Abstract
Background
Transcatheter aortic valve-in-valve (VIV) procedure is a safe alternative to conventional reoperation for bioprosthetic dysfunction. Balloon-expandable valve (BEV) and self-expanding valve (SEV) are the 2 major types of devices used. Evidence regarding the comparison of the 2 valves remains scarce.
Methods
A systematic review and meta-analysis was conducted to compare the outcomes of BEV and SEV in transcatheter VIV for aortic bioprostheses dysfunction. A computerized search of Medline, PubMed, Embase, and Cochrane databases was performed. English-language journal articles reporting SEV or BEV outcomes of at least 10 patients were included.
Results
In total, 27 studies were included, with 2,269 and 1,671 patients in the BEV and SEV groups, respectively. Rates of 30-day mortality and stroke did not differ significantly between the 2 groups. However, BEV was associated with significantly lower rates of postprocedural permanent pacemaker implantation (3.8% vs. 12%; P < 0.001). Regarding echocardiographic parameters, SEV was associated with larger postprocedural effective orifice area at 30 days (1.53 cm2 vs. 1.23 cm2; P < 0.001) and 1 year (1.55 cm2 vs. 1.22 cm2; P < 0.001).
Conclusions
For patients who underwent transcatheter aortic VIV, SEV was associated with larger postprocedural effective orifice area but higher rates of permanent pacemaker implantation. These findings provide valuable information for optimizing device selection for transcatheter aortic VIV.
1. Introduction
The use of bioprosthetic valves in surgical aortic valve replacement (AVR) has increased considerably during the last few decades [1], particularly in middle-aged patients, largely driven by patients’ wish of avoiding lifelong anticoagulation. However, bioprosthesis degenerates, requiring reoperation, which remains a relatively high risk. The evolution of transcatheter aortic valve replacement (TAVR) has enabled a safe and feasible alternative, the transcatheter valve-in-valve (VIV) procedure, which is less invasive than conventional redo surgery and has comparable outcomes [2–6]. Considering the possibility of future transcatheter VIV, the trend of increasing use of bioprostheses in surgical AVR is likely to persist, and the need of aortic VIV is expected to grow exponentially in the future.
Balloon-expandable valve (BEV) and self-expanding valve (SEV) are the two major types of transcatheter heart valves (THVs). These two THV types are different in valve height, implantation depth, relative position of the valve and the annulus, radial force, deployment mechanism, and valve geometry and therefore may result in different outcomes and rates of complication, such as postprocedural transvalvular pressure gradient, conduction block, or paravalvular leak (PVL). Currently, there is no randomized study comparing the two types of THVs, and only few observational studies have been published, with the observation that SEV was associated with better postprocedural hemodynamic performance but higher rates of postprocedural permanent pacemaker (PPM) implantation and aortic regurgitation [7, 8]. However, the most recent publication is a single-center study with limited number of patients and thus may not represent the whole population well [7]. Large cohort studies exist but are relatively outdated [8, 9]. Hence, a meta-analysis of the most recent studies is warranted to guide physicians in selecting the optimal device for VIV candidates.
2. Material and methods
We conducted this systematic review and meta-analysis in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. A PRISMA checklist used for this review is provided in the S1 Table. The study has been registered on PROSPERO (CRD42018111178).
2.1. Literature search
We performed a computer search of the Medline, PubMed, Embase, and Cochrane databases using the following keywords: “transcatheter”, “aortic”, “valve”, “failed”, “failing”, “degenerated”, “degeneration”, “degenerative”, “deterioration”, and “valve in valve”. The detailed search strategy is provided in the S1 Appendix. All relevant studies published until April 2020 were identified. Review articles and meta-analyses were screened for additional studies from the cited references. The processes of searching and reviewing were independently performed by 2 evaluators (H.-A. Lee and S.-W. Chen). Discrepancies were discussed to achieve a consensus.
2.2. Study selection
Inclusion criteria were as follows: (1) original article with full-length content available in English, (2) at least 10 patients who underwent aortic VIV procedures for failed surgical aortic bioprosthesis using either Edwards Lifesciences or Medtronic THVs were enrolled, and (3) results of patients who underwent aortic VIV procedures with BEV or SEV were reported. Studies were excluded if they met any of the following conditions: (1) study population overlapped with another study, including subgroup studies of a main study; (2) devices other than Medtronic valves (Medtronic, Minneapolis, MN) and Edwards Lifesciences valves (Edwards Lifesciences, Irvine, CA) were used; and (3) VIV for failed THVs. If studies were suspected of involving an overlapping cohort, only data of the most recent publication were included for analysis.
2.3. Data extraction
Data extracted were characteristics of the enrolled studies and characteristics of patients reported, including baseline information and outcomes. Study-level characteristics included year of publication, study period, location of the study conducted, number of hospitals, and number of patients included. Baseline patient-level information included age, Society of Thoracic Surgery (STS) score, European System for Cardiac Operative Risk Evaluation (EuroSCORE) II, logistic EuroSCORE, comorbidities, left ventricular ejection fraction, devices used, and characteristics of previous bioprosthesis. Thirty-day and 1-year outcomes were extracted, including death of any cause, cardiovascular death, stroke, coronary artery obstruction, major vascular complications, PPM implantation, major or life-threatening bleeding, acute kidney injury, second valve required, conversion to traditional surgery, and hemodynamics of the implanted valves.
2.4. Quality assessment
The Newcastle–Ottawa Scale (NOS) [10] was used to assess the quality of included studies, with scores ranging from 0 (lowest quality) to 8 (highest quality). Two reviewers (H.-A. Lee and S.-W. Chen) assessed the scores of each study separately; disagreements between the 2 reviewers were discussed until a consensus was achieved.
2.5. Statistical analysis
The estimates of primary and secondary outcomes derived from individual studies for each arm (Medtronic or Edwards Lifesciences valves) were pooled using the random-effects model. In contrast to the fixed-effects model, a random-effects model enables the true underlying effect to vary among individual studies. I2 values >25%, >50%, and >75% were considered to represent low, moderate, and high heterogeneity across the studies, respectively [11]. The pooled estimates between the BEV and SEV were compared using the mixed-effects model. In a further subgroup analysis, we compared outcomes between the Evolut R (Medtronic) and Sapien 3 (Edwards Lifesciences) valves. Statistical significance was set at P < 0.05 with a two-tailed test. Data were analyzed using the software Comprehensive Meta-Analysis (version 3.3; Biostat, Inc., Englewood, NJ, USA).
3. Results
3.1. Literature search
The literature screened, excluded, reviewed, and included for analysis is illustrated in Fig 1. Of the 398 articles yielded by computer search, 293 were excluded after titles and abstracts were screened. Full texts of 105 articles were reviewed to evaluate eligibility; of them, 5 were excluded because they were meta-analysis or review articles, 12 because their case numbers were <10, 9 because they included duplicated cohorts, and 52 because they did not report outcomes of patients who underwent VIV with BEV or SEV. Hence, 27 studies were included for the final quantitative meta-analysis [7, 8, 12–36]. All 27 studies were observational. Five of the studies reported outcomes of both BEV and SEV, while the other 22 studies enrolled only 1 of the 2 types of THV. Basic information of the 27 studies is shown in Table 1. Three studies derived from Valve-In-Valve International Database were included because each of them has data that was not reported in the other articles. For items that were reported by more than 1 of the studies, only those reported by the latest publication were included in our analysis. Quality assessment was performed using the NOS, with scores of the 27 studies ranging 5–9 points (S2 Table).
Table 1. Study data.
First author | Year | Valve types | Study type | Locations/country | No. of centers | study period | Patient number |
---|---|---|---|---|---|---|---|
Woitek [34] | 2020 | BEV, SEV | Single center | Germany | 1 | 2006–2017 | 146 |
Ribeiro [36] | 2018 | BEV, SEV | Multi-center | Global | 135 | 2007–2014 | 1324 |
Ochiai [7] | 2018 | BEV, SEV | Single center | California, USA | 1 | 2012–2017 | 74 |
Dvir [8] | 2014 | BEV, SEV | Multi-center | Global | 55 | 2007–2013 | 459 |
Ihlberg [22] | 2013 | BEV, SEV | Multi-center | Nordic | 11 | 2008–2012 | 45 |
Stankowski [29] | 2020 | SEV | Single center | Germany | 1 | 2003–2018 | 68 |
Pascual [33] | 2019 | SEV | Single center | Spain | 1 | 2012–2017 | 45 |
Schwerg [13] | 2018 | SEV | Single center | Germany | 1 | 2013–2017 | 26 |
Scholtz [14] | 2018 | SEV | Single center | Germany | 1 | 2009–2016 | 37 |
Sang [31] | 2018 | SEV | Single center | Michigan, USA | 1 | 2014–1016 | 22 |
Deeb [17] | 2017 | SEV | Multi-center | USA | NA | 2013–2015 | 227 |
Chhatriwalla [18] | 2017 | SEV | Single center | Michigan, USA | 9 | NA | 12 |
Duncan [20] | 2015 | SEV | Single center | UK | 1 | 2009–2014 | 22 |
Ong [23] | 2012 | SEV | Multi-center | Germany | 3 | NA | 18 |
Linke [24] | 2012 | SEV | Single center | Germany | 1 | NA | 27 |
Bedogni [27] | 2011 | SEV | Multi-center | Italy | 8 | NA | 25 |
Murdoch [30] | 2020 | BEV | Multi-center | Global | 46 | 2012–2015 | 339 |
Stankowski [32] | 2019 | BEV | Single center | Germany | 1 | 2010–2018 | 27 |
Seiffert [12] | 2018 | BEV | Multi-center | Global | NA | NA | 514 |
Webb [15] | 2017 | BEV | Multi-center | Worldwide | 34 | 2012–2014 | 365 |
Nielsen-Kudsk [16] | 2017 | BEV | Single center | Denmark | 1 | 2015–2017 | 10 |
Ye [19] | 2015 | BEV | Single center | Canada | 1 | 2007–2013 | 42 |
Bapat [21] | 2014 | BEV | Single center | UK | 1 | 2010–2014 | 10 |
Seiffert [35] | 2012 | BEV | Single center | Germany | 1 | 2008–2011 | 11 |
Bapat [25] | 2012 | BEV | Single center | UK | 1 | 2009–2011 | 23 |
Pasic [26] | 2011 | BEV | Single center | Germany | 1 | NA | 14 |
Kempfert [28] | 2010 | BEV | Single center | Germany | 1 | 2007–2009 | 11 |
Basic information of studies included in the meta-analysis.
BEV, balloon-expandable valve; SEV, self-expanding valve
3.2. Baseline and procedural characteristics
Table 2 shows the pooled baseline and procedural characteristics of all patients in the 27 included studies. A total of 2,269 and 1,671 patients in the BEV and SEV groups were included. Mean age (78.0 ± 1.6 years in BEV vs. 75.6 ± 10.0 years in SEV), STS score (9.0 ± 2.5 in BEV vs. 9.2 ± 2.2 in SEV), left ventricular ejection fraction (50.0 ± 2.7% in BEV vs. 51.1 ± 3.0% in SEV), and other baseline echocardiographic parameters appeared to be similar between the 2 groups. The proportion of small degenerated surgical bioprostheses (≤21mm) appeared slightly lower in BEV (25.6%) than in SEV (30.7%) groups; however, the proportion of small THVs (≤23 mm) used was much higher in the BEV group (67.5%) than in the SEV group (26.9%). Transfemoral access was more frequently used in the SEV group (95%) than in the BEV group (61.3%).
Table 2. Baseline and procedural characteristics of patients (number of included studies = 27).
BEV (Edwards) | SEV (Medtronic) | |||
---|---|---|---|---|
Variable | Available data, n | Weighted % or mean ± SD | Available data, n | Weighted % or mean ± SD |
Age (year) | 1087 | 78.0 ± 1.6 | 788 | 75.6 ± 10.0 |
Male (%) | 1097 | 63.3% | 751 | 55.3% |
Log EuroSCORE | 730 | 21.7 ± 9.9 | 586 | 26.2 ± 3.4 |
EuroSCORE II | 53 | 19.6 ± 5.7 | 188 | 11.3 ± 2.9 |
STS score | 1087 | 9.0 ± 2.5 | 678 | 9.2 ± 2.2 |
CAD (%) | 491 | 65.2% | 431 | 61.7% |
Prior stroke (%) | 1065 | 13.7% | 666 | 11.7% |
Prior Afib (%) | 392 | 46.2% | 467 | 42.0% |
Prior PPM (%) | 943 | 17.2% | 377 | 22.5% |
PAD (%) | 1050 | 23.7% | 674 | 22.7% |
CKD (%) | 1009 | 37.4% | 674 | 34.3% |
AR ≥moderate (%) | 648 | 43.4% | 509 | 54.0% |
Bioprosthesis age (year) | 732 | 10.5 ± 1.6 | 657 | 9.9 ± 1.3 |
Stented valve (%) | 1078 | 84.2% | 867 | 75.1% |
Stentless valve (%) | 809 | 13.6% | 867 | 19.6% |
Bioprosthesis size (%) | 1097 | 746 | ||
≤21 mm | 25.6% | 30.7% | ||
21–24.9 mm | 40.4% | 37.9% | ||
≥25 mm | 31.4% | 28.9% | ||
Unknown | 2.3% | 3.2% | ||
Mode of failure (%) | ||||
AS | 1126 | 45.8% | 632 | 53.2% |
AR | 1126 | 28.0% | 620 | 27.3% |
Mix | 1116 | 26.6% | 583 | 21.8% |
LVEF (%) | 829 | 50.0 ± 2.7 | 527 | 51.1 ± 3.0 |
AV area (cm2) | 699 | 0.90 ± 0.08 | 671 | 0.95 ± 0.09 |
AVA index (cm2/m2) | 900 | 0.53 ± 0.06 | 213 | 0.55 |
Max PG (mmHg) | 341 | 62.6 ± 8.5 | 423 | 61.0 ± 9.2 |
Mean PG (mmHg) | 1022 | 34.6 ± 3.8 | 754 | 36.0 ± 4.4 |
Fluoroscopic time (min) | 436 | 18.3 ± 3.4 | 92 | 19.6 ± 9.8 |
THV size ≤23 mm (%) | 1299 | 67.5% | 490 | 26.9% |
TF access (%) | 1076 | 68.5% | 561 | 95.0% |
Abbreviations: Afib, atrial fibrillation; AR, aortic regurgitation; AS, aortic stenosis; AV, aortic valve; AVA, aortic valve area; CAD, coronary artery disease; CKD, chronic kidney disease; LVEF, left ventricular ejection fraction; PAD, peripheral artery disease; PG, pressure gradient; PPM, permanent pacemaker; PVL, paravalvular leak; STS, Society of Thoracic Surgery; TF, transfemoral; THV, transcatheter heart valve.
3.3. Clinical and echocardiographic outcomes
The event rates of all-cause mortality, cardiovascular death, and stroke at 30 days did not differ significantly between the BEV and SEV groups (Fig 2A). However, BEV was associated with significantly lower rates of major vascular complications (4.7% vs. 8.7%; P = 0.012), PPM implantation (3.8% vs. 12%; P < 0.001), and second valve requirement (2.9% vs. 6.2%; P = 0.004). One-year all-cause mortality and stroke rates were similar between the 2 groups (Fig 2B).
Regarding echocardiographic outcomes, SEV was associated with better hemodynamic performance than BEV, with significantly larger postoperative effective orifice area (EOA) at 30 days (1.53 cm2 vs. 1.23 cm2; P < 0.001) and 1 year (1.55 cm2 vs. 1.22 cm2; P < 0.001; Fig 3A and 3C) and lower maximal and mean pressure gradients at 1 year (respectively, 23.0 mm Hg vs. 33.3 mm Hg, P = 0.001; and 13 mm Hg vs. 18.4 mm Hg, P = 0.002; Fig 3C).
3.4. Subgroup analysis for newer devices
We also compared the outcomes with the Sapien 3 (Edwards Lifesciences) and Evolut R (Medtronic) valves. These are the newest generation of the 2 types of THVs with published data available for analyses. Although no statistical significance was found, Evolut R seemed to be associated with a lower mean pressure gradient than Sapien 3 (Fig 4).
4. Discussion
4.1. Major findings
With nearly 4000 patients included, the present meta-analysis is the largest sample used for comparing BEV and SEV outcomes in patients with failed aortic valve bioprostheses thus far. Our major findings were as follows: (1) all-cause mortality and cardiovascular death did not differ significantly between the 2 groups; (2) BEV was associated with lower rates of new PPM implantation and major vascular complications; and (3) SEV was associated with larger postprocedural EOA than BEV, both at 30 days and at 1 year.
4.2. New PPM implantation
SEV use is an independent risk factor for PPM implantation in the overall TAVR population [37–39]. However, previous aortic valve procedures, including surgical AVR, seemed to be protective against post-TAVR PPM implantation [37, 39], possibly because the previously implanted bioprosthesis restricted the expansion of the THV. In the present meta-analysis, the pooled PPM implantation rate after SEV implantation was 10.7%, which is nearly 3-fold that in the BEV group (3.6%; P < 0.001). In other words, even under the potential protection of the old bioprosthesis, SEV is still associated with significantly higher risk of postprocedural PPM implantation. This finding is consistent with previous studies focusing on aortic VIV procedure [8, 9].
4.3. EOA
A major concern of the aortic VIV procedure is the relatively small postprocedural aortic valve area and high transvalvular pressure gradient, mainly resulting from restricted expansion of the THVs by the old valves. Several publications, including studies using an in vitro model [40, 41], large cohort studies [8, 9], and propensity-matched analysis [7], reported that SEV was associated with larger postprocedural aortic valve area and lower transvalvular gradient than BEV after aortic VIV procedures. The current meta-analysis further supported these findings in the largest sample size to date.
In the SEV we analyzed, the functioning part is positioned above the aortic annulus (i.e., the “supra-annular design,” which is thought to lessen the detrimental impact on postprocedural EOA by the old valve). The theory was supported by a study using in vitro model in which researchers found that when the CoreValve was positioned deeper than normal, the leaflets were more constrained, and EOA decreased; and when the SAPIEN was placed more supra-annularly, the leaflets expanded more completely, and postprocedural EOA became larger [41].
One may argue that the higher percentage of small THVs (≤23mm) used in the BEV group alone can explain the smaller postprocedural EOA in BEV. However, the proportions of small degenerated surgical bioprostheses (≤21mm) were similar between the 2 groups (Table 2), so why were small THVs more often used in the BEV group? We believe that the supra-annular design of Medtronic SEV allows a relatively larger size, while the intra-annular design of Edward BEV results in marked leaflet distortion if the size is too large [41]. According to the ViV Aortic app, for 19 or 21mm degenerated bioprostheses, a 23mm Medtronic THV, or a 20mm Edward THV is suggested. A study using the Valve-in-Valve International Data Registry also found that elevated postprocedural pressure gradient were more common after BEV-VIV implantation than after SEV-VIV; for small surgical valves (internal diameter < 20mm) and intermediate-sized valves (internal diameter ≥20mm and <23mm) [8].
Insufficient EOA and elevated transvalvular pressure gradient not only diminish patients’ physical activity and quality of life but also predict early structural valve degeneration in bioprosthetic heart valves [42]. In addition, incomplete THV expansion itself leads to localized high stress within the leaflets, which may accelerate valve degeneration [43].
4.4. SEV versus BEV
According to the present meta-analysis, SEV was associated with significantly better postprocedural EOA, which can reduce the risk of patient–prosthesis mismatch and improve quality of life, particularly in patients with larger body size or whose old bioprosthesis is small. Lower transvalvular gradient and better THV expansion may also lead to superior durability of the THV, which is important in patients with life expectancy of 20 years or longer. Nevertheless, higher EOA and lower gradient of SEV did not translate in to lower mortality. Moreover, SEV was associated with higher rates of postprocedural PPM implantation, which is detrimental to late outcome [37, 44].
Therefore, CoreValve may be beneficial in patients whose previous surgical valve is small and those at high risk of patient–prosthesis mismatch. However, Edwards valves may be preferred to Medtronic valves for patients with adequate surgical valve size, particularly those who are prone to encounter postprocedural PPM implantation or PVL, including patients who are older [37] and those who have prior conduction disturbances [38] or a prolonged PR interval [45]. For every transcatheter aortic VIV candidate, particularly younger patients, the valve selection decision should be made carefully after thorough consideration of device characteristics and patient condition and preference, as well as detailed explanation and discussion.
4.5. Study limitations
The study has several limitations. First, this meta-analysis was based on published articles; therefore, data quality and availability are limited. Second, owing to a lack of randomized controlled trials in this area, all studies included were observational, so our results can only be interpreted as “associations,” rather than as “causations.” However, the absence of randomized studies warrants the present meta-analysis to help in optimizing device selection. Third, THV devices continue advancing rapidly, so the outcomes of the present study may differ from those of the newest device.
5. Conclusion
The present systematic review and meta-analysis found that for patients who underwent transcatheter aortic VIV, SEV was associated with significantly larger postprocedural EOA but higher rates of PPM implantation and PVL of moderate or higher degree. These findings provide valuable information in guiding proper management for patients with degenerated aortic bioprostheses.
Supporting information
Acknowledgments
The authors thank Alfred Hsing-Fen Lin and Zoe Ya-Jhu Syu for their assistance with the statistical analysis.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
This work was supported by a grant from Chang Gung Memorial Hospital, Taiwan CMRPG3H1511 (SWC), CMRPG 3J0661 (SWC), and BMRPD95 (SWC). This work was also supported by Ministry of Science and Technology grants MOST 107-2314-B-182A-152 and MOST 108-2314-B-182A-141 (SWC). The authors are thankful for the statistical assistance provided by and acknowledge the support of the Maintenance Project of the Center for Big Data Analytics and Statistics (grant CLRPG3D0045) at Chang Gung Memorial Hospital for statistical consultation and data analysis. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Isaacs AJ, Shuhaiber J, Salemi A, Isom OW, Sedrakyan A. National trends in utilization and in-hospital outcomes of mechanical versus bioprosthetic aortic valve replacements. J Thorac Cardiovasc Surg. 2015;149(5):1262–9 e3. Epub 2015/03/21. 10.1016/j.jtcvs.2015.01.052 . [DOI] [PubMed] [Google Scholar]
- 2.Tam DY, Vo TX, Wijeysundera HC, Dvir D, Friedrich JO, Fremes SE. Transcatheter valve-in-valve versus redo surgical aortic valve replacement for the treatment of degenerated bioprosthetic aortic valve: A systematic review and meta-analysis. Catheter Cardiovasc Interv. 2018;92(7):1404–11. Epub 2018/07/20. 10.1002/ccd.27686 . [DOI] [PubMed] [Google Scholar]
- 3.Takagi H, Mitta S, Ando T. Meta-analysis of Valve-in-Valve Transcatheter versus Redo Surgical Aortic Valve Replacement. Thorac Cardiovasc Surg. 2018;(67):243–50. Epub 2018/08/17. 10.1055/s-0038-1668135 30114716 [DOI] [PubMed] [Google Scholar]
- 4.Neupane S, Singh H, Lammer J, Othman H, Yamasaki H, Rosman HS, et al. Meta-Analysis of Transcatheter Valve-in-Valve Implantation Versus Redo Aortic Valve Surgery for Bioprosthetic Aortic Valve Dysfunction. Am J Cardiol. 2018;121(12):1593–600. Epub 2018/05/20. 10.1016/j.amjcard.2018.02.054 . [DOI] [PubMed] [Google Scholar]
- 5.Nalluri N, Atti V, Munir AB, Karam B, Patel NJ, Kumar V, et al. Valve in valve transcatheter aortic valve implantation (ViV-TAVI) versus redo-Surgical aortic valve replacement (redo-SAVR): A systematic review and meta-analysis. J Interv Cardiol. 2018;31(5):661–71. Epub 2018/05/22. 10.1111/joic.12520 . [DOI] [PubMed] [Google Scholar]
- 6.Gozdek M, Raffa GM, Suwalski P, Kolodziejczak M, Anisimowicz L, Kubica J, et al. Comparative performance of transcatheter aortic valve-in-valve implantation versus conventional surgical redo aortic valve replacement in patients with degenerated aortic valve bioprostheses: systematic review and meta-analysis. Eur J Cardiothorac Surg. 2018;53(3):495–504. Epub 2017/10/14. 10.1093/ejcts/ezx347 . [DOI] [PubMed] [Google Scholar]
- 7.Ochiai T, Yoon SH, Sharma R, Miyasaka M, Nomura T, Rami T, et al. Outcomes of Self-Expanding vs. Balloon-Expandable Transcatheter Heart Valves for the Treatment of Degenerated Aortic Surgical Bioprostheses- A Propensity Score-Matched Comparison. Circ J. 2018;82(10):2655–62. Epub 2018/08/03. 10.1253/circj.CJ-18-0157 . [DOI] [PubMed] [Google Scholar]
- 8.Dvir D, Webb JG, Bleiziffer S, Pasic M, Waksman R, Kodali S, et al. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA 2014;312(2):162–70. 10.1001/jama.2014.7246 [DOI] [PubMed] [Google Scholar]
- 9.Dvir D, Webb J, Brecker S, Bleiziffer S, Hildick-Smith D, Colombo A, et al. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: Results from the global valve-in-valve registry. Circulation. 2012;126(19):2335–44. 10.1161/CIRCULATIONAHA.112.104505 [DOI] [PubMed] [Google Scholar]
- 10.Wells GA, Shea B, O'Connell D, Peterson J, Welch V, Losos M, et al. The Newcastle-Ottawa Scale (NOS) for assessing the quality of nonrandomised studies in meta-analyses. Available from: http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp.
- 11.Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. Bmj. 2003;327(7414):557–60. Epub 2003/09/06. 10.1136/bmj.327.7414.557 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Seiffert M, Treede H, Schofer J, Linke A, Wöhrle J, Baumbach H, et al. Matched comparison of next- and early-generation balloonexpandable transcatheter heart valve implantations in failed surgical aortic bioprostheses. EuroIntervention. 2018;14(4):e397–e404. 10.4244/EIJ-D-17-00546 [DOI] [PubMed] [Google Scholar]
- 13.Schwerg M, Stangl K, Laule M, Stangl V, Dreger H. Valve in valve implantation of the CoreValve Evolut R in degenerated surgical aortic valves. Cardiol J. 2018;25(3):301–7. Epub 2018/03/24. 10.5603/CJ.a2018.0004 . [DOI] [PubMed] [Google Scholar]
- 14.Scholtz S, Piper C, Horstkotte D, Gummert J, Ensminger SM, Borgermann J, et al. Valve-in-valve transcatheter aortic valve implantation with CoreValve/Evolut R((c)) for degenerated small versus bigger bioprostheses. J Interv Cardiol. 2018;31(3):384–90. Epub 2018/03/01. 10.1111/joic.12498 . [DOI] [PubMed] [Google Scholar]
- 15.Webb JG, Mack MJ, White JM, Dvir D, Blanke P, Herrmann HC, et al. Transcatheter Aortic Valve Implantation Within Degenerated Aortic Surgical Bioprostheses: PARTNER 2 Valve-in-Valve Registry. J Am Coll Cardiol. 2017;69(18):2253–62. Epub 2017/05/06. 10.1016/j.jacc.2017.02.057 . [DOI] [PubMed] [Google Scholar]
- 16.Nielsen-Kudsk JE, Andersen A, Therkelsen CJ, Christensen EH, Jensen KT, Krusell LR, et al. High-pressure balloon fracturing of small dysfunctional Mitroflow bioprostheses facilitates transcatheter aortic valve-in-valve implantation. EuroIntervention. 2017;13(9):e1020–e5. 10.4244/EIJ-D-17-00244 [DOI] [PubMed] [Google Scholar]
- 17.Deeb GM, Chetcuti SJ, Reardon MJ, Patel HJ, Grossman PM, Schreiber T, et al. 1-Year Results in Patients Undergoing Transcatheter Aortic Valve Replacement With Failed Surgical Bioprostheses. JACC Cardiovasc Interv. 2017;10(10):1034–44. Epub 2017/05/20. 10.1016/j.jcin.2017.03.018 . [DOI] [PubMed] [Google Scholar]
- 18.Chhatriwalla AK, Allen KB, Saxon JT, Cohen DJ, Aggarwal S, Hart AJ, et al. Bioprosthetic Valve Fracture Improves the Hemodynamic Results of Valve-in-Valve Transcatheter Aortic Valve Replacement. Circ Cardiovasc Interv. 2017;10(7). Epub 2017/07/13. 10.1161/circinterventions.117.005216 . [DOI] [PubMed] [Google Scholar]
- 19.Ye J, Cheung A, Yamashita M, Wood D, Peng D, Gao M, et al. Transcatheter Aortic and Mitral Valve-in-Valve Implantation for Failed Surgical Bioprosthetic Valves: An 8-Year Single-Center Experience. JACC Cardiovasc Interv. 2015;8(13):1735–44. Epub 2015/10/20. 10.1016/j.jcin.2015.08.012 . [DOI] [PubMed] [Google Scholar]
- 20.Duncan A, Davies S, Di Mario C, Moat N. Valve-in-valve transcatheter aortic valve implantation for failing surgical aortic stentless bioprosthetic valves: A single-center experience. J Thorac Cardiovasc Surg. 2015;150(1):91–8. Epub 2015/04/15. 10.1016/j.jtcvs.2015.03.021 . [DOI] [PubMed] [Google Scholar]
- 21.Bapat V, Davies W, Attia R, Hancock J, Bolter K, Young C, et al. Use of balloon expandable transcatheter valves for valve-in-valve implantation in patients with degenerative stentless aortic bioprostheses: Technical considerations and results. J Thorac Cardiovasc Surg. 2014;148(3):917–22; discussion 22–4. Epub 2014/07/20. 10.1016/j.jtcvs.2014.05.029 . [DOI] [PubMed] [Google Scholar]
- 22.Ihlberg L, Nissen H, Nielsen NE, Ruck A, Busund R, Klaarborg KE, et al. Early clinical outcome of aortic transcatheter valve-in-valve implantation in the Nordic countries. J Thorac Cardiovasc Surg. 2013;146(5):1047–54; discussion 54. Epub 2013/09/04. 10.1016/j.jtcvs.2013.06.045 . [DOI] [PubMed] [Google Scholar]
- 23.Ong SH, Bauernschmitt R, Schuler G, Mueller R. Short- and mid-term safety and effectiveness of transcatheter aortic valve implantation in a failing surgical aortic bioprosthesis. Eur J Cardiothorac Surg. 2012;42(2):268–76; discussion 76. Epub 2012/02/16. 10.1093/ejcts/ezs014 . [DOI] [PubMed] [Google Scholar]
- 24.Linke A, Woitek F, Merx MW, Schiefer C, Mobius-Winkler S, Holzhey D, et al. Valve-in-valve implantation of Medtronic CoreValve prosthesis in patients with failing bioprosthetic aortic valves. Circ Cardiovasc Interv. 2012;5(5):689–97. Epub 2012/10/11. 10.1161/circinterventions.112.972331 . [DOI] [PubMed] [Google Scholar]
- 25.Bapat V, Attia R, Redwood S, Hancock J, Wilson K, Young C, et al. Use of transcatheter heart valves for a valve-in-valve implantation in patients with degenerated aortic bioprosthesis: technical considerations and results. J Thorac Cardiovasc Surg. 2012;144(6):1372–9; discussion 9–80. Epub 2012/11/13. 10.1016/j.jtcvs.2012.07.104 . [DOI] [PubMed] [Google Scholar]
- 26.Pasic M, Unbehaun A, Dreysse S, Buz S, Drews T, Kukucka M, et al. Transapical aortic valve implantation after previous aortic valve replacement: clinical proof of the "valve-in-valve" concept. J Thorac Cardiovasc Surg. 2011;142(2):270–7. Epub 2010/11/26. 10.1016/j.jtcvs.2010.09.049 . [DOI] [PubMed] [Google Scholar]
- 27.Bedogni F, Laudisa ML, Pizzocri S, Tamburino C, Ussia GP, Petronio AS, et al. Transcatheter valve-in-valve implantation using Corevalve Revalving System for failed surgical aortic bioprostheses. JACC Cardiovasc Interv. 2011;4(11):1228–34. Epub 2011/11/26. 10.1016/j.jcin.2011.10.002 . [DOI] [PubMed] [Google Scholar]
- 28.Kempfert J, Van Linden A, Linke A, Borger MA, Rastan A, Mukherjee C, et al. Transapical off-pump valve-in-valve implantation in patients with degenerated aortic xenografts. Ann Thorac Surg. 2010;89(6):1934–41. Epub 2010/05/25. 10.1016/j.athoracsur.2010.02.061 . [DOI] [PubMed] [Google Scholar]
- 29.Stankowski T, Aboul-Hassan SS, Seifi Zinab F, Herwig V, Stępiński P, Grimmig O, et al. Femoral transcatheter valve-in-valve implantation as alternative strategy for failed aortic bioprostheses: A single-centre experience with long-term follow-up. Int J Cardiol. 2020. Epub 2020/02/24. 10.1016/j.ijcard.2020.02.035 . [DOI] [PubMed] [Google Scholar]
- 30.Murdoch DJ, Sathananthan J, Hensey M, Alu MC, Liu Y, Crowley A, et al. Mitral regurgitation in patients undergoing transcatheter aortic valve implantation for degenerated surgical aortic bioprosthesis: Insights from PARTNER 2 Valve-in-Valve Registry. Catheter Cardiovasc Interv. 2020. Epub 2020/03/03. 10.1002/ccd.28811 . [DOI] [PubMed] [Google Scholar]
- 31.Sang SLW, Beute T, Heiser J, Berkompas D, Fanning J, Merhi W. Early Outcomes for Valve-in-valve Transcatheter Aortic Valve Replacement in Degenerative Freestyle Bioprostheses. Semin Thorac Cardiovasc Surg. 2018;30(3):262–8. Epub 2017/11/22. 10.1053/j.semtcvs.2017.11.001 . [DOI] [PubMed] [Google Scholar]
- 32.Stankowski T, Aboul-Hassan SS, Seifi-Zinab F, Herwig V, Kubikova M, Harnath A, et al. Severe structural deterioration of small aortic bioprostheses treated with valve-in-valve transcatheter aortic valve implantation. J Card Surg. 2019;34(1):7–13. Epub 2019/01/10. 10.1111/jocs.13976 . [DOI] [PubMed] [Google Scholar]
- 33.Pascual I, Avanzas P, Hernández-Vaquero D, Díaz R, Del Valle R, Padrón R, et al. Self-expanding transcatheter aortic valve implantation for degenerated Mitroflow bioprosthesis: Early outcomes. Int J Cardiol. 2019;287:53–8. Epub 2019/02/18. 10.1016/j.ijcard.2019.01.094 . [DOI] [PubMed] [Google Scholar]
- 34.Woitek FJ, Stachel G, Kiefer P, Haussig S, Leontyev S, Schlotter F, et al. Treatment of failed aortic bioprostheses: An evaluation of conventional redo surgery and transfemoral transcatheter aortic valve-in-valve implantation. Int J Cardiol. 2020;300:80–6. Epub 2019/11/20. 10.1016/j.ijcard.2019.09.039 . [DOI] [PubMed] [Google Scholar]
- 35.Seiffert M, Conradi L, Baldus S, Knap M, Schirmer J, Franzen O, et al. Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses. J Thorac Cardiovasc Surg. 2012;143(3):617–24. Epub 2011/12/16. 10.1016/j.jtcvs.2011.11.004 . [DOI] [PubMed] [Google Scholar]
- 36.Ribeiro HB, Rodes-Cabau J, Blanke P, Leipsic J, Kwan Park J, Bapat V, et al. Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry. Eur Heart J. 2018;39(8):687–95. Epub 2017/10/12. 10.1093/eurheartj/ehx455 . [DOI] [PubMed] [Google Scholar]
- 37.Fadahunsi OO, Olowoyeye A, Ukaigwe A, Li Z, Vora AN, Vemulapalli S, et al. Incidence, Predictors, and Outcomes of Permanent Pacemaker Implantation Following Transcatheter Aortic Valve Replacement: Analysis From the U.S. Society of Thoracic Surgeons/American College of Cardiology TVT Registry. JACC Cardiovasc Interv. 2016;9(1876–7605 (Electronic)):2189–99. [DOI] [PubMed] [Google Scholar]
- 38.Siontis GC, Juni P, Pilgrim T, Stortecky S, Bullesfeld L, Meier B, et al. Predictors of permanent pacemaker implantation in patients with severe aortic stenosis undergoing TAVR: a meta-analysis. J Am Coll Cardiol. 2014;64(2):129–40. Epub 2014/07/12. 10.1016/j.jacc.2014.04.033 . [DOI] [PubMed] [Google Scholar]
- 39.Ledwoch J, Franke J, Gerckens U, Kuck KH, Linke A, Nickenig G, et al. Incidence and predictors of permanent pacemaker implantation following transcatheter aortic valve implantation: analysis from the German transcatheter aortic valve interventions registry. Catheter Cardiovasc Interv. 2013;82(4):E569–77. Epub 2013/03/12. 10.1002/ccd.24915 . [DOI] [PubMed] [Google Scholar]
- 40.Sedaghat A, Sinning JM, Utzenrath M, Ghalati PF, Schmitz C, Werner N, et al. Hydrodynamic Performance of the Medtronic CoreValve and the Edwards SAPIEN XT Transcatheter Heart Valve in Surgical Bioprostheses: An In Vitro Valve-in-Valve Model. Ann Thorac Surg. 2016;101(1):118–24. Epub 2015/09/14. 10.1016/j.athoracsur.2015.06.047 . [DOI] [PubMed] [Google Scholar]
- 41.Midha PA, Raghav V, Condado JF, Okafor IU, Lerakis S, Thourani VH, et al. Valve Type, Size, and Deployment Location Affect Hemodynamics in an In Vitro Valve-in-Valve Model. JACC Cardiovasc Interv. 2016;9(15):1618–28. Epub 2016/08/06. 10.1016/j.jcin.2016.05.030 . [DOI] [PubMed] [Google Scholar]
- 42.Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121(19):2123–9. Epub 2010/05/05. 10.1161/CIRCULATIONAHA.109.901272 . [DOI] [PubMed] [Google Scholar]
- 43.Abbasi M, Azadani AN. Leaflet stress and strain distributions following incomplete transcatheter aortic valve expansion. J Biomech. 2015;48(13):3663–71. Epub 2015/09/05. 10.1016/j.jbiomech.2015.08.012 . [DOI] [PubMed] [Google Scholar]
- 44.Arnold SV, Zhang Y, Baron SJ, McAndrew TC, Alu MC, Kodali SK, et al. Impact of Short-Term Complications on Mortality and Quality of Life After Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv. 2019;12(4):362–9. Epub 2019/02/21. 10.1016/j.jcin.2018.11.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Tichelbacker T, Bergau L, Puls M, Friede T, Mutze T, Maier LS, et al. Insights into permanent pacemaker implantation following TAVR in a real-world cohort. PLoS One. 2018;13(10):e0204503 Epub 2018/10/18. 10.1371/journal.pone.0204503 [DOI] [PMC free article] [PubMed] [Google Scholar]