Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2020 Jun 1;15(6):e0233894. doi: 10.1371/journal.pone.0233894

Balloon-expandable versus self-expanding transcatheter aortic valve replacement for bioprosthetic dysfunction: A systematic review and meta-analysis

Hsiu-An Lee 1,2, An-Hsun Chou 3, Victor Chien-Chia Wu 4, Dong-Yi Chen 4, Hsin-Fu Lee 4, Kuang-Tso Lee 4, Pao-Hsien Chu 4, Yu-Ting Cheng 1, Shang-Hung Chang 4, Shao-Wei Chen 1,5,*
Editor: Corstiaan den Uil6
PMCID: PMC7263630  PMID: 32479546

Abstract

Background

Transcatheter aortic valve-in-valve (VIV) procedure is a safe alternative to conventional reoperation for bioprosthetic dysfunction. Balloon-expandable valve (BEV) and self-expanding valve (SEV) are the 2 major types of devices used. Evidence regarding the comparison of the 2 valves remains scarce.

Methods

A systematic review and meta-analysis was conducted to compare the outcomes of BEV and SEV in transcatheter VIV for aortic bioprostheses dysfunction. A computerized search of Medline, PubMed, Embase, and Cochrane databases was performed. English-language journal articles reporting SEV or BEV outcomes of at least 10 patients were included.

Results

In total, 27 studies were included, with 2,269 and 1,671 patients in the BEV and SEV groups, respectively. Rates of 30-day mortality and stroke did not differ significantly between the 2 groups. However, BEV was associated with significantly lower rates of postprocedural permanent pacemaker implantation (3.8% vs. 12%; P < 0.001). Regarding echocardiographic parameters, SEV was associated with larger postprocedural effective orifice area at 30 days (1.53 cm2 vs. 1.23 cm2; P < 0.001) and 1 year (1.55 cm2 vs. 1.22 cm2; P < 0.001).

Conclusions

For patients who underwent transcatheter aortic VIV, SEV was associated with larger postprocedural effective orifice area but higher rates of permanent pacemaker implantation. These findings provide valuable information for optimizing device selection for transcatheter aortic VIV.

1. Introduction

The use of bioprosthetic valves in surgical aortic valve replacement (AVR) has increased considerably during the last few decades [1], particularly in middle-aged patients, largely driven by patients’ wish of avoiding lifelong anticoagulation. However, bioprosthesis degenerates, requiring reoperation, which remains a relatively high risk. The evolution of transcatheter aortic valve replacement (TAVR) has enabled a safe and feasible alternative, the transcatheter valve-in-valve (VIV) procedure, which is less invasive than conventional redo surgery and has comparable outcomes [26]. Considering the possibility of future transcatheter VIV, the trend of increasing use of bioprostheses in surgical AVR is likely to persist, and the need of aortic VIV is expected to grow exponentially in the future.

Balloon-expandable valve (BEV) and self-expanding valve (SEV) are the two major types of transcatheter heart valves (THVs). These two THV types are different in valve height, implantation depth, relative position of the valve and the annulus, radial force, deployment mechanism, and valve geometry and therefore may result in different outcomes and rates of complication, such as postprocedural transvalvular pressure gradient, conduction block, or paravalvular leak (PVL). Currently, there is no randomized study comparing the two types of THVs, and only few observational studies have been published, with the observation that SEV was associated with better postprocedural hemodynamic performance but higher rates of postprocedural permanent pacemaker (PPM) implantation and aortic regurgitation [7, 8]. However, the most recent publication is a single-center study with limited number of patients and thus may not represent the whole population well [7]. Large cohort studies exist but are relatively outdated [8, 9]. Hence, a meta-analysis of the most recent studies is warranted to guide physicians in selecting the optimal device for VIV candidates.

2. Material and methods

We conducted this systematic review and meta-analysis in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. A PRISMA checklist used for this review is provided in the S1 Table. The study has been registered on PROSPERO (CRD42018111178).

2.1. Literature search

We performed a computer search of the Medline, PubMed, Embase, and Cochrane databases using the following keywords: “transcatheter”, “aortic”, “valve”, “failed”, “failing”, “degenerated”, “degeneration”, “degenerative”, “deterioration”, and “valve in valve”. The detailed search strategy is provided in the S1 Appendix. All relevant studies published until April 2020 were identified. Review articles and meta-analyses were screened for additional studies from the cited references. The processes of searching and reviewing were independently performed by 2 evaluators (H.-A. Lee and S.-W. Chen). Discrepancies were discussed to achieve a consensus.

2.2. Study selection

Inclusion criteria were as follows: (1) original article with full-length content available in English, (2) at least 10 patients who underwent aortic VIV procedures for failed surgical aortic bioprosthesis using either Edwards Lifesciences or Medtronic THVs were enrolled, and (3) results of patients who underwent aortic VIV procedures with BEV or SEV were reported. Studies were excluded if they met any of the following conditions: (1) study population overlapped with another study, including subgroup studies of a main study; (2) devices other than Medtronic valves (Medtronic, Minneapolis, MN) and Edwards Lifesciences valves (Edwards Lifesciences, Irvine, CA) were used; and (3) VIV for failed THVs. If studies were suspected of involving an overlapping cohort, only data of the most recent publication were included for analysis.

2.3. Data extraction

Data extracted were characteristics of the enrolled studies and characteristics of patients reported, including baseline information and outcomes. Study-level characteristics included year of publication, study period, location of the study conducted, number of hospitals, and number of patients included. Baseline patient-level information included age, Society of Thoracic Surgery (STS) score, European System for Cardiac Operative Risk Evaluation (EuroSCORE) II, logistic EuroSCORE, comorbidities, left ventricular ejection fraction, devices used, and characteristics of previous bioprosthesis. Thirty-day and 1-year outcomes were extracted, including death of any cause, cardiovascular death, stroke, coronary artery obstruction, major vascular complications, PPM implantation, major or life-threatening bleeding, acute kidney injury, second valve required, conversion to traditional surgery, and hemodynamics of the implanted valves.

2.4. Quality assessment

The Newcastle–Ottawa Scale (NOS) [10] was used to assess the quality of included studies, with scores ranging from 0 (lowest quality) to 8 (highest quality). Two reviewers (H.-A. Lee and S.-W. Chen) assessed the scores of each study separately; disagreements between the 2 reviewers were discussed until a consensus was achieved.

2.5. Statistical analysis

The estimates of primary and secondary outcomes derived from individual studies for each arm (Medtronic or Edwards Lifesciences valves) were pooled using the random-effects model. In contrast to the fixed-effects model, a random-effects model enables the true underlying effect to vary among individual studies. I2 values >25%, >50%, and >75% were considered to represent low, moderate, and high heterogeneity across the studies, respectively [11]. The pooled estimates between the BEV and SEV were compared using the mixed-effects model. In a further subgroup analysis, we compared outcomes between the Evolut R (Medtronic) and Sapien 3 (Edwards Lifesciences) valves. Statistical significance was set at P < 0.05 with a two-tailed test. Data were analyzed using the software Comprehensive Meta-Analysis (version 3.3; Biostat, Inc., Englewood, NJ, USA).

3. Results

3.1. Literature search

The literature screened, excluded, reviewed, and included for analysis is illustrated in Fig 1. Of the 398 articles yielded by computer search, 293 were excluded after titles and abstracts were screened. Full texts of 105 articles were reviewed to evaluate eligibility; of them, 5 were excluded because they were meta-analysis or review articles, 12 because their case numbers were <10, 9 because they included duplicated cohorts, and 52 because they did not report outcomes of patients who underwent VIV with BEV or SEV. Hence, 27 studies were included for the final quantitative meta-analysis [7, 8, 1236]. All 27 studies were observational. Five of the studies reported outcomes of both BEV and SEV, while the other 22 studies enrolled only 1 of the 2 types of THV. Basic information of the 27 studies is shown in Table 1. Three studies derived from Valve-In-Valve International Database were included because each of them has data that was not reported in the other articles. For items that were reported by more than 1 of the studies, only those reported by the latest publication were included in our analysis. Quality assessment was performed using the NOS, with scores of the 27 studies ranging 5–9 points (S2 Table).

Fig 1. Flow diagram depicting study selection process.

Fig 1

Table 1. Study data.

First author Year Valve types Study type Locations/country No. of centers study period Patient number
Woitek [34] 2020 BEV, SEV Single center Germany 1 2006–2017 146
Ribeiro [36] 2018 BEV, SEV Multi-center Global 135 2007–2014 1324
Ochiai [7] 2018 BEV, SEV Single center California, USA 1 2012–2017 74
Dvir [8] 2014 BEV, SEV Multi-center Global 55 2007–2013 459
Ihlberg [22] 2013 BEV, SEV Multi-center Nordic 11 2008–2012 45
Stankowski [29] 2020 SEV Single center Germany 1 2003–2018 68
Pascual [33] 2019 SEV Single center Spain 1 2012–2017 45
Schwerg [13] 2018 SEV Single center Germany 1 2013–2017 26
Scholtz [14] 2018 SEV Single center Germany 1 2009–2016 37
Sang [31] 2018 SEV Single center Michigan, USA 1 2014–1016 22
Deeb [17] 2017 SEV Multi-center USA NA 2013–2015 227
Chhatriwalla [18] 2017 SEV Single center Michigan, USA 9 NA 12
Duncan [20] 2015 SEV Single center UK 1 2009–2014 22
Ong [23] 2012 SEV Multi-center Germany 3 NA 18
Linke [24] 2012 SEV Single center Germany 1 NA 27
Bedogni [27] 2011 SEV Multi-center Italy 8 NA 25
Murdoch [30] 2020 BEV Multi-center Global 46 2012–2015 339
Stankowski [32] 2019 BEV Single center Germany 1 2010–2018 27
Seiffert [12] 2018 BEV Multi-center Global NA NA 514
Webb [15] 2017 BEV Multi-center Worldwide 34 2012–2014 365
Nielsen-Kudsk [16] 2017 BEV Single center Denmark 1 2015–2017 10
Ye [19] 2015 BEV Single center Canada 1 2007–2013 42
Bapat [21] 2014 BEV Single center UK 1 2010–2014 10
Seiffert [35] 2012 BEV Single center Germany 1 2008–2011 11
Bapat [25] 2012 BEV Single center UK 1 2009–2011 23
Pasic [26] 2011 BEV Single center Germany 1 NA 14
Kempfert [28] 2010 BEV Single center Germany 1 2007–2009 11

Basic information of studies included in the meta-analysis.

BEV, balloon-expandable valve; SEV, self-expanding valve

3.2. Baseline and procedural characteristics

Table 2 shows the pooled baseline and procedural characteristics of all patients in the 27 included studies. A total of 2,269 and 1,671 patients in the BEV and SEV groups were included. Mean age (78.0 ± 1.6 years in BEV vs. 75.6 ± 10.0 years in SEV), STS score (9.0 ± 2.5 in BEV vs. 9.2 ± 2.2 in SEV), left ventricular ejection fraction (50.0 ± 2.7% in BEV vs. 51.1 ± 3.0% in SEV), and other baseline echocardiographic parameters appeared to be similar between the 2 groups. The proportion of small degenerated surgical bioprostheses (≤21mm) appeared slightly lower in BEV (25.6%) than in SEV (30.7%) groups; however, the proportion of small THVs (≤23 mm) used was much higher in the BEV group (67.5%) than in the SEV group (26.9%). Transfemoral access was more frequently used in the SEV group (95%) than in the BEV group (61.3%).

Table 2. Baseline and procedural characteristics of patients (number of included studies = 27).

BEV (Edwards) SEV (Medtronic)
Variable Available data, n Weighted % or mean ± SD Available data, n Weighted % or mean ± SD
Age (year) 1087 78.0 ± 1.6 788 75.6 ± 10.0
Male (%) 1097 63.3% 751 55.3%
Log EuroSCORE 730 21.7 ± 9.9 586 26.2 ± 3.4
EuroSCORE II 53 19.6 ± 5.7 188 11.3 ± 2.9
STS score 1087 9.0 ± 2.5 678 9.2 ± 2.2
CAD (%) 491 65.2% 431 61.7%
Prior stroke (%) 1065 13.7% 666 11.7%
Prior Afib (%) 392 46.2% 467 42.0%
Prior PPM (%) 943 17.2% 377 22.5%
PAD (%) 1050 23.7% 674 22.7%
CKD (%) 1009 37.4% 674 34.3%
AR ≥moderate (%) 648 43.4% 509 54.0%
Bioprosthesis age (year) 732 10.5 ± 1.6 657 9.9 ± 1.3
Stented valve (%) 1078 84.2% 867 75.1%
Stentless valve (%) 809 13.6% 867 19.6%
Bioprosthesis size (%) 1097 746
    ≤21 mm 25.6% 30.7%
    21–24.9 mm 40.4% 37.9%
    ≥25 mm 31.4% 28.9%
    Unknown 2.3% 3.2%
Mode of failure (%)
    AS 1126 45.8% 632 53.2%
    AR 1126 28.0% 620 27.3%
    Mix 1116 26.6% 583 21.8%
LVEF (%) 829 50.0 ± 2.7 527 51.1 ± 3.0
AV area (cm2) 699 0.90 ± 0.08 671 0.95 ± 0.09
AVA index (cm2/m2) 900 0.53 ± 0.06 213 0.55
Max PG (mmHg) 341 62.6 ± 8.5 423 61.0 ± 9.2
Mean PG (mmHg) 1022 34.6 ± 3.8 754 36.0 ± 4.4
Fluoroscopic time (min) 436 18.3 ± 3.4 92 19.6 ± 9.8
THV size ≤23 mm (%) 1299 67.5% 490 26.9%
TF access (%) 1076 68.5% 561 95.0%

Abbreviations: Afib, atrial fibrillation; AR, aortic regurgitation; AS, aortic stenosis; AV, aortic valve; AVA, aortic valve area; CAD, coronary artery disease; CKD, chronic kidney disease; LVEF, left ventricular ejection fraction; PAD, peripheral artery disease; PG, pressure gradient; PPM, permanent pacemaker; PVL, paravalvular leak; STS, Society of Thoracic Surgery; TF, transfemoral; THV, transcatheter heart valve.

3.3. Clinical and echocardiographic outcomes

The event rates of all-cause mortality, cardiovascular death, and stroke at 30 days did not differ significantly between the BEV and SEV groups (Fig 2A). However, BEV was associated with significantly lower rates of major vascular complications (4.7% vs. 8.7%; P = 0.012), PPM implantation (3.8% vs. 12%; P < 0.001), and second valve requirement (2.9% vs. 6.2%; P = 0.004). One-year all-cause mortality and stroke rates were similar between the 2 groups (Fig 2B).

Fig 2.

Fig 2

Forest plot comparing 30-day (A) and 1-year (B) clinical outcomes between BEV and SEV. BEV = balloon-expandable valve; SEV = self-expanding valve.

Regarding echocardiographic outcomes, SEV was associated with better hemodynamic performance than BEV, with significantly larger postoperative effective orifice area (EOA) at 30 days (1.53 cm2 vs. 1.23 cm2; P < 0.001) and 1 year (1.55 cm2 vs. 1.22 cm2; P < 0.001; Fig 3A and 3C) and lower maximal and mean pressure gradients at 1 year (respectively, 23.0 mm Hg vs. 33.3 mm Hg, P = 0.001; and 13 mm Hg vs. 18.4 mm Hg, P = 0.002; Fig 3C).

Fig 3. Forest plot comparing echocardiographic outcomes between BEV and SEV.

Fig 3

Thirty-day continuous outcomes (A), 30-day binary outcomes (B), and 1-year continuous outcomes (C) of BEV and SEV were compared.

3.4. Subgroup analysis for newer devices

We also compared the outcomes with the Sapien 3 (Edwards Lifesciences) and Evolut R (Medtronic) valves. These are the newest generation of the 2 types of THVs with published data available for analyses. Although no statistical significance was found, Evolut R seemed to be associated with a lower mean pressure gradient than Sapien 3 (Fig 4).

Fig 4.

Fig 4

Subgroup analysis comparing 30-day outcomes of Sapien 3 and Evolut R valves for 30-day continuous outcomes (A) and 30-day binary outcomes (B).

4. Discussion

4.1. Major findings

With nearly 4000 patients included, the present meta-analysis is the largest sample used for comparing BEV and SEV outcomes in patients with failed aortic valve bioprostheses thus far. Our major findings were as follows: (1) all-cause mortality and cardiovascular death did not differ significantly between the 2 groups; (2) BEV was associated with lower rates of new PPM implantation and major vascular complications; and (3) SEV was associated with larger postprocedural EOA than BEV, both at 30 days and at 1 year.

4.2. New PPM implantation

SEV use is an independent risk factor for PPM implantation in the overall TAVR population [3739]. However, previous aortic valve procedures, including surgical AVR, seemed to be protective against post-TAVR PPM implantation [37, 39], possibly because the previously implanted bioprosthesis restricted the expansion of the THV. In the present meta-analysis, the pooled PPM implantation rate after SEV implantation was 10.7%, which is nearly 3-fold that in the BEV group (3.6%; P < 0.001). In other words, even under the potential protection of the old bioprosthesis, SEV is still associated with significantly higher risk of postprocedural PPM implantation. This finding is consistent with previous studies focusing on aortic VIV procedure [8, 9].

4.3. EOA

A major concern of the aortic VIV procedure is the relatively small postprocedural aortic valve area and high transvalvular pressure gradient, mainly resulting from restricted expansion of the THVs by the old valves. Several publications, including studies using an in vitro model [40, 41], large cohort studies [8, 9], and propensity-matched analysis [7], reported that SEV was associated with larger postprocedural aortic valve area and lower transvalvular gradient than BEV after aortic VIV procedures. The current meta-analysis further supported these findings in the largest sample size to date.

In the SEV we analyzed, the functioning part is positioned above the aortic annulus (i.e., the “supra-annular design,” which is thought to lessen the detrimental impact on postprocedural EOA by the old valve). The theory was supported by a study using in vitro model in which researchers found that when the CoreValve was positioned deeper than normal, the leaflets were more constrained, and EOA decreased; and when the SAPIEN was placed more supra-annularly, the leaflets expanded more completely, and postprocedural EOA became larger [41].

One may argue that the higher percentage of small THVs (≤23mm) used in the BEV group alone can explain the smaller postprocedural EOA in BEV. However, the proportions of small degenerated surgical bioprostheses (≤21mm) were similar between the 2 groups (Table 2), so why were small THVs more often used in the BEV group? We believe that the supra-annular design of Medtronic SEV allows a relatively larger size, while the intra-annular design of Edward BEV results in marked leaflet distortion if the size is too large [41]. According to the ViV Aortic app, for 19 or 21mm degenerated bioprostheses, a 23mm Medtronic THV, or a 20mm Edward THV is suggested. A study using the Valve-in-Valve International Data Registry also found that elevated postprocedural pressure gradient were more common after BEV-VIV implantation than after SEV-VIV; for small surgical valves (internal diameter < 20mm) and intermediate-sized valves (internal diameter ≥20mm and <23mm) [8].

Insufficient EOA and elevated transvalvular pressure gradient not only diminish patients’ physical activity and quality of life but also predict early structural valve degeneration in bioprosthetic heart valves [42]. In addition, incomplete THV expansion itself leads to localized high stress within the leaflets, which may accelerate valve degeneration [43].

4.4. SEV versus BEV

According to the present meta-analysis, SEV was associated with significantly better postprocedural EOA, which can reduce the risk of patient–prosthesis mismatch and improve quality of life, particularly in patients with larger body size or whose old bioprosthesis is small. Lower transvalvular gradient and better THV expansion may also lead to superior durability of the THV, which is important in patients with life expectancy of 20 years or longer. Nevertheless, higher EOA and lower gradient of SEV did not translate in to lower mortality. Moreover, SEV was associated with higher rates of postprocedural PPM implantation, which is detrimental to late outcome [37, 44].

Therefore, CoreValve may be beneficial in patients whose previous surgical valve is small and those at high risk of patient–prosthesis mismatch. However, Edwards valves may be preferred to Medtronic valves for patients with adequate surgical valve size, particularly those who are prone to encounter postprocedural PPM implantation or PVL, including patients who are older [37] and those who have prior conduction disturbances [38] or a prolonged PR interval [45]. For every transcatheter aortic VIV candidate, particularly younger patients, the valve selection decision should be made carefully after thorough consideration of device characteristics and patient condition and preference, as well as detailed explanation and discussion.

4.5. Study limitations

The study has several limitations. First, this meta-analysis was based on published articles; therefore, data quality and availability are limited. Second, owing to a lack of randomized controlled trials in this area, all studies included were observational, so our results can only be interpreted as “associations,” rather than as “causations.” However, the absence of randomized studies warrants the present meta-analysis to help in optimizing device selection. Third, THV devices continue advancing rapidly, so the outcomes of the present study may differ from those of the newest device.

5. Conclusion

The present systematic review and meta-analysis found that for patients who underwent transcatheter aortic VIV, SEV was associated with significantly larger postprocedural EOA but higher rates of PPM implantation and PVL of moderate or higher degree. These findings provide valuable information in guiding proper management for patients with degenerated aortic bioprostheses.

Supporting information

S1 Appendix. Detailed search strategy.

(DOCX)

S1 Table. Prisma 2009 checklist.

(DOC)

S2 Table. Newcastle-Ottawa Scale quality assessment of included studies.

(DOCX)

Acknowledgments

The authors thank Alfred Hsing-Fen Lin and Zoe Ya-Jhu Syu for their assistance with the statistical analysis.

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

This work was supported by a grant from Chang Gung Memorial Hospital, Taiwan CMRPG3H1511 (SWC), CMRPG 3J0661 (SWC), and BMRPD95 (SWC). This work was also supported by Ministry of Science and Technology grants MOST 107-2314-B-182A-152 and MOST 108-2314-B-182A-141 (SWC). The authors are thankful for the statistical assistance provided by and acknowledge the support of the Maintenance Project of the Center for Big Data Analytics and Statistics (grant CLRPG3D0045) at Chang Gung Memorial Hospital for statistical consultation and data analysis. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Isaacs AJ, Shuhaiber J, Salemi A, Isom OW, Sedrakyan A. National trends in utilization and in-hospital outcomes of mechanical versus bioprosthetic aortic valve replacements. J Thorac Cardiovasc Surg. 2015;149(5):1262–9 e3. Epub 2015/03/21. 10.1016/j.jtcvs.2015.01.052 . [DOI] [PubMed] [Google Scholar]
  • 2.Tam DY, Vo TX, Wijeysundera HC, Dvir D, Friedrich JO, Fremes SE. Transcatheter valve-in-valve versus redo surgical aortic valve replacement for the treatment of degenerated bioprosthetic aortic valve: A systematic review and meta-analysis. Catheter Cardiovasc Interv. 2018;92(7):1404–11. Epub 2018/07/20. 10.1002/ccd.27686 . [DOI] [PubMed] [Google Scholar]
  • 3.Takagi H, Mitta S, Ando T. Meta-analysis of Valve-in-Valve Transcatheter versus Redo Surgical Aortic Valve Replacement. Thorac Cardiovasc Surg. 2018;(67):243–50. Epub 2018/08/17. 10.1055/s-0038-1668135 30114716 [DOI] [PubMed] [Google Scholar]
  • 4.Neupane S, Singh H, Lammer J, Othman H, Yamasaki H, Rosman HS, et al. Meta-Analysis of Transcatheter Valve-in-Valve Implantation Versus Redo Aortic Valve Surgery for Bioprosthetic Aortic Valve Dysfunction. Am J Cardiol. 2018;121(12):1593–600. Epub 2018/05/20. 10.1016/j.amjcard.2018.02.054 . [DOI] [PubMed] [Google Scholar]
  • 5.Nalluri N, Atti V, Munir AB, Karam B, Patel NJ, Kumar V, et al. Valve in valve transcatheter aortic valve implantation (ViV-TAVI) versus redo-Surgical aortic valve replacement (redo-SAVR): A systematic review and meta-analysis. J Interv Cardiol. 2018;31(5):661–71. Epub 2018/05/22. 10.1111/joic.12520 . [DOI] [PubMed] [Google Scholar]
  • 6.Gozdek M, Raffa GM, Suwalski P, Kolodziejczak M, Anisimowicz L, Kubica J, et al. Comparative performance of transcatheter aortic valve-in-valve implantation versus conventional surgical redo aortic valve replacement in patients with degenerated aortic valve bioprostheses: systematic review and meta-analysis. Eur J Cardiothorac Surg. 2018;53(3):495–504. Epub 2017/10/14. 10.1093/ejcts/ezx347 . [DOI] [PubMed] [Google Scholar]
  • 7.Ochiai T, Yoon SH, Sharma R, Miyasaka M, Nomura T, Rami T, et al. Outcomes of Self-Expanding vs. Balloon-Expandable Transcatheter Heart Valves for the Treatment of Degenerated Aortic Surgical Bioprostheses- A Propensity Score-Matched Comparison. Circ J. 2018;82(10):2655–62. Epub 2018/08/03. 10.1253/circj.CJ-18-0157 . [DOI] [PubMed] [Google Scholar]
  • 8.Dvir D, Webb JG, Bleiziffer S, Pasic M, Waksman R, Kodali S, et al. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA 2014;312(2):162–70. 10.1001/jama.2014.7246 [DOI] [PubMed] [Google Scholar]
  • 9.Dvir D, Webb J, Brecker S, Bleiziffer S, Hildick-Smith D, Colombo A, et al. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: Results from the global valve-in-valve registry. Circulation. 2012;126(19):2335–44. 10.1161/CIRCULATIONAHA.112.104505 [DOI] [PubMed] [Google Scholar]
  • 10.Wells GA, Shea B, O'Connell D, Peterson J, Welch V, Losos M, et al. The Newcastle-Ottawa Scale (NOS) for assessing the quality of nonrandomised studies in meta-analyses. Available from: http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp.
  • 11.Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. Bmj. 2003;327(7414):557–60. Epub 2003/09/06. 10.1136/bmj.327.7414.557 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Seiffert M, Treede H, Schofer J, Linke A, Wöhrle J, Baumbach H, et al. Matched comparison of next- and early-generation balloonexpandable transcatheter heart valve implantations in failed surgical aortic bioprostheses. EuroIntervention. 2018;14(4):e397–e404. 10.4244/EIJ-D-17-00546 [DOI] [PubMed] [Google Scholar]
  • 13.Schwerg M, Stangl K, Laule M, Stangl V, Dreger H. Valve in valve implantation of the CoreValve Evolut R in degenerated surgical aortic valves. Cardiol J. 2018;25(3):301–7. Epub 2018/03/24. 10.5603/CJ.a2018.0004 . [DOI] [PubMed] [Google Scholar]
  • 14.Scholtz S, Piper C, Horstkotte D, Gummert J, Ensminger SM, Borgermann J, et al. Valve-in-valve transcatheter aortic valve implantation with CoreValve/Evolut R((c)) for degenerated small versus bigger bioprostheses. J Interv Cardiol. 2018;31(3):384–90. Epub 2018/03/01. 10.1111/joic.12498 . [DOI] [PubMed] [Google Scholar]
  • 15.Webb JG, Mack MJ, White JM, Dvir D, Blanke P, Herrmann HC, et al. Transcatheter Aortic Valve Implantation Within Degenerated Aortic Surgical Bioprostheses: PARTNER 2 Valve-in-Valve Registry. J Am Coll Cardiol. 2017;69(18):2253–62. Epub 2017/05/06. 10.1016/j.jacc.2017.02.057 . [DOI] [PubMed] [Google Scholar]
  • 16.Nielsen-Kudsk JE, Andersen A, Therkelsen CJ, Christensen EH, Jensen KT, Krusell LR, et al. High-pressure balloon fracturing of small dysfunctional Mitroflow bioprostheses facilitates transcatheter aortic valve-in-valve implantation. EuroIntervention. 2017;13(9):e1020–e5. 10.4244/EIJ-D-17-00244 [DOI] [PubMed] [Google Scholar]
  • 17.Deeb GM, Chetcuti SJ, Reardon MJ, Patel HJ, Grossman PM, Schreiber T, et al. 1-Year Results in Patients Undergoing Transcatheter Aortic Valve Replacement With Failed Surgical Bioprostheses. JACC Cardiovasc Interv. 2017;10(10):1034–44. Epub 2017/05/20. 10.1016/j.jcin.2017.03.018 . [DOI] [PubMed] [Google Scholar]
  • 18.Chhatriwalla AK, Allen KB, Saxon JT, Cohen DJ, Aggarwal S, Hart AJ, et al. Bioprosthetic Valve Fracture Improves the Hemodynamic Results of Valve-in-Valve Transcatheter Aortic Valve Replacement. Circ Cardiovasc Interv. 2017;10(7). Epub 2017/07/13. 10.1161/circinterventions.117.005216 . [DOI] [PubMed] [Google Scholar]
  • 19.Ye J, Cheung A, Yamashita M, Wood D, Peng D, Gao M, et al. Transcatheter Aortic and Mitral Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves: An 8-Year Single-Center Experience. JACC Cardiovasc Interv. 2015;8(13):1735–44. Epub 2015/10/20. 10.1016/j.jcin.2015.08.012 . [DOI] [PubMed] [Google Scholar]
  • 20.Duncan A, Davies S, Di Mario C, Moat N. Valve-in-valve transcatheter aortic valve implantation for failing surgical aortic stentless bioprosthetic valves: A single-center experience. J Thorac Cardiovasc Surg. 2015;150(1):91–8. Epub 2015/04/15. 10.1016/j.jtcvs.2015.03.021 . [DOI] [PubMed] [Google Scholar]
  • 21.Bapat V, Davies W, Attia R, Hancock J, Bolter K, Young C, et al. Use of balloon expandable transcatheter valves for valve-in-valve implantation in patients with degenerative stentless aortic bioprostheses: Technical considerations and results. J Thorac Cardiovasc Surg. 2014;148(3):917–22; discussion 22–4. Epub 2014/07/20. 10.1016/j.jtcvs.2014.05.029 . [DOI] [PubMed] [Google Scholar]
  • 22.Ihlberg L, Nissen H, Nielsen NE, Ruck A, Busund R, Klaarborg KE, et al. Early clinical outcome of aortic transcatheter valve-in-valve implantation in the Nordic countries. J Thorac Cardiovasc Surg. 2013;146(5):1047–54; discussion 54. Epub 2013/09/04. 10.1016/j.jtcvs.2013.06.045 . [DOI] [PubMed] [Google Scholar]
  • 23.Ong SH, Bauernschmitt R, Schuler G, Mueller R. Short- and mid-term safety and effectiveness of transcatheter aortic valve implantation in a failing surgical aortic bioprosthesis. Eur J Cardiothorac Surg. 2012;42(2):268–76; discussion 76. Epub 2012/02/16. 10.1093/ejcts/ezs014 . [DOI] [PubMed] [Google Scholar]
  • 24.Linke A, Woitek F, Merx MW, Schiefer C, Mobius-Winkler S, Holzhey D, et al. Valve-in-valve implantation of Medtronic CoreValve prosthesis in patients with failing bioprosthetic aortic valves. Circ Cardiovasc Interv. 2012;5(5):689–97. Epub 2012/10/11. 10.1161/circinterventions.112.972331 . [DOI] [PubMed] [Google Scholar]
  • 25.Bapat V, Attia R, Redwood S, Hancock J, Wilson K, Young C, et al. Use of transcatheter heart valves for a valve-in-valve implantation in patients with degenerated aortic bioprosthesis: technical considerations and results. J Thorac Cardiovasc Surg. 2012;144(6):1372–9; discussion 9–80. Epub 2012/11/13. 10.1016/j.jtcvs.2012.07.104 . [DOI] [PubMed] [Google Scholar]
  • 26.Pasic M, Unbehaun A, Dreysse S, Buz S, Drews T, Kukucka M, et al. Transapical aortic valve implantation after previous aortic valve replacement: clinical proof of the "valve-in-valve" concept. J Thorac Cardiovasc Surg. 2011;142(2):270–7. Epub 2010/11/26. 10.1016/j.jtcvs.2010.09.049 . [DOI] [PubMed] [Google Scholar]
  • 27.Bedogni F, Laudisa ML, Pizzocri S, Tamburino C, Ussia GP, Petronio AS, et al. Transcatheter valve-in-valve implantation using Corevalve Revalving System for failed surgical aortic bioprostheses. JACC Cardiovasc Interv. 2011;4(11):1228–34. Epub 2011/11/26. 10.1016/j.jcin.2011.10.002 . [DOI] [PubMed] [Google Scholar]
  • 28.Kempfert J, Van Linden A, Linke A, Borger MA, Rastan A, Mukherjee C, et al. Transapical off-pump valve-in-valve implantation in patients with degenerated aortic xenografts. Ann Thorac Surg. 2010;89(6):1934–41. Epub 2010/05/25. 10.1016/j.athoracsur.2010.02.061 . [DOI] [PubMed] [Google Scholar]
  • 29.Stankowski T, Aboul-Hassan SS, Seifi Zinab F, Herwig V, Stępiński P, Grimmig O, et al. Femoral transcatheter valve-in-valve implantation as alternative strategy for failed aortic bioprostheses: A single-centre experience with long-term follow-up. Int J Cardiol. 2020. Epub 2020/02/24. 10.1016/j.ijcard.2020.02.035 . [DOI] [PubMed] [Google Scholar]
  • 30.Murdoch DJ, Sathananthan J, Hensey M, Alu MC, Liu Y, Crowley A, et al. Mitral regurgitation in patients undergoing transcatheter aortic valve implantation for degenerated surgical aortic bioprosthesis: Insights from PARTNER 2 Valve-in-Valve Registry. Catheter Cardiovasc Interv. 2020. Epub 2020/03/03. 10.1002/ccd.28811 . [DOI] [PubMed] [Google Scholar]
  • 31.Sang SLW, Beute T, Heiser J, Berkompas D, Fanning J, Merhi W. Early Outcomes for Valve-in-valve Transcatheter Aortic Valve Replacement in Degenerative Freestyle Bioprostheses. Semin Thorac Cardiovasc Surg. 2018;30(3):262–8. Epub 2017/11/22. 10.1053/j.semtcvs.2017.11.001 . [DOI] [PubMed] [Google Scholar]
  • 32.Stankowski T, Aboul-Hassan SS, Seifi-Zinab F, Herwig V, Kubikova M, Harnath A, et al. Severe structural deterioration of small aortic bioprostheses treated with valve-in-valve transcatheter aortic valve implantation. J Card Surg. 2019;34(1):7–13. Epub 2019/01/10. 10.1111/jocs.13976 . [DOI] [PubMed] [Google Scholar]
  • 33.Pascual I, Avanzas P, Hernández-Vaquero D, Díaz R, Del Valle R, Padrón R, et al. Self-expanding transcatheter aortic valve implantation for degenerated Mitroflow bioprosthesis: Early outcomes. Int J Cardiol. 2019;287:53–8. Epub 2019/02/18. 10.1016/j.ijcard.2019.01.094 . [DOI] [PubMed] [Google Scholar]
  • 34.Woitek FJ, Stachel G, Kiefer P, Haussig S, Leontyev S, Schlotter F, et al. Treatment of failed aortic bioprostheses: An evaluation of conventional redo surgery and transfemoral transcatheter aortic valve-in-valve implantation. Int J Cardiol. 2020;300:80–6. Epub 2019/11/20. 10.1016/j.ijcard.2019.09.039 . [DOI] [PubMed] [Google Scholar]
  • 35.Seiffert M, Conradi L, Baldus S, Knap M, Schirmer J, Franzen O, et al. Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses. J Thorac Cardiovasc Surg. 2012;143(3):617–24. Epub 2011/12/16. 10.1016/j.jtcvs.2011.11.004 . [DOI] [PubMed] [Google Scholar]
  • 36.Ribeiro HB, Rodes-Cabau J, Blanke P, Leipsic J, Kwan Park J, Bapat V, et al. Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry. Eur Heart J. 2018;39(8):687–95. Epub 2017/10/12. 10.1093/eurheartj/ehx455 . [DOI] [PubMed] [Google Scholar]
  • 37.Fadahunsi OO, Olowoyeye A, Ukaigwe A, Li Z, Vora AN, Vemulapalli S, et al. Incidence, Predictors, and Outcomes of Permanent Pacemaker Implantation Following Transcatheter Aortic Valve Replacement: Analysis From the U.S. Society of Thoracic Surgeons/American College of Cardiology TVT Registry. JACC Cardiovasc Interv. 2016;9(1876–7605 (Electronic)):2189–99. [DOI] [PubMed] [Google Scholar]
  • 38.Siontis GC, Juni P, Pilgrim T, Stortecky S, Bullesfeld L, Meier B, et al. Predictors of permanent pacemaker implantation in patients with severe aortic stenosis undergoing TAVR: a meta-analysis. J Am Coll Cardiol. 2014;64(2):129–40. Epub 2014/07/12. 10.1016/j.jacc.2014.04.033 . [DOI] [PubMed] [Google Scholar]
  • 39.Ledwoch J, Franke J, Gerckens U, Kuck KH, Linke A, Nickenig G, et al. Incidence and predictors of permanent pacemaker implantation following transcatheter aortic valve implantation: analysis from the German transcatheter aortic valve interventions registry. Catheter Cardiovasc Interv. 2013;82(4):E569–77. Epub 2013/03/12. 10.1002/ccd.24915 . [DOI] [PubMed] [Google Scholar]
  • 40.Sedaghat A, Sinning JM, Utzenrath M, Ghalati PF, Schmitz C, Werner N, et al. Hydrodynamic Performance of the Medtronic CoreValve and the Edwards SAPIEN XT Transcatheter Heart Valve in Surgical Bioprostheses: An In Vitro Valve-in-Valve Model. Ann Thorac Surg. 2016;101(1):118–24. Epub 2015/09/14. 10.1016/j.athoracsur.2015.06.047 . [DOI] [PubMed] [Google Scholar]
  • 41.Midha PA, Raghav V, Condado JF, Okafor IU, Lerakis S, Thourani VH, et al. Valve Type, Size, and Deployment Location Affect Hemodynamics in an In Vitro Valve-in-Valve Model. JACC Cardiovasc Interv. 2016;9(15):1618–28. Epub 2016/08/06. 10.1016/j.jcin.2016.05.030 . [DOI] [PubMed] [Google Scholar]
  • 42.Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121(19):2123–9. Epub 2010/05/05. 10.1161/CIRCULATIONAHA.109.901272 . [DOI] [PubMed] [Google Scholar]
  • 43.Abbasi M, Azadani AN. Leaflet stress and strain distributions following incomplete transcatheter aortic valve expansion. J Biomech. 2015;48(13):3663–71. Epub 2015/09/05. 10.1016/j.jbiomech.2015.08.012 . [DOI] [PubMed] [Google Scholar]
  • 44.Arnold SV, Zhang Y, Baron SJ, McAndrew TC, Alu MC, Kodali SK, et al. Impact of Short-Term Complications on Mortality and Quality of Life After Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv. 2019;12(4):362–9. Epub 2019/02/21. 10.1016/j.jcin.2018.11.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Tichelbacker T, Bergau L, Puls M, Friede T, Mutze T, Maier LS, et al. Insights into permanent pacemaker implantation following TAVR in a real-world cohort. PLoS One. 2018;13(10):e0204503 Epub 2018/10/18. 10.1371/journal.pone.0204503 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Corstiaan den Uil

31 Mar 2020

PONE-D-20-05542

Balloon-expandable versus self-expanding transcatheter aortic valve replacement for bioprosthetic dysfunction: a systematic review and meta-Analysis

PLOS ONE

Dear Dr Chen,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

We would appreciate receiving your revised manuscript by May 15 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Corstiaan den Uil

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements:

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.plosone.org/attachments/PLOSOne_formatting_sample_main_body.pdf and http://www.plosone.org/attachments/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We noticed that the search of your systematic review was last performed in October 2018. Please ensure that the search is up to date and that the systematic review/meta-analysis includes any new studies published since then.

3. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: N/A

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: This is a very interesting review and meta-analysis of VIV-treatment. However, there are some points to discuss:

1. Comparison of EOA between SEV and BEV:

It should be better explained why the EOA of SEV is bigger than BEV. The used BEV-Size was smaller than the SEV-Sizes (THV size ≤ 23mm 65,7% vs 31,7%), this alone can explain the difference. How was the gradient-/EOA- difference in treated small surgical bioprosthesis (<21mm), are there data available?

2. There was a recent publication about measuring error of measured echo gradients in intra annular and supra annular valves (Abbas AE and Pibarot P, CCI 2019). Are there data about invasive measured gradients after VIV available.

3. For a recommendation which valve should be used in which situation, the coronary access possibility after VIV should be discussed, as there are differences between BEV and SEV.

Reviewer #2: it is a very important subject that the authors have performed a metanalysis on given the sparse RCT data comparing these 2 valves. Some questions that need to be answered:

1. Why is there such a large difference between the % of baseline >=moderate MR between the 2 groups (25.5% vs 68.3%)

2. Could the higher number of stented valves in the BEV group explain the higher PVL after the procedures?

3. Given that transfemoral access is the most common modality why were there only 67.8% in the BEV group? Especially given that alternate access is associated with higher periprocedural complication rates.

4. Higher EOA and Lower gradients have been shown in prior studies with SEV, although they did not translate into lower mortality with SEV compared to BEV. Can the authors elaborate on that in their discussion?

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Jun 1;15(6):e0233894. doi: 10.1371/journal.pone.0233894.r002

Author response to Decision Letter 0


14 May 2020

The following is a point-by-point response to reviewers comments.

_______________________________________________________________

2. We noticed that the search of your systematic review was last performed in October 2018. Please ensure that the search is up to date and that the systematic review/meta-analysis includes any new studies published since then

Response: We have updated the search to April 2020.

Reviewer #1: This is a very interesting review and meta-analysis of VIV-treatment. However, there are some points to discuss:

1. Comparison of EOA between SEV and BEV:

It should be better explained why the EOA of SEV is bigger than BEV. The used BEV-Size was smaller than the SEV-Sizes (THV size ≤ 23mm 65,7% vs 31,7%), this alone can explain the difference. How was the gradient-/EOA- difference in treated small surgical bioprosthesis (<21mm), are there data available?

Response: Thanks for the great question. Regarding gradient-/EOA- difference in treated small surgical bioprostheses, a study using the Valve-in-Valve International Data Registry found that elevated postprocedural pressure gradient (mean ≥ 20mmHg) were more common with BEVs in comparison with SEVs; for small surgical valves (internal diameter [ID] < 20mm), 41.2% vs 23.4% (P = .04) and for intermediate-sized valves (≥20 and < 23), 35.8% vs 19.4% (P = .01), respectively [1](eFigure 2C). Moreover, 11.8% of Edwards SAPIEN VIV procedures performed inside small bioprosthesis had very high postprocedural gradients (mean ≥ 40mmHg), while no cases of CoreValve VIV procedures resulted in very high gradients (P = 0.005, eFigure 2D).

We believe that the supra-annular design of Medtronic SEV allows a relatively larger size, while an intra-annular design of Edwards BEV results in marked leaflet distortion if the size is too large. This can explain why the ViV Aortic app suggests 20mm Edwards THVs inside19 or 21mm degenerated bioprosthesis, but 23mm Medtronic THVs inside same-sized bioprostheses.

In conclusion, the supra-annular design of Medtronic CoreValve allows better expansion of the leaflets, and allows a relatively larger sized THV to be inserted inside a small bioprosthesis, hence results in better EOA/gradient.

We have added the above explanation and citation in the revised manuscript.

2. There was a recent publication about measuring error of measured echo gradients in intra annular and supra annular valves (Abbas AE and Pibarot P, CCI 2019). Are there data about invasive measured gradients after VIV available.

Response: Abbas et. al. demonstrated the catheterization/echocardiography discordance after native TAVR and after Valve-in-Valve TAVR[2]. The echocardiography mean gradient is significantly higher than catheterization gradient after both procedures.

In all the studies we reviewed, only 1 reported postprocedural catheterization gradients[3]. The postprocedural mean echocardiographic gradient and mean invasive gradient were 22.8 and 13.6 mmHg, respectively (P < 0.001).

3. For a recommendation which valve should be used in which situation, the coronary access possibility after VIV should be discussed, as there are differences between BEV and SEV.

Response: Yes, the coronary access possibility after VIV is a very important concern. Allali et. al. reported their experience of coronary intervention after TAVR[4]. However, the aim of the current meta-analysis is to collect evidences and analyze. Since we did not find sufficient data regarding coronary-related outcomes after ViV to perform meta-analysis, we did not include this point in our manuscript. Nevertheless, if you consider the point suitable for our article, we would like to add it in our manuscript according to your suggestion.

_________________________________________________________________

Reviewer #2: it is a very important subject that the authors have performed a metanalysis on given the sparse RCT data comparing these 2 valves. Some questions that need to be answered:

1. Why is there such a large difference between the % of baseline >=moderate MR between the 2 groups (25.5% vs 68.3%)

Response: In the original analysis, very few study reported the percentage of baseline ≥moderate AR. A single study of high or low % made a large difference in the results. After collecting more data, the % of baseline ≥moderate AR was 43.4% and 54.0 in the BEV and SEV groups, respectively. However, it should be noted that only a small portion of studies reported baseline ≥moderate AR, so the result does not well represent the whole population.

2. Could the higher number of stented valves in the BEV group explain the higher PVL after the procedures?

Response: In studies comparing outcomes following VIV inside stentless versus stented bioprostheses, stentless bioprostheses appeared to be related to higher PVL than stented bioprostheses[5, 6]. Hence, the higher percentage of stentless valves in the SEV group in our study could contribute to the higher rates of postprocedural PVL.

However, in the revised manuscript, after updating research with more recent data included in the meta-analysis, the difference of postprocedural PVL between BEV and SEV groups was no longer significant.

3. Given that transfemoral access is the most common modality why were there only 67.8% in the BEV group? Especially given that alternate access is associated with higher periprocedural complication rates.

Response: The proportion of transfemoral (TF) access reported by the two largest registry regarding Edwards VIV, VIVID Registry and PARTNER 2 VIV Registry, were 66.7% and 75.4%, respectively[7, 8]. In VIVID Registry, the TF ratio of Sapien XT was only 58.5%. Although nowadays approximately 5% of TAVR candidates require a non-femoral access, in the earlier era, 10~20% of patients require non-femoral access because previous-generation devices had larger profile. Transapical access, which could only be performed using BEV, was the first-developed non-femoral access, and had been the most commonly used non-femoral access for quite a few years. This may explain the lower percentage of TF access in BEV group, especially in earlier era. After updating the search to April, 2020, the TF ratio of BEV group slightly increased to 68.5%

4. Higher EOA and Lower gradients have been shown in prior studies with SEV, although they did not translate into lower mortality with SEV compared to BEV. Can the authors elaborate on that in their discussion?

Response: Thanks for the suggestion. We have discussed EOA and gradients in sections 4.3 and 4.4 of discussion in the original manuscript. We also added a sentence in the first paragraph of section 4.4 to emphasize that higher EOA and lower gradient of SEV did not translate into lower mortality. If there is any further suggestion from the reviewer, we will be glad to revise our manuscript accordingly.

References

1. Dvir D, Webb JG, Bleiziffer S, Pasic M, Waksman R, Kodali S, et al. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA 2014;312(2):162-70. doi: 10.1001/jama.2014.7246.

2. Abbas AE, Mando R, Hanzel G, Gallagher M, Safian R, Hanson I, et al. Invasive Versus Echocardiographic Evaluation of Transvalvular Gradients Immediately Post-Transcatheter Aortic Valve Replacement. Circ Cardiovasc Interv. 2019;12(7):e007973. Epub 2019/07/06. doi: 10.1161/CIRCINTERVENTIONS.119.007973. PubMed PMID: 31272227.

3. Scholtz S, Piper C, Horstkotte D, Gummert J, Ensminger SM, Borgermann J, et al. Valve-in-valve transcatheter aortic valve implantation with CoreValve/Evolut R((c)) for degenerated small versus bigger bioprostheses. J Interv Cardiol. 2018;31(3):384-90. Epub 2018/03/01. doi: 10.1111/joic.12498. PubMed PMID: 29490430.

4. Allali A, El-Mawardy M, Schwarz B, Sato T, Geist V, Toelg R, et al. Incidence, feasibility and outcome of percutaneous coronary intervention after transcatheter aortic valve implantation with a self-expanding prosthesis. Results from a single center experience. Cardiovasc Revasc Med. 2016;17(6):391-8. Epub 2016/07/12. doi: 10.1016/j.carrev.2016.05.010. PubMed PMID: 27396607.

5. Duncan A, Moat N, Simonato M, de Weger A, Kempfert J, Eggebrecht H, et al. Outcomes Following Transcatheter Aortic Valve Replacement for Degenerative Stentless Versus Stented Bioprostheses. JACC Cardiovasc Interv. 2019;12(13):1256-63. Epub 2019/06/17. doi: 10.1016/j.jcin.2019.02.036. PubMed PMID: 31202944.

6. Choi CH, Cheng V, Malaver D, Kon N, Kincaid EH, Gandhi SK, et al. A comparison of valve-in-valve transcatheter aortic valve replacement in failed stentless versus stented surgical bioprosthetic aortic valves. Catheter Cardiovasc Interv. 2019;93(6):1106-15. Epub 2018/12/28. doi: 10.1002/ccd.28039. PubMed PMID: 30588736; PubMed Central PMCID: PMCPMC6590419.

7. Seiffert M, Treede H, Schofer J, Linke A, Wöhrle J, Baumbach H, et al. Matched comparison of next- and early-generation balloonexpandable transcatheter heart valve implantations in failed surgical aortic bioprostheses. EuroIntervention. 2018;14(4):e397-e404. doi: 10.4244/EIJ-D-17-00546.

8. Webb JG, Mack MJ, White JM, Dvir D, Blanke P, Herrmann HC, et al. Transcatheter Aortic Valve Implantation Within Degenerated Aortic Surgical Bioprostheses: PARTNER 2 Valve-in-Valve Registry. J Am Coll Cardiol. 2017;69(18):2253-62. Epub 2017/05/06. doi: 10.1016/j.jacc.2017.02.057. PubMed PMID: 28473128.

Attachment

Submitted filename: Response to Reviewers 20200512.docx

Decision Letter 1

Corstiaan den Uil

15 May 2020

Balloon-expandable versus self-expanding transcatheter aortic valve replacement for bioprosthetic dysfunction: a systematic review and meta-analysis

PONE-D-20-05542R1

Dear Dr. Chen,

We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.

Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.

Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

With kind regards,

Corstiaan den Uil

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Corstiaan den Uil

22 May 2020

PONE-D-20-05542R1

Balloon-expandable versus self-expanding transcatheter aortic valve replacement for bioprosthetic dysfunction: a systematic review and meta-analysis

Dear Dr. Chen:

I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

For any other questions or concerns, please email plosone@plos.org.

Thank you for submitting your work to PLOS ONE.

With kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Corstiaan den Uil

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Appendix. Detailed search strategy.

    (DOCX)

    S1 Table. Prisma 2009 checklist.

    (DOC)

    S2 Table. Newcastle-Ottawa Scale quality assessment of included studies.

    (DOCX)

    Attachment

    Submitted filename: Response to Reviewers 20200512.docx

    Data Availability Statement

    All relevant data are within the paper and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES