Skip to main content
Saudi Journal of Medicine & Medical Sciences logoLink to Saudi Journal of Medicine & Medical Sciences
. 2020 Apr 17;8(2):80–86. doi: 10.4103/sjmms.sjmms_410_19

Clinical Efficacy of Photobiomodulation on Dental Implant Osseointegration: A Systematic Review

Sara Mahmoud Zayed 1,, Ahmed Adel Abdel Hakim 1
PMCID: PMC7305678  PMID: 32587488

Abstract

Background:

Photobiomodulation (PBM) has been shown to have a positive effect on dental implant osseointegration and stability in in vitro and animal studies; however, its usefulness in dental implant clinical practice is yet unclear.

Objective:

The objective was to assess the clinical effectiveness of PBM on dental implants’ osseointegration.

Methods:

Two reviewers independently conducted a comprehensive electronic search of articles published from inception up to January 10, 2020, in PubMed, Cochrane Library and Embase databases following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. Randomized clinical trials (RCTs) and nonrandomized clinical studies that compared the effect of PBM on dental implant stability with control groups were included. Animals and in vitro studies studies as well as studies with confounders such as application of orthodontic were excluded. Risk of bias (using Cochrane Risk of Bias tool for RCTs and Risk of Bias in Non-Randomized Studies of Interventions tool for nonrandomized studies) was assessed by both authors. Owing to substantial heterogeneity, only a narrative synthesis of the included studies is presented.

Results:

Seven relevant clinical studies were included, and they used a variety of PBM parameters and devices. The posterior region of the jaw was found to be more frequently evaluated. For assessing the effect of PBM on implant stability, five studies used resonance frequency analysis and two used periotest; three studies additionally used biomarkers for assessment. Four studies found that PBM has a potential positive effect on the outcome of dental implant stability, whereas three studies reported that PBM has no effect on implant stability.

Conclusion:

The findings of this systematic review suggest that postoperative application of PBM may potentially have some positive effect on dental implant’s osseointegration and stability. However, additional studies are required with uniformity in methods to provide a more robust assessment of this effect.

Keywords: Dental implants, implant stability, low-level laser/light therapy, osseointegration, photobiomodulation

INTRODUCTION

Developments in photonic technologies and an improved knowledge of light–tissue interactions have brought about considerable advances in mainstream medicine. Using high-dose lasers and light devices in medicine and dentistry is nowadays common for various clinical applications.[1,2,3] However, a less prevalent clinical application is the use of low-dosage biophotonics for therapy, referred to as photobiomodulation (PBM) or formerly as low-level laser therapy (LLLT), which is intended for relieving pain and inflammation, regulating immune responses as well as stimulating wound healing and tissue regeneration.[3,4] In this manner, it presents an opportunity for a paradigmatic shift from conventional approaches to regenerative modalities in clinical dentistry. Although PBM initially received mixed responses,[4] it is increasingly being recognized based on the accepted principles of photochemistry and cellular and molecular biology. The growing realization that PBM has a broad range of systemic and regional effects led to a striking increase in the number of its applications, with a special focus on its applications in dentistry.[4,5,6]

In the field of oral implantology, research has been directed toward the potential of PBM to reduce the healing time following implant placement and to improve the potential of bone regeneration. Experimental studies have reported that PBM stimulates proliferation and differentiation of osteoblasts as well as it enhances bonding to titanium implant. In these studies, application of PBM in the early postoperative application was shown to improve the mechanical strength of the bone–implant interface and stimulate the bone matrix production.[7,8,9,10,11] However, the usefulness of PBM application in dental implant clinical practice is not yet clear. Thus, this systematic review was conducted to answer the following research question formulated based on the population, intervention, comparator and outcome (PICO) criteria: “In patients receiving dental implants, does postoperative application of PBM improve osseointegration?”

MATERIALS AND METHODS

This systematic review was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA).[12] Based on the research question, the population was adult patients who had received dental implantation, the intervention was postoperative application of PBM and the comparator was control group. The principal outcome was successful osseointegration of dental implants in terms of improved implant stability, and secondary outcomes were anti-inflammatory and accelerated healing effects.

Search strategy

Two reviewers independently conducted an electronic search of PubMed, Cochrane Library and Embase for articles from inception up to January 10, 2020, using the following search terms: (dental implant OR dental implantology) AND (photobiomodulation OR low level laser therapy OR low level light therapy OR LLLT OR low intensity laser OR soft laser OR laser bio-stimulation) AND (osseointegration OR implant stability). No language filter was applied during the search. The title and abstracts of all articles in the searches were screened against the inclusion criteria to identify potentially eligible articles. Full texts of potential articles were read and assessed independently by the two reviewers. Any discrepancy was resolved through discussion until a consensus was reached. The inter-reviewers’ reliability was calculated based on Cohen’s kappa statistics. To detect any potential unidentified studies, bibliographies and reference lists of the included studies were also searched.

Inclusion/exclusion criteria

Randomized clinical trials (RCTs) and nonrandomized clinical studies that compared LLLT/PBM and other modalities such as ozone application on dental implant stability were included. Exclusion criteria were in vitro and animal studies, case reports, review articles, clinical studies on augmented implant sites and studies assessing a single type of treatment in the absence of a comparison group. In addition, to allow for reliable comparison, studies that applied orthodontic force on mini-implants were excluded to avoid its confounding effect on osseointegration and implant stability.

Data extraction

The two authors manually and independently extracted the data from each of the included studies; any disagreements in data extraction were resolved through discussion. The extracted data fields included first author name, year of publication, country where research was performed, study design, control group details, sample size, gender and mean age of patients, ethical approval, implant-related parameters (type, dimension and position), PBM parameters, number of PBM applications, analysis conducted, duration of follow-up and results.

Quality assessment

Methodological quality of the included studies was evaluated using the Cochrane Risk of Bias tool[13] for RCTs and the Risk of Bias in Nonrandomized Studies-of Interventions (ROBINS-I) tool[14] for nonrandomized studies.

RESULTS

The initial database search yielded a total of 202 articles. All relevant articles were only published in English. After removal of duplicates and applying the inclusion/exclusion criteria, full-text articles of 15 relevant studies were thoroughly evaluated independently by the two reviewers. Eight studies[15,16,17,18,19,20,21,22] did not meet the eligibility criteria and were excluded [Table 1]. Eventually, seven studies were included for data synthesis [Figure 1].

Table 1.

List of articles excluded after full-text assessment and reason for exclusion

Study Reason for exclusion
Sleem et al., 2019[15] PBM combined with platelet-rich fibrin - which may influence PBM effect
Flieger et al., 2019[16] Orthodontic mini-implants
Marañón-Vásquez et al., 2019[17] Orthodontic mini-implants
Arakeeb et al., 2019[18] PBM combined with growth factors
Abohabib et al., 2018[19] Orthodontic mini-implants
Mikhail et al., 2018[20] PBM combined with Vitamin C, omega-3 and calcium therapy
Awad et al., 2017[21] Diabetic patients included in the study
Yanaguizawa et al., 2017[22] Orthodontic mini-implants

PBM – Photobiomodulation

Figure 1.

Figure 1

Flow diagram showing the screening and selection of articles

The Cohen’s kappa value for inter-investigator agreement on the selected studies was 0.85. Owing to inconsistent data and substantial heterogeneity in the included studies, a meta-analysis was not possible, and thus a narrative synthesis of the included studies is presented.

General characteristics of the included studies

Of the seven studies included, five[23,24,25,26,27] were RCTs and two were prospective clinical studies.[28,29] Three studies[23,24,25] used the “split-mouth” experimental design, while the remaining studies separately included individuals as controls.[26,27,28,29] All studies were conducted between 2012 and 2019. Two studies were conducted in Turkey;[28,29] two in Iran;[25,26] and one each in Serbia,[24] Poland[27] and Brazil.[23] All the included studies were conducted in academic research institutions after ethical approval; only one study[25] did not report the ethical approval. The number of participants in the studies ranged from 8 to 25, while the age ranged from 20 to 64 years; one study[25] did not report the mean age. Six studies enrolled healthy, nonsmoking patients with no systemic alterations, whereas the study by Gokmenoglu et al.[28] did not exclude smokers. All the included studies evaluated the efficacy of PBM by comparing against a control group. Four studies[23,24,26,27] compared laser PBM against control group; one study each compared laser PBM and light-emitting diode (LED) PBM against control group,[25] LED PBM against a control group,[28] and ozone therapy to laser PBM and control group [Table 2].[29]

Table 2.

Description of the included studies

Point of comparison Author

García-Morales et al., 2012[23] Mandić et al., 2014[24] Memarian et al., 2017[25] Torkzaban et al., 2018[26] Matys et al., 2019[27] Gokmenoglu et al., 2014[28] Karaca et al., 2018[29]
Country Brazil Serbia Iran Iran Poland Turkey Turkey
Study design Split mouth (RCT) Split mouth (RCT) Split mouth (RCT) Different individuals (RCT) Different individuals (RCT) Different individuals (non-RCT) Different individuals (non-RCT)
Control Placebo Placebo Placebo Placebo Placebo Placebo Placebo
Sample size 8 12 12 19 24 15 25
Gender 2 males 6 females 6 males 6 females Not reported 9 males 10 females 16 males 8 females 9 males 6 females 11 males 14 females
Mean age (years), range 36, 20-55 61.28, 55-75 Not reported Male: 40.8 Female: 43 46.7±8.7 Control: 45.87+13.46 LED: 50.43+9.25 51.2±2.3, 36-64
Ethical approval Yes Yes Not reported Yes Yes Yes Yes
Implant type and dimensions XiVE-S implants (Dentsply, Germany) Diameter: 3.8 mm Length: 11 mm Self-tapping implants BlueSky® (Bredent, Germany) Diameter: 4 mm Length: 10 mm DIO implants (invasive fungal infectionstissue level) with resorbable blast media surface (Korea) Dimensions not reported DIO UF implants (Korea) Diameter: 4-4.5 mm Length: 10-11.5 mm Superline (Dentium, Korea) Diameter: 4.5 mm Length: 10-12 mm XiVE-S implants (Dentsply, Germany) Control diameter: 3.8 (3.5-4.5) mm Length: 11 mm LED: Diameter: 4.5 (3.8-4.5) mm Length: 11.0 (10.6-11.0) DTI Implant Systems (Turkey) Diameter: 4-4.5 mm Length: 10 mm
Implant position Posterior mandible Posterior maxilla Anterior mandible Posterior maxilla Posterior mandible Not reported Posterior mandible
Laser type and wavelength GaAlAs 830 nm GaAlAs 637 nm Diode 810 nm, LED 626 nm Diode 940 nm Diode 635 nm LED 626 nm GaAlAs 830 nm
PBM parameters 86mW, CW 0.25 J, 3s/point, 20 points 0.0028 cm2, 3.71 W/cm2 92.1 J/cm2, in contact 40mW, CW, 6.26 J/cm2 per implant Noncontact (1 cm) 50mw, CW 50 mw/cm2 20 J/cm2 In contact 185 mW, 38.5 mW/cm2, 46.2 J/cm2 Transcutaneous 100 mw, CW 354.6 mw/cm2, 14.18 J/cm2 In contact 100 mw, CW 0.5 cm2 199.04 mW/cm2 8 J/cm2 In contact 185 mW; 222 J; 20 min 38.5 mW/cm2; 46.2 J/cm2 Transcutaneous 86 mW, CW 0.0028 cm2, 0.25J per point, 3.71 W/cm2 92.1 J/cm2, in contact
Number of applications 7 7 5 6 6 9 7
Analysis conducted ISQ assessed by RFA ISQ by RFA+ALP activity and early implant success rate Implant stability By periotest + IL1β, PGE2 ISQ assessed by RFA Implant stability By periotest & Bone Density by CBCT ISQ by RFA and IL-1β, TGF-b, PGE2, NO levels in peri-implant crevicular fluid ISQ assessed by RFA
Follow up 12 weeks 6 weeks 8 weeks 12 weeks 12 weeks 12 weeks 6 months
Results No effect on implant stability No significant influence on osseointegration of self-tapping Implants placed into low-density bone Positive effect on the stability of the implants 3 weeks and no effect on the level of IL1β and PGE2 in 4 and 8 weeks No significant effect on dental implant stability PBM enhanced secondary implant stability and bone density Positive effect on osseointegration and implant stability In LED group, a negative correlation found between PGE2 and ISQ values Positive effect on dental implant stability

RCT – Randomized clinical trial; LED – Light-emitting diodes; ISQ – Implant stability quotient; RFA – Resonance frequency analysis; IL-1β – Interleukin-1β; TGF-β – Transforming growth factor-β; PGE2 – Prostaglandin-E2; NO – Nitric oxide; ALP – Alkaline-phosphatase; CBCT – Cone-beam computed tomography

According to the Cochrane Risk of Bias tool, only two of the five randomized studies[23,24] were of low risk of bias, while the other three studies[25,26,27] were of moderate risk of bias [Table 3]. The risk of bias of the two nonrandomized trials[28,29] was found to be moderate, according to the ROBINS-I tool [Table 4].

Table 3.

Risk-of-bias assessment of the included randomized clinical trials according to Cochrane Risk of Bias Assessment Tool

Domain Authors

García-Morales et al., 2012[23] Mandić et al., 2014[24] Memarian et al., 2017[25] Torkzaban et al., 2018[26] Matys et al., 2019[27]
Random sequence generation (selection bias) + + + + +
Allocation concealment (selection bias) + + ? ? ?
Blinding of participants and personnel (performance bias) + + ? ? ?
Blinding of outcome assessment (detection bias) + + ? ? ?
Incomplete outcome data (attrition bias) + + + + +
Selective outcome reporting (reporting bias) + + + + +
Other potential threats to validity + + ? + +
Risk of bias Low Low Unclear Unclear Unclear

+ – Low risk of bias; ? – Unclear risk of bias

Table 4.

Risk-of-bias assessment of the included nonrandomized clinical trials with the ROBINS-I tool

ROBIN-I item Study

Gokmenoglu et al., 2014[28] Karaca et al., 2018[29]
Bias due to confounding Moderate Moderate
Bias in the selection of participants Moderate Low
Bias in the classification of interventions Low Low
Bias due to deviations from intended interventions Low Low
Bias due to missing data Moderate Low
Bias in the measurement of outcomes Moderate Moderate
Bias in selection of the reported result Low Low
Overall bias Moderate Moderate

Characteristics of treatment and methods used

In terms of the implant systems used, two studies each used XiVE®-s (Dentsply, USA)[23,28] and DIO implant (South Korea),[25,26] and one study each used BlueSKY (Bredent, Germany),[24] DTI implant system (Turkey)[29] and Superline (Dentium, South Korea).[27] Nevertheless, the implant dimensions used were very close; the length ranged from 10 to 12 mm and the diameter from 3.8 to 4.5 mm. Only one study[25] did not report the implant dimensions used.

Mostly, the gallium aluminum arsenide[23,24,29] and semi-conductor diode[25,26,27] lasers were used, and LED was used in one study.[28] A comparison between LED and diode was done in the study by Memarian et al.[25]

Most studies used lasers in the near-infrared spectrum; however, the wavelengths differed: two studies used 830 nm[23,29] and one study each used 810 nm[25] and 940 nm.[26] Three studies used visible-range lasers with wavelengths of 626 nm and 637 nm.[24,27,28] Follow-up period varied from 6 to 24 weeks, and the number of PBM applications ranged from 5 to 9. In terms of the jaw regions evaluated, three studies were performed on the posterior area of the mandible,[23,27,29] two on the maxillary posterior[24,26] and one on the anterior mandible;[25] one study did not state the specific irradiated area.[28] Five studies examined the effect of PBM in terms of implant stability quotient by resonance frequency analysis (RFA),[23,24,26,28,29] whereas the remaining two studies used periotest.[25,27] Three studies additionally examined biomarkers to augment their results.[24,25,28]

Outcomes

Four studies[25,27,28,29] reported that PBM application had a potential positive effect on the outcome of dental implant stability. On the other hand, three studies[23,24,26] reported no significant influence, and there were no reports of adverse or negative effects of PBM. Therefore, a significant positive influence of PBM on osseointegration clinically was not established [Table 2].

DISCUSSION

Dental implant stability depends on the implant’s capacity to successfully osseointegrate, that is, successful bone-to-implant union while preserving the structural and functional integrity of the host site.[30,31,32] PBM is increasingly being acknowledged in the literature for its potential beneficial effects in medicine and dentistry.[3,4,33] Experimental studies have evaluated the use of PBM to stimulate osteoblast activity in vitro and concluded that PBM enhances the stability of dental implants. Furthermore, it was found to be capable of boosting the healing process around the surgical site by increasing adenosine triphosphate synthesis and angiogenesis, in addition to increasing the proliferation of osteoblast and reducing inflammation.[31,32,33] Conversely, other articles had reported contradictory results.[34,35] Such discrepancies may be attributed to variations in the irradiation protocols and/or the experimental models used.[4,6,36] However, given that humans have a different and more complex biological nature than experimental models, results from in vitro and animal models may not necessarily apply clinically.[33] Accordingly, the current systematic review was envisioned to address a focused research question related to the clinical efficacy of PBM on dental implant osseointegration and stability.

Three studies used the split-mouth study design.[23,24,25] Of these, García-Morales et al.[23] reported that there was no evidence of PBM influencing the dental implant stability when measured by RFA. Nonetheless, the authors attributed the high primary stability and favorable bone quality to the rigid bone–implant interface, which may have concealed the effect of PBM. Similarly, Mandić et al.[24] also reported no positive impact on self-tapping implants’ osseointegration when placed into the low-density bone of the posterior maxilla. In this study, the spot size, irradiance and exposure duration were not reported. These parameters may have influenced the results. Moreover, the distance (10 mm) between the probe and tissues may have caused significant amount of laser energy to be reflected, thereby reducing the effectiveness of LLLT or rendering it ineffective. In contrast, Memarian et al.[25] found that PBM by LLL or LED had a positive influence on dental implant stability 3 weeks post surgery. Although this study is of moderate risk of bias, its finding supports previous in vivo study on rabbit mandible, which used 830 nm wavelength at 20 J/cm2 and demonstrated a significant positive effect on new bone formation around dental implants inserted in the rabbit mandible.[8]

Research using the split-mouth design for PBM is arguable owing to its potential systemic effect.[37,38,39] Nonetheless, other studies were also found in the searches that used different individuals as control groups to exclude such systemic effect.[26,27,28,29] Karaca et al.[29] used the same laser parameters previously used by García-Morales et al.,[23] but in different individuals, and reported a positive influence on the healing of the bone surrounding immediately loaded implants and showed a clinically insignificant increase in implant stability. Nevertheless, the results of the later García-Morales et al.[23] may be considered more reliable because of the more appropriate blinding and randomization, rendering it a high-quality study with lower risk of bias.

Gokmenoglu et al.[28] also showed a positive impact of PBM on implant’s osseointegration and stability. They reported that in the control group during the evaluation period, the primary stability revealed significant reduction. Conversely, an insignificant decrease in the primary stability was found in the LED group. In other words, PBM using LED may have maintained the primary implant stability. In the same vein, a more recent RCT[27] concluded that PBM insignificantly improved secondary implant stability – measured by periotest – and significantly improved the bone density after 12 weeks at the middle and apical parts of the implant, as calculated by the cone-beam computed tomography. It is also worth noting that this study used a dose of 8 J/cm2 in the contact mode, which is within the recommended therapeutic window of PBM.[4] In contrast, Torkzaban et al.[26] found that postoperative application of PBM had no significant influence on the stability of dental implants. It is worth noting that they used irradiance of 354.6 mW/cm2, but the values recommended for stimulation and healing range from 5 to 50 mW/cm2;[4] therefore, the insignificant effect may be explained by the biphasic response phenomenon.

The energy dose used in the studies included ranged from 6.2 to 92.1 J/cm2, thereby reflecting an absence of uniformity in the ideal dose for PBM. Moreover, the results are heterogeneous, indicating that the effectiveness of PBM on dental implants’ stability and osseointegration remains controversial. This may be due to the diversity in the parameters used: energy density, number of applications, wavelength and power. This information is of importance for readers and researchers to be able to appraise the studies’ results and the specific treatment applied.

The lack of consensus on the optimal PBM protocol for dental implants osseointegration may be due to different devices and different application protocols used. In addition, the results of this review should be interpreted cautiously owing to the presence of uncontrolled factors in the included studies such as intra- and inter-subject variations; for instance, the thickness and the light absorption characteristics of the intervening tissue affect the dose delivered at the target tissue. Nevertheless, a positive insignificant influence of PBM application on implant stability was reported in four studies,[25,27,28,29] which cannot be disregarded and warrants further research that controls for confounding factors to provide a clearer understanding. In addition, substantial efforts are necessary to achieve standardized clinical dosing and delivery protocols for PBM therapy to warrant the maximum efficacy and safety for PBM and allow for more conclusive studies with meta-analysis.

Regarding the impact of PBM on biomarkers and the inflammatory factors, Memarian et al.[25] reported a significant decrease of both prostaglandin-E2 and interleukin-1β after 2 months. This is a natural phenomenon due to the decline in the inflammatory processes while healing is completed. However, PBM had no impact on the level of these inflammatory markers at different time intervals compared with the control group. Their results are consistent with those of Gokmenoglu et al.,[28] which indicates that the PBM had no influence on the inflammatory factors. A possible explanation is that the influence of PBM on these factors may be considered effective in real infectious situations such as periimplantitis and not natural healing phenomenon of inflammation, i.e. it has no effect on the cells in homeostasis.[40] Moreover, Mandić et al.[24] showed that there was no statistically significant difference in the alkaline-phosphatase (ALP) activity between the test and control groups at all observation points. However, the pattern of ALP activity changes over time was different. In contrast to the control group, where continuous decrease of ALP activity was recorded, in the test group, after the initial decline, an increase was observed in the 4th week. The increase in ALP activity in the laser group might be interpreted as an indication of enhanced osteoblast activity, and thus improved bone neo-formation and mineralization. This biochemical result may suggest a positive influence of PBM on osseointegration.

CONCLUSION

This systematic review found that postoperative application of PBM may have a potential positive clinical influence on dental implants’ osseointegration and stability. However, there were limited number of high-quality studies in addition to the diversity in methodologies and laser parameters used. Therefore, PBM protocols need to be standardized and additional RCTs with larger sample size and less intersubject variations should be performed to confirm the effectiveness of PBM on dental implants’ osseointegration and stability.

Peer review

This article was peer reviewed by three independent and anonymous reviewers.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

REFERENCES

  • 1.Verma SK, Maheshwari S, Singh RK, Chaudhari PK. Laser in dentistry: An innovative tool in modern dental practice. Natl J Maxillofac Surg. 2012;3:124–32. doi: 10.4103/0975-5950.111342. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Sgolastra F, Petrucci A, Severino M, Gatto R, Monaco A. Lasers for the treatment of dentin hypersensitivity: A meta-analysis. J Dent Res. 2013;92:492–9. doi: 10.1177/0022034513487212. [DOI] [PubMed] [Google Scholar]
  • 3.Arany PR. Craniofacial wound healing with photobiomodulation therapy: New insights and current challenges. J Dent Res. 2016;95:977–84. doi: 10.1177/0022034516648939. [DOI] [PubMed] [Google Scholar]
  • 4.Hamblin MR. Handbook of Low-Level Laser Therapy. Singapore: Pan Stanford Publishing Pte, Ltd; 2017. [Google Scholar]
  • 5.Carroll JD, Milward MR, Cooper PR, Hadis M, Palin WM. Developments in low level light therapy (LLLT) for dentistry. Dent Mater. 2014;30:465–75. doi: 10.1016/j.dental.2014.02.006. [DOI] [PubMed] [Google Scholar]
  • 6.Chung H, Dai T, Sharma SK, Huang YY, Carroll JD, Hamblin MR. The nuts and bolts of low-level laser (light) therapy. Ann Biomed Eng. 2012;40:516–33. doi: 10.1007/s10439-011-0454-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Blay A, Blay CC, Tunchel S, Gehrke SA, Shibli JA, Groth EB, et al. Effects of a Low-Intensity Laser on Dental Implant Osseointegration: Removal Torque and Resonance Frequency Analysis in Rabbits. J Oral Implantol. 2016;42:316–20. doi: 10.1563/aaid-joi-D-15-00064. [DOI] [PubMed] [Google Scholar]
  • 8.Massotti FP, Gomes FV, Mayer L, de Oliveira MG, Baraldi CE, Ponzoni D, et al. Histomorphometric assessment of the influence of low-level laser therapy on peri-implant tissue healing in the rabbit mandible. Photomed Laser Surg. 2015;33:123–8. doi: 10.1089/pho.2014.3792. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Mayer L, Gomes FV, Carlsson L, Gerhardt-Oliveira M. Histologic and resonance frequency analysis of peri-implant bone healing after low-level laser therapy: An In vivo study. Int J Oral Maxillofac Implants. 2015;30:1028–35. doi: 10.11607/jomi.3382. [DOI] [PubMed] [Google Scholar]
  • 10.Mayer L, Gomes FV, de Oliveira MG, de Moraes JF, Carlsson L. Peri-implant osseointegration after low-level laser therapy: Micro-computed tomography and resonance frequency analysis in an animal model. Lasers Med Sci. 2016;31:1789–95. doi: 10.1007/s10103-016-2051-3. [DOI] [PubMed] [Google Scholar]
  • 11.Kim YD, Kim SS, Hwang DS, Kim SG, Kwon YH, Shin SH, et al. Effect of low-level laser treatment after installation of dental titanium implant-immunohistochemical study of RANKL, RANK, OPG: An experimental study in rats. Lasers Surg Med. 2007;39:441–50. doi: 10.1002/lsm.20508. [DOI] [PubMed] [Google Scholar]
  • 12.Moher D, Liberati A, Tetzlaff J, Altman DG PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement. Int J Surg. 2010;8:336–41. doi: 10.1016/j.ijsu.2010.02.007. [DOI] [PubMed] [Google Scholar]
  • 13.Higgins JP, Green S. Cochrane Handbook for Systematic Reviews of Interventions. Chichester: Wiley-Blackwell; 2009. [Google Scholar]
  • 14.CvfcSterne JA, Hernán MA, Reeves BC, Savović J, Berkman ND, Viswanathan M, et al. ROBINS-I: A tool for assessing risk of bias in non-randomised studies of interventions. BMJ. 2016;355:i4919. doi: 10.1136/bmj.i4919. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Sleem SS, Zayet MK, El-Ghareeb TI, Saleh HA. Evaluation of the bio-stimulatory effect of platelet rich fibrin augmented by diode LASER compared to platelet rich fibrin alone on dental implant replacing posterior mandibular teeth. Randomised Clinical Trial: Split Mouth Study. Open Access Maced J Med Sci. 2019;7:869–75. doi: 10.3889/oamjms.2019.183. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Flieger R, Gedrange T, Grzech-Leśniak K, Dominiak M, Matys J. Low-level laser therapy with a 635 nm diode laser affects orthodontic mini-implants stability: A randomized clinical split-mouth trial. J Clin Med. 2019;9:1–11. doi: 10.3390/jcm9010112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Marañón-Vásquez GA, Lagravère MO, Borsatto MC, de Souza SS, Watanabe PC, Matsumoto MA, et al. Effect of photobiomodulation on the stability and displacement of orthodontic mini-implants submitted to immediate and delayed loading: A clinical study. Lasers Med Sci. 2019;34:1705–15. doi: 10.1007/s10103-019-02818-0. [DOI] [PubMed] [Google Scholar]
  • 18.Arakeeb MA, Zaky AA, Harhash TA, Salem WS, El-Mofty M. Effect of combined application of growth factors and diode laser bio-stimulation on the osseointegration of dental implants. Open Access Maced J Med Sci. 2019;7:2520–7. doi: 10.3889/oamjms.2019.672. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Abohabib AM, Fayed MM, Labib AH. Effects of low-intensity laser therapy on the stability of orthodontic mini-implants: A randomised controlled clinical trial. J Orthod. 2018;45:149–56. doi: 10.1080/14653125.2018.1481710. [DOI] [PubMed] [Google Scholar]
  • 20.Mikhail FF, El-Din M, Ibrahim T, Zekry K, Nemat A, Nasry S. Effect of laser therapy on the osseointegration of immediately loaded dental implants in patients under vitamin C, Omega-3 and calcium therapy. Open Access Maced J Med Sci. 2018;6:1468–74. doi: 10.3889/oamjms.2018.291. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Awad SM, Mounir RM, Dine E, Salah M, Nasry SA. Effect of laser irradiation on bony implant sites in diabetic patients: A preliminary study. Res J Pharm Biol Chem Sci. 2017;8:1484–95. [Google Scholar]
  • 22.Yanaguizawa MS, Suzuki SS, Martinez EF, Suzuki H, Pelegrin MC, Garcez AS. Effects of low-level laser therapy in orthodontic patients on immediate inflammatory response after mini-implants insertion: A preliminary report. Photomed Laser Surg. 2017;35:57–63. doi: 10.1089/pho.2015.3959. [DOI] [PubMed] [Google Scholar]
  • 23.García-Morales JM, Tortamano-Neto P, Todescan FF, de Andrade JC, Jr, Marotti J, Zezell DM. Stability of dental implants after irradiation with an 830-nm low-level laser: A double-blind randomized clinical study. Lasers Med Sci. 2012;27:703–11. doi: 10.1007/s10103-011-0948-4. [DOI] [PubMed] [Google Scholar]
  • 24.Mandić B, Lazić Z, Marković A, Mandić B, Mandić M, Djinić A, et al. Influence of postoperative low-level laser therapy on the osseointegration of self-tapping implants in the posterior maxilla: A 6-week split-mouth clinical study. Vojnosanit Pregl. 2015;72:233–40. doi: 10.2298/vsp131202075m. [DOI] [PubMed] [Google Scholar]
  • 25.Memarian J, Ketabi M, Amini S. The effect of low-level laser 810 nm and light-emitting diode photobiomodulation (626 nm) on the stability of the implant and inflammatory markers interleukin-1 beta and prostaglandin E2, around implants. Dent Res J. 2018;15:283–8. [PMC free article] [PubMed] [Google Scholar]
  • 26.Torkzaban P, Kasraei S, Torabi S, Farhadian M. Low-level laser therapy with 940 nm diode laser on stability of dental implants: A randomized controlled clinical trial. Lasers Med Sci. 2018;33:287–93. doi: 10.1007/s10103-017-2365-9. [DOI] [PubMed] [Google Scholar]
  • 27.Matys J, Świder K, Grzech-Leśniak K, Dominiak M, Romeo U. Photobiomodulation by a 635nm diode laser on peri-implant bone: Primary and secondary stability and bone density analysis-a randomized clinical trial. Biomed Res Int. 2019;2019:1–8. doi: 10.1155/2019/2785302. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Gokmenoglu C, Ozmeric N, Erguder I, Elgun S. The effect of light-emitting diode photobiomodulation on implant stability and biochemical markers in peri-implant crevicular fluid. Photomed Laser Surg. 2014;32:138–45. doi: 10.1089/pho.2012.3473. [DOI] [PubMed] [Google Scholar]
  • 29.Karaca IR, Ergun G, Ozturk DN. Is Low-level laser therapy and gaseous ozone application effective on osseointegration of immediately loaded implants? Niger J Clin Pract. 2018;21:703–10. doi: 10.4103/njcp.njcp_82_17. [DOI] [PubMed] [Google Scholar]
  • 30.Raghavendra SJ, Dhinakarsamy V. Osseointegration. J Pharm Bioallied Sci. 2015;7:S226–9. doi: 10.4103/0975-7406.155917. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Bengtson AL, Bengtson NG, Bengtson CR, Mendes FM, Pinheiro SR. Histological and radiographic evaluation of the muscle tissue of rats after implantation of bone morphogenic protein (rhBMP-2) in a scaffold of inorganic bone and after stimulation with low-power laser light. Indian J Dent Res. 2010;21:420–4. doi: 10.4103/0970-9290.70816. [DOI] [PubMed] [Google Scholar]
  • 32.Khadra M, Kasem N, Lyngstadaas SP, Haanaes HR, Mustafa K. Laser therapy accelerates initial attachment and subsequent behaviour of human oral fibroblasts cultured on titanium implant material. A scanning electron microscope and histomorphometric analysis. Clin Oral Implants Res. 2005;16:168–75. doi: 10.1111/j.1600-0501.2004.01092.x. [DOI] [PubMed] [Google Scholar]
  • 33.Tang E, Arany P. Photobiomodulation and implants: Implications for dentistry. J Periodontal Implant Sci. 2013;43:262–8. doi: 10.5051/jpis.2013.43.6.262. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Jakse N, Payer M, Tangl S, Berghold A, Kirmeier R, Lorenzoni M. Influence of low-level laser treatment on bone regeneration and osseointegration of dental implants following sinus augmentation. An experimental study on sheep. Clin Oral Implants Res. 2007;18:517–24. doi: 10.1111/j.1600-0501.2007.01369.x. [DOI] [PubMed] [Google Scholar]
  • 35.Dörtbudak O, Haas R, Mailath-Pokorny G. Effect of low-power laser irradiation on bony implant sites. Clin Oral Implants Res. 2002;13:288–92. doi: 10.1034/j.1600-0501.2002.130308.x. [DOI] [PubMed] [Google Scholar]
  • 36.Bashardoust Tajali S, Macdermid JC, Houghton P, Grewal R. Effects of low power laser irradiation on bone healing in animals: A meta-analysis. J Orthop Surg Res. 2010;5:1–11. doi: 10.1186/1749-799X-5-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Hopkins JT, McLoda TA, Seegmiller JG, David Baxter G. Low-level laser therapy facilitates superficial wound healing in humans: A triple-blind, sham-controlled study. J Athletic Training. 2004;39:223. [PMC free article] [PubMed] [Google Scholar]
  • 38.Byrnes KR, Waynant RW, Ilev IK, Wu X, Barna L, Smith K, et al. Light promotes regeneration and functional recovery and alters the immune response after spinal cord injury. Lasers Surg Med. 2005;36:171–85. doi: 10.1002/lsm.20143. [DOI] [PubMed] [Google Scholar]
  • 39.Weber JB. Effect of three different protocols of low-level laser therapy on thyroid hormone production after dental implant placement in an experimental rabbit model. Photomed Laser Surg. 2014;32:11–8. doi: 10.1089/pho.2014.3756. [DOI] [PubMed] [Google Scholar]
  • 40.Yaghobee S, Khorsand A, Rasouli Ghohroudi AA, Sanjari K, Kadkhodazadeh M. Assessment of interleukin-1beta and interleukin-6 in the crevicular fluid around healthy implants, implants with peri-implantitis, and healthy teeth: A cross-sectional study. J Korean Assoc Oral Maxillofac Surg. 2014;40:220–4. doi: 10.5125/jkaoms.2014.40.5.220. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Saudi Journal of Medicine & Medical Sciences are provided here courtesy of Wolters Kluwer -- Medknow Publications

RESOURCES