Skip to main content
Journal of Radiation Research logoLink to Journal of Radiation Research
. 2020 Feb 26;61(3):457–463. doi: 10.1093/jrr/rraa007

Analysis of local setup errors of sub-regions in cone-beam CT-guided post-mastectomy radiation therapy

Jidan Zhou 1, Shuai Li 1, Chengwei Ye 1, Konglong Shen 1, An Li 1, Gang Chen 1, Xiaoyu Li 1, Sen Bai 1, Weifeng Wang 2, Renming Zhong 1,
PMCID: PMC7299271  PMID: 32100830

ABSTRACT

The purpose of the study was to quantify local setup errors and evaluate the planning target volume (PTV) margins for sub-regions in cone-beam computed tomography (CBCT)-guided post-mastectomy radiation therapy (PMRT). The local setup errors of 20 patients undergoing CBCT-guided PMRT were analysed retrospectively. Image registration between CBCT and planning CT was performed using four sub-regions of interest (ROIs): the supraclavicular area (SROI), ipsilateral chest wall region (CROI), ipsilateral chest wall plus supraclavicular region (SROI + CROI) and vertebral region (TROI). Bland–Altman analysis, correlation, local setup errors and PTV margins among these ROIs were evaluated. There was no significant consistency or correlation for registration results between the TROI and the CROI or SROI regions on any translational axis. When using the SROI + CROI as the ROI, the systematic error (Σ) and random error (σ) of the local setup errors for the CROI region were 1.81, 1.19 and 1.76 mm and 1.84, 2.64 and 3.00 mm along the medial–lateral (ML), superior–inferior (SI) and anterior–posterior (AP) directions, respectively. The PTV margins for the CROI region were 5.80, 4.82 and 6.50 mm. The Σ and σ of the local setup errors for the SROI region were 1.29, 1.15 and 0.77 mm and 1.96, 2.65 and 2.2 mm, respectively, and the PTV margins were 4.59, 4.73 and 3.47 mm. Large setup errors and local setup errors occur in PMRT. The vertebral body should not be a position surrogate for the supraclavicular region or chest wall. To compensate for the local setup errors, different PTV margins are required, even with CBCT guidance.

Keywords: local setup errors, CBCT guidance, post-mastectomy radiation therapy

INTRODUCTION

Post-mastectomy radiation therapy (PMRT) is the main treatment for breast cancer patients. With precise radiotherapy, the local control rate and the overall survival rate of patients can be improved [1–3]. Radiotherapy techniques such as intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy can increase the dose of the target area while protecting the surrounding normal tissues [4]. To verify the accuracy of positioning, electronic portal images (EPIs), cone beam computed tomography (CBCT) and optical surface imaging are commonly used. In total breast radiotherapy, Michalski et al. [5] reported setup errors ranging from 2.6 to 22.9 mm by EPIs. Batumalai et al. [6] reviewed different studies that depended on different registration methods and found that the ranges of the systematic errors were 1.2–5.7, 1.3–3.8 and 0.5–5.7 mm, and the random errors were 1.0–7.3, 1.2–4.1 and 0.9–4.0 mm in the medial–lateral (ML), superior–inferior (SI) and anterior–posterior (AP) direction, respectively. Sonmez et al. [7] concluded that setup errors in breast cancer radiotherapy should be strictly controlled below 5 mm. These data show that there are large differences in setup errors in breast cancer radiotherapy.

The treatment area for PMRT also includes the chest wall and supraclavicular, axillary and internal mammary lymphoid areas, and the related organs at risk are the cervical vertebrae, lungs, heart, and so forth. The relative positions among these regions may vary, resulting in local deformation errors for each region. In clinical practice, the supraclavicular region plus the chest wall (SROI + CROI) are deemed as a single region of interest (ROI) for registration in CBCT-guided radiotherapy. The registration results would be averaged for these two sub-regions (SROI and CROI region) using SROI + CROI as an ROI, resulting in an underestimation of local setup errors and dose uncertainty in the region of the supraclavicular area (SROI) or chest wall (CROI). Meanwhile, the thoracic vertebrae (TROI) are far from the SROI + CROI region, and the difference between TROI and other ROIs should be evaluated. Therefore, to ensure position accuracy and set the appropriate clinic target volume to planning target volume (CTV–PTV) margin, the local setup errors should be quantified for these sub-regions using different ROIs in CBCT-guided radiotherapy for PMRT.

MATERIALS AND METHODS

Patients

Twenty post-mastectomy breast cancer patients receiving radiotherapy from March 2015 to March 2017 in our department were randomly enrolled. Written consent was obtained from all patients prior to treatment in the study. For 10 patients the left supraclavicular region and chest wall were treated. For the other 10 patients the right supraclavicular region and chest wall were treated. The median age of the patients was 46 years (range, 34–61 years).

Immobilization

The CIVCO Breast Board Cushion (vacuum bag, size 90 × 55 cm, volume 15 L; CIVCO Radiotherapy, Orange City, IA, USA) was laid on a 15-degree wedge plate, and the upper boundary of the cushion was formed to fit the shape of the patient’s head and to limit the longitudinal movement of the head. The abduction of the upper limb on the involved side was greater than 90°. The vacuum pads on both sides of the arm ensured that the arms were secured as much as possible to prevent them from moving and pulling the skin of the adjacent chest, which would result in deformation of the skin mark (Fig. 1).

Fig. 1.

Fig. 1.

The patient was immobilized with a CIVCO Breast Board Cushion (vacuum bag, size 90 × 55 cm, volume 15 L).

Simulation and planning

Free-breathing patients were simulated using Philips Gemini 64-slice CT with a 3-mm slice thickness from the chin to the lower edge of the liver with intravascular contrast. According to the RTOG Breast Cancer Atlas for Radiation Therapy Planning: Consensus Definitions[8], the target (chest wall, supraclavicular lymph nodes) and organs at risk (i.e., heart, ipsilateral lung, contralateral breast and the spinal cord) were delineated. The treatment plan was generated using the RayStation 4.7 (RaySearch, Stockholm, Sweden) treatment-planning system.

CBCT scan

The CBCT images were obtained by XVI 4.5 (Synergy, Elekta, Crawley, UK). The number of frames per CBCT acquisition was ~400. The scanning parameters were tube voltage 100 kV, tube current 36.1 mA, S20 F1 filter plate, acquisition speed 5.5 frames/s and acquisition angle 50–210°.

Local setup analysis

The local setup errors were analysed offline using four sub-ROIs (see Fig. 2 for details). These four sub-regions included (i) the supraclavicular region (SROI), from the cricoid to the lower edge of the collarbone head in the longitudinal direction and from the cervical vertebra to the humeral head in the lateral direction; (ii) the ipsilateral chest wall region (CROI), from the humeral head to the lower edge of the PTV in the longitudinal direction and from the midline of the body to the outer 1 cm of the chest wall in the lateral direction; (iii) the ipsilateral chest wall plus supraclavicular region (SROI + CROI), the combination of SROI and the CROI region; and (iv) the vertebral region (TROI), from ~C5–T8 in the longitudinal direction, from the anterior of the vertebral body to the posterior boundary of the spinous in the vertical direction. The image registration protocols of our institution are as follows. The automatic registration “gray value (T + R)” (3 translational directions and 3 rotations directions) was used to obtain the coarse registration first, and then manual fine adjusting followed. Regarding SROI, we focused on the registration of the clavicle and cervical vertebrae. We mainly considered rib registration for CROI. Automatic registration was mainly used for SROI + CROI, unless there was a significant registration error. The registration of thoracic vertebrae was used for TROI. The systematic errors (Σ) and random errors (σ) were computed along the ML, SI and AP axes, and the rotation errors along the pitch, roll and yaw axes were also recorded. To quantify local setup errors, the region SROI + CROI was set as the reference ROI. The differences in registration results between SROI + CROI and SROI or CROI appeared to be local setup errors. Meanwhile, the differences in registration results between TROI and other ROIs were evaluated.

Fig. 2.

Fig. 2.

Four sub-regions of interest (ROI) for the local setup error analysis in the same patient. Rows represent the ipsilateral chest wall region (CROI), the supraclavicular region (SROI), the ipsilateral chest wall plus supraclavicular region (SROI + CROI) and the vertebral region (TROI). Columns represent coronal planes, sagittal planes and transverse planes.

Statistical analysis

One-way analysis of variance with Games-Howell post hoc tests was used to compare the setup errors among the four sub-regions (SROI, CROI, SROI + CROI, and TROI). Pearson’s correlation coefficient was used to analyse the correlation of the local setup errors in the four sub-regions. SPSS 19.0 was used to perform the above analysis. Bland–Altman analysis (using Python 3.7.4) was used to assess the ranges of agreement between the registration results of these ROIs.

CTV–PTV margin

According to ICRU Report 50 and Report 62 [9, 10], to reduce placement error and the impact of patient and organ movement on the target area, a ring should be placed outside the CTV. To ensure the CTVs obtain at least 95% of the prescribed dose for 90% of patients, this study used the following simplified margin calculation by Van Herk et al. [11]: 2.5 Σ + 0.7, where Σ is the standard deviation (SD) of the individual means for each ROI registration result per patient, and σ is the root mean square of the individual SD for each ROI registration result per patient.

RESULTS

Table 1 shows that in the AP direction, the difference in registration results between the SROI and CROI, CROI and TROI, SROI + CROI, and TROI regions was statistically significant. In the roll direction, the difference in registration results between the SROI and TROI regions was statistically significant. In the yaw direction, the difference between the SROI and CROI, SROI and SROI + CROI, SROI and TROI, and CROI and SROI + CROI registration results was statistically significant.

Table 1.

Multiple comparisons between groups [Games-Howell (A)*, P-value]

SROI CROI SROI + CROI TROI SROI CROI SROI + CROI TROI
ML Pitch
SROI 0.96 0.98 0.26 SROI 1.00 0.67 0.14
CROI 0.96 0.82 0.57 CROI 1.00 0.74 0.15
SROI + CROI 0.98 0.82 0.13 SROI + CROI 0.67 0.74 0.59
TROI 0.26 0.57 0.13 TROI 0.14 0.15 0.59
SI Roll
SROI 1.00 0.98 0.51 SROI 0.11 0.44 0.01
CROI 1.00 0.99 0.59 CROI 0.11 0.80 0.70
SROI + CROI 0.98 0.99 0.73 SROI + CROI 0.44 0.80 0.18
TROI 0.51 0.59 0.73 TROI 0.01 0.70 0.18
AP Yaw
SROI 0.01 0.51 0.54 SROI 0.00 0.00 0.00
CROI 0.01 0.18 0.00 CROI 0.00 0.00 0.23
SROI + CROI 0.51 0.18 0.05 SROI + CROI 0.00 0.00 0.23
TROI 0.54 0.00 0.05 TROI 0.00 0.23 0.23

*Multiple comparisons between the registration results of four ROIs were done by using the Games-Howell procedure. A P value of 0.05 was considered statistically significant.

The correlation analysis of the registration results among different ROIs showed that the SROI, CROI, and SROI + CROI regions had a very strong correlation in all translational directions. There was a strong correlation between the registration results of SROI + CROI and CROI in the pitch direction, SROI + CROI and SROI in the roll direction, and SROI + CROI and SROI in the yaw direction. There was no significant correlation between the registration results of TROI and any other ROIs (maximum correlation coefficient r = −0.30, P < 0.01). See Table 2 for details.

Table 2.

Correlation of setup errors in difference region

SROI CROI SROI + CROI TROI SROI CROI SROI + CROI TROI
ML Pitch
SROI 1.00 0.80** 0.89** −0.06 SROI 1.00 0.28** 0.50** −0.02
CROI 0.80** 1.00 0.87** −0.02 CROI 0.28** 1.00 0.62** −0.13**
SROI + CROI 0.89** 0.87** 1.00 −0.04 SROI + CROI 0.50** 0.62** 1.00 −0.17**
TROI −0.06 −0.02 −0.04 1.00 TROI −0.02 −0.13** −0.17** 1.00
SI Roll
SROI 1.00 0.85** 0.87** −0.06 SROI 1.00 0.20** .61** −0.07
CROI 0.85** 1.00 0.89** −0.06 CROI 0.20** 1.00 .39** 0.02
SROI + CROI 0.87** 0.89** 1.00 −0.10** SROI + CROI 0.61** 0.39** 1.00 −0.04
TROI −0.06 −0.06 −0.10** 1.00 TROI −0.07 0.02 −0.04 1.00
AP Yaw
SROI 1.00 0.86** 0.92** −0.25** SROI 1.00 0.27** .65** −0.06
CROI 0.86** 1.00 0.084** −0.25** CROI 0.27** 1.00 .58** −0.08
SROI + CROI 0.92** 0.84** 1.00 −0.30** SROI + CROI 0.65** 0.58** 1.00 −0.08
TROI −0.25** −.25** −0.30** 1.00 TROI −0.06 −0.08 −0.08 1.00

**The correlation was statistically significant, P = 0.01, two tails.

The registration results of ROIs, the difference among these ROIs, and the corresponding PTV margins are shown in Table 3. The difference in registration results between TROI and other ROIs was considerable. The systematic and random errors ranged from 4.19 to 8.90 mm, and the corresponding PTV margins ranged from 14.59 to 22.13 mm. The local setup errors of the SROI (the difference in registration results between SROI + CROI and SROI) and CROI (the difference in registration results between SROI + CROI and CROI) were smaller than the registration results between TROI and other ROIs. The Σ and σ of the local setup error for CROI were 1.81, 1.19 and 1.76 mm and 1.84, 2.64 and 3.00 mm, respectively, and the corresponding PTV margins were 5.80, 4.82 and 6.50 mm. The Σ and σ of the local setup error for SROI were 1.29, 1.15 and 0.77 mm and 1.96, 2.65 and 2.2 mm, respectively, and the corresponding PTV margins were 4.59, 4.73 and 3.47 mm.

Table 3.

The registration results in different ROIs and the difference of these errors among ROIs

ROIs Local setup errors Directions ROIs Local setup errors Directions
ML SI AP ML SI AP
SROI Σ 3.30 3.10 4.34 CROI-TROI Σ 4.59 4.82 6.36
σ 3.51 4.52 4.30 σ 5.86 7.53 8.90
Margin (mm) 10.71 10.91 13.87 Margin (mm) 15.59 17.31 22.13
CROI Σ 3.81 3.61 4.40 (SROI + CROI)-TROI Σ 4.19 4.50 6.07
σ 3.53 4.78 4.64 σ 5.90 7.63 8.85
Margin (mm) 12.00 12.37 14.25 Margin (mm) 14.59 16.60 21.38
SROI + CROI Σ 3.27 3.21 4.01 (SROI + CROI)-SROI Σ 1.29 1.15 0.77
σ 3.52 4.67 4.26 σ 1.96 2.65 2.20
Margin (mm) 10.64 11.29 13.00 Margin (mm) 4.59 4.73 3.47
TROI Σ 1.65 2.24 2.42 (SROI + CROI)-CROI Σ 1.81 1.19 1.76
σ 4.98 5.85 7.04 σ 1.84 2.64 3.00
Margin (mm) 7.62 9.70 10.97 Margin (mm) 5.80 4.82 6.50
SROI-TROI Σ 4.24 4.34 6.29
σ 6.01 7.39 8.65
Margin (mm) 14.81 16.02 21.76

The Bland–Altman analysis revealed good agreement in registration results between CROI and SROI + CROI and between SROI and SROI + CROI. However, there was no significant consistency in registration results between the TROI region and the CROI or SROI region on any translational axis. Figure 3 shows the Bland–Altman analysis between CROI and SROI + CROI and between SROI and SROI + CROI in the translational directions.

Fig. 3.

Fig. 3.

Bland–Altman error analysis for the SROI and CROI vs SROI + CROI registration results in translational directions. (a), (b) and (c) SROI vs SROI + CROI registration results in the ML, SI and AP directions, respectively. (d), (e) and (f) CROI vs SROI + CROI registration results in the ML, SI and AP directions, respectively. The abscissa indicates the mean registration results and the ordinate indicates the difference between registration results. The upper and lower dashed black lines are the 95% confidence interval. The middle dashed black line is the mean of the difference. The number of points exceeding the 95% confidence interval was <24 (5% of the total registration number 473), which indicated that the SROI and SROI + CROI, CROI, and SROI + CROI have strong registration consistency.

DISCUSSION

Setup errors in breast cancer radiotherapy are larger than the rigid structures, and the present study found that the setup errors of the SROI + CROI region ranged from 3.21 to 4.67 mm. The corresponding CTV–PTV margin was >10 mm, similar to the study by Popja et al. [4]. Strydhorst et al. found systematic and random errors in the ML, SI and AP directions of the chest wall of 2.7, 9.8 and 4.1 mm and 4.0, 12.0 and 4.5 mm [12]. Koseoglu et al. reported that the setup errors ranged from 4 to 13 mm [13]. Meanwhile, Zhang et al. found that the ipsilateral SROI + CROI region had larger translation errors than the CROI region alone in the AP direction [14]. The setup errors of breast radiotherapy were related to factors such as immobilization device, patient posture, radiotherapist experience and patient body mass index [15–19]. Therefore, it is necessary to use various methods to improve treatment accuracy, such as using X-ray image guidance, modifying the immobilization device, adding additional skin markers [6, 18, 20] and using optical surface imaging guidance to set up and correct posture errors in real time [21–23].

In some centres, the thoracic vertebral bone is considered as the position surrogate in thoracic cancer radiotherapy. To verify whether the thoracic vertebral bone can be a position surrogate in PMRT, we analysed the correlation and difference of the setup errors between the TROI and the CROI, SROI and SROI + CROI regions. The results showed that the thoracic vertebral bone and other ROIs have no significant correlation but do have considerable differences (the maximum random error was 8.9 mm, and the corresponding PTV margin was >20 mm; see Table 3). Meanwhile, there was significantly less consistency in the registration results between TROI and CROI or SROI on any translational axis. All of these results indicate that the thoracic vertebral bone should not be used as a position surrogate in breast cancer radiotherapy.

Local setup errors in sub-regions are an important source of error [24] and can be evaluated through the ROIs of sub-regions in CBCT image guidance, but they are hard to correct online. In routine CBCT-guided practice, image registration is performed using SROI + CROI as a whole ROI. In our study, the Bland–Altman analysis revealed good agreement in registration results between CROI and SROI + CROI and between SROI and SROI + CROI. This means that using SROI + CROI as a whole ROI is reasonable and acceptable in clinical practice. However, when using SROI + CROI as a single ROI, the local setup errors of the CROI region ranged from 1.19 to 3.00 mm, and the corresponding PTV margins were 5.80, 4.82 and 6.50 mm in the ML, SI and AP directions, respectively. The local setup errors of the SROI region ranged from 0.77 to 2.65 mm, with 4.59, 4.73 and 3.47 mm PTV margin in the ML, SI and AP directions, respectively. This indicated that a PTV margin of 5 mm can cover the local setup errors of the SROI region. Meanwhile, to compensate for the local setup errors of the CROI region, a 7-mm PTV margin is required, even with CBCT guidance. However, considering the buildup effect, the thickness of the chest wall and the dose limitations of the lung and heart, a smaller PTV margin is preferred for PMRT. In other words, the larger PTV margin would increase the dose coverage of the target but compromise the dose limitation of organs at risk. Meanwhile, the dose coverage of the target using a complete IMRT plan is more sensitive to setup errors than using a field-in-field plan [5]. Therefore, more caution should be used when using an IMRT plan if there are larger setup errors. Furthermore, a more effective immobilization method for PRMT is urgently required.

It is worth noting that the margin calculation by van Herk [11] was used only for translational errors. In the larger target region such as the SROI + CROI region, even small rotational errors may result in dose uncertainty [25]. In clinical practice, it is difficult to correct the rotation errors and local setup errors using CBCT image guidance. Fortunately, the optical surface guiding system can be used to adjust the posture of the patient and reduce rotational and local setup errors. Meanwhile, to cover the intrafractional error (mainly considering the patient’s respiratory motion for PMRT), we should consider the internal margin (IM) in the PTV margin. In our institution, we placed six metal markers on the patient’s chest wall and measured the respiratory motion using fluoroscopy for another 20 patients (unpublished results). We found that the average motion was ~1 mm in all translational directions (ranging from 0 to 1 mm in the ML direction, 0 to 2 mm in the SI direction and 0 to 4 mm in the AP direction). The results were consistent with the report by Harris et al. [26]. Furthermore, in the CBCT image, we did not find image blur in the chest wall caused by respiratory motion. Michalski et al. [5] reported that interfraction motion has a larger effect on dose distributions than intrafraction motion. According to these findings, we believe that we can ignore the IM or consider only a 1 mm IM for PMRT.

There are some limitations in this study. First, this was the first single-institution study of local setup errors in PMRT patients immobilized with a CIVCO Breast Board Cushion, so it is difficult to tell whether these results are large or small. Second, the size of the CIVCO Breast Board Cushion is too small to adequately indicate the position of the patient’s head, arms and body, which may reduce the reproducibility of the patient’s position. Third, the low-dose CBCT scan mode (small field of view, low kV and low milliampere seconds (mAS)) was used, resulting in poor quality of the CBCT and a small image field. Therefore, we cannot use these CBCT images to recalculate the actual dose distribution and analyse the position of the heart. In the future, more effective immobilization, the optical surface guiding system and actual dose distribution should be studied.

This study confirmed that there are large setup errors and local setup errors in PMRT patients immobilized with the CIVCO vacuum bag. Thus, a more effective immobilization device and image guidance are urgently required. The vertebral body should not be the position surrogate for the supraclavicular region or chest wall. To compensate for the local setup errors of the chest wall or supraclavicular area, different PTV margins are required, even with CBCT guidance.

ACKNOWLEDGMENTS

The authors wish to thank Bin Li and Wan Li for their assistance with data collection.

CONFLICT OF INTEREST

None declared.

FUNDING

This work was supported by the Science and Technology Support Program of Sichuan province, China (Grant number 2016FZ0086).

AUTHOR CONTRIBUTIONS

J.Z. and S.L. collected data and drafted the manuscript. C.Y., K.S., A.L., G.C., X. L. and W.W. helped to collect the data. S.B. helped to design the study. R.Z. designed the study, and revised and approved the final manuscript. All authors read and approved the manuscript.

REFERENCES

  • 1. Alluri P, Jagsi R. Postmastectomy Radiotherapy In: Bland KI, Copeland EM, Klimberg et al. (eds). The Breast: Comprehensive Management of Benign and Malignant Diseases, 5th edn. Amsterdam: Elsevier, 2018, 688–92.e2. [Google Scholar]
  • 2. McGale P, Taylor C, Correa C et al. Effect of radiotherapy after mastectomy and axillary surgery on 10-year recurrence and 20-year breast cancer mortality: Meta-analysis of individual patient data for 8135 women in 22 randomised trials. Lancet 2014;383:2127–35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Wang J, Shi M, Ling R et al. Adjuvant chemotherapy and radiotherapy in triple-negative breast carcinoma: A prospective randomized controlled multi-center trial. Radiother Oncol 2011;100:200–4. [DOI] [PubMed] [Google Scholar]
  • 4. Jain P, Marchant T, Green M et al. Inter-fraction motion and dosimetric consequences during breast intensity-modulated radiotherapy (IMRT). Radiother Oncol 2009;90:93–8. [DOI] [PubMed] [Google Scholar]
  • 5. Michalski A, Atyeo J, Cox J et al. Inter- and intra-fraction motion during radiation therapy to the whole breast in the supine position: A systematic review. J Med Imaging Radiat Oncol 2012;56:499–509. [DOI] [PubMed] [Google Scholar]
  • 6. Batumalai V, Holloway L, Delaney GP. A review of setup error in supine breast radiotherapy using cone -beam computed tomography. Med Dosim 2016;41:225–9. [DOI] [PubMed] [Google Scholar]
  • 7. Sonmez A, Onal C, Sonmez S et al. Effects of setup errors on dose distribution for tangential wedge field and field-in-field techniques during breast irradiation. Uhod-Uluslar Hematol 2014;24:130–8. [Google Scholar]
  • 8. White J, Tai A, Arthur D. Breast Cancer Atlas for Radiation Therapy Planning. http://www.rtog.org/CoreLab/ContouringAtlases/BreastCancerAtlas.aspx (Jul 27 2015, date last accessed). [Google Scholar]
  • 9. ICRU ICRU report 50: Prescribing, recording and reporting photon beam therapy. J ICRU 1993;os26:15–8. [Google Scholar]
  • 10. ICRU ICRU report 62: Prescribing, recording and reporting photon beam therapy (supplement to ICRU report 50). J ICRU 1999;os32:1–52. [Google Scholar]
  • 11. Herk M, Remeijer P, Rasch C et al. The probability of correct target dosage: Dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys 2000;47:1121–35. [DOI] [PubMed] [Google Scholar]
  • 12. Strydhorst J, Caudrelier J, Clark B et al. Evaluation of a thermoplastic immobilization system for breast and chest wall radiation therapy. Med Dosim 2011;36:81–4. [DOI] [PubMed] [Google Scholar]
  • 13. Koseoglu FG, Tuncel N, Kizildag AU et al. Assessment of setup accuracy in patients receiving postmastectomy radiotherapy using electronic portal imaging. Radiat Med 2007;25:45–52. [DOI] [PubMed] [Google Scholar]
  • 14. Zhang H-J, Zhang C, Ge R-G et al. Evaluation of positional error during radiotherapy for breast cancer after modified radical mastectomy. Biomed Res-india 2017;28:7526–33. [Google Scholar]
  • 15. Lee P-Y, Lin C-Y, Chen S-W et al. A topology-based method to mitigate the dosimetric uncertainty caused by the positional variation of the boost volume in breast conservative radiotherapy. Radiat Oncol 2017;12:55. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Raza W, Agarwal S, Das KJM et al. Comparison of set-up errors by breast size on wing board by portal imaging. Rep Pract Oncol Radiother 2016;21:447–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Mulliez T, Gulyban A, Vercauteren T et al. Setup accuracy for prone and supine whole breast irradiation. Strahlenther Onkol 2016;192:254–9. [DOI] [PubMed] [Google Scholar]
  • 18. Harris EJ, Mukesh MB, Donovan EM et al. A multicentre study of the evidence for customized margins in photon breast boost radiotherapy. Br J Radiol 2016;89:20150603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Chung MJ, Lee GJ, Suh YJ et al. Setup error and effectiveness of weekly image-guided radiation therapy of TomoDirect for early breast cancer. Cancer Res Treat 2015;47:774–80. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Kawamura M, Maeda Y, Yamamoto K et al. Development of the breast immobilization system in prone setup: The effect of bra in prone position to improve the breast setup error. J Appl Clin Med Phys 2017;18:155–60. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Ma Z, Zhang W, Su Y et al. Optical surface management system for patient positioning in Interfractional breast cancer radiotherapy. Biomed Res Int 2018;6415497. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Sa AC, Fermento A, Neves D et al. Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience. Rep Pract Oncol Radiother 2018;23:61–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Stanley DN, McConnell KA, Kirby N et al. Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis. J Appl Clin Med Phys 2017;18:58–61. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Zhong R-M, Song Y, Yan Y-Y et al. Analysis of which local set-up errors can be covered by a 5-mm margin for cone beam CT-guided radiotherapy for nasopharyngeal carcinoma. Br J Radiol 2018;91. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Harron EC, McCallum HM, Lambert EL et al. Dosimetric effects of setup uncertainties on breast treatment delivery. Med Dosim 2008;33:293–8. [DOI] [PubMed] [Google Scholar]
  • 26. Harris EJ, Donovan EM, Coles CE et al. How does imaging frequency and soft tissue motion affect the PTV margin size in partial breast and boost radiotherapy? Radiother Oncol 2012;103:166–71. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Radiation Research are provided here courtesy of Oxford University Press

RESOURCES