Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2021 Apr 1.
Published in final edited form as: IEEE Trans Instrum Meas. 2019 May 13;69(4):1690–1697. doi: 10.1109/tim.2019.2916242

Remote characterization of resonance frequency with a wirelessly powered parametric oscillator

Wei Qian 1, Chunqi Qian 2,*
PMCID: PMC7351332  NIHMSID: NIHMS1602557  PMID: 32655185

Abstract

It is of both fundamental importance and practical value to measure the frequency of an LC resonator beyond the near-field region, especially when the resonator is used as a standalone capacitive sensor embedded inside a closed environment. To improve the coupling efficiency between the resonator and the external sniffer loop, we propose a novel method to integrate the LC resonator with a wirelessly-powered parametric resonator whose oscillation signal can be remotely identified in a noisy background. By measuring the minimum power level that is required for oscillation at different pumping frequencies, the resonator can be indirectly characterized by the frequency response curve. Starting from the basic principle of parametric oscillation, we will predict the measurable extremities in the frequency-dependent power curve under various circumstances that are classified based on the relative ratio between the lower and higher resonance frequencies. Our analytical models are validated by on-bench measurements performed on several parametric resonators with different circuit topologies. Their ability for remote characterization will make parametric resonators useful in structural health sensors or biomedical implants.

Index Terms: parametric, oscillation, resonator, varactor, frequency-dependent pumping power, coupling

I. INTRODUCTION

It is of fundamental importance to measure the resonance frequency of LC resonators, the basic building blocks of RF electronic circuits. Conventionally, the resonance frequency of an LC resonator can be inductively measured by a pair of partially overlapped loops connected to the output and input ports of a network analyzer. By properly adjusting the overlapped area for minimized interactions between the excitation and receiving loops, peaks in the transmission curve (S21) can be attributed to resonance frequencies of the LC resonator. Because the resonator’s frequency response can be characterized in a contactless way, it can operate as a batteryless transceiver [14] embedded inside a closed environment. When integrated with a sensing capacitor, the LC resonator can encode environmental signals as capacitance changes, which can be measured wirelessly as shift in resonance frequencies. This scheme for telemetric capacitive sensing can be useful in many applications, including structural mechanical monitoring [59], environmental sensing [1018], food quality control [1921] and physiology monitoring [2227]. However, telemetric capacitive sensor is less effective when the distance separation [2830] between the sniffer loops and the LC resonator is much larger than its own dimension, making it challenging to identify peak frequencies of the S21 curve in a noisy background. This problem will be exacerbated when the resonator is embedded inside lossy biological tissues that may further reduce RF transmission efficiency and twist the S21 baseline. Although magneto-inductive repeaters [3134] or waveguides [35] can greatly increase the effective range of wireless sensors, the requirement for intermediate stage devices mandates the use of catheters [36], which is still different from the application scenarios of embedded sensors.

In this work, we are going to demonstrate a compact circuit to increase the effective distance range for resonance frequency measurement. Utilizing the principle of parametric oscillation [37], the pumping signal generates an oscillation signal at half of the frequency that can be easily identified from the noisy background. By adjusting the required pumping power for marginal oscillation over a range of frequencies, the optimal oscillation frequency can be utilized to characterize the resonator’s frequency response. Unlike transistor-based oscillators [3840] that require rectifiers to convert electromagnetic [41] or acoustic energy [42] into DC power, the parametric oscillator has a much simpler circuitry to directly convert the wirelessly provided pumping power into the oscillation signal, enabling more possibilities for further miniaturization.

II. Methods

A. Parametric oscillation

Parametric oscillator can be implemented as a multi frequency resonator (Fig. 1) with at least one nonlinear component (varactor). This nonlinear resonator can utilize its highest resonance mode to receive wirelessly-provided pumping power and convert it into sustained oscillation signals at lower resonance modes. Under the influence of an electromotive force ζ3(t) that is induced inside the resonator by an external pumping loop, there is a voltage V3 induced across the varactor C2 at ω3. Due to frequency mixing, a much smaller signal V1 at ω1 will interact with the strong pumping signal V3 at ω3 across the nonlinear capacitor C2 to generate an idler signal V2 at the difference frequency ω2 = ω3ω1. As a result, the time-dependent capacitance [43] is:

C2(t)=C20(1V(t)ϕ)λC20(1+λ(V1+V2+V3)ϕ) (1)

where Φ is the contact potential of diode junction. λ is normally between 1/3 and 2, which describes the abruptness of charge distribution variation across the diode’s depletion layer. As a result, the total current in the varactor C2 is:

I1+I2+I3=d(C2(t)(V1+V2+V3))dt (2)

Fig. 1.

Fig. 1.

The schematic diagram of a parametric resonator containing a nonlinear varactor (C2). If the circuit’s linear part consists of a Foster network with a frequency dependent impedance R+jS, multiple resonance modes will appear.

The three voltage terms in Eq. (2) can be expressed in the complex forms:

V1=(Vsejω1t+Vs*ejω1t)/2V2=(Viejω2t+Vi*ejω2t)/2V3=(Vpejω3t+Vp*ejω3t)/2 (3)

Similarly, current terms in Eq. (2) can be expressed as:

I1=(Isejω1t+Is*ejω1t)/2I2=(Iiejω2t+Ii*ejω2t)/2I3=(Ipejω3t+Ip*ejω3t)/2 (4)

By plugging Eqns (4) and (3) to both sides of Eq. (2) and retaining frequency components only at ω1, ω2 and ω3:

Isejωt+Is*ejωt+Iiejω2t+Ii*ejω2t+Ipejω3t+Ip*ejω3t=jC20Vsω1ejωt+jC20ω1λVi*VpejωtϕjC20Vs*ω1ejωtjC20ω1λViVp*ejωtϕ+jC20Viω2ejω2t+jC20ω2λVs*Vpejω2tϕjC20Vi*ω2ejφ2tjC20ω2λVsVp*ejω2tϕ+jC20Vpω3ejω3t+jC20ω3λViVsejω3tϕjC20Vp*ω3ejω3tjC20ω3λVi*Vs*ejω3tϕ+jC20ω1λVs2ej2atϕ+jC20ω2λVi2ej2a1tϕjC20ω1λVs*2ej2atϕjC20ω2λVi*2ej2a2tϕ (5)

By comparing the left and right sides of Eq. (5), the following admittance matrix can be obtained.

(Ii*Is)=(jC20ω2jC20ω2λVp*/ϕjC20ω1λVp/ϕjC20ω1)(Vi*Vs) (6)

and subsequently the impedance matrix is

(Vi*Vs)=1C20ω1ω2(ϕ2λ2|Vp|2)(jϕ2ω1jλϕVp*ω2jλϕVpω1jϕ2ω2)(Ii*Is) (7)

Using Kirchhoff’s Voltage Law, the equivalent voltage source ζ1 applied at frequency ω1 can be expressed as

ξ1=Vs+Is(R1+jω1L1)=jλϕVpω1Ii*jϕ2ω2IsC20ω1ω2(ϕ2λ2|Vp|2)+Is(R1+jω1L1)Is(R1+jX1)jλϕVpω1Ii*C20ω1ω2(ϕ2λ2|Vp|2) (8)

where R1+1L1 is the effective circuit impedance excluding the varactor at frequency ω1, R1+jX1 is the effective circuit impedance including the varactor. Similarly, the equivalent voltage source ζ2 at frequency ω2 is

ξ2*=Vi*+Ii*(R2jω2L2)=jϕ2ω1Ii*+jλϕVp*ω2IsC20ω1ω2(ϕ2λ2|Vp|2)+Ii*(R2jω2L2)jλϕVp*ω2IsC20ω1ω2(ϕ2λ2|Vp|2)+Ii*(R2jX2) (9)

where R2+2L2 is the effective circuit impedance excluding the varactor at frequency ω2, R2+jX2 is the effective circuit impedance containing the varactor at frequency ω2. Sustained oscillation requires Is and Ii* to be non-zero when ξ1 = ξ2 = 0. Therefore, the following relation can be derived from Eqns. (8) and (9):

λ2ϕ2|VP|2ω1ω2C202ω12ω22(ϕ2λ2|Vp|2)2(R1+jX1)(R2jX2)=0 (10)

Eq. (10) describes the general requirement of a multi-frequency parametric resonator that can effectively receive the external pumping signal at ω3 and convert it to sustained oscillation at the signal and the idler frequencies (ω1 and ω2). To simplify circuit design, ω1 and ω2 can share the same resonance mode, i.e. ω1 = ω2 = ω3/2, R1 = R2, X1 = X2. When the varactor is undergoing small signal modulation, i.e. λVp/Φ ≪ 1, Eq (10) is simplified to:

|VP|2(R12+X12)C202ω12ϕ2/λ2 (11)

Similar to the derivation of Eq. (6), by comparing terms with ω3 frequency components on the left and right sides of Eq. (5),

Ip=jC20Vpω3+C20ω1λ(Vi+Vs)2/ϕ (12)

The equivalent voltage source ζ3 at frequency ω3 is

ξ3=Vp+Ip(R3+jω3L3)=Vp+(jC20Vpω3+C20ω1λ(Vi+Vz)2/ϕ)(R3+jω3L3)jC20Vpω3(R3+jX3)+C20ω1λ(Vi+Vz)2(R3+jω3L3)/ϕ (13)

where R3+3L3 is the effective circuit impedance excluding the varactor at frequency ω3, R3+jX3R3+3L3+1/(3C3) is the effective circuit impedance containing the varactor. ζ3 is electromotive force induced by a current IL in the pumping loop whose mutual inductance with the parametric resonator is m. Therefore, ξ3=3mIL=j2ω1mIL and Eq. (13) is equivalated to:

IL=C20Vp(R3+jX3)m+C20λ(Vi+Vs)2(R3+jω3L3)2jmϕ (14)

The second term in Eq. (13) originates from sustained oscillation signal, which is normally much smaller than the first term originating from the pumping signal. By neglecting the second term and plugging Eq. (11) into Eq. (14),

|IL|2ω12ϕ2C204R12R32λ2m2(1+X12R12)(1+X32R32)=ω32ϕ2C204R12R324λ2m2(1+4Q12(ω1ω101)2)(1+4Q32(ω3ω301)2) (15)

where ω10 and ω30 are the circuit’s lower and higher resonance frequencies, ω3 is the externally-provided pumping frequency that is twice the oscillation frequency ω1. For operating frequencies ω1 and ω3 that are deviated from resonance frequencies ω10 and ω30, the reactance is X1 ≈ 2R1Q1(ω1 / ω10 −1) and X3 ≈ 2R3Q3(ω3 / ω30 −1), where Q1 and Q3 are circuit’s quality factor near resonance at ω10 and ω30. By defining the relative pumping frequency as fω3 / ω30 and the resonance ratio deviation as k ≡ 2ω10 / ω30 −1, Eq. (15) is rearranged as

|IL|2f2ϕ2C204R12R32ω3024λ2m2(1+4Q12(f(1+k)1)2)(1+4Q32(f1)2) (16)

Quite often, we can construct a double frequency resonator whose higher resonance frequency ω30 is known a prior, Eq. (16) describes the quantitative relation to estimate the lower resonance frequency ω10 (that contains sensing information) from the frequency sweep. At each externally applied pumping frequency ω3, we can scale the required amount of pumping power (that is proportional to |IL|2) by the relative pumping frequency fω3 / ω30. By plotting |IL|2/f2 with respect to f, we can utilize Eq. (16) to estimate k, the resonance ratio deviation that correlates the lower resonance frequency ω10 with the higher resonance frequency ω30 via ω10 = (k + 1)ω30/2.

B. Extreme value analysis

It is possible to perform a complete fitting of Eq. (16), but in the following, we will exploit the possibility to estimate k from the curve’s local extremities by calculating the derivative of the second and third terms on the r.h.s. with respect to f:

8(f1)(1+4Q12(f(1+k)1)2)Q32+8(f(1+k)211+k)(1+4Q32(f1)2)Q12 (17)

To make the derivative zero, fi − 1 ≡ Si has three possible values.

S1=124(12k+(1+3j)62/3A(9B+81B2+6A3)1/3+61/3(1+3j)(9B+81B2+6A3)1/3) (18)
S2=124(12k(1+3j)62/3A(9B+81B2+6A3)1/3+61/3(13j)(9B+81B2+6A3)1/3) (19)
S3=112(6k62/3A(9B+81B2+6A3)1/3+61/3(9B+81B2+6A3)1/3) (20)

where

A(1/Q32+(k+1)2/Q122k2),Bk(1/Q32(k+1)2/Q12). (21)

By calculating the average of Eqns. (18) to (20)

(f1+f2+f33)/3=k/2 (22)

where f1, f2 and f3 correspond to the three local extremities on the right-hand-side of Eq. (16). Eq. (22) provides an important relation to estimate the resonance ratio deviation k from local extremities that are directly measurable from frequency sweep. If the higher resonance frequency ω30 is known a prior and is made to be independent of the sensing capacitor, the lower resonance frequency (that reflects the sensing capacitance) can be estimated from

ω10ω30(1+k)/2=ω30((f1+f2+f3)/31/2). (23)

Of course, estimation of ω10 based on Eq. (23) requires Eqns. (18)–(20) to be real-valued. This will happen only if

81B2+6A30, (24)

where A and B are defined in Eq. (21). As a result, the second and third terms in Eqns. (18)–(20) would cancel their imaginary parts. Because Eq. (24) also implicitly requires A < 0, Eq. (18) can be rearranged to identify the approximate linear relation between Si and k under certain conditions:

S1=124(12k(1+3j)62/3|A|(9B+j6|A|381B2)1/3+61/3(1+3j)(9B+j6|A|381B2)1/3)=124(12k(1+3j)62/3|A|((6|A|3)1/2ejθ)1/3+61/3(1+3j)((6|A|3)1/2ejθ)1/3)=124(12k4cos(πθ)361/2|A|1/2), (25)

where cosθ9B/6|A|3 and sinθ81B2+6|A|3/6|A|3.

Similarly, Eq. (20) is rearranged as

S2=124(12k(+1+3j)62/3|A|(9B+j6|A|381B2)1/3+61/3(13j)(9B+j6|A|381B2)1/3)=124(12k(1+3j)61/2|A|1/2(ejθ)1/361/2(1+3j)|A|1/2(ejθ)1/3)=124(12k4cos(π+θ)361/2|A|1/2). (26)

And Eq. (21) is rearranged as

S3=112(6k+62/3|A|(9B+j(6|A|381B2)1/3+61/3(9B+j(6|A|381B2))1/3)=112(6k+61/3(9Bj(6|A|381B2))1/3+61/3(9B+j(6|A|381B2))1/3)=112(6k+61/2|A|1/2(2)cos(θ/3)). (27)

Further simplifications can be made when Q1 and Q3 are large enough and when 6|A3| ≫ 81B2,

6|A|381B216k427(1/Q32(k+1)2/Q12)20 (28)

This also requires |k| to be sufficiently large to make θ close to π/2. As a result, Eq. (25) can be approximated as

S1124(12k4cos(π6)61/2|A|1/2)=124(12k218(2k21/Q32(k+1)2/Q12))124(12k12|k|) (29)

The last approximation holds when

2k21/Q32+(k+1)2/Q12, (30)

which is a stronger condition than Eq. (28). Similarly, Eq. (27) can be simplified to

S3112(6k+18|2k21/Q32(k+1)2/Q12|1/2)112(6k+6|k|). (31)

Therefore, when ω10 = (k + 1)ω30/2 is much deviated from ω30/2, Eq. (29) is simplified to f1k + 1 (for k <0), and Eq. (31) is simplified to f3k + 1 (for k >0), both of which correspond to an optimal pumping frequency of 2ω10. In another word, the lower resonance frequency ω10 can be directly estimated from half of the optimal pumping frequency, provided that ω10 is sufficiently different from ω30/2.

C. On-bench measurements

First, we use the conventional method to characterize the resonator’s frequency response, by measuring the transmission coefficient (S21) between the excitation and receiving loops that are connected to the output and input ports of a network analyzer. The overlapped area of the loop pair is empirically adjusted until the S21 reached minimum in the absence of resonator, thus minimizing direct interaction between the excitation and receiving loops. When the resonator is placed beneath the pair of loops to inductively couple with them, its frequency response can in principle be identified from the change in S21 curve. But when the resonator is separated from the double pick-up loop by 4 cm, which is 13-fold larger than its own dimension, the change in S21 curve is no larger than the network analyzer’s noise baseline (− 100 dBm for 1 kHz acquisition bandwidth). To improve the measurement sensitivity, a pumping loop is connected to a frequency synthesizer to provide the pumping signal. The pumping frequency ω3 is swept over an anticipated range. At each pumping frequency, the required pumping power is empirically adjusted until marginal oscillation is observed on the network analyzer. To minimize direct interference from the pumping signal, a low-pass filter can be optionally incorporated between the receiving loop and the input port of a network analyzer. In the following, the same measurement method will be applied to three resonators with different design features. For each case, the resonator will be sufficiently separated from the receiving loop to emulate deep-lying capacitive sensors that would introduce unobservable change in S21 curve by conventional measurement method. Subsequently, pumping power will be gradually increased until the resonator marginally oscillates as observed on the network analyzer. Because the observation frequency is half of the pumping frequency, the receiving signal can be easily separated from the excitation signal, enabling more sensitive characterization of the resonator’s frequency response beyond the near-field region.

III. RESULTS

A. Fabricated resonator #1 (One frequency extremity)

Fig. 3a shows the circuit diagram of a nonlinear double frequency resonator. It is constructed by wrapping a continuous copper wire around a 3.6-mm diameter cylindrical rod for two turns and around a rectangular rod with a 1.5 × 2.5 mm2 cross-section for four turns (Fig. 3b). The cylindrical rod and the rectangular rod are placed perpendicular to each other, so that the two constituting inductors L1 and L2 can effectively couple to the external pumping field from different orientations. C1 is a 2.4-pF chip capacitor, and C2 is a varactor (BBY51, Infineon, Germany) to perform signal mixing. When the pumping frequency is swept over the range between 606 MHz and 634 MHz at 1-MHz interval, there is only one optimal pumping frequency at 622 MHz, which doesn’t correlate to 2ω10 in a straightforward manner. This is because the resonator’s higher resonance frequency ω30 is not sufficiently separated from twice the lower resonance frequency 2ω10, which is evidenced by direct measurement of ω10 to be 307.48 MHz (Q1 = 57) and ω30 to be 624.6 MHz (Q3 = 67). Based on these measured values, the resonance deviation ratio is k = 2ω10/ω30-1=−0.0154, which will make A(1/Q32+(k+1)2/Q122k2)>0. Because the left-hand side of Eq. (24) is larger than 0, the resonator is indeed predicted to have only one optimal pumping frequency whose value is described by Eq. (20). This example is chosen to demonstrate the condition for single-valued frequency extremity. In the following, we will illustrate two more examples where ω10 is easier to estimate based on its simpler relation with the measurable frequency extremities.

Fig. 3.

Fig. 3.

(a) The circuit diagram and (b) the picture of a double-frequency parametric resonator, where the varactor C2 is placed in parallel with the inductor L2. (c) For each integral frequency point swept between 606 MHz and 634 MHz, the required pumping power for marginal oscillation is scaled by f2 (which is the square of the relative frequency) and displayed in units of dBw. The measured power levels (shown in dots) indeed correlate well with the simulated results (shown in continuous line) predicted by Eq. (16).

B. Fabricated resonator #2 (Three frequency extremities)

Fig. 4a is the circuit diagram of another implementation of a parametric resonator. This circuit consists of a parallel resonant circuit (L1 and C1) placed in series with L2 and C2, where C2 is the varactor diode to perform frequency mixing. The resonator is constructed by wrapping a continuous copper wire around a rectangular rod with a 1.6 × 1.6 mm2 cross-section for two turns and around a 1.8-mm diameter cylindrical rod for three turns (Fig. 4b). C1 is a 5.1-pF chip capacitor and C2 is a varactor (BBY52, Infineon, Germany). These component values are chosen so that twice the lower resonance frequency 2ω10 is sufficiently separated from the higher resonance frequency ω30. When the pumping frequency is swept, the required pumping power for marginal oscillation has two local minima at 1013 MHz and 1032 MHz and one local maxima at 1024 MHz (Fig. 4c). Therefore, the average of these three frequency extremities is calculated to be 1023 MHz. According to Eq. (23), the lower resonance and higher resonance frequencies should satisfy the relation ω10 = 1023 − ω30/2. This simple relation is indeed confirmed by direct measurement of ω10 to be 517 MHz (Q = 84) and ω30 to be 1012 MHz (Q = 103). In another word, it is possible to estimate the lower resonance frequency from the higher resonance frequency along with the three extremities of the frequency-dependent pumping curve.

Fig. 4.

Fig. 4.

(a) The circuit diagram and (b) the picture of a double-frequency parametric resonator, where the varactor C2 is placed in series with the inductor L2. (c) For each integral frequency point swept between 1010 MHz and 1035 MHz, the required pumping power for marginal oscillation is scaled by f2 (which is the square of the relative frequency) and displayed in units of dBw. The measured power levels (shown in dots) indeed correlate well with the simulated results (shown in continuous line) predicted by Eq. (16).

C. Fabricated resonator #3 (Three frequency extremities with a higher resonance frequency ω30 that remains invariant.)

Up till now, we have presented two examples to demonstrate the necessary conditions to observe three frequency extremities on the frequency response curve. We have also validated Eq. (23) that describes the quantitative relation between these three extremities and the lower resonance frequency ω10. Here, we will present another circuit design that can be practically used for capacitive sensing, by making the higher resonance frequency ω30 independent of changes in ω10.

Figs. 5a shows the circuit diagram of a cylindrical parametric resonator mounted on a plastic cylinder. Its higher resonance mode responsible for receiving the pumping signal is created by the top end-ring split by a pair of identical varactors, so that the higher resonance frequency ω30 is only determined by the diameter of the top end-ring and the zero-biased capacitance of varactors C2. Meanwhile, the circuit’s lower resonance mode responsible for sustaining the oscillation signal is created by the vertical legs along with the top and bottom end-rings. If the capacitors C1 on the bottom end-rings are sensing capacitors, changes in the lower resonance frequency ω10 will be indicative of changes in C1, without affecting the higher resonance frequency ω30. Because ω30 can be measured a prior and will remain constant, it can effectively function as an internal frequency reference based on which the lower resonance frequency ω10 can be estimated from Eq. (23). Moreover, estimation of ω10 will be even simpler if k ≡ 2ω10 / ω30 −1 is large enough to make Eq. (31) valid, so that ω10 is approximately half of the frequency extremity value. This simpler relation is validated on the circuit implemented in Fig. 5b, where the cylindrical resonator has a diameter of 2.5 mm and a length of 5.9 mm. By splitting the resonator’s top end-ring with a pair of identical varactors (PMEG2005EB, NXP, Netherland) and the bottom end-ring with a pair of 39-pF chip capacitors, the resonator has a transverse resonance frequency at ω10 = 299.7 MHz (Q = 60) and a longitudinal resonance frequency at ω30 = 550.3 MHz (Q = 23). By sweeping the pumping frequency over the range between 592 MHz and 605 MHz, the optimal pumping frequency is 599 MHz (Fig. 8c), which is indeed very close to 2ω10. In principle, we can also measure all the three frequency extremities. But because 2ω10 is too far away from ω30, the local maximum that is approximately half way between 2ω10 and ω30 would require an impractically high level of pumping power for oscillation.

Fig. 5.

Fig. 5.

(a) The circuit diagram and (b) the picture of a cylindrically-symmetric parametric resonator. In this design, a pair of varactors C2 are placed around the top end-ring whose resonance frequency is independent of the capacitors at the bottom end-ring. (c) For each integral frequency swept between 592 MHz and 605 MHz, the required pumping power for marginal oscillation is scaled by f2 (which is the square of the relative frequency) and displayed in units of dBw. The measured power levels (shown in dots) indeed correlate well with the simulated results (shown in continuous line) predicted by Eq. (16).

D. Measurements under different distance separations

Based on the above examples, it is easier to evaluate the resonance behavior of resonators #2 and #3 whose higher resonance frequencies ω30 are sufficiently separated from twice their lower resonance frequencies 2ω10. Here, we measure the frequency extremities of these two resonators when their distance separation from the external pumping loop are gradually increased. For each distance separation, we measure the frequency extremities five times and calculate the averages:

As shown in Table 1, highly consistent frequency extremities are obtained over a range of distance separations, even at a distance separation of 16 cm that is more than 60-fold larger than the resonator’s own width. This large distance separation is already larger than the half-thickness of human torso, indicating the possibility to use parametric resonators as physiological sensors deep inside the body.

Table 1.

Frequency extremities measured over several distance separations

D Resonator #2 (MHz) # 3 (MHz)
4 cm 1012.0±0.0 1024.0±0.0 1032.0±0.0 599.0±0.0
8 cm 1012.0±0.0 1024.0±0.0 1032.0±0.0 599.0±0.0
12 cm 1011.8±0.4 1023.8±0.4 1032.4±0.5 598.8±0.4
16 cm 1012.2±0.5 1023.6±0.5 1032.2±0.4 599.2±0.4

IV. Discussion

In this work, we demonstrate the use of parametric oscillation to increase the effective interaction between an LC resonator and the sniffer loop. Three versions of parametric oscillators are fabricated based on nonlinear multi-frequency resonators. Utilizing the nonlinearity of their constituting varactors, these resonators can convert wirelessly provided pumping signal at ω3 into sustained oscillation signal at ω3/2. By sweeping the frequency of the pumping signal, we can measure the minimal power for marginal oscillation and obtain local extremities along the frequency-dependent power curve. When the resonator is constructed in a way so that its lower resonance frequency ω10 reflects sensing capacitance and its higher resonance frequency ω30 remains invariant, it is possible to estimate ω10 from ω30 using Eq. (23), given that three extremities are measurable along the frequency-dependent curve. Estimation of lower resonance frequency ω10 can be further simplified if 2ω10 is distantly separated from ω30 to make Eq. (31) valid, so that ω10 can be approximated as half of the optimal pumping frequency.

To oscillate a parametric resonator, it is energy-efficient to have the higher resonance frequency approximately twice the lower resonance frequency, i.e. k = 2ω10/ω30−1 ~ 0, so that both quadratic terms in the right-hand side of Eq. (16) become zero. On the other hand, when |k| is very large to make Eq. (31) valid, the optimal pumping frequency is approximately twice the lower resonance frequency, i.e. fr–1=k, making the right-hand side of Eq. (16) (1+4Q32k2)-fold larger than the condition when k = 0. Therefore, the convenience to directly estimate ω10 from half of the optimal pumping frequency is obtained at a cost of larger pumping power. In order to estimate the resonance frequency with better energy efficiency, it is also possible to fabricate the resonator in a way to make the left side of Eq. (24) not too smaller than 0, so that three local extremities can be obtained within a moderate frequency range. In this case, although estimation of ω10 is contingent upon prior knowledge about ω30, there are potential applications where the relative change of ω10 is a direct indication of the relative change in capacitance, so that long-term variation of sensing parameters can be obtained without the need for measuring ω30 a prior.

Conventional methods use a pair of loops to measure the resonator-induced perturbation on power transmission curve (S21) between the excitation and receiving loops. If the mutual coupling between the loops and the resonator is much small than the direct coupling from the excitation to the receiving loops, it is very hard to measure the resonator’s frequency response, because the resonator’s impact on the S21 curve is buried below the noise baseline. However, because parametric resonator can generate oscillation signal that is different from the input signal, emitted signals from the resonator can stand above the background, enabling resonance characterization at much larger distance separations. Of course, a larger distance separation necessitates a larger oscillation signal, which corresponds to a larger value for the second term in Eq. (14). But when the circuit has large enough quality factor, i.e. R3ω3L3, this second term in Eq. (14) only has a slow-varying frequency component 3L3, making its contribution to the |IL|2/f2 curve almost frequency independent. In another word, even though the resonator is operating above marginal oscillation, as long as its oscillation signal is maintained at a constant level over the frequency sweep, the position of frequency extremities will not be affected by the marginal oscillation approximation. The validity of this approximation is evident from the negligible changes in frequency extremities measured under different distance separations (Table 1). Besides the marginal oscillation approximation, another possible source of systematic error is the assumption for small signal modulation of the varactor, i.e. λV/Φ ≪ 1, leading to the approximation described in Eq. (1) and (11). This assumption is self-consistent because the required pumping voltage is Vp ~ Φ/(λQ3) according to Eq. (11). If the resonator’s quality factor Q3 is sufficiently large, λV/Φ ≪ 1 can always be ensured.

Experimentally, a larger source of error may come from the positional variation of the resonator during measurement. Because the required pumping power is measured at individual frequency points, it is important to hold the resonator static with respect to the pumping loop throughout the frequency sweep. For real-world sensors operating in a moving environment (due to physiological motion for example), it is important develop external control apparatus to automatically adjust the pumping frequency and to simultaneously track the oscillation signal so that frequency sweep can be finished within a shorter period compared to the motion time scale. More work is going on towards this direction.

V. Conclusion

In this work, an analytical framework is proposed to improve the effective distance range for resonance frequency characterization based on wirelessly powered parametric oscillator. Three different circuits have been constructed to demonstrate the versatile design of double-frequency parametric resonators and to validate analytical predictions under different resonance conditions. Instead of using multiple excitation signals to interrogate the circuit’s resonance behavior [4447], a parametric oscillator is simple to operate with a single signal source. Since the oscillation frequency is different from the pumping frequency, it is easy to measure the oscillation signal in both the time domain and frequency domain, thus alleviate special requirements on the external read-out circuitry [3, 4851].

Fig. 2.

Fig. 2.

The frequency response of a parametric oscillator activated by an external pumping loop can be characterized beyond near-field regions, for example, at a distance separation of at least 4 cm, which is more than 13-fold larger the resonator’s width.

Acknowledgments

This work was supported in part by the National Institutes of Health (NIH) under Grant number R00EB016753.

Biography

Wei Qian received her B.S. and M.S. degrees from Wuhan University, Wuhan, China, in 2003 and 2006, respectively, and a Ph.D. degree from Michigan State University, East Lansing, in 2011, all in electrical engineering. From 2011 to 2016, she was a Power Electronics Engineer in Magna Electronics (the former E-Car Systems). She is now an assistant professor at Michigan State University. Her research interests include dc-dc converters, dc-ac inverters, vehicle electrification and energy harvest.

Chunqi Qian received his BS degree in chemistry from Nanjing University and his PhD in physical chemistry from the University of California, Berkeley, in 2007. Following postdoctoral trainings at the National High Magnetic Field Laboratory and the National Institutes of Health, he became a principle investigator in the department of Radiology in Michigan State University. His research interest includes the development and application of RF sensor technology for biomedical research.

Contributor Information

Wei Qian, Department of Electrical and Computer Engineering, Michigan State University, East Lansing, MI, 48824, USA..

Chunqi Qian, Department of Radiology, Michigan State University, East Lansing, MI, 48824, USA..

References

  • [1].Collins CC, “Miniature Passive Pressure Transensor for Implanting in Eye,” Ieee Transactions on Biomedical Engineering, vol. Bm14, pp. 74–+, 1967. [DOI] [PubMed] [Google Scholar]
  • [2].Huang QA, Dong L, and Wang LF, “LC Passive Wireless Sensors Toward a Wireless Sensing Platform: Status, Prospects, and Challenges,” Journal of Microelectromechanical Systems, vol. 25, pp. 822–841, October 2016. [Google Scholar]
  • [3].Nopper R, Niekrawietz R, and Reindl L, “Wireless Readout of Passive LC Sensors,” Ieee Transactions on Instrumentation and Measurement, vol. 59, pp. 2450–2457, September 10 2010. [Google Scholar]
  • [4].Zhai J, How TV, and Hon B, “Design and modelling of a passive wireless pressure sensor,” Cirp Annals-Manufacturing Technology, vol. 59, pp. 187–190, 2010. [Google Scholar]
  • [5].Cheng Y, Gao F, Hanif A, Lu Z, and Li ZJ, “Development of a capacitive sensor for concrete structure health monitoring,” Construction and Building Materials, vol. 149, pp. 659–668, September 15 2017. [Google Scholar]
  • [6].Butler JC, Vigliotti AJ, Verdi FW, and Walsh SM, “Wireless, passive, resonant-circuit, inductively coupled, inductive strain sensor,” Sensors and Actuators a-Physical, vol. 102, pp. 61–66, December 1 2002. [Google Scholar]
  • [7].Nabipoor M and Majlis BY, “A new passive telemetry LC pressure and temperature sensor optimized for TPMS,” International Mems Conference 2006, vol. 34, pp. 770–775, 2006. [Google Scholar]
  • [8].Matsuzaki R and Todoroki A, “Wireless flexible capacitive sensor based on ultra-flexible epoxy resin for strain measurement of automobile tires,” Sensors and Actuators a-Physical, vol. 140, pp. 32–42, October 1 2007. [Google Scholar]
  • [9].Kisic MG, Blaz NV, Babkovic KB, Maric AM, Radosavljevic GJ, Zivanov LD, et al. , “Passive Wireless Sensor for Force Measurements,” Ieee Transactions on Magnetics, vol. 51, January 2015. [Google Scholar]
  • [10].Ong KG and Grimes CA, “A resonant printed-circuit sensor for remote query monitoring of environmental parameters,” Smart Materials & Structures, vol. 9, pp. 421–428, August 2000. [Google Scholar]
  • [11].Horton BE, Schweitzer S, DeRouin AJ, and Ong KG, “A Varactor-Based, Inductively Coupled Wireless pH Sensor,” Ieee Sensors Journal, vol. 11, pp. 1061–1066, April 2011. [Google Scholar]
  • [12].Ando B, Baglio S, Pitrone N, Savalli N, and Trigona C, ““Bent beam” MEMS Temperature Sensors for Contactless Measurements in Harsh Environments,” 2008 Ieee Instrumentation and Measurement Technology Conference, Vols 1–5, pp. 1930–1934, 2008. [Google Scholar]
  • [13].Wang LF, Ma JJ, Huang Y, Tang D, and Huang QA, “Passive Wireless Hermetic Environment Monitoring System for Spray Painting Workshop,” Sensors, vol. 16, August 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Harpster TJ, Stark B, and Najafi K, “A passive wireless integrated humidity sensor,” Sensors and Actuators a-Physical, vol. 95, pp. 100–107, January 1 2002. [Google Scholar]
  • [15].Harpster TJ, Hauvespre S, Dokmeci MR, and Najafi K, “A passive humidity monitoring system for in situ remote wireless testing of micropackages,” Journal of Microelectromechanical Systems, vol. 11, pp. 61–67, February 2002. [Google Scholar]
  • [16].Ong JB, You ZP, Mills-Beale J, Tan EL, Pereles BD, and Ong KG, “A Wireless, Passive Embedded Sensor for Real-Time Monitoring of Water Content in Civil Engineering Materials,” Ieee Sensors Journal, vol. 8, pp. 2053–2058, Nov-Dec 2008. [Google Scholar]
  • [17].Fernandez-Salmeron J, Rivadeneyra A, Rodriguez MAC, Capitan-Vallvey LF, and Palma AJ, “HF RFID Tag as Humidity Sensor: Two Different Approaches,” Ieee Sensors Journal, vol. 15, pp. 5726–5733, October 2015. [Google Scholar]
  • [18].Mannoor MS, Tao H, Clayton JD, Sengupta A, Kaplan DL, Naik RR, et al. , “Graphene-based wireless bacteria detection on tooth enamel (vol 3, 763, 2012),” Nature Communications, vol. 4, May 2013. [DOI] [PubMed] [Google Scholar]
  • [19].Ong KG, Bitler JS, Grimes CA, Puckett LG, and Bachas LG, “Remote query resonant-circuit sensors for monitoring of bacteria growth: Application to food quality control,” Sensors, vol. 2, pp. 219–232, June 2002. [Google Scholar]
  • [20].Tan EL, Ng WN, Shao R, Pereles BD, and Ong KG, “A wireless, passive sensor for quantifying packaged food quality,” Sensors, vol. 7, pp. 1747–1756, September 2007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Potyrailo RA, Nagraj N, Tang ZX, Mondello FJ, Surman C, and Morris W, “Battery-free Radio Frequency Identification (RFID) Sensors for Food Quality and Safety,” Journal of Agricultural and Food Chemistry, vol. 60, pp. 8535–8543, September 5 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Weber J, Lagua D, Williamson A, Husar P, and Schober A, “Capacitive sensor concept for monitoring neuronal activity in vitro,” Biomed Tech, vol. 57, pp. 1055–1058, 2017. [Google Scholar]
  • [23].Chatzandroulis S, Tsoukalas D, and Neukomm PA, “A miniature pressure system with a capacitive sensor and a passive telemetry link for use in implantable applications,” Journal of Microelectromechanical Systems, vol. 9, pp. 18–23, March 2000. [Google Scholar]
  • [24].Sauer S, Marschner U, Adolphi B, Clasbrummel B, and Fischer WJ, “Passive Wireless Resonant Galfenol Sensor for Osteosynthesis Plate Bending Measurement,” Ieee Sensors Journal, vol. 12, pp. 1226–1233, May 2012. [Google Scholar]
  • [25].Leonardi M, Pitchon EM, Bertsch A, Renaud P, and Mermoud A, “Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes,” Acta Ophthalmol, vol. 87, pp. 433–7, June 2009. [DOI] [PubMed] [Google Scholar]
  • [26].Puers R, Vandevoorde G, and De Bruyker D, “Electrodeposited copper inductors for intraocular pressure telemetry,” Journal of Micromechanics and Microengineering, vol. 10, pp. 124–129, June 2000. [Google Scholar]
  • [27].Allen MG, “Micromachined endovascularly-implantable wireless aneurysm pressure sensors: From concept to clinic,” Transducers ‘05, Digest of Technical Papers, Vols 1 and 2, pp. 275–278, 2005. [Google Scholar]
  • [28].Zhang X and Chae J, “Working Distance Comparison of Inductive and Electromagnetic Couplings for Wireless and Passive Underwater Monitoring System of Rinsing Process in Semiconductor Facilities,” Ieee Sensors Journal, vol. 11, pp. 2932–2939, November 2011. [Google Scholar]
  • [29].Schantz HG, “Near field propagation law & a novel fundamental limit to antenna gain versus size,” IEEE International Symposium on Antennas and Propagation, vol. 3A, pp. 237–240, 2005. [Google Scholar]
  • [30].Marioli D, Sardini E, Serpelloni M, and Taroni A, “A new measurement method for capacitance transducers in a distance compensated telemetric sensor system,” Measurement Science and Technology, vol. 16, pp. 1593–1599, August 2005. [Google Scholar]
  • [31].Dong L, Wang LF, Zhang C, and Huang QA, “A Cyclic Scanning Repeater for Enhancing the Remote Distance of LC Passive Wireless Sensors,” Ieee Transactions on Circuits and Systems I-Regular Papers, vol. 63, pp. 1426–1433, September 2016. [Google Scholar]
  • [32].Zhang C, Wang LF, and Huang QA, “Extending the remote distance of LC passive wireless sensors via strongly coupled magnetic resonances,” Journal of Micromechanics and Microengineering, vol. 24, December 2014. [Google Scholar]
  • [33].Dong L, Wang LF, and Huang QA, “A Passive Wireless Adaptive Repeater for Enhancing the Readout of LC Passive Wireless Sensors,” Ieee Microwave and Wireless Components Letters, vol. 26, pp. 543–545, July 2016. [Google Scholar]
  • [34].Sanz DA, Mitrosbaras C, Unigarro EA, and Segura-Quijano F, “Passive resonators for wireless passive sensor readout enhancement,” Applied Physics Letters, vol. 103, September 23 2013. [Google Scholar]
  • [35].Syms R, Shamonina E, and Solymar L, “Magneto-inductive waveguide devices,” IEE PROCEEDINGS-MICROWAVES ANTENNAS AND PROPAGATION, vol. 153, pp. 111–121, April 2006. [Google Scholar]
  • [36].Ahmad MM, Syms RRA, Young IR, Mathew B, Casperz W, Taylor-Robinson SD, et al. , “Catheter-based flexible microcoil RF detectors for internal magnetic resonance imaging,” Journal of Micromechanics and Microengineering, vol. 19, July 2009. [Google Scholar]
  • [37].Collin RE, “Foundations for Microwave Engineering,” pp. 541570, 1966. [Google Scholar]
  • [38].Wang Y, Goh WL, Lee JH, Chai KTC, and Je M, “Resonant-Based Capacitive Pressure Sensor Read-Out Oscillating at 1.67 GHz in 0.18 μm CMOS,” International Journal of Electronics and Communication Engineering, pp. 812–815, 2013. [Google Scholar]
  • [39].Piso D, Veiga-Crespo P, and Vecino E, “Modern Monitoring Intraocular Pressure Sensing Devices Based on Application Specific Integrated Circuits,” Journal of Biomaterials and Nanobiotechnology, vol. 3, pp. 301–309, 2017. [Google Scholar]
  • [40].Turner M and Naber J, “The Development of a RFID Based Mixed Signal ASIC for the Wireless Measurement of Intraocular Pressure,” 2010 18th Biennial University/Government/Industry Micro-Nano Symposium, 2010. [Google Scholar]
  • [41].Islam MM, Rasilainen K, and Viikari V, “Implementation of Sensor RFID: Carrying Sensor Information in the Modulation Frequency,” Ieee Transactions on Microwave Theory and Techniques, vol. 63, pp. 2672–2681, August 2015. [Google Scholar]
  • [42].Kim A, Powell CR, and Ziaie B, “An Implantable Pressure Sensing System With Electromechanical Interrogation Scheme,” Ieee Transactions on Biomedical Engineering, vol. 61, pp. 2209–2217, July 2014. [DOI] [PubMed] [Google Scholar]
  • [43].Pozar DM, “Microwave engineering,” p. 537, 2012. [Google Scholar]
  • [44].Pichorim SF and Abatti PJ, “A novel method to read remotely resonant passive sensors in biotelemetric systems,” Ieee Sensors Journal, vol. 8, pp. 6–11, Jan-Feb 2008. [Google Scholar]
  • [45].Viikari V, Seppa H, and Kim DW, “Intermodulation Read-Out Principle for Passive Wireless Sensors,” Ieee Transactions on Microwave Theory and Techniques, vol. 59, pp. 1025–1031, April 2011. [Google Scholar]
  • [46].Song JS, Salmi J, Viikari VV, and Pesonen NMP, “Maximum-Likelihood Estimation for Passive Wireless Intermodulation Communication Sensors,” Ieee Sensors Journal, vol. 15, pp. 2280–2286, April 2015. [Google Scholar]
  • [47].Song JS, Viikari V, Pesonen N, Marttila I, and Seppa H, “Optimization of Wireless Sensors Based on Intermodulation Communication,” Ieee Transactions on Microwave Theory and Techniques, vol. 61, pp. 3446–3452, September 2013. [Google Scholar]
  • [48].Bao KK, Chen DY, Shi Q, Liu LJ, Chen J, Li J, et al. , “A readout circuit for wireless passive LC sensors and its application for gastrointestinal monitoring,” Measurement Science and Technology, vol. 25, August 2014. [Google Scholar]
  • [49].Peterson BJ, Olson AV, and Kaiser TJ, “A Wireless Sensor Interrogator Design for Passive Resonant Frequency Sensors Using Frequency Modulation Spectroscopy,” Ieee Sensors Journal, vol. 10, pp. 1884–1890, December 2010. [Google Scholar]
  • [50].Sajeeda and Kaiser TJ, “Passive telemetric readout system,” Ieee Sensors Journal, vol. 6, pp. 1340–1345, October 2006. [Google Scholar]
  • [51].Jacquemod G, Nowak M, Colinet E, Delorme N, and Conseil F, “Novel architecture and algorithm for remote interrogation of battery-free sensors,” Sensors and Actuators a-Physical, vol. 160, pp. 125–131, May 2010. [Google Scholar]

RESOURCES