Skip to main content
PLOS One logoLink to PLOS One
. 2020 Jul 9;15(7):e0234899. doi: 10.1371/journal.pone.0234899

Multi-laboratory validation of the xMAP—Food Allergen Detection Assay: A multiplex, antibody-based assay for the simultaneous detection of food allergens

Eric A E Garber 1, Chung Y Cho 1,*, Prasad Rallabhandi 1, William L Nowatzke 2,¤a, Kerry G Oliver 2, Kodumudi Venkat Venkateswaran 3,¤b, Neeraja Venkateswaran 4
Editor: Katerina Kourentzi5
PMCID: PMC7347184  PMID: 32645020

Abstract

The increasing prevalence of individuals with multiple food allergies and the need to distinguish between foods containing homologous, cross-reactive proteins have made the use of single-analyte antibody-based methods (e.g., ELISAs) sometimes insufficient. These issues have resulted in the need to conduct multiple analyses and sometimes employ orthogonal methods like mass spectrometry or DNA-based methods for confirmatory purposes. The xMAP Food Allergen Detection Assay (xMAP FADA) was developed to solve this problem while also providing increased throughput and a modular design suitable for adapting to changes in analytical needs. The use of built-in redundancy provides the xMAP FADA with built-in confirmatory analytical capability by including complementary antibody bead sets and secondary analytical end points (e.g., ratio analysis and multi-antibody profiling). A measure of a method’s utility is its performance when employed by analysts of varying expertise in multiple laboratory environments. To gauge this aspect, a multi-laboratory validation (MLV) was conducted with 11 participants of different levels of proficiency. The MLV entailed the analysis of incurred food samples in four problematic food matrices, meat sausage, orange juice, baked muffins, and dark chocolate. Except for a couple of instances, involving two confirmatory components in the analysis of baked muffins, the allergenic foods were detected by all participants at concentrations in the analytical samples comparable to ≤ 10 μg/g in the original food sample. In addition, despite high levels of inter-lab variance in the absolute intensities of the responses, the intra-laboratory reproducibility was sufficient to support analyses based on the calibration standards and direct comparison controls (DCCs) analyzed alongside the samples. In contrast, ratio analyses displayed inter-laboratory %CV (RSDR) values < 20%; presumably because the ratios are based on inherent properties of the antigenic elements. The excellent performance of the xMAP FADA when performed by analysts of varying proficiency indicates a reliability sufficient to meet analytical needs.

Introduction

Over 15 million Americans have at least one food allergy [1]. The only way these individuals can avoid an allergic reaction entails not consuming products that contain the allergenic food. In 2004 the Food Allergen Labeling and Consumer Protection Act (FALCPA) was passed to facilitate the consumer’s ability to determine which products to avoid while maintaining a diverse, healthy diet [2]. As part of the enforcement of FALCPA, the FDA must be able to analyze a diverse group of foods for the presence of any of the allergenic foods specified in the act. Complicating the analytical process is the increasing prevalence of people with multiple food allergies [3, 4], as well as the need to detect unknown amounts of allergenic food ranging from trace levels (e.g., micrograms per oral portion) to substantial levels (e.g., >10%). In addition, the complexity of the world market and need to distinguish between related foods containing homologous cross-reactive proteins makes single-analyte methods, such as the commonly available commercial ELISA test kits, insufficient for many circumstances. Alternative, orthogonal methods, such as PCR and mass spectrometry are still under development and not yet universally recognized as suitable for routine regulatory enforcement and as such are not routinely used by contract and governmental laboratories.

In 2014, the FDA with Radix BioSolutions developed a novel xMAP-based multiplex assay for the simultaneous detection of 14 food allergens plus gluten [5], and sesame [6] based on principles associated with ELISA technology. The xMAP Food Allergen Detection Assay (xMAP FADA) entailed two extraction protocols, buffered-detergent (using either Phosphate Buffered Saline with 0.05% Tween®-20 or UD Buffer) and reduced-denatured (0.5% SDS/2% β-mercaptoethanol). The buffered-detergent extracts are interrogated using a cocktail consisting of 29 antibodies conjugated to different color-coded bead sets. Specifically, two antibodies (bead set numbers denoted as -x, -y) for almond (-12, -13), Brazil nut (-14, -15), cashew (-18, -19), coconut (-20, -21), egg (-25, -26), gluten (-27, -28), hazelnut (-29, -30), macadamia (-33, -34), milk (-35, -36), peanut (-37, -38), pine nut (-39, -42), pistachio (-43, -44), soy (-45, -46), and walnut (-47, -48) and one antibody for crustacean seafood (-22). The reduced-denatured extracts are analyzed using a cocktail containing one antibody for egg (-65), peanut (-72), and gluten (-73) and two for milk (-66, -67).

The xMAP FADA uses built-in redundancy by having two or more bead sets for each allergen target in a single simultaneous assay, which is not possible in conventional ELISAs. By ensuring concurrence of results for the two or more bead sets per allergen target in a given sample, the probability of false positives and false negatives can be lowered. This is by virtue of its design and incorporation of the AssayChex bead sets designed by Radix BioSolutions, Ltd., (Georgetown, TX). The AssayChex bead sets assess four technical aspects of xMAP performance; namely, instrumental performance, ‘non-specific’ binding, detector antibody, and streptavidin-phycoerythrin binding. As such, the need to obtain control material samples, identical to the sample being analyzed but certified as allergen-free, to rule out false positives and false negatives is eliminated. Further, through the use of ratio analysis between complementary bead sets (e.g., almond-12:almond-13) and multi-antibody profiling it is possible to detect and distinguish between homologous, cross-reactive antigenic foods. Several papers describing the performance of the xMAP FADA as both a research tool and for the analysis of regulatory samples have been published [710].

The xMAP FADA has also undergone extensive single lab validation (SLV) examining the performance of the assay to detect each of the targeted food allergens individually or as a mixture in the presence of food extracts and when incurred into buffer, orange juice, dark chocolate, pancake batter, and baked muffins [11], and assay robustness study under various experimental conditions [12]. The xMAP FADA successfully detected the analytes in almost all cases with limits of detection (LoDs) considerably less than the lowest calibration standard (S1), indicative of a potential flexibility to extend the dynamic range should greater sensitivity be desired. As observed with commercial ELISA test kits that employ a buffered-detergent extraction protocol, the recoveries for analyte spiked into food extracts varied typically between 50–150% and decreased when incurred into processed foods, but on only three occasions was it impossible to detect analyte. In contrast, less problems were observed in detecting incurred analytes upon extracting the samples using the reduced-denatured protocol; similar to the performance observed with ELISAs based on a reduced-denatured extraction protocol.

To examine inter-laboratory / inter-analyst variability in the performance of the xMAP FADA, an 11 laboratory multi-laboratory validation (MLV) was performed of analytes incurred in meat-sausage, orange juice, baked muffin, and dark chocolate. To gauge inter-analyst performance, the participants were deliberately chosen to reflect a diverse level of experience. Only two participants were proficient in performing the xMAP FADA. The remaining nine participants were either novices or inexperienced with either the xMAP FADA or xMAP technology; specifically, five participants had prior experience running ELISAs and were provided with two days training on the xMAP FADA and two practice samples, two participants were experts on xMAP technology with no experience with food allergen analysis, one participant had prior experience running ELISAs and limited experience on xMAP technology, and one was new to both xMAP and food allergen analysis. All eleven laboratories generated data that was included in the analyses. The only substantive problems were an inability to ship the meat samples to one of the participants in a foreign country and another laboratory inadvertently failed to specify in the data collection program that the instrument should monitor the results for two analytes/four of the 29 bead sets in the buffered-detergent protocol. Otherwise, the MLV was a success with only minimal analytical problems, all involving the baked muffin samples. As expected, variance across the laboratories was considerable when comparing the absolute median fluorescent intensity (MFI) responses. However, ratio analysis, which is based on inherent physical properties of the antigen, displayed inter-laboratory variations for the analytes at 0.5 μg/mL in the analytical samples derived from meat at 15%, orange juice at 15%, baked muffins at 26%, and dark chocolate at 17%. At lower concentrations the %CV values increased with some analytes not being detected with an MFI in excess of the lowest calibration standard (S1).

Materials and methods

Reagents

Phosphate buffered saline (PBS, cat# P5368), Tween®-20 (cat# P9416), and wheat gluten (cat# G5004) were purchased from Sigma-Aldrich Inc. (St. Louis, MO). BD Difco skim milk powder (cat# DF0032-17-3) and Sodium Dodecyl Sulfate (SDS, cat# 28312) were purchased from Fisher Scientific (Waltham, MA). Whole, raw nuts and legumes were acquired from nuts.com as previously described [5]. All other reagents were of the highest technical grade available. Ingredients (all allergen-free) to prepare the food samples are described below.

Food samples

All food samples were prepared from ingredients acquired locally and selected based on being allergen-free. The primary focus in preparing the food samples was assuring consistent, accurate allergenic food content while exposing the allergenic foods to processing and conditions comparable to consumed products. Thus, the muffin batter was baked at the manufacturer’s recommended temperature of 350°C to generate an edible product.

Meat samples were prepared from ground pure beef, skinless, filler-free hot dogs. The hot dogs were ground in a Robot Coupe® Blixer® 3 Series D (Robot Coupe U.S.A., Inc., Ridgeland, MS) and aliquoted into one-gram portions into 50 mL conical tubes. The allergenic food mixtures were mixed into the meat and stored at -80°C prior to shipping overnight on dry ice.

Orange juice was pulp-free, not from concentrate, pure orange juice. One mL aliquots were pipetted into 50 mL conical tubes to which the allergenic food mixtures were added, mixed, and stored at -80°C prior to shipping overnight on dry ice.

Dark chocolate was pure dark chocolate pellets that consistently tested milk and allergen-free. One-gram aliquots were placed into 50 mL conical tubes, placed for 15 min into a 60°C water bath to melt the chocolate, the allergenic food mixture added and vortexed for 2–3 seconds before being allowed to cool and solidify at room temperature (approx. 22°C). The samples were stored at room temperature and shipped overnight.

Baked muffin samples were prepared using a using a commercial gluten-free rice flour-based cake mix, soy & dairy-free buttery spread (as a substitute for butter), and canola oil. The ingredients were mixed at the ratios on the package instructions (omitting the addition of an egg) to generate the batter. The batter was aliquoted into one-gram portions on mini-cupcake paper wrappers, the allergenic food mixtures mixed into the batter, and baked on a metal pan at 350°C for nine minutes. The baked muffins were stored at 4°C in sealed 50 mL conical tubes until shipped overnight on ice.

The allergen food mixtures incurred into the foods were derived from nonfat dried milk (NFDM) that was multi-analyst validated with multiple commercial ELISA test kits to perform comparable to NIST SRM 1549; NIST SRM 2387 peanut butter; wheat gluten (cat # G5004, Sigma-Aldrich Co.) that has been previous used in analytical studies [13, 14]; organic, dry soy beans a product of ChoripDong (distributed by Seoul Shik Poom, Inc., Flushing NY, USA). Raw, organic (no shell) almonds, hazelnut, and walnuts, and the organic, shredded coconut; were acquired from Nuts.com. The tree nuts and soybeans were ground using an IKA A11 Basic Analytical Mill (Wilmington, NC). The milk and egg stock solutions were prepared at 10% m/v (100,000 ppm) in PBST using a Potter-Elvehjem Tissue Grinder to achieve homogeneity. The legume and tree nut stock solutions were prepared at 1% w/v (10,000 ppm). The wheat gluten was added as a solid to the most concentrated allergenic food mixture (sub-stock 5, SS-5), vortexed and allowed to sit for 2 hours at room temperature prior to four-fold dilution of an aliquot to prepare sub-stock 4 (SS-4), 2.5-fold of an aliquot of SS-4 to prepare SS-3, four-fold dilution of an aliquot of SS-3 to make SS-2, and 2.5-fold dilution of an aliquot of SS-2 to make SS-1. The SS-5 allergenic food mixture used in preparing the meat samples was prepared by mixing a ratio of 1 mL 1% soy: 0.1 mL 10% egg: 0.1 mL 10% NFDM: 10 mg wheat gluten: 1.3 mL PBST. The SS-5 used to prepare the orange juice samples was prepared at a ratio of 1 mL 1% almond: 0.1 mL NFDM: 1 mL 1% soy: 0.4 mL PBST. The SS-5 used to prepare the baked muffin samples was prepared at a ratio of 1 mL 1% coconut: 1 mL 1% walnut: 0.1 mL 10% egg: 0.1 mL 10% NFDM: 10 mg wheat gluten: 0.3 mL PBST. The SS-5 used to prepare the dark chocolate samples was prepared at a ratio of 1 mL 1% hazelnut: 0.1 mL 10% NFDM: 1 mL 1% peanut: 0.4 mL PBST. The 100, 25, 10, 2.5, and 1 ppm (either μg/g or μg/mL) incurred food samples were prepared using 25 μL of SS-5, SS-4, SS-3, SS-2, and SS-1, respectively. By using only 25 μL per gram (or mL) of food the goal was to minimize any artificial dilution of the food matrix. The above selected allergens are possibly present more commonly in similar types of commercially available foods, and we varied the critical concentrations of the appropriate allergens for multi-laboratory validation of the xMAP FADA platform.

Food sample shipping and storage

All food samples were coded and supplied in 50 mL sealed conical tubes. The meat and orange juice samples were stored frozen and shipped overnight on dry ice. The baked muffin samples were stored at 4°C and shipped on ice, while the dark chocolate samples were stored and shipped at room temperature. In addition to the coded, incurred food samples, all participants were supplied with analyte-free food samples and reference materials (i.e., NIST SRM 2387 peanut reference material, non-fat dried milk reference material multi-analyst validated as comparable to NIST SRM 1549 using ELISA test kits, and wheat gluten reference material used for ELISA studies [13, 14]) for preparing the DCCs.

XMAP Food Allergen Detection Assay

The xMAP FADA was performed according to the manufacturer’s instructions, which also provides a detailed description of the antibodies [5, S1 Appendix]. Three of the food samples (meat, orange juice, baked muffins) were extracted according to the PBST-based extraction protocol by adding 20 mL of PBST to 1 g sample, vortex, and extracted for 2 hours at room temperature. The dark chocolate samples were extracted using the 40 mL of UD buffer (105 mM sodium phosphate/75 mM NaCl/ 2.5% Difco skim milk powder/0.05% Tween®-20) by following the UD buffer extraction protocol. All samples were extracted in triplicate on the day of analysis. The PBST extracts were diluted 10-fold prior to mixing with the bead cocktail, while the UD buffer extracts were diluted 5-fold with UD buffer; net dilution of all food samples being 200-fold prior to mixing with the bead cocktail. The calibration standards, bead cocktail, detection antibody cocktail, streptavidin-phycoerythrin, and direct comparison controls (see S1 Appendix instruction) were prepared on the day of use.

Calibrations standards S0 (buffer, background), S1, S2, S5, and S7 were included in all analyses alongside the samples. The concentrations of the various allergenic food protein extracts in the calibration standards are tabulated in Table 1 along with the equivalent concentrations necessary in the food samples, which undergo 200-fold dilution prior to mixing with the bead cocktail, to generate the same amount as in the calibration standards. Table 2 lists the protein content of the various allergenic foods used to calculate between protein and allergenic food.

Table 1. Calibration standards for buffered-detergent analyses & equivalence in food unitsa.

ALMOND BRAZIL NUT CASHEW COCONUT CRUST
as‐is b dil corr c food corr d as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr
ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm
S1 2.9 0.6 2.7 2.9 0.6 4.4 0.5 0.1 0.6 0.6 0.1 1.7 15 3 17
S2 5.2 1 4.9 5.2 1.0 7.8 0.8 0.2 1 1.1 0.2 3.1 28 5.6 31
S5 30 6 28 30.0 6.0 45 5 1 5.8 6.7 1.3 19 162 32 181
S7 99 20 93 99 20 149 16 3.2 19 22 4.4 62 525 105 588
EGG GLUTEN HAZELNUT MACADAMIA MILK
as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr
ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm
S1 5.8 1.2 2.4 3.7 0.7 0.9 1.2 0.2 1.8 4.8 1 12 1.5 0.3 0.8
S2 10 2 4.2 6.6 1.3 1.7 2.1 0.4 3.2 8.7 1.7 22 2.6 0.5 1.5
S5 61 12.2 25 39 7.8 9.8 12.0 2.4 18 51 10 129 15 3 8.5
S7 198 40 83 125 25 31 40 8 60 165 33 418 49 9.8 28
PEANUT PINE NUT PISTACHIO SOY WALNUT
as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr as‐is dil corr food corr
ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm ng/mL μg/mL ppm
S1 1.5 0.3 1.2 8.1 1.6 11 2.9 0.6 2.9 9.2 1.8 5.2 9.2 1.8 11
S2 2.8 0.6 2.2 15.0 3.0 21 5 1 5.2 17 3.4 9.5 17 3.4 20
S5 16 3.2 13 85 17 119 30 6 30 96 19 54 96 19 115
S7 53 11 42 274 55 383 99 20 98 313 63 175 313 63 376

a Concentration of calibration standards after preparation (200-fold dilution of frozen stock and 1.8-fold serial dilutions. S0 is analyte-free, either PBST buffer or UD buffer. Not included in MLV are S6 (1.8-fold dilution of S7), S4 (1.8-fold dilution of S5), and S3 (1.8-fold dilution of S4).

b Calibration standards are supplied as ng extractable protein per mL.

c Calibration standards are not extracted nor diluted (net 200-fold), thus equivalent concentration a food sample would have to contain.

d Equivalent ppm of allergenic food the food sample would have to contain based on the protein content of the allergenic food (see Protein content table).

Table 2. Protein content of allergenic foods used as calibrants.

% Protein Reference a
Almond 21 raw almonds UPC:014113210638
Brazil Nut 13 raw Brazil nuts UPC 708820008175
Cashew 17 raw cashews UPC 019061198014
Coconut 7 raw coconut UPC 033674100110
Crustacean 18 raw black tiger UPC 857536005029
Egg 48 NIST RM 8445
Gluten 80 Sigma‐Aldrich G5004
Hazelnut 13 raw hazelnuts UPC 03003491412
Macadamia 8 raw macadamia UPC 072989767519
Milk 36 1093, milk dry nonfat (Carnation 36%)
Peanut 25 16095, raw, Virginia; standardized vs NIST SRM 2387
Pine Nut 14 raw pine nuts UPC 041497132430
Pistachio 20 12151, raw pistachio nuts
Soy 36 soybeans UPC 019061190308
Walnut 17 raw walnuts UPC 070038645740

a Legumes, tree nuts, and crustacean referenced to entries in USDA, ARS FoodData Central, formerly National Nutrient Database (accessed November 20, 2019; rest as delineated.

Instrumentation

Inasmuch as the results obtained from xMAP FADA analysis using Luminex® 100/200 (Luminex Corp., Austin, TX), Bio-Plex 200 (Bio-Rad Laboratories, Inc., Hercules, CA), and MagPix® (Luminex Corp., Austin, TX) instrumentation are indistinguishable [11], no restriction was placed on the instruments used in the validation. Thus, nine of the participants employed MagPix® instrumentation and two employed Luminex® 100/200 instrumentation.

Validation design

The multi-laboratory validation was conducted as a Level IV: Full Collaborative Study according to the Guidelines for the Validation of Chemical Methods in Food, Feed, Cosmetics, and Veterinary Products, 3nd Edition; U.S. Food and Drug Administration Foods Program. October 2019 as delineated in Table 3 [15]. The validation exceeded the requirements in that 11 laboratories participated in this quantitative validation, 4 food matrices of differing physical and chemical properties were used (meat, orange juice, baked muffin, and dark chocolate), allergens were incurred at 5 levels plus one blank, and everything was done in triplicate. To ensure that the MLV reflected inter-laboratory variability in the preparation, extraction, and analysis of the samples, each sample was individually incurred with the appropriate allergens as one gram, the size of an analytical portion, prior to processing. As such, changes in mass associated with processing (e.g., loss of water during muffin baking) did not result in variability of allergen content available for extraction. Specifically, muffin batter samples incurred with 0, 1, 2.5, 10, 25, or 100 μg analyte resulted in 0, 1, 2.5, 10, 25, or 100 μg of potentially extractable analyte in the encoded samples analyzed by the participants. The focus on increased reliability in the amount (μg) of allergenic food present in the test portion is of importance since the dose (μg) ingested, rather than the concentration per unit volume, determines any potential health risk.

Table 3. Key validation parameter requirements for chemical methods.

Level One: Emergency/ Limited Use Level Two: Single Laboratory Validation Level Three: Multi-Laboratory Validation Level Four: Full Collaborative
Number participating labs 1 1 ≥ 2 8 (quantitative)
10 (qualitative)
Number of matrix sources per matrix* ≥1 ≥3 recommended where available ≥3 recommended where available ≥3 recommended where available
Number of analyte(s) spike levels for at least one matrix source** ≥2 spike levels + 1 matrix blank ≥3 spike levels + 1 matrix blank ≥3 spike levels +1 matrix blank ≥3 spike levels +1 matrix blank
Replicates required per matrix source at each level tested/lab ≥2 (quantitative)
≥2 (qualitative)
≥2 (quantitative)
≥3 (qualitative)
≥2 (quantitative)
≥3 (qualitative)
≥2 (quantitative)
≥3 (qualitative)
Replicates required at each level tested per laboratory if only one matrix source used ≥4 (quantitative)
≥6 (qualitative)
≥6 (quantitative)
≥9 (qualitative)
≥3 (quantitative)
≥6 (qualitative)
≥2 (quantitative)
≥6 (qualitative)

*If a variety of food matrices with differing physical and chemical properties are selected, the number of sources for each food sample matrix may be one or more, but if only one food matrix is studied then ≥3 sources are recommended where available. The number of matrix sources may be reduced, particularly if it is difficult to obtain blank matrix sources, as long as the total number of spike levels and matrix combinations are adequate (e.g., 6 replicates or greater at each spike level for quantitative methods and 9 replicates or greater for qualitative methods).

** Number of spike levels is recommended for at least one source of matrix. Other similar sources of matrix (e.g., within the same category; see Appendix 4) may be studied at one or two spike levels (e.g., at an action/guidance or tolerance level or close to the lower limit of quantitation/detection).

The eleven laboratories that participated in the validation were chosen to represent a variety of expertise. Thus, no individual laboratory was excluded from inclusion in the data analyses despite an analysis of potential outliers using probabilistic models might support dropping the data from up to three lower performing laboratories (the least proficient). Instead, only on rare occasions were specific data points not included in the processed data based on Extreme Value Analysis. Thus, data was only omitted on the rarest of occasions, in which it displayed characteristics of catastrophic failure that would have been easily spotted by an analyst and by definition could not be used (e.g., generation of negative MFI when adjusted for background or the lack of a dynamic response).

To assist the analysts and re-enforce uniformity, the participants were supplied with the same written instructions (see S1 Appendix). This included a copy of the product insert, instructions on how to configure the instrumentation (the SLV demonstrated that both the Luminex/BioPlex and MagPix instruments generated comparable results [11]), and instructions regarding which coded samples were to be extracted using the PBST Buffered-detergent extraction protocol (meat, orange juice, baked muffins) and which using the UD Buffered-detergent extraction protocol (dark chocolate). The analysts were also instructed to perform a (normally optional) 10-fold dilution with PBST of the PBST-based extracts and a 5-fold dilution with UD Buffer of the UD Buffer-based extracts; the difference reflecting that the PBST extractions entailed a 20-fold dilution of the samples, while the UD Buffer extractions entailed a 40-fold dilution. Included in the analyses, the participants were asked to prepare direct comparison control samples (DCC) which entailed spiking one-gram portions of analyte-free foods with either 50 μg of milk, 10 μg of peanut, or 20 μg of wheat gluten; all done according to detailed procedures provided to the participants. In order to accommodate triplicate direct comparison control samples (DCCs), only the S0, S1, S2, S5, and S7 calibration standards were employed. Lastly, the analysts were instructed to use a microtiter plate as a template to facilitate pipetting and place the diluted extracts into alternating columns to reduce the possibility of inadvertent cross contact. Ten of the eleven laboratories were shipped (overnight) at the same time the food samples and xMAP FADA reagents; the eleventh joined the validation approximately two months later. The FADA reagents were from the same production run (Lot 5) and provided in sufficient quantity for 240 analyses (wells of microtiter plates). Problems were only encountered for international shipping of the meat samples.

Results and discussion

Calibration standards

Calibration standards serve a minimum of two essential roles. The one obvious purpose is for determining analyte concentration from standard curve interpolation. Table 1 lists the ng extractable soluble protein per mL for the calibration standards (S3, S4, and S6 were not included in the MLV) and the equivalents of allergenic food (ppm) based on conversion factors taken from the USDA FoodData Central database [16], assuming a 200-fold dilution prior to mixing with the bead set cocktail. Of possibly greater importance is the role calibration standards play as a check on assay performance. Since, the calibration standards are a constant between all production lots and preparation entails just dilution of a stock sample that is stored frozen until use, any changes in performance reflect either changes in the reagents or the ability of the analyst to perform the assay. As such, the performance of the calibration standards in three lots produced over a five-year period provided information on lot-to-lot variability. Further, the calibration standards in the MLV provided an opportunity to examine variability in analyst proficiency without the complexity associated with food sample preparation-extraction and any complications associated with analyzing complex, processed food samples.

Lot-to-lot variability

The performance of the calibration standards by three lots produced during a five-year period as analyzed by a highly proficient analyst, the same for Lots 3 and 5, is presented in Tables 4 and 5. Tabulated in Table 4 for each lot are the average background MFI for each bead set; MFI generated by the calibration standards after subtracting background (S0); and the MFI generated by DCCs containing 20 ppm gluten (G20), 10 ppm peanut (P10), and 50 ppm non-fat dried milk (50M). The three lots are from the first production run (Lot 1), the third production run used for the SLV, and the fifth production run which was prepared for the MLV; all analyses conducted within six months of production. The Lot 1 and Lot 3 data were derived from triplicate samples while the Lot 5 data represents the average of triplicate samples from each of four separate experiments (n = 3 x 4). Calibration standards S3, S4, and S6 were not included in the Lot 5 analyses. Also included purely for informational purposes in the tables, in orange font, are the averages across the three lots; though such should not be used to analyze data. As expected, all bead sets displayed excellent dynamic responses.

Table 4. Lot‐to‐lot variability of calibration standards over five yearsa.

ALMOND b BRAZIL NUT CASHEW COCONUT CRUST EGG GLUTEN HAZELNUT
12 13 14 15 18 19 20 21 22 25 26 27 28 29 30
AVERAGE BACKGROUND MFI
S0 Lot 1 394 193 199 128 243 192 107 133 207 242 245 223 141
S0 Lot 3 34 41 58 39 88 50 77 37 56 101 1111 83 179 88 59
S0 Lot 5 65 53 81 51 130 36 122 48 105 398 3634 102 292 207 48
AVERAGE INTENSITIES (MFI) CALIBRATION STANDARDS ABOVE BACKGROUND (minus average S0)
S1 Lot 1 2409 1168 403 84 2948 369 1978 79 367 405 2225 719 136
S1 Lot 3 2146 801 1363 611 2160 706 1757 70 492 1318 2185 275 775 961 285
S1 Lot 5 2506 990 977 460 2484 1200 1817 68 427 2294 3393 243 563 380 329
S1 av c 2354 986 914 385 2530 758 1851 72 428 1806 2789 308 1188 686 250
S2 Lot 1 3619 1639 649 135 4564 658 3814 121 632 599 3264 1326 253
S2 Lot 3 3595 1379 2609 1112 3540 1241 3374 124 834 2226 3916 400 1163 1688 473
S2 Lot 5 4526 1706 1811 828 4359 2047 3728 119 742 4215 6204 290 664 657 570
S2 av 3913 1574 1690 692 4154 1315 3639 121 736 3220 5060 430 1697 1224 432
S3 Lot 1 5528 2862 1335 267 6957 1133 7663 232 1261 851 4284 2432 452
S3 Lot 3 5533 2329 4607 1957 5658 2103 6151 199 1434 3802 6778 554 1682 2845 853
S3 av 5531 2596 2971 1112 6308 1618 6907 216 1348 3802 6778 702 2983 2638 652
S4 Lot 1 7452 4431 2441 481 9923 1821 12234 356 2228 1194 6188 4355 810
S4 Lot 3 7498 3525 7792 3161 7806 3442 9420 318 2145 5791 10212 752 2421 4737 1411
S4 av 7475 3978 5117 1821 8864 2632 10827 337 2186 5791 10212 973 4305 4546 1110
S5 Lot 1 9939 7042 5549 1380 13761 3516 17126 537 4331 1832 9445 6976 1497
S5 Lot 3 9420 5023 11633 4935 10368 5129 12917 444 3343 7994 13781 982 3368 6822 2078
S5 Lot 5 15814 7344 11895 4813 12710 8768 16803 466 3264 13589 20555 852 2425 2052 2412
S5 av 11724 6469 9692 3709 12280 5804 15615 482 3646 10791 17168 1222 5079 5283 1996
S6 Lot 1 11980 9248 10441 2635 17006 5542 19627 757 7603 2470 13316 9643 2359
S6 Lot 3 11373 6652 16208 7161 13022 7311 15515 616 5164 10944 17899 1319 4964 9654 3315
S6 av 11677 7950 13324 4898 15014 6427 17571 686 6383 10944 17899 1894 9140 9648 2837
S7 Lot 1 14146 11823 16522 4324 19358 8324 21740 1021 11113 2977 15646 12129 3558
S7 Lot 3 12943 8191 20166 9261 15302 9848 17092 785 7775 12234 19028 1586 6151 11263 4357
S7 Lot 5 21666 12109 23495 10225 18867 15831 21942 782 7886 17578 23921 1398 4778 3487 4835
S7 av 16251 10708 20061 7937 17842 11334 20258 863 8925 14906 21475 1987 8858 8960 4250
MFI of DIRECT COMPARISON CONTROLS 9DCCs) d
G20 Lot 5 16 5 30 0 5 14 13 1 1 ‐19 ‐239 1893 4858 4 18
P10 Lot 5 8 8 31 6 4 17 13 2 ‐1 ‐22 ‐308 10 28 0 23
M50 Lot 5 5 7 34 2 16 18 18 1 ‐2 ‐17 ‐237 9 35 2 23
MACADAMIA MILK PEANUT PINE NUT PISTACHIO SOY WALNUT
33 34 35 36 37 38 39 42 43 44 45 46 47 48
AVERAGE BACKGROUND MFI
S0 Lot 1 207 139 203 231 471 128 206 182 171 250 476 182
S0 Lot 3 116 65 3176 2284 42 53 305 42 60 43 47 45 114 51
S0 Lot 5 131 67 126 293 134 101 511 78 89 59 106 75 411 128
AVERAGE INTENSITIES (MFI) CALIBRATION STANDARDS ABOVE BACKGROUND (minus average S0)
S1 Lot 1 5323 234 966 1197 897 109 1304 86 3181 2293 1344 887
S1 Lot 3 4770 219 1823 1118 1623 1141 662 72 2945 228 208 1662 528 254
S1 Lot 5 7968 317 797 655 1120 749 1042 116 4605 313 207 115 1620 1080
S1 av c 6020 257 1310 886 1236 1029 867 99 2951 209 1199 1357 1164 740
S2 Lot 1 8329 378 1736 2224 1186 158 2208 167 4867 3494 1958 1238
S2 Lot 3 7435 379 1887 1053 3100 2249 924 118 4969 420 368 2744 922 432
S2 Lot 5 11110 558 1602 1231 1908 1531 1383 185 8131 556 347 191 2939 1934
S2 av 8958 438 1744 1142 2248 2001 1164 154 5103 381 1860 2143 1939 1201
S3 Lot 1 11147 727 3204 4201 1520 245 3441 283 8594 6030 3540 2469
S3 Lot 3 9375 706 2875 1669 5911 4328 1220 180 7885 696 631 4457 1667 774
S3 av 10261 716 2875 1669 4558 4264 1370 213 5663 490 4613 5244 2603 1622
S4 Lot 1 13991 1284 5337 7187 1858 353 5067 491 12159 8593 5710 4536
S4 Lot 3 11269 1195 4350 2807 10396 7455 1572 282 11022 1182 1043 6938 2841 1287
S4 av 12630 1239 4350 2807 7867 7321 1715 318 8044 837 6601 7766 4276 2911
S5 Lot 1 16309 2542 8128 11365 2360 663 7537 1040 19470 12402 7903 6583
S5 Lot 3 12725 1821 5988 3839 16014 11412 1869 418 13060 1875 1668 10194 4702 1958
S5 Lot 5 18774 2907 8740 5726 11064 11398 2509 569 18861 2354 1611 824 13393 8914
S5 av 15936 2423 7364 4782 11735 11392 2246 550 13153 1756 7583 7807 8666 5818
S6 Lot 1 18673 3941 11163 15661 2725 950 10243 1633 23234 15995 10028 9543
S6 Lot 3 15277 3208 8178 5629 19745 14526 2174 612 14816 2927 2500 13766 6848 2976
S6 av 16975 3575 8178 5629 15454 15093 2449 781 12529 2280 12867 14881 8438 6260
S7 Lot 1 19713 5759 12928 19007 3203 1361 12210 2585 23884 17320 11630 11517
S7 Lot 3 16221 4234 9176 6395 21257 16561 2314 781 15875 4002 3396 16317 8534 4017
S7 Lot 5 22201 7068 12432 8866 20573 16723 2877 867 21306 4935 3400 1535 17805 13755
S7 av 19378 5687 10804 7630 18253 17430 2798 1003 16464 3841 10226 11724 12656 9763
MFI of DIRECT COMPARISON CONTROLS 9DCCs) d
G20 Lot 5 9 4 20 19 8 10 ‐48 3 5 2 5 5 32 7
P10 Lot 5 8 2 108 86 754 369 ‐47 5 4 4 18 4 24 6
M50 Lot 5 8 3 13255 8645 11 7 ‐45 2 13 0 4 4 28 4

a Intensities of the average MFI, after subtracting the average background (S0), of the calibration standards analyzed using three lots of reagents produced over a five year period and used within six months of production. Lot 1 ‐ first production run in 2014 (black font); Lot 3 ‐ production run used for SLV (blue font); Lot 5‐ production run used for MLV (green font). Lot 1 and Lot 3 data are the average of triplicate analyses, Lot 5 is the average of data collected in four experiments, each performed in triplicate. All data were generated by proficient analysts, Lots 3 and 5 by the same analyst.

b Target food allergen of antibodies conjugated to the specified bead sets. Lot 1 did not include in its repertoire beads sets 25, 26, 35, and 36.

c Averages (orange font) across the Lots are presented for informational purposes since variations between production runs may mathematically result in the appearance of inconsistencies, though each lot displayed proper increases in intensities with concentration.

d Direct comparison controls (DCCs) containing 20 ppm gluten (G20), 10 ppm peanut (P10), and 50 ppm NFDM (M50). DCC analyses were only performed using reagents from Lot 5 and the MFI are the averages of DCCs prepared by spiking the specified amount of analyte into analyte‐free food samples, each in triplicate. The meat, orange juice, and baked muffin samples were extracted using the PBST‐based Buffered‐detergent protocol followed by 10‐fold dilution with PBST. The dark chocolate samples were extracted using the UD Buffer Buffered‐detergent protocol followed by 5‐fold dilution with UD buffer. Due to the presence of NFDM in the UD Buffer, the M50 results from the dark chocolate samples were not included in the M50 average. The boxes indicate the bead sets expected to generate a positive response for the specific DCC.

Table 5. Lot‐to‐lot variation in signal / noise above background (S/N ‐ 1) of calibration standards over five yearsa.

Calib Std LOT ALMOND b BRAZIL NUT CASHEW COCONUT CRUST EGG GLUTEN HAZELNUT
12 13 14 15 18 19 20 21 22 25 26 27 28 29 30
Signal‐to‐Noise Above Bakground (/N‐1) c
S1 Lot 1 6 6 2 1 12 2 18 1 2 2 9 3 1
S1 Lot 3 63 20 24 16 25 14 23 2 9 13 2 3 4 11 5
S1 Lot 5 39 19 12 9 19 34 15 1 4 6 1 2 2 2 7
S1 av 36 15 13 9 19 17 19 1 5 9 1 2 5 5 4
S2 Lot 1 9 8 3 1 19 3 36 1 3 2 13 6 2
S2 Lot 3 106 34 45 29 40 25 44 3 15 22 4 5 6 19 8
S2 Lot 5 70 32 22 16 34 57 30 2 7 11 2 3 2 3 12
S2 av 62 25 24 15 31 29 37 2 8 16 3 3 7 9 7
S3 Lot 1 14 15 7 2 29 6 72 2 6 4 17 11 3
S3 Lot 3 163 57 79 51 64 42 80 5 26 38 6 7 9 32 15
S3 av 88 36 43 26 46 24 76 4 16 38 6 5 13 22 9
S4 Lot 1 19 23 12 4 41 9 114 3 11 5 25 20 6
S4 Lot 3 221 86 134 82 89 69 122 9 38 57 9 9 14 54 24
S4 av 120 54 73 43 65 39 118 6 25 57 9 7 19 37 15
S5 Lot 1 25 36 28 11 57 18 160 4 21 8 39 31 11
S5 Lot 3 277 123 201 128 118 103 168 12 60 79 12 12 19 78 36
S5 Lot 5 243 139 147 94 98 246 137 10 31 34 6 8 8 10 50
S5 av 182 99 125 78 91 122 155 9 37 57 9 9 22 40 32
S6 Lot 1 30 48 52 21 70 29 183 6 37 10 54 43 17
S6 Lot 3 335 162 279 186 148 146 201 17 92 108 16 16 28 110 57
S6 av 182 105 166 103 109 88 192 11 65 108 16 13 41 77 37
S7 Lot 1 36 61 83 34 80 43 203 8 54 12 64 55 25
S7 Lot 3 381 200 348 241 174 197 222 21 139 121 17 19 34 128 74
S7 Lot 5 333 228 291 200 146 444 179 16 75 44 7 14 16 17 100
S7 av 250 163 240 158 133 228 202 15 89 83 12 15 38 66 67
MFI of DIRECT COMPARISON CONTROLS (DCCs) CONTAIING 20 PPM GLUTEN, 10 PPM PEANUT, AND 50 PPM MILK d
G20 Lot 5 0 0 0 0 0 0 0 0 0 0 0 19 17 17 0 0
P10 Lot 5 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
M50 Lot 5 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
Calib Std LOT MACADAMIA MILK PEANUT PINE NUT PISTACHIO SOY WALNUT
33 34 35 36 37 38 39 42 43 44 45 46 47 48
Signal‐to‐Noise Above Bakground (/N‐1) c
S1 Lot 1 26 2 5 5 2 1 6 0 19 9 3 5
S1 Lot 3 41 3 1 0 39 22 2 2 49 5 4 37 5 5
S1 Lot 5 61 5 6 2 8 7 2 1 52 5 2 2 4 8
S1 av 43 3 3 1 17 11 2 1 36 4 8 16 4 6
S2 Lot 1 40 3 9 10 3 1 11 1 28 14 4 7
S2 Lot 3 64 6 1 0 74 42 3 3 83 10 8 61 8 8
S2 Lot 5 85 8 13 4 14 15 3 2 91 9 3 3 7 15
S2 av 63 6 7 2 32 22 3 2 62 7 13 26 6 10
S3 Lot 1 54 5 16 18 3 2 17 2 50 24 7 14
S3 Lot 3 81 11 1 1 141 82 4 4 131 16 13 99 15 15
S3 av 67 8 1 1 78 50 4 3 74 9 32 62 11 14
S4 Lot 1 68 9 26 31 4 3 25 3 71 34 12 25
S4 Lot 3 97 18 1 1 248 141 5 7 184 27 22 154 25 25
S4 av 82 14 1 1 137 86 5 5 104 15 47 94 19 25
S5 Lot 1 79 18 40 49 5 5 37 6 114 50 17 36
S5 Lot 3 110 28 2 2 381 215 6 10 218 44 35 227 41 38
S5 Lot 5 144 44 70 20 83 112 5 7 212 40 15 11 33 70
S5 av 111 30 36 11 168 126 5 7 155 30 55 96 30 48
S6 Lot 1 90 28 55 68 6 7 50 9 136 64 21 52
S6 Lot 3 132 49 3 2 470 274 7 15 247 68 53 306 60 58
S6 av 111 39 3 2 263 171 6 11 148 39 95 185 41 55
S7 Lot 1 95 42 64 82 7 11 59 14 140 69 24 63
S7 Lot 3 140 65 3 3 506 312 8 19 265 93 72 363 75 79
S7 Lot 5 170 106 99 30 154 165 6 11 239 83 32 20 43 107
S7 av 135 71 51 17 241 187 7 13 188 63 81 151 48 83
MFI of DIRECT COMPARISON CONTROLS (DCCs) CONTAIING 20 PPM GLUTEN, 10 PPM PEANUT, AND 50 PPM MILK d
G20 Lot 5 0 0 0 0 0 0 0 0 0 0 0 0 0 0
P10 Lot 5 0 0 0 0 6 4 0 0 0 0 0 0 0 0
M50 Lot 5 0 0 105 29 29 0 0 0 0 0 0 0 0 0 0

a The Signal / Noise above background (S/N‐1, background equaling S0) for the calibrations standards analyzed using three lots of reagents produced over a five year period and used within six months of production. Lot 1 ‐ first production run in 2014; Lot 3 ‐ production run used for SLV; Lot 5‐ production run used for MLV. Lot 1 and Lot 3 data are the average of triplicate analyses, Lot 5 is the average of data collected in four experiments, each performed in triplicate. All data were generated by proficient analysts, Lots 3 and 5 by the same analyst.

b Target food allergen of antibodies conjugated to the specified bead sets. Lot 1 did not include in its repertoire beads sets 25, 26, 35, and 36.

c S/N-1 values were rounded to the nearest integer. S/N‐1 < 1 highlighted in pink; 1 to 2 white; 2‐ 5 yellow; 5 ‐ 10 light green; >10 green. Data presented rounded off to the nearest whole number but color highlighting baed on values to the first decimal. Thus a S/N‐1 of 1.9 will appear as '2' but w/o any highlight while 2.1 would appear as a '2' with yellow highligh.

d Direct comparison controls (DCCs) containing 20 ppm gluten (G20), 10 ppm peanut (P10), and 50 ppm NFDM (M50). DCC analyses were only performed using reagents from Lot 5 and the MFI are the averages of DCCs prepared by spiking the specified amount of analyte into analyte‐free food samples, each in triplicate. The meat, orange juice, and baked muffin samples were extracted using the PBST‐based Buffered‐detergent protocol followed by 10‐fold dilution with PBST. The dark chocolate samples were extracted using the UD Buffer Buffered‐detergent protocol followed by 5‐fold dilution with UD buffer. Due to the presence of NFDM in the UD Buffer, the M50 results from the dark chocolate samples were not included in the M50 average. The boxes indicate the bead sets expected to generate a positive response for the specific DCC.

Lot-to-lot variances are not surprising considering the complexity of assembling the reagents. Thus, a more appropriate comparison of the performance by the various lots may be to examine the Signal-to-Noise ratios above background (S/N-1). This data is presented in Table 5 in which the S/N-1 values are tabulated, rounded off to whole numbers, and highlighted based on the S/N-1 values (to the first decimal); S/N-1 ratios < 1 in pink, between 1 and 2 no highlight, between 2 and 5 in yellow, between 5 and 10 in light green, and > 10 in green. As depicted, the data from all three lots displayed a strong robust, dynamic response with increasing analyte that is suitable for quantitative measurements and distinguishing between subtle changes in concentration, if such information is required. Further, should it become necessary to extend the dynamic range to lower concentrations of analyte, it is feasible for 24 of the bead sets, but probably not for five bead sets (coconut-21, egg-26, milk-36, pine nut -39 and -42) which consistently displayed S/N-1 values of 2 or less for the first calibration standard (S1).

Analysis of calibration standards by MLV participants

Figs 1 and 2 depict the average MFI intensities across the participating labs plotted versus the concentration of analyte in the calibration standards analyzed in a single experiment using PBST or UD Buffer, respectively, as the solvent milieu. The solid lines represent the first of the complementary bead sets (e.g., almond-12, Brazil nut-14, …) and the dashed lines represent the second (e.g., almond-13, Brazil nut-15, ….). The inserts in the coconut and pistachio plots expend the vertical scale to better illustrate the dynamic properties and MFI of the less intense bead sets (coconut-21 and pistachio-44). As previously observed, the use of UD Buffer did not significantly affect the performance properties of the bead sets with the target analytes. In all cases, the bead sets displayed excellent dynamic responses, with variances across the participating labs, represented by the error bars which equaled one standard deviation, that were acceptable. Fig 3 depicts the overall average MFI intensities across all 11 labs for the data averaged from four experiments (3 PBST and 1 UD buffer), typically performed on separate days, as a function of analyte concentration in the calibration standards. In the lower left corner, the average S7 MFI and associated %CV (RSDR) values are tabulated. In terms of MFI intensities, the RSDR values ranged from 10–54%; ten (34%) were less than 20% and twelve (41%) displayed RSDR values greater than 30%.

Fig 1. Calibration standards prepared in PBST.

Fig 1

Plots of the calibration curves for each bead set, grouped by complementary pairs, for the average across all participating labs from an individual experiment using PBST as the solvent to prepare the calibration standards. The average MFI intensities were plotted as a function of the concentration of analyte in the calibration standards. The first numerical bead set of a complementary pair (i.e., almond-12, Brazil nut-14, cashew-18, coconut-20, egg-25, gluten-27, hazelnut-29, macadamia-33, milk-35, peanut-37, pine nut-39, pistachio-43, soy-45, and walnut-47) represented by solid lines (_______) and the complementary bead sets (almond-13, Brazil nut-15, cashew-19, coconut-21, egg-26, gluten-28, hazelnut-30, macadamia-34, milk-36, peanut-38, pine nut-42, pistachio-44, soy-46, and walnut-48) by dashed lines (- - - -). The MFI intensities averaged across the 11 laboratories were the values after subtracting the background (S0) with the error bars representing one standard deviation.

Fig 2. Calibration standards prepared in UD buffer.

Fig 2

Similar to Fig 1 except UD buffer used as the solvent.

Fig 3. Average of the calibration standards from four experiments, three in PBST and one in UD buffer.

Fig 3

Similar to Fig 1 except the average of four sets of calibration standards (3 PBST and 1 UD buffer, n = 11 labs* x 4 exp x triplicate analyses) generated by each participant was used to generate the final data plotted. The lower right corner tabulates the overall average MFI of S7 for each bead set along with the %CV (RSDR) values across the 11 labs. *The average data from Lab 05 represented only three (2 PBST and 1 UD buffer) experiments, Lab 09 did not collect any data for the egg (-25, -26) or milk (-35, -36) bead sets and the gluten data generated by Lab 10 were classified as outliers (exceptionally high background resulted in background adjusted MFI <0 for S1 and S2) and not included.

A better demonstration of within lab-consistency is illustrated in S1 Fig. For each of the 11 laboratories the average MFI intensities (above background), across four experiments, each typically performed on separate days, are plotted along with the error bars representative of the standard deviations across the 12 samples (four sets of triplicates, 9 samples for Lab 05) for each of the bead sets, with complementary antibodies on the same graph. Lab 01 displayed virtually no variance between the replicate samples for all 29 analytes. The other laboratories displayed different degrees of variance with the vast majority comparable. Only three of the 163 graphs (315 titration curves, 29x10 + 25), were calibration curves generated with unacceptable levels of variance, all three for milk-35 and -36 (Labs 06, 07, and 08). Four titration plots for soy, two for gluten, and one for walnut (14 titration curves) displayed greater than desirable levels of variance which might impact the ability to quantitatively distinguish between fine differences in analyte concentration. Interestingly, in these cases both complementary antibodies (e.g., bead sets milk-35 and milk-36) were problematic, never a single antibody bead set out of a complementary pair. This problematic behavior with both complementary antibody bead sets may indicate a problem with the preparation of the detector antibody cocktail, but a problem in the handling and preparation of the bead cocktails cannot be ruled out. Altogether, 94% of all calibration curves generated by the 11 participating laboratories were suitable despite the differences in analyst proficiency.

A major strength of the xMAP FADA is the incorporation of built-in controls that make it possible to detect false positives, false negatives, and even apply a quantitative measure of such as may be useful as in the case of some products that display low levels of ‘non-specific’ binding [7, 10]. The use of such is dependent on the relative reliability of the performance by the various bead sets and how such may fluctuate with analyst proficiency. This is examined in Table 6 in which the inter-bead set variability for each participant as a function increasing calibration standards (S1, S2, S5, S7), is itemized for each of the four experiments. Specifically, tabulated in Table 6 are the average %CV (RSDr) values for the triplicate analyses for each calibration standard, averaged across all 29 bead sets with associated standard deviations for each experiment. In addition, the overall averages across all participants are presented in the lower corner (RSDR). As expected, with increasing concentration, the %CV values decreased. Further, for each participant the variances across all bead sets and associated standard deviations were acceptable. Of the 86 entries for calibration standards S5 and S7, only 20 displayed RSDr values ≥ 5% of which only five exceeded 10%. Labs 03, 04, and 07 had no RSDr values above 4%, and Labs 01 and 10 had only one entry with an RSDr of 6%, the remainder all ≤ 4%. Indeed, only one data entry displayed an RSDr > 20%, Lab 02, experiment 2, calibration standard S7. As such, the built-in controls based on monitoring the various bead sets in the assay displayed acceptable levels of reliability irrespective of the proficiency of the analyst. In contrast, inter-lab MFI variability exceeded 25% with %CV (RSDR) values of ≥ 50%. The inter-lab %CV (RSDR) values provides a gauge for possible use when establishing identical standards for analyst performance.

Table 6. Average coefficient of variation across bead setsa.

EXP 1 b EXP 2 EXP 3 EXP 4 EXP 1 EXP 2 EXP 3 EXP 4 EXP 1 EXP 2 EXP 3 EXP 4
LAB 1 c LAB 5 LAB 9
S1 5 ± 6 2 ± 2 2 ± 2 2 ± 1 S1 30 ± 72 47 ± 46 535 ± 2194 S1 10 ± 9 15 ± 6 11 ± 10 10 ± 8
S2 2 ± 2 2 ± 1 2 ± 1 2 ± 1 S2 14 ± 38 12 ± 114 27 ± 38 S2 9 ± 8 10 ± 4 5 ± 3 11 ± 7
S5 2 ± 2 2 ± 1 6 ± 4 2 ± 1 S5 7 ± 8 8 ± 21 11 ± 27 S5 6 ± 3 4 ± 2 4 ± 3 6 ± 3
S7 1 ± 1 1 ± 1 4 ± 3 1 ± 1 S7 7 ± 13 6 ± 15 7 ± 13 S7 12 ± 4 8 ± 4 8 ± 4 14 ± 5
LAB 2 LAB 6 LAB 10
S1 3 ± 3 3 ± 3 11 ± 13 5 ± 7 S1 8 ± 9 9 ± 99 4 ± 4 15 ± 36 S1 36 ± 155 7 ± 9 11 ± 31 26 ± 86
S2 2 ± 2 2 ± 1 2 ± 1 3 ± 2 S2 9 ± 10 0 ± 26 3 ± 4 12 ± 14 S2 2 ± 11 5 ± 7 5 ± 7 4 ± 5
S5 1 ± 1 2 ± 1 2 ± 1 1 ± 1 S5 5 ± 4 2 ± 23 2 ± 1 6 ± 6 S5 3 ± 8 2 ± 1 6 ± 4 2 ± 2
S7 2 ± 3 85 ± 5 1 ± 1 1 ± 0 S7 3 ± 2 18 ± 22 2 ± 1 4 ± 4 S7 2 ± 2 3 ± 1 1 ± 1 2 ± 1
LAB 3 LAB 7 LAB 11
S1 3 ± 4 16 ± 113 9 ± 11 6 ± 18 S1 6 ± 8 ‐31 ± 194 5 ± 9 5 ± 10 S1 7 ± 9 7 ± 11 6 ± 9 37 ± 388
S2 5 ± 4 3 ± 2 7 ± 5 4 ± 35 S2 5 ± 10 17 ± 23 2 ± 4 5 ± 6 S2 4 ± 3 5 ± 5 5 ± 6 33 ± 99
S5 2 ± 2 1 ± 1 2 ± 1 3 ± 4 S5 4 ± 7 3 ± 3 1 ± 2 3 ± 4 S5 1 ± 2 2 ± 1 2 ± 2 5 ± 15
S7 7 ± 3 1 ± 1 1 ± 1 3 ± 1 S7 3 ± 5 3 ± 2 0 ± 4 2 ± 1 S7 2 ± 1 1 ± 1 2 ± 1 6 ± 9
LAB 4 LAB 8 OVERALL d, e
S1 3 ± 4 10 ± 18 5 ± 6 9 ± 18 S1 7 ± 9 16 ± 46 9 ± 7 9 ± 295 S1 55 ± 44 45 ± 35 59 ± 41 77 ± 38
S2 2 ± 4 7 ± 8 3 ± 5 5 ± 6 S2 4 ± 9 12 ± 12 6 ± 9 4 ± 3 S2 48 ± 26 38 ± 19 50 ± 32 70 ± 33
S5 3 ± 2 3 ± 1 2 ± 1 4 ± 2 S5 6 ± 7 2 ± 17 3 ± 2 2 ± 1 S5 35 ± 14 33 ± 17 37 ± 13 50 ± 24
S7 1 ± 1 1 ± 1 1 ± 1 1 ± 1 S7 2 ± 1 3 ± 3 2 ± 3 4 ± 6 S7 27 ± 12 29 ± 15 30 ± 12 35 ± 19

a Average %CV across all bead sets, for each calibration standard ran alongside the food samples. Outlier data for bead sets 27 & 28 not included for Lab 10, neither were data for bead sets 25, 26, 35, 36 not obtained by Lab 09. The Exp 1 data was based on only 10 laboratories due to the inability to ship the meat samples to an international participant (Lab 05).

b Experiments 1, 2, 3, 4 refer to the analysis of Meat, Orange Juice, Baked Muffins, and Dark Chocolate samples, respectively.

C Lab specific averages of %CV (RSDr) values and associated standard deviations for each bead set, determined for each calibration standard and food, each ran in triplicate. Provides a measure of intra‐lab consistency of intensities.

d Overall %CV (RSDR) values across all 11 Laboratories: calculated from the average of the intensities for each bead set (minus background, S0) and averaged across the bead sets to determine an overall average and associated standard deviation to generate the Overall %CV (RSDR) values, for each calibration standard and food matrix. Provides a measure of inter‐lab consistency of intensities.

e Averages across Lab Specific %CV values for S1, S2, S5, S7 generated %CV of 13±17, 8±13, 6±10, & 6±8% for Exp 1; 11±18, 10±10, 5±9, & 13±24% for Exp 2; 15±18, 9±13, 6±10, & 5±8% for Exp 3; and 61±151, 15±20, 8±14, & 7±10% for Exp 4, respectively.

Another measure of inter-lab variability is to compare the signal-to-noise above background data relative to the average across all 11 participating laboratories. This data is presented in Table 7. Tabulated are the averages across four experiments for each participant, for each calibration standard and bead set, rounded off to the first decimal place. To facilitate comparisons, values < 0.80 were highlighted in pink, between 0.80 and 1.20 in yellow, and > 1.20 in green. As indicated, Lab 01 had only 8% of the entries below 0.8 (80%) of the study average with most associated with pine nut-39, soy-45, -46, and walnut-47 and the vast majority (96 out of 116) above 1.2 (120% of the average). In contrast, Labs 05, 09, 10, and 11 generated responses consistently below the average. Labs 02, 03, 04, 06, 07, and 08 displayed average or better S/N-1 values. This trend is consistent with the xMAP proficiency of the analysts with seven labs at or above average and 4 labs below average.

Table 7. Calibration standards compared based upon the signal / noise, relative to the overall average across all laboratoriesa, b.

ALMOND b BRAZIL NUT CASHEW COCONUT CRUST EGG GLUTEN HAZELNUT
12 13 14 15 18 19 20 21 22 25 26 27 28 29 30
LAB 01 / AV ALL LABS c
S1 3.7 1.9 2.7 3.0 2.3 2.4 1.6 1.7 2.6 7.8 6.2 5.9 5.9 1.9 1.9
S2 2.6 1.9 2.9 3.0 2.4 2.4 1.8 1.7 2.5 7.9 6.0 4.8 5.1 2.0 1.9
S5 2.4 1.9 2.8 3.0 2.0 2.3 1.5 1.6 2.1 7.4 4.7 4.3 4.7 1.6 1.8
S7 2.1 1.7 2.2 2.4 1.9 2.1 1.3 1.5 1.9 6.9 3.9 3.9 4.2 1.5 1.7
LAB 02 / AV ALL LABS
S1 1.9 1.1 0.8 0.8 0.8 0.8 0.9 0.8 0.8 0.2 0.3 0.9 0.7 0.7 0.9
S2 1.2 1.1 0.9 0.8 0.8 0.8 0.9 0.8 0.8 0.2 0.3 1.0 1.0 0.7 0.9
S5 1.2 1.1 0.9 0.8 0.8 0.8 0.9 0.8 0.9 0.2 0.5 1.1 1.0 0.8 0.9
S7 1.3 1.2 0.9 0.9 0.8 0.8 0.9 0.9 0.9 0.3 0.5 1.2 1.1 0.8 0.9
LAB 03 / AV ALL LABS
S1 2.4 1.5 1.2 1.1 1.3 1.0 1.7 1.6 0.8 0.4 0.7 0.3 0.5 1.3 1.1
S2 1.6 1.5 1.2 1.1 1.2 0.9 1.7 1.5 0.9 0.4 0.8 0.4 0.6 1.3 1.1
S5 1.3 1.3 1.2 1.0 1.0 0.8 1.2 1.3 1.1 0.4 0.8 0.4 0.7 1.0 1.0
S7 1.1 1.1 1.1 0.9 0.9 0.7 1.0 1.2 1.3 0.4 0.8 0.5 0.7 0.9 0.8
LAB 04 / AV ALL LABS
S1 1.5 1.0 0.8 0.7 0.9 0.8 1.0 0.9 0.5 0.3 0.5 0.2 0.3 1.0 0.8
S2 1.0 1.0 0.8 0.7 0.9 0.8 1.0 0.9 0.5 0.3 0.4 0.2 0.3 1.0 0.8
S5 1.0 1.0 0.8 0.7 0.9 0.8 1.0 0.9 0.7 0.4 0.6 0.3 0.4 1.0 0.8
S7 1.0 1.0 0.9 0.8 0.9 0.8 0.9 1.0 0.9 0.4 0.6 0.3 0.4 0.9 0.8
LAB 05 / AV ALL LABS
S1 0.1 0.2 0.7 0.1 0.4 0.3 0.5 0.3 0.5 0.2 0.4 0.1 0.3 0.5 0.2
S2 0.1 0.2 0.3 0.2 0.2 0.1 0.5 0.4 0.6 0.3 0.4 0.1 0.1 0.5 0.2
S5 0.1 0.2 0.2 0.2 0.2 0.1 0.6 0.5 0.7 0.4 0.8 0.1 0.1 0.6 0.2
S7 0.1 0.2 0.3 0.2 0.2 0.1 0.7 0.5 0.7 0.6 1.1 0.1 0.1 0.6 0.2
LAB 06 / AV ALL LABS
S1 2.1 1.3 0.8 0.9 1.0 1.1 1.1 1.4 1.3 0.4 1.2 2.3 1.5 1.2 1.0
S2 1.4 1.3 0.9 0.9 1.1 1.1 1.1 1.3 1.3 0.3 0.8 2.4 1.7 1.2 1.0
S5 1.5 1.3 0.8 0.8 1.1 1.1 1.2 1.3 1.2 0.3 0.6 2.4 1.5 1.2 1.0
S7 1.7 1.5 1.0 0.9 1.2 1.2 1.3 1.3 1.1 0.3 0.7 2.4 1.4 1.2 1.1
LAB 07 / AV ALL LABS
S1 1.1 0.9 1.2 1.2 1.2 1.1 1.4 1.3 1.9 0.2 0.5 0.5 0.7 1.2 1.2
S2 0.7 0.8 1.2 1.2 1.3 1.1 1.4 1.3 1.9 0.2 0.6 0.6 0.8 1.2 1.2
S5 0.8 0.9 1.4 1.2 1.4 1.2 1.5 1.2 1.8 0.3 0.6 0.6 0.8 1.3 1.2
S7 0.8 0.9 1.5 1.3 1.3 1.2 1.5 1.2 1.6 0.3 0.6 0.6 0.9 1.3 1.2
LAB 08 / AV ALL LABS
S1 0.6 0.6 1.0 1.4 1.0 1.5 0.9 0.9 0.8 0.2 0.3 1.1 0.7 0.9 1.5
S2 0.4 0.6 1.1 1.4 1.1 1.6 0.8 1.0 0.8 0.1 0.2 1.3 0.8 0.9 1.6
S5 0.5 0.7 1.1 1.5 1.2 1.7 1.2 1.7 1.0 1.1 0.9 0.3 0.6 1.4 1.0 1.1 1.7
S7 0.6 0.8 1.2 1.7 1.2 1.7 1.7 1.2 1.7 1.0 1.1 1.0 0.4 0.7 1.4 1.2 1.2 1.8
LAB 09 / AV ALL LABS
S1 0.7 0.7 0.4 0.5 0.6 0.7 0.5 0.7 0.6 na c 0.2 0.3 0.9 0.7
S2 0.4 0.7 0.4 0.4 0.6 0.7 0.4 0.6 0.5 0.1 0.3 0.9 0.6
S5 0.5 0.7 0.4 0.4 0.8 0.7 0.6 0.7 0.5 0.2 0.4 1.0 0.7
S7 0.7 0.9 0.5 0.5 1.0 0.9 0.9 0.8 0.5 0.2 0.4 1.1 0.7
LAB 10 / AV ALL LABS
S1 1.1 0.9 0.5 0.5 0.6 0.6 0.7 0.6 0.5 0.0 ‐0.4 ‐0.4 ‐0.2 0.5 0.6
S2 0.8 0.9 0.6 0.6 0.7 0.6 0.7 0.7 0.6 0.1 0.1 ‐0.1 0.0 0.7 0.7
S5 0.7 0.9 0.6 0.6 0.7 0.6 0.7 0.7 0.6 0.1 0.3 0.0 0.1 0.7 0.7
S7 0.7 0.9 0.6 0.6 0.6 0.5 0.7 0.7 0.6 0.2 0.4 0.0 0.1 0.7 0.7
LAB 11 / AV ALL LABS
S1 1.3 0.9 0.8 0.8 0.9 0.9 0.7 0.8 0.6 0.2 0.5 0.1 0.2 0.8 1.0
S2 0.8 0.9 0.8 0.8 0.8 0.8 0.7 0.8 0.5 0.2 0.4 0.1 0.3 0.7 0.9
S5 0.9 1.0 0.9 0.8 0.9 0.9 0.8 0.8 0.5 0.2 0.5 0.1 0.4 0.8 1.0
S7 0.9 1.0 0.9 0.8 0.9 0.9 0.8 0.8 0.5 0.3 0.6 0.2 0.4 0.8 1.0
AVERAGE S/N‐1 ACROSS LABS d
S1 16 10 4 3 8 14 9 1 2 1 0 0 0 1 4
S2 27 17 8 5 14 24 17 1 3 1 0 1 0 2 6
S5 103 72 52 31 48 106 89 6 15 5 1 2 2 6 27
S7 161 137 135 83 77 212 139 11 40 6 2 4 4 11 58
STDEV OF S/N‐1 DATA ACROSS ALL LABS d
S1 10.3 4.4 2.8 2.3 4.2 7.9 3.8 0.4 1.1 1.8 0.3 0.7 0.5 0.4 1.6
S2 18.6 7.7 5.4 4.1 7.8 14.2 7.8 0.6 1.8 3.3 0.5 0.9 0.6 0.6 2.8
S5 62.7 31.4 36.4 23.7 22.3 62.5 29.9 2.0 7.7 10.4 1.6 2.6 2.3 1.8 12.3
S7 88.9 52.5 69.1 51.3 33.4 115.9 36.8 3.2 18.1 13.3 1.7 4.3 4.5 3.2 26.8
Bkgd e 65 53 81 51 130 36 122 48 105 398 3634 102 292 207 48
SD 357 71 31 18 150 71 41 17 41 2192 3571 470 954 71 20
MACADAMIA MILK PEANUT PINE NUT PISTACHIO SOY WALNUT
33 34 35 36 37 38 39 42 43 44 45 46 47 48
LAB 01 / AV ALL LABS c
S1 1.7 1.4 3.0 1.8 2.1 1.9 0.9 1.4 2.1 1.8 0.7 0.9 0.7 1.1
S2 1.5 1.4 3.3 2.0 2.0 1.9 0.9 1.4 2.0 2.1 0.6 0.9 0.7 1.1
S5 1.2 1.4 3.6 2.1 2.1 1.9 0.8 1.3 1.5 2.1 0.5 0.7 0.9 1.2
S7 1.1 1.5 3.1 1.9 1.7 1.5 0.7 1.1 1.4 2.1 0.5 0.6 0.9 1.2
LAB 02 / AV ALL LABS
S1 1.2 1.0 1.5 1.5 1.2 1.0 1.0 1.0 0.9 0.7 1.4 1.2 0.9 1.0
S2 1.2 1.0 1.7 1.8 1.2 0.9 1.0 1.0 0.9 0.8 1.4 1.1 1.0 1.0
S5 1.2 1.0 2.0 2.3 1.2 1.0 1.0 1.0 0.9 0.8 1.3 1.0 0.9 1.0
S7 1.1 1.1 2.0 2.2 1.2 1.0 1.1 1.1 0.9 0.8 1.2 1.1 0.8 0.9
LAB 03 / AV ALL LABS
S1 1.4 1.3 1.1 1.5 1.4 1.6 1.4 1.4 1.5 0.8 1.7 1.6 1.4 1.8
S2 1.2 1.3 1.2 1.7 1.4 1.6 1.3 1.3 1.4 0.9 1.7 1.7 1.8 2.3
S5 1.0 1.1 1.0 1.6 1.5 1.2 1.2 1.2 0.9 0.7 1.7 1.9 0.8 1.3
S7 1.0 1.0 0.9 1.3 1.1 0.9 1.1 1.1 0.9 0.6 1.2 1.6 0.7 1.0
LAB 04 / AV ALL LABS
S1 0.9 1.0 1.1 1.1 0.9 1.1 1.2 1.0 0.9 0.6 1.0 1.1 0.9 1.1
S2 0.9 1.0 1.2 1.2 0.9 1.1 1.2 0.9 0.9 0.7 1.0 1.1 0.8 1.1
S5 0.8 1.0 1.4 1.4 0.8 1.1 1.1 1.0 1.0 0.7 1.1 1.1 0.8 1.1
S7 0.9 0.9 1.4 1.4 0.8 1.0 1.1 1.0 1.0 0.7 1.1 1.2 0.7 1.0
LAB 05 / AV ALL LABS
S1 0.6 0.5 0.0 ‐0.1 0.0 0.2 0.1 0.2 0.3 1.7 0.6 0.4 0.7 0.5
S2 0.6 0.5 0.0 0.1 0.2 0.4 0.1 0.3 0.2 0.3 0.6 0.4 0.6 0.5
S5 0.7 0.5 0.0 0.0 0.1 0.3 0.2 0.4 0.2 0.3 0.7 0.5 0.8 0.6
S7 0.7 0.6 0.0 0.0 0.2 0.3 0.2 0.4 0.2 0.3 0.9 0.6 0.8 0.6
LAB 06 / AV ALL LABS
S1 1.1 1.3 0.7 1.2 1.3 1.2 1.2 1.6 1.2 1.1 1.8 1.5 3.6 1.0
S2 1.2 1.3 1.0 2.0 1.3 1.2 1.2 1.6 1.3 1.3 1.8 1.5 3.3 0.9
S5 1.3 1.3 0.5 1.2 1.2 1.2 1.2 1.6 1.5 1.2 1.7 1.5 3.7 1.0
S7 1.4 1.3 0.9 1.8 1.6 1.5 1.2 1.7 1.6 1.2 1.9 1.5 3.8 1.1
LAB 07 / AV ALL LABS
S1 1.3 1.0 0.0 0.0 0.9 1.2 1.4 1.2 1.4 1.1 0.7 0.7 0.6 0.9
S2 1.4 1.0 0.0 0.0 0.9 1.2 1.5 1.2 1.5 1.2 0.6 0.7 0.6 0.9
S5 1.3 1.1 0.0 0.0 1.0 1.4 1.5 1.1 1.5 1.2 0.5 0.6 0.8 1.1
S7 1.3 1.1 0.0 0.0 1.1 1.4 1.4 1.1 1.5 1.2 0.5 0.6 0.9 1.2
LAB 08 / AV ALL LABS
S1 0.9 1.3 ‐0.1 ‐0.2 1.4 0.8 1.3 1.2 0.8 1.3 0.9 1.4 0.7 1.2
S2 1.0 1.3 ‐0.1 ‐0.2 1.3 0.8 1.1 1.2 0.9 1.4 0.9 1.4 0.7 1.1
S5 1.1 1.4 0.0 0.0 1.3 1.0 1.3 1.2 1.1 1.6 1.0 1.4 0.8 1.4
S7 1.1 1.5 0.0 0.1 1.5 1.1 1.4 1.3 1.1 1.7 1.0 1.5 0.8 1.4
LAB 09 / AV ALL LABS
S1 0.4 0.7 na 0.5 0.5 1.0 0.8 0.5 0.6 0.6 0.5 0.3 0.5
S2 0.5 0.7 0.5 0.4 1.0 0.7 0.5 0.7 0.6 0.6 0.3 0.4
S5 0.7 0.6 0.4 0.5 1.1 0.7 0.9 0.8 0.6 0.6 0.5 0.6
S7 0.7 0.6 0.5 0.8 1.1 0.8 1.1 0.9 0.7 0.7 0.7 0.8
LAB 10 / AV ALL LABS
S1 0.7 0.6 1.8 2.3 0.7 0.7 0.6 0.6 0.7 0.4 0.6 0.7 0.5 0.8
S2 0.7 0.7 0.6 0.7 0.6 0.7 0.6 0.8 0.7 0.5 0.7 0.7 0.5 0.8
S5 0.8 0.7 0.2 0.2 0.6 0.7 0.6 0.8 0.7 0.5 0.7 0.7 0.5 0.8
S7 0.8 0.6 0.2 0.2 0.7 0.7 0.6 0.8 0.7 0.5 0.8 0.8 0.5 0.7
LAB 11 / AV ALL LABS
S1 0.9 0.9 1.0 0.7 0.6 0.8 1.0 0.6 0.8 0.8 1.1 0.9 0.6 1.0
S2 0.9 0.9 1.1 0.8 0.6 0.7 1.0 0.6 0.8 0.9 1.1 0.9 0.6 1.0
S5 0.9 0.8 1.3 1.1 0.7 0.8 1.0 0.7 0.9 1.0 1.2 0.9 0.6 1.0
S7 0.9 0.8 1.5 1.1 0.8 0.8 1.1 0.7 0.9 1.0 1.1 0.9 0.5 1.0
AVERAGE S/N‐1 ACROSS LABS d
S1 36 3 2 1 4 4 2 1 24 3 3 2 5 7
S2 57 6 4 2 7 8 3 2 45 4 5 3 10 14
S5 117 30 19 9 39 58 6 6 137 19 30 15 37 56
S7 150 70 32 16 93 109 8 10 172 40 63 32 50 89
STDEV OF S/N‐1 DATA ACROSS ALL LABS d
S1 13.8 1.1 2.0 1.0 2.2 1.9 0.9 0.5 12.7 1.3 1.3 0.7 4.7 2.7
S2 18.1 1.8 3.9 1.8 3.6 3.7 1.1 0.7 23.2 2.2 2.4 1.2 8.4 6.9
S5 28.1 9.0 22.2 8.0 21.7 26.6 2.1 2.0 55.2 9.7 12.9 6.8 33.4 14.5
S7 35.7 23.1 32.7 13.6 42.9 39.4 2.7 3.6 66.9 21.2 24.8 12.5 47.2 20.4
Bkgd e 131 67 126 293 134 101 511 78 89 59 106 75 411 128
SD 31 23 3135 2600 215 75 298 34 115 24 33 38 154 36

a Comparisons between Signal /Noise above background (S0, S/N ‐ 1) by each laboratory versus the average S/N‐1 calculated across all 11 participating laboratories. The S/N‐1 ratios were calculated for each laboratory based on the average MFI (above background) for each bead set, for calibration standards S1, S2, S5, S7 as observed in four separate experiments (three for Lab 05); S3, S4, S6 were not included in the experiments.

b Comparison ratios < 0.80 highlighted in pink; 0.80 < ratio < 1.20 highlighted in yellow; ratios > 1.20 highlighted in green. Ratios displayed rounded‐off to the first decimal but highlight colors based on the values to the second decimal place; 0.79 and 0.81 would both be displayed as 0.8 but highlighted differently.

c Ratio between the S/N‐1 by the specified lab versus the overall average S/N‐1 calculated across all labs. 'na' refers to data not available.

d Average S/N‐1 across all labs rounded‐off to whole units. Not included in the overall averages are Gluten ‐27 & ‐28 data by Lab 10 (outlier); Egg ‐25 &‐26 and Milk‐35 & ‐36 data by Lab 09 not available. Lab 05 data based on only 3 experiments.

e Average background MFI across all 11 participating laboratories, each an average of the data gathered from four (three for Lab 05) experiments, with each sample analyzed in triplicate. Standard deviation (SD) calculated across the results by each laboratory.

To better evaluate the quality of the data generated by the various participants and its analytical utility entailed comparing the signal-to-noise, above background (S/N-1) across the dynamic range as defined by the calibration standards S1 and S7 for each bead set. This data is presented in Table 8. Included in the table are also the results associated with Lots 1 and 3 along with the average across the participating labs, sans outliers indicated by either red or pink highlight (Lab 10 egg-25, -26, gluten -27, -28; Lab 05 milk -35, -36, peanut -37, -38; and Lab 08 milk -35, -36). Otherwise, 18 of the 29 bead sets displayed ≥ 10-fold increases across the dynamic range, with three (Brazil nut-15, macadamia-34, and peanut-38) displaying > 20-fold increases. Only macadamia-33 and pine nut-39 displayed ratios of 3; though not large, still adequate for quantitatively distinguishing changes in analyte concentration.

Table 8. Comparison signal / noise, above background, spanning the calibration standards [(S/N‐1)S7 / (S/N‐1)S1]a.

LOT b Analyst c ALMOND d BRAZIL NUT CASHEW COCONUT CRUST EGG GLUTEN HAZELNUT
12 13 14 15 18 19 20 21 22 25 26 27 28 29 30
Lot 1 na 6 10 41 51 7 23 11 13 30 na na 7 7 17 26
Lot 3 na 6 10 15 15 7 14 10 11 16 9 9 6 8 12 15
Lot 5 adj av 11 15 32 30 9 15 17 14 28 12 16 11 15 12 17
stdev 3 3 8 5 2 4 5 4 8 4 9 3 5 2 2
%CV 24 17 24 18 26 26 32 27 28 36 52 26 36 17 14
Lot 5 Lab 01 e 9 12 24 22 8 13 12 11 18 8 7 6 8 9 15
Lot 5 Lab 02 11 15 31 30 10 16 14 16 28 14 19 12 18 12 16
Lot 5 Lab 03 8 11 28 24 7 12 9 9 40 8 13 14 18 8 11
Lot 5 Lab 04 10 14 33 31 9 15 15 15 42 11 15 14 18 11 17
Lot 5 Lab 05 8 12 13 42 5 5 20 24 33 20 34 8 4 13 20
Lot 5 Lab 06 13 16 35 27 11 17 18 13 21 7 7 9 11 12 17
Lot 5 Lab 07 11 15 39 30 10 16 16 12 22 12 14 12 15 13 17
Lot 5 Lab 08 14 19 36 34 11 17 19 15 34 16 25 12 19 15 20
Lot 5 Lab 09 16 17 37 29 14 20 30 14 23 na na 14 14 13 18
Lot 5 Lab 10 10 14 39 34 9 15 14 15 28 80 -10 -1 -5 15 18
Lot 5 Lab 11 11 14 32 28 10 15 16 13 24 11 14 14 21 12 17
LOT Analyst MACADAMIA MILK PEANUT PINE NUT PISTACHIO SOY WALNUT
33 34 35 36 37 38 39 42 43 44 45 46 47 48
Lot 1 na 4 25 na na 13 16 4 12 9 30 8 8 9 13
Lot 3 na 3 19 5 6 13 15 3 11 5 18 16 10 16 16
Lot 5 adj av 4 21 14 13 23 30 4 10 8 15 22 21 10 13
stdev 1 3 7 7 4 8 1 3 3 5 5 6 4 3
%CV 29 17 53 51 16 27 35 33 43 33 22 27 40 26
Lot 5 Lab 01 3 22 16 14 18 22 3 7 5 16 16 13 11 13
Lot 5 Lab 02 4 21 19 19 24 28 4 11 7 15 19 18 8 11
Lot 5 Lab 03 3 15 12 11 17 16 3 7 4 10 15 19 5 7
Lot 5 Lab 04 4 18 19 17 22 27 3 9 7 17 23 21 8 11
Lot 5 Lab 05 5 27 -412 -5 106 55 7 20 5 2 30 34 11 14
Lot 5 Lab 06 5 21 20 20 27 35 3 10 9 16 23 20 10 13
Lot 5 Lab 07 4 23 5 6 28 33 3 8 8 15 16 16 13 16
Lot 5 Lab 08 5 23 ‐4 ‐4 24 37 4 10 9 19 23 21 10 14
Lot 5 Lab 09 8 18 na na 23 46 3 9 16 20 27 25 20 20
Lot 5 Lab 10 4 21 1 1 23 29 3 12 7 17 27 23 9 11
Lot 5 Lab 11 4 17 21 20 27 30 3 10 8 17 21 19 8 12

a Ratio between the Signal/Noise (above background, S/N‐1) spanning the dynamic range of the calibration standards from S1 to S7 [(S/N‐1)S7 / (S/N-1)S1].

b Lot production number, Lot 1 first, Lot 3 used for single lab validation, Lot 5 used for multi‐laboratory validation (MLV). All analyses within 6 months of production.

c Analyst information: 'na' not applicable, 'adj av'‐ adjusted average is the average across the laboratories participating in the MLV (Labs 01 ‐ 11, below) omitting outlier data indicative of operational problems (highlighted in red and pink). 'stdev' and %CV' are the standard deviations and percent coefficient of variation (%CV) associated with the data used to calculate the adjusted averages. Lab 01 ‐ 11 refer to the participating laboratories in the MLV.

d Target analyte and associated bead set number.

e MLV laboratory (Labs 01 ‐ 11) data are the average across 4 experiments (3 for Lab 05) each triplicate analyses of the calibration standards. Outlier data highlighted in orange (negative indicative of a serious error) or pink (indicative of exceptionally high values). Exceptionally low ratios (< 2) highlighted in yellow, but included in the adjusted average. 'na' refers to not applicable, Lab 09 did not collect data for bead sets ‐25, ‐26, ‐35, and ‐36.

Overall, all participants generated calibration curves suitable for analyte quantitation. These results with the calibration standards indicate an acceptable level of proficiency in preparing and handling the reagents, performing the xMAP assay (post sample extraction), and operating the instrumentation.

Food samples

The food samples provide a measure of the inter-laboratory reliability of the assay upon including additional steps of sample preparation and extraction. To better gauge this variance, four types of food samples were chosen to reflect different physical-chemical properties; specifically, sausage-hot dog meat (high in fat and chemical preservatives), orange juice (acidic and contains a reductant, ascorbic acid), baked muffins (heating in a moist environment that undergoes dehydration), and dark chocolate (high in polyphenols and alkaloids). The concentrations of analyte incurred into the foods were chosen to span the dynamic range of the assay and reflect the lower concentration range of interest when analyzing regulatory samples. The coded food samples were prepared by incurring 25 μL of a mixture of the allergenic foods per gram food to generate final concentrations of 0, 1, 2.5, 10, 25, and 100 ppm of each allergenic food in the food sample. By incorporating an optional dilution step, 10-fold for the PBST extracts and 5-fold for the UD buffer extracts (net dilution of the food 200-fold), the analyses served to both span the dynamic range of the assay and demonstrate the ability to analyze samples containing one tenth the incurred concentrations. The applicability of the MLV data to represent performance at these lower concentrations is partly based on the minimal change in matrix carry-over due to the extensive dilution with buffer during extraction.

The food samples provided to the participating labs were:

  • Ground pure beef sausage-hot dog containing in equal mass proportion Egg, Gluten, Milk, and Soy.

  • Orange Juice containing equal mass proportions of Almond, Milk, and Soy.

  • Baked Muffin incurred with equal mass proportions of Coconut, Egg, Gluten, Milk, and Walnut.

  • Dark Chocolate incurred in equal mass proportions with Hazelnut, Milk, and Peanut.

These samples were extracted using UD buffer (contains 2.5% m/v NFDM) and therefore not analyzed for milk. The first and most important feature of an analytical method is its ability to qualitatively detect the analyte. Since presumptive false positives undergo a secondary confirmation this is not a significant problem. Secondary confirmation of presumptive positives with the xMAP FADA is derived from concurrence between complementary antibody bead sets, the presence of antibody bead sets that do not recognize the target analyte (e.g., crustacean-22), ratio analysis, and multi-antibody profiling. Thus, the major concern is not false positives but false negatives. The xMAP FADA incorporates multiple controls to prevent false negatives due to technical aspects of the assay (e.g., the AssayChex bead sets). However, when working with food samples false negatives (and positives) are often due to variances in the data as may be associated with analyst proficiency and/or complexity of the food matrix.

The results of the MLV indicate that despite relatively high levels of variance between the absolute MFI measured by the participants in different laboratories, failures to correctly qualitatively detect the incurred analytes were rare and limited to the baked muffins. Specifically, two labs (05 and 10) generated poor response curves for gluten and the confirmatory responses with coconut-21 and walnut -48 were problematic for two labs each (10 & 11, 06 & 09, respectively). In all other cases the analytes were detected, though at times the intensities of the MFI were not ideal necessitating looking for positive increases with analyte concentration. Fig 4 depicts the overall average MFI responses across the 11 labs and the positive responses generated by three labs (Figs 57). The three labs (01, 03, and 11) were chosen as representatives of different levels of experience and proficiency. The three labs correctly detected the incurred analytes, with the only exception being the inability of Lab 11 to observe confirmatory responses with coconut -21. In addition, low positive responses for cashew (-18 and -19) in orange juice and dark chocolate were observed. Though the standard deviations based on MFI across all labs were large (Fig 4), as was observed with the calibration standards, the data generated within a lab displayed excellent reproducibility (Figs 57). The two weak data sets with the cashew bead sets (observed by many analysts) represents cross-reactivity as evidenced by the low intensity, bordering on background, and the ratio between the two bead sets. Specifically, in both instances, the relative magnitude of the MFI generated by the two complementary cashew bead sets changed from cashew-18 > cashew-19 observed with the calibration standards (black solid line > black dashed line) to cashew-18 ≤ cashew-19 for the orange juice and dark chocolate samples (colored solid line ≤ colored dashed line).

Fig 4. Average MFI across all 11 laboratories of bead sets generating MFI with incurred food samples.

Fig 4

Plotted are the average of concurring positive bead set analyses, each performed in triplicate across all 11 laboratories. The colored lines are the data derived from the food samples (red—meat, blue–orange juice, green–baked muffins, and purple–dark chocolate) and the black lines are the calibration standards ran concurrently. Solid lines indicate the lower numerical bead set (e.g., egg-25) and the dashed lines the higher (e.g., egg-26) of complementary pairs. Graphs i, ii, iii, iv depict the egg, gluten, milk, and soy bead set analyses of the meat samples; Graphs v, vi, vii, viii depict the almond, cashew, milk, and soy bead set analyses of the orange juice samples; Graphs ix, x, xi, xii, xvi depict the coconut, egg, gluten, milk, and walnut bead set analyses of the baked muffin samples; Graphs xiii, xiv, xv depict the hazelnut, peanut, and cashew bead set analyses of dark chocolate. The two cashew results are included despite the responses being ‘classical’ cross-reactivity with very low measurable MFI and the ratios inconsistent with cashew. The meat data entailed only the results from 10 laboratories.

Fig 5. Average MFI of bead sets generating MFI with incurred food samples by Lab 01.

Fig 5

Same as Fig 4 except the average of only the triplicate analyses generated by Lab 01.

Fig 7. Average MFI of bead sets generating MFI with incurred food samples by Lab 11.

Fig 7

Same as Fig 4 except the average of only the triplicate analyses generated by Lab 11.

Fig 6. Average MFI of bead sets generating MFI with incurred food samples by Lab 03.

Fig 6

Same as Fig 4 except the average of only the triplicate analyses generated by Lab 03.

Ratio analysis

A key feature of the xMAP FADA is the use of complementary antibody bead sets and requiring concurrence. Though this feature can be applied on a qualitative basis, a quantitative approach is superior and can be used to distinguish between cross-reactive homologues and the effects of different forms of processing. This quantitative approach is the basis of ratio analysis and multi-antibody profiling; profiling being the ratios between more than just the complementary antibody bead sets. In generating ratios and profiles, the more intense bead set is referred to as the anchor and becomes the denominator when calculating ratios and profiles. Focus on the higher avidity antibody reduces the likelihood of its undergoing saturation by assuring that its intensity is within the dynamic range of the assay. As such, any reduced precision is associated with the lower avidity antibody and the onset of its dynamic response. Thus, the possibility of underestimating allergen content due to the onset of saturation of the higher avidity antibody is reduced and instead the possibility of reduced precision (error) is shifted to the lower avidity antibody and it is not being in the dynamic range. As a result, the error is shifted to having less accuracy at low concentrations where the likelihood of a potential health risk is significantly reduced.

A key premise of ratio analysis is that a positive response that increases with analyte concentration indicates the presence of an antigenic analyte. The antigenic analyte may be the target analyte or a homologous, cross-reactive agent. It is unlikely that two different proteins would contain the same antigenic elements in identical environments (surrounding conformation of the protein) and thus the avidities would be different. As a result, the ratio between the MFI generated by complementary bead sets is unlikely to be the same for two different antigenic agents, provided the concentrations are not saturating. This approach provides a powerful analytical tool provided the target analyte and the reference material have undergone similar modifications (e.g., processing, covalent modification, conformational changes). Thus, ratio analysis provides a reliable secondary endpoint which is enhanced by also comparing the multi-antibody profiles generated with the other bead sets present in the cocktail. If the antigenic element was modified due to processing, it is important to compare the ratio and profile to appropriately, similarly processed standards. While this may complicate the use of such secondary endpoints, it is no different from what is observed with the commonly used ELISAs, with the advantage that it is less likely to generate false negative results due to the built-in redundancy. Further, as research characterizing the effects of processing on antigenicity increases [7], it should ultimately be possible to use these secondary endpoints to determine the type and extent of food processing. This is a potentially important since it is known that various forms of food processing affect allergenicity, and until a biological activity assay is developed, such may provide a useful tool.

Ratio analysis and multi-antibody profiling have been successfully applied to the detection of undeclared food allergens (e.g., raw peanut in garlic) [8], distinguishing between peanut, soy, and other legumes that cross-react with the peanut and soy antibodies at high concentrations [7], detection of allergenic foods (e.g., pecan) without having specific target antibodies [9], and for detecting and distinguishing between botanicals from the same plant families [10].

The calibration standards used in the xMAP FADA are derived from raw reference materials. As such, the ratio analysis data generated with the incurred, processed food samples may differ from the ratios generated by the calibration samples. This difference does not adversely affect the purpose of the MLV, which focuses on the examination of inter-laboratory ability to generate reliable data by analysts of varying proficiency. Indeed, the data gathered from the MLV may add to the database being developed regarding the effects of food processing on complementary bead set ratios and multi-antibody profiles.

Table 9 compares the effects of changing analyte concentration on the average, across all participating labs, of the complementary antibody pair MFI ratios. Of particular interest, are the standard deviations and %CV (RSDR) values describing inter-lab variance. As expected, the variances improved with increasing concentration of analyte with %CV (RSDR) values ≤47% for S1, S2, S5, and S7 of 18%, 14%, and 12% of the entries, respectively. Indeed, 29% of the complementary bead sets for S7 displayed %CV values of ≤5% and the percentage of ratios with %CV (RSDR) values in excess of 20% decreased from 57% for S1 to 7% for S7.

Table 9. Ratios between complementary antibody bead sets by calibration standardsa.

Complementary ALMOND BRAZIL NUT CASSHEW COCONUT EGG c GLUTEN c HAZELNUT MACADAMIA MILK c PEANUT PINE NUT PISTACHIO SOY WALNUT
Ratios b 13/12 15/14 19/18 21/20 25/26 27/28 30/29 34/33 36/35 38/37 42/39 44/43 46/45 48/47
Ratio (brown second antibody/first)
AVERAGE S0 0.59 0.61 0.35 0.44 0.58 0.39 0.27 0.55 1.22 0.57 0.24 0.58 0.97 0.42
S1 0.39 0.37 0.46 0.04 0.59 d 0.37 0.83 0.06 0.94 0.77 0.10 0.08 0.46 0.60
S2 0.38 0.38 0.45 0.04 1.12 0.38 0.85 0.06 0.74 0.79 0.12 0.06 0.48 0.60
S5 0.43 0.34 0.56 0.03 0.82 0.33 0.98 0.14 0.66 1.1 0.21 0.08 0.44 0.64
S7 0.51 0.33 0.70 0.03 0.84 0.28 1.15 0.25 0.69 0.86 0.28 0.13 0.43 0.76
STDEV S0 0.15 0.08 0.08 0.06 0.17 0.10 0.06 0.08 0.47 0.20 0.08 0.14 0.21 0.16
S1 0.03 0.10 0.05 0.01 1.92 0.10 0.14 0.01 0.67 0.31 0.02 0.04 0.04 0.08
S2 0.03 0.06 0.05 0.01 0.33 0.10 0.12 0.01 0.07 0.11 0.02 0.02 0.04 0.09
S5 0.02 0.06 0.06 0.00 0.10 0.07 0.15 0.02 0.06 0.12 0.03 0.02 0.04 0.04
S7 0.03 0.05 0.06 0.00 0.08 0.05 0.18 0.04 0.03 0.07 0.04 0.04 0.02 0.02
%CV S0 26 13 23 14 29 27 21 14 39 35 33 24 21 39
S1 8 28 11 24 323. 27 17 24 71 40 17 52 8 14
S2 8 16 10 23 29 26 14 18 10 14 17 36 9 15
S5 5 18 12 16 13 21 16 16 9 11 14 30 8 7
S7 5 15 9 7 9 20 16 16 4 9 17 34 5 3

a Ratios between MFI by complementary antibody bead sets as averaged for 4 experiments (3 for Lab 05) and then across the 11 participating laboratories.

b Ratio is defined as the MFI by the bead set generating the smaller response divided by the MFI generated by the bead set generating the larger response for the reference material/calibration standards. Ratios in which the numerically first bead set is divided by the second are in black font (e.g., Gluten‐27/‐28), the reverse in brown (Almond‐13/‐12). No Ratio for Crustacean, only a single antibody ‐bead set (‐22).

c Did not include Lab 10 gluten (‐27 & ‐28) data; Lab 07 Milk (‐35 & ‐36) data. Lab 09 did not collect egg (‐25 & ‐26) or milk (‐35 & ‐36) data. Lab 05 only performed 3 experiments.

d Standard deviations, and %CV, for the average Ratios calculated for S1 of Egg and Milk high (red font); indicates unreliability in the Ratio for Egg and Milk at concentrations equivalent to S1.

Ratio analysis provides a stringent measure of analyte detection and inter-laboratory variance. Table 10 lists the ratios determined for the various analytes detected in the food samples. The ratios are only presented when the average responses (across all 11 labs) of the complementary antibody bead sets when both exceed the MFI of their respective S1 calibration standards, the lower limit of the calibration curves. The participants all correctly detected the presence of walnut in the baked muffins, though two displayed poor confirmatory results with walnut-48. However, the walnut data is not included in the table because the average (across all 11 labs) MFI for one of the walnut bead sets (-47 or -48) did not exceed S1 (equivalent to 11 ppm in the food sample, Table 1); within the dynamic range defined by the calibration standards. Using this stringent requirement, concurrence between complementary antibody bead sets within the dynamic range defined by S1 and S7, yielded highly reproducible limits of detection and quantification.

Table 10. Ratio analysis of complementary antibody pairsa, b, c, d.

ng/mL in analytical sample ALMOND 13/12 COCONUT 21/20 EGG 25/26 GLUTEN 27/28 HAZELNUT 30/29 MILK 36/35 PEANUT 38/37 SOY 46/45 WALNUT 48/47
av stdev av stdev av stdev av stdev av stdev av stdev av stdev av stdev av stdev
CALIBRATION STANDARDS
S1 0.39 ± 0.03 0.038 ± 0.009 0.59 ± 1.92 0.37 ± 0.10 0.83 ± 0.14 0.94 ± 0.67 0.77 ± 0.31 0.46 ± 0.04 0.60 ± 0.08
S2 0.38 ± 0.03 0.036 ± 0.009 1.12 ± 0.33 0.38 ± 0.10 0.85 ± 0.12 0.74 ± 0.07 0.79 ± 0.11 0.48 ± 0.04 0.60 ± 0.09
S5 0.43 ± 0.02 0.027 ± 0.004 0.82 ± 0.10 0.33 ± 0.07 0.98 ± 0.15 0.66 ± 0.06 1.10 ± 0.12 0.44 ± 0.04 0.64 ± 0.04
S7 0.51 ± 0.03 0.031 ± 0.002 0.84 ± 0.08 0.28 ± 0.05 1.15 ± 0.18 0.69 ± 0.03 0.86 ± 0.07 0.43 ± 0.02 0.76 ± 0.02
MEAT e 5 <dyn g 0.30 ± 0.10 <dyn
12.5 0.34 ± 0.08 <dyn
50 0.86 ± 0.11 0.33 ± 0.05 0.52 ± 0.29
125 0.79 ± 0.07 0.33 ± 0.04 0.46 ± 0.29 0.37 ± 0.06
500 0.84 ± 0.09 0.36 ± 0.05 0.58 ± 0.12 0.37 ± 0.05
ORANGEJUICE e
5 <dyn <dyn
12.5 <dyn <dyn
50 0.65 ± 0.12
125 0.10 ± 0.03 0.67 ± 0.08 0.49 ± 0.09
500 0.15 ± 0.04 0.67 ± 0.07 0.44 ± 0.05
BAKEDMUFFIN e
5 <dyn <dyn
12.5 <dyn
50 <dyn
125 0.67 ± 1.54 0.82 ± 0.27 0.77 ± 0.03 h
500 0.07 ± 0.05 0.68 ± 0.08 0.73 ± 0.16 0.83 ± 0.03
DARKCHOCOLATE f
5 <dyn n.a. i <dyn
12.5 <dyn
50 0.72 ± 0.11
125 0.81 ± 0.13
500 1.10 ± 0.2 0.74 ± 0.13

a Average (±stdev) Ratios between complementary antibody pairs calculated across the participating laboratories. The anchor bead sets (larger MFI response bead sets) were used in the denominator to calculate the ratios. The calibration standards entailed averaging each participant's results (MFI) across all four experiments.

b All data for which the Overall Average MFI, across all partiipants, exceeded the response generated by S1 with the appropriate bead set (lower limit of the quantitative dynamic range) for both complementary antibodies were used to calculate ratios.

c Due to import problems, Lab 05 did not receive meat samples. Lab 09 failed to program the monitoring of egg (25,26) and milk (35,36). Further, Lab 05 egg and Lab 07 milk data for muffins were dropped as outliers. Lastly, for the calibration standards, the gluten data generated by lab‐10 and the Milk data by Lab‐07 were omitted as outliers.

d Brackets represent analyte incurred in the food.

e Meat, Orange Juice, and Muffin samples contining 1, 2.5, 10, 25, or 100 ppm analyte were extracted using PBST Protocol which entailed a 20‐fold dilution with PBST. A secondary, optional 10‐fold dilution with PBST was applied, thus the results are representative of the analysis of 0.1, 0.25, 1, 2.5, and 10 ppm analyte if the optional 10‐fold dilution was omitted.

f Dark Chocolate samples containing 1, 2.5, 10, 25, or 100 μg/g analyte extracted using UD Buffer Protocol which entailed a 1:40 dilution. A secondary, optional 5‐fold dilution with UD Buffer was applied, thus the results represent the analysis of 0.2, 0.5, 2, 5, and 20 μg/g analyte if the optional 5‐fold dilution omitted. Since, UD Buffer contains milk, milk analysis was not applicable (n.a.).

g '< dyn' indicates that the concentration of the allergenic food in the analytical sample is insufficient to exceed. S1 and would thereby, even at 100% recovery, not met the criteria for inclusion.

h Ten of the 11 labs detected walnut. However, the overall average MFI, across all labs, for both bead sets ‐47 and ‐48 did not exceed the average S1 MFI for each bead set.

i 'n.a.' not applicable. The chocolate samples were analyzed using UD buffer (extraction and dilution) which contains NFDM and therfore the milk bead set responses are not listed.

The limits for the various analytes in the analytical samples of the diluted food extracts varied from 5 to 125 ng/mL for meat, 50 to 125 ng/mL for orange juice, 50 to 500 ng/mL for dark chocolate, and 125 to 500 ng/mL for baked muffins (except for walnut). If the optional 10-fold (5-fold for UD buffer) dilution step was omitted this means the limits are 20-times the concentration in the analytical sample; in all but one case ≤ 10 ppm in the original food samples.

Of interest is how the derived ratios of the incurred food samples compare to those of the calibration standards and the associated standard deviations. Not surprisingly, the ratios of the analytes mixed into the ground (uncooked) beef were comparable to the calibration standards. The ratios observed for milk and soy in orange juice were comparable to the calibration standards while almond showed a significant change from 0.4–0.5 to 0.1–0.15. The change in the almond ratio appears to be primarily due to a decrease in the MFI generated by almond bead set-13 (Figs 47). It is not clear whether this decrease in the almond-13 MFI was due to the acidic environment (e.g., deamidation), reducing conditions (e.g., ascorbic acid), or other factors that might affect the conformation of the target analyte or result in modifications of an antigenic epitope involved in almond-13 detection of the analyte. As expected, baking caused changes in the ratios for all analytes except for milk, for which the ratios were comparable. No major change was observed for either hazelnut or peanut in dark chocolate. Instead of looking at the standard deviations, a better representation of the variances is presented in Table 11.

Table 11. Percent coefficient of variation (%CV) of complementary antibody ratiosa.

ng/mL in analytical sample b Almond c Coconut Egg Gluten Hazelnut Milk Peanut Soy Walnut av
S1 8 24 323 27 17 71 40 8 14 59
CALIBRATION S2 8 23 29 26 14 10 14 9 15 16
STANDARDS S5 5 16 13 21 16 9 11 8 7 12
S7 5 7 9 20 16 4 9 5 3 9
MEAT d
5 33
12.5 24
50 13 16 55
125 9 13 63 15
500 11 14 21 15
ORANGEJUICE d
5
12.5
50 18
125 27 12 17
500 24 10 10
BAKEDMUFFIN d
5
12.5
50 f
125 230 32 4
500 66 12 21 4
DARKCHOCOLATEe
5
12.5
50 15 na
125 16
500 18 17

a %CV values derived from average MFI and associated standard deviations calculated across all 11 labs for the incurred food samples. Included are only those ratios derived from data in which the overall average MFI for both complementary bead sets exceeded the average MFI of S1 for each bead set. Brackets represent that which alergenic foods were incurred into the food samples.

b Concentration of allergenic food in the analytical sample following extraction and subsequent dilution, a net 200‐fold dilution of the original food sample. Ommision of optional dilution of extract provides 10X (5X for UD Buffer extracts) greater sensitivity.

c Ratios calcualted as the ratio of Almond ‐13/‐12; Coconut ‐21/‐20; Egg ‐25/‐26; Gluten ‐27/‐28; Hazelnut ‐30/‐29; Milk‐36/‐35; Peanut ‐38/‐37; Soy ‐46/‐45; Walnut ‐48/‐47.

d Meat, orange juice, and baked muffin samples prepared containing each analyte at either 0, 1, 2.5, 10, 25, or 100 ppm. These samples were diluted 20‐fold with PBST during extraction and diluted an (optional) aditional 10‐fold prior to analyis.

e Dark Chocolate samples prepared containing each analyte at either 0, 1, 2.5, 10, 25, or 100 ppm. These samples were diluted 40‐fold with UD Buffer during extraction and diluted an (optional) aditional 5‐fold prior to analyis.

f Ten of the 11 labs detected walnut. However, the overall average MFI, across all labs, forf both bead sets ‐47 and ‐48 did not exceed the average S1 MFI for each bead set.

Tabulated in Table 11 are the %CV (RSDR) values for the data presented in Table 10. The average %CV for the meat samples progressively decreases from 33% to 15% with increasing concentration. Similar patterns were observed for the other food samples with the %CV values not significantly changing for hazelnut in chocolate (approx. 16%) or milk in baked muffins which was 4% for both the 125 ng/mL and 500 ng/mL samples. The relatively low variances observed in the ratios, compared to the variances observed in the MFI, reflects how the ratios are based on the binding avidities to the complementary antibody bead sets, an inherent property of the analytes. Further, as the concentration of analyte increases, any contribution by noise becomes a smaller proportion, thereby improving the ratio’s reflection of the binding constants and not inter-laboratory performance.

Recovery

To gauge a true estimate of the effects of matrix on analyte recovery, the single lab validation measured the recoveries associated with analyte detection from samples prepared by incurring each of the 15 allergenic foods individually and as a mixture of all 15 into five matrices (four food samples and buffer) and extracted versus spiked into extracts prepared from the same five (analyte-free) matrices at comparable concentrations. This approach circumvented the need to rely on the calibration standards to calculate recovery and the variance that might be associated with any slight cross-reactivities that may be associated with using a mixture of the allergenic foods as calibration standards. Further, it distinguished between the effects of food extracts and the ability to mobilize the allergens from the food matrix. Thus, the goal of the recovery measurements of the MLV was not to ascertain the reliability of the xMAP FADA nor its performance when used exclusively by highly proficient analysts but instead the variance that might be observed by a range of analysts, of whom 9-out-of-11 were not proficient with the assay.

Tables 1214 tabulate the ppm of the allergenic foods detected in the various food samples by each laboratory. Quantitation was performed by a step-wise linear approach using two calibration points bracketing the measured MFI (minus background) of a given sample. The dynamic range applied to calculating the ppm of allergenic food detected was defined as the background (S0) plus 10-times the standard deviation (S0+10D) as the lower limit and the S7 calibration standard for the upper limit. The use of S0+10D, and not S1, was based on xMAP FADA S/N-1 data (Tables 5 and 8) and published data [11] that indicated that the lower end of the dynamic range could be extended. In conducting extrapolation analyses, there are potential disadvantages regarding measurement accuracy and possible distortions in any observed inter-laboratory variances. Average MFI responses by a bead set that were < S0+10D or > S7, were noted in the table as ‘under’ or ‘over’, respectively. Since, the food samples were designed to assess performance across the dynamic range of the xMAP FADA, ‘over’, ‘>‘, responses were expected for food samples incurred with 100 ug/g almond, coconut, egg, or hazelnut if the recovery was 100% and similarly for gluten, milk, and peanut if the recovery exceeded 50% (see Table 1). When performing food sample analyses, it is expected that the analyst might further dilute (or omit the optional dilution step) to optimize quantification, if such is desired.

Table 12. Interpolated analyte recovered from incurred food samples: Meat / Sausagea.

μg / gram food b Lowest Level c Lab number EGG d GLUTEN MILK SOY
‐25 ‐26 av % rec e ‐27 ‐28 av % rec ‐35 ‐36 av % rec ‐45 ‐46 av % rec
1 0.1 Lab01 0.79 0.82 81 0.9 1.0 93 0.6 0.6 57 0.9 0.7 84
1 0.1 Lab02 under under 1.7 2.2 192 0.3 0.4 37 0.3 under
1 0.1 Lab03 0.33 under 1.4 1.5 146 0.58 0.64 61 0.35 under
1 0.1 Lab04 under under 3.2 3.9 354 0.32 under 0.55 under
1 0.1 Lab05 f
1 0.1 Lab06 under under 4.2 4.5 433 4.6 2.3 347 2.9 2.0 241
1 0.1 Lab07 under under 1.3 1.2 128 under 0.9 under under
1 0.1 Lab08 under under 1.4 2.3 184 7.3 7.3 731 0.2 under
1 0.1 Lab09 f 1.4 1.6 152 under under
1 0.1 Lab10 2.8 3.8 326 8.4 3.3 584 under under 0.3 under
1 0.1 Lab11 under under 1.4 2.2 180 6.0 7.7 684 0.4 under
av g < h < 2.5 2.4 245 2.8 2.8 320 0.7 <
stdev g 2.3 1.2 159 3.1 3.2 322 0.9
%CV g 91 49 65 108 115 101 123
2.5 0.25 Lab01 2.0 2.1 82 4.1 4.0 160 1.4 1.3 54 2.2 1.8 80
2.5 0.25 Lab02 2.1 under 4.0 4.4 169 1.0 1.1 42 0.8 0.5 25
2.5 0.25 Lab03 0.9 under 3.5 3.2 134 0.7 0.8 31 0.5 0.4 19
2.5 0.25 Lab04 2.0 under 6.6 6.2 256 1.0 1.1 42 1.6 1.1 54
2.5 0.25 Lab05
2.5 0.25 Lab06 under under 11.3 11.0 448 OVER 21.9 0.9 0.6 30
2.5 0.25 Lab07 2.3 under 5.5 4.6 203 1.0 1.0 40 under under
2.5 0.25 Lab08 under under 5.5 5.4 219 8.0 8.0 320 1.4 2.3 74
2.5 0.25 Lab09 4.5 4.7 184 under under
2.5 0.25 Lab10 4.1 7.6 233 1.6 under under under 0.9 under
2.5 0.25 Lab11 under under 5.5 6.7 245 0.8 0.9 34 0.8 under
av 2.2 < 5.2 5.6 224 2.0 4.5 80 1.1 <
stdev 1.0 2.6 2.3 93 2.7 7.4 106 0.6
%CV 45 49 41 41 134 164 132 50
10 1 Lab01 11 12 115 30 23 261 5.7 5.0 53 11 8.2 95
10 1 Lab02 9 11 100 OVER OVER 4.6 3.8 42 3 2.3 29
10 1 Lab03 3 4 37 25 21 230 2.2 1.9 21 2 2.0 21
10 1 Lab04 9 11 100 OVER 30 4.0 3.8 39 6 3.9 49
10 1 Lab05
10 1 Lab06 12 12 117 OVER OVER 11.8 6.9 93 6 4.0 49
10 1 Lab07 10 10 98 OVER 22 1.1 1.0 11 3 2.8 30
10 1 Lab08 7 8 74 OVER 30 24 24 241 3 2.0 22
10 1 Lab09 30 24 267 3 under
10 1 Lab10 14 22 179 OVER 24 under under 4 4.1 39
10 1 Lab11 9 10 95 OVER 31 4.4 4.1 43 4 2.8 35
av 9 11 101 > h 25 7.2 6.4 68 4.5 3.6 41
stdev 3 5 38 4 7.4 7.5 74 2.5 1.9 23
%CV 32 43 37 16 103 118 109 56 53 55
25 2.5 Lab01 23 24 93 OVER OVER 13 8.1 42 27 22 98
25 2.5 Lab02 21 24 90 OVER OVER 8 6.9 31 9 6 31
25 2.5 Lab03 12 14 52 OVER OVER 6 4.3 20 6 5 21
25 2.5 Lab04 21 25 91 OVER OVER 8 6.7 29 13 9 46
25 2.5 Lab05
25 2.5 Lab06 28 31 119 OVER OVER OVER OVER 14 10 49
25 2.5 Lab07 20 21 83 OVER OVER under under 6 5 22
25 2.5 Lab08 16 16 63 OVER OVER OVER OVER 6 5 22
25 2.5 Lab09 OVER OVER 9 8 34
25 2.5 Lab10 36 62 197 OVER OVER under under 10 8 37
25 2.5 Lab11 21 24 90 OVER OVER 11 7.7 10 7 34
av 22 27 98 > > na h na 11 9 39
stdev 7 14 42 6 5 23
%CV 32 53 43 56 60 57
100 10 Lab01 OVER OVER OVER OVER OVER 26 171 144 158
100 10 Lab02 82 OVER OVER OVER 23 17 20 34 26 30
100 10 Lab03 48 65 56 OVER OVER 16 9 13 20 17 19
100 10 Lab04 81 OVER OVER OVER 25 21 23 50 37 44
100 10 Lab05
100 10 Lab06 OVER OVER OVER OVER OVER OVER 65 48 56
100 10 Lab07 OVER OVER OVER OVER 6 1 3 29 24 26
100 10 Lab08 56 66 61 OVER OVER OVER OVER 23 20 22
100 10 Lab09 OVER OVER 37 34 35
100 10 Lab10 OVER OVER OVER OVER 16 6 11 41 35 38
100 10 Lab11 79 OVER OVER OVER OVER 27 36 28 32
av > > > > 17 15 14 51 41 46
stdev 8 10 8 44 37 41
%CV 45 67 56 88 90 89

a Interpolated ppm using the calibration standards (S1, S2, S5, S7) in a stepwise, linear analysis of the measured MFI (after subtracting background). S7 defines the upper limit of the dynamic range and 10‐times the standard deviation (10D) of the background (S0) defines the lower limit of quantitation. ppm calculated by converting the interpolated ng extractable protein/mL derived from the calibration standards, to ppm whole allergenic food using the protein content as referenced in Table 1. 'under' and 'OVER' indicate that the measured MFI was either less than or greater than the defined limits of the dynamic range (10D and S7).

b μg incurred per gram of food to generate the test portion.

c 'Lowest Level' refers to the lowest concentration of analyte that would generate the same analytical sample; by eliminating the optional dilution following extraction.

d Incurred analyte and associated bead sets.

e 'av % rec' is the average percent recovery by both complementary bead sets; if either is ''under' or 'OVER', no average is reported.

f Lab05 did not analyze meat samples due to shipping / import problems. Lab09 failed to monitor bed sets ‐25, ‐26, ‐35, and ‐36 for egg and milk.

g 'under' or 'OVER' entries are ignored when calculating averages, standard deviations, and %CV values. No entry indicates data insufficient to support calculation.

h '<', '>' indicate MFI responses by more than 3 labs either too low ('under') or too high ('over'). 'na' indicates more than 3 labs generating conflicting, out of range data with at least 2 'under' and 2 'over'.

Table 14. Interpolated analyte recovered from incurred food samples: Baked muffinsa.

μg / gram food b Lowest Level c Lab number Coconut d EGG GLUTEN MILK WALNUT
‐20 ‐21 av % rec e ‐25 ‐26 av % rec ‐27 ‐28 av % rec ‐35 ‐36 av % rec ‐47 ‐48 av % rec
1 0.1 Lab01 under under 0.1 under under under 0.0 under 0.8 under
1 0.1 Lab02 under under under under under under under under under under
1 0.1 Lab03 under 1.2 under under under under under under under 0.26
1 0.1 Lab04 under under under under under under under under under under
1 0.1 Lab05 f under under under under under 2.4 under under under under
1 0.1 Lab06 under under under under under under 0.7 0.7 69 under under
1 0.1 Lab07 0.1 11.7 592 1.4 3.5 247 3.7 3.3 354 OVER OVER 1.6 2.1 186
1 0.1 Lab08 under under under under under under under under under under
1 0.1 Lab09 f under under under under under under
1 0.1 Lab10 under under under under under under under under under under
1 0.1 Lab11 under under under under under under under under under under
av g < h < < < < < < < < <
stdev g
%CV g
2.5 0.25 Lab01 0.0 under 0.2 under 0.2 under 0.1 0.1 5 1.2 under
2.5 0.25 Lab02 under under under under under under 0.1 0.1 4 0.2 under
2.5 0.25 Lab03 0.0 under under under under under under under under 0.3
2.5 0.25 Lab04 0.0 under under under 1.0 1.2 45 under under under under
2.5 0.25 Lab05 under under under under 5.0 12.2 344 4.5 6.1 213 under under
2.5 0.25 Lab06 under under under under 0.3 under 0.0 under under under
2.5 0.25 Lab07 0.2 under 1.4 2.6 78 1.6 under OVER OVER 1.8 under
2.5 0.25 Lab08 under under under under under under 3.6 3.6 145 under 0.5
2.5 0.25 Lab09 under under under under under under
2.5 0.25 Lab10 under under under under 2.5 under 0.7 0.7 under under
2.5 0.25 Lab11 under under under under under under 0.1 under under under
av < < < < < < 1.3 < < <
stdev 1.9
%CV 145
10 1 Lab01 0.1 under 0.6 0.5 6 0.6 0.3 5 0.6 0.6 6 3.4 0.7 21
10 1 Lab02 0.1 under 0.8 0.4 6 under 0.2 0.6 0.6 6 1.1 0.5 8
10 1 Lab03 0.1 1.1 6 0.4 under 1.1 under 0.2 0.3 2 1.1 0.7 9
10 1 Lab04 0.1 under under under under under 0.3 0.3 3 1.0 0.7 9
10 1 Lab05 under under under under under 5.4 under under under under
10 1 Lab06 0.2 1.1 6 under under 2.4 1.0 17 0.2 0.3 2 under under
10 1 Lab07 0.6 15 77 3.3 3.5 34 7.1 1.3 42 OVER OVER 7.6 6.4 70
10 1 Lab08 0.2 3.1 16 under under under under 6.1 6.3 62 1.0 1.0 10
10 1 Lab09 under under under under under under
10 1 Lab10 0.1 under under under 1.6 under 0.5 0.4 5 under under
10 1 Lab11 0.1 under under under under under 0.7 0.7 7 under under
av 0.2 < < < < < 1.2 1.2 12 < <
stdev 0.2 2.0 2.1 20
%CV 94 173 177 175
25 2.5 Lab01 0.4 1.5 4 1.9 2.2 8 5.5 1.6 14 1.4 1.4 6 7.6 3.0 21
25 2.5 Lab02 0.3 3.5 8 4.1 1.0 10 0.5 0.7 2 1.3 1.4 5 2.7 1.4 8
25 2.5 Lab03 0.2 2.0 5 1.0 under 1.5 under 0.9 0.9 4 2.0 1.3 7
25 2.5 Lab04 0.2 under 1.2 1.2 5 2.6 1.3 8 1.0 1.0 4 1.8 1.3 6
25 2.5 Lab05 0.2 under under under under 5.2 2.7 3.5 12 1.9 under
25 2.5 Lab06 0.3 0.9 2 under under 3.4 1.0 9 0.6 0.7 3 under under
25 2.5 Lab07 1.8 under 16 18 68 16 6.0 44 OVER OVER 19 9 57
25 2.5 Lab08 0.4 2.7 6 under under 3.5 1.4 10 1.5 1.4 6 3.4 2.1 11
25 2.5 Lab09 0.4 3.2 7 1.2 under 9.5 under
25 2.5 Lab10 0.3 under 2.0 3.5 11 under under 1.0 1.0 4 2.0 under
25 2.5 Lab11 0.3 under 2.0 2.7 9 3.1 0.9 8 1.9 2.2 8 2.7 1.3 8
av 0.4 < 4.0 < 4.2 2.2 14 1.4 1.5 6 5.3 <
stdev 0.5 5.3 4.7 2.1 14 0.6 0.9 3 5.6
%CV 107 133 114 93 103 47 57 52 106
100 10 Lab01 1.8 5.5 4 8.7 12 11 OVER 8.5 3.6 4.5 4 31 12 22
100 10 Lab02 1.3 4.5 3 18.6 6.1 12 0.3 1.2 1 4.3 5.0 5 9 5 7
100 10 Lab03 1.3 4.1 3 10 15 13 12 4.4 8 3.8 4.6 4 9 6 7
100 10 Lab04 1.6 6.1 4 12 20 16 28 7.6 18 4.3 5.2 5 9 5 7
100 10 Lab05 1.3 5.2 3 13.3 18 16 3.5 7.8 6 6.2 8.3 7 8 5 7
100 10 Lab06 1.6 5.6 4 8.5 10 9 OVER 8.3 0.6 0.7 1 17 under
100 10 Lab07 9.6 3.0 6 41 54 48 OVER 31 OVER OVER 67 47 57
100 10 Lab08 1.2 2.3 2 6.4 9 8 8.4 4.0 6 6.0 6.7 6 8 4 6
100 10 Lab09 1.4 2.1 2 6.3 3.5 5 11 under
100 10 Lab10 1.6 under 17.6 37 27 13.8 1.5 8 4.5 5.7 5 9 6 7
100 10 Lab11 1.3 under 8.5 14 11 11 4.1 8 5.4 6.9 6 7 4 6
av 2.2 4.3 3 14 20 17 10 7.4 7 4.3 5.3 5 17 11 14
stdev 2.5 1.5 1 10 15 12 8.4 8.2 5 1.7 2.1 2 18 14 17
%CV 112 35 41 71 76 71 81 110 66 39 40 39 106 133 120

a Interpolated ppm using the calibration standards (S1, S2, S5, S7) in a stepwise, linear analysis of the measured MFI (after subtracting background). S7 defines the upper limit of the dynamic range and 10‐times the standard deviation (10D) of the background (S0) defines the lower limit of quantitation. ppm calculated by converting the interpolated ng extractable protein/mL derived from the calibration standards, to ppm whole allergenic food using the protein content as referenced in Table 1. 'under' and 'OVER' indicate that the measured MFI was either less than or greater than the defined limits of the dynamic range (10D and S7).

b μg incurred per gram of food to generate the test portion.

c 'Lowest Level' refers to the lowest concentration of analyte that would generate the same analytical sample; by eliminating the optional dilution following extraction.

d Incurred analyte and associated bead sets.

e 'av % rec' is the average percent recovery by both complementary bead sets; if either is 'under' or 'OVER', no average is reported.

f Lab05 did not analyze meat samples due to shipping / import problems. Lab09 failed to monitor bed sets ‐25, ‐26, ‐35, and ‐36 for egg and milk.

g 'Over' or 'under' entries are ignored when calculating averages, standard deviations, and %CV values. No entry indicates data insufficient to support calculation.

h '<', '>' indicate MFI responses by more than 3 labs either too low ('under') or too high ('over'). 'na' indicates more than 3 labs generating conflicting, out of range data with at leat 2 'under' and 2 'over'.

Table 13. Interpolated analyte recovered from incurred food samples: Orange juice and dark chocolatea.

μg / gram food b Lowest Level c Lab number ALMOND d MILK SOY
-12 -13 av % rec e -12% e -35 -36 av % rec -45 -46 av % rec
1 0.1 Lab01 0.45 0.17 31 45 0.48 0.47 47 under under
1 0.1 Lab02 0.79 0.27 53 79 0.56 0.52 54 0.48 0.54 51
1 0.1 Lab03 0.58 under 58 0.70 under 0.34 under
1 0.1 Lab04 0.66 0.22 44 66 0.05 0.05 5 under under
1 0.1 Lab05 f under under 0.89 0.85 87 0.64 under
1 0.1 Lab06 0.5 under 46 8.7 16 1224 under under
1 0.1 Lab07 0.7 0.3 46 65 0.9 1.1 101 under under
1 0.1 Lab08 0.9 0.3 58 86 under 20.2 under under
1 0.1 Lab09 f 0.8 under 75 under under
1 0.1 Lab10 0.7 under 69 under under under under
1 0.1 Lab11 0.8 0.3 51 78 0.4 0.5 46 under under
av g 0.7 < h 67 1.6 4.9 223 < <
stdev g 0.1 14 2.9 8.1 442
%CV g 20 20 181 166 198
2.5 0.25 Lab01 1.1 0.4 30 44 1.4 1.4 57 0.6 under
2.5 0.25 Lab02 1.9 0.6 51 76 1.6 1.6 63 1.1 1.1 43
2.5 0.25 Lab03 1.4 under 55 1.2 1.3 50 0.7 0.9 33
2.5 0.25 Lab04 1.7 0.5 43 67 0.1 under 1.0 under
2.5 0.25 Lab05 under under 5.7 4.6 205 1.4 under
2.5 0.25 Lab06 1.2 under 49 15 18 653 under under
2.5 0.25 Lab07 1.3 0.4 35 54 0.9 1.0 38 under under
2.5 0.25 Lab08 1.8 0.6 48 73 OVER OVER under under
2.5 0.25 Lab09 1.8 0.6 49 73 under under
2.5 0.25 Lab10 1.7 0.5 44 70 0.8 0.9 34 0.9 under
2.5 0.25 Lab11 2.0 0.6 51 79 1.4 1.5 57 1.1 under
av 1.6 0.5 44 64 3.1 3.8 145 < <
stdev 0.3 0.1 8 12 4.6 5.9 213
%CV 19 18 17 19 149 156 147
10 1 Lab01 4.9 1.8 34 49 7.2 6.9 70 2.8 2.7 28
10 1 Lab02 9.8 2.3 61 98 8.2 8.1 82 4.8 4.7 48
10 1 Lab03 6.7 1.2 39 67 6.0 5.9 60 2.9 3.1 30
10 1 Lab04 9.2 1.7 55 92 1.6 1.4 15 4.6 4.2 44
10 1 Lab05 5.3 under 53 12 12 122 5.7 8.3 70
10 1 Lab06 5.3 0.7 30 53 OVER OVER 2.3 under
10 1 Lab07 18 4.1 109 176 25 27 263 25 23 240
10 1 Lab08 8.7 2.3 55 87 OVER OVER 4.7 5.9 53
10 1 Lab09 9.1 2.4 57 91 4.1 5.1 46
10 1 Lab10 8.4 1.9 51 84 7.2 6.7 70 4.4 4.8 46
10 1 Lab11 11.0 2.0 65 110 8.1 8.1 81 4.7 4.2 45
av 8.7 2.1 56 87 9.5 9.6 95 6.0 6.6 65
stdev 3.6 0.9 22 36 7.0 7.8 74 6.4 5.9 63
%CV 41 44 40 41 74 81 78 107 89 96
25 2.5 Lab01 14 2.8 34 57 27 25 104 6.7 6.4 26
25 2.5 Lab02 23 4.9 56 92 OVER OVER 12 12 47
25 2.5 Lab03 19 2.6 44 77 19 18 75 8.2 8.7 34
25 2.5 Lab04 22 4.4 53 89 15 11 53 12 11 45
25 2.5 Lab05 20 under 81 OVER 27 14 11 50
25 2.5 Lab06 18 1.5 39 73 OVER OVER 6.6 7.7 29
25 2.5 Lab07 21 4.1 51 85 25 25 98 9.2 8.3 35
25 2.5 Lab08 22 4.7 52 86 OVER OVER 11.1 10.0 42
25 2.5 Lab09 25 5.8 62 101 9.8 11 42
25 2.5 Lab10 21 4.3 51 85 20 19 77 11 11 43
25 2.5 Lab11 25 4.6 60 102 25 24 97 11.8 10.7 45
av 21 4.0 50 84 > h 21 10 9.8 40
stdev 3.1 1.3 9 13 5.5 2.3 1.7 8
%CV 15 32 17 15 26 23 18 19
100 10 Lab01 56 11 33 56 OVER OVER 34 30 32
100 10 Lab02 OVER 18 OVER OVER 44 45 44
100 10 Lab03 89 12 50 89 OVER OVER 36 37 36
100 10 Lab04 90 19 54 90 OVER OVER 44 41 42
100 10 Lab05 91 12 51 91 OVER OVER 45 36 41
100 10 Lab06 80 5 42 80 OVER OVER 33 29 31
100 10 Lab07 87 16 52 87 OVER OVER 50 45 47
100 10 Lab08 88 18 53 88 OVER OVER 40 40 40
100 10 Lab09 OVER 23 41 43 42
100 10 Lab10 OVER 17 OVER OVER 38 41 40
100 10 Lab11 OVER 17 OVER OVER 43 43 43
av > 15 83 > > 41 39 40
stdev 4.9 13 5.1 5.7 5
%CV 32 15 13 15 13
μg / gram food b Lowest Level c Lab number μg / gram food b Lowest Level c Lab number HAZELNUT d PEANUT
-29 -30 av % rec e -37 -38 av % rec
1 0.1 Lab01 1 0.2 Lab01 under 0.2 0.17 0.14 15
1 0.1 Lab02 1 0.2 Lab02 under 0.4 0.1 0.0 5
1 0.1 Lab03 1 0.2 Lab03 under under under under
1 0.1 Lab04 1 0.2 Lab04 0.50 0.47 49 under under
1 0.1 Lab05 f 1 0.2 Lab05 2.2 2.0 210 1.2 under
1 0.1 Lab06 1 0.2 Lab06 under under under under
1 0.1 Lab07 1 0.2 Lab07 0.5 0.3 40 under under
1 0.1 Lab08 1 0.2 Lab08 under under under under
1 0.1 Lab09 f 1 0.2 Lab09 under under under under
1 0.1 Lab10 1 0.2 Lab10 0.9 0.6 79 0.1 under
1 0.1 Lab11 1 0.2 Lab11 3.0 under 0.9 under
av g av g < h < < <
stdev g stdev g
%CV g %CV g
2.5 0.25 Lab01 2.5 0.5 Lab01 0.7 0.6 25 0.1 0.1 4
2.5 0.25 Lab02 2.5 0.5 Lab02 1.3 1.0 47 0.2 0.1 6
2.5 0.25 Lab03 2.5 0.5 Lab03 under under under under
2.5 0.25 Lab04 2.5 0.5 Lab04 1.6 1.4 61 0.2 under
2.5 0.25 Lab05 2.5 0.5 Lab04 2.8 2.7 109 1.2 under
2.5 0.25 Lab06 2.5 0.5 Lab06 under 0.9 under under
2.5 0.25 Lab07 2.5 0.5 Lab07 1.0 0.6 32 under under
2.5 0.25 Lab08 2.5 0.5 Lab08 under under under under
2.5 0.25 Lab09 2.5 0.5 Lab09 under under under under
2.5 0.25 Lab10 2.5 0.5 Lab10 2.5 1.8 84 0.3 under
2.5 0.25 Lab11 2.5 0.5 Lab11 1.9 under under under
av av < < < <
stdev stdev
%CV %CV
10 1 Lab01 10 2 Lab01 3.4 2.8 31 0.7 0.5 6
10 1 Lab02 10 2 Lab02 6.0 4.6 53 0.8 0.4 6
10 1 Lab03 10 2 Lab03 8.9 7.9 84 0.6 under
10 1 Lab04 10 2 Lab04 5.8 4.8 53 0.7 0.5 6
10 1 Lab05 10 2 Lab04 5.7 5.8 58 1.3 under
10 1 Lab06 10 2 Lab06 6.6 4.9 58 1.1 0.5 8
10 1 Lab07 10 2 Lab07 4.2 2.5 34 0.5 0.3 4
10 1 Lab08 10 2 Lab08 5.3 2.8 40 under under
10 1 Lab09 10 2 Lab09 1.7 1.3 15 under under
10 1 Lab10 10 2 Lab10 8.2 5.6 69 1.4 1.0 12
10 1 Lab11 10 2 Lab11 8.5 5.7 71 2.3 under
av av 5.8 4.4 51 1.0 <
stdev stdev 2.2 1.9 20 0.6
%CV %CV 38 43 39 55
25 2.5 Lab01 25 5 Lab01 9.3 7.9 34 1.7 1.2 6
25 2.5 Lab02 25 5 Lab02 14 12 53 2.0 1.0 6
25 2.5 Lab03 25 5 Lab03 23 22 91 1.5 1.1 5
25 2.5 Lab04 25 5 Lab04 18 16 67 2.2 1.5 7
25 2.5 Lab05 25 5 Lab04 17 18 71 1.6 1.3 6
25 2.5 Lab06 25 5 Lab06 16 13 57 1.9 1.2 6
25 2.5 Lab07 25 5 Lab07 11 7.0 36 1.2 0.8 4
25 2.5 Lab08 25 5 Lab08 12 7.9 40 5.9 5.2 22
25 2.5 Lab09 25 5 Lab09 4.0 3.0 14 under under
25 2.5 Lab10 25 5 Lab10 17 13 61 2.9 2.1 10
25 2.5 Lab11 25 5 Lab11 16 12 56 3.7 2.4 12
av av 14 12 53 2.5 1.8 8
stdev stdev 5.1 5.5 21 1.4 1.3 5
%CV %CV 35 46 40 57 72 63
100 10 Lab01 100 20 Lab01 42 50 46 7.1 5.9 7
100 10 Lab02 100 20 Lab02 OVER OVER 6.1 4.1 5
100 10 Lab03 100 20 Lab03 OVER OVER 7.6 6.8 7
100 10 Lab04 100 20 Lab04 OVER OVER 8.7 5.9 7
100 10 Lab05 100 20 Lab04 OVER OVER 4.8 2.0 3
100 10 Lab06 100 20 Lab06 OVER OVER 7.0 4.5 6
100 10 Lab07 100 20 Lab07 54 38 46 5.6 4.0 5
100 10 Lab08 100 20 Lab08 48 32 40 5.7 3.8 5
100 10 Lab09 100 20 Lab09 16 15 15 2.1 2.0 2
100 10 Lab10 100 20 Lab10 OVER 55 9.3 5.9 8
100 10 Lab11 100 20 Lab11 56 43 49 9.5 6.0 8
av av > h > 6.7 4.6 6
stdev stdev 2.2 1.6 2
%CV %CV 32 35 33

a Interpolated ppm using the calibration standards (S1, S2, S5, S7) in a stepwise, linear analysis of the measured MFI (after subtracting background). S7 defines the upper limit of the dynamic range and 10‐times the standard deviation (10D) of the background (S0) defines the lower limit of quantitation. ppm calculated by converting the interpolated ng extractable protein/mL derived from the calibration standards, to ppm whole allergenic food using the protein content as referenced in Table 1. 'under' and 'OVER' indicate that the measured MFI was either less than or greater than the defined limits of the dynamic range (10D and S7).

b μg incurred per gram of food to generate the test portion.

c 'Lowest Level' refers to the lowest concentration of analyte that would generate the same analytical sample; by eliminating the optional dilution following extraction.

d Incurred analyte and associated bead sets.

e 'av % rec' is the average percent recovery by both complementary bead sets; if either is 'under' or 'OVER', no average is reported. '‐12%' is the percent recovery calcualted based on Almond‐12; only for Orange Juice samples.

f Lab05 did not analyze meat samples due to shipping / import problems. Lab09 failed to monitor bed sets ‐25, ‐26, ‐35, and ‐36 for egg and milk.

g 'Over' or 'under' entries are ignored when calculating averages, standard deviations, and %CV values. No entry indicates data insufficient to support calculation

h '<', '>' indicate MFI responses by more than 3 labs either too low ('under') or too high ('over'). 'na' indicates more than 3 labs generating conflicting, out of range data with at leat 2 'under' and 2'over'.

Further, since a goal of this validation was to appraise xMAP FADA utility when performed by analysts of diverse expertise, poor performing laboratories were not statistically dropped from the analyses. Indeed, to accentuate notation of potential problem areas, if within any sample group (e.g., 1 ppm egg in meat) more than three labs generated MFI outside the dynamic range, no average was calculated and instead noted with either ‘<‘or ‘>‘for the bead set.

In addition, tabulated are the averages of the complementary bead sets. If two complementary bead sets generated comparable ppm levels, it indicates that both bead sets were equally affected. Consistent with the ratio analyses, the interpolated amount of almond in the orange juice samples interpolated to different concentrations with the almond-13 values considerably less than either the almond-12 values or the incurred concentrations of allergenic food. Table 15 tabulates the average calculated concentrations of allergenic food across the 11 participating laboratories, except for when data was not submitted (e.g., Lab 05 did not receive meat samples for analysis, Lab 09 did not collect MFI data for either the egg or milk bead sets). Extending the approach employed in Tables 1214, when more than three labs generated average MFI outside the dynamic range defined by S0+10D and S7, the fraction of labs outside the dynamic range were indicated with ‘<‘or ‘>‘in place of an average. For example, the detection of 1 ppm egg in meat generated average MFI below the dynamic range for 6 of the 9 labs and therefore the notation ‘6/9<‘in the table. Similarly, 100 ppm milk incurred in orange juice generated for all 11 labs average MFI exceeding S7 and is noted in the table as >11/11. In addition, all data that exceeded the lower limit of the dynamic range, defined by S0+10D, are highlighted in yellow to help illustrate the utility of the xMAP FADA to detect the analytes. As expected, with increasing concentration, the %CV values decreased. Since, several of the 100 μg/g incurred samples of meat, orange juice, and dark chocolate exceeded the dynamic range, the %CV (RSDR) values of the 25 μg/g samples were compared. Half of the meat, all orange juice, and one of the dark chocolate bead sets displayed %CV values ≤ 35%. The allergenic foods incurred at 100 μg/g into the baked muffins displayed %CV values ≥ 50% for 7 of the 10 bead sets. The order for worsening %CV of orange juice, meat, dark chocolate, and baked muffins is consistent with the increasing food sample processing and manipulation. Preparation of the orange juice and meat samples entailed only melting the supplied samples and adding extraction buffer, though pipetting of the centrifuged meat extract required avoiding any fat layer. The dark chocolate samples required adding pre-warmed buffer to dissolve the dark chocolate to conduct the extraction. In contrast, the muffin samples entailed baking the incurred allergenic foods and subsequently the analysts had to remove the mini muffins and dice them with a clean razor blade before adding extraction buffer. It is therefore possible that the %CV associated with the recovery of the allergenic foods relate to the degree of processing of the food samples and /or proficiency in analyst handling of the food samples.

Table 15. Average interpolated allergenic food concentrations incurred in food samplesa.

MEAT EGG c, d GLUTEN MILK
‐25 ‐26 ‐27 ‐28 ‐35 ‐36
μg/g b equiv b ppm %CV ppm %CV ppm %CV ppm %CV ppm %CV ppm %CV
1 0.1 6/9 < na 7/9 < na 2.5 91 2.4 49 2.8 108 3.2 115
2.5 0.25 2.2 45 7/9 < na 5.2 49 5.6 41 2 134 4.5 164
10 1 9 32 5 43 28 9 25 16 7.2 103 6.4 118
25 2.5 22 32 27 53 > 10/10 na > 10/10 na 9 31 7 22
100 10 > 4/9 na > 7/9 na > 10/10 na > 10/10 na > 4/9 na 15 67
ORANGE JUICE ALMOND MILK SOY
μg/g equiv ‐12 ‐13 ‐35 ‐36 ‐45 ‐46
ppm %CV ppm %CV ppm %CV ppm %CV ppm %CV ppm %CV
1 0.1 0.7 20 5/11 < na 1.6 181 4.9 166 8/11 < na 10/11 < na
2.5 0.25 1.6 19 0.5 18 3.1 149 3.8 156 4/10 < na 9/11 < na
10 1 8.7 41 2.1 44 9.5 74 9.6 81 6 107 6.6 89
25 2.5 21 15 4 32 > 4/11 na 21 26 10 23 9.8 18
100 10 > 4/11 na 15 32 > 11/11 na > 11/11 na 41 13 39 15
BAKED MUFFIN COCONUT EGG GLUTEN
μg/g equiv ‐20 ‐21 ‐25 ‐26 ‐27 ‐28
ppm %CV ppm %CV ppm %CV ppm %CV ppm %CV ppm %CV
1 0.1 10/11 < na 9/11 < na 8/10 < na 9/10 < na 10/11 < na 9/11 < na
2.5 0.25 7/11 < na 11/11 < na 8/10 < na 9/10 < na 5/11 < na 9/11 < na
10 1 0.2 94 7/11 < na 6/10 < na 7/10 < na 6/11 < na 6/11 < na
25 2.5 0.4 107 5/11 < na 4 133 4/10 < na 4.2 114 2.2 93
100 10 2.2 112 4.3 35 14 71 15 76 10 81 7.4 110
DARK CHOCOLATE HAZELNUT PEANUT
μg/g equiv ‐29 ‐30 ‐37 ‐38
ppm %CV ppm %CV ppm %CV ppm %CV
1 0.2 6/11 < na 5/11 < na 6/11 < na 9/11 < na
2.5 0.5 4/11 < na 4/11 < na 6/11 < na 9/11 < na
10 2 5.8 38 4.4 43 1 55 0.5 47
25 2.5 14 35 12 46 2.5 57 1.8 72
100 20 > 6/11 na > 5/11 na 6.7 32 4.6 35
MEAT SOY
‐45 ‐46
μg/g b equiv b ppm %CV ppm %CV
1 0.1 0.7 123 1.3 65
2.5 0.25 1.1 50 1.1 68
10 1 4.5 56 3.6 53
25 2.5 11 56 9 60
100 10 51 88 41 90
ORANGE JUICE
μg/g equiv
1 0.1
2.5 0.25
10 1
25 2.5
100 10
BAKED MUFFIN MILK WALNUT
μg/g equiv ‐35 ‐36 ‐47 ‐48
ppm %CV ppm %CV ppm %CV ppm %CV
1 0.1 7/9 < na 8/10 < na 9/11 < na 9/11 < na
2.5 0.25 1.3 145 2.1 125 8/11 < na 9/11 < na
10 1 1.2 173 1.2 177 5/11 < na 5/11 < na
25 2.5 1.4 47 1.5 57 5.3 106 4/11 < na
100 10 4.3 39 5.3 40 17 106 11 133
DARK CHOCOLATE
μg/g equiv
1 0.2
2.5 0.5
10 2
25 2.5
100 20

a Concentrations (ppm) derived from ng protein/mL calculated from calibration standards S0, S1, S2, S5, and S7 in PBST (UD Buffer for chocolate) assuming linearity between the different calibrants and using 10‐times the standard deviation of the background (S0+10D) as a lower limit of quantitation. Conversion from interpolated ng protein/mL based on analyte protein content with the calculated ppm averaged across participating laboratories. All samples were extracted in triplicate and the MFI, after subtraction of the background (S0), to calculate ppm of analyte.

b 'μg/g' incurred into a gram of food, 'equiv' is comparable concentration of analyte that would generate similar responses by omitting optional dilution of extract (10X for PBST and 5X for UD Buffer).

c If more than three of the contributing laboratories generated responses outside the dynamic range, the fraction of such labs is reported. Specifically, '6/9 <' indicates 6 of 9 below 10‐time the standard deviation of the background (10D) and '> 4/9' indicates 4 of 9 above S7.

d Light yellow highlight indicates detection of analyte more than the lower limit S0+10D.

The percent recoveries are presented in Table 16 with those instances where more than three labs were outside the dynamic range indicated by the fraction outside the dynamic range. The average percent recoveries were generally consistent with what was observed in the single lab validation and typically observed with ELISAs employing buffered-detergent extraction protocols. The baked muffins displayed the lowest levels of recovery, followed by the chocolate samples, the orange juice samples, and lastly the meat samples. The baked muffins averaged 10% for incurred levels of 10, 25, and 100 μg/g. This percent recovery was approximately one-third the level observed in the single lab validation for baked muffins (33%). This may be due to a combination of less experienced analysts and differences in preparing the muffins for extraction (e.g., degree of dicing). The chocolate samples displayed average recoveries for hazelnut greater than peanut of 57% and 9%, respectively. In the single lab validation, the two analytes behaved comparably with an average percent recovery for all analytes of approximately 12% for the chocolate incurred samples relative to the allergenic foods extracted from PBST samples. The reproducibly higher recovery of hazelnut versus peanut in the MLV may reflect a feature of this particular sample since a similar elevated recovery was not observed in the single lab validation [11]. Recoveries were excellent for the meat and orange samples except for almond-13 which were 15–20%; about one-quarter the recoveries observed with almond-12 and consistent with the change in the almond-13/almond-12 ratio.

Table 16. Percent allergenic food recovered from food samplesa.

MEAT EGG c, d GLUTEN MILK SOY
‐25 ‐26 ‐27 ‐28 ‐35 ‐36 ‐45 ‐46
μg/g b equiv b % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery
1 0.1 6/9 < 7/9 < 250 240 280 320 70 130
2.5 0.25 88 7/9 < 208 224 80 180 44 44
10 1 90 50 280 250 72 64 45 36
25 2.5 88 108 > 10/10 > 10/10 36 28 44 36
100 10 > 4/9 > 7/9 > 10/10 > 10/10 > 4/9 15 51 41
ORANGE JUICE ALMOND MILK SOY
‐12 ‐13 ‐35 ‐36 ‐45 ‐46
μg/g equiv % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery
1 0.1 70 5/11 < 160 490 8/11 < 10/11 <
2.5 0.25 64 20 124 152 4/10 < 9/11 <
10 1 87 21 95 96 60 66
25 2.5 84 16 > 4/11 84 40 39
100 10 > 4/11 15 > 11/11 > 11/11 41 39
BAKED MUFFIN COCONUT EGG GLUTEN MILK WALNUT
‐20 ‐21 ‐25 ‐26 ‐27 ‐28 ‐35 ‐36 ‐47 ‐48
μg/g equiv % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery % Recovery
1 0.1 10/11 < 9/11 < 8/10 < 9/10 < 10/11 < 9/11 < 7/9 < 8/10 < 9/11 < 9/11 <
2.5 0.25 7/11 < 11/11 < 8/10 < 9/10 < 5/11 < 9/11 < 52 84 8/11 < 9/11 <
10 1 2 7/11 < 6/10 < 7/10 < 6/11 < 6/11 < 12 12 5/11 < 5/11 <
25 2.5 2 5/11 < 16 4/10 < 17 9 6 6 21 4/11 <
100 10 2 4 14 15 10 7 4 5 17 11
DARK CHOCOLATE HAZELNUT PEANUT
‐29 ‐30 ‐37 ‐38
μg/g equiv % Recovery % Recovery % Recovery % Recovery
1 0.2 6/11 < 5/11 < 6/11 < 9/11 <
2.5 0.5 4/11 < 4/11 < 6/11 < 9/11 <
10 2 58 44 10 5
25 2.5 56 48 10 7
100 20 > 6/11 > 5/11 7 5

a Percent recoveries based on calculated ppm divided by the amount of allergenic food incurred in the food samples.

b 'μg/g' incurred into a gram of food, 'equiv' is comparable concentration of analyte that would generate similar responses by omitting optional dilution of extract (10X for PBST and 5X for UD Buffer).

c If more than three of the contributing laboratories generated responses outside the dynamic range, the fraction of such labs is reported. Specifically, '6/9 <' indicates 6 of 9 below 10‐time the standard deviation of the background (10D) and '> 4/9' indicates 4 of 9 above S7.

d Light yellow highlight indicates detection of analyte more than the lower limit S0+10D.

Direct comparison controls (DCCs)

Direct comparison controls (DCCs) provide a second measure of analyst proficiency, assay performance, as well as a direct tool for ascertaining whether a food sample contains a target analyte at amounts exceeding a specified level. The DCCs are prepared on the day of analysis by spiking the specified allergenic food into analyte-free samples and analyzing the samples alongside. As such, DCCs avoid any assumptions and conversion factors associated with interpolating analyte concentrations from the calibration standards. Specifically, the DCCs can be prepared by spiking the analyte-free food with the allergenic food as may be inadvertently present in the food samples. This is exemplified by using suspensions of ground raw hazelnut, not a protein extract as employed in the calibration standards. Further, since the detection of spiked allergenic food typically displays higher levels of recovery versus incurred analyte, the observation of MFI characteristic of the analyte and greater than the DCC indicates a definitively greater amount of analyte in the food sample.

The DCCs are handled and extracted alongside the food samples. As such, DCCs also provide a measure of the analyst’s ability to perform the extraction protocol. DCCs also serve as controls for matrix effects, supplementing the roles played by the built-in redundancy, multiplex design, and AssayChex bead sets included in the bead set cocktail. Tables 4 and 5 present the MFI and S/N-1 data for DCCs containing 20 ppm gluten (G20), 10 ppm peanut (P10), and 50 ppm NFDM (M50) analyzed using reagents from Lot 5. Though 20 ppm intact gluten is a regulatory threshold used to define ‘gluten-free’ to safeguard consumers with Celiac Disease, in the USA thresholds have not been established for food allergens. It is envisioned that if threshold levels are adopted, such would influence the amounts of allergenic food used to make the DCCs. Further, the DCCs employed in an experiment would be chosen to represent the allergenic foods of interest; for example, if testing for tree nuts, DCC’s of almond, hazelnut, and walnut might be employed.

The DCC MFI data tabulated in Tables 4 and 5 are the average of DCCs prepared in four different foods, each in triplicate. The meat, orange juice, and baked muffin samples were extracted (1:20) using the PBST Buffered-detergent Protocol and diluted 10-fold prior to mixing with the bead cocktail (net, 200-folddilution). The dark chocolate samples were extracted (1:40) according to the UD Buffer Protocol and diluted 5-fold with UD Buffer prior to analysis (net, 200-fold). Inasmuch as UD Buffer contains milk, the dark chocolate samples were not included in the M50 data sets. As expected, the gluten, peanut, and milk DCCs generated intense responses with the gluten, peanut, and milk bead sets, respectively with virtually no significant responses with the other bead sets. As a result, the S/N-1 data in Table 5 are zeros for all bead sets except the spiked analyte. Not surprisingly, the milk DCC displayed the largest S/N-1 consistent with its 50 ppm concentration being considerably greater than the concentration of gluten (20 ppm) and peanut (10 ppm) in the other two types of DCC with differences between the complementary bead sets as previously observed (see section on ratio analysis above and discussions regarding analyte, bead-specific sensitivities in Cho et al. 2015 [5]).

The MFI observed for the various food samples compared to the MFI response generated by the appropriate DCC are depicted in a series of figures; for gluten bead sets -27 and -28 compared to G20 (S2 Fig), milk bead sets -35 and -36 compared to M50 (S3 Fig), and peanut bead sets -37 and -38 compared to P10 (S4 Fig). The data for each participant is plotted on separate graphs with the results obtained with the meat (red), orange juice (blue), baked muffin (green) and dark chocolate (purple) presented as the difference after subtracting the MFI generated by the DCC. The only exceptions were omission of the dark chocolate samples with bead sets -35 and -36 due to the presence of milk in the UD extraction buffer, meat analyses were not conducted by Lab 05 and Lab 09 did not collect data for bead sets -35 and -36. The detection of an analyte is indicated by a positive slope with crossing of the abscissa expected for samples incurred at levels greater than the amount spiked into the food (by the analyst) to make the DCCs.

S2 Fig depicts the MFI by gluten bead sets -27 and -28 minus the responses generated by the 20 ppm gluten DCCs (G20). All participants correctly detected the presence of gluten in the meat and baked muffin samples with all, but one of the labs, correctly displaying flat lines for the orange juice and dark chocolate samples; ten of the dark chocolate data points properly overlapped the abscissa for gluten-27. Further, eight of the 10 labs displayed appropriate transitions over the abscissa for the meat samples with gluten-27 and six correctly crossed the abscissa for gluten-28, another two being shifted. Reduced recovery has been extensively documented with baked food samples. It was therefore not surprising that the baked muffin samples displayed less intense MFI and only crossed the abscissa appropriately six times with gluten-27 and was greatly shifted in the gluten-28 plots.

Milk was incurred into all samples, but since the dark chocolate was extracted using UD buffer, it was not included in the plots for milk bead sets -35 and -36. Also, one of the labs inadvertently failed to monitor bead sets -35 and -36; therefore, there were only 10 participants for these analyses. Seven of the 10 labs correctly indicated the presence of milk using bead set -35; eight correctly indicated the presence of milk using bead set -36 (S3 Fig). Only three labs displayed results with the meat and orange juice samples appropriately crossing the abscissa. Not surprisingly, none of the baked muffin samples crossed the abscissa and for eight of the ten labs, the baked muffin samples displayed poorer performance than the meat or orange juice samples, with the meat and orange juice data overlapping six times.

Peanut was only incurred into the dark chocolate samples. All 11 of the laboratories correctly detected the presence of peanut in the dark chocolate with seven appropriately crossing the abscissa for peanut bead sets -37 and -38 (S4 Fig). Interestingly, three of the labs displayed slight increases for either the orange juice or baked muffin samples. Otherwise, the eight remaining labs displayed no significant responses for the other food samples, with four of the labs properly displaying indistinguishable flat lines for the meat, orange juice, and baked muffin samples with bead set -37, five labs with bead set -38.

Overall the analysts were able to successfully perform the DCC part of the MLV with only a few minor errors, possibly related to the complexity associated with preparing the DCCs (see INSTR file, pages on DCC preparation). A question critical for the end-user to address is the level of proficiency desired which relates to quantitative variance. The lower quantitative performance of the baked muffin samples is in agreement with the reduced recovery typically observed with baked products using buffered-detergent extraction protocols as observed for the xMAP FADA (above, single lab validation) and commercial ELISA test kits. Despite this variability, the DCCs provided an excellent tool for direct qualitative detection and ascertaining whether a sample contained analyte in excess of a specified target level.

Overview

The xMAP FADA displayed excellent performance reliability despite the MLV entailing 9 out of 11 participants not experienced in the assay. Absolute differences in MFI values were observed but intra-lab variability was low and comparable to ELISA analysis. Further, despite sometimes high levels of variance in the inter-lab data, the strong signal-to-noise enabled reliable qualitative detection and avoidance of false negatives. Specifically, the excellent signal-to-noise at the start of the dynamic range and its large increase across the dynamic range for each bead set made erroneous false negatives due to variance in the data less likely. Further, ratio analyses (i.e., ratios between complementary bead sets and multi-antibody profiling) distinguished between target analytes and cross-reactive homologues. Inter-lab variances of the ratio analyses varied with the type of incurred food samples. The worst (largest) variances were observed with the baked muffin samples. The baked muffin samples containing 100 μg/g of the allergenic foods averaged %CV (RSDR) values of 26% in the ratio analyses and displayed an average recovery of 9% ±5% (n = 10 bead sets). In contrast, the allergenic foods incurred at 100 μg/g in the meat, orange juice, and chocolate samples displayed %CV values in the ratio analyses of 15%, 15%, and 17%, respectively. Recovery patterns were comparable to those published in the single lab validation and basically comparable to what has been typically observed with ELISA test kits that employ buffered-detergent extraction protocols [11, 17]. The baked muffins displayed the poorest recoveries and though only approximately 10%, still generated highly reproducible MFI. Thus, the five incurred allergenic foods were reliably detected except for two labs generating questionable results for gluten and two labs not observing confirmatory responses with coconut-21 and walnut-48. The only potentially serious concern was the inability of the labs to generate MFI in excess of the S1 calibration standard for both walnut bead sets (primarily-48); however, the MFI did exceed the S0+10D in all but two instances for walnut-48, and displayed an appropriate increase in intensity with concentration. It is not clear whether the poor sensitivity in detecting walnut was entirely due to baking or in part related to the preparation of the walnut stock and subsequent sub-stock solutions and the inability to generate a ground powder to generate uniform suspensions. Lastly, preparation of the samples included an optional dilution of the food extracts. Thus, the concentration of analyte in the analytical samples could also be generated from samples containing 0, 0.1, 0.25, 1, 2.5, and 10 μg/g (0.2, 0.5, 2, 5, and 20 μg/g for the chocolate samples) with minimal changes in matrix carry over. As such, the data collected in the MLV represents worse case scenarios since it is expected that typically samples exceeding 10 ppm will be the focus of any analyses and the optional dilution step can be omitted.

The xMAP FADA is currently the only commercial antibody-based method capable of reliably detecting and distinguishing between allergenic foods and cross-reactive homologues for seven of the ‘big+ eight’ classes of food allergens regulated in the USA and many other countries. The xMAP FADA was also multi-lab validated for its ability to quantify detected allergenic foods and using DCCs can provide a second measure of whether an allergenic food was present at concentrations exceeding a chosen target level. Variability in performance with the DCCs was greater than typically observed with experienced analysts and probably reflects the complexity in preparing the DCCs which basically doubles the assay time of < 4 hrs. Despite this variability, the xMAP was successfully performed by all participants despite the lack of expertise. This does not mean that it is not necessary to maximize the proficiency of analysts performing the xMAP FADA, instead all effort should be made to maximize proficiency and hence assay capability. Lastly, by virtue of its modular design the xMAP FADA is compatible to changes in the repertoire to meet special and future needs.

Supporting information

S1 Fig. Lab specific average calibration curves.

(PDF)

S2 Fig. Comparisons between the MFI generated by the incurred food samples and the 20 ppm gluten DCCs (G20).

(PDF)

S3 Fig. Comparisons between the MFI generated by the incurred food samples and the 50 ppm milk DCCs (M50).

(PDF)

S4 Fig. Comparisons between the MFI generated by the incurred food samples and the 10 ppm peanut DCCs (P10).

(PDF)

S1 Appendix. Instructional items supplied to the laboratories.

(PDF)

Acknowledgments

The authors are extremely grateful to the participants without whom the MLV would not be possible: William L. Nowatzke, Ph.D. (Radix BioSolutions); LCDR Chung Y. Cho (FDA); Lauren S. Jackson, Ph.D. (FDA); Nicholas R. Paganella, Marie Kelleher, Sam Zipperer (USDA, FSIS); Terry B. Koerner, Ph.D. (Health Canada); Kodumudi Venkat Venkateswaran, Ph.D. (Omni Array Biotechnology, LLC); Neeraja Venkateswaran, Ph.D. (Tetracore, Inc.); Yolanda Drake (FDA, ORA-Denver Laboratory); Roehl L. Valcos (FDA, ORA-NFFL); LieuChi Phan (FDA, ORA-PSFFL); Tammara Stephens (FDA, ORA-SFFL). Gratitude is also expressed to Carl Coward and Martin G. Wright (FDA) for shipping the samples. Our gratitude is expressed to Masahiro Shoji, Ph.D. (Morinaga Institute of Biological Sciences, Inc., retired); Mansour Samadpour, Ph.D. (IEH Laboratories and Consulting Group); Thomas Grace and John Leslie, Sundaravel Damodaran, Elliott Zell (BiaDiagnostics, LLC; Elution Technologies—3M Food Safety) for their openness and willingness to make resources available for the xMAP FADA. Gratitude is also expressed to Shaun A. MacMahon, Ph.D. and Gregory O. Noonan, Ph.D. (FDA) for their support, without which the xMAP FADA would not have been developed or this MLV conducted. Lastly, but not least, gratitude is expressed to Lynn L-B Rust, Ph.D. whose support was invaluable.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

The author(s) received no specific funding for this work. All costs were covered by the FDA, US Government as part of an internal research project. As indicated in the by-line, the authors WLN, KGO, KVV, and NV are affiliated with the commercial companies: Radix BioSolutions, Georgetown, TX, Omni Array Biotechnology LLC. and Tetracore, Inc., Rockville, MD. These commercial companies provided support only in the form of salaries for authors [WLN, KGO, KVV, and NV], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.

References

  • 1.Gupta RS, Waren CM, Smith BM, Jiang J, Blumenstock JA, Davis MM, et al. Prevalence and severity of food allergies among US adults. JAMA Netw Open 2019; 2(1):e185630 10.1001/jamanetworkopen.2018.5630 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Food Allergen Labeling and Consumer Protection Act of 2004 (FALCPA) Public Law 108–282, Title II. https://www.fda.gov/downloads/Food/GuidanceRegulation/UCM179394.pdf. Accessed December 28, 2019.
  • 3.Gupta RS, Springston EE, Warrier MR, Smith B, Kumar R, Pongracic J, et al. The prevalence, severity, and distribution of childhood food allergy in the United States. Pediatrics 2011; 128(1):e9–17. 10.1542/peds.2011-0204 [DOI] [PubMed] [Google Scholar]
  • 4.Smeekens JM, Bagley K, Kulis M. Tree nut allergies: Allergen homology, cross-reactivity, and implications for therapy. Clin Exp Allergy 2018; 48(7):762–772. [DOI] [PubMed] [Google Scholar]
  • 5.Cho CY, Nowatzke W, Oliver K, Garber EAE. Multiplex detection of food allergens and gluten. Anal Bioanal Chem 2015; 407: 4195–4206, erratum 2016; 408: 6509–6510. 10.1007/s00216-015-8645-y [DOI] [PubMed] [Google Scholar]
  • 6.Cho CY, Ivens KO, Nowatzke WL, Robotham J, Samadpour M, Grace T, et al. Extension of xMAP Food Allergen Detection Assay to Include Sesame. J Food Prot. 2020; 83:129–135 10.4315/0362-028X.JFP-19-304 [DOI] [PubMed] [Google Scholar]
  • 7.Cho CY, Oles C, Nowatzke WL, Oliver KG, Garber EAE. Cross-reactivity profiles of legumes and tree nuts using the xMAP® Multiplex Food Allergen Detection Assay. Anal Bioanal Chem. 2017; 409: 5999–6014. [DOI] [PubMed] [Google Scholar]
  • 8.Pedersen RO, Peters T, Panda R, Wehling P, Garber EAE. Detection and antigenic profiling of undeclared peanut in imported garlic using an xMAP® multiplex immunoassay for food allergens. J Food Prot. 2017; 80: 1204–1213. [DOI] [PubMed] [Google Scholar]
  • 9.Cho CY, Puente-Lelievre C, Jones GD, Stadig S, Taylor DA, Eischeid AC, et al. Application of multi-antigen profiling to detect pecan. J Food Prot. 2018; 81: 700–704. [DOI] [PubMed] [Google Scholar]
  • 10.Pedersen RO, Nowatzke WL, Cho CY, Oliver KG, Garber EAE. Cross-reactivity by botanicals used in dietary supplements and spices using the multiplex xMAP Food Allergen Detection Assay (xMAP FADA). Anal Bioanal Chem. 2018; 410: 5791–5806. 10.1007/s00216-018-1187-3 [DOI] [PubMed] [Google Scholar]
  • 11.Nowatzke WL, Oliver KG, Cho CY, Rallabhandi P, Garber EAE. Single laboratory validation of the multiplex xMAP Food Allergen Detection Assay (xMAP FADA) with incurred food samples. J Agric Food Chem. 2019; 67: 484–498. [DOI] [PubMed] [Google Scholar]
  • 12.Rallabhandi P, Cho CY, Nowatzke WL, Oliver KG, Garber EA. Robustness Testing of the xMAP Food Allergen Detection Assay: A Multiplex Assay for the Simultaneous Detection of Food Allergens. J Food Prot. 2020; 83:1050–1056. 10.4315/JFP-19-531 [DOI] [PubMed] [Google Scholar]
  • 13.Panda R, Boyer M, Garber EAE. A multiplex competitive ELISA for the detection and characterization of gluten in fermented and hydrolyzed foods. Anal Bioanal Chem. 2017; 409: 6959–6973. [DOI] [PubMed] [Google Scholar]
  • 14.Cao W, Watson D, Bakke M, Panda R, Bedford B, Kande PS, et al. Detection of gluten during the fermentation process to produce soy sauce. J Food Prot. 2017; 80: 799–808. [DOI] [PubMed] [Google Scholar]
  • 15.U.S. Food and Drug Administration Foods Program Guidelines for the Validation of Chemical Methods in Food, Feed, Cosmetics, and Veterinary Products, 3rd Edition, 2019. https://www.fda.gov/media/81810/download. Accessed December 28, 2019.
  • 16.U.S. Department of Agriculture, Agricultural Research Service. FoodData Central. https://fdc.nal.usda.gov/index.html. Accessed December 28. 2019.
  • 17.Tiwari RS, Venkatachalam M, Sharma GM, Su M, Roux KH, Sathe SK. Effect of food matrix on amandin, almond (prunus dulcis l.) major protein, immunorecognition and recovery. LWT-Food Science and Technology. 2010. 43:675–683. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

S1 Fig. Lab specific average calibration curves.

(PDF)

S2 Fig. Comparisons between the MFI generated by the incurred food samples and the 20 ppm gluten DCCs (G20).

(PDF)

S3 Fig. Comparisons between the MFI generated by the incurred food samples and the 50 ppm milk DCCs (M50).

(PDF)

S4 Fig. Comparisons between the MFI generated by the incurred food samples and the 10 ppm peanut DCCs (P10).

(PDF)

S1 Appendix. Instructional items supplied to the laboratories.

(PDF)

Data Availability Statement

All relevant data are within the manuscript and its Supporting Information files.


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES