Development and evaluation of machine learning models for voxel dose predictions in online adaptive magnetic resonance guided radiation therapy

M Allan Thomas 1,2,, Yabo Fu 1,3, Deshan Yang 1,
PMCID: PMC7386189  PMID: 32306535

Abstract

Purpose

Daily online adaptive plan quality in magnetic resonance imaging guided radiation therapy (MRgRT) is difficult to assess in relation to the fully optimized, high quality plans traditionally established offline. Machine learning prediction models developed in this work are capable of predicting 3D dose distributions, enabling the evaluation of online adaptive plan quality to better inform adaptive decision‐making in MRgRT.

Methods

Artificial neural networks predicted 3D dose distributions from input variables related to patient anatomy, geometry, and target/organ‐at‐risk relationships in over 300 treatment plans from 53 patients receiving adaptive, linac‐based MRgRT for abdominal cancers. The models do not include any beam related variables such as beam angles or fluence and were optimized to balance errors related to raw dose and specific plan quality metrics used to guide daily online adaptive decisions.

Results

Averaged over all plans, the dose prediction error and the absolute error were 0.1 ± 3.4 Gy (0.1 ± 6.2%) and 3.5 ± 2.4 Gy (6.4 ± 4.3%) respectively. Plan metric prediction errors were −0.1 ± 1.5%, −0.5 ± 2.1%, −0.9 ± 2.2 Gy, and 0.1 ± 2.7 Gy for V95, V100, D95, and Dmean respectively. Plan metric prediction absolute errors were 1.1 ± 1.1%, 1.5 ± 1.5%, 1.9 ± 1.4 Gy, and 2.2 ± 1.6 Gy. Approximately 10% (25) of the plans studied were clearly identified by the prediction models as inferior quality plans needing further optimization and refinement.

Conclusion

Machine learning prediction models for treatment plan 3D dose distributions in online adaptive MRgRT were developed and tested. Clinical integration of the models requires minimal effort, producing 3D dose predictions for a new patient’s plan using only target and OAR structures as inputs. These models can enable improved workflows for MRgRT through more informed plan optimization and plan quality assessment in real time.

Keywords: adaptive radiation therapy, dose prediction, knowledge‐based prediction, magnetic resonance image guidance

1. INTRODUCTION

The efficacy of radiation therapy (RT) for abdominal cancers is limited by a prescription dose (Rx) that can be tolerated without toxicity in nearby organs‐at‐risk (OARs). The abdominal OARs (liver, kidney, and all GI organs, etc.) are highly deformable in nature and their daily motion throughout treatment is substantial. Interfraction motion over 3 cm is possible. 1 , 2 , 3 GI OARs (stomach, duodenum, esophagus, small and large intestines) are also sensitive to high doses of radiation. 4 , 5 As a result, tumors associated with such difficult abdominal sites have not been treated to the same curative doses that have been achieved for other sites, ie a biologically effective dose (BED) >100 Gy. 6 , 7 , 8 Daily, online adaptive treatment with magnetic resonance image guided RT (MRgRT) can provide an opportunity to simultaneously escalate gross tumor volume (GTV) dose while meeting OAR constraints in these difficult cases. 9 Both modeling studies 3 , 10 , 11 and completed phase I clinical trials 12 , 13 , 14 have shown the feasability of ablative dose escalation without significant adjacent organ toxicity in the central thorax and abdomen. Such trials have shown an increase in both local control and overall survival while decreasing radiation‐induced toxicity events, leading to a prospective clinical trial for further evaluation in pancreatic cancer (SMART: NCT03621644).

As the use of online adaptive MRgRT has grown recently across a variety of clinics, both the potential benefits and practical difficulties of its utilization are beginning to be understood. 12 , 13 , 14 , 15 , 16 , 17 , 18 , 19 , 20 , 21 But ample opportunity remains for assessment and improvement of online adaptive RT as this unique treatment method continues to develop clinically. Analysis of a collection of patient data and their treatment plans in order to better understand and improve a particular RT process can proceed in a variety of ways. But the most common method used recently falls under the general term of knowledge‐based (KB) prediction models. 22 KB methods have been used to predict achievable dose‐volume histograms (DVHs), 23 , 24 plan metrics, 25 and full 3D dose distributions. 26 , 27 , 28 , 29 , 30 , 31 In fact, using information from previously treated patients and plans to better inform future procedures is now present in nearly all aspects of RT. 32 , 33 , 34 , 35 Much of the recent work related to KB methods for improving RT has focused on 3D dose prediction: voxel‐by‐voxel prediction of dose for a “future” patient based on a collection of previously treated patients’ plans and/or other relevant data. Such a focus is understandable because the ability to accurately and robustly predict 3D dose volumes for future patients is significant. It produces in a single model the ability to predict a wide variety of clinically relevant data such as target and OAR dose‐volume metrics as well as conformity/homogeneity indices which may also be important. Furthermore, 3D dose predictions could serve as novel inputs for treatment plan optimization that may completely alter the paradigm under which RT planning currently occurs. 26 , 28 , 30

Despite the growth in applications of KB methods to solving interesting RT problems, a majority of the most relevant work has focused on treatment sites such as prostate, 23 , 24 , 26 , 28 , 31 head and neck, 25 , 29 , 30 breast, 28 and lung. 36 There are limited studies dedicated to abdominal cancers 15 , 27 because they are notoriously difficult to treat and there is minimal consensus regarding the best RT treatment approaches. 37 These sites are however expected to benefit most directly from MRgRT and online plan adaptation. Application of KB methods to online adaptive RT is also new, with very few previous studies. 15 KB methods typically require sufficient data collection from previously treated patients and plans, and online adaptive RT is a mostly new technique.

In this work, we have developed artificial neural network (ANN) models to predict 3D dose distributions within the GTV for MRgRT of abdominal cancers. The clinical goals are to enable real‐time guidance for achievable plan quality of each online adaptive case, identify inferior plans, and assure plan quality. Unlike some previous KB works, 26 , 27 our ANN models do not require beam related variables such as beam angles and fluence. They therefore can predict a 3D dose distribution for a new patient’s plan using only target and OAR structures as inputs. In addition, the ANN models are simple in design and rely on small datasets. They do not require significant training time and are not computationally demanding (no GPU required). As a result, they can be integrated clinically with minimal effort. To our knowledge, this work presents the development and results of the first 3D dose prediction models for online adaptive MRgRT.

2. MATERIALS AND METHODS

2.A. Patient Characteristics

As shown in Table 1, datasets from a total of 53 patients with abdominal cancers treated at Washington University in St. Louis with online adaptive, linac‐based MRgRT were utilized. The patient data were combined from two main treatment site groups, pancreas (n = 34) and nonpancreas (n = 19). All patients were treated with one of two high BED protocols: stereotactic body RT (SBRT, n = 48), having prescriptions of 50 Gy in 5 fractions (BED ~ 100 Gy) and critical OAR (OARCRIT) dose constraints of less than 0.5 cubic centimeters (cc) receiving 36 Gy or more (V36 < 0.5 cc); or hypofractionation (HYP, n = 5) having prescriptions of 67.5 Gy in 15 fractions (BED ~ 100 Gy) and OARCRIT dose constraints of V50 < 0.5 cc. Overall, 260 out of 315 (82.5%) treatment fractions were adapted.

Table 1.

Statistics for the plans and GTV V95% and D95% values.

Patient Group Type of Plan # of Plans V95% D95%
SBRT HYP Total Mean ± σ Mean ± σ
Pancreas Simulation a 29 4 33 86.2 ± 8.7 80.7 ± 12.6
Adapted 133 56 189 86.2 ± 8.8 78.5 ± 15.4
NonPancreas Simulation 16 1 17 94.6 ± 11.0 98.9 ± 13.0
Adapted 56 15 71 92.5 ± 10.3 94.1 ± 14.5
All Patients Simulation 45 5 50 89.3 ± 10.4 87.5 ± 15.5
Adapted 189 71 260 87.9 ± 9.6 82.7 ± 16.6
a

Simulation plans are created on the patient's simulation image, receive normal planning time, attention, optimization, and checks just like traditional IMRT plans, and serve as the starting point of plan adaptation for the first fraction. V95% is the percentage of the GTV volume receiving 95% or more of Rx. D95% is the minimum dose received by 95% of the GTV, given as a percentage of Rx.

2.B. Treatment Plan Adaptation Techniques

Detailed descriptions of the specific workflows and planning strategies for adaptive MRgRT using the ViewRay linac system (Viewray, Cleveland, OH) have been presented previously. 12 , 13 , 21 , 38 , 39 , 40 , 41 A brief summary, with particular focus on the details related to the online adaptive process, follows.

Table 2 includes key definitions for structures associated with the planning process. Additional OARs beyond those in OARCRIT may have been included in the plan optimization such as the spinal cord, liver, aorta, and one or both kidneys. PTVOPT is used as the target in the plan optimization instead of the PTV. It is often observed in these plans that, after subtracting OAR5 mm, not all of the GTV will be inside PTVOPT. The planning strategy is quite different from traditional IMRT. PTV V95>=95% was still one planning goal, but the hard OARCRIT dose constraints caused this target coverage goal to be rarely achievable. Hotspots were generally limited to <150% of the Rx inside the GTV.

Table 2.

Structure names and definitions used throughout this work.

Planning Structure Definition
GTV Gross tumor volume
PTV GTV + 5 mm isotropic expansion
OARCRIT All abdomen organs in close proximity of GTV (esophagus, stomach, duodenum, large bowel, small bowel)
OAR5mm OARCRIT + 5 mm isotropic expansion
PTVOPT

PTV minus OAR5 mm:

used as target in plan optimization

The adaptive process workflow is shown visually in Fig. 1. To start the adaptive workflow, on each treatment day the most recently used treatment plan is loaded and a new volumetric MRI with exhale‐breath‐hold is scanned after patient setup. The couch table is shifted according to manual registration of the visible GTV in the new image with the GTV in the loaded plan. Following this registration, all other contours in the loaded plan are then rigidly copied onto the new images. All OARs within 30 mm of the PTV are re‐contoured on the new MR image by the attending physician. The overwhelming majority (>90%) of plans in this study had no GTV contour change. If the GTV is recontoured, most commonly at a patient’s first adapted treatment fraction, the PTV is regenerated by expanding the GTV by an isotropic 5 mm margin. The attending medical physicist reviews the manual OAR segmentation and assesses for other organ changes such as known areas of heterogeneity and skin surface position. Note that other institutions may choose to use the original simulation plan as the starting point for all online adaptive plans. 15 The use of the most recently treated plan for this purpose is simply the chosen workflow at our institution.

Fig. 1.

Fig. 1

Workflow showing the steps and estimated timeline to the online adaptive MRgRT process.

The new radiation dose distribution is then calculated by applying the loaded plan to the current day’s anatomy. Target coverage and OAR doses are reassessed relative to plan goals and constraints. Only if target coverage can clearly be improved or any critical OAR dose constraint is not met will a new plan be created. The new plan is developed by reoptimization using the same beam angles and objective functions established in the currently loaded plan. In rare cases (~4%), attempts are made to adjust optimization objectives to further reoptimize the plan. After the plan is reoptimized, the attending physician evaluates and approves the plan. Physics quality assurance steps are taken next, 42 and then the plan is delivered at free breathing, with real‐time, continuous 2D cine MR guidance for gating at the planned end‐of‐exhalation phase. 41

All plans discussed in this work were developed with OAR isotoxicity prioritized over target coverage. Essentially, the OARCRIT constraints discussed above were hard constraints that must not be exceeded, regardless of target coverage. Therefore the goal of the model predictions in this work is identifying improvements to GTV coverage while maintaining the same OAR avoidance. By contrast, most previous KB prediction models aimed at identifying planning improvements related to OAR dose sparing while maintaining uniform target coverage, eg PTV V95% > 95%. 23 , 24 , 26 , 27 , 31

2.C. GTV 3D Dose Distribution Prediction ANN Model

In previously published KB methods, parameters associated with target size/shape, distance relationships between the target and OARs, and patient size were deemed important for predicting OAR 3D dose, with the most critical parameter the minimum distance from a voxel inside an OAR to the target surface. 23 , 26 , 27 We have a similar approach in this study to predict 3D dose distributions inside the GTV. Our model input parameters include the minimum distances from a voxel inside the GTV to surfaces of various OAR structures (OARCRIT, OAR5mm, etc.). To improve the prediction model accuracy and robustness, additional parameters derived from the distance and geometrical relationships between OARs and the GTV were added and are explained in Table 3. GTV V95 and D95 were found to strongly correlate with these additional parameters. It is important to note that we did not include any beam parameters (beam angles or fluence) so that the developed models could predict 3D dose before a treatment beam plan is devised. This is different from some previous work using ANN models to make 3D dose predictions. 26 , 27

Table 3.

List of 16 ANN model input variables.

# ANN Input Variables # ANN Input Variables
1 % of all voxels within 5 mm occupied by OAR5 mm 9 Square root of min distance to GTV surface
2 % of all voxels within 10 mm occupied by OAR5 mm 10

Min distance to OAR5 mm

(>=0 only) b

3 % of all voxels within 15 mm occupied by OAR5 mm 11

Min distance to OAR5mm

(<=0 only)

4 % of all voxels within 20 mm occupied by OAR5 mm 12 Square root of min distance to OAR5mm
5 % of GTV occupied by OAR5 mm a 13

Min distance to PTVOPT

(>=0 only)

6 Distance to GTV centroid 14

Min distance to PTVOPT

(<=0 only)

7 Max distance from GTV centroid to GTV surface a 15

Min distance to (OAR5 mm‐OARCRIT)

(>=0 only)

8 Min distance to GTV surface 16

Min distance to (OAR5 mm‐OARCRIT)

(<=0 only)

Each variable was calculated for all voxels inside the GTV for each plan unless indicated otherwise. Some ANN model input variables could have both positive and negative values, with negative values corresponding to an overlap between the GTV and the OAR structure associated with the variable.

a

Every voxel for the GTV had the same value.

b

A linear correlation was observed between this distance metric and dose but only up to a distance of ~ 16 mm. So this metric has a maximum value of 16 mm for any voxel.

In a separate study, we determined that for linac‐based MRgRT plans, plan quality comparisons between online adapted plans and fully optimized offline simulation plans showed no statistically significant differences. Therefore, we developed our models using both offline (simulation) and adapted plans. The model prediction output was per voxel dose value normalized to Rx. By tracking the model prediction accuracy in preliminary experiments, we identified the most relevant input variables, eliminated the less significant ones, and simplified the ANN models while avoiding overfitting. Our final ANN model was a two‐layer feed forward neural network design (one hidden layer: 2 nodes, one output layer: 1 node), shown in Fig. 2. It was implemented using MATLAB’s neural network toolbox. The default Levenberg‐Marquardt backpropagation was used for model training. Other models with more nodes in the hidden layer (eg up to ~ 5) provided very similar results if they were similarly optimized. Continuing to increase the number of nodes further led to model overfitting and clearly reduced model performance. Adding more hidden layers was similar. As a result, we chose to keep the simplest model possible that provided equal performance which was a model with two nodes in a single hidden layer. A wide range of different loss functions were also tested, including some that weighted different types of voxels differently. 26 But the loss function that provided the most consistent prediction accuracy among the many that were tested was mean square error.

Fig. 2.

Fig. 2

The ANN models used in this work, with three distinct layers: (1) input layer with 17 (16 inputs + 1 bias) nodes, (2) hidden layer with two activation nodes and one bias node, and (3) output layer with one activation node. The arrows represent the connections between nodes in each layer and are mathematically associated with the weighting parameters that get updated and optimized during model training.

A flowchart for ANN model training, testing, and development is shown in Fig. 3. The simple ANN model design we eventually adopted allowed for very fast model training. So it was straightforward to utilize cross validation to train and validate the ANN models. We were not limited to training potential models and then testing on a small group of separate test patients/plans only once. The total 310 plans (50 simulation, 260 adapted) were separated into training and testing groups as follows: the 53 patients were split into ten groups of five and one group of three, then each group was cycled through as the test group. For each iteration of the cross validation, all of the plans from the test group patients made up the testing data while all the plans from the other ten groups served as training data. The cross validation was done on a patient‐by‐patient basis instead of a plan‐by‐plan basis to avoid testing the models on any plan coming from a patient whose plans were already used for model training.

Fig. 3.

Fig. 3

Process flowchart for ANN model training and testing

The ANN model datasets in each cross validation iteration were quite large, with the typical training group composed of ~250 plans from 48 patients. The total number of voxels available in such a training group dataset was >2 million. It was not necessary to use all available voxels in any given cross validation to produce optimized and consistent model performance. So GTV voxels for each plan from the training group were randomly sampled at a 1/20 ratio and then accumulated to make the training dataset. This data sampling was chosen in an effort to fairly sample each set of GTV voxels regardless of GTV size differences in the group while achieving reduced model training time and consistent prediction results. Each cross validation iteration utilized ~120,000 voxels after data sampling.

Once the training dataset from randomly sampled GTV voxels was created, it was further separated into “training” and “validation” subsets at 85% and 15% respectively. The model was trained with the “training” dataset until six consecutive error increases were observed on the “validation” dataset or a maximum of 200 epochs was reached. Many different epoch cut‐offs were also tested, but the results did not strongly depend on this variable. The final training method was chosen based on allowing the model to either naturally find its optimal number of epochs based on error increases in the “validation” dataset, or run through a sufficient number of epochs to reach model convergence.

In an effort to ensure consistent model prediction performance, the model training sequence was repeated 50 times and the predicted dose results on each plan in the test group were averaged. The mean, mean absolute, and standard deviation errors of each entire 3D dose distribution in the test group were calculated, as well as mean dose (Dmean), V95, V100, and D95. DVHs for each real and predicted dose distribution and their differences were determined (% GTV volume error as a function of dose). This process was repeated for each iteration of the cross validation until all patients were part of the test group, and then the overall results from the cross validation were accumulated.

2.D. Outlier Detection, Rejection, and Model Refinement

One main application of the prediction models is to identify inferior plans by comparing the predicted plan quality metrics to those computed in the actual plans. This concept also allows model refinement by excluding inferior plans from training. 25 , 26 To do so, inferior plans were identified as any with an observed plan metric (V95, V100, or D95) outside the 95% prediction interval (PI) of the original ANN model, with:

95%PI=meanerror+1.96standarddeviationoferror

The inferior plans outside the 95% PI were then discarded and the model was retrained. After two rounds, a total of 25 inferior plans were identified and discarded. The model errors and results are included only for the refined model with 285 plans.

3. RESULTS

The prediction dose error statistics from the cross validation analysis are shown in Table 4. Dose prediction error was computed as ΔDose = DactualDpredicted, then averaged across all plans of all patients. Many different combinations of input variables could produce similar prediction dose errors yet distinct plan metric prediction errors. So it was important to track plan metric prediction errors during model optimization as well. In the end the most optimized ANN model was chosen based on the best balance between dose and plan metric error. As shown in Table 5, overall plan metric prediction errors for V95, V100, D95, and Dmean were generally low with mean absolute errors of 1.1%, 1.5%, 1.9 Gy, and 2.2 Gy respectively. Removing inferior plans to make the refined model helped to enhance its predictive ability for dose values relevant to plan quality, despite only marginal improvements in overall dose prediction errors.

Table 4.

ANN model 3D dose prediction results.

Dose Prediction Error

Mean ± σ (% Rx)

Absolute Dose

Prediction Error

Mean ± σ (% Rx)

0.1 ± 3.4 Gy (0.1 ± 6.2%) 3.5 ± 2.4 Gy (6.4 ± 4.3%)

The listed errors represent the averages on a plan‐by‐plan basis.

Table 5.

ANN model plan metric prediction results.

Metric Prediction Error

Absolute Prediction

Error

95% R2
V95 (%) −0.1 ± 1.5 1.1 ± 1.1 3.3 0.97
V100 (%) −0.5 ± 2.1 1.5 ± 1.5 5.0 0.95
D95 (Gy) −0.9 ± 2.2 1.9 ± 1.4 4.6 0.95

Dmean

(Gy)

0.1 ± 2.7 2.2 ± 1.6 5.1 0.91

Averages across all plans for both raw and absolute prediction errors are given. 95%: the value required to encompass the absolute error from 95% of all plans in the model. R2: coefficient of determination values for the linear fit of actual vs predicted plan metric values.

The dose prediction errors observed here are similar to previous efforts at 3D dose predictions using ANN models, 26 , 27 where errors ranged from 2‐10% of Rx generally but had some contributions up to 20% and greater. Care must be taken to recall that the fair comparison to previous results with this work would be in relating published OAR dose errors to GTV dose errors here. Previously published models focusing on OAR dose distributions also made predictions for GTV/PTV dose, 26 , 27 but due to the key differences in treatment planning strategy and design discussed in Methods, a direct comparison to previous GTV/PTV dose prediction results is not fair. Furthermore, the doses being predicted in this work are high, with the majority over Rx. As a result, the raw and % of Rx errors will appear to be higher than models focusing on OAR dose predictions.

Three axial slices of the planned dose, predicted dose, and the prediction error for a representative pancreas SBRT plan are shown in Fig. 4. It is evident that the predicted dose distribution is smoother than the planned dose because the prediction models did not consider beam angles. As a result, the dose distribution heterogeneity and hot spots are not easily predicted. Nevertheless, the prescription isodose lines are very similar between the planned and predicted doses, suggesting that the model can accurately predict dose values near Rx as well as clinical DVH metrics. Similar effects were observed generally throughout the model predictions. Note that the example in Fig. 4 was chosen as representative of typical predictions but not necessarily matched with all average metrics tracked.

Fig. 4.

Fig. 4

Three axial slices (inferior, center, and superior of GTV) of planned dose, model predicted dose, and ΔDose of a representative patient plan (SBRT: Rx = 50 Gy, OARCRIT constraint = 36 Gy). All OARs (stomach, duodenum, small bowel, large bowel, and spinal cord) are shown in white. The prescription isodose lines are shown in blue for the planned and predicted dose views. Dose prediction errors = 0.7 ± 3.9 Gy, absolute errors = 3.3 + 2.3 Gy. The plan metrics (planned/predicted) are V95 = 90.4%/88.5%, V100 = 86.8%/85.1%, D95 = 42.6 Gy/41.4 Gy, Dmean = 57.2 Gy/56.6 Gy.

The results of some predicted plan metrics vs their true plan values are shown in Fig. 5. Overall, V95, V100, and D95 are well‐predicted, with R2 values of 0.97, 0.95, and 0.95, respectively, and linear fit slopes all near unity. The 25 inferior plans were well‐identified by their divergence from the 95% PI for the V95 and V100 plan metrics. Figure 6(a) shows results for comparisons between the real and predicted DVH metrics: error in Vx as a function of x where x is the percentage of Rx (eg V95 = % of GTV volume receiving 95% of Rx). It is clear that Vx is predicted well for x < 100% and large errors only occur for x > 100%. Figure 6(b) presents the overall dose prediction error distributions in % Rx also as a function of x, confirming that per voxel dose values are predicted well from x = 60%–100%.

Fig. 5.

Fig. 5

Results for (a) V95, (b) V100, and (c) D95 predictions calculated from the model‘s 3D dose predictions. The 45° dashed lines in each plot represent where the predicted and actual values are equal. The linear fit line and coefficient of determination (R2) are included. The outside boundary lines represent the 95% PI. The 25 inferior plans identified during model refinement are shown as *.

Fig. 6.

Fig. 6

Distributions of prediction error for (a) Vx: the percentage of the GTV volume receiving dose of at least x percent of Rx (i.e. V20 through V160), and (b) dose in % Rx. The error distributions are shown as boxplots with a horizontal red line for the median, blue box encompassing the 25% and 75% interquartile ranges, and dotted line extending to the 95% range. The OARCRIT constraint (dotted) and Rx (dashed) levels are shown in the figures.

4. DISCUSSION

ANN models were designed in this study to predict 3D dose distributions based on the average of prior plans used for model training. Our models allow a direct, 3D dose comparison between the history of previously treated plans and upcoming plans for future patients without needing to take the time and effort to create an actual treatment plan. This is possible because our models are based on inputs that require only target and OAR structure data, not planned beam parameters. Because beam angles and related information were not included as input variables, the models generally do not predict beam geometry‐dependent dose distribution heterogeneity and hot spots for each individual patient’s plan. While overall raw dose errors were not ideally low, the larger errors were generally associated with localized dose heterogeneity and high dose regions above Rx (Fig. 6). From a clinical standpoint, such errors are less relevant and do not affect predictions for important DVH metrics such as V95, or regions of the GTV receiving dose below Rx due to OAR sparing and constraint requirements. Such overall plan metrics and the predicted balance between OAR sparing and GTV coverage are the most important to guide adaptive decision‐making.

Due to the time constraints associated with the online adaptive process, adapted plans are not afforded the same level of optimization and scrutiny as high‐quality offline plans. The 3D dose prediction models establish benchmarking that can enable a rapid evaluation of online adaptive plans in terms of both plan quality and the overall dose distribution. Examples that showcase this idea are evident in Fig. 5. The 25 plans deemed inferior during model refinement are distinct from the general model predictions, with predicted values outside the 95% PI of at least one plan quality metric. Future treatment plans that are inferior to the history of previously treated plans could likely be identified similarly by testing with the model.

Several of the 25 plans identified as inferior were retrospectively replanned to assess for potential improvement. Conditions that would be present during real‐time adaptation were simulated for the replanning: no OAR constraints could be exceeded, optimization parameters could be manually tweaked and adjusted, and time constraints for the reoptimization process were kept in mind (~5 min maximum time allowed). Table 6 shows the results of the replanning analysis. While the degree of improvement varied across the plans tested, all plans that were reoptimized are closer to the quality claimed as possible by the ANN prediction model. Generally, from 40–100% of the difference between the predicted plan metrics and the original values were recovered after replanning. These results help to further showcase the clinical relevance and utility of the dose prediction models.

Table 6.

Results of the replanning by manual reoptimization for several plans from the group of 25 identified as inferior by the ANN prediction models.

Metric Type Plan 1 Plan 2 Plan 3 Plan 4
V95 (%) Clinical 52.6 82.3 72.6 80.4
RePlan 63.3 87.9 76.3 83.1
ANN 61.9 91.5 81.0 86.0
V100 (%) Clinical 45.6 70.9 65.0 70.7
RePlan 60.0 84.5 70.4 75.6
ANN 60.7 89.0 79.3 82.4
D95 (Gy) Clinical 23.0 36.6 29.3 35.6
RePlan 22.7 38.9 35.2 37.3
ANN 23.5 42.3 38.0 39.1

Despite the potential applications of the models developed in this work, there are limitations that must be addressed. First, the models were developed using only patient and plan data from a single institution. Additionally, the inability of the models to account for hotspots or other dramatic dose distribution changes prevents clinical applications that relate directly to hotspots. 43 , 44 Dose prediction models would need to include more detailed plan information related to beam angles and/or fluence 26 , 27 in order to more accurately predict such metrics.

In future applications, the predicted 3D dose distributions could serve as an alternative input to the treatment plan development process. 26 , 30 This is a particularly desirable strategy for online adaptive plans, much in the same vein as it is being pursued for automated treatment planning. An estimated 3D dose prediction tailored to the specific anatomy of the day could provide a much improved starting point for subsequent adapted plan development and optimization each fraction. The fact that our ANN models can provide a 3D dose prediction using contour information alone (a fully developed treatment plan is not needed) helps to bolster their potential use as novel inputs for alternative treatment planning strategies.

Additional future work that could extend this study relates mostly to improving the 3D dose prediction models. ANN model development and performance are based mainly on the ability to identify key input data that strongly relates to making accurate output predictions. Most likely, the best 3D dose prediction models would be based on convolutional neural networks (CNNs), 29 , 30 , 31 which have shown improved dose prediction performance relative to what was achieved in this study. However, CNN models are well known to require vast amounts of training data to learn the necessary relationships to make accurate and robust predictions. 29 , 30 , 31 As the number of patients treated with online adaptive MRgRT continues to increase, future work may be possible related to CNN based prediction models with improved capabilities.

5. CONCLUSION

We developed ANN models to predict 3D dose distributions and overall plan quality metrics for abdomen cancers in adaptive MRgRT. The prediction models will be useful to improve adaptive planning strategies and workflows through more informed plan optimization and evaluation in real time.

CONFLICT OF INTEREST

The authors have no conflicts to disclose.

ACKNOWLEDGMENT

Research reported in this study was partially supported by the Agency for Healthcare Research and Quality (AHRQ) under award R01‐HS022888. The content is solely the responsibility of the authors and does not necessarily represent the official views of the Agency of Healthcare Research and Quality.

Contributor Information

M. Allan Thomas, Email: mathomas3@mdanderson.org.

Deshan Yang, Email: yangdeshan@wustl.edu.

REFERENCES

  • 1. Chen I, Mittauer KE, Henke LE, et al. Quantification of interfractional gastrointestinal tract motion for pancreatic cancer radiation therapy. Int J Radiat Oncol Biol Phys. 2016;96:E144. [Google Scholar]
  • 2. Abbas H, Chang B, Chen Z. Motion management in gastrointestinal cancers. Int J Radiat Oncol Biol Phys. 2014;5:223–235. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Liu F, Erickson B, Peng C, Li XA. Characterization and management of interfractional anatomic changes for pancreatic cancer radiotherapy. J. Med Imaging Radiat Sci. 2012;83:e423–429. [DOI] [PubMed] [Google Scholar]
  • 4. Bentzen SM, Constine LS, Deasy JO, et al. Quantitative analyses of normal tissue effects in the clinic (QUANTEC): an introduction to the scientific issues. Int J Radiat Oncol Biol Phys. 2010;76:S3–S9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Thomas TO, Hasan S, Small W Jr, et al. The tolerance of gastrointestinal organs to stereotactic body radiation therapy: what do we know so far? J Gastrintest Oncol. 2014;5:236–246. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Comito T, Clerici E, Tozzi A, D’Agostino G. Liver metastases and SBRT: a new paradigm? Rep Pract Oncol Radiother. 2015;20:464–471. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Palma D, Lagerwaard F, Rodrigues G, Haasbeek C, Senan S. Curative Treatment of Stage I Non‐Small‐Cell Lung Cancer in Patients with Severe COPD: Stereotactic Radiotherapy Outcomes and Systematic Review. Int J Radiat Oncol Biol Phys. 2012;82:1149–1156. [DOI] [PubMed] [Google Scholar]
  • 8. Crane CH. Hypofractionated ablative radiotherapy for locally advanced pancreatic cancer. J Radiat Res. 2016;57:i53–i57. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Acharya S, Fischer‐Valuck BW, Kashani R, et al. Online magnetic resonance image guided adaptive radiation therapy: first clinical applications. Int J Radiat Oncol Biol Phys. 2016;94:394–403. [DOI] [PubMed] [Google Scholar]
  • 10. Li XA, Liu F, Tai A, et al. Development of an online adaptive solution to account for inter‐ and intra‐fractional variations. Radiother Oncol. 2011;100:370–374. [DOI] [PubMed] [Google Scholar]
  • 11. Henke L, Kashani R, Yang D, et al. Simulated online adaptive magnetic resonance‐guided stereotactic body radiation therapy for the treatment of oligometastatic disease of the abdomen and central thorax: characterization of potential advantages. Int J Radiat Oncol Biol Phys. 2016;96:1078–1086. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Henke L, Kashani R, Robinson C, et al. Phase I trial of stereotactic MR‐guided online adaptive radiation therapy (SMART) for the treatment of oligometastatic or unresectable primary malignancies of the abdomen. Radiother Oncol. 2018;126:519–526. [DOI] [PubMed] [Google Scholar]
  • 13. Henke LE, Olsen JR, Contreras JA, et al. Stereotactic MR‐guided online adaptive radiation therapy (SMART) for ultracentral thorax malignancies: results of a phase 1 trial. Adv Radiat Oncol. 2018;4:201–209. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Rudra S, Jiang N, Rosenberg SA, et al. Using adaptive magnetic resonance image‐guided radiation therapy for treatment of inoperable pancreatic cancer Cancer Med. 2019; 8: 2123–2132. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Bohoudi O, Bruynzeel AME, Senan S, et al. Fast and robust online adaptive planning in stereotactic MR‐guided adaptive radiation therapy (SMART) for pancreatic cancer. Radiother Oncol. 2017;125:439–444. [DOI] [PubMed] [Google Scholar]
  • 16. El‐Bared N, Portelance L, Spieler BO, et al. Dosimetric benefits and practical pitfalls of daily online adaptive MRI‐guided stereotactic radiation therapy for pancreatic cancer. Pract Radiat Oncol. 2019;9:e46–e54. [DOI] [PubMed] [Google Scholar]
  • 17. Tyran M, Jiang N, Cao M, et al. Retrospective evaluation of decision‐making for pancreatic stereotactic MR‐guided adaptive radiotherapy Radiother Oncol. 2018;129:319–325. [DOI] [PubMed] [Google Scholar]
  • 18. Olberg S, Green O, Cai B, et al. Optimization of treatment planning workflow and tumor coverage during daily adaptive magnetic resonance imaging guided radiation therapy of pancreatic cancer. Radiat Oncol. 2018;13:51. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Rosenberg SA, Henke LE, Shaverdian N, et al. A multi‐institutional experience of MR‐guided liver stereotactic body radiation therapy. Adv Radiat Oncol. 2018;4:142–149. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Finazzi T, Palacios MA, Spoelstra FOB, et al. Role of on‐table plan adaptation in MR‐guided ablative radiation therapy for central lung tumors. Int J Radiat Oncol Biol Phys. 2019;104:933–941. [DOI] [PubMed] [Google Scholar]
  • 21. Green OL, Henke LE, Hugo GD. Practical clinical workflows for online and offline adaptive radiation therapy. Semin Radiat Oncol. 2019;29:219–227. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Ge Y, Wu QJ. Knowledge‐based planning for intensity‐modulated radiation therapy: a review of data‐driven approaches. Med Phys . 2019;46:2760–2775. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Appenzoller LM, Michalski JM, Thorstad WL, Mutic S, Moore KL. Predicting dose‐volume histograms for organs‐at‐risk in IMRT planning Med Phys. 2012;39:7446–7461. [DOI] [PubMed] [Google Scholar]
  • 24. Ma M, Kovalchuk N, Buyyounouski MK, Xing L, Yang Y. Dosimetric features‐driven machine learning model for DVH prediction in VMAT treatment planning. Med Phys. 2019;46:857–867. [DOI] [PubMed] [Google Scholar]
  • 25. Shiraishi S, Tan J, Olsen LA, Moore KL. Knowledge‐based prediction of plan quality metrics in intracranial stereotactic radiosurgery. Med Phys. 2015;42:908. [DOI] [PubMed] [Google Scholar]
  • 26. Shiraishi S, Moore KL. Knowledge‐based prediction of three‐dimensional dose distributions for external beam radiotherapy. Med Phys. 2016;43:378–387. [DOI] [PubMed] [Google Scholar]
  • 27. Campbell WG, Miften M, Olsen L, et al. Neural network dose models for knowledge‐based planning in pancreatic SBRT. Med Phys. 2017;44:6148–6158. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. McIntosh C, Purdie TG. Contextual atlas regression forests: multiple‐atlas‐based automated dose prediction in radiation therapy. IEEE Trans. Med Imaging. 2016;35:1000–1012. [DOI] [PubMed] [Google Scholar]
  • 29. Nguyen D, Jia X, Sher D, et al. 3D radiotherapy dose prediction on head and neck cancer patients with a hierarchically densely connected U‐net deep learning architecture. Phys Med Biol. 2019;64:065020. [DOI] [PubMed] [Google Scholar]
  • 30. Fan J, Wang J, Chen Z, Hu C, Zhang Z, Hu W. Automatic treatment planning based on three‐dimensional dose distribution predicted from deep learning technique. Med Phys. 2019;46:370–381. [DOI] [PubMed] [Google Scholar]
  • 31. Kearney V, Chan JW, Haaf S, Descovich M, Solberg TD. DoseNet: a volumetric dose prediction algorithm using 3D fully‐convolutional neural networks. Phys Med Biol. 2018;63:235022. [DOI] [PubMed] [Google Scholar]
  • 32. Sun B, Lam D, Yang D, et al. A machine learning approach to the accurate prediction of monitor units for a compact proton machine. Med Phys. 2018;45:2243–2251. [DOI] [PubMed] [Google Scholar]
  • 33. Kang J, Schwartz R, Flickinger J, Beriwal S. Machine learning approaches for predicting radiation therapy outcomes: a clinician’s perspective. Int J Radiat Oncol Biol Phys. 2015;93:1127–1135. [DOI] [PubMed] [Google Scholar]
  • 34. Carlson JN, Park JM, Park S‐Y, Park JI, Choi Y, Ye S‐J. A machine learning approach to the accurate prediction of multi‐leaf collimator positional errors. Phys Med Biol. 2016;61:2514. [DOI] [PubMed] [Google Scholar]
  • 35. Valdes G, Scheuermann R, Hung C, Olszanski A, Bellerive M, Solberg T. A mathematical framework for virtual IMRT QA using machine learning. Med Phys. 2016;43:4323–4334. [DOI] [PubMed] [Google Scholar]
  • 36. Barragán‐Montero AM, Nguyen D, Lu W, et al. 2019 Three‐dimensional dose prediction for lung IMRT patients with deep neural networks: robust learning from heterogeneous beam configurations. Med Phys. 2019;46:3679–3691. [DOI] [PubMed] [Google Scholar]
  • 37. Hammel P, Huguet F, van Laethem JL, et al. and Louvet C. Effect of Chemoradiotherapy vs Chemotherapy on Survival in Patients With Locally Advanced Pancreatic Cancer Controlled After 4 Months of Gemcitabine With or Without Erlotinib: The LAP07 Randomized Clinical Trial. JAMA. 2016;315:1844–1853. [DOI] [PubMed] [Google Scholar]
  • 38. Fischer‐Valuck BW, Henke L, Green O, et al. Two‐and‐a‐half year clinical experience with the world’s first magnetic resonance image guided radiation therapy system. Adv Radiat. Oncol. 2017;2:485–493. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Li HH, Rodriguez VL, Green OL, et al. Patient‐specific quality assurance for the delivery of Co‐60 intensity modulated radiation therapy subject to a 0.35‐T lateral magnetic field. Int J Radiat Oncol Biol Phys. 2015;91:65–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Wang Y, Mazur TR, Green O, et al. A GPU‐accelerated Monte Carlo dose calculation platform and its application toward validating an MRI‐guided radiation therapy beam model. Med Phys. 2016;43:4040–4052. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Greene OL, Rankine LJ, Cai B, et al. First clinical implementation of real‐time, real anatomy tracking and radiation beam control. Med Phys. 2018;45:3728–3740. [DOI] [PubMed] [Google Scholar]
  • 42. Bin C, Green OL, Kashani R, et al. A Practical Implementation of Physics Quality Assurance for Photon Adaptive Radiotherapy. Z Med Phys. 2018;28:211–223. [DOI] [PubMed] [Google Scholar]
  • 43. Krishnan S, Chadha AS, Suh Y, et al. Focal radiation therapy dose escalation improves overall survival in locally advanced pancreatic cancer patients receiving induction chemotherapy and consolidative chemoradiation. Int J Radiat Oncol Biol Phys. 2016;94:755–765. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Reyngold M, Parikh P, Crane CH. Ablative radiation therapy for locally advanced pancreatic cancer: techniques and results. Radiat Oncol. 2019;14:95. [DOI] [PMC free article] [PubMed] [Google Scholar]