Abstract
Although Raman spectroscopy is a major analytical tool in modern chemical experiments, commercial Raman spectrometers remain very pricey for educational and research purposes in individual university laboratories. Thus, this study focused on the structural similarity between the Raman spectrometer and an optical pickup unit (OPU), which is an inexpensive compact optical device used for a part of optical discs. The study investigated whether or not a full set of Raman spectrometer can be developed at a cost of less than 1,000 US$. The OPU-based Raman spectrometer was fabricated using 3D printer-made components, a Raman edge filter, and a laser diode with a wavelength of 520 nm as the light source. A function generator was used as a pulsed power source to analyze the characteristics of the OPU Raman spectrometer according to various frequencies and duty ratios. When using a pulsed DC power supply, the laser wavelength tended to move to a longer wavelength with increases in duty ratios. That is, the higher the frequency at the same duty ratio, the weaker the background light intensity compared with the scattered Raman signal intensity. The findings illustrate that Raman signal strength can be adjusted by adjusting the focal length of the objective lens of the OPU through an external adjustment of an additional DC power. In the Raman spectra of all solid and liquid samples used, the maximum error rate reached approximately 11 cm−1, whereas the maximum intensity deviation reached approximately ± 6%. The cost of the complete OPU Raman spectrometer is less than 1,100 US$ using a function generator as power source and less than 930 US$ using a DC adapter. If the optical density (OD) 6 filter can be replaced with the OD 4 filter, then the costs are expected to decrease to approximately 730 US$.
Subject terms: Chemistry, Materials science
Introduction
Raman spectrometers are very powerful tools used for the simple and rapid identification of chemical species in chemical-related laboratories. In particular, it is considered one of the essential spectroscopic methods along with the research and development of low-dimensional carbon compounds, such as carbon nanotubes, graphene, and carbon dots1–6. Since C. V. Raman discovered Raman scattering in 19287, various Raman spectrometers have been developed after extensive research and improvement. Light scattering can be classified into Mie and Rayleigh scatterings dependent on the size of scattering materials8. Mie scattering is a phenomenon in which the light of all colors is scattered evenly regardless of the wavelength due to particle sizes that are larger than the incident wavelengths of lights. When the size of the scattering matter is much smaller than the wavelength of light, the difference lies in the degree of scattering according to the wavelength of the incident light. The second form, namely, Rayleigh scattering, causes the sky to appear blue. Most photon incidents on a molecule undergo Rayleigh scattering at the same wavelength as the incident light. However, a few photons scatter inelastically, where the wavelength of the scattered light changes due to the interaction with the molecules. When such a Raman scattering occurs, the wavelengths of the scattered light may decrease or increase, and such a scattering is called Stokes Raman scattering or anti-Stokes Raman scattering, respectively (Fig. 1). Anti-Stokes Raman scattering consistently holds lesser intensity than Stokes Raman scattering because it is a phenomenon that occurs in molecules in the excitation state. Therefore, most Raman spectrometers are manufactured using the Stokes Raman scattering, which occurs in ground-state molecules. When Raman scattering occurs, the wavelength of the incident light shifts due to the difference in the intrinsic vibrational energy of the molecule. Each molecule has a specific vibrational energy, such that this frequency difference has its own value. Therefore, the composition or structure of a chemical can be confirmed through Raman scattered light. Figure 2a provides the structure of a simplified Raman spectrometer.
Raman spectrometers have been commercialized in various means dependent on the wavelength of laser light sources, optical paths, or slightly different arrangements in the interior space. However, commercialized Raman spectrometers are difficult to use in individual university labs or educational purposes because of their prices. To solve the cost problem, various studies have been conducted to directly manufacture a Raman spectrometer at a low price9–12. Mohr et al. studied a Raman spectrometer and assembled optical components and a green laser point at 532 nm applicable to undergraduate education and research purposes9. In addition, Rossi et al. studied a Raman spectrometer using a commercially available camera and various optical filters12. Although previous studies reported cheaper prices than commercially available Raman spectrometers, cost and structural problems remained due to problems in the light path control and use of expensive spectrometers.
The optical pickup unit (OPU) used in the optical disk in Fig. 2b is an extremely small and precise optical device. Many studies have applied it as an optical device, such as an optical microscope13, for measuring the thermal vibration of a microcantilever for atomic force microscopy14. The OPU is also useful for refractive index measurement at the femtoliter scale15 and a biological sensor at the nano- and micro-scale levels16 due to its various functions. It has also been applied in micropatterning devices17, optical fluid scanning to detect protein aggregation in microfluidics18, and DNA replication using magnetic nano-beads19. As previously described, the OPU has been applied to various optical devices and research, but it has been rarely used on a Raman spectrometer. In fact, the structure of the OPU has a large similarity with that of a Raman probe. Therefore, the present study fabricated a Raman spectrometer using the OPU to overcome the aforementioned problems, such as difficulty in controlling optical paths and cost.
Results and discussion
Figure 3 shows the process of manufacturing the Raman spectrometer using the OPU (KSS-213C) for a compact disc player. Figure 3 shows that the OPU Raman was manufactured using a laser diode (LD), essential optical parts, and connecting parts made of 3D printers, while maintaining the optical path, lens, and focal length adjustment functions of the OPU. We analyzed the OPU Raman spectrometer using various electrical signals formed by the function generator as the power source (see the Experiments section and Supplementary Information for detailed descriptions of the fabrication method). The LD is frequently operated by a direct current power source. However, we controlled the electrical signal to reduce the power consumption of the LD, which leads to a reduced risk of eye damage. In addition, this feature can enhance resolution and reproducibility by reducing heat formation during the operation of LD, which causes a wavelength shift in the laser. Toward this end, the Raman spectrometer was analyzed using a square-pulsed DC made from a function generator as the power source. First, the frequency was fixed at 100 Hz to analyze the effect of the duty ratio on the OPU Raman, and the duty ratio was adjusted from 10 to 80% to analyze the shift in the center wavelength of the laser and benzene spectrum according to each duty ratio.
After the measurement, the horizontal axis indicated by the wavelength was changed in wavenumber units based on the laser wavelength. Raman spectra were acquired after baseline fitting using MagicPlot Student (ver. 2.7.2, Magicplot Systems, LLC; see Supplementary Information on how to correct the baseline). Figure 4a shows that the center wavelength of the laser increases with the increase in the duty ratio. This result is considered a phenomenon that occurs as the current amount increases, whereas the internal temperature of the LD increases with the increase in the duty ratio. Although this case differs from the LD, similar wavelength changes have been reported in light-emitting diodes in terms of the duty ratio20.
The graph in Fig. 4b denotes the Raman spectra of benzene obtained with the increase in the duty ratio. When comparing the position of the 992 cm−1 peak as the reference peak, the intensity of the peak gradually increased with the increase in the duty ratio, and the position of the peak indicated a shift similar to the wavelength shift of the laser.
To understand the characteristics of the OPU Raman according to the frequency, the duty ratio was fixed. By contrast, the frequency was varied to measure the change in the center wavelength of the LD and reference peak position of benzene. The deformation of the voltage waveform was observed at a high frequency. Thus, the frequency dependency of the OPU Raman was analyzed within a range of 500 Hz or less.
First, after setting the duty ratio to 10%, frequency was increased by 50 Hz to measure the change in the center wavelength of the laser. Figure 4c illustrates that the laser wavelength had slight changes with the increase in the frequency. In contrast to the Raman spectrum of benzene, which is dependent on the duty ratio, the main peak of benzene at 992 cm−1 was nearly unchanged with the increase in the frequency (Fig. 4d). However, although the position and relative intensity of the Raman peak were the same, the intensity of the Rayleigh scattering was reduced under the high-frequency conditions. These results indicate that Raman scattering intensity increases at the same laser power when measured at high frequencies at the same duty ratio.
Any device using an OPU has the advantage of focusing through vertical and horizontal distance adjustments when voltage is applied to the coil around the objective lens of the OPU. The study investigated the effect of focal length adjustment by controlling the applied voltage using a variable resistor (100 Ω) and a battery (3 V). Experiments were conducted using liquid benzene and the solid powder form of anthracene. The sample was placed at a distance similar to the known focal length of the objective lens (an aspheric lens with a focal length of 3.86 mm, working distance of 2.1 mm, and 0.45 numerical aperture), and the Raman spectrum according to resistance change was analyzed. Figure 5a shows changes in the Raman spectra of liquid benzene with the increase in resistance, where little change was observed within a certain focal length. Conversely, in the case of anthracene, a nonuniform solid sample was noted in Fig. 5b, where the intensity of the peak at 1,402 cm−1 varies depending on the resistance value. After passing the critical resistance (25 Ω), the figure shows that nearly the same intensity of the peak is obtained. This result shows that the focal length of the OPU Raman can be adjusted by simply adjusting the external voltage for the coil of the OPU.
Considering the characteristics of the OPU Raman obtained in the basic experiment, Raman spectra were obtained for solid (anthracene and naphthalene) and liquid (benzene and acetone) samples. Afterward, Raman spectra were analyzed after a baseline correction of the spectrum for each sample. The Raman spectrum was obtained for the liquid sample under the following conditions: 400 Hz, duty ratio of 10%, and 20,000 ms integral time. Figure 6a illustrates the baseline corrected Raman spectrum of benzene, and Table 1 provides data on the major peaks of the known Raman peaks of benzene21. The spectrum shows a difference of up to 6 cm−1 compared with known Raman peaks. In the case of acetone, the Raman signal was weakly measured at a low duty ratio and under a duty cycle of 40%, 100 Hz, and 15,000 ms integral time. After measurement, a baseline correction of the Raman spectrum was carried out after the correction of the laser wavelength measured at a 40% duty ratio. A difference of up to 3 cm−1 was observed compared with each reference peak position in Fig. 6b22. For solid samples, measurements were performed under 400 Hz, duty ratio of 10%, and 6,000 ms integration time. Raman spectra were obtained and analyzed after the baseline adjustment in the same manner as the liquid samples. The spectrum for anthracene in Fig. 7a shows a difference of up to 7 cm−1 at each major peak compared with that of the reference23. In addition, naphthalene in Fig. 7b shows a positional change of up to 11 cm−1.
Table 1.
Parts | Supplier (or where to buy) |
Price (USD) |
---|---|---|
OPU (KSS-213C) | Smartech Electronics and Machinery Manufacturing Co., LTD | 4 |
520 nm laser (PL520) | Osram opto semiconductors | 20 |
Raman Edge Filter (532 nm) |
Edmund Optics, 12.5 mm, OD 6 or Replaceable with OD 4 |
445 245 |
Collimator lens (PMMA) | Shenzhen Tyson Tech. Co | < 1 |
Optical fiber (200 µm) | Thunder Optics | 47 |
Spectrometer (R spectrometer) | Thunder Optics | 350 |
Printed parts (PLA 500 g) | Plasil | ~ 10 |
Function generator (AG051) |
Owon or a DC adapter |
199 30 |
Others | Variable Resistor, 1.5 V Battery, Adhesive, etc | ~ 20 |
The reproducibility of the OPU Raman system was tested using anthracene, which has a relatively high Raman signal among the samples. The spectrum was obtained five times under the same condition of 400 Hz, duty ratio of 10%, and an integral time of 10,000 ms. After each measurement, the OPU Raman system was powered off, and Raman spectra were measured at intervals of 10 min. Figure 8a illustrates the spectra measured after each interval, whereas Fig. 8b provides a graph of the deviation (absolute value) of the position and intensity changes between spectra for the main peak, that is, 1,402 cm−1. The maximum deviation of the peak position is measured within ± 3 cm−1, whereas the peak intensity is measured within ± 6%. Considering the resolution of the commercial Raman spectrometer at 1 nm (10 cm−1), the OPU Raman system displays reasonable reproducibility.
The total cost used to manufacture the OPU Raman system is calculated as approximately 1,100 US$ or less when using a function generator as power source or an optical density (OD) 6 Raman edge filter. The cost was lower at approximately 930 US$ or less when a DC adapter is used to replace the power supply. If the OD 6 filter can be replaced with the OD 4 filter, then the costs are expected to reach approximately 730 US$. Based on the results, the OPU Raman system is expected to be widely built and utilized even in an educational research environment where budget is largely insufficient.
In summary, for the development of the Raman spectrometer under 1,000 US$, the study explored a Raman spectrometer consisting of the commercial OPU and a pulsed DC power. As a result of the experiment, Raman spectra with reasonable resolutions and indicated error rates of up to 11 cm−1 were obtained from the liquid and solid samples. In addition, the focal length of the OPU objective lens demonstrated controllability, and Raman signal intensity can be increased by controlling the voltage applied to the coil through a variable resistance adjustment. When the pulsed DC signal is used as power, the LD wavelength tended to move to a longer wavelength with the increase in the duty ratio. The higher the frequency at the same duty ratio, the weaker the intensity of the background signal compared with the Raman signal. The reproducibility of the OPU Raman spectrometer showed a maximum deviation in the peak position within ± 3 cm−1 with a peak intensity of ± 6%. This result showed excellent properties even when compared with a commercial Raman system using a 1 nm (10 cm−1) spectrometer. In terms of cost, the proposed system can be produced within a range of 730 US$ to 1,100 US$. Thus, the OPU Raman spectrometer is expected to be widely used in a low-budget education and research environment.
Methods
Except for the frame and electromagnet coils of the OPU (KSS-213C), other parts, such as LD, circuit connections, covers, and a beam splitter, were removed. Subsequently, a 520 nm LD, collimator lens, Raman edge filter, and reflective mirror were bonded to the OPU frame. A 3D printer (CreatBot F160) was used to build the component holder or connections, an OPU Raman support, and sample holder (see CAD design in Supplementary Information). Finally, the power source for adjusting the focal length was connected, and the optical fiber was connected to a spectrometer to complete it. The components used for the fabrication of the OPU Raman were a 520-nm wavelength LD (PL 520, 50 mW, OSRAM), Raman edge filter with OD 6 (Ø12.5 mm, 537.3 nm, Raman edge filter, Edmund Optics), optical fiber (SMA-type, 200-µm core diameter), and Ag dot mirror with a diameter of 3 mm formed on a 0.1-mm-thick glass substrate using a silver mirror reaction. For solid samples, a sample container with a glass cover was prepared using a 3D printer and a piece of slide glass. For liquid samples, the sample was placed in a small vial with a diameter of 1.2 cm (to minimize air bubbles during sampling) and measured through the glass surface of the bottom of the vial. Benzene (Samchun, 99.5%) and acetone (Daejung, 99.8%) as liquid samples and naphthalene (Daejung, 97%) and anthracene (Daejung, 97%) as solid samples were measured. The Raman signal was measured by changing the duty ratio (10% to 80%) and frequency (50–500 Hz) using a function generator (Owon, AG051F) as the power source for the LD. The optical spectrometer (R spectrometer, Thunder Optics) was calibrated based on the positions of 992 and 3,062 cm−1, and the characteristic Raman peaks of benzene using least squares fitting were built into the software and provided with the optical spectrometer. Subsequently, the wavelength of the LD was measured according to the duty ratio and frequency characteristics using an ultraviolet–visible spectrometer (K-mac, 2100 V). The programs used to build and measure the OPU Raman spectrometers are all freeware: FreeCAD (ver. 0.17, Open Source), Ultimaker Cura for 3D printers (ver. 3.1.0, Ultimaker BV), and MagicPlot Student (ver. 2.7.2, MagicPlot Systems, LLC, for the baseline correction of Raman spectra).
Supplementary information
Acknowledgements
This research was supported by Basic Science Research Program through the National Research Foundation (No. 2018R1A5A1025594 and No. 2019R1A2C1006771) by the Ministry of Science, ICT of Korea.
Author contributions
S.I.A. designed the experiments. S.I.A. and Y.C. conducted the experiments and analyzed the data. S.I.A. and Y.C. drafted the manuscript. Both authors discussed the results and commented on the manuscript.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
is available for this paper at 10.1038/s41598-020-68650-7.
References
- 1.Ferrante C, et al. Raman spectroscopy of graphene under ultrafast laser excitation. Nat. Commun. 2018;9:1–8. doi: 10.1038/s41467-017-02508-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Nanda SS, et al. A new device concept for bacterial sensing by Raman spectroscopy and voltage-gated monolayer graphene. Nanoscale. 2019;11:8528–8537. doi: 10.1039/C8NR10010A. [DOI] [PubMed] [Google Scholar]
- 3.Kato H, Itagaki N, Im HJ. Growth and Raman spectroscopy of thickness-controlled rotationally faulted multilayer graphene. Carbon N. Y. 2019;141:76–82. doi: 10.1016/j.carbon.2018.09.017. [DOI] [Google Scholar]
- 4.Chernyak SA, et al. Oxidation, defunctionalization and catalyst life cycle of carbon nanotubes: a Raman spectroscopy view. Phys. Chem. Chem. Phys. 2017;19:2276–2285. doi: 10.1039/C6CP04657F. [DOI] [PubMed] [Google Scholar]
- 5.Hummer T, et al. Cavity-enhanced Raman microscopy of individual carbon nanotubes. Nat. Commun. 2016;7:1–7. doi: 10.1038/ncomms12155. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Dervishi E, Ji Z, Htoon H, Sykora M, Doorn SK. Raman spectroscopy of bottom-up synthesized graphene quantum dots: Size and structure dependence. Nanoscale. 2019;11:16571–16581. doi: 10.1039/C9NR05345J. [DOI] [PubMed] [Google Scholar]
- 7.Raman CV, Krishnan KS. A new type of secondary radiation. Nature. 1928;121:501–502. doi: 10.1038/121501c0. [DOI] [Google Scholar]
- 8.McCreery RL. Raman Spectroscopy for Chemical Analysis. Hoboken: Wiley; 2005. [Google Scholar]
- 9.Mohr C, Spencer CL, Hippler M. Inexpensive Raman spectrometer for undergraduate and graduate experiments and research. J. Chem. Educ. 2010;87:326–330. doi: 10.1021/ed800081t. [DOI] [Google Scholar]
- 10.Bandyopadhyay AK, Dilawar N, Vijayakumar A, Varandani D, Singh D. A low cost laser-Raman spectrometer. Bull. Mater. Sci. 1998;21:433–438. doi: 10.1007/BF02744931. [DOI] [Google Scholar]
- 11.Ismail N, et al. Raman spectroscopy with an integrated arrayed-waveguide grating. Opt. Lett. 2011;36:4629. doi: 10.1364/OL.36.004629. [DOI] [PubMed] [Google Scholar]
- 12.Rossi EM, Aurelio A-L, Óscar B-E. A homemade cost effective Raman spectrometer with high performance. J. Lab. Chem. Educ. 2015;3:67–75. [Google Scholar]
- 13.Lee YC, Chao S. A compact and low-cost optical pickup head-based optical microscope. IEEE Trans. Magn. 2014;50:1–4. [Google Scholar]
- 14.Ulčinas A, Picco LM, Berry M, Heinrich Hörber JK, Miles MJ. Detection and photothermal actuation of microcantilever oscillations in air and liquid using a modified DVD optical pickup. Sens. Actuators A Phys. 2016;248:6–9. doi: 10.1016/j.sna.2016.07.007. [DOI] [Google Scholar]
- 15.Dunn RC. Compact, inexpensive refractive index detection in femtoliter volumes using commercial optical pickup technology. Anal. Methods. 2019;11:2303–2310. doi: 10.1039/C9AY00369J. [DOI] [Google Scholar]
- 16.Te Hwu EE, Boisen A. Hacking CD/DVD/Blu-ray for Biosensing. ACS Sens. 2018;3:1222–1232. doi: 10.1021/acssensors.8b00340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Lee Y-C, Chao S, Huang C-C, Cheng K-C. A compact optical pickup head in blue wavelength with high horizontal stability for laser thermal lithography. Opt. Express. 2013;21:23556. doi: 10.1364/OE.21.023556. [DOI] [PubMed] [Google Scholar]
- 18.Uddin, R. et al. Integration of agglutination assay for protein detection in microfluidic disc using blu-ray optical pickup unit and optical fluid scanning. in MicroTAS 2015 - 19th Int. Conf. Miniaturized Syst. Chem. Life Sci. 1807–1809 (2015).
- 19.Donolato M, et al. Quantification of rolling circle amplified DNA using magnetic nanobeads and a Blu-ray optical pick-up unit. Biosens. Bioelectron. 2015;67:649–655. doi: 10.1016/j.bios.2014.09.097. [DOI] [PubMed] [Google Scholar]
- 20.Moreira MC, Prado R, Campos A. Application of High Brightness LEDs in the Human Tissue and Its Therapeutic Response. Appl. Biomed. Eng. 2011;1(InTech):3–20. [Google Scholar]
- 21.Wahadoszamen, M., et al. Laser Raman spectroscopy with different excitation sources and extension to surface enhanced raman spectroscopy. J. Spectrosc.2015 (2015).
- 22.Cossee P, Schachtschneider JH. Vibrational analysis of acetone, acetaldehyde, and formaldehyde. J. Chem. Phys. 1966;44:97–111. doi: 10.1063/1.1726509. [DOI] [Google Scholar]
- 23.Abasbegović N, Vukotić N, Colombo L. Raman spectrum of anthracene. J. Chem. Phys. 1964;41:2575–2577. doi: 10.1063/1.1726325. [DOI] [Google Scholar]
- 24.Nedungadi TMK. Analysis of the raman spectrum of single crystals of naphthalene. Proc. Indian Acad. Sci. Sect. A. 1942;15:376–389. doi: 10.1007/BF03046032. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.