To the editor: The rapid development of open diagnostic methods for the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has been key to build capacity for efficient molecular diagnostic tests in laboratories worldwide. These methods based on real-time RT-PCR were recognised as reference protocols since mid-January, at the onset of the pandemic in China, and relayed – although not validated – by the World Health Organization [1]. We would first like to acknowledge the contribution of Corman et al. [2] who were among the first to provide primers and probes for three genes (E, N and RdRp), that have been widely implemented across the globe to tackle the coronavirus disease (COVID-19) pandemic. They recommended the use of the PCR targeting the E gene followed by confirmation with RdRp primers combined to a SARS-CoV-2 specific probe.
Our diagnostic laboratory also implemented and used the above-mentioned assays [2] throughout the first months of the pandemic on its automated platform [3]. RNA was extracted from clinical samples with the MagNA Pure 96 System (Roche, Basel, Switzerland) and the real-time reverse-transcription PCR (RT-PCR) was performed on a QuantStudio 7 system (Applied Biosystems, Waltham, United States). RT-PCRs targeting E and RdRp were used routinely in parallel for 893 samples. In 115 cases with positive amplification by both RT-PCRs, the RdRp assay showed a significantly (paired Wilcoxon rank test, p value < 0.001) higher average cycle threshold (CT) (25.0 CT, interquartile range (IQR): 24.6–27.9) than the E target (22.6 CT, IQR: 19.3–25.8). Furthermore, positive E and negative RdRp results were obtained in 10 cases (1%), triggering further investigations, since patients were unlikely to be infected by other SARS-related viruses that can also be amplified in the E assay. Conversely, no negative E and positive RdRp RT-PCR results were observed.
After careful review of the initial manuscript and analysis of SARS-CoV-2 and other coronavirus sequences, it appeared that the proposed RdRp reverse primer contained an incorrect degenerate base (S), that does not match with the SARS-CoV-2 RNA sequence, as shown in the alignment of Corman et al. Figure 2 [2]. Indeed, the proposed RdRp_SARSr-R sequence (CARATGTTAAASACACTATTAGCATA, R = [AG], S = [GC]) does not match any of the 1,623 SARS-CoV-2 complete genome sequences publicly available in the National Center for Biotechnology Information (NCBI) database as at 7 May 2020. Instead, the corrected RdRp_SARSr-R2 sequence (CARATGTTAAARACACTATTAGCATA, R = [AG]) should allow to amplify SARS-CoV-2 genetic material, including loosely related bat and human sequences, with improved efficiency. Two additional nucleotides in the pan-Sarbecovirus probe RdRP_SARSr-P1 (CCAGGTGGWACRTCATCMGGTGATGC, W = [AT], R = [AG], M = [AC]) should also be replaced (RdRP_SARSr-P1b, CCAGGTGGWACMTCATCMGGWGATGC, M = [AC], W = [AT]) to improve similarity with the SARS-CoV-2 and bat coronavirus genetic sequence while retaining the pan-Sarbecovirus compatibility. These observations based on in silico alignments should be confirmed by wet-laboratory experiments, but they could explain the lower sensitivity of the RdRp RT-PCR also shown by Vogels et al. [4] and point towards potential improvements.
As the pandemic spreads, many laboratories worldwide, including in low-resource countries that may not rely on expensive commercial kits, implement routine diagnostic tests. Thus, we think that such information is critical to ensure a proper detection of SARS-CoV-2 infections, allowing efficient isolation and preventing further transmission of the virus.
Acknowledgements
We gratefully acknowledge the work of all technicians who performed the molecular diagnostic of SARS-CoV-2 at Lausanne University Hospital during the past months.
Conflict of interest: None declared.
Authors’ contributions: Conceived and designed the manuscript: CB, VS, TP, GG. Contributed to the analysis: CB, TP, VS. Implemented and supervised the molecular diagnostic tests: KJ, GG. Drafted the manuscript: CB. Reviewed and approved the manuscript: all authors.
References
- 1.World Health Organization (WHO). Molecular assays to diagnose COVID-19. Geneva: WHO; 24 Jan 2020. Available from: https://www.who.int/who-documents-detail/molecular-assays-to-diagnose-covid-19-summary-table-of-available-protocols
- 2. Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DK, et al. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020;25(3):2000045. 10.2807/1560-7917.ES.2020.25.3.2000045 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Greub G, Sahli R, Brouillet R, Jaton K. Ten years of R&D and full automation in molecular diagnosis. Future Microbiol. 2016;11(3):403-25. 10.2217/fmb.15.152 [DOI] [PubMed] [Google Scholar]
- 4.Vogels CBF, Brito AF, Wyllie AL, Fauver JR, Ott IM, et al. Analytical sensitivity and efficiency comparisons of SARS-COV-2 qRT-PCR primer-probe sets. medRxiv 2020;2020.03.30.20048108; (Preprint). 10.1101/2020.03.30.20048108 [DOI] [PMC free article] [PubMed]