Skip to main content
Springer Nature - PMC COVID-19 Collection logoLink to Springer Nature - PMC COVID-19 Collection
. 2020 Jun 27;35(9):2069–2073. doi: 10.1007/s10103-020-03086-z

Resumption of laser/IPL skin services post COVID-19 lockdown—British Medical Laser Association (BMLA) guidance document

Vishal Madan 1,2,
PMCID: PMC7319908  PMID: 32592132

Abstract

The COVID-19 pandemic has had a significant negative impact on the global health economies. As health care resources have been prioritised to cater for patients affected by COVID-19, routine health care services have remained suspended. In an effort to slow the spread of SARS-CoV-2 virus, the UK introduced a country-wide lockdown which came into effect on the 23rd March 2020. Since then, clinics offering laser and intense pulsed light services have remained closed for business. As the rate of COVID-19 infection slows, the UK government has issued guidance for lifting of the lockdown in a cautious and phased manner. With this in view and to facilitate safe resumption of laser and intense pulsed light services, the British Medical Laser Services has issued this guidance document, based on best available and current scientific evidence.

Keywords: Aerosol, Coronavirus, COVID 19, Guidance, Laser, Pandemic, Plume, SARS-CoV-2 virus

Introduction

As strategies to relax the COVID-19 lockdown that commenced in the UK on the 23rd March 2020 are being discussed and implemented, many laser practitioners will seek to resume their services after a period of 10–12 weeks.

This document specifically addresses challenges laser practitioners and clinic managers face while reopening their clinics to offer laser and IPL services, taking into account that the reproduction rate ‘R0 number’ for transmission of COVID-19 is 0.7–1 at the time of issuance of this guidance1.

Several clinics offer a complement of aesthetic treatments of which lasers and IPL form a component. As this document specifically addresses the challenges in resumption of laser/IPL skin services, the reader should refer to general guidance/publications issued by other organisations such as the Joint council for cosmetic practitioners on commencing non-laser treatments and non-dermatological laser/IPL services (https://www.jccp.org.uk/NewsEvent/covid-19-preparing-for-return-to-work)2.

Pre-treatment screening

Proper pre-treatment screening should substantially reduce the risk of contact with symptomatic patient displaying the commonest signs and symptoms of COVID-19 infection such as high temperature, new and persistent cough, loss or change to sense of smell or taste3. Until such time that antibody testing becomes routinely available, all patients attending for laser treatments should be presumed COVID-19 positive.

Excluded from this document are precursory checks and screening of potential patients for signs of COVID-19, general and COVID-19-specific hygiene in clinic reception etc. which are covered in detail elsewhere (https://www.jccp.org.uk/NewsEvent/covid-19-preparing-for-return-to-work)2.

Guidance

  1. Risk assessment: lasers, aerosols and COVID

  2. Risk management

  • A.

    General hygiene

  • B.

    PPE

  • C.

    Ventilation

  • D.

    Smoke evacuation systems

  • E.

    Laser equipment

Risk assessment: lasers, aerosols and COVID

Aerosols are particles of respirable size generated by both human and environmental sources and that can remain viable and airborne for extended periods in indoor air4.

The SARS-CoV-2 virus which is the causative agent of COVID-19 pandemic is thought to spread mainly from person-to-person through respiratory droplets produced when an infected person breathes, coughs, sneezes or talks. Infection with SARS-CoV-2 occurs primarily by inhalation of aerosolised virus or secondarily by contact with droplets and contaminated fomites such as garments, instruments and furniture.

SARS-CoV-2 virus has a size of 0.06–0.14 μm with mean size of 0.1 μm5. During a sneeze or a cough or while talking, “droplet sprays” of virus laden respiratory tract fluid, typically greater than 5 μm in diameter, impact directly on a susceptible individual. Alternatively, a susceptible person can inhale microscopic aerosol particles (droplet nuclei) consisting of the residual solid components of evaporated respiratory droplets, which are tiny enough (< 5 μm) to remain airborne for hours, particularly in enclosed spaces with poor or no ventilation4.

Lasers and intense pulsed light (IPL) treatment of tissues generates plumes and aerosols which include both combustion and non-combustion-generated products including tissue(s), gases, particulate materials, steam and carbonised material (smoke). All ablative and non-ablative laser procedures can generate potentially hazardous plumes.

In a recent study, gas chromatography-mass spectrometry of plume during laser hair removal showed presence of 377 chemical compounds comprising suspected carcinogens and environmental toxins6. Ablative laser-generated plume has been shown to contain intact human papillomavirus DNA, viable bacteriophages and viable human immunodeficiency virus7. Similarly, micron-sized tattoo ink particles potentially contaminated with aerosolised blood products have been detected following laser tattoo removal8.

Results of polymerase chain reaction (PCR) and viral RNA testing for SARS-CoV-2 from blood samples of most patients with COVID-19 infection have been negative and viraemia is very uncommon9. This is an important finding which should reassure practitioners that the risk of viable SARS-CoV-2 in aerosol generated from laser treatments of asymptomatic COVID-19 patients should be very low. As of date, no studies looking at SARS-CoV-2 viral RNA on intact skin or hair follicles or the effect of laser treatments on viral particles have been published. For this reason, all laser treatments should be considered potentially COVID-19 aerosol-generating procedures (AGP) and all necessary precautions should be followed.

Based on the above, and until such time that evidence to the contrary is available, one could assume that the main route of COVID-19 infection in laser/IPL procedures remains patient-generated respiratory aerosol but still considers laser-generated plume/aerosol as potentially infective.

Risk management

General hygiene and enhanced infection control procedures

In any setting, hand hygiene remains the most important defence against spread of COVID-19. Practitioners should also focus on surface decontamination procedures. These form the basis of reduction of virus transmission. Guidance regarding this is freely available2, 3.

PPE—personal protection equipment comprises face masks, gloves, gowns/aprons, face shields and caps

Barrier precautions such as masks and respirators are regarded as the last line of defence against viral transmission secondary to hand washing and other hygiene measures. In the case of laser/IPL treatments, proper eye protection is imperative and should not be ignored. Resources detailing correct methods of donning and doffing of PPE are freely available10.

Face masks and respirators

Face masks protect against aerosol spread from inside out. They are tested in the direction of expiration (from inside to outside). They offer minimal protection to the wearer from inhalation of droplets. Face masks can simply be classified as surgical and non-surgical. Surgical masks worn by the practitioner protect the patient and the environment (air, surfaces, equipment, surgical site). If worn by patients, they prevent contamination of the patients’ surroundings and environment. Standard surgical masks offer no protection to the practitioner undertaking laser procedures. If available, surgical masks or three ply cotton masks should be offered to patients undergoing non-facial laser procedures.

Filtering facepiece respirators (FFP), which are sometimes called disposable respirators protect from aerosol inhalation. FFP are tested in the direction of inspiration (from outside to inside). The tests take into account the efficiency of the filter and leakage to the face. FFP are subject to various regulatory standards around the world. FFP2 and FFP3 conform to EU standard EN149:2001. The FFP3 standard is often considered broadly equivalent to the US N99 standard and Chinese KN99 standard. The FFP2 standard is often considered broadly equivalent to the US N95 standard and Chinese KN95 standard.

Respirators are often more comfortable for the wearer when fitted with a valve exhalation feature, but this feature has the effect of elevating wearer safety over that of patients and others in the vicinity, and therefore is generally discouraged (Table 1)11.

Table 1.

Differences between FFP2 and FFP3 filtering face piece respirators (source; https://www.finder.com/uk/ffp2-vs-ffp3-face-masks] FFP2 FFP3

FFP2 FFP3
Example
  Conforms to EUROPE: EN 149:2001+A1:2009 EUROPE: EN 149:2001+A1:2009
  Minimum filter efficiency requirement 94% 99%
  Filter efficiency tested using Sodium chloride and paraffin oil Sodium chloride and paraffin oil
  Filter efficiency test flow rate 95 l/min 951/min
  Filter efficiency test particle diameter 0.3 μm (approx.) 0.3 μm (approx.)
  Maximum total inward leakage requirement 8% 2%
  Maximum permitted inhalation resistance 0.7 mbar at 301/min 2.4 mbar at 95 l/min 1.Ombar at 30I/min 3.0 mbar at 951/min
  Maximum permitted exhalation resistance 3.0 mbar at 160I/min 3.0 mbar at 1601/min

The World Health Organisation recommends that health care workers should wear a particulate respirator at least as protective as a N95/ FFP2, or equivalent, when performing aerosol-generating procedures on patients suspected or confirmed of being infected with COVID-1912.

The BMLA also recommends that until proven otherwise, all patients should be considered suspected of being infected with COVID-19. All practitioners should wear N95 respirators as a minimum when undertaking any laser or IPL procedures. This should in addition be complemented by a reusable cleanable face shield. For all above-clavicle procedures, where risk of exposure to patient-generated respiratory aerosol is higher, where available, FFP3 respirators should be used instead (Chart 1).

Chart 1.

Chart 1

Recommendations for facial personal protection equipment during laser/IPL procedures

Practitioners should be fit-tested for all respirators and should receive PPE training comprising proper hand hygiene practices, correct fit, donning and doffing to avoid cross-contamination(https://www.hse.gov.uk/respiratory-protective-equipment/fit-testing-basics.htm).

Concerns about availability and costs of respirators should be taken into account. Unlike surgical masks which are single use (3–8 h maximum), FFP can be reusable or disposable. While the BMLA does not recommend extended use or reuse of disposable FFPs, when availability is an issue and, if unavoidable, it may be possible to extend the life of single use FFPs. The use of reusable, cleanable face shields may enhance the life of single-use FFPs. Thorough decontamination and safe storage of FFP are incumbent upon the user and are beyond the scope of this guidance. Practices ranging from UV radiation (260–285 nm), 70 °C dry heat, 70% ethanol and vaporised hydrogen peroxide (VHP) can reduce SARS-CoV-2 on N95 respirator with VHP treatment exhibiting the best combination of rapid inactivation of SARS-CoV-2 and preservation of N95 respirator integrity13, 14.

Ventilation

As the SARS-CoV-2 spreads mainly through aerosol during AGP in enclosed spaces and as particle aerosol (< 10 μm) remains airborne for long durations in the laser treatment room and adjacent rooms, adequate ventilation is important to ensure appropriate air handling, containment and evacuation of contaminated air15.

Standard in-room air cleaners alone are not effective at protecting staff and preventing the spread of COVID-19. A HEPA (high-efficiency particulate air) filter uses mechanical filtration to remove airborne particles greater than or equal to 0.3 μm in diameter at a minimum 99.97% efficiency and such filters are used in vacuum cleaners and in office buildings air management systems. Used alone, they are not adequate for medical practices.

ULPA (ultra-low particulate air) filters offer up to 99.9995% efficiency on particles down to 0.12 μm. ULPA filters in air filtration systems and ductless fume hoods may help drawing in the airborne drops to capture and remove most of them from the airflow.

Where possible, air conditioning units should be serviced and set at exhaust to extract air from the room to outside the building rather than in air circulation modes. If required, air filters should be replaced in line with device manufacturers’ recommendations. Negative-pressure rooms help control the spread of airborne-transmitted infections in health care facilities such as hospitals, but will not be readily available in high street clinics. If available, laser treatments should ideally be undertaken in negative pressure rooms11.

Smoke evacuation systems

Smoke evacuation systems are useful to reduce aerosol and plume generated during laser procedures but should be considered as an adjunct to hand hygiene, PPE and adequate ventilation. Laser smoke evacuation systems should have sub-micron filtration capability. Several smoke evacuation systems exist in the market but all offer certain common features such as ULPA filters and minimum flow rate of 25 cfm (cubic feet per minute) with variable flow rate to accommodate various levels of smoke.

Some devices offer multistage filtration to ensure adequate removal of all contaminants. Charcoal filter comprises activated charcoal which absorbs gas and vapour. It helps in elimination of strong-smelling gases such as those released from laser hair removal. The optimised primary HEPA filter collects over 99.9% of all vaporised tissue and secondary ULPA filter removes solid and biological particles down to 0.01 μm16, 17.

It is important that smoke capture device (e.g. smoke evacuation pencil capture port, tubing) is positioned as close to the surgical site as possible to effectively collect all traces of surgical smoke. It has been shown that when the smoke extraction tip is moved only 2 cm from the treatment area, up to 50% of the particulate matter escaped into the local environment18, 19. Used smoke evacuator filters, tubing and wands must be handled using standard precautions and disposed of as biohazardous waste20.

Smoke evacuation systems should be serviced as per manufacturer’s recommendations to ensure that they function at maximum efficiency.

Laser equipment and treatments

Prior to resuming clinical services, practitioners much endeavour to switch on the lasers to check for any faults that may be addressed in good time. Additionally, as the lockdown has lasted for well over 10 weeks, it would be prudent for practitioners to spend some time for reorientation with laser protocols.

As SARS-CoV-2 can persist on inanimate surfaces such as metal, glass or plastic for up to 9 days, cleaning of laser equipment, display unit, hand pieces, guiding tips, patient goggles and practitioners’ laser eyewear should be meticulously decontaminated after every treatment and as per manufacturer’s recommendation21. SARS-CoV-2 is efficiently inactivated by surface disinfection procedures with 62–71% ethanol, 0.5% hydrogen peroxide or 0.1% sodium hypochlorite within 1 min. Practitioners must refer to manufacturer-issued COSHH advice to ensure safety of use of these products on laser equipment21.

Although COVID-19 is not known to transmit through skin, pre-treatment skin cleansing should be meticulous as per local guidelines to reduce aerosolisation of the virus, if present on treatment site.

Additional measures can be employed to reduce the aerosol generation during laser treatments. To avoid the excessive dispersal of aerosol and plume during laser treatment, wherever possible, positive cold air flow used during laser treatments for the purpose of skin cooling should be substituted by alternative methods such as gels and disposable ice packs, which should in turn be appropriately disposed of as clinical waste and sapphire tips thoroughly cleansed post-procedure. It has been demonstrated that cold sapphire skin cooling with gel suppresses plume during laser hair removal22.

Similarly, use of hydrogel packs or cling film can reduce tissue splatter and aerosolisation during laser tattoo removal23.

Acknowledgements

Mike Murphy-Laser Protection Adviser; Gen. Sec. of the UK Council on Surgical Plumes; Gen. Sec. of the Association of Laser Safety Professionals.

Dr. Godfrey Town Ph.D.- RPA2000 Accredited Laser Protection Adviser, Committee Member IEC TC76 WG4, TC61 WG30/MT16/WG05/WG39 & ANSI SSC-3, GCG Healthcare Ltd.

Jon Exley, Honorary Secretary-BMLA.

Vasant Oswal, Honorary Vice-President-BMLA.

Harry Moseley, Emeritus Professor and Past President-BMLA.

Reem Hana, Executive Committee Member-BMLA.

Tom Lister, Executive Committee Member-BMLA.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

Not applicable.

Declaration of interests

Vishal Madan servers as honorary president of the British Medical Laser Association.

Disclaimer

Dr. Vishal Madan and the British Medical Laser Association (BMLA) make no guarantee or warranty as to the accuracy or completeness of any information published herein. Dr. Vishal Madan and the BMLA disclaim liability for any injury to persons or property or other damages of any nature whatsoever, whether special, indirect, consequential or compensatory, directly or indirectly resulting from the publication, use of, or reliance on this document. In issuing and making this document available, Dr. Vishal Madan and the BMLA are not undertaking to render professional or other services for or on behalf of any person or entity. Nor are Dr. Vishal Madan and the BMLA undertaking to perform any duty owed by any person or entity to someone else. Anyone using this document should rely on his or her own independent judgement or, as appropriate, seek the advice of a competent professional in determining the exercise of reasonable care in any given circumstances. Dr. Vishal Madan and the BMLA have no power, nor do they undertake, to police or enforce compliance with the contents of this document. Nor do Dr. Vishal Madan and the BMLA list, certify, test or inspect products, designs, or installations for compliance with this document. Any certification or statement of compliance with the requirements of this document shall not be attributable to Dr. Vishal Madan and the BMLA and is solely the responsibility of the certifier or maker of the statement.

Footnotes

Consortium (BMLA)

2Jon Exley, Honorary Secretary; 2Vasant Oswal, Honorary Vice-President; 2Harry Moseley, Emeritus Professor and Past President; 2Reem Hana, Executive Committee Member; 2Tom Lister, Executive Committee Member; Vishal Madan, President

Summary

Aerosols can carry viruses—these can be generated by coughs, sneezes and talking.

Laser/IPL procedures may also generate aerosols.

All patients should wear surgical masks (or similar) when in the clinic.

All laser/IPL operators should wear FFP2/N95 or FFP3/N99 respirators during treatments, preferably with face shields.

High efficiency smoke evacuation systems should be used to reduce plume generated during laser procedures.

All equipment and work surfaces should be decontaminated between procedures.

Good air ventilation should be used to clean the air between patients.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References


Articles from Lasers in Medical Science are provided here courtesy of Nature Publishing Group