Abstract
This study aimed to analyze the connection between corneal biomechanics (corneal hysteresis, CH) and endothelial cell density of cornea (mean endothelial cell density, MCD) in patients diagnosed with cataract.
This retrospective, observational study was performed in the Ophthalmology Clinic of the University Emergency Hospital in Bucharest. Of 60 patients (120 eyes) with cataract, who were included in this study, we analyzed the CH values obtained using with the Ocular Response Analyzer (ORA) and the MCD values obtained using the specular microscopy. The study groups comprised both men and women with ages ranging from 45 to 63 years.
Patients were divided into three study groups according to CH values. In each batch, the CH values obtained with the Ocular Response Analyzer (ORA) were correlated with age, gender and MCD, then the subgroups were compared. All the data gathered showed no correlation to be statistically significant regarding the biomechanical properties of the cornea and the corneal endothelial cell density in patients with cataract.
Keywords: corneal endothelial cell density, specular microscopy, corneal biomechanics, corneal hysteresis, Ocular Response Analyzer
Introduction
This study aimed to determine a possible relationship between the biomechanical properties of the cornea and the corneal endothelial cell density in patients with cataract.
Cataract is a clouding of the lens of the eye that leads to a progressive loss of vision. This is a common ophthalmological disorder that is caused by opacification of the lens and it generally develops in both eyes, but not evenly.
Numerous studies suggest that corneal hysteresis (CH), an indicator of corneal viscoelasticity, is associated with biomechanical properties of the cornea. Some studies showed that CH may be related to the biomechanical characteristics of peripapillary sclera and lamina cribrosa, as they are embryologically developed from the mesoderm and also because the collagen fibers continue with the collagen fibers from the corneal stroma [1-8].
The ORA (Ocular Response Analyzer) is a tool that determines corneal hysteresis (CH) at the indentation by a rapid jet of air. Corneal hysteresis is defined as the difference between the tension at which the cornea bends inward during an air jet applanation and the tension at which it bends out again. The two pressure values (P1 and P2) are obtained at the end of the measurement period, which lasts approximately 20 milliseconds.
Thus, ORA can be used for the assessment of corneal biomechanics in vivo [9,10].
Corneal hysteresis (CH) represents the difference between the two pressure determinations, P1 and P2, reflecting the capacity for absorption of kinetic energy in the tissue, an indicator of corneal viscosity. The mean of the two pressure measurements (P1 and P2) determines the Goldmann correlated intraocular pressure (IOPg) [11-13].
Specular microscopy is a photographic, noncontact technique for visualizing and analyzing the shape, size and number of the population of endothelial cells, by computer-assisted analysis of endothelial morphology. The instrument projects light onto the corneal surface and captures image from the endothelial/ aqueous interface.
Specular microscopy is a noninvasive method for analyzing the corneal endothelium and facilitates the rapid and correct diagnosis of corneal endotheliopathy affecting the structure and physiology of this corneal layer [14-17].
Although there have been several studies analyzing the relationship between eyes with cataract and the biomechanical properties of the cornea or endothelial cell density, the results are still under debate.
Materials and methods
This retrospective, observational study was performed in the Ophthalmology Clinic of University Emergency Hospital in Bucharest, during the year 2019.
All patients involved in the study were informed about the use of personal data, and the study was organized in accordance with the ethical principles stated in the Declaration of Helsinki, developed by the WMA (World Medical Association) [18].
60 patients were included in this study, patients known with changes of transparency of the lens in any of its layers, of at least 3+, pathology for which they underwent surgery. All subjects included in the study were Caucasian, aged between 45 and 63 years, with an equal gender distribution.
Data was gathered from medical records, including a complete patient history, as well as visual acuity with or without correction, refraction, biomicroscopic examination of the anterior segment and posterior pole, keratometry, pachymetry, specular microscopy, tonometry with the help of Goldman applanation and Ocular Response Analyzer (ORA).
Another inclusion criterion was a normal value of intraocular pressure (12–22 mmHg).
The exclusion criteria from the study were the following: positive history for the administration of topical eye medications and/ or for rigid or soft contact lenses use, glaucoma or glaucoma suspect, inflammation and/ or ocular infection in the background, dystrophies/ corneal pathology (Fuchs endothelial dystrophy, keratoconus, pellucid marginal corneal degeneration, bullous keratopathy) and last but not least, without any other ocular surgery procedures performed in medical history, as for example refractive surgery or laser ablation of the cornea and lens, corneal cross-linking (CXL) and intrastromal corneal ring segment (ICRS).
Paraclinical investigations were performed using the equipment of the Ophthalmology Clinic in the University Emergency Hospital, Bucharest.
Images reflecting endothelial cell density were captured using a specular microscope (Topcon, SP-3000P). Three consecutive measurements were made with a specular reflex light, a correct centering of the fixation target, with a calibration and a favorable image resolution. The measurement technique used was the center to center method. 45-50 corneal endothelial cells were counted and cell density (CD) was calculated by automatic analysis. The average of the three CD values was considered.
In vivo data of corneal biomechanics was obtained using the Ocular Response Analyzer (Reichert Ocular Response Analyzer G3 AutoTonometer, evaluation software v.1.01). The correct positioning of the patient was verified throughout the examination and three consecutive measurements (Single Measure) were performed. The average of recordings with a score higher than 7 (Waveform Score) was taken into consideration.
Goldmann applanation tonometry (GAT) was performed for IOP measurements.
This study included 60 patients, aged between 45 and 63 years, females and males, with cataract. We analyzed and compared the following parameters: corneal hysteresis (CH) and mean endothelial cell density (MCD) in both eyes.
Patients have been divided into three distinct study groups, according to the value of corneal hysteresis: batch no. 1 CH at RE = 8.33 ± 0.29 mm Hg and LE CH = 8.48 ± 0.26 mmHg, batch no 2. CH at RE = 9.33 ± 0.275 mmHg and CH at LE = 9.46 ± 0.275, batch no. 3 CH at RE = 10.62 ± 0.345 and LE CH = 10.55 ± 0.266.
SPSS statistics package 20 was used for statistical analysis.
Pearson’s correlation coefficient is the test statistics used to measure the statistical relationship, or association, between CCT (central corneal thickness), mean corneal endothelial cell density and the properties of corneal biomechanics.
For statistical significance we have considered P values ≤ 0.05.
Results
60 patients aged between 45 and 63 years, equal in proportion, women and men, were included in the study group.
Three groups were realized according to corneal hysteresis (CH) values. The data in Table 1 represent the description of the parameters analyzed in the study.
Table 1.
Parameters analyzed
| Parameters | RE | LE | ||||
| Mean ±SD | AM | Min-Max | Mean ±SD | AM | Min-Max | |
| CH Batch 1 | 8.33 ± 0.29 | 8.2 | 8-8.9 | 8.48 ± 0.26 | 8.5 | 8-8.9 |
| MCD Batch 1 | 2199 ± 327.18 | 2106.5 | 1814-2805 | 2232.5 ± 349.34 | 2078 | 1800-2872 |
| CH Batch 2 | 9.33 ± 0.275 | 9.3 | 9-9.9 | 9.46 ± 0.275 | 9.5 | 9-9.9 |
| MCD Batch 2 | 2077.2 ± 328.84 | 1991.5 | 1668-2688 | 2093.3 ± 345.93 | 2002.5 | 1612-2805 |
| CH Batch 3 | 10.62 ± 0.345 | 10.7 | 10.1-11.2 | 10.55 ± 0.266 | 10.5 | 10.1-11 |
| MCD Batch 3 | 2110.25 ± 327.47 | 2007.5 | 1707-2812 | 2141.6 ± 319.05 | 2101 | 1602-2713 |
There were correlations between corneal hysteresis and endothelial cell density (MCD) in each batch. Most variables had a non-parametric distribution according to the Shapiro-Wilk test (p <0.05) (Tables 2-4, Fig. 1-6).
Table 2.
Correlation between corneal hysteresis and endothelial cell counts in patients in group 1
| Correlation | p* |
| CH RE (p=0.005**) x MCD RE (p=0.011**) | 0.972, R=0.008 |
| CH LE (p=0.284**) x MCD LE (p=0.019**) | 0.633, R= -0.114 |
| *Shapiro-Wilk Test, **Spearman’s rho Correlation Coefficient |
Table 4.
Correlation between corneal hysteresis and endothelial cell counts in patients in group 3
| Correlation | p* |
| CH RE (p=0.094**) x MCD RE (p=0.108**) | 0.350, R= -0.221 |
| CH LE (p=0.368**) x MCD LE (p=0.265**) | 0.897, R=0.031*** |
| *Shapiro-Wilk Test, **Pearson Correlation Coefficient |
Fig. 1.

Correlations between CH RE and MCD RE in batch 1
Fig. 6.

Correlations between CH LE and MCD LE in batch 3
Table 3.
Correlation between corneal hysteresis and endothelial cell counts in patients in group 2
| Correlation | p* |
| CH RE (p=0.062**) x MCD RE (p=0.007**) | 0.962, R= -0.011 |
| CH LE (p=0.192**) x MCD LE (p=0.200**) | 0.906, R=0.028*** |
| *Shapiro-Wilk Test, **Spearman’s rho Correlation Coefficient, ***Pearson Correlation Coefficient |
Fig. 2.

Correlations between CH LE and MCD LE in batch 1
Fig. 3.

Correlations between CH RE and MCD RE in batch 2
Fig. 4.

Correlations between CH LE and MCD LE in batch 2
Fig. 5.

Correlations between CH RE and MCD RE in batch 3
According to the correlations established using Spearman’s rho coefficients, the association between corneal hysteresis and endothelial cell density was not statistically significant in the right or left eye for patients in group 1, 2 or 3.
The data in Tables 5-7 compare the groups together by analyzing corneal hysteresis and endothelial cell density from patients in group 1, group 2 and respectively 3. The variables were tested for distribution according to the Shapiro-Wilk test (Fig. 7-18).
Table 5.
Correlations between corneal hysteresis and endothelial cell density in patients in batch 1 vs. batch 2
| Correlation | p* |
| CH RE L1(p=0.005**) x MCD RE L2 (p=0.007**) | 0.525, R= -0.151 |
| CH LE L1 (p=0.284**) x MCD LE L2 (p=0.200**) | 0.626, R= -0.116*** |
| CH RE L2 (p=0.062**) x MCD RE L1 (p=0.011**) | 0.052, R= 0.441 |
| CH LE L2 (p=0.192**) x MCD LE L1 (p=0.019**) | 0.056, R= -0.434 |
| *Shapiro-Wilk Test, **Spearman’s rho Correlation Coefficient, ***Pearson Correlation Coefficient |
Table 7.
Correlations between corneal hysteresis and endothelial cell density in patients in batch 2 vs. batch 3
| Correlation | p* |
| CH RE L2(p=0.062**) x MCD RE L3 (p=0.108**) | 0.102, R= -0.376*** |
| CH LE L2 (p=0.192**) x MCD LE L3 (p=0.265**) | 0.018, R=0.522*** |
| CH RE L3 (p=0.094**) x MCD RE L2 (p=0.007**) | 0.765, R= -0.071 |
| CH LE L3 (p=0.368**) x MCD LE L2 (p=0.200**) | 0.116, R= 0.363*** |
| *Shapiro-Wilk Test, **Spearman’s rho Correlation Coefficient, ***Pearson Correlation Coefficient |
Fig. 7.

Correlation between CH RE Batch 1 and MCD RE Batch 2
Fig. 18.

Correlation between CH LE Batch 3 and MCD LE Batch 2
Table 6.
Correlations between corneal hysteresis and endothelial cell density in patients in batch 1 vs. batch 3
| Correlation | p* |
| CH RE L1 (p=0.005**) x MCD RE L3 (p=0.108**) | 0.636, R= -0.113 |
| CH LE L1 (p=0.284**) x MCD LE L3 (p=0.265**) | 0.069, R= 0.415*** |
| CH RE L3 (p=0.094**) x MCD RE L1 (p=0.011**) | 0.684, R= 0.097 |
| CH LE L3 (p=0.368**) x MCD LE L1 (p=0.019**) | 0.697, R= 0.093 |
| *Shapiro-Wilk Test, **Spearman’s rho Correlation Coefficient, ***Pearson Correlation Coefficient |
Fig. 8.

Correlations between CH LE Batch 1 and MCD LE Batch 2
Fig. 9.

Correlation between CH RE Batch 2 and MCD RE Batch 1
Fig. 10.

Correlation between CH LE Batch 2 and MCD LE Batch 1
Fig. 11.

Correlation between CH RE Batch 1 and MCD RE Batch 3
Fig. 12.

Correlations between CH LE Batch 1 and MCD LE Batch 3
Fig. 13.

Correlation between CH RE Batch 3 and MCD RE Batch 1
Fig. 14.

Correlation between CH LE Batch 3 and MCD LE Batch 1
Fig. 15.

Correlation between CH RE Batch 2 and MCD RE Batch 3
Fig. 16.

Correlations between CH LE Batch 2 and MCD LE Batch 3
Fig. 17.

Correlation between CH RE Batch 3 and MCD RE Batch 2
According to the correlations established using Spearman’s rho/ Pearson coefficients, the associations between corneal hysteresis and endothelial cell density were not statistically significant. The exception was the correlation between the corneal hysteresis value in the left eye for group 2 of patients and the value of the number of cells in the left eye in group 3 of patients (p = 0.018, R = 0.522), thus showing a small value of CH from LE in group 2 of patients (close to 9 mmHg), which was significantly associated with a low value of the number of endothelial cells from LE in group 3 of patients and vice versa.
The data in Table 8 and Fig. 19,20 represent the correlation between corneal hysteresis and endothelial cell density in patients throughout the study group. According to the Shapiro-Wilk test (p <0.05), all variables had a non-parametric distribution.
Table 8.
Correlation between corneal hysteresis and endothelial cell density in patients from the entire study group
| Correlation | p* |
| CH RE (p=0.002**) x MCD RE (p<0.001**) | 0.233, R= -0.156 |
| CH LE (p=0.009**) x MCD LE (p=0.008**) | 0.429, R= -0.104 |
| *Shapiro-Wilk Test, **Spearman’s rho Correlation Coefficient |
Fig. 19.

Correlations between CH RE and MCD RE throughout the study group
Fig. 20.

Correlation of CH LE and MCD LE across the study group
Taking into account the correlations established using Spearman’s rho coefficients, the association between corneal hysteresis and endothelial cell density was not statistically significant in the right eye (p = 0.233) or the left eye (p = 0.429) for patients in the entire study group.
There were correlations between the age/ sex of the patients with the values of corneal hysteresis and the density of corneal endothelial cells. According to the Shapiro-Wilk test (p <0.05), the distribution of variables was non-parametric (Table 9-16, Fig. 21-Fig. 28).
Table 9.
Right eye: relationship between CH and patients age
| Correlation | p* |
| Age (p=0.015**) x CH RE (p=0.002**) | 0.127, R=0.199 |
| *Spearman’s rho Correlation Coefficient, **Shapiro-Wilk Test |
Table 16.
Left eye: relationship between MCD and patient gender
| Gender | Mean ± SD | Median Rank | p* |
| Female (p=0.033**) | 2208.84 ± 322.43 | 33.71 | 0.141 |
| Male (p=0.076**) | 2099.1 ± 349.84 | 27.07 | |
| *Mann-Whitney U Test, **Shapiro-Wilk Test |
Fig. 21.

Right eye: relationship between CH and patients age
Fig. 28.

Left eye: relationship between MCD and patient gender
Table 10.
Left eye: relationship between CH and patients age
| Correlation | p* |
| Age (p=0.015**) x CH LE (p=0.009**) | 0.084, R=0.225 |
| *Spearman’s rho Correlation Coefficient, **Shapiro-Wilk Test |
Table 11.
Right eye: relationship between MCD and patients age
| Correlation | p* |
| Age (p=0.015**) x MCD RE (p<0.001**) | 0.388, R= -0.114 |
| *Spearman’s rho Correlation Coefficient, **Shapiro-Wilk Test |
Table 12.
Left eye: relationship between MCD and patients age
| Correlation | p* |
| Age (p=0.015**) x MCD LE (p=0.008**) | 0.369, R= -0.118 |
| *Spearman’s rho Correlation Coefficient, **Shapiro-Wilk Test |
Table 13.
Right eye: relationship between CH and patient gender
| Gender | Mean ± SD | Median Rank | p* |
| Female (p=0.075**) | 9.39 ± 0.985 | 29.92 | 0.790 |
| Male (p=0.021**) | 9.465 ± 1.015 | 31.12 | |
| *Mann-Whitney U Test, **Shapiro-Wilk Test |
Table 14.
Left eye: relationship between CH and patient gender
| Gender | Mean ± SD | Median Rank | p* |
| Female (p=0.026**) | 9.51 ± 0.946 | 30.89 | 0.859 |
| Male (p=0.263**) | 9.482 ± 0.848 | 30.09 | |
| *Mann-Whitney U Test, **Shapiro-Wilk Test |
Table 15.
Right eye: relationship between MCD and patient gender
| Gender | Mean ± SD | Median Rank | p* |
| Female (p=0.004**) | 2197.84 ± 313.26 | 35.74 | 0.016 |
| Male (p=0.001**) | 2055.66 ± 329.51 | 24.9 | |
| *Mann-Whitney U Test, **Shapiro-Wilk Test |
Fig. 22.

Left eye: relationship between CH and patients age
Fig. 23.

Right eye: relationship between MCD and patients age
Fig. 24.

Left eye: relationship between MCD and patients age
Fig. 25.

Right eye: relationship between CH and patient gender
Fig. 26.

Left eye: relationship between CH and patient gender
Fig. 27.

Right eye: relationship between MCD and patient gender
According to Spearman’s rho correlation coefficient, these associations were not statistically significant. According to the Mann-Whitney U test, the differences between women and men were not statistically significant for corneal hysteresis values, but the women in the study had significantly higher corneal endothelial cells (mean range = 35.74) than men (mean range = 24.9) (p=0.016).
Discussions
Many medical studies have underlined a link between the lower values of CH and different corneal pathologies like keratoconus, corneal dystrophy, corneal edema, post LASIK status, etc. These conditions reflect the disorganization of the collagen fibers at the corneal stroma level.
Low values of the CH have also been identified in patients with primary open-angle glaucoma and normal-tension glaucoma. Thus, they sustain the hypothesis that changes at the Lamina Cribrosa level are directly linked with the corneal biomechanical alteration. Because of this, the CH measurements are useful in the normal-tension glaucoma and early keratoconus [19-22].
Numerous studies have found a clear correlation between corneal hysteresis and CCT (central corneal thickness). The hypothesis that the corneal thickness growth is associated with viscosity growth is a consequence of the fact that CH reflects the viscoelastic properties of the cornea.
Concerning open angle glaucoma patients, many studies have pointed out the link between the CH and the intraocular pressure (IOP). For these patients, the CH value has been significantly lower than the control group. After the IOP decrease under 27 mmHg, the CH level has been partially normalized. Another observation of the authors was that, while in normal, physiological conditions, CH modifications were not correlated with IOP, for IOP values of over 21 mmHg CH modifications being associated.
This is justified by the structural modifications of the collagen fibers that suffer a significant elongation under high IOP values, thus the difference between the P1 and P2 parameters at ORA is decreasing [23-26].
Age-related changes have shown there is a positive correlation between viscoelastic properties of the cornea and CCT (central corneal thickness). Thereby, the CH and CRF (corneal resistance factor) values are decreasing, this being linked directly with the augmentation of the corneal hydration after a certain age [27-30].
For hyperopic and for female patients, higher values for CRF have been observed. For highly myopic patients, other studies have registered lower values of the CH and CRF and higher values of IOPcc (Corneal Compensated Intraocular Pressure) and of IOPg (Goldmann Correlated IOP value) compared to the emmetropic, hyperopic patients or patients with medium or low myopia [31-34].
The recent studies (since 2015), which have analyzed the mechanical corneal properties in relation with the cell density of the corneal endothelium, have been unable to show a statistically significant correlation [35,36].
The IOP growth during the cataract surgery can justify the biomechanical properties of eye globe appearing after surgery. In cataract surgery, the IOP value in the anterior chamber can grow up to approximately 50-60 mmHg; these high pressures predispose the eye tissue to degradation [37-39].
The cataract intervention through phacoemulsification leads to important endothelial corneal damage, which is extremely significant in patients with a low number of endothelial cells. Specular microscopy records a leakage of approximately 14% (11.4% -16.6%) of corneal endothelial cells.
The thickness of the central cornea grows significantly after cataract surgery for the first seven days but after that period, up to 3 months, it decreases to a value smaller than the one pre-surgery [40,41].
In the studies undergone until present, significant differences have been observed between the biomechanical corneal values pre- and post-surgery. The CH value has decreased in the immediate post-surgically period and went back to the pre-surgically level in 1-2 weeks; no statistically significant change has been identified in the CRF values [42-44].
Conclusions
During this study, we have analyzed the following parameters: corneal hysteresis (CH) and the mean endothelial cell density (MCD) in the right and the left eye in 60 patients, males and females, aged between 45 and 63 years, with cataract.
There were no statistically significant variations of the corneal hysteresis for the right eye and for the left eye (p = 0.204) and there were no statistically significant differences between the cell count for the right eye and the left eye (p= 0.138) in all the study groups.
According to the Pearson correlation coefficient in the study batch no. 1, the correlation between the corneal hysteresis value (CH) and the mean endothelial cell density (MCD), was not statistically significant (RE p = 0.972 and LE p = 0.633).
Like batch 1, batches 2 and 3 have shown no statistically significant relationship between the value of the CH and MCD (batch 1: RE p = 0.962 and LE p =0.906, batch 2 RE p=350 and LE p=0.897).
Applying the Pearson correlation coefficient in order to determine if there was a statistically significant association between study group no. 1 and no. 2, we found that in RE p ˃ 0.05 and in LE p ˃ 0.05, therefore the association was not statistically significant.
There was no statistically significant association between study group no. 1 and no. 3 (RE p ˃ 0.05 and LE p ˃ 0.05), according to Pearson correlation coefficient.
However, we found a statistically significant correlation (p = 0.018, R = 0.522) between the CH of the LE measured in batch no. 2. and the MCD in LE in the study group no. 3.
According to the Pearson correlation, we did not find any statistically significant associations (RE p = 0.233 and LE p = 0.429) between the CH and the MCD in patients in groups no. 1, 2 and 3 in one place.
In conclusion, there was no statistically significant correlation between the biomechanical properties of the cornea and the endothelial cornea cell in patients with cataract.
Acknowledgements
The authors gratefully acknowledge Programul Operațional Capital Uman 2014–2020 for the data provided by Reichert Ocular Response Analyzer supported through POCU/91/4/8/109169 (Cod SMIS 2014+: 109169) “Diagnosticul și terapia bolilor rare sistemice cu afectare oculară – OCURARE”.
References
- 1.Esporcatte LPG, Salomão MQ, Lopes BT, et al. Biomechanical diagnostics of the cornea. Eye Vis (Lond) 2020;7:9. doi: 10.1186/s40662-020-0174-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Dupps WJ Jr, Wilson SE. Biomechanics and wound healing in the cornea. Exp Eye Res. 2006;83(4):709–720. doi: 10.1016/j.exer.2006.03.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Roberts CJ, Dupps WJ Jr. Biomechanics of corneal ectasia and biomechanical treatments. J Cataract Refract Surg. 2014;40(6):991–998. doi: 10.1016/j.jcrs.2014.04.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ma J, Wang Y, Wei P, Jhanji V. Biomechanics and structure of the cornea: implications and association with corneal disorders. Surv Ophthalmol. 2018;63(6):851–861. doi: 10.1016/j.survophthal.2018.05.004. [DOI] [PubMed] [Google Scholar]
- 5.Roberts C. Concepts and misconceptions in corneal Biomechanics. J Cataract Refract Surg. 2014;40:862–869. doi: 10.1016/j.jcrs.2014.04.019. [DOI] [PubMed] [Google Scholar]
- 6.Lee KM, Kim TW, Lee EJ, Girard MJA, Mari JM, Weinreb RN. Association of Corneal Hysteresis With Lamina Cribrosa Curvature in Primary Open Angle Glaucoma. Invest Ophthalmol Vis Sci. 2019;60(13):4171–4177. doi: 10.1167/iovs.19-27087. doi:10.1167/iovs.19-27087. [DOI] [PubMed] [Google Scholar]
- 7.Germann JA, Martinez-Enriquez E, Marcos S. Quantization of collagen organization in the stroma with a new order coefficient. Biomed Opt Express. 2017;9(1):173–189. doi: 10.1364/BOE.9.000173. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Park K, Shin J, Lee J. Relationship between corneal biomechanical properties and structural biomarkers in patients with normal-tension glaucoma: a retrospective study. BMC Ophthalmol. 2018;18:7. doi: 10.1186/s12886-018-0673-x. https://doi.org/10.1186/s12886-018-0673-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156–162. doi: 10.1016/j.jcrs.2004.10.044. [DOI] [PubMed] [Google Scholar]
- 10.Terai N, Raiskup F, Haustein M, Pillunat LE, Spoerl E. Identification of biomechanical properties of the cornea: the ocular response analyzer. Curr Eye Res. 2012;37(7):553–562. doi: 10.3109/02713683.2012.669007. [DOI] [PubMed] [Google Scholar]
- 11.Jóźwik A, Kasprzak H, Kuczma M. New Approach to the Analysis of Raw Data from the Ocular Response Analyzer. Biomed Hub. 2016;1:1–11. doi: 10.1159/000452124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Blackburn BJ, Jenkins MW, Rollins AM, Dupps WJ. A Review of Structural and Biomechanical Changes in the Cornea in Aging, Disease, and Photochemical Crosslinking. Front. Bioeng. Biotechnol. 2019;7:66. doi: 10.3389/fbioe.2019.00066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Zhang H, Sun Z, Li L, et al. Comparison of intraocular pressure measured by ocular response analyzer and Goldmann applanation tonometer after corneal refractive surgery: a systematic review and meta-analysis. BMC Ophthalmology. 2020;20:23. doi: 10.1186/s12886-019-1288-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Feizi S. Corneal endothelial cell dysfunction: etiologies and management. Therapeutic advances in ophthalmology. 2018 Dec 7;10:2515841418815802. doi: 10.1177/2515841418815802. doi: 10.1177/2515841418815802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Sanchis-Gimeno JA, Lleó-Pérez A, Alonso L, Rahhal MS, Martínez Soriano F. Corneal endothelial cell density decreases with age in emmetropic eyes. Histology and Histopathology. 2005 Apr;20(2):423–427. doi: 10.14670/HH-20.423. [DOI] [PubMed] [Google Scholar]
- 16.Szentmary N, Butskhrikidze T, Hager T, Langenbucher A. Analysis and interpretation of specular microscopy measurements of the corneal endothelium. Acta Ophthalmologica. 2018;96:122–122. [Google Scholar]
- 17.McCarey BE, Edelhauser HF, Lynn MJ. Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices, and new intraocular drugs and solutions. Cornea. 2008;27(1):1–16. doi: 10.1097/ICO.0b013e31815892da. doi: 10.1097/ICO.0b013e31815892da. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.World Medical Association World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA. 2013 ;310(20):2191–2194. doi: 10.1001/jama.2013.281053. [DOI] [PubMed] [Google Scholar]
- 19.Rosa N, Lanza M, De Bernardo M, Signoriello G, Chiodini P. Relationship Between Corneal Hysteresis and Corneal Resistance Factor with Other Ocular Parameters. Semin Ophthalmol. 2015;30(5-6):335–339. doi: 10.3109/08820538.2013.874479. [DOI] [PubMed] [Google Scholar]
- 20.Detry-Morel M, Jamart J, Pourjavan S. Evaluation of corneal biomechanical properties with the Reichert Ocular Response Analyzer. Eur J Ophthalmol. 2011;21:138–148. doi: 10.5301/ejo.2010.2150. [DOI] [PubMed] [Google Scholar]
- 21.Johnson RD, Nguyen MT, Lee N, Hamilton DR. Corneal biomechanical properties in normal, forme fruste keratoconus, and manifest keratoconus after statistical correction for potentially confounding factors. Cornea. 2011;30:516–523. doi: 10.1097/ICO.0b013e3181f0579e. [DOI] [PubMed] [Google Scholar]
- 22.Berdahl J, Medeiros F, Radcliffe N, Grover D. A New Glaucoma “VITAL SIGN” How Corneal Hysteresis is Improving Glaucoma Patient Management. Review of Ophthalmology. 2019 [Google Scholar]
- 23.Souza MB, de Medeiros FW, Villela FF, Alves MR. Relationship between the biomechanical properties of the cornea and anterior segment measurements. Clinics (Sao Paulo) 2018 ;73:e491. doi: 10.6061/clinics/2018/e491. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Grzybowski A, Och M, Kanclerz P, Leffler C, Moraes CG. Primary Open Angle Glaucoma and Vascular Risk Factors: A Review of Population Based Studies from 1990 to 2019. J Clin Med. 2020;9(3):761. doi: 10.3390/jcm9030761. doi:10.3390/jcm9030761. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Vinciguerra R, Elsheikh A, Roberts CJ, Ambrósio R Jr, Kang DS, Lopes BT, et al. Influence of pachymetry and intraocular pressure on dynamic corneal response parameters in healthy patients. J Refract Surg. 2016;32:550–561. doi: 10.3928/1081597X-20160524-01. [DOI] [PubMed] [Google Scholar]
- 26.Touboul D, Roberts C, Kerautret J. Correlations between corneal hysteresis, intraocular pressure, and corneal central pachymetry. J Cataract Refract Surg. 2008 ;34(4):616–622. doi: 10.1016/j.jcrs.2007.11.051. [DOI] [PubMed] [Google Scholar]
- 27.Sharifipour F, Panahi-Bazaz M, Bidar R, Idani A, Cheraghian B. Age-related variations in corneal biomechanical properties. J Curr Ophthalmol. 2016 ;28(3):117–122. doi: 10.1016/j.joco.2016.05.004. doi:10.1016/j.joco.2016.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Lee H, Kang DS, Ha BJ, et al. Biomechanical Properties of the Cornea Measured With the Dynamic Scheimpflug Analyzer in Young Healthy Adults. Cornea. 2017 ;36(1):53–58. doi: 10.1097/ICO.0000000000001001. [DOI] [PubMed] [Google Scholar]
- 29.Lee H, Kang DSY, Ha BJ, et al. Biomechanical Properties of the Cornea Using a Dynamic Scheimpflug Analyzer in Healthy Eyes. Yonsei Med J. 2018 ;59(9):1115–1122. doi: 10.3349/ymj.2018.59.9.1115. doi:10.3349/ymj.2018.59.9.1115. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Allam RS, Khalil NM. Evaluation of sex differences in corneal hysteresis. Eur J Ophthalmol. 2015 ;25(5):391–395. doi: 10.5301/ejo.5000572. [DOI] [PubMed] [Google Scholar]
- 31.Qiu K, Lu X, Zhang R, Wang G, Zhang M. Corneal Biomechanics Determination in Healthy Myopic Subjects. J Ophthalmol. 2016 ;2016:2793516. doi: 10.1155/2016/2793516. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Wu W, Dou R, Wang Y. Comparison of Corneal Biomechanics Between Low and High Myopic Eyes-A Meta-analysis. Am J Ophthalmol. 2019 ;207:419–425. doi: 10.1016/j.ajo.2019.07.007. [DOI] [PubMed] [Google Scholar]
- 33.Wan K, Cheung SW, Wolffsohn JS, Orr JB, Cho P. Role of corneal biomechanical properties in predicting of speed of myopic progression in children wearing orthokeratology lenses or single-vision spectacles. BMJ Open Ophthalmol. 2018;3(1):e000204. doi: 10.1136/bmjophth-2018-000204. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Qiu K, Lu X, Zhang R, Wang G, Zhang M. Relationship of corneal hysteresis and optic nerve parameters in healthy myopic subjects. Sci Rep. 2017;7(1):17538. doi: 10.1038/s41598-017-15386-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Akova-Budak B, Kıvanç SA. Does corneal hysteresis correlate with endothelial cell density? Med Sci Monit. 2015;21:1460–1463. doi: 10.12659/MSM.893589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Dong J, Zhang Y, Zhang H, et al. Corneal densitometry in high myopia. BMC Ophthalmol. 2018;18(1):182. doi: 10.1186/s12886-018-0851-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Kato Y, Nakakura S, Asaoka R, Matsuya K, Fujio Y, et al. Cataract surgery causes biomechanical alterations to the eye detectable by Corvis ST tonometry. PLOS ONE. 2017;12(2):e0171941. doi: 10.1371/journal.pone.0171941. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Hejsek L, Kadlecova J, Sin M, Velika V, Jiraskova N. Intraoperative intraocular pressure fluctuation during standard phacoemulsification in real human patients. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2019;163(1):75–79. doi: 10.5507/bp.2018.065. [DOI] [PubMed] [Google Scholar]
- 39.Yune Z, Xingyu L, Aizhu T, Jianhua W, Fan L. Intraocular Pressure and Calculated Diastolic Ocular Perfusion Pressure during Three Simulated Steps of Phacoemulsification In Vivo. Invest. Ophthalmol. Vis. Sci. 2009;50(6):2927–2931. doi: 10.1167/iovs.08-2996. [DOI] [PubMed] [Google Scholar]
- 40.Bamdad S, Bolkheir A, Sedaghat MR, Motamed M. Changes in corneal thickness and corneal endothelial cell density after phacoemulsification cataract surgery: a double-blind randomized trial. Electron Physician. 2018;10(4):6616–6623. doi: 10.19082/6616. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Hirasawa K, Nakakura S, Nakao Y, et al. Changes in Corneal Biomechanics and Intraocular Pressure Following Cataract Surgery. Am J Ophthalmol. 2018;195:26–35. doi: 10.1016/j.ajo.2018.07.025. [DOI] [PubMed] [Google Scholar]
- 42.Faramarzi A, Feizi S, Maghsoodlou A. Factors influencing intraocular pressure, corneal thickness and corneal biomechanics after congenital cataract surgery. British Journal of Ophthalmology. 2017;101:1493–1499. doi: 10.1136/bjophthalmol-2016-310077. [DOI] [PubMed] [Google Scholar]
- 43.Zhang Z, Yu H, Dong H, Wang L, Jia YD, Zhang SH. Corneal biomechanical properties changes after coaxial 2.2-mm microincision and standard 3.0-mm phacoemulsification. Int J Ophthalmol. 2016;9(2):230–234. doi: 10.18240/ijo.2016.02.08. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Ahmed MS, Hassanin OA, Abou El-Enin AI, Rashwan AH. Corneal biomechanics and intraocular pressure changes after uneventful phacoemulsification surgery. J Egypt Ophthalmol Soc. 2019 ;112:145–153. [Google Scholar]