Dear Sir,
Unanticipated behaviours of the Airseal Insufflation and Access System (Conmed, Utica, NY, USA), in the public domain since 2018 [1], have recently been restated by the manufacturer [2] in the light of the COVID‐19 pandemic and widespread concerns regarding aerosolization hazards during surgery [3]. Video S1 illustrates this device’s tendency for intra‐abdominal gas effluvium to be continually blown into the operating room during use as well as it's phenomenon of air entrainment (i.e. the tendency for room air to be sucked into the abdomen) via the device during high pressure intra‐operative suctioning.
We used a combination of assessment technologies in a high‐fidelity simulation model (fresh porcine cadaver) as well as during clinical surgery to examine gas flow through the Airseal 12‐mm valveless trocar with the Airseal IFS carbon dioxide (CO2) insufflator in Airseal mode. Schlieren Imaging (an optical imaging technology that identifies differences in gas densities) as well as a specific near‐infrared CO2 visualization system (FLIR GF343; Flir Systems Ltd, Kent, ME, UK) were used to dynamically visualize gas flow around the trocar. A specific laparoscopic nebulizer (Aeroneb Solo; Aerogen, Galway, Ireland) enabled abdominal gas and droplet egress visualization by transillumination in a darkened room [4]. A flowmeter (TSI Series 5000; TSI Inc., Shoreview, MI, USA) measured directional velocity of flow just outside of the trocar.
By these methods, a continuous CO2 plume flowing at a rate of 1–2 l/min is shown arising from the trocar in Airseal mode. This continues, albeit to a lesser extent, with placement of the cap packaged along with the trocar. The nebulization model shows this gas vortex contains unfiltered gas exiting from the peritoneum. This flow reverses with suctioning (whether via a separate port or the Airseal trocar itself) reaching a peak inflow velocity in these tests of 8 l/min. At the same time, the pressure of the pneumoperitoneum is relatively maintained without compensatory increase in Airseal insufflation consistent with room air being sucked directly into the abdomen via the valveless trocar.
The video illustrates a mechanism of direct operating room pollution by intra‐abdominal gas and associated airborne particles, corroborated by direct clinical observation of the CO2 flue during live surgery [5]. This accounts for some of the excess CO2 consumption seen with use of Airseal insufflation [6]. Contamination of the intra‐abdominal space with non‐medical grade air during suctioning dilutes the purity of the CO2 pneumoperitoneum risking infection, combustion and persistence of gas beyond pure CO2 timeframes (prolonging interstitial emphysema or hollow viscus distension). It may also hazard air embolism during surgery (including transanal total mesorectal excision [7,8]).
Conflicts of interest
RC receives speaker fees from Stryker (visualisation), consultancy fees from Touch Surgery and Distal Motion, and holds research funding from Intuitive Corporation and with IBM Ireland (from the Irish Government). JD, FZ and KN have no conflicts of interest to declare.
Supporting information
Acknowledgements
The authors gratefully acknowledge the following equipment loans: high speed Phantom camera for Schlieren Imaging by Bell Labs, Ireland; the FLIR GF 343 from Mr Steve Beynon, Flir Systems Ltd, UK, supplied with thermographic training by Mr David Doyle, Butler Technologies, Ireland; the TSI series 5000 flowmeter supplied with expertise by Mr John O’Dea, Palliare, Galway, Ireland, and the Aeroneb Solo with expertise by Mr Barry Russell, Aerogen, Galway, Ireland.
References
- 1. Lexion Medical, LLC . Jury Awards $10 million in punitive damages against surgiquest, a subsidiary of CONMED corporation for false advertising in dispute with lexion medical. 2017. Available at https://www.prnewswire.com/news‐releases/lexion‐medical‐receives‐12‐3‐million‐on‐july‐10–2018–of‐which‐10‐million‐was‐a‐punitive‐damage‐award‐from‐surgiquest‐a‐subsidiary‐of‐conmed‐the‐manufacturer‐of‐the‐airseal‐insufflator‐for‐false‐advertising‐300679577.html (accessed 20 August 2020).
- 2. Conmed Advanced Surgical . Airseal SAGES Response. Conmed Advanced Surgical.. 2020. Available at https://www.conmed.com/‐/media/CONMED/Documents/AS%20Education/MC20191599_SAGES_Response_Customer_Letter.pdf (accessed 20 August 2020).
- 3. Francis N, Dort J, Cho E et al. SAGES and EAES recommendations for minimally invasive surgery during COVID‐19 pandemic. Surg Endosc 2020; 34: 2327–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Khan MF, Dalli J, Cahill RA. Gas aerosol jetstreams from trocars during laparoscopic surgery. Colorectal Dis 2020; 22: 1028–9. 10.1111/codi.15215. [DOI] [PubMed] [Google Scholar]
- 5. Cahill R, Dalli J, Khan MF, Flood M, Nolan K. Solving the problems of gas leakage at laparoscopy. Br J Surg 2020. (Accepted). 10.1002/bjs.11977. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Horstmann M, Horton K, Kurz M, Padevit C, John H. Prospective comparison between the AirSeal® System valve‐less Trocar and a standard Versaport™ Plus V2 Trocar in robotic‐assisted radical prostatectomy. J Endourol 2013; 27: 579–82. [DOI] [PubMed] [Google Scholar]
- 7. Dickson EA, Penna M, Cunningham C et al. Carbon dioxide embolism associated with transanal total mesorectal excision surgery: a report from the international registries. Dis Colon Rectum 2019; 62: 794–801. [DOI] [PubMed] [Google Scholar]
- 8. Huntington CR, Prince J, Hazelbaker K et al. Safety first: significant risk of air embolism in laparoscopic gasketless insufflation systems. Surg Endosc 2019; 33: 3964–9. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.