Abstract
Introduction
Divers with suspected decompression illness require high concentration oxygen (O2). There are many different O2 delivery devices, with few data comparing their performance. This study evaluated O2 delivery, using tissue O2 partial pressure (PtcO2), in healthy divers breathing O2 via three different delivery devices.
Methods
Twelve divers had PtcO2 measured at six limb sites. Participants breathed O2 from: a demand valve using an intraoral mask with a nose clip (NC); a medical O2 rebreathing system (MORS) with an oronasal mask and with an intraoral mask; and a non-rebreather mask (NRB) at 15 or 10 L·min-1 O2 flow. In-line inspired O2 (FIO2) and nasopharyngeal FIO2 were measured. Participants provided subjective ratings of device comfort, ease of breathing, and overall ease of use.
Results
PtcO2 values and nasopharyngeal FIO2 were similar with the demand valve with intraoral mask, MORS with both masks and the NRB at 15 L·min-1. PtcO2 and nasopharyngeal FIO2 values were significantly lower with the NRB at 10 L·min-1. The NRB was rated as the most comfortable to wear, easiest to breathe with, and overall the easiest to use.
Conclusion
Of the commonly available devices promoted for O2 delivery to injured divers, similar PtcO2 and nasopharyngeal FIO2 values were obtained with the three devices tested: MORS with an oronasal or intraoral mask, demand valve with an intraoral mask and NRB at a flow rate of 15 L·min-1. PtcO2 and nasopharyngeal FIO2 values were significantly lower when the flow rate using the NRB was decreased to 10 L·min-1.
Keywords: Decompression sickness, Decompression illness, First aid, Masks, Medical kits, Transcutaneous oximetry, Scuba diving
Introduction
High concentration oxygen (O2) therapy is an important early first aid treatment for injured divers. Complete relief or improvement of the symptoms of decompression illness (DCI) has been seen in divers receiving pre-hospital normobaric O2 therapy.[ 1] The current pre-hospital care recommendation for divers with symptoms and signs of DCI is for O2 delivery at the highest possible inspired fraction (close to 100%).[ 2] However, there are many factors that need to be considered when choosing the most appropriate O2 delivery system for a dive operation.[ 3 , 4]
A variety of portable O2 delivery units have been designed to provide divers with pre-hospital O2.[ 3 , 5] These units incorporate one of two basic operating configurations: (1) a constant O2 flow configuration used with a non-rebreather mask (NRB), medical O2 rebreathing system (MORS) or other constant flow delivery devices; and (2) a patient triggered demand valve configuration. The recommended initial O2 flow rate with the NRB mask for divers with suspected DCI has long been 15 L∙min-1.[ 6] Divers Alert Network (DAN) America reduced its recommended O2 flow rate to between 10 to 15 L∙min-1, to extend the duration of often limited O2 supplies in the field, while still providing high levels of oxygenation.[ 7] However, the effect of lower flow rates on tissue oxygenation is unknown. A previous study comparing tissue oxygenation found that the NRB at 15 L·min-1 performed better than the demand valve with an oronasal mask.[ 8] However, a subsequent study showed that the demand valve provided the best tissue oxygenation when used with an intraoral mask and nose clip (NC);[ 9] almost certainly because the intraoral mask eliminated leaks that were occurring with the oronasal mask.
The present study used transcutaneous oximetry measurement (TCOM) to determine tissue oxygenation at multiple standardised sites in participants breathing O2 from a demand valve using an intraoral mask with a NC; a MORS with an oronasal mask and with an intraoral mask; and a NRB at 15 and 10 L·min-1. The primary null hypothesis was that there would be no clinically significant difference in the partial pressure of transcutaneous tissue O2 (PtcO2) achieved after 10 min of breathing O2 with any of the different O2 delivery devices or flow rates.
Methods
Ethics approval was granted from The Townsville Health Service District Human Research Ethics Committee (HREC16/QTHS/196). Healthy, non-smoking, adult certified scuba divers of both sexes were recruited for the study. Facial hair or anatomical abnormality that may impair mask seal, any medical condition or medication that may affect tissue oxygenation, or an allergy to topical anaesthetic were exclusion criteria. Written informed consent was provided by all participants prior to their participation.
Participants refrained from consuming food or caffeine or performing heavy exercise for six hours prior to participating in the study. Demographic data, anthropometric measurements, and resting baseline measurements were collected. Tidal volume (VT) was measured at rest using the EasyOne Spirometer (ndd Medical Technologies, Andover MA, USA) according to the manufacturer’s instructions (average over 3–5 breaths). The participants rested in a supine position with their head on one pillow for the duration of the study. The test-room temperature was maintained between 22.4 and 22.9°C; to limit any vasoconstrictive effects of being cold, participants were covered with a blanket.
An 8 French paediatric feeding tube (ConvaTec Ltd., Deeside, UK) was inserted into the right nares after application of topical lignocaine (5%) and phenylephrine (0.5%) (Co-Phenylcaine™ forte spray, ENT Technologies Pty Ltd., Hawthorne East, Australia). Tube position was visually verified with the tip just proximal to the soft palate.[ 9] The tube was then attached to the E-sCO-OO module of a bedside monitor (GE Carescape Monitor B650, GE Healthcare Finland OY, Helsinki, Finland) allowing for both inspired O2 (FIO2, paramagnetic) and end-tidal carbon dioxide (ETCO2, infrared) measurements via a water trap (D-fend Pro+ Water Trap™, GE Healthcare Finland OY, Helsinki, Finland). Room air gas calibration was completed before each breathing system was used. The gas sampling rate was 120 ml·min-1.
TCOM is a non-invasive technique that uses heated electrodes on the skin to measure the PtcO2[ 10] and was thought to provide a relevant measurement of tissue O2 delivery in a study drawing an inference about tissue inert gas elimination. PtcO2 was measured using the TCM400 transcutaneous (tc) PO2 Monitoring System (Radiometer, Copenhagen, Denmark) with tc Sensor E5250. Zero current calibration of the PtcO2 electrode was performed using CAL2 gas (10% CO2 with N2 as balance) prior to commencement of the study, and calibration with atmospheric air occurred prior to each monitoring period. A ‘humidity correction factor’ was entered into the machine prior to each monitoring period. All assessments were performed by the same technician. The TCM400 displayed PtcO2 values in units of mmHg (average of previous monitoring intervals).
Six sensors were used: three on the left arm and three on the left leg.[ 9] Arm sensors were placed on the upper arm, lateral aspect of the lower arm, and the palm of the hand. Leg sensors were placed on the lateral leg, lateral ankle, and dorsum of the foot. Participants rested quietly while the sensors were placed. They were requested to minimise talking during the study as a method of control but were not allowed to sleep. Initial normobaric room air readings from all sensors were recorded after a minimum 20-min equilibration period that allowed all sensors to stabilize.
The participants were then asked to breathe O2 for 10 min from the following devices in randomized order determined using the random number generator in Microsoft® Excel (Microsoft Corporation, Redmond Washington, USA):
Demand valve (L324-020, Life Support Products, Allied Healthcare Products, St. Louis, MO, USA) with intraoral mask and NC held in place by the participant (NuMask® Inc., Woodlands Hills, CA, USA) (Figure 1);
MORS (Wenoll-System, EMS GmbH, Möhrendorf, Germany) (Figure 2) with intraoral mask and NC held in place by the participant;
MORS with air-cushion oronasal mask and a 4-strap mask holder (Figure 3);
NRB mask at 15 L·min-1 with elastic strap (Sturdy Industrial Co. Ltd, New Taipei City, Taiwan);
NRB mask at 10 L·min-1 with elastic strap.
Figure 1.
Configuration of demand valve, spacer with side port (allowing pressure and gas measurements), and intraoral mask
Figure 2.
Medical oxygen rebreathing system (Wenoll-System)
Figure 3.

Oronasal mask provided with the Wenoll-System and four-strap holder; oronasal mask secured in position on participant’s face
The demand inhalator valve provided in portable DAN O2 units was used for this study. A flexible high-pressure O2 hose was used to connect the demand valve to the hospital wall medical grade O2 outlet (415 kPa delivery pressure). The demand valve was attached to a spacer with a side port allowing pressure and gas measurements (Figure 1). A pressure line was attached to the side port and then to the bedside monitor via a BD DTXPlus™ pressure transducer (Argon Medical Devices Inc., Frisco, TX, USA). The monitor was configured to settings used for central venous pressure monitoring to give a high sensitivity in the lower range and zeroed before each participant. A single, new demand valve was used in the study and verification of inspiratory opening pressure required to trigger the valve and the expiratory resistance pressure was made prior to the commencement of each new participant. The demand valve configuration with intraoral mask and NC, from previous optimization trials, was used in this current research.[ 9] Mask and circuit dead space was determined by measuring the amount of water required to fill each device. Mask fill levels were estimated by filling the masks with water and then placing a mannequin’s face into the mask.
The NRB was examined to ensure there were three one-way valves in place and then primed with O2 to inflate the reservoir bag. The NRB was positioned and adjusted to obtain the best seal possible. Participants were asked to breathe normally, and the reservoir bag was monitored for persistent inflation during the breathing periods.
When using the demand valve participants were asked to breathe deeply enough to trigger the valve as outlined in DAN educational material.[ 6 , 7]
The Wenoll MORS system was primed with 40 L·min-1 of O2 until the rebreathing bag was completely filled, and the oronasal mask was attached with a four-strap holder. The O2 flow was 1.5 L·min-1 during the 10-min breathing periods as outlined in the Wenoll-System operation manual.
In-line FIO2, nasopharyngeal FIO2, PtcO2, and other respiratory measures were recorded at the end of the 10-min breathing period, once PtcO2 had stabilized. In-line and nasopharyngeal FIO2 measurements were performed to determine if there is a difference between O2 delivered by a device (in-line) and the O2 reaching the upper airway (nasopharyngeal). Nasopharyngeal gas sampling was intermittent (every two min) throughout the O2 breathing periods to prevent clogging of the catheter and to capture peak values. After each 10-min O2 breathing period, participants breathed room air for 10 min, allowing all PtcO2 levels to return to baseline before the next device was trialled.[ 11] At the end of the data collection period all participants used a five-point Likert scale to rate each configuration on mask comfort, ease of breathing, and overall ease of use of each device. A final open-ended question asked about any adverse effects while breathing O2.
ANALYSIS
All collected data were de-identified and entered into an Excel worksheet, and subsequently exported into Statistical Package for the Social Sciences version 25.0.0 (SPSS, IBM® Corporation, Armonk, New York, USA) for analysis.
Based on previous research when participants breathed 100% O2, mean PtcO2 values between 199 mmHg (26 kPa) (dorsum of foot) and 454 mmHg (60 kPa) (upper arm) were expected.[ 12] Each sensor site generally has slightly different values, however, a decrease of 75 mmHg (10 kPa) across any of these sites was assumed to be clinically significant. Based on the values above and allowing for substantial correlation (r = 0.90) between the repeated measures, a sample size of 12 participants would provide a power of 90% (with α = 0.05) to detect clinically significant reductions in tissue oxygenation. In this context, ‘clinically significant’ was intended to mean a reduction in tissue O2 delivery sufficient to indicate a potentially important corresponding reduction in the diffusion gradient for inert gas from tissues into blood. There are no published data which demonstrate how such gradients can be inferred from changes in tissue oxygenation, so a threshold tissue oxygenation change of 75 mmHg (10 kPa) (smallest increase in PtcO2 at one sensor site breathing 100% O2 with a hood[ 12]) was agreed by consensus of the physiologists and clinicians involved in the study.
The Shapiro-Wilk test was used to evaluate normality of data distribution. None of the data were normally distributed. Differences between median, ETCO2, in-line, and peak and across device over time nasopharyngeal FIO2 readings using the various devices and flow rates were analysed using the Friedman test with post hoc paired analyses completed using the Wilcoxon signed-rank test with Bonferroni correction. For the post hoc tests, a corrected P-value of 0.005 (0.05/10) was considered significant for the PtcO2 values and a corrected P-value of 0.01 (0.05/4) for the 2-min nasopharyngeal FIO2 values.
The primary outcome measure was a comparison of the median PtcO2 measurements recorded across the six sensor sites after breathing O2 for 10 min using each device and flow rate. Secondary outcome measures included in-line and nasopharyngeal FIO2, ETCO2, and participant-rated mask comfort, ease of breathing and overall use of each device.
Results
Twelve healthy volunteers, nine females and three males, met all inclusion criteria and completed the study protocol. Their demographic and baseline measures breathing room air are shown in Table 1.
Table 1. Demographic and baseline measurements for the 12 participants breathing air. Optimal waist-to-hip ratios are < 0.82 for males and < 0.71 for females. BP = blood pressure. IQR = inter-quartile range .
| Characteristic | Median (IQR) | Range |
| Age (years) | 30 (28, 32) | 21–34 |
| Body mass index (kg·m2) | 19 (15, 22) | 15–24 |
| Waist-to-hip ratio | ||
| Males | 0.86 (0.82, 0.87) | 0.82–0.87 |
| Females | 0.74 (0.72, 0.80) | 0.67–0.93 |
| Heart rate (beats·min-1) | 67 (59, 70) | 50–94 |
| Systolic BP (mmHg) | 112 (111, 119) | 100–128 |
| Diastolic BP (mmHg) | 66 (57, 73) | 54–82 |
| Respiratory rate (breaths·min-1) | 14 (12, 16) | 12–16 |
| Tidal volume (ml) | 745 (533, 913) | 470–1130 |
| Oxygen saturation (%) | 97 (96, 98) | 96–99 |
| End-tidal CO2 (mmHg) | 39 (37, 45) | 36–46 |
Figure 4 displays the median PtcO2 readings across all sensor sites and breathing devices and flow rates. Baseline PtcO2 values, median and IQR for each sensor site after breathing O2 for 10 min are presented in Table 2. PtcO2 values were statistically different across each breathing device and flow rate for each sensor site (Table 2). Post hoc analysis showed there were no significant differences in PtcO2 values between the NRB 15 L·min-1, demand valve and MORS with intraoral or oronasal mask. Some differences in median PtcO2 readings between devices at the same sites met the 75-mmHg (10 kPa) threshold for clinical significance, but only in comparisons between the NRB 10 L·min-1 with other devices. The median PtcO2 readings achieved using the NRB 10 L·min-1 were more than 75 mmHg (10 kPa) less than all other devices (including the NRB 15 L·min-1) at the upper and lower arm sites, and at the lateral leg and ankle sites in comparison to the MORS with intraoral mask. No comparisons of the median PtcO2 between other devices met the threshold for clinical significance.
Figure 4.
Median transcutaneous oxygen partial pressures (mmHg) after breathing oxygen for 10 min with different devices and flow rates; NRB = non-rebreather mask; MORS = medical oxygen rebreathing system
Table 2. Transcutaneous oxygen partial pressures (median and inter-quartile range shown in mmHg) while breathing oxygen using the different devices and flow rates; * statistically significantly greater than NRB 10 L·min-1 based on Wilcoxon signed-rank test with Bonferroni correction; # P -values based on the Friedman test; NRB = non-rebreather mask; MORS = medical oxygen rebreathing system .
| Site | Baseline (room air) | 10 L·min-1 NRB | 15 L·min-1 NRB | Demand with intraoral mask | MORS with intraoral mask | MORS with oronasal mask | P-value# |
| Upper arm | 70 (61,77) | 333 (285, 382) | 429 (408, 464)* | 420 (373, 465) | 428 (385, 476)* | 451 (375, 480)* | 0.002 |
| Lower arm | 65 (58,70) | 241 (217, 304) | 327 (296, 405)* | 348 (264, 370) | 324 (286, 405)* | 329 (313, 370)* | 0.004 |
| Palm hand | 66 (63,74) | 192 (151, 266) | 245 (202, 275) | 212 (179, 239) | 236 (190, 293)* | 242 (163, 289) | 0.008 |
| Lateral leg | 57 (49,64) | 201 (172, 233) | 251 (217, 335)* | 261 (208, 353) | 278 (224, 350)* | 270 (219, 330)* | 0.004 |
| Lateral ankle | 62 (48, 67) | 231 (152, 260) | 299 (228, 336)* | 277 (235, 342) | 318 (221, 361) | 302 (246, 319)* | 0.002 |
| Dorsum foot | 54 (50, 65) | 119 (88, 149) | 162 (133, 226)* | 152 (118, 220) | 156 (93, 205) | 168 (125, 198)* | 0.008 |
Peak nasopharyngeal FIO2 was highest breathing O2 at 15 L·min-1 with the NRB and lowest when breathing O2 at 10 L·min-1 with the NRB (Table 3). One participant’s nasopharyngeal FIO2 value at 10 min while breathing using the demand valve was unattainable due to catheter clogging. There was a significant effect of time on nasopharyngeal FIO2 for the MORS (Figure 5). It was shown that for the MORS with intraoral mask, nasopharyngeal FIO2 was significantly higher at time points 3, 4 and 5 (6, 8 and 10 min) compared to time 1 (P < 0.01). Nasopharyngeal FIO2 at all time points (4, 6, 8 and 10 min) was significantly higher than time point 1 breathing with the MORS and oronasal mask (P < 0.01). Both sets of results reflect a roughly linear increase in FIO2 over O2 administration time. There was no statistical difference between the nasopharyngeal FIO2 values at each time point for any of the other breathing devices. Figure 6 illustrates the rise in PtcO2 values over the 10-min O2 breathing periods, mirroring the rise in FIO2 values for the MORS.
Table 3. Inspired oxygen and respiratory measures while breathing oxygen using different devices and flow rates (median and inter-quartile range) and estimated mask and circuit dead space; NRB = non-rebreather mask; MORS = medical oxygen rebreathing system; n/a = not applicable; ETCO2 = end-tidal carbon dioxide; *P -values based on the Friedman test .
| Parameter | 10 L·min -1 NRB | 15 L·min-1 NRB | Demand with intraoral mask | MORS with intraoral mask | MORS with oronasal mask | *P-value |
| In-line FIO2 (%) | n/a | n/a | 95 (92, 95) | 93 (89, 95) | 91 (88, 93) | 0.045 |
| Peak nasopharyngeal FIO2 (%) | 89 (75, 93) | 97 (94, 98) | 92 (88, 94) | 91 (88, 95) | 90 (88, 92) | 0.013 |
| ETCO2 (mmHg) | 39 (35, 43) | 38 (34, 43) | 38 (33, 41) | 38 (32, 44) | 39 (36, 43) | 0.743 |
| Respiratory rate (breaths·min-1) | 12 (10, 15) | 12 (10, 16) | 10 (8, 12) | 11 (8, 12) | 12 (8, 14) | 0.055 |
| Mask + circuit dead space (ml) | 95 | 95 | 14 | 14 + 350 | 136 + 350 | n/a |
Figure 5.
Median (IQR) nasopharyngeal inspired oxygen percentage recorded every two min for each delivery device and flow rate; NRB = non-rebreather mask; MORS = medical oxygen rebreathing system
Figure 6.
Transcutaneous oxygen partial pressures (mmHg) for one participant while breathing O2 using different devices and flow rates over a complete iteration of the study. First 20-min equilibration period, followed by alternating 10-min O2 and air breathing periods on different devices in randomized order. NRB = non-rebreather mask; MORS = medical oxygen rebreathing system
ETCO2 was similar for all devices and flow rates. In-line FIO2 did not exceed 97% with any of the devices and was lowest using the MORS with oronasal mask (80%; Table 3). Estimated mask assembly and circuit dead space is presented in Table 3. Actual individual NRB and oronasal mask volumes would vary slightly depending on each participant’s facial features.
Participant ratings for mask comfort are presented in Table 4. Ease of breathing rating for each device is listed in Table 5. The NRB was rated as overall easiest to use (Table 6). On post hoc analysis no statistical difference was found between each device.
Table 4. Mask comfort rating for each delivery device (n (%)). MORS = medical oxygen rebreathing system; *P-value = 0.052, Friedman test .
| Comfort assessment | Non-rebreather mask | Demand: intraoral mask | MORS: intraoral mask | MORS: oronasal mask |
| 1. Very uncomfortable | 0 | 0 | 0 | 0 |
| 2. Uncomfortable | 0 | 4 (33.3) | 3 (25.0) | 3 (25.0) |
| 3. Neither | 2 (16.7) | 2 (16.7) | 1 (8.3) | 3 (25.0) |
| 4. Comfortable | 3 (25.0) | 4 (33.3) | 6 (50.0) | 5 (41.7) |
| 5. Very comfortable | 7 (58.3) | 2 (16.7) | 2 (16.7) | 1 (8.3) |
| Median score (IQR)* | 5.0 (4.0–5.0) | 3.5 (2.0–4.0) | 4.0 (2.3–4.0) | 3.5 (2.3–4.0) |
Table 5. Ease of breathing rating for each delivery device (n (%)); MORS = medical oxygen rebreathing system; *P-value = 0.061, Friedman test .
| Breathing assessment | Non-rebreather mask | Demand: intraoral mask | MORS: intraoral mask | MORS: oronasal mask |
| 1. Very difficult | 0 | 0 | 0 | 0 |
| 2. Difficult | 0 | 3 (25.0) | 2 (16.7) | 1 (8.3) |
| 3. Neither | 0 | 2 (16.7) | 1 (8.3) | 4 (33.3) |
| 4. Easy | 3 (25.0) | 1 (8.3) | 5 (41.7) | 3 (25.0) |
| 5. Very easy | 9 (75.0) | 6 (50.0) | 4 (33.3) | 4 (33.3) |
| Median score (IQR)* | 5.0 (4.3–5.0) | 4.5 (2.3–5.0) | 4.0 (3.3–5.0) | 4.0 (3.0–5.0) |
Table 6. Overall ease of use for each delivery device (n (%)); MORS = medical oxygen rebreathing system; * P-value = 0.009, Friedman test .
| Overall ease of use assessment | Non-rebreather mask | Demand: intraoral mask | MORS: intraoral mask | MORS: oronasal mask |
| 1. Very difficult | 0 | 0 | 0 | 0 |
| 2. Difficult | 0 | 3 (25.0) | 1 (8.3) | 1 (8.3) |
| 3. Neither | 0 | 3 (25.0) | 1 (8.3) | 2 (16.7) |
| 4. Easy | 3 (25.0) | 2 (16.7) | 4 (33.3) | 6 (50.0) |
| 5. Very easy | 9 (75.0) | 4 (33.3) | 6 (50.0) | 3 (25.0) |
| Median score (IQR)* | 5.0 (4.3–5.0) | 3.5 (2.3–5.0) | 4.5 (4.0–5.0) | 4.0 (3.3–4.8 |
Discussion
High concentration O2 is the primary first aid treatment for divers suspected of having DCI.[ 2 , 3 , 13] O2 has been shown in retrospective reviews to improve symptoms and decrease the subsequent number of hyperbaric treatments required.[ 1] Of the tested commercially available O2 delivery systems designed for diver first aid, our study has shown that all systems can provide similar levels of tissue oxygenation and nasopharyngeal FIO2. However, when breathing with the NRB, an O2 flow rate of 15 L·min-1 is required to reach these levels.
Peak nasopharyngeal FIO2 was highest with the NRB with a flow rate of 15 L·min-1 (Table 3) though 10-min PtcO2 values were similar for each device. This probably reflects the variability in breathing patterns of each participant and flow direction of the O2. Nose breathing during use of the NRB may explain the favourable nasopharyngeal FIO2 results. When breathing with the MORS on the oronasal mask, some participants stated that they kept their mouths slightly open to ensure good fit of the mask, potentially bypassing the nasopharyngeal catheter through mouth breathing. Using the intraoral mask, the O2 flow may have been directed slightly below the nasopharyngeal catheter through obligatory mouth breathing. This illustrates the importance of clearly describing the position of sampling ports in a research protocol and the potential variability in results if O2 is measured at different sites. PtcO2 better reflects actual O2 delivery to the body tissues whereas the nasopharyngeal FIO2 is subject to the above possible confounders.
Portable O2 delivery units can provide a constant flow capability or operate as a pressure-triggered demand valve. The demand valve only delivers O2 when the diver inhales and therefore allows for conservation of O2, dependent on the respiratory minute volume of the user. The ease of use, familiarity for divers, potential to deliver high inspired O2 concentrations,[ 14] as well as the potential for O2 supply conservation, has led to the recommendation of the demand valve as the O2 delivery method of choice in the prehospital treatment of DCI.[ 3] However, previous research unexpectedly showed that the demand valve with oronasal mask provided less tissue O2 than a constant flow NRB.[ 8] In the present study, PtcO2 readings whilst breathing O2 via the demand valve with an intraoral mask and NC were similar to those achieved with a NRB at 15 L·min-1. The previous contradictory findings[ 8] were almost certainly explained by poor fit of the oronasal mask and subsequent entrainment of ambient air.[ 9]
The MORS provided similar oxygen levels regardless of the mask used. The oronasal mask provided with the system has an adjustable air-filled cushion to optimize mask fit and seal. The 4-point mask strap held the mask onto the face to aid with the seal. Fitting of the mask onto the participants’ face prior to the study allowed for a good seal to limit if not eliminate any entrainment of ambient air. Oronasal masks supplied with a demand valve system do not have the mask strap system, and the diver or first aider must therefore apply pressure to the mask to ensure an adequate seal. The technical difficulty of this almost certainly leads to breaks in the mask seal and entrainment of ambient air, especially because the user must generate negative pressure inside the mask to trigger the demand valve. While both oronasal or intraoral masks provide good O2 delivery with the MORS, an intraoral mask may provide the highest levels of oxygenation with a demand valve.[ 9]
The NRB functions as a variable performance device with better oxygenation at a higher flow rate.[ 15] Unfortunately, this means a greater consumption of O2. Demand regulators using an intraoral mask and NC behave as fixed performance devices,[ 9] with O2 consumption based on minute ventilation. The NRB and demand valve are both open systems with exhaled gas lost into the environment. A closed system is most beneficial for O2 conservation as only low flow O2 is required.[ 5 , 16] MORS for first aid O2 delivery are not commonly used by recreational divers, largely due to increased complexity and operational requirements[ 3 , 17] and generally limited availability. However, it is possible that the growing popularity of closed-circuit rebreathers for diving will drive an increased interest in MORS systems for first aid use.
The Wenoll System MORS comes with an air-cushion mask (like a pocket face mask) held tightly in place with a 4-strap holder and a regulator mouthpiece. In this study an intraoral mask rather than the regulator mouthpiece was used with the MORS for better comparison with the demand system. Both masks provided good peak O2 levels at 10 min, but the MORS took longer to reach peak inspired O2 than the NRB and demand valve (Figure 5). This is consistent with previous research showing a seven-minute time frame to reach 98–100% inspired O2.[ 16] It is possible that the MORS may have delivered greater fractions of O2 more quickly if a procedure to remove nitrogen (N2) from the participants’ lungs had been employed at the start of breathing on the MORS (typically, by exhaling to atmosphere completely, then inhaling O2 from the system and exhaling it to the atmosphere for several breaths before breathing exclusively on the MORS). Examination of the PtcO2 values while breathing O2 with the MORS showed a plateau at eight to 10 min and therefore the data reported here probably accurately reflect peak values.
Breathing high concentration O2 eliminates N2 from the inspired gas and enhances N2 elimination from the body.[ 3] Open circuit systems release exhaled gas into the environment, allowing for the elimination of N2. In MORS, the higher O2 flow rate in the first hour not only improves oxygenation[ 16] but allows for excess gas in the breathing circuit to be automatically vented through the over-pressure valve, which serves to purge accumulated N2 into the environment.[ 5] Other suggestions for purging excess N2 from the MORS circuit when used in injured divers include: periodic increase in O2 flow rates and periodic use of a purge button, if equipped.[ 5 , 16] Monitoring for colour change of the carbon dioxide (CO2) absorber and spontaneous increase in tidal volume are ways to evaluate scrubber function.[ 5] Limiting the usage time of the absorber can prevent CO2 intoxication.[ 5] Formal training in the use of a MORS for O2 delivery is recommended.
The NRB was rated as the overall easiest to use (Table 6), even though divers are accustomed to breathing from a demand valve with a mouthpiece. Some participants commented on the change in their breathing patterns when using the demand valve, they used their 'diving breathing pattern' of slower deeper breaths. Although not statistically significant, there was a trend towards a slower respiratory rate when breathing with the demand valve (P = 0.055) (Table 3).
Three commonly available pre-hospital O2 delivery systems were evaluated for O2 delivery, comfort, ease of breathing, and overall use. There are many other factors that need to be considered when selecting the most appropriate O2 delivery system for a dive operation.[ 3 , 15] Remoteness of diving operations, and therefore a protracted time to arrive at medical care, may increase the need for a system that is more comfortable for the diver but also a system that provides a longer duration of O2 delivery and CO2 elimination. Cost, availability and O2 supplies in different countries may also play an important part in the decision-making process. Knowledge of individual country guidelines and training requirements are necessary to make educated decisions about appropriate O2 delivery system selection.
LIMITATIONS
There was a low number of male participants in this study due to a predominance of facial hair. Facial hair was an exclusion factor, as it was thought it could contribute to mask leak.[ 18] In real world use, facial hair in males may reduce the efficacy of O2 delivery by a NRB in comparison to a device that does not rely on a facial seal, such as the demand valve or MORS used with an intraoral mask and NC. Previous research shows no significant difference in PtcO2 results by sex.[ 12]
Even though a higher PtcO2 value likely indicates a greater drive for tissue inert gas elimination, bubble resolution and oxygenation of hypoxic tissues,[ 3 , 19 , 20] this study did not address the clinical efficacy of these O2 devices in treating DCI or achieving bubble resolution.[ 21] Similarly, the arbitrary nature of the consensus decision to use a 75 mmHg (10 kPa) PtcO2 threshold to power the study and to indicate a meaningful difference in O2 delivery / outgassing gradient between devices is acknowledged. Therefore, although perhaps indicative, these data do not prove that one device will be associated with greater clinical efficacy than another.
The nasopharyngeal catheter provided valuable information on the oxygenation provided by each delivery system but may have compromised the seal of both the NRB and oronasal mask. The catheter was secured to the nares and laid against the face, passing under the edge of the masks. The oronasal mask has an air-filled cushion which can easily mould around irregular facial features. The NRB has a more rigid edge and may have been more affected by the presence of the catheter.
The oxygen breathing period was limited to 10 min based on previous research[ 8 , 11] where PtcO2 values had stabilized at that time point. The TCM400 machine has a built-in arrow indicator that depicts upward or downward trends to help clinicians to identify stable peak values (when the arrows disappear). However, in visualizing the nasopharyngeal FIO2 values it seems that the values were still rising for the MORS and NRB at 10 L·min-1. Although there was no statistical difference in the values at eight and 10 min, extending the monitoring time beyond the 10 min O2 breathing period could provide additional information.
Conclusion
The three tested O2 delivery systems used to treat injured divers (MORS with an oronasal or intraoral mask, demand valve with an intraoral mask and NRB at a flow rate of 15 L·min-1) delivered similar PtcO2 and nasopharyngeal FIO2 values. PtcO2 and nasopharyngeal FIO2 values were lower when the flow rate using the NRB was decreased from 15 to 10 L·min-1. O2 delivery and supply conservation are important factors to be considered when selecting an O2 delivery system for a dive operation.
Footnotes
Acknowledgements
The authors gratefully acknowledge the financial assistance received from the Emergency Medicine Foundation (EMF), Queensland and the in-kind support from Divers Alert Network Asia-Pacific. We also thank our participants, especially those who shaved their facial hair.
Conflict of interest and funding:
Professor Mitchell and Associate Professor Pollock are members of the editorial board of Diving and Hyperbaric Medicine but were not involved in the peer review or publication decision-making process for this article. This process was managed by the European Editor Dr Lesley Blogg.
Contributor Information
Denise F Blake, Emergency Department, The Townsville Hospital, Townsville, Queensland, Australia; Marine Biology and Aquaculture, James Cook University, Townsville, Queensland, Australia.
Melissa Crowe, Sport and Exercise Science, James Cook University, Townsville, Queensland, Australia.
Daniel Lindsay, College of Public Health, Medical and Veterinary Sciences, James Cook University, Townsville, Queensland, Australia.
Annie Brouff, Hyperbaric Medicine Unit, The Townsville Hospital, Townsville, Queensland, Australia.
Simon J Mitchell, Department of Anaesthesiology, University of Auckland, Auckland, New Zealand; Department of Anaesthesia, Auckland City Hospital, Auckland, New Zealand.
Peter A Leggat, College of Public Health, Medical and Veterinary Sciences, James Cook University, Townsville, Queensland, Australia.
Neal W Pollock, Department of Kinesiology, Faculty of Medicine, Université Laval, Québec, Canada; Service de médecine hyperbare, Centre de médecine de plongée du Québec, Levis, Québec, Canada.
References
- Longphre JM, Denoble PJ, Moon RE, Vann RD, Freiberger JJ. First aid normobaric oxygen for the treatment of recreational divers. Undersea Hyperb Med. 2007; 34: 43- 9. Available from: http://archive.rubicon-foundation.org/5515. [cited 2018 May 17]. [PubMed] [Google Scholar]
- Mitchell SJ, Bennett MH, Bryson P, Butler FK, Doolette DJ, Holm JR, et al. Pre-hospital management of decompression illness: expert review of key principles and controversies . Diving Hyperb Med. 2018;48:45–55. doi: 10.28920/dhm48.1.45-55. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lippmann J. First aid oxygen administration for divers. SPUMS Journal. 2003; 33: 192- 8. Available from: http://archive.rubicon-foundation.org/10050. [cited 2018 May 17]. [Google Scholar]
- Sayer MDJ, Küpper FC, van West P, Wilson CM, Brown H, Azzopardi E. Managing scientific diving operations in a remote location: The Canadian high Arctic. Diving Hyperb Med. 2013; 43: 239- 43. Available from: https://www.researchgate.net/publication/260130983_Managing_scientific_diving_operations_in_a_remote_location_The_Canadian_high_Arctic. [cited 2019 April 18]. . [PubMed] [Google Scholar]
- Wendling J. Normobaric oxygenation in dive accidents: a challenge for the developers of oxygen delivery systems. SPUMS Journal. 1997; 27: 101- 4. Available from: http://archive.rubicon-foundation.org/6063. [cited 2018 May 17]. [Google Scholar]
- Lippmann J. Oxygen first aid. 4th ed. Melbourne: Submariner Publications; 2011. [Google Scholar]
- Bird N, Nochetto M. Emergency oxygen for scuba diving injuries: student handbook. 7th Durham (NC): Divers Alert Network; 2012. [Google Scholar]
- Blake DF, Naidoo P, Brown LH, Young D, Lippmann J. A comparison of the tissue oxygenation achieved using different oxygen delivery devices and flow rates. Diving Hyperb Med. 2015; 45: 79- 83. Available from: http://www.dhmjournal.com/images/Journals/45/DHM_Vol45_No2.pdf. [cited 2018 May 17] . [PubMed] [Google Scholar]
- Blake DF, Crowe M, Lindsay D, Brouff A, Mitchell SJ, Pollock NW. Comparison of tissue oxygenation achieved breathing oxygen from a demand valve with four different mask configurations . Diving Hyperb Med. 2018;48:209–17. doi: 10.28920/dhm48.4.209-217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Eberhard P, Mindt W, Jann F, Hammacher K. Continuous pO2 monitoring in the neonate by skin electrodes . Med Biol Eng. 1975;13:436–42. doi: 10.1007/BF02477116. [DOI] [PubMed] [Google Scholar]
- Brown JT, Schur MS, McClain BC, Kafer ER. In vivo response time of transcutaneous oxygen measurement to changes in inspired oxygen in normal adults. Can Anaesth Soc J. 1984; 31: 91- 6. Available from: https://link.springer.com/article/10.1007/BF03011489. [cited 2019 May 14] . [DOI] [PubMed] [Google Scholar]
- Blake DF, Young DA, Brown LH. Transcutaneous oximetry: variability in normal values for the upper and lower limb . Diving Hyperb Med. 2018;48:2–9. doi: 10.28920/dhm48.1.2-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vann RD, Butler FK, Mitchell SJ, Moon RE. Decompression illness . Lancet. 2011;377:153–64. doi: 10.1016/S0140-6736(10)61085-9. [DOI] [PubMed] [Google Scholar]
- Hobbs GW, Natoli MJ, Pollock NW. Divers Alert Network emergency oxygen demand regulator validation trials. Centre for Environmental Physiology and Environmental Medicine, Divers Alert Network Internal Report . Durham: Duke University Medical Centre; 2000. [Google Scholar]
- Davis M. Oxygen therapy equipment a theoretical review. SPUMS Journal. 1998; 28: 165- 72. Available from: http://archive.rubicon-foundation.org/5951. [cited 2018 October 13]. [Google Scholar]
- Komesaroff D. Oxygen administration in diving accidents. SPUMS Journal. 1998; 28(Suppl): 20- 5. Available from: http://archive.rubicon-foundation.org/5974. [cited 2018 May 17]. [Google Scholar]
- Pollock NW, Natoli MJ. Performance characteristics of the second-generation remote emergency medical oxygen closed-circuit rebreather . Wilderness Environ Med. 2007;18:86–94. doi: 10.1580/06-WEME-OR-032R.1. [DOI] [PubMed] [Google Scholar]
- Stobbe TJ, daRoza RA, Watkins MA. Facial hair and respirator fit: a review of the literature . Am Ind Hyg Assoc J. 1988;49:199–204. doi: 10.1080/15298668891379594. [DOI] [PubMed] [Google Scholar]
- Hyldegaard O, Møller M, Madsen J. Effect of He-O2, O2, and N2O-O2 breathing on injected bubbles in spinal white matter. Undersea Biomed Res. 1991; 18: 361- 71. Available from: http://archive.rubicon-foundation.org/2580. [cited 2018 May 17] . [PubMed] [Google Scholar]
- Brubakk A. Surface oxygen is an acceptable definitive treatment. Diving Hyperb Med. 2000; 30: 155- 61. Available from: http://archive.rubicon-foundation.org/5866. [cited 2018 May 17] . 1746064 [Google Scholar]
- Blatteau J-E, Pontier J-M. Effect of in-water recompression with oxygen to 6 msw versus normobaric oxygen breathing on bubble formation in divers . Eur J Appl Physiol. 2009;106:691–5. doi: 10.1007/s00421-009-1065-y. [DOI] [PubMed] [Google Scholar]





