IN VIVO GLX AND GLU MEASUREMENTS FROM GABA-EDITED MRS AT 3T

Tiffany Bell 1,2, Elodie S Boudes 1,2, Rachelle S Loo 1,2, Gareth J Barker 3, David J Lythgoe 3, Richard AE Edden 4,5, R Marc Lebel 6, Martin Wilson 7, Ashley D Harris 1,2
PMCID: PMC7384936  NIHMSID: NIHMS1570322  PMID: 31990112

Abstract

In vivo quantification of glutamate (Glu) and GABA using magnetic resonance spectroscopy (MRS) is often achieved using two separate sequences; a short-echo PRESS acquisition for Glu and a MEGA-PRESS acquisition for GABA. The purpose of this study was to examine the agreement of Glu and Glx (the combined signal of glutamate + glutamine) quantified from two different GABA-edited MEGA-PRESS acquisitions (GABA+ TE=68 ms, and macromolecule suppressed, MMSup, TE=80 ms) with Glu and Glx quantified from a short-echo PRESS (PRESS-35, TE=35 ms) acquisition. Fifteen healthy male volunteers underwent a single scan session in which data were acquired using the three acquisitions (GABA+, MMSup and PRESS-35) in both the sensorimotor and anterior cingulate cortices using a voxel size of 3 × 3 × 3 cm3. Glx and Glu were quantified from the MEGA-PRESS data using both the OFF sub-spectra and the difference (DIFF) spectra. Agreement was assessed using correlation analyses, Bland-Altman plots and intraclass correlation coefficients (ICC). Glx quantified from the OFF sub-spectra from both the GABA+ and MMSup acquisitions showed poor agreement with PRESS-35 in both brain regions. In the sensorimotor cortex, Glu quantified from the OFF sub-spectra of GABA+ showed moderate agreement with PRESS-35 data, but this finding was not replicated in the anterior cingulate cortex. Glx and Glu quantified using the DIFF spectra of either MEGA-PRESS sequence were in poor agreement with the PRESS-35 data in both brain regions. In conclusion, Glx and Glu measured from MEGA-PRESS data generally showed poor agreement with Glx and Glu measured using PRESS-35.

Keywords: Glx, Glutamate, PRESS, MEGA-PRESS, GABA-edited MRS

Graphical Abstract

Glutamate and Glx (Glutamate+Glutamine) quantified from the off sub-spectrum and difference spectrum from two different MEGA-PRESS protocols (GABA+, TE=68ms, and macromolecule suppressed, MM-Sup, TE=80ms) generally shows poor agreement with Glx and Glutamate measured using PRESS (TE=35ms).

graphic file with name nihms-1570322-f0001.jpg

INTRODUCTION

There is increasing interest in measuring both glutamate (Glu) and γ-aminobutyric acid (GABA) in the human brain in vivo due to their prominent roles in brain function. It is well accepted that these two neurotransmitters maintain the excitation/inhibition balance which is necessary for complex cognitive functions. Both Glu and GABA levels are related to cognitive functioning1 and learning and plasticity 2,3, and altered GABA and Glu levels have been shown in numerous disorders such as schizophrenia,4 autism,5 psychosis6 and Tourette’s syndrome.7

Both Glu and GABA can be measured in vivo with proton magnetic resonance spectroscopy (MRS). Glu is most commonly measured using a standard short-echo point resolved spectroscopy sequence (PRESS, TE=30–35 ms). This sequence is chosen due to its availability and ease of implementation, relatively short scan time, high signal-to-noise ratio, and because it can provide data on multiple other metabolites.8 In a typical PRESS spectrum, Glu appears as three multiplets at 3.75 ppm, 2.3 ppm and 2.1 ppm, and these resonances overlap with signals of other metabolites, in particular glutamine (Gln). Due to the challenges in separating Glu from Gln at 3 T, they are often quantified as Glx (the combined signal of Glu + Gln).9

GABA is at much lower concentration in the brain and the GABA resonances are overlapped by resonances from metabolites of higher concentration (such as creatine at 3.0 ppm), therefore the Mescher-Garwood PRESS (MEGA-PRESS) editing technique10 is typically used. GABA also appears as three multiplets in the spectrum (3.0 ppm, 2.3 ppm and 1.9 ppm) which are coupled to one another. MEGA-PRESS consists of alternating two acquisition schemes which exploit this coupling to generate “ON” and “OFF” sub-spectra. The “ON” sub-spectrum is generated by applying an editing pulse at 1.9 ppm. As the GABA resonance at 1.9 ppm is coupled to the 3.0 ppm resonance, this pulse modulates the evolution of the 3.0 ppm GABA resonance, without affecting the overlapping creatine. The “OFF” sub-spectra are generated by applying this pulse elsewhere (usually at 7.5 ppm) so that the evolution of the GABA resonance (and those of other metabolites, including creatine) are not affected by this pulse. Subtraction of the ON sub-spectrum from the OFF sub-spectrum to produce the difference spectrum (DIFF) removes resonances that are not affected by the 1.9 ppm editing pulse, therefore revealing the GABA at 3.0 ppm.10,11

Though the majority of metabolites are unaffected by the 1.9 ppm editing pulse, this pulse also affects a macromolecule resonance at 1.7 ppm which is coupled to a macromolecule resonance at 3.0 ppm. Therefore, in a typical GABA-edited MEGA-PRESS acquisition, macromolecules “co-edit” with GABA (i.e. their behaviour is similarly affected); they are therefore also present in the DIFF spectrum at 3.0 ppm and are thought to account for roughly half of the GABA signal. For this reason, this GABA measure is often referred to as GABA+, to represent GABA+macromolecules.12 An alternative approach is the macromolecule-suppressed (MMSup) MEGA-PRESS sequence. In this sequence, editing pulses are placed symmetrically around the 1.7 ppm macromolecule resonance peak (1.9 ppm for the ON sub-spectrum and 1.5 ppm for the OFF sub-spectrum). Therefore, the 1.7 ppm macromolecule signal is equally refocused in both the ON and OFF sub-spectra so the macromolecules resonances cancel out in the DIFF spectra.12 The use of more selective editing pulses (20 ms for MMSup versus 14 ms for GABA+) requires a longer echo time (TE=80 ms) compared to TE = 68 ms typically used in GABA+ measures. The MMSup acquisition provides a “purer” measure of GABA, but the MMSup GABA signal is roughly 50% smaller than the GABA+ signal and subsequently has a lower signal-to-noise ratio (SNR) and a higher model fit error. Additionally, the more selective editing pulses used in MMSup MEGA-PRESS are more sensitive to frequency drift.13 Therefore, the selection of sequence to measure GABA (GABA+ versus MMSup) will depend on study-specific factors such as the region of interest (which will affect overall data quality) and the maximum acceptable scan time (which will affect SNR).

Due to the need to use a specialised sequence to measure GABA, GABA and Glu (or Glx) are often measured with separate acquisitions. However, due to the complementary roles of GABA and Glu (GABA is an inhibitory neurotransmitter, Glu is an excitatory neurotransmitter), as well as the fact that Glu is a precursor to GABA,14 the ability to measure GABA and Glu simultaneously is relevant to many studies. Though not designed to detect Glu (or Glx), this information is nevertheless present in MEGA-PRESS data. The MEGA-PRESS OFF sub-spectrum is similar to a PRESS spectrum (i.e., contains Glu signal), but has a longer echo time to accommodate the editing pulses. Therefore, the OFF sub-spectrum can be used to quantify Glu. Additionally, the editing pulses are not perfectly selective. Consequently, the editing pulse applied at 1.9 ppm for the ON sub-spectrum also edits the Glu and Gln resonances at 2.0–2.1 ppm, resulting in a co-edited Glu (and Glx) resonance at 3.75 ppm in the DIFF spectrum.10,11 This co-edited Glu (and Glx) resonance has been shown to be stable across scan sessions.15 Several studies have reported Glu or Glx measures using both the OFF sub-spectra and the DIFF spectra (e.g 1618), however there has been little validation for either method.

Using a phantom, Nezhad et al. (2017) demonstrated a linear relationship between the known concentration of Glu and its measurement from the DIFF spectrum of a MEGA-PRESS sequence.19 This represents ideal conditions in which there is very little noise, no other metabolites or macromolecules present and no subject movement; in vivo Glu is more difficult to quantify. Indeed, van Veenendaal et al. (2018) showed strong correlations between levels of Glu and Glx measured using both the OFF sub-spectrum and the DIFF spectrum of a GABA+ sequence compared to known concentrations in a phantom. However, when comparing in vivo measures of Glu and Glx acquired using MEGA-PRESS with measures acquired using PRESS, the correlations were substantially weaker.20 Van Veenendaal et al. looked at Glu and Glx measured from a GABA+ sequence in the occipital-parietal lobe, however agreement may differ with voxel location and between MEGA-PRESS implementations (GABA+ versus MMsup).

The objective of the current study is to determine the agreement of Glu and Glx measured from the OFF and DIFF spectra from two GABA-edited MEGA-PRESS sequences (GABA+, TE=68 ms, and MMSup, TE=80 ms) with Glu and Glx measured from a typical short echo PRESS sequence (TE=35 ms, herein referred to as PRESS-35). The fundamental motivation for the study was the need for shorter scanning times when studying clinical populations, which can be achieved if only one of the two approaches is required. Though other sequences may be more accurate for measuring Glu (e.g. STEAM, Carr-Purcell PRESS, TE averaged PRESS21, short echo phase-rotated STEAM22), PRESS-35 was chosen as it is the most popular sequence for clinical use due to its availability on MR scanners, relatively short scan time and its ability to measure multiple metabolites at once. Therefore, comparison of Glu and Glx values from MEGA-PRESS data (both the OFF-sub-spectra and the DIFF spectra) with PRESS-35 data will provide information on whether a separate PRESS sequence is needed to measure Glu, or whether MEGA-PRESS alone can be employed.

While this manuscript is not directly concerned with the measurement of GABA, we will follow standard abbreviations in the field; as a convention, in this manuscript we will refer to the Glu and Glx measurements using the OFF sub-spectra from the GABA+ MEGA-PRESS data as OFF-GABA+, and the measurements from the difference spectra from the GABA+ MEGA-PRESS data as DIFF-GABA+. Similarly, we will refer to the measurements using the OFF and the DIFF spectra from the MMSup MEGA-PRESS data as OFF-MMsup and DIFF-MMsup. Finally, as mentioned above, we will refer to the typical PRESS data acquired with an echo time of 35 ms as PRESS-35.

METHODS

Participants

Fifteen healthy, right-handed male participants between the ages of 18–30 were recruited. Data were collected with approval from the local research ethics board and with written informed consent from volunteers.

Data Acquisition

Data were collected on a 3T GE MR750w scanner (General Electric Healthcare, USA) with a 32-channel head coil. The scanning protocol included acquisition of a T1-weighted BRAVO (BRAin VOlume imaging) scan for voxel placement (TR/TE=7.3/2.7 ms, 1 mm3 isotropic voxels, flip angle=10°, TI=600 ms). Voxels were placed in the anterior cingulate cortex and the left sensorimotor cortex (Figure 1). The anterior cingulate cortex voxel was placed anterior to the genu of the corpus callosum, with the bottom edge parallel to the anterior cingulate-posterior cingulate line. The sensorimotor cortex voxel was centred at the hand-knob of the motor cortex and rotated such that the coronal and sagittal planes aligned with the cortical surface.23 Voxel rotation for the sensorimotor voxel was achieved using a separate imaging prescription (that allows for rotation in all planes) to obtain a rotation and linking it to the voxel which thus enables an MRS voxel to be rotated in all planes.

Figure 1:

Figure 1:

Voxel location and example spectra from a single volunteer in the sensorimotor cortex (A) and the anterior cingulate cortex (B). GABA and Glx resonances in the DIFF spectra have been scaled for visualisation.

The MEGA-PRESS implementation used was developed at King’s College London and kindly made available by the co-authors from this institution. For each voxel location, two separate GABA-edited measurements were made, first, GABA+24 (14 ms editing pulses placed at 1.9 ppm and 7.46 ppm in the ON and OFF conditions, respectively; TR/TE=1800/68 ms) and second, MMSup25 (20 ms editing pulses placed at 1.9 ppm and 1.5 ppm in the ON and OFF conditions, respectively; TR/TE=1800/80 ms). Both sequences used a voxel size of 3×3×3 cm3, 320 averages and 4096 data points sampled at 5 kHz. This was followed by the PRESS-35 sequence (voxel size=3×3×3 cm3; TR/TE=1800/35 ms; 64 averages; 4096 data points). Sixteen acquisitions without water suppression were also acquired for each sequence.

Data Processing

The FID-A toolbox26 was used to apply the following pre-processing steps to all data: coil combination, removal of motion-corrupted averages, frequency drift correction and zero order phase correction. FID-A was also used to isolate and average the OFF sub-spectra and calculate the DIFF spectra from the MEGA-PRESS data. LCModel Version 6.3–1J27 was used to apply eddy current correction and quantification relative to water for all spectra. To account for the T2 decay of water at each echo time the ATTH2O parameter settings for each acquisition were as follows: MMSup=0.37, GABA+=0.43, PRESS-35=0.65, as per the LCModel manual. The concentration of water (WCONC) was set to 55510. Basis sets for quantification were simulated using the FID-A toolbox based on exact sequence timings and RF pulse shapes for each sequence. Basis sets for the OFF-GABA+, OFF-MMSup and PRESS-35 data included the following metabolites: alanine, aspartate, glycerophosphocholine, phosphocholine, creatine, phosphocreatine, GABA, Glu, Gln, lactate, inositol, n-acetyl aspartate, n-acetylaspartylglutamate, scyllo-inositol, glutathione, glucose and taurine. Basis sets for the DIFF-GABA+ and DIFF-MMSup data included: GABA, Glu, Gln, n-acetyl aspartate, n-acetylaspartylglutamate and glutathione. Additionally, the setting ‘mega-press-3’ was used for the DIFF-GABA+ and DIFF-MMSup data. Macromolecules were fitted using the default parameterized macromolecule resonances in LCModel. All other LCModel parameters (in particular ATTMET=1.0) were left at their default values.

A voxel mask was generated and segmented using the tissue segmentation functions in ‘Gannet’.28 The grey matter, white matter and cerebrospinal fluid fractions were subsequently used to perform tissue correction as per the Gasparovic method29 according to the equation:

Cmetab=metLCMx(fGMxRGM+fWMxRWM+fCSFxRCSF)(1RCSF)xRmet

where

Rx=eTRT2xx(1eTET1x)

and

fx=fx_volxdxfGM_volxdGM+fWM_volxdWM+fCSF_volxdCSF

Cmetab is the molal concentration of a given metabolite, metLCM is the metabolite level obtained from LCModel, T1x and T2x are the assumed T1 and T2 relation times for each tissue type and metabolite, TR and TE are the repetition and echo times, respectively, fx_vol is the tissue fraction obtained from the segmentation analysis, dx refers to the density of MR-visible water (GM = 0.78, WM = 0.65, CSF = 0.97).

Data quality was assessed by visual inspection and metabolite linewidth; spectra with a linewidth over 0.07 ppm (9 Hz) were excluded based on the data quality recommendations as per Kreis.30

Statistical Analysis

Statistical analyses were performed using SPSS (IBM Corp. Released 2017. IBM SPSS Statistics for Macintosh, Version 25.0. Armonk, NY: IBM Corp). Correlation analysis was used as an initial step to assess the strength of the relationship between Glx and Glu measured using the two MEGA-PRESS acquisitions compared to Glx and Glu measured from PRESS-35. Agreement between Glx and Glu from the MEGA-PRESS data compared to Glx and Glu from the PRESS-35 data was assessed using Bland-Altman plots and the intraclass correlation coefficient (ICC).31 Bland-Altman plots were used to visualise differences between the two values32 and were used in addition to regression analysis to test for a systematic and proportional bias.33 The ICC is a modification of the Pearson correlation coefficient that takes into account both the degree of correlation and the overall agreement between measurements. A mixed-effects ICC model was used and the level of agreement was assessed using the range of the 95% confidence intervals, values below 0.5 are considered poor agreement, values between 0.5–0.7 are considered moderate agreement, values above 0.7 are considered good agreement.34

RESULTS

Data Quality

See Table 1 for quality metrics. All spectra acquired from the sensorimotor cortex had sufficient quality for further analysis. Quality inspection resulted in exclusion of PRESS-35 spectra from the anterior cingulate cortex data in three participants. Subsequently all data for these participants were excluded. Additionally, quality inspection resulted in exclusion of the following spectra from the anterior cingulate cortex: participant 8, DIFF-GABA+, OFF-MMSup, DIFF MM-Sup; participant 9, OFF-MMSup, DIFF MM-Sup; participant 11, DIFF-MMsup.

Table 1:

Summary of data acquired in the sensorimotor and anterior cingulate cortices, including number of datasets included, average linewidth, mean and standard deviation of the quantified Glx and Glu concentrations, and fit metrics (CRLB) for each metabolite.

PRESS-35 OFF-GABA+ OFF-MM-Sup DIFF-GABA+ DIFF-MM-Sup
SM n 15 15 15 15 15

FWHM (Hz) 4.2 ± 0.9 4.5 ± 0.9 4.0 ± 0.6 5.5 ± 1.8 3.8 ± 1.0

Glx
Mean ± SD (mol/kg) 13.1 ± 1.5 13.8 ± 1.2 9.6 ± 1.0 6.1 ± 0.5 2.8 ± 0.4
%CRLB Range 5–8 6–9 6–10 2–4 3–7
CoV 0.12 0.10 0.10 0.08 0.15

Glu
Mean ± SD (mol/kg) 10.9 ± 1.3 12.8 ± 1.1 9.1 ± 0.9 5.0 ± 0.5 2.5 ± 0.4
%CRLB Range 5–7 6–9 6–9 4–7 4–15
CoV 0.12 0.08 0.09 0.11 0.14

ACC n 12 12 10 11 9

FWHM (Hz) 6.1 ± 0.9 5.5 ± 1.3 6.0 ± 1.4 6.4 ± 1.3 6.5 ± 1.4

Glx
Mean ± SD (mol/kg) 18.8 ± 2.3 15.6 ± 3.0 12.8 ± 2.2 7.4 ± 0.7 2.9 ± 0.5
%CRLB Range 4–7 6–12 6–11 3–4 4–6
CV 0.12 0.19 0.18 0.09 0.19

Glu
Mean ± SD (mol/kg) 15.5 ± 2.0 13.8 ± 3.0 11.8 ± 1.7 5.3 ± 0.8 2.6 ± 0.5
%CRLB Range 4–6 6–12 6–11 4–10 4–9
CoV 0.13 0.21 0.14 0.16 0.20

Mean ± standard deviation, FWHM = full width half maximum estimated by LCModel, CRLB = Cramér-Rao lower bounds as calculated by LCModel, CoV = coefficient of variation, SM = Sensorimotor Cortex, ACC = Anterior Cingulate Cortex, Glx = Glutamate+Glutamine, Glu = Glutamate, GABA+ =GABA plus macromolecules (TE=68 ms), MMSup=macromolecule suppressed (TE=80 ms), OFF=data quantified from the OFF sub-spectra, DIFF=data quantified from the difference spectra.

Sensorimotor Cortex Data

Glx measures in the sensorimotor cortex from the OFF sub-spectra from either MEGA-PRESS sequence did not significantly correlate with PRESS-35 Glx measures at a significance level of p≤0.05, though a trend towards a positive correlation was seen between OFF-MMSup and PRESS-35 (p=0.058; Figure 2A). Glu measures from both OFF-GABA+ and OFF-MMSup significantly correlated with PRESS-35 Glu measures (Figure 2B). Glx (Figure 2C) and Glu (Figure 2D) measured from both DIFF-GABA+ and DIFF-MMSup also significantly correlated with Glx and Glu measured using PRESS-35.

Figure 2:

Figure 2:

Scatter plots comparing Glx and Glu quantification methods in the sensorimotor cortex, with significant p-values in bold. (A) Glx and (B) Glu measurements from OFF-GABA+ and OFF-MMSup compared to the PRESS-35 measurements. (C) Glx and (D) Glu measured from the DIFF-GABA+ and DIFF-MMSup acquisitions compared to the PRESS-35 measurements. GABA+=GABA plus macromolecules (TE=68 ms), MMSup=macromolecule suppressed (TE=80 ms).

The Bland-Altman analysis indicates no systemic bias when comparing both Glx and Glu measured in the sensorimotor cortex using OFF-GABA+ and Glu measured using OFF-MMSup with PRESS-35 measures, as zero is within the limits of agreement (Figure 3A). Glx measured using OFF-MMSup is on average 3.5 units lower than PRESS-35 values (Figure 3B). All measures from the MEGA-PRESS DIFF spectra were systematically lower than the PRESS-35 measurements (Figure 3C and Figure 3D).

Figure 3:

Figure 3:

Bland-Altman plots comparing Glx and Glu quantification methods in the sensorimotor cortex. (A) Glx quantified from OFF MEGA-PRESS compared with PRESS-35, (B) Glu quantified from OFF MEGA-PRESS compared with PRESS-35, (C) Glx quantified from DIFF MEGA-PRESS compared with PRESS-35, (D) Glu quantified from DIFF MEGA-PRESS compared with PRESS-35. The difference between the two paired measurements (PRESS-35 – MEGA-PRESS) is shown on the y-axis and the mean of the two measures is shown on the x-axis. Grey triangles show data from the GABA+ acquisition and black circles show data from the MMSup acquisition. Solid lines represent the overall mean difference and the dashed lines represent the limits of agreement based on the 95% confidence intervals.

To further investigate the results of the Bland-Altman plots, regression analysis was used to examine the relationship between the difference of the two measures (i.e. the y-axis of the Bland-Altman plot, which shows the values from the OFF-MEGA-PRESS acquisitions subtracted from the values from the PRESS-35 acquisition) and the mean of the two measures (i.e. the x-axis of the Bland-Altman plot). A significant relationship between the measures indicates a proportional bias; as the mean of the two measures increases, the difference between the two measures also increases. There was no proportional bias in either Glx (Figure 3A) or Glu (Figure 3B) values measured from OFF-GABA+ (Glx: β=0.302, p = 0.274; Glu: β=0.354, p=0.195; Figure 3A). There was, however, a significant proportional bias in both Glx and Glu values measured from OFF-MMSup (Glx: β=0.543, p=0.036, Figure 3A Glu: β=0.566, p=0.028, Figure 3B). A significant proportional bias was also seen in both Glx (Figure 3C) and Glu (Figure 3D) measured using the DIFF sub-spectra (DIFF-GABA+ Glx: β=0.871, p<0.001; Glu: β=0.864, p<0.001; DIFF-MMSup Glx: β=0.916, p<0.001; Glu: β=0.917 p<0.001).

Due to the large systematic and proportional bias observed in Glx and Glu values from the DIFF spectra, ICC values were only calculated for measures from the OFF-GABA+ and OFF-MMSup sub-spectra compared to the PRESS-35 data. Glx and Glu measured from OFF-GABA+ and Glu measured from OFF-MMSup all show poor-to-moderate agreement with PRESS-35 measures, as the mean was below an ICC of 0.5 but the 95% confidence interval crosses 0.5. The mean and the confidence intervals of the ICC comparing Glx measured from OFF-MMSup and Glx from PRESS-35 were below 0.5 indicating poor agreement (Figure 4).

Figure 4:

Figure 4:

Intraclass correlation coefficient values and 95% confidence intervals comparing Glx and Glu quantified from OFF MEGA-PRESS data and PRESS-35 data for the sensorimotor cortex. Values of 0.5 (dashed line) indicate moderate agreement.

Anterior Cingulate Cortex Data

Glx and Glu measured in the anterior cingulate cortex using OFF-GABA+ or OFF-MMsup did not significantly correlate with PRESS-35 values (Figure 5A and Figure 5B, respectively). Only Glu measured using DIFF-GABA+ significantly correlated with PRESS-35 values, all other Glx and Glu values obtained from the DIFF spectra did not correlated with PRESS-35 values (Figure 5C and Figure 5D, respectively).

Figure 5:

Figure 5:

Scatter plots comparing Glu and Glx quantification methods in the anterior cingulate cortex. (A) Glx and (B) Glu measurements from OFF-GABA+ and OFF-MMSup compared to the PRESS-35 measurements. (C) Glx and (D) Glu measured from the DIFF-GABA+ and DIFF-MMSup acquisitions compared to the PRESS-35 measurements.

The Bland-Altman analysis indicates no systemic bias when comparing both Glx and Glu measured in the anterior cingulate cortex using OFF-GABA+ with PRESS-35 measures, as zero is within the limits of agreement, however, there is a large range in the limits of agreement. Glx measured using OFF-MMSup was on average 6 units lower than the PRESS-35 values. There was no systematic bias in Glu measured using OFF-MMSup (Figure 6A and Figure 6B). All measures from the DIFF spectra (both DIFF-GABA+ and DIFF-MM-Sup) were systematically lower than the PRESS-35 measurements (Figure 6C and Figure 6D). Regression analysis showed no proportional bias in either Glx or Glu values measured using OFF-GABA+ (Glx: β=−0.300, p=0.344, Figure 6A; Glu; β=−0.241, p=0.451, Figure 6B). In contrast, there was a significant proportional bias in Glx measured using OFF-MMSup (Glx: β=0.060, p=0.003, Figure 6A). There was no proportional bias in Glu measured using OFF-MMSup (Glu: β=0.060, p=0.870, Figure 6B). Additionally, proportional biases were seen for both Glx (Figure 6C) and Glu (Figure 6D) measures from DIFF-GABA+ and DIFF-MMSup (DIFF-GABA+ Glx: β=0.843, p=0.001; Glu: β=0.831, p=0.002; DIFF-MMSup Glx: β=0.894, p=0.001; Glu β=0.853, p=0.003; Figure 6C and D).

Figure 6:

Figure 6:

Bland-Altman plots comparing Glx and Glu quantification methods in the anterior cingulate cortex. (A) Glx quantified from OFF MEGA-PRESS compared with PRESS-35, (B) Glu quantified from OFF MEGA-PRESS compared with PRESS-35, (C) Glx quantified from DIFF MEGA-PRESS compared with PRESS-35, (D) Glu quantified from DIFF MEGA-PRESS compared with PRESS-35. The difference between the two paired measurements (PRESS-35 – MEGA-PRESS) is shown on the y-axis and the mean of the two measures is shown on the x-axis. Grey triangles show data from the GABA+ acquisition and black circles show data from the MM-Sup acquisition. Solid lines represent the overall mean difference and the dashed lines represent the limits of agreement based on the 95% confidence intervals.

Due to the large systematic and proportional bias observed in the values measured using the DIFF spectra, ICC values were only calculated for measures from the OFF sub-spectra. The ICC indicates poor-moderate agreement between Glx and Glu measured using OFF-GABA+ with the PRESS-35 values, and poor agreement between Glx and Glu measured using OFF-MMSup and the PRESS-35 values (Figure 7).

Figure 7:

Figure 7:

Intraclass correlation coefficient values and 95% confidence intervals comparing Glx and Glu quantified from OFF MEGA-PRESS data and PRESS-35 data for the sensorimotor cortex. Values of 0.5 (dashed line) indicate moderate agreement.

DISCUSSION

In the present study, we examined the agreement of Glx and Glu measured from the OFF and DIFF spectra from two GABA-edited MEGA-PRESS sequences (GABA+: TE=68 ms, MMSup: TE=80 ms) across two voxels (sensorimotor cortex and anterior cingulate cortex). Overall, there is limited agreement when measuring Glx and Glu from GABA-edited MEGA-PRESS data compared to the Glx and Glu measures obtained from a PRESS-35 acquisition, which leads us to conclude that quantifying Glx and/or Glu from either the OFF-sub-spectrum or the co-edited resonance from the DIFF spectrum of GABA-edited MEGA-PRESS data cannot replace the measure obtained from a short echo PRESS acquisition.

When considering the results from the correlation analyses, Bland-Altman plots and the ICC analyses, the strongest agreement with the PRESS-35 measures was for Glu measured using OFF-GABA+ in the sensorimotor cortex; however, this was not replicated in the anterior cingulate cortex data. This regional difference is likely in part due to the overall lower data quality in the anterior cingulate cortex, resulting from poorer B0 and B1 homogeneity and challenges shimming in this region, particularly for large voxels. For example, the linewidth of the DIFF-MMSup data from the anterior cingulate cortex is almost double the linewidth of the DIFF-MMSup data from the sensorimotor cortex. Therefore, if data quality is sufficient, Glu measured using OFF-GABA+ could potentially be used as a substitute for a PRESS-35 acquisition, however this method should not be applied when data quality is expected to be lower, such as in the anterior cingulate cortex.

All Glx measures from the MEGA-PRESS data (OFF-GABA+ and OFF-MMSup, DIFF-GABA+ and DIFF-MMSup) showed poor agreement with PRESS-35 data in both brain regions. This discrepancy between Glu and Glx may be due to the phase evolution of Gln at the different echo times. Using simulations, Mullins et al. (2008) showed suppression of Gln resonance intensity at TE=40–80 ms, with the greatest Gln suppression at 80 ms.9 Therefore, the overall Glx signal (Glu+Gln) will vary across echo times due to the different Gln signal contributions. Consistent with these simulations, Schubert et al. (2004) showed that Glx resonances obtained using a PRESS sequence with TE = 80 ms in vivo resembled a Glu spectrum obtained from a phantom.35 Interestingly, our data show Glu measured using OFF-MMSup (TE=80 ms) was slightly lower than Glu measured using PRESS-35 (though this was not classed as a significant bias as zero was still within the limits of agreement). This may be due to certain factors related to working with in vivo data. The OFF-MMSup data has an editing pulse applied at 1.5 ppm, although this shouldn’t affect the evolution of the Glu signal, it likely will affect the evolution of the underlying macromolecules. Additionally, the contamination of the macromolecule baseline is likely to be higher at shorter echo-time acquisition (i.e. PRESS-35) due to macromolecules’ short T2 relaxation times36, which may explain the difference seen here. Indeed, there is evidence that PRESS with TE=35 ms overestimates Glu.21 This reiterates the need in the MRS field as a whole for a better understanding of the contribution of macromolecule baseline to MRS data.

Our findings are consistent with van Veenendaal et al. (2018) who also showed a correlation between glutamate measured from the DIFF spectrum and a PRESS-35 acquisition.20 Here we additionally show systematic and proportional biases are present in Glx and Glu measures from the DIFF spectra, as seen in the Bland-Altman plots. The presence of bias in the data from the DIFF spectra is unsurprising because this is an edited measure. The 1.9 ppm editing pulse is broad and therefore also modulates the 2.1 ppm Glu multiplet that is coupled at 3.75 ppm.11,37 As the editing pulse does not directly target 2.1 ppm, the Glu and Gln resonances are only partially refocussed, resulting in a fraction of the possible signal present in the DIFF spectra.10,11,15 As the concentration of Glu increases, only a proportion of this increase will be reflected in the DIFF spectra, but all of this increase will be directly reflected in the PRESS-35 measures. Therefore, at higher concentrations, Glu (and Glx) measurements will be lower in the DIFF spectra, resulting in a proportional bias in addition to the systematic bias. Furthermore, Gln is edited less effectively than Glu as it is further away from the editing pulse,19 which may explain the stronger correlations for the Glu data compared to the Glx data. The effects of the editing pulse not completely targeting Glu and Gln are augmented in the MMSup results as this editing pulse is more selective.

One limitation of this study is that the ground truth of the Glx/Glu concentration is not known, making it impossible to say for certain which sequence represents the correct value. Alternative sequences (e.g. STEAM, Carr-Purcell PRESS, TE averaged PRESS,21 short echo phase-rotated STEAM22), and different PRESS echo times (e.g. 40 ms, 80 ms9,21) have been suggested to be more optimal for measuring Glu. In an alternative comparison, Maddock et al. (2018) demonstrated strong correlations between Glx and Glu measured using PRESS with TE=80 ms and OFF-GABA+ in the dorsolateral prefrontal cortex.39 The suppression of Gln is more similar at echo times of 68 ms and 80 ms compared to the difference in suppression of Gln at 35 ms and 68 ms, which explains the strong Glx correlation reported by Maddock et al. compared to the weak correlations observed in this current study. We chose PRESS-35 (a standard PRESS acquisition with TE=35 ms) as our reference as it is the most commonly used approach in the literature to measure Glx and Glu. Hancu (2009) showed Glx and Glu measured using PRESS-35 data have lower coefficients of variation and CRLBs than when measured using a PRESS acquisition with a TE=80 ms,21 suggesting that PRESS-35 Glx/Glu measurements are more reproducible than PRESS data with TE=80 ms.

A second limitation of this study is that the order of the sequences was not counterbalanced. There was always a time delay between the GABA+ acquisition and the PRESS-35, and the sensorimotor cortex was always scanned after the anterior cingulate cortex. Lastly, our sample size was small and included only a healthy, male population; quantification of Glx and Glu may be more difficult in clinical populations.

In conclusion, the results of this study generally show poor agreement between Glx and Glu measured using the OFF sub-spectra or the DIFF spectra from GABA+ or MMSup data and Glx and Glu measured using PRESS-35. Therefore, we recommend that a separate PRESS-35 be used when studying Glx, Glu and GABA, rather than using Glx or Glu metrics from the MEGA-PRESS data, particularly in areas of the brain with lower signal quality.

Acknowledgments

SPONSORS: This project was supported by The Natural Sciences and Engineering Research Council of Canada and University of Calgary University Research Grants Committee. RAEE receives salary support from NIH grants R01 EB016089, R01 EB023693 and P41 EB015909.

ABBREVIATIONS

GABA

γ-aminobutyric acid

Glu

Glutamate

Gln

Glutamine

MRS

Magnetic Resonance Spectroscopy

PRESS

Point RESolved Spectroscopy

MEGA-PRESS

Mescher-Garwood PRESS

GABA+

GABA plus macromolecules

MM-Sup

Macromolecule Suppressed

TE

echo time

Footnotes

CONFLICTS OF INTEREST: GJB receives Honoraria for teaching from General Electric Healthcare, and acts as a consultant for IXICO. RML is an employee of General Electric Healthcare.

REFERENCES

  • 1.Ende G Proton Magnetic Resonance Spectroscopy: Relevance of Glutamate and GABA to Neuropsychology. Neuropsychol Rev. 2015;25(3):315–325. doi: 10.1007/s11065-015-9295-8 [DOI] [PubMed] [Google Scholar]
  • 2.Schmidt-Wilcke T, Fuchs E, Funke K, et al. GABA—from Inhibition to Cognition: Emerging Concepts. Neuroscientist. 2018;24(5):501–515. doi: 10.1177/1073858417734530 [DOI] [PubMed] [Google Scholar]
  • 3.Stagg CJ, Bachtiar V, Johansen-Berg H. What are we measuring with GABA Magnetic Resonance Spectroscopy? Commun Integr Biol. 2011;4(5):573–575. doi: 10.4161/cib.16213 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ramadan S, Lin A, Stanwell P. Glutamate and Glutamine: A Review of In Vivo MRS in the Human Brain. NMR Biomed. 2014;26(12):1–36. doi: 10.1002/nbm.3045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Puts NAJ, Wodka EL, Harris AD, et al. Reduced GABA and altered somatosensory function in children with autism spectrum disorder. Autism Res. 2017;10(4):608–619. doi: 10.1002/aur.1691 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.de la Fuente-Sandoval C, León-Ortiz P, Azcárraga M, et al. Glutamate Levels in the Associative Striatum Before and After 4 Weeks of Antipsychotic Treatment in First-Episode Psychosis. JAMA Psychiatry. 2013;70(10):1057. doi: 10.1001/jamapsychiatry.2013.289 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Puts NAJ, Harris AD, Crocetti D, et al. Reduced GABAergic inhibition and abnormal sensory symptoms in children with Tourette syndrome. J Neurophysiol. 2015;114(2):808–817. doi: 10.1152/jn.00060.2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Bertholdo D, Watcharakorn A, Castillo M. Brain Proton Magnetic Resonance Spectroscopy: Introduction and Overview. Neuroimaging Clin N Am. 2013;23(3):359–380. doi: 10.1016/j.nic.2012.10.002 [DOI] [PubMed] [Google Scholar]
  • 9.Mullins PG, Chen H, Xu J, Caprihan A, Gasparovic C. Comparative reliability of proton spectroscopy techniques designed to improve detection of J-coupled metabolites. Magn Reson Med. 2008;60(4):964–969. doi: 10.1002/mrm.21696 [DOI] [PubMed] [Google Scholar]
  • 10.Harris AD, Saleh MG, Edden RAE. Edited 1 H magnetic resonance spectroscopy in vivo: Methods and metabolites. Magn Reson Med. 2017;77(4):1377–1389. doi: 10.1002/mrm.26619 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Mullins PG, McGonigle DJ, O’Gorman RL, et al. Current practice in the use of MEGA-PRESS spectroscopy for the detection of GABA. Neuroimage. 2014;86:43–52. doi: 10.1016/j.neuroimage.2012.12.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Harris AD, Puts NAJ, Barker PB, Edden RAE. Spectral-editing measurements of GABA in the human brain with and without macromolecule suppression. Magn Reson Med. 2015;74:1523–1529. doi: 10.1002/mrm.25549 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Mikkelsen M, Harris AD, Edden RAE, Puts NAJ. Macromolecule-suppressed GABA measurements correlate more strongly with behavior than macromolecule-contaminated GABA+ measurements. Brain Res. 2018;1701(September):204–211. doi: 10.1016/j.brainres.2018.09.021 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Rae CD. A guide to the metabolic pathways and function of metabolites observed in human brain 1H magnetic resonance spectra. Neurochem Res. 2014;39(1):1–36. doi: 10.1007/s11064-013-1199-5 [DOI] [PubMed] [Google Scholar]
  • 15.Geramita M, van der Veen JW, Barnett AS, et al. Reproducibility of prefrontal γ-aminobutyric acid measurements with J-edited spectroscopy. NMR Biomed. 2011;24(9):1089–1098. doi: 10.1002/nbm.1662 [DOI] [PubMed] [Google Scholar]
  • 16.Öngür D, Prescot A, McCarthy J. Elevated gamma-aminobutyric acid levels in chronic schizophrenia. Biol Psychiatry. 2010;68(7):667–670. doi: 10.1016/j.biopsych.2010.05.016.ELEVATED [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Cleve M, Gussew A, Reichenbach JR . In vivo detection of acute pain-induced changes of GABA+ and Glx in the human brain by using functional 1H MEGA-PRESS MR spectroscopy. Neuroimage 2015;105:67–75. doi: 10.1016/j.neuroimage.2014.10.042 [DOI] [PubMed] [Google Scholar]
  • 18.De La Fuente-Sandoval C, Reyes-Madrigal F, Mao X, et al. Cortico-striatal GABAergic and glutamatergic dysregulations in subjects at ultra-high risk for psychosis investigated with proton magnetic resonance spectroscopy. Int J Neuropsychopharmacol. 2015;19(3):1–10. doi: 10.1093/ijnp/pyv105 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Nezhad FS, Anton A, Michou E, Jung JY, Parkes LM, Williams SR. Quantification of GABA, glutamate and glutamine in a single measurement at 3 T using GABA-edited MEGA-PRESS. NMR Biomed. 2017;31(1):1–11. doi: 10.1002/nbm.3847 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.van Veenendaal TM, Backes WH, van Bussel FCG, et al. Glutamate quantification by PRESS or MEGA-PRESS: Validation, repeatability, and concordance. Magn Reson Imaging. 2018;48(September 2017):107–114. doi: 10.1016/j.mri.2017.12.029 [DOI] [PubMed] [Google Scholar]
  • 21.Hancu I Optimized glutamate detection at 3T. J Magn Reson Imaging. 2009;30(5):1155–1162. doi: 10.1002/jmri.21936 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wijtenburg SA, Knight-Scott J. Very short echo time improves the precision of glutamate detection at 3T in 1H magnetic resonance spectroscopy. J Magn Reson Imaging. 2011;34(3):645–652. doi: 10.1002/jmri.22638 [DOI] [PubMed] [Google Scholar]
  • 23.Yousry TA, Schmid UD, Alkadhi H, et al. Localization of the motor hand area to a knob on the precentral gyrus A new landmark. Brain. 1997:141–157. [DOI] [PubMed] [Google Scholar]
  • 24.Mescher M, Merkle H, Kirsch J, Garwood M, Gruetter R. Simultaneous in vivo spectral editing and water suppression. NMR Biomed. 1998;11(6):266–272. [DOI] [PubMed] [Google Scholar]
  • 25.Edden RAE, Puts NAJ, Barker PB. Macromolecule-suppressed GABA-edited magnetic resonance spectroscopy at 3T. Magn Reson Med. 2012;68(3):657–661. doi: 10.1002/mrm.24391 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Simpson R, Devenyi GA, Jezzard P, Hennessy TJ, Near J. Advanced processing and simulation of MRS data using the FID appliance (FID-A)—An open source, MATLAB-based toolkit. Magn Reson Med. 2017;77(1):23–33. doi: 10.1002/mrm.26091 [DOI] [PubMed] [Google Scholar]
  • 27.Provencher SW. Automatic quantitation of localized in vivo 1H spectra with LCModel. NMR Biomed. 2001;14:260–264. [DOI] [PubMed] [Google Scholar]
  • 28.Harris AD, Puts NAJ, Edden RAE. Tissue correction for GABA-edited MRS: Considerations of voxel composition, tissue segmentation, and tissue relaxations. J Magn Reson Imaging. 2015;42(5):1431–1440. doi: 10.1002/jmri.24903 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Gasparovic C, Song T, Devier D, et al. Use of tissue water as a concentration reference for proton spectroscopic imaging. Magn Reson Med. 2006;55(6):1219–1226. doi: 10.1002/mrm.20901 [DOI] [PubMed] [Google Scholar]
  • 30.Kreis R Issues of spectral quality in clinical 1H-magnetic resonance spectroscopy and a gallery of artifacts. NMR Biomed. 2004;17(6):361–381. doi: 10.1002/nbm.891 [DOI] [PubMed] [Google Scholar]
  • 31.Watson PF, Petrie A. Method agreement analysis: A review of correct methodology. Theriogenology. 2010;73(9):1167–1179. doi: 10.1016/j.theriogenology.2010.01.003 [DOI] [PubMed] [Google Scholar]
  • 32.Bland MJ, Altman DG. Statistical Methods for Assessing Agreement Between Two Methods of Clinical Measurement. Lancet. 1986;327(8476):307–310. doi: 10.1016/S0140-6736(86)90837-8 [DOI] [PubMed] [Google Scholar]
  • 33.Giavarina D Understanding Bland Altman analysis. Biochem Med. 2015;25(2):141–151. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016;15(2):155–163. doi: 10.1016/j.jcm.2016.02.012 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Schubert F, Gallinat J, Seifert F, Rinneberg H. Glutamate concentrations in human brain using single voxel proton magnetic resonance spectroscopy at 3 Tesla. Neuroimage. 2004;21(4):1762–1771. doi: 10.1016/j.neuroimage.2003.11.014 [DOI] [PubMed] [Google Scholar]
  • 36.Cudalbu C, Mlynárik V, Gruetter R. Handling macromolecule signals in the quantification of the neurochemical profile. J Alzheimers Dis. 2012;31 Suppl 3: S101–15. doi: 10.3233/JAD-2012-120100 [DOI] [PubMed] [Google Scholar]
  • 37.Govindaraju V, Young K, Maudsley AA. Proton NMR chemical shifts and coupling constants for brain metabolites. NMR Biomed. 2000;13:129–153. doi: [pii] [DOI] [PubMed] [Google Scholar]
  • 38.Mikkelsen M, Singh KD, Sumner P, Evans CJ. Comparison of the repeatability of GABA-edited magnetic resonance spectroscopy with and without macromolecule suppression. Magn Reson Med. 2016;75(3):946–953. doi: 10.1002/mrm.25699 [DOI] [PubMed] [Google Scholar]
  • 39.Maddock RJ, Caton MD, Ragland JD. Estimating glutamate and Glx from GABA-optimized MEGA-PRESS: Off-resonance but not difference spectra values correspond to PRESS values. Psychiatry Res - Neuroimaging. 2018;279(June):22–30. doi: 10.1016/j.pscychresns.2018.07.003 [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES