Abstract
MRI developed during the last half-century from a very basic concept to an indispensable non-ionising medical imaging technique that has found broad application in diagnostics, therapy control and far beyond. Due to its excellent soft-tissue contrast and the huge variety of accessible tissue- and physiological-parameters, MRI is often preferred to other existing modalities. In the course of its development, MRI underwent many substantial transformations. From the beginning, starting as a proof of concept, much effort was expended to develop the appropriate basic scanning technology and methodology, and to establish the many clinical contrasts (e.g., T1, T2, flow, diffusion, water/fat, etc.) that MRI is famous for today. Beyond that, additional prominent innovations to the field have been parallel imaging and compressed sensing, leading to significant scanning time reductions, and the move towards higher static magnetic field strengths, which led to increased sensitivity and improved image quality. Improvements in workflow and the use of artificial intelligence are among many current trends seen in this field, paving the way for a broad use of MRI. The 125th anniversary of the BJR is a good point to reflect on all these changes and developments and to offer some slightly speculative ideas as to what the future may bring.
Introduction
Up until a century ago, the living human body was almost opaque to the human eye. This changed radically with the advent of medical imaging technologies, which dramatically revolutionised medicine and in particular diagnosis, but in so doing also influenced the arts, culture and indeed modern society.1 Among these emerging technologies, MRI is now firmly established as a key modality. It is safe for the patient from the perspective of radiation exposure and offers a range of excellent soft-tissue image contrasts, covering a huge diagnostic bandwidth that supports modern medical diagnostics, making MRI an indispensable tool in diagnosis and therapy monitoring.
Born in 1973, roughly 50 years ago, MRI started with Lauterbur’s pioneering idea,2 that established a proof of principle imaging prototype. It showed early promise in the ability to differentiate between healthy and tumorous tissue via tissue specific MR relaxation parameters (T1, T2),3 but at that stage had an uncertain future. This promise triggered a number of British research groups (primarily Nottingham, Hammersmith and Aberdeen) to construct first scanner prototypes, which were based on resistive magnets. However, by modern standards, the early image quality was low, and data acquisition speeds excruciatingly slow, acquiring a single echo approximately every second and scanning a few minutes for a single low-resolution image.4 The multi-slice multi-echo spin-echo imaging sequence introduced shortly afterwards greatly improved efficiency.5 Furthermore, it was at the time not clear whether the young MRI branch would ever achieve whole body- or high-field imaging as experienced today.6 From the beginning, the development of MR technology was characterised by a strong interplay between basic research and hardware development. The former established the concept of k-space,7,8 the space in which actual MR data acquisition takes place, and the underlying principles of image formation, mainly founded on Nyquist’s sampling theorem and the Fourier transform, encompassing slice-selective excitation and gradient-/spin-echo-based acquisition techniques. On the hardware side, the basic system architecture was developed, which has remained conceptually largely unchanged, but has evolved enormously in the system components and their details over the last three decades (c.f. Figure 1).
The initial systems were quite simple and consisted of a homogeneous magnet, responsible for creating the measurable polarisation within the body. The inner bore hosts a magnetic field gradient system that facilitates spatial signal encoding by allowing manipulation of a linear magnetic field, the gradient, along all three spatial axes. Within this, the antennas of the transmit and receive systems are placed, allowing excitation of the spin system and subsequent reception of the tiny MR signal responses (Figure 1b).
Since then much has changed. Hardware, methodology and clinical needs have continuously cross-fertilised each other, triggering many waves of scientific innovation, which can be grouped into several general trends:
towards higher magnetic fields, to boost the available polarisation and thus the signal-to-noise ratio (SNR),
to higher acquisition speed, boosting scanning efficiency, because MRI is by its nature a rather slow technique,
towards parallelisation to increase overall system performance (including receivers, transmitters, computer cores, see Figure 1), which enabled a number of key applications,
the quest for new diagnostic contrasts including potential biomarkers, which goes along with the trend to team up with other diagnostic modalities, like PET,9
the move towards therapy where MRI acts as a camera in interventional set-ups like the MR-LINAC,10 and
the trend of specialisation and standardisation in the MR community. Large sub-communities and disciplines cover areas such as spectroscopy, animal research, functional MRI (fMRI) quantitative MR and interventional MRI, to name but few examples.
Although MRI may be approaching maturity, it is, and it will remain, a field of constant innovation. This will be reflected in this brief review, which aims to sketch the trajectory of selected key technical innovations of the last 30 years and to extrapolate these into the future.
From the beginning
MRI actually started with many challenges, as the radial imaging scheme that was initially proposed2 gave insufficient image quality owing to the poor B0 homogeneity of the early systems. The concept of slice selective excitation11 which constrained the signal to a 2D plane and the introduction of Cartesian spin warp imaging4 which uniformly samples k-space in the presence of a constant read-gradient, offered for the first time images of diagnostic quality. The simplicity and the robustness of this imaging scheme allowed also a straight-forward image reconstruction requiring only an inverse fast Fourier transform (FFT) to be applied to the sampled echo signals in k-space to yield the final images. The FFT was often run on dedicated hardware such as array processors, precursors to modern GPUs. Spin-warp is even today the dominant k-space sampling scheme, due to its straight-forward extension to 3D and high robustness to imperfections such as slight eddy currents. This paved the way for initial, clinical adoption using simple scanners, which became commercially available in the early 80s, characterised by limited speed and sequence flexibility, and offering only a small number of image contrasts (T1 and T2). They were operating at very low field strength,12 with the drawback of limited SNR and limited contrast.
Magnets
The SNR scales supra linearly with B0 (13) and this is one of the major driving forces behind the trend to increase field strength. Superconducting magnets were an essential innovation on the road towards this goal. They are more stable against field drifts, but were technologically more demanding, requiring superconducting wire maintained at liquid helium temperature by a combination of electromagnetics, vacuum physics and refrigeration technology. These initially bulky, very heavy, claustrophobia-inducing magnets became within two decades very compact and patient-friendly, with wider bores and a more open design. The first clinical 1.5T machines (60 cm bore) were in use by the mid 1980s and became 'wide bore' (70 cm) in the early 2000s.
Modern magnets are self-shielded, to reduce the fringe field to ease siting. However, this further increases the gradient of the static magnetic field close to the magnet, which increases the safety risk when ferro-magnetic objects are brought close to the magnet, making defined safety procedures for staff and patients necessary to reduce risk of injury. Magnets, furthermore, became significantly lighter, so that siting, even in the upper floors of hospitals is often no longer an issue. Although the external appearance of magnets has not changed very much, inside they are continuously being improved. Recently, ‘cryo-free’ magnet technology became available, allowing magnets with less than 10 litres liquid helium (a huge reduction from the previous value of about 1000 litres). This lowers costs in the unlikely case of a quench and further eases siting constraints because the quench line will soon be obsolete.
Meanwhile researchers are considering exploring the real human high-field limits with field strengths up to 14T currently being planned. Such systems are at present completely beyond clinical applicability and will probably remain forever in the research realm, because of their extremely high price, limited magnet production capacity and complex operation, including the management of unwanted physiological side effects. However, there is also a thriving research community operating at 7T and it will be interesting to see how this makes the transition to the clinic in the coming years, because some applications recently received approval from the US Food and Drug Administration (FDA). As anticipated in the early years of MRI,6 RF wave propagation effects can seriously compromise high-field image quality and corresponding contrast. Parallel transmission,14,15 schematically illustrated in Figure 1d, can mitigate those adverse effects, supporting RF shimming, which homogenises the RF transmit field, paving the way for high-quality 3T whole body-16 or ultra-high-field head/torso imaging.
Despite this trend to pursue increased sensitivity by means of increasing the static magnetic field strength, there has recently been a small counter trend, to reconsider the use of lower field strengths.17 One potential driver is the option to lower system costs, in some part also due to the lower interaction between patient and RF transmit fields, eliminating the need for parallel transmission technology, to mitigate wave propagation effects at high fields.15–17 In addition, there is also the potential to reduce risk especially in interventional applications.18
Gradients
The room temperature bore of the magnet contains the gradient coil that performs the primary spatial encoding. The performance of the entire gradient system (c.f. Figure 1b) substantially influences the resulting image quality and has significantly improved with the introduction of self-shielded gradient coils19–21 and the use of gradient pre-emphasis systems.22 However, to speed-up data acquisition, higher sampling rates are necessary, which demands higher imaging gradient strength. The vendors answered this need by offering gradient systems with higher specifications (strength and slew rate), which often made a redesign of the gradient chain necessary, including the gradient coil interior,23 the active cooling system and also the power amplifier operating principle, which moved to fully digitised switched mode devices. Today’s gradient systems are high-tech/high-power electronic devices, running often at peak power levels close to a 1 MW.24 This trend was compounded by the recent advent of wide bore systems, because the stored energy and thus the power requirements for a gradient coil increase with the fifth power of the diameter, and hence more than factor of 2 when the diameter just increases from 60 to 70cm.25 Vendors have coped with these constraints by making compromises on gradient linearity, which have their main effect far from the iso-centre. However, the simple notion that higher gradient strengths offer higher speed is often not true anymore, because the maximum slew rate in imaging is often constrained by peripheral nerve stimulation (PNS) limits,26 which get more critical in wide bore environments. Currently, there are some initial efforts to directly include PNS aspects as an additional constraint in the general coil design process27 or to build dedicated high-end gradient coils restricting their effective fields only to small portions of the body like the head and thus minimising PNS.28 It is interesting to note here that one of the main drivers of increased gradient strengths is the use of pulsed-field gradients to generate diffusion contrast. Higher gradients reduce the minimum echo time (TE) and thus reduce signal loss caused by T2 decay.29
Basic image acquisition
Better gradients and faster spectrometers have enabled improved scanner performance, with shorter TRs, TEs and echo spacing and higher sampling rates—supporting modern acquisition techniques. This supported the development of accelerated multi-echo RF-refocused fast spin echo imaging,30 which reduced the time associated with T2 weighted acquisitions considerably. Similarly, for fast gradient echo imaging,31 technological improvements have led to faster acquisitions of strongly T1 weighted images.
Improved system performance32 allowed the implementation of single shot Echo Planar Imaging33 (EPI), for the first time breaking the dogma of sampling only in the presence of a constant read-out gradient. EPI has been the fastest imaging scheme available since the early 90s and is still the workhorse in many demanding applications: like fMRI, diffusion and perfusion imaging. This development represented in essence the first wave in terms of accelerating MRI. It was driven by the wish to shorten total examination time and to enhance patient throughput but also to reduce motion artefacts which increase with total scan time and or breath-hold duration. To further enhance the scope of these new sequences, numerous magnetisation preparation experiments34,35 were designed to benefit from the new and more efficient gradient-echo and multi-spin-echo sampling schemes of the early 90s.
From the early 1990s, the improved main field homogeneity of magnets made possible the exploration of advanced alternative non-Cartesian k-space sampling schemes including spirals,36,37 radial38 and also hybrid schemes, mixing radial and Cartesian,39 see Figure 2. These demanded non-uniform FFT or efficient regridding routines and often required dedicated field map field and concomitant field40 gradient-based data corrections to cope with their increased off-resonance sensitivity.41 Techniques like radial scanning gave an impetus to auto-correction42 and auto-navigation,43 which were developed to their full in the radial–Cartesian hybrid approach.39
Contrast
Many clinical applications were developed for almost all anatomies throughout the body. In the beginning, there was a focus on the brain and the central nervous system, followed by musculo-skeletal applications including extremities and joints. These organs were the easiest to image as they are relatively easy to immobilise for the duration of an MRI scan. As techniques for motion correction improved, body applications including spine, liver, lung, heart and pelvis became more common. Today there is almost no body part, which is not covered by MRI. Static diagnostic imaging was initially the primary focus, reflecting anatomy or different stages of pathology, but also dynamic/fluoroscopic or real-time imaging approaches44,45 were developed to capture temporal changes such as heart motion or to utilise contrast media in the circulation to form MR angiograms.
The richness of the potential contrasts MRI can create is immense. Contrasts can be based on the Bloch equations, mainly driven by T1 and T2, the chemical shift effect, the chemical exchange of the water protons/magnetisation transfer, local magnetic susceptibility changes, but also on physiology-driven effects like flow, perfusion and diffusion. Thus, T1 weighted and T2 weighted imaging were performed using the fast gradient and turbo spin echo sequences. New sequences were designed like fluid-attenuated inversion recovery,46 which can visualise abnormalities caused by subtle pathologies in the brain. Unlike in CT, where fat gives only a small signal, in MRI fat signal is often seen as a serious confounding factor. Therefore, fat suppression, performed by pre-saturation,47 selective excitation, or by chemical shift encoding48 has become a valuable tool to increase image contrast. However, the modern chemical shift encoding approaches49 allow the separation of water and fat signals, leveraging also the information present in the fat to better understand obesity, the metabolic syndrome and to derive biomarkers in fat-related liver diseases.50
Another important contrast is diffusion, because it allows us to probe the tissue microstructure at a scale that is orders of magnitude below the actual image resolution.51 The diffusion-weighted signal is read out preferably by single shot EPI32 or TSE variants.35 Diffusion helps to characterise tissue and tumours in type, state and grade non-invasively.51 Applying different diffusion sensitisations (b-values) allows the calculation of an apparent diffusion coefficient (ADC) that can act as a bio-maker for stroke,52 whereas probing diffusion in different spatial directions facilitates the estimation of the diffusion tensor that can map fibre orientations in the white matter of the brain53 or in muscle. For brain imaging, the tensor formalism has largely been replaced by more sophisticated analysis techniques54 that can interrogate multiple fibre orientations within a voxel. These require high angular resolution data acquisition, sometimes at multiple b-values. Knowledge of fibre distribution makes it possible to identify the major nerve bundles within the brain, which can be important for surgical planning, and to estimate the connectivity between grey matter regions, which is of great interest for neuroscientists.54–56Another extension of DWI is to examine the deviation from Gaussian diffusion, by using kurtosis imaging, which has proven to be very sensitive for tumour microstructure characterisation.57Other MR active nuclei (3He, 13C, 19F, 129Xe, etc.) can also be imaged giving access to new contrasts. If those nuclei have no or neglectable natural background signal in the body, they can be used as tracers, measuring MR properties of the compounds they label. This is a research branch on its own within MR which has already shown clinically relevant applications in lung imaging.58To illustrate the large range of contrasts accessible with MRI, some examples are given in Figure 3.
Around the turn of the millennium, the Radiological Society of North America (RSNA) declared the goal of transforming many imaging modalities, including MRI, into quantitative ones. Numbers were getting important! Numbers that can help to identify pathology and to quantify its state and extent and which can act as reliable biomarkers.59 Numbers that can further help to bridge the gap between the contrasts of different MR sequences, different field strengths and between the images of different vendors when using the same nominal acquisition technique. Numbers that can also help to assess potential therapy success or disease progression, like quantitative fat fraction mapping as a biomarker for hepatic steatosis60 or amide protons targeting magnetisation transfer approaches, that can help to grade tumours.61 Although initial studies, performed in the early days of MR,62 revealed a large spread in the values of relaxation times in normal tissue, making reliable detection of pathology rather questionable, improved technology made it worthwhile reconsidering such concepts. This triggered the need for fast quantitative imaging to obtain values for MR measurable parameters like T1, T2, T2* and M0. Next to accelerated versions of classical approaches for MR parameter-mapping, new techniques were developed very recently, among them MR finger printing63 and MR stat.64 To efficiently encode tissue properties, MR finger printing uses gradient-echo trains, preceded by spin inversion, with sequence parameters varying in time (such as TE, TR and the flip angle). The spatial information is encoded along those trains using heavily under sampled incoherent acquisition schemes, with spirals currently the preferred readout.63 Using appropriate dictionaries, based on Bloch simulations, tissue characterisation and identification can be performed. This concept can be applied not only to T1 and T2 contrasts, but also to other applications like perfusion, water/fat imaging or other parameters which can be retrieved by an appropriate reconstruction.
Concepts of synthesising MR contrasts were revived65 to synthesise via Bloch simulations’ multiple contrasts with the hope to better visualise pathology.66 However, it is not clear yet, how much these approaches will influence future clinical practice, and substantial further clinical work will be necessary to validate them and determine their ultimate value.
Parallel imaging
In the quest to further improve SNR either the available signal can be increased or the noise reduced. At clinical field strengths, the patient dominates the noise present in the receiver coil and thus in the receiver chain. Using smaller coils instead of a large body coil can help to reduce the detected noise level. The limited spatial coverage is then compensated by using an array of such small coils receiving the MR signal in parallel. This led to the birth of parallel imaging in the early 90s,67 then understood simply as the use of multiple receiver coils independently contributing to the final image (c.f. Figure 1c). High-impedance pre-amplifier technology and geometric coil element overlapping approaches helped to reduce mutual coil coupling, considerably boosting SNR compared to conventional single body coil reception.67–69 Manufacturers supported this trend rather quickly in two ways: first by offering MRI platforms that support many parallel receiving channels, and second by dedicated coil arrays for different anatomies, moving the fieldfrom 16 to 32 or even more channels for reception. Miniaturisation of necessary components, new coil designs and manufacturing principles helped to push these limits even further to massive parallel reception.68,69 The use of up to 128 elements has been shown as a proof of principle, just to probe the real limits.70,71 However, it has still to be proven whether this is the right balance between technical complexity, clinical usability, reliability/serviceability and affordability for MRI in the future. Apart from this simple number of elements- and channels-driven race, patient comfort is another important factor. Flexible72 and surface adaptable, lightweight coil arrays73 have been proposed to support the examination of children74 and very sick patients who are not able to stand bulky heavy coils. This is even today a field of active research and technological development.
Sense/GRAPPA
It took until the turn of the millennium before the hardware-dominated parallel imaging concept really revolutionised the entire MR acquisition methodology. This development represented in essence the second wave in terms of accelerating MRI, leading also to a new interpretation of the term parallel imaging, which today mainly means accelerated acquisition. Making the individual coils smaller, makes the detectable MR signal also dependent on the actual coil element positions and shape which gives access to a new and alternative spatial encoding principle that is complementary to the classical Fourier encoding process hitherto used in MR. Sensitivity encoding75–77 became a powerful additional spatial encoding engine which quickly found adoption in commercial platforms mainly because of its simplicity, beauty and general applicability to all MR contrasts. The SNR gained by parallel reception could immediately be traded into acquisition speed by performing appropriate (mostly uniform k-space) under-sampling and corresponding unfolding of the reconstructed images either in the spatial76 or in the k-space77 domain. This second scan acceleration methodology was applicable to both MRI and MR spectroscopic imaging.78
Due to the simplicity of the phase encoding concept used in spin warp imaging, acceleration could be applied in one, two and more dimensions (spatial-, temporal-, parameter-encoded dimensions) immediately. However, successful image reconstruction relies on sufficient differentiation in the sensitivity profiles of the individual receiver coils, which can make the reconstruction ill-posed particularly at the centre of the object where receiver coil sensitivity is at its lowest, and the differential sensitivity between individual coils at its weakest. This led to the invaluable G-factor concept76 that gives a measure for the conditioning of the parallel imaging reconstruction, which should be kept close to unity for satisfactory imaged quality. Coil compression concepts79 can speed-up reconstruction, which is especially important in non-uniform sampling schemes.80 G-factor noise propagation in the reconstruction process was reduced by smart and dedicated k-space sub-sampling schemes, which enabled better control of the aliasing in parallel imaging81,82 and appropriate regularisation83 in the reconstruction.
Recently the use of additional encoding gradients, applied simultaneously with the Cartesian read-out gradients, was proposed to enhance the controlled aliasing. This approach allows for an even better conditioning of the parallel imaging reconstruction,84 bridging the conceptual gap to the sensitivity encoded accelerated non-Cartesian sampling world.80
However, apart from acceleration, the sensitivity encoding concept gave also new freedom for acoustic noise reduction by reducing the amount of gradient switching.85 This burden for the patient can be lowered, while keeping the scanning time unaltered.
Multi-band/Simultaneous Multi-Slice
Parallel imaging can also be applied along the slice selection direction in multi-slice applications.86 By taking advantage from the 3D variation in coil sensitivities, the signals of simultaneously excited multiple slices can be disentangled by parallel imaging. However, it took almost a full decade before robust sub-sampling schemes87 and widespread availability of multi-channel receive coils helped to better pre-condition the unfolding problem, finally facilitating simultaneous multi-band imaging which shows the benefit of almost no SNR penalty due to the low G-factor burden. To mitigate potential RF peak power and specific absorption rate-related issues, optimal phase design for the multi-band RF pulses involved has been proposed.88 The multi-band option nicely bridged the gap between multi-slice and 3D acquisitions, because clinically there are many sampling schemes, which are often not performed in 3D, for instance fMRI or single shot diffusion EPI, and are not appreciably accelerated by standard parallel imaging, as they have to be performed at a fixed TE.89
Model-based reconstruction
Parallel imaging increased the quantity of MR data substantially and turned MR reconstruction into an over determined problem, that is, there are simply more measured data points than pixels to be finally reconstructed. The redundancy in the multi-channel data was primarily used to boost SNR, but offers additional options to enforce data consistency, ruling out potential outliers or corrupted data90,91 to make the best out of the available data and finally improve the overall image quality including the SNR. These model-based reconstruction approaches92 still form an active field of research and are supported by recent advances in linear and non-linear numerical solvers. Depending on the signal model assumed for the image formation process and the cost function, relevant model parameters like coil sensitivities93 can be estimated or updated along with the final image. This could help to mitigate inconsistencies and resulting image quality problems when actual and pre-measured coil sensitivities differ due to patient motion. Other confounding scanning conditions can also be modelled, such as for instance gross motion,94 physiology-induced phase changes95 or motion artefacts in segmented diffusion acquisition96 or B0 fluctuations in gradient-echo imaging.97
Not all of these proposed methods have yet found a place in daily clinical scanning practice, because they also change the image reconstruction process significantly, from simple pipelines into iterative loops which are often rather time consuming and could benefit from further numerical and/or computer hardware speed-ups.
Compressed Sensing
The sampling community had not yet assimilated all aspects of accelerated parallel imaging, when the next wave of scan acceleration arrived. Previously MR sampling schemes were based on the Nyquist sampling theorem. However, the robustness of radial imaging against under-sampling, seen already in the early days, made it plausible that alternatives might be viable. Compressed sensing (CS)98 is a new sampling paradigm, proven by mathematicians, demonstrating that images can correctly be reconstructed from even less data than usually sampled. To explore CS, the underlying signal distribution must be sparse, what seems to be the case in many MR applications, an incoherent sampling scheme has to be used and a sparsity enforcing (iterative) reconstruction has to be employed to reduce the pseudo-noise introduced by the unconventional sampling process, which can be realised using a dedicated cost function model-based reconstruction.92 It turned out that CS enables additional acceleration on top of parallel imaging by roughly a factor two and potentially even more, when more sparse encoding dimensions are involved.99,100 This substantial extra acceleration, the compatibility to parallel imaging (c.f. Figure 4) and with many current clinical protocols, made CS very attractive so that it rapidly found FDA approval101 and early adoption in clinical practice.
Deep learning and artificial intelligence
Since about 2015 another trend emerged when it was shown that artificial intelligence (AI) can outperform humans in selected tasks.102 So, registration and segmentation, typical classical image processing tasks have been shown to be better performed by AI.102 Many typical MR artefacts, with the motion-induced ones being the most critical, can be learned, identified and removed from MRI images.103,104 In addition to those post-processing applications, AI approaches have already been applied to reconstruct the image directly from under-sampled k-space data.105 One can argue whether it is useful to learn the FFT for which we have already very efficient algorithms, or whether the huge memory needed to form the fully connected layers that are necessary for that purpose are really justified, but at least it shows the huge potential of AI, even when it is just as an element in our current model-based reconstructions.106,107 Iterative algorithms can also be seen as a recurrent neural network consisting of different layers with simple operations in between optimised in a data-driven/learning-based fashion.108This is the current situation, at a cusp between model—driven to data—driven approaches which can be expected to nicely complement each other in the future (Figure 5).
Networks can learn the optimal sparsifying transform and the corresponding regularisation parameters, often critical in many modern model-based reconstructions109 and can also be helpful in final image de-noising.110 Furthermore, they have another benefit when the appropriately trained reconstruction model is in place, the actual computation for reconstructing a new data set can be extremely fast.
However, despite the improvements achievable with machine learning compared to classical or iterative approaches, there is a concern that neural networks have a black-box nature and that there is no closed theory foundation about them available.106 However, there is no doubt, deep learning in medical image formation, data analysis and augmented image interpretation and diagnosis is here to stay. As a community we need more experience in the machine learning area just to determine which problems can be completely solved using deep learning techniques, which are better handled by a combination of deep learning with other techniques, or for which no deep learning component is needed at all.102
Workflow
One further important aspect of operating an MRI system today is workflow to guarantee highest diagnostic value, optimal scan efficacy, and best patient comfort and safety. In many aspects here, the MR radiographer is key. Back in the early days, technicians were MR experts, but today they have to have other skills too. They manage multiple tasks, sometimes multiple systems, often multiple modalities, and most importantly, they often handle many different patients simultaneously. Today’s radiographers have to fix scheduling issues, instruct and prepare the patient, ensuring that the patient is metal-free (especially ferro magnetic), add physiology sensors, place coils properly, ensure safe patient positioning in the magnet and finally run the different clinical scans, not only maximising image quality but also making decisions which potential scan should be added or skipped to appropriately support final diagnosis.
This will become an area of substantial work in the future, and it will lead to concepts of complexity-reduced and auto-driven scanners, with real-time auto-triaging capabilities, freeing the operator from boring standard work to better deal with the patient, increasing patient-comfort and diagnostic confidence.
In that respect, we can also expect changes in the way MRI will be used in the future. Currently MR often comes at the end of the diagnostic chain with the aim to finalise an often unclear diagnosis. This results in long total examination times adding more scans to get a comprehensive radiological view of the patients’ problem. However, this can change in the future, because recently also 5 min comprehensive exams arrived on the radiologist’s wish list, next to the more extensive traditional scanning protocols. These fast exams can further reduce table times, paving the way for applying MR as a test, and can further change the way we perform MRI.
Concluding remarks
Looking back on those 50 years, MRI has already come a long way, and there is still lots of room for considerable development in the future. With the ability to probe the local tissue environment, via multiple MR contrasts, and with the development of fast and motion-still sequences, we went as a community a long way towards improving the diagnostic image quality and value of MRI. We improved acquisition speed with strong gradients, parallel imaging and compressed sensing and in the future advanced model-driven and AI supported reconstruction algorithms will allow us to obtain more reliable and better diagnostic information, increasing diagnostic confidence and offering a convenient and highly patient-oriented service.
Again, it needs a single human to have a great idea, but a whole community of medical physicists, MR-physicists, engineers, radiographers, physicians and all patients to make it really fly!
Footnotes
Acknowledgment: Many thanks to Dr Christian Stehning and Dr Hendrik Kooijman who helped to compile Figure 3.
Funding: Peter Börnert, David G Norris: no relevant grants have to be mentioned.
Contributor Information
Peter Börnert, Email: peter.boernert@philips.com.
David G. Norris, Email: david.norris@donders.ru.nl.
REFERENCES
- 1.Holtzmann Kevles B, Holtzmann Kevles B. Naked to the bone: medical imaging in the twentieth century: Rutgers University Press; 1998. 978–201328332. [Google Scholar]
- 2.Lauterbur PC. Image formation by induced local interactions: examples employing nuclear magnetic resonance. Nature 1973; 242: 190–1. doi: 10.1038/242190a0 [DOI] [PubMed] [Google Scholar]
- 3.Damadian R. Tumor detection by nuclear magnetic resonance. Science 1971; 171: 1151–3. doi: 10.1126/science.171.3976.1151 [DOI] [PubMed] [Google Scholar]
- 4.Edelstein WA, Hutchison JMS, Johnson G, Redpath T. Spin warp NMR imaging and applications to human whole-body imaging. Phys Med Biol 1980; 25: 751–6. doi: 10.1088/0031-9155/25/4/017 [DOI] [PubMed] [Google Scholar]
- 5.Crooks L, Arakawa M, Hoenninger J, Watts J, McRee R, Kaufman L, et al. . Nuclear magnetic resonance whole-body imager operating at 3.5 KGauss. Radiology 1982; 143: 169–74. doi: 10.1148/radiology.143.1.7063722 [DOI] [PubMed] [Google Scholar]
- 6.Mansfield P, Morris PG. Advances in Magnetic Resonance : NMR Imaging in Biomedicine. New York: Academic Press; 1982. 187. [Google Scholar]
- 7.Twieg DB. The k -trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methods. Med Phys 1983; 10: 610–21. doi: 10.1118/1.595331 [DOI] [PubMed] [Google Scholar]
- 8.Ljunggren S. A simple graphical representation of fourier-based imaging methods. Journal of Magnetic Resonance 1983; 54: 338–43. doi: https://doi.org/ 10.1016/0022-2364(83)90060-4 [Google Scholar]
- 9.Ehman EC, Johnson GB, Villanueva-Meyer JE, Cha S, Leynes AP, Larson PEZ, et al. . Pet/Mri: where might it replace PET/CT? J. Magn. Reson. Imaging 2017; 46: 1247–62. doi: 10.1002/jmri.25711 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hall WA, Paulson ES, van der Heide UA. Atlantic Consortium and the ViewRay C2T2 research Consortium.cThe transformation of radiation oncology using real-time magnetic resonance guidance: a review. Eur J Cancer 2019; 122: 42–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Mansfield P, Maudsley AA. Planar spin imaging by NMR. J. Phys. C: Solid State Phys. 1976; 9: L409–12. doi: 10.1088/0022-3719/9/15/004 [DOI] [Google Scholar]
- 12.Hutchinson JMS, Sutherland RJ, Mallard JR. Nmr imaging: image recovery under magnetic fields with large non-uniformities. J. Phys. E: Sci. Instrum. 1978; 11: 217–22. doi: 10.1088/0022-3735/11/3/012 [DOI] [PubMed] [Google Scholar]
- 13.Redpath TW. Signal-To-Noise ratio in MRI. Br J Radiol 1998; 71: 704–7. doi: 10.1259/bjr.71.847.9771379 [DOI] [PubMed] [Google Scholar]
- 14.Katscher U, Börnert P, Leussler C, van den Brink JS. Transmit sense. Magn. Reson. Med. 2003; ; 49: 144–50Jan. doi: 10.1002/mrm.10353 [DOI] [PubMed] [Google Scholar]
- 15.Zhu Y. Parallel excitation with an array of transmit coils. Magn. Reson. Med. 2004; 51: 775–84. doi: 10.1002/mrm.20011 [DOI] [PubMed] [Google Scholar]
- 16.Nelles M, König RS, Gieseke J, Guerand-van Battum MM, Kukuk GM, Schild HH, et al. . Dual-source parallel rf transmission for clinical MR imaging of the spine at 3.0 T: Intraindividual comparison with conventional single-source transmission. Radiology 2010; 257: 743–53. doi: 10.1148/radiol.10092146 [DOI] [PubMed] [Google Scholar]
- 17.Marques JP, Simonis FFJ, Webb AG. Low‐field MRI: an Mr physics perspective. J. Magn. Reson. Imaging 2019; 49: 1528–42. doi: 10.1002/jmri.26637 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Campbell-Washburn AE, Ramasawmy R, Restivo MC, Bhattacharya I, Basar B, Herzka DA, et al. . Opportunities in interventional and diagnostic imaging by using high-performance Low-Field-Strength MRI. Radiology 2019; 293: 384–93. doi: 10.1148/radiol.2019190452 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Roemer PB, Hickey JS. Self-shielded gradient coils for nuclear magnetic resonance imaging. European Patent Application 1986; 87: 198. [Google Scholar]
- 20.Turner R. A target field approach to optimal coil design. J. Phys. D: Appl. Phys. 1986; 19: L147–51. doi: 10.1088/0022-3727/19/8/001 [DOI] [Google Scholar]
- 21.Mansfield P, Chapman B. Active magnetic screening of gradient coils in NMR imaging. Journal of Magnetic Resonance 1986; 66: 573–6. doi: 10.1016/0022-2364(86)90205-2 [DOI] [Google Scholar]
- 22.Van Vaals JJ, Bergman AH. Optimization of eddy-current compensation. Journal of Magnetic Resonance 1990; 90: 52–70. doi: 10.1016/0022-2364(90)90365-G [DOI] [Google Scholar]
- 23.Turner R. Gradient coil design: a review of methods. Magn Reson Imaging 1993; 11: 903–20. doi: 10.1016/0730-725X(93)90209-V [DOI] [PubMed] [Google Scholar]
- 24.Hidalgo-Tobon SS. Theory of gradient coil design methods for magnetic resonance imaging. Concepts in Magnetic Resonance Part A 2010; 36A: 223–42. doi: 10.1002/cmr.a.20163 [DOI] [Google Scholar]
- 25.Golay MJE. Field Homogenizing coils for nuclear spin resonance instrumentation. Review of Scientific Instruments 1958; 29: 313–5. doi: 10.1063/1.1716184 [DOI] [Google Scholar]
- 26.Davids M, Guérin B, Malzacher M, Schad LR, Wald LL. Predicting Magnetostimulation thresholds in the peripheral nervous system using realistic body models. Sci Rep 2017; 7: 5316. doi: 10.1038/s41598-017-05493-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Davids M, Guérin B, vom Endt A, Schad LR, Wald LL. Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations. Magn. Reson. Med 2019; 81: 686–701. doi: 10.1002/mrm.27382 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Weiger M, Overweg J, Rösler MB, Froidevaux R, Hennel F, Wilm BJ, et al. . A high-performance gradient insert for rapid and short-T 2 imaging at full duty cycle. Magn. Reson. Med 2018; 79: 3256–66. doi: 10.1002/mrm.26954 [DOI] [PubMed] [Google Scholar]
- 29.Burdette JH, Elster AD, Ricci PE. Acute cerebral infarction: quantification of spin-density and T2 shine-through phenomena on diffusion-weighted Mr images. Radiology 1999; 212: 333–9. doi: 10.1148/radiology.212.2.r99au36333 [DOI] [PubMed] [Google Scholar]
- 30.Hennig J, Nauerth A, Friedburg H. Rare imaging: a fast imaging method for clinical Mr. Magn Reson Med 1986; 3: 823–33. doi: 10.1002/mrm.1910030602 [DOI] [PubMed] [Google Scholar]
- 31.Haase A, Frahm J, Matthaei D, Hanicke W, Merboldt K-D. Flash imaging. rapid NMR imaging using low flip-angle pulses. Journal of Magnetic Resonance 1986; 67: 258–66. doi: 10.1016/0022-2364(86)90433-6 [DOI] [PubMed] [Google Scholar]
- 32.Rzedzian R, Doyle M, Mansfield P, Chapman B, Guilfoyle C, Coupland P, et al. . Lancet 1983; 2: 1281–2. [DOI] [PubMed] [Google Scholar]
- 33.Mansfield P. Multi-planar image formation using NMR spin echoes. J. Phys. C: Solid State Phys. 1977; 10: L55–8. doi: 10.1088/0022-3719/10/3/004 [DOI] [Google Scholar]
- 34.Haase A, Snapshot FLASH MRI. Applications to T1, T2, and chemical-shift imaging. MagnReson Med 1990; 13: 77–89. doi: 10.1002/mrm.1910130109 [DOI] [PubMed] [Google Scholar]
- 35.Norris DG, Börnert P, Reese T, Leibfritz D. On the application of ultra-fast rare experiments. Magn. Reson. Med. 1992; 27: 142–64. doi: 10.1002/mrm.1910270114 [DOI] [PubMed] [Google Scholar]
- 36.Ahn CB, Kim JH, Cho ZH. High-Speed Spiral-Scan echo planar NMR Imaging-I. IEEE Trans Med Imaging 1986; 5: 2–7. doi: 10.1109/TMI.1986.4307732 [DOI] [PubMed] [Google Scholar]
- 37.Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn. Reson. Med. 1992; 28: 202–13. doi: 10.1002/mrm.1910280204 [DOI] [PubMed] [Google Scholar]
- 38.Glover GH, Pauly JM. Projection reconstruction techniques for reduction of motion effects in MRI. Magn. Reson. Med. 1992; 28: 275–89. doi: 10.1002/mrm.1910280209 [DOI] [PubMed] [Google Scholar]
- 39.Pipe JG. Motion correction with propeller MRI: application to head motion and free-breathing cardiac imaging. Magn. Reson. Med. 1999; 42: 963–9. doi: [DOI] [PubMed] [Google Scholar]
- 40.Norris DG, Hutchison JMS. Concomitant magnetic field gradients and their effects on imaging at low magnetic field strengths. Magn Reson Imaging 1990; 8: 33–7. doi: 10.1016/0730-725X(90)90209-K [DOI] [PubMed] [Google Scholar]
- 41.Börnert P, Schomberg H, Aldefeld B, Groen J. Improvements in spiral MR imaging. MAGMA 1999; 9(1-2): 29–41. doi: 10.1007/BF02634590 [DOI] [PubMed] [Google Scholar]
- 42.Rasche V, Holz D, Proksa R. Mr fluoroscopy using projection reconstruction multi-gradient-echo (prMGE) MRI. Magn. Reson. Med. 1999; 42: 324–34. doi: [DOI] [PubMed] [Google Scholar]
- 43.Schäffter T, Rasche V, Carlsen IC. Motion compensated projection reconstruction. Magn Reson Med 1999; 41: 954–63. doi: [DOI] [PubMed] [Google Scholar]
- 44.Riederer SJ, Tasciyan T, Farzaneh F, Lee JN, Wright RC, Herfkens RJ. Mr fluoroscopy: technical feasibility. Magn. Reson. Med. 1988; 8: 1–15. doi: 10.1002/mrm.1910080102 [DOI] [PubMed] [Google Scholar]
- 45.Pearlman JD, Edelman RR. Ultrafast magnetic resonance imaging. segmented turboflash, echo-planar, and real-time nuclear magnetic resonance. Radiol Clin North Am 1994; 32: 593–612. [PubMed] [Google Scholar]
- 46.Hajnal JV, Bryant DJ, Kasuboski L, Pattany PM, Coene BD, Lewis PD, et al. . Use of fluid attenuated inversion recovery (FLAIR) pulse sequences in MRI of the brain. J Comput Assist Tomogr 1992; 16: 841–4. Nov-Dec;. doi: 10.1097/00004728-199211000-00001 [DOI] [PubMed] [Google Scholar]
- 47.Haase A, Frahm J, Hänicke W, Matthaei D. 1 H NMR chemical shift selective (CHESS) imaging. Phys Med Biol 1985; 30: 341–4. doi: 10.1088/0031-9155/30/4/008 [DOI] [PubMed] [Google Scholar]
- 48.Dixon WT. Simple proton spectroscopic imaging. Radiology 1984; 153: 189–94. doi: 10.1148/radiology.153.1.6089263 [DOI] [PubMed] [Google Scholar]
- 49.Reeder SB, Wen Z, Yu H, Pineda AR, Gold GE, Markl M, et al. . Multicoil Dixon chemical species separation with an iterative least-squares estimation method. Magn. Reson. Med. 2004; 51: 35–45. doi: 10.1002/mrm.10675 [DOI] [PubMed] [Google Scholar]
- 50.Hu HH, Börnert P, Hernando D, Kellman P, Ma J, Reeder S, et al. . ISMRM workshop on fat-water separation: insights, applications and progress in MRI. Magn Reson Med 2012; 68: 378–88. doi: 10.1002/mrm.24369 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval-Jeantet M. Mr imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology 1986; 161: 401–7. doi: 10.1148/radiology.161.2.3763909 [DOI] [PubMed] [Google Scholar]
- 52.Moseley ME, Butts K, Yenari MA, Marks M, Crespigny AD. Clinical aspects of DWI. NMR Biomed 1995; 8(7-8): 387–96. Nov-Dec. doi: 10.1002/nbm.1940080712 [DOI] [PubMed] [Google Scholar]
- 53.Mori S, Kaufmann WE, Davatzikos C, Stieltjes B, Amodei L, Fredericksen K, et al. . Imaging cortical association tracts in the human brain using diffusion-tensor-based axonal tracking. Magn. Reson. Med. 2002; 47: 215–23. doi: 10.1002/mrm.10074 [DOI] [PubMed] [Google Scholar]
- 54.Jbabdi S, Johansen-Berg H. Tractography: where do we go from here? Brain Connect 2011; 1: 169–83. doi: 10.1089/brain.2011.0033 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Jeurissen B, Descoteaux M, Mori S, Leemans A. Diffusion MRI fiber tractography of the brain. NMR Biomed 2019; 32: e3785. doi: 10.1002/nbm.3785 [DOI] [PubMed] [Google Scholar]
- 56.Minati L, Węglarz WP. Physical foundations, models, and methods of diffusion magnetic resonance imaging of the brain: a review. Concepts in Magnetic Resonance Part A 2007; 30A: 278–30737. doi: 10.1002/cmr.a.20094 [DOI] [Google Scholar]
- 57.Szczepankiewicz F, van Westen D, Englund E, Westin C-F, Ståhlberg F, Lätt J, et al. . The link between diffusion MRI and tumor heterogeneity: mapping cell eccentricity and density by diffusional variance decomposition (divide. Neuroimage 2016; 142: 522–32. doi: 10.1016/j.neuroimage.2016.07.038 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Kern AL, Vogel-Claussen J. Hyperpolarized gas MRI in pulmonology. Br J Radiol 2018; 53: 20170647. doi: 10.1259/bjr.20170647 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. https://www.rsna.org/en/research/quantitative-imaging-biomarkers-alliance
- 60.Reeder SB, Hu HH, Sirlin CB. Proton density fat-fraction: a standardized mr-based biomarker of tissue fat concentration. J. Magn. Reson. Imaging 2012; 36: 1011–4. doi: 10.1002/jmri.23741 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Togao O, Yoshiura T, Keupp J, Hiwatashi A, Yamashita K, Kikuchi K, et al. . Amide proton transfer imaging of adult diffuse gliomas: correlation with histopathological grades. Neuro Oncol 2014; 16: 441–8. doi: 10.1093/neuonc/not158 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Bottomley PA, Hardy CJ, Argersinger RE, Allen-Moore G. A review of 1H nuclear magnetic resonance relaxation in pathology: are T1 and T2 diagnostic? Med Phys 1987; 14: 1–37. doi: 10.1118/1.596111 [DOI] [PubMed] [Google Scholar]
- 63.Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, et al. . Magnetic resonance fingerprinting. Nature 2013; ; 495: 187–9214. doi: 10.1038/nature11971 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Sbrizzi A, Heide Ovander, Cloos M, Toorn Avander, Hoogduin H, Luijten PR, et al. . Fast quantitative MRI as a nonlinear tomography problem. Magn Reson Imaging 2018; 46: 56–63. doi: 10.1016/j.mri.2017.10.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Bobman SA, Riederer SJ, Lee JN, Suddarth SA, Wang HZ, Drayer BP, et al. . Cerebral magnetic resonance image synthesis. AJNR Am J Neuroradiol 1985; 6: 265–9. Mar-Apr. [PMC free article] [PubMed] [Google Scholar]
- 66.Andica C, Hagiwara A, Hori M, Kamagata K, Koshino S, Maekawa T, et al. . Review of synthetic MRI in pediatric brains: basic principle of Mr quantification, its features, clinical applications, and limitations. Journal of Neuroradiology 2019; 46: 268–75. doi: 10.1016/j.neurad.2019.02.005 [DOI] [PubMed] [Google Scholar]
- 67.Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med 1990; 16: 192–225. doi: 10.1002/mrm.1910160203 [DOI] [PubMed] [Google Scholar]
- 68.Wiesinger F, Boesiger P, Pruessmann KP. Electrodynamics and ultimate SNR in parallel MR imaging. Magn. Reson. Med. 2004; 52: 376–90. doi: 10.1002/mrm.20183 [DOI] [PubMed] [Google Scholar]
- 69.Keil B, Wald LL. Massively parallel MRI detector arrays. Journal of Magnetic Resonance 2013; 229: 75–89. doi: 10.1016/j.jmr.2013.02.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Schmitt M, Potthast A, Sosnovik DE, Polimeni JR, Wiggins GC, Triantafyllou C, et al. . A 128-channel receive-only cardiac coil for highly accelerated cardiac MRI at 3 tesla. Magn. Reson. Med. 2008; 59: 1431–9. doi: 10.1002/mrm.21598 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Hardy CJ, Giaquinto RO, Piel JE, Rohling AAS KW, Marinelli L, Blezek DJ, et al. . 128-channel body MRI with a flexible high-density receiver-coil array. J. Magn. Reson. Imaging 2008; 28: 1219–25. doi: 10.1002/jmri.21463 [DOI] [PubMed] [Google Scholar]
- 72.Nordmeyer-Massner JA, De Zanche N, Pruessmann KP. Stretchable coil arrays: application to knee imaging under varying flexion angles. Magn. Reson. Med. 2012; 67: 872–9. doi: 10.1002/mrm.23240 [DOI] [PubMed] [Google Scholar]
- 73.McGee KP, Stormont RS, Lindsay SA, Taracila V, Savitskij D, Robb F, et al. . Characterization and evaluation of a flexible MRI receive coil array for radiation therapy Mr treatment planning using highly decoupled rf circuits. Phys. Med. Biol. 2018; 63: 08NT02.08NT02. doi: 10.1088/1361-6560/aab691 [DOI] [PubMed] [Google Scholar]
- 74.Zhang T, Grafendorfer T, Cheng JY, Ning P, Rainey B, Giancola M, et al. . A semiflexible 64-channel receive-only phased array for pediatric body MRI at 3T. Magn. Reson. Med. 2016; 76: 1015–21. doi: 10.1002/mrm.25999 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn. Reson. Med. 1997; 38: 591–603. doi: 10.1002/mrm.1910380414 [DOI] [PubMed] [Google Scholar]
- 76.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. Sense: sensitivity encoding for fast MRI. Magn. Reson. Med. 1999; 42: 952–62. doi: [DOI] [PubMed] [Google Scholar]
- 77.Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. . Generalized autocalibrating partially parallel acquisitions (grappa. Magn. Reson. Med. 2002; 47: 1202–10. doi: 10.1002/mrm.10171 [DOI] [PubMed] [Google Scholar]
- 78.Dydak U, Weiger M, Pruessmann KP, Meier D, Boesiger P. Sensitivity-encoded spectroscopic imaging. Magn. Reson. Med. 2001; 46: 713–22. doi: 10.1002/mrm.1250 [DOI] [PubMed] [Google Scholar]
- 79.Buehrer M, Pruessmann KP, Boesiger P, Kozerke S. Array compression for MRI with large coil arrays. Magn. Reson. Med. 2007; 57: 1131–9. doi: 10.1002/mrm.21237 [DOI] [PubMed] [Google Scholar]
- 80.Pruessmann KP, Weiger M, Börnert P, Boesiger P. Advances in sensitivity encoding with arbitrary k -space trajectories. Magn. Reson. Med. 2001; 46: 638–51. doi: 10.1002/mrm.1241 [DOI] [PubMed] [Google Scholar]
- 81.Breuer FA, Blaimer M, Mueller MF, Seiberlich N, Heidemann RM, Griswold MA, et al. . Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA. Magn. Reson. Med. 2006; 55: 549–56. doi: 10.1002/mrm.20787 [DOI] [PubMed] [Google Scholar]
- 82.Tsao J, Kozerke S, Boesiger P, Pruessmann KP. Optimizing spatiotemporal sampling fork-t blast andk-t sense: application to high-resolution real-time cardiac steady-state free precession. Magn. Reson. Med. 2005; 53: 1372–82. doi: 10.1002/mrm.20483 [DOI] [PubMed] [Google Scholar]
- 83.Lin F-H, Kwong KK, Belliveau JW, Wald LL. Parallel imaging reconstruction using automatic regularization. Magn. Reson. Med. 2004; 51: 559–67. doi: 10.1002/mrm.10718 [DOI] [PubMed] [Google Scholar]
- 84.Bilgic B, Gagoski BA, Cauley SF, Fan AP, Polimeni JR, Grant PE, et al. . Wave-CAIPI for highly accelerated 3D imaging. Magn. Reson. Med. 2015; 73: 2152–62. doi: 10.1002/mrm.25347 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.de Zwart JA, van Gelderen P, Kellman P, Duyn JH. Reduction of gradient acoustic noise in MRI using SENSE-EPI. Neuroimage 2002; 16: 1151–5. doi: 10.1006/nimg.2002.1119 [DOI] [PubMed] [Google Scholar]
- 86.Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging 2001; 13: 313–7. doi: [DOI] [PubMed] [Google Scholar]
- 87.Breuer FA, Blaimer M, Heidemann RM, Mueller MF, Griswold MA, Jakob PM. Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magn. Reson. Med. 2005; 53: 684–91. doi: 10.1002/mrm.20401 [DOI] [PubMed] [Google Scholar]
- 88.Hennig J. Chemical shift imaging with phase-encoding rf pulses. Magn. Reson. Med. 1992; 25: 289–98. doi: 10.1002/mrm.1910250207 [DOI] [PubMed] [Google Scholar]
- 89.Moeller S, Yacoub E, Olman CA, Auerbach E, Strupp J, Harel N, et al. . Multiband multislice GE-EPI at 7 Tesla, with 16-fold acceleration using partial parallel imaging with application to high spatial and temporal whole-brain fMRI. Magn. Reson. Med. 2010; 63: 1144–5364. doi: 10.1002/mrm.22361 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Bydder M, Atkinson D, Larkman DJ, Hill DLG, Hajnal JV. SMASH navigators. Magn. Reson. Med. 2003; ; 49: 493–500Mar. doi: 10.1002/mrm.10388 [DOI] [PubMed] [Google Scholar]
- 91.Winkelmann R, Börnert P, Dössel O. Ghost artifact removal using a parallel imaging approach. Magn. Reson. Med. 2005; 54: 1002–9. doi: 10.1002/mrm.20640 [DOI] [PubMed] [Google Scholar]
- 92.Fessler J. Model-Based image reconstruction for MRI. IEEE Signal Process Mag 2010; ; 27: 81–91. doi: 10.1109/MSP.2010.936726 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Ying L, Sheng J. Joint image reconstruction and sensitivity estimation in sense (JSENSE. Magn. Reson. Med. 2007; 57: 1196–202. doi: 10.1002/mrm.21245 [DOI] [PubMed] [Google Scholar]
- 94.Odille F, Vuissoz P-A, Marie P-Y, Felblinger J. Generalized reconstruction by inversion of coupled systems (GRICS) applied to free-breathing MRI. Magn. Reson. Med. 2008; 60: 146–57. doi: 10.1002/mrm.21623 [DOI] [PubMed] [Google Scholar]
- 95.Guo H, Ma X, Zhang Z, Zhang B, Yuan C, Huang F. POCS-enhanced inherent correction of motion-induced phase errors (POCS-ICE) for high-resolution multishot diffusion MRI. Magn. Reson. Med. 2016; 75: 169–80. doi: 10.1002/mrm.25594 [DOI] [PubMed] [Google Scholar]
- 96.Steinhoff M, Nehrke K, Mertins A, Börnert P. Segmented diffusion imaging with iterative motion‐corrected reconstruction (sediment) for brain echo‐planar imaging. NMR Biomed 2019; 13: e4185. doi: 10.1002/nbm.4185 [DOI] [PubMed] [Google Scholar]
- 97.Meineke J, Nielsen T. Data consistency-driven determination of B0-fluctuations in gradient-echo MRI. Magn Reson Med 2019; 81: 3046–55. doi: 10.1002/mrm.27630 [DOI] [PubMed] [Google Scholar]
- 98.Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 2007; 58: 1182–95. doi: 10.1002/mrm.21391 [DOI] [PubMed] [Google Scholar]
- 99.Murphy M, Alley M, Demmel J, Keutzer K, Vasanawala S, Lustig M. Fast $\ell_1$-SPIRiT Compressed Sensing Parallel Imaging MRI: Scalable Parallel Implementation and Clinically Feasible Runtime. IEEE Trans Med Imaging 2012; 31: 1250–62. doi: 10.1109/TMI.2012.2188039 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Doneva M, Börnert P, Eggers H, Stehning C, Sénégas J, Mertins A. Compressed sensing reconstruction for magnetic resonance parameter mapping. Magn. Reson. Med. 2010; 64: 1114–20. doi: 10.1002/mrm.22483 [DOI] [PubMed] [Google Scholar]
- 101.Fessler JA. Optimization methods for Mr image reconstruction. arXiv preprint arXiv 2019;, . 035101903. [Google Scholar]
- 102.Lundervold AS, Lundervold A. An overview of deep learning in medical imaging focusing on MRI. Z Med Phys 2019; 29: 102–27. doi: 10.1016/j.zemedi.2018.11.002 [DOI] [PubMed] [Google Scholar]
- 103.Küstner T, Liebgott A, Mauch L, Martirosian P, Bamberg F, Nikolaou K, et al. . Automated reference-free detection of motion artifacts in magnetic resonance images. MAGMA 2018; 31: 243–56. doi: 10.1007/s10334-017-0650-z [DOI] [PubMed] [Google Scholar]
- 104.Sommer K, Saalbach A, Brosch T, Hall C, Cross NM, Andre JB. Correction of motion artifacts using a multiscale FullyConvolutional neural network. AJNR Am J Neuroradiol 2020; 41: 416–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Zhu B, Liu JZ, Cauley SF, Rosen BR, Rosen MS. Image reconstruction by domain-transform manifold learning. Nature 2018; 555: 487–92. doi: 10.1038/nature25988 [DOI] [PubMed] [Google Scholar]
- 106.Haskell MW, Cauley SF, Bilgic B, Hossbach J, Splitthoff DN, Pfeuffer J, et al. . Network accelerated motion estimation and reduction (NAMER): Convolutional neural network guided retrospective motion correction using a separable motion model. Magn Reson Med 2019; 82: 1452–61. doi: 10.1002/mrm.27771 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Lee J, Han Y, Ryu J-K, Park J-Y, Ye JC. k-Space deep learning for reference-free epi ghost correction. Magn Reson Med 2019; 82: 2299–313. doi: 10.1002/mrm.27896 [DOI] [PubMed] [Google Scholar]
- 108.Wang G, Ye JC, Mueller K, Fessler JA. Image reconstruction is a new frontier of machine learning. IEEE Trans Med Imaging 2018; 37: 1289–96. doi: 10.1109/TMI.2018.2833635 [DOI] [PubMed] [Google Scholar]
- 109.Hammernik K, Klatzer T, Kobler E, Recht MP, Sodickson DK, Pock T, et al. . Learning a variational network for reconstruction of accelerated MRI data. Magn Reson Med 2018; 79: 3055–71. doi: 10.1002/mrm.26977 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.Kaur P, Singh G, Kaur P. A review of denoising medical images using machine learning approaches. Curr Med Imaging Rev 2018; 14: 675–85. doi: 10.2174/1573405613666170428154156 [DOI] [PMC free article] [PubMed] [Google Scholar]