Abstract
Background
The spread of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which appeared in late 2019, has been limited by isolating infected individuals. However, identifying such individuals requires accurate diagnostic tools.
Objective
This study evaluates the capacity of the Aptima™ Transcription-Mediated Amplification (TMA) assay (Hologic® Panther System) to detect the virus in clinical samples.
Study design
We compared the Aptima™ assay to two in-house real-time RT-PCR techniques, one running on the Panther Fusion™ module and the other on the MagNA Pure 96 and Light-Cycler 480 instruments. We included a total of 200 respiratory specimens: 100 tested prospectively and 100 retrospectively (25 -ve/75 +ve).
Results
The final Cohen’s kappa coefficients were: κ = 0.978 between the Aptima™ and Panther Fusion™ assays, κ = 0.945 between the Aptima™ and MagNA/LC480 assays and κ = 0.956 between the MagNA/LC480 and Panther Fusion™ assays.
Conclusion
These findings indicate that the Aptima™ SARS-CoV-2 TMA assay data agree well with those obtained with our routine methods and that this assay can be used to diagnose coronavirus disease 2019 (COVID-19).
Keywords: SARS-CoV-2, COVID-19, Coronavirus, Hologic® Panther System
1. Background
The emergence of a new severe acute respiratory syndrome coronavirus (SARS-CoV-2) in late 2019 and its rapid spread worldwide led to over 6 million cases of coronavirus disease 2019 (COVID-19) by 1 June 2020 [1,2]. As the rapid identification of infected individuals and their isolation is essential to prevent the spread of the disease, it became necessary to develop accurate, easy-to-use molecular diagnostic assays. Virology laboratories first developed and published in-house techniques [3,4], which were soon followed by commercial assays. These must be validated on clinical specimens in addition to preliminary tests on culture lysates/supernatants or serially diluted RNA transcripts. Nasopharyngeal swabs are generally used for diagnosing viral respiratory tract infections, while deeper samples, such as tracheal aspirations or broncho-alveolar lavage fluids are used only in severe cases [5].
2. Objectives
This study evaluates the performance of the automated commercial Aptima™ SARS-CoV-2 assay (Hologic® Panther System) on clinical samples. This test, which uses transcription-mediated amplification (TMA) to detect SARS-CoV-2, was compared to two in-house reverse transcriptase – polymerase chain reaction (RT-PCR) assays we presently use to diagnose COVID-19.
3. Study design
3.1. Samples
We tested 200 respiratory samples (198 nasopharyngeal swabs and 2 tracheal aspirations) from patients suspected of having COVID-19 in April and May 2020 (54.5 % men; median [IQR] age: 61 [38–83] years). Half of the samples were tested with all three assays in parallel. These were the prospective samples (98 nasopharyngeal swabs + 2 tracheal aspirations). The remaining 100 nasopharyngeal swabs were selected based on the results obtained with the Panther Fusion™ module in open access (see below). They included 25 samples that were negative, 25 that were positive with a cycle threshold (Ct) value > 35, 25 that were positive with a Ct of 25–35, and 25 that were positive with a Ct < 25. These retrospective samples had been stored at −20 °C for up to 43 days.
3.2. Methods
The Aptima™ SARS-CoV-2 assay was performed following the manufacturer’s instructions. Virus transport medium (500 μL) was manually placed in the Panther™ tube containing 710 μL of lysis buffer. The instrument used 360 μL of this mix for the lysis and capture of nucleic acids. The Aptima™ assay targets two virus sequences located on the ORF1ab gene. An internal control was included.
Our two in-house assays that both target virus sequences on the SARS-CoV-2 RNA-dependent RNA-polymerase (RdRp) gene, use primers and probes (IP-2 and IP-4) and amplification programs from the Institut Pasteur, Paris [4]. The first assay uses the MagNA Pure 96 instrument (Roche Diagnostics) to extract nucleic acids from 200 μL of virus transport medium eluted in 100 μL. RT-PCR is then performed on the Light-Cycler 480 instrument (LC480) (Roche Diagnostics) using the AccuStart™ Taq DNA polymerase (QuantaBio) and 2 μL of nucleic acids (total reaction volume: 10 μL). The second assay uses the same primers and probes on the Panther Fusion™ module (Hologic®) in open access. Samples were prepared as described above for the Aptima™ assay.
3.3. Data analysis
Because there is no reference standard method for the detection of SARS-CoV-2, we calculated the concordance between assays by using Cohen's kappa coefficient (κ), the positive percent agreement, the negative percent agreement and overall percent agreement, all with 95 % confidence intervals [6,7].
4. Results and discussion
4.1. Prospective study
The prospective study provided 2 positive and 98 negative samples, and there was perfect concordance between results of the Aptima™ assay and those of the Panther Fusion™ assay and very good concordance between the Aptima™ assay and the MagNA/LC480 assay. One sample was faintly positive with the MagNA/LC480 assay (virus IP-4 target, Ct = 39.97 and IP-2 -ve) and negative with the other two assays (sample A2 0141 0134, cf. Table S1 in Supplementary materials).
4.2. Retrospective study
All positive samples with Ct values < 35 (n = 50) with the Panther Fusion™ assay tested positive with the other two assays. The MagNA/LC480 assay detected only one virus target (IP-4) in 6 of these samples (initial Ct: 32–35) (cf. Table S2 in Supplementary materials).
The 25 (Ct > 35) positive samples detected by the Panther Fusion™ assay included 20 (80 %) samples that were positive with The Aptima™ assay and 17 (68 %) that were positive with the MagNA/LC480 assay. These latter included 13 (65 %) in which only one virus target (IP-4) was detected. Five (20 %) samples were negative with the Aptima™ assay and eight (32 %) with the MagNA/LC480 assay, four of these were negative with both assays. These nine samples were tested again with the Panther Fusion™ assay since sample storage can result in the partial degradation of virus RNA, and thus to reduced assay sensitivity. Four were negative and five were positive or inconclusive (three with only IP-2, one with only IP-4 and one with both targets) (cf. Table S2). These discrepancies could also be due to selection bias - samples were chosen based on the Panther Fusion™ assay results.
All the Panther Fusion™ assay negative samples were also negative with the Aptima™ assay, except for one invalid sample. One sample was positive with the MagNA/LC480 assay, but displayed atypical PCR curves. A second RT-PCR on the same nucleic acid extract was negative (sample A2 0139 3655, cf. Table S2).
4.3. Concordance between assays
The invalid rate, all results included, with the Aptima™ assay was 0.5 % (1/200). The Aptima™ and Panther Fusion™ assays gave a Cohen’s coefficient κ = 0.978 [0.949–1.000], a positive percent agreement = 97.3 % [95.0–99.6 %], a negative percent agreement = 100 % [100−100%] and an overall percent agreement = 99.0 % [97.6–100 %] after re-testing of discrepancies. The Aptima™ and MagNA/LC480 assays had a Cohen’s coefficient κ = 0.945 [0.898−0.993], a positive percent agreement = 98.6 % [96.9–100 %], a negative percent agreement = 96.9 % [94.5–99.3 %] and an overall percent agreement = 97.5 % [95.3–99.7 %]. The MagNA/LC480 and Panther Fusion™ assays had a Cohen’s coefficient κ = 0.956 [0.914−0.999], a positive percent agreement = 94.5 % [91.4–97.7 %], a negative percent agreement = 100 % [100−100%] and an overall percent agreement = 98.0 % [96.1–99.9 %] (cf. Table 1 ).
Table 1.
A. Aptima™ and Panther Fusion™ assays | |||
---|---|---|---|
Aptima™\Panther Fusion™ | Positive | Negative | Total |
Positive | 72 | 0 | 72 |
Negative | 2 | 125 | 127 |
Total | 74 | 125 | 199a |
B. Aptima™ and MagNA/LC480 assays | |||
---|---|---|---|
Aptima™\MagNA/LC480 | Positive | Negative | Total |
Positive | 68 | 4 | 72 |
Negative | 1 | 126 | 127 |
Total | 69 | 130 | 199* |
C. MagNA/LC480 and Panther Fusion™ assays | |||
---|---|---|---|
MagNA/LC480\Panther Fusion™ | Positive | Negative | Total |
Positive | 69 | 0 | 69 |
Negative | 4 | 127 | 131 |
Total | 73 | 127 | 200 |
One of the 200 samples was invalid with the Aptima™ assay.
The Aptima™ SARS-CoV-2 TMA assay was given emergency use authorization by the United States Food and Drug Administration [8] and the CE mark for in vitro diagnostic use in Europe.
We conclude that the results of the commercial Aptima™ SARS-CoV-2 TMA assay agreed very closely with those obtained with our in-house assays. It includes an internal control to detect inhibition of amplification. The Panther™ instrument also provides random access, which allows urgent samples prioritization to obtain results within 3.5 h with the Aptima™ assay, and can deliver up to 60 results per hour. This can help laboratories provide rapid, high-throughput diagnosis during the COVID-19 pandemic crisis.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Declaration of Competing Interest
None declared.
Footnotes
Supplementary material related to this article can be found, in the online version, at doi:https://doi.org/10.1016/j.jcv.2020.104541.
Appendix A. Supplementary data
The following is Supplementary data to this article:
References
- 1.Zhu N., Zhang D., Wang W., Li X., Yang B., Song J., Zhao X., Huang B., Shi W., Lu R., Niu P., Zhan F., Ma X., Wang D., Xu W., Wu G., Gao G.F., Tan W. China Novel Coronavirus Investigating and Research Team. 2020. A novel coronavirus from patients with pneumonia in China. N. Eng. J. Med. 2019;382:727–733. doi: 10.1056/NEJMoa2001017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.World Heanlth Organisation . 2020. 01 June 2020 2020. Coronavirus (COVID-19) Situation Report - 133.https://www.who.int/docs/default-source/coronaviruse/situation-reports/ Accessed 01 June 2020. [Google Scholar]
- 3.Corman V.M., Landt O., Kaiser M., Molenkamp R., Meijer A., Chu D.K., Bleicker T., Brunink S., Schneider J., Schmidt M.L., Mulders D.G., Haagmans B.L., van der Veer B., van den Brink S., Wijsman L., Goderski G., Romette J.L., Ellis J., Zambon M., Peiris M., Goossens H., Reusken C., Koopmans M.P., Drosten C. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020;25 doi: 10.2807/1560-7917.ES.2020.25.3.2000045. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.World Heanlth Organisation . 2020. 24 January 2020 2020. Molecular Assays to Diagnose COVID-19: Summary Table of Available Protocols.https://www.who.int/docs/default-source/coronaviruse/whoinhouseassays.pdf?sfvrsn=de3a76aa_2&download=true Accessed 10 February 2020. [Google Scholar]
- 5.Tang Y.W., Schmitz J.E., Persing D.H., Stratton C.W. Laboratory diagnosis of COVID-19: current issues and challenges. J. Clin. Microbiol. 2020;58 doi: 10.1128/JCM.00512-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.McHugh M.L. Interrater reliability: the kappa statistic. Biochem. Med. (Zagreb) 2012;22:276–282. [PMC free article] [PubMed] [Google Scholar]
- 7.2020. U.S. Food and Drug Administration. 27 March 2018 2007. Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests - Guidance for Industry and FDA Staff.https://www.fda.gov/regulatory-information/search-fda-guidance-documents/statistical-guidance-reporting-results-studies-evaluating-diagnostic-tests-guidance-industry-and-fda Accessed 27 May 2020. [Google Scholar]
- 8.2020. U.S. Food and Drug Administration. 21 May 2020 2020. Coronavirus Disease 2019 (COVID-19) Emergency Use Authorization Information - In Vitro Diagnostic Products.https://www.fda.gov/emergency-preparedness-and-response/mcm-legal-regulatory-and-policy-framework/emergency-use-authorization-covidinvitrodev Accessed 05 June 2020. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.