Abstract
The hollow cylindrical nanohydroxyapatite/polyamide 66 strut (n-HA/PA66) has been used clinically for anterior cervical reconstruction. However, rates of occurrence of a “radiolucent gap” between the dense strut and adjacent endplates were reported. The aim of this in vivo study was to evaluate the viability and advantages of the novel porous n-HA/PA66 strut. The goat C3/4 partial discectomy and fusion model was built, and two groups of n-HA/PA66 struts were implanted into C3/4: group 1, porous n-HA/PA66 strut; and group 2, hollow cylindrical n-HA/PA66 strut filled with autogenous cancellous bone. CT evaluation was performed to assess the fusion status after 12 and 24 weeks. The cervical spines were harvested. Histomorphological analysis was performed to determine new bone formation. Biomechanical testing was performed to determine range of motion (ROM). CT confirmed the disappearance of the boundary of the porous strut and host bone, while the radiolucent gap remained clearly discernible in the dense strut group. The mean CT fusion scores of the porous group were significantly higher. Histologic evaluation showed that the porous struts promoted better osteointegration. Calcein fluorochrome labelling indicated faster bone ingrowth in the porous struts. Biomechanical tests revealed that the porous struts had significantly reduced micromotion. The porous n-HA/PA66 strut could offer interesting potential for cervical reconstruction after corpectomy.
Subject terms: Electron microscopy, Bioinspired materials
Introduction
Anterior cervical discectomy and fusion (ACDF) has been highly successful in treating cervical discogenic diseases associated with radiculopathy or myepathy1. Many struts have been used to reconstruct the stability of the anterior cervical column, including structural autografts, allografts, struts and synthetic substitutes2,3. However, the optimal surgical procedure remains controversial4,5. Autogenous tricortical iliac crest grafts have been reported to cause donor site morbidity in approximately 25% of cases, such as infection, persistent donor site pain and hematoma6. Lower fusion rates, higher rates of breakage, poor biocompatibility, immunologic rejection and the risk of disease transmission greatly limited the use of structural allografts7,8. The complications of subsidence, stress shielding, and radiopacity of titanium mesh struts are still of great concern, hindering them from becoming ideal reconstructing devices9.
Inorganic HA and organic polymers constituted by various composite materials have been explored as bone substitutes over the years2. Anterior cervical reconstruction was performed using a hollow cylindrical nanohydroxyapatite/polyamide 66 strut (n-HA/PA66, Sichuan National Nano Technology Co., Ltd. Chengdu, Sichuan) for many years. The n-HA/PA66 strut is a non-metallic strut device; it is a composite material of nanohydroxyapatite and polyamide 66, and the device is similar to an apatite blend with collagen organisms in natural bone10. Some studies have reported satisfactory clinical outcomes1,2,5,11. However, a previous clinical study recorded the occurrence rate of the “radiolucent gap” between this dense n-HA/PA66 strut with a hollow design and the adjacent endplate at the one-year and last follow-up, with rates of 56% (28/50) and 62% (31/50), respectively2. The subsidence rate was 4% (2/50) at the one-year follow-up, and at the final follow-up, the rate was increased to 8% (4/50). The possible reason may be that the osteoconductivity was insufficient in the n-HA/PA66 composite and the radiographic penetrability of the fibrous tissue exhibited a “radiolucent gap”2.
In order to solve these shortcomings, a novel porous n-HA/PA66 composite has been developed in recent years12–14. Structural 3D bioactive scaffolds with suitable interpenetrating porosity and mechanical stability meet several requirements considered essential for bone regeneration. High porosity and proper pore size may facilitate osteocyte seeding, differentiation, growth, survival and proliferation15. The objective of this study was 2-fold: first, we evaluated the biocompatibility, osteogenesis and biomechanical properties of the porous n-HA/PA66 strut in vivo using a goat cervical discectomy model. The second objective was to explore comparative assessments of the porous n-HA/PA66 strut with the dense hollow cylindrical n-HA/PA66 strut.
Methods and Materials
Ethics statement
The use of animals and the experimental protocols were approved by the Institutional Review Board and the Animal Care Committee of Chongqing Medical University (Approval No. 201213). All methods were carried out in accordance with the approval ethical guidelines.
Materials
The dense n-HA/PA66 strut with hollow cylindrical design and the porous n-HA/PA66 composite strut were designed and fabricated in collaboration with the Chinese Sichuan University Institute of Materials Science and Technology department and our department. The hollow n-HA/PA66 strut has been approved for clinical use since 2005 by the State Drug and Food Administration of China1,2,5,11. The n-HA/PA composites were developed with a bioactive apatite (n-HA) and organic polymer (PA) to mimic natural bone12. The hollow cylindrical strut was designed with a 9-mm outer diameter, a 3-mm inner diameter and a 10-mm length, with several 0.5-mm holes and slots around the outside (Fig. 1A). Each strut had grooves at each end to increase the friction between the strut and vertebra.
The dimensions of the porous n-HA/PA66 struts used in the experiment were 9 mm in diameter and 10 mm in length (Fig. 1B). The porosity of the struts was 80%, the average pore diameter was 300 μm, and the HA content was 40 wt%. The previous results of mechanical testing showed that the strengths of bending, tensing, and compressing were 68, 50 and 95 MPa, respectively, and those strengths were matched with human natural cortical bone.
Animals
Thirty skeletally mature goats (females, age 2-3 years, body weight 20-25 kg, provided by the Experimental Animal Center of Chongqing Medical University) were divided into two groups: one group for the implantation of dense n-HA/PA66 struts with hollow cylindrical design and another for the novel porous n-HA/PA66 composite struts.
Surgical technique and group management
All surgeries were performed by one senior spine surgeon. A right-sided anterior cervical approach was used. Ketamine (10 mg/kg) and diazepam (0.15 mg/kg) were used to induce anaesthesia through intravenous administration, maintained with endotracheal inhalation of 1.5% isoflurane throughout the operation. All animals were placed in the supine position with neck extension. Following the discectomy of the cervical disc between C3 and C4, discectomy was performed in line with the technique of Smith and Robinson16,17. The superior and inferior endplates were elaborately made by high-speed deburring and scraping to remove covered cartilage. The dense n-HA/PA66 strut was filled with autogenous cancellous bone from the resected vertebra, and the strut was implanted into the intervertebral space after corpectomy (Fig. 2A). The appropriately sized porous n-HA/PA66 strut was inserted into the prepared intervertebral space (Fig. 2B). To achieve immediate stabilization, a 3-hole compression plate was used to fix the C3 and C4 vertebral bodies (Fig. 2C). To reduce the perioperative infection risk, the goats received antibiotic prophylaxis (penicillin solution, 800,000 units per day, muscular injection). After surgery, all animals were required to wear a soft cervical collar for approximately four weeks. The diet, activity level, wound healing, and range of motion were assessed for each goat.
Radiographic analysis
Radiographic assessment was completed by two independent observers. We evaluated the presence of a “radiolucent gap” between the n-HA/PA66 struts and their contacted endplates to assess the osteoconductivity and osseointegration. Each group of fifteen samples at 12 weeks and each group of ten samples at 24 weeks underwent computed tomography (CT) scans of C3 and C4 using a 64-detector row CT system (Somatom Sensation 64, Siemens Medical Imaging, Erlangen, Germany) postoperatively. Four sagittal planes were reconstructed with 2-mm slice thickness. The interface between the implant and vertical bone on each plane was divided into 1, 2 and 3 zones for the degree of fusion status (Fig. 3). Fusion was defined by continuous bridging of trabecular bone and the absence of radiolucent lines on CT18,19. All the radiographs were encoded and reviewed in a blinded fashion to assess interbody fusion according to a three-point radiographic score (RS) that was described by van Dijk et al.5,20,21. RS0 indicates pseudoarthrosis; RS1 indicates ingrowth of bone with the strut securely fixed to vertebral bone above and below but with a radiolucent discontinuity in the fusion mass; and RS2 indicates arthrodesis with solid bone bridging the fusion area5. The total points for a section was 24 points5.
Gross and histological observations
Five goats randomly selected from each group were humanely euthanized with diazepam and KCI overdose at 12 and 24 weeks after implantation. Their cervical columns were harvested, including segment C3-C4 with all ligamentous structures. General photography was conducted to observe the interface between implants and vertebral bone. The C3/C4 motion segments were fixed in 10% normal buffered formalin for 2–4 weeks, dehydrated in ascending concentrations of ethanol and cleared in xylene for 4 days, followed by embedding undecalcified in isobornyl methacrylate (Technovit 1200 VLC, Kulzer, Algol, Germany). Specimens were cut into approximately 5-μm-thick sections with a Leica heavy duty sliding microtome (Leica, Wetzlar, Germany), and the sections were stained with haematoxylin and eosin (H&E) and Masson’s trichrome stains22,23. The Olympus B51 virtual microscope with a U-CMAD3 camera attached (Olympus Optical Co. Ltd, Tokyo, Japan) were used to image the histologic sections.
Quantification of newly formed bone
We statistically analysed the histological sections of different implantation periods to quantitatively determine the amount of newly formed bone from the randomly euthanized five goats of each group. Three pieces were randomly chosen from both the hollow cylindrical strut group and porous strut group from every implantation period histological section (12 and 24 weeks). After H&E staining, each section was observed under a light microscope at 100 x magnification, and at least 10 images were randomly obtained from one section. New bone volume (NBV) was expressed as the percentage of newly formed bone area in the implants and analysed by Image-Pro Plus (Media Cybernetics, USA) software.
Calcein labelling observation
Before histological evaluation of bone formation rates on the 7th day, fluorochrome labels (calcein), when bound to calcium ions, can be incorporated at sites of the mineralization front in the form of hydroxyapatite crystals. Calcein injections (C-0875; Sigma-Aldrich Chemie, Germany) were given twice, subcutaneously at 20 mg/kg23. Hard tissue sections were left unstained to study calcein labelling under UV light.
Biomechanical test
Five goats selected from each group were sacrificed at 24 weeks after surgery. The C3/C4 motion segment was isolated from the cervical spine and cleaned of all residual musculature, metallic plates and screws, with care taken to preserve all ligamentous structures. Biomechanical testing was completed within 24 hours. We performed the test as described by Zhou24. A mechanical testing system (MTS) machine (Instron, Model 8874, USA) was used to perform nondestructive biomechanical testing. The vertebrae were mounted in pots using polymethylmethacrylate. The lower pot was rigidly attached to the base of the testing apparatus (Fig. 4). The weight of the upper fixation pot represented the average weight of the goat head, resulting in a compressive preload of 25 N. Pure bending moments induced flexion, extension, left and right lateral bending, and left and right axial rotation. Moments were applied in a quasistatic manner in increments of 1 Nm to a maximum of 6 Nm. Specimens were preconditioned with three cycles of 6-Nm load with a velocity of 1.2 mm/s of the mobile bar. We measured the fourth cycle. The motional analysis system tracked the resultant three-dimensional range of motion (ROM) of each segment.
Statistical analysis
We performed all quantitative experiments three times. The data are expressed as the mean ± s.d., and statistical significance between two groups was determined by one-way analysis of variance and Student’s t-test. The level of significance was determined at P < 0.05.
Results
Macroscopic observation
No goats died in any of the groups. Food intake in each goat gradually improved. All wounds healed to grade A. No postoperative infections, allergic reaction, or material displacement were encountered. At 12 weeks post-implantation, the strut and the vertebral body were closely integrated in both groups. In the dense strut group, fibrous callus formation was observed at the interface, and some gaps between the strut and the host bone were still visible. In the porous strut group, the boundary between the strut and host bone became unclear, the sufficient formation of mature bone tissues was observed, and the bone tissues were ingrown into the pores of the artificial scaffold and bonded tightly with the material (Fig. 5A a and b). At 24 weeks, the interface between the material and host bone formed a close union in both groups. In the dense strut group, more mature callus was detectable. In the porous strut group, new bone regenerated and penetrated through the interconnective pores to the scaffolds (Fig. 5B a and b).
Radiographic analysis
The evaluation of radiographic outcomes included the radiolucent gap and fusion rate. In both groups, the interface between the implants and the host bone was clearly discernible at 12 weeks after the operation (Fig. 6A a and b). At 24 weeks after implantation of the porous n-HA/PA66 strut, the disappearance of the boundary of the porous strut and host bone indicated that new bone had grown into the strut, and the density of newly formed bone was similar to that of host bone. Almost complete osteointegration of the material/bone boundary had been achieved. In the dense control group, the longitudinal hole in the centre of the strut became unidentifiable, which illustrated that more bone tissue grew. However, the radiolucent gap remained clearly discernible (Fig. 6B a and b). CT microradiographic analysis also confirmed the histological study results.
The mean CT fusion scores of each group at 12 and 24 weeks after surgery are shown (Fig. 7). Noticeable differences in the fusion rate were found between the two groups. In comparison with the results of the dense group, those of the porous group were superior (Fig. 7).
Histological observation
All the struts implanted after 12 weeks were encapsulated by fibrous callus. A large quantity of fibrous tissue and osteoblasts were observed along the interface, proving the formation of new bone. Moreover, in the porous n-HA/PA66 strut group, a large proportion of osteoblasts penetrated the interconnective pores, which may accelerate the mineralization and regeneration of bone. The interface between the porous material and the host bone was hardly detectable and formed a close union without any gap (Fig. 8A a and b). In the dense n-HA/PA66 strut group, the neighbouring bone tissue surrounding the strut seemed conspicuous. Few qualities of new bone had ingrown into the strut (Fig. 8A c and d).
At 24 weeks after surgery, more mature trabecular tissues were deposited at the interface of the struts and host bone, and active osteoblasts were present in both groups (Fig. 8B). In the porous group, the sufficient formation of mature osseous tissue crept into the interconnected porosity and connected tightly with the material directly (Fig. 8B a and b).
Quantification of newly formed bone
We performed a quantitative determination of newly formed bone through statistical analysis of histological sections. Figure 9 shows the new bone volume (NBV) at each implantation period. Obviously, at 12 and 24 weeks post-implantation, the amounts of newly formed bone in the porous n-HA/PA66 strut group increased dramatically, much more than that of the dense n-HA/PA66 strut with hollow cylindrical design. The NBV in the inside pores was greater in the porous n-HA/PA66 strutsthan in the dense n-HA/PA66 struts. New bone formation gradually grew in the porous n-HA/PA66 struts. Histological assessment revealed that bone formation was more active in the porous group than in the dense group.
Calcein staining of new bone formation
To further discover the difference in new bone development on the struts, we applied fluorescent calcein green to label the newly formed bone. Both groups exhibited varying amounts of woven bone formation characterized by continuous fluorescent lines at 12 weeks (Fig. 10A a and b). In the 24-week specimens, the diffuse fluorescence areas of calcein green partially decreased (Fig. 10B a and b). Calcein-labelled bones showed that progressive bone remodelling continued at 3 months. A substantial slowing from the 6-month time point was observed from the number of active sites noted on bone sections. For the porous n-HA/PA66 strut group, the fluorochrome bands were wider and loosely mixed. Furthermore, the labelling was not limited to the outer regions of the porous struts adjacent to the strut-bone interface but also appeared multifocal at different pore areas of the implants. However, the fluorescent bands were weakly and sparsely marked and limited to the outer regions of the dense strut group. Calcein fluorochrome labelling (green) of all unstained histologic sections illustrated the dynamic deposition of mineralized bone matrix at the interface and internal pore structure of the porous HA/PA66 strut.
Biomechanical evaluation
The mechanical stability of cervical fusion was evaluated by measuring the relative micromotion of two fused vertebral bodies through the ROM assay. The extent of this motion decrease was associated with the fusion stability. After 24 weeks, the implantation of the porous HA/PA66 strut showed a significantly lower ROM than the dense n-HA/PA66 strut with hollow cylindrical design for all the axial rotation, flexion-extension, and lateral bending testing modes (P < 0.05), demonstrating that the porous HA/PA66 strut produced a much better fusion quality and a higher mechanical stability (Fig. 11).
Discussion
Cervical spine reconstruction is much more difficult after corpectomy than discectomy. Many surgeons prefer the struts, titanium mesh cages (TMCs), packed with cancellous bone for cervical fusion. TMCs can provide effective support and satisfactory fusion without using structural iliac autografts. Donor-site morbidity and the risk of disease transmission associated with allografts have been prevented7,10,25. However, the typical complication, TMC subsidence, has been frequently observed in the early postoperative period1,8–10,26,27. Chen et al. reported that TMC subsidence (more than one millimetre) occurred in 79.7% of patients, and 19% of cases who underwent TMC reconstruction after anterior cervical corpectomy had severe subsidence (more than three millimetres)9. Weber et al. found that 18 patients who underwent primary cervical corpectomy had an average graft subsidence of 2.9 mm, and 11.1% needed a revision program28. It seems that TMCs lead to height loss and segmental alignment, and they also increase the possibility of fusion failure and serious neurological complications. However, a systematic review of the incidence and clinical relevance of TMC subsidence in anterior cervical discectomy and fusion reported that the mean incidence of subsidence was 21% using strut screw combinations, PEEK, titanium, or PMMA struts29. The subsidence rate of titanium struts was 24.9%, and the risk for subsidence using titanium struts also seems lower29. Titanium mesh struts subsidence-related questions remain controversial, but subsidence has resulted in serious consequences, such as buckling of the cervical ligamentum flavum, foraminal stenosis and re-compression of the cervical spinal cord and nerve roots26,30–32.
Anterior reconstruction biomaterials have been developed for better clinical outcomes33. The aim of artificial struts is not only to provide sufficient mechanical strength, great histocompatibility and appropriate elasticity but also to integrate with host bone. Requisite durability and functionality are governed by the properties of biomaterials. The composite, n-HA/PA66, is made by infiltrating nano-HA into PA66. Because it mimics natural bone in that apatite is distributed within a collagen matrix5,34,35, the n-HA/PA66 composite has the mechanical strength of HA and the elastic properties of PA665,13,33. The dense hollow cylindrical n-HA/PA66 strut was designed for use in 2005, and clinical follow-up has gradually been reported in recent years1,2,5,11. Zhao et al. reported a fusion rate of 94.3% in their 35 patients with a hollow cylindrical n-HA/PA66 strut fusion following cervical corpectomy1. Zhang et al. found that using a n-HA/PA66 strut could result in a lower subsidence incidence, better fused segment height maintenance and similar bony fusion as TMCs in 117 patients with multilevel cervical spondylotic myelopathy who underwent anterior cervical corpectomy and fusion5. However, Zhang et al. noted that the incidences of a “radiolucent gap” at the interface between this dense n-HA/PA66 strut with a hollow design and the vertebra were 56% (28/50) at one year and 62% (31/50) at the last follow-up2. He thought the radiographic penetrability of the fibrous tissue exhibited a “radiolucent gap”2. Xiong et al. indicated that it had a high loss rate if the fibrous tissue growth wrapped the interface of the bone implant, which could lead to failure12.
Bone tissue engineering using porous biomimetic matrices coupled with various modifications has been a concern in many studies36. A porous bone substitute provides a three-dimensional interconnected pore structure for cell adhesion, migration and proliferation of new bone tissue37. Our result was in accordance with the finding of Li et al, who showed that the cylindrical n-HA/PA66 scaffold, designed as parallel linear pores with a diameter of 300 μm, was the best choice for bone regeneration and infiltration38. Jie and Li mentioned that the more than 50% porosity of the porous n-HA/PA66 strut had valuable effects on osteocyte adherence and extracellular matrix deposition on its inner surface39. Li et al. stated that the inner porous surface enables cells such as osteoblasts to colonize it40. It also acts as an organizer of vascular canals. The interconnective porous structure increased the transmission of oxygen and nutrients and induced the growth of blood vessels and nerves12,40. Xiong and Wang mentioned that a three-dimensional structure with interconnective pores could promote cell adhesion, differentiation, and proliferation; it also promoted fibrovascular and nerve colonization, which caused the new bone “crawls” into the material to achieve early osteoinduction12,13. The outer layer, which is dense or less porous, can prevent osteoblast ingrowth and mesenchymal stem cell growth into biomaterials. It was suggested that the porous n-HA/PA66 strut with high interpenetrating porosity appeared to promote rapid tissue infiltration and bone tissue regeneration, whereas this process was delayed in the dense strut. Zhong et al. found a common n-HA/PA66 strut subsidence in patients undergoing anterior cervical corpectomy decompression and fusion. They found that postoperative fused segments have height reduction and that height reduction is an independent risk factor for n-HA/PA66 strut subsidence41.
The pore structure theoretically reduces the rigidity of the biomaterials that are close to the elastic modulus of the bone tissue, which ensures that the interface is stable15. The designed increase in pore size, but not the porosity, decreases the mechanical strength. The average pore size of the porous n-HA/PA66 strut material used in the study was 300 µm, and the porosity was 80%. Our mechanical results of the porous n-HA/PA strut showed that the composite has a compressive strength of 95 MPa, which is similar to the lower value of the cortical bone. The porosity rate was 80%, which was an appropriate porosity that satisfied the needs of cell infiltration and proliferation and ingrowth. We found that when the porous n-HA/PA66 struts were implanted, the pores of the struts and host bone gradually integrated into the cross-linked form. The porous HA/PA66 strut presented better biocompatibility and stronger osteogenesis of bone formation than the dense strut. The highly interconnected pore structure may be beneficial for forming mechanical locking and anchoring it with the surrounding bone37. However, Zhong et al. suggested that the n-HA/PA66 strut should be further optimized and that more detailed measurements of the distraction should be performed before clinical applications41. Much scrutiny needs to be executed in future work to avoid poor clinical outcomes.
Conclusion
Compared with the dense n-HA/PA66 strut with a hollow cylindrical design, the porous n-HA/PA66 strut effectively improved intervertebral bony fusion and offers interesting potential for cervical reconstruction after corpectomy.
Acknowledgements
The authors wish to thank Dr. Fajin Lv of the Radiology Department at The First Affiliated Hospital of Chongqing Medical University for his assistance and advice on imaging analysis. This work was supported by research grants from the National Natural Science Foundation of China (NSFC) (Grant numbers 81672167, 81572142, 81672118, and U1601220), the Major Project of Chongqing (cstc2017zdcy-zdyfX0062) and the Science and Health Joint Medical Research Project Fund of Chongqing (2019QNXM016).
Author contributions
Study design: N.H. and X.L. Study conduct: X.L. Drafting and revising of the manuscript: N.H. Data collection: X.G., J.L., S.Y., Q.L., Z.L. and F.L. Data analysis: F.L., Q.L., J.L. and S.Y. Figure preparation: X.G. and Z.Z. Data interpretation: N.H. and X.L. Approval of the final version of the manuscript: N.H., X.L., F.L. and W.H. Funding acquisition: N.H., W.H., X.T. and F.L. X.L. takes responsibility for the integrity of the data analysis.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
These authors contributed equally: Xi Liang and Feilong Li.
References
- 1.Zhao Z, et al. A hollow cylindrical nano-hydroxyapatite/polyamide composite strut for cervical reconstruction after cervical corpectomy. J Clin Neurosci. 2012;19:536–540. doi: 10.1016/j.jocn.2011.05.043. [DOI] [PubMed] [Google Scholar]
- 2.Zhang Y, et al. Long-term results of anterior cervical corpectomy and fusion with nano-hydroxyapatite/polyamide 66 strut for cervical spondylotic myelopathy. Sci Rep. 2016;6:26751. doi: 10.1038/srep26751. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Finkelstein JA, Chapman JR, Mirza S. Anterior cortical allograft in thoracolumbar fractures. Journal of spinal disorders. 1999;12:424–429. doi: 10.1097/00002517-199912050-00013. [DOI] [PubMed] [Google Scholar]
- 4.Lin Q, et al. A comparison of anterior cervical discectomy and corpectomy in patients with multilevel cervical spondylotic myelopathy. Eur Spine J. 2012;21:474–481. doi: 10.1007/s00586-011-1961-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Zhang Y, et al. Evaluation of anterior cervical reconstruction with titanium mesh cages versus nano-hydroxyapatite/polyamide66 cages after 1- or 2-level corpectomy for multilevel cervical spondylotic myelopathy: a retrospective study of 117 patients. PLoS One. 2014;9:e96265. doi: 10.1371/journal.pone.0096265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Siddiqui AA, Jackowski A. Cage versus tricortical graft for cervical interbody fusion. A prospective randomised study. The Journal of bone and joint surgery. British volume. 2003;85:1019–1025. doi: 10.1302/0301-620x.85b7.13398. [DOI] [PubMed] [Google Scholar]
- 7.Tomford WW. Transmission of disease through transplantation of musculoskeletal allografts. The Journal of bone and joint surgery. American volume. 1995;77:1742–1754. doi: 10.2106/00004623-199511000-00017. [DOI] [PubMed] [Google Scholar]
- 8.Yang X, et al. Comparison of anterior cervical fusion by titanium mesh cage versus nano-hydroxyapatite/polyamide cage following single-level corpectomy. Int Orthop. 2013;37:2421–2427. doi: 10.1007/s00264-013-2101-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Chen Y, et al. Subsidence of titanium mesh cage: a study based on 300 cases. Journal of spinal disorders & techniques. 2008;21:489–492. doi: 10.1097/BSD.0b013e318158de22. [DOI] [PubMed] [Google Scholar]
- 10.Yang X, et al. Anterior reconstruction with nano-hydroxyapatite/polyamide-66 cage after thoracic and lumbar corpectomy. Orthopedics. 2012;35:e66–73. doi: 10.3928/01477447-20111122-10. [DOI] [PubMed] [Google Scholar]
- 11.Ou Y, Jiang D, Quan Z, An H, Liu B. Application of artificial vertebral body of biomimetic nano-hydroxyapatite/polyamide 66 composite in anterior surgical treatment of thoracolumbar fractures. Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery. 2007;21:1084–1088. [PubMed] [Google Scholar]
- 12.Xiong Y, et al. Analyzing the behavior of a porous nano-hydroxyapatite/polyamide 66 (n-HA/PA66) composite for healing of bone defects. Int J Nanomedicine. 2014;9:485–494. doi: 10.2147/IJN.S52990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wang H, et al. Biocompatibility and osteogenesis of biomimetic nano-hydroxyapatite/polyamide composite scaffolds for bone tissue engineering. Biomaterials. 2007;28:3338–3348. doi: 10.1016/j.biomaterials.2007.04.014. [DOI] [PubMed] [Google Scholar]
- 14.Hong LiYL. Yonggang Yan, Jidong Li, Aiping Yang Hongzhao Xiang. Fabrication of Porous n-HA/PA66 Composite for Bone Repair. Key Engineering Materials Vols. 2007;330–332:321–324. [Google Scholar]
- 15.Xu Q, et al. Tissue engineering scaffold material of porous nanohydroxyapatite/polyamide 66. Int J Nanomedicine. 2010;5:331–335. doi: 10.2147/ijn.s9869. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Robinson RA. Fusions of the cervical spine. The Journal of bone and joint surgery. American volume. 1959;41-a:1–6. doi: 10.2106/00004623-195941010-00001. [DOI] [PubMed] [Google Scholar]
- 17.Smith GW, Robinson RA. The treatment of certain cervical-spine disorders by anterior removal of the intervertebral disc and interbody fusion. The Journal of bone and joint surgery. American volume. 1958;40-a:607–624. doi: 10.2106/00004623-195840030-00009. [DOI] [PubMed] [Google Scholar]
- 18.Hou Y, et al. Fabrication of an integrated cartilage/bone joint prosthesis and its potential application in joint replacement. J Mech Behav Biomed Mater. 2016;59:265–271. doi: 10.1016/j.jmbbm.2016.02.001. [DOI] [PubMed] [Google Scholar]
- 19.Huang, D. et al. Interfacial and biological properties of the gradient coating on polyamide substrate for bone substitute. J R Soc Interface11, 10.1098/rsif.2014.0101 (2014). [DOI] [PMC free article] [PubMed]
- 20.van Dijk M, Smit TH, Sugihara S, Burger EH, Wuisman PI. The effect of cage stiffness on the rate of lumbar interbody fusion: an in vivo model using poly(l-lactic Acid) and titanium cages. Spine (Phila Pa 1976) 2002;27:682–688. doi: 10.1097/00007632-200204010-00003. [DOI] [PubMed] [Google Scholar]
- 21.van Dijk M, Smit TH, Burger EH, Wuisman PI. Bioabsorbable poly-L-lactic acid cages for lumbar interbody fusion: three-year follow-up radiographic, histologic, and histomorphometric analysis in goats. Spine (Phila Pa 1976) 2002;27:2706–2714. doi: 10.1097/00007632-200212010-00010. [DOI] [PubMed] [Google Scholar]
- 22.Hu N, et al. BMP9-regulated angiogenic signaling plays an important role in the osteogenic differentiation of mesenchymal progenitor cells. Journal of cell science. 2013;126:532–541. doi: 10.1242/jcs.114231. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Hu N, et al. Inhibition of histone deacetylases potentiates BMP9-induced osteogenic signaling in mouse mesenchymal stem cells. Cellular physiology and biochemistry: international journal of experimental cellular physiology, biochemistry, and pharmacology. 2013;32:486–498. doi: 10.1159/000354453. [DOI] [PubMed] [Google Scholar]
- 24.Chunguang Z, et al. Evaluation of bioabsorbable multiamino acid copolymer/alpha-tri-calcium phosphate interbody fusion cages in a goat model. Spine (Phila Pa 1976) 2011;36:E1615–1622. doi: 10.1097/BRS.0b013e318210ca32. [DOI] [PubMed] [Google Scholar]
- 25.Banwart JC, Asher MA, Hassanein RS. Iliac crest bone graft harvest donor site morbidity. A statistical evaluation. Spine (Phila Pa 1976) 1995;20:1055–1060. doi: 10.1097/00007632-199505000-00012. [DOI] [PubMed] [Google Scholar]
- 26.Bilbao G, et al. Surgical results and complications in a series of 71 consecutive cervical spondylotic corpectomies. Acta neurochirurgica. 2010;152:1155–1163. doi: 10.1007/s00701-010-0660-3. [DOI] [PubMed] [Google Scholar]
- 27.Dvorak MF, et al. Effectiveness of titanium mesh cylindrical cages in anterior column reconstruction after thoracic and lumbar vertebral body resection. Spine (Phila Pa 1976) 2003;28:902–908. doi: 10.1097/01.brs.0000058712.88053.13. [DOI] [PubMed] [Google Scholar]
- 28.Weber MH, et al. Graft Subsidence and Revision Rates Following Anterior Cervical Corpectomy: A Clinical Study Comparing Different Interbody Cages. Clinical spine surgery. 2017;30:E1239–e1245. doi: 10.1097/bsd.0000000000000428. [DOI] [PubMed] [Google Scholar]
- 29.Noordhoek I, Koning MT, Jacobs WCH, Vleggeert-Lankamp CLA. Incidence and clinical relevance of cage subsidence in anterior cervical discectomy and fusion: a systematic review. Acta neurochirurgica. 2018;160:873–880. doi: 10.1007/s00701-018-3490-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Hee, H. T., Majd, M. E., Holt, R. T., Whitecloud, T. S., III & Pienkowski, D. Complications of multilevel cervical corpectomies and reconstruction with titanium cages and anterior plating. Journal of spinal disorders & techniques16, 1-8; discussion 8-9, 10.1097/00024720-200302000-00001 (2003). [DOI] [PubMed]
- 31.Lim TH, et al. Effect of endplate conditions and bone mineral density on the compressive strength of the graft-endplate interface in anterior cervical spine fusion. Spine (Phila Pa 1976) 2001;26:951–956. doi: 10.1097/00007632-200104150-00021. [DOI] [PubMed] [Google Scholar]
- 32.Chibbaro S, et al. Anterior cervical corpectomy for cervical spondylotic myelopathy: experience and surgical results in a series of 70 consecutive patients. J Clin Neurosci. 2006;13:233–238. doi: 10.1016/j.jocn.2005.04.011. [DOI] [PubMed] [Google Scholar]
- 33.Yang X, et al. Outcome of single level anterior cervical discectomy and fusion using nano-hydroxyapatite/polyamide-66 cage. Indian J Orthop. 2014;48:152–157. doi: 10.4103/0019-5413.128753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Qiao B, et al. Bone plate composed of a ternary nano-hydroxyapatite/polyamide 66/glass fiber composite: biomechanical properties and biocompatibility. Int J Nanomedicine. 2014;9:1423–1432. doi: 10.2147/IJN.S57353. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Li J, et al. Bone regeneration and infiltration of an anisotropic composite scaffold: an experimental study of rabbit cranial defect repair. J Biomater Sci Polym Ed. 2016;27:327–338. doi: 10.1080/09205063.2015.1128248. [DOI] [PubMed] [Google Scholar]
- 36.Khadka A, et al. Evaluation of hybrid porous biomimetic nano-hydroxyapatite/polyamide 6 and bone marrow-derived stem cell construct in repair of calvarial critical size defect. J Craniofac Surg. 2011;22:1852–1858. doi: 10.1097/SCS.0b013e31822e8274. [DOI] [PubMed] [Google Scholar]
- 37.Zhang JC, et al. The repair of critical-size defects with porous hydroxyapatite/polyamide nanocomposite: an experimental study in rabbit mandibles. Int J Oral Maxillofac Surg. 2010;39:469–477. doi: 10.1016/j.ijom.2010.01.013. [DOI] [PubMed] [Google Scholar]
- 38.Chang BS, et al. Osteoconduction at porous hydroxyapatite with various pore configurations. Biomaterials. 2000;21:1291–1298. doi: 10.1016/s0142-9612(00)00030-2. [DOI] [PubMed] [Google Scholar]
- 39.Wei J, Li Y. Tissue engineering scaffold material of nano-apatite crystals and polyamide composite. European Polymer Journal. 2004;40:509–515. doi: 10.1016/j.eurpolymj.2003.10.028. [DOI] [Google Scholar]
- 40.Li J, et al. In Vitro and In Vivo Evaluation of a nHA/PA66 Composite Membrane for Guided Bone Regeneration. J Biomater Sci Polym Ed. 2011;22:263–275. doi: 10.1163/092050609X12602753096279. [DOI] [PubMed] [Google Scholar]
- 41.Zhong W, et al. Nanohydroxyapatite/polyamide 66 strut subsidence after one-level corpectomy: underlying mechanism and effect on cervical neurological function. Sci Rep. 2018;8:12098. doi: 10.1038/s41598-018-30678-1. [DOI] [PMC free article] [PubMed] [Google Scholar]