Abstract
Corrosion at modular junctions of total hip replacement (THR) remains a major concern today. Multiple types of damage modes have been identified at modular junctions, correlated with different corrosion characteristics that may eventually lead to implant failure. Recently, within the head-taper region of the CoCrMo retrieval implants, cell-like features and trails of etching patterns were observed that could potentially be linked to the involvement of cells of the periprosthetic region. However, there is no experimental evidence to corroborate this phenomenon. Therefore, we aimed to study the potential role of periprosthetic cell types on corrosion of CoCrMo alloy under different culture conditions, including the presence of CoCrMo wear debris. Cells were incubated with and without CoCrMo wear debris (obtained from a hip simulator) with an average particle size of 119±138nm. Electrochemical impedance spectroscopy (EIS) was used to evaluate the corrosion tendency, corrosion rate, and corrosion kinetics using the media after 24h of cell culture as the electrolyte. Results of the study showed that there was lower corrosion resistance (p<0.02) and higher capacitance (p<0.05) within cell media from macrophages challenged with particles when compared to the other media conditions studied. The potentiodynamic results were also in agreement with the EIS values, showing significantly higher corrosion tendency (low Ecorr) (p<0.0001) and high Icorr (p<0.05) in media from challenged macrophages compared to media with H2O2 solution. Overall, the study provides in-vitro experimental evidence for the possible role of macrophages in altering the chemical environment within the crevice and thereby accelerating corrosion of CoCrMo alloy.
Keywords: macrophages, periprosthetic cells, corrosion, CoCrMo, electrochemical impedance spectroscopy (EIS), modular junction, THR
Graphical Abstract
This study has successfully established an in-vitro test protocol to characterize the cell accelerated corrosion behavior of CoCrMo alloy in conditioned cell culture media of different periprosthetic cell types. Macrophages activated with wear particles had a significant effect on the increased corrosion rate and the overall corrosion kinetics. Thus, macrophages can alter the electrochemical environment that drives the passivation and preferential dissolution processes at the microstructure level of the alloy, thus accelerating damage caused by corrosion.
1.0. Introduction
Fretting and corrosion within modular taper junctions of total hip replacements (THR) is a persisting problem in orthopedics. Adverse local tissue reactions (ALTR) due to tribocorrosion debris is recognized as a frequent cause of premature implant failure1–3. Reports of the prevalence of ALTR related failures vary between 2 and 4%4,5. Considering the increasing number of THRs performed annually, it is important to reduce this number by preventing the occurrence of fretting and corrosion processes. A unified model of fretting and corrosion within modular junctions has been described as mechanically assisted crevice corrosion (MACC)6. The crevice environment is given by the interface between the head and stem taper surfaces. The mechanical aspect is primarily provided by micro-motion that leads to fretting wear and corrosion. Yet, retrieval studies have shown that the resulting damage cannot be explained by one single damage mode6–9. Multiple underlying damage modes occur which can be either mechanically or chemically driven9. Many in vitro fretting studies have been conducted to simulate in vivo fretting processes10–13. To our knowledge, no study was able to recreate all damage features that have been observed in vivo. Specifically, chemically dominated damage features such as etching, pitting, intergranular corrosion, and phase boundary corrosion have not been reproduced using a fretting apparatus. Therefore, it appears that the chemistry of the environment is a significant factor that needs to be considered for in vitro testing. Gilbert et al. presented a model of inflammatory cell-induced corrosion suggesting that reactive oxygen species released by macrophages or other cell types may cause corrosion damage to the implant surface14. However, so far there is little evidence that macrophages or other cell types of the joint environment could significantly contribute to the corrosive process within THR modular taper junctions.
The hypothesis of this study was that among periprosthetic cell types, macrophages will have the most dominant impact on the generation of a corrosive environment under inflammatory conditions. Hence, the major goal of this study was to generate in vitro experimental evidence for cell-accelerated corrosion by creating in vivo-like conditions in lab-scale and develop an in vitro model for the evaluation of the molecular mechanisms of cell-mediated corrosion.
2.0. Materials and methods
2.1. Overview of sample groups
The investigation of the corrosion behavior of CoCrMo alloy under normal and inflammatory media conditions was carried out by means of electrochemical analysis. Three different cell types (cell lines) from the periprosthetic environment—monocytes, macrophages, and osteoblasts—were used in this study. Normal medium condition (conditioned medium) was the medium collected after 24 h of cell culture (monocytes, macrophages, or osteoblasts). Inflammatory medium conditions were generated by challenging cell cultures with CoCrMo wear debris. The viability of cells under inflammatory conditions was evaluated by cytotoxicity and particle uptake analysis. Cell culture media from both normal and inflammatory conditions were employed as an electrolyte in an electrochemical corrosion chamber. The corrosion behavior of CoCrMo alloy within different electrolytes was determined by open circuit potential (OCP) measurements, cyclic potentiodynamic polarization tests, and electrochemical impedance spectroscopy (EIS) (Figure 1).
Figure 1:
Schematic representation of the overall study outline.
2.2. Sample Material and Preparation
Low carbon wrought CoCrMo alloy (Carpenters Ltd, CA, USA) as specified by the ASTM15 was used to prepare sample discs 12 mm in diameter with a thickness of 7 mm. These discs were used for both the corrosion tests and the generation of wear particles to challenge cells in culture. For the corrosion tests, the sample surface was mechanically ground using 320 grit silicon carbide grinding paper. This rougher surface finish was chosen to resemble head taper surfaces15 more closely. Prior to testing all samples were ultrasonically cleaned in 75% 2-propanol and distilled water for a duration 15 minutes each. For wear particle generation, the prepared discs were polished to a mirror finish.
2.3. Wear particle generation
In order to generate inflammatory conditions in the cell culture tests CoCrMo particles with a known size range were added to the different types of cell cultures. These particles were generated using a tribocorrosion test rig with a pin-on-ball configuration16 (Figure 2a). The experimental conditions were previously developed and optimized by Mathew et al17,18 (See Supplementary file). In order to determine the wear particle size range and chemical composition, dynamic light scattering (DLS; Zeta sizer, Nano ZS, Melvern instruments GmbH, Germany) and scanning electron microscopy (SEM, Hitachi SU8030 FE-SEM, Tokyo, Japan) with energy dispersive x-ray spectroscopy (EDS) were employed.
Figure 2:
a) Schematic representation of the tribocorrosion hip-simulator and b) corrosion apparatus connected to the potentiostat through a three-electrode system, where the CoCrMo disc is the working electrode fully immersed in the electrolyte.
2.4. Cell Culture
2.4.1. Cell lines
The THP-1 human monocytes cell line (ATCC, Virginia, USA) was cultured in RPMI 1640 supplemented with L-Glutamine, penicillin & streptomycin (100IU/ml), 25 mM Hepes buffer and 10% fetal bovine serum. Cells were subcultured in 1:3 ratio after 80% confluency. RAW 264.7 (mouse macrophage cell line), and MG63 (human osteosarcoma cell line) were obtained from ATCC. Both RAW 264.7 and MG-63 cell lines were maintained in DMEM supplemented with L-glutamine and (Gibco, USA), 10% FBS (Sigma Aldrich, USA) and 100IU/ml penicillin–streptomycin under standard conditions. Cells were detached from the culture plate at 80–85% confluency. Once confluent, they were passaged in a 1:3 ratio.
2.4.2. Cytotoxicity
The cells (monocytes, macrophages, and osteoblasts) were seeded onto 96 well plates at a seeding density of 5000 cells/well. Wear particles at concentrations in the range of 10 g/ml to 0.1 g/ml were added to cell cultures for 24 h of incubation. In a previous study by Bijukumar et al., we observed that concentrations above 10 g/ml caused a reduction in cell growth to less than 40% in the case of neuronal cells19. Hence, in this experiment, we selected concentrations from 10 g/ml to 0.1 μg/ml to evaluate their effect on the cells of interest in this study. The aim of this test was to identify the concentration of wear particles showing lower growth inhibition (IC50) after 24h compared to control in order to challenge the cells. After 24 h, the cells were incubated with 10% alamarBlue (Invitrogen, USA) in complete medium for 4 h. After incubation, the optical density at 570nm with 600nm as reference was recorded using a microplate.
2.4.3. Determination of particle uptake
To study the cellular uptake of particles 50,000 RAW, THP1 and MG63 cells were seeded in 12 well plates. Each type of cells was treated with 1 μg/ml of wear particles. Controls of each type were maintained containing the same number of cells in the plain medium. All experiments were conducted in triplicate. The plates were incubated for 24h, after which the cells were trypsinized, collected, and vortexed to obtain a single-cell suspension. The percentage of the cell population that had taken up the particles were analyzed using a flow cytometer (BD FACS Calibur, USA). In the flow cytometer, a laser light beam of 499nm illuminated the cells in the sample stream. The forward and side scatter light intensities were considered to be proportional to the size of the cells and the granularity of the cells, respectively.
2.5. pH measurement
The pH of the media collected for the experiments were measured just before the electrochemical experiments. The control medium, medium with wear particles (freshly prepared) and the medium with H2O2 (freshly prepared) were taken from the refrigerator and warmed to 37° C in a water bath using a closed container. Conditioned medium and medium from the cells challenged with wear particles after 24h (already in 37 °C incubator) were collected in a closed container, and the pH was also measured just before the electrochemical testing. The time duration between the collection of samples and pH measurements was 15 to 20 minutes.
2.6. Electrochemical tests
CoCrMo samples were mounted onto a custom-made poly sulfonate multi-well corrosion chamber which exposed an area 0.95 cm2 of the sample surface20 (Figure 2b). Alloy samples were connected to a single channel of an electrochemical multiplexer, which was controlled by a Gamry interface 1000E potentiostat in a 3-electrode configuration. The setup included a reference electrode (SCE), a counter electrode (graphite rod), and the CoCrMo alloy sample served as the working electrode. The sequence of analysis included the monitoring of the open circuit potential (OCP), a potentiostatic scan (PS), repeated OCP monitoring, electrochemical impedance spectroscopy (EIS), cyclic potentiodynamic polarization (CP) and a final monitoring of the OCP. Measurements were conducted at 37 °C using different electrolytes. Electrolytes included cell culture media used for the three different types of cell cultures conducted under normal and inflammatory conditions. Additionally, several controls were used: 1) pure cell culture medium, 2) cell culture medium that was incubated with cells for 24h (conditioned medium), 3) cell culture medium that was incubated for 24h with the addition of 1μg/ml of wear particles, 4) cell culture medium with a 0.03M concentration of H2O2 (Sigma Aldrich, USA). The latter was tested as it was suggested to be one of the active agents released by macrophages under inflammatory conditions14,21.
The potentiodynamic test was performed from −0.8 V to 1.8 V vs. SCE with a scan rate of 1mV/s. After the experiment, OCP, EIS, and potentiodynamic polarization data were used to analyze the corrosion tendency, corrosion rate, and corrosion kinetics. The corrosion tendency (Icorr) and corrosion potential (Ecorr) were obtained from the potentiodynamic data. EIS tests were performed at OCP at the frequency from 100 kHz to 0.01 Hz with an amplitude of ± 10mV. A modified Randle’s circuit was used as an equivalent electrical circuit to model the EIS data from which resistance to polarization (Rp) and capacitance (Cdl) were obtained. All experiments were repeated three times (N=3), and all media used in the three experiments were generated in three independent cell culture experiments. Normality of the data from electrochemical tests was confirmed with QQ-plots. If data was normal, a one-way ANOVA was used with Bonferroni correction (significance level p=0.05). If data was not normal, a Kruskal-Wallis test was conducted, followed by a Mann-Whitney U test for pairwise comparison against control.
3.0. Results
3.1. Wear particle characterization
The DLS spectrum showed that the CoCrMo wear particles ranged from the nanoscale to the microscale with an average particle size of 119 ± 138.4 nm (Figure 3a) with a median value of a 164.2nm and interquartile range (Q3-Q1) of 399.93nm (1st Quartile - 58.77 and 3rd Quartile – 458.7). Sixty-five percent of particles were under 150nm in diameter. SEM and EDS analysis of larger particles and/or particle agglomerates revealed the presence of both Co and Cr species within the particles (Figure 3b, c).
Figure 3:
a) DLS spectrum showing the particle size range of CoCrMo wear debris generated from the hip simulator. b) Scanning electron microscopic image of wear debris. c) Results of the EDS analysis of CoCrMo wear debris to confirm the presence of cobalt and chromium. The results are qualitatively due to the fact that lighter elements such as carbon and oxygen cannot be accurately quantified by this EDS system19. However, there appeared to be a fairly large amount of oxygen indicating the presence of metal oxides, likely chromium oxide. The overall dominant presence of cobalt compared to chromium indicates that the majority of debris is still cobalt-based alloy. The presence of carbon, sodium, and chlorine indicates the presence of organic residue and salt from the wear testing solution.
3.2. Cytotoxicity and wear particle uptake analysis
Cell cultures were incubated with different wear particle concentrations (10 μg/ml, 5 μg/ml, 1 μg/ml, 0.1 μg/ml) for a period of 24 h in order to optimize the dose required to challenge cell cultures without causing necrosis due to toxicity. The results showed that higher concentrations of wear debris (5 and 10 μg/ml) were toxic to all tested cell types after 24 h. In the case of 10 g/ml concentration of wear particles, cell viability was reduced by 48.9 ± 2.6% (p=0.0008), 13.9 ± 2.1% (p=0.025), and 39.2 ± 1.41% (p=0.005) in osteoblasts, macrophages, and monocytes respectively. Similarly at 5 g/ml the reduction in cell viability was observed as 31.4 ± 2.3% (p=0.001), 14.2 ± 4.1% (p=0.112), and 30.5 ± 2.5% (p=0.02) in osteoblasts, macrophages, and monocytes respectively. There was no significant toxicity observed at 0.1 μg/ml [osteoblasts 95.1 ± 3.2 % (p=0.125), macrophages 100.9 ± 2.2 % (p=0.55) and monocytes 97.33 ± 6.1% (p=0.66)] and 1 μg/ml [osteoblasts 88.5 ±5.7 % (p=0.125), macrophages 104.9 ± 5.5 % (p=0.55) and monocytes 97.1 ± 2.4% (p=0.27)] (Figure 4a). Furthermore, flow cytometry confirmed the cell uptake of wear particles at 1 μg/ml concentration (Figure 4b). The side scatter (Q4) vs. forward scatter (Q2+Q3) plots illustrated that cellular granularity increased over time, which suggest that the cells had readily taken up the wear particles. However, from the cytogram and the corresponding bar graph, it was clear that particle uptake was higher in macrophages compared to monocytes and osteoblasts. In addition, the forward scatter of macrophages was significantly (p <0.001 for challenged monocytes vs. challenged macrophages and p< 0.0004 for challenged osteoblasts vs. challenged macrophages) higher suggesting that the increase in cell size occurred due to the presence of wear debris.
Figure 4:
a) Graph showing cytotoxicity of wear particles to osteoblasts, macrophages, and monocytes at different concentrations after 24 h. Data are presented as Mean ± SD. b) Dot plot obtained from particle uptake by osteoblasts, macrophages, and monocytes after 24 h incubation with wear debris (concentration: 1g/ml) using flow cytometry compared to –ve control (cells without particle treatment). c) Corresponding graph showing side scatter events by the cells. Data are presented as Mean ± SD. *p<0.05. N=3 and all experiments were repeated three times.
3.3. Electrochemical analysis
3.3.1. pH variation
The pH of the different media such as control medium, DMEM medium from 24h cell culture of macrophages, and DMEM medium with H2O2 was analyzed just before the experiment using a pH meter (Table 1). All media conditions except medium collected after 24 h of incubation of macrophages with wear particles had an average pH in the range of 7.30± 0.07 (conditioned medium) to 7.62±.0.01 (DMEM with 0.03M H2O2). However, in the case of challenged macrophages, the pH value was slightly acidic at 6.74 ± 0.14 (Table 1).
Table 1:
pH values from different media after incubation (media without cells) or 24h cell culture (medium with cells) and before electrochemical analysis. The media includes control, medium with DP, conditioned medium, and challenged medium from macrophages, monocytes, and osteoblasts.
| Samples | Control Medium (G1) | Medium with DP(G2) | Medium with H2O2(G3) | Conditioned Medium(G4) | Challenged Medium Treated with 1 μg/ml of DP | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| DMEM medium, stored in the fridge and warmed to 37°C | Medium, stored in the fridge, warmed to 37°C and treated with 1 μg/ml of DP | Medium stored in the fridge and warmed to 37°C and added 0.03M H2O2 | 24 h | ||||||||
| Macrophages (G5) | Monocytes (G6) | Osteoblasts (G7) | |||||||||
| DMEM | RPMI | DMEM | RPMI | DMEM | RPMI | DMEM | RPMI | DMEM | RPMI | DMEM | |
| N1 | 7.62 | 7.26 | 7.61 | 7.36 | 7.63 | 7.35 | 7.22 | 7.33 | 6.79 | 7.23 | 7.39 |
| N2 | 7.60 | 7.35 | 7.54 | 7.32 | 7.61 | 7.36 | 7.34 | 7.35 | 6.58 | 7.31 | 7.52 |
| N3 | 7.59 | 7.31 | 7.59 | 7.31 | 7.64 | 7.35 | 7.35 | 7.39 | 6.87 | 7.21 | 7.35 |
| AVG | 7.60 | 7.34 | 7.58 | 7.33 | 7.62 | 7.35 | 7.30 | 7.35 | 6.74 | 7.25 | 7.42 |
| STDEV | 0.01 | 0.02 | 0.03 | 0.02 | 0.01 | 0.005 | 0.07 | 0.03 | 0.14 | 0.05 | 0.08 |
AVG-Average, STDEV-Standard Deviation, G1-G7- Group1-Group7
3.3.2. Open circuit potential (OCP)
The evolution of the OCP as a function of time differed broadly depending on the media (Figure 5). The potential of the alloy in the presence of control medium and the medium collected after 24 h of cell culture under normal conditions ranged from −0.4 V to −0.5 V vs. SCE. However, the OCP measured in media collected after 24 h incubation of cells (macrophages, monocytes, and osteoblasts) under inflammatory conditions was reduced, ranging from −0.7V to −0.8V vs. SCE. There was no statistical significance between the challenged media groups (osteoblasts vs. macrophages p=0.265, osteoblasts vs. monocytes p=0.55, macrophages vs. monocytes p=0.06). However compared to control medium and conditioned medium, challenged macrophage medium showed a significant reduction in OCP value (p<0.05 and p<0.008 respectively). No significance was observed with monocytes medium and osteoblasts medium in comparison to control and conditioned media (control vs. challenged monocytes, p= 0.06 and challenged vs. osteoblasts p=0.421; conditioned medium vs. challenged monocytes, p=0.06 and conditioned medium vs. challenged osteoblasts p=0.33).
Figure 5:
OCP of CoCrMo alloy samples within different media. H2O2 had the highest OCP, whereas media from cells cultured under inflammatory conditions had the lowest OCP. N=3 and the experiments were repeated three times.
3.3.3. Electrochemical impedance spectroscopy (EIS)
Representative Nyquist (real part of impedance (Zreal) vs. imaginary part of impedance (Zimg)) and Bode plot (Impedance modulus (IZI value and phase angle vs. frequency) obtained from the EIS data for CoCrMo alloy under different media conditions are shown in Figure 6a & b. The Nyquist plots revealed a higher impedance for all the control media conditions studied. Whereas, for the media from different challenged cell cultures (macrophages, monocytes, and osteoblasts), the impedance values of CoCrMo alloy were lower, with macrophages challenged medium showing the lowest value. Challenged monocytes medium also provided a lower impedance value compared to control medium, but higher than osteoblasts and macrophages. In addition, medium with H2O2 also provided lower impedance compared to control media studied, suggesting high corrosion kinetics of CoCrMo alloy in the presence of an inflammatory medium. The Bode plot also revealed consistent results with low phase angles and impedance vs. frequency values with challenged macrophages medium (Figure 6b).
Figure 6:
a) Nyquist and b) Bode plot of CoCrMo alloy based on EIS data from CoCrMo alloy tested within different media. (c) Resistance to polarization (Rp) and (d) capacitance (Cdl) of CoCrMo alloy within different media. N=3 and all experiments were repeated three times.
A modified Randle’s equivalent circuit was used to model the EIS data, to determine to polarization resistances (Rp) and capacitance (C) of the electrochemical double layer at metal-solution interface with different media conditions were analyzed (Figure 6 c&d). The elements of the electrical circuit were Rp, and Q (Q=Cdl n and Q=C, if n>0.9) are in parallel, and in series with solution resistance (Rs) between the working electrode (WE) and the reference electrode (RE). The ANOVA for the comparison of Rp between different test media was significant (p=0.002). Resistance to polarization (Rp) in the presence of all the media conditions except challenged osteoblasts and macrophages was approximately 6×106 Ώ.cm2. However, the challenged macrophages and osteoblasts media decreased the Rp significantly compared to the control medium with values of 1×104 Ώ.cm2 (p=0.02) and 4 ×104 Ώ.cm2 (p=0.021), respectively. Such low Rp values indicated higher corrosion kinetics in the presence of medium from challenged macrophages and osteoblasts than other media conditions studied. The ANOVA for the comparison of the capacitance was also highly significant (p<0.001). Pairwise comparison also exhibited higher capacitance for medium from macrophages (3.5×10−5 Fˑcm2, p=0.047) compared to control medium.
3.3.4. Potentiodynamic studies
Representative potentiodynamic curves for all the electrolyte conditions are shown in Figure 7a. It was evident that the curves for different control media conditions were shifted towards lower current densities compared to inflammatory media conditions. The corrosion potential (Ecorr) and evolution of the corrosion current density (Icorr) values were obtained by Tafel’s estimation from potentiodynamic curves (Figure 7b &c). The Kruskal-Wallis test for the comparison of the corrosion potential between different test media was significant (p=0.007). The Ecorr value was higher (p=0.046) with media containing H2O2 (0.151±0.009 V vs. SCE) compared to the control medium, which was also in agreement with the OCP measurements. No significant difference was observed in the Ecorr values from challenged media conditions of macrophages, monocytes, and osteoblasts with that of the control medium. The ANOVA of the comparison of the corrosion current density was also significant (p=0.001). In contrast to Ecorr, Icorr values were significantly (p<0.008) higher with challenged macrophages (2.18×10−6 ± 8.7×10−7 A·cm2) compared to control medium and all other conditions (p<0.05).
Figure 7:
a) Potentiodynamic polarization curves of CoCrMo alloy within the different cell culture media and control electrolytes. Bar graphs showing b) ECorr and c) Icorr values of CoCrMo alloy within different cell culture media and control electrolytes. N=3 and all experiments were repeated three times.
4.0. Discussion
Fretting and corrosion processes within THR modular junctions are complex and multifactorial22,23. It was the goal of this study to generate in vitro experimental evidence for cell-accelerated corrosion. The results support our hypothesis that macrophages challenged with wear particles can generate an environment that negatively impacts the corrosion kinetics and corrosion rate of CoCrMo alloy. Thus, it appears possible that macrophages could contribute to corrosion processes occurring in THR modular junctions.
Our results have shown that monocytes, macrophages, and osteoblasts could be challenged with wear particles generated from CoCrMo alloy at 1 g/ml concentration with an average cell viability of > 88% after 24 h of incubation. The resulting media, after 24 h of culture, was then used as an electrolyte to test the corrosion kinetics and corrosion rate of CoCrMo alloy using standard electrochemical methods. This approach has several limitations. First, wear particles used to simulate inflammatory conditions were generated under sliding wear. This approach was chosen because wear particles can be generated at a higher rate, making it likely that the majority of particles are released into the wear testing solution, whereas particles from fretting tests are more likely to form films and agglomerates of oxide particles that stick to the sample surfaces and therefore are more difficult to isolate9. However, wear particles generated under sliding wear may differ from wear debris and corrosion products generated under fretting-corrosion within modular junctions. To date, there are no studies that can accurately describe the size and composition of fretting particles generated within modular junctions24,25. Most studies focus on larger flakes that can be removed from the taper surfaces8,24, yet the majority of particles are likely much smaller. It is also likely that such particles dwell in the crevice and form thick deposit layers. Such layers consist predominantly of chromium-oxide, CoCr particles, and metallo-organic complexes8,9. Most cases with ALTRs also exhibit chromium phosphate particles, but the conditions which form these particles are unknown. The type of in vivo fretting particles that is generated strongly depends on the fretting regime (partial slip or gross slip)26 and the underlying mechanism (abrasion, surface fatigue or tribochemical reactions)27. So far there is no consensus on which fretting regime is predominant, although retrieval findings suggest that at least initially, partial slip occurs9. But the fretting regime and wear mechanism may change over time. The resulting particle composition may further differ due to the transformation from the moment of initial generation to the moment of phagocytosis based on particle size and chemical environment28,29. The particles generated here appear to contain both CoCrMo alloy and chromium-oxide particles. As particles were generated within the bovine serum, metal-protein complexes will likely be present as well. Thus, particles generated in this study will at least partially resemble those generated under fretting within the modular junction. Another limitation of this study is that only macrophages, monocytes, and osteoblasts were considered to represent periprosthetic cell types, but other cell types, such as osteoclasts and fibroblasts may also have a role.
4.1. Generation of a corrosive environment by macrophages
Activation of cells by the addition of CoCrMo alloy wear particles clearly resulted in changes of the corrosion kinetics of CoCrMo within the conditioned media. In terms of different cell types, the corrosion tendency was noticeably higher in media from challenged macrophages and osteoblasts (compared to monocytes), and was significantly higher when compared to control media. Similarly, EIS and potentiodynamic results further confirmed that the cell medium from challenged macrophages negatively influences the corrosion rate and kinetics of CoCrMo alloy compared to other cell culture conditions. The results of this study demonstrate that under inflammatory conditions, macrophages can alter the chemical environment to promote corrosion and compromise the corrosion behavior of CoCrMo alloy. The effect of other cell types was negligible, although medium from challenged osteoblasts did result in a lower polarization resistance and higher capacitance compared to control media.
These results confirm the model of cell-accelerated corrosion proposed by Gilbert et al., suggesting that macrophages release reactive oxygen species under inflammatory conditions and thus promote corrosion14,30,31. One of the potential reactive oxygen species occurring in vivo is H2O2. In this study, we have used 0.03M H2O2 containing medium as a positive control for reactive oxygen species. Consistent with a previous report21, the corrosion tendency (OCP and Ecorr) of CoCrMo alloy in the presence of medium with 0.03M H2O2 was significantly higher (towards noble potential), compared to other media conditions. This effect could be attributed to the changes in the passivation kinetics in the presence of H2O2, potentially protecting the CoCrMo metal surface21. However, this effect was not observed in medium from challenged macrophages. It is, therefore, possible that H2O2 is not the primary acting reactive oxygen species responsible for the observed corrosion kinetics. For example, macrophages could release inflammatory cytokines, chemokines, inflammatory enzymes, or proteolytic enzymes such as matrix metalloproteinases. Hence, the practice of H2O2 containing solutions as a positive control for such experiments might need additional consideration. It is important to note that media from challenged macrophages had the lowest pH. However, a pH of 6.74±0.14 is not low enough to explain the significant shift in OCP, Rp, Cp, and Icorr that was observed here or the severe corrosion features observed on retrieved implants9. It is more likely that metal wear debris activates macrophages through a unique cellular mechanism to generate a corrosive environment. The exact combination of the different reactive oxygen species and other chemical species secreted by macrophages is unknown and needs to be determined in detail. However, the results of our study demonstrate that the release of this combination of chemical species is able to compromise the corrosion behavior of CoCrMo alloy within the confined microenvironment of THR modular junctions. One of the limitations in the evaluation of potentiodynamic data is the subjective nature of the Tafel’s estimation. Therefore, experiments were conducted in triplicates, and the mean values are reported. The fairly large deviation of the results is typical for this evaluation method, yet significant differences in Ecorr and Icorr could be detected. Also, it is important to note that all the electrochemical experiments were conducted under atmospheric conditions, not in CO2-enriched atmosphere. Therefore, a slight variation in pH can occur as reported in a previous study32, but this pH change is inconsequential for the corrosion behavior of CoCrMo, and all tests were conducted under the same conditions.
4.2. Model of cell-accelerated corrosion
In a previous study, cell-like features were found on the head taper surfaces of some femoral heads9. These features were clearly separate from one another, had a size range from 20 to 50 micrometers, and had a comparable morphology to that of cells such as macrophages in a scanning electron microscope (SEM). In an earlier study, some surface damage features on retrieved implant surfaces—in areas outside of modular interfaces—have been attributed to cell-induced corrosion14, but most of these features were later explained to be caused by the use of electrocautery during revision surgery33,34. In contrast, the previously described cell-like features are located within the contact zone of the head/stem taper surfaces. Specifically, these features were usually seated within areas of column damage. Column damage is characterized by long etched troughs running along the taper axis9,35, and is associated with banding of the CoCrMo alloy microstructure36. Considering our findings in this study, it is probable that fretting within modular taper junctions will eventually release particles from the crevice. Such particles may attract monocytes and macrophages to the crevice. Under optimal head-neck seating conditions, the interface is too tight to allow fluid to pass into the crevice formed by the modular junction interface. However, if the femoral head is not seated properly, or if the taper surfaces engage proximally in the head taper because of the taper design, an opening could occur and cells could gain access to the crevice. It is also possible that ongoing fretting processes may lead to a local widening of the crevice. Macrophages that possibly enter the crevice will be exposed to particles and particle agglomerates generated during fretting. Activated cells may be able to adhere to the taper surfaces and continuously release reactive oxygen species and cytokines under their cell membrane (cell-metal interface) resulting in eventual chemical changes of the microenvironment of the metal surfaces. Depending on the number of cells and considering the confined environment of the crevice, the concentration of chemically active species may be high enough to detrimentally change the corrosion kinetics on the CoCrMo alloy surfaces and increase the corrosion rate. The change of the crevice chemistry may explain harsh damage modes such as column damage, etching and pitting on the head taper surfaces, and intergranular corrosion phase boundary corrosion and pitting on the stem taper surfaces as earlier reported6,9,37,38. These damage modes can contribute significantly to fretting-induced material loss. It is also likely that macrophages will undergo necrosis due to the high particle and metal ion burden. The release of cell contents into the crevice during necrosis may further contribute to the harsh chemical environment within the crevice. Ongoing fretting will continue to recruit more inflammatory cells to the site, which will, in turn, accelerate corrosion processes within the taper junction. The process continues synergistically in a cyclic manner, thereby assisting the occurrence of corrosion modes that eventually accelerate the damage modes, as illustrated in Figure 8.
Figure 8:
Schematic representation of the pathway of cell assisted corrosion (CAC) within the neck taped modular junction of THRs. Wear particles released due to the micromotion of the head and neck modular junction attract inflammatory cells, including monocytes and macrophages, to the site. Wear debris causes polarization of monocytes to specific macrophage lineage (phenotype is unknown) and activate macrophages at the head-neck interface. These activated cells will secrete factors (ROS, cytokines, chemokines, and other unknown factors), which cause a corrosive environment at the cell-material interface. This may lead to a cascade of events which changes the overall chemical nature of the microenvironment within the crevice, further accelerating the corrosion damage.
Our previous retrieval findings and the electrochemical data of this study suggest that the contribution of macrophages to corrosion occurs as a result of fretting. Macrophages can only contribute to the corrosion process after fretting occurs, and a pathway for cells into the crevice is provided. Therefore, the term, cell-accelerated corrosion (CAC) may be more appropriate than cell-induced corrosion. The contribution of CAC to material loss cannot be easily quantified and may differ broadly. At least in the case of column damage, one can estimate that the contribution is significant as it covers large areas and leaves troughs with a depth of several tens of micrometers9,36. Column damage is also inherently related to the alloy microstructure as are other damage modes such as phase boundary corrosion and intergranular corrosion 35,37,38. Thus, the potential impact of CAC is also dictated by the alloy microstructure.
Although the results of this study support the theory that CAC may be an important contributor to corrosion processes within THR modular junctions, further studies are required to determine the transitions in the electrochemical mechanisms/kinetics and overall impact on the material loss. Therefore, a detailed evaluation of molecular pathways and cellular phenotypic differences in macrophage cells during corrosion are the future perspectives of this study. In addition, the study utilized low carbon CoCrMo alloy instead of high carbon CoCrMo alloy for all the electrochemical tests, which will be considered in future evaluations.
5.0. Conclusions
In conclusion, this study has successfully established an in vitro test to characterize the corrosion behavior of CoCrMo alloy in conditioned cell culture media of different periprosthetic cell types under normal and inflammatory conditions. The results demonstrated that macrophages had the highest particle uptake and had a significant effect on the corrosion rate as measured by Icorr and the overall corrosion kinetics as measured by the OCP, Rp, and the Nyquist plots. Thus, macrophages can alter the chemical environment in the presence of wear debris and compromise the corrosion behavior of CoCrMo alloy. These results and earlier retrieval findings suggest the following in vivo model: fretting wear and corrosion induced by micromotion may release wear debris that recruits monocytes/macrophages to the site of the modular junction. If the crevice is wide enough, macrophages may enter the crevice and change the chemical environment, give rise to chemically dominated damage modes, and subsequently accelerate material loss. Further studies are needed to determine the exact mechanisms of CAC and its overall impact on metal degradation and metal ion release. A detailed evaluation of molecular pathways and cellular phenotypic differences in macrophages during corrosion needs to be further investigated.
Supplementary Material
6.0. Acknowledgments
The authors acknowledge financial support from NIH R01 AR070181 and the Blazer foundation for the Regenerative Medicine and Disability Research (RMDR) lab at the Department of Biomedical Sciences, UIC College of Medicine at Rockford. The authors also would like to thank Robert M. Urban and Hannah J. Lundberg (Department of Orthopedics, Rush University Medical Center) for valuable comments and discussions.
Footnotes
Conflict of interests
The authors confirm that there is no conflict of interest.
References
- 1.Carli A, Reuven A, Zukor DJ, Antoniou J. 2011. Adverse soft-tissue reactions around non-metal-on-metal total hip arthroplasty - a systematic review of the literature. Bull. NYU Hosp. Jt. Dis. 69 Suppl 1:S47–51. [PubMed] [Google Scholar]
- 2.John Cooper H, Della Valle CJ, Berger RA, et al. 2012. Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty. J. Bone Jt. Surg. 94(18):1655–1661. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Lindgren JU, Brismar BH, Wikstrom AC. 2011. Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J. Bone Joint Surg. Br. 93(10):1427–1430. [DOI] [PubMed] [Google Scholar]
- 4.Hussey DK, McGrory BJ. 2017. Ten-Year Cross-Sectional Study of Mechanically Assisted Crevice Corrosion in 1352 Consecutive Patients With Metal-on-Polyethylene Total Hip Arthroplasty. J. Arthroplasty 32(8):2546–2551. [DOI] [PubMed] [Google Scholar]
- 5.McGrory BJ, MacKenzie J, Babikian G. 2015. A High Prevalence of Corrosion at the Head-Neck Taper with Contemporary Zimmer Non-Cemented Femoral Hip Components. J. Arthroplasty 30(7):1265–1268. [DOI] [PubMed] [Google Scholar]
- 6.Gilbert JL, Buckley CA, Jacobs JJ. 1993. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. J. Biomed. Mater. Res. 27(12):1533–1544. [DOI] [PubMed] [Google Scholar]
- 7.Gilbert JL, Mali S, Urban RM, et al. 2011. In vivo oxide-induced stress corrosion cracking of Ti-6Al-4V in a neck-stem modular taper: Emergent behavior in a new mechanism of in vivo corrosion. J. Biomed. Mater. Res. B Appl. Biomater. 100(2):584–94. [DOI] [PubMed] [Google Scholar]
- 8.Hall DJ, Pourzal R, Della Valle CJ, et al. 2015. Corrosion of Modular Junctions in Femoral and Acetabular Components for Hip Arthroplasty and Its Local and Systemic Effects. In: STP1591, ASTM International. West Conshohocken, PA: p 410–427. [Google Scholar]
- 9.Hall DJ, Pourzal R, Lundberg HJ, et al. 2018. Mechanical, chemical and biological damage modes within head-neck tapers of CoCrMo and Ti6Al4V contemporary hip replacements. J. Biomed. Mater. Res. B Appl. Biomater. 106(5):1672–1685. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Dufils J, Laurent MP, Kunze J, et al. 2015. A Servoelectric Apparatus with Potentiostat to Study the Fretting Corrosion of Cobalt-Chromium–Titanium Alloy Couples. In: Greenwald AS, Kurtz SM, Lemons JE, Mihalko WM, editors. Modularity and Tapers in Total Joint Replacement Devices. 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA: 19428–2959: ASTM International; p 303–320 [cited 2016 Jun 8] Available from: http://www.astm.org/doiLink.cgi?STP159120140143. [Google Scholar]
- 11.Fischer A, Janssen D, Wimmer MA. [date unknown]. The Influence of Molybdenum on the Fretting Corrosion Behavior of CoCr/TiAlV Couples. Biotribology [cited 2017 May 30] Available from: http://www.sciencedirect.com/science/article/pii/S2352573816300555.
- 12.Royhman D, Mathew MT, Runa MJ, et al. 2014. Electrochemical And Fretting-corrosion Behavior Of Mixed Metal Contacts (Ti6Al4V-CoCrMo) In A Hip Modular Junctions. Trans ORS 60:0055. [Google Scholar]
- 13.Hallab NJ, Messina C, Skipor A, Jacobs JJ. 2004. Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacements. J. Orthop. Res. 22(2):250–259. [DOI] [PubMed] [Google Scholar]
- 14.Gilbert JL, Sivan S, Liu Y, et al. 2015. Direct in vivo inflammatory cell-induced corrosion of CoCrMo alloy orthopedic implant surfaces. J. Biomed. Mater. Res. A 103(1):211–223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Lundberg HJ, Ha NQ, Hall DJ, et al. 2015. Contact Mechanics and Plastic Deformation at the Local Surface Topography Level After Assembly of Modular Head-Neck Junctions in Modern Total Hip Replacement Devices In: Greenwald AS, Kurtz SM, Lemons JE, Mihalko WM, editors. Modularity and Tapers in Total Joint Replacement Devices. West Conshohocken, PA: ASTM International; p 59–82 [cited 2016 Feb 17]. [Google Scholar]
- 16.Wimmer MA, Laurent MP, Mathew MT, et al. 2015. The effect of contact load on CoCrMo wear and the formation and retention of tribofilms. Wear 332–333:643–649. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Mathew MT, Uth T, Hallab NJ, et al. 2011. Construction of a tribocorrosion test apparatus for the hip joint: Validation, test methodology and analysis. Wear 271(9):2651–2659. [Google Scholar]
- 18.Mathew MT, Jacobs JJ, Wimmer MA. 2012. Wear-corrosion synergism in a CoCrMo hip bearing alloy is influenced by proteins. Clin. Orthop. 470(11):3109–3117. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Bijukumar DR, Segu A, Mou Y, et al. 2018. Differential toxicity of processed and non-processed states of CoCrMo degradation products generated from a hip simulator on neural cells. Nanotoxicology [cited 2018 Sep 25] Available from: https://www.tandfonline.com/doi/abs/10.1080/17435390.2018.1498929. [DOI] [PMC free article] [PubMed]
- 20.Mathew MT, Runa MJ, Laurent M, et al. 2011. Tribocorrosion behavior of CoCrMo alloy for hip prosthesis as a function of loads: a comparison between two testing systems. Wear Int. J. Sci. Technol. Frict. Lubr. Wear 271(9–10):1210–1219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Bearinger JP, Orme CA, Gilbert JL. 2003. Effect of hydrogen peroxide on titanium surfaces: In situ imaging and step-polarization impedance spectroscopy of commercially pure titanium and titanium, 6-aluminum, 4-vanadium. J. Biomed. Mater. Res. A 67A(3):702–712. [DOI] [PubMed] [Google Scholar]
- 22.Pourzal R, Lundberg HJ, Hall DJ, Jacobs JJ. 2018. What Factors Drive Taper Corrosion? J. Arthroplasty 33(9):2707–2711. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Jacobs JJ. 2016. Corrosion at the Head-Neck Junction: Why Is This Happening Now? J. Arthroplasty 31(7):1378–1380. [DOI] [PubMed] [Google Scholar]
- 24.Crainic AM, Callisti M, Palmer MR, Cook RB. 2018. Investigation of nano-sized debris released from CoCrMo secondary interfaces in total hip replacements: Digestion of the flakes. J. Biomed. Mater. Res. B Appl. Biomater.. [DOI] [PubMed] [Google Scholar]
- 25.Catelas I, Bobyn JD, Medley JB, et al. 2003. Size, shape, and composition of wear particles from metal-metal hip simulator testing: effects of alloy and number of loading cycles. J. Biomed. Mater. Res. A 67(1):312–327. [DOI] [PubMed] [Google Scholar]
- 26.Fouvry S, Kapsa Ph, Vincent L. 1995. Analysis of sliding behaviour for fretting loadings: determination of transition criteria. Wear 185(1–2):35–46. [Google Scholar]
- 27.Zum Gahr K-H. 1987. Microstructure and Wear of Materials. Amsterdam, Netherlands: Elsevier. [Google Scholar]
- 28.Pourzal R, Catelas I, Theissmann R, et al. 2011. Characterization of Wear Particles Generated from CoCrMo Alloy under Sliding Wear Conditions. Wear 271(9–10):1658–1666. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Gill HS, Grammatopoulos G, Adshead S, et al. 2012. Molecular and immune toxicity of CoCr nanoparticles in MoM hip arthroplasty. Trends Mol. Med. 18(3):145–155. [DOI] [PubMed] [Google Scholar]
- 30.Zhang J, Hiromoto S, Yamazaki T, et al. 2016. Effect of macrophages on in vitro corrosion behavior of magnesium alloy. J. Biomed. Mater. Res. A 104(10):2476–2487. [DOI] [PubMed] [Google Scholar]
- 31.Atrens A, Liu M, Zainal Abidin NI. 2011. Corrosion mechanism applicable to biodegradable magnesium implants. Mater. Sci. Eng. B 176(20):1609–1636. [Google Scholar]
- 32.Pourzal R, Cichon R, Mathew MT, et al. 2014. Design of a tribocorrosion bioreactor for the analysis of immune cell response to in situ generated wear products. J. Long. Term Eff. Med. Implants 24(1):65–76. [DOI] [PubMed] [Google Scholar]
- 33.Campbell P, Yuan N, Luck J, et al. 2017. Re-Examining the Concept of Inflammatory Cell-Induced Corrosion. Bone Jt. J 99-B(SUPP 3):57–57. [Google Scholar]
- 34.Aldinger P, Pawar V. 2017. Similarities Between Reported Inflammatory Cell-Induced Corrosion Features and Electrocautery Damage. Bone Jt. J 99-B(SUPP 3):9–9. [Google Scholar]
- 35.Pourzal R, Hall DJ, Ehrich J, et al. 2017. Alloy Microstructure Dictates Corrosion Modes in THA Modular Junctions. Clin. Orthop. Relat. Res. 475(12):3026–3043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Hall D, McCarthy S, Ehrich J, et al. 2018. Imprinting and Column Damage on CoCrMo Head Taper Surfaces in Total Hip Replacements. Implant Retr. Anal. Methods Implant Surveill. [cited 2019 Feb 25] Available from: http://www.astm.org/DIGITAL_LIBRARY/STP/PAGES/STP160620170121.htm.
- 37.Lin A, Hoffman EE, Marks LD. 2017. Effects of Grain Boundary Misorientation and Chromium Segregation on Corrosion of CoCrMo Alloys. CORROSION 73(3):256–267. [Google Scholar]
- 38.Panigrahi P, Liao Y, Mathew MT, et al. 2014. Intergranular pitting corrosion of CoCrMo biomedical implant alloy. J. Biomed. Mater. Res. - Part B Appl. Biomater. 102(4):850–859. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.









