Abstract
Purpose
To demonstrate the feasibility of a new method for measuring T1 of short T2 species based on an adiabatic inversion recovery-prepared three-dimensional ultrashort echo time Cones (3D IR-UTE-Cones) sequence.
Methods
T1 values for short T2 species were quantified using 3D IR-UTE-Cones data acquired with different repetition times (TRs) and inversion times (TIs). An inversion efficiency factor Q was introduced into the fitting model to accurately calculate T1 values for short T2 species. Experiments were performed on twelve MnCl2 aqueous solution phantoms with a wide range of T1 values and T2* values on a 3T clinical MR system to verify the efficacy of the proposed method. For comparison, a variable flip angle UTE (VFA-UTE) sequence, a variable TR UTE (VTR-UTE) sequence, and a conventional 2D IR fast spin echo (IR-FSE) sequence were also used to quantify T1 values of those phantoms. T1 values were compared between all performed sequences.
Results
The proposed 3D IR-UTE-Cones method provided higher contrast images of short T2 phantoms and measured much shorter T1 values than the 2D IR-FSE, VFA-UTE, and VTR-UTE methods. T1 values as short as 2.95 ms could be measured by the 3D IR-UTE-Cones sequence. The 3D IR-UTE-Cones methods with different TRs were applied to different ranges of T1 measurement, and the scan time was significantly decreased by using 5 TIs along the recovery curves to perform fitting with comparable accuracy.
Conclusion
The 3D IR-UTE-Cones sequence could accurately measure short T1 values while providing high contrast images of short T2 species.
Keywords: adiabatic inversion recovery, UTE-Cones, T1 measurement, MnCl2 phantom
Graphical Abstract

INTRODUCTION
Spin-lattice relaxation time (T1) is an intrinsic MR tissue property. T1 is also an important biomarker for many diseases, thus playing a key role in MR signal intensity and imaging contrast. For example, T1 mapping of the brain can be used for the assessment of multiple sclerosis (MS) and Parkinson’s disease [1–4]. Osteoarthritis (OA) can be detected at an early stage with T1 mapping in delayed gadolinium-enhanced MRI of cartilage (dGEMRIC) [5, 6]. In nanoparticle-based hyperthermia, T1 can be used as a biomarker to evaluate the concentration of iron-oxide nanoparticles (IONPs) [7–10]. In MRI-guided thermal ablation procedures, such as radiofrequency, laser, or focused ultrasound, T1 mapping can be used to monitor the temperature [11–13].
A variety of T1 measurement techniques have been reported, including inversion recovery (IR), Look-Looker (LL), variable flip angles (VFA), and variable repetition time (VTR)-based methods [14–18]. IR is the gold standard for T1 measurement. The combinations of these methods with conventional spin echo (SE), fast spin echo (FSE), or gradient recalled echo (GRE) sequences have been used to measure T1 values of primarily long T2 species [1, 19–23], such as gray matter (GM), white matter (WM), cerebrospinal fluid (CSF), muscle, the superficial layers of articular cartilage, etc. However, for species with short T2s, such as the deep layers of articular cartilage, menisci, tendons, ligaments, bone, and iron overload, their longitudinal magnetizations cannot be accurately rotated as their T2s are around or shorter than the duration of the preparation (e.g., inversion or saturation) pulses [22]. The transverse magnetizations may decay significantly during the preparation pulses. Furthermore, conventional sequences may have too long echo times (TEs) to detect short T2 signals. As a result, accurate T1 mapping cannot be achieved for short T2 species using conventional T1 measurement methods.
Ultrashort echo time (UTE) based-methods are promising alternatives to overcome this problem. For example, the combination of variable flip angle (VFA) and UTE (VFA-UTE), variable TR and UTE (VTR-UTE), and adiabatic inversion recovery and UTE (IR-UTE) methods have been reported to provide accurate measurement of T1 values of short T2 species in the musculoskeletal (MSK) system [22,24–27]. VFA-UTE and VTR-UTE could provide volumetric T1 mapping of short T2 species, but they suffer from high sensitivity to B1 inhomogeneity, with poor image contrast for short T2 species. Adiabatic IR-UTE pulse sequences have been used for high contrast imaging of short T2 species by suppressing signals from long T2 species [27, 28]. The adiabatic IR-UTE method is able to provide not only high contrast morphological imaging but also quantitative evaluation, such as proton density (PD) and T2* and T1 mapping of short T2 species [24, 25]. This is especially important since species with short T2* values typically have short T1 values, which are difficult or impossible to evaluate with conventional clinical sequences.
In this study, we aimed to develop a novel 3D IR-UTE based method to quantitatively measure T1 values of short T2* species. Data were acquired using an adiabatic inversion recovery prepared 3D UTE Cones (3D IR-UTE-Cones) sequence with a series of repetition times (TRs) and inversion times (TIs). An inversion efficiency factor Q was introduced to account for partial inversion of the longitudinal magnetization, and further incorporated into the fitting model for accurate T1 mapping. Experiments were performed on twelve MnCl2 aqueous solution phantoms with a wide range of T1 values and T2* values on a 3T clinical MR system to verify the efficacy of the proposed method.
THEORY and METHOD
Phantom Preparation
As reported in [29], the T2* values of the MnCl2 aqueous solutions had a linear relationship with their concentrations. Based on the linear relationship between concentrations and T2* values, a wide range of concentrations of MnCl2 aqueous solutions were made to explore the shortest T1 values the proposed method could measure.
Twelve 5-mL polypropylene tubes filled with MnCl2 aqueous solutions with concentrations ranging from 0 to 89.16 mM (details in Table 1) were placed inside a cylindrical container (8.5 cm in diameter, 9.5 cm in height) filled with agarose gel (1% by weight). During the scan, the tubes were aligned parallel to the B0 field to minimize susceptibility artifact.
Table 1.
T2* and T1 results of different imaging protocols
| Tube Numbers | Concentration (mM) | T2* (ms) | Estimated T1 values (ms) | ||||
|---|---|---|---|---|---|---|---|
| 3D IR-UTE (TR = 400 ms) | 3D IR-UTE (TR = 50 ms) | VFA-UTE | VTR-UTE | IR-FSE | |||
| 1 | 0 | - | 3133.23±794.74 | - | 2418.00±148.26 | 1433.08±487.12 | 2953.34±16.86 |
| 2 | 0.08 | - | 2639.91±304.17 | - | 1710.69±98.14 | 1337.63±334.19 | 2092.08±3.51 |
| 3 | 0.24 | - | 863.31±20.15 | - | 662.89±32.35 | 695.72±25.30 | 782.66±1.08 |
| 4 | 1.43 | 10.61±0.42 | 158.43±0.48 | 231.34±39.83 | 143.56±71.32 | 159.93±4.03 | 166.52±0.49 |
| 5 | 2.86 | 4.99±0.16 | 76.53±0.14 | 90.09±6.84 | 71.32±14.65 | 82.55±2.42 | 73.13±14.65 |
| 6 | 4.29 | 3.44±0.11 | 51.86±0.15 | 58.29±2.43 | 53.29±3.36 | 60.23±1.97 | 59.33±2.34 |
| 7 | 6.20 | 2.34±0.05 | 36.08±0.09 | 37.56±0.64 | 41.23±3.01 | 46.26±1.96 | - |
| 8 | 11.92 | 1.16±0.03 | 19.09±0.07 | 18.88±0.03 | 29.27±3.95 | 32.33±3.07 | - |
| 9 | 23.84 | 0.55±0.02 | 9.55±0.58 | 9.39±0.01 | - | - | - |
| 10 | 35.76 | 0.35±0.02 | 6.33±0.13 | 6.31±0.01 | - | - | - |
| 11 | 77.72 | 0.16±0.02 | - | 3.17±0.04 | - | - | - |
| 12 | 89.16 | 0.14±0.03 | - | 2.95±0.07 | - | - | - |
Pulse sequence and fitting model
The diagram of the 3D IR-UTE-Cones pulse sequence is illustrated in Figure 1 [30–32]. An adiabatic inversion recovery preparation pulse with a specific TI is followed by N number of spokes (Ns) separated with an equal time interval τ for fast data acquisition (Figure 1a). In this situation, TI is defined as the time from the center of the adiabatic inversion pulse to the center of the spoke cluster. A short rectangular pulse (duration of 26–52 μs) is used for non-selective signal excitation in each spoke (Figure 1b) followed by 3D spiral trajectories with conical view ordering (Figure 1c).
Figure 1.
3D IR-UTE-Cones sequence diagram for T1 measurement. (a) The pulse sequence scheme showing an adiabatic inversion preparation pulse followed by a 3D UTE-Cones data acquisition. (b) In the basic 3D UTE-Cones sequence, a rectangular pulse is used for signal excitation followed by 3D spiral sampling with TE of 32 μs. (c) Conical view ordering of the spiral trajectories. (d) Longitudinal magnetization scheme of three representative tissues (yellow, tissues with ultrashort T2* saturated during the IR pulse; blue, tissues with short T2* partially inverted during the IR pulse; purple, tissues with long T2* fully inverted during the IR pulse) during the course of inversion recovery.
Adiabatic inversion pulses can effectively invert the longitudinal magnetizations of long T2 species, while the longitudinal magnetizations of short T2 species can only be partially inverted because of their fast decay during the relatively long adiabatic inversion process. To accurately describe the inversion status of species with different T2s, we introduced an inversion efficiency factor Q with a range of −1 (full inversion, for species with long T2s, e.g., T2 > 100 ms) to 1 (no disturbance to the longitudinal magnetization, for species with extremely short T2s, e.g., T2 < 1 μs), and Q is 0 when the magnetization is completely saturated (for tissues with ultrashort T2s, e.g., T2 ~ 0.1 ms), as shown in Figure 1d.
Within a steady-state, the longitudinal magnetization of the spoke is as follows:
| (1) |
Where
| (2) |
| (3) |
And
| (4) |
In Equations (2)–(4), c1 = exp(−τ/T1), which is related to the longitudinal T1 relaxation during each spoke τ in the multi-spoke IR-UTE data acquisition; , which is related to the longitudinal T1 relaxation during the whole multi-spokes (Ns-1) τ; and C3 = exp {– [TR – TI – τ(Ns – 1)/2]/T1}, which is related to the longitudinal T1 relaxation during the interval between the last spoke and next adiabatic IR pulse [TR – TI – τ(Ns – 1)/2]. S0 is the steady-state magnetization, τ is the time interval between two spokes, and θ is the flip angle. Sp is the longitudinal magnetization after the adiabatic IR pulse. The explicit derivation of Sp can be found in appendix I.
As the UTE-Cones sequence can achieve echo time (TE) as short as 32 μs, which is much shorter than the T2 of most species. Theoretically, the 3D IR-UTE-Cones sequence can detect signals from short T2 species before they decay to near zero. According to Equations (1)–(4), when TR is kept constant and TIs are set incrementally along the recovery curve of the longitudinal magnetization, T1 can be calculated by nonlinear fitting of equations (1)–(4).
Imaging Parameters
All sequences were implemented on a 3T clinical MRI scanner (MR750, GE Healthcare Technology, Milwaukee, WI) with maximum gradient amplitude of 50 mT/m and 200 T/m/s slew rate. An eight-channel transmit/receive knee coil was used for RF transmission and signal reception.
The 3D IR-UTE-Cones sequence used an adiabatic inversion pulse (Silver-Hoult with a pulse duration of 6.048 ms, a bandwidth of 1.643 kHz, and a maximum B1 amplitude of 17 μT) for robust inversion. The duration of the hard RF pulse is 30 μs. As T1 values of the phantoms distributed on a wide range, two TRs of 50 ms and 400 ms were used to compare the effects of different TRs. For each TR, we used a series of TR/TI combinations (for TR = 50 ms, TI = 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 20, 24, 28, 32, 36, 40 ms; for TR = 400 ms, TI = 13, 15, 18, 25, 30, 40, 50, 70, 100, 130, 160, 200, 240, 280, 320, 360 ms). Other parameters were the same: TE = 32 μs; flip angle (FA) = 10°; field of view (FOV) = 120 × 120 × 128 mm3; acquisition matrix size = 160 × 160 × 32; readout bandwidth = ± 125 kHz. Total scan times were about 145 min and 362 min, respectively.
For the purpose of comparison, the commonly used 2D IR-FSE sequence and two previously reported T1 measurement methods (18, 21), UTE Cones-based variable flip angle (VFA-UTE) sequence and variable repetition time (VTR-UTE) sequence, were performed as well. To minimize the signal loss caused by fast T2* decay, the shortest TE achievable for each sequence was selected. The parameters for 2D IR-FSE were: TE/TR = 6.52/4000 ms; FOV = 120 × 120 mm2; slice thickness = 5 mm; matrix size = 192 × 192; readout bandwidth = ± 62.5 kHz; 13 TIs (TI = 50, 75, 100, 200, 300, 400, 600, 900, 1500, 2000, 2500, 3000, 3500 ms) were utilized for single slice imaging. The total scan time was about 30 min. Since both VFA-UTE and VTR-UTE are sensitive to the inhomogeneity of the transmitted B1 field, the 3D dual TR UTE-Cones sequence was employed for actual flip angle imaging (AFI) and therefore B1 field mapping. Parameters for AFI were: TE = 32 μs; TR = 20/100 ms; FOV = 120 × 120 × 128 mm3; matrix size = 160 × 160 × 32; FA = 45°; bandwidth = ± 125 kHz; scan time was about 20 min. Parameters for VFA were: TE = 32 μs; TR = 20 ms; FOV = 120 × 120 × 128 mm3; matrix size = 160 × 160 × 32; FA = 4°, 8°, 12°, 16°, 20°, 24°, 28°, 32°, 36°; bandwidth = ± 125 kHz; scan time was about 31 min. VTR-UTE used similar parameters except for the following changes: TR = 20, 40, 60, 80 ms; FA = 45°; scan time was about 34 min.
Furthermore, in order to estimate the range of T2* values of phantoms, conventional 3D UTE-Cones acquisitions with 15 different TEs were executed. Parameters for T2* measurement were: TE = 0.032, 0.2, 0.4, 0.8, 1.4, 2.2, 4.4, 6.6, 8.8, 11, 22, 33, 44, 66, 88 ms; TR = 100 ms; FOV = 120 × 120 × 128 mm3; matrix size = 160 × 160 × 32; FA = 20°; bandwidth = ± 125 kHz; scan time was about 86 min.
Data Analysis
The data analysis algorithm was written in MATLAB (MathWorks, Natick, MA). Levenberg-Marquard method was used for nonlinear least-squares curve fitting and executed offline on DICOM images obtained by the protocols described above. Regions of interest (ROIs) were placed in the center of each tube to avoid susceptibility artifacts near the interface. ROIs were then copied automatically to the corresponding site for all the images performed with different TIs at each TR. T1 was calculated by nonlinear fitting of equations (1)–(4). The mean and standard deviation of T1 were calculated for each tube. Linear regression analysis was performed between T1 and MnCl2 concentration for each method. The total scan time for the 3D IR-UTE-Cones method can be reduced by using a smaller number of TIs for each TR. We investigated the accuracy in T1 estimation by using a smaller number of TIs, e.g., 5 vs. 16 or 17 TIs. Percent errors were calculated for each phantom by dividing the difference between the two T1 values and the T1 with more TIs, which was considered the reference standard.
RESULTS
Figure 2 displays representative images acquired with the 3D IR-UTE-Cones and 2D IR-FSE sequences with different TR/TI combinations. Tubes with higher MnCl2 concentrations are “visible” with the 3D IR-UTE-Cones sequence but “invisible” with the 2D IR-FSE sequence with most TIs and TRs. In contrast, tubes with lower MnCl2 concentrations (e.g., <= 0.24 mM) are more “visible” with the 2D IR-FSE sequence, and less “visible” with the 3D IR-UTE-Cones sequence, especially when a shorter TR of 50 ms is used. Tubes 7–12 were dark for all TIs with the 2D IR-FSE sequence, largely because of the high MnCl2 concentrations and thus very short T2* values (i.e., T2*s were shorter than the TE of the 2D IR-FSE sequence). The 3D IR-UTE-Cones sequence captured signals of tubes with short T2*s (tubes 7–12), which were undetectable (dark at any TI) with the 2D IR-FSE sequence. As a result, T1 values of those tubes cannot be measured with the conventional 2D IR-FSE sequence but can be measured with the 3D IR-UTE-Cones sequence. Meanwhile, T1 values of tubes with very low concentrations of MnCl2 (e.g., tubes 1–3) can be measured with the 2D IR-FSE sequence but cannot be reliably measured with the 3D IR-UTE-Cones sequence, especially when a short TR (e.g., 50 ms) is used. The adiabatic IR pulse together with a short TR leads to efficient long T2 suppression (27), precluding accurate quantification of T1s for phantoms with low concentrations of MnCl2.
Figure 2.
Images acquired with 3D IR-UTE-Cones and 2D IR-FSE sequences. The first and middle row are representative images acquired with the 3D IR-UTE protocol at different TR/TI combinations: TR/TI = 50/8, 16, 24, and 36 ms (first row), TR/TI = 400/13, 50, 200, and 360 ms (middle row). The bottom row shows the images acquired with the 2D IR-FSE protocol at TR/TI = 4000/50, 300, 1500, and 3500 ms. The numbers of the tubes are labeled on the first image of the middle row.
The averaged T1 and T2* values, as well as standard deviations of each tube calculated from different protocols, are listed in table 1. The dashes in table 1 indicate data using those specific settings were not suitable for fitting (too big fitting errors due to inappropriate imaging parameters, for example, too narrow range of TIs/TEs to measure long T1 values/T2* values). The 3D IR-UTE-Cones method could measure T1 values as short as 2.95 ms, with corresponding T2* values as short as 0.14 ms, which was unachievable with the other three methods.
Figure 3 shows the fitting curves of each tube with the 3D IR-UTE-Cones method. Two different TRs of 50 ms and 400 ms were investigated. Excellent fitting was achieved for tubes 4–12 with a TR of 50 ms, and tubes 1–10 with a TR of 400 ms. Less accurate fittings were achieved for tubes with lower MnCl2 concentrations, and thus longer T2* values, when a shorter TR of 50 ms and a narrower range of TIs of 8–40 ms were used. On the other hand, increased fitting errors were observed for tubes with high MnCl2 concentrations when a longer TR of 400 ms and a broader range of TIs of 13–360 ms were used. A shorter TR and narrower range of TIs are required for a more accurate estimation of extremely short T1 values.
Figure 3.
The fitting curves of each tube under the 3D IR-UTE-Cones method. (a) The fitting curves of the TR = 50 ms set. (b) The fitting curves of the TR = 400 ms set.
Figure 4 presents the linear fittings between the MnCl2 concentration and the relaxivity rate R1 (1/T1) calculated by the 3D IR-UTE-Cones method. The R1 values show strong correlations (R equaled to 0.99904 and 0.99995 for TR of 50 ms and 400 ms respectively) with the MnCl2 concentrations, suggesting that the 3D IR-UTE-Cones method could reliably measure T1 values, especially short T1 values for phantoms with high MnCl2 concentrations and thus short T2* values.
Figure 4.
The correlations between the R1 values calculated using the 3D IR-UTE-Cones method and MnCl2 solution concentrations. (a) The correlation between R1 values from TR = 50 ms set (tube 4–12) and concentrations. (b) The correlation between R1 values from TR = 400 ms set (tube 1–10) and concentrations.
Figure 5 shows nearly perfect correlations between MnCl2 concentrations and R1 values calculated from different methods, and correlations between the 3D IR-UTE-Cones and other methods (i.e., VTR-UTE, VFA-UTE, and IR-FSE) in terms of T1 values. Those results demonstrate that the 3D IR-UTE-Cones method could accurately measure a broader range of T1 values than the other three methods.
Figure 5.
Correlations between R1s obtained from different methods and MnCl2 solution concentrations and the correlations between the 3D IR-UTE-Cones method and other methods. (a) The correlations between R1s obtained from 3D IR-UTE-Cones, VTR-UTE, VFA-UTE, and IR-FSE and MnCl2 solution concentrations. (b) The enlargement of the region marked with a yellow dashed line in (a). (c) The correlations between T1 values calculated by 3D IR-UTE-Cones and VTR-UTE, VFA-UTE, and 2D IR-FSE.
Table 2 shows the mean value and standard deviation of T1 values calculated by fitting with 5 TIs at both TRs, and the T1 percent errors between T1 values calculated with 5 TIs and 16 TIs (when TR was 400 ms) or 17 TIs (when TR was 50 ms). Most of the percent errors were less than 5%, except for tubes 1 and 2 when TR was 400 ms (TR was too short to accurately calculate long T1 values), tube 10 when TR was 400 ms (TR was too long to accurately calculate short T1 values), and tubes 1–4 when TR was 50 ms (TR was too short to accurately calculate long T1 values). The results demonstrated that five TIs were enough to calculate T1 values accurately with the 3D IRUTE-Cones method when the parameters were appropriately set.
Table 2.
T1 values calculated by fitting with five TIs under 3D IR-UTE-Cones method.
| Tube Numbers | TR = 400 ms | TR = 50 ms | ||
|---|---|---|---|---|
| T1_5 (ms)a | Percent Error (%)b | T1_5 (ms)a | Percent Error (%)b | |
| 1 | 5252.08±1423.90 | 67.63% | - | - |
| 2 | 3173±628.66 | 20.20% | - | - |
| 3 | 847.94±57.16 | 1.78% | - | - |
| 4 | 159.07±2.23 | 0.40% | 181.22±1.74 | 21.67% |
| 5 | 76.59±0.31 | 0.08% | 87.6±3.19 | 2.76% |
| 6 | 51.87±0.33 | 0.02% | 55.53±0.22 | 4.73% |
| 7 | 37.03±0.05 | 2.63% | 37.21±0.88 | 0.93% |
| 8 | 19.10±2.87 | 0.05% | 18.98±0.03 | 0.53% |
| 9 | 9.68±0.55 | 1.36% | 9.41±0.01 | 0.21% |
| 10 | 6.86±2.26 | 8.37% | 6.32±0.01 | 0.16% |
| 11 | - | - | 3.16±0.10 | 0.32% |
| 12 | - | - | 2.99±0.24 | 1.36% |
T1_5 represents the T1 values calculated with five TIs at both TRs (TR/TI = 50 / 8, 12, 16, 24, 40 ms; TR/TI = 400 / 13, 40, 130, 240, 360 ms ).
DISCUSSION
This phantom study demonstrated that the 3D IR-UTE-Cones method can be used as a valid method for T1 measurement, especially for species with short T2* values and short T1 values. For phantoms with relatively long T2* values and T1 values, all four methods, including 2D IR-FSE, 3D VFA-UTE, 3D VTR-UTE, and 3D IR-UTE-Cones methods, provided similar results where excellent linear correlations were observed between T1 relaxation rates (R1s) and MnCl2 concentrations. For phantoms with short T2* values and short T1 values, the 2D IR-FSE method was incapable of either imaging or T1 quantification. Compared with 3D VFA-UTE and VTR-UTE methods, the 3D IR-UTE-Cones method produced a more accurate measurement of shorter T1 values (phantoms with higher concentrations of MnCl2). For the 3D IR-UTE-Cones method, the scan time was shortened by fitting with fewer TIs at an acceptable error (i.e., less than 5% error). However, the 3D IR-UTE-Cones method was not suitable for measuring long T1 values, unless a very long TR was used at the cost of scan time. The method worked best for measuring short or extremely short T1 values of short T2 species.
Benefitting from the combination of the inversion recovery method and the 3D UTE-Cones acquisition scheme, the proposed 3D IR-UTE-Cones method provided both excellent imaging contrast and accurate measurement of T1 values for phantoms with short T2* values. This was a significant advantage over the clinical 2D IR-FSE method, as well as the new 3D VFA-UTE and VTR-UTE methods. The echo time of the 3D UTE-Cones sequence is as short as 32 μs, which was short enough to capture signals from short T2* phantoms (e.g., T2* of 0.14 ms for tube 12), as shown in Table 1. On the other hand, for the clinical 2D IR-FSE sequence, the shortest echo time was 6.52 ms, during which the signals from tubes 6 to 12 had completely decayed. Previously reported 3D UTE-Cones-based VFA and VTR methods have drawbacks, such as high sensitivity to the B1 field inhomogeneity [26] and poor image contrast for short T2 species. Both VFA and VTR methods need to be combined with AFI to do B1 correction, but one of the fundamental assumptions which general AFI relies on is that the two TRs (a shorter TR1 and a longer TR2) are much shorter than the T1 [33]. TRs for the AFI-VFA-UTE-Cones and AFI-VTR-UTE-Cones protocols could not satisfy the condition TR << T1 values when the T1 values were very short. Consequently, the ultrashort T1 values of tube 9 to 11 were unable to be measured with these two methods. Furthermore, the fitting model of the proposed 3D IR-UTE-Cones method introduced an inversion recovery efficiency factor Q, which describes the inversion status of tissues with different T2* values more precisely than conventional SE- and GRE-based IR methods, making the measured short T1 values of short T2* species more accurate.
The 3D IR-UTE-Cones method also provided high contrast images for short T2 species, another significant advantage over the other three methods. In 3D IR-UTE-Cones imaging, long T2 and long T1 species were excellently suppressed, especially when a shorter TR was used, providing excellent contrast for short T2 species. The short T2 species appear with much reduced image contrast with both VFA-UTE and VTR-UTE due to the high signals from species with longer T1 and T2 values. The advantage of the 3D IR-UTE-Cones method in terms of high contrast imaging is even more obvious for short T2 species, such as cortical bone, as these species are nearly “invisible” to both VFA-UTE and VTR-UTE imaging due to their low proton density and short T2*, resulting in much lower signal than the surrounding species which have higher proton densities and longer T2s, such as muscle and marrow fat.
As shown in Table 1 and Figure 3, TR plays an important role in the accurate quantification of T1 relaxation when using the 3D IR-UTE-Cones method. The general rule is that a longer TR is preferred for more accurate quantification of longer T1s, while a shorter TR is preferred for more accurate quantification of shorter T1s. When the T1 is much longer than the TR, the fitting is unreasonable and the resulting T1 value is deemed unacceptable due to the considerable fitting error. The nulling points of the fitting curves of tubes 1–3 should be greater than the curves of tubes 4–12 in Figure 3(a) and the nulling point of the fitting curve of tube 1 should be greater than tubes 2–12 in Figure 3(b), indicating that the TRs of 50 ms and 400 ms were inappropriate for T1 measurements of tubes 1–3 and tube 1, respectively. When the T1 is much shorter than the TR, measurement with a relatively long TR would be inaccurate, as well. As shown in Figure 3(b), the data sample points of tubes 11–12 deviated significantly from the fitting curves, which resulted in large fitting errors and unreasonable fitting results, illustrating that the data acquired with a long TR of 400 ms were not suitable for short T1 calculations. In IR-UTE imaging, long T2 signals will be suppressed more efficiently by using a shorter TR (e.g., 50 ms in this study) (27). Tubes 4–5 have relatively long T2*s of 10.61 and 4.99 ms, respectively, as shown in Table 1. The IR pulse will partially invert their longitudinal magnetizations as the duration of the adiabatic IR pulse is on the order of their T2*s [27, 34]. When a short TR (e.g., 50 ms) is used, tubes 4–5 show as low signal in 3D IR-UTE-Cones imaging, leading to increased errors in T1 quantification. A longer TR (e.g., 400 ms) is expected to provide more accurate T1 quantification for tubes 4–5, consistent with results shown in Tables 1 and 2. A longer T1 means much increased scan time with the 3D IR-UTE-Cones method. Furthermore, tissues with longer T1s may also be imaged and quantified with conventional clinical sequences. And thus the 3D IR-UTE-Cones method is of less importance. On the other hand, tissues with very short T1s, such as iron overload in the liver or spine [35, 36], or hemosiderin in the hemophilia joint [37], cannot be imaged and quantified with conventional clinical sequences. The 3D IR-UTE-Cones sequence with a longer TR (e.g, 400 ms) may not work well when the iron concentration is too high and T2* too short (e.g., T2* < 0.3 ms). In such cases, a shorter TR (e.g., 50 ms) is expected to work better, providing more accurate quantification of T1 relaxation. Consequently, the 3D IR-UTE-Cones method based on short TRs is extremely suitable for short T1 measurements of short T2 species. However, the shortest TR is limited by the acceptable specific absorption ratio (SAR).
With the advantage of being able to measure short T1 values of short T2 species, the proposed 3D IR-UTE-Cones method could be promising in a variety of applications. For instance, as the accumulation of iron shortens both T1 and T2* values of local tissue, T1 could be used as a biomarker of iron overload, both spatially (location) and quantitatively (concentration). In several pathologies, accumulation of iron occurs after frequent blood transfusions or in the case of excess iron absorption [38, 39]. Recently, quantitative susceptibility mapping (QSM) was supposed to quantify the liver iron concentration (LIC), but the QSM and LIC only showed correlation at low LIC levels [35, 40]. The 3D IR-UTE-Cones method can be used to quantify LIC with T1 mapping at high LIC levels. Similarly, the presence of hemorrhage modulates T1 and T2 values. T2 values have previously been used to identify hemorrhage [41]; analogously, T1 mapping could be used as a probe for hemorrhage detection. The proposed 3D IR-UTE-Cones T1 mapping might therefore be a useful method. In the iron oxide nanoparticles (IONPs)-based hyperthermia therapy and stem cell therapy, the 3D IR-UTE-Cones method could be utilized to visualize the locations and concentrations of the IONPs or labeled stem cells with T1 mapping [10, 42]. One of our recent studies showed that the iron accumulation in the cartilage of hemophilic populations decreased T2* values of cartilage [37], presenting another potential application of the proposed 3D IR-UTE-Cones T1 measurement method.
This study has several limitations. First, we have only demonstrated the technical feasibility of the 3D IR-UTE-Cones method in measuring short T1 values of short T2* species by studying MnCl2 aqueous solution phantoms. No ex vivo or in vivo experiments were completed in this study. Second, the efficiency and accuracy of the 3D IR-UTE-Cones protocol in terms of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were not assessed or compared with the other three T1 measurement methods mentioned above. However, it is quite obvious that the 3D IR-UTE-Cones method provides improved CNR for phantoms with high MnCl2 concentrations over 2D IR-FSE, VFA-UTE, and VTR-UTE methods. Third, Nss were set to 1 for both TRs, and the number of TIs along each inversion recovery curve for a TR of 50 ms and 400 ms were set to 17 and 16, respectively. These were the main reason for the long total scan time. Multi-spoke acquisitions (e..g, Ns of more than 30) and fewer TIs could be used in future study to reduce the scan time. Fourth, as there are many potential clinical applications, such as iron quantification in diseases like hemophilia, thalassemia, liver diseases, etc., and T1 mapping for cortical bone, ligaments, tendons, etc., clinical evaluation of this novel technique remains to be done.
CONCLUSION
In this study, we demonstrated the feasibility of using the inversion recovery-prepared 3D ultrashort echo time Cones sequence to accurately measure short T1 values while providing high contrast images of short T2 species.
Supplementary Material
Capable of quantifying the T1 of short T2 species;
Can measure T1 as short as 2.95 ms;
Introduce an inversion efficiency factor Q to accurately measure a wide range of T1.
ACKNOWLEDGEMENTS
The authors achknowledge grant support from the GE Healthcare, NIH (2R01 AR062581, 1R01 NS092650, and 1R01 AR068987) and the scholarship support from the Joint PhD Training Program of University of Chinese Academy of Sciences (UCAS).
Footnotes
Declaration of interests
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
REFERENCE
- [1].Vrenken H, Geurts JJG, Knol DL, Noor Van Dijk L, Dattola V, Jasperse B, Van Schijndel RA, Polman CH, Castelijns JA, Barkhof F, Pouwels PJW, Whole-brain T1 mapping in multiple sclerosis: Global changes of normal-appearing gray and white matter, Radiology. 240 (2006) 811–820. 10.1148/radiol.2403050569. [DOI] [PubMed] [Google Scholar]
- [2].Manfredonia F, Ciccarelli O, Khaleeli Z, Tozer DJ, Sastre-Garriga J, Miller DH, Thompson AJ, Normal-appearing brain T1 relaxation time predicts disability in early primary progressive multiple sclerosis, Arch. Neurol. 64 (2007) 411–415. 10.1001/archneur.64.3.411. [DOI] [PubMed] [Google Scholar]
- [3].Papadopoulos K, Tozer DJ, Fisniku L, Altmann DR, Davies G, Rashid W, Thompson AJ, Miller DH, Chard DT, TI-relaxation time changes over five years in relapsing-remitting multiple sclerosis, Mult. Scler 16 (2010) 427–433. 10.1177/1352458509359924. [DOI] [PubMed] [Google Scholar]
- [4].Baudrexel S, Nürnberger L, Rüb U, Seifried C, Klein JC, Deller T, Steinmetz H, Deichmann R, Hilker R, Quantitative mapping of T1 and T2 discloses nigral and brainstem pathology in early Parkinson’s disease, Neuroimage. 51 (2010) 512–520. 10.1016/j.neuroimage.2010.03.005. [DOI] [PubMed] [Google Scholar]
- [5].Tiderius CJ, Olsson LE, Leander P, Ekberg O, Dahlberg L, Delayed gadolinium-enhanced MRI of cartilage (dGEMRIC) in early knee osteoarthritis, Magn. Reson. Med. 49 (2003) 488–492. 10.1002/mrm.10389. [DOI] [PubMed] [Google Scholar]
- [6].Burstein D, Velyvis J, Scott KT, Stock KW, Kim YJ, Jaramillo D, Boutin RD, Gray ML, Protocol issues for delayed Gd(DTPA)2--enhanced MRI (dGEMRIC) for clinical evaluation of articular cartilage, Magn. Reson. Med. 45 (2001) 36–41. . [DOI] [PubMed] [Google Scholar]
- [7].Mahmoudi K, Bouras A, Bozec D, Ivkov R, Hadjipanayis C, Magnetic hyperthermia therapy for the treatment of glioblastoma: a review of the therapy’s history, efficacy and application in humans, Int. J. Hyperth. 34 (2018) 1316–1328. 10.1080/02656736.2018.1430867. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Zhang J, Chamberlain R, Etheridge M, Idiyatullin D, Corum C, Bischof J, Garwood M, Quantifying iron-oxide nanoparticles at high concentration based on longitudinal relaxation using a three-dimensional SWIFT look-locker sequence, Magn. Reson. Med. 71 (2014) 1982–1988. 10.1002/mrm.25181. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Lv YB, Chandrasekharan P, Li Y, Liu XL, Avila JP, Yang Y, Chuang KH, Liang XJ, Ding J, Magnetic resonance imaging quantification and biodistribution of magnetic nanoparticles using: T1-enhanced contrast, J. Mater. Chem. B. 6 (2018) 1470–1478. 10.1039/c7tb03129g. [DOI] [PubMed] [Google Scholar]
- [10].Vallabani NVS, Singh S, Recent advances and future prospects of iron oxide nanoparticles in biomedicine and diagnostics, 3 Biotech. 8 (2018) 1–23. 10.1007/s13205-018-1286-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Woodrum DA, Kawashima A, Gorny KR, Mynderse LA, Prostate cancer: state of the art imaging and focal treatment, Clin. Radiol. 72 (2017) 665–679. 10.1016/j.crad.2017.02.010. [DOI] [PubMed] [Google Scholar]
- [12].Kuroda K, MR techniques for guiding high-intensity focused ultrasound (HIFU) treatments, J. Magn. Reson. Imaging. 47 (2018) 316–331. 10.1002/jmri.25770. [DOI] [PubMed] [Google Scholar]
- [13].Zhu M, Sun Z, Ng CK, Image-guided thermal ablation with MR-based thermometry, Quant. Imaging Med. Surg. 7 (2017) 356–368. 10.21037/qims.2017.06.06. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Deoni SCL, Rutt BK, Peters TM, Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state, Magn. Reson. Med. 49 (2003) 515–526. 10.1002/mrm.10407. [DOI] [PubMed] [Google Scholar]
- [15].Techawiboonwong A, Song HK, Leonard MB, Wehrli FW, Cortical Bone Water : In Vivo Quantification with Ultrashort Echo-Time MRI Imaging, Radiology. 248 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Cheng H-LM, Wright GA, Rapid High-Resolution T1 mapping by Variable Flip Angles: Accurable and Precise Measurements in the Presence of Radiofrequency field Inhomogeneity, Magn. Reson. Med. 55 (2006) 566–574. 10.1002/mrm.20791. [DOI] [PubMed] [Google Scholar]
- [17].Stikov N, Boudreau M, Levesque IR, Tardif CL, Barral JK, Pike GB, On the accuracy of T1 mapping: Searching for common ground, Magn. Reson. Med. 73 (2015) 514–522. 10.1002/mrm.25135. [DOI] [PubMed] [Google Scholar]
- [18].Chen J, Chang EY, Carl M, Ma Y, Shao H, Chen B, Wu Z, Du J, Measurement of bound and pore water T1 relaxation times in cortical bone using three-dimensional ultrashort echo time cones sequences, Magn. Reson. Med. 77 (2017) 2136–2145. 10.1002/mrm.26292. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Zaharchuk G, Martin AJ, Rosenthal G, Manley GT, Dillon WP, Measurement of cerebrospinal fluid oxygen partial pressure in humans using MRI, Magn. Reson. Med. 54 (2005) 113–121. 10.1002/mrm.20546. [DOI] [PubMed] [Google Scholar]
- [20].Marty B, Carlier PG, Physiological and pathological skeletal muscle T1 changes quantified using a fast inversion-recovery radial NMR imaging sequence, Sci. Rep. 9 (2019) 1–9. 10.1038/s41598-019-43398-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [21].Marty B, Coppa B, Baudin P, Carlier P, Monitoring skeletal muscle chronic fatty degenerations using fast NMR T1-mapping, Neuromuscul. Disord. 27 (2017) S126–S127. 10.1016/j.nmd.2017.06.128. [DOI] [Google Scholar]
- [22].Ma YJ, Zhao W, Wan L, Guo T, Searleman A, Jang H, Chang EY, Du J, Whole knee joint T 1 values measured in vivo at 3T by combined 3D ultrashort echo time cones actual flip angle and variable flip angle methods, Magn. Reson. Med. 81 (2019) 1634–1644. 10.1002/mrm.27510. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [23].Nieminen MT, Rieppo J, Silvennoinen J, Töyräs J, Hakumäki JM, Hyttinen MM, Helminen HJ, Jurvelin JS, Spatial assessment of articular cartilage proteoglycans with Gd-DTPA-enhanced T1 imaging, Magn. Reson. Med. 48 (2002) 640–648. 10.1002/mrm.10273. [DOI] [PubMed] [Google Scholar]
- [24].Du J, Carl M, Bydder M, Takahashi A, Chung CB, Bydder GM, Qualitative and quantitative ultrashort echo time (UTE) imaging of cortical bone, J. Magn. Reson. 207 (2010) 304–311. 10.1016/j.jmr.2010.09.013. [DOI] [PubMed] [Google Scholar]
- [25].Du J, Carl M, Bae WC, Statum S, Chang EY, Bydder GM, Chung CB, Dual inversion recovery ultrashort echo time (DIR-UTE) imaging and quantification of the zone of calcified cartilage (ZCC), Osteoarthr. Cartil. 21 (2013) 77–85. 10.1016/j.joca.2012.09.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Ma YJ, Lu X, Carl M, Zhu Y, Szeverenyi NM, Bydder GM, Chang EY, Du J, Accurate T 1 mapping of short T 2 tissues using a three-dimensional ultrashort echo time cones actual flip angle imaging-variable repetition time (3D UTE-Cones AFI-VTR) method, Magn. Reson. Med. 80 (2018) 598–608. 10.1002/mrm.27066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Ma YJ, Chen Y, Li L, Cai Z, Wei Z, Jerban S, Jang H, Chang EY, Du J, Trabecular bone imaging using a 3D adiabatic inversion recovery prepared ultrashort TE Cones sequence at 3T, Magn. Reson. Med. (2019) 1–12. 10.1002/mrm.28027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [28].Jang H, Wei Z, Wu M, Ma YJ, Chang EY, Corey-Bloom J, Du J, Improved volumetric myelin imaging in human brain using 3D dual echo inversion recovery-prepared UTE with complex echo subtraction, Magn. Reson. Med. (2019) 1–10. 10.1002/mrm.28082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [29].Kurmis AP, Barber C, Slavotinek JP, A MnCl2-based MR Signal Intensity Linear Response Phantom, Radiol. Technol. 79 (2007) 119–125. [PubMed] [Google Scholar]
- [30].Ma YJ, Zhu Y, Lu X, Carl M, Chang EY, Du J, Short T 2 imaging using a 3D double adiabatic inversion recovery prepared ultrashort echo time cones (3D DIR-UTE-Cones) sequence, Magn. Reson. Med. 79 (2018) 2555–2563. 10.1002/mrm.26908. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [31].Carl M, Bydder GM, Du J, UTE imaging with simultaneous water and fat signal suppression using a time-efficient multispoke inversion recovery pulse sequence, Magn. Reson. Med. 76 (2016) 577–582. 10.1002/mrm.25823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [32].Gurney PT, Hargreaves BA, Nishimura DG, Design and analysis of a practical 3D cones trajectory, Magn. Reson. Med. 55 (2006) 575–582. 10.1002/mrm.20796. [DOI] [PubMed] [Google Scholar]
- [33].Yarnykh VL, Actual flip-angle imaging in the pulsed steady state: A method for rapid three-dimensional mapping of the transmitted radiofrequency field, Magn. Reson. Med. 57 (2007) 192–200. 10.1002/mrm.21120. [DOI] [PubMed] [Google Scholar]
- [34].Larson PEZ, Conolly SM, Pauly JM, Nishimura DG, Using adiabatic inversion pulses for long-T2 suppression in ultrashort echo time (UTE) imaging, Magn. Reson. Med. 58 (2007) 952–961. 10.1002/mrm.21341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Lin H, Wei H, He N, Fu C, Cheng S, Shen J, Wang B, Yan X, Liu C, Yan F, Quantitative susceptibility mapping in combination with water-fat separation for simultaneous liver iron and fat fraction quantification, Eur. Radiol. 28 (2018) 3494–3504. 10.1007/s00330-017-5263-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Hanrahan CJ, Shah LM, MRI of spinal bone marrow: Part 2, T1-weighted imaging-based differential diagnosis, Am. J. Roentgenol. 197 (2011) 1309–1321. 10.2214/AJR.11.7420. [DOI] [PubMed] [Google Scholar]
- [37].von Drygalski A, Barnes RFW, Jang H, Ma Y, Wong JH, Berman Z, Du J, Chang EY, Advanced magnetic resonance imaging of cartilage components in haemophilic joints reveals that cartilage hemosiderin correlates with joint deterioration, Haemophilia. 25 (2019) 851–858. 10.1111/hae.13802. [DOI] [PubMed] [Google Scholar]
- [38].St. Pierre TG, Clark PR, Chua-Anusorn W, Measurement and mapping of liver iron concentrations using magnetic resonance imaging, Ann. N. Y. Acad. Sci. 1054 (2005) 379–385. 10.1196/annals.1345.046. [DOI] [PubMed] [Google Scholar]
- [39].Berdoukas V, Chouliaras G, Moraitis P, Zannikos K, Berdoussi E, Ladis V, The efficacy of iron chelator regimes in reducing cardiac and hepatic iron in patients with thalassaemia major: A clinical observational study, J. Cardiovasc. Magn. Reson. 11 (2009). 10.1186/1532-429X-11-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [40].Sharma SD, Fischer R, Schoennagel BP, Nielsen P, Kooijman H, Yamamura J, Adam G, Bannas P, Hernando D, Reeder SB, MRI-based quantitative susceptibility mapping (QSM) and R2* mapping of liver iron overload: Comparison with SQUID-based biomagnetic liver susceptometry, Magn. Reson. Med. 78 (2017) 264–270. 10.1002/mrm.26358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [41].Ghugre NR, Ramanan V, Pop M, Yang Y, Barry J, Qiang B, Connelly KA, Dick AJ, Wright GA, Quantitative tracking of edema, hemorrhage, and microvascular obstruction in subacute myocardial infarction in a porcine model by MRI, Magn. Reson. Med. 66 (2011) 1129–1141. 10.1002/mrm.22855. [DOI] [PubMed] [Google Scholar]
- [42].Haedicke IE, Loai S, Cheng HLM, An efficient T1 contrast agent for labeling and tracking human embryonic stem cells on MRI, Contrast Media Mol. Imaging. 2019. (2019) 1–11. 10.1155/2019/3475786. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.





