Abstract
Recurrent instability remains a common cause of failure after revision total hip arthroplasty (THA). Dual mobility (DM) cups and designs with constrained acetabular liners (CAL) have both been developed to help overcome this challenge. The aim of this report is to compare safety and efficacy outcomes of both designs based on the literature. A comprehensive literature review including published evidence on the results of DM and CAL in revision THA was performed and summarized. Available literature focusing on overall survival, dislocation, loosening, intra-prosthetic dislocation (IPD), and functional outcomes were analysed. Forty-six reports including an evaluation of 5,617 total hips were evaluated were included in the review. The included studies were divided into two distinct groups based on implantation approach: i) CAL (n=15) and ii) DM (n=31). The DM group had higher overall survival rates (94.7% vs. 81.0%), lower dislocation rates (2.6% vs. 11.0%), and lower acetabular loosening rates (1.0% vs. 2.0%) compared to the CAL group. IPDs were reported in 6 studies (mean rate, 0.6%). No differences in functional outcomes were identified due to incomplete reports. Our observations reveal that designs with CAL have poorer outcomes as compared to DM cups in revision THA. Currently, the use of DM seems more appropriate since they offer lower rates of dislocations, loosening and re-revisions in the short- and mid-term. Concerns regarding the potential of increased wear in a younger, high-demand population require additional data and evaluation by long-term studies for the DM design.
Keywords: Dual mobility, Total hip arthroplasty, Constrained, Dislocation, Revision
INTRODUCTION
Total hip arthroplasty (THA) is regarded as one of the most successful and cost-effective surgical procedures in end-stage hip arthritis1,2,3). THA effectively reduces pain, improves function, mobility and quality of life3). As health care continues to improve and life expectancy increases, the demand for total joint arthroplasty will rise to reflect this more active, aging population2). Internationally, the amount of THAs is expected to increase by 170% between now and 2030. In addition, an increasing number of THA require revision; total hip revisions are also projected to increase by 137% by 20302,4,5,6,7). Complications after THA can be very challenging for both the surgeon and the patient6). Thromboembolic disease, infection, peri-prosthetic fracture, aseptic loosening and dislocation are the most common postoperative complications.
Postoperative dislocation remains a persistent, disabling complication after THA. Recurrent instability after revision THA represents an even greater burden and is the leading indication for re-revision surgery8). Dislocation rates range between 0.3% and 10% in primary THA5,9). Depending on the reasons for revision, dislocation rates after revision THA varies from 5.0% to 30.0%2,4,5,9,10,11,12). Contributing factors to instability include component malposition, impingement, inadequate soft tissue tension, abductor insufficiency, neuromuscular disorders, cognitive dysfunctions, and non-compliance. In many cases, the etiology is multifactorial1,6,9). Several prosthetic options and surgical approaches have been tested to both prevent and treat instability (e.g., component reorientation, trochanteric advancement, soft tissue reinforcement, capsulorraphy, large-diameter femoral heads, constrained liners, dual mobility constructs)3,12,13,14,15), however, none are without risk9,16).
The purpose of this study is to review literature comparing outcomes of constrained acetabular components with those of dual mobility (DM) cups in revision THA. In view of the increased importance of revision THAs, this study aims to compare the outcomes of both implants including overall survival, dislocation rates, implant migration/loosening, DM implant-specific complications (IPD), and functional outcomes (Harris hip score, HHS). Conversion to a DM design and a constrained liner are both procedures for recurrent instability. In current clinical practice, constrained liners are only used in a salvage procedure17,18); DM cups, on the other hand, are also used in primary THA.
The principle feature of constrained acetabular components is to hold the femoral head captive within the cup by means of a locking ring mechanism8,13,19,20). There are two prevailing designs for constrained implants: the constrained tripolar and the locking ring5,13). In this study, constrained tripolar devices are not included. A constrained liner is a device consisting of a polyethylene (PE) liner and a reinforcing ring. As shown in Fig. 19,18,21), the metal locking ring fits into a groove on the outer surface of the liner to reinforce the capture of the femoral head within the liner22). These constrained components are designed to physically resist dislocation by avoiding re-expansion of the liner. The metal locking ring allows immediate stability, but may be associated with a reduced range of motion (ROM) due to impingement of the neck on the PE liner8,22). The restricted movement of the components may increase the forces at the bone/implant interface, increasing the risk of aseptic loosening and the necessity for re-revision16). Implant failure can be classified to five distinct types: Type I failure is failure at the implant-bone interface. Type II and III failures are respectively disconnection of the PE liner from the cup or locking ring failure. Type IV failure is dislocation of the femoral head from the liner. Type V failure is infection (Table 1)1,8,9,12,23,24,25).
Fig. 1. Example of constrained acetabular liners with locking ring mechanism. (A) Duraloc constrained system (DePuy Synthes®). (B) Trilogy Acetabular System Constrained Liner (Zimmer®)9,18,21).
Table 1. Implant Failure Classifications8,9,12,23,24,25).
| Type I | Failure at bone/cup interface (cementless) or bone/cement interface (cemented) |
| Type II | Liner/cup (cementless) or liner/cement (cemented) disengagement |
| Type III | Locking ring failure |
| Type IV | Dislocation of femoral head from PE Liner |
| Type V | Infection (deep/superficial) |
PE: polyethylene.
DM cups ensure stability using a different biomechanical concept than the constrained locking ring mechanism. While the DM design has been utilized for more than 25 years in Europe, it only became available in North-America in 2009 when FDA approval was achieved1,2,26). A DM cup is made up of two different articulations, one between the femoral head and the PE liner, and a second between the liner and the acetabular shell2). Thus, the first motion arises between the small femoral head and the inner/concave surface of the PE liner, until the neck of the femoral stem reaches the liner. As shown in Fig. 227), the secondary motion occurs between the outer/convex surface of the PE liner and the acetabular metal cup, when a higher ROM is required1,28). DM is therefore expected to improve the ROM to impingement29). The purpose of DM cups is to provide improved head-to-neck ratio, a larger effective head size and a higher jump distance, all of which lead to enhanced stability1,10,26,30,31). Major concerns with DM cups include accelerated PE wear due to an additional bearing surface, aseptic loosening and IPD1,2,3,29). IPD is a unique failure mechanism for DM constructs, in which the inner femoral head dislodges from the outer PE liner10,32). The metallic head can then articulate directly with the acetabular socket, causing devastating complications, including rapid wear and severe soft tissue metallosis, that will necessitate re-revision2,14,32). To prevent the mentioned concerns about accelerated wear on the additional bearing surface, cross-linked ultra-high molecular weight polyethylene (UHMWPE) doped with vitamin-E (an anti-oxidant) have been integrated into DM implants2,29,33).
Fig. 2. (A) Articular motion of the dual mobility (DM) design. The femoral head rotates within the liner until contact is made (α), at which point the liner then rotates within acetabular shell (β). (B) Example of DM cup (MDM Stryker)27).
MATERIALS AND METHODS
PubMed/Medline, Embase, Cochrane Library and Trip Database were searched for articles published up to May 2018. A search was conducted using various combinations of the following search terms: “total hip arthroplasty”, “total hip replacement”, “revision hip arthroplasty”, “revision hip replacement”, “constrained”, “constrained liners”, “constrained acetabular liners”, “constrained components”, “dual mobility”, “dual-mobility”, “bousquet”, “complications”, “outcome”, “failure”, “instability”, “dislocation”, “intra-prosthetic dislocation”, “loosening”, “aseptic loosening”, “acetabular loosening”. There were no set limitations on the year of publication1). Only English language papers were evaluated. No limitations were set concerning the category of study (retro-/prospective design), follow-up period or the number of patients enlisted. Inclusion criteria were articles investigating the outcomes of revision THA with constrained acetebular components or a DM design. Patients undergoing primary THA or revision with tripolar constrained devices (Omnifit Osteonics; Trident Stryker, Kalamazoo, MI, USA) were not included in the review. Duplicates, literature reviews, animal studies, in vitro investigations, cadaveric studies, technical notes and instructional courses were excluded. Additionally, reference lists of the included papers were manually inspected by the author for missed articles1,32). A flow chart is enclosed according to the PRISMA guidelines (Fig. 3).
Fig. 3. Literature search/flow chart according to PRISMA guidelines.
DM: dual mobility, CAL: constrained acetabular liners.
A variety of outcomes were assessed. For each study, we collected data including the number of hips undergoing revision THA, the mean follow-up duration and the mean age of the study population. First, we assessed the overall survival rates. Survival rate or cup survivorship was defined as “1-surgical re-revision due to all causes”. Second, we assessed the incidence rates of the most common complications in revision THA. The first complication evaluated in this study was dislocation rate. Thereafter, we assessed the rate of acetabular loosening, classically described as deterioration of bone implant interface with a measurable migration of the cup34,35). Acetabular loosening is mostly evaluated radiographically according to the DeLee and Charnley classification (Fig. 4)23,36,37,38,39,40).
Fig. 4. (A) DeLee and Charnley acetabular zones. The acetabulum is divided into three equal zones. Acetabular components are considered loose if they have migrated >3 mm in one zone, or if there is a circumferential peri-acetabular radiolucent line of at least 2 mm in all three zones23,37,38,39,40). (B) Radiolucent zone around acetabular component in zone I and zone II. (C) Acetabular inclination, the angle between the transischial line and a line through the lateral and medial margins of the acetabular cup.
Finally, the rate of IPD was analysed in all studies involving DM constructs.
Clinical evaluations were assessed using the HHS, a clinician-based outcome scale conducted by a well-qualified, professional health care provider41,42). This standardized outcome measure is a rating scale of 100 points with domains of pain, function, ROM and deformity. Higher scores represent better outcomes and less dysfunction41,42).
RESULTS
The literature search resulted in a total of 46 references. We divided the studies that met the inclusion criteria into two distinct groups. The first group included 15 studies reporting use of constrained acetabular liners (CAL) components in revision THA. Of the 15 includes articles, 14 had a retrospective design. Studies were published starting from 1994, the latest in 2017. A total of 1,924 hips were analysed in this group. The mean follow-up period ranged from 19 months to 120 months. The average age at the time of the procedure was 69.0±7.0 years. The mean dislocation rate was 11.0% (range, 0–29.0%; standard deviation [SD], 8.7) and average survival rate was 81.0% (range, 54.7–97.2%; SD, 10.9). Evaluating all cases of acetabular loosening revealed a mean rate of 2.0% (range, 0–8.3%; SD, 2.8).
The second group included 31 studies dealing with DM devices in the setting of revision THA; of these, 23 were retrospective. Analysis of the 31 reports included 3,693 hips. The first reports included in this review were published in 2009, at the time when FDA approval for DM devices was obtained in the United States. The average follow-up time ranged from 16 to 87 months. Mean age at surgery of the overall study population was 68.8±6.2 years. Dislocation rates vary from 0–10.4% with a mean dislocation rate of 2.6±2.5%. Overall survival ranged between 88.9% and 100%, with a mean rate of 94.7±3.5%. Analysis of acetabular loosening revealed a rate of 1.0% (range, 0–6.4%; SD, 1.5). Of the 31 studies, 6 report one or more cases of IPD (range, 0–4.0%). The average rate of IPD was 0.6±1.2%. These data reconfirm that this unique complication encountered with DM cups is extremely rare3,15,32).
Table 2, 3, 44,9,10,11,12,14,16,19,21,22,23,24,28,30,31,32,36,39,43,44,45,46,47,48,49,50,51,52,53,54,55,56,57,58,59,60,61,62,63,64,65,66,67,68,69,70,71,72,73) summarize the extracted data. The results of this study led to the development of some general considerations about both designs32).
Table 2. Reports of Revision THA with Constrained Acetabular Components.
| Study (year) | Implant (manufacturer) | No. of hip | Mean follow-up (mo) | Mean age (yr) | Dislocation rate (%) | Survival rate (%) | Acetabular loosening (%) | HHS (preope ration) | HHS at last follow-up |
|---|---|---|---|---|---|---|---|---|---|
| Anderson et al. (1994)39) | S-ROM (Depuy) | 21 | 31 | 65.5 | 29 | 71 | 0 | - | 76 |
| Stanton et al. (2002)48) | S-ROM (Depuy), Mallory-Head CL (Biomet) | 13 | 43 | 73 | 0 | 92.3 | 0 | - | 72 |
| Della Valle et al. (2005)19) | Duraloc CL (Depuy) | 55 | 19 | 62 | 16 | 84 | 0 | - | - |
| Berend et al. (2005)47) | S-Rom Poly-Dial (Depuy), Ringloc (Biomet) | 565 | 128 | 67.2 | 18.6 | 54.7 | 8.3 | 44.1 | 64.6 |
| Knudsen et al. (2007)49) | Trilogy CL (Zimmer) | 40 | 27 | - | 10 | 87.5 | 0 | - | - |
| Rady et al. (2010)22) | Duraloc CL (Depuy), Trilogy CL (Zimmer) | 15 | 26 | 57.4 | 6.6 | 93.4 | 0 | 22 | 85 |
| Pattyn et al. (2010)21) | Trilogy CL (Zimmer), Ringloc (Biomet) | 46 | 66 | 59.6 | 21 | 74 | 4.3 | - | 59 |
| Andersen et al. (2013)9) | Trilogy CL & LCL (Zimmer) | 32 | 21 | 74 | 12 | 79 | 3.1 | - | 81 |
| Munro et al. (2013)24) | Trilogy LCL (Zimmer) | 72 | 34 | 68 | 3.7 | 86.6 | 2.4 | - | - |
| Mäkinen et al. (2016)23) | Trilogy LCL (Zimmer) | 98 | 38 | 69.4 | 5 | 84 | - | 69.8 | 85.4 |
| Hernandez-Vaquero et al. (2016)46) | Lefèvre retentive cup (Lepine) | 36 | 25 | 83 | 2.8 | 97.2 | 0 | - | - |
| Clavé et al. (2016)16) | Lefèvre retentive cup (Lepine) | 121 | 78 | 75.9 | 8 | 87.5 | 1.8 | - | - |
| Chalmers et al. (2016)12) | Trilogy LCL (Zimmer) | 58 | 42 | 69 | 19 | 76 | 3.5 | - | 74 |
| Lewis et al. (2017)44) | Mixed | 700 | 120 | - | - | 73.1 | 7.3 | - | - |
| Karvonen et al. (2017)50) | Freedom CL (Biomet) | 52.0 | 28.0 | 73.4 | 7.7 | 74.0 | 3.8 | - | - |
THA: total hip arthroplasty, HHS: Harris hip score.
Table 3. Reports of Revision THA with Dual Mobility Design (I).
| Study (year) | Implant (manufacturer) | No. of hip | Mean follow-up (mo) | Mean age (yr) | Dislocation rate (%) | Survival rate (%) | IPD (%) | Acetabular loosening (%) | HHS (preope ration) | HHS at last follow-up |
|---|---|---|---|---|---|---|---|---|---|---|
| Guyen et al. (2009)53) | Saturne (Amplitude) | 54 | 47 | 66.5 | 1.9 | 90.8 | 3.7 | 0 | 68.8 | 83.7 |
| Philippot et al. (2009)11) | Novae (Serf) | 163 | 60 | 68.7 | 3.7 | 96.1 | 0 | 1.2 | - | - |
| Massin and Besnier (2010)54) | Collégia (Wright) | 23 | 54 | 68 | 8.6 | 95.7 | 0 | 4.3 | - | - |
| Leiber-Wackenheim et al. (2011)36) | Novae-1/Novae-E (Serf) | 59 | 96 | 68 | 1.7 | 98 | 0 | 0 | - | 86.7 |
| Schneider et al. (2011)55) | Novae (Serf) | 96 | 42 | 69.9 | 10.4 | 95.6 | 0 | 1 | - | - |
| Pattyn and Audenaert (2012)56) | Apogée (Biotechni) | 37 | 16 | 70.4 | 5.4 | 97.3 | 0 | 0 | 39.9 | - |
| Mertl et al. (2012)57) | - | 180 | 43 | 67.4 | 4.8 | 92.6 | 1.4 | 1.4 | 76.9 | 83.9 |
| Civinini et al. (2012)58) | Avantage (Biomet) | 33 | 36 | 69 | 0 | 97 | 0 | 0 | 48 | 86 |
| Hailer et al. (2012)43) | Avantage (Biomet) | 228 | 24 | 75 | 2 | 93 | - | 2 | - | - |
| Saragaglia et al. (2013)59) | Mixed | 29 | 46 | 75.6 | 3.4 | 100 | - | 0 | - | - |
| Mukka et al. (2013)60) | Avantage (Biomet) | 34 | 18 | 75.7 | 6 | - | - | - | - | 67 |
| Prudhon et al. (2014)61) | Ades (Dedienne), Integra (Lépine) | 79 | 24 | 75.5 | 1.3 | 97.3 | - | 2.7 | - | - |
| Jakobsen et al. (2014)62) | Saturne (Amplitude) | 56 | 44 | 72 | 1.8 | 94.7 | 1.8 | 1.8 | 76 | 87 |
| van Heumen et al. (2015)28) | Avantage (Biomet) | 50 | 29 | 67 | 0 | 93 | - | 2 | - | - |
| Snir et al. (2015)63) | ADM/MDM (Stryker), AA E1 (Biomet) | 18 | 17 | 51 | 0 | 100 | 0 | 0 | 28 | 66 |
THA: total hip arthroplasty, IPD: intra-prosthetic dislocation, HHS: Harris hip score.
Table 4. Reports of Revision THA with Dual Mobility Design (II).
| Study (year) | Implant (manufacturer) | No. of hip | Mean follow-up (mo) | Mean age (yr) | Dislocation rate (%) | Survival rate (%) | IPD (%) | Acetabular loosening (%) | HHS (preope ration) | HHS at last follow-up |
|---|---|---|---|---|---|---|---|---|---|---|
| Wegrzyn et al. (2015)45) | Saturne (Amplitude) | 994 | 87 | 70 | 1.5 | - | 0.2 | 0 | - | - |
| Mohammed et al. (2015)14) | Novae (Serf) | 24 | 22 | 70.8 | 0 | 100 | 0 | - | - | - |
| Simian et al. (2015)70) | DMS/Evora (SEM), Mobilité (Tornier) | 72 | 87 | 67.9 | 1.4 | 90 | 0 | 1.4 | - | 80.4 |
| Haen et al. (2015)64) | Saturne (Amplitude) | 66 | 50 | 79.8 | 0 | 98 | 0 | 1.5 | - | - |
| Viste et al. (2017)4) | Novae (Serf) | 334 | 60 | - | 3.3 | - | 0 | 3 | - | - |
| Jauregui et al. (2016)10) | MDM (Stryker) | 60 | 30 | 57 | 1.7 | - | 0 | 1.7 | - | 89 |
| Dangin et al. (2016)71) | Novae E (Serf) | 91 | 33 | 71 | 3.5 | 91.4 | 0 | 0 | - | - |
| Carulli et al. (2016)72) | Avantage (Biomet) | 31 | 46 | 75.4 | 0 | 100 | 0 | 0 | 62.2 | 76 |
| Plummer et al. (2016)65) | ADM (Stryker) | 36 | 29 | 64 | 2.7 | 88.9 | 0 | 0 | 45 | 90 |
| Gonzalez et al. (2017)66) | Polarcup (S&N), Versafit (Medacta) | 150 | 31 | 73 | 2.7 | 91 | - | 0.6 | - | - |
| Plummer et al. (2016)67) | Saturne (Amplitude) | 25 | 29 | 61.7 | 0 | 96 | 4 | 0 | 57 | 87 |
| Lebeau et al. (2017)68) | Quattro (Lépine) | 62 | 77 | 70.5 | 1.6 | 91.9 | 0 | 6.4 | 49 | 73 |
| Mohaddes et al. (2017)69) | Avantage (Biomet) | 436 | 48 | 75 | 1.6 | 91 | - | - | - | - |
| Sutter et al. (2017)31) | MDM (Stryker) | 64 | 36 | 59 | 3 | 91 | - | 1.5 | - | - |
| Lange et al. (2018)30) | ADM (Stryker), MDM (Stryker) | 40 | 36 | 64 | 5 | 90 | 2.5 | 0 | - | - |
| Brüggeman et al. (2018)73) | Avantage (Biomet) | 69 | 59 | 67 | 1.4 | 96 | 0 | 2.9 | - | 77 |
THA: total hip arthroplasty, IPD: intra-prosthetic dislocation, HHS: Harris hip score.
DISCUSSION
As seen from the aforementioned results, our main finding was that DM cups seem preferable to constrained cups. This literature review revealed higher survival rates for the DM designs compared to the CAL designs (94.7% vs. 81.0%). Overall dislocation rates and loosening rates using DM in revision THA are lower than those reported in revision THA with a constrained implant (2.6% vs. 11.0% and 1.0% vs. 2.0%, respectively). We emphasize that the success of these implants are reported with great variation. It is important to bear in mind that any type of prosthesis (including DM constructs or CAL constructs) cannot compensate for poor surgical technique or technical errors, such as incorrect implant orientation or inadequate restoration of soft-tissue tension5,17,26,29). Surgeons must consider all the involved parameters including implant features, patient characteristics, positioning and soft tissues conditions. Risk factors for re-revision were described as younger age (50–59 yr) at the time of index surgery, revision for recurrent dislocation, revision following deep infection and the number of previous revision surgeries at the same joint11,43,51). We report that most studies assess an elderly population (>65 yr), with few data on younger more active patients3). It is necessary to exercise caution when using DM or constrained implants in this high-demand population, in whom increased wear could be problematic2,3).
Our results are in line with previous reviews. The general limitations of constrained designs are the decreased ROM to impingement, risk of breakage of the liner or retaining ring, increased component loosening due to high interfacial stress and excessive volumetric PE wear5,8,16,17,20,29,52). Considering the high failure rates revealed in this review, we suggest that clinicians maintain the use of constrained liners limited to salvage scenarios in selected patients in whom other operative options are not practical or have not been successful5,8,17,18). The most documented indications for use of a CAL design nowadays are senile, low-activity patients, including those with neuromuscular disorders, cognitive impairment, idiopathic instability, abductor insufficiency, or insufficient soft tissue tension5,8,12,13,18,44,45,46).
As mentioned above, DM devices provide a greater effective head size and improve the head-to-neck ratio of the implant10,38). Therefore, DM is expected to increase the arc of motion of the hip joint contrary to constrained devices. However, the additional bearing surface in DM cups raises major concerns about accelerated wear and loosening as well as the newly encountered failure method IPD. Our findings demonstrate that IPD is exceedingly rare and the favorable low rates of loosening indicate that DM cups do not result in high stresses at the bone-implant interface70).
DISLOCATION
Our findings demonstrate great variation between the dislocation rate in DM and CAL designs. DM devices seem to lower the risk of dislocation in revision THA to 0.0–10.4% (mean, 2.6%) compared to standard designs (5.0–30.0%)2,4,5,9,10,11,12,74). Constrained designs have been reported with moderate success revealing dislocation rates up to 29.0% (mean, 11.0%). These results suggest that DM cups are effective in minimising the risk of dislocation after revision THA and are probably going to become the gold standard in treating THA dislocations1,17,71,75).
FOLLOW-UP TIME
Another issue to be considered is the fact that all but two studies involving the CAL design, have a short or medium follow-up time (<10 yr)44,47). In the DM cup reports specifically, there are limited medium-term and no longer-term studies, likely because DM was unavailable in the United States until FDA approval in 2009. Since DM implants are utilized with increasing frequency in North America, further and longer-term follow-up will be needed to confirm the medium-term results.
INFECTION
Surprisingly, few studies evaluate postoperative infectious complications (e.g., superficial soft tissue colonization, deep peri-implant infection). Currently there is no scientific evidence suggesting that one of the discussed designs leads to significantly higher infection rates.
SURGICAL APPROACH
An additional factor that may influence the outcome of revision THA, in particular the dislocation rate, is the surgical approach. Multiple reports demonstrated that the posterior approach to the hip joint, requiring detachment of the posterior capsule and short external rotators, is correlated with an increased dislocation rate compared to the anterior, anterolateral, and lateral approaches6,76,77,78). We were unable to draw significant conclusions here due to incomplete reports and substantial differences in indications for revision THA in the included studies.
FUNCTIONAL OUTCOMES
Another finding of this study is the lack of functional results in accordance with a clinical scale28). Few reports (<50%) include clinician- and/or patient-reported functional outcomes in their results. The HHS is most commonly used, however, many studies report incomplete functional results without comparative preoperative scores. Scores established only at clinical follow-up but not at baseline are insufficient to allow for conclusions related to the potential impact of the treatment on quality of life73). In a limited number of French studies, the Postel-Merle d'Aubigné score was used. Additionally, few patient-reported outcome measures are described (e.g., Oxford hip score, WOMAC score). We therefore suggest using universal clinical measurement instruments (clinician- and patient-reported) with comparative preoperative and follow-up postoperative results to provide information about the functional performance of the two implants28).
LIMITATIONS
This review has several limitations1,70,72). The main limitation is the inconsistent quality of the included studies. First, most papers included in this review have a retrospective design. Second, the majority of the studies are not randomized and lack a control group. Thirdly, many included reports have a small cohort size with an often heterogeneous patient population70). Indications for revision THA are not taken into account individually and there is great variability among the patient population regarding demographics and diagnosis. Furthermore, various modifications and improvements have been made over the years to the mechanics, metallurgy and materials of the original DM or CAL designs2). Subsequent generations of the DM constructs with highly cross-linked UHMWPE and vitamin-E impregnated PE, dual hydroxyapatite and titanium coating instead of alumina, modular shells for screw fixation, cemented designs have all been introduced to improve cup fixation and decrease complications such as PE wear, dislocation, IPD and psoas irritation2,3,26). For this reason it may be recommended to assess and compare the outcomes of different DM and CAL implant designs mutually in the future. No meta-analysis was performed. Finally, we should mention selection bias as a weakness of the present study as specific data was selectively extracted from the articles1). These limitations confine the level of evidence of this paper.
CONCLUSION
In conclusion, this comparative literature review revealed that designs with constrained acetabular components have poorer outcomes as compared to DM cups in revision THA. The use of DM seems more appropriate at the current time since they impose lower rates of dislocations, loosening and re-revisions in the short- and mid-term. While IPD is a new failure mechanism related to DM designs, it is extremely uncommon26,71). Additional data will be needed to assess long-term survivorship of DM cups and concerns regarding increased wear in a younger, high-demand patient population. Constrained designs are still an alternative option but only as salvage procedure for selected, low-demand patients in case of failure of previous treatments12).
Footnotes
CONFLICT OF INTEREST: The authors declare that there is no potential conflict of interest relevant to this article.
References
- 1.De Martino I, D'Apolito R, Soranoglou VG, Poultsides LA, Sculco PK, Sculco TP. Dislocation following total hip arthroplasty using dual mobility acetabular components: a systematic review. Bone Joint J. 2017;99-B(ASuppl1):18–24. doi: 10.1302/0301-620X.99B1.BJJ-2016-0398.R1. [DOI] [PubMed] [Google Scholar]
- 2.Plummer DR, Haughom BD, Della Valle CJ. Dual mobility in total hip arthroplasty. Orthop Clin North Am. 2014;45:1–8. doi: 10.1016/j.ocl.2013.08.004. [DOI] [PubMed] [Google Scholar]
- 3.De Martino I, Triantafyllopoulos GK, Sculco PK, Sculco TP. Dual mobility cups in total hip arthroplasty. World J Orthop. 2014;5:180–187. doi: 10.5312/wjo.v5.i3.180. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Viste A, Desmarchelier R, Fessy MH. Dual mobility cups in revision total hip arthroplasty. Int Orthop. 2017;41:535–542. doi: 10.1007/s00264-016-3363-4. [DOI] [PubMed] [Google Scholar]
- 5.Abdel MP, Padgett DE. Constrained liners in revision total hip replacement. Bone Joint J. 2013;95 Suppl A doi: 10.1302/0301-620X/95B.33361. [DOI] [Google Scholar]
- 6.Dargel J, Oppermann J, Brüggemann GP, Eysel P. Dislocation following total hip replacement. Dtsch Arztebl Int. 2014;111:884–890. doi: 10.3238/arztebl.2014.0884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Wera GD, Ting NT, Moric M, Paprosky WG, Sporer SM, Della Valle CJ. Classification and management of the unstable total hip arthroplasty. J Arthroplasty. 2012;27:710–715. doi: 10.1016/j.arth.2011.09.010. [DOI] [PubMed] [Google Scholar]
- 8.Jones SA. Constrained acetabular liners. J Arthroplasty. 2018;33:1331–1336. doi: 10.1016/j.arth.2018.01.026. [DOI] [PubMed] [Google Scholar]
- 9.Andersen AV, Kjersgaard AG, Solgaard S. Trilogy-constrained acetabular component for recurrent dislocation. ISRN Orthop. 2013;2013:629201. doi: 10.1155/2013/629201. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Jauregui JJ, Pierce TP, Elmallah RK, Cherian JJ, Delanois RE, Mont MA. Dual mobility cups: an effective prosthesis in revision total hip arthroplasties for preventing dislocations. Hip Int. 2016;26:57–61. doi: 10.5301/hipint.5000295. [DOI] [PubMed] [Google Scholar]
- 11.Philippot R, Adam P, Reckhaus M, et al. Prevention of dislocation in total hip revision surgery using a dual mobility design. Orthop Traumatol Surg Res. 2009;95:407–413. doi: 10.1016/j.otsr.2009.04.016. [DOI] [PubMed] [Google Scholar]
- 12.Chalmers BP, Arsoy D, Sierra RJ, Lewallen DG, Trousdale RT. High failure rate of modular exchange with a specific design of a constrained liner in high-risk patients undergoing revision total hip arthroplasty. J Arthroplasty. 2016;31:1963–1969. doi: 10.1016/j.arth.2016.02.021. [DOI] [PubMed] [Google Scholar]
- 13.Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop Relat Res. 2004;(420):122–129. doi: 10.1097/00003086-200403000-00017. [DOI] [PubMed] [Google Scholar]
- 14.Mohammed R, Hayward K, Mulay S, Bindi F, Wallace M. Outcomes of dual-mobility acetabular cup for instability in primary and revision total hip arthroplasty. J Orthop Traumatol. 2015;16:9–13. doi: 10.1007/s10195-014-0324-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Batailler C, Fary C, Verdier R, Aslanian T, Caton J, Lustig S. The evolution of outcomes and indications for the dual-mobility cup: a systematic review. Int Orthop. 2017;41:645–659. doi: 10.1007/s00264-016-3377-y. [DOI] [PubMed] [Google Scholar]
- 16.Clavé A, Maurer D, Tristan L, Dubrana F, Lefèvre C, Pandit H. Midterm survivorship of the Lefèvre constrained liner: a consecutive multisurgeon series of 166 cases. J Arthroplasty. 2016;31:1970–1978. doi: 10.1016/j.arth.2016.02.031. [DOI] [PubMed] [Google Scholar]
- 17.Zagra L, Caboni E. Total hip arthroplasty instability treatment without dual mobility cups: brief overview and experience of other options. Int Orthop. 2017;41:661–668. doi: 10.1007/s00264-016-3383-0. [DOI] [PubMed] [Google Scholar]
- 18.Williams JT, Jr, Ragland PS, Clarke S. Constrained components for the unstable hip following total hip arthroplasty: a literature review. Int Orthop. 2007;31:273–277. doi: 10.1007/s00264-006-0191-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Della Valle CJ, Chang D, Sporer S, Berger RA, Rosenberg AG, Paprosky WG. High failure rate of a constrained acetabular liner in revision total hip arthroplasty. J Arthroplasty. 2005;20(7 Suppl 3):103–107. doi: 10.1016/j.arth.2005.05.005. [DOI] [PubMed] [Google Scholar]
- 20.Khan RJ, Fick D, Alakeson R, Li MG, Nivbrant B, Wood D. The constrained acetabular component for hip instability. J Arthroplasty. 2007;22:377–382. doi: 10.1016/j.arth.2006.04.020. [DOI] [PubMed] [Google Scholar]
- 21.Pattyn C, De Haan R, Kloeck A, Van Maele G, De Smet K. Complications encountered with the use of constrained acetabular prostheses in total hip arthroplasty. J Arthroplasty. 2010;25:287–294. doi: 10.1016/j.arth.2008.10.010. [DOI] [PubMed] [Google Scholar]
- 22.Rady AE, Asal MK, Bassiony AA. The use of a constrained cementless acetabular component for instability in total hip replacement. Hip Int. 2010;20:434–439. doi: 10.1177/112070001002000404. [DOI] [PubMed] [Google Scholar]
- 23.Mäkinen TJ, Fichman SG, Rahman WA, et al. The focally constrained liner is a reasonable option for revision of unstable total hip arthroplasty. Int Orthop. 2016;40:2239–2245. doi: 10.1007/s00264-015-3082-2. [DOI] [PubMed] [Google Scholar]
- 24.Munro JT, Vioreanu MH, Masri BA, Duncan CP. Acetabular liner with focal constraint to prevent dislocation after THA. Clin Orthop Relat Res. 2013;471:3883–3890. doi: 10.1007/s11999-013-2858-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Guyen O, Lewallen DG, Cabanela ME. Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am. 2008;90:1553–1560. doi: 10.2106/JBJS.G.00317. [DOI] [PubMed] [Google Scholar]
- 26.Abdel MP. Dual-mobility constructs in revision total hip arthroplasties. J Arthroplasty. 2018;33:1328–1330. doi: 10.1016/j.arth.2018.01.030. [DOI] [PubMed] [Google Scholar]
- 27.Moore C, Orlandini L. The dual mobility concept in total hip arthroplasty. Bone Joint Sci. 2012;3:1–8. [Google Scholar]
- 28.van Heumen M, Heesterbeek PJ, Swierstra BA, Van Hellemondt GG, Goosen JH. Dual mobility acetabular component in revision total hip arthroplasty for persistent dislocation: no dislocations in 50 hips after 1-5 years. J Orthop Traumatol. 2015;16:15–20. doi: 10.1007/s10195-014-0318-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Guyen O. Constrained liners, dual mobility or large diameter heads to avoid dislocation in THA. EFORT Open Rev. 2017;1:197–204. doi: 10.1302/2058-5241.1.000054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Lange JK, Spiro SK, Westrich GH. Utilizing dual mobility components for first-time revision total hip arthroplasty for instability. J Arthroplasty. 2018;33:505–509. doi: 10.1016/j.arth.2017.09.029. [DOI] [PubMed] [Google Scholar]
- 31.Sutter EG, McClellan TR, Attarian DE, Bolognesi MP, Lachiewicz PF, Wellman SS. Outcomes of modular dual mobility acetabular components in revision total hip arthroplasty. J Arthroplasty. 2017;32:S220–S224. doi: 10.1016/j.arth.2017.03.035. [DOI] [PubMed] [Google Scholar]
- 32.De Martino I, D'Apolito R, Waddell BS, McLawhorn AS, Sculco PK, Sculco TP. Early intraprosthetic dislocation in dual-mobility implants: a systematic review. Arthroplast Today. 2017;3:197–202. doi: 10.1016/j.artd.2016.12.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Oral E, Christensen SD, Malhi AS, Wannomae KK, Muratoglu OK. Wear resistance and mechanical properties of highly cross-linked, ultrahigh-molecular weight polyethylene doped with vitamin E. J Arthroplasty. 2006;21:580–591. doi: 10.1016/j.arth.2005.07.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Rastogi S, Trikha V, Bhan S, Bal C. Aseptic loosening of the acetabular component in relation to the positioning of prosthesis. Indian J Orthop. 2002;36:3 [Google Scholar]
- 35.Picado CHF, Garcia FL, Kanabushi C, Carvalho S. Pain as symptom of threaded acetabular cup loosening. Acta Ortop Bras. 2005;13:2004–2006. [Google Scholar]
- 36.Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabrion A, Mertl P. Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: a 59 cases series with a mean 8 years' follow-up. Orthop Traumatol Surg Res. 2011;97:8–13. doi: 10.1016/j.otsr.2010.08.003. [DOI] [PubMed] [Google Scholar]
- 37.Williams J, Neep M. Radiographic evaluation of hip replacements. Spectrum. 2012;19:8–12. Available from: https://www.minnisjournals.com.au/articles/rad%20hip%20assess%20spec%20sept%2012.pdf. [Google Scholar]
- 38.Chalmers BP, Ledford CK, Taunton MJ, Sierra RJ, Lewallen DG, Trousdale RT. Cementation of a dual mobility construct in recurrently dislocating and high risk patients undergoing revision total arthroplasty. J Arthroplasty. 2018;33:1501–1506. doi: 10.1016/j.arth.2017.11.055. [DOI] [PubMed] [Google Scholar]
- 39.Anderson MJ, Murray WR, Skinner HB. Constrained acetabular components. J Arthroplasty. 1994;9:17–23. doi: 10.1016/0883-5403(94)90133-3. [DOI] [PubMed] [Google Scholar]
- 40.Watt I, Boldrik S, Langelaan EV, Smithuis R. Hip-arthroplasty [Internet] The Radiology Assistant; 2006. [cited 2018 Nov 24]. Available from: http://www.radiologyassistant.nl/en/p431c8258e7ac3/hip-arthroplasty.html. [Google Scholar]
- 41.Nilsdotter A, Bremander A. Measures of hip function and symptoms: Harris Hip Score (HHS), Hip Disability and Osteoarthritis Outcome Score (HOOS), Oxford Hip Score (OHS), Lequesne Index of Severity for Osteoarthritis of the Hip (LISOH), and American Academy of Orthopedic Surgeons (AAOS) hip and knee questionnaire. Arthritis Care Res (Hoboken) 2011;63 Suppl 11:S200–S207. doi: 10.1002/acr.20549. [DOI] [PubMed] [Google Scholar]
- 42.Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755. [PubMed] [Google Scholar]
- 43.Hailer NP, Weiss RJ, Stark A, Kärrholm J. Dual-mobility cups for revision due to instability are associated with a low rate of re-revisions due to dislocation: 228 patients from the Swedish Hip Arthroplasty Register. Acta Orthop. 2012;83:566–571. doi: 10.3109/17453674.2012.742395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Lewis PL, Graves SE, de Steiger RN, Cuthbert AR. Constrained acetabular components used in revision total hip arthroplasty: a registry analysis. J Arthroplasty. 2017;32:3102–3107. doi: 10.1016/j.arth.2017.04.043. [DOI] [PubMed] [Google Scholar]
- 45.Wegrzyn J, Tebaa E, Jacquel A, Carret JP, Béjui-Hugues J, Pibarot V. Can dual mobility cups prevent dislocation in all situations after revision total hip arthroplasty? J Arthroplasty. 2015;30:631–640. doi: 10.1016/j.arth.2014.10.034. [DOI] [PubMed] [Google Scholar]
- 46.Hernandez-Vaquero D, Llorens De Los Rios J, Perez-Coto I, Iglesias-Fernandez S, Fernandez-Lombardia J, Cuervo-Olay MDC. Constrained acetabular liners in hip revision surgery. A low-cost solution for senile patients. Acta Orthop Belg. 2016;82:563–569. [PubMed] [Google Scholar]
- 47.Berend KR, Lombardi AV, Jr, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty. 2005;20(7 Suppl 3):93–102. doi: 10.1016/j.arth.2005.06.001. [DOI] [PubMed] [Google Scholar]
- 48.Stanton DA, Bruce WJ, Goldberg JA, Walsh W. Salvaging unstable or recurrent dislocating total hip arthroplasty with the constrained acetabular component. J Orthop Surg (Hong Kong) 2002;10:165–169. doi: 10.1177/230949900201000210. [DOI] [PubMed] [Google Scholar]
- 49.Knudsen R, Ovesen O, Kjaersgaard-Andersen P, Overgaard S. Constrained liners for recurrent dislocations in total hip arthroplasty. Hip Int. 2007;17:78–81. doi: 10.1177/112070000701700204. [DOI] [PubMed] [Google Scholar]
- 50.Karvonen M, Karvonen H, Seppänen M, Liukas A, Koivisto M, Mäkelä KT. Freedom constrained liner for the treatment and prevention of dislocation in total hip arthroplasty. Scand J Surg. 2017;106:165–172. doi: 10.1177/1457496916660035. [DOI] [PubMed] [Google Scholar]
- 51.Jo S, Jimenez Almonte JH, Sierra RJ. The cumulative risk of re-dislocation after revision THA performed for instability increases close to 35% at 15years. J Arthroplasty. 2015;30:1177–1182. doi: 10.1016/j.arth.2015.02.001. [DOI] [PubMed] [Google Scholar]
- 52.Fricka KB, Marshall A, Paprosky WG. Constrained liners in revision total hip arthroplasty: an overuse syndrome: in the affirmative. J Arthroplasty. 2006;21(4 Suppl 1):121–125. doi: 10.1016/j.arth.2006.02.100. [DOI] [PubMed] [Google Scholar]
- 53.Guyen O, Pibarot V, Vaz G, Chevillotte C, Béjui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res. 2009;467:465–472. doi: 10.1007/s11999-008-0476-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Massin P, Besnier L. Acetabular revision using a press-fit dual mobility cup. Orthop Traumatol Surg Res. 2010;96:9–13. doi: 10.1016/j.rcot.2009.11.006. [DOI] [PubMed] [Google Scholar]
- 55.Schneider L, Philippot R, Boyer B, Farizon F. Revision total hip arthroplasty using a reconstruction cage device and a cemented dual mobility cup. Orthop Traumatol Surg Res. 2011;97:807–813. doi: 10.1016/j.otsr.2011.09.010. [DOI] [PubMed] [Google Scholar]
- 56.Pattyn C, Audenaert E. Early complications after revision total hip arthroplasty with cemented dual-mobility socket and reinforcement ring. Acta Orthop Belg. 2012;78:357–361. [PubMed] [Google Scholar]
- 57.Mertl P, Combes A, Leiber-Wackenheim F, Fessy MH, Girard J, Migaud H. Recurrence of dislocation following total hip arthroplasty revision using dual mobility cups was rare in 180 hips followed over 7 years. HSS J. 2012;8:251–256. doi: 10.1007/s11420-012-9301-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Civinini R, Carulli C, Matassi F, Nistri L, Innocenti M. A dual-mobility cup reduces risk of dislocation in isolated acetabular revisions. Clin Orthop Relat Res. 2012;470:3542–3548. doi: 10.1007/s11999-012-2428-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Saragaglia D, Ruatti S, Refaie R. Relevance of a press-fit dual mobility cup to deal with recurrent dislocation of conventional total hip arthroplasty: a 29-case series. Eur J Orthop Surg Traumatol. 2013;23:431–436. doi: 10.1007/s00590-012-1002-3. [DOI] [PubMed] [Google Scholar]
- 60.Mukka SS, Mahmood SS, Sjödén GO, Sayed-Noor AS. Dual mobility cups for preventing early hip arthroplasty dislocation in patients at risk: experience in a county hospital. Orthop Rev (Pavia) 2013;5:48–51. doi: 10.4081/or.2013.e10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Prudhon JL, Steffann F, Ferreira A, Verdier R, Aslanian T, Caton J. Cementless dual-mobility cup in total hip arthroplasty revision. Int Orthop. 2014;38:2463–2468. doi: 10.1007/s00264-014-2448-1. [DOI] [PubMed] [Google Scholar]
- 62.Jakobsen T, Kappel A, Hansen F, Krarup N. The dislocating hip replacement-revision with a dual mobility cup in 56 consecutive patients. Open Orthop J. 2014;8:268–271. doi: 10.2174/1874325001408010268. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Snir N, Park BK, Garofolo G, Marwin SE. Revision of failed hip resurfacing and large metal-on-metal total hip arthroplasty using dual-mobility components. Orthopedics. 2015;38:369–374. doi: 10.3928/01477447-20150603-04. [DOI] [PubMed] [Google Scholar]
- 64.Haen TX, Lonjon G, Vandenbussche E. Can cemented dual-mobility cups be used without a reinforcement device in cases of mild acetabular bone stock alteration in total hip arthroplasty? Orthop Traumatol Surg Res. 2015;101:923–927. doi: 10.1016/j.otsr.2015.09.027. [DOI] [PubMed] [Google Scholar]
- 65.Plummer DR, Christy JM, Sporer SM, Paprosky WG, Della Valle CJ. Dual-mobility articulations for patients at high risk for dislocation. J Arthroplasty. 2016;31(9 Suppl):131–135. doi: 10.1016/j.arth.2016.03.021. [DOI] [PubMed] [Google Scholar]
- 66.Gonzalez AI, Bartolone P, Lubbeke A, et al. Comparison of dual-mobility cup and unipolar cup for prevention of dislocation after revision total hip arthroplasty. Acta Orthop. 2017;88:18–23. doi: 10.1080/17453674.2016.1255482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Plummer DR, Botero HG, Berend KR, Pritchett JW, Lombardi AV, Della Valle CJ. Salvage of monoblock metal-on-metal acetabular components using a dual-mobility bearing. J Arthroplasty. 2016;31:846–849. doi: 10.1016/j.arth.2015.08.016. [DOI] [PubMed] [Google Scholar]
- 68.Lebeau N, Bayle M, Belhaouane R, et al. Total hip arthroplasty revision by dual-mobility acetabular cup cemented in a metal reinforcement: a 62 case series at a minimum 5 years' follow-up. Orthop Traumatol Surg Res. 2017;103:679–684. doi: 10.1016/j.otsr.2017.04.009. [DOI] [PubMed] [Google Scholar]
- 69.Mohaddes M, Cnudde P, Rolfson O, Wall A, Kärrholm J. Use of dual-mobility cup in revision hip arthroplasty reduces the risk for further dislocation: analysis of seven hundred and ninety one first-time revisions performed due to dislocation, reported to the Swedish Hip Arthroplasty Register. Int Orthop. 2017;41:583–588. doi: 10.1007/s00264-016-3381-2. [DOI] [PubMed] [Google Scholar]
- 70.Simian E, Chatellard R, Druon J, Berhouet J, Rosset P. Dual mobility cup in revision total hip arthroplasty: dislocation rate and survival after 5 years. Orthop Traumatol Surg Res. 2015;101:577–581. doi: 10.1016/j.otsr.2015.05.002. [DOI] [PubMed] [Google Scholar]
- 71.Dangin A, Boulat S, Farizon F, Philippot R. Prevention of dislocation risk during hip revision surgery with the dual mobility concept; study of a new generation of dual mobility cups. Surg Technol Int. 2016;29:314–319. [PubMed] [Google Scholar]
- 72.Carulli C, Macera A, Matassi F, Civinini R, Innocenti M. The use of a dual mobility cup in the management of recurrent dislocations of hip hemiarthroplasty. J Orthop Traumatol. 2016;17:131–136. doi: 10.1007/s10195-015-0365-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Brüggemann A, Mallmin H, Hailer NP. Do dual-mobility cups cemented into porous tantalum shells reduce the risk of dislocation after revision surgery? Acta Orthop. 2018;89:156–162. doi: 10.1080/17453674.2018.1432927. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Noble PC, Durrani SK, Usrey MM, Mathis KB, Bardakos NV. Constrained cups appear incapable of meeting the demands of revision THA. Clin Orthop Relat Res. 2012;470:1907–1916. doi: 10.1007/s11999-011-2212-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Wirth T, Stratmann L, Hinrichs F. Evolution of late presenting developmental dysplasia of the hip and associated surgical procedures after 14 years of neonatal ultrasound screening. J Bone Joint Surg Br. 2004;86:585–589. [PubMed] [Google Scholar]
- 76.Rosenlund S, Broeng L, Jensen C, Holsgaard-Larsen A, Overgaard S. The effect of posterior and lateral approach on patient-reported outcome measures and physical function in patients with osteoarthritis, undergoing total hip replacement: a randomised controlled trial protocol. BMC Musculoskelet Disord. 2014;15:354. doi: 10.1186/1471-2474-15-354. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2005;87:2456–2463. doi: 10.2106/JBJS.D.02860. [DOI] [PubMed] [Google Scholar]
- 78.Berend KR, Sporer SM, Sierra RJ, Glassman AH, Morris MJ. Achieving stability and lower-limb length in total hip arthroplasty. J Bone Joint Surg Am. 2010;92:2737–2752. [PubMed] [Google Scholar]