Abstract
Modern Holter devices are very trendy tools used in medicine, research, or sport. They monitor a variety of human physiological or pathophysiological signals. Nowadays, Holter devices have been developing very fast. New innovative products come to the market every day. They have become smaller, smarter, cheaper, have ultra-low power consumption, do not limit everyday life, and allow comfortable measurements of humans to be accomplished in a familiar and natural environment, without extreme fear from doctors. People can be informed about their health and 24/7 monitoring can sometimes easily detect specific diseases, which are normally passed during routine ambulance operation. However, there is a problem with the reliability, quality, and quantity of the collected data. In normal life, there may be a loss of signal recording, abnormal growth of artifacts, etc. At this point, there is a need for multiple sensors capturing single variables in parallel by different sensing methods to complement these methods and diminish the level of artifacts. We can also sense multiple different signals that are complementary and give us a coherent picture. In this article, we describe actual interesting multi-sensor principles on the grounds of our own long-year experiences and many experiments.
Keywords: multi-sensors, Holter, electrocardiography, electromyography, electrodermal activity, inertial measurement unit, pulse-oximetry
1. Introduction
The human body is a good conductor, allowing easy recording of electrical signals generated by the human body. That fact was applied by William Einthoven when he measured electric potentials of the heart at the beginning of the 20th century. An easier way to measure the heart electric activity was shown by Norman Jeffrey Holter in the second half of the 20th century. He started his trials with the long-term recording of human physiological parameters on active people [1]. He fixed his experiment on portable devices for recording heart activity–electrocardiographic (ECG) Holter [2]. The first Holters were large, uncomfortable, and required a chest lead unit, including a radio broadcasting unit with an antenna system [3]. Thanks to technical progress, nowadays, Holters are small portable devices [4] with wireless signal transmission, integrated memory for data recording, and modern electrodes. They are used for patient monitoring in the comfort of their own home where they are not scared by doctors. The so-called white coat syndrome has disappeared.
Holters are great for intercepting different heart or breath abnormalities [4,5,6,7,8,9,10], fetal arrhythmias [11], or simple physiological human states [12,13,14,15]. Holters showing daily ECG records are the perfect tools for long-term monitoring of the patient’s physiological state, able to notice deviations, which are not frequently repeated. Many patients often come to the ambulance saying they were sick the day before but feel fine at present, with ECG records showing no anomalies. This frequently happens in the case of cardiac fibrillation, which can eventually create cerebral defeat. Therefore, a classic ECG, which is just like a short flash in a patient’s history, is insufficient.
With the advent of telecommunication and internet technology, medicine could expand its borders, creating a new field called telemedicine. First medical data were communicated by telegraph, and people and doctors later started to use telephone and, in the 20th century, internet [16]. For today’s communication and data transfer, low-power wide-area networks (LPWAN) are very promising [17,18,19,20]. Telemedicine is a very fast evolving and transforming sphere where Holter devices are used in home care or for monitoring human physiological parameters by mobile healthcare assistants [5,21,22,23]. Human health is very important and early accurate diagnosis is essential to sustaining a high quality of life. The hectic lifestyle of today’s world causes people to live in stress without enough rest, undermining their self-care of health. Hence, enhancing and inventing new diagnostic methods, which record human physiology and show diseases in the early stages, are necessary for the development of medicine and more effective treatments for patients at long distance. Innovative methods could inform people about their own physiological state nonstop, and if some health problems arise, doctors would be directly informed of the person’s health state. Data measured by Holters sent immediately to the doctor would allow the doctor to control their patient from different places [24,25]. In addition, in the case of health problems, the doctor could instantly consult the health state with a specialist, or act fast and send an ambulance, if necessary. In addition, telemedicine helps to disburden doctors from little work, as the market is literally overwhelmed by numerous new medical products aimed at everyday people. The conditions for data recording, however, can vary in different environments, and are not comparable to predefined hospital conditions, and the records of the signals suffer from lower quality. These disadvantages could be removed by utilization of advanced methods and the newest technical devices implementing integrated chips together with the increasing computing power of new microprocessors, and creating cutting-edge integrated medical devices. In the near future, we can expect various new modern integrated medical devices for home examination and medicine self-diagnostics considering the amounts of capital invested in this field.
In this article, we would like to summarize the actual state in modern Holters and describe the perspective and innovative sensing principles usable in multi-sensor monitoring. Likewise, we would like to outline short sections of selected experiments we accomplished using our own designed devices during a longer period of time. These experiments aim to widen the knowledge in the multisensory field.
2. Capability of Modern Multi-Sensor Holters
The advantages of multi-sensors and multi-channel sensing are in the possibility to measure the same variable to achieve greater reliability in a more difficult environment. For example, it is possible to measure the heart rate (HR) electrically from an ECG [26,27,28,29,30,31], optically using pulse-oximetry [26,32,33,34], mechanically by vibration from seismocardiography (SCG) [7,30,34,35,36,37,38,39], or from minor variations in electrodermal activity (EDA) [40,41].
2.1. Electrocardiography and Respiration
Most famous and historically, the ECG Holter is one of the first introduced wearable monitoring devices. Today, the market offers single-channel or multi-channel Holters, with many complementary sensors, with or without a hospital certificate. Their price varies from tens of euros [42] to several thousands (EC-12H 12-Channel Holter ECG system by LabTech, Hungary or CardioMera ECG Holter Monitor, Medusoft, Australia) [43,44,45,46,47] depending on the quality and properties offered.
However, there are still challenges in this area. When considering a simple ECG record, with the doctor, the patient is in a quiet environment and is supervised, the quality of the electrode contacts is verified, the patient is in the prescribed position (mostly lying), and, for a brief moment of signal recording, they usually hold their breath to reduce artifacts, and the doctor, if necessary, receives immediate patient feedback. On the other hand, in real-life ECG recording using a standard Holter, the doctor is not able to control the quality of the skin–electrode contact or the detailed position of electrodes if they are not wrongly set. He/she is not familiar with the patient’s physical or mental activity and, thus, cannot evaluate if the increased heart pulse is objective instead of tachycardia. Moreover, human posture (patient bent in an unnatural position or traveling in a car and the fluctuating ECG amplitude is related to vehicle shaking), as well as environment (increased breathlessness may be caused by elevated temperature and humidity), have to be taken in account. Therefore, the ECG recording in the home environment has to be supplemented by the recording of physical parameters of the environment, and it is necessary to correlate all these effects to the final evaluation. This requires increased work activity from the doctor, and it is necessary to extend his skills in analyzing such modified ECGs. Hence, the development of auxiliary software, diagnostic applications, implementation of databases, and neural networks, which simplify this specific and more complicated diagnostics, must go hand in hand [48,49]. Long-term monitoring also produces an enormous quantity of data, which requires efficient data coding and communication of measured signals to the remote clinical back-end systems, and therefore, several modern communication protocols like Constrained Application Protocol (CoAP), Message Queuing Telemetry Transport (MQTT), Message Queuing Telemetry Transport for Sensor Networks (MQTT-SN), and Advanced Message Queuing Protocol (AMQP) have been introduced [50], and there is an urgent need for evolution in automatized screening [51,52,53].
For research purposes, our laboratory has developed an ECG Holter platform (Figure 1) [54]. In order to incorporate a wide range of different scientific tasks, the Holter was designed to be as versatile as possible. The heart of this platform is based on an analog front-end ADS1292R (Texas Instruments, Texas, USA) and ATxmega 128A3 (Microchip Technology, Arizona, USA) microcontroller. The analog front-end also includes circuits for impedance respiration measurement. The overall device battery consumption starts at 3 mA, which is required for portable electronics. The ECG Holter is complemented by inertial measurement units (IMU): Accelerometer with magnetometer LSM303D STMicroelectronics, gyroscope L3GD20 STMicroelectronics, barometer with temperature sensor BMP180 Bosh and notch filter. The gain, range, and sample frequencies for all units can be set in a wide range using a configuration file or Bluetooth control. Data are stored to a built-in 16 GB SD card in Comma-separated values (CSV) format, with the possibility of conversion to European Data Format Plus (EDF+) format. Measured data can be transferred offline via a USB connector or online via a Bluetooth Low Energy 4.0 (BLE) interface.
In Figure 2, we performed a short validation measurement where we compared our ECG Holter with a laboratory instrument. The aim was to verify the detailed parameters of the used Holter, including how the signal is affected when different electrode placing is used, if commercially used automated software correctly evaluates our output signals, and, of course, the progress in the process of hospital certification. We were also interested in the overall behavior and reliability of the innovative impedance respiration sensing [54], how it will stand compared to indirect measurement through pressure sensors, and the resistance of the circuit rib cage - chest belts [55,56,57] in different life situations. The ECG Holter signal was post-processed using a digital band-pass filter in range of 1–100 Hz with an auto-adjustable transition width. The respiration was filtered using the same filter in the range of 0.05–3 Hz. We observed that the results were comparable, and even in some situations such as low physical load or conversation, the impedance monitoring performed better. This can also be supported by the fact that the respiration impedance sensing technique has started to be widely used today [58,59,60].
2.2. Inertial Measurement Units and Seismocardiography
Inertial measurement units (IMU) are mostly used to record the movement, position, and posture of the human body [61,62,63,64,65,66] or for classification of human daily activities [67]. The first versions were used as simple pedometers, but at present, the range of applications is fast expanding. The main reason is their very low cost and easy implementation. Currently, they are inserted into almost all wearable instruments [7,36,68] and can also be used for indoor navigation.
In today’s highly industrialized time, the overall physical activity of the population falls below the recommended levels. As a result, obesity and diabetes are spreading all over the world. Obesity is often labelled as the epidemic of the 21st century. Monitoring childhood obesity, where accelerometers offer an objective measure of habitual activity independent of self-reporting [69,70], or homecare monitoring of elderly persons, where the system records their activities, events, and potentially important medical symptoms [71,72], is very well known.
As mentioned earlier, it is very important to correlate the ECG, heart rate variability (HRV) signal, and physical activity (IMU). This minimizes the number of false alarms and gives a more complete overview of the ECG record and its abnormalities.
The next application that seems to be interesting is seismocardiography (SCG). SCG measures small thorax movement and vibrations that contain information related to the cardiovascular and respiratory system [7,35,36,37,38]. The thorax accelerometer signal contains a low-frequency component corresponding to the motion of the chest wall due to respiration and a higher-frequency component corresponding to the heartbeat. Actual research teams are either trying to develop extracting algorithms of these signals [73] or are focused on increasing the quality of daily-life ECG monitoring using quantitative analysis of motion artifacts [74,75]. In different studies, the quality of the obtained SCG is enhanced using a novel adaptive recursive least-squares filter [76], time–frequency distribution analysis [77], or even using two cooperating accelerometers [78]. Very interesting are the studies where relations between ECG and SCG waveforms are analyzed [79], and then the simultaneous acquisition of ECG and SCG signals followed by mechanisms for the automatic delineation of relevant feature points can distinguish between normal and abnormal morphology [80], detect critical cardiac behaviors, and build early warning systems [81].
Figure 3a shows an example of signals obtained using our Holter with its IMU fixed to the thorax [68]. Except the overall movement and position of a human, a detailed look of the heart and respiration activity (mechanical heart activity) can be captured. Using a simple 20 Hz high-pass filtering, HR can be obtained even during demanding exercise when the ECG signal is unreadable due to heart bouncing (Figure 3b). Detailed accelerometer signals can be used, e.g., for cough and swallow analysis [82].
2.3. Muscle Activity
Electromyography (EMG) is used in motion analysis, physiotherapy, clinical research, and sport training [83,84,85,86,87,88]. EMG signals can often be used in automation for prosthetic devices such as prosthetic hands or lower limbs [89]. However, sensing in this area has certain specifics. Myoelectric prostheses expect comfortable and reliable electrodes without interfering with the user’s daily life. They must be ventilated, flexible, and foldable. Polymers, like polysiloxane [90], conductive fabric [91], or textile electrodes made by screen printing technology [92] meet the requirements.
In measuring muscle activity, the application of accelerometers seems to be very interesting, and the relevant field of science is called mechanomyography (MMG) [93]. MMG refers to the surface measuring of small vibrations of loaded muscles (amyostasia) or even a single motor unit [94]. Although there exist alternative methods of sensing muscle vibration, like piezoelectric resonance-based sensors [95] or laser Doppler [96], accelerometer sensing of MMG is still the most common. MMG can be a useful alternative to the electromyogram (EMG). It has a higher signal-to-noise ratio (SNR) than surface EMG and it can monitor the activity of even deeper muscles. This technique is, therefore, often utilized for evaluation of the muscular fatigue [97,98,99,100,101] or of the mechanical delay of muscle contraction [102,103]. The combination of MMG and EMG was already investigated for rehabilitation, control of prostheses, and in robotics. If MMG is added as a second detector to EMG monitoring, the total error of devices and prostheses can decrease up to 50% [89].
Electrical impedance myography (EIM), which measures the impedance of the electrical potential generated by muscles and neuron cells, can also be classified among innovative myographic methods. It can be used for observing muscle health and conditions. For example, there is a report for usage in the diagnosis of neuromuscular disorders [104], or for capturing changes in muscle composition [105]. In the study by Ma et al. [106], a wearable motion capture and measurement system combining an EMG, MMG, and ultrasound probe for understanding locomotion was developed.
When looking at our results, we demonstrated the capability of the combination of EMG (electromyography), MMG (mechanomyography), and EIM (electrical impedance myography) for examining muscle activity [107]. These variables can be measured simultaneously only by redesigning the electrode contacts and software setting of previously introduced Holter parameters. Instead of using cable contacts, we mounted the Ag/AgCl electrodes with clips directly to the bottom part of the housing in a distance of 2 cm (Figure 4). Compared to ECG measurements, we increased the sampling frequency of biopotential and impedance measurements from 1000 to 2000 Hz and for vibration measurement from 100 to 400 Hz. When measuring vibrations, the range was also increased from ±2 to ±4 g. Obtained signals were again software-filtered. The EMG signal used a digital band-pass filter in the range of 1–500 Hz, EIM, and MMG using a 0.1 Hz high-pass filter. Measurement was performed on the biceps brachii muscle. In Figure 5, the total time response of all signals is presented. The exercise comprised six series (each five lifts) of gradually increasing isotonic load (20–70% maximum voluntary contractions (MVC)), followed by 30 s of isometric exercise with 50% MVC (maximum voluntary contractions). Detailed analysis (Figure 6) shows that there is visible delay between the muscle activation (EMG start) and movement (MMG shift). These phases represent the transition of isotonic movement to isometric. This delay is interesting from the perspective of the evaluation of reflexes and reactions. For robotic and prostheses control, it is helpful that EIM is sensitive only in the case of isotonic signals and not isometric signals (Figure 5 and Figure 6). Figure 7 interprets the typical power signal density (PSD) of an isometric load. The comparison shows the amplitude and mean frequency (MNF) shift due to the fatigue factor.
The capability of muscle activity Holters can grow rapidly when connected together in a synchronous network, demonstrated, for example, in the case of EMG Holters by Delsys Incorporated (MA, USA) in their Trigno platform or by BTS Bioengineering Corp. (MA, USA) in their FREEEMG series [108,109,110,111,112]. As interesting research study for the precise control of prosthetic devices, the surface potential mapping [113] or stretchable EMG patch sensor integrated with the miniaturized wireless system modules, should also be mentioned [90].
2.4. Electrodermal Activity (EDA)
Further, there is growing interest and promise for wearable devices for electrodermal activity (EDA) recording, often also known as the electrodermal response (EDR) or psychogalvanic reflex (PGR) [114]. Laboratory devices commonly measure electrodermal activity between the down part of two fingers of the non-dominant hand. Such devices are ideal for psychological research, lie and stress detectors, etc. For daily life, more suitable EDA meters are in the form of watches, wristbands [115,116,117,118], or even smart eyewear [119] and bras [120], which are more practical and comfortable. The imperfection is in decreased sensitivity, while the nervous reactions and change in skin conductivity are more significant in the palm areas, where the stratum lucidum and potential barrier are presented [121,122,123]. Another problem is that the signal can be, due to a high movement artifact, obtained only in the calm state, so again, cooperation with IMU or at least the accelerometer is required.
We tried to avoid these problems by constructing an EDA Holter in the form of a practical ring [124] (Figure 8), which does not sense the EDA between the fingers like a conventional device, but locally on a small place in an area of 1 × 1 cm (Figure 9a). Such an EDA ring is more suitable for daily wear and the signal artifacts are lower. The Holter generator generates a sinus signal with an amplitude of 1.6–3 V and frequency of 1 kHz to a gold-plated interdigitated array of electrodes (IDAE). Their size is 200/200 m, so the electric field enters only into the neuroactive areas (stratum lucidum–potential barrier) and the impedance changes are quicker and more bound to the psychogalvanic response. Total dimensions of the EDA Holter are 20 × 20 × 5 mm and the device is connected to a smartphone with Bluetooth 2.0. The data are stored in CSV format. To minimize polarization effects in the skin, also called electrodermal phenomena, we used software filtering of this drift based on periodical recalibration with an exponential function. An exponential approximation was performed every minute, and this exponential was subtracted from the measured data (Figure 9b). This experiment also led to a very important result: The microelectrode probes are able to monitor the electrodermal response, as well as the heart pulses, simultaneously. They are present as small variations in the EDA signal related to blood pulsating in the bloodstream–plethysmography.
2.5. Pulse-Oximetry–Photoplethysmogram
The next sensor principle we describe in this article is pulse-oximetric, which is closely related to the photoplethysmogram (PPG). Classical sensors are commonly used in hospitals to monitor heart rate (HR) and blood oxygen level (spO2) mostly using the transmittance principle on fingers or earlobes [125,126,127], but there are also less traditional variants like PPG sensors placed in the human trachea during anesthesia [128]. In homecare, they are usually known to be implemented in all types of smart watches [129,130,131,132,133], and they work on a little more complicated but suitable reflectance principle for nonstop wearing [134,135,136,137]. They use mostly red, infrared, or green light to monitor HR, and their combination is used to determine spO2 levels. In fact, this method is arguably the most used for measuring HR. To remove false values and increase reliability, PPGs are sometimes enhanced by accelerometers [138] or multiplied [139,140] so they can be deployed for detailed heart rate variability (HRV) analysis [141], fibrillation classification [142], arterial status (ageing) monitoring [140], or in demanding applications like automotive applications where the PPG signal is measured from the palm of the hands [143].
Our team have developed a pulse-oximetric Holter (Figure 10a), based on the reflective principle, and enhanced it using two synchronized pulse-oximeters at a predetermined distance of 2 cm, so it is possible to also measure the local blood flow rate. Our measuring system (Figure 10b) consists of two pairs of light-emitting diodes (LED). The used RED LED is a Vishay VLMR51Z1AA and IR LED VSMY2943RG. The wavelengths of the LEDs are chosen matching the optical properties of blood. Oxygenated hemoglobin has a maximum light absorption of 940 nm and, on the other side, reduced hemoglobin has this maximum of 660 nm. Thanks to this difference, we are able to determine the ratio of oxygenated and reduced hemoglobin. This ratio determines the level of spO2 directly. Both pairs of LEDs are coupled with Vishay BWP34 photodiodes, which have comparable sensitivity to both of the used wavelengths of the emitted lights. The signal from the photodiodes is hardware-filtered using a band-pass filter of 0.5–5 Hz. All of the measurement processes are controlled by a Holter, which is based on a JN5148 microcontroller with an integrated ZigBee communication module and on a 24-bit analog-to-digital converter with software-adjustable sample rate and gain. PC communication is provided by one receiver module (USB dongle), which can simultaneously communicate with 4 to 6 of such Holters [144].
In the last years, a huge effort has been made to use PPG for cuffless blood pressure (BP) determination. There are known algorithms where systolic and diastolic BP are calculated from the shape of the PPG curve [145]. Thanks to modern neural learning methods [146,147], the accuracy of the results is continuously increasing [148,149,150].
For accurate estimation of BP, the combination of PPG with other physiological parameters such as ECG, SCG, ballistocardiogram (BCG), impedance cardiogram (ICG), etc. seems to be very promising [151]. Blood flow rate partially correlates with blood pressure [152,153]. In practice, plenty of devices are now coming that measure the approximate blood pressure by the phase shift between the ECG and the PPG curve [154,155,156,157]. The principle is simple. It is assumed that the ECG signal spreads across the body at the speed of light, so it is recorded immediately, and the PPG signal reacts to the blood flow itself, making it “a little” slower. These devices are mostly in the form of smart watches, where there is an optical PPG sensor with one ECG electrode (one bottom electrode can also be added for noise reduction) on the bottom, and a second ECG electrode on the top of watches. Such a smart watch on the market, for example, is Glutrac, where they even try to determine the blood glucose from an advanced PPG [158]. A sensing method like this, of course, needs planned human cooperation, as they must hold the top electrode with their second hand in the I bipolar Einthoven lead for about 10–60 s. The devices fusing PPG and ECG with deep neural networks seem to be very promising for the future, combining together PPG shape recognition and PTT to increase the accuracy of results [159,160]. What is also interesting seems to be a device combining an ECG, PPG, and SCG placed on the human chest [161]. In this principle, we estimate that the blood pressure resolution will be decreased by the low separation distance between the PPG sensing location and the heart. Smartwatch CareUp [162] uses two pulse oximeters for obtaining BP: One placed on the back and the second on the front of the watch. BP is calculated from the time delay between them. The Omron company is trying to miniaturize the classical blood pressure cuff into smart watch bands [163].
However, returning back to our device that uses only an optical PPG principle, it has to be clarified why it is relevant to continue in this sensing principle. Our main goal is to synchronize ECG Holters on the chest with a smart watch enhanced by our double PPG sensor. Thanks to this concept, we will obtain not only the overall blood speed but also the local speed and, using the appropriate algorithm, we can suppress artifacts like vascular elasticity, hydration, drug effects, etc. and increase accuracy. Figure 11 shows a part of our results, where we see shifted PPG signals (output signals of photodiode, only RED diode active) from the middle finger and wrist, which corresponds to blood speeds of 2.1 and 10.1 cm/s. The heart rate was about 70 BPM in both cases.
2.6. Technolgies of Wearable Devices and Smart Clothing
All wearable technologies have a few common pitfalls that they must improve. These include battery life, miniaturization associated with non-invasiveness, and, last but not least, price. Cost-effectiveness is one of the most important goals for the adoption and implementation of wearable technologies. The more sophisticated the system, the higher their price, and therefore, only if they are available and reasonably priced, they can really have the potential to improve health care. However, the smaller the device developed while maintaining the functionality, the greater the price, and the same applies to batteries. The biggest problem is that there is always a demand for longer-lasting but smaller batteries at the same time. It seems that the solution is not in the batteries themselves but in various settings and optimizations of data transfer, components, or software for the best possible battery management. In the coming years, batteries may have twice the density, but due to the chemical laws of energy storage, they have no potential to improve over time, as we can see from semiconductors. The solution to this key problem of wearable sensor devices is, therefore, in the development and selection of the right sensors and components with low consumption, as well as the use of low-energy technologies for data transmission such as the LoRa, Sigfox, or NB-IOT network. It is these networks that have low bandwidth and can transmit data over long distances. There exist many studies and reviews about smart wearables, and recent advances and future challenges [164] about the creation of cooperative systems based on wearable devices are directed to the field research context [165] or the energy challenges for wearable sensing technologies [166]. In addition, studies about battery-free wearables are very usable, as demonstrated by Orfanidis et al. [167], where the authors demonstrated how LPWAN wearables can operate by using energy-harvesting, which illustrates that the LoRa radio is able to operate by a combination of solar and mechanical energy on a smart shoe prototype outdoors.
The current trend is the connection of Holters with clothing containing conductive or piezoelectric textiles and threads. All this can be categorized as smart clothes. New Holters can be worn on the body in a variety of ways such as a chest strap or belt. As they may be visible under light clothing or under shirts with a low neckline, current researchers, including our team, are trying to create and implement invisible Holters directly into smart clothes for unlimited wearing [168,169]. The market already provides several examples. The sports bra from Movesense contains textile electrodes and a detachable Holter containing HR, IMU, and Bluetooth [170]. Alternatively, the MyHeart Instrumented Shirt [171] is equipped with sensors from conducting and piezo-resistive materials on a textile structure integrated in fabric, which is able to monitor ECG, on-arm EMG, respiration rate, skin temperature, and body movements. Blood pressure and oxygen saturation can also be obtained on demand. Some research institutes are trying to measure body composition and hydration. In this research, a self-powered smart patch for sweat conductivity monitoring [172] or this wearable potentiostat [173] can also be helpful. Interesting also is the ECG-derived respiration, which is a method for determining the respiratory information from ECG. In our research [174], we optimized it for miniaturized Holters with a low inter-electrode separation distance (few centimeters), often available in the form of patches [75,175,176]. We found that for measuring ECG and derived respiration, it is best to use the position under the pectoral muscle and, if possible, with electrodes orientated parallel to the heart axis. The suggested electrode setup is suitable for common daily activities, if no high physical activity is present. The possibilities seem almost unlimited.
All new technologies also bring various challenges with them and problematic features that need to be addressed or improved. The basic problem of all sensor systems recording physiological data of the human body is the privacy and security of data. Wearable devices thus record, process, and store sensitive user data. Users are generally unaware of the real risks of losing privacy and often underestimate this form of security. The basic form of increased security is data encryption during recording and storage in the internal memory of measuring devices. Encryption is usually integrated in the processor/on the chip, what most of the reliable developers of such systems currently do. For example, the choice of on-chip encryption can affect processor selection and, thus, battery consumption. After measurement and storage, the encrypted data are moved, e.g., to a server where other protection elements can be used [177,178,179].
An inseparable part of the research of portable devices is also the connection of their production with the selection of materials and technologies so that they are practical, cleanable, and fashionable [180].
3. Conclusions
Nowadays, due to huge technical developments, telemedicine applications have ascended. New sensors, software diagnostics, database systems, and innovative sensing principles have been developing fast. Development of information technologies gives us a great opportunity to improve health systems and facilitate doctor’s work.
Another advantage lies in the measurement of human body activity in home conditions, where people are calmer, more at rest, and not frightened by doctors. In these conditions, the results could be more appropriate. In addition, many people can suffer from diseases, when their heart demonstrates irregular activities often unnoticeable during ECG measurement in the ambulance; therefore, longer 24/7 recording with a Holter device is needed to catch the discrepancies.
Currently, in our laboratory, we are trying to apply these findings to modern smart clothes. We place the Holter on the chest, sew a suitably positioned electrode (using an optimized position of the electrodes for ECG recording), and place the improved pulse oximeter in the wrist or shoulder bracelet. If this is all combined, we could ultimately create a modern and robust device for a comprehensive record of human physiology. An integrated part of our research is the development of applications, access points, screening, diagnostic algorithms, and development of neural networks, in the form of powerful computers, as well as integrated directly inside Holters.
Taken together, in this work, we outlined some substantial options. Yet, there is a huge variety of other methodologies and applications, and their number will continue to grow with time. Likewise, Holter principles will be implanted in further devices of everyday use (e.g., our e-health mouse patent). We believe that we have shown a very broad application of multi-sensor Holters and that they have a bright future in diagnosis and treatment. We used our Holters introduced here, for example, in psychological research, for cognitive monitoring of the relationship of anxiety and allergies, in the investigation of ADHD attention in young children, in monitoring the impact of the working environment on performance, in apnea research, in determining the anaerobic threshold of rowers, or in research on the impact of architecture and design on human physiology [181].
Acknowledgments
In this section you can acknowledge any support given which is not covered by the author contribution or funding sections.
Funding
This work was supported by the Slovak Research and Development Agency under contract APVV-18-0550 Printed sensor elements for monitoring of human health using the IoT, APVV-16-0626 Flexible system for Internet of things with use of SMART sensors and the Ministry of Education, Science, Research and Sport of the Slovak Republic under grant VEGA 1/0733/20.
Conflicts of Interest
The authors declare no conflict of interest.
References
- 1.Kennedy H.L. The evolution of ambulatory ECG monitoring. Prog. Cardiovasc. Dis. 2013;56:127–132. doi: 10.1016/j.pcad.2013.08.005. [DOI] [PubMed] [Google Scholar]
- 2.Roberts W.C., Silver M.A. Norman Jefferis Holter and ambulatory ECG monitoring. Am. J. Cardiol. 1983;52:903–906. doi: 10.1016/0002-9149(83)90439-3. [DOI] [PubMed] [Google Scholar]
- 3.Makarov L.M. To the centennial of Norman Holter (1914–1983) Kardiologiia. 2015;55:59–61. doi: 10.18565/cardio.2015.1.59-61. [DOI] [PubMed] [Google Scholar]
- 4.Nguyen A., Ansari S., Hooshmand M., Lin K., Ghanbari H., Gryak J., Najarian K. Comparative Study on Heart Rate Variability Analysis for Atrial Fibrillation Detection in Short Single-Lead ECG Recordings; Proceedings of the 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Honolulu, HI, USA. 18–21 July 2018; pp. 526–529. [DOI] [PubMed] [Google Scholar]
- 5.Penders J., Altini M., de Molengraft J., van Romero I., Yazicioglu F., Van Hoof C. A low-power wireless ECG necklace for reliable cardiac activity monitoring on the move; Proceedings of the 33rd Annual International IEEE EMBS Conference; Boston, MA, USA. 30 August–3 September 2011. [Google Scholar]
- 6.Ruangsuwana R., Velikic G., Bocko M. Methods to extract respiration information from ECG signals; Proceedings of the 2010 IEEE International Conference on Acoustics, Speech and Signal Processing; Dallas, TX, USA. 14–19 March 2010; pp. 570–573. [Google Scholar]
- 7.Marco T. De Novel Wearable Seismocardiography and Clinical Status of Heart Failure Patients. Circ. Hear. Fail. 2018;11:1–11. doi: 10.1161/CIRCHEARTFAILURE.117.004313. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Fernández-mariño A.I., Harpole T., Oelstrom K., Delemotte L., Chanda B. Gating interaction maps reveal a noncanonical electromechanical coupling mode in the Shaker K+ channel. Nat. Struct. Mol. Biol. 2018;25:320–326. doi: 10.1038/s41594-018-0047-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Gao Y., Yan H., Xu Z., Xiao M., Song J. A principal component analysis based data fusion method for ECG-derived respiration from single-lead ECG. Australas. Phys. Eng. Sci. Med. 2018;41:59–67. doi: 10.1007/s13246-017-0612-9. [DOI] [PubMed] [Google Scholar]
- 10.Kimura T., Aizawa Y., Kurata N., Nakajima K., Kashimura S., Kunitomi A., Nishiyama T., Katsumata Y., Nishiyama N., Fukumoto K., et al. Assessment of atrial fibrillation ablation outcomes with clinic ECG, monthly 24-h Holter ECG, and twice-daily telemonitoring. Heart Vessel. 2017;32:317–325. doi: 10.1007/s00380-016-0866-2. [DOI] [PubMed] [Google Scholar]
- 11.Behar J.A. Non-invasive fetal electrocardiography for the detection of fetal arrhythmias. Prenat. Diagn. 2019;3:178–187. doi: 10.1002/pd.5412. [DOI] [PubMed] [Google Scholar]
- 12.Abreu R., Nunes S., Leal A., Figueiredo P. Physiological noise correction using ECG-derived respiratory signals for enhanced mapping of spontaneous neuronal activity with simultaneous EEG-fMRI. Neuroimage. 2017;154:115–127. doi: 10.1016/j.neuroimage.2016.08.008. [DOI] [PubMed] [Google Scholar]
- 13.Przystup P., Polinski A., Bujnowski A., Kocejko T., Wtorek J. A body position influence on ECG derived respiration; Proceedings of the 2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Jeju Island, Korea. 11–15 July 2017; pp. 3513–3516. [DOI] [PubMed] [Google Scholar]
- 14.Boyle J., Bidargaddi N., Sarela A., Karunanithi M. Automatic Detection of Respiration Rate from Ambulatory Single-Lead ECG. IEEE Trans. Inf. Technol. Biomed. 2009;13:890–896. doi: 10.1109/TITB.2009.2031239. [DOI] [PubMed] [Google Scholar]
- 15.Norman S.E., Eager R.A., Waran N.K., Jeffery L., Schroter R.C., Marlin D.J. Recording of ECG signals on a portable MiniDisc recorder for time and frequency domain heart rate variability analysis. Physiol. Behav. 2005;83:729–738. doi: 10.1016/j.physbeh.2004.09.007. [DOI] [PubMed] [Google Scholar]
- 16.Ryu S., Rochelle N., Mary N.Y., Liebert A. History of Telemedicine: Evolution, Context, and Transformation. Healthc. Inform. Res. 2010;16:65–66. doi: 10.4258/hir.2010.16.1.65. [DOI] [Google Scholar]
- 17.Chaudhari B.S., Zennaro M., Borkar S. LPWAN technologies: Emerging application characteristics, requirements, and design considerations. Futur. Internet. 2020;12:46. doi: 10.3390/fi12030046. [DOI] [Google Scholar]
- 18.Rubio-Aparicio J., Cerdan-Cartagena F., Suardiaz-Muro J., Ybarra-Moreno J. Design and Implementation of a Mixed IoT LPWAN Network Architecture. Sensors. 2019;19:675. doi: 10.3390/s19030675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Zemrane H., Baddi Y., Hasbi A. Ehealth smart application of WSN on WWAN. ACM Int. Conf. Proc. Ser. 2019:1–8. [Google Scholar]
- 20.Cilfone A., Davoli L., Belli L., Ferrari G. Wireless mesh networking: An IoT-oriented perspective survey on relevant technologies. Futur. Internet. 2019;11:99. doi: 10.3390/fi11040099. [DOI] [Google Scholar]
- 21.Chuo Y., Marzencki M., Hung B., Jaggernauth C., Tavakolian K., Lin P., Kaminska B. Mechanically Flexible Wireless Multisensor Platform for Human Physical Activity and Vitals Monitoring. IEEE Trans. Biomed. Circuits Syst. 2010;4:281–294. doi: 10.1109/TBCAS.2010.2052616. [DOI] [PubMed] [Google Scholar]
- 22.Boquete L., Ascariz J.M.R., Cantos J., Barea R., Miguel J.M., Ortega S., Peixoto N. A portable wireless biometric multi-channel system. Measurement. 2012;45:1587–1598. doi: 10.1016/j.measurement.2012.02.018. [DOI] [Google Scholar]
- 23.Carmo J.P., Correia J.H. RF CMOS transceiver at 2.4 GHz in wearables for measuring the cardio-respiratory function. Measurement. 2011;44:65–73. doi: 10.1016/j.measurement.2010.09.027. [DOI] [Google Scholar]
- 24.Waller M., Stotler C. Telemedicine: A Primer. Curr. Allergy Asthma Rep. 2018;18:54. doi: 10.1007/s11882-018-0808-4. [DOI] [PubMed] [Google Scholar]
- 25.Belardinelli A., Muratori L., Corazza I., Magnalardo M., Marangoni F., Zannoli R. Multi-functional device for cardiologic telemedicine and diagnostic holter Multi-Functional Device for Cardiologic Telemedicine and Diagnostic Holter. Comput. Cardiol. 2008:985–987. [Google Scholar]
- 26.Passler M. Senner In-Ear Pulse Rate Measurement: A Valid Alternative to Heart Rate Derived from Electrocardiography? Sensors. 2019;19:3641. doi: 10.3390/s19173641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Fereniec M., Stix G., Kania M., Mroczka T., Janusek D., Maniewski R. Risk assessment of ventricular arrhythmia using new parameters based on high resolution body surface potential mapping. Med. Sci. Monit. 2011;17:20. doi: 10.12659/MSM.881436. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Imam M.H., Karmakar C.K., Khandoker A.H., Palaniswami M. Effect of ECG-derived respiration (EDR) on modeling ventricular repolarization dynamics in different physiological and psychological conditions. Med. Biol. Eng. Comput. 2014;52:851–860. doi: 10.1007/s11517-014-1188-0. [DOI] [PubMed] [Google Scholar]
- 29.Lacko A., Hruboň A., Bestvina D., Mokáň M. Diagnostika kardiálnej autonómnej dysfunkcie. Súčasná Klinická Prax. 2005;3:19–24. [Google Scholar]
- 30.Siecinski S., Tkacz E.J., Kostka P.S. Comparison of HRV indices obtained from ECG and SCG signals from CEBS database. Biomed. Eng. Online. 2019;18:69. doi: 10.1186/s12938-019-0687-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Thakur R.K., Anoop C.S. A non-contact capacitance based electrocardiograph and associated heart-rate detection using enhanced Fourier interpolation method; Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society; Milan, Italy. 25–29 August 2015; pp. 849–852. [DOI] [PubMed] [Google Scholar]
- 32.Horton J.F., Stergiou P., Fung T.S., Katz L. Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during Exercise. Med. Sci. Sports Exerc. 2017;49:2600–2607. doi: 10.1249/MSS.0000000000001388. [DOI] [PubMed] [Google Scholar]
- 33.Her L., Care I., Her E.L., Guyen Q.T.N., Pateau V., Bodenes L., Lellouche F. Photoplethysmographic determination of the respiratory rate in acutely ill patients: Validation of a new algorithm and implementation into a biomedical device. Ann. Intensive Care. 2019;9:11. doi: 10.1186/s13613-019-0485-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Li K.H.C., White F.A., Tipoe T., Liu T., Wong M.C.S., Jesuthasan A., Baranchuk A., Tse G., Yan B.P. The current state of mobile phone apps for monitoring heart rate, heart rate variability, and atrial fibrillation: Narrative review. J. Med. Internet Res. 2019;21:1–23. doi: 10.2196/11606. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Tadi M.J., Lehtonen E., Hurnanen T., Koskinen J., Eriksson J., Pänkälä M., Teräs M., Koivisto T. A real-time approach for heart rate monitoring using a Hilbert transform in seismocardiograms. Physiol. Meas. 2016;37:1885–1909. doi: 10.1088/0967-3334/37/11/1885. [DOI] [PubMed] [Google Scholar]
- 36.Tadi M.J., Lehtonen E., Koivisto T., Pankaala M., Paasio A., Teras M. Seismocardiography: Toward heart rate variability (HRV) estimation; Proceedings of the 2015 IEEE International Symposium on Medical Measurements and Applications (MeMeA) Proceedings; Torino, Italy. 7–9 May 2015; pp. 261–266. [Google Scholar]
- 37.Taebi A., Solar B., Bomar A., Sandler R., Mansy H. Recent Advances in Seismocardiography. Vibration. 2019;2:64–86. doi: 10.3390/vibration2010005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Lee H., Lee H., Whang M. An Enhanced Method to Estimate Heart Rate from Seismocardiography via Ensemble Averaging of Body Movements at Six Degrees of Freedom. Sensors. 2018;18:238. doi: 10.3390/s18010238. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.D’Mello Y., Skoric J., Xu S., Roche P.J., Lortie M., Gagnon S., Plant D.V. Real-Time Cardiac Beat Detection and Heart Rate Monitoring from Combined Seismocardiography and Gyrocardiography. Sensors. 2019;19:3472. doi: 10.3390/s19163472. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Mukhopadhyay S.C. Wearable Sensors for Human Activity Monitoring: A Review. IEEE Sens. J. 2015;15:1321–1330. doi: 10.1109/JSEN.2014.2370945. [DOI] [Google Scholar]
- 41.Drachen A., Nacke L.E., Yannakakis G., Pedersen A.L. Sandbox ′10: Proceedings of the 5th ACM SIGGRAPH Symposium on Video Games. ACM Press; New York, NY, USA: 2010. Correlation between heart rate, electrodermal activity and player experience in first-person shooter games; pp. 49–54. [Google Scholar]
- 42.Batista D., da Silva H.P., Fred A., Moreira C., Reis M., Ferreira H.A. Benchmarking of the BITalino biomedical toolkit against an established gold standard. Healthc. Technol. Lett. 2019;6:32–36. doi: 10.1049/htl.2018.5037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Fujita H., Takeuchi I.T., Ohe K. A novel cloud-based mobile 12-lead ECG significantly improved onset-to-balloon time for STEMI patients. Eur. Heart J. 2013;34:2771. doi: 10.1093/eurheartj/eht309.2771. [DOI] [Google Scholar]
- 44.Yufu K., Shimomura T., Fujinami M., Nakashima T., Saito S., Ayabe R., Kawano K., Ishii Y., Okada N., Akioka H., et al. Impact of Mobile Cloud Electrocardiography System on Door-to-Balloon Time in Patients with Acute Coronary Syndrome in Oita Prefecture. Circ. Rep. 2019;1:241–247. doi: 10.1253/circrep.CR-19-0020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Frigy A., Varga I., Fogarasi Z., Belényi B., Kocsis I. The Influence of Sleep Apnea on 24-Hour and Nocturnal ECG and Blood Pressure Parameters in Patients with Acute Heart Failure. Med. Princ. Pract. 2019;28:150–157. doi: 10.1159/000496148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Shakhmatova K., Zalzman A., Antropova O., Osipova I., Lobanova N. Electric heart instability in men of stress professions with arterial hypertension. J. Hypertens. 2011;29:e534. doi: 10.1097/00004872-201106001-01618. [DOI] [Google Scholar]
- 47.O’Brien E., Dolan E. Ambulatory Blood Pressure Measurement in the Elderly. Hypertension. 2019;73:961–964. doi: 10.1161/HYPERTENSIONAHA.118.11977. [DOI] [PubMed] [Google Scholar]
- 48.Wu W.H., Bui A.A.T., Batalin M.A., Au L.K., Binney J.D., Kaiser W.J. MEDIC: Medical embedded device for individualized care. Artif. Intell. Med. 2008;42:137–152. doi: 10.1016/j.artmed.2007.11.006. [DOI] [PubMed] [Google Scholar]
- 49.Alemdar H., Ersoy C. Wireless sensor networks for healthcare: A survey. Comput. Netw. 2010;54:2688–2710. doi: 10.1016/j.comnet.2010.05.003. [DOI] [Google Scholar]
- 50.Hamida S., Hamida E., Ahmed B. A New mHealth Communication Framework for Use in Wearable WBANs and Mobile Technologies. Sensors. 2015;15:3379–3408. doi: 10.3390/s150203379. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Mond H.G. The Spectrum of Ambulatory Electrocardiographic Monitoring. Hear. Lung Circ. 2017;26:1–15. doi: 10.1016/j.hlc.2017.02.034. [DOI] [PubMed] [Google Scholar]
- 52.Sanders D., Ungar L., Eskander M.A., Seto A.H. Ambulatory ECG monitoring in the age of smartphones. Clevel. Clin. J. Med. 2019;86:483–493. doi: 10.3949/ccjm.86a.18123. [DOI] [PubMed] [Google Scholar]
- 53.Deserno T.M., Marx N. Computational electrocardiography: Revisiting Holter ECG monitoring. Methods Inf. Med. 2016;55:305–311. doi: 10.3414/ME15-05-0009. [DOI] [PubMed] [Google Scholar]
- 54.Vavrinsky E., Daricek M., Moskalova D., Horinek F., Donoval M. Design of Very Precise and Miniature Low Power ECG Holter; Proceedings of the 41st International Congress on Electrocardiology; Bratislava, Slovakia. 4–7 June 2014; pp. 257–260. [Google Scholar]
- 55.Reinvuo T., Hannula M., Sorvoja H., Alasaarela E., Myllylä R. Measurement of respiratory rate with high-resolution accelerometer and EMFit pressure sensor; Proceedings of the 2006 IEEE Sensors Applications Symposium; Houston, TX, USA. 7–9 February 2006; pp. 192–195. [Google Scholar]
- 56.Brady S., Dunne L.E., Tynan R., Diamond D., Smyth B., O’Hare G.M.P. Garment-based monitoring of respiration rate using a foam pressure sensor; Proceedings of the International Symposium on Wearable Computers (ISWC); Osaka, Japan. 18–21 October 2005; pp. 214–215. [Google Scholar]
- 57.Widjaja D., Taelman J., Vandeput S., Braeken M.A., Otte R.A., Van Den Bergh B.R., Van Huffel S. ECG-Derived Respiration: Comparison and New Measures for Respiratory Variability. Comput. Cardiol. 2010;37:149–152. [Google Scholar]
- 58.Chan H.-L., Lin S.-H., Wang F.-T., Hsu W.-Y., Wang C.-L. ECG-derived respirations based on phase-space reconstruction of single-lead ECG: Validations over various physical activities based on parallel recordings of ECG, respiration, and body accelerations; Proceedings of the 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society; Chicago, IL, USA. 26–30 August 2014; pp. 2282–2285. [DOI] [PubMed] [Google Scholar]
- 59.Ansari S., Ward K.R., Najarian K. Motion Artifact Suppression in Impedance Pneumography Signal for Portable Monitoring of Respiration: An Adaptive Approach. IEEE J. Biomed. Heal. Inform. 2017;21:387–398. doi: 10.1109/JBHI.2016.2524646. [DOI] [PubMed] [Google Scholar]
- 60.Wang F.-T., Chan H.-L., Wang C.-L., Jian H.-M., Lin S.-H. Instantaneous Respiratory Estimation from Thoracic Impedance by Empirical Mode Decomposition. Sensors. 2015;15:16372–16387. doi: 10.3390/s150716372. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.McGrath T., Fineman R., Stirling L. An Auto-Calibrating Knee Flexion-Extension Axis Estimator Using Principal Component Analysis with Inertial Sensors. Sensors. 2018;18:1882. doi: 10.3390/s18061882. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Du S., Sun W., Gao Y. Improving Observability of an Inertial System by Rotary Motions of an IMU. Sensors. 2017;17:698. doi: 10.3390/s17040698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Xiao Y., Ruan X., Chai J., Zhang X., Zhu X. Online IMU Self-Calibration for Visual-Inertial Systems. Sensors. 2019;19:1624. doi: 10.3390/s19071624. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Zimmermann T., Taetz B., Bleser G. IMU-to-Segment Assignment and Orientation Alignment for the Lower Body Using Deep Learning. Sensors. 2018;18:302. doi: 10.3390/s18010302. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Niu X., Wang Q., Li Y., Zhang Q., Jiang P. An IMU Evaluation Method Using a Signal Grafting Scheme. Sensors. 2016;16:854. doi: 10.3390/s16060854. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Vargas-Valencia L., Elias A., Rocon E., Bastos-Filho T., Frizera A. An IMU-to-Body Alignment Method Applied to Human Gait Analysis. Sensors. 2016;16:2090. doi: 10.3390/s16122090. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Rahim K.N.K.A., Elamvazuthi I., Izhar L., Capi G. Classification of Human Daily Activities Using Ensemble Methods Based on Smartphone Inertial Sensors. Sensors. 2018;18:4132. doi: 10.3390/s18124132. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Vavrinský E., Moskal’vá D., Darříček M., Donoval M., Horínek F., Popovič M., Miklovič P. Application of Acceleration Sensors in Physiological Experiments. J. Electr. Eng. 2014;65:304–308. doi: 10.2478/jee-2014-0049. [DOI] [Google Scholar]
- 69.Robertson W., Stewart-Brown S., Wilcock E., Oldfield M., Thorogood M. Utility of accelerometers to measure physical activity in children attending an obesity treatment intervention. J. Obes. 2011;2011:1–8. doi: 10.1155/2011/398918. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Elmesmari R., Martin A., Reilly J.J., Paton J.Y. Comparison of accelerometer measured levels of physical activity and sedentary time between obese and non-obese children and adolescents: A systematic review. BMC Pediatr. 2018;18:106. doi: 10.1186/s12887-018-1031-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Manchanda S., Ehsanullah M. Suspected cardiac syncope in elderly patients: Use of the 12-lead electrocardiogram to select patients for Holter monitoring. Gerontology. 2001;47:195–197. doi: 10.1159/000052798. [DOI] [PubMed] [Google Scholar]
- 72.Gokalp H., Clarke M. Monitoring activities of daily living of the elderly and the potential for its use in telecare and telehealth: A review. Telemed. e-Health. 2013;19:910–923. doi: 10.1089/tmj.2013.0109. [DOI] [PubMed] [Google Scholar]
- 73.Pandia K., Inan O.T., Kovacs G.T.A., Giovangrandi L. Extracting respiratory information from seismocardiogram signals acquired on the chest using a miniature accelerometer. Physiol. Meas. 2012;33:1643–1660. doi: 10.1088/0967-3334/33/10/1643. [DOI] [PubMed] [Google Scholar]
- 74.Han D.K., Hong J.H., Shin J.Y., Lee T.S. Accelerometer based motion noise analysis of ECG signal; Proceedings of the World Congress on Medical Physics and Biomedical Engineering; Munich, Germany. 7–12 September 2009; pp. 198–201. [Google Scholar]
- 75.Marcelli E., Capucci A., Minardi G., Cercenelli L. Multi-Sense CardioPatch. ASAIO J. 2017;63:73–79. doi: 10.1097/MAT.0000000000000446. [DOI] [PubMed] [Google Scholar]
- 76.Yu S., Liu S. A Novel Adaptive Recursive Least Squares Filter to Remove the Motion Artifact in Seismocardiography. Sensors. 2020;20:1596. doi: 10.3390/s20061596. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Taebi A., Mansy H.A. Time-frequency distribution of seismocardiographic signals: A comparative study. Bioengineering. 2017;4:32. doi: 10.3390/bioengineering4020032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Luu L., Dinh A. Artifact Noise Removal Techniques on Seismocardiogram Using Two Tri-Axial Accelerometers. Sensors. 2018;18:1067. doi: 10.3390/s18041067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Mora N., Cocconcelli F., Matrella G., Ciampolini P. Detection and Analysis of Heartbeats in Seismocardiogram Signals. Sensors. 2020;20:1670. doi: 10.3390/s20061670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Sahoo P., Thakkar H., Lin W.-Y., Chang P.-C., Lee M.-Y. On the Design of an Efficient Cardiac Health Monitoring System through Combined Analysis of ECG and SCG Signals. Sensors. 2018;18:379. doi: 10.3390/s18020379. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Sahoo P., Thakkar H., Lee M.-Y. A Cardiac Early Warning System with Multi Channel SCG and ECG Monitoring for Mobile Health. Sensors. 2017;17:711. doi: 10.3390/s17040711. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Lee K.H., Ni X., Lee J.Y., Arafa H., Pe D.J., Xu S., Avila R., Irie M., Lee J.H., Easterlin R.L., et al. Mechano-acoustic sensing of physiological processes and body motions via a soft wireless device placed at the suprasternal notch. Nat. Biomed. Eng. 2020;4:148–158. doi: 10.1038/s41551-019-0480-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Kimoto A., Yamada Y. A new layered sensor for simultaneous measurement of EMG, MMG and oxygen consumption at the same position. Med. Biol. Eng. Comput. 2015;53:15–22. doi: 10.1007/s11517-014-1208-0. [DOI] [PubMed] [Google Scholar]
- 84.González-Izal M., Malanda A., Navarro-Amézqueta I., Gorostiaga E.M., Mallor F., Ibañez J., Izquierdo M. EMG spectral indices and muscle power fatigue during dynamic contractions. J. Electromyogr. Kinesiol. 2010;20:233–240. doi: 10.1016/j.jelekin.2009.03.011. [DOI] [PubMed] [Google Scholar]
- 85.Cifrek M., Medved V., Tonković S., Ostojić S. Surface EMG based muscle fatigue evaluation in biomechanics. Clin. Biomech. 2009;24:327–340. doi: 10.1016/j.clinbiomech.2009.01.010. [DOI] [PubMed] [Google Scholar]
- 86.Konrad P. The ABC of EMG. Noraxon Inc.; Scottsdale, AZ, USA: 2006. [Google Scholar]
- 87.De Luca C.J. The use of surface electromyography in biomechanics. J. Appl. Biomech. 1997;13:135–163. doi: 10.1123/jab.13.2.135. [DOI] [Google Scholar]
- 88.De Luca C.J. Electromyography. In: Webster J.G., editor. Encyclopedia of Medical Devices and Instrumentation. John Wiley Publisher; Hoboken, NJ, USA: 2006. pp. 98–109. [Google Scholar]
- 89.Zhang X., Li X., Samuel O.W., Huang Z., Fang P., Li G. Improving the robustness of electromyogram-pattern recognition for prosthetic control by a postprocessing strategy. Front. Neurorobotics. 2017;11:1–15. doi: 10.3389/fnbot.2017.00051. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Lee S., Yoon J., Lee D., Seong D., Lee S., Jang M., Choi J., Yu K.J., Kim J., Lee S., et al. Wireless Epidermal Electromyogram Sensing System. Electronics. 2020;9:269. doi: 10.3390/electronics9020269. [DOI] [Google Scholar]
- 91.Pylatiuk C., Muller-Riederer M., Kargov A., Schulz S., Schill O., Reischl M., Bretthauer G. Comparison of surface EMG monitoring electrodes for long-term use in rehabilitation device control; Proceedings of the 2009 IEEE International Conference on Rehabilitation Robotics; Kyoto, Japan. 23–26 June 2009; pp. 300–304. [Google Scholar]
- 92.Zhang H., Tian L., Zhang L., Li G. Using textile electrode EMG for prosthetic movement identification in transradial amputees; Proceedings of the 2013 IEEE International Conference on Body Sensor Networks; Cambridge, MA, USA. 6–9 May 2013; pp. 1–5. [Google Scholar]
- 93.Islam M.A., Sundaraj K., Ahmad R.B., Ahamed N.U., Ali M.A. Mechanomyography Sensor Development, Related Signal Processing, and Applications: A Systematic Review. IEEE Sens. J. 2013;13:2499–2516. doi: 10.1109/JSEN.2013.2255982. [DOI] [Google Scholar]
- 94.Uchiyama T., Hashimoto E. System identification of the mechanomyogram from single motor units during voluntary isometric contraction. Med. Biol. Eng. Comput. 2011;49:1035–1043. doi: 10.1007/s11517-011-0752-0. [DOI] [PubMed] [Google Scholar]
- 95.Han H., Jo S., Kim J. Comparative study of a muscle stiffness sensor and electromyography and mechanomyography under fatigue conditions. Med. Biol. Eng. Comput. 2015;53:577–588. doi: 10.1007/s11517-015-1271-1. [DOI] [PubMed] [Google Scholar]
- 96.Casaccia S., Scalise L., Casacanditella L., Tomasini E.P., Rohrbaugh J.W. Non-contact assessment of muscle contraction: Laser Doppler Myography; Proceedings of the 2015 IEEE International Symposium on Medical Measurements and Applications (MeMeA) Proceedings; Torino, Italy. 7–9 May 2015; pp. 610–615. [Google Scholar]
- 97.Al-Mulla M., Sepulveda F. Novel Pseudo-Wavelet Function for MMG Signal Extraction during Dynamic Fatiguing Contractions. Sensors. 2014;14:9489–9504. doi: 10.3390/s140609489. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Hill E.C., Housh T.J., Smith C.M., Cochrane K.C., Jenkins N.D.M., Cramer J.T., Schmidt R.J., Johnson G.O. Effect of sex on torque, recovery, EMG, and MMG responses to fatigue. J. Musculoskelet. Neuronal Interact. 2016;16:310–317. [PMC free article] [PubMed] [Google Scholar]
- 99.Bilgin G., Hindistan E., Ozkaya Y.G., Koklukaya E., Polat O., Colak O.H. Determination of Fatigue Following Maximal Loaded Treadmill Exercise by Using Wavelet Packet Transform Analysis and MLPNN from MMG-EMG Data Combinations. J. Med. Syst. 2015;39:108. doi: 10.1007/s10916-015-0304-5. [DOI] [PubMed] [Google Scholar]
- 100.Woodward R.B., Stokes M.J., Shefelbine S.J., Vaidyanathan R. Segmenting Mechanomyography Measures of Muscle Activity Phases Using Inertial Data. Sci. Rep. 2019;9:1–10. doi: 10.1038/s41598-019-41860-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Orizio C., Gobbo M., Diemont B., Esposito F., Veicsteinas A. The surface mechanomyogram as a tool to describe the influence of fatigue on biceps brachii motor unit activation strategy. Historical basis and novel evidence. Eur. J. Appl. Physiol. 2003;90:326–336. doi: 10.1007/s00421-003-0924-1. [DOI] [PubMed] [Google Scholar]
- 102.Cè E., Rampichini S., Monti E., Venturelli M., Limonta E., Esposito F. Changes in the electromechanical delay components during a fatiguing stimulation in human skeletal muscle: An EMG, MMG and force combined approach. Eur. J. Appl. Physiol. 2017;117:95–107. doi: 10.1007/s00421-016-3502-z. [DOI] [PubMed] [Google Scholar]
- 103.Smith C.M., Housh T.J., Hill E.C., Johnson G.O., Schmidt R.J. Changes in electromechanical delay during fatiguing dynamic muscle actions. Muscle Nerve. 2017;56:315–320. doi: 10.1002/mus.25502. [DOI] [PubMed] [Google Scholar]
- 104.Sanchez B., Rutkove S.B. Present Uses, Future Applications, and Technical Underpinnings of Electrical Impedance Myography. Curr. Neurol. Neurosci. Rep. 2017;17:86. doi: 10.1007/s11910-017-0793-3. [DOI] [PubMed] [Google Scholar]
- 105.Li X., Shin H., Li L., Magat E., Li S., Zhou P. Assessing the immediate impact of botulinum toxin injection on impedance of spastic muscle. Med. Eng. Phys. 2017;43:97–102. doi: 10.1016/j.medengphy.2017.01.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Ma C.Z.-H., Ling Y.T., Shea Q.T.K., Wang L.-K., Wang X.-Y., Zheng Y.-P. Towards Wearable Comprehensive Capture and Analysis of Skeletal Muscle Activity during Human Locomotion. Sensors. 2019;19:195. doi: 10.3390/s19010195. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Vavrinsky E., Svobodova H., Donoval M., Daricek M., Kopani M., Miklovic P., Horinek F., Telek P. Application of single wireless holter to simultaneous EMG, MMG and EIM measurement of human muscles activity. Lek. Tech. 2018;48:52–58. [Google Scholar]
- 108.Samulski B., Prebor J., Armitano C., Morrison S. Coupling of motor oscillators—What really happens when you chew gum and walk? Neurosci. Lett. 2019;698:90–96. doi: 10.1016/j.neulet.2019.01.016. [DOI] [PubMed] [Google Scholar]
- 109.Harmon K.K., Girts R.M., Maclennan R.J., Stock M.S. Is the motor unit mean firing rate versus recruitment threshold relationship linear? Physiol. Meas. 2019;40:095002. doi: 10.1088/1361-6579/ab4025. [DOI] [PubMed] [Google Scholar]
- 110.Miller J.D., Marissa C.J.L., Kyle D.G., Mandy L.S., Trent E.W. The effect of rate of torque development on motor unit recruitment and firing rates during isometric voluntary trapezoidal contractions. Exp. Brain Res. 2019;237:2653–2664. doi: 10.1007/s00221-019-05612-0. [DOI] [PubMed] [Google Scholar]
- 111.Hsu W.-C., Chang C.-C., Lin Y.-J., Yang F.-C., Lin L.-F., Chou K.-N. The Use of Wearable Sensors for the Movement Assessment on Muscle Contraction Sequences in Post-Stroke Patients during Sit-to-Stand. Sensors. 2019;19:657. doi: 10.3390/s19030657. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Kordi M., Folland J., Goodall S., Barratt P., Howatson G. Reliability of traditional and task specific reference tasks to assess peak muscle activation during two different sprint cycling tests. J. Electromyogr. Kinesiol. 2019;46:41–48. doi: 10.1016/j.jelekin.2019.03.008. [DOI] [PubMed] [Google Scholar]
- 113.Tian L., Zimmerman B., Akhtar A., Yu K.J., Moore M., Wu J., Larsen R.J., Lee J.W., Li J., Liu Y., et al. Large-area MRI-compatible epidermal electronic interfaces for prosthetic control and cognitive monitoring. Nat. Biomed. Eng. 2019;3:194–205. doi: 10.1038/s41551-019-0347-x. [DOI] [PubMed] [Google Scholar]
- 114.Posada-Quintero H.F., Chon K.H. Innovations in Electrodermal Activity Data Collection and Signal Processing: A Systematic Review. Sensors. 2020;20:479. doi: 10.3390/s20020479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Onorati F., Regalia G., Gaborni C., Picard R. Improvement of a convulsive seizure detector relying on accelerometer and electrodermal activity collected continuously by a wristband; Proceedings of the 2016 Epilepsy Pipeline Conference; San Francisco, CA, USA. 26 February 2016. [Google Scholar]
- 116.Regalia G., Onorati F., Migliorini M., Picard R.W. An improved wrist-worn convulsive seizure detector based on accelerometry and electrodermal activity sensors. Am. Epilepsy Soc. Annu. Meet. 2015;2015:1. [Google Scholar]
- 117.Poh M.Z., Loddenkemper T., Reinsberger C., Swenson N.C., Goyal S., Sabtala M.C., Madsen J.R., Picard R.W. Convulsive seizure detection using a wrist-worn electrodermal activity and accelerometry biosensor. Epilepsia. 2012;53:93–97. doi: 10.1111/j.1528-1167.2012.03444.x. [DOI] [PubMed] [Google Scholar]
- 118.Caborni C., Regalia G., Onorati F., Picard R.W. Clinical evaluation of the Embrace smart watch detection capability of generalized tonic-clonic seizures recorded at the ankles; Proceedings of the American Epilepsy Society annual meeting; Houston, TX, USA. 30 November–4 December 2018. [Google Scholar]
- 119.Zheng D., Chernyshov G., Kunze K. Electrodermal activity sensing using smart eyewear; Proceedings of the 2019 ACM International Joint Conference on Pervasive and Ubiquitous Computing and Proceedings of the 2019 ACM International Symposium on Wearable Computers—UbiComp/ISWC ′19; London, UK. 16 September 2019; pp. 653–656. [Google Scholar]
- 120.Carroll E.A., Czerwinski M., Roseway A., Kapoor A., Johns P., Rowan K., Schraefel M.C. Food and mood: Just-in-time support for emotional eating; Proceedings of the 2013 Humaine Association Conference on Affective Computing and Intelligent Interaction; Geneva, Switzerland. 2–5 September 2013; pp. 252–257. [Google Scholar]
- 121.Vavrinsky E., Stopjakova V., Majer L., Tvarozek V., Weis M., Marman P. Monitoring of Psychosomatic Properties of Human Body by Skin Conductivity Measurements using Thin Film Microelectrode Arrays; Proceedings of the 2006 International Conference on Advanced Semiconductor Devices and Microsystems; Smolenice, Slovakia. 16–18 October 2006; pp. 275–278. [Google Scholar]
- 122.Aldosky H.Y.Y., Bari D.S. Electrochemical Impedance Spectroscopy. Volume 395. IntechOpen; London, UK: 2019. Electrodermal Activity: Simultaneous Recordings; pp. 116–124. [Google Scholar]
- 123.Malmivuo J., Plonsey R. Biomagnetism. Principles and Applications of Bioelectric and Biomagnetic Fields. Oxford University Press; Oxford, UK: 1995. The Electrodermal Response. [Google Scholar]
- 124.Vavrinsky E., Stopjakova V., Donoval M., Daricek M., Svobodova H., Mihalov J., Hanic M., Tvarozek V. Design of sensor systems for long time electrodermal activity monitoring. Adv. Electr. Electron. Eng. 2017;15:184–191. doi: 10.15598/aeee.v15i2.2021. [DOI] [Google Scholar]
- 125.Tannheimer M. The Use of Pulse Oximetry at High Altitude. Res. Inves. Sports. Med. 2020;6:10–13. [Google Scholar]
- 126.Mannheimer P.D. The light-tissue interaction of pulse oximetry. Anesth. Analg. 2007;105:10–17. doi: 10.1213/01.ane.0000269522.84942.54. [DOI] [PubMed] [Google Scholar]
- 127.Nitzan M., Romem A., Koppel R. Pulse oximetry: Fundamentals and technology update. Med. Devices Evid. Res. 2014;7:231–239. doi: 10.2147/MDER.S47319. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.May J.M., Phillips J.P., Fitchat T., Ramaswamy S., Snidvongs S., Kyriacou P.A. A novel photoplethysmography sensor for vital signs monitoring from the human trachea. Biosensors. 2019;9:119. doi: 10.3390/bios9040119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Kumar S., Buckley J.L., Barton J., Pigeon M., Newberry R., Rodencal M., Hajzeraj A., Hannon T., Rogers K., Casey D., et al. A Wristwatch-Based Wireless Sensor Platform for IoT Health Monitoring Applications. Sensors. 2020;20:1675. doi: 10.3390/s20061675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Elsamnah F., Bilgaiyan A., Affiq M., Shim C.H., Ishidai H., Hattori R. Reflectance-based organic pulse meter sensor for wireless monitoring of photoplethysmogram signal. Biosensors. 2019;9:87. doi: 10.3390/bios9030087. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131.Tedesco S., Sica M., Ancillao A., Timmons S., Barton J., O’Flynn B. Accuracy of consumer-level and research-grade activity trackers in ambulatory settings in older adults. PLoS ONE. 2019;14:1–17. doi: 10.1371/journal.pone.0216891. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132.GARMIN Foreruner 945 Owner’s Manual. Pulse Oximeter. [(accessed on 27 March 2020)]; Available online: https://www8.garmin.com/manuals/webhelp/forerunner945/EN-US/GUID-4D425925-D4EE-4C26-B974-5375D0670860.html.
- 133.Husain I.M. How Apple Watch’s Ability to Measure Blood Oxygen Saturation Can Be Used in Medicine. [(accessed on 26 March 2020)]; Available online: https://www.imedicalapps.com/2015/04/how-apple-watchs-ability-to-measure-blood-oxygen-saturation-can-be-used-in-medicine/
- 134.König V., Huch R., Huch A. Reflectance pulse oximetry—Principles and obstetric application in the Zurich system. J. Clin. Monit. Comput. 1998;14:403–412. doi: 10.1023/A:1009983010772. [DOI] [PubMed] [Google Scholar]
- 135.Jørgensen J.S., Schmid E.R., König V., Faisst K., Huch A., Huch R. Limitations of forehead pulse oximetry. J. Clin. Monit. 1995;11:253–256. doi: 10.1007/BF01617520. [DOI] [PubMed] [Google Scholar]
- 136.Mendelson Y., Ochs B.D. Noninvasive Pulse Oximetry Utilizing Skin Reflectance Photoplethysmography. IEEE Trans. Biomed. Eng. 1988;35:798–805. doi: 10.1109/10.7286. [DOI] [PubMed] [Google Scholar]
- 137.Lee H., Ko H., Lee J. Reflectance pulse oximetry: Practical issues and limitations. ICT Express. 2016;2:195–198. doi: 10.1016/j.icte.2016.10.004. [DOI] [Google Scholar]
- 138.Wójcikowski M., Pankiewicz B. Photoplethysmographic Time-Domain Heart Rate Measurement Algorithm for Resource-Constrained Wearable Devices and its Implementation. Sensors. 2020;20:1783. doi: 10.3390/s20061783. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139.Lee J., Kim M., Park H.-K., Kim I.Y. Motion Artifact Reduction in Wearable Photoplethysmography Based on Multi-Channel Sensors with Multiple Wavelengths. Sensors. 2020;20:1493. doi: 10.3390/s20051493. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Perpetuini D., Chiarelli A.M., Vinciguerra V., Vitulli P., Rinella S., Cardone D., Bianco F., Perciavalle V., Gallina S., Fallica G., et al. Integrated Multi-channel PPG and ECG System for Cardiovascular Risk Assessment. Proceedings. 2019;27:8. doi: 10.3390/proceedings2019027008. [DOI] [Google Scholar]
- 141.Georgieva-Tsaneva G., Gospodinova E., Gospodinov M., Cheshmedzhiev K. Portable Sensor System for Registration, Processing and Mathematical Analysis of PPG Signals. Appl. Sci. 2020;10:1051. doi: 10.3390/app10031051. [DOI] [Google Scholar]
- 142.Millán C.A., Girón N.A., Lopez D.M. Analysis of relevant features from photoplethysmographic signals for atrial fibrillation classification. Int. J. Environ. Res. Public Health. 2020;17:498. doi: 10.3390/ijerph17020498. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 143.Rundo F., Petralia S., Fallica G., Conoci S. A Nonlinear Pattern Recognition Pipeline for PPG/ECG Medical Assessments. In: Andò B., Baldini F., Di Natale C., Ferrari V., Marletta V., Marrazza G., editors. Convegno Nazionale Sensori. Springer International Publishing; Cham, Switzerland: 2019. pp. 473–480. [Google Scholar]
- 144.Jagelka M., Jeleň M., Vavrinský E., Daříček M., Donoval M. Implementation of pulse oximetry measurement to wireless biosignals probe. Lek. Tech. 2014;44:37–40. [Google Scholar]
- 145.Hosanee M., Chan G., Welykholowa K., Cooper R., Kyriacou P.A., Zheng D., Allen J., Abbott D., Menon C., Lovell N.H., et al. Cuffless Single-Site Photoplethysmography for Blood Pressure Monitoring. J. Clin. Med. 2020;9:723. doi: 10.3390/jcm9030723. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 146.He K., Zhang X., Ren S., Sun J. Deep residual learning for image recognition. Proc. IEEE Comput. Soc. Conf. Comput. Vis. Pattern Recognit. 2016;2016:770–778. [Google Scholar]
- 147.Hardoon D.R., Szedmak S., Shawe-Taylor J. Canonical correlation analysis: An overview with application to learning methods. Neural Comput. 2004;16:2639–2664. doi: 10.1162/0899766042321814. [DOI] [PubMed] [Google Scholar]
- 148.Rundo F., Ortis A., Battiato S., Conoci S. Advanced bio-inspired system for noninvasive cuff-less blood pressure estimation from physiological signal analysis. Computation. 2018;6:46. doi: 10.3390/computation6030046. [DOI] [Google Scholar]
- 149.Liang Y., Chen Z., Ward R., Elgendi M. Photoplethysmography and deep learning: Enhancing hypertension risk stratification. Biosensors. 2018;8:101. doi: 10.3390/bios8040101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 150.Slapničar G., Mlakar N., Luštrek M. Blood Pressure Estimation from Photoplethysmogram Using a Spectro-Temporal Deep Neural Network. Sensors. 2019;19:3420. doi: 10.3390/s19153420. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Welykholowa K., Hosanee M., Chan G., Cooper R., Kyriacou P.A., Zheng D., Allen J., Abbott D., Menon C., Lovell N.H., et al. Multimodal Photoplethysmography-Based Approaches for Improved Detection of Hypertension. J. Clin. Med. 2020;9:1203. doi: 10.3390/jcm9041203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 152.Chen Y., Wen C., Tao G., Bi M., Li G. Continuous and noninvasive blood pressure measurement: A novel modeling methodology of the relationship between blood pressure and pulse wave velocity. Ann. Biomed. Eng. 2009;37:2222–2233. doi: 10.1007/s10439-009-9759-1. [DOI] [PubMed] [Google Scholar]
- 153.Padilla J.M., Berjano E.J., Sáiz J., Fácila L., Díaz P., Mercé S. Assessment of relationships between blood pressure, pulse wave velocity and digital volume pulse. Comput. Cardiol. 2006;33:893–896. [Google Scholar]
- 154.Yoon Y., Cho J.H., Yoon G. Non-constrained blood pressure monitoring using ECG and PPG for personal healthcare. J. Med. Syst. 2009;33:261–266. doi: 10.1007/s10916-008-9186-0. [DOI] [PubMed] [Google Scholar]
- 155.Liu W., Fang X., Chen Q., Li Y., Li T. Reliability analysis of an integrated device of ECG, PPG and pressure pulse wave for cardiovascular disease. Microelectron. Reliab. 2018;87:183–187. doi: 10.1016/j.microrel.2018.06.008. [DOI] [Google Scholar]
- 156.Shin W., Cha Y.D., Yoon G. ECG/PPG integer signal processing for a ubiquitous health monitoring system. J. Med. Syst. 2010;34:891–898. doi: 10.1007/s10916-009-9304-7. [DOI] [PubMed] [Google Scholar]
- 157.Jeon G.-R., Jung D.-K., Kim G.-R., Shin B.-J. The Development of Integrated Sensor System for Measuring Simultaneously ECG, PPG and PPW. J. Korea Acad. Coop. Soc. 2009;10:992–999. [Google Scholar]
- 158.GlutracTM. [(accessed on 26 March 2020)]; Available online: https://www.add-care.net/
- 159.Rundo F., Conoci S., Ortis A., Battiato S. An Advanced Bio-Inspired PhotoPlethysmoGraphy (PPG) and ECG Pattern Recognition System for Medical Assessment. Sensors. 2018;18:405. doi: 10.3390/s18020405. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 160.Liu S.-H., Liu L.-J., Pan K.-L., Chen W., Tan T.-H. Using the Characteristics of Pulse Waveform to Enhance the Accuracy of Blood Pressure Measurement by a Multi-Dimension Regression Model. Appl. Sci. 2019;9:2922. doi: 10.3390/app9142922. [DOI] [PubMed] [Google Scholar]
- 161.Di Rienzo M., Rizzo G., Işilay Z.M., Lombardi P. SeisMote: A Multi-Sensor Wireless Platform for Cardiovascular Monitoring in Laboratory, Daily Life, and Telemedicine. Sensors. 2020;20:680. doi: 10.3390/s20030680. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 162.Lazazzera R., Belhaj Y., Carrault G. A New Wearable Device for Blood Pressure Estimation Using Photoplethysmogram. Sensors. 2019;19:2557. doi: 10.3390/s19112557. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 163.OMRON HeartGuideTM. [(accessed on 26 March 2020)]; Available online: https://omronhealthcare.com/products/heartguide-wearable-blood-pressure-monitor-bp8000m/
- 164.Niknejad N., Ismail W.B., Mardani A., Liao H., Ghani I. A comprehensive overview of smart wearables: The state of the art literature, recent advances, and future challenges. Eng. Appl. Artif. Intell. 2020;90:103529. doi: 10.1016/j.engappai.2020.103529. [DOI] [Google Scholar]
- 165.Silva M.C., Amorim V.J.P., Ribeiro S.P., Oliveira R.A.R. Field Research Cooperative Wearable Systems: Challenges in Requirements, Design and Validation. Sensors. 2019;19:4417. doi: 10.3390/s19204417. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 166.Williamson J., Liu Q., Lu F., Mohrman W., Li K., Dick R., Shang L. Data sensing and analysis: Challenges for wearables; Proceedings of the 20th Asia and South Pacific Design Automation Conference; Chiba, Japan. 19–22 January 2015; pp. 136–141. [Google Scholar]
- 167.Orfanidis C., Dimitrakopoulos K., Fafoutis X., Jacobsson M. No Batteries Needed: Providing Physical Context with Energy-Harvesting Beacons; Proceedings of the 7th International Workshop on Energy Harvesting & Energy-Neutral Sensing Systems—ENSsys′19; New York, NY, USA. 10 November 2019; pp. 52–53. [Google Scholar]
- 168.Steinberg C., Philippon F., Sanchez M., Fortier-Poisson P., O’Hara G., Molin F., Sarrazin J.F., Nault I., Blier L., Roy K., et al. A novelwearable device for continuous ambulatory ECG recording: Proof of concept and assessment of signal quality. Biosensors. 2019;9:17. doi: 10.3390/bios9010017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 169.Fouassier D., Roy X., Blanchard A., Hulot J.S. Assessment of signal quality measured with a smart 12-lead ECG acquisition T-shirt. Ann. Noninvasive Electrocardiol. 2020;25:1–7. doi: 10.1111/anec.12682. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 170.Movesense Sports Bra. [(accessed on 26 March 2020)]; Available online: https://www.movesense.com/product/movesense-sports-bra/
- 171.Soh P.J., Vandenbosch G.A.E., Mercuri M., Schreurs D.M.M.P. Wearable wireless health monitoring: Current developments, challenges, and future trends. IEEE Microw. Mag. 2015;16:55–70. doi: 10.1109/MMM.2015.2394021. [DOI] [Google Scholar]
- 172.Ortega L., Llorella A., Esquivel J.P., Sabaté N. Self-powered smart patch for sweat conductivity monitoring. Microsyst. Nanoeng. 2019;5:3. doi: 10.1038/s41378-018-0043-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 173.Ahmad R., Wolfbeis O.S., Alshareef H.N., Salama K.N., Surya S.G., Batista Sales J., Mkaouar H., Cavalcanti Catunda S.Y., Rodrigues Belfort D., Lei Y., et al. KAUSTat: A Wireless, Wearable, Open-Source Potentiostat for Electrochemical Measurements. Proc. IEEE Sensors. 2019;2019:1–4. [Google Scholar]
- 174.Svobodova H., Vavrinsky E., Turonova D., Donoval M., Daricek M., Telek P., Kopani M. Optimization of the position of single-lead wireless sensor with low electrodes separation distance for ECG-derived respiration. Adv. Electr. Electron. Eng. 2018;16:528–537. doi: 10.15598/aeee.v16i4.2773. [DOI] [Google Scholar]
- 175.Kabir M.M., Perez-Alday E.A., Thomas J., Sedaghat G., Tereshchenko L.G. Optimal configuration of adhesive ECG patches suitable for long-term monitoring of a vectorcardiogram. J. Electrocardiol. 2017;50:342–348. doi: 10.1016/j.jelectrocard.2016.12.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 176.Dias D., Cunha J.P.S. Wearable Health Devices—Vital Sign Monitoring, Systems and Technologies. Sensors. 2018;18:2414. doi: 10.3390/s18082414. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 177.Motti V.G., Caine K. Users’ Privacy Concerns about Wearables. In: Brenner M., Christin N., Johnson B., Rohloff K., editors. Financial Cryptography and Data Security. Springer; Berlin/Heidelberg, Germany: 2015. pp. 231–244. [Google Scholar]
- 178.Perez A.J., Zeadally S. Privacy Issues and Solutions for Consumer Wearables. IT Prof. 2018;20:46–56. doi: 10.1109/MITP.2017.265105905. [DOI] [Google Scholar]
- 179.Schaub F., Balebako R., Durity A.L., Cranor L.F. A design space for effective privacy notices; Proceedings of the Eleventh Symposium on Usable Privacy and Security (SOUPS); Santa Clara, CA, USA. 11–13 August 2019; pp. 1–17. [Google Scholar]
- 180.Guler S.D., Gannon M., Sicchio K. Crafting Wearables. Apress; Berkeley, CA, USA: 2016. [Google Scholar]
- 181.Kotradyova V., Vavrinsky E., Kalinakova B., Petro D., Jansakova K., Boles M., Svobodova H. Wood and Its Impact on Humans and Environment Quality in Health Care Facilities. Int. J. Environ. Res. Public Health. 2019;16:3496. doi: 10.3390/ijerph16183496. [DOI] [PMC free article] [PubMed] [Google Scholar]