Abstract
Focused ultrasound (FUS) neuromodulation has shown that mechanical waves can interact with cell membranes and mechanosensitive ion channels, causing changes in neuronal activity. However, the thorough understanding of the mechanisms involved in these interactions are hindered by different experimental conditions for a variety of animal scales and models. While the lack of complete understanding of FUS neuromodulation mechanisms does not impede benefiting from the current known advantages and potential of this technique, a precise characterization of its mechanisms of action and their dependence on experimental setup (e.g., tuning acoustic parameters and characterizing safety ranges) has the potential to exponentially improve its efficacy as well as spatial and functional selectivity. This could potentially reach the cell type specificity typical of other, more invasive techniques e.g., opto- and chemogenetics or at least orientation-specific selectivity afforded by transcranial magnetic stimulation. Here, the mechanisms and their potential overlap are reviewed along with discussions on the potential insights into mechanisms that magnetic resonance imaging sequences along with a multimodal stimulation approach involving electrical, magnetic, chemical, light, and mechanical stimuli can provide.
Keywords: central nervous system, focused ultrasound, magnetic resonance imaging, peripheral nervous system, therapeutic ultrasound, ultrasound neuromodulation
Introduction
The ability to probe spatially specific brain regions enable understanding brain functioning and connectivity. In turn, this can unlock a wealth of potential investigative and therapeutic applications. Focused Ultrasound (FUS) has been proven capable of eliciting excitatory and inhibitory effects non-invasively and locally in the central nervous system (CNS) and peripheral nervous system (PNS), depending on the adopted pulsing regime [1]. Several studies have demonstrated the elicitation of motor responses in rodents obtained from the FUS stimulation of cortical brain regions [2–6]. Furthermore, the capability of FUS to reach deeper brain structures (which is one of the major challenges of other current- or voltage-controlled neuromodulation techniques) can provide access to subcortical areas of the brain. For example, the stimulation of deep-seated structures such as locus coeruleus and superior colliculus caused pupil dilation and eyeball movements in mice [6]. Also, FUS stimulation of the putamen induced improvements in speed and accuracy of visual-motor tasks in non-human primates (NHPs) [7]. In humans, targeting the head of caudate resulted in hemodynamic responses visible through functional magnetic resonance imaging (fMRI) [8] and stimulating the thalamus induced changes in somatosensory evoked responses [9].
The mechanisms proposed to explain the FUS neuromodulation effects are based on multiple hypothesis on how ultrasound interferes with depolarization through mechanical deformation of the cell membrane. In addition, experimental evidences have shown that ultrasound can activate mechanosensitive ion channels in neurons [10–13] and other brain cell types like astrocytes [14], providing additional avenues for FUS to interfere with the membrane potential. Despite the advances provided by in vitro, ex vivo, and in vivo experiments, the high variability in experimental conditions and setups, as well as partially conflicting results, has led to somewhat contradictory interpretations and a variety of possible hypotheses about underlying physiological mechanisms, which may be acting concurrently in a dynamic interplay every time FUS is applied. Moreover, most current animal experiments are performed under anesthesia. The interaction of pharmacological sedation with FUS neuromodulation is not entirely understood and may partially obfuscate the interpretation of a number of FUS neuromodulation experiments [15].
The use of magnetic resonance imaging (MRI) can provide insights into brain structure and activity and hence support FUS-based neuromodulation through targeting, safety evaluation, and the evaluation of brain function and mechanisms. In this context, multimodal stimulation coupled with neuroelectric or MRI may present a better opportunity to understanding of the multiple factors that play a role in neuron functioning as well as how FUS interferes with it.
In this review, the proposed mechanisms for ultrasound neuromodulation and interactions of FUS with tissue are revisited, and current contradictory findings are discussed in light of varying experimental conditions and anesthesia effects. Finally, the potential of multimodal stimulation and the use of MRI is discussed as a promising future avenue for spatiotemporally selective, non-invasive neuromodulation.
Mechanisms of ultrasound neuromodulation
Ultrasound propagation in biological tissue is characterized by vibrational waves traveling with frequencies above the hearing range (>20 kHz). In the compressional phase, ultrasound displaces tissue particles and fluid molecules, generating an elastic restoring force. As the tissue and fluids return to their normal configurations, molecules experience a rarefaction phase. During this process, waves propagate through the tissue giving rise to an acoustic radiation force (ARF) where part of the energy is stored in the tissue in the form of elastic deformation, and part is dissipated as heat due to viscous frictional forces. When acoustic wave flow experiences opposition due to acoustic impedance discontinuities, parts of the wave are transmitted, reflected, and refracted. Both scattering and heating dissipation are frequency-dependent where energy deposition in the medium occurs through absorption. The scattered waves can be subsequently partially absorbed and partially re-scattered multiple times. Other effects during the rarefaction phase can occur, such as cavitation (nucleation) [16], which has a higher probability of occurring at higher pressures and lower frequencies. Potential mechanisms for ultrasound neuromodulation are associated with changes in membrane potential due to ultrasound-induced neuronal membrane deformation and the activation of mechanosensitive channels (Table 1) (see Jerusalem et al. [15] for a review). In this context, both theoretical and experimental studies have proposed that mechanical deformations induced by strain gradients produce a membrane polarization, giving rise to a flexoelectric effect [17,18]. A study using a model lipid bilayer membrane demonstrated that the displacement of the membrane caused by the ARF results in changes in the membrane area and its capacitance, which in turn creates capacitive currents measured with voltage-clamp techniques [19]. Another recent study evoked neuronal calcium responses obtained from local mechanical indentation delivered by a piston in cultured rat cortical and hippocampal neurons [20], giving experimental evidences that neurons are sensitive to mechanical stress. Also, 196 Muratore et al. [21] have shown that ARF can deform the cell 197 membrane. Intriguingly, other theoretical studies have proposed that the rarefactional phase of ultrasound waves can pull apart the two membrane lipid leaflets, leading the formation of bubbles in the intramembrane space, which in turn induces currents by modulating membrane capacitance in an oscillatory manner [22,23]. However, an ex vivo study has shown that micron-scale tissue displacements consistent with ARF generation triggered spiking activity that remained unchanged to a broad acoustic frequency range (0.5 to 43 MHz), hence excluding a potential cavitation-related effect at least in an ex vivo setting [24]. Moreover, a new theory known as the soliton model proposes that the action potential (AP) involves an adiabatic process, where a mechanical pulse propagates in phase with an electrical pulse along the axon [25]. The reversed pathway could mean that deformations of the neuronal membrane induced by the ARF could potentially both annihilate or enhance axonal electrophysiology [26]. Also, for specific regimes (high pulse repetition frequency, high duty cycle, high pressure), ultrasound may increase temperature and alter the electrical capacitance of the plasma membrane [27], as demonstrated through light-induced temperature increase [28]. Interestingly, a behavioral study using mutants C. elegans model demonstrated that knocking out mechanosensitive ion channels abolishes neuronal responses to mechanical stimulation, while knocking out thermosensitive ion channels kept responses unaffected [29]. In this context, Thompson et al. (1985) have demonstrated a temperature dependence of neuronal membrane conductance and synaptic potentials [30], while recent studies have shown that ultrasound can directly drive a number of mechanosensitive ion channels (K+ channel family TREK-1, TREK-2, and TRAAK [11], voltage gated Na+ and Ca+ [10], and piezo type mechanosensitive channel Piezo1 [12,13] and Piezo2 [31]) as well as and indirectly control neuronal responses via modulation of TRPA1 (transient receptor potential ankyrin 1) channels in astrocytes with glutamate-releasing bestrophin-1 (Best1) as a mediator of glia-neuron interaction [14]. Therefore, it is highly likely that, depending on the pulse regime, different combinations of partially overlapping mechanisms would concur to the final result of the interaction between ultrasound and the cell membrane.
Table 1.
Mechanisms | Description |
---|---|
Membrane deformation causing capacitance changes | Capacitance changes have been observed during artificial membrane deflection [19] and deformation of in vitro membranes [20,21] and modeled in simulations [26]. Capacitance can be altered by membrane volume changes or be associated with a flexoelectric effect (a property of the membrane that causes a spontaneous electric polarization when submitted to a mechanical strain gradient [18]). |
Soliton model | Changes in membrane conformation could arise from interfering with a thermodynamic process involved in electromechanical pulse traveling during AP [25]. |
Intramembrane cavitation model | Ultrasound-induced intramembrane cavitation within the bilayer membrane induces a current through membrane capacitance changes [22]. |
Mechanosensitive ion channels modulation | A number of mechanosensitive ion channels has seen in vitro to be sensitive to ultrasound waves (TREK-1, TREK-2, TRAAK [11]; voltage-gated Na+ and Ca+ [10]; Piezo1 [12,13]; and Piezo2 [31]). |
Modulation of TRPA1 channels in astrocytes | Ultrasound opens TRPA1 channels in astrocytes, inducing glutamate- releasing Best1 as a mediator of glia-neuron interaction [14]. |
Thermal modulation | Heating reversibly alters the membrane capacitance, resulting in depolarization [27,28]. FUS can increase temperature at specific regimes. Neuronal membrane conductance and synaptic potentials are altered by temperature changes [30]. |
AP, action potential; FUS, focused ultrasound; TRPA, transient receptor potential ankyrin.
Ex vivo/in vitro versus in vivo
Despite the advances provided by ex vivo and in vitro studies, contradictory results regarding the absence [24] or presence [32,33] of cavitation and its role [22] in ultrasound neuromodulation have been reported. These conflicts may potentially be due to differences in experimental conditions. For instance, the oxygenation process inherent to culturing cells may introduce bubbles in in vitro preparations [32]. Furthermore, in vivo translation of in vitro and ex vivo results is hampered by differences in a number of parameters and effects such as cavitation threshold, the rapid cooling effects associated with brain perfusion [34], the contribution of different cells to the neuromodulatory effect [14] and skull-related effects such as attenuation due to absorption and scattering, and shear wave from mode conversion [35]. Indirect confounding effects may also include activation through auditory pathways [36,37]. Nevertheless, all ultrasound neuromodulation studies have demonstrated that the paradigm of framing neural activity within and electromagnetic perspective is too simplistic, confirming that ultrasound neuromodulation studies can be of great aid in all applications requiring fast and painless interference of brain function, both in investigative and in therapeutic contexts.
In vivo studies - Anesthesia effects
Anesthesia effects have long represented a major confounding factor in neuromodulation studies. It has been shown that motor-evoked potentials induced by electrical stimulation are suppressed by isoflurane in a dose-dependent manner [38]. Similarly, ketamine blocks cortical neuron activity, which suppresses ultrasound-elicited motor responses [39]. In this context, in FUS neuromodulation studies, the isoflurane dose was reduced down to 0.1%, which corresponds to operating on a semi awake animal [4]. However, some experiments have reported auditory artifacts and audible buzzing sounds generated by the ultrasound transducer, which may affect experiments in animals [36,37], as well as in humans [40–42]. Therefore, the use of low-level anesthesia to maintain animal semi alert requires careful considerations in the setup and techniques such as signal smoothing [43] to avoid confounds. From deep brain stimulation (DBS) studies, it is known that anesthesia affects the spontaneous background firing and the neuronal spike activity patterns, as well as potentiates the inhibitory actions of gamma-aminobutyric acid (GABA) and causes a global depression in neuronal discharge, among other effects [44,45]. In a repetitive transcranial magnetic stimulation (rTMS) study in rats, isoflurane, dexmedetomidine, and propofol caused significant different effects on functional connectivity, particularly between the sensorimotor cortex and thalamus [46]. In general, as reviewed by Jerusalem et al. [15], anesthetics lead to unconsciousness, immobility, amnesia, and analgesia without a complete understanding of the mechanisms underlying loss of consciousness and depth of anesthesia, which is mirrored by an even more partial insights into the implication of sedation and deep anesthesia in humans [47] to the extent that anesthesia itself can be considered an instrument to explore the neural substrates of cognitive processes [48]. Importantly, several hypotheses about how anesthetic drugs modulate membrane excitability overlap with potential mechanisms of FUS neuromodulation. These include membrane deformation, changes in the thermodynamic properties of the membrane, and bubble formation. Therefore, awake studies are needed for a more precise characterization of the neural underpinning of FUS neuromodulation.
MRI
MRI can help advance neuromodulation technologies in a number of ways [49]. Importantly, MRI and ultrasound neuromodulation share similar spatial resolutions, which lies in the order of millimeters or sub-millimeters. MRI resolution depends on several factors, including magnetic field strength [50–52], coil performance, and subsequent imaging gradients [53,54]. High magnetic field strengths from 3 to 7 T for humans [55] and above 7 T for preclinical studies [56] dedicated multi-transmit head coils [57,58], and strong imaging gradients up to 100 mT/m for human scanners and 1000 mT/m for preclinical systems [59–61] can provide spatial resolutions ranging from 1–2 mm3 to submillimeter (depending on imaging modality) for human and animals studies, respectively [60,62].
On the other side, the lateral resolution (L) of ultrasound neuromodulation for a concave transducer can be characterized as L=1.4λF/A, where λ is the wavelength (equal the ratio of the speed of sound in the medium and the ultrasound frequency), F is the focal length, and A is the aperture size (F/A is also known as the f-number). However, the frequency dependence of the ultrasound attenuation factor, mainly influenced by the skull, imposes a trade-off in the frequency choice. The attenuation factor is given by α0fn, where α0 is a temperature-dependent attenuation factor at 1 MHz, f is the ultrasound frequency, and n lies in the range of 0.9 to 2.1 for the human skull bone and 1.05 to 1.1 for brain [63]. Typically, ultrasound neuromodulation delivers submillimetric to millimetric resolution that employs frequencies in the kHz range for humans (i.e. f=250 kHz, L=7 mm [64]) and non-human primates (i.e. f=320 kHz, L=5 mm [64]), and kHz to MHz range for rodents (i.e. f=1.9 MHz, L=1mm [6], and f=5 MHz, L= 0.29 mm [65]).
Motion sensitizing gradients can detect phase shifts in MR data that encode brain tissue displacements caused by FUS application [66,67]. This specific MRI technique, called magnetic resonance-acoustic radiation force imaging (MR-ARFI), has been shown to be safe despite the need for high-intensity FUS pulses to displace tissue [68]. Currently, just like in transcranial magnetic stimulation (TMS, which employs pulsed magnetic fields to induce eddy currents in the brain) or transcranial direct current stimulation (tDCS), neuromodulation studies rely on numerical simulations to perform targeting. However, a confirmation of tissue engagement through MR-ARFI would be highly desirable, especially for small brain structures. In this context, targeting accuracy can be improved by using low-frequency ranges and normal incidence angles [69] both minimizing FUS beam distortions and by adopting neuronavigation systems based on MR images [70].
MR phase-difference images can also be used for temperature monitoring during FUS [71,72] in order to avoid artifacts that would arise from local temperature measurements based on thermocouples [73,74]. While no significant temperature elevation has been detected in low-intensity neuromodulation protocols [75], higher intensity protocols [6] may cause physiologically relevant temperature elevations [74,76], and monitoring temperature may provide insights into FUS neuromodulation mechanisms. Other MRI modalities, such as T2-weighted and T2*-weighted imaging, can provide safety evaluation such as the detection of potential hemorrhages and edema formation [77]. Also, T2-weighted Fluid Attenuated Inversion Recovery (FLAIR) can provide safety assessment with better differentiation between CSF and abnormal tissue [78]. In addition, Diffusion Weighted Imaging (DWI) is highly sensitive to both reversible and irreversible changes in brain microstructure [79]. Moreover, in order to reveal intentional [7] or unintentional breakdown in the blood-brain barrier in the context of neuromodulation, T2- or T1-weighted MR images can be used to evaluate T2 or T1 contrast agents deposition in brain tissue after ultrasound application [80–82]. Finally, fMRI has been used in NHP to identify brain areas to be modulated [64] or to reveal the extent and connectivity of spatial changes in hemodynamic responses caused by FUS [8,83–87].
Other neuromodulation techniques: multimodal stimulation
In general, amongst the numerous techniques available for neuromodulation, keeping more and more of the neurophysiology under experimental control goes hand in hand with an increase in invasiveness and biotechnological constraints. For example, neuronal activity of specific neuronal populations could be reversibly silenced by genetic approaches [88] while FUS would probe specific brain structures. This could potentially reveal the spatial and temporal scales of the different mechanisms of action, the contribution of FUS neuromodulation in different brain cells, and the contribution of defined projection pathways to neuronal network dynamics and animal behavior. On the noninvasive side, which is more immediately translatable to humans, combining magnetic and ultrasound stimuli is capable of enhancing the effect of FUS [89]. It has been proposed that ions in motion under a static magnetic field could be subjected to a Lorentz force, giving rise to electric currents that would contribute to the neuromodulatory effect of FUS [89,90]. For example, in humans, concurrent FUS and TMS applied to the primary motor cortex (M1) attenuated motor evoked potential amplitude, reduced intracortical facilitation, and slightly shortened (10 μs) the response time in visual tasks [42]. Notably, FUS parameters can span a range of values that has been shown capable of inducing mechanical [91] or thermal effects to obtain excitatory or inhibitory effects on mice sciatic nerve [92]. In general, the ability of FUS to probe spatially specific brain regions enables understanding of, e.g., brain functioning and connectivity in non-invasive and spatially selective manner, with little or no cell type specificity. In this respect, FUS is somewhat similar to TMS, although it may offer better focusing of deeper structures (at least with a single coil) [93,94]. Interestingly, TMS and FUS still share the large potential for noninvasive brain enhancing and silencing, as well as the lack of a thorough understanding of the mechanisms of action underlying the diversity of effects observed throughout the literature, which may include involuntary cell type specificity, axonal stimulation [95], uncontrolled/uncontrollable activation at different loci of the neuron, distributed stimulation peaks [96] and complex interplay of modulating inhibitory and excitatory synaptic potentials [97]. In this context, current-controlled “priming” techniques such as tDCS can be used in conjunction with time-localized TMS [98] (or possibly FUS) to modify the underlying neuronal activity substrate and possibly enhance specificity.
Conversely, techniques such as optogenetics [99] and chemogenetics/pharmacogenetics [100] can provide cell-type specific, selectively inhibitory, excitatory or combined control of neuronal activity by expressing light-sensitive ion channels called opsins, which can be either excitatory (e.g., channelrhodopsin) or inhibitory (e.g., halorhodopsin). The specificity [100] may be selectively activated [101] with light at different frequencies allowing a virtually infinite combination of stimuli, which can open/close ion channels with extremely high frequencies (up to 30 Hz). The drawbacks of such techniques lie both in practical challenges like, e.g., the implantation of fiber optics for stimulation (which may interfere with behavioral experiments and limit human translational potential) and the high spatial selectivity (200 microns) of light delivery (which, interestingly may not suffice to inhibit the function of a particular brain region and hence examine its function) and in neurobiological effects such the need to genetically modify the organisms to achieve cell-type specificity, nonphysiological hyperpolarization (which in turn can generate rebound phenomena) and in the potential generation of antidromic potentials (which may blur the physiological significance of the stimulation). While the first set of constraints may be partially solved by pharmacogenetics approaches (which employ chemical stimuli to activate the opsins, and hence eliminate the surgical requirements and the need for constant stimulation when envisaging future treatment strategies in humans), the second may not. This calls for a new generation of combined biotechnological and physical neuromodulation techniques in order to achieve successful translation to the human context, especially in the therapeutic and clinical trial arena.
Interestingly, novel paradigms have been proposed involving the combination of genetic approaches with either magnetic or ultrasound stimulation. In magnetogenetics, thermo-sensitive and mechanosensitive ion channels (typically transient receptor potential vanilloid class receptors TRPV, which are selective calcium Ca2+ transporters) are genetically engineered to be tightly coupled to the ironstorage protein ferritin (or another paramagnetic protein), so that they can be activated by external magnetic fields [102]. In sonogenetics, through a similar approach, it has been demonstrated that neuron-specific misexpression of TRP-4 (a pore-forming subunit of a mechanotransduction channel) can sensitize neurons to US stimuli with detectable behavioral outputs [103]. It appears, therefore, evident that combined, multimodal strategies are the principal future avenue for tailoring neuromodulation intervention to an application-specific and possibly patient-specific context within a precision medicine paradigm.
Discussion
In this review, we have summarized potential mechanisms underlying the neural substrates of FUS neuromodulation and outlined conflicting hypotheses of the current literature. Similar to what has been shown for TMS, it is our opinion that apparent contradictions observed in some experimental and modeling studies could be resulted mainly due to variability from different experimental conditions in vitro, ex vivo, and in vivo applications and that they could be reconciled by detailed standardization and translation studies. In turn, this would allow drawing more informed conclusions on the FUS neuromodulation mechanisms. Additionally, the lack of a complete understanding of anesthesia effects on neurons encourages further awake FUS neuromodulation studies, which with the aid of MRI in assessing brain activity, targeting, and safety, will provide a clearer picture of both the neurophysiological underpinnings and of the potential translational applications of FUS, whether alone or in a multimodal context.
A number of experimental evidences show that the AP involves an electro-mechanical process and that the deformation of tissues induced by the ARF plays a crucial role in neuromodulation through potentially capacitance changes modulation or a flexoelectric effect triggering. Another possibility is that FUS could cause a neuronal membrane deformation capable of interfering with membrane electrical depolarization by mechanical coupling with the endogenous mechanical waves (soliton) associated with action potentials. In addition, ultrasound propagation can deform tissues elastically while the pulse energy is lost through heating due to viscous frictional forces. Whether the thermal effect is detectable or important in the context of neuromodulation will depend on the temperature level that is reached, neuronal sensitivity to temperature transients, tissue diffusion, and perfusion capability. In the soliton model, the membrane temperature is a crucial factor, and it should be noted that the membrane melting point is slightly below physiological temperature. Therefore, small temperature elevations caused by viscous frictional forces during ultrasound propagation may cause an interference with electromechanical membrane physiology. Most studies have consistently strived to avoid thermal effects from FUS effects, which is important to separate ablative from non-ablative effects. However, the mechanical and low-temperature increase generated by FUS could also potentially improve the neuromodulatory effects [104]. In this context, animal experiments based on sedation or anesthesia need to take thermal effects into account as mild hypothermia is common during deep sedation [105].
While the lack of a complete understanding of the FUS neuromodulation mechanisms does not currently impede reaping potential benefits in a more application-driven context, it is reasonable to expect that a better mechanistic understanding will immediately reverberate onto the applicability and efficacy of FUS-based neuromodulation. Importantly, as the technology continues to gain ground and acceptance, safety must remain a prime concern. Therefore, overcoming current limitations in both target confirmation and safety monitoring through the human skull is imperative. Techniques such as mapping of cavitation, temperature, and displacement will ensure a successful clinical translation of US neuromodulation, at the same time providing more control over the acoustic parameters. This will allow employing precisely determined mechanism combinations for achieving targeted neuronal excitation and inhibition. While more and more studies are being planned, several investigations have already demonstrated the existence of a wide range of safe parameters [106–108].
While fMRI is undoubtedly the gold standard for functional brain imaging, other techniques can provide complementary information on brain function. Recent studies have combined fMRI and optical imaging to show that US neuromodulation induces cerebral hemodynamic changes in different animals at variable peak latencies: mice ~2.5 s [109,110], rabbits ~3.2 s [111], and NHP ~6.5 s [86]. Furthermore, a technique termed functional ultrasound (fUS) is capable of detecting transient changes in blood volume [112], and it has been demonstrated capable of providing deep brain functional images with high spatial resolution (from 50 to 200 um) [113] and temporal resolution of less <1 s [114]. In addition, fUS features high sensitivity and portability, which enable awake experiments with freely moving subjects. The development of 2D transducer arrays [117], capable of generating images and steerable, highly focused beams, potentially with multifrequency capability [118], may facilitate human fUS during the US neuromodulation.
Crucially, while in vitro and ex vivo studies are necessary for understanding mechanisms, in vivo brain activity studies are essential to gather mesoscale and macroscale information about the effect of FUS on brain functioning. In small animals, experiments in ultra-high-field (UHF) MRI (7 to 21 T) can provide higher signal-to-noise and contrast-to-noise ratios as well as in the case of fMRI increased susceptibility [60,119]. In turn, this will unlock more in-depth insights into how the intact brain works and into the available windows in interfering with its activity in a non-invasive or minimally invasive manner, accelerating the translation towards human applications and especially the empowerment of clinical trials. This may include applications to neurological diseases like epilepsy and chronic pains, psychiatry [obsessive compulsive 633 disorder (OCD), pharmacoresistant depression, agoraphobia], as well as fostering neuronal plasticity in, e.g., rehabilitation or slowing the progression of degenerative brain diseases. Especially in this latter context, multimodal stimulation (as electrical, magnetic, chemical, light, mechanical), possibly coupled with a state-of-the-art monitoring tool like UHF MRI for non-invasive techniques and calcium imaging [120] may enable simultaneous, multi-scale, brain structure- or cell type-specific silencing or excitation, allowing the exploration of both brain-wide pathways as well as specific cognitive, emotional and pathological mechanisms. This can provide a significant step change in keeping more and more neurophysiological aspects under experimental control, and hence ultimately approaching the neurobiological goal of neuromodulation in a more precise, targeted, painless, and direct manner.
Funding
HK and EK were supported in part by National Institutes of Health (NIH) under award numbers R01CA228275 and R01EB027576, and the Defense Advanced Research Projects Agency (DARPA) Biological Technologies Office (BTO) Electrical Prescriptions (ElectRx) under award number HR0011-15-2-0054. This study received funding from SoundStim Therapeutics and Google X. The funders were not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication.
Footnotes
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
References
- [1].Blackmore J, Shrivastava S, Sallet J, Butler CR and Cleveland RO 2019. Ultrasound Neuromodulation: A Review of Results, Mechanisms and Safety Ultrasound Med. Biol 45 1509–36 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Yoo SS, Kim H, Min BK, Franck E and Park S 2011. Transcranial focused ultrasound to the thalamus alters anesthesia time in rats Neuroreport 22 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Younan Y, Deffieux T, Larrat B, Fink M, Tanter M and Aubry J-F 2013. Influence of the pressure field distribution in transcranial ultrasonic neurostimulation Med. Phys 40 082902. [DOI] [PubMed] [Google Scholar]
- [4].King RL, Brown JR, Newsome WT and Pauly KB 2013. Effective parameters for ultrasound-induced in vivo neurostimulation Ultrasound Med. Biol 39 312–31 [DOI] [PubMed] [Google Scholar]
- [5].Kim H, Chiu A, Lee SD, Fischer K and Yoo S-S S 2014. Focused ultrasound-mediated non invasive brain stimulation: Examination of sonication parameters Brain Stimul. 5 181–204 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Kamimura HAS, Wang S, Chen H, Wang Q, Aurup C, Acosta C, Carneiro AAO and Konofagou EE 2016. Focused ultrasound neuromodulation of cortical and subcortical brain structures using 1.9 MHz Med. Phys 43 5730–5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Downs ME, Teichert T, Buch A, Karakatsani ME, Sierra C, Chen S, Konofagou EE and Ferrera VP 2017. Toward a Cognitive Neural Prosthesis Using Focused Ultrasound Front. Neurosci 11 607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Ai L, Mueller JK, Grant A, Eryaman Y and Legon W 2016. Transcranial focused ultrasound for BOLD fMRI signal modulation in humans 2016 38th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc 2016-Octob 1758–61 [DOI] [PubMed] [Google Scholar]
- [9].Legon W, Ai L, Bansal P and Mueller JK 2018. Neuromodulation with single-element transcranial focused ultrasound in human thalamus Hum. Brain Mapp 39 1995–2006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Tyler WJ, Tufail Y, Finsterwald M, Tauchmann ML, Olson EJ and Majestic C 2008. Remote Excitation of Neuronal Circuits Using Low-Intensity, Low-Frequency Ultrasound ed Tanimoto H PLoS One 3 e3511. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Kubanek J, Shi J, Marsh J, Chen D, Deng C and Cui J 2016. Ultrasound modulates ion channel currents Sci. Rep 6 24170. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Prieto ML, Firouzi K, Khuri-Yakub BT and Maduke M 2018. Activation of Piezo1 but Not NaV1.2 Channels by Ultrasound at 43 MHz Ultrasound Med. Biol 44 1217–32 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Qiu Z, Guo J, Kala S, Zhu J, Xian Q, Qiu W, Li G, Zhu T, Meng L, Zhang R, Chan HC, Zheng H and Sun L 2019. The Mechanosensitive Ion Channel Piezo1 Significantly Mediates In Vitro Ultrasonic Stimulation of Neurons iScience 21 448–57 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Oh S-J, Lee JM, Kim H-B, Lee J, Han S, Bae JY, Hong G-S, Koh W, Kwon J, Hwang E-S, Woo DH, Youn I, Cho I-J, Bae YC, Lee S, Shim JW, Park J-H and Lee CJ 2019. Ultrasonic Neuromodulation via Astrocytic TRPA1 Curr. Biol 29 3386–3401.e8 [DOI] [PubMed] [Google Scholar]
- [15].Jerusalem A, Al-Rekabi Z, Chen H, Ercole A, Malboubi M, Tamayo-Elizalde M, Verhagen L and Contera S 2019. Electrophysiological-mechanical coupling in the neuronal membrane and its role in ultrasound neuromodulation and general anaesthesia Acta Biomater. 97 116–40 [DOI] [PubMed] [Google Scholar]
- [16].Leighton T. The Acoustic Bubble. London: Academic Press; (1994) [Google Scholar]
- [17].Petrov AG, Miller BA, Hristova K and Usherwood PN 1993. Flexoelectric effects in model and native membarnes containing ion channels Eur Biophys J 22 [DOI] [PubMed] [Google Scholar]
- [18].Petrov AG. Electricity and mechanics of biomembrane systems: flexoelectricity in living membranes. Anal Chim Acta. 2006;568 doi: 10.1016/j.aca.2006.01.108. [DOI] [PubMed] [Google Scholar]
- [19].Prieto ML, Oralkan Ö, Khuri-Yakub BT and Maduke MC 2013. Dynamic Response of Model Lipid Membranes to Ultrasonic Radiation Force ed W Phillips PLoS One 8 e77115. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Gaub BM, Kasuba KC, Mace E, Strittmatter T, Laskowski PR, Geissler SA, Hierlemann A, Fussenegger M, Roska B and Müller DJ 2019. Neurons differentiate magnitude and location of mechanical stimuli. Proc. Natl. Acad. Sci. U. S. A [DOI] [PMC free article] [PubMed] [Google Scholar]
- [21].Muratore R, LaManna J, Szulman E, Kalisz MSA, Lamprecht M, Simon MSM, Yu MSZ, Xu N, Morrison B and Ebbini ES 2009. Bioeffective Ultrasound at Very Low Doses: Reversible Manipulation of Neuronal Cell Morphology and Function in Vitro AIP Conference Proceedings vol 1113 (American Institute of Physics) pp 25–9 [Google Scholar]
- [22].Plaksin M, Shoham S and Kimmel E 2014. Intramembrane Cavitation as a Predictive Bio-Piezoelectric Mechanism for Ultrasonic Brain Stimulation Phys. Rev. X 4 011004 [Google Scholar]
- [23].Lemaire T, Neufeld E, Kuster N and Micera S 2019. Understanding ultrasound neuromodulation using a computationally efficient and interpretable model of intramembrane cavitation J. Neural Eng 16 046007. [DOI] [PubMed] [Google Scholar]
- [24].Menz MD, Ye P, Firouzi K, Nikoozadeh A, Pauly KB, Khuri-Yakub P and Baccus SA 2019. Radiation Force as a Physical Mechanism for Ultrasonic Neurostimulation of the Ex Vivo Retina J. Neurosci 39 6251 LP–6264 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [25].Heimburg T and Jackson AD 2005. On soliton propagation in biomembranes and nerves Proc. Natl. Acad. Sci 102 9790–5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Chen H, Garcia-Gonzalez D and Jérusalem A 2019. Computational model of the mechanoelectrophysiological coupling in axons with application to neuromodulation Phys. Rev. E 99 032406. [DOI] [PubMed] [Google Scholar]
- [27].Plaksin M, Shapira E, Kimmel E and Shoham S 2018. Thermal Transients Excite Neurons through Universal Intramembrane Mechanoelectrical Effects Phys. Rev. X 8 11043 [Google Scholar]
- [28].Shapiro MG, Homma K, Villarreal S, Richter C-P and Bezanilla F 2012. Infrared light excites cells by changing their electrical capacitance Nat. Commun 3 736. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [29].Kubanek J, Shukla P, Das A, Baccus SA and Goodman MB 2018. Ultrasound Elicits Behavioral Responses through Mechanical Effects on Neurons and Ion Channels in a Simple Nervous System J. Neurosci 38 3081 LP–3091 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Thompson S, Masukawa L and Prince D 1985. Temperature dependence of intrinsic membrane properties and synaptic potentials in hippocampal CA1 neurons in vitro J. Neurosci 5 817–24 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [31].Baba Y, Hoffman BU, Lee SA, Konofagou EE, Lumpkin E. A Focused ultrasound excites action potential firing in mammalian peripheral neurons in intact tissue, Proc. Natl. Acad. Sci. U.S.A, (under review). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [32].Wright CJ, Rothwell J and Saffari N 2015. Ultrasonic stimulation of peripheral nervous tissue: an investigation into mechanisms J. Phys. Conf. Ser 581 012003 [Google Scholar]
- [33].Wright CJ, Haqshenas SR, Rothwell J and Saffari N 2017. Unmyelinated Peripheral Nerves Can Be Stimulated in Vitro Using Pulsed Ultrasound Ultrasound Med. Biol 43 2269–83 [DOI] [PubMed] [Google Scholar]
- [34].Wang H, Kim M, Normoyle KP and Llano D 2016. Thermal Regulation of the Brain—An Anatomical and Physiological Review for Clinical Neuroscientists Front. Neurosci 9 528. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Pinton G, Aubry J-F, Bossy E, Muller M, Pernot M and Tanter M 2011. Attenuation, scattering, and absorption of ultrasound in the skull bone Med. Phys 39 299–307 [DOI] [PubMed] [Google Scholar]
- [36].Sato T, Shapiro MG and Tsao DY 2018. Ultrasonic Neuromodulation Causes Widespread Cortical Activation via an Indirect Auditory Mechanism Neuron 98 1031–1041.e5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [37].Guo H, Hamilton M, Offutt SJ, Gloeckner CD, Li T, Kim Y, Legon W, Alford JK and Lim HH 2018. Ultrasound Produces Extensive Brain Activation via a Cochlear Pathway Neuron 98 1020–1030.e4 [DOI] [PubMed] [Google Scholar]
- [38].Kawaguchi M, Shimizu K, Furuya H, Sakamoto T, Ohnishi H and Karasawa J 1996. Effect of Isoflurane on Motor-evoked Potentials Induced by Direct Electrical Stimulation of the Exposed Motor Cortex with Single, Double, and Triple Stimuli in Rats Anesthesiology 85 1176–83 [DOI] [PubMed] [Google Scholar]
- [39].Han S, Kim M, Kim H, Shin H and Youn I 2018. Ketamine Inhibits Ultrasound Stimulation-Induced Neuromodulation by Blocking Cortical Neuron Activity Ultrasound Med. Biol 44 635–46 [DOI] [PubMed] [Google Scholar]
- [40].Mueller J, Legon W, Opitz A, Sato TF and Tyler WJ 2014. Transcranial Focused Ultrasound Modulates Intrinsic and Evoked EEG Dynamics Brain Stimul. 7 900–8 [DOI] [PubMed] [Google Scholar]
- [41].Legon W, Sato TF, Opitz A, Mueller J, Barbour A, Williams A and Tyler WJ 2014. Transcranial focused ultrasound modulates the activity of primary somatosensory cortex in humans Nat. Neurosci 17 322–9 [DOI] [PubMed] [Google Scholar]
- [42].Legon W, Bansal P, Tyshynsky R, Ai L and Mueller JK 2018. Transcranial focused ultrasound neuromodulation of the human primary motor cortex Sci. Rep 8 10007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [43].Mohammadjavadi M, Ye PP, Xia A, Brown J, Popelka G and Pauly KB 2019. Elimination of peripheral auditory pathway activation does not affect motor responses from ultrasound neuromodulation Brain Stimul. 12 901–10 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [44].Chakrabarti R, Ghazanwy M and Tewari A 2014. Anesthetic challenges for deep brain stimulation: A systematic approach N. Am. J. Med. Sci 6 359. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [45].Grant R, Gruenbaum SE and Gerrard J 2015. Anaesthesia for deep brain stimulation Curr. Opin. Anaesthesiol 28 505–10 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [46].Boonzaier J, van Tilborg GAF, Straathof M, Petrov PI, van Heijningen CL, van Vliet G, Smirnov N, van der Toorn A, Neggers SF and Dijkhuizen RM 2017. Differential outcomes of rTMS and anesthesia effects on functional connectivity in the rat brain Brain Stimul. 10 418 [Google Scholar]
- [47].Purdon PL, Pierce ET, Mukamel EA, Prerau MJ, Walsh JL, Wong KFK, Salazar-Gomez AF, Harrell PG, Sampson AL, Cimenser A, Ching S, Kopell NJ, Tavares-Stoeckel C, Habeeb K, Merhar R and Brown EN 2013. Electroencephalogram signatures of loss and recovery of consciousness from propofol Proc. Natl. Acad. Sci 110 E1142–51 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [48].Bonhomme V, Staquet C, Montupil J, Defresne A, Kirsch M, Martial C, Vanhaudenhuyse A, Chatelle C, Larroque SK, Raimondo F, Demertzi A, Bodart O, Laureys S and Gosseries O 2019. General Anesthesia: A Probe to Explore Consciousness Front. Syst. Neurosci 13 36. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [49].Downar J, Davis KD. Magnetic resonance imaging in neuromodulation In: Hamani C, Holtzheimer P, Lozano AM, Mayberg H, editors. Neuromodulation in Psychiatry. Wiley Online Library, West Sussex, UK: (2015). p. 49–79. doi: 10.1002/9781118801086.ch4 [DOI] [Google Scholar]
- [50].Ladd ME, Bachert P, Meyerspeer M, Moser E, Nagel AM, Norris DG, Schmitter S, Speck O, Straub S and Zaiss M 2018. Pros and cons of ultra-high-field MRI/MRS for human application Prog. Nucl. Magn. Reson. Spectrosc 109 1–50 [DOI] [PubMed] [Google Scholar]
- [51].Galante A, Sinibaldi R, Conti A, De Luca C, Catallo N, Sebastiani P, Pizzella V, Romani GL, Sotgiu A and Della Penna S 2015. Fast Room Temperature Very Low Field-Magnetic Resonance Imaging System Compatible with MagnetoEncephaloGraphy Environment ed F-H Lin PLoS One 10 e0142701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].Guidotti R, Sinibaldi R, De Luca C, Conti A, Ilmoniemi RJ, Zevenhoven KCJ, Magnelind PE, Pizzella V, Del Gratta C, Romani GL and Della Penna S 2018. Optimized 3D co registration of ultra-low-field and high-field magnetic resonance images ed C-T Chen PLoS One 13 e0193890. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [53].Ibrahim TS, Kangarlu A and Chakeress DW 2005. Design and Performance Issues of RF Coils Utilized in Ultra High Field MRI: Experimental and Numerical Evaluations IEEE Trans. Biomed. Eng 52 1278–84 [DOI] [PubMed] [Google Scholar]
- [54].Mangrum W, Hoang QB, Amrhein TJ, Duncan SM, Maxfield CM, Merkle E, et al. Duke Review of MRI Principles: Case Review Series E-Book. Elsevier Health Sciences, Philadelphia, PA, USA: (2012). [Google Scholar]
- [55].Olman CA and Yacoub E 2011. High-Field fMRI for Human Applications: An Overview of Spatial Resolution and Signal Specificity Open Neuroimag. J 5 74–89 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [56].Moser D, Zadicario E, Schiff G and Jeanmonod D 2013. MR-guided focused ultrasound technique in functional neurosurgery: targeting accuracy J. Ther. Ultrasound 1 1–10 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [57].Sengupta S, Roebroeck A, Kemper VG, Poser BA, Zimmermann J, Goebel R and Adriany G 2016. A Specialized Multi-Transmit Head Coil for High Resolution fMRI of the Human Visual Cortex at 7T ed A Leemans PLoS One 11 e0165418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [58].Kaza E, Klose U and Lotze M 2011. Comparison of a 32-channel with a 12-channel head coil: Are there relevant improvements for functional imaging? J. Magn. Reson. Imaging 34 173–83 [DOI] [PubMed] [Google Scholar]
- [59].Silva AC and Merkle H 2003. Hardware considerations for functional magnetic resonance imaging Concepts Magn. Reson 16A 35–49 [Google Scholar]
- [60].Ciobanu L, Solomon E, Pyatigorskaya N, Roussel T, Le Bihan D and Frydman L 2015. fMRI contrast at high and ultrahigh magnetic fields: Insight from complementary methods Neuroimage 113 37–43 [DOI] [PubMed] [Google Scholar]
- [61].Dickson SL and Mercer JG 2016. Neuroendocrinology of Appetite (Chichester, UK: John Wiley & Sons, Ltd; ) [Google Scholar]
- [62].Smith SM, Beckmann CF, Andersson J, Auerbach EJ, Bijsterbosch J, Douaud G, Duff E, Feinberg DA, Griffanti L, Harms MP, Kelly M, Laumann T, Miller KL, Moeller S, Petersen S, Power J, Salimi-Khorshidi G, Snyder AZ, Vu AT, Woolrich MW, Xu J, Yacoub E, Uğurbil K, Van Essen DC and Glasser MF 2013. Resting-state fMRI in the Human Connectome Project Neuroimage 80 144–68 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [63].Cobbold RSC. Foundations of Biomedical Ultrasound. Oxford university press, New York, NY, USA: (2015) [Google Scholar]
- [64].Lee W, Kim H, Jung Y, Song I-U, Chung YA and Yoo S-S 2015. Image-Guided Transcranial Focused Ultrasound Stimulates Human Primary Somatosensory Cortex Sci. Rep 5 8743. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [65].Li G-F, Zhao H-X, Zhou H, Yan F, Wang J-Y, Xu C-X, Wang C-Z, Niu L-L, Meng L, Wu S, Zhang H-L, Qiu W-B and Zheng H-R 2016. Improved Anatomical Specificity of Non-invasive Neuro-stimulation by High Frequency (5 MHz) Ultrasound Sci. Rep 6 24738. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [66].McDannold N and Maier SE 2008. Magnetic resonance acoustic radiation force imaging Med. Phys 35 3748–58 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [67].Larrat B, Pernot M, Aubry J-F, Dervishi E, Sinkus R, Seilhean D, Marie Y, Boch A-L, Fink M and Tanter M 2010. MR-guided transcranial brain HIFU in small animal models Phys. Med. Biol 55 365–88 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [68].Phipps MA, Jonathan SV, Yang P-F, Chaplin V, Chen LM, Grissom WA and Caskey CF 2019. Considerations for ultrasound exposure during transcranial MR acoustic radiation force imaging Sci. Rep 9 16235. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [69].Karakatsani ME, Samiotaki G, Downs ME, Ferrera VP and Konofagou EE 2015. Targeting effects on the volume and gray-to-white-matter ratio of the focused-ultrasound induced blood brain barrier opening in non-human primates in vivo 15th Int. Symp. Ther. Ultrasound 3–6 [Google Scholar]
- [70].Wu S-Y, Aurup C, Sanchez CS, Grondin J, Zheng W, Kamimura H, Ferrera VP and Konofagou EE 2018. Efficient Blood-Brain Barrier Opening in Primates with Neuronavigation-Guided Ultrasound and Real-Time Acoustic Mapping Sci. Rep 8 7978. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [71].Poorter J De 1995. Noninvasive MRI thermometry with the proton resonance frequency method: Study of susceptibility effects Magn. Reson. Med 34 359–67 [DOI] [PubMed] [Google Scholar]
- [72].Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K and Suzuki Y 1995. A precise and fast temperature mapping using water proton chemical shift Magn. Reson. Med 34 814–23 [DOI] [PubMed] [Google Scholar]
- [73].Morris H, Rivens I, Shaw A and Haar G Ter 2008. Investigation of the viscous heating artefact arising from the use of thermocouples in a focused ultrasound field Phys. Med. Biol 53 4759–76 [DOI] [PubMed] [Google Scholar]
- [74].Kamimura HAS, Aurup C, Bendau EV, Saharkhiz N, Kim MG and Konofagou EE 2020. Iterative Curve Fitting of the Bioheat Transfer Equation for Thermocouple-Based Temperature Estimation $In~ Vitro$ and $In~ Vivo$ IEEE Trans. Ultrason. Ferroelectr. Freq. Control 67 70–80 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [75].Dallapiazza RF, Timbie KF, Holmberg S, Gatesman J, Lopes MB, Price RJ, Miller GW and Elias WJ 2018. Noninvasive neuromodulation and thalamic mapping with low-intensity focused ultrasound J. Neurosurg 128 875–84 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [76].Constans C, Mateo P, Tanter M and Aubry J-F 2017. Potential impact of thermal effects during ultrasonic neurostimulation: retrospective numerical estimation of temperature elevation in seven rodent setups Phys. Med. Biol [DOI] [PubMed] [Google Scholar]
- [77].Ho M-L, Rojas R and Eisenberg RL 2012. Cerebral Edema Am. J. Roentgenol 199 W258–73 [DOI] [PubMed] [Google Scholar]
- [78].Lee EK, Lee EJ, Kim S and Lee YS 2016. Importance of Contrast-Enhanced Fluid Attenuated Inversion Recovery Magnetic Resonance Imaging in Various Intracranial Pathologic Conditions Korean J. Radiol 17 127–41 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [79].Kamiya K, Hori M, Irie R, Miyajima M, Nakajima M, Kamagata K, Tsuruta K, Saito A, Nakazawa M, Suzuki Y, Mori H, Kunimatsu A, Arai H, Aoki S and Abe O 2017. Diffusion imaging of reversible and irreversible microstructural changes within the corticospinal tract in idiopathic normal pressure hydrocephalus NeuroImage Clin. 14 663–71 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [80].Hynynen K, McDannold N, Vykhodtseva N and Jolesz F a 2001. Noninvasive MR Imaging– guided Focal Opening of the Blood-Brain Barrier in Rabbits Radiology 220 640–6 [DOI] [PubMed] [Google Scholar]
- [81].Moon HW, Fung PY, Whipp SC and Isaacson RE 1978. Effects of age and ambient temperature on the responses of infant mice to heat-stable enterotoxin of Escherichia coli: assay modifications Infect. Immun 20 36–9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [82].Conti A, Magnin R, Gerstenmayer M, Tsapis N, Dumont E, Tillement O, Lux F, Le Bihan D, Mériaux S, Della Penna S and Larrat B 2019. Empirical and Theoretical Characterization of the Diffusion Process of Different Gadolinium-Based Nanoparticles within the Brain Tissue after Ultrasound-Induced Permeabilization of the Blood-Brain Barrier Contrast Media Mol. Imaging 2019 1–13 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [83].Legon W, Rowlands A, Opitz A, Sato TF and Tyler WJ 2012. Pulsed Ultrasound Differentially Stimulates Somatosensory Circuits in Humans as Indicated by EEG and fMRI ed M Ptito PLoS One 7 e51177. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [84].Lee W, Kim H-C, Jung Y, Chung YA, Song I-U, Lee J-H and Yoo S-S 2016. Transcranial focused ultrasound stimulation of human primary visual cortex Sci. Rep 6 34026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [85].Ai L, Bansal P, Mueller JK and Legon W 2018. Effects of transcranial focused ultrasound on human primary motor cortex using 7T fMRI: a pilot study BMC Neurosci. 19 56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [86].Yang P-F, Phipps MA, Newton AT, Chaplin V, Gore JC, Caskey CF and Chen LM 2018. Neuromodulation of sensory networks in monkey brain by focused ultrasound with MRI guidance and detection Sci. Rep 8 7993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [87].Verhagen L, Gallea C, Folloni D, Constans C, Jensen DE, Ahnine H, Roumazeilles L, Santin M, Ahmed B, Lehericy S, Klein-Flügge MC, Krug K, Mars RB, Rushworth MF, Pouget P, Aubry J-F and Sallet J 2019. Offline impact of transcranial focused ultrasound on cortical activation in primates Elife 8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [88].Wiegert JS, Mahn M, Prigge M, Printz Y and Yizhar O 2017. Silencing Neurons: Tools, Applications, and Experimental Constraints Neuron 95 504–29 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [89].Yuan Y, Pang N, Chen YD, Wang Y and Li XL 2017. Theoretical analysis of the effects of transcranial magneto-acoustical stimulation on neuronal firing rhythm and Ca 2+ concentration with Chay neuron model Biomed. Phys. Eng. Express 3 055006 [Google Scholar]
- [90].Yuan Y, Chen Y and Li X 2016. Theoretical Analysis of Transcranial Magneto-Acoustical Stimulation with Hodgkin-Huxley Neuron Model Front. Comput. Neurosci 10 35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [91].Downs ME, Lee SA, Yang G, Kim S, Wang Q and Konofagou EE 2018. Non-invasive peripheral nerve stimulation via focused ultrasound in vivo Phys. Med. Biol 63 035011. [DOI] [PubMed] [Google Scholar]
- [92].Kim MG, Kamimura HAS, Lee SA, Aurup C, Kwon N and Konofagou EE 2020. Image guided focused ultrasound modulates electrically evoked motor neuronal activity in the mouse peripheral nervous system in vivo J. Neural Eng [DOI] [PMC free article] [PubMed] [Google Scholar]
- [93].Goetz SM and Deng Z-D 2017. The development and modelling of devices and paradigms for transcranial magnetic stimulation Int. Rev. Psychiatry 29 115–45 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [94].Heller L and van Hulsteyn DB 1992. Brain stimulation using electromagnetic sources: theoretical aspects Biophys. J 63 129–38 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [95].Aberra AS, Wang B, Grill WM and Peterchev AV. 2020. Simulation of transcranial magnetic stimulation in head model with morphologically-realistic cortical neurons Brain Stimul. 13 175–89 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [96].Toschi N, Welt T, Guerrisi M and Keck ME 2009. Transcranial magnetic stimulation in heterogeneous brain tissue: Clinical impact on focality, reproducibility and true sham stimulation J. Psychiatr. Res 43 255–64 [DOI] [PubMed] [Google Scholar]
- [97].Esser SK, Hill SL and Tononi G 2005. Modeling the Effects of Transcranial Magnetic Stimulation on Cortical Circuits J. Neurophysiol 94 622–39 [DOI] [PubMed] [Google Scholar]
- [98].Dayan E, Censor N, Buch ER, Sandrini M and Cohen LG 2013. Noninvasive brain stimulation: from physiology to network dynamics and back Nat. Neurosci 16 838–44 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [99].Boyden ES, Zhang F, Bamberg E, Nagel G and Deisseroth K 2005. Millisecond-timescale, genetically targeted optical control of neural activity Nat. Neurosci 8 [DOI] [PubMed] [Google Scholar]
- [100].Jiang J, Cui H and Rahmouni K 2017. Optogenetics and pharmacogenetics: principles and applications Am. J. Physiol. Integr. Comp. Physiol 313 R633–45 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [101].Spangler SM and Bruchas MR 2017. Optogenetic approaches for dissecting neuromodulation and GPCR signaling in neural circuits Curr. Opin. Pharmacol 32 56–70 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [102].Barbic M. Possible magneto-mechanical and magneto-thermal mechanisms of ion channel activation in magnetogenetics. Elife. 2019;8 doi: 10.7554/eLife.45807. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [103].Ibsen S, Tong A, Schutt C, Esener S and Chalasani SH 2015. Sonogenetics is a non-invasive approach to activating neurons in Caenorhabditis elegans Nat. Commun 6 8264. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [104].Darrow DP, O’Brien P, Richner TJ, Netoff TI and Ebbini ES 2019. Reversible neuroinhibition by focused ultrasound is mediated by a thermal mechanism Brain Stimul. 12 1439–47 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [105].Díaz M and Becker DE 2010. Thermoregulation: Physiological and Clinical Considerations during Sedation and General Anesthesia Anesth. Prog 57 25–33 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [106].Pasquinelli C, Hanson LG, Siebner HR, Lee HJ and Thielscher A 2019. Safety of transcranial focused ultrasound stimulation: A systematic review of the state of knowledge from both human and animal studies Brain Stimul. 12 1367–80 [DOI] [PubMed] [Google Scholar]
- [107].Legon W, Bansal P, Ai L, Mueller JK, Meekins G and Gillick B 2018. Safety of transcranial focused ultrasound for human neuromodulation bioRxiv 314856 [Google Scholar]
- [108].Legon W, Adams S, Bansal P, Patel PD, Hobbs L, Ai L, Mueller JK, Meekins G and Gillick BT 2020. A retrospective qualitative report of symptoms and safety from transcranial focused ultrasound for neuromodulation in humans Sci. Rep 10 5573. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [109].Kim E, Anguluan E and Kim JG 2017. Monitoring cerebral hemodynamic change during transcranial ultrasound stimulation using optical intrinsic signal imaging Sci. Rep 7 13148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [110].Yuan Y, Wang Z, Liu M and Shoham S 2020. Cortical hemodynamic responses induced by low-intensity transcranial ultrasound stimulation of mouse cortex Neuroimage 211 116597. [DOI] [PubMed] [Google Scholar]
- [111].Yoo S-S, Bystritsky A, Lee J-H, Zhang Y, Fischer K, Min B-K, McDannold NJ, Pascual-Leone A and Jolesz FA 2011. Focused ultrasound modulates region-specific brain activity Neuroimage 56 1267–75 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [112].Macé E, Montaldo G, Cohen I, Baulac M, Fink M and Tanter M 2011. Functional ultrasound imaging of the brain Nat. Methods 8 662–4 [DOI] [PubMed] [Google Scholar]
- [113].Mace E, Montaldo G, Osmanski B-F, Cohen I, Fink M and Tanter M 2013. Functional ultrasound imaging of the brain: theory and basic principles IEEE Trans. Ultrason. Ferroelectr. Freq. Control 60 492–506 [DOI] [PubMed] [Google Scholar]
- [114].Deffieux T, Demene C, Pernot M and Tanter M 2018. Functional ultrasound neuroimaging: a review of the preclinical and clinical state of the art Curr. Opin. Neurobiol 50 128–35 [DOI] [PubMed] [Google Scholar]
- [115].Gesnik M, Blaize K, Deffieux T, Gennisson J-L, Sahel J-A, Fink M, Picaud S and Tanter M 2017. 3D functional ultrasound imaging of the cerebral visual system in rodents Neuroimage 149 267–74 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [116].Rau R, Kruizinga P, Mastik F, Belau M, de Jong N, Bosch JG, Scheffer W and Maret G 2018. 3D functional ultrasound imaging of pigeons Neuroimage 183 469–77 [DOI] [PubMed] [Google Scholar]
- [117].Kamimura HAS, Urban MW, Carneiro AAO, Fatemi M and Alizad A 2012. Vibro acoustography beam formation with reconfigurable arrays IEEE Trans. Ultrason. Ferroelectr. Freq. Control 59 1421–31 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [118].Constans C, Deffieux T, Pouget P, Tanter M and Aubry J-F 2017. A 200–1380-kHz Quadrifrequency Focused Ultrasound Transducer for Neurostimulation in Rodents and Primates: Transcranial In Vitro Calibration and Numerical Study of the Influence of Skull Cavity IEEE Trans. Ultrason. Ferroelectr. Freq. Control 64 717–24 [DOI] [PubMed] [Google Scholar]
- [119].Ugurbil K 2014. Magnetic Resonance Imaging at Ultrahigh Fields IEEE Trans. Biomed. Eng 61 1364–79 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [120].Grienberger C and Konnerth A 2012. Imaging Calcium in Neurons Neuron 73 862–85 [DOI] [PubMed] [Google Scholar]