Abstract
Background:
Polycaprolactone (PCL) implants show isovolemic degradation during phase-1 degradation; they maintain their volume as their molecular weight decreases. Phase-2 begins with PLC volume being reduced by bulk degradation with autocatalysis. Isovolemic degradation of PCL particles during phase 1 and their longevity should be established in humans. PCL particle size can be mathematically calculated through cross-sectioned PCL particles in biopsy slides.
Methods:
Biopsy specimens were obtained from humans after giving them a subdermal injection for 4 years to measure cross-section diameters of PCL particles. In all (160) biopsy slides, all cross-sections of PCL particles were measured in size in microscopic photographs, and the real size of PCL particles was calculated through Equation of a circle (Equationcircle) and mean value theorem for integrals (IntegralTheorem). Diameters of Ellansé particles were measured with particle size analyzer.
Results:
On average, the calculated PCL particle size using IntegralTheorem was 42.83 (immediately), then 43.18(1), 42.62(2), 40.90(3), and 34.46 µm(4 years), respectively. These results were similar to the diameters calculated using the Equationcircle. PCL size remained unchanged until 3 years, which began to decrease from the fourth year, making the transition point in between. In particle size analyzer, the mean diameter was 42.42 µm.
Conclusions:
PCL particle size was mathematically calculated for 4 years in an in vivo biopsy study. Until 3 years after the injection, PCL particle diameter remained at 95.47% and showed phase-1 isovolemic degradation. From 4 years after the injection, particles decreased in size, showing phase-2 bulk degradation. PCL particles were smooth and circular for 3 years, and from the fourth year, the surface became very rough. The Ellansé-M longevity was longer than 4 years.
INTRODUCTION
Polycaprolactone (PCL) has been used as an implantation material since the early 1930s.1 It also became an environment-friendly agricultural product, as it could be degraded by microorganisms in vitro and in vivo.2 PCL (a hydrophobic semicrystalline polymer) displays the ideal solubility and plasticity needed for various applications and can be easily manufactured for medical uses. The melting point of PCL is from 55°C to 60°C.3–5 PCL implants and scaffolds have been reported to exhibit a promising tunable porosity and controlled degradability.6 This tunable degradability is settled by the molecular weight (Mw) of PCL: longer-high Mw lasted longer than short-lower Mw.7 Therefore, PCL-M2 (Ellansé-M, Sinclair Pharmaceuticals, London, UK) was longer lasting than PCL-S1 (Ellansé S-version) after being injected into faces. Additionally, PCL filler showed neocollagenesis, neoelastinogenesis, and neovascularization in human studies.8,9
For human studies, PCL-M2 was injected into the subdermis of the patients and biopsy specimens were obtained 1, 2, 3, and 4 years after the injections. Then, PCL diameter and PCL surface erosion were evaluated. PCL particles revealed various cross-sections in the biopsy slides. Using various cutting cross sections, the real diameter (x100) was calculated using Equation of a circle (Equationcircle)10 and Mean Value Theorem for Integrals (IntegralTheorem).11 The real PCL particle size can be calculated mathematically, and isovolemic phase-1 degradation property (how long the diameter lasts without changing in size) can be evaluated in in vivo human biopsy studies.
With molecular weight, phase-1 degradation can be calculated using the formula Mn = Mw:exp(−k′t).12 Phase-1 degradation times (t) of PCL fillers were 1, 2, 3, and 4 years in manufacturer’s planning. Therefore, Mw of PCL-S1, -M2, -L3, and -E4 can be estimated using this formula.
METHOD
All 8 patients underwent clinical evaluation after satisfying all study criteria related to previous medical and surgical history (including physical examinations). All patients, who had been to the author’s clinic in the past, were women, with the average age of 38.25 years. The patients were previously injected with botulinum toxin or were treated with an energy-based device such as high intensity–focused ultrasound (HIFU) and radiofrequency (RF). However, they were not treated within 6 months before the commencement of the study. Patients who had previously received any filler injections were excluded from this study. All 8 patients agreed to have not taken filler injection, have not been treated with HIFU, and have not undergoneRF procedures during the follow-up period. These criteria were mandated because these would change the PCL particle size. In relation to these criteria, only Botox injection was allowed for the patients and was applied in the author’s clinic during the follow-up. Informed consent was obtained from the patients, in compliance with the Korean Society of Plastic and Reconstructive Surgeons standards. This study protocol conformed to the ethical guidelines of the 1975 Declaration of Helsinki and to the applicable local regulations. This is a single-clinic prospective study conducted from January 2013 to June 2019. Using subdermal scraping skin tightening technique (SSSTT), PCL-M2 was injected for all patients just under the dermis with the help of a 25G-5 cm blunt cannula. To evaluate immediately after injections, PCL-M2 fillers were injected under the skin and in the muscle layer on fresh chicken breasts, and biopsy studies were conducted.
Biopsy Study
Specimens were obtained from 8 patients who had planned to undergo a face-lifting surgery after 1, 2, 3, and 4 years (random) after PCL-M2 injections. A 0.5 × 3 cm piece of soft tissue was excised from below the temporal hairline or preauricular area during the surgery.
A biopsy study was performed to measure the size of PCL particle in biopsy specimens, using a light microscope (Olympus BX41; Olympus, Tokyo, Japan) and a microscope digital camera (eXcope T300; Touptek, Zhejiang, China). Ten biopsy slides were selected from each of the 8 patients, amounting to a total of 80 slides. For each of the 80 slides, 5 photographs were taken in different areas, after observing them under a light-microscope (H&E stain) with 100× magnification so as to measure PCL size. Additionally, 5 photographs were taken with 200× magnification to evaluate PCL shape. Slide photographs with 100× magnification contained 30–100 PCL particles, and slide photographs with 200× magnification contained 5–45 PCL particles.
In 80 biopsy slides, thousand diameters from cutting cross sections follow normal distribution because measuring value was large enough. The author wanted to mathematically estimate the actual PCL particle size by cutting diameter of particles, and how long the diameter lasts without a change in diameter length (isovolemic degradation property). Also surface erosion without volume loss (phase-1 degradation) and bulk degradation showing volume loss (phase-2 degradation) were evaluated in long-term follow-up from human biopsy studies. The degradation features and shape of PCL particles were observed. (See Video [online], which displays the degradation features and shape of PCL particles.)
Video 1. This video displays the degradation features and shape of PCL particles.
Using the Image-J program, all PCL particles were measured by 400 photographs (100×) from 80 slides. ImageJ is a Java-based image-processing program developed at the National Institutes of Health and the Laboratory for Optical and Computational Instrumentation (LOCI, University of Wisconsin).13,14 Using cross-section diameters of PCL particles in biopsy slides, real particle diameter was calculated mathematically using Equationcircle10 and IntegralTheorem.11
To measure immediately after injection, PCL-M2 fillers were injected under the skin and in muscle layer on fresh chicken breasts purchased from a usual market and biopsy slides were made (Tables 1 and 2).
Table 1.
PCL Particle Size Analysis from Biopsy Slide Using the Equation of a Circle
| Immediate* | 1 Year | 2 Years | 3 Years | 4 Years | |
|---|---|---|---|---|---|
| Top20 (µm) SD | 42.5525 ± 5.4737 | 42.8995 ± 6.6820 | 42.3365 ± 4.1930 | 40.6269 ± 4.1001 | 34.2361 ± 11.0927 |
| × parameterCircle | × 1.006752540 | ||||
| Calculated PSize (µm) equation of a circle | 42.7668 | 43.1156 | 42.5497 | 40.8315 | 34.4085 |
| Changes in size | 100.00% | 100.81% | 99.49% | 95.47% | 80.45% |
The average of top 20% measured diameter (Top20) and the calculated particle size (PSize) using Equation of a Circle from Top20.*To measure immediately after injection, PCL-M2 fillers were injected under the skin and in muscle layer on fresh chicken breasts purchased from a usual market and biopsy slides were made.
Table 2.
PCL Particle Size Analysis from Biopsy Slide Using Integraltheorem.
| Immediate* | 1 Year | 2 Years | 3 Years | 4 Years | |
|---|---|---|---|---|---|
| Top20 (µm) SD | 42.5525 ±5.4737 | 42.8995 ± 6.6820 | 42.3365 ± 4.1930 | 40.6269 ± 4.1001 | 34.2361 ± 11.0927 |
| × parameterInteg | × 1.006752540 | ||||
| Calculated PSize (µm) Integraltheorem | 42.8398 | 43.1892 | 42.6223 | 40.9013 | 34.4673 |
| Changes in size | 100.00% | 100.81% | 99.49% | 95.47% | 80.45% |
Measured Top20 in all slides, and PSize was calculated using Integraltheorem from Top20.*To measure immediately after injection, PCL-M2 fillers were injected under the skin and in muscle layer on fresh chicken breasts purchased from a usual market and biopsy slides were made.
Mathematical Calculation to Estimate PCL Particle Size
Particle sizes (all values) in 400 photographs (5 photographs from 80 biopsy slides) captured under 100× magnification were measured, and only top 20% of the largest particles can be used to calculate the actual size of PCL particle (PSize). Among all values, top 20% of the largest diameters were selected using Excel program and the average (Top20) was calculated. The measured diameters (all values) were normally Gaussian distributed because the population size was large enough.
There are two methods to calculate PSize from Top20 in biopsy slides. One method is Equation of a circle and other is Mean Value Theorem for Integrals (see Video [online], which displays the degradation features and shape of PCL particles).
Equation of a Circle
Since all measured radii followed normal distribution, top 20% of the largest sizes (from x100 to x80) were selected. The actual radius of Ellansé was assumed to be 10 in the Equation of a circle. Then parameterEq can be calculated by the average diameter of top 20% (Top20) using equation of a circle. By multiplying parameterEq to Top20, it is the actual size of Ellansé particle (PSize).
Equation of a circle
If the radius of particle was assumed 10, x100 = 10
If actual particle size is assumed 10, then x100 is 10 in above graph of Figure 1.
Fig. 1.

Equation of a circle.
“y” is the measured radius.
If the top 20% of the largest in size from all values were selected, “x” is between 0 (x100) to 2 (x80)
x100 is the biggest radius among selected measured radius
x80 is the smallest radius among selected measured radius
(Fig. 1)
The mean diameter of Top20 = x90
Assumed radius of Ellansé: (10) = x90 × parameterEq
![]() |
When top 20% of diameters were selected among all the measured values, the mean diameter (Top20) turned out to be 99.4987437% of the actual diameter (PSize). When top 20% of diameters were selected from all values, particles were selected from 97.97958971% size (x80) to 100% size (x100) in normal distribution. If hypothetic radius is 10, the mean value of the selected top 20% will lie on top 10% (x90). And parameterEq can be calculated as above. Therefore, by multiplying 1.0050378 (parameterEq) to Top20 in biopsy slides, it is possible to calculate the actual particle size (PSize).
Mean Value Theorem of Integrals
Integraltheorem Using Top 20% Values
Actual Ellansé radius can also be calculated by Integraltheorem with top 20% largest particles in biopsy slides instead of the Equationcircle. Using Integraltheorem, mean diameter from x80 to x100 can be calculated. The resulting integral can be computed using integration by parts or a double angle formula followed by one more substitution. One can also note that the function being integrated is the upper right quarter of a circle with a radius of one, and hence integrating the upper right quarter from zero to one is the geometric equivalent to the area of one quarter of the unit circle, or π/4. Assuming that the radius of a circle is 10, the average of the top 20% of the radius is Top20 (Fig. 2).
Fig. 2.

Mean value theorem for integrals using top 20% values.
With top 20% cross-section radius, real radius can be mathematically calculated using Integraltheorem. The result is obtained through integral mean value theorem, where a is 0 and b is 2.
Integral Mean Value Theorem
![]() |
![]() |
Manually computed antiderivative for
![]() |
To calculate the definite integral
, the result follows:
![]() |
![]() |
![]() |
To achieve the final result in
![]() |
The calculated mean is 9.932927505324625, the Top20.
One can note that the function being integrated is the upper right quarter of a circle with a radius of 10. (Though needing the top 20%, a circle is symmetrical to the radius; therefore, it is logical to use only one quarter of the circle.) The actual radius of the PCL particle can be obtained through the ratio below after calculation of parameterInteg.
![]() |
Integraltheorem Using All Measured Values
To calculate the real PCL size, the average of all measured value (Allaver) can be used instead of the top 20% data, and these 2 results are the same. When calculating all measured size, a is 0 and b is 10 (Fig. 3).
Fig. 3.

Mean value theorem for integrals using all values. With all cross-section radii, real radius can be calculated mathematically using Integraltheorem.
Integral Mean Value Theorem
![]() |
![]() |
![]() |
If x100 assumed 10, integration of all radius is 7.853982.
![]() |
Laser Diffraction Particle Size Analysis
Additionally, PCL-M2 particle sizes were measured using a laser diffraction particle size analyzer (PSA) (PSA-1190, Anton-Paar, Graz, Austria). The PSA determines the particle size distribution based on laser diffraction exhibited by the sample. The angle of diffraction depends on the particle size. This diffraction pattern is detected and analyzed. The analysis compares the measured values with the expected theoretical values. This analyzer allows the measure of particles between 0.04 and 2500 μm by the diffraction pattern, using Fraunhofer or Mie theory15–18 (Fig. 4).
Fig. 4.

Diffraction particle size analysis (PSA). Diffraction is when a wave goes through a small hole and has a flared-out geometric shadow of the slit. Reflection is when waves, whether physical or electromagnetic, bounce from a surface back toward the source. A mirror reflects the image of the observer. Refraction is when waves, whether physical or electromagnetic, are deflected when they go through a substance. The wave generally changes the angle of its general direction.
RESULTS
PCL particles maintained their diameters. The diameters after 1, 2, and 3 years were almost the same as the values obtained immediately after injection, showing phase-1 degradation. The changes in size were 100.81% (1 year), 99.49% (2 years), 95.47% (3 years), and 80.45% (4 years). Four years after injection, the diameter became 80%, showing phase-2 degradation. These changes in size were exactly the same in Equation of a circle and in Integraltheorem (Tables 1 and 2).
On average, the calculated PCL size (PSize) was 42.76 (immediately), 43.11 (1 year), 42.54 (2 years), 40.83 (3 years), and 34.40 µm (4 years) using Equation of a circle (Table 1 and Figs. 5–7).
Fig. 5.

Particle size and shape 2 years after PCL injection in subdermis (H&E ×200). All particles were circular with smooth surfaces.
Fig. 7.

Particle size and shape 4 years after PCL injection in subdermis (H&E ×200). All particles had a rough surface showing surface erosion, and particle size decreased, showing bulky degradation (phase-2 degradation). Matured collagen bundles were seen, which were found to mimic a hypertrophic scar tissue pattern.
Fig. 6.

Particle size and shape 3 years after PCL injection in the subdermis (H&E ×200). Some particles had a smooth surface (arrow head). Other particles had a slightly rough surface showing surface erosion (arrow). F, fibrocyte; FGC, foreign body giant cell; Mp, macrophage.
And PSize was 42.83 (immediately), 43.18 (1 year), 42.62 (2 years), 40.90 (3 years), and 34.46 µm (4 years) using Integraltheorem (Table 2).
In laser diffraction particle size analysis, the mean diameter was 42.42 µm (29.10–59.72 µm), with the median at 41.26 µm. And this value was close to the calculated size, that is, 42.76 µm (Equationcircle) and 42.82 µm (Integraltheorem) (Fig. 5) (Tables 1 and 2).
The results obtained using mathematical calculation by Equationcircle and IntegralTheorem in biopsy slides were almost the same as the result gained through the laser diffraction method in vitro.
Equation of a Circle
The average of top 20% measured diameters in biopsy slides (Top20) were 42.5525 (immediately after injection), 42.8995 (1 year), 42.3365 (2 years), 40.6269 (3 years), and 34.236 µm (4 years).
The calculated diameter is Top20 × parameterCircle.
PSize = Top20 × parameterCircle = Top20 × 1.006752540
The diameter calculated using the equation of a circle was 42.8398 (immediately), 43.1892 (1 year), 42.6223 (2 years), 40.9013 (3 years), and 34.4673 µm (4 years) (Table 1).
Mean Value Theorem for Integrals
Integraltheorem Using Top 20% Values
Top20 were 42.55 ± 5.47, 42.89 ± 6.68, 42.33 ± 4.19, 40.62 ± 4.10, and 34.23 ± 11.09.
The calculated diameter is Top20 × parameterInteg.
PSize = Top20 × parameterInteg = Top20 × 1.006752540
The calculated diameter using mean value theorem for integrals was 42.8398 (immediately), 43.1892 (1 year), 42.6223 (2 years), 40.9013 (3 years), and 34.4673 (4 years) (Table 2).
Integral Mean Value Theorem Using All Values
Diameters of PCL particles (PSize) can be calculated with all measured PCL sizes in biopsy slides.
![]() |
![]() |
![]() |
If x100 assumed 10, integration of all radius is 7.853982.
![]() |
![]() |
The average of all measured particle sizes (immediately after the injection) was 32.2893617 ± 10.76050592 µm. Using all measured values, calculated particle size was 41.11 µm. There was a very small difference between calculations using all measured values (41.11 µm) and using the top 20% (42.83 µm).
Laser Diffraction Particle Size Analysis
In laser diffraction PSA, the actual mean diameter of PCL particles was 42.42 µm (29.10–59.72 µm), and median diameter was 41.26 µm (Fig. 5). The measured diameter (42.42 µm) using PSA in vitro was almost the same as the diameter (42.76 µm by Equationcircle and 42.83 µm by Integraltheorem) calculated from biopsy slides (Fig. 8).
Fig. 8.

PCL-M2 particle size measurement using laser diffraction particle size analyzer.
DISCUSSION
From molecular weight phase 1, phase-1 degradation time (t) can be calculated using the formula Mn = Mw:exp(−k′t). Also reversely, if phase-1 degradation can be 1, 2, 3, and 4 years, molecular weight can be calculated as wished.12
where
This kinetic expression holds surprisingly well for an extended degradation period, during which Mw decreases to <10% of its initial value, and the crystallinity increases from 45% to about 80%. The observation that the kinetic expression (1) incorporates the first power of the carboxylic acid end group concentration is unexpected. The hydrogen ion concentration of a weak acid is proportional to the square root of the acid concentration.12
Regardless of the initial Mw or its geometry of the sample, a linear relationship between In M or In [v] and the time until the Mw decreases to about 5000. Thereafter, there was a consistent but slower increase in crystallinity with chain cleavage to a value of about 80% after 120 weeks (Mn = 4600). The rate constant k′ derived from the initial part of the linear plot in Figure 2 (up to 110 weeks) is 3.07 × 10−3 day-l, that is, the number of ester groups cleaved per day corresponds to about 0.31% of the carboxylic acid end groups present (Fig. 9).
Fig. 9.

In vivo degradation of PCL capsule in Pitt’s study12: (•) PCL-FRL-3 with Silastic insert; (▴) PCL-RRL-5B containing norgestrel insesame oil; (♦) PCL*-1 unfilled.
![]() |
Using the given formula and putting 784 days (112 weeks) for t,
Therefore, the equation is matched to time and molecular weight.
The consistency of the data in Figure 9 demonstrates that the value of the rate constant is not affected by the presence of drug and dispersing agent.12 From this formula, manufacturer’s developmental background can be guessed. If they wanted to make 2 year of phase-1 degradation, it will be 730 days (2 year). The original molecular weights (Mw) can be calculated by 1, 2, 3, and 4 years of phase-1 degradation (t) inverting the confidence interval procedure. The unit of t is day.
In the manufacturer guideline, PCL-S1 lasts for 1 year, PCL-M2 for 2 years, PCL-L3 for 3 years, and PCL-E4 for 4 years. They must have used the formula above. Then, we can derive the molecular weight of PCL particles.
Using the method above, the author’s estimation of the original molecular weight (Mw) of Ellansé product series follows:
To make exactly 1, 2, 3, and 4 years of phase-1 degradation time (t)
The result follows: for 1 year, MwS1 will be 14,106; for 2 years, 43,265; for 3 years, 132,648; and for 4 years, 406,773.
But, no powder with the exact molecular weight exists. The company would normally use ready-made powder with the molecular weight (Mw) of 10k, 15k, 20k, 25k, 30k, 37k, 40k, 43k, 50k, 60k, 80k, 100k, and 190k. The longevity of powders can be found in Figures 10 and 11.
Fig. 10.

Estimated “phase-1 degradation time (t)” from commonly sold PCL powders. (The unit of t is day.) If molecular weight (Mw) of 15,000 was used to make PCL-S1 (Ellanse-S version; Sinclair Pharmaceuticals, London, UK), phase-1 degradation time (t) will be 385.01 days (1.05 years). For 40,000 Mw, it will be 704.50 days (1.99 years); for 100,000 Mw, 1002.97 days (2.74 years); and for 1,900,000 Mw, 1212.04 days (3.32years) (Figs. 10 and 11).
Fig. 11.

Estimated phase-1 degradation time from commonly sold PCL powders. Though the line should be linear, the nonlinear line is due to calculation through only the powders sold publicly (Mw, molecular weight) (Mw unit: 103).
Biodegradables are solid polymeric materials that break down due to macromolecular degradation with dispersion in vivo but no proof for the elimination from the body. Bioresorbables show bulk degradation and further resorb in vivo. Bioerodibles show surface degradation and resorb in vivo. Bioresorbables and bioerodibles reflect total elimination of the initial foreign material and of surface degradation by-product with no residual side effects. Bioabsorbables dissolve in body fluids without any polymer chain cleavage or molecular mass decrease.19,20
PCLs can be biodegraded by outdoor living organisms with enzyme.21 The process in human body, which can be divided into 2 phases, takes much longer.22 Phase 1 involves nonenzymatic hydrolytic cleavage of the ester linkages. The mass and volume of the implant remain unchanged during this process.12,23,24 The phase shows controlled, predictable, first-order, and linear bioresorption pattern. Phase 2 starts at a specific chain length with approximate chain molecular weight of about 5000 Da.25 Bulk degradation occurs in this phase when water penetrates the entire polymer bulk, causing hydrolysis throughout the entire polymer matrix. In this case, the internal concentration of auto-catalysis product can produce an acidic gradient.26 The nonbacterial inflammatory reactions in vivo such as swelling or redness after the PCL filler injection is due to this acidic gradient and the rapid release of the oligomers. This also results in accelerated speed of internal degradation compared with that in the surface; an outer-layer’s molecular weight becomes higher than that of the interior. When the inner oligomers become small enough, they diffuse rapidly through the outer layer, accompanied by an onset of weight loss and a decrease in the rate of chain scission. During this phase, rate of chain scission decreases. It is also characterized by the onset of controlled and predictable total mass loss via bioresorption of the microspheres and excretion through the normal metabolic pathways.7,23
Under isothermal conditions, the mass of PCL remained constant for 2 hours at temperatures ranging from 100 to 250°C.27 Therefore, temperature cannot be used to remove PCL fillers because 50–60°C do not effect thermal degradation.
PCL particles lasted for 3 years, maintaining its diameter, and isovolemic phase-1 degradation property of PCL was proved by this study. In cases such as HA fillers and calcium hydroxyapatite filler (Radiesse), they have a property of constant volume reduction immediate after injection. Therefore, the isovolemic degradation property is very different from other fillers; thus it is crucial for the surgeon to be aware of the change in volume after each filler injection (see Video [online], which displays the degradation features and shape of PCL particles).
CONCLUSIONS
For 3 years in subdermis, PCL particles maintained 95.47% (40 µm) of their original diameter (42 µm). This proved for the first time the isovolemic degradation property of PCL particles in human biopsy study. PCL particles maintained their diameter and volume for a minimum of 3 years in phase-1 degradation.
Four years after the injection, the PCL diameter became smaller, to 80% of its size (34 µm), showing phase-2 degradation, and the surface of particles became very rough and eroded. The transition from phase 1 to 2 is between 3 and 4 years. The longevity of PCL-M2 was greater than 4 years.
ACKNOWLEDGMENT
The author thanks Yeonoo Kim (a medical writer) for helping the author complete the article.
Footnotes
Published online 26 June 2020.
Disclosure: The author has no financial interest to declare in relation to the content of this article.
Related Digital Media are available in the full-text version of the article on www.PRSGlobalOpen.com.
REFERENCES
- 1.Van Natta FJ, Hill JW, Carruthers WH. Polymerization and ring formation, caprolactone and its polymers. J Am Chem Soc. 1934;56:455–459. [Google Scholar]
- 2.Huang S. Mark F, Bikales N, Overberger C, Menges G, Kroshwitz J. Biodegradable polymers. In: Encyclopedia of Polymer Science and Engineering. 1985:New York, NY: John Wiley and Sons; 220–243. [Google Scholar]
- 3.Chandra R, Rustgi R. Biodegradable polymers. Progr Polym Sci. 1998;23:1273–1335. [Google Scholar]
- 4.Okada M. Chemical syntheses of biodegradable polymers. Progr Polym Sci. 2002;27:87–133. [Google Scholar]
- 5.Nair LS, Laurencin CT. Biodegradable polymers as biomaterials. Progr Polym Sci. 2007;32:762–798. [Google Scholar]
- 6.Guarino V, Causa F, Taddei P, et al. Polylactic acid fibre-reinforced polycaprolactone scaffolds for bone tissue engineering. Biomaterials. 2008;29:3662–3670. [DOI] [PubMed] [Google Scholar]
- 7.Sun H, Mei L, Song C, et al. The in vivo degradation, absorption and excretion of PCL-based implant. Biomaterials. 2006;27:1735–1740. [DOI] [PubMed] [Google Scholar]
- 8.Kim JS. Changes in dermal thickness in biopsy study of histologic findings after a single injection of polycaprolactone-based filler into the dermis. Aesthet Surg J. 2019;39:NP484–NP494. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kim JA, Van Abel D. Neocollagenesis in human tissue injected with a polycaprolactone-based dermal filler. J Cosmet Laser Ther. 2015;17:99–101. [DOI] [PubMed] [Google Scholar]
- 10.Equation of a Circle. Standard form equation of a circle. Math warehouse. Available at https://www.mathwarehouse.com/geometry/circle/equation-of-a-circle.php. Accessed August 2, 2019.
- 11.Scherfgen D. 2019 integral calculator. Available at https://www.integral-calculator.com. Accessed August 2, 2019.
- 12.Pitt CG, Chaasalow FI, Hibionada YM, et al. Aliphatic polyesters. I. the degradation of poly(e-caprolactone) in vivo. J Appl Polym Sci. 1981;26:3779–3787. [Google Scholar]
- 13.Schneider CA, Rasband WS, Eliceiri KW. NIH Image to ImageJ: 25 years of image analysis. Nat Methods. 2012;9:671–675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Collins TJ. ImageJ for microscopy. Biotechniques. 2007;431 suppl25–30. [DOI] [PubMed] [Google Scholar]
- 15.Mudroch A. Manual of Physico-chemical Analysis of Aquatic Sediments. 1997:Boca Raton: Lewis Publishers; 30. [Google Scholar]
- 16.Ye Zi, Jiang X, Wang Z. Measurements of particle size distribution based on mie scattering theory and markov chain inversion algorithm. J Softw. 2012;7:2309–2316. [Google Scholar]
- 17.Francesco MG. Physico-Mathesis De Lumine, Coloribus, Et Iride, Aliisque Annexis Libri. 1665:Bernia: 2. [Google Scholar]
- 18.John M. Diffraction Physics. 1975North-Holland: American Elsevier Publishing Company; [Google Scholar]
- 19.Woodruff MA, Hutmacher DW. The return of a forgotten polymer—polycaprolactone in the 21st century. Prog Polym Sci. 2010;35:1217–1256. [Google Scholar]
- 20.Vert M, Li SM, Spenlehauer G, et al. Bioresorbability and biocompatibility of aliphatic polyesters. J Mater Sci Mater Med. 1992;3:432–446. [Google Scholar]
- 21.Vert M. Degradable and bioresorbable polymers in surgery and in pharmacology: beliefs and facts. J Mater Sci Mater Med. 2009;20:437–446. [DOI] [PubMed] [Google Scholar]
- 22.Ginde R, Gupta R. In vitro chemical degradation of poly(glycolic acid) pellets and fibers. J Appl Polym Sci. 1987;33:2411–2429. [Google Scholar]
- 23.Woodward SC, Brewer PS, Moatamed F, et al. The intracellular degradation of poly(epsilon-caprolactone). J Biomed Mater Res. 1985;19:437–444. [DOI] [PubMed] [Google Scholar]
- 24.Chen DR, Bei JZ, Wang SG. Polycaprolactone microparticles and their biodegradation. Polym Degrad Stab. 2000;67:455–459. [Google Scholar]
- 25.Wiggenhauser PS, Balmayor ER, Rotter N, et al. In vivo evaluation of a regenerative approach to nasal dorsum augmentation with a polycaprolactone-based implant. Eur J Med Res. 2019;24:6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Bergsma JE, de Bruijn WC, Rozema FR, et al. Late degradation tissue response to poly(L-lactide) bone plates and screws. Biomaterials. 1995;16:25–31. [DOI] [PubMed] [Google Scholar]
- 27.Lam CX, Hutmacher DW, Schantz JT, et al. Evaluation of polycaprolactone scaffold degradation for 6 months in vitro and in vivo. J Biomed Mater Res A. 2009;90:906–919. [DOI] [PubMed] [Google Scholar]