Abstract

Pure gases and mixtures containing stable isotopes are used in a wide variety of applications including health care, food authentication, geochemistry, and environmental monitoring. It is therefore important to understand the role of moisture, which is one of the most critical impurities in compressed gas mixtures and pure gases, in their stability. Gaseous carbon dioxide (CO2) was used as a proxy for the evaluation of the effects of its isotopic composition, when in contact with moisture throughout the depletion of a cylinder’s contents, as well as pressure regulation and long-term stability. To accentuate the detrimental effects of moisture on CO2 isotopic stability, enriched 18O-water was added to natural-abundance, gaseous carbon dioxide. The δ18O–CO2 changed from −23.16‰ vs Vienna Pee Dee Belemnite (VPDB) to +109‰ vs VPDB. It was further demonstrated that with appropriate cylinder preparation to reduce residual moisture, source material purity with low moisture content, and pressure regulation (from 57.0 down to 0.5 bar), both δ13C and δ18O remained consistent within ±0.04 and ±0.06‰, respectively, throughout the entire cylinder contents. Pressure reduction using a dual-stage regulator yielded statistically consistent results at the 99% confidence level from delivered pressures of 0.1–0.8 bar. Furthermore, the isotopic values remained consistent during a 1 year shelf-life study, illustrating the ability to utilize and regulate pressurized gases as working reference standard gases.
Introduction
Stable isotopes have been an integral part of the toolbox of many branches of science and technology and have seen an increase in areas such as food authentication and environmental research.1 At the heart of the toolbox is the requirement for stable calibration and working gas mixtures used for analyzer calibration. Calibration gases are often used as working gases or calibration gases for stable isotope measurements. However, these calibration and working gases are sometimes thought of as unstable, and the longer the gas remains within a cylinder, the larger the discrepancy will be with the original isotopic values.2 Furthermore, it is believed that as the cylinder is depleted or the pressure is regulated, the isotopic ratio of the gas will change (due to fractionation).3,4
Examining the catalogs of industrial and specialty gas manufacturers, it is evident that there is a wide range of specifications available for all pure gases. Stated purities range from 99.9 to 99.999%. At a purity level of 99.999%, no more than 10 ppm total specified impurities will be present in the gas, whereas 99.9% contains no more than 1000 ppm total specified impurities. These impurities can include moisture, hydrocarbons, carbon dioxide, carbon monoxide, oxygen, and other impurities. The cylinder preparation, which includes the use of high-purity cylinders, vacuum baking, and measurement of trace impurities are not as rigorous for industrial quality gases or those with lower purity specifications.
In a compressed gas cylinder, the molecules of each component within the gas mixture are constantly in motion according to the kinetic theory of gases. This helps maintain the homogeneity of the mixture. However, this is also one of the primary reasons that reactive components in a mixture decompose.5 They will come in contact with the cylinder material or impurities present in the mixture and react or adsorb, thus forming new compounds. By understanding the factors that affect the stability, those having an impact can be minimized with appropriate material selection,6,7 cylinder treatment,8 and quality control of the source material used to prepare the mixtures.5 Stable isotope mixtures used for calibration of analytical equipment have an added degree of complexity: the isotopic composition, in addition to the molecular concentration, must remain stable. Either or both of these may change with inappropriate cylinder preparation methods or source material that may contain critical impurities, such as moisture.
Since accurate and reliable calibration mixtures and working gases are important to all stable isotope measurements, experiments were designed to access the effect of moisture on the isotopic stability. A previous work by Gonfiantini et al.9 focused on δ13C-CO2 ratios and how they changed with the addition of 13C-enriched CO2 into natural-abundance CO2. In compressed gas cylinders, moisture is ubiquitous and known to cause stability issues;5,8 therefore, these experiments focused on using added moisture to highlight the 18O-16O exchange between the moisture and CO2. Gaseous carbon dioxide was chosen as a proxy for these evaluations as it is known that the oxygen in CO2 and the oxygen in water (H2O) can undergo exchange reactions10 as illustrated
For stable isotope mixtures to be stable (both in composition and isotopic signatures), there must be no residual moisture to participate in exchange reactions.3 This exchange rate increases with both temperature and pressure.10 Therefore, any residual moisture in the compressed gas cylinder or source gas can participate in this exchange reaction, altering the isotopic composition of both the residual moisture and CO2. It was therefore decided to use gaseous CO2 as a model stable isotope compound to understand the sensitivity of this exchange reaction to moisture, which may be contained within the compressed gas cylinder after cylinder preparation or as an impurity in the gaseous CO2.
Furthermore, the stability with high-purity carbon dioxide was verified with depletion studies to verify that fractionation in appropriately prepared compressed gas cylinders (which eliminates trace moisture) would not occur during usage. The achievement of cylinder nonreactivity and source material purity was validated with a shelf-life study conducted over 1 year.
Additionally, to use a compressed gas cylinder, it is necessary to reduce and regulate the pressure to a safe and suitable inlet pressure for an isotope ratio mass spectrometer (IRMS) or any other instrument used to measure isotope ratios. Experiments were designed to compare δ13C-CO2 and δ18O-CO2 values during pressure reduction by both sequential volume expansion (pressure reduction without the use of a pressure regulator) and with the use of a dual-stage pressure regulator.
Results and Discussion
Exchange of Oxygen in CO2 and O in Water
Oxygen-enriched water was added to a cylinder of research grade CO2. The cylinder was then heated for 67 h to favor the kinetics of the exchange reaction to achieve a new thermodynamic equilibrium. It was then allowed to sit for 121 days at room temperature. Samples of the CO2 were obtained both before and after moisture addition and analyzed for δ18O-CO2 by IRMS. The initial δ18O–CO2 value was −23.16‰ vs Vienna Pee Dee Belemnite (VPDB) and the value after 121 days was +109‰ vs VPDB; thus validating the significant effect that moisture could have in the stability of the δ18O values of CO2 in compressed gas cylinders.
Pressure Regulation
The ability to regulate pressure from a compressed gas cylinder without fractionation is important to ensure that the highest quality data are obtained. To evaluate this, pressure from a compressed gas cylinder was reduced through sequential expansions to avoid the use of a pressure regulator and compared with the values after pressure reduction using a regulator. A dual-stage, low-dead-volume piston regulator was selected for the comparison over a single-stage regulator as the dual-stage regulator is designed to lower the pressure in two stages and will provide a more consistent delivery pressure.
The experimental setup is illustrated in Figure 1. All measurements were averages of three individual measurements. Background levels obtained by sequential expansion for δ13C and δ18O were (−24.50 ± 0.05)‰ and (−24.99 ± 0.05)‰, respectively. Measurements of pressure-reduced samples obtained using a pressure regulator were done in the range of 0.1–0.8 bar. The triplicate measurements are illustrated in Table 1. To assess the effect of pressure reduction, the measurement values at the designated pressures are subtracted from the background. These values are reported as Δ13C and Δ18O in Table 1. Background levels were obtained by sequentially expanding the gas through a series of successive volume expansions as described in the Materials and Methods section. This facilitated reduction of the pressure to a usable level without the use of a pressure regulator. At the 99% confidence level, the measurement values as a function of pressure are statistically the same as the background indicating that the use of a regulator does not statistically alter the stable isotopic values.
Figure 1.

Experimental configuration to study the effect of pressure regulation vs a background that was obtained by sequential gas expansion (without the use of a regulator). Stepwise pressure reduction/expansion: Sequential expansion through Segments a, b, and c. The volume of the system is approximately 3024 mL. Effect of pressure reduction: Segments a and d with an in-line pressure regulator. The volume of the aliquoter in Segment d is 5 mL.
Table 1. Background Analysis of δ13C–CO2 (VPDB) and δ18O–CO2 (VPDB) Without the Use of a Pressure Regulator (Gas Expansion Only) and of δ13C–CO2 E(VPDB) and δ18O–CO2 (VPDB) Obtained with the Use of a Pressure Regulatora.
| δ13C (‰) | δ18O (‰) | Δ13C (‰) | Δ18O (‰) | |
|---|---|---|---|---|
| background | –24.50 ± 0.05 | –24.99 ± 0.05 | ||
| 0.14 bar | –24.48 ± 0.03 | –25.01 ± 0.06 | 0.00 ± 0.08 | 0.02 ± 0.11 |
| 0.21 bar | –24.49 ± 0.04 | –25.00 ± 0.07 | 0.01 ± 0.09 | 0.01 ± 0.12 |
| 0.34 bar | –24.47 ± 0.04 | –24.98 ± 0.06 | 0.00 ± 0.09 | –0.01 ± 0.14 |
| 0.54 bar | –24.47 ± 0.04 | –24.98 ± 0.06 | 0.01 ± 0.09 | –0.01 ± 0.11 |
| 0.82 bar | –24.51 ± 0.05 | –25.01 ± 0.05 | 0.01 ± 0.10 | 0.02 ± 0.10 |
Δ13C and Δ18O represent the difference between pressure reduction by sequential volumetric expansion and pressure reduction with a pressure regulator.
Stability of δ13C–CO2 and δ18O–CO2 During Cylinder Depletion
Typically, as the cylinder pressure decreases, species which may be present on the surface of the cylinder, such as moisture, can desorb from the cylinder surface.8 Trace moisture is known to desorb from compressed gas cylinder walls below approximately 4 bar;11 therefore, if there is any residual moisture remaining in the compressed gas cylinder, the CO2–oxygen isotopic ratios will change as the cylinder is depleted, with more pronounced effects below approximately 4 bar.
To validate the absence of moisture in the cylinder after vacuum baking, a 16 L cylinder was filled to approximately 33 bar with gaseous research grade CO2. The δ18O and δ13C isotopic ratios were measured by IRMS as the cylinder was depleted at a rate of approximately 50 mL/min (Figure 2). The withdrawal took approximately 35 days. The initial δ13C and δ18O values at 33 bar were −39.28 and −24.13‰, respectively. The final values were −39.28‰ for δ13C and −24.14‰ for δ18O. The average values over the entire cylinder depletion were −39.27 and −24.15‰ for δ13C and δ18O, respectively, with an overall precision of better than ±0.05‰ for δ13C and ±0.07‰ for δ18O. This stability of δ18O over the entire cylinder contents validated the absence of residual moisture in the cylinder or source gas.
Figure 2.

δ13C(VPDB) and δ18O(VPDB) vs CO2 cylinder pressure in bar of a gas-phase CO2 cylinder, where 35 bar represents a full cylinder. The flow rate of CO2 throughout the duration of the experiment (∼5 weeks) was kept constant at ∼50 mL/min.
Shelf Life
To validate the stability of the CO2 over time, δ13C and δ18O isotopic values were measured periodically over 1 year (Figure 3). Over the period, δ13C and δ18O registered the values of (−10.85 ± 0.05)‰ and (−12.64 ± 0.03)‰, respectively, validating the stability of the research grade CO2 with low moisture levels (≤3 ppm) in appropriately prepared cylinders. Proper cylinder preparation is discussed in the Materials and Methods section. Note that these isotope measurements were made from a different CO2 source than those made during the CO2 depletion experiment, hence the C and O isotopic differences between the two.
Figure 3.

Stability of gaseous δ13C–CO2 (VPDB) and δ18O–CO2 (VPDB) over 1 year.
Conclusions
It was demonstrated that with the carbon dioxide source material containing low levels of moisture, proper cylinder preparation, which included achieving vacuum levels <1.3 × 10–5 bar and pressure regulation, the stable isotope composition of both C and O will remain consistent to within ±0.05 and ±0.07‰, respectively, throughout depletion of the entire cylinder contents. The belief that gaseous stable isotopic standards are unstable is likely due to mixtures or pure gases with high levels of moisture and poorly prepared cylinders.
Additionally, it was demonstrated that pressure reduction using a dual-stage regulator yielded statistically consistent results at the 99% confidence level from a delivery pressure of 0.1–0.8 bar. It was further shown that the isotopic values remained stable during a 1 year shelf-life study, dispelling long-held beliefs that gaseous components tend to fractionate upon storage and fractionate upon use. This work also illustrates the ability to utilize and regulate pressurized gases as working reference standard gases.
Materials and Methods
The cylinders used in this study were 6061 aluminum alloys, which had internal volumes of 47, 30, and 16 L and were obtained from Luxfer (Riverside, CA). All cylinders were visually examined prior to use to ensure that the interior cylinder surface was free from obvious contamination. Ceodeux (Mt. Pleasant, PA) brass diaphragm valves were used. Cylinders were prepared using the Aculife 3 cylinder preparation technology prior to filling, which entails vacuum baking at vacuum levels <8 × 10–6 bar. Stable isotope ratio measurements were made on a Thermo Fisher Scientific MAT 253+ isotope ratio mass spectrometer (Thermo Fisher Scientific, Bremen, Germany) analyzed in dual inlet mode. Working reference CO2 materials were calibrated against NBS-18 (δ13CVPDB = −5.014‰, δ18OVPDB = −23.2‰) and NBS-19 (δ13CVPDB = +1.95‰, δ18OVPDB = −2.20‰) and NIST 8562 (δ13CVPDB = −3.72‰, δ18OVPDB = −8.43‰), NIST 8563 (δ13CVPDB = −41.59‰, δ18OVPDB = −23.61‰), and NIST 8564 (δ13CVPDB = −10.45‰, δ18OVPDB = +0.06‰). Final 13C and 18O values using the per mil (δ‰) deviation of the isotope ratios are expressed relative to Vienna Pee Dee Belemnite (VPDB).
All calibration standards and gases used in this study were obtained from Airgas (Plumsteadville, PA.). Three different sources of research grade gaseous carbon dioxide were used in this study (99.999%, containing ≤3 ppm moisture, Airgas, Plumsteadville, PA). Regulators used in this study were Airgas Model 14 dual-stage, low-dead-volume piston regulators (Airgas, Plumsteadville, PA).
Exchange of Oxygen in CO2 and O in Water
The ability of gaseous CO2 and H2O to undergo an exchange reaction in a compressed gas cylinder was first validated by adding approximately 1 g of 18O enriched water (98% atom purity, Berry & Associates, Dexter, MI) to a 47 L aluminum compressed gas cylinder using a 2 mL syringe, which had first been purged with dry nitrogen and heated in a lab oven set to 35 °C for 1 h prior to water injection. The injected quantity of moisture was selected to ensure that the dew point was not reached inside the cylinder, thus preventing droplet or cloud formation. The cylinder itself was placed in a cylinder oven set to 80 °C and purged with dry nitrogen several times before evacuating to 8 × 10–6 bar. The cylinder was vacuum-purged several times with the final evacuation performed overnight to 8 × 10–6 bar. The water injection apparatus was evacuated after it was attached to the cylinder prior to quantitative addition of water. After the water injection, the apparatus was purged with dry gaseous research grade CO2 to ensure that residual water was swept into the cylinder. Approximately 20 barg of gaseous research grade CO2 was added into the cylinder, which was then inverted on a cylinder inverter multiple times to ensure a homogeneous mixture. To favor the thermodynamics and kinetics of the exchange reaction, the cylinder was heated in an oven at 80 °C for 67 h. It was subsequently removed from the oven and allowed to sit at room temperature for 121 days.
Pressure Reduction with Sequential Volumetric Expansion
Prior to studying the dependence of cylinder pressure on the δ value uniformity throughout the compressed gas cylinder contents, it was necessary to validate the ability to reduce the cylinder pressure to that used by the IRMS without potential fractionation caused by pressure regulation. A 57 bar cylinder filled with research grade gaseous CO2(moisture ≤3 ppm) was used for the comparison. In order to reduce the cylinder pressure to a suitable pressure without the use of a regulator, a system was designed (Figure 1) to facilitate pressure reduction through stepwise gas expansion. All tubing segments were constructed of 1/4 in. electropolished stainless steel tubing, with the exception of the segment between the cylinder valve and valve 9, which was 1/8 in. The tubing segments between the cylinder valve and valve 9 had a volume of 2 mL. The tubing segments between valves 9 and 8 also had an internal volume of approximately 2 mL. The tee’s between valves 8, 6, and 7 as well as 3, 1, and 2 had an internal volume of approximately 10 mL. The expansion vessel in “Segment b” was 1 L. The total volume of the sequential volumetric expansion apparatus was approximately 3024 mL. The aliquot tube in “Segment d” was 5 mL.
Volumetric expansion was achieved by first closing the cylinder valve and valve 7 to the regulator. The entire system was then evacuated to 1.3 × 10–6 bar with a vacuum pump prior to expansion. After sufficient vacuum was reached, all valves were closed. The 57 bar CO2 cylinder valve was opened and the gas was allowed to expand into the 2 mL tubing segment (the small volume between the cylinder valve and valve 9, 1/4 in. SS diaphragm valve, model SS-DSS-4, Swagelok, Solon, OH). After expanding and allowing the gas to equilibrate for 30 s the cylinder valve was closed and valves 9, 8, and 6 were opened, allowing the gas to expand into the volume within the expansion vessel between valves 6 and 3. The gas was again allowed to equilibrate for 30 s after which valves 6 and 2 were closed and valves 3, 1, and the valve to the 2 L flask were opened. The gas was again allowed to equilibrate for 30 s. The valve to the 2 L flask was then closed as well as valve 1. The 2 L flask (now containing ∼0.5 bar of CO2 gas) was removed from the line and carried to the dual inlet of the IRMS for isotope analysis.
Direct Pressure Regulation
The effect of pressure regulation was obtained by regulating the pressure after the first expansion (“Segment a”) with a Model 14 dual-stage, low-dead-volume piston regulator. Valve 6 was closed; the valves in Segment d were opened as well as the valve to the CO2 cylinder and valves 9, 8, and 7. CO2 was then purged through the regulator and the gas aliquot tube (shown as Segment d) by opening valves 5 and 4 at a rate of 2 L/min for approximately 30 s to achieve a steady state. After purging was complete, valves 4 and 5 were closed (in that order). The pressure was dialed back to zero outlet pressure; valves 7, 8, and 9, as well as the cylinder valve were closed. The gas aliquot tube (Segment d) was removed from the regulator and carried to the dual inlet of the IRMS for isotope analysis. This procedure was repeated as the pressure was regulated at approximately 0.5 bar increments.
Stability of δ13C–CO2 and δ18O–CO2 During Cylinder Depletion
If any adsorbed species are present on the surface of the cylinder, they will tend to be desorbed as the cylinder pressure is decreased. The isotopic concentration uniformity as a function of cylinder pressure was evaluated by depleting a gas-phase carbon dioxide cylinder at a rate of approximately 50 mL/min until the cylinder was depleted. The cylinder, which was equipped with a CGA 320 connector, was placed on a cylinder scale to gravimetrically measure the amount of depleted CO2. A 1/8″ stainless steel tubing was used to connect the CGA to the regulator (Figure 4). The gas distribution system was evacuated to remove any residual contaminants. The cylinder pressure was regulated to approximately 0.2 barg. Gas aliquots were taken periodically by collecting a sample in a gas aliquot tube, “Segment e”. The gas aliquot tube was removed from the regulator and carried to the dual inlet of the IRMS for isotope analysis.
Figure 4.

Experimental configuration to evaluate the dependence of δ13C–CO2 (VPDB) and δ18O–CO2 (VPDB) on cylinder pressure.
Shelf Life
The stability of CO2 was evaluated by preparing cylinders using the Aculife 3 cylinder treatment process, which included vacuum baking to a vacuum level of <1.3 × 10–5 bar prior to filling the 30 L aluminum cylinders to 35 bar with research grade CO2. The cylinders were kept at room temperature and sampled periodically to assess the stability.
Acknowledgments
The authors thank Dr. Takuya Shimosaka and Dr. Shohei Murayama at the National Institute of Advanced Industrial Science and Technology (Japan) for making the isotopic measurements for the shelf-life study.
The authors declare no competing financial interest.
References
- Moore J. C.; Spink J.; Lipp M. Development and application of a data base of food ingredient fraud and economically motivated adulteration from 1980-2010. J. Food Sci. 2012, 77, R118–R126. 10.1111/j.1750-3841.2012.02657.x. [DOI] [PubMed] [Google Scholar]
- Meier-Augenstein W.; Schimmelmann A. A guide for proper utilization of stable isotope reference materials. Isot. Environ. Health Stud. 2019, 55, 113–128. 10.1080/10256016.2018.1538137. [DOI] [PubMed] [Google Scholar]
- Werner R. A.; Roth M.; Brand W. A. Extraction of CO2 from air samples for isotopic analysis and limits to ultra high precision δ18O determination in CO2 gas. Rapid Commun. Mass Spectrom. 2001, 15, 2152–2167. 10.1002/rcm.487. [DOI] [PubMed] [Google Scholar]
- Liu L. X.; Zhou L. X.; Vaughn B.; Miller J. B.; Brand W. A.; Rothe M.; Xia J. L. Background variations of atmospheric CO2 and carbon-stable isotopes at Waliguan and Shangdiazi stations in China. J. Geophys. Res.: Atmos. 2014, 119, 5602–5612. 10.1002/2013JD019605. [DOI] [Google Scholar]
- Jacksier T.; Weterings W. In Gaseous Calibration Standards: Manufacturing, Stability, Traceability and Uncertainty, Offshore Technology Conference, OTC 27614-MS, Houston, TX, May 1–4, 2017.
- Jacksier T.; Benesch R.. Increased Stability Low Concentration Gases, Products, Comprising Same and Methods of Making Same. U.S. Patent US77991502010.
- Benesch R.; Haouchine M.; Jacksier T.. Articles of Manufacture Containing Increased Stability Low Concentration Gases and Methods of Making and Using the Same. U.S. Patent US77948412010.
- Benesch R.; Haouchine M.; Jacksier T. The stability of 100 ppb hydrogen sulfide standards. Anal. Chem. 2004, 76, 7396–7399. 10.1021/ac049016o. [DOI] [PubMed] [Google Scholar]
- Gonfiantini R.; Valkiers S.; Taylor P. D. P.; De Bievre P. Using isotopic disequilibrium of CO2 to model gas adsorption in mass spectrometric measurements. Int. J. Mass Spectrom. Ion Processes 1997, 161, 15–26. 10.1016/S0168-1176(96)04527-2. [DOI] [Google Scholar]
- Cole D. R.; Chakraborty S. Rates and Mechanisms of Isotopic Exchange. Rev. Mineral. Geochem. 2001, 43, 83–223. 10.2138/gsrmg.43.1.83. [DOI] [Google Scholar]
- Li Y.-E.; Rizos J.; Vassalo D.; Giacobbe F. Behavior of moisture in high pressure inert cylinder gases. Vacuum 1993, 44, 433–434. 10.1016/0042-207X(93)90065-I. [DOI] [Google Scholar]