Abstract
Purpose:
Gadoxetic acid uptake rate (k1) obtained from DCE MRI (Dynamic, Contrast Enhanced MRI) is a promising measure of regional liver function. Clinical exams are typically poorly temporally characterized, as seen in a low temporal resolution (LTR) compared to high temporal resolution (HTR) experimental acquisitions. Meanwhile, clinical demands incentivize shortening these exams. This study develops a neural network based approach to quantification of k1, for increased robustness over current models such as the linearized single-input, two-compartment (LSITC) model.
Methods:
30 Liver HTR DCE MRI exams were acquired in 22 patients with at least 16 minutes of post-contrast data sampled at least every 13 seconds. A simple neural network (NN) with 4 hidden layers was trained on voxel-wise LTR data to predict k1. LTR data was created by subsampling HTR data to contain 6 time points, replicating the characteristics of clinical LTR data. Both the total length and the placement of points in the training data was varied considerably to encourage robustness to variation. A GAN (Generative Adversarial Network) was used to generate arterial and portal venous inputs for use in data augmentation based on the dual-input, two-compartment, pharmacokinetic model of gadoxetic acid in the liver. The performance of the NN was compared to direct application of LSITC on both LTR and HTR data. The error was assessed when subsampling lengths from 16 to 4 minutes, enabling assessment of robustness to acquisition length.
Results:
For acquisition lengths of 16 min NRMSE (Normalized Root-Mean-Squared Error) in k1 was 0.60, 1.77, and 1.21, for LSITC applied to HTR data, LSITC applied to LTR data, and GAN augmented NN applied to LTR data, respectively. As the acquisition length was shortened, errors greatly increased for LSITC approaches by several folds. For acquisitions shorter than 12 minutes the GAN augmented NN approach outperformed the LSITC approach to a statistically significant extent, even with HTR data.
Conclusions:
The study indicates that data length is significant for LSITC analysis as applied to DCE data for standard temporal sampling, and that machine learning methods, such as the implemented NN, have potential for much greater resilience to shortened acquisition time than directly fitting to the LSITC model.
Keywords: Liver Function, Quantitative Imaging, GAN
1. INTRODUCTION
Gadoxetic acid enhanced dynamic MRI has been shown to have promising applications in the assessment of liver function1–6 and diagnosis of various pathologies in the liver7–12. Gadoxetic acid provides utility as a hepatobiliary contrast, allowing interrogation of the uptake of contrast into the hepatocytes as well as liver perfusion parameters. Various pharmacokinetic parameters have been used as a measure of regional liver function1,13–15 with gadoxetic acid uptake rate being among the most direct due to its correspondence with the number of functioning hepatocytes, making it a reasonable quantitative measure of regional liver function6,16. Quantification of regional liver function is important in functional avoidance therapy, where radiation therapy is optimized to spare highly functional regions of the liver17,18. Many models exist for the analysis of contrast kinetics in MRI19–22. Fewer models are specifically applicable for determining gadoxetic acid uptake rate in the liver, including the dual-input, two-compartment (DITC) model of gadoxetic acid kinetics, and the DITC derived linearized single-input, two-compartment model (LSITC)3,23. Most models are applicable to the high temporal resolution (HTR) dynamic, contrast enhanced (DCE), scans that collect volumes regularly enough to well characterize the concentration across time in the relevant regions, typically sampling every 5 to 15 seconds. However, the most common clinical gadoxetic acid enhanced MRI exams do not sample this comprehensively. Clinical multiphase scans are obtained for metastases detection and diagnosis. These clinical exams typically have low temporal resolution (LTR), with as few as 6 volumes irregularly sampling 20 minutes of contrast kinetics. It should also be noted that clinical demands inevitably incentivize shortening exams. If quantification accuracy can be maintained or improved while shortening total acquisition time and eliminating the need for constant acquisition (e.g. LTR style acquisitions), the patient can be given equivalent care with less inconvenience and discomfort, and minimal change to common clinical workflows.
This motivates the development of methods for accurate quantification of regional liver function from short and poorly characterized DCE MRI exams in a robust manner. This study develops an artificial neural network (NN) approach to predict k1 from LTR data. Furthermore, this approach uses data augmentation from a generative adversarial network (GAN) implemented to allow realistic and varied simulation of gadoxetic acid dynamics from the DITC model of gadoxetic acid kinetics in the liver. These approaches are compared to least squares fitting of the LSITC model3 as applied to both HTR and LTR data. We hypothesize that the new NN approach allows faster and more convenient acquisition without a sacrifice to the accuracy of functional maps sufficient to compromise treatment guidance.
2. METHODS
A NN based approach is developed to predict k1 from LTR data derived from DCE scans. To counter the inherent granularity of the underlying input functions a GAN is used to generate input functions for the augmentation of NN training. The NN based approaches are compared to LSITC analysis for both well characterized HTR data, and the more limited LTR data with varied acquisition duration to assess robustness of the approaches.
2.1. Models
The dual-input, two-compartment (DITC) model (Figure 1) of gadoxetic acid in the liver describes the contrast concentration dynamics in the liver at a given time as determined by the uptake rate (k1), distribution volume (vdis), arterial rate (ka), portal venous rate (kpv), and the respective portal venous and arterial blood arrival delays (Tpv and Ta)3,23. This allows simulation of concentration for any given set of parameters and inputs, or fitting of the observed output to find the likely input parameters.
If prediction of uptake is of chief interest, a simpler linearized single-input, dual-compartment (LSITC) model can fit to the observed data. This LSITC model is derived from the DITC model, but allows for more robust and rapid analysis over more limited datasets. Whereas fitting the DITC model involves 6 tunable parameters, with appropriate assumptions it collapses to the LSITC model described by the following 2 parameter linear equation3.
(1) |
where Ct is the measured contrast concentration in the region of interest, Ca is the concentration in the arterial blood supply, k1 is the contrast uptake rate, vdis is the volume-normalized volume of distribution, and Hct is the hematocrit. Since Ct, Ca and Hct can be measured or estimated, the parameters to fit are k1 and vdis. This model applies after some point in time t0 when the model assumptions hold. Thus, after t0, k1 and vdis can be easily computed through a linear regression of the relevant data formulated as the vectors x and y.
2.2. Data acquisition
In order to assess error across analysis types and data characteristics, 3D volumetric DCE MRI of the liver were acquired during the intravenous injection of a single standard dose of gadoxetic acid using a Golden-Angle Radial sampling VIBE sequence on a 3T scanner (Skyra, Siemens Healthineer) in a prospective protocol approved by University of Michigan Institutional Review Board. 30 exams were acquired over a set of 22 patients (Age: 50 to 82 years, 6 female) with hepatocellular carcinoma. The 3D free-breathing DCE images of the liver were acquired using a 3D golden-angle radial stack-of-stars VIBE sequence. This sequence over-samples the center of k-space, and allows greater resilience to motion effects than other sequences24. The time-series images were co-registered within the liver VOI using an over-determined, rigid-body transformation approach25. All acquisitions continued for 16–20 minutes after injection of a single-dose gadoxetic acid contrast and had temporal resolutions of at least 5 samples per minute.
The acquired HTR data was subsampled to produce corresponding LTR data (Figure 2). This was done by interpolating (1) a pre-contrast volume, (2) 3 volumes spaced 25 seconds apart designed to capture the arterial and portal venous phases, and (3) two volumes at the end and midpoint of the acquisition (roughly 20 and 10 min, respectively).
Ca, Cpv, and Ct were obtained as described in a prior study3. In brief, the arterial concentration (Ca) was defined by the mean 100 voxels with the maximum value just prior to the arterial peak, and selected from the three inches of aorta just prior to the aortic split to the liver.
The portal venous concentration (Cpv) was defined analogously based on a contour of the portal vein. In both cases relative enhancement was used to create the input functions:
(2) |
where C(iT) is the relevant concentration at time point i, given a sampling interval of T, and SIi and SIprecontrast are the average signal intensities in the given region at time point i, and prior to contrast enhancement respectively. The same calculation was performed for each voxel in the liver to obtain the tissue concentration (Ct).
2.3. Least squares fitting of LSITC model
LSITC analysis involved linear regression for the best fit to equation (1). For HTR data t0 was selected to maximize the linearity of the time range being fit, as described in prior work3. In the analysis of the synthetic LTR data, t0 was chosen 75 seconds after the initial upswing of the arterial peak. In both cases the resulting estimate of the k1 was the intercept normalized slope of the least squares linear fit from t0 to the final point. This allowed the linear fit to incorporate 3 points for the LTR data.
2.4. Neural network – rationale and implementation
Given a reasonable set of patients with k1 estimated from HTR data, a machine learning approach is a natural means for creating a prediction from a subset of that data, e.g., multiphase LTR data. To this end, a simple fully connected neural network (NN) with 4 hidden layers (10,10, 5 and 5 neurons) was trained on voxel-wise LTR data to predict k1 (Figure 3). Both the total acquisition length and the placement of points in the training data were varied considerably to encourage robustness to variation. This was performed by having the arterial and portal venous phase points sampled uniformly 15 to 50 seconds apart, with uniformly distributed perturbation up to 10% of the sampling period. The endpoint tend was randomly selected from a uniform distribution from 5 minutes after the arterial upswing until the end of the acquisition. The midpoint sample was selected from a uniform distribution from 0.25tend to 0.75tend. Each voxel then consisted of 5 pairs of values representing the x and y vectors calculated from equation (1) based on 5 post-contrast time points (as in the right panel of Figure 2).
Training was performed by randomly selecting 3 million voxels in the livers from 30 exams, holding 3/5ths for training, 1/5th for validation, and 1/5th for testing. Training and validation data did not have patients that overlapped with the patients in the data held for testing.
2.5. GAN
2.5.1. GAN - rationale
No matter how many voxels are used for training, if we have only a pool of 30 exams, and 22 patients, each voxel will come from one of 30 categories defined by the precise input functions that corresponded to that exam. This inspires data augmentation for the set of input functions to ensure the training data is better spread across the reasonable space of input functions. A GAN is a reasonable choice for this generative task. This approach trains both a generator and a discriminator, who act as adversaries to one another. The generator seeks to generate artificial input functions that are in the space of real input functions. The discriminator attempts to discriminate between the real examples and those generated artificially. Eventually, the generated examples should be essentially indistinguishable from examples drawn from the true dataset. GANs have been applied in a number of circumstances, involving both temporal biological signal26 and medical image27,28 generation, including generation for data augmentation29. Here we use a generative adversarial neural network to generate arterial and portal venous input functions for gadoxetic acid dynamics in the liver.
2.5.2. GAN design and implementation
The GAN consisted of a simple network for conversion of a random vector (length 20) into outputs corresponding to arterial (Ca) and portal venous (Cpv) input functions (two vectors of length 100) along with an indicator of the sampling period T. The network architecture can be seen in figure 4.
The generated input functions are then used as to create tissue concentration curves (Ct) using the DITC model.
2.5.3. NN augmentation from GAN data
Training using the GAN generated data serves a dual purpose – firstly it acts as a confirmation that the GAN generated data is actually representative of the real Ca and Cpv curves, secondly, it could improve prediction accuracy with comparatively minimal chance of overfitting, based on the increased variability in Ca and Cpv for the training data. This dataset then has ground truth DITC defined uptake rates with input functions replicating the variation observed empirically. This data can be used to augment the real data in training neural models to determine uptake from restricted datasets.
In order to train a network to generated Ca and Cpv curves from a random vector, training data was created by first generating 1 million random Ca and Cpv pairs with corresponding T. This was performed for 5 holdout groups of patients corresponding to the training holdout groups described in 2.4 to ensure the learned sets were not influenced by testing patients’ own data. For each of these sets of Ca and Cpv curves, k1 and vdis values were randomly selected from the relevant patient set (excluding holdout patients), while ka, kpv, Ta and Tpv were randomly selected from roughly physiologically reasonable ranges (see table 1). Ct curves were then generated from the DITC model using the GAN generated Ca and Cpv functions along with the random parameters described in table 1 as inputs to the model. Finally, gaussian distributed noise was added such that the measured SNR would be 40 dB.
Table 1.
Parameter | Distribution | |
---|---|---|
k1,vdis | Randomly drawn from patient set | mL/100mL/min, mL/mL |
kpvp+ kap | U(50,300) | mL/100mL/min |
kpvp | N(0.75, (1/16)2)(kpvp+ kap), | mL/100mL/min |
2.5.4. LSITC optimization from GAN data
Finally, consideration was given to minimize the error in LSITC analysis. The two obvious “tunable” parameters are t0 and sampling time. The parameter t0 refers to the first time point considered to satisfy the conditions of the LSITC model and thus used as the first point in the linear fit of the model. This is currently selected through a maximization of linearity as calculated by the ratio of singular values3. Determination of the sampling times is more complex, particularly if we implement irregular sampling as in LTR collection. This study uses the GAN simulated data to optimize t0 and the sampling times, discretized in 30 second increments, for the LSITC analysis. Optimization is performed using a genetic algorithm to search for t0 and sampling times. Breaking the signal into 30 seconds intervals increased the tractability of the problem for this discrete genetic algorithm. This resulted in each of the sampling points being chosen from 32 intervals of 30 seconds in the 16 minute datasets, where the first and last points are required. This was performed for 1 to 10 additional points, where the choice of points was optimized to minimize MSE error in a set of GAN based DITC generated synthetic voxels.
2.6. Error metric for evaluation of analysis methods and acquisition paradigms
For each method and dataset used to estimate k1, the error was measured as NRMSE with the results of least squares fitting of the LSITC model for the full length (16–21 min) HTR dataset as the reference. NRMSE is defined here as RMSE normalized on an exam by exam basis by the interquartile range of the reference values as:
(3) |
Mean NRMSE is merely the mean across all exams analyzed.
The reference values were restricted to the values with a relative uncertainty below the 75th percentile. This minimizes the likelihood of performing the comparison with outliers and artifacts, such as those seen on some edges, but will also tend to decrease the denominator in the NRMSE calculation.
Relative uncertainty was measured as the expected standard deviation in k1 estimation for the fit in a given voxel divided by the predicted k1 for that voxel. Here the variance in k1 is estimated by the Taylor expansion of the variation of K1/vdis (where K1 is the slope in equation 1) as:
(4) |
where
(5) |
(6) |
Where σa and μa are the respective standard deviations and means of any given measure a. x and y are defined in equation 1.
All results from five methods and datasets were compared to the k1 estimated by fitting the LSITC model for HTR data at maximum length (at least 16 minutes and no more than 21 minutes), which are summarized in table 2.
Table 2.
Method/Data Abbreviation | Method Description | Input Data Description |
---|---|---|
LSITC-HTR | Fitting of LSITC model with t0 chosen to maximize linearity | HTR data, with the data length truncated to a maximum length of 4 to 16 minutes |
LSITC-LTR | Fitting of LSITC model with t0=75 seconds | LTR data, with the data length truncated to a maximum length of 4 to 16 minutes. The initial points spaced at 25 second intervals. |
NN-LTR | Application of the NN model trained by k1 resulting from LSITC-HTR for full HTR datasets | LTR data, with the data length truncated to a maximum length of 4 to 16 minutes. The initial points spaced at 25 second intervals. |
Augmented NN-LTR | Application of the NN model trained by DITC based data using input functions generated by GAN. | LTR data, with the data length truncated to a maximum length of 4 to 16 minutes. The initial points spaced at 25 second intervals. |
OPT LSITC-LTR | Fitting of LSITC model with algorithmically chosen sampling times and t0 | 8 points selected algorithmically to minimize error in augmented dataset. Truncated to a maximum length of 8 to 16 minutes. |
LSITC HTR t0 = OPT | Fitting of LSITC model with HTR data but t0 set to the optimum found in OPT LSITC-LTR | HTR data, with the data length truncated to a maximum length of 4 to 16 minutes |
3. RESULTS
3.1. Fitting of LSITC model
As expected, directly fitting the LSITC model to HTR data yielded more accurate k1 values than fitting to LTR data. For both datasets the errors grew rapidly with a decrease in the acquisition length of the data (see figure 5). At full acquisition length (16 minutes), LSTIC-HTR and LSITC-LTR resulted in an average NRMSE across exams of 0.60 (SD 0.38) and 1.77 (0.99), respectively. At an acquisition length of 10 min the average NRMSE increased to 2.59 (1.34) and 3.09 (1.54) for HTR and LTR datasets, respectively, as seen in table 3. A visual comparison at 10 minutes can be seen in figure 6.
Table 3.
Series Duration (min) | NRMSE - mean (standard deviation) | |||||
---|---|---|---|---|---|---|
LSITC HTR | LSITC LTR | NN LTR | Augmented NN LTR | OPT LSITC LTR | LSITC HTR t0 = OPT | |
4 | 7.17 (4.39) | 7.21 (3.93) | 2.44 (2.06)* | 2.15 (1.78)* | 14.64 (9.44) −* | |
5 | 5.86 (3.47) | 6.21 (3.24) | 2.21 (1.79)* | 1.91 (1.43)* | 7.78 (4.58) | |
6 | 4.68 (2.72) | 5.01 (2.95) | 2.04 (1.52)* | 1.82 (1.16)* | 5.27 (3.18) | |
8 | 3.27 (1.79) | 4.02 (2.35) | 1.71 (1.15)* | 1.52 (0.85)* | 1.97 (1.39)* | 3.05 (1.78) |
10 | 2.59 (1.34) | 3.09 (1.54) | 1.54 (0.93)* | 1.41 (0.75)* | 1.38 (0.72)* | 2.23 (1.17) |
12 | 1.81 (1.08) | 2.57 (1.39) −* | 1.44 (0.79) | 1.32 (0.67)* | 1.07 (0.57)* | 1.60 (0.99) |
14 | 1.31 (1.01) | 2.05 (1.27) −* | 1.32 (0.71) | 1.24 (0.62) | 0.90 (0.53) | 1.14 (0.90) |
15 | 0.92 (0.61) | 1.79 (1.07) −* | 1.28 (0.68)−* | 1.24 (0.63) | 0.86 (0.62) | |
15.5 | 0.78 (0.52) | 1.80 (1.02) −* | 1.25 (0.64) −* | 1.20 (0.58) −* | 0.76 (0.54) | |
16 | 0.60 (0.38) | 1.77 (0.99) −* | 1.22 (0.69) −* | 1.21 (0.66) −* | 0.77 (0.42) | 0.68 (0.50) |
Max | 0.00 (0.00) | 1.39 (0.80) −* | 1.14 (0.58) −* | 1.06 (0.56) −* | 0.72 (0.33) −* | 0.42 (0.26) −* |
3.2. NN model
The NN model yielded significantly reduced error rates in k1 estimation over direct fitting of the LSITC model to the LTR data at all tested acquisition lengths (4–20 min). When the acquisition length was less than 14 min, the NN model applied to the LTR data resulted in the errors less than directly fitting of the LSITC model to the HTR data. This difference became significant for acquisitions of 10 minutes or less. The errors yielded by the NN model increased slowly with the acquisition length reduction, suggesting the NN model was resilient to data length. In contrast, direct fitting of the LSITC model yielded quickly increased errors with the data length reduction, regardless of the temporal resolution of the data (figure 5).
3.3. GAN augmented NN model
On visual inspection randomly selected curves generated by the trained GAN seemed to replicate the basic features of the measured curves without being direct copies of individual examples. For randomly selected GAN generated Ca curves, the nearest normalized neighbor was found from the measured set of input curves. Three examples are shown in figure 7. In each column the top plot is a randomly selected generated Ca and Cpv pair, and the bottom plot is the real Ca and Cpv pair whose normalized Ca curve is the nearest neighbor to the generated Ca curve based on a sum of squares difference. The comparisons did not show evidence of direct replication of the specifics of particular measured curves.
In addition to visual inspection, the distribution created by the GAN was assessed by producing histograms approximating the probability distribution of the pairwise Euclidean differences between examples within the measured data, as well as the pairwise differences in data generated by each GAN. Figure 8 displays these distributions of pairwise differences for each GAN, superimposed over the distribution of pairwise differences for the measured data. The difference between the mean distance for each GAN and the measured data is shown, along with the earth-movers-distance (EMD) to better represent the differences between distributions. In all cases the distribution of differences in GAN data visually mirrors that of the full dataset, with the smoothing we would expect from a larger number of samples from a similar dataset.
Augmentation with GAN generated data gave mixed results. Training on only synthetic data resulted in improvement over prediction error from training only on real data (figure 5). With a statistically significant improvement in error over LSITC HTR for all datasets of length 12 minutes or less, and no significant drop in error up to 15 minutes. However, combining the real data with additional synthetic data did not meaningfully improve the prediction error. The results of augmented NN model trained by synthetic data only are shown in figure 5 and table 3 (Augmented NN-LTR).
3.4. Optimization of time points for the LSITC model fitting
When selecting the optimum sampling points for the LSITC model fitting, as additional points were selectively added to the set, optimization yielded a t0 of 3 minutes in every case, without any sampling point prior to t0. The sampling times chosen tended to group just after t0, and near the end of the dataset. The error leveled off near 8 points in the simulated data, as seen in figure 9. As a result, 8 points were used when testing this approach, apart from the pre-contrast and final points.
Implementation of the GAN data for LSITC optimization (OPT-LSITC LTR) yielded errors significantly lower than direct fitting of the LSITC model to HTR data with acquisition lengths of 12 min or less, and lower than NN models for data lengths greater than 10 minutes (figure 5). This suggests that optimization of the time of data point acquisition could improve the performance of the LSITC model, but the NN model with non-optimized data still could perform better at a short acquisition length.
A further test of the optimal t0 (3 min) was performed with full HTR data. As seen in figure 5, the LSITC model fitting to HTR with a dynamic t0 (LSITC-HTR) and an optimal t0 (LSITC-HTR t0=OPT) yielded similar results, but worse results than the LSITC model fitting to the optimal 8-point LTR data (OPT LSITC LTR), indicating that the robustness of performance of the optimized LSITC is not merely due to the choice of t0 but due to the particular set of points selected.
4. DISCUSSION
In this study, we developed NN models for estimation of k1 and compared the results of the NN models to those from direct fitting of the LSITC model for various acquisition lengths and temporal resolutions of Gadoxetic acid enhanced dynamic MRI of the liver. Overall, the NN models are more resilient to the acquisition length reduction. The augmented input functions using GAN can further improve the performance of the NN models. For direct fitting of the LSITC model, ten optimized time points in the Gadoxetic acid enhanced dynamic data can significantly out-perform the HTR data (5–10 sec per volume) for acquisition lengths of 12 minutes or less, and the NN method for acquisition length not shorter than 8 minutes. Our study suggests that the NN approach can be used to enhance the performance of k1 estimation and optimize the data acquisition.
A key element of modeling liver pharmacokinetics is obtaining arterial and portal venous input functions. These input functions have been estimated using combinations of exponentials and other simplifications, but this involves either profound simplification or the development of models of increased complexity without a guarantee of successfully capturing the relevant features of the input functions. Use of measured input functions has notable advantages in capturing the true empirical characteristics of these input functions. However, when employing data driven methods this will practically limit the researcher to a relatively small number of example cases. When machine learning methods are applied to millions of voxels but the guiding input functions consist of a few dozen examples, we may fear overlearning these limited underlying examples, rather than a more useful learning of the underlying relationships between our relevant parameters and input functions in general. Addition of noise or variation in sampling time may make this underlying granularity less starkly memorable. However, a more ideal solution would be the construction of arbitrary or random input functions from the feature space the input functions inhabit. A promising means for this generative task is a generative adversarial neural network.
One difficulty in generative networks, where the network is not cyclic (generating corresponding examples in another space rather than arbitrary or random examples in the desired space) is assessment of the quality of the generative model. One approach is the usage of these examples as augmentation data for a relevant learning task. If the augmentation helps, it is more reasonable that the generative model is representing the variation in the underlying set appropriately, or at least in a way that helps the trained network to better understand the relevant relationships. Here we used a generative adversarial neural network to generate arterial and portal venous input functions for gadoxetic acid kinetics in the liver.
The augmented NN that was trained only on GAN generated data resulted in superior results as compared to the NN trained using any fraction of the measured data with HTR-LSITC as the reference. There are various possible causes of the decrease in performance with the addition of real data. It is likely that the very few input functions were not useful in further generalizing the solution over the training from the GAN and DITC generated data. It may also have skewed the solution towards those measured input functions. It should be noted that since the GAN itself is trained from measured data, the generated examples will include characteristics caused by sampling noise, movement and other variations in the data. Because of this the input to the DITC model generated from this GAN has variation that would not be expected in the underlying input functions in reality.
In addition to the already mentioned benefits, the GAN derived data and DITC model defined reference values allowed the simulated dataset to be used to evaluate independent models relative to the DITC model. This allowed us to use references of not only our best estimate (whether DITC or LSITC) to complete (16+ min) real datasets, but also to the ground truth inputs to the DITC model without fitting error in the reference k1 values. This helps quantify possible error in these estimates and gives a parallel reference measure for restricted methods. This is of particular interest when attempting to assess optimum, or at least improved, acquisition times for the image volumes used to estimate k1. Use of these model defined input parameters made this optimization less susceptible to a mere reproduction of the linear fit of the LSITC model (along with any limitations or errors in this method), and helped to assess the best timing (giving the variability observed in the input functions) to acquire points for LSITC without bias to the timing used in the measured reference set.
The optimal sampling points for OPT-LSITC LTR essentially followed the expected weights for a linear regression, in that points near the end were preferred, with successive trials adding points closer to the center as those at the ends were already included. The selection of t0 is perhaps more salient, indicating that the addition of a point near the 3 min mark would aid LSITC accuracy when applied to LTR data. This time roughly corresponds to the equilibrium phase30, which would logically initiate the portion of the data where the assumptions of the LSITC model hold true. This approach resulted in lower error than even LSITC applied to HTR data from 15.5 to 8 min, for the real dataset, even though the reference was used HTR data with a variable t0. This also casts doubt on the use of 75 seconds as t0 in LTR data. If 3 minutes is the location of the equilibrium phase, then voxel-wise LSITC analysis of most LTR data has only 2 data points to work with, since none of the arterial or portal venous phase points will fall after that point. Without an overdetermined fit the error rates will likely be large, and concurrent error quantification will rely on assumptions regarding the similarity of nearby points. However, the selection of t0 was not the primary factor in the improvement over other LSITC methods. This is apparent from the small difference between LSITC-HTR and LSITC-HTR where t0 = OPT. This indicates that the specific selection of points was helpful in improving the fit. It is possible that some of the improvement came from selecting no points prior to t0. This does not change which points are fit, but does change the x and y vectors since the integral of Ca will differ in equation 1. It may be that the discrepancy of Ca from Cpv increases the error in datasets that include pre-t0 sampling points.
Regardless of the method used the error was greater for shorter datasets. Data length was especially significant for LSITC analysis, for both LTR and HTR data. With a fixed best t0 and careful choice of sampling points this was reduced somewhat, perhaps making acquisitions as short as 12 minutes practical. Below this level the NN methods worked best, showing relatively little change in error with data length in time. This indicates that the underlying information is sufficient for a comparatively accurate prediction even with relatively short collection time used by the NN. However, the results did not outperform LSITC-HTR for long datasets. In each of these cases it is important not to interpret the error in absolute terms, particularly near the maximum length. Remember that the error measures will be impacted by error in the results of LSITC applied to HTR.
Use of the LSITC model as the reference allowed rapid analysis and comparison with regard to k1, even for LTR data. In a previous study, k1 values estimated from the LSITC and DITC have been compared and the results are very similar3. However, this model does omit parameters present in the DITC model, notably ka and kpv. Previous studies have correlated portal venous perfusion to liver function13 and arterial perfusion to tumor presence23. Theoretically, simultaneous quantification of k1, ka and kpv from a single dynamic MRI acquisition using the DITC is advantageous. Practically, there are some limitations. The FDA approved standard dose of Gadoxetic acid only contains a quarter of the Gadolinium in a standard dose of Gd-DTPA or Multihance. This results in a weak contrast enhancement and a low signal-to-noise ratio in the arterial phase signals, thereby challenging reliable quantification of arterial perfusion. Therefore, in practice, if tumor diagnosis and assessment are the primary interest, Gd-DTPA or Multihance is used. If liver function measurement is the primary interest, Gadoxetic acid is used. If both tumor assessment and liver function are of interest, a trade-off has to be made. Compared to the DITC and LSITC models, the Tofts model only considers the contrast transport between the intra-vascular and the extra-cellular space, which can only be applied for an extra-cellular contrast agent, but not an intra-cellular agent, like Gadoxetic acid.
5. CONCLUSIONS
Data length is significant for LSITC analysis as applied to DCE data for standard temporal sampling. With a fixed best t0 and careful choice of sampling points this can be reduced somewhat, particularly for acquisitions at least 12 minutes in length. Below this level the NN worked best, indicating that NN methods may be helpful in improving the robustness of uptake analysis in temporally short datasets. Combination of a GAN with DITC model created data contributed to the training of the NN, indicating the variation in input functions was being appropriately represented. Further work should assess the impact on functional avoidance therapy dependent on the means used to create functional maps.
7. ACKNOWLEDGMENTS
This work is supported in part by NIH grants of R01 CA132834 and P01 CA059827. The authors thank Siemens Healthineer for providing the Radial VIBE pulse sequence
Table of abbreviations appearing in text with corresponding definitions.
- 3D
Three-dimensional
- DCE
Dynamic, Contrast Enhanced
- DITC
Dual-Input, Two-Compartment
- EMD
Earth Mover’s Distance
- GAN
Generative Adversarial Network
- Hct
Hematocrit
- HTR
High Temporal Resolution
- LTR
Low Temporal Resolution
- LSITC
Linearized Single-Input, Two-Compartment
- MRI
Magnetic Resonance Imaging
- MSE
Mean Squared Error
- NN
Neural Network
- NRMSE
Normalized Root Mean Squared Error
- SD
Standard Deviation
- SNR
Signal to Noise Ratio
6 REFERENCES
- 1.Yoon JH, Lee JM, Kang H, et al. Quantitative Assessment of Liver Function by Using Gadoxetic Acid–enhanced MRI: Hepatocyte Uptake Ratio. Radiology. 2018;290(1):125–133. doi: 10.1148/radiol.2018180753 [DOI] [PubMed] [Google Scholar]
- 2.Georgiou L, Penny J, Nicholls G, et al. Quantitative Assessment of Liver Function Using Gadoxetate-Enhanced Magnetic Resonance Imaging. Invest Radiol. 2017;52(2):111–119. doi: 10.1097/RLI.0000000000000316 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Simeth J, Johansson A, Owen D, et al. Quantification of liver function by linearization of a two-compartment model of gadoxetic acid uptake using dynamic contrast-enhanced magnetic resonance imaging. NMR in Biomedicine. 31(6):e3913. doi: 10.1002/nbm.3913 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Leporq B, Daire J-L, Pastor CM, et al. Quantification of hepatic perfusion and hepatocyte function with dynamic gadoxetic acid-enhanced MRI in patients with chronic liver disease. Clinical Science. 2018;132(7):813–824. doi: 10.1042/CS20171131 [DOI] [PubMed] [Google Scholar]
- 5.Nilsson H, Blomqvist L, Douglas L, Nordell A, Jonas E. Assessment of liver function in primary biliary cirrhosis using Gd-EOB-DTPA-enhanced liver MRI. HPB. 2010;12(8):567–576. doi: 10.1111/j.1477-2574.2010.00223.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Haimerl M, Schlabeck M, Verloh N, et al. Volume-assisted estimation of liver function based on Gd-EOB-DTPA–enhanced MR relaxometry. Eur Radiol. 2016;26(4):1125–1133. doi: 10.1007/s00330-015-3919-5 [DOI] [PubMed] [Google Scholar]
- 7.Joo I, Lee JM. Recent Advances in the Imaging Diagnosis of Hepatocellular Carcinoma: Value of Gadoxetic Acid-Enhanced MRI. Liver Cancer. 2016;5(1):67–87. doi: 10.1159/000367750 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Nishie A, Asayama Y, Ishigami K, et al. MR prediction of liver fibrosis using a liver-specific contrast agent: Superparamagnetic iron oxide versus Gd-EOB-DTPA. Journal of Magnetic Resonance Imaging. 2012;36(3):664–671. doi: 10.1002/jmri.23691 [DOI] [PubMed] [Google Scholar]
- 9.Inchingolo R, Faletti R, Grazioli L, et al. MR with Gd-EOB-DTPA in assessment of liver nodules in cirrhotic patients. World J Hepatol. 2018;10(7):462–473. doi: 10.4254/wjh.v10.i7.462 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Liu X, Jiang H, Chen J, Zhou Y, Huang Z, Song B. Gadoxetic acid disodium–enhanced magnetic resonance imaging outperformed multidetector computed tomography in diagnosing small hepatocellular carcinoma: A meta-analysis. Liver Transplantation. 2017;23(12):1505–1518. doi: 10.1002/lt.24867 [DOI] [PubMed] [Google Scholar]
- 11.Li J, Li X, Weng J, et al. Gd-EOB-DTPA dynamic contrast-enhanced magnetic resonance imaging is more effective than enhanced 64-slice CT for the detection of small lesions in patients with hepatocellular carcinoma. Medicine (Baltimore). 2018;97(52). doi: 10.1097/MD.0000000000013964 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ba‐Ssalamah A, Bastati N, Wibmer A, et al. Hepatic gadoxetic acid uptake as a measure of diffuse liver disease: Where are we? Journal of Magnetic Resonance Imaging. 2017;45(3):646–659. doi: 10.1002/jmri.25518 [DOI] [PubMed] [Google Scholar]
- 13.Cao Y, Wang H, Johnson TD, et al. Prediction of Liver Function by Using Magnetic Resonance-based Portal Venous Perfusion Imaging. Int J Radiat Oncol Biol Phys. 2013;85(1):258–263. doi: 10.1016/j.ijrobp.2012.02.037 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Yamada A. Quantitative Evaluation of Liver Function Within MR Imaging In: El-Baz AS, Saba L, Suri J, eds. Abdomen and Thoracic Imaging. Springer; US; 2014:233–251. doi: 10.1007/978-1-4614-8498-1_9 [DOI] [Google Scholar]
- 15.Truhn D, Kuhl CK, Ciritsis A, Barabasch A, Kraemer NA. A New Model for MR Evaluation of Liver Function with Gadoxetic Acid, Including Both Uptake and Excretion. Eur Radiol. 2019;29(1):383–391. doi: 10.1007/s00330-018-5500-5 [DOI] [PubMed] [Google Scholar]
- 16.Verloh N, Haimerl M, Zeman F, et al. Assessing liver function by liver enhancement during the hepatobiliary phase with Gd-EOB-DTPA-enhanced MRI at 3 Tesla. Eur Radiol. 2014;24(5):1013–1019. doi: 10.1007/s00330-014-3108-y [DOI] [PubMed] [Google Scholar]
- 17.Bennink RJ, Cieslak KP, van Delden OM, et al. Monitoring of Total and Regional Liver Function after SIRT. Front Oncol. 2014;4. doi: 10.3389/fonc.2014.00152 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Wu VW, Epelman MA, Wang H, et al. Optimizing global liver function in radiation therapy treatment planning. Phys Med Biol. 2016;61(17):6465–6484. doi: 10.1088/0031-9155/61/17/6465 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Sourbron SP, Buckley DL. Tracer kinetic modelling in MRI: estimating perfusion and capillary permeability. Phys Med Biol. 2012;57(2):R1–33. doi: 10.1088/0031-9155/57/2/R1 [DOI] [PubMed] [Google Scholar]
- 20.Sourbron SP, Buckley DL. Classic models for dynamic contrast-enhanced MRI. NMR Biomed. 2013;26(8):1004–1027. doi: 10.1002/nbm.2940 [DOI] [PubMed] [Google Scholar]
- 21.Ewing JR, Bagher-Ebadian H. Model selection in measures of vascular parameters using dynamic contrast-enhanced MRI: experimental and clinical applications. NMR Biomed. 2013;26(8):1028–1041. doi: 10.1002/nbm.2996 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Khalifa F, Soliman A, El-Baz A, et al. Models and methods for analyzing DCE-MRI: a review. Med Phys. 2014;41(12):124301. doi: 10.1118/1.4898202 [DOI] [PubMed] [Google Scholar]
- 23.Sourbron S, Sommer WH, Reiser MF, Zech CJ. Combined quantification of liver perfusion and function with dynamic gadoxetic acid-enhanced MR imaging. Radiology. 2012;263(3):874–883. doi: 10.1148/radiol.12110337 [DOI] [PubMed] [Google Scholar]
- 24.Chandarana H, Block TK, Rosenkrantz AB, et al. Free-Breathing Radial 3D Fat-Suppressed T1-Weighted Gradient Echo Sequence: A Viable Alternative for Contrast-Enhanced Liver Imaging in Patients Unable to Suspend Respiration. Investigative Radiology. 2011;46(10):648–653. doi: 10.1097/RLI.0b013e31821eea45 [DOI] [PubMed] [Google Scholar]
- 25.Johansson A, Balter J, Feng M, Cao Y. An Overdetermined System of Transform Equations in Support of Robust DCE-MRI Registration with Outlier Rejection. Tomography: A Journal for Imaging Research. 2016;2(3):188–196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Hartmann KG, Schirrmeister RT, Ball T. EEG-GAN: Generative adversarial networks for electroencephalograhic (EEG) brain signals. arXiv:180601875 [cs, eess, q-bio, stat]. June 2018. http://arxiv.org/abs/1806.01875. Accessed June 19, 2019. [Google Scholar]
- 27.Emami H, Dong M, Nejad-Davarani SP, Glide-Hurst CK. Generating synthetic CTs from magnetic resonance images using generative adversarial networks. Med Phys. June 2018. doi: 10.1002/mp.13047 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ren Y, Zhu Z, Li Y, et al. Mask Embedding for Realistic High-Resolution Medical Image Synthesis In: Shen D, Liu T, Peters TM, et al. , eds. Medical Image Computing and Computer Assisted Intervention – MICCAI 2019. Vol 11769 Cham: Springer International Publishing; 2019:422–430. doi: 10.1007/978-3-030-32226-7_47 [DOI] [Google Scholar]
- 29.Frid-Adar M, Diamant I, Klang E, Amitai M, Goldberger J, Greenspan H. GAN-based Synthetic Medical Image Augmentation for increased CNN Performance in Liver Lesion Classification. Neurocomputing. 2018;321:321–331. doi: 10.1016/j.neucom.2018.09.013 [DOI] [Google Scholar]
- 30.Niendorf E, Spilseth B, Wang X, Taylor A. Contrast Enhanced MRI in the Diagnosis of HCC. Diagnostics (Basel). 2015;5(3):383–398. doi: 10.3390/diagnostics5030383 [DOI] [PMC free article] [PubMed] [Google Scholar]