Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2021 Apr 2.
Published in final edited form as: Audiol Neurootol. 2020 Apr 2;25(4):215–223. doi: 10.1159/000506666

Assessment of hearing-aid benefit using patient-reported outcomes and audiologic measures

James R Dornhoffer 1, Ted A Meyer 1, Judy R Dubno 1, Theodore R McRackan 1
PMCID: PMC7371552  NIHMSID: NIHMS1573675  PMID: 32241007

Abstract

Purpose:

To determine the contributions to hearing aid benefit of patient-reported outcomes and audiologic measures.

Methods:

Independent review was conducted on audiologic and patient-reported outcomes of hearing aid benefit collected in the course of a middle-ear implant FDA clinical trial. Unaided and aided data were extracted from the preoperative profiles of 95 experienced hearing aid users, and the relationships between a patient-reported outcome and audiologic measures were assessed. The following data were extracted: unaided and aided pure-tone or warble-tone thresholds (PTA), word recognition in quiet (NU-6), Speech Perception in Noise (low/high context SPIN), and patient-reported benefit (Abbreviated Profile of Hearing Aid Benefit, APHAB). Hearing aid benefit was defined as the difference in thresholds or scores between unaided and aided conditions, as measured in the soundfield. Correlations were computed among audiologic measures and global APHAB and subscale scores of hearing aid benefit.

Results:

Significant improvements in all audiologic measures and APHAB scores were observed comparing unaided to aided listening (all p<0.001). However, correlations between audiologic and patient reported measures of aided performance or hearing aid benefit were low-to-weak or absent. No significant correlations were found between aided audiologic measures (PTA, NU-6, SPIN) and any aided APHAB scores (all p>0.0125), and significant relationships for hearing aid benefit were absent with only few exceptions. Hearing aid benefit defined by global APHAB using NU-6 and SPIN scores showed significant but weak positive correlations (r=0.37, p<0.001; r=0.28, p=0.005, respectively) and ease of communication APHAB subscale scores (r=0.32, p<0.001; r=0.33, p=0.001, respectively).

Conclusion:

Hearing aid benefit assessed with audiologic measures were poor predictors of patient-reported benefit. Thus, patient-reported outcomes may provide a unique assessment of patient-perceived benefit from hearing aids, which can be used to direct hearing aid programming, training, or recommendations of alternative hearing services.

Keywords: APHAB, hearing aid benefit, audiologic measures, patient-reported outcome measure

INTRODUCTION

Approximately 38 million individuals in the United States over the age of 12 years have some form of hearing loss (Goman & Lin, 2016; Huddle et al, 2017). The majority of individuals have mild to moderate hearing loss, which is typically treated by amplification with hearing aids, which can result in significant improvement in quality of life (Humes & Krull, 2012; Goman & Lin, 2016). New hearing aid technologies and complex fitting algorithms have become available over the past several years, but with limited evidence to guide the selection and fitting of specific hearing-aid features. Hearing-aid fitting is typically determined by pure-tone thresholds, with the fitting verified by other audiologic measures, such as speech-recognition scores and real-ear probe-microphone measures (Carhart, 1946; Dillon, James & Ginis, 1997). Some audiologists may also use patient-reported outcome measures (PROMs) of hearing aid benefit and quality of life to supplement audiologic measures, such as: the Abbreviated Profile of Hearing Aid Benefit (APHAB), Clincally Oriented Scale of Improvement (COSI), Speech Spatial and Qualities of Hearing Scale (SSQ), and International Outcomes Inventory-Hearing Aids (IOI-HA) (Dillon, James & Ginis, 1997; Cox & Alexander, 1992; Cox et al, 2000; Gatehouse, 2001; Killion & Gudmundsen, 2005; Mendel, 2007). However, the relationship between audiologic measures used for hearing aid fitting and PROMs that assess hearing aid benefit remains uncertain, which limits the ability of providers to assess the appropriateness of the fit and predict hearing aid use and satisfaction.

Research in this area is lacking and, at times, inconsistent (Cox et al, 2000; Humes & Krull, 2012). A review of the literature indicates that audiologic measures tend to poorly correlate with PROMs evaluating hearing aid benefit (Kapteyn, 1977; Parving, 1991; Cox et al, 2000; Humes, 2003; Killion & Gudmundsen, 2005; Chang et al, 2016;). However, many of these studies have small sample sizes – often fewer than 30 subjects (Mendel, 2007; Cox & Alexander, 1992; Cox et al, 2000) – and report results from non-validated, study specific, instruments that are not commonly used in clinical or research settings (Killion & Gudmundsen, 2005; Kapteyn, 1977; Parving, 1991). Those reporting results from validated PROMs with reasonable sample sizes examined only the relationship of PROMs with unaided audiologic measures and not hearing aid benefit, which the current study aims to address (Humes, 2003; Humes, Wilson & Humes, 2003; Chang et al, 2016).

The APHAB is one of the most commonly used PROMs to evaluate hearing aid benefit and is often cited for its ease of delivery and interpretation (Cox & Alexander, 1995; Löhler et al, 2017). Despite widespread use, little is known regarding its relationship to hearing aid benefit as indicated by audiologic measures (Cox et al, 2000; Cox et al, 2003; Killion & Gudmundsen, 2005; Löhler et al, 2017). Cox et al, 2003, demonstrated a moderate correlation of unaided APHAB scores with pure-tone average (PTA) and unaided speech recognition (Northwestern University Test Number 6; NU-6) testing in quiet and in noise; however, the audiologic outcomes were measured with headphones and subjects were not tested under aided conditions. As such, the relationship of aided audiologic measures with aided APHAB scores is unknown, as is the relationship between hearing aid benefit assessed with audiologic measures and PROMs.

The purpose of this study is to assess the relationship between patient self-report of aided performance and hearing aid benefit, as measured by the APHAB, and audiologic measures, including pure-tone and warble-tone thresholds, speech recognition scores using the NU-6, and the Speech Perception in Noise text (SPIN) (Tillman & Carhart, 1966). This information may demontrate the relative value of each type of measure as a component in a hearing aid evaluation battery and aid in interpreting results to maximize hearing aid benefit.

MATERIALS AND METHODS

Data

Data for this study were collected by Ototronix (Houston, TX) as part of multi-center Phase III Food and Drug Administration (FDA) clinical trial for the Soundtec Direct Drive Hearing System (now Maxum Hearing Implant). A Material Transfer Agreement was signed between our institution and Ototronix before de-identified data were shared with the authors. Raw data for individual subjects were provided without prior statistical manipulation. Ototronix personnel did not participate in the planning, execution, or composition of this project and did not review the manuscript.

Subjects

The study sample included 95 subjects (34.7% female; median age: 67 years; age range: 21–80 years) with mild to moderate hearing loss who were experienced users of traditional hearing aids electing to pursue evaluation for middle ear implantation due to dissatisfaction with their hearing aids. These subjects were previously described in McRackan et al, 2016 and 2018. Data on the effects of middle ear implantations were not collected or assessed. Rather, this study took advantage of preoperative audiologic and quality of life data from a large sample of experienced hearing aid users, measured without and with hearing aids. The clinical trial inclusion and exclusion criteria were as follows. Subjects were required to be native English speakers with a minimum 2-year history of stable hearing loss. They were required to have bilateral symmetrical sensorineural hearing loss with <15 dB difference in PTA (average of 1000, 2000, and 4000 Hz) between ears, and <10 dB difference between air and bone conduction thresholds. Frequencies for calculating this PTA were part of the clinical trial protocol and were not used in analysis for the current study. Preoperative, bilateral pure-tone thresholds were required to fall within the following ranges: 0–50 dB HL at 250 Hz, 0–60 dB HL at 500 Hz, 10–70 dB HL at 1000 Hz, 35–75 dB HL at 2000 Hz, 50–75 dB HL at 3000 Hz, 50–80 dB HL at 4000 Hz, 40–100 dB HL at 6000 Hz; PTA was required to range from 35–70 dB HL. Word recognition scores measured under earphones in quiet using 50-word lists from the NU-6 were required to be >60% in both ears. As determined by the clinical trial protocol, the poorer hearing ear was selected as the test ear.

Prior to data collection, subjects were required to have worn properly fitted hearing aids, defined according to NAL-R targets (Byrne & Dillon, 1986) for a minimum of 6 months. Hearing aids that had not been fitted using NAL-R targets were refitted by clinical trial audiologists and subjects were required to use their hearing aids with their new settings for a minimum of 45 days before enrollment. Patients utilized their hearing aids for an average of 7.2 years prior to data collection; ranging from 6 months to 30 years. Subjects with conductive or retrocochlear pathology were excluded from the study.

Audiologic measures (thresholds and speech-recognition scores)

Pure-tone or warble-tone thresholds, NU-6 word-recognition scores, and SPIN scores were the primary audiologic outcome measures. Conditions under which each were tested are presented in Table 1. Thresholds from 250–6000 Hz were measured under three conditions: pure tones with either supra-aural or insert earphones, warble tones unaided in the soundfield, and warble tones with hearing aids in the soundfield. These conditions will be referred to as earphone, unaided, and aided thresholds, respectively. For statistical analysis, PTAs (500, 1000, and 2000 Hz) were calculated for each subject and in each condition. Hearing aid benefit was defined as the reduction in PTA with hearing aid use. This was calculated as unaided PTA in soundfield minus aided PTA in soundfield, so improvements in threshold with hearing aid use yield positive values.

Table 1.

Conditions Under Which Outcomes Were Measured

Outcome Measure Earphone Unaided Aided
Pure-Tone Thresholds X
Warble-Tone Thresholds X X
NU-6* X X X
SPIN X X
APHAB X X
*

Northwestern University Auditory Test No.

Speech Perception in Noise Test

Abbreviated Profile of Hearing Aid Benefit

Word recognition scores using the NU-6 was also measured under the same three conditions: supra-aural or insert earphones with words presented at 40 dB above speech recognition threshold (earphone), unaided in the soundfield with words presented at 63 dB SPL (unaided), and with hearing aids in the soundfield at 63 dB SPL (aided). Sentence recognition using the SPIN was measured under unaided and aided conditions. The SPIN is composed of 8 lists of 50 sentences each with multitalker babble presented at a +8 dB signal-to-noise ratio. Listeners are instructed to repeat the last word of each sentence. Half of the sentences are high context sentences in which the final word was predictable from the earlier part of the sentence, and half of the sentences were low context sentences in which the final word cannot be determined from the earlier part of the sentence. Hearing aid benefit was defined as improvement in scores with hearing aid use. This was calculated as aided scores in the soundfield subtracted by unaided scores in the soundfield. As with PTA, improvements in scores with hearing aid use yield positive values.

All soundfield testing was performed in sound attenuated booths that met ANSI standards for sound attenuation, with subjects sitting 1 meter from the loudspeaker and speech (and babble, if present) delivered at 0 degrees azimuth. The non-test ear was occluded with a foam plug during testing.

Patient-reported outcome measure (APHAB)

The APHAB (Cox & Alexander, 1995) is a self-report questionnaire in which individuals rate their frequency of problems in various situations, which are scored in 4 subscales: ease of communication in quiet (EC), in background noise (BN), and in situations of reverberation such as gymnasiums or classrooms (RV). Subjects also rate the frequency with which they experience negative reactions or aversion to environmental sounds (AV) (Humes, 2003). Global scores are calculated as the average of the EC, BN, and RV subscale scores. Higher scores reflect greater frequency of problems. As part of the FDA clinical trial, subjects provided responses to the APHAB that reflected their frequency of problems without and with hearing aids (unaided and aided). Hearing aid benefit as indicated by APHAB was calculated as unaided scores minus aided scores, so improvement with hearing aids yields positive values (Cox & Alexander, 1995).

Statistical analyses

All analyses were performed with SPSS 25.0 (IBMCorp., Armonk, NY). All continuous variables were tested for normal distribution as determined by the Kolmogorov-Smirnov test. Nominal variables were summarized by frequency, percentage, and/or range. Continuous variables were summarized by mean (standard deviation, SD) where appropriate. PTAs and NU-6 scores in earphone, unaided, and aided conditions were analyzed with a one-way ANOVA or Kruskal-Wallis test, with Dunn-Bonferroni post hoc testing employed as necessary for pairwise assessment of continuous variables. Unaided and aided SPIN, APHAB global and APHAB subscale scores were compared using repeated samples t-test for repeated measures in the same subject. Pearson’s correlations were used to test for the presence of significant associations among APHAB and audiologic measures. During correlational analysis, aided APHAB global and subscores and APHAB hearing aid benefit were analyzed independently against aided thresholds, NU-6 scores, SPIN scores, and each estimate of hearing aid benefit.

A p-value <0.05 was used to determine statistical significance in non-correlational analyses. For Pearson correlation testing against multiple independent variables, p-values underwent Bonferroni correction by a factor of 4, with significance defined as p<0.0125. The strength of correlations was defined as follows: very weak=0.00–0.19; weak=0.20–0.39; moderate=0.40–0.59; strong=0.60–0.79; very strong; 0.80–1.00.22

RESULTS

Audiologic measures: Thresholds and speech recognition scores

Audiologic measures are summarized in Table 2. As expected, subjects’ mean PTAs were higher in the earphone and unaided conditions as compared to the aided condition (p<0.001 for both). Earphone and unaided PTAs were not significantly different (p=0.309). Scores on the NU-6 were better in the aided and earphone conditions as compared to the unaided condition (p<0.001 for both). NU-6 scores in the earphone and aided conditions were not significantly different (p=0.427). Similarly, scores were higher for both low and high context SPIN in the aided condition as compared to the unaided condition (p<0.001 for both). Significant decreases in thresholds and significant improvements in scores between aided and unaided conditions reflect significant hearing aid benefit.

Table 2.

Comparison of Audiologic and Patient Reported Outcome Measures

Earphone Unaided Aided p-value
PTA* dB HL mean (SD) 40.9 (8.8) 40.9 (10.6) 25.4 (6.7) p<0.0010
Earphone vs. Unaided p=0.309
Earphone vs. Aided p<0.001
Unaided vs. Aided p<0.001
NU-6 % mean (SD) 81.5 (10.7) 42.8 (28.9) 77.4 (16.5) p<0.0010
Earphone vs. Unaided p<0.001
Earphone vs. Aided p=0.427
Unaided vs. Aided p<0.001
SPIN % mean (SD)
Low Context n/a 3.76 (4.2) 9.96 (5.4) p<0.001
High Context n/a 11.97 (9.1) 21.4 (4.4) p<0.001
APHAB§ mean (SD)
Ease of Communication n/a 60.7 (21.7) 26.1 (14.3) p<0.001
Background Noise n/a 73.8 (15.1) 40.0 (15.5) p<0.001
Reverberation n/a 73.1 (14.9) 36.1 (15.6) p<0.001
Aversiveness n/a 21.1 (18.4) 47.0 (26.4) p<0.001
Global n/a 69.2 (14.4) 34.0 (12.5) p<0.001
*

Pure Tone Average

Northwestern University Auditory Test No. 6

Speech Perception in Noise Test

§

Abbreviated Profile of Hearing Aid Benefit

0

ANOVA of earphone, unaided, and aided conditions; subsequent pairwise analysis is done post-hoc

Patient-reported outcome measure (APHAB)

APHAB subscale and global scores under unaided and aided conditions are reported in Table 2. Significant reductions in reported frequency of problems (hearing aid benefit) in ease of communication, background noise, reverberation subscales and global APHAB scores were observed in the aided condition as compared to the unaided condition (p<0.001). In contrast, significant increases in reported frequency of negative reactions or aversions to environmental sounds were observed in the aided condition as compared to the unaided condition (p<0.001). Length of hearing aid usage was not significantly associated with improvement in global APHAB scores (r=0.002, p=0.978) or any APHAB subset score (p>0.1 for all).

The aided values shown in this patient population correspond to previously identified normative APHAB scores for modern, wide-dynamic range compression hearing aids. The 50th percentile APHAB scores for modern hearing aids are EC=23, BN=40, RV=37, AV=38, and global=33 (Johnson, Cox, & Alexander, 2010). The scores for the current study’s patient population fall at the 50th to 65th percentile scores globally and in each category.

Correlations among audiologic and patient-reported outcome measures

Correlations among aided audiologic measures (PTA, NU-6, SPIN) and aided APHAB subscale and global scores showed no significant relationships (Table 3). Correlations among audiologic and patient-reported measures of hearing aid benefit showed several weak correlations (Table 4). Hearing aid benefit as indicated by NU-6 and high context SPIN scores showed weak positive correlations with global APHAB benefit (r=0.37, p<0.001; r=0.28, p=0.005; respectively) and ease of communication APHAB benefit (r=0.32, p=0.002; r=0.33, p=0.001; respectively). Hearing aid benefit as indicated by APHAB RV subscores showed a weak positive correlation with benefit as indicated by PTA and with benefit as indicated by NU-6 scores (r=0.26, p=0.011; r=0.33, p=0.001; respectively). Figure 1 shows global APHAB benefit plotted against benefit as determined by NU-6 and is the strongest associationobserved between patient-reported and audiologic measures of hearing aid benefit (r=0.37). Benefit as indicated by the APHAB BN and AV subscale scores were not significantly associated with any audiologic measures of hearing aid benefit.

Table 3.

Pearson Correlations among Aided APHAB (Subscale and Global Scores) and Aided Audiologic Measures (PTA, NU-6, SPIN)

Aided Measures PTA* NU-6 Low Context SPIN High Context SPIN
APHAB§ Ease of Communication r (95% CI) −0.07 (−0.26 to 0.13) 0.06 (−0.14 to 0.26) 0.15 (−0.07 to 0.32) 0.05 (−0.15 to 0.24)
Sig. (2-tailed) 0.525 0.572 0.154 0.649
APHAB§ Reverberation r (95% CI) −0.05 (−0.25 to 0.14) −0.19 (−0.37 to 0.01) −0.09 (−0.29 to 0.10) −0.03 (−0.22 to 0.17)
Sig. (2-tailed) 0.612 0.071 0.383 0.808
APHAB§ Background Noise r (95% CI) −0.07 0.26 to 0.13) −0.03 (−0.23 to 0.17) −0.02 (−0.22 to 0.18) 0.05 (−0.15 to 0.25)
Sig. (2-tailed) 0.478 0.742 0.869 0.610
APHAB§ Aversiveness r (95% CI) −0.08 (−0.27 to 0.12) 0.13 (−0.06 to 0.33) 0.09 (−0.10 to 0.29) 0.21 (0.02 to 0.39)
Sig. (2-tailed) 0.472 0.198 0.399 0.039
APHAB§ Global r (95% CI) −0.08 (−0.27 to 0.12) −0.07 (−0.27 to 0.12) 0.01 (−0.18 to 0.22) 0.03 (−0.16 to 0.23)
Sig. (2-tailed) 0.455 0.503 0.912 0.777

P-value <0.0125 is significant. None of these associations were statistically significant.

*

Pure Tone Average

Northwestern University Test Auditory Test No. 6

Speech Perception in Noise Test

§

Abbreviated Profile of Hearing Aid Benefit

Table 4.

Pearson Correlations Among Hearing Aid Benefit (Improvement in Measure with Hearing Aid Use) As Indicated by APHAB (Global and Subscale) and Audiologic Measures (PTA, NU-6, SPIN).

Measures of Hearing Aid Benefit PTA* NU-6 Low Context SPIN High Context SPIN
APHAB§: Ease of Communication r (95% CI) 0.16 (−0.04 to 0.33) 0.32 (0.13 to 0.49) 0.15 (−0.05 to 0.33) 0.33 (0.14 to 0.49)
Sig. (2-tailed) 0.133 0.002 0.155 0.001
APHAB§: Reverberation r (95% CI) 0.26 (0.44 to 0.07) 0.33 (0.13 to 0.49) 0.02 (−0.17 to 0.22) 0.23 (0.03 to 0.41)
Sig. (2-tailed) 0.011 0.001 0.814 0.027
APHAB§: Background Noise r (95% CI) 0.02 (−0.20 to 0.21) 0.16 (−0.04 to 0.34) 0.13 (−0.07 to 0.31) 0.05 (−0.14 to 0.25)
Sig. (2-tailed) 0.864 0.123 0.209 0.607
APHAB§: Aversiveness r (95% CI) −0.08 (−0.25 to 0.14) −0.08 (−0.27 t0 0.12) −0.01 (−0.21 to 0.19) −0.09 (−0.27 to 0.12)
Sig. (2-tailed) 0.864 0.123 0.209 0.607
APHAB§: Global r (95% CI) 0.20 (0.39 to 0.01) 0.37 (0.18 to 0.52) 0.13 (−0.05 to 0.33) 0.28 (0.09 to 0.46)
Sig. (2-tailed) 0.051 <0.001 0.205 0.005

P-value <0.0125 is significant. Bold text shows significant associations.

*

Pure Tone Average

Northwestern University Auditory Test No. 6

Speech Perception in Noise Test

§

Abbreviated Profile of Hearing Aid Benefit

Figure 1.

Figure 1.

Hearing aid benefit as indicated by Global APHAB score plotted against hearing aid benefit as indicated by word recognition score in quiet (NU-6).

DISCUSSION

In this study, we showed that audiologic measures and PROMs of aided performance were not significantly correlated, but some weak correlations were observed among audiologic and self-report measures of hearing aid benefit. Specifically, hearing aid benefit as indicated by NU-6 and high context SPIN scores correlated weakly and positively with benefit as indicated by global and ease of communication APHAB scores, whereas PTA and NU-6 benefit correlated weakly and positively with benefit assessed by the APHAB reverberation subscale. No single audiologic measure of hearing aid benefit predicted patient self-report of hearing aid benefit as measured with the APHAB. Therefore, given that the APHAB appears to provide unique information about hearing aid benefit that is not provided by audiologic outcomes, administering PROMs such as the APHAB should be considered when evaluating benefit for hearing aid users.

Evidence of the strength of relationship between audiologic measures and PROMs under aided conditions is limited. Many studies have small sample sizes (Cox & Alexander, 1992; Cox et al, 2000; Mendel, 2007) and/or used experimental and non-standardized audiologic measures and non-validated PROMs (Kapteyn, 1977; Parving, 1991; Killion & Gudmundsen, 2005). Those studies with large sample sizes and validated PROMs include primarily unaided measures (Humes, 2003; Humes, Wilson & Humes, 2003; Chang et al, 2016). However, these studies show similar outcomes to the current study with absent to low correlations between PROMs and unaided audiologic measures (Kapteyn, 1977; Parving, 1991; Killion & Gudmundsen, 2005; Humes & Krull, 2012;). In a multiple regression analysis, Chang et al, 2016, reported a significant but small negative association of unaided word recognition score and PROMs at one month but a weak positive correlation at 3 months (P=−0.68 and 1.01, respectively). Humes, 2003, and Humes et al, 2003, showed no correlation in a similar sample as well as in a large meta-analysis. In addition, absent to low correlations between patient self-report and speech recognition have been reported in cochlear implant users (Brendel et al, 2014; McRackan et al, 2016; Ramakers et al, 2017; McRackan et al, 2018; McRackan et al, 2018; Moberly et al, 2018). Together, these data support the added value of PROMs as an independent measure of patient-perceived benefit from hearing interventions.

The most likely explanation for the low to absent correlations seen in the current study is that hearing aid users’ real-world listening environments are more varied than can be predicted by simple audiologic measures. By extension, the benefit users achieve from their hearing aids in these environments are difficult to predict from benefit estimated from audiologic measures, including speech recognition in noise (in this case, SPIN). For example, no significant correlations were seen between aided performance or benefit as indicated by audiologic measures and by the APHAB background noise subscale, which asks about communication ability in specific listening situations (i.e., in the grocery store, in a crowd). Likewise, the aversiveness subscale asks about patients’ perceptions of particular environments (i.e., construction work, traffic) being too loud. Although there is value in optimizing hearing aid benefit as indicated by audiologic measures, low to absent correlations suggest that this benefit may not necessarily translate to self-perceived hearing aid benefit. Patient-perceived benefit is important as it is one of the most accurate predictors of hearing aid acquisition and consistent use (Fischer et al, 2011; McCormack & Fortnum, 2013) with an odds ratio of hearing aid acquisition 2 to 3 times greater than audiologic evidence of hearing loss or communication difficulties (Fischer et al, 2011). Of patients who are candidates for hearing aids, only 1 in 5 acquire them (Fischer et al, 2011; Chien & Lin, 2012; McCormack & Fortnum, 2013). Moreover, patients wait an average of 8.9 years to aquire hearing aids after becoming hearing aid candidates (Simpson et al., 2019) and, of those patients who acquire hearing aids, up to 24% do not routinely wear them (Chien & Lin, 2012; McCormack & Fortnum, 2013). Similar to data on acquisition, a systematic review of the literature found that the most commonly cited reasons for inconsistent hearing aid use were related to perceived benefit and general quality of life issues, such as hearing aid ease of use, in contrast to audiologic evidence of benefit (McCormack & Fortnum, 2013).

Hearing aid success is often defined as consistent and proper use of hearing aids (Fischer et al, 2011; Chien & Lin, 2012; McCormack & Fortnum, 2013; Gallagher & Woodside, 2018). Given the link between patient-perceived benefit and hearing aid acquisition and use, PROMs may help guide patients toward successful hearing rehabilitation. PROMs may also help providers direct changes in hearing aid use to maximize perceived benefit and examine nonaudiologic factors that are commonly cited as leading to increased success with hearing aids. These include increased family support, access to counselling, hearing aid comfort, ease of hearing aid use, and a favorable cost-benefit ratio of the hearing aid device (McCormack & Fortnum, 2013; Gallagher & Woodside, 2018). These factors are more difficult to predict using audiologic measures of hearing aid benefit alone. PROMs may also be used to advise patients about alternate means of rehabilitation such as a middle ear or cochlear implants.

The large, muilticenter sample size is a major strength of the current study. The rigourous nature of the prospective FDA trial that provided these data is additional strength. Specifically, all hearing aids were optimally fit (met NAL-R targets for at least 4 weeks prior to enrollment) and subjects included in this study were experienced hearing aid users (minimum of 6 months of hearing aid use). In addition, audiologic data included earphone, unaided and aided conditions, speech recognition in quiet and babble, along with a validated PROM (APHAB), which supported the comprehensive analyses performed in the current study.

One limitation of the current study is that participants in the FDA clinical trial were recruited for evaluation for middle ear implantation on the presumption of dissatisfaction with their current hearing aids. A potential limitation is that such patients may rate their patient-reported hearing aid benefit relatively low given their dissatisfaction, or may have poor audiologic outcomes with their hearing aids as a cause of such dissatisfaction. Nevertheless, hearing aid benefit as indicated by audiologic measures was robust and consistent with estimates reported in the literature (Humes & Krull, 2012; Goman & Lin, 2016). Similarly, the aided APHAB scores and APHAB benefit scores fall well within the population norm as defined by Johnson, Cox, & Alexander 2010 for modern hearing aids, as reported in the Results section.

A second limitation of this study is the absence of certain subject demographics, which may contribute to hearing aid use and benefit. Studies have linked consistent use of hearing aids with perceived cost-benefit ratio, social pressure/encouragement, comfort, and ease of hearing aid use rather than specific audiologic outcomes (McCormack & Fortnum, 2013; Gallagher & Woodside, 2018). As such, subject demographics, such as income, education, comfort with technology, and perceived social support, may play a role in determining self-perceived hearing aid benefit or may modify the relationship between audiologic measures and these outcomes. Future studies should consider the contribution of these factors to audiologic outcomes and PROMs related to hearing aid benefit. These analyses could help identify patterns of audiologic and demographic factors that correlate with the patient-perceived benefit and help further individualize hearing aid selection and fitting and selection of other hearing services and technologies.

CONCLUSIONS

Hearing aid benefit assessed with audiologic outcomes shows absent to low correlation with patient self-report assessments of hearing aid benefit. As such, PROMs provide a unique view of patients’ perceived benefit from the use of hearing aids, independent of audiologic measures. This is notable because perceived benefit rather than audiologic benefit is a strong independent predictor of hearing aid acquisition and consistent use. Thus, results of the current study support the use of PROMs as supplements to audiologic measures of hearing aid outcomes to help direct recommendations for continued hearing aid use or alternative hearing services or technologies.

Acknowledgement

Data for this study were collected by Ototronix (Houston, TX) as part of multi-center Phase III Food and Drug Administration (FDA) clinical trial for the Soundtec Direct Drive Hearing System (now Maxum Hearing Implant). A Material Transfer Agreement was signed between our institution and Ototronix before de-identified data were shared with the authors.

Funding Sources

This research was supported (in part) by a K12 award through the South Carolina Clinical & Translational Research (SCTR) Institute, with an academic home at the Medical University of South Carolina, NIH/NCATS Grant Number UL1 TR001450 and a grant from the National Institutes of Health (K23 DC016911). The funding agencies had no influence on the analysis or preparation of data for publication.

Footnotes

Statement of Ethics

All data analyzed in this manuscript were de-identifed before data access, handling, or analysis. Due to the de-identified nature of the source data, the current study was exempted from review by the Institutional Review Board of the Medical University of South Carolina

Disclosure Statement

The authors have no conflicts of interest to declare.

REFERENCES

  • 1.Brendel M, Frohne-Buechner C, Lesinski-Schiedat A, Lenarz T, Buechner A (2014). Everyday listening questionnaire: correlation between subjective hearing and objective performance. Cochlear Implants International, 15(1):13–9. [DOI] [PubMed] [Google Scholar]
  • 2.Byrne D, Dillon H. (1986). The National Acoustic Laboratories’ (NAL) new procedure for selecting the gain and frequency response of a hearing aid. Ear and Hearing, 7(4): 257–65. [DOI] [PubMed] [Google Scholar]
  • 3.Carhart R (1946). Tests for selection of hearing aids. Laryngoscope, 56(12): 780–94. [PubMed] [Google Scholar]
  • 4.Chien W, Lin FR. (2012). Prevalence of hearing aid use among older adults in the United States. Archives if Internal Medicine, 172(3):292–293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Cox R, Hyde M, Gatehouse S, Noble W, Dillon H, Bentler R, Stephens D, Arlinger S, Beck L, Wilkerson D, Kramer S, Kricos P, Gagne JP, Bess F, Hallberg L. (2000). Optimal outcome measures, research priorities, and international cooperation. Ear and Hearing, 21(4 Suppl):106S–115S. [DOI] [PubMed] [Google Scholar]
  • 6.Cox RM, Alexander GC (1995). The abbreviated profile of hearing aid benefit. Ear and Hearing, 16(2):176–86. [DOI] [PubMed] [Google Scholar]
  • 7.Cox RM, Alexander GC, Gray GA. (2003). Audiometric correlates of the unaided APHAB. Journal of the American Academy of Audiology, 14(7):361–71. [PubMed] [Google Scholar]
  • 8.Cox RM, Alexander GC. (1992). Maturation of hearing aid benefit: objective and subjective measurements. Ear and Hearing, 13(3): 131–41 [DOI] [PubMed] [Google Scholar]
  • 9.Dillon H, James A, Ginis J. (1997). Client Oriented Scale of Improvement (COSI) and its relationship to several other measures of benefit and satisfaction providedby hearing aids. Journal of the American Academy of Audiology, 8(1): 27–43 [PubMed] [Google Scholar]
  • 10.Fischer ME, Cruickshanks KJ, Wiley TL, Klein BE, Klein R, Tweed TS. (2011). Determinants of hearing aid acquisition in older adults. American Journal of Public Health,101(8):1449–55. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Gallagher NE, Woodside JV. (2018). Factors affecting hearing aid adoption and use: qualitative study. Journal of the American Academy of Audiology, 29(4):300–312. [DOI] [PubMed] [Google Scholar]
  • 12.Gatehouse S (2001). Self-report outcome measures for adult hearing aid services: some uses, users, and options. Trends in Amplification, 5(3):91–110. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Goman AM, Lin FR. (2016). Prevalence of hearing loss by severity in the United States. American Journal of Public Health, 106(10):1820–1822. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Gurgel RK, Jackler RK, Dobie RA, Popelka GR. (2012). A new standardized format for reporting hearing outcome in clinical trials. Otolaryngology Head Neck Surgery, 147(5):803–7. [DOI] [PubMed] [Google Scholar]
  • 15.Huddle MG, Goman AM, Kernizan FC, Foley DM, Price C, Frick KD, Lin FR. (2017). The economic impact of adult hearing loss: a systematic review. JAMA Otolaryngology Head and Neck Surgery, 143(10):1040–1048. [DOI] [PubMed] [Google Scholar]
  • 16.Humes LE, Wilson DL, Humes AC. (2003) Examination of differences between successful and unsuccessful elderly hearing aid candidates matched for age, hearing loss and gender. Internation Journal of Audiology, 42, 432–441. [DOI] [PubMed] [Google Scholar]
  • 17.Humes LE. (2003) Modeling and predicting hearing aid outcome. Trends in Amplification, 7(2):41–75. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Humes LE and Krull V Hearing aids for adults In Wong L & Hickson L (Eds). (2012) Evidence-based practice in audiology: evaluating interventions or children and adults with hearing impairment. San Diego, CA; Plural Publishing: 61–92. [Google Scholar]
  • 19.Johnson JA, Cox RM, Alexander GC. (2010). Development of APHAB norms for WDRC hearing aids and comparisons with original norms. Ear and Hearing, 31(1):47–55. [DOI] [PubMed] [Google Scholar]
  • 20.Kapteyn TS. (1977) Factors in the appreciation of a prosthetic rehabilitation. Audiology, 16:446–452 [DOI] [PubMed] [Google Scholar]
  • 21.Killion MC, Gudmundsen GI. (2005). Fitting hearing aids using clinical prefitting speech measures: an evidence-based review. Journal of the American Academy of Audiolology, 16(7):439–47. [DOI] [PubMed] [Google Scholar]
  • 22.Löhler J, Grabner F, Wollenberg B, Schlattmann P, Schonweiler R. (2017). Sensitivity and specificity of the abbreviated profile of hearing aid benefit (APHAB). European Archives of Oto-Rhino-Laryngology, 274(10):3593–3598 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.McCormack A, Fortnum H. (2013). Why do people fitted with hearing aids not wear them? International Journal of Audiology, 52(5):360–368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.McRackan TR, Ahlstrom JB, Clinkscales WB, Meyer TA, Dubno JR. (2016). Clinical implications of word recognition differences in earphone and aided conditions. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology, 37(10):1475–1481. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.McRackan TR, Bauschard M, Hatch JL, Franko-Tobin E, Droghini HR, Nguyen SA, Dubno JR. (2018). Meta-analysis of quality-of-life improvement after cochlear implantation and associations with speech recognition abilities. Laryngoscope, 128(4):982–990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.McRackan TR, Bauschard M, Hatch JL, Franko-Tobin E, Droghini HR, Velozo CA, Nguyen SA, Dubno JR (2018). Meta-analysis of cochlear implantation outcomes evaluated with general health-related patient-reported outcome measures. Otology and Neurotology, 39(1):29–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Mendel LL. (2007). Objective and subjective hearing aid assessment outcomes. American Journal of Audiology,16(2):118–29 [DOI] [PubMed] [Google Scholar]
  • 28.Moberly AC, Harris MS, Boyce L, Vasil K, Wucinich T, Pisoni DB, Baxter J, Ray C, Shafiro V. (2018). Relating quality of life to outcomes and predictors in adult cochlear implant users: Are we measuring the right things? Laryngoscope, 128(4):959–966. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Parving A (1991) The value of speech audiometry in hearing-aid rehabilitation. Scandanavian Journal of Audiology, 20:159–164. [DOI] [PubMed] [Google Scholar]
  • 30.Ramakers GGJ, Smulders YE, van Zon A, Van Zanten GA, Grolman W, Stegeman I. (2017). Correlation between subjective and objective hearing tests after unilateral and bilateral cochlear implantation. BMC Ear, Nose, and Throat Disorders, 17:10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Simpson AN, Matthews LJ, Cassarly C, Dubno JR. (2019). Time from hearing aid candidacy to hearing aid adoption: a longitudinal cohort study. Ear and Hearing. 40(3):468–476. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Tillman TW, Carhart R. (1966). An expanded test for speech discrimination utilizing CNC monosyllabic words. Northwestern University Auditory Test No. 6. Tech Rep SAM-TR, 1–12. [DOI] [PubMed] [Google Scholar]
  • 33.Chang Young-Soo, Choi Jeesun, Moon Joon Il, Hong Sung Hwa, Chung Won-Ho & Cho Yang-Sun (2016) Factors associated with self-reported outcome in adaptation of hearing aid, Acta Oto-Laryngologica, 136:9, 905–911 [DOI] [PubMed] [Google Scholar]

RESOURCES