Abstract

The strong absorption of liquid water in the infrared (IR) molecular fingerprint region constitutes a challenge for applications of vibrational spectroscopy in chemistry, biology, and medicine. While high-power IR laser sources enable the penetration of ever thicker aqueous samples, thereby mitigating the detrimental effects of strong attenuation on detection sensitivity, a basic advantage of heterodyne-measurement-based methods has—to the best of our knowledge—not been harnessed in broadband IR measurements to date. Here, employing field-resolved spectroscopy (FRS), we demonstrate in theory and experiment fundamental advantages of techniques whose signal-to-noise ratio (SNR) scales linearly with the electric field over those whose SNR scales linearly with radiation intensity, including conventional Fourier-transform infrared (FTIR) and direct absorption spectroscopy. Field-scaling brings about two major improvements. First, it squares the measurement dynamic range. Second, we show that the optimum interaction length with samples for SNR-maximized measurements is twice the value usually considered to be optimum for FTIR devices. In order to take full advantage of these properties, the measurement must not be significantly affected by technical noise, such as intensity fluctuations, which are common for high-power sources. Recently, it has been shown that subcycle, nonlinear gating of the molecular fingerprint signal renders FRS robust against intensity noise. Here, we quantitatively demonstrate this advantage of FRS for thick aqueous samples. We report sub-μg/mL detection sensitivities for transmission path lengths up to 80 μm and a limit of detection in the lower μg/mL range for transmission paths as long as 200 μm.
Molecular vibrational spectroscopies, such as Raman and infrared (IR) spectroscopies, are versatile tools for delivering chemical and molecular information on complex samples in a fast, reliable, and label-free manner.1 Every molecule exhibits a unique spectrum of vibrational eigenstates,2 leading to highly specific signals when a molecular ensemble (for instance biofluids or tissue) is subjected to spectroscopic interrogation. This “spectral fingerprint” contains a wealth of information about the molecular composition, structure, and conformation.1,2 This underlies numerous applications in biology and medicine, such as comprehending protein dynamics and folding,3 identifying cell phenotypes,4 quantifying blood-based clinical parameters,5 and detecting cancerous states in tissues6 and biofluids.7,8
Typically, the study of biological systems is preferably performed in their natural aqueous environment. While Raman spectroscopy is well suited for transmission measurements thereof, it suffers from weak signals.1 Broadband IR spectroscopy, on the other hand, profits from large interaction cross sections, potentially affording a unique combination of detection sensitivity and molecular coverage.1 However, the strong absorbance of (liquid) water in this spectral range has severely limited the applicability of IR vibrational spectroscopy (and microscopy) in transmission geometry so far. In fact, in table-top Fourier-transform IR (FTIR) spectrometry setups, the most commonly employed technology, the transmission path length needs to be limited to <10 μm due to the modest brightness of the source and the lack of high-sensitivity mid-infrared (MIR) detectors.3,9 This implies severe practical complications and limitations.
First, when studying living cells or (complex) biofluids, sophisticated cell holders and/or microfluidic systems have to be utilized, with small flow cross sections, complicating sample handling and increasing the danger of clogging.10 Second, strong water absorption prevents the investigation of thick samples in transmission, such as large biological cells, cell complexes (e.g., 3D cell cultures), and (aqueous) tissue. To circumvent the strong water absorption, attenuated total reflection techniques11 are often applied, albeit at the cost of even smaller penetration depths. Alternatively, the sample can be dried, which excludes the study of live organisms and strongly alters the sample.12
In order to work with larger path lengths and to mitigate the loss in signal-to-noise ratio (SNR) due to strong water absorption, it is necessary to use high-brightness sources like synchrotrons or quantum cascade lasers (QCL). By utilizing their increased MIR power, determination of glucose, lactate, and triglycerides in blood serum with path lengths above 100 μm (1030–1230 cm–1)5 and spectroscopy of proteins in aqueous solution in the Amide I and II region with path lengths as large as 32 μm13 have been demonstrated. However, power scaling is limited, because the strong water absorption eventually results in heating of the samples. Therefore, a further increase of the path length and/or sensitivity is not readily achievable when the remaining transmitted intensity is directly measured with (noisy) MIR photodetectors.
In this work, we demonstrate in theory and experiment the potential of field-resolved infrared spectroscopy (FRS)14 to overcome these long-standing limitations. FRS relies on the excitation of resonant molecular vibrations with waveform-stable, broadband MIR pulses and electric-field-resolved detection of the emerging fingerprint waveforms. This brings about two major advantages. First, the measurement signal scales linearly with the electric field and, therefore, it decreases with the square root of intensity attenuation. This effectively squares the measurement intensity dynamic range, which is usually limited by the detector dynamic range.15−17 Second, temporal isolation of the resonant molecular signal from the impulsive excitation renders FRS robust against fluctuations of the latter.14
Together, these properties result in linear scaling of the measurement SNR with the electric field. In the following, we refer to this scaling behavior as field-scaling in order to distinguish it from intensity-scaling techniques whose SNR scales linearly with the light intensity reaching the detector (e.g., conventional FTIR spectrometry). Although field-scaling is not unique to FRS, with other methods, the presence of the time-integrated sample signal (containing all noise contribution from the source) on the detector(s) makes it much more challenging to achieve this regime, especially in conjunction with high-power MIR sources. When applicable, field-scaling of the SNR brings about fundamental advantages for measurements of strongly absorbing samples.
For each SNR-scaling regime, we derive scaling laws for the limit of detection (LOD) of a given analyte in a matrix material (e.g., liquid water) having the absorption coefficient αM and considering the total interaction length x. We show that for field-scaling—in comparison to intensity-scaling—a lower LOD can be obtained for a significantly wider range of interaction lengths, which effectively increases the useable range of path lengths in applications (Figure 1). We find that the theoretical optimum sample thickness is twice the length considered optimum for FTIR measurements.18,19 Finally, we experimentally validate this formula, demonstrating that field-scaling and intensity-noise robust detection can be achieved in practice.
Figure 1.

Theoretical limit of detection (LOD) of a given analyte embedded within a matrix substance for field- and intensity-scaling techniques. The theoretical LOD values were calculated assuming a spectrometry device for both approaches offering the same SNR for a non-attenuated beam. For weak attenuations by the matrix, the LOD of field-scaling and intensity-scaling techniques are comparable. For strong attenuations, field-scaling is advantageous.
Scaling Laws and Optimum Iteraction Length
For a spectroscopic measurement, the optimum path length depends on the SNR-scaling with intensity I. Typically, a distinction into one of the three following scaling-regimes is possible:2,20
-
(1)
Square-root-scaling (
) -
(2)
Linear scaling (SNR ∼ I)
-
(3)
Constant (SNR ∼ const)
The majority of MIR spectroscopies operate either in the second or the third regime. For a given setup, the “operating regime” often depends on the average signal power reaching the detector.20 Usually, a system is intensity–noise limited at high optical powers and, for decreasing power, eventually becomes detector–noise limited. This has to be considered when choosing the interaction length. By contrast, FRS has been demonstrated to afford field-scaling operation for a range of powers extending from a few hundred photons per second to tens of milliwatts14 and, potentially, to Watt-level average powers.21
For all scaling regimes, derivations of the optimum sample thickness can be found in the Supporting Information. In the following, we focus on the discussion of scaling laws for FRS (i.e., field-scaling). Subsequently, we discuss to what extent these apply to other spectroscopic techniques and how they compare to those of conventional FTIR spectroscopy.
For any spectroscopic measurement, it is desirable to record the molecular signal with the highest possible SNR. A general approach to gaining a stronger signal is to increase the interaction length with the analyte under test. However, this brings about an attenuation of the light intensity due to interaction with the matrix containing the analyte and, for large path lengths, eventually decreases the (relative) measurement SNR. This is particularly severe for aqueous solutions, as the strong IR absorption of water is, unlike the molecules dissolved in it, mostly of little spectroscopic interest. Therefore, it is desirable to optimize the interaction length with the sample under this condition.
According to Beer’s absoprtion law, the light intensity I (at a certain frequency) after passing through a medium with absorption coefficient α and propagation length x is given by I = I0 e–αx, where I0 denotes the intensity before the medium. For the case of a field-scaling time-domain measurement, we have previously derived a simple formula for the minimum value of (αx)min detectable for a particular absorption line of a molecule:14
| 1 |
Here, the electric-field dynamic range DRE is defined as the ratio of the spectral amplitude of the electric field to the detection noise for a given instrument setting (total measurement time, etc.); TL denotes the dephasing time of the considered absorption line. For a Lorentzian-shaped absorption with a spectral width υ, the decay time is TL = (πcv)−1, with c being the speed of light. The parameter tB is defined as the instant when the temporal window for an infrared-background-free measurement begins.14
Now, let αM denote the absorption
of water (or any other matrix substance) and αA, the absorption of analyte molecules. By increasing
the interaction length x, the MIR beam is mainly
attenuated due to water absorption according to Beer’s law.
The electric field dynamic range for a certain optical frequency v scales as
with DRE0 being the
dynamic range for the measurement of the unattenuated radiation. Equation 1 now writes:
| 2 |
The minimum for αA is reached for (see the Supporting Information):
| 3 |
According to eq 3, the optimum interaction length is independent of the system’s dynamic range and source intensity noise. Therefore, for any FRS system, the choice of an optimum liquid cuvette will depend not on the particular specifications of the system but on only the absorption coefficient αM of the matrix substance at the desired wavelength.
For broadband spectroscopy, the choice of interaction length becomes more complicated as αM might vary within the spectral range of interest. This is for instance the case for aqueous samples, for which MIR spectroscopy is most informative between 1000 and 3000 cm–1. Due to the strong water absorption band at 1640 cm–1 (HOH bending vibration), the optimum interaction length for FRS is 7.4 μm, while at 2630 cm–1, it is 178 μm (Figure 2). In order to maintain good measurement performance over the entire spectral range, we suggest choosing a path length of ∼25 μm. This keeps the expected LOD within 33% of the optimum value at all wavenumbers (see the Supporting Information).
Figure 2.

(a) Optimum sample thickness for measurements in aqueous media in dependence of the wavenumber for field-scaling (blue) and intensity-scaling (red) techniques. The absorption coefficient of water (top panel) determines the path length for optimum sensitivity. (b) Theoretical limit of detection (LOD) of a given analyte dissolved in water in dependence of the thickness of the measurement cuvette at the different wavenumbers.
It is noteworthy that similar scaling laws apply to any spectroscopic method whose SNR scales linearly with the field strength. For example, Withayachumnankul et al.22 obtained a similar expression for the optimum path length in THz time-domain spectroscopy.23 However, their derivation assumes that multiplicative noise (i.e., relative intensity noise) from the source is negligible, without discussing under what circumstances this condition is met.
Similar considerations can be applied to asymmetric FTIR spectrometers where the sample is placed in one arm of the interferometer.20 However, due to limited dynamic range and poor noise performance of the available MIR detectors as well as due to source intensity noise, the favorable scaling behavior with intensity attenuation may either not be reached at all or only within a smaller range.
The discussed scaling laws and optimum interaction lengths are in sharp contrast to spectroscopic methods whose SNR is linearly proportional to the light intensity, as is the case for conventional (i.e., symmetric) FTIR spectrometers.2 For the comparison of the two scaling regimes, in the following, we consider field- and intensity-scaling devices (e.g., a symmetric and an asymmetric FTIR) with the same performance in terms of minimum detectable absorption difference for non-attenuated beams.
For weak attenuation (i.e., short path length and/or weakly absorbing matrix), the LOD is comparable (Figures 1 and 2b). For strong attenuation by the matrix (i.e., long path length and/or strongly absorbing matrix), field-scaling is clearly advantageous. This becomes particularly important for samples with nonuniform thickness (e.g., biological tissue or living cells) or varying total absorption and/or water content. In addition, the field-scaling device inherently reaches a twice as low LOD at an optimum path length that is twice the one of the intensity-scaling devices (Figure 2a and the Supporting Information).
In principle, SNR-scaling with field strength can also be obtained with direct absorption spectroscopy (DAS) in the MIR range. However, in practice, for DAS, this is only applicable when MIR shot-noise-limited performance can be reached over the entire measurement dynamic range. This is technologically very challenging due to typically orders of magnitude stronger source intensity noise. Additionally, shot-noise-limited performance can only be achieved for high MIR powers, as the noise-equivalent shot-noise power has to overcome the noise of the MIR photodetectors. When Brandstetter et al.5 investigated the path length with maximum measurement SNR for a QCL-based DAS setup, they found it to be 140 μm, determined by the intensity noise of the source. In general, when a spectroscopic system is intensity-noise limited, the full potential of the light source (set by photon shot noise) cannot be exploited and the SNR-maximized path length in experiment will be longer than the optimum value in a detection-limited setting (see the Supporting Information).
Experimental Setup
In the remainder of the paper, we verify these theoretical considerations experimentally, employing the mid-infrared field-resolving spectrometer described in detail in our previous work14 (see Figure 3). In short, a Kerr-lens mode-locked Yb:YAG thin-disc oscillator emits a 28 MHz repetition-rate train of 220 fs pulses spectrally centered at 1030 nm, with an average power of 100 W. The near-infrared (NIR) pulses are spectrally broadened in three nonlinear multipass bulk-transmission stages24 and temporally compressed to 16 fs.25 Few-cycle mid-infrared pulses, spectrally covering the 910–1530 cm–1 range at −20 dB spectral intensity with an average power of 50 mW, are obtained via intrapulse difference frequency generation (IPDFG) in a LiGaS2 crystal. After the IPDFG stage, the NIR pulses are separated from the MIR beam via a custom-designed dichroic mirror, recompressed to 16 fs, and optically delayed with respect to the MIR transients via a mechanical stage. After passing through a chopper wheel modulating the signal at 10 kHz for lock-in detection, the MIR radiation traverses a liquid cuvette for transmission measurements. Custom multilayer optics were used to compensate for the dispersion of the MIR pulse upon propagation through the ZnSe windows of the liquid cuvette and temporally compressed the MIR pulse to 59 fs (intensity envelope full width at half-maximum).26
Figure 3.

(a) Layout of the field-resolved spectrometer. For this experiment, measurement cuvettes with 7, 34, and 50 μm path lengths were used. (b) Intensity dynamic range of the FRS spectrometer after passing through water layers of 7, 34, and 50 μm thicknesses, measured over 20 s, with a spectral resolution of 10 cm–1. The gray dashed lines show the expected dynamic range after transmission through 100, 200, and 300 μm of water, respectively.
The MIR and NIR beams are temporally and spatially recombined at a germanium plate and sent to an electro-optic sampling detection system14,27,28 that allows for recording the MIR waveform. To increase measurement stability, the entire MIR beam path was put under vacuum conditions and an active intensity noise stabilization29 as well as interferometric delay tracking30 of the interferometer were implemented.
Results
The power spectrum (Figure 3b) was obtained via Fourier transforming the recorded time trace (Figure 4a). After passing through liquid cuvettes with path lengths of 7, 34, and 50 μm, we obtained peak intensity dynamic range values of 5 × 1010, 2 × 1010, and 0.5 × 1010 for a total measurement time of 20 s with a spectral resolution of 10 cm–1, respectively. The wide intensity dynamic range renders the instrument advantageous for measurements of strongly absorbing (aqueous) samples. This is also illustrated in Figure 3b, showing the calculated power spectral density of the spectrum after transmission through 100, 200, and 300 μm of water.
Figure 4.
FRS measurement of a solution of DMSO2 in water. (a) EOS traces of a reference and a test sample (liquid cuvette filled with pure water and one with a 1 mg/mL solution of DMSO2 in water, respectively). The difference between the two traces is miniscule (cannot be discerned by eye). (b) Amplitude and phase spectra obtained via Fourier transform (FT) of the traces in (a), permitting the calculation of conventional absorption and phase spectra of the analyte DMSO2 in its aqueous environment. For comparison, an absorption spectrum of the same solution, measured with a commercial FTIR spectrometer device, is shown. (c) Time-domain molecular response of DMSO2 was obtained via numerical subtraction of the sample from the reference trace, as shown in (a). Due to fluctuations in the excitation remaining after transmission through the sample (induced by intensity and phase noise of the source, by interferometer fluctuations etc.), the reconstruction of the molecular response is noisy in time windows with large EOS signal—in particular during the excitation pulse centered around 0 fs. After several hundred femtoseconds, this detrimental influence of source noise on the subtraction result becomes negligible because of the rapid decay of the excitation pulse. This opens up the window for background-free measurements of the molecular signal. (d) Similar to the conventional spectra shown in (b), one can obtain analyte-specific amplitude and phase spectra of the time-filtered sample response via FT of the truncated time-domain difference trace.
In addition to the instrument’s ability to detect electric fields with high sensitivity, the time-domain nature of FRS renders the detection of molecular emission robust against fluctuations of the excitation. This advantage is described in detail in our previous work14 and will be illustrated here for the example of dimethyl sulfone (DMSO2), a test molecule used for the measurements presented in this work. The spectral absorption and phase information on DMSO2 can be obtained from the corresponding reference and sample measurements (Figure 4a,b). However, this procedure becomes unstable for small absorptions due to signal fluctuations in the range of the excitation, affecting the (complex) Fourier-transformed spectra.
Measuring the time-domain response to an ultrashort MIR excitation pulse allows for the temporal separation of the sample response of the molecules under investigation and the—orders of magnitude stronger—excitation energy remaining after transmission through the sample. Consequently, the noise power carried by the excitation can be temporally separated from the molecular signal. For substances in aqueous environments, the resonant molecular response usually spans over a few picoseconds. In the case of DMSO2, the dephasing time of the absorption line at 1139 cm–1 is ∼0.8 ps, which is significantly longer than the full-width at half-maximum of 59 fs of the excitation pulse (Figure 4a). Already 500 fs after the pulse, the signal from the reference pulse drops to ∼0.5% of the peak value. Within this strongly reduced background, a significantly weaker signal of the excited molecules can be detected, as the detrimental influence of the excitation fluctuations is strongly reduced.
This becomes apparent when extracting the resonant molecular response by subtracting a reference from a sample measurement (Figure 4c). During the excitation, source intenstiy fluctuations hinder the clean subtraction, i.e., referencing, of two strong signals, leading to noise in the signal of interest in this time window. After a few hundred femtoseconds, the excitation pulse has decayed either below the detection noise floor or it has become weak enough to be subtracted (or referenced out). Note that a signal of the excited matrix molecules might be spanning over several picoseconds and therefore potentially masking the molecular signal of the analyte. However, vibrations of liquid water dephase within tens of femtoseconds and therefore do not affect the detection of the resonant analyte response in the wake of the excitation. In practice, this means that after a certain time tB, the detection of the molecular signal is only limited by the detection noise and not by any technical noise of the driving laser or of the interferometer.14
In our case, the window for background-free access to resonant analyte response opens up approximately 500 fs after the peak of the excitation pulse. Fourier-transforming the time-filtered molecular signal delivers a sample-specific “resonant fingerprint”, i.e., a power and a phase spectrum (Figure 4d). Even though its appearance differs from that of conventional spectra, the main spectral features are contained and can be used for quantitative identification of the analyte.
In order to validate eq 2, we evaluated the LOD for DMSO2 dissolved in water in a dilution series, using liquid cuvettes with thicknesses of 7 (commonly used in FTIR spectrometers), 34, and 50 μm. To retrieve the concentration of DMSO2 for each sample, a calibration measurement at a concentration of 1 mg/mL was performed for each thickness and used for 1-parameter fits yielding the concentrations of the individual samples.14
For a spectral resolution of 4 cm–1 (corresponding to an 8.3 ps time window) and a total effective measurement time of 90 s per injected sample (45 s each for reference and sample measurement), we experimentally obtained LOD values of 1.6, 0.6, and 0.7 μg/mL for increasing cuvette thickness, respectively (Figure 5). The theoretical LOD curve was calculated using the absorption coefficient of DMSO2 of 12.92 cm–1 for 1 mg/mL at 1139 cm–1, a decay time TL of 770 fs, the absorption coefficient of water of 537 cm–1,31 and a field dynamic range DRE0 of the non-attenuated beam of 3.0 × 105. This theoretical curve is in very good agreement with the experimentally obtained values for the LOD.
Figure 5.

Theoretical vs experimental limits of detection (LOD) for DMSO2 dissolved in water in dependence of the thickness of the measurement cuvette.
Conclusion and Outlook
We have demonstrated that broadband field-resolved MIR spectroscopy can offer an intensity dynamic range in excess of 1010 over more than 460 cm–1 (without measurement cuvette) for a measurement time of less than 1 min, rendering it applicable for the investigation of strongly absorbing (aqueous) samples. The ability of FRS to measure the temporally retarded sample response separated from the nonresonant response (i.e., excitation remaining after transmission through the sample) grants access to the—orders of magnitude smaller—molecular signal in a background-reduced manner, rendering the measurement SNR virtually independent from excitation fluctuations. This enables detection sensitivities in the sub-microgram/milliliter range for molecules solved in water. Furthermore, the signal strength and SNR scale linearly with the electric field of the excitation source. Based on these properties, we theoretically derived and experimentally confirmed that in FRS the optimum interaction length with a sample, xopt = 2/αM, only depends on the absorption coefficient αM of the matrix substance.
For many biological samples, the matrix substance is water, therefore, we further analyzed this case and evaluated the measurement performance in dependence of wavelength and thickness. We found that sub-microgram/milliliter detection sensitivities can be maintained for samples as thick as 80 μm and that, even for 0.2 mm thick samples, LOD values in the range of 10 μg/mL are feasible, which is the level achieved by state-of-the-art research-grade FTIR instruments,14 albeit under the stringent condition of sub-10 μm sample thickness. In the fingerprint region from 1000 to 3000 cm–1, the measurement performance can be kept within 33% of the peak performance at all wavenumbers by choosing a sample thickness around 25 μm. Rapid advances of femtosecond technology hold promise for the extension of FRS to the coverage of the entire molecular fingerprint region in the near future.32,33 At the same time, further improvement of the sensitivity of electro-optic sampling will push the limit of detection to the low nanogram/milliliter range or below.34
Compared to conventional (i.e., symmetric) FTIR spectrometers, the scaling laws of FRS for the sensitivity in dependence of the sample thickness and total absorption are highly advantageous. The relaxed requirements to the sample thickness enable a more flexible design of liquid cuvettes and microfluidic chips for MIR spectroscopic applications. In addition, the increased SNR for the measurement of thick aqueous samples will be beneficial for MIR transmission spectroscopy and spectroscopy–microscopy of biological samples, such as living cells, bulk cells, and tissue cultures as well as biological tissues, in their natural (hydrated) state.
Supporting Information Available
The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acs.analchem.9b05744.
Discussions of scaling laws for the limit of detection in aqueous media for intensity- and field-scaling spectroscopic methods and relative-intensity noise dominated measurements (PDF)
The authors declare no competing financial interest.
Supplementary Material
References
- Diem M.; Griffiths P. R.; Chalmers J. M.. Vibrational Spectroscopy for Medical Diagnosis; Wiley: Chichester, 2008; Vol. 40. [Google Scholar]
- Griffiths P. R.; De Haseth J. A.. Fourier Transform Infrared Spectrometry; John Wiley & Sons, 2007; Vol. 171. [Google Scholar]
- Yang H.; Yang S.; Kong J.; Dong A.; Yu S. Obtaining Information about Protein Secondary Structures in Aqueous Solution Using Fourier Transform IR Spectroscopy. Nat. Protoc. 2015, 10 (3), 382–396. 10.1038/nprot.2015.024. [DOI] [PubMed] [Google Scholar]
- Martin F. L.; Kelly J. G.; Llabjani V.; Martin-Hirsch P. L.; Patel I. I.; Trevisan J.; Fullwood N. J.; Walsh M. J. Distinguishing Cell Types or Populations Based on the Computational Analysis of Their Infrared Spectra. Nat. Protoc. 2010, 5 (11), 1748–1760. 10.1038/nprot.2010.133. [DOI] [PubMed] [Google Scholar]
- Brandstetter M.; Volgger L.; Genner A.; Jungbauer C.; Lendl B. Direct Determination of Glucose, Lactate and Triglycerides in Blood Serum by a Tunable Quantum Cascade Laser-Based Mid-IR Sensor. Appl. Phys. B: Lasers Opt. 2013, 110 (2), 233–239. 10.1007/s00340-012-5080-z. [DOI] [Google Scholar]
- Movasaghi Z.; Rehman S.; Ur Rehman D. I. Fourier Transform Infrared (FTIR) Spectroscopy of Biological Tissues. Appl. Spectrosc. Rev. 2008, 43 (2), 134–179. 10.1080/05704920701829043. [DOI] [Google Scholar]
- Ollesch J.; Theegarten D.; Altmayer M.; Darwiche K.; Hager T.; Stamatis G.; Gerwert K. An Infrared Spectroscopic Blood Test for Non-Small Cell Lung Carcinoma and Subtyping into Pulmonary Squamous Cell Carcinoma or Adenocarcinoma. Biomed. Spectrosc. Imaging 2016, 5 (2), 129–144. 10.3233/BSI-160144. [DOI] [Google Scholar]
- Hands J. R.; Clemens G.; Stables R.; Ashton K.; Brodbelt A.; Davis C.; Dawson T. P.; Jenkinson M. D.; Lea R. W.; Walker C.; Baker M. J. Brain Tumour Differentiation: Rapid Stratified Serum Diagnostics via Attenuated Total Reflection Fourier-Transform Infrared Spectroscopy. J. Neuro-Oncol. 2016, 127 (3), 463–472. 10.1007/s11060-016-2060-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rogalski A.Infrared Detectors; CRC press, 2010. [Google Scholar]
- Vaccari L.; Birarda G.; Businaro L.; Pacor S.; Grenci G. Infrared Microspectroscopy of Live Cells in Microfluidic Devices (MD-IRMS): Toward a Powerful Label-Free Cell-Based Assay. Anal. Chem. 2012, 84 (11), 4768–4775. 10.1021/ac300313x. [DOI] [PubMed] [Google Scholar]
- Kazarian S. G.; Chan K. L. A. ATR-FTIR Spectroscopic Imaging: Recent Advances and Applications to Biological Systems. Analyst 2013, 138 (7), 1940. 10.1039/c3an36865c. [DOI] [PubMed] [Google Scholar]
- Cameron J. M.; Butler H. J.; Palmer D. S.; Baker M. J. Biofluid Spectroscopic Disease Diagnostics: A Review on the Processes and Spectral Impact of Drying. J. Biophotonics 2018, 11 (4), e201700299 10.1002/jbio.201700299. [DOI] [PubMed] [Google Scholar]
- Schwaighofer A.; Montemurro M.; Freitag S.; Kristament C.; Culzoni M. J.; Lendl B. Beyond Fourier Transform Infrared Spectroscopy: External Cavity Quantum Cascade Laser-Based Mid-Infrared Transmission Spectroscopy of Proteins in the Amide i and Amide II Region. Anal. Chem. 2018, 90 (11), 7072–7079. 10.1021/acs.analchem.8b01632. [DOI] [PubMed] [Google Scholar]
- Pupeza I.; Huber M.; Trubetskov M.; Schweinberger W.; Hussain S. A.; Hofer C.; Fritsch K.; Poetzlberger M.; Vamos L.; Fill E.; Amotchkina T.; Kepesidis K. V.; Apolonski A.; Karpowicz N.; Pervak V.; Pronin O.; Fleischmann F.; Azzeer A.; Žigman M.; Krausz F. Field-Resolved Infrared Spectroscopy of Biological Systems. Nature 2020, 577 (7788), 52–59. 10.1038/s41586-019-1850-7. [DOI] [PubMed] [Google Scholar]
- Snyder J. J. Wide Dynamic Range Optical Power Measurement Using Coherent Heterodyne Radiometry. Appl. Opt. 1988, 27 (21), 4465. 10.1364/AO.27.004465. [DOI] [PubMed] [Google Scholar]
- Jarry G.; Poupinet L.; Watson J.; Lepine T. Extinction Measurements in Diffusing Mammalian Tissue with Heterodyne Detection and a Titanium:Sapphire Laser. Appl. Opt. 1995, 34 (12), 2045. 10.1364/AO.34.002045. [DOI] [PubMed] [Google Scholar]
- Liukaityte S.; Lequime M.; Zerrad M.; Begou T.; Amra C. Broadband Spectral Transmittance Measurements of Complex Thin-Film Filters with Optical Densities of up to 12. Opt. Lett. 2015, 40 (14), 3225. 10.1364/OL.40.003225. [DOI] [PubMed] [Google Scholar]
- Jensen P. S.; Bak J. Near-Infrared Transmission Spectroscopy of Aqueous Solutions: Influence of Optical Pathlength on Signal-To-Noise Ratio. Appl. Spectrosc. 2002, 56 (12), 1600–1606. 10.1366/000370202321115878. [DOI] [Google Scholar]
- Mark H. L.; Griffiths P. R. Analysis of Noise in Fourier Transform Infrared Spectra. Appl. Spectrosc. 2002, 56 (5), 633–639. 10.1366/0003702021955196. [DOI] [Google Scholar]
- Newbury N. R.; Coddington I.; Swann W. Sensitivity of Coherent Dual-Comb Spectroscopy. Opt. Express 2010, 18 (8), 7929. 10.1364/OE.18.007929. [DOI] [PubMed] [Google Scholar]
- Butler T. P.; Gerz D.; Hofer C.; Xu J.; Gaida C.; Heuermann T.; Gebhardt M.; Vamos L.; Schweinberger W.; Gessner J. A.; Siefke T.; Heusinger M.; Zeitner U.; Apolonski A.; Karpowicz N.; Limpert J.; Krausz F.; Pupeza I. Watt-Scale 50-MHz Source of Single-Cycle Waveform-Stable Pulses in the Molecular Fingerprint Region. Opt. Lett. 2019, 44 (7), 1730. 10.1364/OL.44.001730. [DOI] [PubMed] [Google Scholar]
- Withayachumnankul W.; Fischer B. M.; Abbott D. Material Thickness Optimization for Transmission-Mode Terahertz Time-Domain Spectroscopy. Opt. Express 2008, 16 (10), 7382. 10.1364/OE.16.007382. [DOI] [PubMed] [Google Scholar]
- Jepsen P. U.; Cooke D. G.; Koch M. Terahertz Spectroscopy and Imaging - Modern Techniques and Applications. Laser Photon. Rev. 2011, 5 (1), 124–166. 10.1002/lpor.201000011. [DOI] [Google Scholar]
- Schulte J.; Sartorius T.; Weitenberg J.; Vernaleken A.; Russbueldt P. Nonlinear Pulse Compression in a Multi-Pass Cell. Opt. Lett. 2016, 41 (19), 4511. 10.1364/OL.41.004511. [DOI] [PubMed] [Google Scholar]
- Fritsch K.; Poetzlberger M.; Pervak V.; Brons J.; Pronin O. All-Solid-State Multipass Spectral Broadening to Sub-20 Fs. Opt. Lett. 2018, 43 (19), 4643. 10.1364/OL.43.004643. [DOI] [PubMed] [Google Scholar]
- Amotchkina T.; Trubetskov M.; Hussain S. A.; Hahner D.; Gerz D.; Huber M.; Schweinberger W.; Pupeza I.; Krausz F.; Pervak V. Broadband Dispersive Ge/YbF 3 Mirrors for Mid-Infrared Spectral Range. Opt. Lett. 2019, 44 (21), 5210. 10.1364/OL.44.005210. [DOI] [PubMed] [Google Scholar]
- Gallot G.; Grischkowsky D. Electro-Optic Detection of Terahertz Radiation. J. Opt. Soc. Am. B 1999, 16 (8), 1204. 10.1364/JOSAB.16.001204. [DOI] [Google Scholar]
- Wu Q.; Zhang X. Free-space Electro-optic Sampling of Terahertz Beams. Appl. Phys. Lett. 1995, 67 (24), 3523–3525. 10.1063/1.114909. [DOI] [Google Scholar]
- Huber M.; Schweinberger W.; Stutzki F.; Limpert J.; Pupeza I.; Pronin O. Active Intensity Noise Suppression for a Broadband Mid-Infrared Laser Source. Opt. Express 2017, 25 (19), 22499. 10.1364/OE.25.022499. [DOI] [PubMed] [Google Scholar]
- Schweinberger W.; Vamos L.; Xu J.; Hussain S. A.; Baune C.; Rode S.; Pupeza I. Interferometric Delay Tracking for Low-Noise Mach-Zehnder-Type Scanning Measurements. Opt. Express 2019, 27 (4), 4789. 10.1364/OE.27.004789. [DOI] [PubMed] [Google Scholar]
- Segelstein D. J.The Complex Refractive Index of Water. University of Missouri: Kansas City, 1981. [Google Scholar]
- Butler T. P.; Lilienfein N.; Xu J.; Nagl N.; Hofer C.; Gerz D.; Mak K. F.; Gaida C.; Heuermann T.; Gebhardt M.; Limpert J.; Krausz F.; Pupeza I. Multi-Octave Spanning, Watt-Level Ultrafast Mid-Infrared Source. J. Phys. Photonics 2019, 1 (4), 044006. 10.1088/2515-7647/ab4976. [DOI] [Google Scholar]
- Wang Q.; Zhang J.; Kessel A.; Nagl N.; Pervak V.; Pronin O.; Mak K. F. Broadband Mid-Infrared Coverage (2–17 Mm) with Few-Cycle Pulses via Cascaded Parametric Processes. Opt. Lett. 2019, 44 (10), 2566. 10.1364/OL.44.002566. [DOI] [PubMed] [Google Scholar]
- Hofer C.; Hussain S. A.; Schweinberger W.; Huber M.; Butler T. P.; Gerz D.; Karpowicz N.; Krausz F.; Pupeza I.. Quantum-Efficiency and Bandwidth Optimized Electro-Optic Sampling. 2019 Conference on Lasers and Electro-Optics Europe & European Quantum Electronics Conference (CLEO/Europe-EQEC); IEEE, 2019; p 1. 10.1109/CLEOE-EQEC.2019.8873083. [DOI]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.