Abstract
Background
Metallic microwave ablation (MWA) antenna-related artifacts are usually created in conventional CT images, and these artifacts can influence the effect of ablation. The aim of this study was to evaluate a new type of metal artifact reduction (MAR+) technique in CT-guided MWA for lung cancer.
Material/Methods
This retrospective study enrolled 30 lung cancer patients who received CT-guided MWA treatment from December 2017 to April 2018. Images after microwave antenna insertion into the tumor were reconstructed by the filter back projection (group A) and MAR+ reconstruction (group B). The CT values and standard deviations of the regions of interest (ROIs) on the chosen image were recorded, including the most significantly hypodense artifact (ROI1), hyperdense artifacts (ROI2), and chest muscles of the same layer (ROI3). The metal artifact indexes based on ROI1 and ROI2 (AI1, AI2) and the overall metal artifact index (AI) were calculated. Subjective image quality was graded on a five-point scale (1=worst, 5=excellent).
Results
The AI1 (74.14±76.32), AI2 (13.75±19.02) and AI (54.12±54.82) of group B were lower than those of group A [(153.33±89.04), (30.63±26.42), (112.00±63.10), respectively] (P<0.001 for all). Both radiologists reported that the subjective image value of group B was significantly higher than that of group A (P<0.001). The subjective image quality scores evaluated by 2 observers showed excellent consistency (ICC=0.829).
Conclusions
The MAR+ imaging reconstruction significantly reduced metal artifacts, which helps radiologists to clearly observe the relationship between the ablation antenna and the lesion.
MeSH Keywords: Artifacts; Lung Neoplasms; Microwaves; Tomography, X-Ray Computed
Background
Lung cancer is the most commonly diagnosed cancer (11.6% of all cancers) and the most common cause of cancer death (18.4% of all cancer deaths) worldwide [1]. In recent years, percutaneous microwave ablation (MWA), as a novel local treatment modality, has been widely used for lung cancer therapy [2–4]. MWA uses electromagnetic waves in the microwave energy spectrum to generate tissue-heating effects. The current MWA manufacturer uses a frequency of 2450 MHz. The polar molecules such as water molecules in tumor tissue generate extremely high-speed vibrations under the action of this microwave electromagnetic field, causing collisions and friction between molecules, and the thermal energy is then transferred to the adjacent tissues. When the temperature reaches 60~150°C in a short time, it causes irreversible damage to cells or coagulative necrosis [5,6]. Moreover, a single MWA probe covered a larger ablation volume than radiofrequency ablation (RFA) [7]. Computed tomography (CT) is currently the most accurate image-guidance technique for lung ablation.
The efficiency of MWA depends on the appropriate and accurate location of the microwave ablation antenna in the tumor. However, metal instruments used in interventional procedures can create significant artifacts in CT images [8]. The severity of metal artifacts depends on the material of the instruments used. MWA antenna-related artifacts are usually created in conventional CT images. Severe artifacts can cover the surrounding tissues and influence the exact location of the ablation antenna [9,10]. Therefore, reducing artifacts is very important for MWA.
To reduce these artifacts, the optimal acquisition and reconstruction method must be used for patients with various types of metal prostheses and implants. However, metal artifact reduction (MAR) techniques are rarely applied in interventional therapy. The aim of this study was to evaluate a new type of metal artifact reduction (MAR+) technique in CT-guided MWA for lung cancer.
Material and Methods
This retrospective study was approved by the Institutional Review Board of our hospital. The requirement for informed consent was waived for all subjects. We retrospectively reviewed patients with lung cancer who underwent CT-guided MWA using the MAR+ technique between December 2017 and April 2018. A total of 30 patients were identified.
CT-guided MWA procedure
All ablation processes for lung cancer were performed with the guidance of a NeuViz 128 CT (Neusoft Medical Company, Shenyang, China). The scanning parameters for the location were 120 kVp tube voltage, automatically modulated tube current (reference mAs: 180), 0.6 second gantry rotation time, 1.2 pitch, 128×0.625 mm collimation width, 360 mm field of view (FOV), 5 mm slice thickness, and 5 mm slice intervals. The effective dose-length product (DLP, mGy×cm) was recorded, and the effective dose (ED, mSv) was calculated for each ablation series using the equation: ED=DLP×k [k=0.014 (mSv×mGy−1×cm−1)]
Patients underwent a CT scan of the area of interest during suspended inspiration. The entry site was prepared and draped in a sterile fashion. All ablations were performed under local anesthesia consisting of an injection of 5–10 mL of 1% lidocaine hydrochloride. All ablations were performed by a microwave ablation system (2450 MHz generator, MICRO TECH, KANG YOU, Nanjing, China) and a 16-gauge antenna (made of stainless steel as biomedical material). The antenna was inserted into the lesion during a breath-hold. Five-millimeters sections with image reconstruction were then obtained during the breath-hold to verify the location of the antenna (Figure 1A–1D). The ablation power and time were adjusted to the size and location of the tumor according to the manufacturer’s recommendations. All ablations were performed at a constant power of 50–60 W. The ablation time per tumor was 5–10 minutes. The procedure aimed to ablate the tumor plus at least a 5-mm margin.
Figure 1.
A 58-year-old man with the primary lung cancer who underwent MWA treatment. (A) An axial plain CT image before the ablation process showing an elliptical nodule of 3.4×2.4 cm in the basal right lower lobe in the lung window (red arrow). (B) The axial plain CT scan after filter back projection reconstruction presenting with poor quality, where the extensive metal artifacts impair the assessment of the antenna’s position within the nodule. (C) The axial plain CT scan after MAR+ reconstruction appearing with excellent quality where the metal artifacts are greatly reduced, allowing better visualization of the relationship of the antenna and lesion. (D) An axial plain CT scan after the biopsy procedure showing patchy areas of ground-glass opacity around the lesion (red arrows).
After the ablation procedure, CT images were immediately acquired at the level of the ablation site to search for the presence of complications, such as pneumothorax or parenchymal hemorrhage.
Image postprocessing
The raw data of all available images after microwave antenna insertion into the tumor site were reconstructed by the filter back projection algorithm (group A) and MAR+ reconstruction (group B). The reconstruction parameters of group A and B were: group A with Clearview closed, MAR+ closed; group B with Clearview 40%, MAR+ open. Both groups were reconstructed with a slice thickness of 5 mm and a slice interval of 5 mm. Actual MWA procedures were performed under the guidance of group B images. The images of groups A and B were transferred to the advanced workstation (AVW 1.0.8, Neusoft Medical Company, Shenyang, China) for further analysis.
MAR+ technique is a kind of normalized MAR, which involves a special method to reduce the effect of the microwave antenna. The MAR+ algorithm was created in the following way: according to the geometric parameters of the system, the characteristics of the electronic components and the detector, the anatomical model, the front projection model, and the noise model are refined. The correction process based on the iterative algorithm removes the noise caused by metal and restores the true anatomical structure around the metal object (Figure 2).
Figure 2.
The flow of the MAR+ technique.
Evaluation methods
The images that included the tip of the microwave antenna and the lesion with the most severe artifacts were chosen for analysis by 2 radiologists (with 5 and 8 years of experience in chest imaging) independently in the soft-tissue window (window width: 300 HU, window level: 40 HU). Any disagreements were resolved by discussion and consensus.
Subjective evaluation
Two radiologists, who were blinded to the imaging reconstruction method, graded on a five-point scale [11] in each patient based on an overall assessment of the severity of metal artifacts. The images were evaluated for subjective evaluation in the soft-tissue window setting (window width: 300 HU, window level: 40 HU). The specific standards were: 1-worst, severe metal-related artifacts that hardly reveals the surrounding lung cancer tissue; 2-bad, limited observation of the surrounding tissue; 3-medium, some artifact but acceptable visualization; 4-good, small enough artifact to clearly observe the relationship between the microwave antenna and the lesion; 5-excellent, almost no artifact.
Objective evaluation
The objective evaluation was performed by another radiologist with 3 years of experience in chest imaging in the soft-tissue window (window width: 300 HU, window level: 40 HU). The CT values in Hounsfield units (CT1, CT2, CT3) and standard deviations (SD1, SD2, SD3) of the elliptical regions of interest (ROIs) with the area of 30–40 mm2 were recorded, including the most significantly hypodense artifacts (ROI1) and hyperdense artifacts (ROI2) of the selected image, along with chest muscles (ROI3) of the same slice. The metal artifact indexes based on ROI1 and ROI2 (AI1, AI2) and the overall metal artifact index (AI) were calculated as shown below [12,13], respectively.
Statistical analysis
Statistical analysis was performed with SPSS (version 20.0, SPSS Inc., Chicago, USA). The discrete variables are expressed as median and interquartile range, and the continuous variables are expressed as mean±SD. A paired t test was used to compare the differences in AI1, AI2, and AI between the 2 groups. The subjective scores were compared using a Wilcoxon signed-rank test. The consistency between the 2 readers was evaluated by the intraclass correlation coefficient (ICC) test with the following scale: poor (<0.40), moderate (0.40–0.60), good (0.60–0.75), and excellent (>0.75) [14]. P<0.05 was considered statistically significant.
Results
General characteristics
The general patient characteristics are summarized in Table 1. All 30 lung cancer patients underwent MWA therapy for a single lesion. Twenty-one cases (70%) were primary pulmonary cancers (adenocarcinoma, n=16; squamous cell carcinoma, n=5), and 9 cases were metastatic cancers (esophageal squamous cell carcinoma, n=1; hepatocellular carcinoma, n=3; colorectal adenocarcinoma, n=5). There were 7 women (23.3%) and 23 men (76.7%) with a mean age of (60.3±11.6) years (range 38–77 years). The median tumor size was (3.5±0.7) cm (range 2.3–4.9 cm). One MWA antenna was used for each patient during the ablation procedure. The mean ED was (2.74±0.43) mSv (range 2.23–3.17 mSv) per scan. The reconstructions utilizing MAR+ algorithms were processed in an average time of (3.2±0.4) seconds.
Table 1.
Clinical characteristics of the patients.
Characteristics | Values |
---|---|
Age (years)* | 60.3±11.6 (38–77) |
Sex | |
Male | 23 (76.7%) |
Female | 7 (23.3%) |
Histopathological types | |
Primary lung cancer | 21 (70%) |
Adenocarcinoma | 15 (71.4%) |
Squamous cell carcinoma | 6 (28.6%) |
Metastatic lung cancer | 9 (30%) |
Esophageal squamous cell cancer | 1 (11.1%) |
Hepatocellular carcinoma | 5 (55.6%) |
Colorectal adenocarcinoma | 3 (33.3%) |
Tumor size (cm)* | 3.5±0.7 (2.3–4.9) |
Complications | |
Pneumothorax | 4 (13.3%) |
Pleural effusion | 2 (6.7%) |
Effective dose (mSv)* | 2.74±0.43 (2.23–3.17) |
Data are the mean±standard deviation, and data in parentheses are the ranges or percentages.
Objective evaluation results
The objective evaluation results are summarized in Table 2. The AI1, AI2, and AI of group B (74.14±76.32, 13.75±19.02, 54.12±54.82) were significantly lower than those of group A (153.33±89.04, 30.63±26.42, 112.00±63.10, all P<0.001).
Table 2.
The values of AI1, AI2 and AI were significantly different between groups A and B.
Group A | Group B | P values | |
---|---|---|---|
AI1 | 153.33±89.04 | 74.14±76.32 | <0.001 |
AI2 | 30.63±26.42 | 13.75±19.02 | <0.001 |
AI | 112.00±63.10 | 54.12±54.82 | <0.001 |
Data are presented as the mean±standard deviation, and P<0.05 indicates statistical significance.
Subjective evaluation results
Radiologist 1 evaluated that the subjective image value of group B was higher than that of group A, and the difference was statistically significant [4 (4, 5) vs. 1 (1, 2), P<0.001], which was consistent with the subjective assessment results of Radiologist 2. The subjective image quality scores evaluated by 2 observers showed excellent consistency (ICC=0.829) (Table 3, Figure 1A–1D).
Table 3.
Subjective image quality assessment between groups A and B by 2 radiologists.
Radiologists | Patients | Score 1 | Score 2 | Score 3 | Score 4 | Score 5 | Mean score* | P value |
---|---|---|---|---|---|---|---|---|
Radiologist 1 | ||||||||
Group 1 | 30 | 20 (66.7%) | 10 (33.3%) | 0 (0%) | 0 (0%) | 0 (0%) | 1 (1, 2) | P<0.001 |
Group 2 | 30 | 0 (0%) | 0 (0%) | 5 (16.7%) | 15 (50%) | 10 (33.3%) | 4 (4, 5) | |
Radiologist 2 | ||||||||
Group 1 | 30 | 20 (66.7%) | 10 (33.3%) | 0 (0%) | 0 (0%) | 0 (0%) | 1 (1, 2) | P<0.001 |
Group 2 | 30 | 0 (0%) | 0 (0%) | 4 (13.3%) | 18 (60%) | 8 (26.7%) | 4 (4, 5) |
Data are the median and interquartile range and data in parentheses are the percentages.
MWA-related results
The technical success rate of MWA treatment for the 30 patients was 100%. The main complications were pneumothorax in 4 patients (13.3%) and pleural effusion in 2 patients (6.7%). No deaths were observed during the ablation process.
Follow-up contrast-enhanced chest CT examinations were used to detect local recurrence. No patients had local recurrence at the ablation site during the short follow-up period (3–6 months).
Discussion
Surgical lobectomy is viewed as the criterion standard treatment for lung cancer [15–17]. However, patients with high-risk conditions, such as cardiopulmonary dysfunction, impaired renal function, and hypertension, or who are elderly, are not good candidates for surgery. Currently, MWA has been widely used in patients with medically inoperable tumors [18]. Several studies have suggested that MWA is safe and effective in the treatment of primary pulmonary malignances and pulmonary metastases [6,16,19,20]. The presence of metallic objects in the imaging field may deteriorate the image quality of CT scans and thus impact treatment [21]. The significance of reducing metallic artifacts is to allow the operator to clearly observe the location of the lesion and the ablation antenna, thus improving the image quality and providing more possibilities for accurate MWA procedures.
The causes of metal artifacts are complex and include beam hardening, photon starvation, and increased scatter and noise [22–24]. To reduce these artifacts, the optimal acquisition and reconstruction method must be used for patients with various types of metal prostheses and implants. Different CT venders have their own specific MAR algorithms. For example, gemstone spectral (GSI) dual-energy CT (DECT) with or without metal artifact reduction software (MARs) from GE (GE Healthcare, USA) used projection-based reconstructions on 40–140 keV monochromatic images to reduce artifacts [24]. Iterative metal artifact reduction (IMAR) algorithms from Siemens (Siemens Healthcare, Germany) used an iterative frequency split technique to suppress metal-related artifacts [25]. Orthopedic metal artifact reduction algorithms (O-MAR) used the iterative reconstruction method developed by Philips (Philips Medical Systems, The Netherlands) [26,27]. Toshiba (Toshiba Medical Systems, Japan) developed the single-energy metal artifact reduction (SEMAR) procedure, which mainly removes the metal artifacts based on a linear interpolation in the blended sinogram (original sinogram and tissue-classified sinogram) [28]. NeuViz 128 CT is configured with the MAR+ technique to clearly remove the metal artifact in a reasonable time, to a certain extent, in case of the presence of complications. The procedure is carried out by creating a front projection model, an anatomy model, and a noise model according to the geometric parameters, electronic components, and detector characteristics, and then clearly removing the metal artifact based on the iterative correction method.
Metal artifact reduction techniques have been applied to evaluate the image visualization of patients with knee replacements [29], total or unilateral hip arthroplasty [30,31], dental implants [32], and spine metallic implants [33]. Few studies have explored how to decrease the artifacts produced by metallic ablation antennas during CT-guided MWA for patients with lung cancer. In our study, the results showed that MAR+ imaging reconstruction could reduce the MWA antenna-related metal artifacts subjectively and objectively. In images reconstructed with the MAR+ technique, the microwave ablation antenna was observed with a clearer delineation and the removal of metal artifacts at the needle edges. The removal of metal artifacts and the visualization of the soft-tissue region surrounding the antenna were critical, as they can enhance needle confirmation within the target lesion. Although switching to the bone window may partially reduce metal artifacts, the surrounding soft-tissue region is unclear. The MAR+ technique in the present study reduced metal artifacts in only approximately 3 seconds per scan, which is nearly a real-time correction of metal artifacts.
The limitations of this study include its retrospective nature and its fairly small sample size of 30 patients, which limits the generalization of the findings. This study mainly focused on image quality, and the follow-up time was relatively short. The survival and outcomes of the patients require further, long-term, follow-up.
Conclusions
In conclusion, the MAR+ imaging reconstruction algorithm from NeuViz 128 CT significantly reduced metal artifacts, which makes it possible for radiologists to clearly observe the relationship between the ablation antenna and the lesion and thus improves the efficiency and accuracy of MWA procedures.
Footnotes
Conflicts of interest
None.
Source of support: This study was funded by National Public Welfare Basic Scientific Research Program of Chinese Academy of Medical Sciences (grant numbers 2018PT32003 and 2017PT32004)
References
- 1.Bray F, Ferlay J, Soerjomataram I, et al. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. Cancer J Clin. 2018;68(6):394–424. doi: 10.3322/caac.21492. [DOI] [PubMed] [Google Scholar]
- 2.Yang X, Ye X, Lin Z, et al. Computed tomography-guided percutaneous microwave ablation for treatment of peripheral ground-glass opacity-Lung adenocarcinoma: A pilot study. J Cancer Res Ther. 2018;14(4):764–71. doi: 10.4103/jcrt.JCRT_269_18. [DOI] [PubMed] [Google Scholar]
- 3.Narsule CK, Sridhar P, Nair D, et al. Percutaneous thermal ablation for stage IA non-small cell lung cancer: long-term follow-up. J Thorac Dis. 2017;9(10):4039–45. doi: 10.21037/jtd.2017.08.142. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Yang X, Ye X, Zhang L, et al. Microwave ablation for lung cancer patients with a single lung: Clinical evaluation of 11 cases. Thorac Cancer. 2018;9(5):548–54. doi: 10.1111/1759-7714.12611. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Sidoff L, Dupuy DE. Clinical experiences with microwave thermal ablation of lung malignancies. Int J Hyperthermia. 2017;33(1):25–33. doi: 10.1080/02656736.2016.1204630. [DOI] [PubMed] [Google Scholar]
- 6.Vogl TJ, Nour-Eldin NA, Albrecht MH, et al. Thermal ablation of lung tumors: focus on microwave ablation. Rofo. 2017;189(9):828–43. doi: 10.1055/s-0043-109010. [DOI] [PubMed] [Google Scholar]
- 7.de Baere T, Tselikas L, Catena V, et al. Percutaneous thermal ablation of primary lung cancer. Diagn Interv Imaging. 2016;97(10):1019–24. doi: 10.1016/j.diii.2016.08.016. [DOI] [PubMed] [Google Scholar]
- 8.David A, Ray A, Hermet PL, et al. Reduction of metallic coil artifacts in CT angiography with metal artefact reduction (MAR) algorithm. Diagn Interv Imaging. 2019;100(6):381–82. doi: 10.1016/j.diii.2019.02.002. [DOI] [PubMed] [Google Scholar]
- 9.Peng C, Qiu B, Li M, et al. Gaussian diffusion sinogram inpainting for X-ray CT metal artifact reduction. Biomed Eng Online. 2017;16(1):1. doi: 10.1186/s12938-016-0292-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Wellenberg RHH, Hakvoort ET, Slump CH, et al. Metal artifact reduction techniques in musculoskeletal CT-imaging. Eur J Radiol. 2018;107:60–69. doi: 10.1016/j.ejrad.2018.08.010. [DOI] [PubMed] [Google Scholar]
- 11.Dong Y, Shi AJ, Wu JL, et al. Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT. Eur Spine J. 2016;25(6):1754–63. doi: 10.1007/s00586-015-4053-4. [DOI] [PubMed] [Google Scholar]
- 12.Hu Y, Pan S, Zhao X, et al. Value and clinical application of orthopedic metal artifact reduction algorithm in CT scans after orthopedic metal implantation. Korean J Radiol. 2017;18(3):526–35. doi: 10.3348/kjr.2017.18.3.526. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wang F, Zhang Y, Xue H, et al. Combined use of iterative reconstruction and monochromatic imaging in spinal fusion CT images. Acta Radiol. 2017;58(1):62–69. doi: 10.1177/0284185116631182. [DOI] [PubMed] [Google Scholar]
- 14.Sekir U, Yildiz Y, Hazneci B, et al. Reliability of a functional test battery evaluating functionality, proprioception, and strength in recreational athletes with functional ankle instability. Eur J Phys Rehabil Med. 2008;44(4):407–15. [PubMed] [Google Scholar]
- 15.Yao W, Lu M, Fan W, et al. Comparison between microwave ablation and lobectomy for stage I non-small cell lung cancer: A propensity score analysis. Int J Hyperthermia. 2018;34(8):1329–36. doi: 10.1080/02656736.2018.1434901. [DOI] [PubMed] [Google Scholar]
- 16.Zhong L, Sun S, Shi J, et al. Clinical analysis on 113 patients with lung cancer treated by percutaneous CT-guided microwave ablation. J Thorac Dis. 2017;9(3):590–97. doi: 10.21037/jtd.2017.03.14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Das M, Abdelmaksoud MH, Loo BW, Jr, Kothary N. Alternatives to surgery for early stage non-small cell lung cancer-ready for prime time? Curr Treat Options Oncol. 2010;11(1–2):24–35. doi: 10.1007/s11864-010-0119-z. [DOI] [PubMed] [Google Scholar]
- 18.Yang X, Ye X, Zheng A, et al. Percutaneous microwave ablation of stage I medically inoperable non-small cell lung cancer: Clinical evaluation of 47 cases. J Surg Oncol. 2014;110(6):758–63. doi: 10.1002/jso.23701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Kurilova I, Gonzalez-Aguirre A, Beets-Tan RG, et al. Microwave ablation in the management of colorectal cancer pulmonary metastases. Cardiovasc Intervent Radiol. 2018;41(10):1530–44. doi: 10.1007/s00270-018-2000-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ni X, Han JQ, Ye X, Wei ZG. Percutaneous CT-guided microwave ablation as maintenance after first-line treatment for patients with advanced NSCLC. Onco Targets Ther. 2015;8:3227–35. doi: 10.2147/OTT.S90528. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Aissa J, Boos J, Sawicki LM, et al. Iterative metal artefact reduction (MAR) in postsurgical chest CT: Comparison of three iMAR-algorithms. Br J Radiol. 2017;90(1079):20160778. doi: 10.1259/bjr.20160778. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Huang JY, Kerns JR, Nute JL, et al. An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol. 2015;60(3):1047–67. doi: 10.1088/0031-9155/60/3/1047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Andersson KM, Nowik P, Persliden J, et al. Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors. Br J Radiol. 2015;88(1052):20140473. doi: 10.1259/bjr.20140473. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Lee YH, Park KK, Song HT, et al. Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol. 2012;22(6):1331–40. doi: 10.1007/s00330-011-2370-5. [DOI] [PubMed] [Google Scholar]
- 25.Diehn FE, Michalak GJ, DeLone DR, et al. CT dental artifact: Comparison of an iterative metal artifact reduction technique with weighted filtered back-projection. Acta Radiol Open. 2017;6(11):2058460117743279. doi: 10.1177/2058460117743279. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Park J, Kim SH, Han JK. Combined application of virtual monoenergetic high keV images and the orthopedic metal artifact reduction algorithm (O-MAR): Effect on image quality. Abdom Radiol (NY) 2019;44(2):756–65. doi: 10.1007/s00261-018-1748-0. [DOI] [PubMed] [Google Scholar]
- 27.Sunwoo L, Park SW, Rhim JH, et al. Metal artifact reduction for orthopedic implants: Brain CT angiography in patients with intracranial metallic implants. J Korean Med Sci. 2018;33(21):e158. doi: 10.3346/jkms.2018.33.e158. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ragusi M, van der Meer RW, Joemai RMS, et al. Evaluation of CT angiography image quality acquired with single-energy metal artifact reduction (SEMAR) algorithm in patients after complex endovascular aortic repair. Cardiovasc Intervent Radiol. 2018;41(2):323–29. doi: 10.1007/s00270-017-1812-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Magarelli N, De Santis V, Marziali G, et al. Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses. Radiol Med. 2018;123(8):593–600. doi: 10.1007/s11547-018-0881-8. [DOI] [PubMed] [Google Scholar]
- 30.Laukamp KR, Lennartz S, Neuhaus VF, et al. CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur Radiol. 2018;28(11):4524–33. doi: 10.1007/s00330-018-5414-2. [DOI] [PubMed] [Google Scholar]
- 31.Yue D, Fan Rong C, Ning C, et al. Reduction of metal artifacts from unilateral hip arthroplasty on dual-energy CT with metal artifact reduction software. Acta Radiol. 2018;59(7):853–60. doi: 10.1177/0284185117731475. [DOI] [PubMed] [Google Scholar]
- 32.Grosse Hokamp N, Laukamp KR, Lennartz S, et al. Artifact reduction from dental implants using virtual monoenergetic reconstructions from novel spectral detector CT. Eur J Radiol. 2018;104:136–42. doi: 10.1016/j.ejrad.2018.04.018. [DOI] [PubMed] [Google Scholar]
- 33.Shen ZL, Xia P, Klahr P, Djemil T. Dosimetric impact of orthopedic metal artifact reduction (O-MAR) on Spine SBRT patients. J Appl Clin Med Phys. 2015;16(5):106–16. doi: 10.1120/jacmp.v16i5.5356. [DOI] [PMC free article] [PubMed] [Google Scholar]