Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2021 Jul 1.
Published in final edited form as: Med Phys. 2020 May 11;47(7):2931–2936. doi: 10.1002/mp.14179

Technical Note: A Deuterated 13C-Urea Reference for Clinical Multiparametric Prostate Cancer Studies Including Hyperpolarized Pyruvate

Collin J Harlan 1, Zhan Xu 1, Keith A Michel 1,2, Christopher M Walker 1, Sanjaya D Lokugama 3,4, Gary V Martinez 1, Mark D Pagel 2,3, James A Bankson 1,2,*
PMCID: PMC7375896  NIHMSID: NIHMS1592397  PMID: 32286689

Abstract

Purpose:

Metabolic MRI using hyperpolarized [1-13C]-pyruvate offers unprecedented new insight into disease and response to therapy. 13C-enriched reference standards are required to enable fast and accurate calibration for 13C studies, but care must be taken to ensure that the reference is compatible with both 13C and 1H acquisitions. The goal of this study was to optimize the composition of a 13C-urea reference for a dual-tuned 13C/1H endorectal coil and minimize imaging artifacts in metabolic and multiparametric MRI studies involving hyperpolarized [1-13C]-pyruvate.

Methods:

Due to a high amount of Gd doping for the purpose of reducing the spin-lattice relaxation time (T1) of urea, the 1H signal produced by a reference of 13C-urea in normal water was rapidly relaxed, resulting in severe artifacts in heavily T1-weighted images. Hyperintense ringing artifacts in 1H images were mitigated by reducing the 1H concentration in a 13C-urea reference via deuteration and lyophilization. Several references were fabricated and their SNR was compared using 1H and 13C imaging sequences on a 3T MRI scanner. Finally, 1H prostate phantom imaging was conducted to compare image quality and 1H signal intensity of normal and deuterated urea references.

Results:

The deuterated 13C-urea reference provides strong 13C signal for calibration and an attenuated 1H signal that does not interfere with heavily T1-weighted scans. Deuteration and lyophilization were fundamental to the reduction of 1H signal and hyperintense ringing artifacts. There was a 25-fold reduction in signal intensity when comparing the non-deuterated reference to the deuterated reference, while the 13C signal was unaffected.

Conclusion:

A deuterated reference reduced hyperintense ringing artifacts in 1H images by reducing the 1H signal produced from the 13C-urea in the reference. The deuterated reference can be used to improve anatomical image quality in future clinical 1H and hyperpolarized [1-13C]-pyruvate MRI prostate imaging studies.

Keywords: MRI, hyperpolarized pyruvate, metabolic imaging

Introduction

The ability to manage and predict outcome for patients with prostate cancer is a critically important unmet clinical need. Currently, a crucial goal for prostate cancer imaging is accurate disease characterization, which can ultimately be achieved through a combination of multi-modality information gathered by minimally invasive imaging techniques.1 The metabolic properties of localized prostate cancer are of great interest due to the high rate of aerobic glycolysis undergone by prostate tumors.2,3 Hyperpolarized (HP) 13C magnetic resonance imaging (MRI) is an emerging metabolic imaging method that can be used to quantify the conversion of HP pyruvate into lactate as an indicator for glycolytic flux in vivo.4 HP 13C MRI can be paired with dynamic contrast-enhanced (DCE-) MRI to inform on vascular function and enhance quantitative analysis of 13C signal evolution.4 HP 13C MRI and DCE-MRI have the ability to provide attractive multiparametric imaging for patients with prostate cancer. In addition to standard metabolic information provided by HP [1-13C]-pyruvate for prostate cancer studies,5,6 anatomical and DCE-MRI can be performed during the same examination, without moving the patient.

Hyperpolarization of 13C labeled agents can result in a greater than 100,000-fold gain in signal intensity compared to thermal equilibrium;7 it also enables real-time, in vivo observation of biochemical processes, thus allowing investigators the ability to probe specific enzymatic pathways. 13C-labeled pyruvate and its downstream metabolites have been shown to be of interest regarding the characterization of tumor metabolism.810 Prior to injection of HP [1-13C]-pyruvate, it is necessary to calibrate to the 13C center frequency and transmit power for imaging. This can be accomplished by using a 13C-labeled urea reference placed inside the coil.11 The most common way to create a 13C-urea reference is by dissolving 13C-urea in H2O. In this way, the reference can provide 13C and 1H signals, which can be detected using 13C/1H dual-tuned coils.

Currently, dual-tuned 13C/1H endorectal coils (ERC) are used for clinical assessment of HP [1-13C]-pyruvate in the setting of prostate cancer,5,1214 as the use of dual-tuned 13C/1H coils improves acquisition efficiency, quantitative analysis, and allows for the collection of 1H images and 13C spectral data.15 A 13C-urea reference standard can be integrated into 13C/1H ERC. However, due to the close proximity to coil elements and high amount of Gd doping for the purpose of reducing the spin-lattice relaxation time (T1) of 13C-urea, the 1H signal produced by the reference is rapidly relaxed and can be much greater than that of nearby tissues. As a result, the large signal produced by the reference can lead to imaging artifacts that greatly reduce 1H image quality, thus limiting the diagnostic utility of 1H anatomical scans. Therefore, the purpose of this research was to develop a dual-tuned 13C-urea reference containing optimized reagents. Optimal 13C-urea reference fabrication was ultimately achieved by replacing H2O with D2O and reducing the 1H signal of urea by deuteration and lyophilization of 13C-urea in D2O.16,17

Materials and Methods

13C-Urea Reference Fabrication

Five candidate 13C-urea references were created for comparison to determine the optimal mixture that would reduce 1H signal intensity. Three references were fabricated by dissolving 244 mg of non-lyophilized, once lyophilized, or twice lyophilized 13C-urea (Cambridge Isotope Laboratories, Andover, MA, USA, Item Number: CLM-311-PK) to create 4M 13C-urea solutions in 1.00 mL D2O (99.9 atom % D, Sigma-Aldrich, Milwaukee, WI, USA, SKU: 151882). A fourth reference was fabricated by dissolving 244 mg non-lyophilized 13C-urea in 1.00 mL distilled H2O. The fifth reference was comprised by dissolving 488 mg 13C-urea to create an 8M 13C-urea reference in a mixture of 0.10 mL glycerol and 0.90 mL H2O. Following lyophilization of 13C-urea in D2O, all five references were also mixed with 0.65 mg of NaN3 (Fisher Scientific, Fair Lawn, NJ, USA, Catalog Number: S227I-25) to obtain a 10 mM NaN3 concentration and 3 μL 1 mmol/mL Gadovist contrast agent (Bayer HealthCare Pharmaceuticals Inc., Whippany, NJ, USA). It is important to note that the addition of Gadovist contributed to additional 1H signal in the references. After vortexing, each sample was added to a 1 mL syringe, which was sealed with hot glue.

Lyophilization was performed as follows: two aliquots of 244 mg 13C-urea were dissolved in 1.00 mL D2O each. After approximately one hour in room temperature, both samples were frozen in liquid nitrogen and lyophilized overnight with a freeze dryer (Labconco FreeZone 4.5 Liter −105°C, Labconco Corporation, Kansas City, MO, USA, Catalog Number: 7382020). This process was repeated for one sample after redissolving in 1.00 mL D2O, resulting in a twice lyophilized sample of 13C-urea.

Magnetic Resonance Imaging Acquisition

All images were acquired with a GE Discovery 3T scanner (MR750, GE Healthcare, Waukesha, WI, USA). To begin, the non-lyophilized urea in H2O, non-lyophilized urea in D2O, once lyophilized, and twice lyophilized deuterated urea references were scanned together in order to compare their relative signal intensity. It was quickly determined that the non-lyophilized urea in H2O and in D2O references severely dominated the dynamic range of the image. Those references were removed and the scan was repeated for the lyophilized references in D2O. The references were scanned using a 1H 3D fast spoiled gradient echo sequence (FSPGR), similar to our clinical DCE-MRI sequence, inside a 13C/1H volume resonator with a 35-mm inner diameter (RAPID Biomedical GmbH, Rimpar, Germany). The 1H 3D FSPGR images were acquired with a 3.87 ms repetition time, 1.41 ms echo time, 22 cm × 22 cm field of view, 3.6 mm slice thickness, and a 20-degree excitation angle.

A 13C gradient echo-planar imaging (EPI) sequence18 was performed for the reference configuration containing all four references to ensure that the references provided similar 13C signal, regardless of if they produced 1H imaging artifacts. 13C EPI was acquired with a 1000 ms repetition time, 25 ms echo time, 6 cm × 6 cm field of view, a single NEX, 40 mm slice thickness, and a 90-degree excitation angle centered on the resonance frequency for 13C-urea, as determined from a non-selective pulse-acquire spectrum.

Next, 1H 3D FSPGR of a prostate phantom (Figure 1) was acquired using the dual tuned 13C/1H ERC (RAPID Biomedical GmbH, Rimpar, Germany) containing the non-deuterated reference (488 mg 13C urea dissolved in a mixture of 0.10 mL glycerol and 0.90 mL H2O). This urea reference was then replaced with the twice lyophilized deuterated reference and the 1H 3D FSPGR scan was repeated. The 1H 3D FSPGR of a prostate phantom for both the non-deuterated urea reference and twice lyophilized deuterated urea reference were acquired with a 3.07 ms repetition time, 1.34 ms echo time, 22 cm × 22 cm field of view, 3.6 mm slice thickness, left-right frequency encoding direction, and a 20-degree excitation angle. To permit quantitative comparison of the relative signal from these references, the mean signal from a region-of-interest (ROI) encompassing the reference was normalized by the root-mean-square noise amplitude measured in an artifact-free background region of the image.

Figure 1.

Figure 1.

Endorectal coil, 13C-urea reference, and prostate phantom setup. (a) The 13C-urea sample, located inside of a 1 mL syringe, can easily be integrated into the ERC. (b) The head of the ERC contains several regions that house coil elements, as well as a bored out region in the center of the coil head where the 13C-urea reference resides. (c) A prostate phantom setup was used to compare the signal intensity of the non-deuterated and deuterated 13C-urea references.

13C T1 Measurements

13C T1 measurements were collected using the GE Discovery 3T scanner inside the 13C/1H volume resonator. Four reference samples were measured individually: non-lyophilized urea in H2O without Gd, non-lyophilized urea in H2O with Gd, twice lyophilized urea in D2O without Gd, and twice lyophilized urea in D2O with Gd. A non-selective spectroscopy method was used to acquire variable flip angle measurements for calculating the T1 for 13C-urea in these mixtures.

Results

1H 3D FSPGR MRI acquired of the ERC embedded with the non-deuterated urea reference inside a prostate phantom (Figure 2) showed hyperintense ringing artifacts in both the phase and frequency encoding directions. This reference yielded a normalized signal of 2.77 × 103.

Figure 2.

Figure 2.

1H 3D FSPGR MRI acquired with ERC containing the non-deuterated 13C-urea reference inside a prostate phantom. There is a clear 1H hyperintense ringing artifact produced by relaxed 13C-urea in H2O. The red ROI is the area used to calculate the normalized signal intensity of the reference.

1H 3D FSPGR MRI of all references (Figure 3a), acquired using the 13C/1H volume resonator, demonstrated the hyperintensity of the non-lyophilized urea in distilled H2O (labeled as reference four in Figure 3a) as it dominated the dynamic range compared to the other three urea in D2O references, especially the two lyophilized references (labeled as reference one and two in Figure 3a). 1H 3D FSPGR MRI of the non-lyophilized, once lyophilized, and twice lyophilized urea in D2O references (Figure 3c) demonstrated that even after removing the non-lyophilized urea in distilled H2O reference, the dynamic range was dominated by the non-lyophilized urea in D2O reference (labeled as reference three in Figure 3c). 1H 3D FSPGR MRI of the once lyophilized and twice lyophilized references (labeled as reference one and two in Figure 3d) was the first image where more than one reference was clearly visible (Figure 3d). 13C T1 measurements showed that deuteration of 13C-urea increased the 13C T1 from 37.7s to approximately 96.7s, but this effect was mitigated by the addition of Gd which reduced the 13C T1 to approximately 0.474s (compared to 0.545s for non-lyophilized urea in H2O). 13C EPI of all references (Figure 3b) showed that the 13C signals detected from the references were comparable. The normalized 1H signal intensity of the non-lyophilized urea in H2O, non-lyophilized urea in D2O, once lyophilized, and twice lyophilized urea references in D2O were 2.80 × 104, 2.84 × 103, 5.65 × 102, and 3.62 × 102, respectively.

Figure 3.

Figure 3.

1H 3D FSPGR MRI and 13C EPI of the samples using a 13C/1H volume resonator. (a) 1H MRI was conducted for the non-lyophilized 13C-urea in distilled H2O (four), non-lyophilized 13C-urea in D2O (three), twice lyophilized 13C-urea in D2O (two), and once lyophilized 13C-urea in D2O (one) reference group. (b) 13C EPI was conducted for the non-lyophilized 13C-urea in distilled H2O, non-lyophilized 13C-urea in D2O, twice lyophilized 13C-urea in D2O, and once lyophilized 13C-urea in D2O reference group in the same position as (a). (c) 1H MRI was conducted for the non-lyophilized 13C-urea in D2O, twice lyophilized 13C-urea in D2O, and once lyophilized 13C-urea in D2O reference group. (d) 1H MRI was conducted for the twice lyophilized and once lyophilized 13C-urea in D2O references.

1H 3D FSPGR MRI acquired of the ERC embedded with the twice lyophilized reference (Figure 4) showed reduction of the hyperintense ringing artifacts seen in Figure 2. The normalized signal intensity of the twice lyophilized reference was 1.13 × 102. Therefore, the signal intensity of the non-deuterated urea reference compared to the twice lyophilized deuterated reference showed a reduction by a factor of approximately 25.

Figure 4.

Figure 4.

1H 3D FSPGR MRI acquired with ERC containing the deuterated urea reference inside a prostate phantom. The hyperintense ringing artifact has been reduced as a result of the twice lyophilized deuterated reference replacing the non-deuterated urea reference. The red ROI is the area used to calculate the normalized signal intensity of the reference.

Discussion

Replacement of H2O with D2O in the reference was a fundamental first step toward the reduction of hyperintense ringing artifacts produced during 1H imaging. However, this action alone did not solve the problem. The exchange of the 1H located on the urea molecule with the surrounding D2O creates H2O in the surrounding solvent. Therefore, urea in D2O alone is insufficient to reduce the hyperintense ringing artifacts because there is still a substantial 1H signal that arises from the urea and surrounding D2O solvent. [D4, 13C]-urea can be created in the laboratory and is a much better alternative when dissolved in D2O because it does not increase 1H signal in the reference. In this work, we found that two cycles of exchange in D2O were sufficient to attenuate the heavily doped 1H signal in the reference to be consistent with signal observed in a tissue-mimicking phantom.

It is important to note that two rounds of lyophilization, in order to achieve optimal 1H signal reduction, is not required. This experiment was carried out by performing two rounds of deuteration and lyophilization of 244 mg 13C-urea in 1 mL of D2O to demonstrate the progressive reduction in 1H signal. However, the same fractional enrichment of amine hydrogen atoms could be achieved by a single lyophilization of 244 mg 13C-urea in 8.75 mL of 99.9% D2O.

The twice lyophilized deuterated urea reference successfully reduced the 1H imaging artifacts produced by the highly Gd doped, undeuterated 13C-urea reference, to below the noise floor of our FSPGR acquisition. The hyperintense ringing artifacts observed from the undeuterated reference limit the diagnostic quality of the 1H images and DCE-MRI scans acquired in conjunction with HP 13C MRI. The twice lyophilized reference was able to successfully reduce the mean signal intensity, which led to the reduction of hyperintense ringing artifacts. Furthermore, the 13C-urea signal from the twice lyophilized reference was unaffected, allowing it to continue to serve as a calibration reference for future 13C studies. The 13C T1 of non-lyophilized urea in H2O with Gd and twice lyophilized urea in D2O with Gd were comparable, and short enough that either can successfully be used as a convenient reference for 13C calibrations. The deuterium-enriched urea reference is a fundamental 13C calibration tool which will not produce imaging artifacts during 1H MRI and thus will support the acquisition of anatomical and functional data to compliment HP 13C imaging.

The twice lyophilized urea reference is not only useful for prostate cancer imaging and implementation in ERC but can also be applied to HP 13C imaging for other anatomical sites and coil configurations. Any laboratory that uses dual-tuned 13C/1H coils for anatomic, functional, and HP 13C MRI could integrate a similar calibration reference to reduce imaging artifacts that stem from excessive 1H signal in their reference. Uncompromised 1H imaging will allow for standard of care imaging alongside HP 13C imaging studies. Furthermore, multiparametric 1H studies can improve the quantification of 13C.4

Conclusion

We designed a reference standard for calibration of 13C scans that does not compromise anatomical or functional 1H MRI. The deuterated urea references showed a reduction in signal intensity when compared to a non-deuterated reference embedded in the ERC, and reduced 1H ringing artifacts in heavily T1-weighted images. The deuterium-enriched 13C-urea reference can be used to improve anatomical image quality in future clinical 1H and HP [1-13C]-pyruvate MRI prostate cancer imaging studies.

Acknowledgements

This work was supported by funding from the National Cancer Institute of the National Institutes of Health (R01CA211150, P30CA016672) and GE Healthcare. The content is solely the responsibility of the authors and does not necessarily represent the official views of the funding agencies.

The authors would also like to thank Dr. Anna Romanowska-Pawliczek for help preparing this manuscript.

Footnotes

Conflict of Interest

The authors have no relevant conflicts of interest to disclose.

References

  • 1.Kelloff GJ, Choyke P, Coffey DS, Prostate Cancer Imaging Working G. Challenges in clinical prostate cancer: role of imaging. AJR Am J Roentgenol. 2009;192(6):1455–1470. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Board M, Humm S, Newsholme EA. Maximum activities of key enzymes of glycolysis, glutaminolysis, pentose phosphate pathway and tricarboxylic acid cycle in normal, neoplastic and suppressed cells. Biochem J. 1990;265(2):503–509. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Eidelman E, Twum-Ampofo J, Ansari J, Siddiqui MM. The Metabolic Phenotype of Prostate Cancer. Front Oncol. 2017;7:131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Bankson JA, Walker CM, Ramirez MS, et al. Kinetic Modeling and Constrained Reconstruction of Hyperpolarized [1–13C]-Pyruvate Offers Improved Metabolic Imaging of Tumors. Cancer Res. 2015;75(22):4708–4717. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Nelson SJ, Kurhanewicz J, Vigneron DB, et al. Metabolic imaging of patients with prostate cancer using hyperpolarized [1-(1)(3)C]pyruvate. Sci Transl Med. 2013;5(198):198ra108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Aggarwal R, Vigneron DB, Kurhanewicz J. Hyperpolarized 1-[(13)C]-Pyruvate Magnetic Resonance Imaging Detects an Early Metabolic Response to Androgen Ablation Therapy in Prostate Cancer. Eur Urol. 2017;72(6):1028–1029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ardenkjaer-Larsen JH, Fridlund B, Gram A, et al. Increase in signal-to-noise ratio of > 10,000 times in liquid-state NMR. Proc Natl Acad Sci U S A. 2003;100(18):10158–10163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Day SE, Kettunen MI, Gallagher FA, et al. Detecting tumor response to treatment using hyperpolarized 13C magnetic resonance imaging and spectroscopy. Nature Medicine. 2007;13:1382. [DOI] [PubMed] [Google Scholar]
  • 9.Gallagher FA, Kettunen MI, Day SE, et al. Magnetic resonance imaging of pH in vivo using hyperpolarized 13C-labelled bicarbonate. Nature. 2008;453(7197):940–943. [DOI] [PubMed] [Google Scholar]
  • 10.Merritt ME, Harrison C, Storey C, Jeffrey FM, Sherry AD, Malloy CR. Hyperpolarized 13C allows a direct measure of flux through a single enzyme-catalyzed step by NMR. Proc Natl Acad Sci U S A. 2007;104(50):19773–19777. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Wilson DM, Keshari KR, Larson PE, et al. Multi-compound polarization by DNP allows simultaneous assessment of multiple enzymatic activities in vivo. J Magn Reson. 2010;205(1):141–147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Gordon JW, Chen HY, Autry A, et al. Translation of Carbon-13 EPI for hyperpolarized MR molecular imaging of prostate and brain cancer patients. Magn Reson Med. 2019;81(4):2702–2709. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Kurhanewicz J, Vigneron DB, Ardenkjaer-Larsen JH, et al. Hyperpolarized (13)C MRI: Path to Clinical Translation in Oncology. Neoplasia. 2019;21(1):1–16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Larson PEZ, Chen H- Y, Gordon JW, et al. Investigation of analysis methods for hyperpolarized 13C-pyruvate metabolic MRI in prostate cancer patients. NMR in Biomedicine. 2018;31(11):e3997. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Gordon JW, Fain SB, Niles DJ, Ludwig KD, Johnson KM, Peterson ET. Simultaneous imaging of 13C metabolism and 1H structure: technical considerations and potential applications. NMR Biomed. 2015;28(5):576–582. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Englander SW, Sosnick TR, Englander JJ, Mayne L. Mechanisms and uses of hydrogen exchange. Curr Opin Struct Biol. 1996;6(1):18–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Liepinsh E, Otting G. Proton exchange rates from amino acid side chains--implications for image contrast. Magn Reson Med. 1996;35(1):30–42. [DOI] [PubMed] [Google Scholar]
  • 18.Gordon JW, Vigneron DB, Larson PE. Development of a symmetric echo planar imaging framework for clinical translation of rapid dynamic hyperpolarized (13) C imaging. Magn Reson Med. 2017;77(2):826–832. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES