Abstract
Purpose
To evaluate small-size hamstring (HS) autografts for biomechanical properties and determine a threshold diameter necessary for appropriate reconstruction.
Methods
In a controlled laboratory setting, biomechanical testing was performed upon 15 hamstring autografts. The grafts were divided into three groups by diameter, with five grafts each at diameter sizes of 6, 7, and 8 mm. Testing of the specimens was performed using an MTS 858 (Materials Testing System, Eden Prairie, MN). We determined load to failure by looking at the maximum load as well as the stiffness of the graft. Statistical analysis was performed via analysis of variance (ANOVA) testing with Tukey's post-hoc test and P-values set at 0.05.
Results
There was a significant difference in ultimate tensile strength for the different size grafts: 1990 +/- 302.42 N for 6 mm grafts (n=5), 2179 +/- 685.36 N for the 7 mm grafts (n=5), and 3074 +/- 781 N for 8 mm grafts (n=5). This was statistically significant for the group overall (p=0.039), as well as between the 6 mm and 8 mm grafts (p=0.044). Graft stiffness for the 6 mm grafts was 317 +/- 85 N (n=5), 288.6 +/- 66 for 7 mm grafts (n=5), and 428.053 +/- 83 for 8 mm grafts (n=5). This achieved statistical significance for the group overall (p =0.037) as well as between the 8 mm and 7 mm grafts.
Conclusions
The biomechanical data presented here demonstrate that graft diameter is highly correlated with ultimate tensile strength and stiffness.
Clinical relevance
When viewing this biomechanical data in conjunction with prior clinical data, consideration should be given for the supplementation of an HS autograft as the size decreases below 8 mm.
Keywords: acl repair, hamstring autografts, biomechanical properties
Introduction
Anterior cruciate ligament (ACL) injuries are common, and surgical reconstruction of anterior cruciate ligament tears is the most frequent knee ligament reconstruction procedure performed in the United States [1]. Modern intra-articular reconstruction techniques have yielded excellent reconstructive results, however, subsequent graft tears and anterior cruciate ligament revision surgery remains a problem. Rates of revision surgery, regardless of graft used, are quoted from 5%-20%, however, the precise mode of failure and indication for revision has not been well-standardized in previous literature [2-4]. Recurrent or persistent instability is one of the leading causes of revision surgery. Young, active patients experience revision surgeries at increased rates, suggesting they are more likely to experience graft failure or are less tolerant of instability symptoms [5]. Magnusson et al. demonstrated in a clinical study that younger age and smaller graft diameter are both independently correlated with the increased rate of revision surgery, suggesting that a larger graft is less likely to need surgical revision [6].
A common reconstruction technique uses a hamstring tendon (HT) autograft for ACL graft reconstruction [7-14]. Historically, bone-patellar-tendon-bone (BPTB) grafts were thought to be biomechanically superior in resistance to tensile strain. However, HT autograft preparation techniques that increase the size of the graft have been shown to be equivocal or even superior to BPTB grafts in terms of strength [13-14]. The most frequent HT graft configuration is a quadruple strand graft consisting of a doubled gracilis tendon paired with a doubled semitendinosus graft. This quadruple strand HT graft has been shown to be biomechanically superior to BPTB grafts in tension when the mean graft diameter is 7.7 mm to 8.5 mm, which, notably, is similar in size to the BPTB graft [14-16].
Biomechanical cadaver studies have demonstrated that the mean biomechanical strength to resist the maximum tensile load of the native ACL is approximately 2160 N in young adults (age 22-35). As adults age, this maximum tensile strength has been shown to decrease. Stiffness of the native ACL, defined as the ultimate load at failure compared to the cross-sectional area of the ACL, also decreases with age [17-20].
Studies of HT autografts show the doubling of the gracilis or semitendinosus approach, with a tensile strength of 1550 and 2630 N, respectively, of the native ACL [19]. The quadruple bundle HT graft has a mean tensile strength of 4090 N, exceeding that of the native ACL. Doubling and quadrupling of hamstring grafts show obvious benefits to tensile strength. Additionally, significant heterogeneity exists in the patient population regarding hamstring autograft size, as well as harvesting and preparation techniques [18-21].
Although these prior studies demonstrate increasing biomechanical strength with increasing graft diameter, previous studies do not provide biomechanical evidence to support a specific ACL graft diameter for reconstruction. Thus, the purpose of this study is threefold: 1) to evaluate small size hamstring graft (6, 7, and 8 mm) for biomechanical qualities (ultimate tensile strength, stiffness, and creep), 2) to determine a threshold diameter of a looped graft necessary for appropriate reconstruction, and, thus, 3) to determine at what diameter should consideration be given to supplement an HS autograft.
Materials and methods
There are no conflicts of interest to disclose from the corresponding authors of this manuscript. This project was funded internally by the University of Florida. This research involves cadaveric specimen testing utilizing tabletop biomechanical devices, and thus is exempt from Institutional Review Board approval.
In a controlled laboratory setting, biomechanical testing was performed upon 15 hamstring grafts. The hamstring grafts were harvested from cadaveric specimens of ages 50-70. Harvesting of the semitendinosus and gracilis tendons was performed using standard autograft techniques. A 3-cm longitudinal incision centered between the tibial tubercle and the posteromedial border of the tibia was utilized. The sartorius fascia was divided parallel to the superior border of the semitendinosus tendon and reflected inferomedially. The semitendinosus and gracilis tendons were identified on the deep surface and harvested with a tendon stripper. All grafts were stripped of muscle and were then pre-soaked in normal saline for 20 minutes. After preparation, the diameter of the graft was identified using standard sizing tubes in an ACL combo set. The grafts were pulled through the smallest-possible diameter tube without damage to the graft. The sizing procedure was confirmed and repeated three separate times.
The 15 hamstrings grafts were divided into three groups by diameter, with five grafts each at diameter sizes of 6, 7, and 8 mm. There were five quadruple strand grafts (four of which had diameter 8 mm, one had diameter 7 mm) and 10 double-strand grafts (five with 6 mm diameter, four with 7 mm diameter, and one with 8 mm diameter). Grafts were prepared to approximate similar tension across all strands.
Testing of the specimens was performed using a validated, commercially available materials testing device (MTS 858, Materials Testing System, Eden Prairie, MN). The hamstring graft specimens were secured by the free end with clamps. They were then frozen into the clamps with dry ice before final tightening (Figure 1). The looped end of the graft was secured with a hook where weight was applied (Figure 2).
Figure 1. MTS 858 used for graft testing.
Testing of the specimens was performed using an MTS 858. The hamstring graft specimens were secured by the free end with clamps. They were then frozen into the clamps with dry ice before final tightening. Pictured is the block of dry ice holding the tendon secured to the hook.
MTS 858: Materials Testing System, Eden Prairie, MN
Figure 2. Close-up of testing device construct.
The hamstring graft specimens were secured by the free end with clamps. They were then frozen into the clamps with dry ice before final tightening. Pictured is a close-up of the block of dry ice holding the tendon secured to the hook.
In order to determine creep, we pre-tensioned the graft. We performed cyclic loading of 50-250 N @ 1 Hz for 1000 cycles. We determined load to failure by recording the peak load prior to graft rupture or subsequent load decrease, and stiffness of the graft was calculated by comparing the maximum load at failure to the cross-sectional area of the graft. All failures were mid substance. Statistical analysis was performed via analysis of variance (ANOVA) testing with Tukey’s post-hoc test and P-value set at 0.05.
Results
There was a significant difference in ultimate tensile strength (UTS) for the different size grafts: 1990+/-302.42N for 6 mm grafts (n=5), 2179 +/- 685.36N for 7 mm grafts (n=5), and 3074+/-781N for 8 mm grafts (n=5) (Figure 3, Table 1). Additionally, the difference between the 6 mm and 8 mm grafts was also found to be statistically significant (p=0.044) (Table 2).
Figure 3. Maximum tensile load vs graft diameter.
Graph depicting the maximum tensile load identified versus the graft diameter
Table 1. Average tensile load by diameter.
Table depicting the average tensile load (N) and standard deviation as determined by graft diameter in millimeters.
Max. force (N) | Size (mm) | Mean | Std. deviation | N |
6 | 1990.320 | 302.4223 | 5 | |
7 | 2179.440 | 685.3587 | 5 | |
8 | 3074.920 | 781.8207 | 5 | |
Total | 2414.893 | 758.1160 | 15 |
Table 2. Comparison of average max force across 6, 7, 8 mm graft diameters.
Table depicting the comparison of graft diameter based upon mean tension. Statistical significance was noted between the smallest (6 mm) and largest (8 mm) grafts.
Dependent variable | (I) Size | (J) Size | Mean difference (I-J) | Std. error | Sig. |
Max Force (N) | 6 | 7 | -189.120 | 395.3765 | 0.883 |
8 | -1084.600* | 395.3765 | 0.044 | ||
7 | 6 | 189.120 | 395.3765 | 0.883 | |
8 | -895.480 | 395.3765 | 0.100 | ||
8 | 6 | 1084.600* | 395.3765 | 0.044 | |
7 | 895.480 | 395.3765 | 0.100 |
Graft stiffness for 6 mm grafts was 317+/-85N (n=5), 288.6+/-66 for 7 mm grafts (n=5), and 428.053+/-83 for 8 mm grafts (n=5) (Figure 4). This achieved statistical significance for the group overall (p=0.037) (Table 3) as well as between the 8 mm and 7 mm grafts (p=0.039) (Table 4).
Figure 4. Graft stiffness by diameter.
Graph depicting graft stiffness by diameter
Table 3. Graft stiffness statistical significance.
Table comparing the statistical significance of graft stiffness measurements.
Dependent variable | Sum of squares | df | Mean square | F | Sig. | |
Stiffness (N/mm) | Contrast | 54704.841 | 2 | 27352.421 | 4.376 | 0.037 |
Error | 74999.076 | 12 | 6249.923 |
Table 4. Comparison of graft stiffness across 6, 7, and 8 mm diameters.
Table comparing graft stiffness among different size grafts with statistical significance observed between the 7/8 mm grafts
Dependent variable | (I) Size | (J) Size | Mean difference (I-J) | Std. error | Sig. |
Stiffness (N/mm) | 6 | 7 | 29.180 | 49.9997 | 0.831 |
8 | -111.000 | 49.9997 | 0.108 | ||
7 | 6 | -29.180 | 49.9997 | 0.831 | |
8 | -140.180* | 49.9997 | 0.039 | ||
8 | 6 | 111.000 | 49.9997 | 0.108 | |
7 | 140.180* | 49.9997 | 0.039 |
Graft creep was measured as the difference in length at max force (250 N) between cycle 10 and cycle 1000. This did not reach statistical significance as a group or when comparing the 6, 7, and 8 mm grafts (Tables 5-7).
Table 5. Measurement of tendon graft creep.
Table showing the average change in graft length at max load after 1000 cycles of variable loading
Graft creep | Size (mm) | Mean | Std. deviation | N |
6 | 0.66406740 | 0.282114554 | 5 | |
7 | 0.81433100 | 0.323660640 | 5 | |
8 | 0.56054680 | 0.301467034 | 5 | |
Total | 0.67964840 | 0.300445577 | 15 |
Table 7. Comparison of tendon graft creep between 6, 7, and 8 mm graft diameters.
Table displaying a comparison of the measurements of tendon graft creep between different diameters. Statistical significance was not found between any graft diameter or as a group.
Dependent variable | (I) Size | (J) Size | Mean difference (I-J) | Std. error | Sig. |
Graft Creep | 6 | 7 | -0.15026360 | 0.191564508 | 0.719 |
8 | 0.10352060 | 0.191564508 | 0.853 | ||
7 | 6 | 0.15026360 | 0.191564508 | 0.719 | |
8 | 0.25378420 | 0.191564508 | 0.409 | ||
8 | 6 | -0.10352060 | 0.191564508 | 0.853 | |
7 | -0.25378420 | 0.191564508 | 0.409 |
Table 6. Statistical analysis of measurements of tendon graft creep.
Table displays statistical analysis of measurements used to assess creep. Statistical difference was not reached as a group.
Dependent variable | Sum of squares | df | Mean square | F | Sig. | |
Graft Creep | Contrast | 0.163 | 2 | 0.081 | 0.887 | 0.437 |
Error | 1.101 | 12 | 0.092 |
Discussion
Previous data demonstrated that quadruple-strand grafts have superior strength versus maximum loads when compared to double or single strand grafts, and graft cross-sectional area has been strongly correlated with graft strength [18]. Studies comparing BPTB to HT grafts have also identified that HT grafts are equivalent in tensile strength to patellar tendon grafts if the diameter of the HT graft is of similar size [13].
Our study demonstrated that the UTS of HT grafts of diameter 8 mm was greater than or equal to that of the native ACL. In addition, the difference in UTS between 6 and 8 mm grafts was statistically significant. The difference in 6 and 7 mm grafts did not reach statistical significance although there was a trend toward suboptimal strength (Table 8). In regards to stiffness, the mean for each size was greater than the native ACL, 242±28 N/mm, as determined by Woo in his biomechanical study [19]. This could possibly be attributed to the age of the grafts as well as an effect of the cooling process.
Table 8. Summary of max force and stiffness across 6, 7, and 8 mm diameters.
Summary of the data including max force and stiffness across the graft diameters of 6, 7, and 8 mm with included 95% confidence intervals.
Dependent variable | 95% confidence interval | ||||
Size (mm) | Mean | Std. Error | Lower Bound | Upper Bound | |
Max Force (N) | 6 | 1990.320 | 279.573 | 1381.182 | 2599.458 |
7 | 2179.440 | 279.573 | 1570.302 | 2788.578 | |
8 | 3074.920 | 279.573 | 2465.782 | 3684.058 | |
Stiffness (N/mm) | 6 | 317.780 | 35.355 | 240.748 | 394.812 |
7 | 288.600 | 35.355 | 211.568 | 365.632 | |
8 | 428.780 | 35.355 | 351.748 | 505.812 |
The biomechanical data of our study would support the clinical data published by Magnusson et al., which demonstrated that smaller diameter HT grafts had increased rates of revision [6]. Grafts >8 mm had a 1.7% revision rate (n=1/58) versus 6.5% (n=9/139) in patients with grafts between 7 mm and 8 mm, and 13.6% (8 of 59) with grafts 7 mm or less in diameter. The population in the study by Magnusson notably used quadruple-strand grafts only in comparison to our study, which used both double and quadruple strand grafts. Previous literature has identified that grafts should be at least 7 mm in diameter, and our study confirms that a 7 mm graft approaches the properties of the native ACL. However, the biomechanical data presented here confirm that an 8 mm graft is stronger. This viewed in conjunction with the prior clinical data by Magnusson et al., which indicated higher revision rates for grafts smaller than 8 mm, would argue for an optimal graft diameter of at least 8 mm.
Some prior studies have also attributed gender as an independent risk factor for revision when undergoing HT autografts [22-25]. Ma et al. demonstrated that female patients undergoing an autograft HT harvest are nearly twice as likely to have grafts smaller than 8 mm in diameter in comparison with male patients (40% of females compared to 19% of males) [26]. Magnusson et al. identified that there was no difference in the ACL revision rate between male and female patients when graft diameter was equivalent [6]. Many of the studies which have reported a higher rate of revision among female patients either do not report HT graft diameter or do not control it. Thus, this information in combination with our biomechanical data could attribute smaller HT graft size as the contributing factor for the increased rate of ACL revision surgery for women [3,23-25].
The limitations of our study include our sample size of harvested tendon-grafts, the cadaver age for harvested tendons and resulting tendon variability, and the method of testing tensile strength. Additionally, prior studies demonstrate that the weakest part of a reconstruction construct are the bony fixation points [27-28]. Our graft failures were mid substance and thus should be taken into consideration when viewing ultimate tensile strength and stiffness.
Conclusions
ACL reconstruction surgery is one of the most common ligamentous reconstruction procedures. Autologous hamstring grafting is an attractive choice for reconstruction. Previous clinical data has shown smaller grafts have an increased rate of revision surgery. The biomechanical data presented here demonstrate that graft diameter is highly correlated with ultimate tensile strength and stiffness. When viewing this biomechanical data in conjunction with prior clinical data, consideration should be given for the supplementation of HS autografts that fall below 8 mm.
The content published in Cureus is the result of clinical experience and/or research by independent individuals or organizations. Cureus is not responsible for the scientific accuracy or reliability of data or conclusions published herein. All content published within Cureus is intended only for educational, research and reference purposes. Additionally, articles published within Cureus should not be deemed a suitable substitute for the advice of a qualified health care professional. Do not disregard or avoid professional medical advice due to content published within Cureus.
The authors have declared financial relationships, which are detailed in the next section.
Kevin W. Farmer declare(s) personal fees from Arthrex (Naples, FL).
Human Ethics
Consent was obtained by all participants in this study
Animal Ethics
Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.
References
- 1.An illustrated history of anterior cruciate ligament surgery. McCulloch PC, Lattermann C, Boland AL, Bach BR Jr. J Knee Surg. 2007;20:95–104. doi: 10.1055/s-0030-1248027. [DOI] [PubMed] [Google Scholar]
- 2.Revision anterior cruciate ligament reconstruction with nonirradiated fresh-frozen patellar tendon allograft. Fox JA, Pierce M, Bojchuk J, Hayden J, Bush-Joseph CA, Bach BR Jr. Arthroscopy. 2004;20:787–794. doi: 10.1016/j.arthro.2004.07.019. [DOI] [PubMed] [Google Scholar]
- 3.Current concepts review: revision anterior cruciate ligament reconstruction. George MS, Dunn WR, Spindler KP. Am J Sports Med. 2006;34:2026–2037. doi: 10.1177/0363546506295026. [DOI] [PubMed] [Google Scholar]
- 4.Revision anterior cruciate surgery with use of bone-patellar tendon-bone autogenous grafts. Noyes FR, Barber-Westin SD. J Bone Joint Surg Am. 2001;83:1131–1143. doi: 10.2106/00004623-200108000-00001. [DOI] [PubMed] [Google Scholar]
- 5.Incidence of subsequent injury to either knee within 5 years after anterior cruciate ligament reconstruction with patellar tendon autograft. Shelbourne KD, Gray T, Haro M. Am J Sports Med. 2009;37:246–251. doi: 10.1177/0363546508325665. [DOI] [PubMed] [Google Scholar]
- 6.Graft size and patient age are predictors of early revision after anterior cruciate ligament reconstruction with hamstring autograft. Magnussen RA, Lawrence JTR, West RL, Toth AP, Taylor DC, Garrett WE. Arthroscopy. 2012;28:526–5312. doi: 10.1016/j.arthro.2011.11.024. [DOI] [PubMed] [Google Scholar]
- 7.The Scandinavian ACL registries 2004-2007: baseline epidemiology. Granan LP, Forssblad M, Lind M, Engebretsen L. Acta Orthop. 2009;80:563–567. doi: 10.3109/17453670903350107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.The first results from the Danish ACL reconstruction registry: epidemiologic and 2 year follow-up results from 5,818 knee ligament reconstructions. Lind M, Menhert F, Pedersen AB. Knee Surg Sports Traumatol Arthrosc. 2009;17:117–124. doi: 10.1007/s00167-008-0654-3. [DOI] [PubMed] [Google Scholar]
- 9.Quadrupled semitendinosus anterior cruciate ligament reconstruction: 5-year results in patients without meniscus loss. Cooley VJ, Deffner KT, Rosenberg TD. Arthroscopy. 2001;17:795–800. doi: 10.1016/s0749-8063(01)90001-5. [DOI] [PubMed] [Google Scholar]
- 10.Anterior cruciate ligament reconstruction with double-looped semitendinosus and gracilis tendon graft directly fixed to cortical bone: 5-year results. Giron F, Aglietti P, Cuomo P, Mondanelli N, Ciardullo A. Knee Surg Sports Traumatol Arthrosc. 2005;13:81–91. doi: 10.1007/s00167-004-0553-1. [DOI] [PubMed] [Google Scholar]
- 11.Clinical evaluation of arthroscopically assisted anterior cruciate ligament reconstruction: patellar tendon versus gracilis and semitendinosus autograft. Ibrahim SA, Al-Kussary IM, Al-Misfer AR, Al-Mutairi HQ, Ghafar SA, El Noor TA. Arthroscopy. 2005;21:412–417. doi: 10.1016/j.arthro.2004.12.002. [DOI] [PubMed] [Google Scholar]
- 12.A 6-year follow-up of the effect of graft site on strength, stability, range of motion, function, and joint degeneration after anterior cruciate ligament reconstruction: patellar tendon versus semitendinosus and gracilis tendon graft. Keays SL, Bullock-Saxton JE, Keays AC, Newcombe PA, Bullock MI. Am J Sports Med. 2007;35:729–739. doi: 10.1177/0363546506298277. [DOI] [PubMed] [Google Scholar]
- 13.A biomechanical analysis of matched bone-patellar tendon-bone and double-looped semitendinosus and gracilis tendon grafts. Wilson TW, Zafuta MP, Zobitz M. Am J Sports Med. 1999;27:202–207. doi: 10.1177/03635465990270021501. [DOI] [PubMed] [Google Scholar]
- 14.In vivo tensile behavior of a four-bundle hamstring graft as a replacement for the anterior cruciate ligament. Wallace MP, Howell SM, Hull ML. J Orthop Res. 1997;15:539–545. doi: 10.1002/jor.1100150409. [DOI] [PubMed] [Google Scholar]
- 15.Predictors for hamstring graft diameter in anterior cruciate ligament reconstruction. Tuman JM, Diduch DR, Rubino LJ, Baumfeld JA, Nguyen HS, Hart JM. Am J Sports Med. 2007;35:1945–1949. doi: 10.1177/0363546507304667. [DOI] [PubMed] [Google Scholar]
- 16.Preoperative magnetic resonance imaging cross-sectional area for the measurement of hamstring autograft diameter for reconstruction of the adolescent anterior cruciate ligament. Bickel BA, Fowler TT, Mowbray JG, Adler B, Klingele K, Phillips G. Arthroscopy. 2008;24:1336–1341. doi: 10.1016/j.arthro.2008.07.012. [DOI] [PubMed] [Google Scholar]
- 17.Biomechanical analysis of human ligament grafts used in knee-ligament repairs and reconstructions. Noyes FR, Butler DL, Grood ES, Zernicke RF, Hefzy MS. https://www.ncbi.nlm.nih.gov/pubmed/6699049. J Bone Joint Surg Am. 1984;66:344–352. [PubMed] [Google Scholar]
- 18.Hamstring tendon grafts for reconstruction of the anterior cruciate ligament: biomechanical evaluation of the use of multiple strands and tensioning techniques. Hamner DL1, Brown CH Jr, Steiner ME, Hecker AT, Hayes WC. J Bone Joint Surg Am. 1999;81:549–557. doi: 10.2106/00004623-199904000-00013. [DOI] [PubMed] [Google Scholar]
- 19.Tensile properties of the human femur-anterior cruciate ligament-tibia complex. The effects of specimen age and orientation. Woo SL, Hollis JM, Adams DJ, Lyon RM, Takai S. Am J Sports Med. 1991;19:217–225. doi: 10.1177/036354659101900303. [DOI] [PubMed] [Google Scholar]
- 20.Anterior cruciate ligament reconstruction with multistranded autogenous semitendinosus tendon. Maeda A, Shino K, Horibe S, Nakata K, Buccafusca G. Am J Sports Med. 1996;24:504–509. doi: 10.1177/036354659602400416. [DOI] [PubMed] [Google Scholar]
- 21.Analysis of the risk factors regarding anterior cruciate ligament reconstruction using multiple-looped semitendinosus tendon. Kim SG, Kurosawa H, Sakuraba K, Ikeda H, Takazawa S, Seto H, Ishijima M. Knee. 2005;12:366–369. doi: 10.1016/j.knee.2004.10.001. [DOI] [PubMed] [Google Scholar]
- 22.Clinical comparison of intraarticular anterior cruciate ligament reconstruction using autogenous semitendinosus and gracilis tendons in men versus women. Noojin FK, Barrett GR, Hartzog CW, Nash CR. Am J Sports Med. 2000;28:783–789. doi: 10.1177/03635465000280060301. [DOI] [PubMed] [Google Scholar]
- 23.A rigorous comparison between the sexes of results and complications after anterior cruciate ligament reconstruction. Barber-Westin SD, Noyes FR, Andrews M. Am J Sports Med. 1997;25:514–526. doi: 10.1177/036354659702500415. [DOI] [PubMed] [Google Scholar]
- 24.Anterior cruciate ligament reconstruction in men and women: An outcome analysis comparing gender. Ferrari JD, Bach BR Jr, Bush-Joseph CA, Wang T, Bojchuk J. Arthroscopy. 2001;17:588–596. doi: 10.1053/jars.2001.24686. [DOI] [PubMed] [Google Scholar]
- 25.Reconstruction of the anterior cruciate ligament in females: a comparison of hamstring versus patellar tendon autograft. Barrett GR, Noojin FK, Hartzog CW, Nash CR. Arthroscopy. 2002;18:46–54. doi: 10.1053/jars.2002.25974. [DOI] [PubMed] [Google Scholar]
- 26.Can preoperative measures predict quadruple hamstring graft diameter? Ma CB, Keifa E, Dunn W, Fu FH, Harner CD. Knee. 2010;17:81–83. doi: 10.1016/j.knee.2009.06.005. [DOI] [PubMed] [Google Scholar]
- 27.Anterior cruciate ligament graft fixation. Comparison of hamstring and patellar tendon grafts. Steiner ME, Hecker AT, Brown CH Jr, Hayes WC. Am J Sports Med. 1994;22:240–247. doi: 10.1177/036354659402200215. [DOI] [PubMed] [Google Scholar]
- 28.Tendon-healing in a bone tunnel. A biomechanical and histological study in the dog. Rodeo SA, Arnoczky SP, Torzilli PA, Hidaka C, Warren RF. J Bone Joint Surg. 1993;75-A:1795–1803. doi: 10.2106/00004623-199312000-00009. [DOI] [PubMed] [Google Scholar]