverage of lipidome, there is still limitation in resolving the large number of lipid species with similar masses in a narrow mass window. We recently reported that two orthogonal separation techniques, hydrophilic interaction liquid chromatography (HILIC) and ion mobility (IM), enhance the resolution of lipid species based on headgroup polarity and gas-phase size and shape, respectively, for various classes of glycerolipids, glycolipids, phospholipids, and sphingolipids. Here we describe the application of our HILIC-IM-MS lipidomics protocol to the analysis of lipid extracts derived from either tissues or cells, to identify significant changes in the lipidome in response to an internal or external stimulus, such as exposure to environmental chemicals.
Keywords: hydrophilic interaction liquid chromatography, ion mobility, mass spectrometry, lipidomics, collision cross section
1. Introduction
In recent years, the advancement of singular and multi-disciplinary ‘omics technologies have quickly taken storm as a systems biology approach is applied to biomedical research. They are highly informative experiments that can reveal molecular changes within some altered state of a system, such as a disease or a treatment condition [1]. The central biological processes of DNA transcription to mRNA, subsequent translation to protein, and eventual production of metabolites correspond to the studies of genomics, transcriptomics, proteomics, and metabolomics, respectively. Lipidomics falls under the category of metabolomics as the lipidome represents a major fraction of the human metabolome [2]. Specifically, lipidomics aims to systematically identify and quantify individual lipid species within cells, tissues, or an organism, as well as map out the complex interactions between lipids and other macromolecules. There is growing evidence that lipids are closely related to the onset and progression of several diseases, including neurodegenerative, neurodevelopmental, cardiovascular, and metabolic disorders. Thus, lipidomics represents a unique niche to address specific questions about the role of altered lipid metabolism within disease etiologies, as well as advance biomarker identifications to aid diagnosis [3].
Lipids cover a diverse range of functions within the cell as structural components of membranes, energy reservoirs, and signaling molecules, which reflects their large chemical diversity. There are eight main lipid classes (FA, fatty acyls; GL, glycerolipids; GP, glycerophospholipids; SP, sphingolipids; ST, sterol lipids; PR, prenol lipids; SL, saccharolipids; PK, polyketides) and multiple subclasses [4, 5]. Within each subclass, further variations are possible from multiple types of head groups, backbones, fatty acyl chains, linkages (ester, ether, or vinyl ether), and other accessory modifications, like sugar molecules [4, 5]. This vast combinatorial space is estimated (by the LIPID MAPS database) to contain over 40,000 distinct lipids, which presents a major analytical challenge for lipidomics because most lipid species occupy a narrow mass range of 600 – 900 m/z. Modern mass spectrometry has greatly advanced the field of lipidomics and increased coverage of the lipidome over the past two decades, but some limitations remain.
Mass spectrometry-based lipidomics can be divided into two categories: shotgun [6, 7] and liquid chromatography (LC)-based experiments [8–10]. In shotgun lipidomics, samples are directly infused into the mass spectrometer without any prior chromatographic separation, aiming to maximize the throughput of analysis. However, this approach is unable to resolve isomeric and isobaric lipids and limits the detection of lower abundance lipids due to ion suppression. LC-based lipidomics involves LC separation, such as normal phase (NP), reverse phase (RP), or hydrophilic interaction liquid chromatography (HILIC), prior to ionization. RP separation utilizes nonpolar C18 columns with polar solvents, where lipids are separated based on hydrophobicity, resulting in separation by chain length and degree of unsaturation of fatty acyl chains. HILIC separation utilizes columns with polar stationary phase in combination with typical reverse phase solvents, such as acetonitrile and water, where lipids partition at the water/silica interface. HILIC separation is mostly dependent on the polarity of lipid head groups [11, 12]. Further separation, although to a lesser extent, occurs within each subclass by chain length and degree of unsaturation of fatty acyl chains.
Ion mobility (IM) is another technology that has gained popularity and shown advantages in lipidomics studies as it contributes to both separation and structural elucidation of analytes and can be coupled to LC separation without affecting throughput [13–16]. Ions are separated based on their size, shape, and charge as they are pushed through a drift tube by an electric field through an opposing flow of inert gas [16–18]. Drift times are related to a physicochemical property, called collisional cross section (CCS), which can be measured either directly on a drift tube-IM (DTIM) instrument using the Mason-Schamp equation [19, 20] or on a traveling wave-IM (TWIM) instrument through calibration [21–23]. Previous papers from our group have shown that in TWIM experiments, matched lipid calibrants produce more accurate and precise CCS values of lipid standards compared to calibration by calibrants from other classes of molecules, such as poly-DL-alanine, tetraalkylammonium compounds, and hexakis(fluoroalkoxy)phosphazines [21]. IM is a gas-phase technique that is orthogonal to LC and enables separation of lipid ions from other classes of macromolecules due to their distinct IM profiles [24, 25]. Even within total lipids, individual lipid classes can be separated from each other, although complete resolution of certain species is still limited by instrumentation [15, 26–28]. Coupled with retention times (RT) and mass-to-charge (m/z) ratios, CCS values greatly increase the confidence in lipid identifications. Recently, the Zhu Lab reported the first platform for lipid identification using multidimensional RT-m/z-CCS data, LipidIMMS, which can utilize both RP-IM-MS and HILIC-IM-MS data [29].
Herein, we discuss the use of our HILIC-IM-MS protocol [14, 15] for the analysis of lipid species in the lipid extracts of biological samples using brain tissues and cells as examples. The workflow involves (1) sample preparation including tissue homogenization, lysis, and liquid phase extraction, (2) data acquisition on a Waters SYNAPT HDMS G2-Si Q-TOF ion mobility-mass spectrometer and (3) data analysis using Progenesis QI and EZ Info software. The data included in Figure 4 comes from a paper by members of our research group, investigating the lipidomic changes in a neuroblastoma cell line exposed to an environmental chemical, benzylalkonium chlorides (BACs), and related compounds known to disrupt cholesterol and lipid homeostasis [14]. Readers are encouraged to consult the original paper for the biological context.
Figure 4 –
Schematic of HILIC-IM-MS data analysis workflow from MS data processing to lipid identification.
2. Materials
General
Ammonium hydroxide (certified ACS reagent), ammonium acetate (Optima LC/MS grade), and sodium chloride (certified ACS crystalline) were purchased from Fisher Scientific. All mobile phase solvents (acetonitrile and water) and extraction reagents (chloroform, methanol, and methylene chloride) were Fisher Optima LC/MS grade. Lipid standards were purchased from Avanti Polar Lipids (Alabaster, AL) and Nu-Chek Prep (Elysian, MN). Phosphate buffer saline (PBS) was purchased from Gibco.
2.1. Sample Preparation
1x phosphate buffer saline.
DC™ Protein Assay Kit II.
2.2. Folch Lipid Extraction
Folch solution: 2:1 (v/v) chloroform/methanol.
Sodium chloride solution: 0.9% (w/v) NaCl aqueous solution.
2.3. HILIC-MS
HILIC A mobile phase: 95% ACN, 5% H2O, 5 mM ammonium acetate.
HILIC B mobile phase: 50% ACN, 50% H2O, 5 mM ammonium acetate.
2.4. Lipid Standards
Lipid internal standard solution: prepare 1 mM stock solutions of lipid standards in chloroform. Prepare a 5 μM mixture in HILIC A (see Table 1 for the list of lipid standards).
Lipid mixture solution: extracts purchased from Avanti Polar Lipids were prepared at 1 mM concentration in chloroform. Prepare a 10 μM mixture in chloroform. For analysis, make a 1:2 dilution of the mixture in HILIC A (see Table 2 for the list of lipid extracts).
CCS calibrants: phosphatidylcholine (PC) and phosphatidylethanolamine (PE) lipid standards (Avanti Polar Lipids). Prepare 1 mM stock solutions of lipid standards. Prepare a mixture of PC 6:0–24:0 in methanol with 0.1% formic acid for positive mode analysis and PE 6:0–24:0 at 5–10 μM in methanol with 50 μM ammonium hydroxide for negative mode analysis. Prior to calibration, make a 1:4 dilution of the PC mixture and 1:2 dilution of the PE mixture in their respective solvents (see Table 3).
Table 1.
Lipid standards included in discrete lipid internal standards mixture.
Lipid common name | Systematic name | Catalog no. |
---|---|---|
DG 13:0/13:0 | Ditridecanoin glyceride | D-136 (Nu-Chek) |
TG 15:0/15:0/15:0 | Tripentadecanoin glyceride | T-145 (Nu-Chek) |
Cer d18:1/17:0 | N-heptadecanoyl-D-erythro-sphingosine | 860517 |
PG 15:0/15:0 | 1,2-dipentadecanoyl-sn-glycero-3-phospho-(1′-rac-glycerol) | 840446 |
PE 15:0/15:0 | 1,2-dipentadecanoyl-sn-glycero-3-phosphoethanolamine | 850704 |
PC 15:0/15:0 | 1,2-dipentadecanoyl-sn-glycero-3-phosphocholine | 850350 |
PA 12:0/12:0 | 1,2-dilauroyl-sn-glycero-3-phosphate | 840635 |
PS 12:0/12:0 | 1,2-dilauroyl-sn-glycero-3-phospho-L-serine | 840038 |
SM d18:1/17:0 | N-heptadecanoyl-D-erythro-sphingosylphosphorylcholine | 860585 |
C12 Sphingosyl PE | N-lauroyl-D-erythro-sphingosyl phosphoethanolamine | 860529 |
Lyso PC 15:0/0:0 | 1-pentadecanoyl-2-hydroxy-sn-glycero-3 -phosphocholine | 855576 |
Lyso PE 13:0/0:0 | 1-tridecanoyl-sn-glycero-3-phosphoethanolamine | 856706 |
Table 2.
Lipid extracts included in lipid mixture.
Lipid class | Catalog no. |
---|---|
Ceramide | 860052P |
Cerebroside | 131303P |
Sphingomyelin | 860062C |
L-α-phosphatidic acid | 840101C |
L-α-phosphatidylcholine | 840051C |
L-α-lysophosphatidylcholine | 830071P |
L-α-phosphatidylethanolamine | 840022C |
L-α-phosphatidylglycerol | 841138C |
L-α-phosphatidylinositol | 840042C |
L-α-phosphatidylserine | 840032C |
Diacylglyceride | Nu-Chek Prep |
Table 3.
PE and PC lipid species included in the lipid CCS calibration mixture.
Lipid common name | Systematic name | Catalog no. | Conc. (μM) |
---|---|---|---|
PE | |||
PE 6:0/6:0 | 1,2-dihexanoyl-sn-glycero-3-phosphoethanolamine | 850697C | 5 |
PE 8:0/8:0 | 1,2-dioctanoyl-sn-glycero-3-phosphoethanolamine | 850699C | 5 |
PE 10:0/10:0 | 1,2-didecanoyl-sn-glycero-3-phosphoethanolamine | 850700C | 5 |
PE 12:0/12:0 | 1,2-dilauroyl-sn-glycero-3-phosphoethanolamine | 850702X | 5 |
PE 14:0/14:0 | 1,2-dimyristoyl-sn-glycero-3-phosphoethanolamine | 850745X | 5 |
PE 15:0/15:0 | 1,2-dipentadecanoyl-sn-glycero-3-phosphoethanolamine | 850704X | 10 |
PE 16:1/16:1 | 1,2-dipalmitoleoyl-sn-glycero-3-phosphoethanolamine | 850706C | 10 |
PE 16:0/16:0 | 1,2-dipalmitoyl-sn-glycero-3-phosphoethanolamine | 850705X | 10 |
PE 16:0/18:1 | 1-palmitoyl-2-oleoyl-sn-glycero-3-phosphoethanolamine | 850757C | 10 |
PE 17:0/17:0 | 1,2-diheptadecanoyl-sn-glycero-3-phosphoethanolamine | 830756X | 10 |
PE 18:1/18:1 | 1,2-dioleoyl-sn-glycero-3-phosphoethanolamine | 850725C | 10 |
PE 18:0/18:1 | 1-stearoyl-2-oleoyl-sn-glycero-3-phosphoethanolamine | 850758C | 10 |
PE 18:0/18:0 | 1,2-distearoyl-sn-glycero-3-phosphoethanolamine | 850715X | 10 |
PE 20:4/20:4 | 1,2-diarachidonoyl-sn-glycero-3-phosphoethanolamine | 850800C | 10 |
PC | |||
PC 6:0/6:0 | 1,2-dihexanoyl-sn-glycero-3-phosphocholine | 850305C | 5 |
PC 8:0/8:0 | 1,2-dioctanoyl-sn-glycero-3-phosphocholine | 850315C | 5 |
PC 10:0/10:0 | 1,2-didecanoyl-sn-glycero-3-phosphocholine | 850325C | 5 |
PC 12:0/12:0 | 1,2-dilauroyl-sn-glycero-3-phosphocholine | 850335C | 5 |
PC 14:0/14:0 | 1,2-dimyristoyl-sn-glycero-3-phosphocholine | 850345C | 5 |
PC 16:0/16:0 | 1,2-dipalmitoyl-sn-glycero-3-phosphocholine | 850355C | 10 |
PC 18:0/18:0 | 1,2-distearoyl-sn-glycero-3-phosphocholine | 850365C | 10 |
PC 20:0/20:0 | 1,2-diarachidoyl-sn-glycero-3-phosphocholine | 850368C | 10 |
PC 22:0/22:0 | 1,2-dibehenoyl-sn-glycero-3-phosphocholine | 850371C | 10 |
PC 24:0/24:0 | 1,2-dilignoceroyl-sn-glycero-3-phosphocholine | 850373C | 10 |
2.5. Supplies and Equipment
Analytical Column: HILIC Phenomenex Kinetex®, 2.1 × 100 mm, 1.7 μm.
Guard Cartridge: Phenomenex SecurityGuard® ULTRA HILIC Cartridge for 2.1mm ID UHPLC columns.
Guard Cartridge Holder: Phenomenex SecurityGuard® ULTRA Holder for 2.1–4.6 mm ID UHPLC columns.
Instrument: Waters Synapt® G2-Si HDMS equipped with an ESI source (Waters Corporation, Milford, MA, USA).
Inlet System: Waters® Acquity I-Class FTN UPLC with Autosampler (Waters Corporation, Milford, MA, USA).
3. Methods
Carry out procedures at room temperature.
3.1. Sample Preparation – Cultured Cells
-
1
Pellet a minimum of 106 cells in Pyrex glass centrifuge tube at 104 x g (750 rpm) for 5 minutes at 4°C (Sorvall® Legend® X1R Centrifuge) and keep samples on ice (see Note 1).
-
2
Lyse in 300 μL chilled 1x PBS using cold sonication for 30 minutes.
-
3
Conduct protein quantitation assay in a 96-well microplate (BioTek® Synergy® HTX Microplate Reader) (see Note 2).
-
4
Store at −80°C or perform lipid extraction immediately.
3.1. Sample Preparation - Tissues
-
5
Measure tissue weight (see Note 2).
-
6
Add 4 mL Folch solution.
-
7
Disrupt and homogenize tissue for 30 seconds using a tissue homogenizer (Polytron® PT 1200 Kinematica).
-
8
Add 1 mL NaCl solution.
-
9
Continue to step 2 of “Folch Lipid Extraction”.
3.2. Folch Lipid Extraction
Add 4 mL Folch solution and 1 mL NaCl solution to cell lysates.
If including lipid internal standard, add to sample (see Note 3).
Vortex for 30 seconds.
Centrifuge at 1660 x g (3000 rpm) for 5 minutes at 4°C.
Using a 9” glass Pasteur pipet with a mechanical pipet pump, pipette through the upper, aqueous phase and protein layer and transfer approximately 2 mL of the lower organic layer to a 10 mL glass tube.
Dry lipid extracts with a speed vacuum concentrator (Thermo Fisher Savant SpeedVac® SC210A).
Reconstitute in 300 μL CH2Cl2 for cells and tissue samples with weights below 100 mg. Reconstitute heavier tissue samples in 1 mL CH2Cl2 and transfer to HPLC vials with screw caps. Store at −80°C until HILIC-MS analysis.
3.3. Data Acquisition
To prepare samples for analysis, make a 1:4 dilution of lipid extracts in an LC glass autosampler vial. Adjust dilutions according to instrument signal intensity (see Note 4).
Prepare a pooled QC sample, combining an equal volume from each final sample into a separate LC autosampler vial.
Equilibrate LC column and instrument (see Table 4–6). The pressure should be around 4000 psi when the 0.5 mL/min flow rate is established.
Perform detector setup through Intellistart interface (see Note 5).
Mass calibration. Infuse sodium formate at 20 μL/min and perform mass calibration through Intellistart interface for both positive and negative modes.
Switch instrument to Mobility TOF mode and allow the IM cell to equilibrate for an hour.
CCS calibration. Direct infuse PC and PE calibrant mixtures at 15 μL/min and calibrate through Intellistart interface (see Figure 3).
LockSpray Source Setup to check calibration curve quality (checks that it gives the right CCS for LeuEnk based on the observed DT).
Inject the lipid extract mix (see Figures 1 and 2 for example chromatograms), standard mix and a blank sample.
Randomize the sample list and increment a QC sample every 10 injections.
Table 4.
HILIC-IM-MS parameters: LC conditions.
LC Method | |
Injection volume | 5 μL for positive mode, 10 μL for negative mode |
Flow rate | 0.5 mL/min |
Column temperature | 40°C |
Sample temperature | 6°C |
Mobile phase gradient | 0–1 min: Hold at 100% HILIC A 4 min: Hold at 90% HILIC A 7–8 min: Linear gradient to 70% HILIC A 9–12 min: Equilibrate at 100% HILIC A |
Table 6.
Ion mobility parameters.
IM/MS Conditions | |
IMS wave velocity | 500 m/s |
IMS wave height | 40.0 V |
Trap wave velocity | 311 m/z |
Trap wave height | 4.0 V |
Trap collision Energy | 4.0 eV |
Transfer wave velocity | 380 m/s |
Transfer wave height | 4.0 V |
Transfer collision energy, function 1 (MS) | 2.0 eV |
Transfer collision energy, function 2 (MSE) | Low, 35 eV High, 45 eV |
Trap Gas Flow | 2 mL/min |
Helium Cell Gas Flow | 180 mL/min |
IMS Gas Flow | 90 mL/min |
Figure 3 –
(A) Example calibration plots and (B) corresponding residual plots for CCS calibration in both positive and negative ionization modes using PC and PE calibrant mixtures, respectively.
Figure 1 –
(A) Drift time versus m/z spectrum showing IM profile of lipids. (B) Chromatogram of the most intense lipids from individual lipid subclasses that are present in the mammalian lipid extract mixture. Ceramide (Cer) d36:1; Diacylglycerol (DG) 36:4; Hexosylceramide (HexCer) d42:2; Phosphatidylglycerol (PG) 34:1; Phosphatidylinositol (PI) 38:4; Phosphatidylethanolamine (PE) 36:1; LysoPE 18:0; Phosphatidylcholine (PC) 34:1; Sphingomyelin (SM) d36:1; LysoPC 16:0.
Figure 2 –
(A) Chromatogram showing separation of phosphatidyl and plasmenyl PEs. M+H adducts shown. (B) Chromatogram showing separation of PGs by fatty acyl chain composition. From left to right: PG 38:5; PG 38:4; PG 36:3; PG 36:2; PG 34:3; PG 34:1; PG 32:1. M+Na adducts shown.
3.4. Data Processing and Analysis (see Figure 4)
Import .raw files into Progenesis® QI (Nonlinear Dynamics) and process data in the chromatographic region between 0.4 and 8.5 minutes: data alignment to a reference QC sample, automatic peak detection of profile (continuum) data, and normalization to the external standard (such as tissue weight or protein weight).
Perform multivariate statistical analysis of the data (EZInfo).
If an internal standard was used, normalize the signals to the signal intensity of the appropriate internal standard in addition to external normalization.
Determine features that are significantly different between groups.
Identify features by matching retention time to the lipid classes, m/z to the METLIN, LIPID MAPS, and LipidBlast databases, and CCS values to LipidCCS/LipidIMMS [28, 29] and our own lipid CCS database [14, 15].
Table 5.
MS source conditions.
Source Conditions | |
Mode | Resolution |
Mass range | 50–1200 m/z |
Capillary voltage | +2.5 kV for positive, −2.0 kV for negative |
Sampling cone | 40 V |
Source offset | 80 V |
Source temperature | 150°C |
Desolvation temperature | 500°C |
Desolvation gas flow | 1000 L/hr |
Cone gas flow | 100 L/hr |
Nebulizer gas flow | 5.0 bar |
Acknowledgements
This work was supported by grants from National Institutes of Health (R00HD073270 and R01HD092659). AL is an appointed trainee of the Pharmacological Sciences Training Program funded by the National Institutes of Health (T32GM007750).
Footnotes
Glass tubes and pipettes are recommended when working with chloroform to prevent leaching of plasticizers, typically eluting between 2–3 minutes, which will interfere with data analysis.
Data should be normalized to protein weight or tissue weight, and internal standard (if used) post-data acquisition.
An example internal standard is d17:1/12:0 sphingo PE, N-lauroyl-D-erythro-sphingosyl phosphoethanolamine, catalog no. 860529. Please refer to the original paper (see ref. [12]) for more information on quantitative analysis using a single internal lipid standard.
Signal intensity of TIC should be on the order of 107 to ensure detection of lower concentration lipids. Preparing samples at two concentrations may be necessary to get the best results for both high and low abundance lipids.
LockSpray should be utilized (leucine-enkephalin; positive mode: m/z 566.2771, CCS 226.0; negative mode: m/z 564.2615, CCS 224.0) to correct shifts in masses and drift times throughout the run.
References
- 1.Jurowski K, Kochan K, Walczak J, Baranska M, Piekoszewski W, and Buszewski B (2017) Analytical Techniques in Lipidomics: State of the Art, Crit Rev Anal Chem 47, 418–437. [DOI] [PubMed] [Google Scholar]
- 2.Wishart DS, Knox C, Guo AC, Eisner R, Young N, Gautam B, Hau DD, Psychogios N, Dong E, Bouatra S, Mandal R, Sinelnikov I, Xia J, Jia L, Cruz JA, Lim E, Sobsey CA, Shrivastava S, Huang P, Liu P, Fang L, Peng J, Fradette R, Cheng D, Tzur D, Clements M, Lewis A, De Souza A, Zuniga A, Dawe M, Xiong Y, Clive D, Greiner R, Nazyrova A, Shaykhutdinov R, Li L, Vogel HJ, and Forsythe I (2009) HMDB: a knowledgebase for the human metabolome, Nucleic Acids Res 37, D603–610. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Rustam YH, and Reid GE (2018) Analytical Challenges and Recent Advances in Mass Spectrometry Based Lipidomics, Anal. Chem 90, 374–397. [DOI] [PubMed] [Google Scholar]
- 4.Fahy E, Subramaniam S, Brown HA, Glass CK, Merrill AH Jr., Murphy RC, Raetz CR, Russell DW, Seyama Y, Shaw W, Shimizu T, Spener F, van Meer G, VanNieuwenhze MS, White SH, Witztum JL, and Dennis EA (2005) A comprehensive classification system for lipids, J. Lipid Res 46, 839–861. [DOI] [PubMed] [Google Scholar]
- 5.Fahy E, Subramaniam S, Murphy RC, Nishijima M, Raetz CR, Shimizu T, Spener F, van Meer G, Wakelam MJ, and Dennis EA (2009) Update of the LIPID MAPS comprehensive classification system for lipids, J. Lipid Res 50 Suppl, S9–14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Han X, and Gross RW (2005) Shotgun lipidomics: electrospray ionization mass spectrometric analysis and quantitation of cellular lipidomes directly from crude extracts of biological samples, Mass Spectrom. Rev 24, 367–412. [DOI] [PubMed] [Google Scholar]
- 7.Schwudke D, Liebisch G, Herzog R, Schmitz G, and Shevchenko A (2007) Shotgun lipidomics by tandem mass spectrometry under data-dependent acquisition control, Methods Enzymol. 433, 175–191. [DOI] [PubMed] [Google Scholar]
- 8.Quehenberger O, et al. (2010) Lipidomics reveals a remarkable diversity of lipids in human plasma, J. Lipid Res 51, 3299–3305. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Merrill AH Jr., Sullards MC, Allegood JC, Kelly S, and Wang E (2005) Sphingolipidomics: high-throughput, structure-specific, and quantitative analysis of sphingolipids by liquid chromatography tandem mass spectrometry, Methods 36, 207–224. [DOI] [PubMed] [Google Scholar]
- 10.Ivanova PT, Milne SB, Byrne MO, Xiang Y, and Brown HA (2007) Glycerophospholipid identification and quantitation by electrospray ionization mass spectrometry, Methods Enzymol. 432, 21–57. [DOI] [PubMed] [Google Scholar]
- 11.Baker PR, Armando AM, Campbell JL, Quehenberger O, and Dennis EA (2014) Three-dimensional enhanced lipidomics analysis combining UPLC, differential ion mobility spectrometry, and mass spectrometric separation strategies, J. Lipid Res, Epub ahead of print. doi: 10.1194/jlr.D051581. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Cifkova E, Holcapek M, Lisa M, Ovcacikova M, Lycka A, Lynen F, and Sandra P (2012) Nontargeted quantitation of lipid classes using hydrophilic interaction liquid chromatography-electrospray ionization mass spectrometry with single internal standard and response factor approach, Anal. Chem 84, 10064–10070. [DOI] [PubMed] [Google Scholar]
- 13.Hinz C, Liggi S, and Griffin JL (2018) The potential of Ion Mobility Mass Spectrometry for high-throughput and high-resolution lipidomics, Curr Opin Chem Biol 42, 42–50. [DOI] [PubMed] [Google Scholar]
- 14.Hines K, Herron J, and Xu L (2017) Assessment of Altered Lipid Homeostasis by HILIC-Ion Mobility-Mass Spectrometry-Based Lipidomics, J. Lipid Res 58, 809–819. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Hines KM, Waalkes A, Penewit K, Holmes EA, Salipante SJ, Werth BJ, and Xu L (2017) Characterization of the Mechanisms of Daptomycin Resistance among Gram-Positive Bacterial Pathogens by Multidimensional Lipidomics, mSphere 2, e00492–00417. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kliman M, May JC, and McLean JA (2011) Lipid analysis and lipidomics by structurally selective ion mobility-mass spectrometry, Biochim. Biophys. Acta, Mol. Cell Biol. Lipids 1811, 935–945. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Fenn LS, and McLean JA (2008) Biomolecular structural separations by ion mobility-mass spectrometry, Anal. Bioanal. Chem. 391, 905–909. [DOI] [PubMed] [Google Scholar]
- 18.McLean JA, Ruotolo BT, Gillig KJ, and Russell DH (2005) Ion mobility-mass spectrometry: a new paradigm for proteomics, Int. J. Mass Spectrom 240, 301–315. [Google Scholar]
- 19.Mason EA, and Schamp HW Jr (1958) Mobility of gaseous lons in weak electric fields, Ann. Phys 4, 233–270. [Google Scholar]
- 20.Mason EA, and McDaniel EW (1988) Transport properties of ions in gases, Wiley, New York. [Google Scholar]
- 21.Hines KM, May JC, McLean JA, and Xu L (2016) Evaluation of Collision Cross Section Calibrants for Structural Analysis of Lipids by Traveling Wave Ion Mobility-Mass Spectrometry, Anal. Chem 88, 7329–7336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Ruotolo BT, Benesch JL, Sandercock AM, Hyung SJ, and Robinson CV (2008) Ion mobility-mass spectrometry analysis of large protein complexes, Nat. Protoc 3, 1139–1152. [DOI] [PubMed] [Google Scholar]
- 23.Forsythe JG, Petrov AS, Walker CA, Allen SJ, Pellissier JS, Bush MF, Hud NV, and Fernandez FM (2015) Collision cross section calibrants for negative ion mode traveling wave ion mobility-mass spectrometry, Analyst 140, 6853–6861. [DOI] [PubMed] [Google Scholar]
- 24.Fenn LS, Kliman M, Mahsut A, Zhao SR, and McLean JA (2009) Characterizing ion mobility-mass spectrometry conformation space for the analysis of complex biological samples, Anal. Bioanal. Chem 394, 235–244. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.May JC, Goodwin CR, Lareau NM, Leaptrot KL, Morris CB, Kurulugama RT, Mordehai A, Klein C, Barry W, Darland E, Overney G, Imatani K, Stafford GC, Fjeldsted JC, and McLean JA (2014) Conformational ordering of biomolecules in the gas phase: nitrogen collision cross sections measured on a prototype high resolution drift tube ion mobility-mass spectrometer, Anal. Chem 86, 2107–2116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Paglia G, Angel P, Williams JP, Richardson K, Olivos HJ, Thompson JW, Menikarachchi L, Lai S, Walsh C, Moseley A, Plumb RS, Grant DF, Palsson BO, Langridge J, Geromanos S, and Astarita G (2015) Ion mobility-derived collision cross section as an additional measure for lipid fingerprinting and identification, Anal. Chem 87, 1137–1144. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Kyle JE, Zhang X, Weitz KK, Monroe ME, Ibrahim YM, Moore RJ, Cha J, Sun X, Lovelace ES, Wagoner J, Polyak SJ, Metz TO, Dey SK, Smith RD, Burnum-Johnson KE, and Baker ES (2016) Uncovering biologically significant lipid isomers with liquid chromatography, ion mobility spectrometry and mass spectrometry, Analyst 141, 1649–1659. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Zhou Z, Tu J, Xiong X, Shen X, and Zhu ZJ (2017) LipidCCS: Prediction of Collision Cross-Section Values for Lipids with High Precision To Support Ion Mobility-Mass Spectrometry-Based Lipidomics, Anal. Chem 89, 9559–9566. [DOI] [PubMed] [Google Scholar]
- 29.Zhou Z, Shen X, Chen X, Tu J, Xiong X, and Zhu ZJ (2018) LipidIMMS Analyzer: Integrating multi-dimensional information to support lipid identification in ion mobility - mass spectrometry based lipidomics, Bioinformatics, In Press. doi: 10.1093/bioinformatics/bty1661. [DOI] [PubMed] [Google Scholar]